The Ar Artt of Phacoemulsification
The Ar t of Phacoemulsification Editors Keiki R Mehta MD MBBS DOMS MS (Ophth) FORCE (India) DO (Ireland)
DO (London) FRSH (London) FIOS (USA) Medical Director: Mehta International Eye Institute Chief: Ophthalmic Services, Colaba Eye Hospital Chief: Surgical Services, Netra Rukshak, Rural Eye Services Wing, Mumbai, India Head: Eye Department, Breach Candy Hospital and Research Centre, Mumbai Hon Ophthalmic Consultant Surgeon: Armed Forces, India Hon Ophthalmic Consultant to the Governor of Maharashtra Past President: All India Ophthalmological Society, Intraocular Implant Society, India and
John J Alpar MD FACS FICS PA Diplomate: American Board of Ophthalmology Diplomate: Hungarian Board of Ophthalmology Clinical Professor: Texas Tech University School of Medicine Medical Director: Panhandle Ophthalmological Research Foundation, Texas, USA Fellow: American Academy of Ophthalmology, Saint Luke Eye Institute Amarilo, Texas, USA
JAYPEE BROTHERS MEDICAL PUBLISHERS (P) LTD New Delhi
Published by Jitendar P Vij Jaypee Brothers Medical Publishers (P) Ltd B-3 EMCA House, 23/23B Ansari Road, Daryaganj Post Box 7193, New Delhi 110 002, India Phones: 3272143, 3272703, 3282021 Fax: 011-3276490 E-mail:
[email protected] Visit our web site: http://www.jpbros.20m.com
Branches • 202 Batavia Chambers, 8 Kumara Kruppa Road, Kumara Park East, Bangalore 560 001, Phones: 2285971, 2382956 Tele Fax: 2281761 E-mail:
[email protected] • 1A Indian Mirror Street, Wellington Square Calcutta 700 013, Phone: 2451926 Fax: 2456075 E-mail:
[email protected] • 282 IIIrd Floor, Khaleel Shirazi Estate, Fountain Plaza Pantheon Road, Chennai 600 008, Phone: 8262665 Fax: 8262331 E-mail:
[email protected] • 106 Amit Industrial Estate, 61 Dr SS Rao Road, Near MGM Hospital Parel, Mumbai 400 012 , Phones: 4124863, 4104532 Fax: 4160828 E-mail:
[email protected]
The Art of Phacoemulsification © 2001, Editors All rights reserved. No part of this publication should be reproduced, stored in a retrieval system, or transmitted in any form or by any means: electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of the editors and the publisher. This book has been published on good faith that the material provided by editors is original. Every effort is made to ensure accuracy of material, but the publisher, printer and editors will not be held responsible for any inadvertent error(s). In case of any dispute, all legal matters to be settled under Delhi jurisdiction only.
First Edition: 2001 ISBN 81-7179-790-3
Publishing Director: RK Yadav Typeset at JPBMP typesetting unit Printed at Gopsons Paper Ltd., Noida
Contributors
Alpar John J
MD FACS FICS PA
Garry P Condon
MD
Clinical Professor Texas Tech University School of Medicine Medical Director Panhandle Ophthalmological Research Foundation, Texas, USA Fellow American Academy of Ophthalmology Saint Luke Eye Institute Amarilo, Texas, USA
Director, Division of Glaucoma Allegheny General Hospital MCP Hahnaman University Pittsburgh, Pennsylvania, USA
Fine Howard I
Masket Samuel
MD
Clinical Assistant Professor Oregon Health Sciences University Portland, Oregon Oregon Eye Assocciates, Eugene Oregon, USA
Richard S Hoffman
MD
Oregon Eye Associates, Eugene, Oregon, USA
Fry Luther L
MD
Director and Chief Fry Associates PA/Ophthalmology 310 East Walnut, Garden City, Kansas, USA
Jonathan P Ellant
MD
Chief St Clare’s Hospital and Health Care Center Assistant Professor Mt Sinai School of Medicine New York, USA
Luis W Lu
MD FACS
Instructor University of Pittsburgh School of Medicine ELK County Eye Clinic Center for Advanced Eye Care St. Marys, Pennsylvania, USA
Louis D Nichamin
MD
Medical Director Laurel Eye Clinic Brookville, Pennsylvania, USA MD
Clinical Professor Jules Stein Eye Institute UCLA, Los Angeles, USA
Allen David E
MD FRCOphth
Consultant Ophthalmologist City Hospitals, Sunderland Sunderland Eye Infirmary Queen Alexandra Road, Sunderland, UK
Arnott Eric J
MD DO FRCS FRCOphth
Consultant Ophthalmologist Arnott Eye Centre, Trottsford Farm Headley, Nr Brandon Hamshire, UK
Packard Richard
MD FRCS FRCOphth
Ophthalmic Surgeon Prince Charles Eye Unit Windsor, UK
Durval Carvalho M
MD
Chief Cataract Department Centro Brasileiro de Cirurljia, de Olhros (CBCO); Member Conselho da Sociedade Brasileria de Cataracta, Brazil
vi
THE ART
OF
Durval Carvalho M Jr
P HACOEMULSIFICATION MD
Doctorate, Departamento de Oftalmologia da Universidade Sao Paulo (USP) Member, Departamento de Oftalmologia da Universidade Federal de Goias (UFG), Brazil
Arshinoff Steve A
MD FRCSC
Enrique Chipont
MD PhD
Ophthalmic Surgeon Instituto Oftalmologico de Alicante IOIS General Secretary, AVDA–DE Denia, Alicante, Spain
Agarwal Amar
MS FRCS FRCOphth (Lon)
Ophthalmic Surgeon York Finch Eye Associates 2115 Finch Avenue W, Suite 316, Toronto, Ontario, Canada
Consultant Ophthalmic Surgeon Dr Agarwal’s Eye Institute 13 Cathedral Road, Chennai, India
Davis Peter L
Consultant Ophthalmic Surgeon Dr Agarwal’s Eye Institute 13 Cathedral Road, Chennai, India
MD, FRCS (C)
Senior Ophthalmologist North Okanagan Health District Vernon BC V1T 2M9, Canada
Aron-Rosa Danielle
MD
Ophthalmic Surgeon 28 Ave, Raphael, Paris, France
Joseph Leon A
MD
Consulting Surgeon Polyclinique Comiti, Dept of Ophthalmology 20000, Ajaccio, France
Claude S Leon
MD
MD
Professor, Dipartimento Di Discipline Chirurgiche, Via Vettoio, Blocco 117A, 6710 Coppito (AQ), Italy
Leopoldo Spadea
MD
University of L’AQuila S Salvatore Hospital, L’AQuila, Italy
Luigi Mosca
MD
University of L’AQuila S Salvatore Hospital, L’AQuila, Italy
Oshika Tetsuro
MD
Ophthalmic Surgeon University of Tokyo School of Medicine 7-3-1 Hongo, Bunkyo-ku, Tokyo, Japan
Alio Jorge L
MD (Path) FRSH (Lon) DO
Agarwal Jaiveer
MD
Director and Chief Dr Agarwal’s Eye Institute 13 Cathedral Road, Chennai, India
Agarwal Sunita
MD FSVH (WG) FRSH (Lon) DO
Director and Chief Dr Agarwal’s Eye Institute 15 Eagle Street, Bangalore, India
Agarwal T
MD
Consultant Ophthalmic Surgeon Dr Agarwal’s Eye Institute 13 Cathedral Road, Chennai, India
France
Emilio Balestrazzi
Agarwal Athiya
MD PhD
Professor and Chairman of Ophthalmology, Medical Director Instituto Oftalmologico de Alicante IOIS General Secretary, AVDA–DE Denia, Alicante, Spain
Col Akhil Bharadwaj
MD
Chief Eye Surgeon, Armed Forces, Asvini, Colaba Mumbai, India
Dada Vijay K
MD DOMS
Chief and Professor Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Namrata Sharma
MD
Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Tanuj Dada
Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Kapoor Shashi
MD
Consulting Ophthalmic Surgeon Kapoor Eye Clinic, 409 Om Chambers Kemps Corner Mumbai, India
C ONTRIBUTORS Kelkar Shrikant
MD
Murthy KR
MD
Director and Chief National Institute of Ophthalmology 1187/30, Ghole Road, Shivaji Nagar Pune, India
Consultant Ophthalmic Surgeon Prabha Eye Clinic 504, 40th Cross, 8th Block Jayanagar, Bangalore, India
Lahane Tatyarao P
Sachdev Mahipal S
MD
Professor and Head Grant Medical College JJ Group of Hospitals Mumbai, India
Maniar Ranjit H
MD
Head, Shushrusha Hospital, Mumbai Honorary Ophthalmic Consultant Jankikund Hospital, Chitrakoot
Mehta Cyres K
MD
Consultant Ophthalmologist Mehta International Eye Institute Colaba Eye Hospital Mumbai, India
Mehta Keiki R
MD
Medical Director Mehta International Eye Institute Chief, Ophthalmic Services Colaba Eye Hospital Chief, Surgical Services Netra Rukshak: Rural Eye Services Wing Head, Eye Department Breach Candy Hospital and Research Centre Sea Side, 147 Shahid Bhagat Singh Road, Mumbai, India
Mody Kirit K
MD FRCS
Medical Director Salil Eye Clinic & Contact Lens Centre, 506 Om Chambers, 123 August Kranti Marg, Kemps Corner, Mumbai, Hon Professor Grant Medical College Hon Eye Surgeon GT General Hospital Conwest Jain Clinic Hospital Smt. Lilavati Hospital Mumbai, India
Murthy Gowri J
MD
Consultant Ophthalmic Surgeon Prabha Eye Clinic 504, 40th Cross, 8th Block Jayanagar, Bangalore, India
vii
MD
Cornea Fellowship (USA), Phacoemulsification, Excimer Laser, Cornea and Contact Lens Specialist New Delhi Centre for Sight A-23 Green Park, Aurobindo Marg New Delhi, India
Pradeep Venkatesh Pool Officer Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Shroff Noshir M
MD
Medical Director Shroff Eye Centre A-9 Kailash Colony New Delhi, India
Ranjan Dutta
MD
Shroff Eye Centre A-9, Kailash Colony New Delhi, India
Gurpreet Singh
DO
Shroff Eye Centre A-9, Kailash Colony New Delhi, India
Vajpayee Rasik B
MD
Professor Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Vishal Gupta
MD
Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences New Delhi, India
Inderjit Singh Professor and Head Coffs Harbour Hospital 69 Albany Street, Coffs Harbour New South Wales 2450, Australia
Preface
Advances in the field of Cataract surgery and intraocular implantation over the last 50 years have been astonishing. Phacoemulsification, had a slow beginning, but in the last 5 years has exploded forwards. Improvements in technique are increasing at a rapid pace, as the advantages of small incision cataract surgery, the instant patient rehabilitation physical and visual are obvious. Nevertheless, the ultimate expression of minimal patient inconvenience and minimal delay in resumption of patient’s lifestyle is the legacy of Phacoemulsification It was in the winter of 1989 when one of us (KRM) performed our first cataract operation utilizing Phacoemulsification. From that moment onwards, we had no doubts that this was a winning combination and in future all cataract operations would be performed with this technique and with this technique only. The procedure has changed since then. It has evolved and improved significantly. The technique has been perfected. The technology has progressed. The quality of the surgery is now virtually unsurpassable, and most importantly, surgeons all over the world now trust and rely on the Phacoemulsification technique. Old concepts change, giving way to new ideas, as fresh advances in all fields of science and medicine forge ahead. Phacoemulsification surgery and its complications are no exception. We have tried to be highly selective in modifying old concepts and including not only those changes that have widespread acceptance but also have included newer developments which will make their mark in the new millennium. The visual needs of patients are dictated by their circumstances, which include age, occupation, leisure interests, and their independence. As an adviser to a patient, the ophthalmic surgeon must consider the individual requirements of the patient and balance these against the potential risks of surgical treatment. As a surgeon, his or her attitude is tempered by his or her experience, knowledge of the experiences of others and confidence in his or her own ability to achieve perfection of results that should shine as a beacon of excellence in the community. The primary intent of the book, The Art of Phacoemulsification is to provide an introduction to the subject of Phacoemulsification as well as the framework on which could be constructed the study of that discipline. Presentation of the material begins with the most fundamental aspects and builds up successive levels of knowledge.
x
THE ART
OF
PHACOEMULSIFICATION
The purpose of this volume is to consider, in-depth, all aspects of Phacoemulsification. It is hoped that it will provide the ophthalmic surgeon who intends to commence Phacoemulsification surgery with the information necessary for effective and safe participation. For those who are already in this exciting field, this book should provide a useful source of reference. It is designed to consider both the problem and its solution. The solution has many variations but all should take into account the vulnerable tissues that are required to be protected during surgery. Every step of Phacoemulsification procedure is critical in determining the final surgical outcome. The surgery is a sequence of steps, each fundamentally important to the entire procedure. Thus while this treatise has designed to allow the readers to refine and enhance their surgical technique, it should reveal the latest advances in both the science and art of the modern technique of Phacoemulsification cataract surgery. The Art of Phacoemulsification has been conceptualized by some of the foremost cataract surgeons in the world. These contributing authors share their preferred techniques and ideas presenting the most advanced methods of surgical procedures, the newest equipment available, and their methods of managing the cataract patient. It will undoubtedly be apparent to many readers that some subjects have received a greater emphasis than others. These represent to some extent, our special interests and experiences. The major task of keeping abreast of the dynamic changes in Phacoemulsification surgery is nearly nondescribable. No single surgeon can speak authoritatively about every subject. Nor can every aspect of the subject be covered. This book covers a fair set of surgical methods and complications. Each chapter thus is an insight into the technique suitable for each surgeon and most importantly; the technique, which best suits that particular case. Each surgeon provides a detailed description directly from his or her experience of the more important Phacoemulsification techniques and his or her reaction to each new development. All contributors are aware that the production of a book having such a large volume of information, takes many months. Time and tide wait for none and certainly; the fast changing field of Phacoemulsification has new developments literally every day. The subject is developing rapidly, but it is our hope that the wealth of experience incorporated here will be valuable not only in the present time, but also for the future. This book is a multi-authored text and is, as such, abounds in differing literary styles. The editors have sought to provide a level of organizational conformity and scientific balance without sacrificing the originality and style of the individual authors. A distinctive feature of this book is the diversity of its many outstanding contributors. Probably the outstanding characteristic of this book is the editors’ ability to select critically important contributions to the management of cataract problems. These are analyzed and evaluated based on their vast experiences as ophthalmic surgeons and consultants.
P REFACE
xi
We would like to thank the contributing authors who are by their own right, considered the outstanding leaders in the field of Cataract surgery. This book would be non-existent without the devoted and tireless efforts of these physicians who contributed to this text while maintaining a rigorous daily practice. With promising laser techniques on the horizon, non-invasive Cataract surgery may soon be feasible. Many of these exciting developments are arriving so rapidly that even senior consultants in the field are often hard-pressed to keep up with the latest advances. What the future holds for us is difficult to portend, but one thing is certain, it will be exciting. Finally, the editors wish to take the liberty to express their deep appreciation to Jaypee Brothers Medical Publishers, India’s premier medical publication company, for their unflagging encouragement and tireless assistance in the production of The Art of Phacoemulsification. Mr Jitender P Vij, the Managing Director of Jaypee is truly a gem of a man, not only for his great faith that I would, one day, finally finish this epitome, but also for the general all-round editorial assistance given, the well laidout pages and the crisp photographs which have made it, truly, a world class book.
Keiki R Mehta John J Alpar
Contents
1. Commencing Phacoemulsification: The Basics .................................................. 1 Keiki R Mehta, Ranjit H Maniar
2. The Phacomachine ................................................................................................... 15 Mahipal S Sachdev, Pradeep Venkatesh
3. New Phacomachines Offer More Control ......................................................... 32 David E Allen
4. Cavitating Microbubbles Create Shock Waves that Emulsify Cataract ...................................................................................................... 45 Peter L Davis
5. Local Anesthesia ....................................................................................................... 51 KR Murthy
6. Ocular Anesthesia for Small-Incision Cataract Surgery ........................................................................................................ 58 Samuel Masket
7. The Limbal Incision ................................................................................................ 64 Shashi Kapoor
8. No Anesthesia Cataract Surgery .......................................................................... 76 Amar Agarwal, Athiya Agarwal, Sunita Agarwal
9. Clear Corneal Cataract Surgery ............................................................................ 86 Keiki R Mehta, Cyres K Mehta
10. Capsulorrhexis: A Beginner’s Guide .................................................................. 94 Shashi Kapoor
11. Capsulorrhexis: Principles and Advanced Techniques ................................103 Shrikant Kelkar
12. Hydrodissection and Hydrodelineation ........................................................... 112 Keiki R Mehta, Cyres K Mehta
xiv
T HE ART
OF
PHACOEMULSIFICATION
13. Phacoemulsification: The Quadrantic Cracking, Chopping and Stuffing Technique ................................................................................................. 118 Noshir M Shroff, Ranjan Dutta, Gurpreet Singh
14. Current Phacoemulsification Techniques ......................................................... 130 Richard Packard
15. Phaco Slice and Separate ..................................................................................... 154 Steve A Arshinoff
16. Cataract Extraction and Lens Implantation: The Implosion Technique .................................................................................... 161 Eric J Arnott
17. Phacoemulsification in Special Situations ...................................................... 170 Rasik B Vajpayee, Tanuj Dada
18. Zen in the Art of Phaco ...................................................................................... 177 Jonathan P Ellant
19. My Personal Technique of Vertical “Hubbing” Phacoemulsification ................................................................................................ 187 Keiki R Mehta
20. Innovative Nucleotomy Techniques .................................................................. 204 Vijay K Dada, Namrata Sharma, Tanuj Dada
21. Phacoemulsification in White Cataracts .......................................................... 214 Rasik B Vajpayee, Tanuj Dada, Vishal Gupta
22. Phacoemulsification in Difficult Cases ............................................................ 223 Inderjit Singh
23. Irrigation and Aspiration following Phacoemulsification .......................... 246 Keiki R Mehta, Cyres K Mehta
24. Foldable Intraocular Implants ............................................................................ 253 Vijay K Dada, Namrata Sharma, Tanuj Dada
25. History of Lens Implantation ............................................................................. 263 Eric J Arnott
26. Implantation Techniques of Acrylic Foldable Intraocular Lens and its Clinical Results ........................................................................................ 268 Tetsuro Oshika
27. The Mini-loop Plate and Accommodating Lenses ....................................... 288 J Stuart Cumming
28. Suprahard Cataracts: Their Evaluation and Management ......................... 299 Keiki R Mehta, Kirit K Mody
C ONTENTS
xv
29. Stretch Pupilloplasty for Small Pupil Management in Cataract Surgery ...................................................................................................... 314 Luther L Fry
30. Management of Glaucoma in Cataract Patients ...........................................322 Gowri J Murthy, KR Murthy
31. Phacoemulsification in the Previously Filtered Eye .................................... 329 Garry P Condon, Luis W Lu
32. Phacoemulsification in Patients with Significant Astigmatism ................ 344 Luis W Lu, Louis D Nichamin
33. Cataracts in Patients with Uveitis .....................................................................353 Enrique Chipont, Jorge L Alio
34. Corneal Endothelium and its Protection in Phacoemulsification ................................................................................................ 365 Keiki R Mehta, Cyres K Mehta
35. Phacoemulsification in the Presence of Pseudoexfoliation: Challenges and Options .......................................................................................381 I Howard Fine, Richard S Hoffman
36. Phacoemulsification in Severe Chronic Obstructive Pulmonary Disease ................................................................................................. 388 I Howard Fine, Richard S Hoffman
37. The Prevention of Complications and their Management in Phacoemulsification ................................................................................................ 393 Keiki R Mehta
38. Management of Posterior Chamber IOL Capture ........................................ 415 Durval M Carvalho, Durval M Carvalho Jr
39. IOL Scleral Fixation in Aphakic Eyes .............................................................422 Durval M Carvalho, Durval M Carvalho Jr
40. Phakonit and Laser Phakonit ............................................................................. 446 Amar Agarwal, Athiya Agarwal, Sunita Agarwal
41. Pharmacology of Intraocular Solutions and Drugs used in Phacoemulsification ................................................................................................ 453 Keiki R Mehta, TP Lahane
42. Triple Procedure with Phocoemulsification before Trephination ............ 462 Emilio Balestrazzi, Leopoldo Spadea, Luigi Mosca
43. Multiport Phaco Tip: A Safer and More Effective Training Device for Phacoemulsification ........................................................ 471 Keiki R Mehta
xvi
THE ART
OF
PHACOEMULSIFICATION
44. Intraocular Lenses Dislocated into the Vitreous .......................................... 479 John J Alpar
45. Favit: A New Technique to Manage Dropped Nuclei ................................ 486 Amar Agarwal
46. Laser Phaco Cataract Surgery ............................................................................. 493 Sunita Agarwal, J Agarwal, T Agarwal
47. Endoscopy-Assisted Phacoemulsification ......................................................... 500 Claude S Leon, Joseph A Leon, Danielle Aron-Rosa
48. Phacoemulsification: The Eye Camp Way ...................................................... 507 Keiki R Mehta, Kirit K Mody, Ranjit H Maniar, Cyres K Mehta, Akhil Bharadwaj
Index ............................................................................................................................ 529
The Ar Artt of Phacoemulsification The primary intent of this book is to provide an introduction to the subject of Phacoemulsification. Presentation of the material begins with the most fundamental aspects and builds up successive levels of knowledge. Every step of Phacoemulsification procedure is critical in determining the final surgical outcome. The surgery is a sequence of steps, each fundamentally important to the entire procedure. This book has been prepared by some of the foremost cataract surgeons in the world. These contributing authors share their preferred techniques and ideas presenting the most advanced methods of surgical procedures, the newest equipment available, and their methods of managing the cataract patient. This is a uniquely up-to-date book which covers scientific principles and current clinical and research trends, with practical information on patient assessment, variable surgical techniques, clinical results and the identification, avoidance and the management of complications. Keiki R Mehta, one of India’s foremost cataract surgeons, commenced intraocular implants in India in 1972, and developed in 1975, for the first time in the world, a soft HEMA Intraocular implant. He initiated Phacoemulsification for cataract surgery in 1979 and is known as the “Father of Phacoemulsification in India“. An exceptional surgeon, he has operated live in workshops, teaching phaco all over India. Dr Mehta is a popular teacher in great demand at conferences and symposia and has written many scientific books and published scientific papers, both at national and international level. He has been awarded eight gold medals and innumerable oration awards from national / international bodies. Extremely innovative, he has been on the cutting edge of technology and has devised multiple new techniques and instruments. John J Alpar is an exceptionally skilled surgeon , edited an extremely popular book, which literally became in the 1990’s, a bible for implant surgeons, termed ” Fechner’s Intraocular Lenses”,. With Professor Fechner of Hanover, Germany, Dr Alpar produced a book, impressive by its well-organized structure, its wealth of information, its practicality and the superb technique of presentation. A prodigious speaker and author, Dr Alpar has delivered 178 lectures, published 158 scientific articles , 11 chapters, 4 books and attended over 320 meetings. Widely traveled, he is a life member of the All India, Mexican and Hungarian Ophthalmological Societies, the Indian and Canadian Implant Societies, and is member of the Medical Societies in Peru, Japan, France. Dr John Alpar is respected worldwide for the quality of his work, the scope of his knowledge, his sharp intellect and brilliance and his keen, incisive analysis of facts.
Contributors include • Alpar John J • Fine Howard I • Fry Luther L • Jonathan P Ellant • Luis Lu W • Masket Samuel • Allen David E • Arnott Eric J • Richard Packard • Durval Carvalho M • Durval Carvalho M Jr • Arshinoff Steve A • Davis Peter L • AronRosa Danielle • Joseph Leon A • Emilio Balestrazzi • Oshika Tetsuro • Alio Jorge L • Enrique Chipont • Agarwal Amar • Agarwal Athiya • Agarwal Jaiveer • Agarwal Sunita • Agarwal T • Col Akhil Bharadwaj • Dada Vijay K • Kapoor Shashi • Kelkar Shrikant • Lahane Tatyarao P • Maniar Ranjit H
ISBN 81-7179-790-3
Rs.
