Bioengineering
Analysis of Orthodontic Mechanics
Bloengi neering
Analysis of Orthodontic Mechanics ROBERT J. NIKOLA...
132 downloads
1794 Views
8MB Size
Report
This content was uploaded by our users and we assume good faith they have the permission to share this book. If you own the copyright to this book and it is wrongfully on our website, we offer a simple DMCA procedure to remove your content from our site. Start by pressing the button below!
Report copyright / DMCA form
Bioengineering
Analysis of Orthodontic Mechanics
Bloengi neering
Analysis of Orthodontic Mechanics ROBERT J. NIKOLAI, Ph.D. Professor, Department of Orthodontics St. Louis University Medical Center St. Louis, Missouri
Foreword by CHARLES J. BURSTONE
LEA & FEBIGER
PHILADELPHIA
1985
LEA & FEBIGER
600 South Washington Square Philadelphia, PA 191064198 U.S.A. (215)922-1330
Library of Congress Cataloging in Publication Data Nikolai, Robert J. Bioengineering analysis of orthodontic mechanics Bibliography: p. Includes index.
1. Orthodontic appliances. 2. Force and energy. 3. Bioengineering. I. Title. [DNLM: 1. Biomedical engineering. 2. Orthodontic appliances. 3. Orthodontics. WU 400 N693b] RK527.N54 1984 617.643 84-5712 ISBN 0-8121-0922-8
Copyright © 1985 by Lea & Febiger. Copyright under the Inter-
national Copyright Union. All Rights Reserved. This book is protected by copyright. No part of it may be reproduced in any manner or by any means without written permission of the Publisher. Prioted in the United States of America
Print Number:
5
4
3
2
1
To My Parents
Foreword
The advancement of orthodontics and the quality of orthodontic treat-
ment are dependent upon a thorough knowledge of basic science. Although there are many orthodontic texts that cover the biologic basic sciences, such as growth and development, neuromuscular physiology, genetics, and anatomy, considerably less attention has been paid to bioengineering. Certainly no basic science could be more relevant to the orthodontic researcher and to the practicing orthodontist than bioengineering. Dr. Nikolai has written the first comprehensive text that can serve as a keystone to the understanding of bioengineering principles in general and their application to clinical orthodontics.
This text develops the fundamental concepts in an orderly way which will allow the orthodontist to understand and use his orthodontic appliances. Since most dentists in their predoctoral training receive little information relative to bioengineering, the development of basic concepts and their application is most important and necessary before moving on to more complicated clinical situations. The sequential presentation of material from the elementary principles to the more complicated force systems of multibanded appliances will allow both student and practitioner to gain a better understanding of how orthodontic appliances work and how to improve clinical therapy based on this understanding. Bioengineering has a broad application to orthodontics and, hence, the scope of the text is very wide. It includes a discussion of the force systems produced from orthodontic appliances, the behavior of the materials that are used, and correlates the relationship between the torce systems and the biologic changes that occur in the dentition and periodontium. The excellent use of clinical problems and examples through-
out the text insures that this is not just another engineering text, but truly a book on orthodontic bioengineering, which can serve both the student and the experienced practitioner. CHARLES J. BURSTONE, D.D.S., M.S. Professor and Head Department of Orthodontics School of Dental Medicine The University of Connecticut vii
Preface
As viewed by the observant bioengineer, the orthodontic appliance is a
unique assembly of small, structural elements. When activated, the appliance exerts a system of forces upon the human dentition. These forces cause, or at least serve to catalyze, biologic remodeling within the dentofacial complex toward achievement of finite, permanent displacements of teeth. From both structural and clinical perspectives, the principal component of the fixed, intraoral appliance is the arch wire. This wire is currently available to the clinician in at least five materials (metal alloys), round and rectangular cross-sections in various sizes, single- and mul-
tistrand formats, and in straight lengths or precontoured into arch forms. Prior to engagement and activation the practitioner may bend or twist the wire, using special tools to place or modify the arch form and detail the wire geometry to the specific patient. The appliance contacts the dentition through crown-affixed brackets and tubes, and the structural assembly may also include various kinds of metallic springs and polymeric elements to exert forces. Moreover, extraoral adjuncts may contribute to the total appliance.
The complexity of the orthodontic appliance as a structure is extended through its dynamic behavior and that of its foundation, the dentition; over time the appliance deactivates slowly in concert with movements within the biologic complex to which it is attached. The overall analysis requires an integrated examination of the appliance and the force system it delivers to the dentition. The characteristics of that delivered force system influence the format of the potential displacements of the teeth. The interested bioengineer now envisions a multifaceted analysis problem. The experienced practitioner, as well as the clinician-in-training, perhaps viewing structural analysis as "foreign territory," should gain some familiarity with the engineering of the orthodontic appliance, and this text has been prepared to assist in that education. Herein the subjects of force and structural analysis are discussed within the framework of practical, orthodontic treatment and at a level appropriate to the
intended audience. The presentation is seemingly theoretical for the most part, although much of the theory has been demonstrated to be ix
x
Preface
well founded through bench and clinical experimentation. The topic
coverage is largely a composite of analytic and graphical mechanics, materials science, and mechanics of deformable bodies (subject titles familiar to the engineering student), but presented with minimal mathematical involvement and with direct application to orthodontics. Substantial attention is given to the interactions between the activated appliance and the biologic systems and processes affected by the transmitted force.
This text is an outgrowth of class notes prepared and carefully revised, over a period in excess of ten years, for a bioengineering course presented to first-year students in an orthodontics graduate program. Not only intended for students, the book should serve the interested, experienced practitioner as well, whether or not previously exposed to formalized graduate or continuing-education coursework in orthodontic biomechanics. Because examples are used throughout, the text as-
sumes at least superficial familiarity with orthodontic therapy that would be gained, for example, through an introductory or survey exposure within the professional dental curriculum. Principles and procedures accumulate from the start in the book, but no preparation in mathematics and science, beyond traditional predental and dental curricula, are required. The text begins with a discussion of needed mathematical "tools." The subsequent three chapters introduce concepts of engineering mechanics in the orthodontic setting. Con-
sidered next is the response to applied force of the dentition and its supporting tissues. Structural analysis procedures are then introduced and applied to simple appliance elements. Chapters are devoted, individually, to bending and torsional analyses of arch wires and to extraoral appliances. Finally, force and structural analyses of entire representative orthodontic mechanics are undertaken.
Orthodontic bioengineering is becoming widely recognized as an important area of formal study for the aspiring clinician. The design and preparation of appliances have been traditionally learned from preceptors, and refined through actual clinical experience. Recent advances in appliances and techniques, however, have resulted from the joint contributions of practitioner expertise and engineering science. Although the latter has demonstrated its potential impact toward overall advance-
ment of orthodontic treatment, the organized application of bioengineering to orthodontics remains relatively obscure, apparently, to the majority of practitioners. Accordingly, the objective in preparing this text was to provide a means whereby a bioengineering approach to the analysis and design of orthodontic appliances and procedures might be outlined to the orthodontic community. Several individuals have intangibly, although importantly, contributed to the completion of this work. Vasil Vasileff, now retired from the School of Dental Medicine at Southern Illinois University, and formerly a faculty member at the St. Louis University Medical Center, presented
the author with his first opportunity to interact with graduate orthodontics students. Kenneth Marshall, former Chairman of the Depart-
xi Preface
ment of Orthodontics, hired and charged the author, trained in engi-
neering rather than in dentistry, to develop the bioengineering course within the graduate curriculum. Lysle Johnston, present Chairman of the Department, through his urging and encouragement, was instrumental in making this text, begun in earnest during a sabbatical leave, a reality. The author would be remiss without acknowledging the influence of
the students in the graduate orthodontics program at St. Louis University, particularly those whose thesis research projects he has supervised; their comments, criticisms, and suggestions were of substantial aid in this effort. Finally, to Charles Burstone, eminent researcher, clinician, and educator, who kindly consented to contribute the foreword to this text, the author expresses his sincere gratitude. ROBERT J. NIKOLAI, Ph.D. St. Louis, Missouri
Contents
1. Mathematical Topics Constants, variables, and functions 1 Frames of reference 3 Displacements of particles and solid bodies Trigonometry
8
10
An introduction to vector algebra 16 Dimensions and units 20 Measurements, computations, and numerical accuracy
21
2. Introduction to Analysis of Orthodontic Force
24
Mechanics 24 Force 25
Vector addition and decomposition of concentrated forces 29 The moment of a concentrated force 36 The couple 40 Resultants of force systems exerted on rigid bodies 46 Distributed forces and their resultants 50 Friction
53
Concepts of mechanical equilibrium and their applicability to orthodontic mechanics 56 Synopsis
69
3. Material Behavior of the Orthodontic Appliance
Internal structure of a solid material 73 Load-deformation behavior of a structural member 78 Mechanical stress 87 Mechanical and structural properties: Standardized testing Chemical and thermal influences 103 Selection of materials 110
71
91
xlii
xiv Contents
4. Energy Analyses in Orthodontics Concepts leading to the process laws Displacement 114
113 114
Mechanical work 115 Energy 117
Heat transfer and thermal energy 121 The conservation-of-energy law 122 Available energy 124 Activation and deactivation processes 125 Strain hardening and heat treatment of metals and alloys A work-energy analysis of the preparation of an arch wire Synopsis 144
5. Response of Dentition and Periodontium to Force
137 141
146
Mechanical response of the individual tooth to applied
force
148
Transverse crown force systems and tooth displacements Extrusion, intrusion, and long-axis rotation 163 Response of the periodontium to force 169 Displacements related to magnitude and duration of force Controlling the force-time pattern 178 Physiologically proper orthodontic forces 182 Dentofacial Orthopedics 189 Synopsis 191
151
177
6. Introduction to Structural Analysis of the Orthodontic
Appliance
194
orthodontic appliance: A structure or a machine? 196 Attachment of the orthodontic appliance to the dentition 197 A continuous-arch-wire appliance model 205 The
An overview of the structural analysis of an orthodontic appliance 208 The activation and deactivation characteristics of representative appliance elements 211 The action and response of tip-back bends: An illustration of the four-step procedure in orthodontic structural analysis 224 "Control" of the orthodontic appliance 228 Synopsis 231 7. Behavior of Orthodontic Wire in Bending
deformations, strains, and stresses 235 Force systems within the beam 241 Beam
Beam stiffnesses 248
233
xv Contents
Additional topics in elastic bending 252
Application of elastic beam theory to orthodontic arch-wire activation in bending 257 Inelastic behavior in bending 261 Orthodontic wire loops 264 Synopsis
269
8. Delivery of Torque by the Orthodontic Appliance Structural theory for the straight, circular shaft
272
273
Extension of theory to shafts having rectangular cross-sections 285
Application of shaft theory to the orthodontic arch wire 288 Isolation of the activating torque 288 Responses of the appliance and the dentition to torsional activation 289
Structural influences on active and responsive force systems 295 Anterior-segment torquing mechanics 299 Rectangular-wire torquing 299 Torquing spurs in the appliance 305 Inelastic behavior in third-order mechanics 310 Wire loops, springs, and torsion 314 Synopsis 318 9. Extraoral Appliances
322
The cervical-pull, face-bow appliance Occiusal-plane analysis 326
325
Buccal-view analysis of the cervical appliance
335
A coronal-plane view and comments on the asymmetric problem 343 Canine retraction with headgear 344 Extraoral force delivered to an anterior segment or an entire arch
348
Delivery of extraoral force to the mandibular arch 358 Dual-force headgear 361 The chin-cap assembly 362 Reverse-pull appliances 365 Synopsis 369
10. Force and Structural Analyses of Representative Orthodontic Mechanics Individual tooth malalignments Leveling displacements 374
374
372
xvi Contents
Rotational corrections
379
Bilateral action 385 Interarch mechanics 388 Intra-arch vertical positioning 392 Intra-arch retraction mechanics 402
Prepared, posterior-segment anchorage 412 Class III mechanics 417 Class II mechanics 424 Synopsis 435
Appendix I Glossary of Terms
437
Appendix II List of Symbols
458
Index
463
Mathematical Topics I Orthodontic bioengineering has emerged as an important subfield of
orthodontics, not only for research but for formal study by the practitioner-in-training as well as the experienced clinician. Bioengineering is appropriately termed an "applied science" and, as such, uses numerous mathematical concepts and procedures. This chapter introduces and explores the mathematical topics required in this text. These developments are not unduly rigorous and most will not be totally new to the reader. Familiarization with the necessary mathematics is undertaken initially in this text, rather than to introduce each topic individually as the need arises. As necessary, the reader may refer back to this chapter for a review of any specific concept. Moreover, a foundation is laid for uninterrupted and continuous study of the principles and applications of bioengineering to orthodontics.
Constants, Variables, and Functions Any object having physical attributes or characteristics may be described
qualitatively or quantitatively. For example, a specific orthodontic wire may be characterized as being shiny, made of stainless steel, round, and having a cross-sectional diameter equal to .018 in. The first three de-
scriptive words or phrases refer to qualitative attributes; the last is a quantitative characteristic. Descriptions, whether qualitative or quantitative, are helpful only insofar as their references or bases are familiar and understood. In describing the orthodontic wire, the adjectives were ordered from the least to the most definitive: "shiny" implies a comparison with some herein undefined luster or smoothness of surface; several types of stainless steel, differing from one another metallurgically, are
used in fabricating orthodontic appliances; and "round" is meant to describe the shape of the wire cross-section (but, alternatively, it might
refer to a curvature placed in the wire). Finally, the reference to the "diameter" of the cross-section with the measurement given in inches is
2
Bioengineering Analysis of Orthodontic Mechanics
the most definite because the dimension is well understood and the inch
is an established unit of length. This example suggests that descriptions in quantifiable terms are generally clearer because of familiarity with the bases of those terms. A particular stainless steel may be described in
terms of the constituent materials of the alloy, given in percent by weight or by volume, or implicitly, for example, by the American Iron and Steel Institute grade-and-type designation. (Given the AISI code for the alloy, its composition may be learned by consulting an appropriate metals handbook.) Because many objective descriptions undergo some form of change in the most general discussion, in practical considerations some limits or bounds on the extent of change are often defined or implied. The magnitude of a force, exerted by an appliance on a tooth, may lessen over a period of days as the tooth moves and the appliance partially or totally deactivates. In a structural analysis undertaken to be valid for only a short time, however, the force may be considered as unchanging. The temperature in the oral cavity may vary in a complex fashion and may be affected temporarily by ingesting hot or cold food or beverages. Often, however, it is sufficient to consider oral temperature as unchanging, but elevated some 30°F relative to normal room temperature. (In a study of the characteristics of a metallic arch-wire material, for example, even the fact that intraoral temperature is elevated with respect to the ambient air may be inconsequential.) A constant, then, is defined as a quantity that retains a fixed, unchanging value or level within the bounds of a particular discussion or investigation. A quantity that cannot be regarded as constant while under study, either because its variation is the subject of the study or the contribution of its changing is not negligible in overall considerations, may experience change associated with movement from one location to another, or directly or indirectly as time passes. Location or position may be related to time. The force in a helically-coiled spring may be altered by changing
the position of one end of the spring with respect to the other. The position or orientation of a malposed tooth under the influence of an activated orthodontic appliance may change with time. A variable, is, in general description, a definable characteristic or parameter under study that takes on more than one quantifiable value during the course of a discussion or investigation. Variables may be categorized as independent or dependent. Although this distinction may often depend on viewpoint, a few variables (e.g., time and temperature) are generally independent, and other parameters assume their values in some manner associated with the magnitude(s) of the independent variable(s). In the analysis of the helical spring, the force in the spring may be said to be dependent on the amount of deformation referenced from the passive (zero-appliedforce) configuration; alternatively, the extent of the deformation may be viewed as dependent on the size of the force applied to the spring. In
the example of tooth location versus time, however, the parameter "time" can never be realistically considered as dependent on the orientation or position of a malposed tooth.
3
Mathematical Topics
Whenever two or more variables are involved in a common discussion or investigation, and whenever these variables are interrelated such that one or more unique values of one variable are assumed upon the assignment of specific values to the remaining variables, the first is dependent on, or said to be a function of, the other (independent) variables. A functional relationship is usually expressed explicitly in the depend-
ent variable. For example, if the symbol, y represents a quantity that takes on a particular value upon the assignment of a value to x, the general relationship is written as y = function (x), or simply y = f(x). If the "data" relating y to x originate from an experiment, the function f(x) is subsequently sought toward the facility of determining all values of y within some range or bounds of the independent variable x. The variation in a parametric value might be expressed mathematically in a smooth, continuous manner, and an analytic relationship, an equation, may be written to represent it. This representation can always be set down in a graphical or tabular form, but there are times when a variable takes on individual, discrete, seemingly unrelated values and an equation cannot be derived easily, if at all. The graph or table is then necessary to visualize a pattern in and the range of the variable values. In this text, the typical equation is simple in form and often derivable from the graphical representation of the relationship among the variables.
Frames of Reference When expressing graphically the change in value of a variable, the values and change may be associated with the length of, or distance along, a reference line or axis. Many variables have the possibility of assuming
the value of zero, and this or another convenient reference point or origin is located on the axis. To associate length or distance with the variable values, a scale is needed. In Figure 1-1, an axis is established for
the parameter "time." The origin of the plot coincides with a relative time of zero (hours). The chosen scale assigned one hour to a specific interval, and twelve adjacent intervals "equals" one-half day. In this particular representation, amounts of time elapsed or time change are of interest and negative time readings have no real meaning. Hence, since time is not an algebraic quantity, there is no need to extend the reference
I
t
i
I
i
i
I
I
I
1
1 hr
FIGuRE 1-1. A one-dimensional plot relating time as a variable, in hours, to distance along an axis.
4
Bloengineering Analysis of Orthodontic Mechanics
line in both directions from the origin, which has, therefore, been placed
at the left end of the axis with time increasing to the right. The graphical relationship of one dependent variable to another, or possibly to two other independent variables, is usually established within a two-dimensional, rectangular-coordinates framework. A pair of reference axes are defined as the basis for the graph; the axes are perpendicular to one another with the origin of each the intersection point,
the now-defined origin of the framework as a whole. Ordinarily, one axis is positioned horizontally and the other vertically. When the variables may only assume positive values, the origin is placed at the lowerleft so that variable values associated with the horizontal axis will increase from left to right and values associated with the vertical axis will increase upward. As with the one-dimensional "plot" of Figure 1-1, a variable name (e.g., force) and units of measurement (e.g., ounces) are appended to each axis. A scale is implied by the distance associated with an individual unit division along each axis. When preparing a graph of the relationship between two variables,
symbolized, for example, by x and y (without specifying the units of measurement of either), points would be individually plotted in the x-y plane and each point would be designated by a number pair, or coordinates, (x,y). When plotting a two-dimensional relationship with respect to a set of horizontal and vertical axes, the x-coordinate of a point is often termed the abscissa and the y-coordinate the ordinate. When all points have been plotted, a curve is sketched as a "best fit" through all of the plotted points. An attempt may then be made to obtain an analytic
expression that relates x and y at or near to every point on the drawn curve. Figure 1-2 shows a straight-line relationship that is defined as to position and orientation by just two (x,y) coordinate pairs. A nonlinear relationship is also displayed that may necessitate the plotting of many points prior to sketching the curve to obtain an accurate representation. The scales chosen for the axes need not be identical, regardless of whether or not similar units of measurement are appropriate to the variables symbolized by x and y. The straight line in Figure 1-2 was drawn after the coordinates (1,2) and (2,4) were plotted. If the lower bound for x and y is zero, the plot appropriately begins at the origin of this x-y frame (0,0). The equation of the line is easily determined to be y = 2x. The curve sketched with respect to the same pair of coordinate axes suggests a functional relationship between x and y that is quite different from that of the straight line. Although the numerical relationship of x to y can be readily obtained for any specific point on the curve, the equation relating the two variables at every point on the curve is not easily derived from the plotted points. The two plots of Figure 1-2 may be unrelated to one another or may
be "members" of a family of curves. If the latter is the case, a third variable is involved and, for each value of interest of that third variable, an individual relationship exists between x and y. Plotting a family of curves in the plane is a method of visually describing the relationship among three variables within a two-dimensional reference frame. Gen-
(2,4)
(1,2)
0
1
2
3
4
x
URE 1-2. Relationships between pairs of variables displayed, referenced to an coordinate framework.
thy,
when plotting one curve or a family of curves with respect to a
me framework consisting of horizontal and vertical axes, the dependI variable, expressed graphically as a function of one or two independ-
I variables, is scaled on the vertical axis.
6
Bioengineering Analysis of Orthodontic Mechanics
y
z
FIGURE 1-3. Relationships of an individual dependent variable to pairs of independent variables, sketched within an x-y-z-coordinate framework.
From a two-dimensional plot, in addition to enabling the determinalion of one or more values of the dependent variable for a given value of the independent variable(s), the slope of the curve at one or more specific
points is often useful. This slope represents the rate of change of the dependent variable with respect to a unit increase in the scaled dependent variable and is, geometrically, the inclination of the tangent to the curve at the specified point. The slope at a point may be positive or negative, depending upon whether the dependent variable is increasing or decreasing in value from that point. If the plot of the relationship between two variables is a straight line, the slope is the same at all points. Along the nonlinear curve in Figure 1-2, the slope decreases to x = 1, then increases with increasing values of x. The rectangular-coordinate framework in two dimensions may be extended to three with the addition of a third axis, mutually perpendicu-
lar to both axes in the plane and intersecting the origin of the plane frame. In the coordinate system in space, a functional relationship among three or possibly four variables may be displayed. Figure 1-3 shows a curve and a surface referenced to an x-y-z framework. Both the curve and the surface might represent functions in which z symbolizes the dependent variable and x and y the independent variables. The surface may serve to display a function for all values of x and y within some bounds; the curve, drawn in the surface, may graphically provide the functional relationship for only those values of the independent variables that are of particular interest, A function of three independent variables may be displayed in the spatial framework as a family of curves or
7
Mathematical Topics
The concept of slope can be extended to three dimensions. Along a surface the rate of change of z, for example, the dependent surfaces.
variable, may be considered at a point of the surface independently with respect to unit changes in x and y or with respect to a coordinate in the direction of a curve in the surface. A unique tangent plane is associated
with each point on a surface, and a slope may be computed for any specified direction in that plane. The interpretation of a functional relationship displayed in a threedimensional reference framework is difficult, particularly when sketched in two dimensions. Although the pictorial representation of analytic functions in three dimensions is generally beyond the needs of this text, coordinate systems have other uses, and an important one is the designation of reference directions in a plane or in space. In the study of dental anatomy, a coordinate frame is often placed in an individual tooth; one axis coincides with the long axis of the dental unit. Mutually perpendicular directions associated with this framework are occlusogingival (or occlusoapical), faciolingual (or labiolingual or buccolingual), and mesiodistal. Such a reference frame is shown in Figure 1-4. Note that these axes extend on both sides of the origin of the framework;
rather than adopting an algebraic scheme, a sense is specified (and the direction is implied) in discussions. For example, as indicated in Figure 1-4, "distal" indicates a specific sense in the mesiodistal direction. In a
l(ingual)
dOstal)
Origin
g(ingival) t(acial)
FIGURE 1-4. Mutually perpendicular coordinate directions used in dentistry.
m(esial)
$
Bioengineering Analysis of Orthodontic Mechanics
framework associated with the entire dental arch or dentition, the refer-
ence directions are generally labeled anteroposterior, lateral or transverse, and vertical. A plane of reference is generally associated directly with the coordinate perpendicular to the plane, often with the additional indication of the "side" of the plane being viewed. For example, the occiusal plane is perpendicular to the occlusogingival direction with the view toward the tooth crowns. A buccal plane is perpendicular to a buccolingual direction; an identical lingual plane exists, but the implied view is toward the lingual surface(s). The faciolingual and mesiodistal planes are perpendicular to one another, references associated with an individual dental unit, and the line common to both planes is coincident with the long axis of the tooth. Note that the common terminology for the planes of reference for the dentition, or each arch, as a whole deviates from this pattern; the coronal, the occlusal (or transverse), and the sagittal planes are the mutually perpendicular references, and two of the associated, perpendicular axes have designations unlike their counterpart planes.
Displacements of Particles and Solid Bodies The distance between two points in space (e.g., A and B) is measured in
length units along a straight line connecting the two points. The displacement "path" of a particle moving from point A to point B, however, need not be along that line. The actual distance traveled by the particle in moving from A to B is measured along the path followed, and
may be sizable compared to the straight-line measurement. The displacement of a particle from point A is in magnitude the distance from A to a specific point on the path, and is dependent upon the elapsed time since its coincidence with point A. Displacements of orthodontic interest
are generally small in absolute measurement, but are comparable to characteristic intraoral dimensions. Such displacements may be difficult to obtain accurately because (1) often few, if any, potential references with the dentofacial complex remain stationary over time, (2) position measurements cannot be monitored continuously, and (3) total move-
ments may be an inseparable combination of orthodontic and orthopedic displacements and growth. Although full coverage of this subject is properly relegated to the study of cephalometrics, several portions of the topic fit into the discussions of this chapter. First, any displacement to be quantified, must be measured with respect to a reference frame. The displacement path of a particle must often be estimated by connecting straight-line distances between successive positions of the particle; the larger the number of position measurements, the more accurate the estimation. The displacement of a solid body as a whole is fundamentally described in terms of the displacement of one particle of the body and the angular movement of a line within the body passing through this particle.
9
Mathematical Topics
C, C
1800
01
I
I arc D"D'
C
FIGURE 1-5. Successive positions of a tooth undergoing planar orthodontic displacement, including a change in angular orientation.
Figure 1-5 depicts the displacement of a tooth as well as the displace-
ments of two particles (or points) of the tooth; in this example, the points and movements are planar. The displacement of point D may be expressed in terms of the displacement of point C and the relative movement of D with respect to C. The angular displacement of the tooth is the difference in orientations, 02 — 01, of the line through points C and D. In
fact, all lines of the tooth in the plane of the sketch, or parallel to it, experience the same angular displacement. Because the distance between points C and D in the tooth does not change during the movement, the displacement path of the particle at D with respect to, or as viewed from, point C is a circular arc. The radius of that circular arc is the line segment in the tooth from C to D, the change in angular position of the segment CD represents the angular displacement of the tooth as a whole, and the relative curvilinear displacement of the particle at D with respect to point C is along a circular path, the arc D'D', and equals the product of the length of the segment CD and the angular change, 02 01. In
this computation the angular change must be expressed in radi-
ans, where 22/7 radians is comparable to 180°. The units of the length CD
and the relative displacement DD' are then identical. The completion of the determination of the "absolute" displacement of the particle coinciding with point D must be deferred until later in this
10
Bioerigineering Analysis of Orthodontic Mechanics
chapter because this displacement is, mathematically, a vector quantity.
This section has, however, introduced or reviewed concepts associated with particle and angular whole-body displacements and recalled the relationship among the size of the angle, the radius, and the length of the circular arc for a circular segment.
Trgonometry As noted in the preceding section, an angle is the measurement of the
orientation of one line with respect to another in the plane containing the lines. The actual quantification implies a circle centered at the point of intersection of the lines; the curvilinear length of the circular arc between the lines divided by the radius of the circle yields the size of the angle in radians. The example of Figure 1-6 shows an angle of 22/21 radians, or 600. (Although the dimensionless units of degrees are more familiar than radians, many computations to obtain an angle as the result yield the answer in radians.) Angles are often categorized by size: acute angles are smaller than 90°, right angles are exactly 90°, obtuse angles are larger than 90° but less than 1800; and straight angles are exactly 1800. Although circles are described as encompassing 360°, the smaller interior angle between two radii never exceeds 180°. The sum of acute and adjacent obtuse angles within the intersection of two lines is 180°; these angles are termed "supplementary" to one another. A triangle is a plane, three-sided, closed figure made up of three line segments. The intersection of each pair of line segments defines an
QA = OB =
S
AB = s 600 = 22/21 radians s = (22/21)r
0 FIGURE 1-6. Angular measurement.
A
11
Mathematical Topics
angle. Each angle is measured interior to the figure and the three angular measurements always sum to 1800 or 22/7 ("pi") radians. Also, for
every triangle, the ratio of the length of a side to a function of the angle opposite that side is a constant. That function is defined as the "sine" of the angle; it is a dimensionless quantity that has bounds for angles between 0° and 180° of 0 and +1. Triangles are categorized according to the size of the largest interior angle. A right triangle contains, opposite to the longest side termed the "hypotenuse," a right angle. The other two angles, necessarily acute, must sum to 90°; such angles are said to be "complementary" to one another. A right triangle is pictured in Figure 1-7. The following trigonometric functions are defined for an acute angle within a right triangle: sine A =
opposite
=
a
hypotenuse c adjacent b cosine A = =— hypotenuse c opposite a tangent A = adjacent
(1-1)
b
"Opposite" and "adjacent" refer to the positions of sides of the triangle with respect to angle A and the ratios are of side lengths. Similar relationships may be written explicitly for angle B. The defined trigonometric functions are generally abbreviated "sin," "cos," and "tan." Table 1-1 contains trigonometric functional values and conversions between degrees and radians for angles between 0° and 90°. Three additional trigonometric functions—the cosecant (csc), the
C
a
b
FIGURE 1-7. A right triangle with angles and sides labeled symbolically.