JAYPEE BROTHERS MEDICAL PUBLISHERS (P) LTD B-3 EMCA House, 23/23B Ansari Road, Daryaganj Post Box 7193, New Delhi 110 002, India
Keiki R Mehta Ranjit H Maniar
Commencing Phacoemulsification: The Basics
1
INTRODUCTION Phacoemulsification is a superb technique, but to be able to conduct it comfortably and effectively all the appliances, solutions and even the personnel in the theatre need to be properly located and trained to respond to any situation which may arise. It is imperative that the surgeon understands the basic applications of each of the different requirements. For maximum efficiency and safety, proper location and arrangements, right from the operating table, the microscope, the surgeon’s chair, the instruments on the trolley, the placement of the staff nurse, the assistants, and even that of the wardboys, have to be carefully planned and worked out in advance. Phacoemulsification is an instrument-based surgery. It is also a high-pressure surgery, with periods of calm alternating with high tension. Moreover, as such there must be adequate space in the theatre so that effective and rapid movement when and if required can take place smoothly. Too compact a theatre is a sure prescription for disaster. THE OPERATING ROOM: REQUIREMENTS AND NECESSITIES The operating room—requirements and necessities of primary importance is a properly designed and functioning operating room or theatre (Fig. 1.1). There are many important points, which need to be considered. Some of them will be felt unnecessary, but the strength of any chain is dependent on the weakest link. Thus every link has to be given its importance and considered on its own merits.
2
THE A RT
OF
PHACOEMULSIFICATION
Lighting Lighting often does not get the importance it deserves. It is important that the theatre can be lighted up properly, but equally important can also be darkened adequately. It should be lighted sufficiently with shadowless theatre lamps of adequate power (minimum 50,000 Lux lamps). Though most of the surgery is done under the operating microscope, there are times when good peripherally focused lights are an advantage as for squint or oculoplasty surgery. Additional, movable goose-neck direct or fiberoptic side theatre lamps are a necessity with good illumination and are especially good for capsulorrhexis in hard opaque cataracts. In addition, in case the theatre is darkened, provision should be made for spot lighting of the instrument trolley, the phaco unit, life-support systems, anesthetic equipment, with a lighting device on a rheostat so that the intensity is adequate for the scrub and circulating nurse and the anesthetist to see the equipment clearly, but is not so bright as to blind the surgeon. Many surgeons need a spot of light focusing on the operating table in addition to the operating microscope lights. General illumination of the theater is also an important requirement. Though the powerful focused lights may be shut off, gentle illumination is needed in a theatre to allow for movement. Tube lights, ideally should never be used in the theatre as they are very distracting especially when the flicker increases as the tube gets a little older. In addition the flicker fusion frequency of the operating staff tend to be affected by the tube light especially when they are tired after a long session, increasing surgeon and staff irritability Darkening the theatre enhances the contrast under the operating microscope and is extremely useful when doing a capsulorrhexis especially in a hard cataract. When the main theatre is darkened, provision needs to be made to gently illuminate the floor so that personnel can still move around without tripping on objects. It is important that spot lighting should be kept off places where high reflectance stainless steel appliances and instruments are placed. This is to prevent extraneous glare from reflection from these shiny surfaces. The colors of the clothing worn by the operating room personnel should be soft and muted avoiding harsh colors. Pastel shades of blue, green or yellow are quite acceptable, however; red and ocher should be avoided Air-conditioning and Ventilation The ventilation of the room should be adequate and the air-conditioning sufficient to compensate for the number of personnel who are going to be in the room. The air-conditioning vents should be so arranged that they do not allow the air to be blown over the operating surfaces and at the same time keep the theatre cool. The ideal operating temperature would vary from surgeon to surgeon, however, a good comfortable temperature level in India is 70o C. The ideal air-conditioning would be one-way, taking in air from outside, filtering it, cooling it, and then expelling it out again after circulating through the operation theatre. Most theatres in India and in most of the smaller hospitals and nursing homes would seem to have window- mounted
COMMENCING PHACOEMULSIFICATION : T HE B ASICS
3
air-conditioning. The size and number should be adequate to provide good cooling with the air intake for fresh air always remaining open. It is very important that every evening following surgery the filters of the air-conditioner should be washed and soaked in a dilute solution of Cetavalon for half an hour prior to being re installed in the air-conditioner. The position of the units should be such that they do not blow over the sterile field, or blow directly onto the operating staff. Noise Level in the Theatre An operating theatre should be an oasis of calm. It is therefore important that the operating room should be located in a quiet area of the hospital or facility, and away from distracting sounds. It is essential that the windows be double-glazed (twin sheets of glass with an air-space between them) to keep the noise level down to a minimal level. It should be imperative that the operating staff learn from the beginning that unnecessary talk be kept to a minimum level and communications as far as possible should be by hand signs. This would guarantee that the surgeon, and the staff enjoy adequate peace and quiet to be able to concentrate on doing a good job. Background soft music should always be played in the operating theatre as it defuses tensions. The music should neither have a harsh beat nor irregular cadences. Electrical Power and Outlets The power points in the operating theatre should use high-quality reputable switches of an adequate output so that they are not overloaded and at the same time, good contacts are obtained between the plugs and the sockets. Often, even in so-called, Five Star facilities, it is seen that from a single outlet, using multipoint extensions, a number of lines are drawn. The wires then are left carelessly on the floor. All power points should be far away from the surgical field and preferably from a central hanging pod so that one cannot accidentally trip over the wires. In case it is required to trail a wire on the floor, it must be well protected with a masking tape so it is not accidentally pulled out in the dark. The power outlets should be rated at a sufficient level to comfortably run the medical equipment. Over loading of the points often leads to failure at a critical time during surgery. It is also imperative that fuses be provided for every media outlet in the theatre of the self-adjusting type which could be reset simply by pushing in a button rather than the older wire-looped fuses. Sensitive instruments like a phacoemulsification machine and life-support evaluation systems (cardiac monitor or oxygen saturation monitors should always be run through an on-line UPS (uninterrupted power supply). This permits the surgery to be completed even if the lights go off or the power supply fluctuates or even trips (Fig. 1.2). Power Generators In India, as with many developing countries, power outage is not uncommon. It is important when the theatre is planned that one should compensate for this problem.
4
THE A RT
OF
PHACOEMULSIFICATION
Though it would be nice to have automatic switch-over power systems where the load is taken temporarily on batteries and then automatically shifted to the generated supply, it is a very costly system which is rarely used. Instead small power generators are utilized, adequate to run the theatre lamps, general lighting and power the instruments, including the Phacoemulsifier and support systems. It is important that the wiring be so organized that all that needs to be done, at the time of a power failure, is to turn on the switch and start the generator. The load on the generator should never exceed 75 percent of its rated output to prevent overload and tripping. The generators, which are usually run off petrol, kerosene or diesel, all have a few common features. They are all noisy, smelly and temperamental. Hence they need to be placed in a room with good ventilation, and isolated so that the sound and smell does not reach in the hospital or theatre complex. They should be serviced regularly, and personnel trained to start and run the units. Scrubbing Facilities The scrubbing room should be separate and kept outside the theatre. There is a specific reason for this. When gloves are worn there is always glove powder scattered around which is then be circulated in the room leading to contamination. Not only does this choke up the filters of the air-conditioner, but leave a patina of dust all over the sterile surfaces of the room. Personnel in the Theatre The ideal theatre room composition should be a scrub nurse, who surgically assists the surgeon, and a circulating nurse, who remains unsterile. In my theatre, where I like to have a turnover of around 12 to 15 cases per day, preferably in a threehour period, I find it best to use two separate teams. The scrub nurse who surgically assists me in the surgery will, after the case is finished, wash the instruments, place them into the sterilizing box and then put it herself into the autoclave. The scrub nurse then washes up again, dons a fresh gown and gloves and commences preparing for the next case preparing the table and opening up the disposables which are handed to her by the circulating nurse. The unsterile circulating nurse will open the presterile disposables, remove the instruments from the sterilizer, and hand them across to the scrub nurse. The second scrub nurse, who has been assisting me with the second cases, finishes, moves out, and the totally prepared first scrub nurse is ready to commence the next case. This technique has a big advantage that the scrub nurse knows all about the instruments, where they are placed and their functional status. In addition it makes for far faster and more efficient application. Theatre Autoclave The autoclave should be a rapid action unit with flash sterilizing ability. A number of sterilizing systems are now available. Statim is a common one in usage (I use
COMMENCING PHACOEMULSIFICATION : T HE B ASICS
5
the Statim Cassette Autoclave, which has an 8-minute cycle, just perfect) and has preset operative timing levels, has adequate safety fail-safe built-in, and even has a small printout which confirms that the autoclaving cycle was complete and effective. The cassette system makes it very simple to insert and remove the instruments. Another good system is the Totawer and the Korean system which work in essentially identical manner. It is important to have a proper place to store the autoclaved instruments and theatre linen. The corridors and wall nooks are not for this purpose. It has to be in a well-ventilated room, far away from any traffic so that sterility is not compromised. The Operating Table The operating table has, as its primary requirement rock-solid stability even under deflecting forces, like inadvertent pressure at the head end, or accidental tail end pressure or lateral pressure. The standard operating table with the rotating axis in the middle is not suitable for ophthalmic surgery as the slightest pressure at either the head end of the table or the tail end of the table causes the entire platform to rock. When an operating microscope is being used, zero movement is permissible with any level of safety and efficiency. A very steady table is mandatory. Some of the operating tables are exceptional, like the Marquette system which is, however extremely costly. Alternately, more economical systems like the EyeTech table seem to work equally well and are sufficiently rigid for ophthalmic use. The table should be motorized permitting free movement both up and down in fine increments so that it could be fine tuned with the surgeon and the microscope in position. There should be the ability to tilt the head end of the patient a little up or down to compensate for those with a larger anterior/posterior diameter of the head, or when little children are being operated. It should wide enough to accommodate the patient, but narrow at the head end so that it does not impede the surgeon especially when temporal; surgery is being undertaken. It is also important that there should be adequate place under the table for the surgeon’s feet, the foot pedal console of the phacoemulsification unit, as well as for the foot pedal console of the operating microscope. Foot-mounted electrical controllers for an operating table should be avoided as the irrigating fluid, be it normal saline, Ringer lactate, or balanced salt solution (BSS) is bound to splash on the floor leading to a short circuit. The operating table should also have the ability to take a right sided arm rest where the arm can be positioned by the anesthesiologist for placing an IV cannula for any intravenous injection or sedation as may be needed the need for inserting a very uncomfortable arm support under the back of the patient (Fig.1.3). The mattress of the operating table should be at least 3 inches thick. The primitive 1 inch hard, unyielding, uncomfortable theatre mattress should be dispensed off with. Unlike general and orthopedic surgery where the patient is deeply sedated or even unconscious, the average ophthalmic patient is wide awake. With the advent of topical anesthesia, with the patient having to lie, totally without moving for
6
THE A RT
OF
PHACOEMULSIFICATION
long minutes at an end, the minimum which could be expected is a comfortable rubber mattress. The discerning and concerned surgeon should try sleeping on his or her own operating table to see its comfort level. There has always been discussion as to whether wrist support is required. Proponents of the wrist support system feel that it helps in stabilizing the wrist, and at the same time permits a little cavity or gully for collection of fluids rather than letting the fluid rundown the face. On the other hand, there are others who feel that it restricts the freedom of movement of the hand around the face and since usually the forehead is already being used to support the fingers the presence of a wrist rest is superfluous. It is basically a surgeon’s choice. I personally feel it interferes more with the surgery than helps, and though I have used it in many operating theatres, have never felt that it was really necessary. Personally I feel it restricts the free movement of the phaco handpiece. However, it is an individual choice. The Surgical Chair Phacoemulsification requires both hands and both feet to be utilized simultaneously. It stands to reason that the surgical chair is an important piece of surgical equipment. It gives the surgeon stability, supports his back, it gives a comfortable seating arrangement. It is important to remember that the feet have to be kept on the pedals for the full time of the surgery, and the surgical chair must be so designed that it prevents any pressure on his thighs. It is imperative that the chair be very comfortable, for the surgeon will need to sit on it for long hours every day if he is to complete his surgical list. Any discomfort, overtime, tends to get magnified, which affects, in the final score, the surgical competence. In the intracapsular days, most surgeons operated without any magnifying aids except for low-powered spectacle magnifiers or head-worn loupes, and operated standing. The advent of extracapsular cataract surgery changed the entire gambit. The necessity of visualizing the red glow meant the use of a coaxial operating microscope became mandatory. With the use of automated irrigation/aspiration units, both feet needed to be utilized. Thus the surgeon had no option except to sit and operate. Ideally the operating chair should have a minimum of five and preferably seven smoothly moving, nylon castors, to give total stability, with a lock on at least two of them, to immobilize the chair. The arm rest should be of adjustable height and properly padded and designed with a slight hollow so that during surgery the resting elbows should not slip off (with, as one may well expect, dire consequences). They should support the elbow, but at the same time, should neither restrict, nor interfere with, the surgeon’s movement. The chair should also fit easily under the operating table, with adequate space for the surgeon’s thighs. The area below the head rest should not be in contact with the chair or its armrests, neither should the base touch against the operating table. The height of the chair could be either electrically or hydraulically adjusted so that the appropriate height for each individual patient and the surgeon can be utilized. Finally, the chair must be grounded.
COMMENCING PHACOEMULSIFICATION : T HE B ASICS
7
Operating Microscope The microscope is perhaps the most important single piece of equipment in the theatre. Without an exceptional microscope, good phacoemulsification is difficult, if not impossible. The basic requirements are as follows: Excellent optics with clear vision at the edge of the optics There should be no blooming or distortion of the image and the lenses need to be color corrected. The latest microscopes (Zeiss) have apochromatic optics. Adequate depth of focus The entire lens should be visible from the front to the back without refocusing. This is very important since when doing phaco the traverse of the tip from the front to the back of the lens is almost 4 to 5 mm and it is important that excellent focus be available at all times. Perfect coaxial optics Are of prime importance if a good red glow is to be visualized. In modern extracapsular cataract extraction (ECCE) and, even more so, phacoemulsification, the surgeon literally operates against the background of the red glow. A good glow from one edge of the pupil to the other is thus a basic requirement. Good X-Y device The advantage is that the position of the microscope can be adjusted during surgery utilizing the foot controls without having to manually push a heavy microscope around. A good X-Y device also compensates for the little head movement which is to be expected during surgery. Automated zoom magnification It is not absolutely essential but is extremely useful as one can zoom in for a difficult situation ( doing rhexis in a hypermature cataract, or to see the edges of the capsule while doing posterior rhexis) and then zoom out with a reduced magnification for more effective surgery. Easily movable without damaging the unit Should be mounted on movable castors so that it could be positioned easily and locked in place in the operating theatre. Proper and stable optics delivery The arm connecting the microscope head to the supporting pillar should have adequate movement but at the same time should possess rock-solid stability. It should be possible to position the microscope head easily and then lock the arms. Tilt optics Not mandatory but makes a great deal of difference in comfort. The horizontal to vertical tilt arrangement (range of 90 degrees) is in the opinion of the author a really useful device as it makes the difference between operating comfortably and struggling and operating. It is particularly useful when operating on patients who cannot lie flat and who have to be literally operated in a 45-degree position. One can position the microscope to be parallel to the plane of the head and then simply tilt the optics to operate comfortably.
8
THE A RT
OF
PHACOEMULSIFICATION
Fig. 1.1: Layout of operation theatre with phaco on right
Fig. 1.2: Showing set-up with video, VCR, cardiac monitor, oxygen saturation monitor and cautery on right side of surgery
Preoperative Microscope Positioning It is imperative that the microscope be positioned accurately at the time of commencing surgery. Ideally the microscope should be on the right side of the patient, the same side at which the phacoemulsification unit is kept. The left side is reserved for
COMMENCING PHACOEMULSIFICATION : T HE B ASICS
9
Fig. 1.3: Taken from the foot end of the patient. Anesthetist on the surgeon’s right side and instrument table on the surgeon’s left side
Fig. 1.4: Showing the double tubing Surge Suppression System attached to the author ’s Alcon Legacy
permitting the patient to be shifted from the gurney or trolley to the table and the subsequent removal after surgery. The ideal place for keeping the instrument trolley is at patient’s left. The surgical assistant stands on the same side. The operating microscope camera should have its monitor placed at the surgeons right, set slightly
10
THE ART
OF
P HACOEMULSIFICATION
Fig. 1.5: The balancing balls designed by Dr Tony Fernandez for softening eye
behind the surgeon so that the scrub nurse and the anesthetist can both follow the progress of the surgery, at the same time it will not distract the surgeon. An X-Y attachment is a very useful adjunct as it allows the stabilization of the optical axis to the patient’s eye during surgery without unnecessary coarse movements of the optical head. The X-Y device should be placed at its zero position prior commencing the surgery. Most microscopes have removable autoclavable plastic, metal or silicone caps for the microscope. Alternatively, cloth covers, which can be autoclaved, can be utilized. It is important that all the arrangements and positioning of the microscope be done prior to commencing the surgery. The luminosity of the microscope should be kept at the lowest level consistent with good vision. It is important to remember that the so-called ‘cold” fibreoptic light is not really cold but simply not too hot. A good heat shield must be fitted in the microscope especially if the microscope has a filament bulb. Keep the light intensity low until required. Prior commencing, the surgeon should place the focus at one-third position, i.e. if the traverse can be visually divided into three parts, it should be fitted in the upper third. This allows the surgeon to have more than adequate range during surgery. The surgeon should commence with the microscope focused at the limbus where the initial incision will be made. Placing the setting at the upper one-third position of the head traverse, enables the available traverse ( up-down movement) of the microscope to be utilized effectively. It is also important to adjust the microscope optics to the surgeon’s ametropia and his interpupillary distance if the microscope is used in a multi-user environment.
C OMMENCING P HACOEMULSIFICATION: T HE B ASICS
11
To enable good coaxiality so as to obtain an excellent red glow the corneal plane must be exactly at right angles to the microscope tube. Be certain to position the eye perfectly prior commencing. Many microscopes come with a small round macular occluder which can be brought into position after the critical part of the surgery is over to diminish the quantum of light entering the macula. A simple alternate technique is to change the angle of the optics immediately after the cortical aspiration is over and to dim the light. Modern phaco surgery is now so fast that sometimes one wonders whether it is really required, however it is a good practice and should be followed. Footwear Use in the Theatre The use of footwear is very much dependent upon the surgeon. I personally prefer to use stocking feet rather than using slippers or shoes as I personally feels that it gives far better control. The use of thin-soled tennis shoes would perhaps work just as well. The thick-soled Nike and Adidas shoes though excellent in the sport field are not really useful as the fine control is lost. Using the X-Y control with stocking feet is a snap as the toes can easily encircle the knob. However, thin-soled shoes do seem to work well. The problem comes about in utilizing the X-Y control on the microscope. Stockinged feet are able to comfortably go around the tip of the X-Y knob permitting exquisite control. The important guidelines to observe are comfort. Be careful not to use loose floppy footwear like rubber slippers or cotton slippers, as they tend to slide over the footswitch area and can, in a critical moment, jam the footswitch and precipitate problems. The Patient in the Operating Theatre Positioning the Patient in the Theatre The patient’s head should be positioned in such a manner that the iris plane is parallel to the floor and perpendicular to the coaxial light of the microscope. In case the patient’s eyebrow is pronounced or the nose is pronounced, both of which would interfere with the surgery, one can easily shift to a full temporal approach. I personally prefer to enter at the 10 o’clock position in the right eye and the same in the left eye. The only time I change positions to a full temporal approach is when space is inadequate for a proper exposure. It is always very tempting, after scrubbing, to enter the operating room and wear the gloves from the instrument trolley, next to the patient. It is important to don the rubber gloves prior to entering the theatre, and as far away from the instruments trolley as possible. Whenever gloves are snapped on, talcum/glove powder, which is on the gloves, tends to be liberated and then falls as a fine patina all over the instruments and the eye. Prior to commencing surgery the surgeon should wipe his or her hands with a sterile dry towel, after putting on the gloves so as to remove the excess gloves powder. Washing is also acceptable but must then be in copious distilled water as otherwise it simply cakes the gloves, making matters worse. The dry towel scrub is the best to remove excess glove powder.
12
THE ART
OF
P HACOEMULSIFICATION
Preparation of the Eye The area around the eye and the eye itself are washed with 10 ml of 50 percent diluted Betadine 5 percent ( povidone-iodine) solution. A cotton bud soaked in full strength Betadine solution is swept along the lashes to make sure that they are well cleaned. Subsequently the eye is flushed out with distilled water or with Ringer lactate to remove all the impurities. There is never any need to cut lashes. Draping the Eye The area around the eye is dried thoroughly with a sterile towel. A sterile self-adhesive plastic drape, either individually, or as a part of a complete drape, must now be placed over the eye. The method of placement is fairly simple. The eyelids are kept widely open either by the surgeon’s left hand or kept open by the surgical assistant using cotton buds. The sterile drape is positioned over the opened eye, the tip of the index finger is allowed to press the drape in between the opened area, gradually letting the drape stick onto the lashes and then onto the area around the eye. Using a blunt-tipped scissors, the plastic drape is tented and then incised down the middle being careful that the cornea is not accidentally touched. A soft wire speculum or a self-retaining speculum is then inserted in such a manner that the incised drape turns over the lashes, and then passes under the lids, held in place by the speculum, isolating them from the sterile field. Another big advantage of draping is that at no time is there any accidental touch at the time of insertion of the phacoemulsification probe or the implant in the eye. Following the application of the drapes, a second cloth drape can be put over the site. It has three functions: (i) it acts as an additional sterile barrier, (ii) cuts down on reflections, and (iii) acts as an absorbent media. Use of Lid Stitches In the days of intracapsular and later extracapsular cataract surgery, the use of lid sutures or superior rectus sutures was almost a routine. In the phaco era, lid stitches are used extremely rarely and are quite unnecessary. The only time any sutures are used is a superior rectus suture placed if the surgeon requires more exposure as when he or she wishes to do a combined glaucoma and cataract procedure. By eliminating the use of a superior rectus stitch, postoperative ptosis incidence is markedly reduced, it is infinitely less traumatic, eliminates the hematomas, which occasionally accompanied the placing of the stitch, and reduces postoperative inflammation. Since more often than not, topical anesthesia is the technique of choice, the eye is kept stable enough by the patient and the use of the superior rectus stitch is thus redundant. Suction Facilities To maintain a dry field during surgery is important. It is very difficult to operate with a pool of liquid reflecting back the microscope light. Rather than repeatedly
C OMMENCING P HACOEMULSIFICATION: T HE B ASICS
13
swabbing the area dry, which leads to conjunctival irritation and interferes with the surgeon, the best is to keep a small suction available. A good option is the use of self-retaining speculum with aspiration ports attached to a small dental suction which gives suction in the range of 5 to 7 mm Hg. An alternate method is to utilize a drainage device like a sterile plastic bag which can be attached on to the side of the eye to hold the excess fluid as it drains out, or to use an absorbent wick drape which permits easy leakage. Which ever device is used, it is important to keep the floor dry. Dripping irrigating solution can be a source of great irritation to the surgeon when it falls on his or her feet, or wets his or her clothes. In addition the dripping liquids tends to cause the phaco foot switch to become slippery and may even jam in time thanks to the dried salt crusts. In addition, the saline is electrically conductive and is thus an electrical hazard. FURTHER READING 1. Mehta KR, Sathe SM, Karyekar SD: Computer Terminal Usage and Eye Fatigue, Xth Congress APAO. Soc Proc 2:946-48, 1985. 2. Mehta KR: Phacoemulsification cataract extraction with foldable IOLS-First 50 cases. All India Ophthl Soc Proc 56-60, 1989. 3. Mehta KR: Progressive corneal endothelial decompensation—extended wear contact lenses with aphakia. All India Ophthl Soc Proc 109-14, 1989. 4. Mehta KR: Endocapsular phacoemulsification and posterior chamber IOL implantation. All India Ophthl Soc Proc 217-20, 1989. 5. Mehta KR: Post-cataract astigmatism: A comparison between phacoemulsification and ECCE procedure: cataract with and without intra-ocular implantation. All India Ophthl Soc Proc 226-29, 1989. 6. Mehta KR: Posterior capsular capsulorrhexis with shallow core vitrectomy following implantation in paediatric cataracts. All India Ophthl Soc Proc 207-10, 1995. 7. Mehta KR: The loop tri suction nonphaco technique of small incision cataract surgery. All India Ophthl Soc Proc 210-12, 1995. 8. Mehta KR: The clear corneal phacoemulsification with injectable silicone lenses. All India Ophthl Soc Proc 218-22, 1995. 9. Mehta KR: An Advanced but simple keratometer for control of postoperative astigmatism. All India Ophthl Soc Proc 122-23, 1990. 10. Mehta KR: Posterior chamber implantation. All India Ophthl Soc Proc 143-44, 1990. 11. Mehta KR: YAG laser damage to intraocular implants—an evaluation. All India Ophthl Soc Proc 14750, 1990. 12. Mehta KR: Phacoemulsification—is it the true III world answer for eye camps. All India Ophthl Soc Proc 301-303. 1990. 13. Mehta KR: An analysis of causative factor leading to eye strain caused by computer monitor screens. All India Ophthl Soc Proc 334-36, 1990. 14. Mehta KR: Single stitch elliptical funnel incision for cataract surgery. All India Ophthl Soc Proc 25354, 1991. 15. Mehta KR: Bifocal intraocular implants—a functional evaluation based on 425 cases. All India Ophthl Soc Proc 271-74, 1991. 16. Mehta KR: Phacoemulsification with flexible PC IOL—is it really a step forward. All India Ophthl Soc Proc 287-88, 1991. 17. Mehta KR: The new phaco cleave technique for hard cataracts. J Intraocular Implant and Refractive Society, India 1(1): 74-75, 1996. 18. Mehta KR, Sathe SN, Karyekar SD: The new soft intraocular lens implant. Am Intraocular Implant Society J4(4):200-05, 1978. 19. Mehta KR, Sathe SN, Karyekar SD: New soft posterior chamber implant, X Congress of the AsiaPacific Academy of Ophthalmology. New Delhi,1985. 20. Mehta KR: Clear corneal phaco with injectable silicone IOL proc. All India Ophthl Soc Proc (Mumbai) 1995.
14
THE ART
OF
P HACOEMULSIFICATION
21. Mehta KR: Phaco with flexible IOL—is it a step forward. All India Ophthl Soc Proc (Bangalore) 1991. 22. Mehta KR: The tripod posterior chamber flexible acrylic implant —the answer to better stability. APIIA Conference, 1997. 23. Mehta KR: Intralenticular “hubbing” technique for simple eye camp phacoemulsification—a simple technique. APIIA Conference, 1997. 24. Mehta KR: Newer techniques for eye camp safe phaco techniques. APIIA Conference, 1997. 25. Mehta KR: Intralenticular “hubbing” phaco technique for safe phaco. Proc of SAARC Conference, Nepal, 1994. 26. Mehta KR: The New Multiport Phaco Tip for Safer, More Effective Phacoemulsification, with Virtually Zero Capsular Damage. Proc of SAARC Conference, Nepal, 1994.
Mahipal S Sachdev
The Phacoemulsifier
2
INTRODUCTION It is very important than a thorough knowledge of the phacoemulsification machine is available to the operating surgeon. There are many machines available in the market, each with their own characteristics. However once the basics of the machine are understood, it becomes simple to analyze them, and having done so, understand how exactly they work. All machines fall into two basic categories, those utilizing a peristaltic pump and those using a Venturi pump. It is critical that every surgeon learns about the machine parameters and their individual effects, how they interrelate and in total how they affect the environment in which the surgery is performed. The Machine: Basic Features The phacoemulsification machine (Fig. 2.1) is essentially a system which generates ultrasound energy transmitted to the tip of the handpiece. The machine console only generates the electrical energy. The conversion
Fig. 2.1: The Laser Phacoemulsifier Machine (Alcon Legacy)
16
THE A RT
OF
PHACOEMULSIFICATION
of electrical energy into ultrasound is at the handpiece level. The body of the machine thus basically has controls which modulate every key requirements, be it diathermy, irrigation-aspiration control, ultrasound energy stability, or even the height of the irrigation bottle, etc. The fine-tuning is done by the foot switch which gives the surgeon more flexibility. Every phacoemulsifier has five basic functions; diathermy, irrigation, irrigationaspiration, ultrasonic fragmentation and vitrectomy. Each of these functions has a handpiece to match them. Irrigation Handpiece The irrigation handpiece is used when only irrigation is required. It is connected to an irrigating cystitome for anterior capsulotomy, or to an irrigating loop for hydrodissection. Many machines have the ability to preset controls so that when only irrigation is required, the foot switch functions purely as an on-off mechanism. Irrigation-Aspiration (I-A) Handpiece The infusion liquid is sent to the anterior chamber through the connected tubes. The basic function of the I-A handpiece is to aspirate liquid and cortical material through the aspiration port, at the same time infusing chamber-maintaining liquid into the anterior chamber. Essentially the irrigation-aspiration (I-A) handpiece, has a either single piece metal (stainless steel or titanium) irrigation-aspiration sleeve or has an aspiration sleeve with a silicone sleeve that fits snuggly around the aspirating tip. The I-A tip differs from the phaco tip in being smooth and rounded with a single aspiration port on the side of the tip and not at the end. The sleeve may be turned to orient the irrigation port in any direction. The irrigation ports in the silicone sleeve should be kept perpendicular to the metallic aspiration port as this helps direct the infusion fluid along the iris plane. This reduces iris flutter during the surgery. Typically the I-A handpiece has a rounded tip with the aspirating port at one side usually 0.75 mm to 1.5 mm away from the tip. The opening can be in a diameter of 0.2, 0.3, 0.4, or 0.5 mm. The overall diameter of the I-A handpiece usually varies from 2.5 to 3.0 mm depending on whether the aspiration sleeve is metal, or of silicone. The angulations of the I-A handpiece can be straight, 45° bent, or has a 90° bend. Most surgeons prefer to utilize the curved I-A tip. Recently Alcon in its Legacy phaco machine has taken out a tip which can be varied as desired termed a “steerable tip”. The commonly used I-A port is a 0.3 nun port. It has the safety feature that it will aspirate the cortex and not the capsule. It is however wise to keep on one’s table a 0.5 port so that at times when you wish to aspirate larger particles it is available. The larger port is also useful when doing a direct aspiration, as is often done in a congenital cataract. Irrigation/aspiration handpieces corne with metal or silicone sleeves, each having their own advantages.
T HE P HACOEMULSIFIER
17
Metal Sleeves Metal sleeves allow a more regular inflow since they neither are compressed by the incision edges nor are they compressed if the tip is moved in a tunnel obliquely when oar-locking can obstruct the flow. Having smoother edges they are easier to introduce into the phaco tunnel. They also do not snag on the edges of the iris. Naturally being metal, they last much longer (Fig. 2.2).
Fig. 2.2: Irrigation/aspiration metal sleeved handpiece curved, and 90 degrees bent
Silicone Sleeves Silicone sleeves have greater flexibility and by molding themselves to the walls of the tunnel (basically, once a tunnel is opened, it is no longer a slit but elliptical in shape) give a better fit, thus diminishing the leakage from the chamber. This is important especially if the eye pressure is a bit high, chamber is shallow, or in children (Fig. 2.3).
Fig. 2.3: Silicone-sleeved bent for irrigation/aspiration bent
The Diathermy Handpiece In diathermy handpiece is a very essential adjunct and is ideal when a blood-free field is required typically in preparation of squared or smile (chexron) scleral or semiscleral incision.
18
THE A RT
OF
PHACOEMULSIFICATION
Diathermy handpieces can be coaxial (Erasertm) or of the forceps type. The coaxial type is excellent in preparing and having a bloodless scleral area. The forceps on the other hand can also be used for sealing the edges of the conjunctiva together (coaptation) at the end of the surgery. It is essential that the minimum quantum of diathermy be utilized. In most modern machines the control of the quantum of diathermy is linear, i.e. it is controlled by depressing the foot pedal. The maximum and minimum values can be preset on the console. Anterior Vitrectomy Handpiece The unit can be either of the guillotine type or of the rotating type with a triangular tip. In the earlier days most machines had the rotating vitrectomy tip, but it was soon recognized that the moment the unit got a little older it tended to entrap and tug on the vitreous and hence the guillotine vitrector has now become a standard in most machines. For anterior vitrectomy, the tip usually comes with a perfusion sleeve which can be removed if so desired. On the console, the essential values of flow rate, cut rate and vacuum can be set to suit individual requirements. Ultrasonic Handpiece Bimanual It has become customary for many surgeons to use separate handpieces for irrigation and aspiration. This helps immensely in cortical removal (Fig. 2.4).
Fig. 2.4: Bimanual hand pieces, separate for irrigation/aspiration
Phacoemulsification of a lens nucleus depends upon ultrasonic power which is the function of the acoustic vibrator that has been incorporated into the ultrasonic handpiece. Attached to this vibrator is a hollow titanium needle or the phaco tip. The acoustic energy produced along the ultrasonic handpiece is then transmitted onto the phaco tip (Fig. 2.5).
T HE P HACOEMULSIFIER
19
Fig. 2.5: Phacoemulsification handpiece with four crystals (Alcon Legacy)
The acoustic vibrator is of two types: magnetostrictive or piezoelectric device. The acoustic vibrator converts electrical energy into mechanical energy under the influence of an electrical signal. The acoustic vibrator oscillates longitudinally at a frequency between 30,000 and 60,000 Hz. This imparts a linear motion to the ultrasonic tip. The stroke amplitude of the linear movement is 3.1000 of an inch and the acceleration 80,000 to 2,40,000 G. Magnetostrictive Handpiece Magnetostrictive handpiece was the first in use, and has now been phased out. It uses an electric current to induce a magnetic field which results in the linear movements of the ultrasonic tip. The electromagnetic field is generated by a coil of wires wrapped around the handpiece. Advantages and Disadvantages of Magnetostrictive Handpiece • • • • • • •
Can be autoclaved repeatedly with no risk to the handpiece Much sturdier. Does not break if dropped Can be repaired easily The handpiece is larger (almost the width of the base of a billiard cue) It is much heavier Needs to be water-cooled The greatest problem is that power delivery is inadequate and often at peak powers tends to be erratic, more so as the handpiece gets older.
Piezoelectric Handpiece Piezoelectric handpiece uses electric energy to reorient the piezoelectric crystal which in turn is translated into linear movement. The piezoelectric transducer requires a direct electrical contact to be made with the crystal.
THE A RT
20
OF
PHACOEMULSIFICATION
Advantages and Disadvantages of Piezoelectric Handpiece • Has a more efficient power delivery. With the use of multiple crystals the full range of delivery can be made very smooth even at very small increments. • It is air-cooled • Is very much lighter, almost featherweight as compared to the magnetostrictive handpieces • It is however very fragile and can break on being dropped • Costly to repair. Some handpieces may need to be calibrated every 1500 phaco procedures for optimal output. Phaco Tip The phaco tip can have various bevel angles ranging from 0° to 60° and comes in various shapes and sizes. The phaco tip is made of titanium and is hollow with the distal opening functioning as the aspiration port. The acoustic energy produced along the ultrasonic handpiece is then transmitted onto the phaco tip. The angle of the tips are for basically two reasons: a flat tip, like the 0° and 15° are excellent for holding but very poor for cutting; on the other hand to make a trench in a hard cataract the 60° tip is ideal, but because of its large surface area of the oblique opening, its holding power is poor. Tips may also be of various types, flared at the end (Cobra tip) or with the tip bent (Mackool tip) or with small ports, termed ABS port (Fig. 2.6). Entering into the anterior chamber is easy with Fig. 2.6: Peristaltic pump of the 60° tip and progressively harder with a 15° Legacy machine or a 0° tip. The commonly used tips are 30° and 45° phaco tips.
Alcon
Analyzing the Tips 0° Tip 15° Tip 30° Tip 45° Tip 60° Tip
Basically a flat, square cut tip with minimum cutting power but excellent holding capacity. Ideal for phaco chop techniques. Less cutting and more holding power. Suitable for improving follow ability. Balanced cutting and holding power. Suitable in most of the phaco procedures. Sharp cutting with good cutting ability and less holding power. Very sharp cutting edge with minimum holding power. Ideal for grooving hard cataracts.
T HE P HACOEMULSIFIER
21
Tuning the Phaco Tip The phaco tip is screwed into the handpiece directly using a wrench. The handpiece is then tuned so as to synchronize the mechanical movement of each tip with the handpiece. Autotuning also allows the handpiece to maintain its frequency irrespective of change in the density of the medium. A loose or a heavily used needle will not tune. It is also customary to tune every time a needle is changed. Some of the newer machine (Sovereign Allergan) can retune the needle in few seconds. The irrigation fluid is made to flow through the two side ports on the silicone sleeve. The silicone hub threads the sleeve onto the outer casing of the handpiece. In some instruments (Alcon), an internal rigid sleeve has also been designed to separate the aspiration and the irrigation fluids. This is also supposed to reduce the bubble formation that is often encountered during the phaco procedure. Phaco Power Settings There is no predetermined “correct” power. Initially the manufacturer’s recommended settings are used. With experience, each surgeon “fine-tunes” his settings. Power variables are adjusted intraoperatively depending on • Density of nucleus where phaco tip is engaged • Amount of tip engaged • Linear velocity of the tip during emulsification. When the power is inadequate, the tip will fail to cut the nucleus, and tend the push excessively on the nucleus which lead to zonular stress and can be dangerous. When the power is too much, rather than holding the nucleus it will cause the nucleus to flyaway from the ultrasound tip, termed chatter. Too much power can also accidentally pierce the nucleus, making a hole in the capsule and leading to a dropped lens, a catastrophe, best avoided. Thus setting a safe power setting prior commencing is important. A safe “standard” setting is as under. The ultrasound power is set to 50 to 70 percent. If the lens is soft, it is decreased to about 30 percent and if it is hard, the power is increased to 80 percent or 90 percent. Power is reduced if the nucleus chatters. At this stage, the linear ultrasound mode is changed to pulse mode, which tends to hold the nucleus better against the tip and by giving a break between each pulse enables the fragments to corne to the tip easier. The third-generation machines, having four crystals per handpiece, have far better fragmentation control and rarely need the power to be turned up above 70 percent. It is best to consult each individual manufacturer regarding their safe “recommended” settings and only after experience is derived on that machine gradually change the values to suit ones individual style of phaco. Ultrasound is inaudible. The buzzing audible sound often mistaken for ultrasound, is simply the harmonic overtones of the handpiece and phaco tip.
22
THE A RT
OF
PHACOEMULSIFICATION
Phacoemulsification Terminology Phaco power Phaco power is the ability of the phaco handpiece to cut or emulsify a cataract. Phaco power is directly related to stroke length, frequency and efficiency of handpiece. Stroke length Stroke length is the distance by which the titanium phaco tip moves to and fro. It is the most important factor in deciding the phaco power. The stroke length can be altered by changing the phaco power setting of the machine. Frequency Frequency is the number of times the tip oscillates and is fixed for a particular phaco handpiece. It is measured in kHz. Preset levels Each surgeon sets his level which he does not wish to exceed during the surgery, both for minimum levels as well as maximum levels. This is done so that the safe levels are not exceeded inadvertently during the stress of surgery. Linear v/s panel In linear control, pressing the foot-pedal leads to gradual rise of the parameters from zero to preset maximum with a linear relation to the footpedal control. In panel mode, the parameter reaches the preset panel maximum on pressing the foot switch without any linear foot pedal control. Essentially it panel is simply on or with no variables in between, panel mode is normally utilized for diathermy, flow rate or for vitrectomy settings, never for power settings or aspiration settings on phacoemulsification. Constant v/s pulse phaco power In constant mode, power is delivered continuously and it can be linear or panel controlled. Pulse mode allows phaco power to be delivered at preset intervals which can be varied. The pulse mode gives relative intervals where there is absence of tIP movement. This improves the flow characteristics and helps in evacuating small nucleus particles towards the end of the surgery. The pulse mode is also relatively safer for the epinucleus because a more consistent and predictable cutting power will provide greater stability in the posterior chamber. Maximum phaco power Maximum phaco power is preset by the surgeon. It determines the maximum obtainable ultrasonic energy when the foot pedal is fully depressed. Actual phaco power Actual phaco power in a machine with a linear foot-pedal control is proportional to foot pedal position and denotes the power actually being delivered at a given time. Effective phaco time (EPT) Effective phaco time is the total phaco time at 100 percent phaco power. It can be less than the total foot-pedal time. EPT is very significant as less EPT indicates proportionately less energy delivered to the eye thereby reducing the side effects of phaco power. When one compares the energy used between different types of procedures, or even between different instruments, one has to compare the EPT which is a much more accurate index.
T HE P HACOEMULSIFIER
23
How does phacoemulsificatlon work? There are various steps involved in the actual phacoemulsification process. • Mechanical contact of the tip with the lens • Acoustical wave transmitted through fluid in front of the tip • Cavitation At the cessation of the forward stroke, the tip has imparted forward momentum to the fluid and the lens particles in front of it. On the tip being retreated, the fluid cannot follow thereby creating a void in front of the tip. The void is collapsed by the implosion (cavitation) of the tip thereby creating additional shock waves. • There is an impact of fluid and lens particles being pushed forward in front of the tip. Considering the mechanics of phaco it is clear that there is attenuation of energy on phacoemulsifying within nuclear material. This reduces the deleterious effects on the corneal endothelium. Therefore, posterior chamber phaco helps to maintain the safety of the procedure by increasing the working distance from the endothelium. Further, if phaco power is used only when the tip is in the nucleus, the safety margin is significantly enhanced. The ultrasonic handle has three functions, namely, irrigation, aspiration and fragmentation. These can be operated separately or simultaneously. The dynamics of irrigation and aspiration are now considered in detail. Irrigation System In most phacomachines, irrigation during phacoemulsification is provided by gravity feed through the space between the titanium phaco tip and the sleeve. The amount of irrigation is determined by the bottle height relative to the patient’s eye, by the sleeve diameter, and most importantly by the loss of fluid from the eye. Stable anterior chamber dynamics: Irrigation = aspiration + leakage from the wound. Rigid sleeves may be preferred over flexible sleeves because the irrigation does not get compromised while manipulating the handpiece in the incision. The height of the irrigation bottle during phaco is usually placed between 65 cmand 75 cm above the eye level. The eye should be at the same level above the floor as the pump (cassette) of the phacoemulsifier. Aspiration System Aspiration is defined as the evacuation of fluid through a closed system. Two important concepts concerning aspiration are flow rate and vacuum level. Flow rate Flow rate is the quantity of the fluid pulled from the eye per minute through the instrument tip and irrigation tubing. Flow rate therefore helps in bringing the material towards the tip. Flow rate is measured in cc/min and is dependent on the level of vacuum created in the aspiration tubing by the aspiration pump
24
THE A RT
OF
PHACOEMULSIFICATION
and surface area of the port of aspirating tip. Flow rate determines the rate of rise of the aspiration vacuum when the aspiration port is occluded. Vacuum Vacuum level is the difference in pressure between atmospheric pressure and the pressure inside the aspiration tubing. This is a negative suction pressure that is created by the pump. Port vacuum (mm Hg/min) = the vacuum created (mm Hg) port area (mrn)2 The vacuum level created at the port therefore varies inversely with the diameter of the tip. The vacuum or negative suction force created helps in holding the material to the tip and its final aspiration. Aspiration Pumps Depending on the machine, three kinds of pumps are used to control aspiration and produce the negative suction pressure, i.e. the vacuum (Fig. 2.7). They are • Peristaltic pump • Venturi pump • Diaphragmatic pump.
Fig. 2.7: The peristaltic pump of the opticon P4000 machine
The peristaltic pump is also known as a “constant flow” pump while the Venturi and the diaphragmatic pumps are of the “constant vacuum” variety. Peristaltic Pump Peristaltic pump (Figs 2.8 and 9) was popularized by the heart-lung machine. In these pumps, a pressure differential is created by compression of the aspiration tubing in a rotatory motion. When the rotational speed is low, vacuum develops only when the aspiration port is occluded. On occlusion, vacuum builds up to preset
T HE P HACOEMULSIFIER
25
Fig. 2.8: Full function display of the Alcon Legacy Machine
value in a stair-stepped pattern. By increasing the rotational speed, as in the newer generation machines, a linear build-up of vacuum occurs even without occlusion of the tip. It can thus be made to simulate a Venturi or a diaphragmatic pump. Advantages of a Peristaltic Pump • • • •
•
Fig. 2.9: Hand-held full function remote control of Alcon which controls the Alcon Legacy Machine
• •
Vacuum build-up occurs only on occlusion of the aspiration port. There is a large safety margin in this pump as it is slower in building up vacuum The peristaltic pump is a dedicated anterior segment system The peristaltic system is a more forgiving system as there is no inadvertent pull on the ocular structure since vacuum builds up only on occlusion The fluidics of the peristaltic pump are more controlled with little or no deflation of the anterior chamber on sudden removal of occlusion Vacuum level and flow rate may be controlled independent of each other Peristaltic pump allows both zero and high vacuum phaco.