12
Bioengineering Analysis of Orthodontic Mechanics TABLE 1-1. Degrees-radians equivalents and trigonometric functions Degrees 0 1
2 3
4 5 6 7 8
9
10
Radians
Sin
Cos
Tan
0.000
0.000
1.000
0.000
0.018 0.035 0.052 0.070 0.087 0.105 0.122 0.140 0.157 0.175
0.018 0.035 0.052 0.070 0.087 0.105 0.122 0.139 0.156 0.174
0.999 0.999 0.999 0.998 0.996 0.995 0.993 0.990 0.987 0.985
0.018 0.035 0.052 0.070 0.088 0.105 0.123
0.192 0.209 0.227 0.244 0.262 0.279 0.297 0.314 0.332 0.349
0.191
0.367 0.384
0358
22
0.375
0.934 0.927
23
0.401
0.391
0.921
24 25
0.419 0.436 0.454
0.407 0.423 0.438 0.454 0.470 0.485 0.500
0.914 0.906 0.899
0.515 0.530 0.545 0.559 0.574 0.588 0.602 0.617 0.629 0.643
11
12 13 14
15 16 17 18 19
20 21
26 27 28
0.471
30
0.489 0.506 0.524
31
0.541
32 33 34
0.559 0.576 0.593
35 36 37
0.611
38
0.663
29
0628 0646
39
0.681
40
0698
41
0.716 0.733
42 43 44 45
0.751
0768 0.785
0.208 0.225 0.242 0.259 0.276 0.292 0.309 0.326 0.342
0.982 0.978 0.974 0.970 0.966
0.141
Cot 1.571
57.3 28.6 19.9 14.3 11.4 9.51
8.14 7.12
0.158 0.176
6.31
0.194 0.213
5.14 4.70 4.33
0.231
5.67
59
1.15
0.855 0.838 0.820 0.803 0.785
49
Radians
Degrees
0.755 0.743 0.719 0.707
0.869 0.900 0.933 0.966 1.000
Cos
Sin
Cot
0.731
0.625 0.649 0.675 0.700 0.727 0.754 0.781
0.810 0.839
79 78
1.361
1.030 1.012 0.995 0.977 0.959 0.943 0.925 0.908 0.890 0.873
0.656 0.669 0.682 0.695 0.707
0.883 0.875 0.866
1.378
80
1.66 1.60 1.54 1.48 1.43 1.38 1.33 1.28 1.24 1.19
0.601
0.891
81
1.80 1.73
0.857 0 848 0.839 0.829 0.819 0.809 0.800 0.788 0.777 0.766
0.940
1.414 1.396
69 68 67
2.61
0946
1.431
89 88 87 86 85 84 83 82
1.204 1.187 1.169 1.152 1.135 1.127 1.100 1.082 1.065 1.047
0.384 0.404 0.425 0.445 0.466 0.488 0.510 0.532 0.554 0.577
0.951
1.484 1.466 1.448
90
77 76
4.01
0.956
1.501
-
1.344 1.327 1.309 1.292 1.274 1.256 1.239 1.222
0.249 0.268 0.287 0.306 0.325 0.344 0.364
0.961
1.553 1.536 1.518
3.73 3.49 3.27 3.08 2.90 2.75 2.48 2.36 2.25 2.14 2.05 1.96 1.88
1.11
1.07 1.04 1.00 Tan
75 74 73 72 71
70
66 65 64 63 62 61
60 58 57
56 55 54 53 52 51
50 48 47 46
45
13
Mathematical Topics
(sec), and the cotangent (cot or ctn)—are defined as the reciprocals of the sine, cosine, and tangent, respectively. The lengths of the secant
sides of a right triangle are interrelated through the Pythagorean theorem;
for the triangle of Figure 1-7 the relationship is = a2 + b2
(1-2)
or, in words, the length of the hypotenuse equals the square root of the
sum of the squares of the other two sides. Also, from the foregoing definitions and because A and B are mutually complementary: sin A = cos cot A = tan
tan A = cot csc A = sec
cos B = sin A sec A = csc B
B B
B
B
(1-3)
An oblique triangle, incorporating no right angle, is sketched in Figure 1-8. The law of sines for this triangle is
sinD
—
sinE
ci
e
—
sinF
(1-4)
f
Another relationship, valid for the oblique triangle, also relating sides and angles, is the law of cosines: d2 =
e2 =f2+
d2
2ef(cos D) —2fd(cos E)
+
e2
2de(cos F)
f2 =
e2
+ f2
(1-5)
Although the oblique triangle of Figure 1-8 is specifically an obtuse triangle, these two laws may also be used in analyzing an acute triangle.
d
e
FIGURE 1-8. An oblique triangle with angles and sides labeled symbolically.
14
Bioengineering Analysis of Orthodontic Mechanics
y L
0
x
FIGURE 1-9. A line segment referred to an x-y-coordinate framework.
For the right triangle, the sine and cosine laws are replaced by the defini-
tions of the sine and cosine functions and the Pythagorean theorem. Figure 1-9 shows a line drawn in the plane of the page with the origin of an x-y-coordinate framework coinciding with one end of the line segment. Because the reference frame is rectangular, the line makes complementary angles with the x- and y-axes. These angles, labeled and are called the direction angles of the line OL. The "shadow" or projection of the line segment on the x-axis is the apparent length of the line seen from a vantage point far out on the y-axis; that length equals OL cos Similarly, the projection of OL on the y-axis is equal to OL cos The cosines of the direction angles are termed direction cosines. Enclosing the line OL in a rectangle, having adjacent sides coincident with the x- and
y-axes and making the segment OL a diagonal of the rectangle, and noting the two equal right triangles formed (which share a common hypotenuse), the following equations may be written: (x-projection)2 + (p-projection)2 = 0L2 + = 1
(1-6)
The development just completed using Figure 1-9 may be extended to three dimensions. In Figure 1-10 the coordinate framework is again
appended to one end of the line segment (although, in general, the segment may be remote with respect to the frame); the end points of the segment are (0,0,0) and (XN,YN,ZN). The direction angles are defined as in Figure 1-9, except there are three; for instance, in Figure 1-10 the angle 0,
15
Mathematical Topics
\
N
0 z
Q
\\ a y
ox
x
FIGURE 1-10. A line segment in space referred to an x-y-z-coordinate framework.
measured between the x-axis and the line segment and that measurement is made in the plane containing ON and the x-axis. The projections of the segment ON on the coordinate axes are is
y-projection = QN(cos x-projection = ON(cos z-projection = ON(cos
0!,)
respectively, and the three projections are also related to the coordinates of end points of the segment and its direction cosines: x-projection
0 XN y-projection = z-projection = ZN — 0
YN
—
0
(1-8)
An alternative two-step process may be used to obtain the projections. First, for example, project the line on the y-axis and into the x-z plane. The angle 4, measured between the segment ON and the x-z plane, is complementary to and the projection in the plane equals ON cos q5. Second, decompose the x-z plane projection, OQ, into x- and z-axis projections using the complementary angles a and y, measured in the x-z plane. The three projections may then be expressed as follows:
16
Bioengineering Analysis of Orthodontic Mechanics
x-projection = ON(cos 4)cos a
y-projection = ON(cos Os,,) z-projection = ON(cos q5)cos y
(1-9)
The length of the line segment is related to the coordinate-axis projections through the three-dimensional form of the Pythagorean theorem: (x-projection)2 + (y-projection)2 + (z-projection)2
(1-10)
Dividing through by the square of the segment length and noting the resulting forms defining the direction cosines, the expression becomes +
+
=
1
(1-11)
The comparable relationship in the two-dimensional situation, previously examined, is the complementary nature of the two direction angles. (No simple relationship among the three direction angles exists spatially.) Before closing this section, it is important to note that a three-dimensional geometry problem may always be decomposed into two or three
two-dimensional problems. The original problem is, seemingly, expanded from one into two or three, but the gain is in terms of each of the
two or three "projections" being in its own plane setting.
An Introduction to Vector Algebra
physical quantities associated with orthodontic bioengineering possess characteristics in addition to magnitude. A most important entity, the point or concentrated force must be known in magnitude, in direction (and sense), and in point of application to be fully described. A multifaceted description is similarly associated with the displacement of a particle or a point of a solid body. Both the point force and displacement obey the laws of vector mathematics. A vector is a mathematical quantity possessing both magnitude and direction. The equations describing vectors are unique, and vector quantities combine with scalars and with other vectors according to specific mathematical laws and procedures. Vector quantities may be expressed in both graphic and analytic formats. The vector is pictured as a directed line segment. When drawn to scale, the length of the segment is proportional to the magnitude of the vector. The inclination or orientation of the segment, the angle between it and a specified reference line or axis, indicates the direction of the vector. The sense of the vector is denoted by an arrowhead affixed to one end of the segment. Given all other characteristics of a vector, with only two possibilities for sense, it may be specified algebraically. Sense is often linked with direction; two vectors having identical angles with a reference axis, but differing in sense, are said Several
17
Mathematical Topics
be opposite in direction to one another. Without being specific as to the physical quantity represented, a vector is shown in Figure 1-11 with an x-y-coordinate frame included in the sketch that serves as a reference to
for direction and sense. The displayed vector, indicated as having a magnitude of 100 units with the directed line segment subdivided into 5 equal length units, implies a linear scale factor of 100 to 5 or 20 to 1. Figure 1-11 is similar to Figure 1-9 except for the absence of the arrowhead in the latter sketch and the coordinate framework not attached to an end of the line segment in the former. The angles labeled in Figure 1-11 are those which the line of the vector, if extended, would make with the x- and y-axes; therefore, these angles are properly termed the direction angles of the vector. Individual multiplications of the vector length by the direction cosines gives the x- and y-projections of the vector. Converting these (scalar) projections to vector quantities by including the senses and multiplying each by the scale factor for the sketch yields the
x-, and y-components of the given vector. The lengths of the x- and y-projections are indicated as 4 and 3 units; with a 20-to-i scale factor, the x- and y-component magnitudes are 80 and 60, respectively. The rectangular components of a vector are, by definition, mutually perpendicular and individually parallel to a coordinate axis. Hence, the sum of the squares of the component magnitudes (or the lengths of their line-
segment representations) equals the square of the magnitude (or the length) of the original vector. The analytic expression for a vector is typically written in terms of its component magnitudes (projections) and unit vectors (dimensionless and of magnitude one). Vector quantities appear in boldface type in this
y
60 units
Scale: 1:20
80 units
0 FIGURE 1-11. A vector positioned in the x-y plane.
X
18
Bioengineering Analysis of Orthodontic Mechanics
text to distinguish them from scalars. The unit vectors i and j, defined as
associated in direction and sense with the x- and y-directions, are shown in Figure 1-11; note the positive senses corresponding to increasing coordinate values. The equation for the vector displayed may be written as V = 80i + 60j
(1-12)
Shown in Figure 1-12 is a vector in space. The ends of the vector are coincident with points located with respect to the chosen x-y-z-coordinate framework; the "tail" of the vector coincides with point D(3, 1,2) and the vector symbol extends to point E(5,5,6). In this example, the vector represents a 60-g force. A parallelepiped, oriented to the coordinate frame, "surrounds" the vector, which lies on a main diagonal. The length DE of the force vector is found through the three-dimensional form of the Pythagorean theorem: DE = [(5
—
3)2
+ (5
—
1)2 + (6
—
2)21112 =
(1-13)
6
The scale factor for the sketch is, then, 60 g to 6 length units or 10 g to 1 length unit. The lengths of the sides DA (2 units), DB (4 units), and DC (also 4 units) of the parallelepiped correspond to the projections of the given vector, which are parallel to the x-, y-, and z-axes, respectively. The direction angles and of the force vector are the angles EDA, EDB, and EDC, respectively. Similarities in and differences between Figures 1-10 and 1-12 should now be apparent. Defining k as the unit vector y
/
/ /7 y,/
//
/'/I / /
4
1/
/11/
x
1 0(3,12) E(55,6)
Scale:
FIGURE 1-12. A vector referenced to a three-dimensional framework.
1
unit = log
19
Mathematical Topics
parallel to the z-axis and with sense corresponding to increasing z, the
vector equation for the 60-g force may now be written:
F=20i+40j+4Okg
(1-14)
The terms of Equation 1-14 are the x-, y-, and z-components of the force vector. By way of application, Figure 1-12 might be displaying a force exerted by a stretched elastic, pulling from a point (3 cm, 1 cm, 2 cm) in a lateral-anteroposterior-vertical reference frame. (Continuing, the point
(5 cm, 5 cm, 6 cm) is on the line of the force, but the "head" (arrowhead) of the vector located at that point in the figure is merely a result of the particular choice of scale factor for the sketch.) Equation 1-14 charac-
terizes the force as to magnitude, direction, and sense at a particular time, but does not indicate the point of application of the force. To complete the displacement discussion begun in an earlier section of this chapter, a vector addition must be carried out. The displacement of point D, the root apex in Figure 1-5, equals the sum of the displacement vector of point C in the crown and the relative displacement, D with respect to C. In Figure 1-11 the vector components having magnitudes of 80 and 60 units "add" to give the resultant vector V with magnitude of 100 units. (The sum of two vectors depends on both the individ-
ual magnitudes and their directions and senses.) In that example, the (rectangular) components were perpendicular to one another; the components of the apex displacement vector are not 900 apart. Note, however, that the parallelogram law of vector addition, when employed graphi-
cally, is easily undertaken regardless of the relative orientations of the vectors to be summed. The vector addition of the two displacement components is shown in Figure 1-13. The displacement of point C, in reality obtained, perhaps, from successive cephalograms, is drawn to a scale; in this example, the scale magnifies the actual displacement size. The magnitude of the relative displacement is the scalar product of the change in long-axis angulation (in radians) and the length of the axis segment CD. The directions and senses are from C to C' and D" to D', respectively, from Figure 1-5. To accomplish the summation, the two vectors are placed in series, with the tail of one meeting the arrowhead of the other, and they form adjaC (or D)
d(isplacement)D
d010
C, (or D")
FIGURE 1-13. Vector addition. The relationship among the displacement vectors of points in Figure 1-5.
20
Bioengineering Analysis of Orthodontic Mechanics
cent sides of a parallelogram. The third and fourth sides of the parallelo-
gram are added (dashed) to complete the figure and the desired vector, the absolute displacement of point D, is coincident with a diagonal. The length of the resultant vector is measured and multiplied by the scale factor to obtain the magnitude of the displacement of point D. Because the two vectors summed in Figure 1-13 are nearly parallel and have comparable senses, the result in this example is almost as if the component magnitudes were added. Parallel components, however, are the exception and not the rule; the 80-unit and 60-unit components in Figure 1-11 sum to give a resultant magnitude of 100 units, not 140. Additional examples of vector addition are presented in Chapter 2.
Dimensions and Units The quantifiable parameters of orthodontic bioengineering are generally
dimensional and their magnitudes depend upon the units in which the parameters are expressed. Four dimensions are basic or fundamental: length, time, mass, and temperature. All other dimensions of interest may
be derived from these four; the most important is the dimension of force. The result of a dimensional analysis of an equation of classical physics yields force as the product of mass, length, and the inverse of time squared [mass(length)/(time2)}. Dimensions help to describe a quantity, and also to distinguish among apparently similar parameters. For example, pressure is dimensionally mass/length!(time2) or force! (length2) and, therefore, is not identical to, or a type of, force, although the two are related. One quantity already discussed, which is dimensionless (or nondimensional), is angular measurement; another, which will be introduced in Chapter 3, is strain. The numerical value assigned to a dimensional quantity (and some nondimensional quantities as well) is meaningful only when units are appended to that value. Each fundamental or derived dimension generally has several units or sets of units that may be associated with it. The numbers 1, 3, and 36 are different from each other, but when used as expressions of length in yards, feet, and inches, respectively, they are identical.
Because of biologic and materials-science influences on their field, dental practitioners are familiar with two systems of units. Metric units are common to the life sciences; length is measured in meters and mass in grams or fractions!multiples thereof (e.g., millimeters [mm] and kilograms [kg]). In this country the physical sciences have long employed the American engineering system of units; typically, lengths are in feet and forces in pounds-force. In both unit systems the basic unit of time is the second. The derived, composite unit of force in the metric system is
the "newton;" the gram-force is more common now in the United States, but is a distortion of the metric mass unit. In the American engi-
21
Mathematical Topics
neering scheme, force is considered a fundamental dimension and the
"slug" is the derived unit of mass (although pound-mass is also used in a manner comparable to gram-force). Conversions from one metric-units set to another is straightforward; "kilo" indicates 1000 and "centi" means "divide by 100," for example. Such conversions in the American engineering system are more difficult (e.g., 5280 ft/mile and 16 oz/lb);
this is probably a principal reason why the scheme has never been widely accepted outside of the United States. Orthodontists, until now purchasing arch wires having cross-seclions measured in inches, but themselves measuring arch lengths and discrepancies in millimeters, should be pleased about the pressures from abroad toward a worldwide move to the Système International d'Unités (French), or the SI units system. A modified metric scheme, the fundamental SI units of length, time, mass, and temperature are the meter, the second, the kilogram, and the degree Kelvin, respectively. Examples of SI units associated with derived dimensions include the newton (force), the pascal (stress or pressure; one newton per square meter), and the joule (energy; one newton-meter). Currently, the conversion to one international system of units remains incomplete. Hence, the clinician must possess some proficiency in converting from one unit
or set of units to another in the same system and in transposing or converting between the metric and American engineering schemes. Table 1-2 is provided to assist the reader with a partial listing of conversion factors within and between unit systems.
ants, Computations, and Numerical Accuracy Although not professing to be mathematicians, orthodontists are con-
fronted with, and make use of, a substantial amount of quantitative information. Treatment planning is influenced by computations performed after measurements are taken from casts and cephalograms. Today's vendors of wires, elastics, and the whole array of appliance ele-
ments are oriented in the materials and engineering sciences, and clinicians must be able to communicate appropriately with them. To understand the mechanical and structural functions expected of their appliances, and to match sets of appliances properly with the treatment plan, the practitioner must be able to undertake at least a cursory analysis of the force systems created by activation of those appliances. Physiologically-proper forces, to be exerted on the dentition, must be approximated. To follow the development and examples in this text, and more importantly to make clinical use of the knowledge to be gained in the study of the bioengineering topics to follow, the mathematical tools discussed in this chapter become indispensable. When possible, emphases will be placed on the appropriate graphic techniques to carry out an analysis, in
particular to help visualize relationships among variables and in the
22
Bioengineering Analysis of Orthodontic Mechanics TABLE 1-2. Conversions within and between units systems
Dimension(s)
Length (L)
Area (L2)
Multiply mils
0.001
centimeters meters inches
100 25.4
inches2 centimeters2
0.155
pounds pounds pounds ounces kilograms-force
Force (F)
by
10
645 16
454
4.45 0.278 2.2
to Obtain inches millimeters millimeters millimeters millimeters2 inches2
ounces grams-force newtons newtons pounds
Energy (F-L)
inch-ounces foot-pounds
720 1.36
gram-millimeters newton-meters (joules)
Stress (FIL2)
pounds/inch2 pounds/inch2 pounds/inch2
0.704 0. 07 04
grams/mm2 kilograms/cm2 newtons/meter2 (pascals)
Temperature (T)
degrees F—32
degrees C
6900 5/9 9/5
degrees C degrees F—32
Time (t)
seconds minutes
3600 1440
hours days
Angle
radians
57.3
degrees
1. Degrees F = °Fahrenheit Degrees C = 2. Angular measurement is nondimen-
sional but is quantified in degrees or radians.
3. Joules (J) and pascals (Pa) are derived units in the SI system. 4. Absolute temperature is measured in °Kelvin = °Celsius + 273 or kine = ¶ahrenheit + 460
manipulations of force and displacement parameters. When necessary, analytic procedures will be used and dimensional and units analyses included as a partial check on the correctness of terms and equations. In any analysis requiring mathematical manipulations, after an overview of the entire undertaking and coming to a decision regarding the appropriate procedure to be followed, the input or given data must be gathered or examined. The techniques of obtaining linear and angular measurements, and the accuracy of these data, are expected to be those ordinarily achieved in acceptable cephalometric exercises. Determinations of force magnitudes should generally be to the nearest ½ oz or 15 g. When input data for a problem to be solved originate from a previous statistical analysis, the accuracy of that data is to be determined; if
mean values of parameters are to be used, their standard deviations should be sought.
23
Mathematical Topics
Analyses within this text are to be undertaken and solutions pursued through principles of force and structural mechanics interfaced with
knowledge obtained from clinical experience and experimental research.
Quantifiable results will be sought by using mathematical procedures
introduced in this chapter. After any solution has been obtained it should be carefully checked. Two possible means of checking results are
(1) to proceed to the solution by an alternate method or approach, if available, or to change the order of steps in the solution procedure followed, or at least to substitute the numerical answers obtained into the mathematical relationships initially used to ensure satisfaction; and (2) to make sure that no mistakes exist in dimensions or units in the procedural steps. A dimensional analysis of an algebraic equation, for example, must show identical, net dimensions and units for each term. The solution to a problem involving mathematical operations is only as accurate as the data employed and the accuracy of the computations performed. Deficiencies in either can render a solution inadequate, even if the procedures are all correct. An accurate set of calculations cannot make up for inaccuracies in initial data. Hand-held, electronic calculators and digital computers will yield numerical solutions to six or more
figures, but no more digits should appear in the answer than can be justified. The evaluation of the accuracy of mathematical procedures is a complex matter and will not be investigated here. It is perhaps sufficient to indicate that most input data in orthodontic bioengineering problems are accurate to no more than three significant figures and, therefore, no more than three significant figures should be recorded in the solutions obtained.
Readings Davis, H.F., and Snider, A.D.: Introduction to Vector Analysis. 4th Ed. Boston,
Allyn & Bacon, 1979, Chapter 1.
Drooyan, I., Hadel, W., and Carico, C.C.: Trigonometry. 3rd Ed. New York, Macmillan, 1979, Chapters 1, 4, 5, and Appendix A. Greener, E.H., Harcourt, J.K., and Lautenschlager, E.P.: Materials Science in Dentistry. Baltimore, Williams & Wilkins, 1972, Chapter 1. Spiegel, M.R.: Mathematical Handbook of Formulas and Tables. Schaum's Outline Series. New York, McGraw-Hill, 1968.
Spiegel, M.R.: Theory and Problems of Vector Analysis. New York, Schaum, 1959, Chapter 1.
Thurow, R.C.: Atlas of Orthodontic Principles. St. Louis, C.V. Mosby, 1970, Chapter 5. Wilson, W.A., and Tracey, J.I.: Analytic Geometry. Boston, D.C. Heath, 1949, Chapters 1—3, 13.
Introduction to Analysis of Orthodontic Force 22 The activation of an orthodontic appliance creates a system of forces that
are transmitted through the members of that appliance to the dentition, there providing the potential to produce displacements of dental units. Relationships exist between force systems and potential displacements of bodies upon which the forces act. Hence, an understanding of the fundamental principles and procedures of force analysis is indispensable to the practitioner who fabricates, places, and activates the appliance toward the desired end: the controlled movement of teeth. A large portion of the analysis of forces to be discussed has its foundations in the subfield of classical physics known as particle mechanics. Several of the basic concepts may be familiar to the reader from an undergraduate survey course in general physics. Nevertheless, the discussion here commences with no assumptions of previous knowledge of the subject.
Mechanics as an area of study within the physical sciences, is concerned with the state of rest or motion of bodies subjected to forces. Mechanics is subdivided by the physicist according to the sizes of the bodies studied (from the smallest subatomic parts of matter—quantum mechanics— to the planets and galaxies—astronomy) and with respect to the degree of motion (from no movement whatsoever to bodies moving at rates of the order of the speed of light). Although evidence exists that the early Greek scientists understood and used some aspects of mechanics several hundred years nc., the fundamental principles of particle mechanics, as understood today, were formulated by Sir Isaac Newton in the seven-
Mechanics,
teenth century. Newtonian mechanics pertain to bodies of moderate size, if moving then not at the excessive speeds where mass and velocity
may become relatively, mutually indistinguishable, and is separable from relativity theory and celestial mechanics. 24
25
Introduction to Analysis of Orthodontic Force
Newtonian mechanics is divided into branches known as statics,
which treats bodies at rest and under the action of forces, and dynamics, which deals with moving bodies. Dynamics is further subdivided into kinematics, the study of motion itself, and kinetics, in which relationships between the force systems and the characteristics of body motion are explored. The subareas of study within mechanics also may be categorized according to interest in whole-body motions and displacements of solids, the deformations that solid bodies experience under loading, and the flow of fluids. The principles of Newtonian mechanics, together with their interactions with the biologic systems and processes involved, govern the study of orthodontic forces. The movements of individual teeth are examples of whole-body displacements. Because the activation of the orthodontic appliance requires deformation of one or more elements of the appliance, applicable principles of mechanics of deformable bodies are discussed in subsequent chapters.
This chapter first introduces the principles of Newtonian partide mechanics. The extensions of these principles, enabling the treatment of bodies that are large in comparison to the particle, are then discussed. Some modeling (idealizing from the actual) and approximations will be
found necessary or beneficial and, in each instance, the validity and limitations of the model or approximation must be assured and understood. Although introduced individually, note that in large measure the concepts and principles "fit" together or accumulate to ultimately form a packaged procedure to be employed in analyses wherein total descrip•tions are required of force systems induced in the orthodontic appliance
and in the dentition to which the appliance is attached.
Force is defined as an act upon a body that changes or tends to change the state of rest or the motion of that body. The categories of forces are numerous, and forces are described in many different ways. Two bodies are associated with every force; one body exerts the force and the other receives the force. A first manner of categorizing forces is according to the proximity of the two involved bodies with respect to each other. Contact Force
forces arise due to actual physical contact between the bodies; body forces exist between bodies that are some distance from one another. The most common, sizable body force is the attraction of the Earth for a body on or near the Earth's surface: the weight of the body. Newton's
gravitational law enables the determination of the weight of a body, which is directly proportional to the product of the masses of the Earth and the body and inversely proportional to the square of the distance between the centers of the Earth and the body. Other body forces include those existing in the presence of electric or magnetic fields, but these do not occur routinely in the study of orthodontic force systems and thus will receive little attention. Notably, the weights of teeth and
26 Bioengineering Analysis of Orthodontic Mechanics
appliance parts are small and may be neglected in comparison to the
contact forces present within the activated appliance. In analysis procedures, forces are associated with the areas or volumes of bodies with which they directly interact, and contact forces are associated with the surface areas over which the contact exists. If the contact area is small in comparison with the total surface area of the body, the force is modeled as exerted on a surface point or particle of the body and is termed a concentrated or point force. On the other hand, if the contact area is relatively large, the pattern of the force intensity over
the area may be an important consideration and, therefore, the force is explicitly termed distributed. For example, the force exerted by an elastic stretched against a hook, soldered to a bracket or to an arch wire, is reasonably concentrated, but the force arising between the periodontal ligament and the root, in response to appliance activation transmitted to the tooth, must be analyzed with regard to its distribution over the sizable surface area of contact. By definition, forces exerted on an individual particle are point forces that may actually or potentially only pull or push the particle. As previously mentioned, two bodies are necessary to the existence of every force. Moreover, whenever one force is created or present, a counterpart exists as well; when one body exerts a force on another, the second body also exerts a force on the first. Between the two bodies there is an action and a reaction, and which force is which depends solely
on point of view. In analysis procedures, however, just one of the two bodies will be examined at a time and that one body will be studied with respect to the forces exerted on it (not those forces exerted by it). Reactive
forces must not be confused with responsive forces; reactions exist as counterparts to both active and responsive forces. In the study of a body subjected to forces, the active forces are those purposely created or those that "load" the body; the responsive forces are also exerted on the body,
but are a direct result of the creation of active forces. The responsive forces are often undesired, but being present they must be included in the analysis. As a final categorization of forces for present purposes, forces are said to be external to a body under study when they are exerted by other bodies not part of the specific analysis. Internal forces arise within a body as a result of loading or activating the body under study. These internal forces "carry" or transmit the action through the body to the responsive locations, for example, to connections with other bodies.
In addition to the gravitational law, Sir Isaac Newton formulated a set of three laws governing the mechanics of a particle. Newton's First Law states that a particle subjected to a balanced system of concentrated forces will remain at rest, if originally at rest, or will move with constant speed in a straight line, if originally in motion. ("Balanced" force system: the net force is zero.) The Second Law indicates that, if the particle is subjected to an unbalanced system of forces, the particle will be accelerated in the direction of the net force exerted; in equation form, the net
27
Introduction to Analysis of Orthodontic Force
force and the product of the particle mass and its acceleration, with the
three quantities expressed in consistent, related units, are identical in magnitude, direction, and sense. (Acceleration is the time rate of change of velocity and, like the concentrated force, is mathematically a vector quantity.) The Third Law states that paired active and reactive forces are
equal in magnitude, but are directly opposed to one another and are exerted on adjacent particles. A corollary to the Third Law is the existence of internal forces, as previously defined, in canceling pairs; the ramification of this corollary is demonstrated later in this chapter. Newton's First Law might be considered a special case of the Second Law, and in orthodontic force analysis a strong argument may be made to the point that the inertia of an appliance element or a tooth is negligibly small, in particular after the almost instantaneous response, in the form of soft-tissue deformations, immediately following the activation of an appliance. If the inertia, the product of mass and acceleration for a particle, may be neglected in comparison with the individual components of the force system created through activation, the orthodontic system may be said to exist in a kind of quasi-static state; the displacements that take place are small and occur over a relatively long period of
time and, at any instant, a force analysis may be carried out without erring appreciably, as if the system was at rest. Clearly, this approximation has profound consequences; for the purpose of most analyses involving the entire force system, a dynamic body or group of bodies is effectively replaced by a static body or assembly. In Chapter 1 the concentrated force was reported to fulfill the mathematical requisites of a vector quantity. In the complete description of a point force, then, its characteristics as a vector must be stated: magni-
tude, direction, and sense. Moreover, from a physical standpoint the location of the force on or in the body must be given. Figure 2-1 shows a point force exerted on a molar crown. From the sketch, the force is seen to have a magnitude of 200 g; in this two-dimensional idealization, the direction of the force (and that of the dashed line of action of the force) is 700 from the long axis of the tooth. The sense of the force may be said to be up and to the left with respect to the sketch as a whole, and the point of application is point B with the force pushing on that point. (If desired,
an equation in vector form may be written for this or any point force, following the establishment of a reference frame, and all of the necessary characteristics to describe the force, except point of application, would be contained in that equation.) Figure 2-1, then, provides a complete visual description of the 200-g force. The potential or actual effects of a force exerted on a body, whether active or responsive, are displacement and/or deformation. Although both effects are of interest in orthodontic structural analysis, only the association of displacement with force is discussed in this chapter. Disregard-
ing deformation resulting from force application is tantamount to as-
suming that the body, on which the force or force system acts, is perfectly rigid. Although this is a reasonable assumption for some bod-
28
Bioengineering Analysis of Orthodontic Mechanics
200 g
FIGURE 2-1. A concentrated force exerted on a tooth crown.
ies, including individual teeth, it is not a valid assumption for others, for example, elastics. In considering effects of force application, however, bodies that are clearly nonrigid may be reasonably analyzed only within a framework that recognizes differences in passive and active configurations.
The rigid body, then, is a model of Newtonian mechanics used in analysis procedures when only completion of the description of the total force system or the whole-body displacement is sought. (In Chapter 1,
the displacement of point D of the tooth was obtained within the assumption that the tooth was rigid; otherwise, the relative displacement, point D with respect to point C in Figure 1-5, could not have been described as it was.) An important feature of rigid-body mechanics allows the "sliding" of a point force to any location on its line of action found most convenient for analysis purposes. This principle of transmissibility states that the mechanical, whole-body effect of a concentrated force, acting at a specific point, is unchanged by replacing the given force by another force having the same magnitude, direction, and sense as the original force, but acting at some different point on the line of action. Figure 2-2 shows an 80-g force in two different locations on its line of action (dashed); the potential for displacement of the tooth as a whole is the same, whether the force is pulling on the lingual side or pushing on the labial crown surface. Before proceeding further, the general dependence of orthodontic
force upon the independent variable "time" must be mentioned. The
29
Introduction to Analysis of Orthodontic Force
—— 80 g
FIGURE 2-2. Two concentrated torces having identical displacement potentials. The principle of transmissibility.
concentrated orthodontic force often changes with time and that change
is typically reflected in the magnitude characteristic, although alteration of direction may also occur with time. Magnitudes of force generated by appliance activation decrease during the between-appointment periods and an accompanying change in displacement potential generally oc-
curs. The changes with time may be divided into short-term occurrences, primarily due to soft-tissue deformations immediately upon acti-
vation, and long-term changes associated with tissue remodeling. Magnitude increases and, perhaps, directional alterations take place when the orthodontist reactivates the appliance. Detailed discussions of these phenomena are undertaken later in this text; in the present chapter, force systems are analyzed as if all involved entities are "frozen" at a particular instant in time.
ition and Decomposition of Concentrated Forces Two
fundamental analysis manipulations with point forces are dis-
cussed in this section. First, toward enabling the prediction of the form and direction of a potential rigid-body displacement associated with a system of active forces, the forces are combined in such a way that the mechanical effect of the original system remains unchanged while the resultant is as simple as possible. Second, often in the process of obtaining the resultant of a force system, the replacement of a force by a set of force components with each component having a designated direction is worthwhile. Beginning the first procedure with the fundamental operation, a pair of concentrated forces having intersecting lines of action has, as its resultant, a single force. The two original forces define a plane; the resultant lies in that plane and its line of action intersects the point where the
30
Bioengineering Analysis of Orthodontic Mechanics
lines of action of the original forces come together. The means of reduc-
ing the two concurrent forces to a single force, mechanically equivalent to the combination of the original forces, is known as the parallelogram law of vector addition. To illustrate, Figure 2-3 shows a pair of point forces: one originating from an interarch elastic (EL) and the other from an extraoral appliance (HG); both are exerted on a maxillary molar bracket or buccal tube. The view is from the buccal on the right side. To obtain the resultant graphi-
cally, the directions and senses are sketched exactly as they exist in a two-dimensional picture of the appliance action. The force magnitudes are represented in the lengths of the vectors according to the given scale; because it is twice the magnitude, the length of the headgear force vector is twice that of the elastic force. The given forces intersect at point 0, the center of the bracket in the buccal view, and their vector representations form adjacent sides of a parallelogram. The dashed lines are added to complete the geometric figure. The vector representing the resultant is drawn as the diagonal of the parallelogram; the "tail" of the resultant vector coincides with the point where the "tails" of the given vectors
meet. Its sense is then known, the direction of the resultant may be determined with respect to a chosen reference line with a protractor, and the length of the vector representation is converted to the force magnitude through the scale as indicated. Note that if one of the forces is seemingly pushing on the point and the other is pulling on the point of intersection of their lines of action, one or the other vector must be moved along its line of action until either both "tails" or both "arrowheads" or "tips" of the vectors meet at the point. Then the parallelogram may be completed and the sense of the resultant is determined; either all three "tails" or all three "tips" meet at the common point.
0
EL = 200 g
HG = 400 g 0=
+ 02
Scale: 1 in. = bOg lengths = 2.8 in. H = 280 g = 28°
FIGURE 2-3. Graphical determination of the resultant of frio concurrent forces. The parallelogram law.
31
Introduction to Analysis of Orthodontic Force
With the aid of a rough sketch, the resultant of two concurrent point forces may be obtained analytically. Seen in the parallelogram construclion of Figure 2-3 is a pair of congruent triangles with a shared side, a
diagonal, coinciding with the resultant vector. Two sides and the included angle of each triangle are known through the characteristics of the given forces. The laws of cosines and sines may then be used to obtain the magnitude of the resultant force and the orientation of its line
of action with that of a given force or some other known reference. Either the vector lengths or the actual magnitudes of the given forces, because length and force size are proportional, may be used in the computations which would proceed as follows: R=
+ HG2 — 2(EL)HG(cos 0) [2002 + 4002 2(200)400(cos 400)]h/2
= 278 g or approximately 10 oz
sin0 EL
=
.
(2-1)
200
R
cb=27.5°
The rough sketch is necessary, accompanying the analytic solution, to determine and illustrate the sense of the resultant force and location of its line of action. If the two forces to be "summed" are perpendicular to one another, the graphical procedure remains essentially unchanged, but the parallelogram becomes a rectangle. The pair of congruent triangles become a set of adjacent right triangles, and the analytic solution is simplified somewhat with the cosine and sine laws replaced by the Pythagorean theorem and the definition of one trigonometric function. An example solulion is given in Figure 2-4. The vector summations of Figures 2-3 and 2-4 may be handled alternatively, using a corollary of the parallelogram law: the triangle law.
R
Q2
= 36 = 6.8 oz
P = 3 oz
I tan 6 =
Q = 6 oz
P
0 = 27°
FIGURE 2-4. The resultant of two concurrent, mutually-perpendicular forces.
=
3
= 0.5
32
Bioengineering Analysis of Orthodontic Mechanics
0
JR
HG
R
P
0 Q
FIGURE 2-5. Resultants obtained for the given forces of Figure 2-3 (top) and Figure 2-4 (bottom) using the triangle law.
Graphical solutions for the examples of Figures 2-3 and 2-4, using the triangle law, are shown in Figure 2-5. Recall that the "tails" or "tips" of the given force vectors must coincide when using the parallelogram law; the solution by the triangle law requires that the "tip" of one given force and the "tail" of the other be coincident. The resultant is drawn from the "tail" of one given force to the "lip" of the other so that the three forces form the sides of a triangle. Note that in summing two concurrent forces by means of the triangle law, one given force may have to be displaced to an artificial location in order that the resultant emerges in the construction along its proper line of action; alternatively, keeping both given forces on their original lines of action in the graphical approach, the resultant obtained is correct in all characteristics except for its actual point of application. Hence, care must be taken in using the triangle law so that the proper line of action of the resultant is ultimately determined and designated correctly.
33
Introduction to Analysis of Orthodontic Force
The resultant of any number of point forces having lines of action all intersecting at one common point is a single, concentrated force; the line of action of the resultant passes through that same point. To determine the resultant of a system of three or more concurrent forces, the parallel-
ogram law or triangle law may be used repeatedly. Two of the given forces may be combined into one, that one combined with a third given force, and so forth until all of the given forces have been included in the process; the resultant emerges from the final combination involving the
last given force. Because any two of the concurrent forces define a unique plane in which their resultant will lie, the two procedures may be used in a three-dimensional problem. If all of the concurrent forces act in the same plane, the triangle law may be extended to a polygonal law.
Graphically, the given forces are drawn to scale with the "tip" of one touching the "tail" of the next until all have been included, in any order desired. The magnitude, direction, and sense of the resultant are determined by closing the polygon with the vector resultant having its "tip" touching that of the last given force drawn (and its "tail" touching that of the force with which the sketch was begun). The number of sides of the polygon is one more than the number of given forces. In this construction all but two of the given forces must be moved from their actual lines of action. The line of action of the resultant is known, though, to pass through the point common to the lines of action of the given forces. An example of the procedure using the polygonal law is illustrated with three given forces in Figure 2-6. In this example the forces are "added"
in the order indicated by the equation in the figure. The dashed lines suggest the graphical procedure by two successive applications of the parallelogram law. In a force-system analysis, the decomposition of given forces into com-
ponents is often desirable. The procedure is the reverse of that of the parallelogram or triangle law of vector addition. In Figure 2-3 the 200and 400-g forces could be considered components of the 278-g force. In analyses of orthodontic force systems, however, the component directions often coincide with the mutually perpendicular, dental-coordinate directions. In Figure 2-4, with proper orientation, the 3- and 6-oz forces become occlusogingival and mesiodistal components of the 6.8-oz force. The number of mutually perpendicular, nonzero components equals, at
most, the number of coordinate dimensions in a given problem: up to two in the plane problem and three in the spatial problem. Although the determination of vector components perpendicular to one another was outlined in Chapter 1, by way of review consider the example of Figure 2-7 in which the occlusal and mesial components of the 300-g force against the molar are obtained. Graphically, the parallelogram (actually a rectangle because perpendicular components are desired) is constructed around the 300-g force as its diagonal, drawn
to scale. The components are adjacent sides of the rectangle; their "tails" touch that of the given vector. The component magnitudes are obtained from their lengths and the scale factor for the sketch. Analyti-
34
Bioengineering Analysis of Orthodontic Mechanics
V A = 4 oz
7-7
'7
7
B = 6 oz
0 R = 7.5 oz
,, 7, C = 5 oz
7,
7
7 R=A+B+C
FIGURE 2-6. Three concurrent, coplanar forces and their resultant obtained through the polygonal law.
cally, each rectangular-component magnitude is the product of its direction cosine and the magnitude of the given force as indicated in the figure. The three-dimensional problem of decomposition into mutually per-
pendicular components cannot be easily handled graphically. The analytic solution, however, can be a direct extension of that used in the plane problem. A rough sketch, particularly helpful toward determining component senses, should be used. In proceeding, first establish the coordinate directions and, perhaps, sketch the corresponding rectangular parallelepiped around the given force. Second, obtain a set of angles, either (1) the direction angles (angles between the given force and the coordinate axes), or (2) the angles to project the given vector into one of the coordinate planes and onto the axis perpendicular to that plane and, subsequently, to project the component in the plane onto the coordinate axes defining the plane. Again, the procedure illustrated by Figure 2-8 is a review of the spatial relationship between the vector and its mutually
35
Introduction to Analysis of Orthodontic Force
0
300 0
280 g
F0 = F cos 70° = 300(0.34) = 1000 Fm = F cos 20° = F sin 700 = 300(0.94) = 280 g FIGURE 2-7. Decomposition of a concentrated force into two rectangular components. The parallelogram law in reverse.
perpendicular components previously discussed in Chapter 1 (see Fig.
1-12). Note that only two of the three direction angles need to be meas-
ured or estimated; the third angle can then be determined from the geometric relationship requiring the sum of the squares of the directions cosines to be unity.
a
d
F
I, i/I
a: apical d: distal 0: occlusal I:
F=l2oz
lingual
Fd = 12 cos 30° = 10.4 oz F0 = 12 cos 75° = 3.1 oz
F, = 12 cos 65° = 5.1 oz
(cos 300)2 + (cos 750)2 + (cos 650)2 = 1
FIGURE 2-8. Three mutually perpendicular components of a concentrated force not located in any of the three coordinate planes.
36
Bioengineering Analysis of Orthodontic Mechanics
The Moment of a Concentrated Force action of a concentrated force upon a particle tends to produce, simply, a push or pull displacement; the particle is small and the force seemingly covers the particle as a whole. The magnitude, direction, and sense of the force, together with characteristics of the particle and its supporting structure, are collectively a measurement of the potential of the force to displace the particle. A force, exerted on a small portion of the total surface area of a relatively large body, tends to produce a displacement of the body which is generally rotational. The moment of a force about, or with respect to, a specified point or line is a measure of the potential of that force to rotate the body, upon which the force acts, about the particular point or line. Consider a rigid body that is hanging from a cord attached to just one point of the body. Label that point 0. A force F is exerted on the body so that its line of action does not pass through point 0. Define the line OB so that point B is on the line of action of the concentrated force F and the two lines intersect at right angles to one another. As depicted in Figure 2-9, the line OB and the line of action of force F define a plane. If the distance from point 0 measured along OB is d to the line of action of F, the magnitude of the moment of force F about point 0 is the product of the force magnitude F and the distance d. Given the manner of support of the body in the figure, the application of force F would displace the body from the position shown and cause an angular change in the initially vertical line through point G (the center of gravity, through which the weight resultant acts). The moment of a force about a point is a vector quantity; the direction of the moment vector is perpendicular to the plane defined by the force vector and the point. The sense of the moment vector is determined by a rule associated with the apparent rotational tendency, clockwise or counterclockwise as viewed from above the plane of the force and the point about which the moment is determined: out of the plane if counterclockwise and into the plane if clockwise. The moment vector is located such that its line of action passes through the point about which the moment is computed, and may be specifically positioned wherever convenient along its line of action. The moment vector of magnitude Fd is shown in Figure 2-9. Force and moment vectors may not be directly combined in a force-system analysis and, because the two types of vectors may appear in the same sketch, the short slash is incorporated near the arrowhead in the symbol for the moment vector. To obtain the moment of a force about a point, three quantities must be completely known: (1) the force itself; (2) the point about which the moment is to be found or moment center (point 0 in Fig. 2-9); and (3) the The
length of the moment arm extending from the moment center and traversing the shortest distance to the line of action of the force (line OB in Fig. 2-9). From the determination of the moment of a force about a point, the
37
Introduction to Malysis of Orthodontic Force
Mzz = F(d) = F(d)(cos 0)
FIGURE 2-9. A concentrated torce and moment vectors with respect to lines (axes) through point 0.
38
Bioengineering Analysis of Orthodontic Mechanics
potential of the force to produce rotation about a particular line through
the point may be obtained. The rotational potential with respect to a line, a moment axis, has the greater application in orthodontic mechanics. In Figure 2-9, the moment of F about point 0 is also the moment of the
given force with respect to the line Z-Z. To obtain the moment about another line through point 0, line L-L in Figure 2-9 for example, the magnitude of the moment about line Z-Z is multiplied by the cosine of the angle between L-L and Z-Z. The vector component is the directed projection of the moment about Z-Z on line L-L. Apparent now is the general, three-dimensional nature of the moment representation and computation. The moment axis typically does not lie in the plane of the force and the moment arm. In fact, if the moment axis is in the plane of the force and moment arm, either the axis intersects the line of action of the force and, therefore, the length of the moment arm is zero, or the axis is parallel to the force; in both cases the tendency for rotation about the specified axis is nonexistent, and, correspondingly, the moment with respect to the specific axis is zero. In the general problem, the force for which the rotational potential is sought
likely will not be perpendicular to the moment axis, and one of two approaches may be used: (1) the procedure followed in Figure 2-9 in which the moment vector with respect to point 0 was projected onto line L-L; or (2) the process of Figure 2-10. In the latter approach, the force R is decomposed into components parallel (P) and perpendicular (Q) to the moment axis S-S. Now, the moment of the resultant of several con-
current forces must be equal to the vector sum of the moments of the individual forces with respect to the same axis (a theorem established by the French mathematician Varignon in the late 1600s, before the advent
of vector algebra). Since the moment of P about axis S-S is zero, the magnitude of the moment of R equals that of Q: Qb; the sense is as shown in the figure according to the rule previously established. In problems in which the forces of interest all exist in the same plane, the moment concept may be put into a two-dimensional format. Rotational potentials are obtained with respect to axes that are perpendicular to the plane of the force. In sketches these axes will appear in end views, or as apparent points in the plane. Moment vectors with respect to these moment centers will all have a common direction: perpendicular to the plane of the forces. They may then be treated as algebraic quantities with the distinction between moments having clockwise or counterclockwise rotational potentials made merely by a sign convention. To determine the net rotational tendency with respect to a particular moment axis, owing to the action of several forces, either of two procedures may be followed. The moment vectors may be obtained for each individual force with respect to a particular point on the moment axis, these concurrent vectors may be combined into one by methods of vector addition already discussed, and the resultant moment vector may then be projected onto the moment axis to obtain the magnitude and sense of the rotational potential of the force system. Alternatively, the resultant of the system of forces may be sought and the moment of the resultant then found with respect to the specified moment axis.
39
Introduction to Analysis of Orthodontic Force
0
0 R
7 I
I
/J
Mss 1
= Q(b)
FIGURE 2-10. A force in space, decomposed into components parallel and perpendicular to a given line, and its moment vector with respect to that line.
The former procedure is employed in Figure 2-11 to obtain the net moment of a pair of concurrent forces with respect to the z-axis of the reference framework. In this instance a two-dimensional solution may be carried out in the plane of the given forces; the z-axis would be perpendicular to the plane of the forces and in the sketch it is, in effect, point 0. The curved-arrow notation indicates the senses of the moments of the 120- and 90-g forces and of the net moment. Two moment computations are displayed with magnitudes and senses obtained with respect to moment axes perpendicular to the plane of the forces through points O and E. The net moment being equal to the moment of the resultant of the two forces enables the determinations of the actual moment arms of the 150-g force with respect to points 0 and E; the moment-arm lengths
are symbolized by o and e, respectively. Viewing the 120- and 90-g forces as x- and y-components of the l50-g force, whether or not to decompose
a force into rectangular components to obtain its moment will depend upon the ease of the actual decomposition process and the establishment of the moment arms of the components versus the determination of the actual moment arm of the given force. The latter can often be a
40
Bioengineering Analysis of Orthodontic Mechanics
y
M0 = 90(3)1 + 120(2) 30 g-mm
= 420 g-mm J M0 30 o = -h-- = = 0.2 mm
d=
\ E(5,0)
ME
=
420
= 2.8 mm
x
0 FIGURE 2-11. Determination of the moment of a force about a point in two dimensions by using rectangular components of that force in the plane.
seemingly difficult geometry program while the moment arms of the
components become clear after the coordinates of the moment center and one point on the line of action of the force are known. The more complex problems of determining the net rotational effect of a system of forces, which may be two-dimensional but nonconcurrent
or three-dimensional in nature, have been outlined as to general approach. They are more easily discussed in detail, however, following the
introduction in the next section, "The Couple."
The Couple A pair of point forces having equal magnitudes and identical directions,
opposite in sense to one another and having noncoincident lines of action, when considered as one mechanical entity is termed a couple. Because the forces have the same magnitude, but are oppositely directed, the net potential of this special force system to translate the body upon which it acts is nil. The lines of action of the two forces are parallel, but are not coincident; as a result, this force system tends to rotate the body acted upon. Also, since the forces are nonconcurrent, they cannot
be combined and, therefore, the couple cannot be simplified further. The couple, occupying a specific plane as defined by the two forces, is illustrated in Figure 2-12. The rotational potential of the couple with respect to point 0, or about a moment axis perpendicular to the plane
through 0, is the sum of the moments of the individual forces with respect to the point. Using the procedure and notation of Figure 2-11,
41
Introduction to Analysis of Orthodontic Force
y
M0 = 300(1) 300 g
+ 300(3)
= 600 g-mm) = 300(2)
= 600 g-mm)
2mm
= 300(1)) + 300(1))
•
=600g-mm) M0 = 300(3)) + 300(1)
C(2,2)
= 600 g-mm) 8(11) 300 g
D(4,0)
0
x
FIGURE 2-12. A couple and computations of its inherent moment.
noting that with respect to point 0 one moment is clockwise and the
other is counterclockwise, the net moment is found to be 600 g-mm counterclockwise. Using the points B, C, and D as individual moment centers, the same moment is obtained in the three computations—the same as that with respect to point 0. In fact, for all moment centers in or moment axes perpendicular to the plane of the couple, the same result emerges. A couple has, then, an inherent moment, and the magnitude of the couple itself is that of the inherent moment, the product of the size
of one of the forces and the perpendicular distance between the two forces. Because the moments of the individual forces are vector quantities, so also is that of the couple and it follows that the couple itself may be represented mathematically as a vector. The direction of the couple is
associated with the plane of the pair of forces and the sense with its rotational potential, either clockwise or counterclockwise, as viewed looking into the plane of the couple. Figure 2-13 displays a couple and its vector representation (and that of its moment). The direction of the vector C is perpendicular to the plane of the couple (and the "moment arm" b between the forces) and the sense is determined according to the convention noted earlier for moment vectors: out of the plane if viewed as counterclockwise, into the plane if clockwise. Note the slash through the symbol, again to differentiate between force and moment vectors. A couple is completely described when specified are its magnitude,
the plane of the two forces, its sense, and, if the body is nonrigid, the actual points of application of the two forces. The value of the couple as
42
Bloengineering Analysis of Orthodontic Mechanics C = Pb
FIGURE 2-13. A couple in its plane and its vector representation perpendicular to that plane.
an analysis tool, however, is in the determination of whole-body displacement potential and, when so used, the specific points of application are unimportant. Although its vector representation in Figure 2-13 is shown midway between the two forces, because the moment of the couple is the same with respect to every moment axis perpendicular to the plane of the forces, the vector has no particular line-of-action location and may be drawn through any point of the plane of the couple. This "freedom" associated with the couple vector has far-reaching implications with respect to certain force-analysis procedures. To review, a force acting on a rigid body has the potential, generally, to both translate and rotate the body. Sliding the force along its line of
action will not change this potential. The tendency to rotate the body about a specific axis is associated with the vector characteristics of the force and also with the location of the line of action of that force because the axis and the line define a moment arm. Moving the line of action of the force with respect to the moment axis alters the force direction or the moment arm and, therefore, changes the translational or rotational potential of the force.
The couple, however, has been shown to possess an inherent rotational potential and no tendency to produce translation. Apparently, then, as long as the characteristics of the couple are unchanged, the forces making up the couple may be altered without affecting wholebody response; it follows, though, that an alteration of one force of the pair necessitates a comparable change in the other. Indeed, the use of properties of and operations on forces already discussed, including the
43
Introduction to Analysis of Orthodontic Force
parallelogram law, demonstrates that the following alterations may be
made without affecting the mechanical effect of the couple on the body as a whole: (1) the forces may be moved ("slid") along their individual lines of action; (2) the forces may be rotated in the plane of the couple, keeping their lines of action mutually parallel and the distance between
them unchanged; (3) the couple may be moved as a whole from the given plane to another parallel plane; and (4) the magnitudes of the forces may be altered, while preserving the equality of force magnitudes
and the moment of the couple, the latter by making a corresponding change in the distance between the lines of action of the forces. Figure 2-14 shows four mechanically-equivalent couples, one being a
curved-arrow representation which is convenient in two-dimensional analyses. These permissible movements of the couple, and the allowable correlated changes in the force magnitudes and the distances between
their lines of action, are embodied in the vector representation of the couple. The couple acting on a rigid body is, mathematically, a "free" vector, having no specific line-of-action location as noted previously. Frce and moment vectors may not be directly combined, and the slash through the moment-vector symbol is a reminder of this fact. Perhaps unfortunately, however, no symbolic distinction is made that would indicate the difference between the vector representing the moment of a single point force and that depicting a couple with its inherent
800 g 200 g .75 mm
-.
mm
200g
600 9-mm
800 g
FIGURE 2-14. Four mechanically-equivalent couples.
300 9
44 Rioengineering Analysis of Orthodontic Mechanics
moment. Note that the moment vector, representing the rotational po-
tential of the concentrated force with respect to a specific moment center or axis, is not a "free" vector. On the other hand, because a couple is a
free vector and through its two-force representation, a couple may be combined with a third force. The form of the resultant from this combination is particularly dependent on the direction of the third force with respect to the couple vector. Figure 2-15 shows the relative position of the force and couple vectors, enabling the placement of the two forces of the couple in a plane containing the third force, which leads to a simple resultant. Figure 2-16 illustrates the combination procedure. With the three forces in the same plane, the couple is represented as two forces having magnitudes equal to that of the single force; the distance between the forces making up the couple is then determined. Next, the couple is translated and rotated in the plane so that one force of the couple shares the line of action of single force and is opposed (in sense) to it. The now directly opposite forces cancel one another and one concentrated force from the original three remains, having the same vector characteristics as the original single force, but with a line of action some distance from it. The procedure of combining a force and a couple may also be carried out in reverse. Given a single force having specific characteristics, including location of line of action, a force and a couple may replace it without altering the whole-body displacement potential. The transla-
800 g-mm
H
FIGURE 2-15. Mutually-perpendicular force and couple vectors referenced to the plane of the force system.