Disadvantages of a Petlstaltlc Pump • The vacuum build-up is directly related to the density of occlusion which in turn would depend upon the bevel angle of the titanium tip and the tissue density • The vacuum build-up is in a stair-stepped pattern
26
THE A RT
OF
PHACOEMULSIFICATION
• Because of the stair-stepped pattern of the vacuum build-up, there could be more pulsations in the anterior chamber • True linear aspiration is not seen, however newer pumps do simulate a linear build-up of vacuum • One has to mechanically approach the nuclear or cortical matter to first achieve occlusion for vacuum to build up in order to aspirate the tissue. However, the rapid rotation mode has significantly improved the followability of the tissue, even in the peristaltic pump. Venturi Pump A Venturi pump uses compressed gas to create inverse pressure. Vacuum generated is related to gas flow which in turn is regulated by a valve (vacuum build-up occurs linearly in a consistent manner from zero to a preset value. The build-up is almost instantaneous on pressing the foot-pedal. Due to this there is an increased risk of iris trauma and posterior capsular rents which make these pumps unsafe, particularly so for beginners. Advantages of a Venturi Pump • There is a good follow ability of the tissue • The vacuum build-up is linear • There is a consistent increase in the vacuum from zero to the preset level on depressing the foot switch • Nuclear and cortical material can be attracted towards the probe on depressing the foot pedal. Disadvantages of a Venturi Pump • This pump has the least safety margin and is not forgiving to the surgeon • The rise time is too fast • There is an immediate rise in the vacuum on pressing the foot switch to position 3 without any linear foot pedal control • The incidence of iris chaffing and posterior capsular rents have been reported to be much higher with this pump as compared to’ the peristaltic pump • Venturi pump does not allow either zero to high . vacuum phaco. Diaphragm Pump A diaphragm pump uses a flexible membrane within a cassette to generate vacuum. Build up of vacuum is more linear and reaches the preset level even without occlusion. This makes it unsafe. However, lens material can be aspirated without having to mechanically approach it. Advantages of Diaphragm Pump • There is an improved linearity of vacuum build-up • The flow rate and aspiration are faster • Tissue can be pulled towards the center as vacuum builds up to preset even without occlusion
T HE P HACOEMULSIFIER
27
• There is a greater control with the diaphragm pump during posterior segment surgery. DIsadvantages of a DIaphragm Pump • This being a faster pump it offers lesser safety margin • Foot pedal depression does not have a very good graded control over vacuum build-up • Rise of vacuum depends on the fluid in the chamber • Vacuum build-up reaches preset level even without occlusion. This leads to inadvertent pull on ocular tissue resulting in a higher complication rate • A Venturi is not a forgiving pump and has to be handled by newcomers with caution though in the hands of an expert it can give excellent results. Physics of Phaco: Certain Aspects Aspiration pressure It is modified depending on the stage of surgery and is inversely proportional to the diameter of the aspirating port. The ultrasonic tip has a port diameter of 1.00 to 1.20 mm with which the maximum vacuum achievable is 70 to 100 mm of Hg. However, in new machines (Alcon’s Legacy, and the Allergan Sovereign series, etc.) the vacuum can be raised to 500 mm of Hg in the phaco mode. The I-A tip has a diameter of 0.3 mm and the aspiration pressure may be increased to 500 mm of Hg. Rise Time and Pump Flow Rise time Rise time is a measure of how rapidly vacuum builds up once the aspiration port is occluded. Pump flow Pump flow is a measure of the rotational speed of the peristaltic pump head (which in turn determines flow rate and aspiration). This changes from machine to machine. RelationshIp of Illse TIme and Pump Flow As the pump flow increases, vacuum builds rapidly as the tip is occluded and therefore the rise time decreases. Pump flow is usually preset by the surgeon and is measured as a percentage. Normally 100 percent is equal to flow rate of 35 cc/minute. It is an overall measurement of fluid turnover in the eye. Pump flow determines rise time and event time. Vacuum Settings Maximum vacuum Determine the maximum obtainable vacuum when the aspiration port is fully occluded. Maximum vacuum is preset by the surgeon and is measured in mm of Hg. Typical settings are 65 to 75 mm Hg for phaco and 400+ mm Hg
28
THE A RT
OF
PHACOEMULSIFICATION
for irrigation aspiration. The speed at which this vacuum is achieved is determined by the pump flow setting and the bore of the aspirating tube. Actual vacuum It indicates pressure at the aspirating port at a given time. This depends on the maximum preset pump flow, degree of tip occlusion and position of the foot pedal when linear control is used. RelationshIp Between Pump Flow, Irrigation and Aspiration With an increase in the pump head rotational speed, the pump flow increases. Due to this, both aspiration rate and irrigation flow also increase. Relationship between pump flow, rise time and vacuum To reach a preset vacuum, as pump flow increases the rise time decreases, e.g. if pump flow is doubled the rise time gets halved. Fluidic Balance Fluidic balance is the balance between inflow of fluid into the eye and the outflow of fluid out of the eye, which helps in maintenance of the IOP. An adequate fluidic balance provides • Constant lOP • Stable anterior chamber • Protects corneal endothelium and posterior capsule. The amount of irrigation is determined by the bottle height relative to the patient’s eye, by the sleeve diameter, and, most importantly, by the loss of fluid from the eye. The following situations may exist: • Balanced anterior chamber dynamics: Irrigation = Aspiration + Leakage • Tight Wound (Irrigation decreases):
Irrigation »Aspiration + Leakage A tight wound can limit irrigation if the space between the flexible sleeve and the phaco tip is compromised when manipulating the handpiece. This problem is not present with a rigid sleeve, although such a sleeve is not generally used because ‘contact with the rapidly vibrating phaco tip can cause serious overheating problems and tip damage if the sleeve is metallic. Excessive leakage between the rigid sleeve and the wound can also occur. Most surgeons have a shelved incision to reduce iris prolapse and create an efficient wound, although some fluid will, and should, always leak through the wound entering the anterior chamber. With an appropriate size keratome, i.e. 3.00 mm to 3.20 mm, proper irrigation is facilitated. Irrigation Bottle Height The irrigation bottle during phaco is usually placed between 65 em and 75 em above the eye. The eye should be at the same level above the floor as the pump (cassette) of the phacoemulsifier. It must be however understood that merely raising the bottle will not cure the problem of anterior chamber collapse caused by an
T HE P HACOEMULSIFIER
29
excessively large wound, since irrigation and leakage are increased concomitantly by the increased force created by the higher bottle. Height of the Tubes Running from the Handpiece to the Machine Often forgotten and a sadly neglected fact that water always runs down hill. If the tubes are not supported but allowed to sag, when the phaco is turned off, the chamber will continue to empty which will lead to a chamber collapse. In addition it leads to air being sucked in the tubes which will lead to a compromised suction when the handpiece is used again. Always support the tubes at the same level as the patient’s eye. Thus a table which can be raised (A Mayo trolley) is an integral part of the set-up for good phaco. Enhancing Fluidics in a Phaco emulsification Machine The various ways to achieve good fluidics in a machine are 1. Narrower bore aspiration tubes Different size of irrigation tubing and aspiration tubing, irrigation being larger than aspiration. By making the aspiration tube with a narrower bore yet with a thicker wall decreases the compliance of the tubes and enhances response time with a more accurate aspiration pressure. It is the use of these high compliance tubing’S, (also called HiVac tubing’s) coupled with computerized controls which have made the new third-generation machines use higher and higher vacuum (and lower and lower ultrasound timings) with great safety. 2. By using software-driven pump systems where aspiration rate slows as the vacuum rises. Aspiration is zero at maximum vacuum, i.e. on total occlusion. As the occlusion clears aspiration rate slowly starts to rise from zero to maximum. A number of pumps at the time of release will actually reverse the pump a bit to make sure that the release is accurate. Foot Pedal The mode of operation in which the instrument is functioning on depressing the foot pedal in a linear manner is shown by the position indicator. Position 0 : Instrument turned on, no fluid flows, no ultrasonic vibration. Position 1 : Only irrigation solution is flowing. Position 2 : Irrigation and aspiration occur simultaneously. Position 3 : Irrigation, aspiration and fragmentation take place simultaneously. There is no position 2 or 3 in irrigation mode. Similarly there is no position 3 in I-A mode. A useful option in some machines is a foot pedal reflux control wherein fluid is pushed from the aspiration line to the eye in order to disengage tissue that has been inadvertently engaged in the tip. Programmability of foot switch In most of the machines these positions are prefixed. But in new advanced machines the positions are programmable, e.g. in Alcon Legacy, and the Allergan Sovereign.
30
THE A RT
OF
PHACOEMULSIFICATION
This provides more space between the position required and allows a better linear control. What Mode is the Machine Functioning? Every phacoemulsifier has various feedbacks to help the surgeon determine the mode in which the machine is presently working. The feedbacks could be tactile (foot) or auditory. Some machines have an electronic feedback. Tactile Feedback Increasing vibration in the handpiece may be indicative of increase in ultrasonic energy. Tactile feedback from the foot pedal depends on the machine. Some may have detents between the various positions while others may have uniform resistance, i.e. no detents. In few machines there is an increasing resistance with increasing depression of the foot pedal, while some machines may have vibrations between pedal positions. AudHoty Feedbacks Various auditory feedbacks are also preset in the phacoemulsifier. Auditory electronic sounds may be an additional feature of some machines, e.g. beeping is indicative of 1/A mode while a bell is indicative of occlusion at the tip. User modification may be possible with regard to type and intensity of the sound. Some machines will actually talk and tell you the mode you are in. And you can even choose your language. Essentials of a Good Phacoemulsifier • Good panel display (Fig. 2.10) with tactile controls parameters display. A continuous display showing all the variables at any time including the effective phaco time are essential. Simple fixed display machines are not really useful. • Linear and pulse mode should both be available in phaco mode. • Motorized infusion pole is preferable to change bottle height in order to alter rate of infusion. • Vacuum and flow rate should have separate controls. • Wet field cautery with a co-optation forceps should always be there with the cautery having the facility of both fixed and linear output. • A full function anterior vitrectomy unit should be in order to manage the complication of vitreous loss if it occurs. • Multifunction foot switch, i.e. it controls all the parameters with different foot switch positions is essential. A reflux mechanism should be present in the foot switch in order to disengage tissue when required. • Programmable foot switch is ideal. This facility available in the Allergen Sovereign and the Alcon Legacy series helps change the detents of the foot position 0, 1, 2 and 3. This enables the surgeon to have a greater play for different foot positions.
T HE P HACOEMULSIFIER
31
• Full function remote is extremely valuable. It can be easily kept in the sterile area to facilitate change of parameters by the surgeon during the procedure. • Multiple programmability facility must be available. This allows individual surgeons to feed in their own parameters for surgeries on various grades of cataract as also for different stages of the surgery in a given case. • Facility for Reusable tubing’s and cassettes The machine should be cost-effective. In Indian circumstances, it is imperative that one be able to reuse the tubing’s, sleeves, etc.
Fig. 2.10: P4000 machine (Opticon)
The ophthalmologists need to carefully look at the various options available to them before purchasing a phacoemulsifier. Though several manufacturers are selling their products, one needs to remember that phaco surgery is significantly machine dependent. A good service back-up is therefore essential. It would always be prudent to take the opinion of a contemporary who is more experienced in the art of phaco surgery regarding which equipment to buy and the reasons for the same before taking a final decision. FURTHER READING 1. Arensten IT et al: Corneal opacification occurring after phacoemulsification and phaco fragmentation. Am J Ophthalmol 73: 794-804, 1977. 2. Barbell A: Health devices: Phacoemulsification systems. ECRI 18: 392, 1989. 3. Benolken RM, Emery JM, Landis OJ: Temperature profiles in the anterior chamber during phacoemulsification. Invest Ophthalmol 13: 71-74, 1974. 4. Binder P: Corneal endothelial damage associated with phacoemulsification. Am J Ophthalmol 82: 4854, 1976. 5. Ito K: Experimental studies on clinical and pathological changes of neighboring tissues of lens by ultrasonic vibrating tip for phacoemulsification. Japanica 74: 725, 1970. 6. Kelman CD: Physics of ultrasound in cataract removal. Int Ophthalmol Clin 9: 739-44, 1969. 7. Maloney WF, Maloney K: Two handed method. Video Journal Ophthalmology 3(4): 1987. 8. Pollack FM, Sugar A: The phacoemulsification procedure. II Corneal endothelial changes. Invest Ophthal 15: 458-69, 1976.
32
THE ART
OF
P HACOEMULSIFICATION
David E Allen
New Phacomachines Offer More Control
3
INTRODUCTION The art of cataract extraction has developed considerably over the past ten years, largely as a result of the increased sophistication of the machines the surgeon now uses to assist the process. This chapter is written with particular reference to four machines of which the author has experience. They are not the only machines to offer some of the features described, and the reader should be able, as a result of the knowledge gained from this chapter, to make appropriate enquiry to the manufacturers of other machines. THE MACHINE IN PRINCIPLE A generic phacoemulsification machine consists of two components: a pump which assists in the flow of fluid through the eye, and a probe which helps to reduce the crystalline lens to pieces of a size that can be removed by the pump. The pump actively controls the volume and rate of removal of some fluid from the eye (other fluid leaves the eye through incisional leakage). Inflow of fluid to the eye to replace that which leaves is passive, and is dependent on gravitational forces, modulated by the inflow resistance of the tubing and handpiece. If the anterior chamber is to be safely maintained, then the maximum potential inflow must exceed the maximum outflow of fluid. Modern machines offer very good control of the active removal of fluid. Phacoemulsification is achieved by either a magnetostrictive mechanism (the original type of phaco generator still used in some handpieces) or by the piezoelectric
NEW PHACOMACHINES OFFER MORE C ONTROL
33
effect of certain types of crystals. Computer-controlled electrical circuits produce vibrations that are transmitted to a tungsten tip. The active mechanisms by which phaco needles work to break down the solid lens include • Ultrasonic effect of the high speed vibration of the tip • Cavitation effects • Particle and fluid pressure waves • Direct mechanical jack-hammer action. Modern phaco machines make use of sophisticated computer electronics to control the phaco-tip, and the tips themselves are being designed to enhance the “cutting” action, particularly using cavitation. SURGEON’S REQUIREMENTS DURING THE PROCEDURE Efficient Removal of Fluid and Particles from the Eye As described above, fluid egress from the eye is largely controlled by the machine, and is the most important factor in maintaining the correct balance. Modern pumps allow the rate of flow from the eye through the handpiece to be varied. Peristaltic pumps generate positive displacement of fluid along the handpiece tubing. Varying the pump speed has a direct effect on the rate of fluid displacement. Vacuumbased pumps (venturi or rotary vane) generate flow as a result of the pressure difference between the needle tip in the eye (at IOP) and the vacuum chamber in the machine. Varying the vacuum level in the cassette indirectly affects flow in these machines. A new type of pump (scroll pump) has characteristics that allows it to work in a way that emulates either of these modes and is currently available in the “Concentrix” module as an option in the Bausch and Lomb “Millennium“ machine. Varying the flow rate allows the surgeon to control the speed with which material is attracted to the tip. Low flow allows the surgeon to safely work close to sensitive structures such as iris or capsule (Fig. 3.1). A high flow rate causes material to be rapidly attracted to the tip, and once the tip is occluded, produces a rapid vacuum build-up. Flow of fluid also, as a secondary effect, tends to cool the phaco tip, both the flow along the lumen, but more importantly the compensatory inflow along the irrigation channel round the outside of the tip. Efficient Removal of Lens, Reducing the Amount of Phaco Energy Used This is achieved by using appropriate vacuum settings. In a flow-based system the maximum vacuum generated at the tip can be adjusted independently from the flow rate. In vacuum-based systems the vacuum in the machine vacuum chamber is the maximum vacuum that can be generated at the tip, as well as the factor controlling flow through the handpiece. Modern phaco machines allow the surgeon to select a vacuum setting that is much higher than was safe with older systems. Many modern systems can now safely use maximum vacuum settings between 300 and 500 mm Hg.
34
THE ART
OF
P HACOEMULSIFICATION
Fig. 3.1: Low flow allows the surgeon to work safely close to sensitive structures such as iris and capsule
High-vacuum settings allow lens material to be removed through the tip with less phaco energy. High vacuum can cause nuclear material to be “molded” through the phaco tip without phaco energy, or allows low phaco power to be assisted by high vacuum in achieving the same effect. Whenever phaco power is applied to nuclear material at the tip the forward movement of the tip repels the material. Material held by a higher vacuum (and/or flow) onto the tip is less likely to be pushed away—increasing the tip’s effectiveness. The efficiency with which a tip functions to ”cut” lens material is another important factor. The tip and the handpiece (particularly the crystals or magnets it contains) tend to heat up during use, and this can have a significant impact on the functioning of the tip. The most efficient phaco machines have electronic circuits that frequently (or even “constantly”) monitor the performance of the tip and adjust the frequency of vibration to take account of this—“auto-tuning”. Some also adjust the voltage (power) applied to the tip to take account of the loading on the tip—taking account for example of the difference in effective weight of a tip if a large nuclear fragment is impaled on it. The shape of the tip has been the subject of developments in the past 10 years. The first really different tip was the “Cobra tip” (Fig. 3.2) produced by Surgical Design, and this has now been copied by several manufacturers. This tip has a conventional external diameter at the external opening, but 2 to 3 mm behind this the tip reduces significantly in diameter. The “shoulders” inside the tip increase the surface area moving back and forth, which has two effects. There is an increase in the amount of acoustic shock waves, but more importantly an increase in the cavitation effects of the tip (Figs 3.3A and B). Alcon subsequently developed the “Kelman tip”, which is bent and generates similar increased acoustic and cavitation
NEW PHACOMACHINES OFFER MORE C ONTROL
35
Fig. 3.2: The Cobra tip
effects. Some manufacturers and surgeons are now reinventing the external shape of the tip to produce chisel, diamond, hexagonal and other external shapes. As a result there are now a group of needles from different manufacturers which are particularly powerful when used with medium or hard nuclei. A needle with a flared tip is also one with a reduced bore behind the tip. This has an indirect effect of reducing postocclusion surge because of increased resistance to outflow. Stable Intraocular Environment The modern cataract operation is characterized by superb control of the intraocular environment. This has resulted in a procedure that produces much less disruption
Figs 3.3A and B: Cavitation effect of Cobra tip
36
THE ART
OF
P HACOEMULSIFICATION
to ocular physiology than older techniques. This is partly because of the reduced incision size, but also partly because of increased anterior chamber (AC) stability. AC stability is enhanced when fluid entry and egress to and from the eye are in balance. Modern machines allow this despite the trend, alluded to above, towards higher flow and/or vacuum settings. Modern peristaltic pumps produce a much smoother flow pattern than the versions used on early phaco machines. Modern microprocessor controls make the pump much more responsive to the surgeon’s demands (expressed through the foot pedal). Higher maximum vacuum settings can now be used as a result of this improved control as well as other improvements. When the phaco tip is occluded, vacuum in the tip and tubing equilibrates and approaches the maximum preset. There is then a pressure difference between the anterior chamber (usually at +30-40 mm Hg) and the lumen of the tubing or vacuum chamber (at the preset maximum, which may be as low as –400 mm Hg). When occlusion breaks, there is a rush of fluid from the AC into the tubing to equilibrate the pressures—postocclusion surge. Soft silicone tubing connecting the handpiece to the machine can tend to collapse when subjected to high vacuum during occlusion, and with the release of that vacuum can be subject to a rebound expansion. This re-expansion of tubing tends to increase the magnitude of postocclusion surge. Modern machines use a combination of different strategies to deal with postocclusion surge. The first machine to actively combat this problem and allow the use of high vacuums was the Surgical Design “Ocusystem“. This machine uses a combination of relatively rigid tubing and a pressure sensor in the aspiration line. When this sensor detects a rapid rise in pressure in the aspiration line (i.e. occlusion had broken), a pinch valve opens allowing a small volume of fluid from a secondary (higher) bottle to enter the aspiration line neutralizing the pressure differential. Other options range from the use of rigid tubing as well as an extremely low compliance pump (Bausch and Lomb ‘Millennium’), through microprocessor control that delays the uptake of pump rotation and sometimes transient reversal of pump (AMO ‘Sovereign’), to adjustments in the phaco tip that reduce the pressure difference (Alcon ‘ABS tips’). These and similar developments mean that modern machines can use much higher vacuum levels (up to 450 or 500 mm Hg during phaco), while still maintaining stability of the AC. This means the lens is removed more efficiently, there is less risk to the integrity of corneal endothelium and posterior capsule, and reduced breakdown of the blood-aqueous barrier (BAB) as a consequence of reduced AC pressure fluctuations. Reduced Flow of Fluid through the Eye Fluid outflow from the eye during phacoemulsification is not just through the aspiration port of the tip. There may be significant incisional leakage through the main incision or any side-port incisions. Traditionally, incisions have been made large relative to the tip, to avoid compression of the silicone irrigation (inflow) tubing. Compression of the inflow tubing has two effects: (i) maintenance of the
NEW PHACOMACHINES OFFER MORE C ONTROL
37
anterior chamber is compromised by restricted inflow, and (ii) the cooling effect of fluid flowing around the vibrating needle is lost. Thermal damage of cornea/sclera then results. A rigid sleeve does not have this disadvantage, and the incision can be made as tight as possible – but there will always be some leakage. An alternative approach is to ensure that complete occlusion of a silicone sleeve cannot occur. This can be achieved by having some rigid elements within the soft, compressible outer shell (Fig. 3.2), or by grooving the external wall of the tip (Fig. 3.4). Fig. 3.4: Grooves on external wall of tip The volume of fluid flowing through the eye during surgery correlates with the risk of endophthalmitis and endothelial damage. It may also correlate with reduced inflammation and reduced BAB breakdown. Safe and efficient sealing of the incision around the probe along with the factors discussed above (higher vacuum, efficient tips) lead to a much reduced volume of fluid flowing through the eye. These all contribute to the improved outcomes of modern phaco procedures compared to earlier results. PRINCIPLES OF LENS REMOVAL AND MAKING BEST USE OF MACHINE PARAMETERS Original phaco techniques were dependent on the use of power to gradually “shave” the nucleus into a smaller and smaller piece. Modern lens removal strategies are based on mechanically breaking the lens into smaller fragments that can then be consumed. If the surgeon uses a technique such as “divide and conquer“ or “stop and chop”, the phaco probe is used to cut one or more grooves through the nucleus. These grooves are normally best cut as deeply as possible, close to the posterior capsule. Often (particularly when dealing with a soft nucleus) the surgeon will also wish to sculpt out close to the equator of the lens. When sculpting close to the posterior capsule, capsulorrhexis edge or iris (Fig. 3.1), the surgeon requires low flow settings on the phaco machine in order to minimize the chances of these structures being drawn into and being damaged by the phaco probe. As well as low flow during nucleus sculpting, the machine should be set with a low vacuum. This ensures that if the wrong structure is accidentally engaged, less damage will result and, for example, the capsule may not be torn if the preset vacuum is, say, 20 mm Hg. It is during the sculpting phase that the surgeon is most likely to require higher levels of phaco power, as the technique uses phaco power to cut through the lens material. The amount of power used should be titrated to the density of the lens. Once the lens has been broken into pieces (either by sculpting or by chopping/
38
THE ART
OF
P HACOEMULSIFICATION
snapping), phaco power is much less important, and lower power levels can be used. Alongside this reduced emphasis on power however is the need for higher flow and higher vacuums when consuming the lens pieces. The higher vacuum holds pieces onto the tip and molds the pieces into the tip as described above. The flow rate determines the rise-time of the vacuum (the speed with which the preset maximum is reached) in a flow-based system, as well as determining the rate at which material is attracted to the unoccluded tip. Control of Power In this discussion I will use the conventional notation about foot pedal “positions” Fig. 3.5A). Position 1 refers to the range of pedal movement that allows simple irrigation. Position 2 refers to irrigation and aspiration, while position 3 adds phaco power. The first phaco machines had power output that was fixed at the control panel. All phaco machines currently in use allow the surgeon to control the power
Figs 3.5A and B: (A) Conventional foot pedal position,and (B) dual linear pedal
NEW PHACOMACHINES OFFER MORE C ONTROL
39
in a continuous linear fashion in position 3 of the foot pedal. The surgeon can therefore titrate the power used according to the density of the nucleus and according to the particular maneuver in hand. One circumstance when a surgeon may choose panel control of power is when dealing with a hard nucleus. Under these circumstances it is easy for a surgeon to tend to try to physically push the tip into/through the nucleus, rather than allowing the ultrasonic power to achieve this. This inadequate use of power can put considerable strain on the zonules. Setting a fixed power level (say 50% or 70% for a very hard nucleus) ensures that maximum power is available to “cut” into the nucleus as soon as foot pedal position 3 is entered. Alternatively, the Bausch and Lomb “Millennium“ phaco machine has the ability to set a minimum as well as a maximum in its linear control of power. A surgeon therefore may choose linear control but power may start at say 35 percent and range up to 70 percent linearly. This retains the advantages of linear control (use of the minimum power level necessary) while reducing the chances of using totally inadequate power levels. While sculpting, the surgeon requires continuous power while the tip is advancing. For consumption of fragments, however, continuous power is not required, as the aim is to try to mold the fragment into a size and shape that will be aspirated. One of the earliest advances in the control of power, shortly after the concept of linear control, was pulsed phaco. The original rationale for this was the fact that applying phaco power to a fragment tends to push it away from the tip by direct hammer action of, and shock-waves from, the needle. After a short burst of phaco, the fluidics draw the fragment back into contact with the tip and re-establish occlusion in preparation for the next burst. Therefore the original drive was to make the power application more efficient. Now, however, the main reason for surgeons using pulsed power is to reduce the total amount of phaco energy used, by using short bursts to assist the vacuum in molding material into the tip. The second reason is that consumption of material is slowed, making it less likely that rapid consumption will lead to the attraction of unwanted material (such as capsule or iris) into the tip. Most implementations of pulse mode allow power to be varied as normal in foot pedal position 3. Some recent machines have an additional power variation known as burst mode. The machines can be set to deliver a single burst of power, at a fixed level, and for a set (but adjustable) duration, or can be made to deliver bursts at increasing frequency depending on the foot pedal traverse in position 3. With burst mode, power is delivered at a fixed level. Burst mode is said to be particularly helpful with techniques that employ embedding the phaco tip into the nucleus for some form of chopping/splitting. The use of a very short burst at fixed power prevents lateral spread of acoustic energy from the tip that can lead to the creation of a space around the tip preventing good occlusion. This type of phaco is said to lead to more efficient embedding into a tight-fitting space in the nucleus, and so a better seal is created, leading to better vacuum rise time, etc.
40
THE ART
OF
P HACOEMULSIFICATION
Control of Flow All phaco machines currently in the market allow the surgeon to control the aspiration flow rate. Those surgeons using vacuum-based machines, however, (mainly Venturi) often do not realize that they have control. The principles were mentioned earlier in this chapter. Flow-based machines give the surgeon the ability to directly control the aspiration flow rate, and have a dial or electronic equivalent, that allows the direct setting of a flow rate in cc/min. With an unoccluded tip, flow in a vacuumbased system is generated by the pressure difference between the anterior chamber and the vacuum chamber in the machine. Therefore as the commanded vacuum in the cassette increases, the pressure gradient along the tubing increases, generating increased flow through the handpiece. Some machines have linear adjustment of flow rate (or vacuum) during traverse through foot pedal position 2. In addition to this feature all machines in current production have various memory settings which allow the surgeon to set different flow rates for different parts of the procedure, or different types of cataract. This ability to adjust flow rate (directly or indirectly) allows the surgeon to control the flow of material to the tip, or the vacuum rise time when the tip is occluded. The surgeon should always remember that all flow of fluid out of the eye must be balanced by an equal inflow if the anterior chamber is to remain stable and formed. In addition, the higher the outflow (and therefore inflow), the greater is the potential turbulence in the anterior chamber. This is important in the presence of an unstable capsule caused by either a break in the rhexis rim, or a rupture in the posterior capsule. Under these circumstances the surgeon should reduce the flow rate as well lowering the bottle height, to minimize stress on, and potential extension of, the torn capsule edge. Control of Vacuum Control of the maximum vacuum level has only recently become something that has featured in the thoughts of surgeons. This is because earlier generations of machines did not allow safe use of vacuum levels above 75 or 100 mm Hg. For the reasons rehearsed earlier in the chapter, higher vacuum levels are now safely used in some modern machines. The vacuum set on a flow-based machine is the maximum vacuum the machine is allowed to generate once the tip is occluded. In unoccluded mode, the pump generates flow through the tip. With an occluded tip, the pump continues to turn and generates a vacuum that rises towards the preset maximum. Once that maximum has been reached the pump stops and/or some venting into the aspiration line is allowed so as to maintain that vacuum level. Once the occlusion breaks and the vacuum reduces, the pump begins to turn again. In a vacuum-based machine, the set vacuum is that which is produced constantly in the vacuum chamber, and this indirectly causes flow until occlusion, and then, as in the flow-based system, the vacuum within the tubing and handpiece rises to that same level.