45
Introduction to Analysis of Orthodontic Force 200 g
200g
200g
/
mm
/
/
200 g
I B
200 g
200g
/
FIGURE 2-16. Reducing the force and couple of Figure 2-15 to a single force.
tional potential remains unchanged if the force portion of the new sys-
tem has the same vector characteristics as the given force. To leave the rotational potential unaltered, the magnitude, direction, and sense of the couple must correlate with the position of the line of action of the force of the substituted system with respect to that of the given force. In Figure 2-17, in occlusal view the force exerted on the edgewise orthodontic bracket is "moved" by means of the addition of the couple to a parallel position through the long-axis of the tooth. Although seemingly adding to the complexity of a force system, the procedure provides a means to "free" the force vector from its specified location and permit lateral movement of its line of action. A couple is a special two-force system. In orthodontic application, active couples are often created upon bracket engagement of arch wires a
C = Qe
FIGURE 2-17. Moving the line of action of a force 0 through introduction of a couple C. The reverse of the process illustrated in Figure 2-16.
46
Bioengineering Analysis of Orthodontic Mechanics
into which permanent bends or twists have been placed. The dimension
of the couple magnitude is the product of force and length. The couple tends to cause only rotation of the body on which it acts. In addition to its importance as a displacement-producing system, the couple is useful in analysis because it alone enables the transfer of a concentrated force, exerted on a rigid body, from its given location to a parallel, alternative line of action associated with the same rigid body.
Resultants of Force Systems Exerted on Rigid Bodies definition of force, presented earlier in this chapter, was stated in terms of its potential to influence the motion of the body acted upon. In
The
discussions of forces, moments, and couples thus far, references to translational and rotational displacements have been made. The word "resultant" was used previously in discussing the replacement of two forces by one through use of the parallelogram or triangle law. In the analysis of a class of particle- and rigid-body-mechanics problems, including certain applications to orthodontics, relationships are sought between an active system of forces and a displacement pattern to be achieved by, or to result from, force application. Because the force systems often are somewhat complicated in their actual formats, the replacement of the given system by another, which is mechanically equivalent but more easily related to the potential or actual displacement of the body, is usually desirable, if not necessary. Generally, this new or replacement system is simpler, and is termed a resultant of the original force system. It is possible to reduce any force system acting on a body, when relationships between the given system and whole-body displacement potentials are sought, to one force and one couple, and in some instances to just a single force or just one couple. Clearly, these are the most simple resultants.
In the process of determining the displacement format of a body subjected to a set of forces, the resultant of the entire, active force system is sought. (Recall that the active forces "load" the body and are distinguished from responsive forces, which come into being because of contact between the body and whatever supports it. Generally, the respon-
sive force system is present in some form, however abbreviated, even before the loading is applied to the body, simply because of the member weights and the physical contact of the body with its supporting structure. Loading the body, then, usually alters the responsive force system in some manner.) Although a more simple resultant of the active force system may exist, and often does, the resultant sought is usually that referenced to a point in the body known as its center of resistance. The location of the center of resistance depends upon the size and shape of the body, the distribution of the weight of the body, and the manner in which the body is supported. In general, the location of the center of resistance of a body must be found experimentally or approximately by
47
Introduction to Analysis of Orthodontic Force
appropriate analytic analysis; simple force systems are applied and displacement responses are observed or determined and specific displacement formats are sought. an
Upon the application of loading to a body, which results in an activeplus-responsive force system, and if the entire force system exerted on the body is not completely self-balancing, one of three forms of displacement occurs. Translation is that displacement within which no line in the body experiences a change in angular position (orientation) with respect to a fixed reference line. If the body is rigid, during translation all points in the body follow parallel paths; although these paths may be straight lines, they may also be curvilinear without violating the requisites for translation. In order for a translational displacement to result from the application of force, the resultant of the applied, active force system mj.ist reduce to a single concentrated force, and the line of action of that force must pass through the center of resistance of the body. When the resultant of the active force system is a couple, the resulting displacement is termed pure rotation. All points in or on the body move in circular paths; the centers of those circles are all on one line or axis that pierces the center of resistance of the body. The vector representing the resultant couple is parallel to the axis of rotation. The most common form of displacement is a combination of translation and pure rotation and is known as generalized rotation. The resultant of the active force system is usually expressed as one force and one couple. The displacement may be difficult to describe except to note that the angulation of the body has changed. If the applied force system is two-dimensional and the resultant of the active force system reduces to a single force, its line of action does not pierce the center of resistance of the body, and for small displacements or if the resultant remains constant in all characteristics with respect to the body acted upon, the motion is circular about an axis perpendicular to the plane of the forces but not passing through the center of resistance. Except when the active force system consists entirely of couples, at least a portion of the resultant problem is the simplification of a concurrent force system. Although the parallelogram or triangle law may be used to reduce the concurrent system to a single force, an alternative procedure may be more simple if numerous forces are involved; its advantage in an analytic approach is organization. Useful in both two- and three-dimensional problems, the process follows a sequence of four steps: (1) a coordinate framework is established; (2) each given force is decomposed into its coordinate components; (3) the algebraic sum of each set of components is determined, thereby obtaining the resultant components; and (4) the resultant components may be combined into one single force, using the parallelogram law or the three-dimensional extension of it. A re-examination of a previous example serves to illustrate this procedure (see Fig. 2-3). Figure 2-18 shows the elastic and headgear forces referenced to a chosen framework in a buccal-plane view, the given forces decomposed into components along the reference axes, the resultant components,
48
Bloengineering Analysis of Orthodontic Mechanics
0
0
m
d
192g
365 g
55 9
200 g
400 9
163 g
0
Rd = 400 sin 66° — 200 sin 74°
=365—192=173g = 400 cos 66°+ 200 cos 74°
Rd
d i
=163+55=218g
//1 I
R
I
= '11732 + 2182 = 278 g
FIGURE 2-18. The resultant of several concurrent forces obtained through the use of force components. A two-dimensional example.
and the single-force resultant together with the computations providing
the force characteristics sought. The resultant components as computed suggest that the displacement should have a strong, occlusally directed aspect, but some potential for distal movement also exists. The resultant force does not pass through the center of resistance of the tooth in this two-dimensional analysis. The procedure for determining the resultant of a general force system, consisting of both point forces and couples exerted on a rigid body, is as follows. First, when the entire active force system is under investigation, a reference point is chosen that often will be the center of resistance of the body if its location is known. A coordinate framework is
established, centered at the reference point, with coordinate axes o* ented according to the force-system layout. Second, each concentrated force is transferred from its given line of action to a parallel line passing through the reference point. Displacing the force in this manner requires
the addition of a couple to the force system. Following the approach shown in Figure 2-17, the magnitude of the couple added equals the product of the force magnitude and the perpendicular distance between the new and given lines of action of the force. The sense of the couple is that of the moment of the original force about the reference point. In a
three-dimensional problem, the couple will be represented in vector fashion with its direction perpendicular to the plane of the given and new force vectors. The given force system has now been transformed into a set of concurrent point forces at the chosen reference point and a set of couples (including those given and those created in "moving" forces).
49
Introduction to Analysis of Orthodontic Force
Third, the forces are combined and the set reduced to a single concentrated force, using the parallelogram or triangle law or following the process outline with Figure 2-18 in which each force is decomposed into coordinate components. The couples, properly represented by free vectors, can all be gathered at any reference point and reduced to one couple. The choice of reduction procedures is the same as that for reducing a system of concurrent force vectors. The given force system now has been replaced by an equivalent system consisting of one point force, or
its two or three rectangular coordinate components, at the reference point and a single couple (or its two or three components). Fourth and finally, to the extent possible or desirable, the force and couple may be
combined; in this part of the process, the direct association with the ref eMnce point is lost (unless, as determined, no couple portion of the resultant exists as associated with the point). In the special circumstance when the force and couple vectors are perpendicular to one another, this pair may be reduced further to a single force. In the plane problem, if
both force and couple, as found with respect to the chosen reference point, are nonzero, these vectors are always mutually perpendicular and, therefore, may be combined to yield the single-force resultant. A two-dimensional resultant problem, using the procedure just outlined, is presented in Figure 2-19. The rectangular plate is viewed from above, initially resting on a frictionless surface. The three forces and the couple are then applied, and the resultant is sought toward description of the displacement produced by the loading. If the plate is homogeneous (its weight uniformly distributed throughout its volume), point 0 at
the geometric center is the center of resistance. By the choice of the 300 g
8mm
2mm
4—
8mm
—
lOOg
200g
L
X
= 300(4)) + 100(0)
= 300— 100 = 200 g
+ 200(4)) + 960)
= —200 g 01 200 g
R=V2002+2002=2800 200 1
°x =
=560g-mm) M0
—h—560 == 2mm
FIGURE 2-19. The resultant of a general two-dimensional force system acting on a solid body.
50
Bioengineering Analysis of Orthodontic Mechanics
orientation of the x-y-coordinate framework, no decomposition of any of
the given forces is necessary in this example. Each force is then moved to point 0; the line of action of one of the given forces passes through 0, so the creation of the concurrent force system there requires the introduction of two additional couples. The resultant force components are determined, then the force portion of the resultant at point 0. The three couples may be combined in algebraic fashion, since they occupy a common plane. Note that the couple portion of the resultant has characteristics identical to those of the sum of the moments of the given forces with respect to point 0; this is always the case, whether the problem is twoor three-dimensional (with vector addition of moments necessary in the spatial situation). The motion resulting from the application of the force system may now be described: point 0 moves in the direction and sense of the 280-g force with accompanying counterclockwise rotation of the plate. The final sketch shows the single-force resultant following combination of the force and couple portions of the resultant (see the procedure shown in Fig. 2-16); the line of action of the single force is fixed and is totally unrelated to the initial choice of point 0 as the reference point for the solution. The typical resultant problems in orthodontic mechanics involve a
single tooth or a dental segment. Because the active force system is ordinarily applied to the labial or lingual crown surface(s), the problem is generally three-dimensional. Although seemingly a somewhat indirect and lengthy process, from the standpoint of ease of viewing, the elimination of the need to express couples in the vector format, and the definitions of orthodontic displacements, it is usually convenient to decom-
pose the actual spatial situation into two or three two-dimensional portions. Employing as the reference the center of resistance as seen in each dental coordinate plane needed, in each plane the force-system components are reduced to force and couple resultant components. If
then desired, as a final step in the solution, the components may be combined and the resultant expressed in the three-dimensional format.
Distributed Forces and Their Resultants Earlier, contact forces were noted as being associated with surface areas
of the bodies acted upon. If such forces directly affect only a small portion of the total surface area of the body, in modeling, that small area is
collapsed around a surface point and a concentrated force is exerted there. On the other hand, when a force system is spread over a relatively sizable portion of the surface area in a continuous manner, the system is termed a distributed force. A resultant, a point force and/or a couple, may
be determined to replace the distributed force system for purposes of examination of whole-body displacement tendencies. When considering associated deformations, however, it is necessary to differentiate among forces according to the sizes of contact areas involved.
51
Introduction to Analysis of Orthodontic Force
Two familiar examples of distributed force are that exerted by a load of grain or coal against the floor and sides of a railroad car and the force of water pushing against a dam. In the analysis of any distributed force, the pattern of distribution must be established and, to this end, the total contact area is partitioned into small portions. The amount of the total
force associated with each small part of the subdivided area is determined; for each partial area the fraction of the total force divided by that
partial area is the intensity of force there (measured, for example, in pounds per square inch). Generally, the intensity of force varies from one location to another over the total area covered by the distribution. In the railroad-car example, the distributed force arises from the weight of the contents; the depth of the grain at any specific location measured vertically from the car floor is proportional to the local intensity of contact force against the floor. Water pressure increases directly with depth,
and the pressure at a point or level beneath the water surface is the intensity of force there. If the car floor and the dam-water interface are plane surfaces, the forces against each partial area are parallel to one another. In problems involving curved surfaces, the distributions may be decomposed into components yielding two or three systems of parallel, distributed forces. The number of parallel forces in a specific force distribution equals the number of parts into which the contact area is divided. The larger the number of partial areas, the better the approximation by a set of discrete forces of a continuous distribution. If the intensity of force at any location can be expressed mathematically in terms of the coordinates of the location within the contact area, an exact representation of the distribution may be available. Two examples of distributed force occurring with the activation of an orthodontic appliance are the "loading" of the periodontal ligament
by a tooth root and the force system generated within the adhesive when an engaged arch wire transfers contact force through a bonded bracket pad (or when a masticatory force is unintentionally exerted on the bracket). Distributed forces can be external or internal to a body under study; in an analysis of the railroad-car floor, the force of the grain
or coal is a distributed, external load, and every activated arch wire "contains" distributed, internal force. The dimensions of force intensity are force divided by length squared and the intensity of internal force is known as stress. Another vector quantity, stress is discussed in detail in Chapter 3. The systems of distributed force of orthodontic interest may be represented or modeled by one or more sets of parallel forces, with all forces in a set having the same sense. The resultant of a set of such forces is a single force having the direction and sense of the set, with its line of action passing through the center of the distribution (generally not the center of the associated contact area). Although the problem is inherently three-dimensional, many force distributions vary in only one direction parallel to the contact area. In such problems, the two-dimensional sketch of interest wifi contain directions referencing the magnitude of the force intensity from point to point and the associated
52
Bioengineering Analysis of Orthodontic Mechanics
coordinate locating points of the contact area; the area itself will appear
only in edge view. In Figure 2-20, two particular distributions for which the determination of resultants is straightforward are displayed with their resultants. The distributions are assumed uniform in the direction perpendicular to the page and, for simplicity, the areas are assumed to be rectangular with the dimension not seen (perpendicular to the sketch) symbolized by w. The independent coordinate x, extending from zero to L, locates points along the edge of the area and, generally, the intensity (at any point represented by the height of the diagram) may be expressed mathematically as a function of the coordinate x. For any distribution of parallel forces, all having the same sense, the resultant magnitude equals the average intensity multiplied by the area covered by the distribution. For the uniformly distributed loading on the left in the figure, the intensity, which is the same everywhere, is clearly the average intensity. For the linearly increasing, triangular distribution, the average intensity is onehalf of the sum of the minimum (zero) and maximum intensities. Note that the resultant force may be computed as the area, within the boundary of the distribution seen in the figure, multiplied by the width w of the rectangular area covered by the distributed force system. In a dimensional analysis, force/(length2) multiplied by length gives force/length for the "area" of the sketched diagram, and this multiplied by another p
h
L
(L,p0)
(L,h0)
x
— — — — ——1
\ \\ P = h0Lw
4
.1
L
L
L
-
J
3
Lw
\\
\\
FIGURE 2-20. The resultants of uniform and linearly-varying, parallel, distributed forces.
53
Introduction to Analysis of Orthodontic Force
length (perpendicular to the plane of the sketch) yields force. Also note
the locations of the resultant forces; their lines of action pierce the cen-
ters of the rectangular and triangular "areas" shown in the figure. Again, this development is restricted to distributions over rectangular areas where the intensifies vary only in one direction along one edge of an area. Do not confuse the two areas that have been mentioned: the rectangular contact area over which the force is distributed and the area within the distribution-of-force sketch.
Friction Two materials touching one another share a contact area. The resistance
to movement tangent to this area, of one material relative to the other, is known as friction. Friction may be generated between dry solid surfaces; the resistance to displacement may be lessened if a lubricant is placed between the contacting surfaces. Friction may exist between two solid surfaces, at a solid-fluid interface, or between fluid layers. Friction be-
tween solids is termed rolling or sliding, depending upon the form of relative movement attempted; rolling resistance is highly dependent on the amount of localized deformation where contact occurs. The resistance that precludes actual motion is termed static friction; that which exists during motion is called dynamic friction. Both the static and dynamics forms of sliding friction are of orthodontic interest. Sliding friction is
generated between arch wire and bracket when the wire "guides" the bracket during mesiodistal movement of an individual tooth or when the arch wire is slipped through posterior crown attachments in, for example, the retraction of an anterior dental segment. A classical model of frictional analysis is depicted in Figure 2-21. Before the active force P is applied, to produce intended horizontal movement to the left, the fixed surface on which the block rests responds only to the weight of the block with an upward force, perpendicular to the plane contact area. This force, often symbolized by N (representing "normal") is actually the resultant of a distributed force system
against the bottom, rectangular face of the block and, if the block is homogeneous, initially the distribution is uniform over the contact area. With the application of the force P at a relatively low magnitude, the
responsive distribution of the force against the bottom surface of the block becomes nonuniform and the magnitude, direction, and point of application of the resultant of the distribution are altered. With P sufficiently small, however, the block remains stationary and the changes in characteristics of the responsive force from the horizontal surface take place because the contacting surfaces are not perfectly smooth. (If absolutely no friction existed, the application of P, however small in magni-
tude, would result in motion with no change from the static (P = 0) configuration in the responsive surface force.) With the application of the force
the now-inclined, responsive, resultant force R may be de-
54
Bioengineering Analysis of Orthodontic Mechanics
Wi
P
f
p ,-p 777'?
N
FIGURE 2-21. Active and responsive forces exerted on a block supported by a rough horizontal surface.
composed into components perpendicular (normal) and parallel (tan-
gential) to the contact area. Since no potential for vertical motion exists, the normal component is N and the tangential component, in the direction of but opposite in sense to the intended displacement, is the resistive force of friction. Increasing the magnitude of P results in a conesponding increase in the magnitude of the frictional force f, but motion does not occur until the force of friction reaches a maximum value corre-
sponding to a critical magnitude of the applied force P. A slight, additional increase in P beyond this critical size results in motion, and the force of friction generally decreases slightly from the "maximum 1' as the static situation becomes dynamic. A plot of the frictional force versus the applied force P for the model described is displayed in Figure 2-22. Controlled experiments in which frictional forces between plane surfaces were measured have determined that the maximum f and the level
of the frictional force following the initiation of motion are highly dependent on the relative roughness of the contacting surfaces and thtmagnitude of the normal force component N. In the classical model, fmax and the dynamic frictional force are individually assumed proportional to N;
the proportionality constants are the static and dynamic coefficients of friction, quantified empirically for pairs of contacting surfaces. Note that in the static model only the maximum frictional-force magnitude is equal
to the product of the static frictional coefficient and the normal force component N and the model does not consider the manner of distribution of contact force between the surfaces. The classical block-on-plane analysis gives a good fundamental basis
for examining the problem of friction between arch wire and bracket slot. The block and plane are analogous to the wire and slot, but while
55
Introduction to Analysis of Orthodontic Force
Frictional Component
(Pent, fmax)
Static
Dynamic
Activating Force P FIGURE 2-22. Relationship between the magnitudes of active and frictional forces, P and f, for the block and surface of Figure 2-2 1.
only one contact area exists in the foregoing example (Fig. 2-21), the
orthodontic problem is three-dimensional in that the wire may contact the bracket slot and the accompanying ligation at as many as four locations simultaneously. The direction of the frictional force is bounded by the angulations of wire and bracket slot; the sense opposes the relative motion and depends upon whether force is exerted on the bracket or the wire (action-reaction). If the bracket is to move distally on the wire, the frictional force against the bracket is generally toward the mesial (and its action-reaction counterpart against the wire is directed distally). If the wire is to slide toward the posterior through a stationary bracket, the frictional force from the wire and against the bracket acts toward the distal (since the force of friction against the wire is directed mesially, opposing the intended motion). Oscillation between static and dynamic forms of friction occurs between the wire and slot in an activated retraction appliance as the relative movement takes place in small "steps." As viewed buccally, one or more of four separate forms of contact between bracket and arch wire may be observed during canine retraction. These four forms of contact are shown in Figure 2-23 and a pair of components, normal and frictional forces, would generally exist at each individual contact area. Influencing the normal component of contact force, and therefore the level of friction, are the angulation between arch wire and bracket and the tightness of the ligation. The wire material and that of the ligature, for a given bracket, also influence the frictional resistance. Beyond the classical model, the difference in contact areas that exists with round versus rectangular wire likely affects the magnitude of frictional resistance. For a more complete discussion of friction between bracket and wire, refer to Frank and Nikolai (1980).
56
Bioengineering Analysis of Orthodontic Mechanics
Zero angulation, clearance
One-edge contact, clearance
f4
Two-edge contact, zero clearance
Snug wire/slot fit, zero clearance
FIGURE 2-23. Bucca! views of possible contact modes between bracket and orthodontic
arch wire.
Concepts of Mechanical Equilibrium and Their Applicability to Orthodontic Mechanics The definition of "mechanical equilibrium" of a solid body represents an extension of Newton's First Law for a particle. In effect, the practical definition says that the net tendencies for translation and rotation of the body are nil: if the magnitudes of the force and couple portions of the entire force system exerted on the body are both the body is in mechanical equilibrium. If the body was stationary before
this "balanced" force system was applied, the system adds zero net potential for whole-body motion. The balanced force system creates no
inertia, where inertia is proportional to the products of mass and the acceleration components of the body. When the total force system is unbalanced, one or the other or both portions of the resultant of that system are nonzero, an extension of Newton's Second Law governs,
57 Introduction to Analysis of Orthodontic Force
and the body undergoes accelerated motion and has been given some
inertia.
An important notion in this discussion is that of the "entire force system": all of the external forces and couples, both active and responsive,
exerted on the body under study. The definition of mechanical equilibrium may be extended to apply to groups of bodies or assemblies. Inter-
nal forces, by definition being within the body or assembly, exist in cancelling pairs and, therefore, collectively contribute nothing to a resultant computation. Internal forces, however, may become external, for
example, when an assembly is disassembled for purposes of analysis and one body or member is studied individually. Pairs of forces associated with the connection(s) of this body to others in the assembly are internal to the assembly, but one of each pair of forces is external to the one member. Under the influence of an activated orthodontic appliance, the dentilion should not be static. Over a finite period of time the teeth are to be moyed and, correspondingly, within the appliance itself, relative movements and changes in as-activated shapes may occur. The concept of the resultant of a system of forces has been discussed toward establishing a procedure by which forms of displacement caused by active forces might be predicted. Also created with the activation of the appliance, however,
are responsive force systems in need of description: the force system transmitted through the tooth and exerted by the root on the periodontal ligament, and that delivered by the appliance to the "anchorage," for example. Force analyses of the entire tooth, segment, and the appliance and its
elements or parts are to be carried out, but exact results need not be obtained. Note, then, that although appliance activation generally results in the creation of some translational and rotational inertia, in the dentition as well as in the appliance, these levels of inertia are not sizable. Both parametric parts of the inertia, the mass and the time rate of change of velocity (the acceleration), are small. Although the entire complex is more dynamic immediately following activation, during the short lime period of soft-tissue deformations, than during the much longer period of tissue remodeling, tooth displacement, and appliance deactivation, the contributions of inertia to the analysis may reasonably be neglected, even in comparison to the smallest of active forces. The pair
of governing vector equations of the orthodontic dynamics problem may, then, be written as Vector sum of all forces
inertia =
(Vector sum of all moments +jot&ti
liriertia)mass center =
0
With the inertia terms deleted, the relationships are those associated with the definition of mechanical equilibrium. The equations are necessary and sufficient conditions for a mechanically static system. Thus, at
58
Bioengineering Analysis of Orthodontic Mechanics
any instant following activation, the dentition and appliance are said to
exist in a quasi-static state and the equations of mechanical equilibrium may be employed to obtain a close approximation to the relative magni-
tudes of the components of the active and responsive force systems. The rationale for the consideration of the mechanical equilibrium problem and its analysis is, therefore, established. The general problem may be stated as follows: Given a body or group of interconnected bodies subjected to a system of external forces and couples, such that the state produced by the system is mechanically static, determine the relationships that must hold among the characteristics of the external force system. Often the problem is one in which several of the characteristics of the force system are quantified and those characteristics that are not given explicitly are to be obtained. Nearly indispensable in proceeding toward the solution of all but the simplest of problems is the use of the free-body diagram, and several of
these diagrams may be needed to solve one problem. In the solution procedure, a body or assembly of bodies is earmarked for study and, as the name suggests, is "freed" from its surroundings (its supporting structure and all of its connections to other bodies); this is depicted in an isolated sketch. All of the external forces and couples exerted, both active and responsive, are shown in the sketch. As has been the convention in previous force diagrams, forces and couples exerted on (not by) the body are displayed. Not only the completed known portions of the force system, but also those forces and couples with unknown characteristics, must be properly depicted. Moreover, to aid in moment computa-
tions, distances and dimensions must be included. Omitting existing parts of the force system, including nonexistent forces or couples, and incorrectly identifying support structures of force-transmission mechanisms (connections) are typical mistakes made in these problems. Once the free-body diagram has been correctly completed, however, the interrelationships among the known and unknown force-system characteristics may be expressed in a rather straightforward manner. For each freebody diagram drawn, a set of scalar equations may be written, obtained from the pair of governing vector relationships written previously. In words, these equations say: All of the forces exerted on the body or group of bodies, pictured and under study, must balance, and the moments of all of the forces and coupies, with respect to any chosen moment center or axis, must also balance.
(This is an alternative way of stating that both the force and couple portions of the resultant of the entire force system must have zero mag-
nitudes.) The number of independent, scalar, algebraic relationships that may be written from each free-body diagram will depend upon the type of force system and whether the problem is one-, two-, or threedimensional.
In the quasi-static orthodontic problem, because body forces are nearly always negligible in comparison with contact forces, understand-
59
Introduction to Analysis of Orthodontic Force
ing the manner of transfer of force through a connection is fundamen-
tally important. Connections of specific interest include those between the tooth crown and appliance element(s) and between individual parts of the orthodontic appliance. Before examining connections within the appliance, however, the displaying of responsive forces and couples having unknown characteristics and several general types of connections are considered. The complete description of a concentrated support or connection force generally requires the definition of several scalar quantities. The magnItude of a force or couple is the most common unknown in the equilibrium problem. To define the direction of the line of action of a force, one scalar quantity, the size of an angle relative to a reference line, is needed in the two-dimensional problem; two angles are required in the spatial problem. Although nearly always known in practical problems of orthodontic interest, the point of application of a concentrated force is described through the dimensions of the body or in terms of the coordinates of the point in relation to a specified reference framework. In approaching an equilibrium problem in which the vector characteristics of a force are sought, with the point of application of the force known the unknowns may be expressed as the magnitude and one or two angles or as the two or three components of the force referenced to an established set of coordinates. Figure 2-24 illustrates the two methods of displaying an unknown force at a point in two dimensions. On the left, the force is completely described by the magnitude, the one angle, and a convention to define sense of the vector; that convention could be an understanding that the force would appear to be pulling on the point
y
x
F
n
-__,,_,, , Scalar unknowns: F, 6
Scalar unknowns:
FIGURE 2-24. Two methods of illustrating a concentrated force in two dimensions, unknown except for point of application.
60
Bioengineering Analysis of Orthodontic Mechanics
and the angle would be measured counterclockwise from the positive
extent of the x-axis, the angle having a value between 0 and 360°. On the right, the force is completely described in terms of its x- and y-components; two algebraic quantities provide the magnitude and sense of each component. Because the sign convention leading to the determination of sense is more straightforward in the component description of a force, expressing the unknowns as the components themselves is, seemingly, the more convenient procedure in an analytic approach to the equilibrium problem. To clarify, then, although sense often is noted as an individual force characteristic, sense is never a separate unknown in the equilibrium problem. A force at a point is totally described in terms of its components. In
picturing an unknown force in a free-body diagram, an assumption of sense may be made for each component; in solving an algebraic equation
for a component magnitude, the sign accompanying the answer will indicate whether the assumed sense was correct (plus sign) or incorrect (minus sign). Shown in Table 2-1 are several types of connections between bodies, the associated force systems exerted on each body (action and reaction), and the number of scalar quantities involved, exclusive of location of the connection. The first is, effectively, one-dimensional (although it may exist within a two- or three-dimensional problem), the next three are two-dimensional, and the final two are three-dimensional. The first two entries in the table indicate push-contact between two bodies that share a contact area. In the general two-dimensional analysis, the resultant of the distributed force system over the contact area would be displayed as two force components, one perpendicular to and the other tangent to the area at the center of the distribution. When it is reasonable to assume that friction between the bodies is negligible, the force of one body upon the other is perpendicular to the contact surface and, possibly, only one characteristic is an unknown: the force magnitude. The tangential component is the frictional force discussed previously, and it generally adds a second unknown at the contact site (except when proportional to the normal component, and then the sense of the frictional force must be correctly indicated). The third entry in the table, depicting a frictionless hinge connection (with the pin through the two members perpendicular to the plane of the sketch), generally transmits a force having unknown magnitude and direction. The two scalar quantities may be represented as the magnitudes of the components of the force; the components must be mutually
perpendicular, but the pair orientation as a whole may be arbitrarily selected (often most convenient with one component coinciding in direction with the long dimension of one of the interconnected members).
Shown as the fourth entry is the fixed support. Whereas the hinged connection possesses no inherent resistance to the angular movement of one body with respect to the other in the plane of the sketch and in the absence of friction, the fixed support or connection provides that resist-
.
61
Introduction to Analysis of Orthodontic Force TABLE 2-1. Connections between members and supports and associated force systems Force system
Description of contact or connection
\\
Action
Bar against fricfiorrless
from
surface
f
Hr
Reactive composaNs of bar against surface
Force components exerted by surface
Bar against rough
:::e
Action and reaction counterpart forces perpendicular to plane of contact. (One componeat magnitude is Unknown.)
Force of bar against surface
Notmal force
\ \ surtace
Remarks
Ft eaction
N2
t
are unknown.)
N2
Components exerted by pin
Bar restrained by frictiortless pin connection
/ p3
Force components from bar against P3I pin
Direction of resultant contact force generalty not longitudinal. (Two component magnitudes are unknown
Force system exerted by support on bar
Couple represents rasistance to rotation inherent in support (Action and reaction interchanged compared to toregoing examples. (Three component magnitudes are unknown.)
—
Action of bar against support
Fixed support of bar in plane setting
J
H4
3
/
I
C4
Frictronless support of shaft in
-
Fixed support of bar
1 C)
C) P)
C) C) P)
Indicated is potential for bar to slide to the right. Friction opposes this tendency. (Two component magnitudes
U
No resistance to longitudinat shaft displacement assumed. (Four component magnifudea are unknown)
Complete resistance to linear and angular dinplacements in all directions. (Six component magnitudes are unknown.)
ance, represented by the couple. In the plane problem only one unknown, its magnitude, is associated with such a couple; the sense of the couple (clockwise or counterclockwise) is obtained through a sign convention in the manner of that of an unknown force component. A number of three-dimensional connections are relatively common within structures and machines, encompassing from one to six potential unknowns, but just two of interest are displayed in Table 2-1. For the bearing-shaft connection, negligible resistance to shaft translation through and rotation within the bearing is assumed. This leaves a force vector and a couple vector, both perpendicular to the shaft axis, generally transmitted through the connection. Expressing each vector in com-
62
Bioengineering Analysis of Orthodontic Mechanics
ponent form, a total of four potential unknowns is present. The force is
associated with the positional resistance of one• body relative to the other, and often equilibrium can exist in the absence of the couple; in such instances the couple is regarded as a secondary effect in primarily giving the connection added stability. The final entry in the table is the totally fixed, rigid connection. This connection, and a weld is a good example, prohibits both translation and rotation of one body with respect to the other in any direction. The force and couple are expressed as sets of three components each, giving a total of six potential unknowns. Before proceeding to a discussion of connections associated with the orthodontic appliance, a straightforward and applicable statics problem is examined. Consider an elastic that is stretched between two connection points in the orthodontic complex. The elastic at either end might be
activated against a hook attached to a bracket, a hook soldered or welded to the arch wire, or a loop bent into the wire, but assume that the elastic is stretched between just two connection locations. A reasonable assumption, based on the flexibility of the elastic, is that a single force is transmitted through the elastic and, therefore, through each connectioti point. Shown in Figure 2-25 are seemingly possible free-body diagrams of the elastic, the first with both forces of apparently unequal magnitudes drawn in arbitrary directions. If, however, the elastic is in mechanical equilibrium, and its weight is comparatively negligible, the moments of the two forces must balance. Choose first a moment center at point A;
with the line of action of P passing through A its moment arm, and therefore its moment, is zero. Hence, the moment of Q with respect to A
//
'4,
0
0
B
A
B
B
A
A
'P P
I
I,
I
t
FIGURE 2-25. Preparation of a free-body diagram for a static elastic stretched between two fixed points.
63
Introduction to Analysis of Orthodontic Force
also be zero, and thus the line of action of Q must also pass through point A. The direction of Q is then determined as shown in the second diagram of Figure 2-25. Now, making a similar argument using point B as the moment center, the direction of the line of action of I? is determined; the two forces, accordingly, must share a common line of action as shown in the third sketch. Finally, because the entire system of forces must balance in any static situation, the forces P and Q must be equal in magnitude and opposite in sense. Their common line of action is determined by the positions of the two connection points, which both lie on the line. Hence, for any fwo-force member in equilibrium, the two forces must be equal in magnitude, opposite in sense, and share one line of action. Generally, the forces may be either pulling away from or pushmust
.ing toward one another, but an elastic has no resistance to pushing forces and, therefore, when activated it must be "in tension." Now, with the background of Table 2-1 and the force analysis of a stretched elastic, a number of connections associated with orthodontic mechanics are analyzed in Table 2-2. Although all are shown in a twodimensional format, with several a third dimension may be necessary to represent the connection force system completely. In the first entry just one potential unknown is present, the magnitude of P1. provided the orientation (angulation) of the stretched elastic has been or can be measured. Because the bracket in the second example has finite mesiodistal width, the couple, due to angulation of the wire within the bracket, as well as the force may be generated. This is somewhat analogous to the fixed support in two dimensions shown in Table 2-1. Friction, not included in this sketch but potentially present, would add a third scalar quantity. Examining this connection in an occlusal-plane view, assuming the presence of ligation, another force-couple pair must be present, and with friction neglected and round wire engaged, this connection is analogous to that of the bearing-and-shaft connection of Table 2-1. The third entry in Table 2-2 assumes that the rectangular wire will not rotate within the bracket slot; the couple shown is torsional in nature. Also generally present is an accompanying force transverse to the wire, and a total of three potential scalar unknowns in this connection not including friction. The fourth entry presents a somewhat simplified representation with friction included; analogous to the second entry in Table 2-1, this is just one of the possible configurations, in a view from the buccal perspective, already displayed. (See Fig. 2-23.) The fifth entry in Table 2-2 shows the resistance to rotation and translation provided by
a buccal tube. The component H5 may arise from a step bend in the engaged arch wire, a stop, and/or from friction. (If the arch wire was tied back, H5 would have the opposite sense.) Two additional components, a buccolingual force and another antirotation couple, would be seen in an occlusal view, reflecting a total of five potential unknowns associated with this connection. Finally, the connection of a J-hook to an arch wire has little; if any, inherent resistance to rotation. This is the three-dimensional extension of the pin connection with three associated potential unknown force components, two of them existing in the buccal-plane
64
Bioengineering Analysis of Orthodontic Mechanics TABLE 2-2. Orthodontic appliance interconnections and contact force systems exerted on the appliance members Force system
Connected members
I-look affixed to buccal tube and elastic
Remarks
Reaction
Action
Force trom elastic against hook
Hook force against elastic
Direct tension Lines of action along the stretched elastic. 1
Force system exerted Ofl
Bracket and arch
bracket
V2
wire
Bracket force system against wire
I
Second-order angulation and contact creates couple
V2
Bracket and rectangular arch
Force system from wire
Force system from bracket Third-order angulation and contact creates tingual root torque.
vof
wire C3
H3
Bracket and arch wire
Force components exerted by wire N,
Bracket-generated force components
Friction opposes relative displacement of wire and bracket, No second-order contact assumed.
f4
Stopped arch wire and buccal tube
Generated by stopped arch wire
Reaction of tube against wire
Slop contact with mesial extent of tube produces horizontal force
S
component. H5
v5
J-hook and hook soldered H6 H6
¶
Representative of highpull-headgear force delivery to maxillary anterior segment of arch wire
view. Note that the most general and complex connection, analogous to the final entry in Table 2-1, can be generated between the ligated bracket and a rectangular arch wire: all three force components (the mésiodistal component from a stop or friction) and all three couple components may exist for a total of six potential scalar unknown quantities. The two vector equations governing the equilibrium problem, which indicated that both the force and couple portions of the resultant of the entire force system must be zero, suggest that a maximum of six scalar relationships may be written among the characteristics of the force system properly displayed in the free-body diagram. Although the majority
of the quasi-static problems to be examined are three-dimensional,
65
Introduction to Analysis of Orthodontic Force
many of these problems may be modeled into two dimensions. Also,
those problems that must be treated in three dimensions may be decom-
posed into several interrelated two-dimensional analyses and approached in parts, each in its own plane. As noted earlier, a relationship exists between the maximum number of independent scalar equations available, relating characteristics of the force system, and the dimensional level and format of that system. Several examples illustrating this relationship and the solution procedures follow. In Figure 2-3, two forces were shown acting on a bracket. Returning to that problem: the equilibrating force exerted on the bracket through its attachment to the band or its bond to the facial crown surface is to be determined. To be pictured in a free-body diagram, then, is the bracket itself and the forces exerted on it. The three existing forces are: the elastic force, the headgear force, and the unknown force. Ignoring the possibility that the two known forces may be slightly out of plane, relative to each other, and neglecting the bracket thickness, the problem then becomes two-dimensional. Moreover, the force system is concurrent with the lines of action of all three forces passing through the bracket center. The free-body diagram is shown in Figure 2-26. All characteristics of the elastic and headgear forces are known. The point of application of
Bg
EL = 200 g
= 400 g
+
Rm =
Rg = B8
97:O=_147O3O0S1fl 66° = 0
400 COS 66° — 200 COS 74° = 0
B8 = 163 = 55 = 218 g F8
+ 2182
278 g
FIGURE 2-26. Free-body diagram of an orthodontic bracket subjected to a twodimensional set of concurrent forces.
66
Bioengineeririg Analysis of Orthodontic Mechanics
the third force is known, but its magnitude and direction are not. There-
fore, two unknown scalar quantities are unknown and are pictured as mesially- and gingivally-directed components in the facial view. In the problem which is two-dimensional with the forces concurrent, only two independent scalar relationships may be written: the algebraic sum of force components in two mutually perpendicular directions in the plane must be zero to ensure a force balance. The solutions for the components of the bracket force and, subsequently, the resultant of those components are carried out in the figure. The solution yielding algebraically positive results indicates correct initial assumptions of the senses of the components. No moment-balance equation is written. (Clearly, a moment equation using the bracket center as the moment center is identically satisfied since the moment arms of all three forces are zero; hence,
no useful information is obtained from the moment-balance relationship. Another moment center might be chosen, and a moment balance written with respect to it, but the equation obtained would only be useful in replacing one of the two scalar force relationships.) The force just obtained is noted to be identical in all characteristics, except sense, to the
resultant of the elastic and headgear forces found in Figure 2-3; in all equilibrium problems involving a concurrent force system, the resultant of the given forces must be balanced by the force(s) that involve the unknowns. The lever is a second example that has applicability to orthodontic appliance analysis. Figure 2-27 shows a straight member that could represent an arch-wire segment balanced on a "fulcrum" by the forces P1 and P2 at the ends of the member. The fulcrum force and the relationship between the forces P1 and P2 to ensure equilibrium are sought. It is
P1
Pl
C
C
.
L1
L1
P0 x
= P0 — Pi — P2 = 0
P0 =
+
McP2L2)+P1L1)0
-
P2L2 = P1L1
FIGURE 2-27. A lever subjected to a two-dimensional parallel force system.
—
67
Introduction to Analysis of Orthodontic Force
assumed that the applied forces exist in a common plane and are both
perpendicular to the long dimension of the member. Hence, the problem is two-dimensional and the free-body diagram reflects a parallel force system because neither of the applied forces has a component in the direction of the long dimension of the member. The responsive (fulcrum) force likewise can have no long-dimension component and reflect a balanced system. The magnitude of the fulcrum force is obtained by balancing the forces, algebraically summing the forces in the one direction of all three lines of action. The relationship between P1 and P2 is obtained through a moment balance using any convenient moment center; in this example, point C, which is immediately above the fulcrum, is chosen.
In general, for a two-dimensional parallel force system, two independent scalar equations of equilibrium may be written. One force and one moment equation were used to analyze the lever example. The general vector moment-balance equation previously described indicates that the center of mass of the body or assembly under study is to be used as the moment center. With the inertia effects absent or disregarded, however, it may be demonstrated that a force balance and a moment balance about the mass center demand that moments balance with respect to any selected moment center. It follows that the two scalar equilibrium equations for the problem of Figure 2-27 could have been two moment equations, and those equations would be independent of one another so long as the line through the two moment centers is not parallel to the
direction common to the three forces. Figure 2-28 shows a straight "beam" subjected to a transverse concentrated "load" of 300 g. The problem of determining the responses at the supports is an inversion of that of Figure 2-27; sketching the free-body diagram and using the equa-
300 g
I—
Pt'
6mm
II
P4
FIGURE 2-28. A structural member supported at its ends and subjected to a transverse, concentrated load. The correct free-body diagram shows a parallel force system in one plane.
68
Bioengineering Analysis of Orthodontic Mechanics
lions of Figure 2-27 with P0 = 300 g, the magnitudes of the left and right
support may be shown to equal 100 g and 200 g, respectively. In the final example of this section, a cantilevered straight member is subjected to a loading consisting of forces at its right end, parallel and perpendicular to its long dimension, and a couple, all in a common plane. The left end of the member is engaged in a fixed support, and the components of the support force system are sought. The free-body diagram of the member includes the support resistances to any movement of its left end in the plane of the sketch. This is a general two-dimensional equilibrium problem and a total of three independent scalar equations may be written to relate characteristics of the total force system. An analysis is undertaken in Figure 2-29. A choice of sets of equations is
available: two force and one moment equations, one force and two 80g
I
I I I
lB
A
100 g
I 2100 g-mm
35mm
80 g CA
B
A
lOOg 2100 g-mm
x
— 100 = 0 = = 100 g
bOg M8 —
+ 2100
+ CA
=0
CA = 2100 —2800 = —700 g-mm 700 g-mm
FIGURE 2-29. A cantilevered structural member loaded at its "free" end. A twodimensional system of active and responsive forces and couples.
69
Introduction to Analysis of Orthodontic Force
moment equations (but the line connecting the two moment centers may
not be perpendicular to the direction of the force-component balance), or three moment equations (but the three moment centers may not lie on one straight line). The free-body diagram in Figure 2-29 contains three unknown forcesystem characteristics: the magnitudes of the two force components and that of the couple. The reference frame adjacent to the free-body diagram defines the chosen directions and assumed positive senses for the force-balance relationships written below the sketches. Note that the assumed senses associated with the unknown magnitudes need not coincide with the sign conventions for the equations. In the solution, note that a negative result was obtained for the support couple; this indicates that the assumed sense designated in the free-body diagram was incorrect. Also, recalling that the moment of a couple is inherent, note that its contribution to a moment balance in a two-dimensional analysis is the same, regardless of the moment-center location. The couple portion of the load on the member was included in the moment-balance relationship written, even though the moment center chosen at the right end of the member eliminated contributions from the other applied forces (because their moment arms with respect to point B are zef o) and the couple and moment center appear to be coincident. A partial check of the solution obtained in the figure may be pursued by writing a moment-
balance equation using point A at the left end of the member as the moment center.
Synopsis This chapter introduced the concept of force. Various force manipula-
tions were discussed, including the determinations of the moment of a force and the resultant of a system of forces. The couple, a special force system, was defined and its use as a tool in force analysis, as well as its occurrence as a load and a mechanical response, was considered. Modeling of actual force systems and the bodies acted upon, toward simplification of analyses, were discussed. The resultant of an active force system is sought in order to describe the whole-body displacement that may be produced by the system. A rationale was developed for analyzing the quasi-static problem, applicable to the orthodontic appliance, slowly deactivating between appointments, using the governing equadons of mechanical equilibrium. Force systems produced by activated orthodontic appliances provide the catalyst for the processes that result in the displacements of teeth. Forces are also generated in orthopedic actions toward alterations of growth patterns and to cause movements of dentofacial bones. Hence, a first step in evaluating the potential of a given appliance, with regard to its tendency to produce desired displacements in a controlled manner,
70 Bioengineering Analysis of Orthodontic Mechanics
or
in the design of a new appliance, is an examination of the forces
created upon activation that are associated with the actual or proposed appliance. This, however, is only the beginning of the total structural analysis. Thus far, deformations have not been considered, nor has the interface between the biologic and mechanical systems. To initiate the action of the orthodontic mechanics, the appliance must be activated; one or more appliance elements must be deformed from their passive configurations, thereby "loading" the appliance and inducing forces throughout it. Because only a part of the total appliance force system can be quantified through direct measurement and the use of equilibrium
equations, relationships between force and deformation must be sought. Such relationships involve the material characteristics of the appliance members as well as their sizes and shapes and will be discussed in the next chapter.
Reference Frank, C.A., and Nikolai, R.J.: A comparative study of frictional resistances be-
tween orthodontic bracket and arch wire. Am. J. Orthod., 78:593-609, 1980.
Suggested Readings F,P., and Johnston, E.R., Jr.: Vector Mechanics for Engineers. 3rd Ed. New York, McGraw-Hill, 1977, Chapters 1 to 4, 6, and 8.
Beer,
Jarabak, J.R., and Fizzell, J.A.: Technique and Treatment with Light-wire Edgewise Appliances. 2nd Ed. St. Louis, C.V. Mosby, 1972, Chapter 1. McLean, W.G., and Nelson, E.W.: Engineering Mechanics: Statics and Dynamics. New York, Schaum Publishing, 1962, Chapters 2, 3, 5, and 8.
Mulligan, T.F.: Common Sense Mechanics. Phoenix, CSM, 1982, Chapters 2 and 3. Smith, R.J., and Burstone, C.J.: Mechanics of tooth movement. Am. J. Orthod., 85:294—307, 1984.
Thurow, R.C.: Atlas of Orthodontic Principles. St. Louis, C.V. Mosby, 1970, Chapter 4.
Thurow, R.C.: Edgewise Orthodontics. 4th Ed. St. Louis, CV. Mosby, 1982, Chapter 2.
Material Behavior of the Orthodontic Appliance
In planning the treatment of a malocclusion, the orthodontist selects a
series of appliances for sequential use in the correction. Each appliance is a kind of structure. Activated through a deformation from its passive, as-prepared configuration, the appliance attempts to return to that passive state, but is prevented, temporarily, by the resistance of the dentilion to which it is affixed. The appliance responds to the activation by exerting a system of forces against those teeth and portions of the dentofacial complex that hold it in an activated state. Creation of contact forces between the appliance and dentition is accompanied by force sys-
tems within the individual members of the appliance, transmitted through connections from one member to another. The responses of the
appliance members to orthodontic "loading" are fundamentally dependent on the materials of those members. The orthodontist uses a variety of materials: metallic-alloy arch wires, bands, brackets, and headgear components, polymeric "elastics" (bands, modules, 0-rings, chains, threads), adhesives, alginates, acrylics, plasters, among others. It is important, then, that the clinician be acquainted with the basic concepts of materials science as applicable to the specialty. When designing and preparing each arch wire, auxiliary, or retainer, the orthodontist must be able to predict the response of each appliance member to "in-service" conditions. The materials are subjected not only to mechanical actions associated with the induced force
systems, but the environment in which the materials must exist also influences their behavior. The basis for explaining or predicting the response of a structural or machine member under specified conditions of use is found in its properties. A property is, in a general sense, a descriptive quantity that gives a specific characteristic to the member. Properties may be categorized in many ways, but initially they should be divided into those associated directly with the material or materials of the member and those termed 71
72
Bioengineering Analysis of Orthodontic Mechanics
properties, which are descriptive of its size, shape, and appearance. Material properties are further subdivided into characteristics that are independent of external influences, simply termed "material" properties, and those that are associated in some way with the conditions of use or the use environment: mechanical, chemical, thermal, and magnetic, for example. Consider, for instance, the decision of selecting a spring to produce a particular orthodontic displacement. Properties that enter into this design decision include the cross-sectional wire size, the coil diameter, and the overall passive length of the spring (physical characteristics), the stiffness and strength of the wire material (mechanical) and the spring itself (structural), the resistance of the material to corrosion (chemical), and the effects of a somewhat elevated and varying temperature (thermal). The materials of the members of the orthodontic appliance are the focus of this chapter. Behavior of ancillary materials, such as plasters, will not be discussed here. The responses of biologic materials during orthodontic treatment will not be overlooked, but are relegated to a subphysical
sequent chapter. Physical, mechanical, chemical, and thermal are the property categories of primary interest to the orthodontist; optical, electrical,
and magnetic properties, although not to be generally disregarded, are of little importance within the scope of this text and will not be discussed in this chapter. Properties may be interdependent and, as previously mentioned, are often influenced by conditions of use of the material and in-service environment. The characteristics of a stationary body, directly associated with its response to applied or induced forces, which include both material and physical contributions, are known as structural properties. The response to activation of an orthodontic wire loop will depend upon the wire material, the wire cross-section, the loop geometry, and the amounts and directions of the bends made in forming the loop. The behavior of an elastic module used intraorally will differ in some respects from the results of an evaluation of its mechanical characteristics on the bench in the laboratory under ordinary conditions of temperature and humidity or its response as part of an extraoral appliance. Material-property values are generally obtained in controlled experimentation. In-service conditions, as necessary, are created in the laboratory and the required measurements are taken. When widespread use of the results is expected, and to ensure reliability of comparisons, testing procedures must be standardized. Much of the standardization in the United States in testing, particularly in the determination of mechanical properties, has been and is established by the American Society for Testing Materials. Property quantification and certification of testing procedures in the dental fields is carried out under the auspices of the American Dental Association. In the sections to follow, the nature and structure of solid materials are surveyed and the interaction between load and deformation is examined and pertinent structural characteristics are described. The concepts of stress and strain are introduced, mechanical and structural properties
73
Material Behavior of the Orthodontic Appliance
obtained through standardized testing are defined and discussed as needed for developments in subsequent chapters, and chemical and as
thermal characteristics of interest are considered. The chapter concludes with a set of guidelines to be used when selecting orthodontic-appliance materials.
Internal Structure of a Solid Material Any fluid or solid material is a conglomerate of particles. Forces, generally attractive, exist between pairs of particles, and their magnitudes are determined by particle densities and distances between particle centers. The forces of attraction in a fluid are relatively small and, as a result, the
material readily conforms to its container. The particles in a solid are dense and relatively closely packed; the associated cohesive forces are large enough that, generally, the solid retains shape in the absence of any containment. The smallest subdivided unit of matter, which is representative of a pure material (chemical element), is the atom. Submicroscopic in size, the atom is envisioned as a nucleus surrounded by electrons in a kind of planetary arrangement. The nucleus contains protons and neutrons; these subparticles have approximately identical masses, the proton possesses a small, positive, electrical charge, and the neutron is electrically neutral. The mass of an electron is approximately 1/2000th that of the proton, yet it possesses a negative charge equal in magnitude, but opposite in sense (sign) to that of the proton. The atomic number of an element equals the number of protons in its nucleus. The atomic weight is approxi-
mately that of the nucleus, neglecting the small contribution from the electrons, a figure that numerically is, in essence, the sum of the protons and neutrons and varies with element from slightly more than 1 to in excess of 250. In the element model, the electrons exist in concentric orbits, or "shells," with the nucleus at the center. Heavier atoms may possess more electron shells than light atoms, but independent of the material element each shell—its position designated radially with respect to the nucleus—has a specific electron capacity. Except for the outside shell, which may be incomplete, all shells are totally "filled." The outer shell contains, at most, 8 electrons; elements with like numbers of electrons in their outer shells have similar chemical behaviors and the number of outer-shell electrons is particularly significant in connection with mechanical, electrical, and magnetic properties. Atoms hold together because of the intra-atomic attraction between elecIrons and protons. Overall, an atom is said to be electrically neutral if the numbers of protons and electrons are equal, regardless of the number of
electron shells and the number of electrons in the outer shell. Although possessing a net charge of zero with numbers of protons and electrons equal, the stability of an atom with regard to its potential to combine with other atoms is largely tied to the outer shell and the
74
Bioengineering Analysis of Orthodonfic Mechanics
number of "valence" electrons in it. Certain materials are inert or chemi-
cally inactive because the pure chemical elements exhibit filled outer shells together with having no net electrical charge. Many elements do not occur in their pure forms in nature because their atomic instability has resulted in chemical reaction or combination with other elements. In short, the most stable form is that achieved in a configuration with the outer shell of electrons completely filled. With the exception of the noble
gases, to attain this stabilized state an element must either effectively gain or lose from one to four electrons. Filling the outer shell results in the atom assuming a net negative charge, and the atom that loses its valence electrons assumes a positive charge; in either instance the result is a transformation from atom to ion. Ionization yields the potential for the formation of primary, interatomic bonds, of which three forms exist. Ionic bonds are formed when the
valence electrons shed by one group of atoms are transferred to, and complete the outer shells of, other atoms. Table salt is the result of sodium losing its one valence electron to chlorine, electrically neutral with
seven electrons in its outer shell. Covalent bonding is the sharing of electrons. The simplest example is the combination of two hydrogen atoms, another is the sharing among four hydrogen atoms and a carbon atom to form methane, and a third is the formation of diamond solely from carbon atoms. Metallic bonds may be formed by atoms of elements having just a few (generally one or two) valence electrons which are shed easily; formed is a structure of positive ions and "free" electrons or an "electron cloud" and between them the bonding occurs. The type of primary bond is usually indicated by the number of valence electrons in one of the elements; in a few materials, more than one form of bonding can exist. Ionic and covalent bonding results in combinations of atoms known as molecules. Although the molecule is electrically neutral, or nearly so, due to the nonuniform distribution of electrical charges, intermolecular bonds (also termed secondary or van der Waals bonds) may form.
Several types of secondary bonds are defined, but for all forms their strengths are substantially less, often an order of magnitude lower than that of the typical primary bond. Interatomic and intermolecular forms of attraction pull the particles toward one another, but spaces also exist within the particle structures, primarily because mutually repulsive forces between their electron shells become dominant when two atoms become too close to each other. Figure 3-1 depicts the forces of attraction and repulsion, both shown in dashed plots, versus the distance between the centers of a pair
of atoms. The solid curve, the net force versus interatomic spacing, crosses the horizontal axis at the point of equal attractive and repulsive forces—the so-called "equilibrium distance." This equilibrium spacing or distance between atoms varies with the material elements because of atomic size and the number of adjacent atoms in the molecular structure. Also, temperature is an important, controlling parameter. As the temperature is raised (lowered) the mean spacing between atoms increases (decreases), accounting for the thermal expansion (contraction) of materials.
75
Material Behavior of the Orthodontic Appliance
Interatomic Force
\ Force of attraction
N ed
0
Net
/
/
Atomic Spacing
/ Force of repulsion
/
ed: Equilibrium distance
I FIGURE 3-1. Plots of interatomic force versus distance between atoms.
Several material properties are related to the bonding characteristics of atoms and molecules. Density is determined by atomic weight and spacing. Overall strength of a material depends upon bond strengths. Material stiffness is related jointly to magnitudes of bond forces and particle spacings. Thermal and electrical conductivities are low in ionicand covalent-bonded materials and high in materials that are metalli-
cally bonded wherein the free electrons serve as carriers of energy. Prop-
erties are also dependent upon the arrangement of atoms within the material, and atomic arrangements are generally classified as molecular, crystalline, or amorphous in structure. The chemical composition of a limited number of atoms forms a molecule, having a net electrical charge of zero due to primary bonding. From the characteristics of the constituent atoms, the numbers of bonds, the
bond lengths, and possible relative positions of the atoms (molecular geometries) may be determined. Often, for a molecular composition, more than one atomic arrangement is possible; such structural variations are termed isomers. Molecular materials, solids in particular, are held
together through secondary bonding. Such solids have low melting points, indicative of the relative weak intermolecular bonding, but the larger molecules have greater strengths because of the increased num-
76
Bioengineering Analysis of Orthodontic Mechanics
bers of such bonds. Basic to the understanding of molecular structures is
the study of hydrocarbons. In a so-called "unsaturated" hydrocarbon, multiple carbon-to-carbon bonds exist that enable the formation of a large molecule from several smaller molecules in a chemical reaction known as polymerization. One form of this reaction, termed "addition polymerization," beginning with like monomers of ethylene, is illustrated in Figure 3-2. The polymeric molecule, or polymer, obtained is made up of many repeating units, called mers. A number of polymeric materials are familiar to the orthodontist: waxes, alginates (impression materials), acrylics (for retainers), adhesives (for direct or indirect bracket bonding),
and the elastomeric (rubber or rubberlike) bands, modules, "chains," and "threads." Although polymerization can develop lengthy molecular "chains," and the formation of van der Waals bonds, or "cross-linking" (vulcanization of rubber, for example) can create two- and three-dimensional molecular networks having some strength, the majority of engineering materials are made up of thrde-dimensional, repeating patterns of atoms called crystals. The primary bonding is typically metallic, but secondary
bonding can result in similar highly ordered patterns such as in the molecular crystal of table salt. The pictorial conception of a crystal is a spatial structure of straight lines with the atomic nuclei (in table salt, the sodium-ion centers) at their intersections. The three-dimensional, repeating element of the crystalline structure is known as the unit cell. Some 14 different unit-cell geometries are defined and the most common cubic crystal representations are depicted in Figure 3-3. Various material properties may be correlated with crystalline structure, and often the unit cell is indicative of the material class of the solid. Some materials, however, can exist as a solid in more than one crystal structure (polymorphism or allotropism, analogous to isomerism in polymers). A prime example is iron, which may exhibit a body- or face-centered-cubic crystal structure, depending on its temperature or the presence of alloying elements. Materials that do not display the regular, repetitive patterns found in crystalline solids and many polymers, but instead reflect a disordered
A
H
I
I
H
H
H
H
H
H
I
I
I
I
I
I
C
C
C—C————C———-—C
I
I
I
I
I
I
I
I
H
H
H
H
H
H
H
H
me
Monomer
Polymer
H: Hydrogen C: Carbon FIGURE 3-2. The ethylene monomer and a portion of the long-chain molecule resulting from addition polymerization.