NEW PHACOMACHINES OFFER MORE C ONTROL
41
However, there is more to vacuum control than just altering the maximum allowable vacuum. The rate at which that vacuum is actually achieved at the tip can be varied. This can be of use to the surgeon who may wish, for example, to have a slow “rise-time” of the vacuum, allowing time to correct any mistakes, rather than allowing a high vacuum to be rapidly reached before any error (such as engagement of capsule) can be corrected. In a vacuum-based system (e.g. Venturi) some form of restriction or limiter can be placed which effectively allows leakage of some of the vacuum as it rises towards the maximum. In a flow-based system the flow rate determines the rate at which the vacuum limit is reached. For example, once occlusion occurs, a given set vacuum is reached in half the time if the flow is set at 20 cc/min compared to the time taken at 10 cc/min. Thus a simple system can easily be adjusted to give the surgeon a vacuum rise-time he or she desires, at the expense of independent control of the flow rate. The Surgical Design Ocusystem was the first machine to offer the surgeon the opportunity of adjusting the rise-time while keeping the basic flow rate independent. For consuming nuclear quadrants, for example, the basic flow rate could be set at 25 cc/min, with a maximum vacuum of 350 mm Hg. The surgeon can, however, set a threshold vacuum (say 100 mm Hg) and when the vacuum reaches this level, a different flow rate is used. The postthreshold flow rate could be much lower— say 15 cc/min—if the surgeon feels there are risks with a rapid rise to higher levels (Fig. 3.6). Alternatively the post threshold rate could be even higher, if the surgeon is happy that by the time 100 mm is reached the chances of the wrong material being engaged has disappeared. Several machines now allow the surgeon to program this type of change in the rate at which the vacuum limit is reached. Sophisticated Foot Pedals As in the earlier section I will use the conventional notation about “positions“. Position 1 refers to the range of pedal movement that allows simple irrigation. Position 2 refers to irrigation and aspiration, and position 3 adds phaco power.
Fig. 3.6: Adjustment of vacuum rise time
42
THE ART
OF
P HACOEMULSIFICATION
Fig. 3.7: Multifunciton foot switch
Foot pedals are no longer simple car accelerator-type pedals. Many machines now offer the opportunity to adjust the proportion of the total range of movement that is used for the three different “positions”, and save these in memory as part of the settings for a particular surgeon. All pedals have had some form of additional toggle switch (usually to the side) that switches on the reflux function. Now, however, it is common to offer even more toggle or rocker switches (Fig. 3.7). These can be used for activating the in-built diathermy on many machines, or for switching to different memory settings or “modes“. Some pedals can be set to activate an “emergency”change (for example rapid lowering of the inflow bottle height in the event of a posterior capsule rupture). While many see the proliferation of switches and rockers on newer foot pedals as a yet another complicating factor, they do undoubtedly offer the surgeon a greater degree of control and independence from support staff if that is desired. They do not have to be used if not desired! Major innovation in foot pedal design leading to even greater control over the procedure has been the introduction of a simultaneous dual linear foot pedal by Bausch and Lomb. This pedal allows the surgeon (again only if desired) to have linear control in position 2 while simultaneously having linear control of position 3. Standard pedals only have travel in one plane (pitch). The dual linear pedal also has the ability to rotate sideways (yaw). The surgeon can select whether ultrasound or flow/vacuum is the function controlled by yaw. If ultrasound is chosen for yaw, then this means that position 2 can be used to linearly control flow rate (say between 15 cc/min and 35 cc/min). At any point in the travel (i.e. at any flow rate) the surgeon can then add yaw and begin phaco (Figs 3.5B and 3.8). The phaco is also linearly controlled in yaw, and so at any flow rate there is also a range of phaco—
NEW PHACOMACHINES OFFER MORE C ONTROL
43
Figs 3.8A and B: Yaw movement of pedal
say ranging from 0 to 30 percent. In addition to the added control this gives it also allows a greater range of pedal travel in position 2 (i.e. the whole range normally allocated to two and three are available in position 2). It is clear that by using this dual linear function there is a much reduced need for multiple preset modes for different types of nucleus or for different stages in the procedure. The surgeon can continuously adjust all the parameters in real time in response to changes in the characteristics of the lens being dealt with. Sophisticated Microprocessors The single most important factor in the development of the sophisticated equipment now available to the cataract surgeon has been the development of microprocessor control of the functions of the machine. Initially electronics allowed control over the power functions of the handpiece—pulse mode and “auto-tuning” being significant developments. More recently the processors have enabled sophisticated ways of enabling postocclusion surge to be controlled and minimized. The surgical design ocusystem was the first machine to introduce active ways of controlling postocclusion surge. This machine has a pressure (vacuum) sensor in the aspiration line. When it detects a significant and rapid drop in vacuum (as when occlusion breaks) fluid is immediately bled into the aspiration line, and this temporarily increases the pressure in the aspiration line, neutralizing and temporarily reversing the postocclusion surge. Later machines have employed a variety of methods to try to reduce postocclusion surge—some of which also employ sensitive pressure sensing devices and rapid electronic responses. The ability to program an adjustable rise time has already been mentioned. The AMO “Diplomax” and more recently “Sovereign” machines have made particular use of sophisticated microprocessors to enable their “occlusion mode phaco”. The
44
THE ART
OF
P HACOEMULSIFICATION
basic concept here is that the surgeon may wish to use one set of phaco parameters while the tip is unoccluded (e.g. continuous phaco at a relatively low flow while sculpting). Once the tip is occluded (which the computer detects by rapid rise in vacuum level) the surgeon may want different parameters to consume a nuclear fragment (e.g. pulse phaco, at lower power, but at a high flow). Exactly how the various parameters change, and at what threshold vacuum limit this occurs is set according to the surgeon’s preferences, and stored in that surgeon’s individual settings file within the computer. Allusion has already been made in earlier sections of this chapter to the ability of the modern computers that control the generation of the currents that power the phaco probe itself. Control circuits within the computer continuously monitor the working of the crystals or magnet stacks in the handpiece and fine tune the frequency and power with which the needle is driven. Ability to adjust the foot pedal settings, range of travel in different modes, assignment of different functions to certain switches are particularly well developed in the unique dual linear foot pedal of the Bausch and Lamb “Millennium“ machine, which has three additional switches. One of these additional controls is a rocker switch to which the surgeon can assign various functions. For example it can be set to increase or decrease flow rate, vacuum limit, bottle height, etc. or be set to move backwards and forwards between program modes. Some surgeons are intimidated by the added sophistication of the latest generation of phaco machines. However they do offer user the ability to control much of the procedure in a way which was not possible with earlier simpler machines, yet they can be used in a “simple” way if that is the surgeon’s wish. Notwithstanding these reservations, the ability to take more control is welcomed by many.
Peter L Davis
Cavitating Microbubbles Create Shock Waves that Emulsify Cataract
4
INTRODUCTION Our ophthalmic literature has frequently presented inaccurate ideas concerning the basic physics of the energy source that emulsifies cataract. Included among them are the following • Cataracts are emulsified by phaco needles moving back and forth at ultrasonic speed, jack hammering and cutting lens nuclei into an emulsate. • A 45-degree angle phaco needle emulsifies hard lens nuclei more readily because a sharper tip cuts best into the nucleus. The above statements have been repeated many times since the early days of phaco. A senior executive of a pioneering manufacturer of phacoemulsification equipment related to me told that his company did not wish to confuse eye surgeons with information about ultrasonic energy. Therefore, the marketing section of his company decided that it would be best for salesmen to inform surgeons the phaco needle was “cutting” thus making the new phaco method a simple extension of scalpel surgery. The knowledge that ultrasonically activated phaco needles are creating massive shock waves the same as those that have been used to scale teeth and clean surgical instruments is often met with skepticism. This chapter reviews library and basic research that shows the shock waves energy that emulsifies cataract are generated by imploding microbubbles that form and collapse in fluid when the titanium needle is ultrasonically activated. History The basic physics study of the massive energy created by imploding bubbles started when the British Royal Navy began using steam turbine engines in the 1890s. These
46
THE ART
OF
P HACOEMULSIFICATION
Fig. 4.1: Charles Parsons’ ship powered by a steam turbine engine
rotary engines had been engineered by Charles Parsons. He fabricated an engine and installed it in a small ship that he then demonstrated to the British Admiralty (Fig. 4.1). These new engines doubled propeller speed but to the consternation of the Navy the torpedo boat “Daring” that was first equipped with the new engines fractured its propellers. Lord Raleigh1 led the research that revealed the rapid propeller revolutions were hydrodynamically creating bubbles that imploded under water generating massive shock waves that destroyed the ship’s propellers. Research has since led us to the current situation when imploding shock waves are generated hydrodynamically, by whistles and by ultrasonic transducers that convert electrical to ultrasonic energy. KS Suslick2 has published a review of the basics of ultrasound and its applications. In the introduction to his text he states the use of ultrasound in industry and clinical medicine is common place but “there had been almost a complete lack of review material on the underlying principles from which such effects originate.” Reviewing Suslick’s 1988 multiauthor book and his article in Scientific American3 and going through the physics literature leads one to the conclusion that our phaco needles are not oscillating chisels.4 Basics of Phaco Transducers A transducer is a device that converts one energy form to another (Fig. 4.2). Handheld phaco transducers are converting electrical energy to ultrasonic (beyond 16,000 cycles/sec) acoustic waves. In the past, vibrating metal wafers were used in magnetostrictive devices notorious for generating and releasing heat that caused corneal thermal injury. Modern designs use piezo (Greek—to press) ceramic crystals to make the conversion. The piezo designs initially incorporated two crystals that were less durable than the magnetostrictive designs but were more efficient and generated less heat. In recent times, the transducers are manufactured with four crystals that makes them even more efficient (Figs 4.3 and 4.4).
CAVITATING M ICROBUBBLES CREATE SHOCK WAVES
THAT
EMULSIFY CATARACT
47
Fig. 4.2: A transducer converts one form of energy to another
Fig. 4.3: Piezoelectric transducers have replaced magnetostrictive designs
Cavitation (Transient) Eye surgeons dislike the gas bubbles that are released during phaco procedures to collect inside the corneal dome making it difficult to view the cataract. These annoying bubbles are air that was dissolved in the balanced salt solution (BSS) and was released from the BSS by ultrasonic wave activity in the eye (degassing fluids with ultrasound has been used in industry for decades).2 Unfortunately, these air-bubbles have been labeled “cavitation bubbles”. In contrast, ultrasound scientists discuss micron-sized bubbles formed by transient cavitation which is the formation of tiny invisible gas bubbles that implode creating energy at the end of acoustic horns in fluid (in ophthalmology—at the end of phaco needles) (Fig. 4.5). Ultrasonic waves released in fluid cause 100 to 150 micron size bubbles to form and then implode within a few acoustic cycles (current phaco transducers are vibrating at 27 to 55 kilohertz). The energy released by this microbubble implosion is in the form of shock and fluid waves. These activities are listed in Table 4.1.
48
THE ART
OF
P HACOEMULSIFICATION
Fig. 4.4: Drawing of both piezoelectric and magnetostrictive design transducers
Fig. 4.5: Svensson’s photo of microbubbles at the end of phaco needles
This energy has been used for many industrial purposes including the cleaning of surgical instruments and contact lenses (the tiny flame at 1500 degree C that forms as the microbubbles implode is too small to raise the temperature of the fluid).3 Activity Around Phaco Needles Research done on phaco needles in vitro has demonstrated the same activity as has been shown in physics laboratories. The first papers using photography to demonstrate the microbubbles and flame were presented in 1991 at the American Table 4.1: Activities of imploding microbubbles (transient cavitation) 1. 2. 3. 4.
Shock waves of 500 atmospheres (enormous force) Fluid waves of 400 km/h A flame at 1500 degree C within the microbubble Radical ions
CAVITATING M ICROBUBBLES CREATE SHOCK WAVES
THAT
EMULSIFY CATARACT
49
Society of Cataract and Refractive Surgery (ASCRS) Meeting. Table 4.2 outlines the initial presentations of re-phaco needle shock wave physics. Table 4.2: Outline of two presentations at ASCRS 1991 showing evidence of transient cavitation with phaco needles Author
Method
Findings
1.
Svensson B
High-speed photography
Microbubbles forming and imploding on the rim of phaco needles (Fig. 4.5)
2.
Tsubota K
Heat-sensitive photography
Demonstrated flame in imploding bubbles at end of phaco needle
In more recent times, the demonstration of shock waves in fluid that have a different index of refraction (Schlieren activity) has been shown by optical and video recording (Fig. 4.6). Table 4.3 summarizes this research. Table 4.3: Documentation of Schlieren waves at the end of phaco needles 1.
Obermaier M
ASCRS, 1996
Demonstrated shock waves in vitro and in cadaver eyes (Fig. 4.6)
2.
Fishkind W
ASCRS, 1997
Video recording of shock waves in vitro
Fig. 4.6: Photo from Fishkind’s video recording of shock waves anterior to a 45-degree phaco needle
In recent years, phaco needles have had their ends enlarged or recessed to increase their surface area and allow the formation of more microbubbles that implode and release more shock wave energy (Fig. 4.7). I demonstrate this idea with an illustration prepared in 1991. Phaco surgeons often notice cataract tissue breakdown anterior to their phaco needles without the tip touching the cataract. This is because the shock waves are
50
THE ART
OF
P HACOEMULSIFICATION
Fig. 4.7: A 0.4-mm phaco needle may have more microbubble formation
focused in front of the phaco needle as shown by Schlieren imaging. The presentation that phaco needles are simply vibrating chisels was a convenient way to market phaco technique when it was introduced. However, a reading of the basics of ultrasonic energy and more recent ophthalmic in vitro and cadaver eye studies outlined in this chapter, reveals the forces that emulsify cataract are massive shock waves formed at the end of phaco needles by imploding microbubbles. REFERENCES 1. 2. 3. 4.
Lord Raleigh: Phil Mag Serv 6:34-94, 1917. Suslick KS (Ed): Ultrasound: It’s Chemical, Physical and Biologic Effects VCH Pub Inc: New York, 1988. Suslick KS: The chemical effects of ultrasound. Scientific American 80-86, 1989. Davis PL: Phaco transducers, basic principles and corneal thermal injury. Eur J Implant Ref Surg 5: 109-12, 1993.
KR Murthy
Local Anesthesia
5
INTRODUCTION Surgical procedures in ophthalmology are usually performed under local anesthesia. General anesthesia is employed when the procedure is expected to be time consuming or if the patient cooperation is not possible due to young age, mental status or extreme apprehension. Even among regional blocks, there has been a change, towards topical anesthesia, as the technique of surgery has been constantly altered. Knowledge of the principles of mechanism of drug action, as well as the administration techniques, recognition of complications, and their management, is essential for the success of the surgical procedure. Mechanism of Action Local anesthetics must be, water soluble as well as lipid soluble in order to traverse the nerve membrane and bind with the phospholipid membrane that surrounds the inner openings of sodium channels. This impedes the access of sodium, to the axons and results in a reversible blockage of the nerve.1 The pH of the solution has an effect, on the onset, and spread of the anesthetic effect.2 Buffered solutions are less painful during administration.3 The duration of the effect depends, on the length of time, the agent is bound to the nerve membrane. The anesthetic drug and its concentration, and its rate of removal have an effect on the duration of action. It is to be remembered that these agents can be, readily absorbed through mucous membranes, and can cause toxic reactions.4 Hence, a surface anesthetic should never be injected. Cardiorespiratory resuscitation facilities should always be ready at hand. The body can metabolize the drug when toxic reaction
THE ART
52
OF
P HACOEMULSIFICATION
Table 5.1: Local anesthetic agents (topical) Agent
Trade Name
Concentration (percentage)
Benoxinate HCl (Combined with fluorescein sodium 0.25%) Cocaine HCl Proparacaine HCl
Fluress
0.4
Ak-taine Alcaine Ophthaine Ophthetic Anacel Pontocaine
1-4 0.5 0.5 0.5 0.5 0.5 0.5
Tetracaine HCl Onset within 1 minute Duration of action 10 to 20 minutes
(Physician’s Desk Reference for Ophthalmology Oradell, NJ, Medical Economics, 1991)
occurs, but resuscitative measures should be, at hand to maintain the ventilation and circulation till the drug is metabolized.5 Action of the local anesthetic can be enhanced by the addition of epinephrine 1:200,000 in addition it also provides vasoconstriction and reduces bleeding. Caution should be exercised when the patient has systemic hypertension and cardiovascular disease or thyrotoxicosis.6 Addition of hyaluronidase, which depolymerizes hyaluronic acid, results in quicker diffusion of the agent. Mechanical pressure is needed to spread the agent through the tissues effectively.7 Local Anesthetic Agents Topical (Table 5.1) Before starting surgical procedure under local anesthesia one should ensure that an intravenous line is started and emergency drugs are at hand. The patient should have a cardiac monitor and also his or her oxygen and carbon dioxide levels should be monitored. Facilities for oxygen administration and intubation should also be available and it is preferable if an anesthetist is available to manage an emergency or to give additional sedation if the patient is in need of it. Injectable See Table 5.2. TECHNIQUES
OF
LOCAL
ANESTHETIC
INJECTIONS
Orbicularis or facial block to paralyze the orbicularis muscle can be achieved by blocking the trunk of the facial nerve at stylomastoid foramen, (Nadbath block) or at the neck of the mandible (Obrien’s technique) in front of the tragus or by injecting at the orbital margin (van lint technique,/Atkinson block). When the peribulbar block with hyaluronidase is administered there is enough diffusion of the agent to block the orbicularis also and this negates a separate orbicularis block. When the trunk of the nerve is being blocked sometimes there could be
L OCAL A NESTHESIA
53
Table 5.2: Local anesthetic agents (injectable) Agent (Trade name)
Chemical Class
Concentration (percentage)
Maximum Dose (mg)
Relative Potency
Onset of Action (min)
Duration of Action
Procaine (Novacaine)
Ester
1-4
500
1
7-8
30-45 min
Choloroprocaine (Nesacaine)
Ester
1-3
800
1
6-12
60 min
Mepivacaine (Carbocaine)
Amide
1-2
500
2
3-5
120 min
Lidocaine (Xylocaine) (Dalcaine)
Amide
1-2
500
2
4-6
40-60 min
Bupivacaine (Marcaine) (Sensorcaine)
Amide
0.25-0.75
175
8
5-11
4-12 hr
Etidocaine (Duranest)
Amide
1-1.5
400
8
3-5
5-10 hr
(Adapted from Raj PP: Handbook of Regional Anesthesia. Churchill Livingstone: New York, 1985; Physician’s Desk Reference for Ophthalmology Ordell NJ, Medical Economics, 1991; Crandall DC: Pharmacology of ocular anesthetics, In Duane TD, Jaeger EA (Eds): Biomedical Foundations of Ophthalmology, Harper and Row: Philadelphia, 1986)
needle injury to the nerve as it is in a relatively fixed position and cannot roll away from the needle and this could cause prolonged weakness of the facial muscles.8 Some times after the Nadbath technique dysphagia and respiratory distress can occur due to paresis of vagus, glossopharyngeal, and spinal accessory nerves, causing aspiration of oral secretion.9 Retrobulbar Injection The technique places the anesthetic in the retrobulbar space, which contains the sensory; and the motor nerves that supply the eye. Except the superior oblique this block paralyzes the rest of the ocular muscles since the trochlear nerve is outside the muscle cone. This injection is given with an 11/4 inch blunt-tipped Atkinson needle. It should ideally be given by the surgeon himself as he or she is familiar with the anatomy and not relegated to others and it should be given without putting the nerve on stretch and after making sure that the needle is not in a blood vessel by withdrawing the piston before injecting. The patient is asked to look medially and then midway between the lateral limbus and canthus at the lower orbital margin, the needle pierces the skin and is passed posteriorly parallel to the floor till it clears the equator of the globe, then the needle is angled upwards and inwards towards the apex of the orbit and pushed further to enter the muscle cone. The bevel of the needle should face the globe to reduce the chances of piercing the globe. One can often detect the passage through the muscle cone
54
THE ART
OF
P HACOEMULSIFICATION
by noticing a slight flick of the eye and one can also rule out globe perforation by making lateral motion to note whether the tip of the needle is fixed or free. Then about 1 cc or 2 cc of the anesthetic is introduced and observation is made of the upper eyelid, which tends to droop if the injection is in the right place. Slight pressure over the eye for few minutes after the injection helps to distribute the anesthetic and within minutes the akinesia and anesthesia is obtained, the pressure applied also helps in reducing the tension of the eye. Immediately after the injection one should check the opposite eye for any paresis of the ocular muscles in the other eye and also enquire whether there is any amaurosis of the opposite eye as this will give an immediate clue to diffusion of the anesthetic along the subarachnoid space around the optic nerve in a posterior direction, and if observed immediate resuscitative measures should be instituted and oxygen administration can be started. Repeated yawning after a retrobulbar injection can also be sign of relative hypoxia and should alert the surgeon to possible respiratory depression. Peribulbar Injection There has been an increasing change to this technique to avoid the complication of the retrobulbar anesthesia. In this technique the agent is placed either in the anterior or posterior extraconal space and allowed to diffuse into the retrobulbar space posteriorly and also anteriorly to obtain an orbicularis block as well. To achieve this much larger volume of the anesthetic has to be placed and also it is preferable to use a diffusion agent like hyaluronidase and apply pressure for some time. It often takes more time than a retrobulbar injection to achieve the desired result and the incidence of incomplete akinesia and anesthesia is higher.10 Procedure of Posterior Peribulbar Block (DUANE)* With an intravenous line running or heparin lock in place and after desired sedation and eyelid preparation, proceed with the following steps; 1. Make a small skin wheal in the lower eyelid 1 cm medial to the lateral canthus over the inferior orbital rim. 2. Make a small skin wheal in the upper eyelid in the skinfold directly inferior to the supraorbital foramen. 3. Through the inferior skin wheal, with a 27-gauge needle, inject 0.5 ml of lidocaine 1% in the orbicularis and inject 1 ml just deep to the muscle. 4. Through the upper lid skin wheal, repeat as in step three. 5. Through the lower lid skin wheal, inject 1 ml of lidocaine 1 percent bupivacaine 3/4 percent hyaluronidase solution in the orbicularis muscle and 1 ml immediately deep to it. Advance along the floor of the orbit to the equator of the eye; aspirate and inject 1 ml; aiming slightly superomedially, advance the needle to its full depth and inject 1 to 1.5 ml. 6. Pushing the globe inferiorly with a free index finger, enter through the upper lid skin wheal and inject 1 ml ½ inch deep to the orbicularis muscle and slightly nearer the canthus than is the original skin wheal; direct the needle along the orbital roof without engaging the periosteum to the equator, where 1 ml is injected, then to the superior orbital fissure and inject a final 1 ml of solution. * (Adapted from Nugent CC: Peribulbar Anesthesia—A Safe, Simple Effective and Relatively Painless Technique)
L OCAL A NESTHESIA 7. Pressure akinesia if lateral is seen. (Always
55
on the globe/orbit and time are essential for a good block; after 8 minutes, if incomplete remains, 3 to 4 ml of additional anesthetic solution is injected by the lower approach or inferior movement is seen, by the superior approach if superior or medial movement perform lower lid injections before upper lid).