77
Material B.ehavior of the Orthodontic Appliance
Ii
I
Simple cubic
Body-centered cubic
Face-centered cubic
FIGURE 3-3. Representations of simple cubic, body-centered cubic, and face-centered cubic unit cells.
molecular structure, are termed amorphous. Gases and liquids are amor-
phous ("without form") as are a number of oxides including glasses. Although such materials are not generally of specific interest to the orthodontist, some aggregates of crystalline or molecular and amorphous materials may be. Plastics are often reinforced with various "fillers." The divisions between crystals (grain boundaries) contain amorphous matter. Polycrystalline materials are, by definition, aggregates of small crys-
tals. Those polymers having highly disordered molecular structures might be categorized as aggregate materials. The internal, submicroscopic structures of many materials are highly ordered, but imperfections exist. Although often cumulatively occupying a small portion or volume of the material, certain imperfections, particularly in crystalline solids, cannot be ignored owing to their substantial
impact upon properties. Within the crystals themselves atoms may be missing from individual cells, foreign atoms may have replaced some of those of the pure material, atoms may have become displaced to "in-between" (interstitial) locations within the cells, or foreign atoms may have entered the structure and assumed interstitial positions. An extra line or plane of atoms, termed a "dislocation," may be present within the crystal. Imperfections inherently exist at external surfaces where the exposure of surface atoms makes them more vulnerable to reactions with matter within the adjacent environment. Within crystalline materials at internal or grain boundaries, amorphous matter up to ten or more atomic distances in thickness is present which generally possesses at least some relevant property values inferior to those of the individual crystals.
Certain imperfections are created to enhance property values. For example, brass is a solid solution of copper and zinc, termed "substitutional" because zinc atoms are substituted for some copper atoms in the crystal structure to produce the alloy. Steel formation involves elevating the temperature of iron to a level such that the relatively small carbon atoms can fit into the middle of the face-centered-cubic cell structure of the iron. As indicated previously, some material properties are highly sensitive to imperfections. Although the weight fraction of carbon in steel is often less than 5%, the presence of the carbon greatly improves its strength properties over those of pure iron. On the other hand, den-
78
Bloengineering Analysis of Orthodontic Mechanics
sity and material stiffness are termed "structure insensitive" properties
because they are affected only in proportion to the imperfection level or numbers. (For example, the density of stainless steel is equal to the sum of the products of the fractional amounts, by mass, and the densities of the constituent elements.) Atoms, molecules, and crystals within a solid are retained in particular positions because of the existence of an overall, internal, force balance. The potential for relative displacement is often present, however, either in the form of some external effect or simply because the particles are being maintained in a strained configuration. Actually, atoms are continually in motion, vibrating about their equilibrium positions with amplitudes dependent upon temperature; apparently, completely static states occur only at absolute zero. If temperatures become sufficiently elevated and the particles are initially in strained positions, the vibra-
tions can result in displacements to less constrained configurations. Mechanical action can cause particle displacements, often to more constrained configurations, as can an electric or magnetic field if the particle
characteristics are such as to respond to the field. A combination of mechanical and thermal inputs can cause a displacement that cannot be achieved individually at the same level as the combined action. The movements of atoms within a crystal often involve point imperfections; with relatively little mechanical thermal input, an atom can be displaced into a neighboring vacancy or can move a smaller foreign atom interstitially and take its position. Displacements of molecules usually require less input than average atomic movements because secondary bonds are weaker than primary bonds; however, bond density is sometimes an overriding factor. The displacement of entire crystals typically occurs with the grains sliding with respect to one another; secondary bonds with grain-boundary matter are broken and then re-established when the resistance to further movement exceeds what remains of the initial input. A somewhat superficial discussion of the internal structure of a solid material is contained in the preceding paragraphs. Much should be familiar from previous dental-materials coursework, and a more complete treatment of these topics may be found in the materials-science texts in the "Suggested Readings" at the end of the chapter. The objective of this section has been to examine internal material structure to the point of recognizing the existence of relationships with pertinent properties of orthodontic appliance materials. Those relationships will be recalled as the associated properties and material behavior are considered in subsequent sections of this chapter.
Load-Deformation Behavior of a Structural Member generation of forces within the members of an orthodontic appliance, in addition to those associated with interatomic and intermolecuThe
79
Material Behavior of the Orthodontic Appliance
bonding, and the transferral of induced forces to and through the dentition and into the periodontium, results in the remodeling of the periodontal support system and the displacements of teeth toward corlar
rection of the malocclusion. Although the rigid-body model was used in the discussion of forces in Chapter 2, the rationale was a simplification in force analyses when deformations could reasonably be neglected in comparison to the dimensions of the body or member under study. In fact, for any stationary, supported body, loading and deformation cannot take place exclusive of one another. To the typical engineering struc-
ture, a "load" (an external force system) is applied which results in a pattern of deformation; although perhaps unwanted, that deformation must be included in design or analysis procedures. On the other hand, the clinician produces a deformation, in activating an orthodontic appliance, to induce the desired force system. Although the cause-and-effect roles seemingly are reversed in loading the appliance compared to the typical structure, the parts of the orthodontic appliance may reasonably be termed "structural members" because they carry and deliver force. Accordingly, certain appliance-member characteristics defined in this section will be referred to as structural properties. The practitioner generally produces and observes deformation before, during, and after activation of the orthodontic appliance. Each deformation, defined as any localized or overall change in shape of a mem-
ber, results in an alteration of the force system within the appliance. Most familiar are the deformations necessary to activate the appliance, producing the initial levels of force to be delivered to the dentition. Over a period of time, with movements of teeth, the initial appliance deformations are lessened with corresponding changes in force magnitudes. At a
succeeding appointment, if the appliance has not totally deactivated (lost all force) in the meantime, the practitioner may unload the appliance with configuration changes accompanying disengagement. An appliance member may be said to be structurally passive when it is subjected only to gravitational attraction and any forces responsive only to its small weight. Prior to appliance engagement and activation, the clinician may affect loading and unloading deformations that take an appliance element, typically a length of arch wire, from one passive configu-
ration to another. In the formation of arches or auxiliaries from as-received wire, the orthodontist must apply (often with pliers), and
then release, force systems in order to produce the desired shape changes. Several types of deformation are to be recognized. First, consider two particles within a structural member. A change in the distance between particles reflects a normal deformation. If the distance is increased, the deformation is termed tensile or extensional; if the distance is decreased, the deformation is compressive. A second fundamental deformation is
associated with a pair of lines of particles of the material. If the angle between the two lines changes, a shearing deformation has occurred. Normal deformations are depicted in Figure 3-4; note the changes in relative positions of particles at right angles to the normal deformation,
80
Bioengineering Analysis of Orthodontic Mechanics
11 Passive
configuration
Compressive deformation
Extensional deformation
FIGURE 3-4. Normal deformations, extension and compression, illustrated on an atomic scale.
associated
with bonding forces and distances, the so-called "Poisson
effect." The schematic of a shearing deformation appears in Figure 3-5.
Both normal and shearing deformations can occur simultaneously within a member—indeed, even at the same location. These fundamental forms of deformation may or may not be visually recognizable, depending upon their sizes and the volumes of material affected; an obvious deformation is the extension associated with the activation of an "elastic." Two additional forms of deformation include both of the fundamental forms. Bending is identified externally by a longitudinal shape change in the member, often from a straight to a curved configuration, as depicted in Figure 3-6. In the analysis of bending, the member is conceptualized as an assembly of longitudinal fibers, all having the same shape.
Passive
configuration
Shearing deformation
FIGURE 3-5. An atomic representation of shear deformation.
81
Material Behavior of the Orthodontic Appliance
I
-
I
Passive bar
Bar deformed in bending FIGURE 3-6. An illustration of bending detormation, together with two measures of this form of deformation.
all fibers are attached to one another in a solid member, bending results in the stretching of some fibers and the compressing of others. In Figure 3-6, if external forces hold the member in the deformed configuration the lower fibers are maintained in compression and the upper fibers in tension. Two external measures of bending deformation are defined; as shown in the figure, a transverse deflection 6 denotes the displacement of a cross-section of the member perpendicular to the longitudinal dimension and an angular coordinate U indicates the rotation of a cross-section. The final type of deformation to be introduced is twisting. Figure 3-7 illustrates that the external measure of a twisting deformation is an angular coordinate 4) which denotes the change in orientation of a transverse reference line in the cross-section. In a twisted member, because the amounts of rotation generally vary from one cross-section to another, and since cross-sections are interconnected, the characteristic internal deformation is a shearing deformation. (Shearing action often also exists within a member subjected to bending activation, but the accompanying deformation is negligible except, possibly, when the longitudinal dimension of the member is small.) Beginning with a structural member in a passive but supported configuration, the application of external force produces an associated deformation. To maintain the deformed configuration induced by the force system, the force system itself must be maintained and the member is said to be constrained. Releasing the load results in a second configuration change, and the member again becomes passive. By definition, the application of force causes an elastic deformation of the member if the passive configurations, before loading and following unloading, are geometrically identical. If, upon unloading, the member does not return Because
to its initial, passive configuration, the loading produces inelastic
behavior.
82
Bioengineeririg Analysis of Orthodontic Mechanics
Passive bar
Fixed end
Twist
deformation
Bar deformed in torsion
FIGURE 3-7. A twist deformation and an angle 4' as a measure of it.
As a first example to study, totally from an external standpoint, the load-deformation behavior of a structural member, the bending of an arch-wire segment is examined. Consider an initially straight piece of wire between two fixed supports. The span, the distance between the supports, is sizable in comparison to the cross-sectional dimensions of the wire. The wire segment has been cut from a longer piece of as-
83
Material Behavior of the Orthodontic Appliance
Applied force P
----Deflection ô
FIGURE 3-8. An arch-wire segment subjected to bending deformation produced by a transverse, midspan force.
received wire. A point force is applied midway between the supports to
generate a bending deformation; specifically, the external deformation of interest is the transverse deflection of the cross-section directly under the applied load. Shown schematically in Figure 3-8 are the undeformed shape (dashed) and a deformed configuration (solid), the latter generated by the applied force. The load-deformation curve, given in Figure 3-9, is the result of the plotting of corresponding values of applied force and deflected position of the midspan cross-section relative to its passive location. The load is slowly increased in small increments and the accu-
mulated deflection grows until the simple structure collapses (is no
Applied Force P 2
/ S: Stiffness p1: Proportional limit
el: Elastic limit
0
Deflection
FIGURE 3-9. A p'ot of applied force versus midspan deflection for the wire segment of Figure 3-8.
84 Bloengineering Analysis of Orthodontic Mechanics
longer able to sustain the load). The overall shape of the curve is charac-
teristic of crystalline materials, regardless of the member geometry or the type of deformation induced. The initial portion of the plot is linear with the slope then decreasing as the loading (plotted as the ordinate on the graph) becomes relatively substantial. The end point of the linear portion of the plot is appropriately termed the proportional limit. The coordinates of the final point of the curve denote the largest values of deformation and force before failure. The application of a relatively light force to the member of Figure 3-8 results in a small deflection; the two values are the coordinates of a point on the curve of Figure 3-9, most likely below the proportional limit. If the load is held constant and the temperature is not elevated substantially, the deformation of the metallic member will generally remain unchanged, virtually independent of time. Upon release of the load the deformation will disappear and, in general, the original straight-line shape will be regained, provided the elastic limit has not been exceeded. The elastic limit is the extreme point on the force-deformation plot from which unloading will occur with no permanent deformation resulting from the load-unload cycle. Points 1 and 2 in Figure 3-9 are below and above the elastic limit, respectively. An unloading plot from point 1 traces the loading plot between the origin and point 1. The dashed unloading curve from point 2 is notably linear and parallel to the linear portion of the loading plot; the deformation coordinate of the point of intersection of this unloading curve with the abscissa (deformation axis) denotes the permanent deformation of the segment and a new passive configuration of the now bent wire. Internally, the loading to point 1 stretches some interatomic bonds and pushes other atoms toward one another, but no bonds are broken and, as the load is released, the atoms return to their prior equilibrium positions. Loading to point 2, however, severs some bonds and results in movements of crystals relative to one another; new bonds are established that are reflected upon unloading in a new, altered, passive configuration. The elastic limit for a crystalline member is found experimentally to be only slightly beyond the proportional limit and, to simplify computations, the elastic limit is often "placed" at the end of the linear portion of the plot, coincident with the proportional limit. The elastic limit divides the structural character of the wire segment and the load-deformation plot into regions of totally elastic behavior and inelastic response. The slope of the linear part of the curve is the bending stiffness at the "midspan" location of the arch-wire segment. In an orthodontic activation, the loading of a wire is intended to be to a point not beyond the proportional limit so that the totally deactivated configuration is known: generally identical to the preload, passive geometry. Hence, in an elastic activation, the stiffness when quantified enables the determination of (1) the magnitude of the load corresponding to the activating deformation—the load is the product of stiffness and - the amount of activation—and (2) the reduction in the as-activated level of
85
Material Behavior of the Orthodontic Appliance
loading during deactivation—the product of stiffness and quantified
deactivation increment. The coordinates of the elastic-limit point on the plot are known as the elastic range and the elastic strength. In this example, the elastic range is the largest transverse bending deformation of the midspan cross-section
that the wire segment can experience and subsequently deactivate totally or be unloaded without sustaining a permanent bend. The elastic strength is the maximum magnitude of applied, midspan, bending force without inelastic material behavior occurring somewhere in the segment. A fourth notable elastic structural property is termed resilience. An energy-related quantity, and as such to be discussed in more detail in Chapter 4, resilience is, quantified, the area under the load-deformation diagram up to the elastic limit; for crystalline materials that area is triangular. Dependent collectively upon both elastic-limit coordinates, which themselves are intradependent with stiffness, a member exhibiting low stiffness but a high elastic range may possess a resilience similar to that of a member having greater stiffness but a lesser elastic range. Although the activation of an arch wire in bending is not intended to take the member beyond its elastic limit, the nonlinear portion of the load-deformation diagram is also important because inelastic bending is necessary by the practitioner to place permanent bends. Taking the example wire segment to point 2 on the load-deformation curve results in the appearance of a "gable bend" midway between the supports. How severe a permanent bend the wire can sustain without fracture depends upon the ductility of the wire in bending. Two measurements of ductility are the deformation to fracture, the horizontal coordinate in Figure 3-9 of the final point of the plot, compared to the elastic range, and the toughness, the total area under the load-deformation diagram, compared to the resilience. Both of these quantities just defined, compared to their elastic counterparts, are substantial for the very ductile ligature wire, for example. In contrast, both comparisons are 1:1 for a perfectly brittle member; the fracture point and elastic limit coincide on the load-deformation diagram. A second, pertinent, structural example is that of a stretched elastomeric module. Schematics showing the passive and an activated configuration are presented in Figure 3-10. One end of the module is fixed and the member is elongated by the application of a concentrated force at the other end. The overall deformation (e) is the extended length (L) minus the passive length (L0). As with the arch wire in bending, the larger the
load, the greater the deformation, but the module exhibits a two- or three-part load-deformation diagram, quite different from that of the arch wire. An elastomer is a rubber or rubberlike, often synthetic, polymeric material. The nature of the bonding of atoms and molecules, and the imperfections, make the material inherently more flexible than a crystalline solid of the same size and shape. A typical two-part diagram is presented in Figure 3-11; the solid curve represents loading to failure by rupture (separation). Along the lengthy, linear portion of the plot,
86 Bioengineering Analysis of Orthodontic Mechanics
Stretching force
F
e
L
0
Passive
Activated
FIGURE 3-10. Activation of an elastomeric module in direct tension.
the elongation under load is reflected principally in the uncoiling of long molecular chains. (A typical three-part curve suggests, in the initial nonlinear portion, that secondary bonds must be broken before the major uncoiling of chain molecules can begin; that first part of the curve exhibits decreasing slope with increasing applied force.) As the uncoiling is
completed, material stiffness increases as primary bonds begin to be stretched; with continued increasing of the load the slope of the curve grows and rupture eventually occurs. Accurate determination of the coordinates of the elastic limit of the module is difficult because of the time-dependent nature of the mechanical behavior of the material (to be discussed subsequently). Most elastomers, however, are relatively brittle; their elastic limits are "high" on the load-deformation curves, beyond the linear initial or middle portions of these plots. Inelastic action,
therefore, is never expected in clinical use, as the name "elastomer" suggests.
The dashed curve in Figure 3-11 is representative of rather rapid unloading of the module. Unloaded from a point below its elastic limit, some deformation remains immediately after removal of the activating force, but over time, full recovery of the preload, passive configuration takes place as secondary bonds are re-established (dashed line). Strktly speaking, orthodontic deactivation cannot be shown in Figure 3-11 be-
cause the force in the module depends upon the time elapsed while
87
Material Behavior of the Orthodontic Appliance
Stretching Force F
/ / / 0
/ / Extensional Deformation e
FIGURE 3-11. The force-deformation diagram for the module of Figure 3-10.
under load, in addition to the amount of elongating activation. Hence, stiffness, as the slope of the load-deformation curve, can only be approximated because it, too, is changing with time. As previously mentioned, time-dependent behavior is considered further in a subsequent section of this chapter.
Mechanical Stress concept of internal force was introduced in Chapter 2. The interatomic and intermolecular bonding forces discussed previously in this chapter are fundamentally part of any internal force system. Superimposed on the bonding forces, however, are internal force distributions The
arising from the external loading of the structure through the application of mechanical force or through the imposition of an activating deformation and maintenance of a constrained configuration. The loading of one member creates internal forces in the whole of the structure; these forces
seemingly travel through the loaded member, across the connections and into adjacent members, and ultimately into the supports or foundation of the structure. These internal forces, both from bonding and from external loading, not only exist throughout the lengths of the members, but are also distributed over the entirety of the member cross-sections.
88
Bioengineering Analysis of Orthodontic Mechanics
In general, the intensity of internal force can vary in all directions within
a member. (In structural analyses, although the internal force system arising from external loading is superimposed on the bonding force system and, likely, also a residual internal-force distribution remaining from manufacturing or preactivation fabrication processes, the analyses often ignore all but the first on the basis of magnitude comparisons.) As a first example in this section, consider the activated elastomeric
module of Figure 3-10. Held in this constrained state, although stretched, the module is in mechanical equilibrium with its supports. The internal force system can be exposed in part by making an imaginary cut through the module and arguing that the resulting two parts are each in equilibrium as is the whole of the module. The specific loca-
tion of the "cut" along the module length is arbitrary in this example and is simply made perpendicular to the longitudinal direction. The two portions of the module are shown in Figure 3-12 together with the activating and responsive forces on the entire module. The external forces at the module ends are equal and opposite from a force balance; the exposed internal forces are equal and opposite because they are action-reaction counterparts. Assumed is a one-material, homogeneous module and, with the line of action of the external forces coinciding with the
longitudinal axis of geometric symmetry, the distribution of internal force, arising from the loading, is essentially uniform over the cross-section as illustrated. Also, with the resultant of the internal force system equal and opposite to the external force (from a longitudinal force balance on either module portion), the internal distribution is wholly pulling against the cross-section of the "cut." Finally, because the external loading is only at the ends of the module, and with its weight negligible, the pattern of internal forces on all cross-sections having identical orientations is the same from one end of the module to the other, regardless
of the longitudinal location of the "cut." The average intensity of the internal force in Fig. 3-12) is the resultant, internal-force magnitude (F) divided by the cross-sectional area (As) exposed by the "cut." In the determination of the initiation of inelastic material behavior or failure, intensity of internal force is the critical parameter. In short, but sometimes oversimplified, the larger the cross-section of a structural member,
the greater is the load that it can carry. Stress is defined as the intensity of internal force, per unit area, at a point within a machine or structural member. It is a vector quantity, obtained by dividing a vector (force) by a scalar (area). The dimensions of stress are force divided by length-squared (FIL2). A "cut" exposes an area internal to the member, and that area is divided into a number of subareas with a portion of the internal-force resultant exerted on each subarea. As the number of subareas increases, the size of each subarea becomes smaller and the pattern of internal-force intensity (stress) becomes more detailed. In the module example, the stress everywhere is normal to the area and is the same on every subarea of the exposed cross-section, but this is a special situation; in general, the stress may be oblique to the area and may vary over the whole of a cross-sectional area
89
Material Behavior of the Orthodontic Appliance
"cut"
F=
Crosssectional area
F A,,
F
FIGURE 3-12. Activation and response at the ends of an elastomeric module, Internal force distribution over a cross-section of the module.
90
Bioengineering Analysis of Orthodontic Mechanics
Stress
S5
Ss
S
Sn Sn
Side view
FIGURE 3-13. A general stress vector decomposed into normal- and shear-stress components.
of a member. Shown in Figure 3-13, within the total cross-section, is a typical subarea (As) (enlarged) with the stress s associated with it. Nearly always convenient is the decomposition of the stress vector into components normal (perpendicular) to and tangential to the area; these components are known as the normal and shear stresses, symbolized by a or s,, and r or s5, respectively. When the normal stress is apparently pulling against the area, as in Figure 3-13, it is called tensile stress. A normal stress that pushes against the area is a compressive stress, and is directly analogous to pressure in a fluid. Note the correspondence between the two categories of normal stress and the forms of deformation previously discussed. As an example illustrating the generation of shear stresses, consider an orthodontic bracket pad bonded to a facial crown surface and the bracket subjected to an external force parallel to the facial surface. Figure 3-14, on the left in perhaps a mesial or distal view, shows schematically the enamel crown surface, an exaggerated sketch of the deformed adhesive, and the loaded bracket. On the right, following a "cut" through the adhesive, which is made parallel to the load Q, a free-body diagram has
been drawn of the bracket, pad, and a portion of the adhesive as a whole. Neglecting weight and the faciolingual thickness, the distributed
internal force system exposed by the "cut" is entirely shear. A force balance indicates that the resultant internal force must equal Q. The average shear stress in the adhesive equals the resultant of the distribu-
tion divided by the area in shear k. In the examples of the module in tension and the adhesive in shear, the resultant internal force equals the product of average stress and the corresponding area. The computation in arch-wire bending, however, is somewhat more complex. Figure 3-15 depicts a wire segment with the "right-section cut" (perpendicular to the longitudinal dimension) exposing an internal cross-sectional area (seen only in edge view in the figure) and a typical internal force system including both forms of normal stress as well as shear stress, with none of the three uniformly distributed. In this segment, no inelastic behavior is present and the resultant of the tensile- and compressive-stress distributions is a pair of equal, opposite, and parallel forces—a "bending couple." The resultant of the shear-
91
Material Behavior of the Orthodontic Appliance
Q
Average shear stress where A,.
=
total area exposed by the cut'
FIGURE 3-14. Shear loading of a bonded bracket and shear-stress distribution within the bonding adhesive.
distribution is an internal force tangent to the area exposed by the "cut" and parallel to the direction of external force causing the bending deformation. A complete discussion of bending is undertaken in Chapter 7. stress
Mechanical and Structural Properties: Standardized Testing The
response of a structural member to external loading depends
upon the size and shape, or geometry, of the member, the relationship of the loading pattern or characteristics to that geometry, and the mechanical properties of the material of the member. Problems of analysis
92
Bioengineering Analysis of Orthodontic Mechanics Compressive stresses
Wire segment Position and direction ot shear stresses
Distribution ot shear stress
FIGURE 3-15. Typical stress patterns internal to a structural member subjected to inplane bending.
or design of structures are often approached by taking experimentally
determined mechanical-property values, substituting them into formulas or equations that express the governing principles of mechanics and
include the contributions of geometric and loading parameters, and thereby obtain or check what might be termed structural potential or capacity. The differences, if any, in bending characteristics of two, .016-
in. -diameter, arch wires, for example, are founded in the mechanical properties of the wire materials. Hence, a familiarity with these material properties is important to the practitioner. Many of the fundamental mechanical properties of a solid material are commonly quantified through the performance of a static tension test. Each test specimen is carefully prepared to minimize experimental varia-
tion or "error" associated with the testing itself. The test volume (the middle portion, longitudinally) of the specimen is uniform in cross-sectional shape and size; the cross-section often exhibits at least one axis of
geometric symmetry. The straight specimen is to be held at its ends, some distance away from the test volume such that the manner of attachment of the loading system to the specimen will not influence the test volume and the data gathered. An example test specimen is shown in the "grips" of the testing apparatus in Figure 3-16. The grips transfer the loading, measured by the testing machine, to the specimen in the resultant form of a longitudinal force having a line of action that passes through the centers of the cross-sections of the test volume. With the weight of the specimen ne-
glected (compared to the load), the specimen is a two-force member, similar to the module of Figure 3-12 and as described in Chapter 2. The grips are attached to the "heads" of the testing machine; one head is movable and the other is stationary. The testing machine has the capability to apply a slowly increasing load (thus the term "static" test) and
Load P
-II
IV
Responsive
forceP
-
FIGURE 3-16. A tensile-test specimen in the grips of a standard test apparatus.
_____ 94 Bioengineering Analysis of Orthodontic Mechanics
to monitor continuously the magnitude of that load. A longitudinal or
axial dimension within the test volume is known as the "gauge length." Its initial value, before external loading is begun, is shown in Figure 3-16 as 4?,,. An "extensometer" is attached to the specimen at the extremes of
the gauge length, or an alternative procedure is employed, to enable instantaneous measurements of the gauge length which increases during the tension test. (The extensometer weight must also be negligible compared to the specimen load or counterbalanced so as not to create notable testing error or destroy the mechanical symmetry.) In the test itself, the loading is begun from zero and gradually increased until rupture of the specimen occurs. Readings of applied load (which equals the internal axial tensile force) and the corresponding extensional deformation (or extended length) of the gauge length are taken, either manually within equal loading increments or, by means of transducers and a stripchart recorder, recorded continuously by the testing machine and plotted as the test is proceeding. The extension of the gauge length during the tension test will depend upon its initial length; therefore, this extension or elongation is "normalized" with respect to the initial length. Axial deformation per unit length is a form of engineering normal strain: 4?
(3-1) e,,
The test under discussion produces longitudinal tensile strain; if the loading sense is reversed, the gauge length is lessened under pushing force and longitudinal compressive strain is induced. Corresponding to the normal strains are normal stresses against cross-sectional areas exposed by "cutting" the specimen perpendicular to its longitudinal axis. The uniformly distributed internal force system, as in the module example, has an intensity that depends on the size of the exposed area. The engineering tensile stress equals the resultant, internal force divided by the pas-
sive, cross-sectional area. Note that by converting force to stress and deformation to strain, the net effect is the dividing out of the volume of material in the test section and, thereby, eliminating the influences of specimen size from the test results. (As the test specimen elongates under load, the cross-sectional area generally reduces; recall the Poisson effect illustrated in Figure 3-4. Engineering stress and strain are obtaine& by dividing out the undefonned, as-prepared area and length of the test volume. True stress and strain are obtained by dividing the instantaneous, deformed area and length into the corresponding internal force and elon-
gallon, respectively. All stresses and strains discussed in this chapter and throughout the text are engineering stresses and strains.) From the paired readings taken during the tension test, converting force to stress and elongation or stretched length to strain, a tensile stress-strain diagram is prepared (either by hand or plotted by the testing machine). With modifications of coordinates of data only through division by constants, the stress-strain diagram in tension for a specific ma-
95
Material Behavior of the Orthodontic Appliance
—'C C,) C,)
C
Stainless, nickel-chrome steel
Mild, structural steel E
x: tracture point
0
Tensile Strain
0
Tensile Strain
FIGURE 3-17. Stress-strain diagrams in tension, obtained from static testing, for mild, structural steel (left) and for a stainless steel (right).
terial has the same shape as the tensile, load-versus-deformation plot for
a structural member of the same material. Shown in Figure 3-17 are tensile stress-strain diagrams depicting the behavior of mild, structural steel, which exhibits a yielding phenomenon, and a stainless steel, which does not notably yield. Yielding is the straining of the material at virtually a constant stress well below the level at which rupture occurs. Illustrated in Figure 3-18 are diagrams for a ceramic material and twoand three-part diagrams for typical polymeric materials. The mechanical properties of interest are obtained from the stress-strain diagram for the material and from measurements taken directly from the specimen, and
are analogous to corresponding structural properties, many of which have been defined previously. A crystalline material obeys Hooke's Law, which states that the stress-
strain ratio is constant up to the proportional limit; the constant in this linear stress-strain relationship is the modulus of elasticity E (Young's modulus) in tension, the elastic stiffness of the material and the slope of the initial portion of the diagram. The ultimate strength of the material is the value of the maximum ordinate of the curve; as indicated in Figure 3-17 this may or may not be the stress level at fracture (but if the high point of the cprve is reached in loading, although data may be taken at greater strains, the imminent fracture generally cannot be prevented). The coordinates of the fracture/rupture point are ordinarily termed the fracture strain and fracture stress or strength. The proportional and elastic limits are close together on the curve; the coordinates of the elastic limit are termed the elastic-limit strain and elastic strength of the material. Accordingly, the simple relationship a = Ec, which is the expression of
96
Bioengineering Analysis of Orthodontic Mechanics
Tensile Stress
A ceramic material
x: fracture point
Polymeric materials
0
Tensile Strain
FIGURE 3-18. Static stress-strain diagrams in tension for ceramic and polymeric materials.
Hooke's Law in direct tension, holds virtually up to the elastic limit. For
those crystalline materials that yield, the yield strength is the stress level at which this phenomenon is exhibited; because yielding immediately follows the initiation of inelastic action in tension, the values of elastic strength and yield strength are, for practical purposes, equal.
In the discussion of mechanical properties, the word "modulus" implies "per unit volume;" accordingly, the modulus of resilience of the material is quantitatively the area under the stress-strain diagram up to the elastic limit and the modulus of toughness is the total area under the diagram to the point of fracture. A material is relatively ductile or brittle, depending on the extent of the stress-strain diagram beyond the elastic limit. Clearly, the ceramic material of Figure 3-18 reflects no ductility at
all; the plot being linear to fracture indicates the nonexistence of an elastic limit and the inability of the material to take a permanent strain. Typical measures of ductility are, analogous to the content of an earlier discussion of structural characteristics, the fracture strain referenced to the elastic-limit strain and the modulus of toughness relative to the modulus of resilience. A third measure is obtained by taking the two por-
tions of the fracture specimen from the testing machine, mating the fractured surfaces, measuring the as-fractured gauge length, determining the corresponding strain value, and finally multiplying by 100 to obtain the percent elongation at fracture. A fourth measure of ductility,
97
Material Behavior of the Orthodontic Appliance
similar to the third, is obtained from the pretest cross-sectional area and
the cross-sectional area of the fractured specimen at the fracture location: the percent reduction in area.
Noncrystalline materials, such as the polymers of Figure 3-18, seldom exhibit any useful level of ductility and often do not obey Hooke's Law. As a result, there is little distinction between elastic strength and ultimate strength. Although not exactly correct, the elastic limit may be taken as the point where the slope of the stress-strain curve begins to increase substantially toward rupture. The stiffness per unit volume is not a constant in those ranges where the curve is nonlinear; hence, the simple relation between stiffness modulus and the coordinates of the elastic limit, (cr/e)ej = E for crystalline materials, is not valid for those materials exhibiting a nonlinear first portion of their stress-strain diagrams. To determine the stress level, for a given activation in tension of an elastic for example, it is best to directly measure the force and divide that magnitude by the cross-sectional area. Mechanical properties of materials are obtained from simple experiments, tension tests for example, which generate the fundamental forms
of deformation cited previously. Because their values often depend upon the form of deformation generated in the test, these properties when quantified must be presented as "in tension," "in compression," or "in shear." In addition, the rate of loading during the test will influence mechanical-property values to some extent; for example, distinctions are made between static and impact tests. Furthermore, environmental conditions, particularly temperature in the test locality, may affect values, as may specimen-storage time and conditions prior to testing. Finally, the as-received internal material configuration will partially
determine the values of some mechanical properties, in particular for a metallic material the coordinates of the elastic limit and, therefore, its modulus of resilience. (The influence on properties of permanent deformations and accompanying "residual stresses" is considered in Chapter 4.) Material behavior in static compression is similar to that exhibited in a tension test, but some differences in test specimens and procedures, and in obtained mechanical-property values, do exist. In an axial compression test, with the sense of the loading reversed from that of the tension test, the as-prepared specimen often has a uniform cross-sectional geometry from one end to the other, and is short and thick to ensure symmetric loading throughout the test and to avoid the possible occurrence of lateral deflection or buckling. Load and deformation data are taken, converted to compressive stress and strain, and the diagram is constructed. For a crystalline material, the modulus of elasticity is generally the same in tension and compression, although the extension of the linear portion of the curve to a greater elastic-strength magnitude in compression than in tension is not uncommon. Noncrystalline materials may exhibit a substantially altered stress-strain diagram in compression compared to the tension plot. With molecular materials not extensively cross-linked, the chain molecules that uncoil at low to mod-
98
Bioengineering Analysis of Orthodontic Mechanics
loads in tension clearly would not do so to any great extent in compression. Although some tensile strain is developed laterally from the Poisson effect accompanying the axial loading, with compressive loading, interatomic bond deformations begin almost immediately. Ultierate
mate strength in compression is defined and computed in the same manner as in tension, but the failure mode must be substantially different from the tensile rupture. The load-carrying capacity of a member in direct compression is lost when one of the following occurs: extensive lateral deformation (buckling); longitudinal splitting (also known as internal buckling); or a "slipping fracture," on a plane inclined with respect to the loading axis, due to the exceeding of the ultimate strength of the material in shear. (Although absent on right cross-sections in members undergoing direct tensile or compressive loading, shear stresses exist on all other planes exposed by "cuts" through the members and are largest on planes at 45° with the loading axis.) The nonlinear portion of the stress-strain curve in compression, seemingly related to ductility, is often reduced comparatively in materials that can sustain at least moderate amounts of inelastic action in tension; the opposite may be true with apparently brittle materials because right-section rupture cannot occur under compressive loading. Examples of actual or potential shear deformation, associated with orthodontic appliances, were cited previously; a schematic was shown in Figure 3-5 and the shear strain, generally symbolized by y or is defined as the tangent of the angle developed as two intersecting line segments at a point within the material, passively perpendicular, are inclined to one another with the member under load. In a test to determine mechanical properties in shear, the specimen and loading are typically similar to the bonded bracket assembly of Figure 3-14 except that
the specimen is quite small, is of one material, and in particular the moment distance of the applied force with respect to the specimen support is minimized. Readings of load and angular deformation obtained are converted to shear stress and strain, and the reduced data are plotted to give the stress-strain diagram. Two example diagrams are presented in Figure 3-19. The form of the plot for the crystalline material is, by now, familiar; mechanical properties as derived from the diagram are defined exactly as in the tension or compression test except each name ends with the phrase "in shear." The slope of the linear part of the shear-test curve, however, is generally often termed the modulus of rigidity G. From stiffness magnitudes, many metallic materials seem to be more flexible in shear; for arch-wire alloys, the ratio of elastic moduli in identical units,
shear to axial tension, is about 0.4. (Although measured differently, both normal and shear strains are nondimensional, so these stiffnesses are dimensionally alike.) Hooke's Law for pure shear loading is written T = G-y and is valid up to the proportional limit in shear. For ductile materials in general, the elastic and ultimate strengths in shear are about one-half of the comparable values in tension; for brittle materials, the ratio of ultimate strength in shear to that in tension approaches unity.
99
Material Behavior of the Orthodontic Appliance
Shear Stress
A crystalline material
Shear Strain
0
FIGURE 3-19. Static stress-strain diagrams in shear for crystalline and polymeric materials.
The forms of the stress-strain diagram in shear and in compression are
similar for most molecular and amorphous materials; the initial nonlinear portion of the three-part curves in tension of some noncrystalline materials, as shown in Figure 3-18, is rarely seen under compressive or shear loading.
Previously depending upon mechanical-property data generated elsewhere, in 1977 the Council on Dental Materials and Devices of the American Dental Association announced the ADA Specification Number 32 for orthodontic wires not containing precious metals. The Specification includes a detailed description of the preparation and testing of arch-wire specimens toward the determination of the modulus of elasticity and "yield strength" for the arch-wire material in flexure (bending). A schematic of the cantilevered-beam test is given in Figure 3-20. A bending couple is applied to one end of the specimen where only rota-
tion is permitted; at the other end of the test span the wire is held against a fixed, knife-edge stop. The angular deformation measured is the rotation of the shaft, which is also the rotation of the end cross-section of the specimen engaging the shaft with respect to the passive orientation. A typical plot of the couple versus the rotational deformation, for a specimen of metallic material not exhibiting a well-defined yield point, is
100
Bioengineering Analysis of Orthodontic Mechanics
Load couple C
FIGURE 3-20. An illustration of the fiexure-test arrangement dictated by ADA Specification No. 32.
presented in Figure 3-21. To obtain an equivalent yield point, approximating the elastic limit, the dashed line is drawn parallel to the initial, linear portion of the plot, intersecting the deformation axis at the specified offset (approximately 2.9° according to Specification No. 32), and the yield strength of the material is determined from the couple magnitude at the point. (The actual computations to quantify the elastic modu-
lus and yield strength in flexure are outlined in Chapter 7.) A third material property of orthodontic interest may then be computed; the ratio of yield strength to modulus of elasticity is termed the springback of the material and is a close approximation to the elastic-limit strain. Table 3-1 contains approximate values of these three properties in bending for five arch-wire materials. In addition to the static flexure test, an experimental procedure that may be terminated after the data confirms exceeding of the elastic limit, another is required which, in part, provides an indication of the ductility
of the wire material. In this second portion of the Specification, a 90° bend is placed in the wire, then removed, and then replaced at the same cross-section and in the same direction. The radius of the bend is specified. The 90-to-0-to-90° bending cycle is continued at a specific rate until fracture occurs. The number of cycles to failure is then compared with the Specification requirements, which vary somewhat with wire size. Before concluding this section, several additional comments associated with testing and mechanical properties are in order. Indicating that a material is relatively "hard" or "soft" does not pertain, strictly speaking, to any previously discussed property. Hardness is defined as the resistance of a material to localized, permanent, compressive deformation (indentation); it is, therefore, largely a surface phenomenon. Hardness numbers are determined experimentally by subjecting a material specimen to an indenter of a specific geometry exerting a designated load. Indices of hardness familiar to engineers are the Brinell and Rockwell hardness scales. Although hardness is very nearly, directly proportional to ultimate tensile strength for a number of metallic materials, this property must not be confused with material stiffness or resilience. Mechanical properties of some materials are difficult to quantify by testing because their stress-strain characteristics change with time under
101
Material Behavior of the Orthodontic Appliance
Load Couple C
Equivalent
yield point
/ /
/
/ Stiffness
/
/ / 0
Angular Deformation 0
FIGURE 3-21. Couple versus angular deformation; typical plot of flexure-test data obtained according to ADA Specification No. 32.
load. Although not a factor at room or oral temperature with most metals, often it is with molecular materials, particularly certain polymers. Subjecting a typical, orthodontic, elastic module, thread, or "chain" to a moderate tensile load, and holding that load constant while monitoring deformation, a continuing elongation will be observed over a period of hours. The material is said to creep and, from data taken, a strain-versustime curve may be generated similar to the plot of Figure 3-22. If the temperature or time period is sufficiently high or long, the straight-line portion of the curve will give way to a nonlinear increase in strain fol-
lowed by rupture, even though the stress level may be substantially below the ultimate strength of the material as gained from a static, tension test at room temperature. A material that creeps will also "relax." Relaxation is the decrease with time of load carried under conditions of constant strain; a relaxation plot is presented in Chapter 4.
102
Bioengineering Analysis of Orthodontic Mechanics
TABLE 3-1. Mechanical pmperties in bending for selected, orthodontic, arch-wire materials
Modulus of elasticity
Yield strength
Springback*
(x106 lb/in.2)
(x103 lb/in.2)
(x102 in/in.)
A gold alloy
15
150
0.94
Type 302 stainless steel
29
280
0.97
Elgiloy (Cr-Co alloy)
28.5
315
1.1
Material
Nitinol (Ni-Ti alloy)
4.8
195.
4.1
Beta titanium (Ti-Mo alloy)
9.5
170
1.8
upon unloading from the (equivalent) yield point: the yieldstrength-to-elastic-modulus ratio. Yield strength and springback values will vary with cold working or heat treatment. *The recovered, unit
The final topic considered in this section is fatigue, the progressive failure of a material undergoing loading that changes with time. Usually
such loading is cyclic, exhibiting a repeating pattern; the maximum stress generated may be substantially below the ultimate strength of the material obtained from a static test, and still result in an eventual fracture. Examples of cyclic loading pertinent to orthodontics include masti-
Strain
Stage 3
0
Time
FIGURE 3-22. A strain-versus-time plot illustrating the three stages of creep.
103
Material Behavior of the Orthodontic Appliance
Stress
Ultimate strength
Endurance limit
0
Number of Cycles to Fracture
FIGURE 3-23. A typical plot of the results of tests to fracture of a metallic member subjected to cyclic loading in bending.
(relatively high-frequency cycling) and the daily activation, continuous-force, deactivation pattern associated with an extraoral cation
appliance and interrupted loading (low-frequency cycling). A common fatigue failure is the progressive fracture of a bent-wire segment emerging from the acrylic which is part of a retainer; the removal and replacement of the appliance represents one cycle. Figure 3-23 shows the re-
suits of a series of fatigue tests in bending. The curve intersects the stress axis at the ultimate-strength level. The inclined portion of the plot is obtained from fractures occurring at maximum stress levels below the ultimate-strength value; the lower the stress, the greater the number of cycles to failure. Below a particular stress level, termed the endurance limit, fracture does not occur regardless of the number of load cycles. With the retainer in which the wire eventually fractures, the maximum
stress in the activated wire ioop is less than the ultimate strength but greater than the endurance limit of the wire material. Besides indicating
level of ductility, the previously described bending test to fracture within ADA Specification No. 32, owing to the cyclic-loading pattern required, also indirectly provides an approximate, relative measure of arch-wire resistance to fatigue.
Chemical and Thermal Influences With orthodontic application in mind, a discussion of the chemical and thermal behaviors of solids is undertaken. The chemical and thermal material properties themselves are not the principal focus of atten-
104
Bioengineering Analysis of Orthodontic Mechanics
lion, but rather the potential for interactions that may result in deleterious effects upon the mechanical or physical characteristics of orthodontic materials. Some, and often all, of the members of the ortho-
dontic appliance must exist and "work" for substantial periods of time in the oral cavity. Materials used in the mouth must be biologically admissible and the potential for interference with acceptability, because of chemical reactions that may occur over time, either before or during intraoral engagement, must be realized and understood. The oral-cavity environment is inherently corrosive. The oral fluids are strong, potential reactants toward oxidation of metals. Saliva is known to contain salts and acids are often liberated during mastication; ingested food and drink vary widely in their levels of acidity or alkalinity (pH). Food debris may become lodged and remain for relatively long periods of time and provide the catalyst for initiation of corrosion; although generally a dental-care matter, the tendency for this occurrence is increased by the presence of appliances. Both metallic materials (e.g., restorations, wires, bands) and molecular solids (e.g., "elastics," cements, adhesives, acrylics) may be increasingly vulnerable to chemical degradation in the moist oral environment. The effects of chemistry and temperature upon the materials themselves, as well as the products of chemical reactions, must be of concern. The extraoral environment may also influence the structural capabilities of orthodontic appliance materials. The time period between manufacture and actual use can be substantial. Storage may occur with the manufacturer, with the vendor, and with the practitioner. Molecular materials, in particular, often have a finite "shelf-life," which may be reduced by high temperature and humidity. Parts of extraoral appliances come into contact with skin and hair, and the oils as well as perspiration, often together with particles suspended in the air, can and do have a degrading effect. Corrosion is defined as a deterioration because of a chemical reaction that results in apparent disappearance of the material attacked. Corrosion may occur by a chemical solution contacting the material; ionization ensues and the material is dissolved in the fluid. For example, silver will be corroded by a solution containing sulfides with silver sulfide formed in the solvent. A more common form of corrosion is known as electrochemical oxidation. Characteristic of metallic materials, the process begins with a reaction initiated by a fluid that removes electrons, thereby form-
ing positive material ions. In turn, chemical combination of these ions with electrically negative ions in the fluid may take place. Rusting starts with the stripping of three electrons from the iron element, forming the
ferric ion. The degree to which electrons are bound to the atoms in metals varies, and so also does their tendency to form compounds in the
presence of nonmetals having incomplete outer electron shells. Gold and platinum are relatively inactive and, in mining, are often found in the pure form; iron and aluminum, for example, as active metals, are continually oxidized and must be purified (reduced) chemically before metallurgical processing.
105
Material Behavior of the Orthodontic Appliance
Generally, corrosion occurs as a combination of the two forms described. The facility for the ionization of a metal is dependent upon the nature of the fluid-solvent and the inherent activity level of the material; the latter is often termed the electrode potential of the metal and is a direct indicator of its corroding tendency. The determination of the ranking of metals by electrode potential is accomplished through use of the "standard cell" shown in Figure 3-24. Immersed in the solution is the test-metal
electrode on the left and a hydrogen electrode on the right, the two connected externally through a potentiometer. Ionization begins at both
electrodes; however, depending upon the electrode potential of the metal with respect to hydrogen, electrons will flow one way or the other, registering a potential difference on the meter. The experiment is
Potentiometer
H2
V TM ±
1-1k
TM
Electrolyte
Positive ion of test metal or alloy H2: Hydrogen gas Hydrogen ion FIGURE 3-24. A standard electrochemical cell.
106
Bioengineering Analysis of O,thodontic Mechanics
repeated, changing only the test metal. With all potential-difference val-
ues recorded, a ranking such as that given in Table 3-2 is obtained with hydrogen as the reference ion. Consider a cell such as that shown in Figure 3-25 with iron as the electrode on the left and water as the solution. From Table 3-2, iron has a
higher electrode potential than tin and, upon ionizing, electrons removed from the iron flow through the connector to the hydrogen electrode. The hydrogen gas is also ionized and water molecules are decom-
posed into hydrogen and hydroxyl (OH) ions. The electrons coming from the iron unite with hydrogen ions and hydrogen gas is liberated. 3-2. Electrode-potential rankings, with respect to hydrogen, of selected metals and alloys TABLE
The ranking to follow is in the order of most anodic, at the top of the list, to most cathodic at the bottom of the list. Magnesium Aluminum, active Titanium Cesium Vanadium Zinc Aluminum, passivated Chromium Iron
Stainless steel, active Cadmium Cobalt Nickel, active Lead-tin solder Tin
Lead
nconel*, active HYDROGEN Brass
Copper Bronze Monelt Nickel, passivated lnconel*, passivated Stainless steel, passivated Silver Palladium Mercury Platinum Gold lnconel is a nickel-chromium alloy. tMonel is a copper-nickel alloy.
107
Material Behavior of the Orthodontic Appliance Electron flow
Fe3t Iron ions OH—: Hydroxyl ion Fe(OH)3: Rust
FIGURE 3-25. A galvanic cell illustrating the electrochemical corrosion of iron.
ferric ions react with the water and oxygen therein to form ferric hydroxide: rust. The rusting of iron will occur in such a galvanic cell, in fact, whenever the iron is connected externally to an electrode having a lower potential (occupying a lower position in Table 3-2) and the elecThe
trodes are immersed in a solution (the electrolyte) wherein hydroxyl ions become available with ionization. In general, the electrode supplying the electrons is called the anode and the receiver of electrons is the cathode. The anode experiences this galvanic corrosion and the location of
the corrosion products will depend upon the relative ease with which the ions, forming those products, are able to diffuse through the electrolyte. In a galvanic cell containing iron and a less active metal, because ferric ions are smaller than hydroxyl ions and because the product requires three OH ions for every ferric ion, rust will usually be found at or near the cathode. The formation of a galvanic cell with the physical contact of an orthodontic-appliance member and an amalgam restoration, in the presence of saliva as the electrolyte, results in momentary ionization and electron flow and the dental "galvanic shock." This is an example of a composition cell established between two dissimilar metals. A tin roof, placed on a
108
Bloengineering Analysis of Orthodontic Mechanics
building having steel siding, becomes cathodic and corrosion begins in
the presence of moisture and direct contact between roof and siding. Important in addition are the more subtle forms of electrochemical cells. A stress cell may form between two different locations on the same material, because of an imbalance in electrode potential. A vivid example of stress corrosion is the etching or corroding of grain boundaries; the
atoms at the boundaries are more loosely bound and, therefore, are potentially anodic with respect to those within the crystals. It follows, then, that a fine-grained metal is potentially more vulnerable to corrosive attack than is a coarse-grained metal. An imbalance in electrode potential at different surface locations of the same material results from localized inelastic action. The region of a permanent bend in an arch wire is anodic in comparison to the passively-straight portions of the wire. A concentration cell may occur involving one material (or two) when the electrolyte is nonhomogeneous. The surface area exposed to an electrolyte is potentially more active where the concentration is lower. In a one-material cell, the region exposed to a diluted electrolyte is anodic with respect to a region in contact with a concentrated electrolyte. The
significance here is the anodic nature of cracks, crevices, and areas under accumulations of debris or other surface contaminations. Moreover, often the deposits of corrosion products aggravate the situation and increase the rate of degradation; localized pitting of a material is a typical manifestation. Electrochemical corrosion requires the presence of two electrodes, physical contact between them for conduction of electrons, and an ionizing, ion-carrying fluid, the electrolyte. Retardation or elimination of corrosion tendencies necessitates minimization of the electrode potential or a break in the "circuit." In two-material systems, corrosion may be prevented by insulating one material from the other or one or both materials from the electrolyte. With one material, isolate it from the electrolyte, keep its surface smooth, clean, and free of sharp edges, and reduce its inherent electrode potential. Steel coated with zinc is "galvanized" to separate the steel from moisture. Because zinc is more active than steel, it is the potential anode should a crack or scratch expose the steel. Orthodontic stainless steel generally resists corrosion well because its surface is protected by passivation with chromium oxide, the nickel in the alloy allows a reduction in carbon content (and a lessening, then of the potential for chromium-carbide formation while retaining strength and ductility properties), and the result is a position in Table 3-2 just above silver.
The foregoing discussion has dealt primarily with "wet" corrosion in which a liquid electrolyte or at least moisture is necessary to the reaction. Dry corrosion is the result of a chemical reaction of the affected material with a gas, often air. With metals the specific gaseous reactant is oxygen and an oxide scale is formed, initially on the metal surface. In general, the rate of increase of the film thickness depends upon the existing thickness, the diffusion coefficients for the reactants, and the temperature. Continuation of the reaction is dependent upon the pene-
109
Material Behavior of the Orthodontic Appliance
trabiity of the scale film by either the gas or the metal ions. An example of dry oxidation is the passivation of stainless steel; the chromium oxide film is very thin and highly impenetrable. Dry corrosion of molecular materials also occurs; of particular interest is the aging of elastomers. In this oxidation reaction the initial result is an increase in cross-linking (secondary bonding), which causes harden-
ing, stiffening, and embrittlement. Heat, light, stress, and ozone concentration influence aging, which takes place because of the existence of
free-end molecular chains. Antioxidants may be incorporated during polymerization to combine in a monofunctional manner with the chain ends to provide some resistance to aging. The practitioner must be aware of aging and its effects, particularly with regard to the time period and location of storage of polymeric, orthodontic-appliance elements. In a moist environment polymeric materials experience a combination of solution and aging corrosion. "Elastics" that have been stored may contain aging cracks that are catalysts for further and accelerated degradation following activation; intraoral placement of these elements tends to quicken deterioration compared to extraoral application. Another type of degradation occurring in the orthodontic setting involves extraoral appliances, in particular the neck pads, head caps, and
other nonmetallic elements that come into contact with the skin and hair. A form of wet corrosion occurs as skin and hair oils, and perspiration attack these molecular materials over time. Although causing some aging, of equal or perhaps more concern is the chemical reaction initi-
ated by the hydrocarbons that ultimately breaks down the material structure into small molecules, thereby drastically reducing mechanical strength. The progress of this deterioration is relatively slow, and material discolorations indicate the start of the process. Frequent cleansings of washable parts of the extraoral appliance, for example, retards this form of corrosion. The effects of temperature on a material are, to a great extent, manifested through influences upon properties already discussed. The fundamental material response to change in temperature is on the level of atomic activity which, in turn, affects volume. A typical unconstrained solid will expand upon experiencing an increase in temperature; such a material will contract as its temperature is lowered. (Some polymeric materials, however, react differently with temperature changes due to accompanying bonding alterations.) A structural member that is constrained against undergoing volumetric change will be subjected to induced thermal stresses if its temperature is altered; these stresses are superimposed upon those that exist due to mechanical loading. The increased vibrational motions of atoms and molecules, accompanying increased temperatures, generally affect mechanical behavior in that ductility characteristics are enhanced but strength properties are reduced. The rates of, or tendencies toward, creep and relaxation grow as the temperature rises. The greater the atomic motions, the more easily materials may be ionized; hence, increased temperatures generally raise the rates of corrosion and oxidation.
110
Bioengineering Analysis of Orthodonfic Mechanics
In general, the effect of an incremental change in temperature is greater upon nonmetals than upon metals. Pertinent to orthodontic application, the influence of the difference between room and oral temperature, or between that of ingested ice cream and coffee, is inconsequential upon arch-wire materials. The effects of such temperature differentials, however, are discernible in many of the elastomeric materials. Although the metallic appliance components do not experience large temperature ranges during actual orthodontic therapy, they may in soldering, welding, or heat-treatment procedures prior to engagement and activation. Note that when the temperature is raised to about 900°F and above, the carbon atoms in stainless steels tend to migrate toward the grain boundaries; they combine with chromium ions to form chromium carbide, and the passivating potential of the alloy is lessened. Hence, care must be taken in soldering or welding this material, and other metals and alloys, to concentrate the heat and hold it for only a very short time.
Selection of Materials In concluding this chapter, comments pertaining to the selection of ma-
terials are appropriate. Although materials and appliance members associated with orthodontic therapy are of specific interest, the same criteria are reviewed and examined in the intelligent selection of materials for any structural or machine application. The relationship of the relevant characteristics or properties of the material to its intended use must be the first and overriding considera-
tion in material selection. The orthodontic appliance is a structural mechanism and, with appliance members available in varieties of physical sizes, attention is focused primarily on mechanical properties. Archwire materials must be ductile if arch fabrication, including the place-
ments of permanent bends and twists, is necessary. Moreover, arch wires must be moderately flexible while possessing substantial elastic strength to provide desired activation and deactivation responses. All parts of the appliance must exhibit strength characteristics to sustain without failure the stresses induced in appliance activation, both at the moment of engagement and over time periods ranging from a few days to, perhaps, several years. Possible chemical reactions, which may affect mechanical behavior or result in unwanted product compounds, must be a factor of serious consideration, particularly in the selection of intra-
oral appliance materials. Ideally, the mechanical characteristics of a structural material should not be influenced by time or by temperature fluctuations that may occur in the service environment; however, with real materials, one or both of these parameters may exert significant impact on mechanical behavior.
111
Material Behavior of the Orthodontic Appliance
Other factors to be evaluated are generally subordinate to those directly associated with in-service use and are given significant weight in the selection decision only when, from principal considerations, two or more materials seem equally suitable or when a compromise is warranted and can be tolerated. The ranking of these additional factors will vary, depending on the specific application and, often, on financial influences. Costs will clearly impact the selection, and they include the initial investment, expenses associated with subsequent fabrication, and costs related to storage of the material. A compromise directly related to initial cost has led to the virtual disappearance of gold alloys as archwire materials. Availability is another factor that may influence initial cost and may, at least temporarily, impact the selection procedure. In any business involving a product, inventories must be kept, and orthodontic practices are no exception. Accordingly, the effects of storage on the material and the needs for handling and working with the material, in transforming it from the condition "as received" to the state of readiness for immediate use, come into consideration. Chemical degradation limits the "shelf life" of a number of materials used by the orthodontist. Moreover, many cements, adhesives, alginates, and acrylics, for example, require storage in controlled environments or careful mixing to produce chemical reactions resulting in optimum material characteristics. Not to be ignored, but often rightfully near or at the end of the list of influencing factors, is appearance. Although advertising may attempt to inflate the importance of external attractiveness, the potential user must place this factor in its proper position in the list of selection criteria. Although often an apparently difficult decision, a sacrifice within the primary selection factor, that of relation of material properties to inservice use, should rarely be made solely to accommodate appearance.