Complications • Retrobulbar hemorrhage sometimes may result (1 to 3%)11,12 due to injury to a blood vessel and this is recognized by immediate tightening of the lids and proptosis of the globe and some times the subconjunctival appearance of the hemorrhage. Pressure over the eyeball for a few minutes may arrest a minor bleed and may allow the surgery to be undertaken. However, it is always prudent to postpone the surgery for a week and take up the case later. On most occasion no permanent damage is done to the eye. But there has been mention of optic atrophy after such an event and some times one may get occlusion of the central retinal artery. • Perforation of the globe (0.075%)13 is a serious complication which can occur with both the retrobulbar as well as peribulbar blocks and this is more often seen when a sharp disposable needle is used and its frequency is increased if the injection is made by a person not very familiar with the anatomical knowledge of the orbit. It produces immediate rise of intraocular pressure (IOP) and may also cause sharp pain. Even double perforation of the globe has been noticed when such sharp needles are used and the complication has come to be recognized only after the surgery when the retinal examination has been done. • Allergic reactions to the local anesthetic are relatively rare. Many of the socalled allergic reactions are often instances of toxicity. After routine testing for allergy to local anesthetic for fifteen years in all patients undergoing ocular surgery there was no instance where an allergic reaction was noted and hence this practice was stopped. But in a patient with known history of allergies and hypersensitivity reactions one can test for evidence for hypersensitivity taking care that emergency drugs and resuscitative measures are available. • Convulsions, tremors and confusion and other central nervous system manifestations can occur especially when inadvertent intravascular injection occurs. Alternative Methods (Table 5.3) Any new LA technique should be as safe, effective and acceptable as its predecessors. Recently, subconjunctival, sub-Tenon’s (parabulbar), and purely topical anesthesia, sometimes with additional intracameral lignocaine have become popular. All these methods of anesthesia are useful and effective if the patient assessment and selection are good and if the surgery is not prolonged. These also necessitate more interaction between the surgeon and the patient and make the surgeon alert and be aware of the patient cooperation and comfort. Patients with mental retardation, deafness, children and where communication between the operator and the subject is not possible are not suitable for such
THE ART
56
OF
P HACOEMULSIFICATION Table 5.3: Alternative methods
Local anesthesia Advantages
Disadvantages
Topical
Alert patient No retrobulbar hemorrhage, No optic nerve injury No diplopia, ptosis No globe rupture Functional vision maintained
No akinesia Inadequate anesthesia Distraction by patient
Sub-Tenon’s
Less painful No retrobulbar hemorrhage No optic nerve injury No IOP increase No globe perforation Low dose, low volume
The anesthetic has to be placed In sub-Tenon’s place correctly Absence of akinesia
Retrobulbar
Reliable Quick Less volume Loss of vision helps in patients Who do not want to see anything of the procedure
Optic nerve injury Retrobulbar hemorrhage Globe perforation Systemic complications
Peribulbar
No optic nerve injury Less chance of retrobulbar hemorrhage All advantages of retrobulbar
All disadvantages of retrobulbar less frequent Akinesia and anesthesia may be incomplete Longer to act Chemosis Ptosis Slower recovery Expensive More risky when systemic diseases of cardiopulmonary nature exist
Comfortable patient Ideal operating condition Method of choice in difficult cases No local anesthetic complication No residual paralysis bilateral surgery Better for teaching
techniques. But these techniques do abolish some of the complications associated with the retro- and peribulbar techniques. Topical anesthesia is achieved by instilling 4 percent lidocaine into the conjunctival cul-de-sac of the eye to be operated at five minute interval for about 15 minutes. A sponge soaked in the local anesthetic can be placed at the limbus so that the area can get the maximum contact with the agent. REFERENCES 1. Dejong RH: Neural blockade by local anesthetics. JAMA 238: 1383, 1977. 2. Zahl K, Jordan A, McGroarty J et al: Peribulbar anesthesia—effect of bicarbonate on mixtures of lidocaine, bupivacaine, and hyaluronidase with or without epinephrine. Ophthalmology 98: 239, 1991. 3. Eccarius SG, Gordon ME, Parelman JJ: Bicarbonate-buffered lidocaine-epinephrine-hyaluronidse for eyelid anesthesia. Ophthalmology 97: 1499, 1990.
L OCAL A NESTHESIA 4. 5. 6. 7. 8. 9. 10. 11. 12. 13.
57
Goodman LS, Gilman A: The Pharmacological Basis of Therapeutics (4th ed), 1970. Mauger TF, Craig EL: Havener’s Ocular Pharmacology (6th ed), Mosby Year Book: St. Louis, 1994. Mauger TF, Craig EL: Havener’s Ocular Pharmacology (6th ed), Mosby Year Book: St. Louis, 1994. Mauger TF, Craig EL: Havener’s Ocular Pharmacology (6th ed), Mosby Year Book: St. Louis, 1994. Atkinson WS: Facial nerve block. Am J Ophthalmology 57: 144, 1964. Koenig SB, Snyder RW, Jonathan K: Respiratory distress after a Nadbath block. Ophthalmology 95: 1285, 1988. Mauger TF, Craig EL: Havener’s Ocular Pharmacology (6th ed), Mosby Year Book: St. Louis, 1994. Linn Jr JG, Smith RB: Intraoperative complications and their management. Int Ophthalmol Clin 13: 149, 1973. Cionni RJ, Osher RH: Retrobulbar hemorrhage. Ophthalmology 98: 1153, 1991. Zaturansky B, Hyams S: Perforation of the globe during the injection of local anesthesia. Ophthalmic Surg 18:585, 1987.
58
THE ART
OF
P HACOEMULSIFICATION
Samuel Masket
Ocular Anesthesia for Small-Incision Cataract Surgery
6
INTRODUCTION Traditional methods of local ocular anesthesia for cataract surgery have employed injection of anesthetics to the periorbital region. It is well recognized that regional infiltration can produce ocular anesthesia, ocular akinesia, orbicularis akinesia, and varying degrees of amaurosis. However, recent trends strongly indicate that only ocular anesthesia is necessary for routine small-incision cataract surgery. The 1997 American Society of Cataract and Refractive Surgery (ASCRS) members survey for the year 1996 suggests that roughly 15 percent of surgeons employ non-injection anesthesia (topical with or without intracameral agents) routinely.1 Nevertheless, the great majority of surgeons continue to use anesthetic injections with some degree of risks, that include damage to the globe, optic nerve, and periocular structures, and central nervous system (CNS) toxicity including brainstem anesthesia, apnea, and death. Very rarely, the patient may sustain bilateral ocular anesthesia as a result of anesthetic spread through the cavernous sinus. Moreover, with anesthetic injection, there is the potential for cosmetic blemish of the lids and conjunctiva. It is worth noting that patients often rate the quality of their cataract surgery by how the eye looks as well as how the eye sees during the early postoperative period. The risks of periorbital anesthetic injections are of some consequence, in that the overall occurrence rate for retrobulbar hemorrhage is in the vicinity of 1 percent of all cases; the likelihood increases with long needles and intraconal injection.2 Furthermore, ocular penetration and optic nerve damage are not terribly rare. The risks of those maloccurrences increase in patients who are uncooperative for injection, those with high myopia, those with prior scleral buckling surgery, and when the injections are administered by non-ophthalmologists. Additionally, no needle types,
OCULAR ANESTHESIA
FOR
S MALL-I NCISION CATARACT S URGERY
59
injection sites, or injection styles are immune to the risk for damage to the globe or other orbital structures.3-5 Another issue regarding the blind passage of sharp needles into the orbit concerns those patients on anticoagulant medications or those with naturally occurring coagulopathies. It should be obvious that these patients are at greater risk for periocular hemorrhage with needle injection, but often the medical necessity for anticoagulation dictates that patients remain on treatment during the perioperative period. Often, the systemic risk to cessation of anticoagulant treatment is greater than the risk of intraoperative bleeding. Indeed, the published guidelines for cataract surgery in the United Kingdom suggest that cataract surgery should proceed up to an INR (International Normalized Ratio) of 4.0 for patients taking Coumadin. It is evident that non-injection forms of local anesthesia are safer for anticoagulated patients. Additional consequences of periocular anesthetic include an inability of the patient to move the eye during and after surgery. While it was once considered essential that the eye be fully still for safe surgery, it is now recognized that purposeful eye movements, on command, can benefit the progress of surgery. As an example, in cases with narrow palpebral fissures, the eye can be moved to facilitate incisions, etc. A further consequence of regional anesthetic infiltration is amaurosis. As a result, the patient cannot see to fixate a target. However, with topical/intracameral anesthesia, the patient can be asked to follow a light source or other visual target to help fixate the globe in a satisfactory position for surgery. Movement away from periocular injection toward topical methods of ocular anesthesia is natural, given the overall changes in small-incision cataract surgery that have progressed to outpatient surgery with methods that allow for immediate ambulation, rapid return to a full lifestyle, and stable optical results of surgery within days.6 The immediate use of the eye after cataract surgery is possible only with topical or intracameral methods and is in keeping with the concepts of modern surgery. Topical anesthesia resurfaced in this decade as a useful tool after Fichman’s suggestion regarding the use of tetracaine 0.5 percent applied to the eye as the only anesthetic for cataract surgery.7 Other agents, such as bupivacaine and lidocaine have been popularized because of a reduced tendency to cause corneal epitheliopathy and to have a longer period of action as compared with tetracaine. However, patients are not universally comfortable with topical anesthesia as the only agent. Many surgeons employ small amounts of intravenous, oral, or sublingual sedation as an adjunct. However, in 1995 Gills suggested the routine use of intracameral nonpreserved lidocaine in addition to topical anesthesia with or without systemic sedation,8 although the concept had been mentioned earlier by Fichman who considered intraocular tetracaine for use in difficult case situations. Safety and efficacy of intracameral lidocaine has been further established by Koch9 and Masket with Gokmen in separate studies.10 In the latter investigation, approximately 40 percent of greater than 300 patients receiving only topical anesthesia required intraoperative conversion to a deeper level of local anesthesia, whereas fewer than 1 percent of 300 cases receiving intracameral lidocaine had need for an additional local anesthetic
60
THE ART
OF
P HACOEMULSIFICATION
method. In the same study safety was measured by comparing the degree of corneal edema on the first postoperative day between the two groups, a reduced likelihood for corneal edema was associated with the use of intracameral nonpreserved lidocaine hydrochloride 1 percent, but this finding may be related to the use of chop style phacoemulsification for the latter group. Nevertheless, based upon the early postoperative appearance of the cornea, non-preserved lidocaine is seemingly non-toxic although Koch reports reduced contrast sensitivity Fig. 6.1: Blunted reusable cannula for subTenon’s (parabulbar) anesthesia (Courtesy and visual acuity in the first few hours Rhein Medical, Tampa, Florida) after surgery. Other methods to provide ocular anesthesia for cataract surgery without the risks of blind pass, sharp needle orbital injection have evolved during the same era as the movement to topical anesthesia. Posterior sub-Tenon’s infiltration employs a blunt cannula (Fig. 6.1) to place local anesthesia directly in the retrobulbar space. A conjunctival button hole incision, performed under topical anesthesia, is necessary for the cannula to gain direct access to the sub-Tenon’s space. This method was suggested as an alternative to sharp needle orbital injection,11 and has been further popularized by Greenbaum as a primary method for cataract anesthesia; he coined the term “parabulbar” anesthesia to describe the concept.12 Additionally, the method may be used for surgeons in transition to topical/intracameral anesthesia and is very useful to convert from topical methods in cases where complications occur, surgery is prolonged, or if the patient is otherwise in need of a deeper level of anesthesia. As long as the cataract incision is self-sealing, the parabulbar infiltration may be given at any time during the surgery. Varying with the nature of the agent used for infiltration, parabulbar anesthesia may provide complete ocular akinesia and amaurosis. Other alternatives include anterior subconjunctival injection given diffusely or only focally in the region of the incision, so-called “pin-point” anesthesia.13 It is evident that traditional ocular anesthesia for cataract surgery, utilizing sharp needles passed blindly through the skin of the lids or the conjunctiva engenders risks (Table 6.1) that are avoidable with topical or intracameral or other recently developed means for local anesthesia. However, in addition to the greater safety associated with newer anesthetic systems, topical and topical or intracameral methods avoid the need for patching and allow patient the use of the eye immediately following surgery in the overwhelming majority of cases. Advantages, therefore, include safety, improved cosmesis, ability to use the eye immediately following surgery, and the ability to move and fixate the eye during surgery in response to the surgeon as an aid to the procedure (Table 6.2). Varying with the experience of the surgeon, certain conditions may contraindicate the use of topical/intracameral anesthesia (Table 6.3). Given the ability to move
OCULAR ANESTHESIA
FOR
S MALL-I NCISION CATARACT S URGERY
61
Table 6.1: Risks of injection anesthesia _______________________________________________________________________________________________________________________________________________________________________________ • Damage to optic nerve • Retrobulbar hemorrhage • Ocular penetration/perforation • Central nervous system anesthesia • Apnea • Unintended bilateral ocular anesthesia • Damage to extraocular muscles/diplopia • Esthetic blemish _______________________________________________________________________________________________________________________________________________________________________________ Table 6.2: Advantages of topical/intracameral anesthesia _____________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________ • Avoids pain, blemish and risk of injection anesthesia • Allows immediate useful vision after surgery • Eliminates need for patch after surgery • Reduces anxiety and/or heavy sedation associated with injection anesthesia • Compatible for patients on anticoagulants • Patients can aid surgeon by moving eye for favorable exposure _____________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________ Table 6.3: Contraindications to topical/intracameral anethesia ____________________________________________________________________________________________________________________________________________________________________________________ A. Relative 1. Language barrier 2. Anticipated difficult surgery 3. Poorly cooperative patient B. Absolute 1. Total deafness 2. Coarse nystagmus ____________________________________________________________________________________________________________________________________________________________________________________
the eye, the patient can aid in the surgery or create significant obstacles; cataract surgery under topical or intracameral anesthesia is, by necessity, interactive. Poor patient cooperation is a relative contraindication, as is the inability of the surgeon and patient to adequately communicate in the same language. Often, an interpreter or bilingual family member can be present in the operating theater in order to facilitate surgery without need for injection anesthesia. However, absolute congenital deafness with speaking difficulty is an absolute contraindication, since the patient may become disoriented under the surgical drapes and cannot be expected to communicate by the usual means of lipreading or sign language; patients of this nature often require general anesthesia. Ocular conditions may also act as relative or absolute contraindications; cataracts too dense to allow fixation on the microscope light, potentially complicated surgery (preoperative zonulysis, etc.), and nystagmus are common examples. Nevertheless, the huge majority of patients may safely experience small-incision cataract surgery under topical/intracameral anesthetic with very limited sedation. Methods The author prefers the use of lidocaine HCl 4.0 percent non-preserved for topical anesthetic. It is long acting and non-mucogenic (Previous experience with 0.75 percent
62
THE ART
OF
P HACOEMULSIFICATION
bupivacaine HCl suggests that it causes undesired mucous production.) Intracameral anesthesia is achieved with unpreserved lidocaine HCl 1.0 percent. Some surgeons advocate diluting the intracameral agent with BSS solution in order to raise the pH and reduce the mild discomfort associated with anterior chamber instillation. • Administer topical proparacaine HCl 0.4 percent to initiate anesthesia with little sting. Administer dilating agents (cyclopentolate or tropicamide and phenylephrine 2.5%), topical antibiotics, a topical NSAID, and lidocaine HCl 4.0 percent four times at five minute intervals prior to surgery. • After the patient is brought into the theater several drops of the 4.0 percent lidocaine are administered prior to the sterile “prep”. The latter begins with instillation of two drops of half strength Betadine solution (not Betadine scrub) directly to the operative eye. At this time very small amounts of intravenous sedation may be given, depending upon the mental and medical status of the patient, the anxiety of the surgeon, and the observations of the anesthetist or equivalent. The author generally asks that 0.5 mg to 1.0 mg of midazolam HCl be administered IV. • During the draping process communicate with the patients about the operative process. Tell them that they will feel slight pressure from the lid speculum and that they will need to fixate on the light of the microscope. Tell them that requests to look up, down, etc. should be achieved by moving the eye and not the head. Reassure them that they will feel no pain. • Begin surgery with the microscope light at low levels of illumination, sufficient only to perform a paracentesis. Place 0.2 cc of non-preserved lidocaine HCl in the anterior chamber and follow that with the viscoelastic of choice. The anesthetic will be washed out as the viscoagent fills the chamber if the lip of the sideport is depressed as the viscoelastic is injected. Slowly increase the microscope light and perform the clear-corneal incision. Continue with routine surgical procedure. • Generally, no further anesthesia is necessary. However, in situations with prolonged surgery or very sensitive patients, additional intracameral anesthetic may be administered for complaints of “pressure“ or intraocular pain. For surface discomfort the conjunctiva may be swabbed with a pledget of any sterile topical anesthetic, but care should be taken to avoid placing the agent near or in the incision if it contains preservatives. Additionally, small increments of intravenous sedation can be added as may be (rarely) necessary. • In the very unlikely case that the patient cannot tolerate the microscope light even at low illumination and continues to squeeze the lids against the speculum, additional doses of IV medicine could be given until the intracameral anesthetic is administered. In the author’s observations, once the eye has received the intracameral agent, all lid squeezing and signs of anxiety or discomfort abate rapidly. However, in extreme situations or should an operative complication occur that will significantly prolong surgery, one can stop surgery, given a self-sealing incision, pressurize the eye to normal, and administer deep sub-Tenon’s local anesthetic with a blunt cannula (Fig. 6.1) through a conjunctival buttonhole entry in the superior or inferior nasal quadrant.
OCULAR ANESTHESIA
FOR
S MALL-I NCISION CATARACT S URGERY
63
Prior to incising the conjunctiva, a pledget of local anesthesia may be placed on the area for a few moments. The cannula should reach the retrobulbar space with ease if the buttonhole opening includes Tenon’s capsule. Only 2.0 cc of local agent is necessary, given direct access to the muscle cone. Relative amaurosis will be achieved in a matter of seconds, and, with strong local agents, akinesia can be established in a few minutes. REFERENCES 1. Leaming DV: Practice styles and preferences of ASCRS members—1996 survey. J Cataract Refract Surg 23:527-35, 1997. 2. Cionni R, Osher R: Retrobulbar hemorrhage. Ophthalmology 98:1153-55, 1991. 3. Duker JS, Belmont, JB, Benson WE et al: Inadvertent globe perforation during retrobulbar and peribulbar anesthesia. Ophthalmology 98:519-26, 1997. 4. Hay A, Flynn Jr HW, Hoffmann JI et al:. Needle penetration of the globe during retrobulbar and peribulbar injections. Ophthalmology 98:1017-24,1991. 5. Grizzard WS, Kirk NM, Pavan PR et al: Perforating ocular injuries caused by anesthesia personnel. Ophthalmology 98:1011-16, 1991. 6. Masket S, Tennen DG: Astigmatic stabilization of 3.0 mm temporal clear corneal cataract incisions. J Cataract Refract Surg, 22(10):1451-55, 1996. 7. Fine IH, Fichman RA, Grabow HR: Clear-corneal Cataract Surgery and Topical Anesthesia. Slack Inc: Thorofare, 1993. 8. Gills JP, Cherchio M, Raanan MG: Unpreserved lidocaine to control discomfort during cataract surgery using topical anesthesia. J Cataract Refract Surg 23:545-50, 1997. 9. Koch PS: Anterior chamber irrigation with unpreserved lidocaine 1% for anesthesia during cataract surgery. J Cataract Refract Surg 23:551-54, 1997. 10. Masket S, Gokmen F: Efficacy and apparent safety of intracameral lidocaine as a supplement to topical anesthesia. J Cataract Refract Surg (24)7:956-60, 1997. 11. Stevens JD : A new local anaesthesia technique for cataract extraction by one quadrant sub-Tenon’s infiltration. Br J Ophthalmol 76:670, 1992. 12. Greenbaum S: Anesthesia in cataract surgery. In Greenbaum S (Ed): Ocular Anesthesia WB Saunders: Philadelphia 1-55, 1997. 13. Fukasaku H, Marron JA: Pinpoint anesthesia—a new approach to local ocular anesthesia. J Cataract Refract Surg 20:468, 1994.
64
THE ART
OF
P HACOEMULSIFICATION
Shashi Kapoor
The limbal Incision
7
INTRODUCTION Kratz is credited as the first surgeon, to move from the limbus posteriorly to the sclera, increasing appositional surfaces to enhance wound healing and attempt to exert less traction on the cornea, thereby controlling surgically induced astigmatism. Girard and Hoffmann were the first to call the posterior incision a “scleral tunnel” incision. In 1989, McFarland and Ernst introduced an incision architecture that allowed the phacoemulsification and implantation of lenses without the need for suturing. Besides lengthening the scleral tunnel, this incision terminated in a decidedly corneal entrance and the posterior lip of the incision, the so-called corneal lip, acted as one-way valve imparting to this incision its self-sealing characteristics. Paul Koch described what he called the `incision funnel’ indicating that there were certain characteristics of self-sealing incisions with respect to length and configuration that imparted not only self-sealability but also astigmatism neutrality to these incisions. The advent of the foldable IOL and advances in device technologies have led to significant refinements in small incision surgery and wound closure that are revolutionizing the practice of cataract surgery. These clinical advances are leading to reductions in iatrogenic ocular trauma and postoperative astigmatism, decreased risk of hyphema, delayed filtering blebs and iris prolapse, increased wound stability, and faster recovery and vision restoration. Posterior Limbal Incision Ernst et al are advocating the use of the temporal posterior limbal incision, which can be made nearly square for foldable IOLs (3.2 mm × 3.0 mm). In a cadaver
T HE LIMBAL INCISION
65
Fig. 7.1: Square corneal wound, 3.5 × 3.5 mm, is impractical since it impinges on the visual axis, and causes disturbing striae, intraoperatively
Fig. 7.2: Square corneal wound, 2 × 2 mm (1.5 mm being the minimal internal corneal lip) is impractical, as it is too small to permit the insertion of current phaco tips and IOLs
eye studies, as rectangular clear corneal incisions were made more square by only 0.5 mm from 3.2 mm × 2.0 mm to 3.0 mm × 2.5 mm, they became more resistant to pressure. Once the length was increased so that the wounds were square (3.2 mm × 3.2 mm) with a 1.5 mm internal corneal lip, they were capable of withstanding maximal external pressures of 525 psi. As yet, square clear corneal incisions are clinically impractical, because they either encroach upon the visual axis (Fig. 7.1) or are too small to permit the insertion of current phaco tips and IOLs (Fig. 7.2). The limbal incision has several advantages over the currently advocated clear corneal incision. Clear corneal incisions, especially those that have a vertical component, are more subject to foreign body sensation than limbal incisions. This irritation occurs because there is edema of both the anterior and posterior aspects of the incision in clear corneal incisions with vertical components, which creates a gape in the most superficial aspect of the incision (Figs 7.3A and B). In turn, this gape forms a ridge that can irritate the patient. Dr Tipperman et al, performed a feline study evaluating limbal versus clear corneal incisions. The incision dimensions were non-square, 3.00 mm in width and 1.75 mm in length. A paracentesis type incision was made in each case. Reformation of the anterior chamber and stromal hydration were not performed. The eyes were observed over a period of 16 days. Pinpoint pressure was applied to each eye at various time intervals. The results showed a statistically significant
66
THE ART
OF
P HACOEMULSIFICATION
Fig. 7.3A: Schematic drawing of a clear corneal incision postphacoemulsification showing corneal edema and ridge
Fig. 7.3B: Schematic diagram of a limbal incision postphacoemulsification showing edema of the corneal tunnel, but not edema of the incision
difference in resistance to pinpoint deformation pressure between the limbal and paracentesis incision (P = 0.0034) (Fig. 7.4). Incision Design The design of modern sutureless, self-sealing cataract incisions, involves the consideration of three design parameters, site, size and shape. Incision Shape Shape is an anatomic parameter in the geometric consideration of cataract incision construction. There are two aspect views of these incisions: sagittal and anteroposterior; and three components: the external incisions, the intratissue tunnel and the internal incision. From the sagittal aspect, limbal incisions may be made in one of following configurations varying, between single-plane, grooved beveled and triplane with a groove and a bevel. The external component of limbal incisions also may be in one of the following theoretical configurations: the single-step “stab” incision as initially introduced by Howard Fine; the two-step grooved incision by Charles Williamson, who felt that the incision should be larger on the outside than the inside, which would allow the injector to fit more easily in the incision without stretching the sclera.
T HE LIMBAL INCISION
67
Fig. 7.4: Graph comparing resistance deformation pressure of clear corneal (CC) and limbal incisions (LB) over time
The ultimate goal is to have no tissue stretch. The trapezoid incision that Dr Williamson describes are about 0.5 mm wider on the outside than the inside. Langerman Hinge and Gills Modification Langerman and Gills have modified clear corneal incisions, and these modifications are applicable to limbal incisions. Although clear corneal incisions appear to be watertight at the end of surgery, Langerman and Ernst have shown that both the straight-in and two-plane clear corneals can leak. This ability to leak indicates a less than perfect seal. John has shown a higher incidence of sterile endophthalmitis in clear corneals compared to scleral tunnels. This increased incidence is probably due to reflux of flora into the anterior chamber through the incision. The lack of culture positive infection can be attributed to John’s aggressive prophylactic antibiotic regimen, after the Gills method. Langerman has developed a hinge technique which involves making the initial vertical groove deeper than the point at which the horizontal shelf is started (Fig. 7.5). The deep groove then functions as a hinge, with the inner aspect of the incision flexible and malleable, allowing it to adhere to the outer portion of the incision.