Suggested Readings Burstone, C.J.: Application of bioengineering to clinical orthodontics. In Current
Orthodontic Concepts and Techniques. 2nd Ed. Edited by T. M. Graber and B. F. Swain. Philadelphia, W.B. Saunders, 1975, Chapter 3. Burstone, C.J., and Goldberg, A.J.: Beta titanium: A new orthodontic alloy. Am. J. Orthod., 77:121—132, 1980.
Greener, E.H., Harcourt, J.K., and Lautenschlager, E.P.: Materials Science in Dentistry. Baltimore, Williams & Wilkins, 1972, Chapters 3—6, 8, 11, and 13.
Hayden, H.W., Moffatt, W.G., and Wuiff, H.C.: The Structure and Properties of Materials. Vol. III. New York, John Wiley & Sons, 1965, Chapters 1—3, 6, 7, and 10. Jarabak, JR., and Fizzell, J.A.: Technique and Treatment with Light-wire Edgewise Appliances. 2nd Ed. St. Louis, C.V. Mosby, 1972, Chapter 3.
112
Bioengineering Analysis of Orthodontic Mechanics
Kusy, R.P., and Greenberg, A.R.: Effects of composition and cross-section on
the elastic properties of orthodontic arch wires. Angle Orthod., 51:325—341, 1981.
Richards, C.W.: Engineering Materials Science. Belmont, CA, Wadsworth Publishing, 1961, Chapters 1—4, 8, 9, 12, and 13. Phillips, R.W.: Skinner's Science of Dental Materials. 7th Ed. Philadelphia, W.B. Saunders, 1973, Chapters 2, 3, 11, 15—19, and 32.
Thurow, R.C.: Edgewise Orthodontics. 4th Ed. St. Louis, C.V. Mosby, 1982, Chapters 3, 7, and 8. Van Vlack, L.H.: Elements of Materials Science. 2nd Ed. Reading, Massachusetts, Addison-Wesley Publishing, 1964, Chapters 1—3, 6, 7, and 12.
Energy Analyses in Orthodontics ]
The theoretical developments within classical mechanics have provided two analysis formulations. The four laws offered by Isaac Newton are
the foundation for one approach toward the solution of problems in mechanics, and two mathematicians—Jean Bernoulli, a Swiss, and J. L. Lagrange, a Frenchmen—are generally given credit for the more con-
temporary hypotheses and methods of work and energy. In Chapter 2 the procedures of Newtonian mechanics were introduced as based upon a set of equations of equilibrium or motion, depending upon the state of rest or movement of the mechanical system. The free-body diagram was introduced and employed as a principal tool in the solution of problems. Generally, the objective was the altering,
simplifying, or completing of the description of the set of forces and couples exerted at a particular instant of time upon or within the body or the elements of a structure or machine. In the analysis procedure employing work and energy, the mechanical system undergoes an actual or
contrived process that may be of short or long duration and results in changes in one or more of the characteristics of the system. To prepare for an analysis through use of the principles of work and energy, the body or set of elements or bodies included within the mechanical system under study must be defined. The initial state of the system is described at a particular time; physical, geometric, kinematic, thermal, chemical, and mechanical properties—any characteristics that may experience change during the ensuing process—are noted. The process that is to take place results from an interaction between the system and its surroundings. Therefore, the surroundings, which will influence
the changes in the mechanical system, must also be characterized. Of particular importance are (1) the set of forces and couples to be exerted by the surroundings upon the mechanical system under study and (2) any heat sources or heat "sinks" in the surroundings that may interact with the system. With the system and its surroundings adequately described to define the initial state, an examination of the process is undertaken. A process is the interaction between a defined system and its surroundings that results in a change of state (alteration of character, changes 113
114
Bloengineering Analysis of Orthodontic Mechanics
in property values) of the system. The final state of the mechanical system is described in terms of the properties of the system following com-
pletion of the process. In summary, the analysis is, in general, toward the solution of the following problem: Given the initial state of a mechanical system to be studied and the description of the interaction be-
tween the system and its surroundings during the defined process, characterize the final state of the system. Several processes associated with orthodontic treatment may be analyzed more directly through energy methods than by Newtonian mechanics. Included, and to be discussed in this chapter, are the processes of activation and deactivation of the orthodontic appliance, the placements of permanent bends and twists in an arch wire prior to intraoral engagement, the various forms of heat treatment to which arch wire and some auxiliaries may be subjected, and the actual process of manufacturing arch wire. Because each process to be analyzed generally involves some displacement or motion of the system, and the interaction may include the transfer of heat toward affecting the thermal as well as the mechanical characteristics of the system, it is appropriate that the body or group of bodies or system elements to be studied be termed a "thermodynamic system."
Concepts Leading to the Process Laws In this section appropriate concepts are introduced or reviewed toward
stating and understanding the process laws that govern the work-andenergy approach to mechanics problems.
Displacement As described in Chapter 1, the displacement of a particle is the change in
its position with respect to a reference framework. A straight-line, particle displacement is shown in Figure 4-1. The initial position of the parti-
cle is defined by the vector r1, drawn from point 0 to point P. The particle then moves to point Q. Labeling the coordinate s along the displacement path, the distance covered may be symbolized as where prefix represents "change in" the parameter. Because the the delta displacement is along a straight line in this example, the distance traveled by the particle equals the magnitude of the displacement vector. The displacement vector may be expressed more generally as the difference between the final and initial position vectors, r2 — r1. The displacement of just one point or particle of the body cannot indicate whether or not the orientation or angulation of the body was concurrently changed. If the body is rigid and is moved, but without an angular change, the displacement of the body is a translation and the displacements of all particles of the body have identical vector characteristics. Alternatively, if the orientation of the reference line in the rigid
115
Energy Analyses in Orthodontics
y
Q .—
p
r2
0
x
FIGURE 4-1. The displacement of a particle along a straight-line path.
body is changed in the displacement, some form of rotation has occurred. If the body is nonrigid, linear displacements of particles or angular displacements of lines of the body may be the result of widespread or
localized deformations. Displacements accompanying deformations, both visible external changes in shape and alterations internal to the body, are of substantial interest in this chapter. Orthodontic displacements to be studied subsequently include that of one end of an "elastic" with respect to the other in an extensional activation, the displacements that occur in the elongation or compression from the passive state of a helical-coiled spring, the occiusogingival or faciolingual bending deactivation of an arch wire accompanying tooth movement, and the placement of a tip-back or toe-in bend in an arch wire. Mechanical Work
Displacements produced mechanically, whether or not the body experiences deformation, are associated with forces or force systems. Both
active and responsive forces may cause displacements; on the other hand, displacements can result in the creation of forces. Figure 4-2
116
Bioengineering Analysis of Orthodontic Mechanics
y
F
P
0
x
FIGURE 4-2. A concentrated force exerted upon, and during its displacement doing work on, a particle.
shows a force exerted on the particle of Figure 4-1. As the particle under-
goes a displacement with the force acting on it, regardless of the relationship between the force and the displacement, the force is said to be doing mechanical work on the particle if the force has a nonzero component in the direction of the displacement. Although, in general, the magnitude or direction of the force may change during the particle displacement, if the vector characteristics of the force shown in Figure 4-2 are all constant, the work done by the force F on the particle as it moves from point P to point Q is equal to F(cos Although computed from two vectors, mechanical work is a scalar quantity. The component of the force involved is that having the direction of the displacement vector. If the force component and displacement vector have the same sense, the angle 4 is less than 900 and the work is algebraically positive; if the senses are opposite (q5 is between 90° and 180°), the work done is negative. If the force is continually perpendicular to the displacement vector, no work is done on the particle by the force. The force of Figure 4-2 would properly be termed an external force if
exerted on an isolated particle or a particle on the outside surface of a body; if the particle is internal to the body, the contact force would also be internal. Internal forces, occurring in canceling pairs, do zero net work on a rigid body; however, within a nonrigid body the net work of a pair of equal and opposite internal forces may be nonzero because the
117
Energy Analyses in Orthodontics
particles on which they act may undergo unlike displacements. In the
absence of friction, the net work done by the force systems connecting one body to another is zero if the parts of the connection are rigid. A couple exerted on a body does work if the body experiences an associated rotation; the work is positive if, in the pertinent plane view, the couple and angular displacement have the same sense. Friction may also do work; if the force of friction is resistive to the displacement, the frictional force and displacement vectors will have opposite senses and the work of friction will be negative.
Work is done by the orthodontist in the placement of permanent bends or twists in arch wires; the forces are transmitted by the hands of the practitioner to and through the pliers and to the wires. In the placement of a tip-back bend, for example, the force system exerted by the pliers on the end of the wire, causing the rotational displacement beyond the elastic limit of the wire resulting in permanent deformation, is
a couple. The clinician does work on an elastic in stretching it, on a spring in compressing it, and on an arch wire in actively engaging it. Appliance elements do work on teeth as their force systems produce desired or unwanted displacements of dental units. In the typical bend or twist placement, activation, or deactivation, the force doing the work changes in magnitude with the displacement, requiring an averaging or a graphical procedure to quantify the work done; examples presented later in the chapter illustrate the methods of computing work in these instances.
The typical process in which mechanical work is done involves a change of state that is reflected partially in the displacements of particles of the mechanical system under study. A cause-and-effect relationship
exists between force and displacement in the process. The external forces that may do work on the system are exerted by elements of the surroundings of the mechanical system, indicating that this form of work is an interaction between system and surroundings. On the other hand, the work of internal forces occurs entirely within the system being
analyzed. In the overall analysis of a process, these two categories of work are often separated because of this difference and its ramifications.
Energy Newton's Second Law for a particle of mass in under the action of a
single, constant force F suggests that this force produces a proportional increase in the speed v of the particle with time t. For a one-dimensional motion, the relationship may be written as (4-1)
where Av and the displacement coordinate s have the same sense, identical to that of the force, as shown in Figure 4-3. The work done by F during the displacement of the particle from point P to point Q may be
118
Bioengineering Analysis of Orthodontic Mechanics m
P
F
Q
0
v+Av
v
As
At FIGURE
4-3. Changes in kinematic characteristics of a particle as work is done on it.
determined. Multiplying both sides of Equation 4-1 by As, noting that
speed is the time rate of displacement, and with appropriate mathematical operations, the following formula for mechanical work is obtained: Work = F(As)
+ Au)2 = + m[(v
—
v2}
(4-2)
The quantity ½mv2 is recognized as the kinetic energy of the particle; Equation 4-2 indicates that the work done on the particle results in a change in the kinetic energy of the particle. This relationship may be expanded to yield the law of kinetic energy: the work of all of the forces, internal as well as external, acting on a mechanical system, equals the change in kinetic energy of the system. The speeds used in the formula must be referenced to a stationary coordinate framework or one moving in a straight line with constant speed; such reference frames are termed Newtonian. Kinetic energy is energy of motion and the kinetic-energy law is a form of, and in one sense the basis for, the more general process law toward which the discussion of this section is directed. A rationale for partitioning the forms of work done in a process by the external and internal forces was suggested earlier. A parallel division separates the work done that is dependent on the displacement path(s) from that which is dependent only on the initial and final states of the mechanical system. An example of path-independent work is that which is performed by the force of gravity (weight). For processes during which the center of gravity of the system undergoes a change in elevation that is small compared to the radius of the Earth, the magnitude of the work done is the product of the weight and the elevation change. Because the force of gravity acts directly downward, gravitational work is negative when the elevation increases. A second example of path-independent work is that which is done on or by a helical-coiled spring in activating or deactivating it in direct tension or compression. In this example the magnitude of the force changes with the amount of spring deformation. Note that the force activating the spring and the resulting displacement always have the same sense; therefore, the work of activation on the spring is positive and by the spring (and on the attached body) is negative.
119
Energy Analyses in Orthodontics Passive state 0
, —
Initial state 1 ______x
Work12 =
Activated state A
Displaced state 2
x
Work done by gravity (left) and within a spring as it is stretched (right).
Figure 4-4 shows the forces and displacements associated with the work done by an external weight force (left) and that by the internal
forces in a spring (right). Because the amounts of work done are dependent only on the initial and final configurations of the system, and with
the form of the law of kinetic energy, such path-independent work is defined as energy change. Specifically, the work done by the weight force is, in magnitude, set equal to the change in gravitational potential energy, and the work done by the internal spring forces is, in absolute value, set equal to the change in strain energy of the spring. Moreover, in the examples of Figure 4-4, increases in energy levels of the bodies, increased elevation of the center of gravity and increased deformation of the spring, are seen to correspond algebraically to negative amounts of
work done on the bodies. To this point in the discussion, two forms of external energy, kinetic and gravitational potential, and one form of internal energy, strain energy, have been introduced. Energy is defined as the capacity of a body or system to produce an effect; thus far that effect has been, exclusively, mechanical work. Energy is a scalar quantity and generally is referenced
so as to be non-negative. Although kinetic energy has an apparently natural reference (zero speed), gravitational potential energy has no such convenient reference level. Although initially it might seem reasonable to set strain energy equal to zero for the passive state of the spring, this form of internal energy is added to the spring in the manufacturing process when it is given its coiled shape. In actual process analyses, only
changes in energy levels are determined. Zero-energy references are unnecessary, but for each individual process such references may be chosen in a reasonable manner, for each energy form, if so desired.
120
Bioengineering Analysis of Orthodontic Mechanics
A work-energy analysis of standard testing procedures may be undertaken to determine certain mechanical-property values. Such tests were discussed initially in the previous chapter. The general process of interest is the loading of the test specimen. Because the loading is slow, the kinetic-energy change for the process is negligible. Thus, the law of kinetic energy reduces to "The total work of all forces, when individual contributions are algebraically summed, equals zero." With the work of
the weight force, the change in gravitational potential energy, also negligible, the process equation becomes the sum of two work contributions, that done by the external force system in loading the specimen (a positive quantity) and the work of the internal, responsive forces (a negative quantity), which must be zero. Substituting for the work of the internal forces, the final form of the process says that the work done in loading the specimen equals the increase in strain energy of the specimen. Figures 4-5 and 4-6 show the stress-strain and load-deflection diagrams associated with the standard uniaxial tension and ADA bending tests, respectively, which were described in Chapter 3. (In obtaining the diagram of Figure 4-5, the initial volume of the gauge-length portion of
the specimen has been divided out, converting force and deformed length into stress and strain.) With the loading force system increasing in magnitude with deformation, and the resulting need to sum quantities of the form F[AsI (in the tension test cr[A€J) to obtain the external work done, the total work in loading the specimen from its passive state to some deformed configuration is the area under the diagram up to that activated state. The process law says this area also equals the increase in strain energy. Therefore, in Figure 4-5, the process that takes the specimen from point 0 to point el, the elastic limit, requires positive, external
Stress
C
ci
area OABO: modulus of resilience area OCDO: modulus of toughness
0 0
B
Strain
FIGURE 4-5. A stress-strain diagram generated from the data of a standard tensile test.
121
Energy Analyses in Orthodontics
p
Load Couple
a!
area OMNO: resilience area OPQO: toughness
Q
0
N
Angular Deformation
FIGURE 4-6. A load-deformation diagram plotted from the data of a bending test of a cantilevered beam.
work per unit volume, and increases the strain energy per unit volume an amount given by the triangular area under the stress-strain plot up to the dashed line. This area is also the magnitude of the mechanical property known as the modulus of resilience in tension ("modulus of" denoting
per-unit-volume). The amount of strain energy per unit volume that may be added to the specimen before fracture occurs is equal to the entire area under the stress-strain diagram, and is known as the modulus of toughness in tension. Analogous areas under the diagram of Figure 4-6 give the resilience and toughness of the arch-wire specimen in bending. Note that both the resilience and toughness magnitudes are dependent upon the "stress history" of the specimen. In their initial states the specimens contain unknown levels of internal energy; these levels are dependent upon the mechanical and/or thermal processes that occurred prior to the tests (including manufacturing operations). Just as a maximum load exists, which the specimen can sustain without fracture, the specimen also possesses a finite capacity for strain energy, and fracture will occur when this maximum level is exceeded. Heat Transfer and Thermal Energy
In general, a process may be initiated or accompanied by the transfer of
heat between the system and the surroundings. When heat is transferred to the system, a heat source must exist in the surroundings. If heat leaves the system, it is absorbed by a heat sink in the surroundings. In the
absence of perfect insulation between system and surroundings, and
122
Bioengineering Analysis of Orthodontic Mechanics
with a temperature differential between them, heat will move from one to the other. Amounts of heat transfer are determinable; heat transfer, like mechanical work, is a scalar quantity, but it has sense, always moving from a region of higher temperature to one of lower temperature.
When heat flows to or from a system of finite size and capacity, the temperature of the system changes; the temperature increases when heat is transferred into the system and decreases when heat flows away from the system. Three forms of heat transfer are distinguishable: (1) conduction, by physical contact between system and surroundings; (2) convection, the transporting of heat between system and surroundings by a third medium, often a fluid; and (3) radiation, the emission and
absorption of rays traveling through space between system and surroundings. If the temperature is uniform within a system at a particular lime, then temperature is a property of the system and is state-related. A change in temperature of a body is reflected in the level of activity of the elemental particles that make up the body, sometimes termed the "particle kinetic energy" of the body. Appropriate, then, is the introduction of the concept of thermal energy, a form of internal energy dependent on the temperature of the body or system. Like other forms of energy, thermal energy is inherently a scalar, non-negative quantity. Its zero level might
be said to coincide with a system temperature of absolute zero, but because only changes in thermal-energy levels occur in process analyses, as with other forms of energy, a fixed reference level is unnecessary. Two types of internal energy now have been mentioned: strain and
thermal energies. Although individually they directly depend upon apparently separate, more fundamental properties, deformation and temperature, strain and thermal energies are not mutually exclusive. For example, the results of the tension tests of two specimens to fracture, initially identical in every characteristic except for a finite difference in the temperatures of the test environments, will generally yield significant differences in resilience and toughness. The lower the test-environ-
ment temperature, the more brittle the response of the specimen. Hence, although the strain and thermal forms of energy may be differentiated from each other, the more important consideration is their contribution to the total internal energy of the system under discussion. Existing experimental evidence suggests that the total internal energy of a body may depend upon other parameters besides deformation and
temperature, but only significant changes in the strain and thermal forms of internal energy occur in processes of orthodontic interest. The Conservation-of-Energy Law
A process occurs when an interaction between the system, defined for
study, and its surroundings takes place. Generally, but not always, a process results in a change of state of the system, usually reflected in an alteration in the total energy level of the system. The law of conservation
123
Energy Analyses in Orthodontics
of energy says, in simple terms, that energy can neither be created nor
eliminated within the system. If, in a process, the total energy of the system increases, energy has been transferred from the surroundings into the system. If the total energy of the system decreases, an energy transfer to the surroundings from the system has taken place. If no net energy change of the system occurred during a process (but perhaps levels of individual forms of energy have changed), the net energy trans-
fer to or from the system is zero. Two forms of energy transfer have been defined: mechanical work of
the external forces exerted on the system and heat. Several types of energy, possessed by or contained within the system have been described: the two external forms of note are gravitational potential and kinetic energy and the two significant forms of internal energy are strain and thermal energy. In the general law to be stated, energy transfers are positive when proceeding from the surroundings to the system; energy changes of the system are positive when the final levels exceed the initial
levels. The work of the weight of the system equals the decrease in gravitational potential energy; one or the other, but not both, is included in the formulation. In orthodontic applications, changes in gravitational potential energy as well as electric and magnetic energy contributions rarely occur. With the foregoing definitions, terminology, assumptions, and sign conventions, a form of the law of conservation of energy, sufficiently general for orthodontic applications, may be narrated as follows: The mechanical work of external, contact forces and couples performed on the system plus the heat transfer into the system during a process equals
the increase in internal energy of the system during that process.
This "equation" is also known as the first law of thermodynamics. It is somewhat similar in form, but is actually independent of the law of kinetic energy in that the external work term may be eliminated between the two formulas, yielding an explicit expression for the work of the internal forces. A dimensional analysis of the first-law equation indicates that energy, energy transfer, and work are dimensionally identical. The dimension of work is the product of force and length, so the same is true for heat and energy change (although a variety of units are employed: pound-feet, newton-meters (joules), gram-millimeters, ounceinches, and British thermal units (Btus, for example). The law of conservation of energy is altered somewhat for application to orthodontic processes. First, changes in gravitational potential and kinetic energies may be reasonably neglected. Second, a third form of energy transfer may be defined, to be termed losses, energy escaping from or released by the system in an uncontrolled and usually undesired manner. Because the sense of losses is always from the system to the surroundings, with the sign convention previously assigned to energy transfers, losses are inherently negative quantities. With these alterations, the law of conservation of energy, modified for strict application to orthodontic processes, may be rewritten as follows:
124
Bioengineering Analysis of Orthodontic Mechanics
The mechanical work of the external, contact forces on the system plus the
heat flow into the system minus the losses equals the net increase in the strain and thermal energies of the system.
A schematic of the orthodontic process law is given in Figure 4-7; therein SE, TE, and En represent strain, thermal, and total energy, respectively.
Available Energy A process generally begins with the system in total (mechanical, ther-
mal, and chemical) equilibrium. The final state as well is usually an equilibrium configuration. Any equilibrium state, however, may be constate (1)
En, = SE, + TE, System
WorkW
Losses L
Surroundings
Heat H
state (1)
Enf = SEf + TEf
FIGURE 4-7. A schematic depicting an orthodontically related process and the workenergy law that governs it.
125
Energy Analyses in Orthodontics
strained such that, if released from its constraints, without any catalyst
an energy transfer to the surroundings occurs as the system proceeds toward a minimum-energy configuration. Orthodontic examples of these constraints are the maintenance of an elastic in a stretched state and an arch wire at an elevated temperature in a furnace. The available energy of a system is that portion of the internal energy that is released by removing all constraints and permitting the system to come to a completely passive equilibrium state. The total internal energy of a system may be expressed as the sum of its available and unavailable ("locked in") parts. The concept of available energy becomes more meaningful in the study of orthodontic processes within the following sections of this chapter.
Activation and Deactivation Processes When activating an orthodontic appliance, the practitioner applles exter-
nal force and does mechanical work on one or more elements of that appliance. Because negligible heat transfer occurs in such processes, this quantity of work equals the increase in internal energy; since little if any
change in temperature of the system takes place, the internal energy increase is reflected in the strain-energy form. The appliance is taken in the process from a passive state to a constrained, activated state. The visible evidence of the process and the added strain energy are the deformations of activation. The activation process occurs over a short time interval and negligible losses accompany it, but losses may occur over the relatively long time period to follow when the appliance is "working" on the dentition. If no losses, as defined previously, occur subsequent to activation, and the activation does not take the material(s) be-
yond their elastic limits anywhere in the appliance,
all
of the
strain-energy "input" is available to produce displacements of teeth or other movements within the dentofacial complex. As a first example, consider the activation of a linear, helical-coiled spring in compression. (This spring might be wound around an arch wire and positioned against the mesial extent of a canine bracket, for example; when activated the spring provides the distal driving force to do the work of retraction.) The design of the spring is such that, even when the compressive force pushes all of the coils into full contact with one another, the elastic limit of the spring-wire material is not exceeded. The laws of kinetic energy and conservation of energy say that the net contribution from the external and internal forces is zero and the work done by the external forces equals the increase in strain energy of the spring. The spring, compressed from the right end is shown in Figure 4-8; the initial configuration is drawn dashed, and the final, activated state is superimposed. The activated spring is "cut" to expose the internal force system and a free-body diagram of the right-hand portion is included in the figure. Note that the responsive force against the left end
126
Bioengineering Analysis of Orthodontic Mechanics
Responsive force
d,
"cut"
force
FIGURE 4-8. The process of compressing an open-coil, helical spring.
of the spring does no work because no displacement occurs there. The relationship between the external force that compresses the spring and the spring deformation is illustrated in Figure 4-9. The spring is termed "linear" when the plot of this relationship is a straight line. The force is seen to increase with the spring deformation and the work done to produce the 8-mm compression of the spring from its passive configuration is the cross-hatched area under the plot:
Work =
=
(160)8 = 640
g-mm
(4-3)
The process law indicates that, with negligible losses during activation, 640 g-mm of strain energy have been gained by the spring. Figure 4-9 shows the loading or activation plot of the linear spring. Because the elastic limit of the spring has not been exceeded, and if the spring was formed from a hard-metal alloy such as stainless steel, no losses of energy will occur over time as an activated configuration of the spring is maintained. As a result, the unloading or deactivation plot will trace the loading curve, indicating that all of the added strain energy of activation is available energy. A second process may now be analyzed in
which the spring is an energy source in the surroundings and under study is the system on which the spring will do work (e.g., a tooth). The
work capacity of the spring in this second process, during which the spring is deactivated to some extent, equals the 640 g-mm of strain energy added to it in the activation process.
127 Energy Analyses in Orthodontics
200
(0
E .2?
160
F1)
0) C-)
0
LI-
-o
0)
a-
a
120
80
40
0
0
2
4
6
8
10
Compressive Deformation (millimeters)
FIGURE 4-9. The force-deformation plot for the linear spring of Figure 4-8.
The analyses of the activation and deactivation processes for an "elastic" are similar to those just discussed for the spring, but notable differences exist in the response of the elastic. Figure 4-10 shows the
activation plot for an elastic band or module, fixed at one end and stretched by an external activating force at the other end. As with the spring, the process laws indicate the equality of magnitudes of work done by the external and internal forces and the increase in internal
P1)
0 0
as
1.0
1
a (inches)
Passive state
FIGURE 4-10. The activation ot an elastic module and the associated torce-elongation diagram.
As noted in Chapter 3, because of the generally amorphous nature of elastic materials, the initial portion of the force-deformation plot is often nonlinear. To simplify the computation of the area under the loading curve, a straight line is sketched that enables equal, partial areas to be cut above and below the actual plot. The obtained triangular area is the mechanical work of the activating, external force,
energy.
(18)1 =
9
oz-in.
(4-4)
and is also the strain energy added to the elastic in the short-term activation process that stretches it, in this example, 1 in. Although springs and activated arch wires may be expected to deactivate partially, if not totally, during between-appointments periods, the activated elastic module often maintains much of its initial stretch for a
substantial length of time. Because of the relatively weak, secondary bonding within certain amorphous materials, when maintained in a stretched configuration many of these materials lose tension over time. The force-versus-time plot for the example polymeric element under constant strain conditions is illustrated in Figure. 4-11. The phenomenon exhibited, defined in Chapter 3, is "relaxation"—loss of force with time and without deactivation. An energy-analysis schematic of this process is provided in Figure 4-12. Like the spring, the activation of the elastic module provides it with available energy to do work on another system; however, although the activating process should not take the material of the module beyond its elastic limit, the amount of available energy may decrease with time due
129
Energy Analyses in Orthodontics
24
U,
0 C
=
0 a)
16
0 0
LL
8
0
0
12
6
18
24
30
36
Time (hours)
FIGURE 4-1 1. A relaxation plot characteristic of some polymeric materials.
to relaxation. In the relatively long-term deactivation process, it is gener-
ally impossible to quantitatively separate the energy transferred from the elastic module into external-work and loss portions. To illustrate the effect, however, consider maintaining (constant extensional strain) the activation depicted graphically in Figure 4-10 for several hours. Figure 4-13 shows the force-deformation relationship upon subsequent total unloading of the module. Note that this plot is also nonlinear and the force at the initiation of unloading is somewhat reduced, due to relaxation, from the level immediately following the activation process. Again,
the straight line is sketched in order to aid in the computation of the area: (13)1
= 6.5 oz-in. = strain energy released
(4-5)
The difference between the initially induced strain-energy increment and the strain energy released are the losses of relaxation and the energy no longer available to do work. In other words, the 6.5 oz-in. was the available energy in the elastic at the time of total deactivation. In general, the magnitude of such losses depends on the material of the