68
THE ART
OF
P HACOEMULSIFICATION
The Gills modification of the Langerman hinge involves a vertical incision that is 85% deep. The next step is to create a deep longitudinal bevel in the stroma in either a single-or twoplane fashion. This portion is deeper and longer than Langerman describes. Because the inner flap is very thin, it allows for greater flexibility and better contact when pressure is applied to any portion of the wound. Fig. 7.5: Schematic drawing demonstrating The key to creating a very deep, thin wound construction incision is to pick up the distal aspect of the limbal wound with forceps. This step is essential for two reasons: (i) it allows the surgeon to have control over fixation of the eye when using topical anesthesia, and (ii) it allows the surgeon to make a deep incision with greater ease. The flap is created by directing the keratome (diamond or metal) toward the dome of the eye. After tunneling through the posterior stroma for a distance of approximately two-thirds the keratome width, the incision is beveled posteriorly through Descemet’s membrane and into the anterior chamber. Using a 2.3 mm to 2.5 mm trapezoidal keratome and beveling it into the anterior chamber approximately 2 mm will create a 2.3 mm to 2.5 mm by 2.0 mm incision. The incision performs similarly to navigational valves, when pressure against the rubber forces the aperture closed. Grooved incisions provide a superficial corneal flap that has a thicker edge, which helps to avoid avulsion when grasping with forceps or when suturing. The deep groove of Langerman is believed to physically separate the sclera from the internal corneal flap sufficiently to allow sclera indentation (patient rubbing) without internal wound separation, thereby reducing the potential for wound leak. The sagittal shape and direction of the tunnel may also vary. Most tunnels are made flat or planar by flat blades advanced in a single plane. Some surgeons believe there may be superior sealability with a convex sagittal curve to the tunnel although it is technically difficult to create and to reproduce. The third component of these incisions, the internal opening, may have one of two sagittal shapes: single-plane “stabs” in the same plane as the limbocorneal tunnel (Fig. 7.6), or bi-planar “steeped”. In addition, Singer proposes that the most Fig. 7.6: Limbocorneal tunnel incision accurate anatomic angle of entry into
T HE LIMBAL INCISION
69
Figs 7.7A to C: Anteroposterior aspect of limbaltunnel (Grabow): (A) convergent trapezoid, (B) parallelogram, and (C) divergent trapezoid
the anterior chamber through the endothelial surface is 17.5 degrees, that which duplicates the average angle of perforation of scleral veins. The anteroposterior aspect of the second component of limbal incisions, the tissue tunnel, is summarized in Figures 7.7A to C. The original configuration was parallelogram introduced by Fine. The second configuration, was a convergent trapezoid, introduced by Williamson. In this design, the external incision is longer than the internal incision, to improve instrument maneuverability while preserving the smallest internal opening for self-sealability. The third configuration, the divergent trapezoid of Hoffer, has its external opening smaller than its internal opening, to reduce the astigmatic effect, while allowing maneuverability internally. The anteroposterior aspect of the third incisional component, the internal opening has only recently come under investigation. Three directional possibilities exist for this incision (Figs 7.8A to C): limbus-parallel, tangential and limbus-antiparallel (“corneal frown”). Limbus-parallel internal incisions, unlike limbus parallel external incisions, have been demonstrated to be the most stable and most resistant to deformation and leakage. Conversely, corneal frown internal incisions are the weakest. Therefore, cutting strokes to enlarge an internal incision should move either in a central direction, by advancing the blade, or in a lateral direction, so as to preserve a limbus-parallel configuration, rather than in a peripheral direction, as when withdrawing the blade.
70
THE ART
OF
P HACOEMULSIFICATION
Figs 7.8A to C: Anteroposterior aspect of internal corneal openings (Grabow): (A) limbus parallel, (B) tangential, and (C) limbusantiparallel “corneal frown”
Site One classification of location of cataract incisions, involves the radial distance from the optic axis. These locations for cataract incisions have traditionally been designated by their anatomic locations; the sclera, the cornea, and the junction of these two—the limbus. Scleral Incisions Scleral incisions, being furthest from the optic axis, when compared to corneal and limbal incisions of equal dimension, have less effect on both corneal astigmatism and the corneal endothelium, and the same is true for peripheral corneal incision as compared to more central corneal incisions. The external entry incision for small, self-sealing sclerocorneal incisions, as originally described was made 3 to 4 mm posterior to the limbus. The distance from the limbus has become progressively less as surgeons have become more conversant with creating the self-sealing internal corneal lip. Now most surgeons are comfortable about 1.5 mm posterior to the limbus, giving a total sclerocorneal tunnel length of approximately 2.5 to 3.0 mm. In 1972, Fine introduced the clear corneal incision, which was originally described as a stab incision temporally located in avascular tissue. Since then, many modifications and variations have been developed, encompassing a wide variety of sutureless incisions.
T HE LIMBAL INCISION
71
Limbal Incision Anatomists describe the anterior vascular arcade as extending 0.5 mm into clear cornea, anterior to the limbus with its external surface covered by conjunctiva. Thus the limbus can be defined as the anterior edge of the conjunctival insertion, with the corneal vascular arcade considered to be in clear cornea. Architecturally, limbal incisions can be described as follows. a. Single plane (without a vertical groove at the external edge of the incision) b. Shallow groove (having a groove perpendicular to the corneal surface at the external edge of the incision, upto 400 µ deep) c. Deeply grooved (where the groove perpendicular to the external edge of the incision is greater than 400 µm). In another classification, the cataract incision is usually described as located at one of four possible sites on the corneal circle superiorly, obliquely, temporally or on steep axis. Superiorly located incisions when not under the influence of sutures, are known to have an against-the-rule (ART) astigmatic effect, greater with longer incisions, due to the effects of gravity and/or lid closure on wound gape. The oblique location, whether nasal or temporal, is espoused by some who prefer this site not only for ergonomic advantage, but also for greater wound stability when compared to superior incisions. These wounds are believed by some to be the only ones not affected by transmitted tractional forces of a rectus muscle, a theory yet to be documented. Superolateral Incisions It has been reported that a lateral or superolateral incision, can decrease and quickly stabilize surgically induced astigmatism (SIA), although the reason for this is unclear. Hayashi et al compared superior (group I) with superolateral incisions (Group 2) in IOL patients who underwent a 6.5 mm beveled incision, 0.5 mm posterior to the surgical limbus, followed by endocapsular phacoemulsification, enlargement of incision to 6.5 mm and insertion of a posterior chamber IOL with a 6.0 mm optic. The wound was closed by 3 interrupted 10-0 nylon sutures. The patients were followed up using keratometer and computerized corneal topography. Figure 7.9 shows the induced changes in the mean keratometric cylinder at each interval, in the two groups. Group 1 has significantly less induced astigmatism throughout the six month follow-up than group 2. The patients in group 1 also had less standard deviation throughout the follow-up than the patients in group 2. This indicates that SIA in group 1 was less variable than in group 2. Two types of corneal topographic maps were used to illustrate the postoperative corneal shape alterations: color-coded maps “averaging” all corneas at each interval and absolute scale maps of individual cases. The averaged maps of Group 1 patients show a slight steepening in the central cornea in the 10 O’clock meridian, one week after surgery. There were no remarkable changes in the peripheral cornea. The surgically induced steepening in the central cornea disappeared within one
72
THE ART
OF
P HACOEMULSIFICATION
Fig. 7.9: (Hayashi) Mean (standard deviation) induced keratometric cylinder following superolateral incision and superior incision surgery using the vector analysis method. The mean induced keratometric cylinders in group 1 were significantly less than those in group 2 throughout the six month follow-up. The standard deviations in group 1 were also smaller than those in group 2
month after surgery. Consequently, in group 1, the averaged maps of the corneas, after 1 month were virtually the same as the map of preoperative corneas. The averaged maps of group 2 patients show a marked steepening of the upper and lower corneas, in the vertical meridian, one week after surgery. This induced steepening gradually decreased but remained upto three months. The averaged map of the cornea at “six“ months after surgery, was almost the same as the preoperative map (as compared to one month, group 1). Absolute scale maps of a patient in Group 1 reveal remarkable changes in the corneal shape after surgery. There was prominent corneal steepening in the 10 O’clock meridian and the map shows an asymmetric horizontal bow-tie configuration one week after surgery. This steepening quickly disappeared; at three and six months after surgery the cornea had almost recovered it’s preoperative shape, although a focal steepening in the horizontal meridian remained (Fig. 7.10). In group 1 19.7 percent of corneas showed this prominent steepening, just after surgery. Figure 7.11 shows a patient from group 2. Marked steepening of the upper and lower corneas occurred at one week and is shown as a vertical bow-tie configuration. The steepening did not change substantially during the six-month follow-up. In group 2, 68.4 percent of corneas, showed this marked corneal steepening. Thus, superolateral incision cataract surgery induced minimum changes in the corneal shape, as well as SIA, and the changes decreased quickly and stabilized in the early postoperative period. One cannot explain why the induced corneal shape changes are less with the superolateral approach. The differences between
T HE LIMBAL INCISION
73
Fig. 7.10: (Hayashi) Absolute scale maps of a cornea in the superolateral incision group: (A) Preoperatively, and (B) at 3 months postoperatively. After 3 months the map was almost same as preoperatively
Fig. 7.11: (Hayashi) Absolute scale maps of a cornea in the superior incision group: (A) Preoperatively and (B) at 3 months postoperatively. At three months the steepening in the upper and lower corneas, did not change substantially
the semilateral and superior incision surgeries were only related to the wound site. There are a few anatomical differences between the horizontal and the superior limbus. Alternatively, some unknown physiologic and/or optical factor may exist. The rapid stabilization with the superolateral incision appeared to be rather easy to explain. Continuous stroking of the upper eyelid causes pressure on the superior corneoscleral wound. Therefore corneal distortion by the superior wound persists longer. A superolateral incision is relatively free from such eyelid pressure and corneal shape changes stabilize faster. Thus, superolateral incision surgery appears to have advantages, although the exact mechanism remains unproven. The temporal location has been shown to be the most astigmatically stable of these three locations, achieving stability almost immediately and maintaining stability for life or until further astigmatic interventions. First, the temporal location is furthest from the visual axis, and so there will be less impact on the corneal curvature at the visual axis. Second the wound is parallel to the effects of both lid blink and gravity. The temporal location has an added advantage in that it provides the easiest anatomic access to the surgical site, being unimpeded by the bony orbital rim.
THE ART
74
OF
P HACOEMULSIFICATION
Fig. 7.12: Clear corneal incision, 1 week postsurgery showing no wound healing
Fig. 7.13: Limbal incision, 1 week postsurgery demonstrating significant fibroblastic activity completely sealing the incision
Fig. 7.14: Two-month postsurger y, clearcorneal incision demonstrating the incision is healed through binding of the stromal keratocytes as seen in corneal transplant surgery
LIMBUS
vs
CLEAR
CORNEA
In their fourth cadaver study of clear corneal incisions Ernst et al reported that when clear corneal incisions of exactly the same dimensions were made at the limbus and anterior to the limbus, the one nearer the limbus demonstrated greater
T HE LIMBAL INCISION
75
Fig. 7.15: Two-month postsurgery, limbal incision, showing the same fibroblastic activity and the same degree of healing as 1 week postsurgery
wound stability. The reason of this is not completely clear. It may be that the architecture of the limbus, with its circumferential fibers allows greater resistance than the radial fibers of the cornea. The limbus also contains more elastic fibers than the cornea, which has none. Histologic evaluations at 1 week showed no healing activity of the clear-corneal incision (Fig. 7.12). The limbal incision, whether it was placed anteriorly or posteriorly, showed a significant fibroblastic response causing the wound to seal (Fig. 7.13). Both limbal and corneal incisions heal by fibroblast response. The key is the timing of the healing process: 7 days for vascular origin (limbal incisions) and 60 days for avascular origin (corneal incisions). Limbal incisions heal by an influx of fibroblasts in the vascular arcade or through the differentiation of stem cells into fibroblasts. They heal faster than corneal incisions, which heal in a way similar to that of corneal transplants, in which binding stromal keratocytes transform into fibroblasts (Figs 7.14 and 7.15). The advantages of moving to a temporal approach and more posterior into a vascular region (limbal incisions) allow an earlier fibroblast response that seals incisions faster than the 60 days required for incisions made in an avascular region (corneal incisions).
76
T HE ART
OF
PHACOEMULSIFICATION
Amar Agarwal Athiya Agarwal Sunita Agarwal
No Anesthesia Cataract Surgery
8
INTRODUCTION On June 13th, 1998 at Ahmedabad, India the authors (Amar Agarwal) did the first No anesthesia cataract surgery at the Phaco and Refractive Surgery Conference. This was performed as a live surgery in front of 250 delegates. This has opened up various new concepts in cataract surgery. Phacoemulsification Since the introduction of phacoemulsification as an alternative to standard cataract extraction techniques, surgeons throughout the world have been attempting to make this new procedure safer and easier to perform while assuring good visual outcome and patient recovery. The fundamental goal of phaco is to remove the cataract with minimal disturbance to the eye using least number of surgical manipulations. Each maneuver should be performed with minimal force and maximal efficiency should be obtained. The latest generation phaco procedures began with Dr Howard Gimbel’s “divide and conquer” nuclear fracture technique in which he simply split apart the nuclear rim. Since then we have evolved through the various techniques namely four quadrant cracking, chip and flip, spring surgery, stop and chop and phaco chop. The phaco chop technique was developed and introduced by Dr Nagahara to reduce total phaco time and power needed to remove the cataractous lens. In this technique the tip is embedded in the superior half of the nucleus using shallow sculpturing motion. The tip of the chopper is introduced into the nucleus at the 6 O’clock position by penetrating the lens cortex and epinucleus. The two
NO ANESTHESIA CATARACT S URGERY
Fig. 8.1: Eye with cataract
77
Fig. 8.2: Left hand injects viscoelastic using a 26-gauge needle
instruments create the nuclear fracture. The nucleus is rotated and the process repeated to get 4 to 6 fragments. Each quadrant is then emulsified. This chopping does peripheral chopping. In other words, the nucleus is chopped from the periphery. Agarwal Karate Chop Technique Incision A temporal clear-corneal section is made. If the astigmatism is plus at 90 degrees then the incision is made superiorly. first of all, in the cataractous eye (Fig. 8.1) a needle with viscoelastic is injected inside the eye in the area where the second site is made (Fig. 8.2). This will distend the eye so that when you make a clear corneal incision, the eye will be tense and one can create a good valve. Now use a straight rod to stabilize the eye with the left hand. With the right hand make the clear-corneal incision (Fig. 8.3). When we started making the temporal incisions, we positioned ourselves temporally. The problem by this method is that, every time the microscope has to be turned which in turn would affect the cables connected to the video camera. Further the theater staff would get disturbed between right eye and left eye. To
78
T HE ART
OF
PHACOEMULSIFICATION
Fig. 8.3: Clear corneal incision. Note the straight rod inside the eye in the left hand. The right hand is performing the clear corneal incision. This is a temporal incision and the surgeon is sitting temporally
Fig. 8.4: Rhexis being done with a needle
solve this problem, we then decided on a different strategy. We have operating trolleys on wheels. The patient is wheeled inside the operation theater and for the right eye the trolley is placed slightly obliquely so that the surgeon does not change his or her position. The surgeon stays at the 12 O’clock position. For the left eye the trolley with the patient is rotated horizontally so that the temporal portion of the left eye comes at 12 O’clock. This way the patient is moved and not the surgeon. Rhexis Capsulorrhexis is then performed through the same incision (Fig. 8.4). While performing rhexis it is important to note that the rhexis is started from the center and the needle moved to the right and then downwards. This is important because today concepts have changed to temporal and nasal. It is better to remember it as superior, inferior, right or left. If we would start the rhexis is generally where you finish it. In other words, the point where you tend to lose the rhexis is near its completion. If you have done the rhexis from the center and moved to the left, then you might have an incomplete rhexis on the left hand side either
NO ANESTHESIA CATARACT S URGERY
79
inferiorly or superiorly. Now, the phaco probe is always moved down and to the left. So every stroke of your hand can extend the rhexis posteriorly creating a posterior capsular rupture. Now, if we perform the rhexis from the center and move to the right and then push the flap inferiorly—then if we have an incomplete rhexis near the end of the rhexis it will be superiorly and to the right. Any incomplete rhexis can extend and create a posterior capsular tear. But in this case, the chances of survival are better. This is because we are moving the phaco probe down and to the left, but the rhexis is incomplete up and to the right. If you are a left-handed person start the rhexis from the center and move to the left and then down. Hydrodissection Hydrodissection is then performed (Fig. 8.5). We watch for the fluid wave to see that hydrodissection is complete. We do not perform hydrodelineation or test for rotation of the nucleus. Viscoelastic is then introduced before inserting the phaco probe. Fig. 8.5: Hydrodissection
Karate Chop—Two Halves We then insert the phaco probe through the incision slightly superior to the center of the nucleus (Fig. 8.6). At that point apply ultrasound and see that the phaco tip gets embedded in the nucleus (Fig. 8.7). The direction of the phaco probe should be obliquely downwards toward the vitreous and not horizontally towards the iris. Then only the nucleus will get embedded. The settings at this stage are 70 percent phaco power, 24 ml/minute flow rate and 101 mm of Hg suction. By the time the phaco tip gets embedded in the nucleus the tip would have reached the middle of the nucleus. We do not turn the bevel of the phaco tip downwards when we do this step, as the embedding is better the other way. We prefer a 15-degree tip but any tip can be used. Now stop phaco ultrasound and bring your foot to position 2 so that only suction is being used. Now lift the nucleus. When we say lift it does not mean lift a lot but just a little so that when we apply pressure on the nucleus with the chopper the direction of the pressure is downwards. If the capsule is a bit thin like in hypermature cataracts you might rupture the posterior capsule and
80
T HE ART
OF
PHACOEMULSIFICATION
Fig. 8.6: Phaco probe placed at the superior end of the rhexis
Fig. 8.7: Phaco probe embedded in the nucleus. We started from the superior end of the rhexis and note it has got embedded in the middle of the nucleus. If we had started in the middle then we would have embedded only inferiorly that is at the edge of the rhexis and chopping would be difficult. Note the tip of the phaco is not seen as it is fully embedded
create a nucleus drop. So when we lift the nucleus the pressure on the posterior capsule is lessened. Now, with the chopper cut the nucleus with a straight downward motion (Fig. 8.8) and then move the chopper to the left when you reach the center of the nucleus. In other words, your left hand moves the chopper like an inverted L. Remember—do not go to the periphery for chopping but do it at the center. Once you have created a crack, split the nucleus till the center. Then rotate the nucleus 180 degrees and crack again so that you get two halves of the nucleus. In brown cataracts, the nucleus will crack but sometimes in the center the nucleus will still be attached. You have to split the nucleus totally in two halves and you should see the posterior capsule throughout. Karate Chop—Further Chopping Now that you have two halves, you have a shelf to embed the probe. So, now place the probe with ultrasound into one half of the nucleus (Fig. 8.9). You can
NO ANESTHESIA CATARACT S URGERY
Fig. 8.8: Left hand chops the nucleus and splits like an inverted L shape, that is downwards and to the left. When the crack is complete, you should see posterior capsule throughout the crack
81
Fig. 8.9: Embed the probe in one-half of the nucleus. Go horizontally and not vertically as you have now a shelf of nucleus to embed
pass the direction of the probe horizontally as now you have a shelf. Embed the probe, then pull it a little bit. This step is important so that you get the extra bit of space for chopping. This will prevent you from chopping the rhexis margin. Apply the force of the chopper downwards. Then move the chopper to the left so that the nucleus gets split. Again, you should see posterior capsule throughout so that you know the nucleus is totally split. Then release the probe, as the probe will still be embedded into the nucleus. Like this create three quadrants in one half of the nucleus. Then make another three halves with the second half of the nucleus. Thus, you now have 6 quadrants or pie-shaped fragments. The settings at this stage are 50 percent phaco power, 24 ml/minute flow rate and 101 mm of Hg suction. Remember 5 words—embed, pull, chop, split and release. Pulse Phaco Once all the pieces have been chopped, take out each piece one by one and in pulse phaco mode aspirate the pieces at the level of the iris. Do not work
82
T HE ART
OF
PHACOEMULSIFICATION
Fig. 8.10: Cortical aspiration done
Fig. 8.11: Cortical aspiration completed. Note the straight rod in the left hand, which helps control the movements of the eye
in the bag unless the cornea is preoperatively shows signs of inadequate endothelial cells, or the patient is very elderly. The setting at this stage can be phaco power 30-50 percent, flow rate 24 ml, and suction 100 mm of Hg. Cortical Washing and Foldable IOL Implantation The next step is to do cortical washing (Fig. 8.10). Always try to remove the subincisional cortex first, as that is the most difficult. In Figure 8.11 note the cortical aspiration complete. Note also the rhexis margins. Note also that everytime the left hand has the straight rod controlling the movements of the eye. If necessary use a bimanual irrigation-aspiration technique. Then inject viscoelastic and implant the foldable IOL. We use the plate haptic foldable IOL (Fig. 8.12) with large fenestrations generally as we find them superior. Take out the viscoelastic with the irrigation-aspiration probe (Fig. 8.13). Stromal Hydration At the end of the procedure, inject the BSS inside the lips of the clear corneal incision (Fig. 8.14). This will create a stromal hydration at the wound. This will
NO ANESTHESIA CATARACT S URGERY
Fig. 8.12: Plate haptic foldable IOL being implanted
83
Fig. 8.13: Viscoelastic removed with the irrigation aspiration probe
create a whiteness, which will disappear after 4 to 5 hours. The advantage of this is that the wound gets sealed better. Pad, S/C Injections No subconjunctival injections or pad are put in the eye. The patient walks out of the theater and goes home. The patient is seen the next day and after a month glasses prescribed. Agarwal Chopper We have devised our own chopper. The other choppers, which cut from the periphery, are blunt choppers. Our chopper is a sharp chopper. It has a sharp cutting edge. It also has a sharp point. The advantage of such a chopper is that you can chop in the center and need not go to the periphery to make the chop. In this method by going directly into the center of the nucleus, without sculpting, ultrasound energy required is reduced. The chopper always remains within the rhexis margin and never goes underneath the anterior capsule. Hence it is easy to work with even small pupils or glaucomatous eyes. Since we don’t have to
84
T HE ART
OF
PHACOEMULSIFICATION widen the pupil, there is little likelihood of tearing the sphincter and allowing prostaglandins to leak out and cause inflammation or cystoid macular edema (CME). In this technique we can easily go into even hard nuclei on the first attempt. Soft Cataracts In soft cataracts, the technique is a bit different. We embed the phaco tip and then cut the nucleus as if we are cutting a piece of cake. This should be done 2 to 3 times in the same area so that the cataract gets cut. It is very tough to chop a soft cataract, so this technique helps in splitting the cataract. Topical Anesthesia Cataract Surgery
All our cases are done under topical anesthesia. Four percent Xylocaine drops Fig. 8.14: Stromal hydration are instilled in the eye about 3 time’s 10 to15 minutes before surgery. No intracameral anesthesia is used. It is not advisable to use Xylocaine drops while operating. This can damage the epithelium and create more trouble in visualization. No stitches and no pad are applied. This is called the—no injection, no stitch, no pad cataract surgery technique. No Anesthesia Cataract Surgery We had been wondering whether any topical anesthesia is required or not. So we then operated patients without any anesthesia. In these patients no Xylocaine drops were instilled. The patients did not have any pain. It sounds funny because we have been taught from the beginning that we should apply Xylocaine. This is possible because we do not touch the conjunctiva or sclera. We never use any one-tooth forceps to stabilize the eye. Instead what we use is a straight rod which is passed inside the eye to stabilize it when we are performing rhexis, etc. The anterior chamber should be well maintained and the amount of ultrasound power used very less. If you tend to use the techniques like trenching then the ultrasound power generated is high, which in turn generates heat. This causes pain to the patient. If you follow these rules one can perform No Anesthesia Cataract Surgery. It is not necessary to do this, as there is no harm in instilling some drops of Xylocaine in the eye. The point is that there is always a discussion
NO ANESTHESIA CATARACT S URGERY
85
which anesthetic drop to use. It does not matter. The technique, which you perform, should not produce pain to the patient. CONCLUSION As in any other field, progress is inevitable in ophthalmology more so in cataract surgery. We have started to look on cataract surgery as a craft and should constantly try to improve our craft and become better everyday. By this, we will be able to provide good vision to more people than any one dared dream a few decades ago. It also goes without saying that we are and will be forever grateful to all our patients because without their faith, we would never have had the courage to proceed. Keeping this in mind, we hope and wish that the effectiveness and the advantages of this “Karate Chop Technique” be realized and practised, thereby making the technique of phacoemulsification safer and easier providing good visual outcome and patient recovery.
86
THE ART
OF
P HACOEMULSIFICATION
Keiki R Mehta Cyres K Mehta
Clear Corneal Cataract Surgery
9
INTRODUCTION Clear corneal cataract incisions are becoming very popular for cataract extraction and IOL implantation throughout the world. Using clear corneal incisions, with the concomitant use of intracameral anesthesia, cataract surgery has now been refined to such an extraordinary level, that it has reached the stage of virtually immediate, visual rehabilitation. Perhaps the greatest advantage of clear corneal incisions is the tremendous safety with relative astigmatic neutrality, coupled with exceptional results. The limbal location for cataract surgery has been in use from time immemorial. In India, couching has been an established technique, which, is still in use in inaccessible parts. Jaques Daviel in 1745 utilized clear corneal incisions in his cataract surgery technique. Albrecht Von Graefe in 1750 developed a clear corneal knife and developed a very successful corneal incision technique that became famous as the “Graefe section”. Kratz in 1980, was the first phacoemulsification surgeon to go more posterior , making scleral incisions commencing far back and tunneling forwards with a view to increase appositional surface, which would enhance wound healing and thus reduce surgically induced astigmatism. Gerard and Hoffman in 1984 were the first to name the posterior incision as “the corneal tunnel”. Moreover, they were the first to make a point of entering the anterior chamber through the cornea, creating for the first time a corneal shelf. Maloney in 1988, popularized this particular corneal shelf, which in the earlier days was made purely as a means of preventing iris prolapse. In 1989 MacFarland recognized the advantages of implantation of foldable lenses with the self-sealing incisions, but it was Ernst in 1990 who recognized that this long scleral entry, with a tunnel configuration, coupled with a corneal shelf acted as a one-way valve and
CLEAR C ORNEAL CATARACT S URGERY
87
thus explained the mechanism for self-apposition and changed the name from corneal shelf to posterior corneal lip. Howard Fine in 1992 presented his self-sealing temporal corneal tunnel incision at the annual meeting of the American Society of Cataract and Refractive Surgery (ASCRS) and from that time onwards it has become an extremely popular technique. Giving appropriate credit to many surgeons all over the world who had favored corneal incisions for cataract surgery, the leaders in this field were Harms and Mackinson in Germany in 1967, Charles Elm in 1968, Troutman in 1973. In Japan, one of the leading proponents was Kimiya Shimuju. In a recent survey in Eye World magazine, 60 percent of the American surgeons surveyed utilize self-sealing clear corneal cataract incisions for phacoemulsification and foldable IOL implantation. It is anticipated the number will perhaps double by the next year. CONTROVERSIES REGARDING CLEAR CORNEAL INCISIONS The indications for clear corneal cataract surgery have expanded significantly since the last few years. Initially the indications were limited to those patients on anticoagulants, with blood dyscrasias, patients with cicatrizing diseases such as ocular pemphigoid, or Stevens Johnson syndrome. However, the greatest advantage of the clear corneal incision has been the ability to do surgery with topical anesthesia. With the ability to avoid an injection into the orbit, with all its attendant risks, or utilization of intravenous medication in patients with cardiovascular insufficiency, or enhanced risk in patients with general debility, clear corneal cataract surgery became automatically the ideal technique. Another big advantage of clear corneal incisions was that the clear corneal incision technique is topographically astigmatism neutral. With inherent astigmatic neutrality, the predictability of additional astigmatic and cylindrical reducing techniques like arcuate or limbal keratotomy became more predictably effective. The advent of phakic IOL as well as multifocal IOLs has led to stress to gain accuracy not only with the basic power, but to avoid any further development of astigmatism and reduce significantly, if not eliminate astigmatism. Advantages of the temporal clear corneal incision include better preservation of pre-existing corneal configuration, with better preservation of the limbal zone at the 12 O’clock position for future filtering surgery. In cases with tight lids where the exposure is poor, the temporal incision makes life much easier for the surgeon. In addition, drainage from the temporal site is easier as the fluid trickles out from the naturally draining side zone. A fair quantum of controversy regarding the clear corneal incisions was the major concern of increased incidence of endophthalmitis secondary to delayed chamber reformation. It was felt that poor wound healing enhanced the possibility of late postoperative infection. Fortunately, a significant number of very meticulous studies have shown that with a well-designed trapdoor incision, a clear corneal incision, be it placed vertically or temporally does not leak, and thus does not in any way
88
THE ART
OF
P HACOEMULSIFICATION
enhance the risks. On the other hand, the relative simplicity, the superb clarity the next day and the quiet eye far outweigh any inconvenience involved. CLASSIFICATION OF CORNEAL TUNNEL INCISIONS (after Fine et al) Location • Corneal tunnel incision—entry posterior to limbus, exit at the cornea-scleral junction • Corneal tunnel incision—entry just posterior to limbus, exit in clear cornea. • Clear corneal tunnel incision—entry and exit in the clear cornea. Architecture • Single plane no groove • Shallow groove < 400 microns • Deep groove > 400 microns Size and Planar Configuration • Single-plane incision 2.5 by 1.5 mm rectangular tunnel • Two-plane incision 2.5 by 1.5 mm rectangular tunnel • Three-plane incision 2.5 by 1.5 mm rectangular tunnel plus a perpendicular arcuate component. STRENGTH OF CLEAR CORNEAL INCISIONS VERSUS LIMBAL OR SCLERAL INCISIONS Paul Ernst in 1994, demonstrated that rectangular clear corneal incisions in animal models show higher resistance to external deformation utilizing pinpoint pressure as compared to square limbal incisions. The question of stability of the corneal tunnel has been the challenge of pinpoint pressure. The concept has been that whether cataract wound strength should be evaluated by challenging one’s incision using a pinpoint pressure on the posterior lip of the incision. Howard Fine and many other surgeons have demonstrated that it is not a correct test, as it is very unlikely that the patient would challenge his or her own wound strength by pressing with so fine an instrument as to pinpoint the exact site of pressure, which may leak. It is more appropriate to check the challenge with a blunt hook, which closely resembles the pressure, induced by a fingertip or knuckle, which is the most likely way the patient, would exert pressure. It is very unlikely that even a small percentage of incisions would leak spontaneously with blunt pressure (Fig. 9.1). Questions have been raised regarding the relative safety of clear corneal incision versus a limbal-based corneal incision. Though, in theory, the limbal-based corneal incision would heal faster, and have a stronger ability to prevent leakage, the corneal incision gives excellent stability against leakage, and against accidental pressure.The one big disadvantage of the limbal-based corneal incision is the greater likelihood of ballooning of the conjunctiva, either at the site of the incision or even sometimes
CLEAR C ORNEAL CATARACT S URGERY
Fig. 9.1: Side port being made with 1.2 mm MVR blade
89
Fig. 9.2: Clear corneal tunnel being prepared with corneal substance
the complete conjunctiva, billowing forwards, making visualization of the anterior chamber structures during the surgical procedure more difficult. A point of caution Park in 1997 has demonstrated that violation of the glaucoma bleb could threaten the integrity, not only of pre-existing filtering bleb but could also make the zone very amenable to subsequent infection. In addition, there is always likelihood that even a minor trauma like a hard rub to the eye could lead to a small hemorrhage. TECHNIQUES Fine et al in 1992 described the first self-sealing corneal tunnel incision for small incision cataract surgery utilizing a 3.00 mm diamond knife. The technique utilized was a two-plane incision. The first incision was performed perpendicular to the plane of the cornea (Fig. 9.2). The depth of the incision is kept at roughly 200 microns. Though ideally done with a calibrated diamond knife, it is more often, than not, done using the unguarded edge of the diamond knife, with the incision being made a little deeper than a superficial scratch. In doing the second plane, the knife enters deep into the primary incision and then continues in a plane parallel to the corneal plane, forming the cleavage in the corneal stroma. In practice it is simple to do, if the operating surgeon places the flat of the diamond blade, opposed to the conjunctiva, and then enters via the primary incision made. It gives a virtually perfect placement in the stromal zone (Fig. 9.3). The tip of the diamond knife after a 2 mm tunnel is constructed, is then allowed to “dimple” the edge of the Descemet’s, and is then simply pushed forward in the same plane, to achieve a cut in a straightline configuration (Fig. 9.4). Alternatively, the surgeon can use a round disk knife to dissect open a shallow stromal tunnel of the required length and then enter the chamber with the knife. Caution should be used in not entering the chamber with the rounded disk knife as it does not give a straight line cut and will compromise the valve function of the incision.
90
THE ART
OF
P HACOEMULSIFICATION
Fig. 9.3: Tip entered into anterior chamber. Tunnel complete
Fig. 9.4: If required the tip can be dimpled down prior insertion into anterior chamber
Paul Ernst has clearly demonstrated that to be astigmatically neutral the incision should be squared. (i.e. the length and the width of the tunnel should be equal). Williamson in 1993 was the first to utilize a 300 to 400 microns primary clear corneal incision. Rationale for the Williamson incision was that a thicker external edge to the roof of the tunnel had a less likelihood of tearing. Langerman in 1994 described the single hinge incision in which the primary incision was made vertically in the cornea for a depth of ¾ of the cornea (400 microns) with the calibrated diamond knife. Subsequently, half way through the depth of the incision a horizontal groove is made in the stroma. The knife is then passed parallel to the corneal plane for a length of 2.00 mm, is dipped down to dimple the Descemet’s membrane and penetrates into the anterior chamber. Langerman felt that this initial “hinge” gave total protection towards accidental leakage of aqueous by pressure on the posterior flange of the incision. This technique led to an improved resistance for leakage from the incisional edges by the application of external pressure leading to deformation. New Blade Technologies The Fine Tri-diamond knife was developed with Mastel so that the entry incision into the cornea and exit from the cornea into the anterior chamber could be made in an extremely sharp thin line without a necessity to depress the tip of the knife down which often results in a tendency for tearing of tissue or scrolling of the Descemet’s membrane. It also makes for a superb self-sealing valve. Rhein Medical developed a 3-D blade of 2.8 mm in width. This blade has differential slopes on the anterior and posterior surfaces, so that the forces of the tissues exerted along the blade will automatically allow the blade to “flow” in the plane of the cornea with no risk of an early entry nor of an inadvertent anterior escape. One simply places the tip of the blade where one desires the external incision. The blade is then pushed in the plane of the cornea with no attempt to applanate the cornea or attempt to enter the eye by dimpling. Perfect incision can be made rapidly, and more importantly, reproducibly produced.
CLEAR C ORNEAL CATARACT S URGERY
91
CLEAR CORNEAL INCISIONS AND ASTIGMATISM One has to clearly understand the rationale of clear corneal incisions • Excellent access to the anterior chamber for proper rhexis performance and access to the cataract and for IOL placement. • Virtually bloodless incision • Enables the formation of a self-sealing incision, resistant to deformation or leakage. • Variable incision architecture to reduce or eliminate pre-existing astigmatism • Faster physical rehabilitation of the patient. Most surgeons will permit immediate postoperative bending over by the patients or even strenuous physical activity without the risk of wound disruption. • Being an anastigmatic incision, the refractive stability is almost perfect enabling additional reading spectacles to be prescribed in 4 days time. • A much quieter eye, with faster healing, virtually no irritation or redness and no “flag” signs of inflammation. TOPOGRAPHIC CONTROL OF CORNEAL ASTIGMATISM Astigmatism has always been an integral part of cataract surgery. Intracapsular cataract extraction (ICCE) is still popular in many parts of the world especially in Asia, Middle East and Africa. The ICCE technique generally utilized the Graefe knife or scissor based, 180-degree corneal incision. Subsequent wound closure by suturing with 10/0 nylon usually saddled the poor patient with an exorbitant quantum of variable astigmatism. The shift to extracapsular cataract surgery (ECCE) did little to reduce the astigmatism, as invariably a two plane or a three plane, large implant had to be inserted. It was the advent of phacoemulsification that has made the surgeon, and the patient, appreciate the advantages of small incision surgery and apply the concept in an endeavor to reduce astigmatism still further. The greatest advantage of clear corneal < 3.00 mm (also termed sub-three) incision, was that it was literally an astigmatic neutral incision. Evaluating corneal changes utilizing the computerized corneal topographer has significantly improved the understanding of how these incisions work and what has to be done to reduce their astigmatic tendencies. Ideally, corneal topography should be done in all cases in an effort to evaluate the astigmatic component and the resting status of the cornea. In order to understand the effect of clear corneal sutureless incisions, one has to comprehend that the shape of the corneal dome is derived partly from lamellar collagen bundles and the relative elasticity of the corneal tissue. Trauma induced to a cornea, be it an injury or even simply, surgery, affects the regularity of the corneal surface. As the lesions heal, alterations are induced to the corneal shape. It must however be clearly appreciated that while a corneal incision made of 2.8 mm size will induce no topographically demonstratable astigmatic changes whatsoever, one of 4.00 mm or 5.00 mm will induce against-the-rule (ART) astigmatism, more so, if the incisions are not properly constructed.
92
THE ART
OF
P HACOEMULSIFICATION
It is perfectly feasible to combine phacoemulsification and astigmatic surgery on the table. However this is acceptable only provided, the cornea is quite regular (< 2.00 D astigmatism) and the maximum width of the incision to be utilized is 3.00 mm, or smaller. If the astigmatic component is more or if one is going to be enhancing the incision to 4.00 or more, it makes more sense to do it as a twostep procedure. Do phaco as a primary procedure, allow the cornea to stabilize postoperatively, and only then after doing a computerized corneal topography , plan and complete the astigmatic procedure. CONCLUSION Clinically, clear corneal incisions have now become the most popular option for cataract extraction IOL implantation throughout the world. Spearheaded by phacoemulsification and the now sub-three (< 3.00 mm ) sized incisions, significant improvements in surgical techniques have resulted in a keen appreciation of astigmatism, how it is induced and what can be done to reduce, if not eliminate it entirely. Being naturally anastigmatic, phacoemulsification incisions serve as an ideal background datum to achieve a zero refractive status, and at the same time, in managing residual astigmatism, obviate the necessity for complex calculations. Clear corneal incisions have had an outstanding record of safety with exceptional cosmetic outcome and should increase in popularity with time. FURTHER READING 1. Kirk S, Burde RLM, Waltman SRL: Minimizing corneal endothelium damage due to intraocular lens contact. Invest Ophthalmol Vis Sci 16:1053, 1977. 2. Ernst PH, Kiessling LA, Lavery KT: Relative strength of cataract incision in cadaver eyes. J Cataract Refract Surg 17(suppl):668-71, 1991. 3. Ernst PH, Lavery KT, Kiessling LA: Relative strengths of scleral corneal and clear corneal incisions constructed in cadaver eyes. J Cataract Refract Surg 21:39-42, 1994. 4. Ernst PH, Neuhann T: Posterior limbal incision. J Cataract Refract Surg 22:78-84, 1996. 5. Ernst PH: The corneal lip tunnel incision. J Cataract Refract Surg 20:154-57, 1994. 6. Ernst PH: The self-sealing sutureless wound: Engineering aspects and experimental studies. In Gills JP, Martin RG, Sanders DR (Eds): Sutureless Cataract Surgery. SLACK Inc: Thorofare 23-39, 1992. 7. Edelhauser HF, Gonnering R, Van Horn DL: Intraocular irrigating solutions—a comparative study of BSS plus and lactated Ringer’s solutions. Arch Ophthalmol 96:516-20, 1978. 8. Edelhauser HF, Van Horn DL, Schultz RO et al: Comparative toxicity of intraocular irrigating solutions on the corneal endothelium. Am J Ophthalmol 81:473-81, 1976. 9. Edelhauser HF, Van Horn DL, Hynduiuk RA et al: Intraocular irrigating solutions—their effect on the corneal endothelium. Arch Ophthalmol 93:648-57, 1975. 10. Edelhauser HF, Rosenfeld SI, Waltman SR et al: Discussion of comparison of intraocular irrigating solutions in pars plana vitrectomy. Ophthalmology 93:114-15, 1986. 11. Fine IH, Finchman RA, Grabow HB (Eds): Clear Corneal Cataract Surgery and Topical Anesthesia Slack Inc: Thorofare 1993 12. Fine IH: The Rhein 3-D diamond knife. Eye World 1:2-24, 1996. 13. Fine IH, Fichman RA, Grabow HB: Clear Corneal Cataract Surgery and Topical Anesthesia Slack Inc: Thorofare 1993. 14. Fine IH: Architecture and construction of a self-sealing incision for cataract surgery. J Cataract Refract Surg 17(suppl): 672-76, 1991. 15. Fine IH: Cortical cleaving hydrodissection. J Cataract Refract Surg 18:508-12, 1992.
CLEAR C ORNEAL CATARACT S URGERY
93
16. Mehta KR: Pitfalls encountered in 1500 consecutive posterior chamber implant. All India Ophthl Soc Proc 165-66,1986. 17. Mehta KR: Phacoemulsification cataract extraction with foldable IOLS—first 50 cases. All India Ophthl Soc Proc 56-60, 1989. 18. Mehta KR: Endocapsular phacoemulsification and posterior chamber IOL implantation. All India Ophthl Soc Proc 217-20, 1989. 19. Keiki R Mehta: Post-cataract astigmatism—a comparison between phacoemulsification and ECCE procedure—cataract with and without intraocular implantation. All India Ophthl Soc Proc 226-29, 1989. 20. Mehta KR: Posterior capsular capsulorrhexis with shallow core vitrectomy following implantation in paediatric cataracts. All India Ophthl Soc Proc 207-10, 1995. 21. Mehta KR: The clear corneal phacoemulsification with injectable silicone lenses. All India Ophthl Soc Proc 218-22, 1995. 22. Mehta KR: The new shelve and shear technique for simplified phacoemulsification. All India Ophthl Soc Proc 222-24, 1995. 23. Mehta KR: Posterior chamber implantation. All India Ophthl Soc Proc 143-44, 1990. 24. Mehta KR: YAG Laser damage to intraocular implants—an evaluation. All India Ophthl Soc Proc 14750, 1990. 25. Mehta KR: Phacoemulsification—is it the true III world answer for eye camps. All India Ophthl Soc Proc 301-03, 1990. 26. Mehta KR: An analysis of causative faction leading to eye strain caused by computer monitor screens. All India Ophthl Soc Proc 334-36, 1990. 27. Mehta KR: Single stitch elliptical funnel incision for cataract surgery. All India Ophthl Soc Proc 25354, 1991. 28. Mehta KR: Bifocal intraocular implants—a functional evaluation based on 425 cases. All India Ophthl Soc Proc 271-74, 1991. 29. Mehta KR: Phacoemulsification with flexible PC IOL—is it really a step forward? All India Ophthl Soc Proc 287-88, 1991. 30. Mehta KR: The New Phaco cleave technique for hard cataracts. J Intraocular Implant and Refractive Society India, 1(1): 74-75, 1996. 31. Mehta KR, Sathe, SN, Karyekar SD: The new soft intraocular lens implant. Am Intra-Ocular Implant Society J 4(4):200-205, 1978. 32. Mehta KR, Sathe SN, Karyekar SD: New soft posterior chamber implant. X Congress of the AsiaPacific Academy of Ophthalmology. New Delhi,1985. 33. Keiki R Mehta: When not to do an anterior chamber implant. All India Ophthl Soc Proc 164 -165,1986. 34. Mehta KR: Mehta Tangential Chop (MTC) technique for phacoemulsification. All India Ophthl Soc Proc (Chandigarh) 1996. 35. Mehta KR: Phaco-levitation—a peaceful way. All India Ophthl Soc Proc (Chandigarh) 1996. 36. Mehta KR: Comparison of centration stability and capsular response to AcrySoft and silicone S130 lenses. All India Ophthl Soc Proc 1998. 37. Mehta KR: Intralenticular “hubbing” technique for simple eye camp phacoemulsification—a simple technique. APIIA Conference 1997. 38. Mehta KR: Management of subincisional cortex in small incision cataract surgery (SICS). Proc of SAARC Conference, Nepal, 1994. 39. Mehta KR: Intralenticular “hubbing” phaco technique for safe phaco. Proc of SAARC Conference, Nepal, 1994. 40. Mehta KR: The new multiport phaco tip for safer, more effective phacoemulsification, with virtually zero capsular damage. Proc of SAARC Conference, Nepal, 1994.
94
THE ART
OF
PHACOEMULSIFICATION
Shashi Kapoor
Capsulorrhexis (CCC): A Beginner’s Guide
10
INTRODUCTION Continuous curvilinear capsulorrhexis (CCC) is one of the revolutionary innovations of modern cataract surgery. It was presented to the ophthalmic, surgical community in 1985 and 1986 by Fercho, Graether, Gimbel, and Neuhann. These ophthalmologists were able to use and appreciate CCC because they had developed methods for performing phacoemulsification totally within the capsular bag, i.e. they were not using an iris plane approach in which the superior pole of the cataract is tipped superiorly out of the bag for tip access. Terminology CCC—continuous, circular, capsulorrhexis. The border does not need to be “circular”, but may be ovoid or elliptic. Hence, the more generic word “curvilinear” has replaced “circular”, leaving the abbreviation the same—CCC. Anatomy of Lens Capsule The posterior zonular fibers are inserted 1.0 to 1.5 mm from the equator, while the anterior zonular fibers are attached approximately 2.0 mm from the equator. Since the diameter of the adult crystalline lens is 9.5 to 10.0 mm, the “zonulefree” area, on the anterior capsule is approximately, 6.0 mm in diameter only. It is therefore ideal to create a tear limited to the zonule-free area, preferably slightly inside the zonular frontier, as zonular fibers have been observed more centrally than has previously been considered.
CAPSULORRHEXIS (CCC): A B EGINNER’ S GUIDE
95
Technique Prerequisites Absence of positive pressure— facilitated by use of viscoelastic agents, air, irrigation, or a “closedchamber” technique, where a bent needle, used to perform the CCC, perforates the preplaced incision, before any other entry is made into the anterior chamber. Instruments Cystotomes, bent needles or forceps, can all be used effectively. Existing ultrasonic or thermal devices, are not believed to offer any advantages. Initiation of Tear Beginning of CCC peripherally, carries a greater liability than beginning centrally. With the initial cuts made centrally, radial stress vectors across the anterior capsule are interrupted, resulting in less tendency for the tear to extend towards the equator. It is always easier to spiral the initial tear out to enlarge a capsulotomy, than it is to pull a peripheral tear back towards the center. Technique Using a bent needle, of 23 G to 25 G, a perforation is made in the Figs 10.1 A to C: Continuous curvilinear capsulorrhexis center of the anterior capsule. By (CCC): (A) central puncture with cystotome and progression extending this with the sharp edge of tear to the 3.0 O’clock position, (B) the curvilinear tear is continued counterclockwise with the capsular flap folded of the needle, a horizontal incision over itself, and (C) the progressing tear is blended at 3.0 is made (Fig. 10.1A). The tip of the O’clock with the cystotome having re-engaged the ventral needle is now used to redirect the surface of the capsule for optimum control (Gimbel) tear in a counterclockwise direction. This creates a flap with a smooth curve as it is beginning. The flap is then pulled along in a circular manner by means of gentle traction with the needle tip. If the tear starts to extend peripherally, it is usually the result of positive vitreous pressure
96
THE ART
OF
PHACOEMULSIFICATION
and can be counteracted by reinflating the anterior chamber with viscoelastic. A light touch is needed, because if one presses too hard on the flap, it creates a responding increase in vitreous pressure, which forces the tear outward. As the flap progresses, large amounts of capsular folds will present and must be pushed out of the way, so that one can visualize the exact point at which to place the tip of the needle (Fig. 10.1C). When completing the CCC, one should overlap the tear such that the last part of the tear joints the first part from the outside towards the center, thus resulting in a continuous edge. If the overlap is created from the center towards the outside, it will result in a small triangular flap, with a tendency to tear towards the equator or beyond. The best control of the progressing tear is achieved by grasping the developing capsular flap, with the desired instrument close to where the capsule is tearing at the time. The direction of the tear can be controlled by the position of the instrument. Placing the tip of the instrument a little peripheral to the advancing tear, will direct it outwards. Placing it a bit central to the tear will direct it towards the center. Tear patterns may vary, and progress clockwise or counterclockwise. There are cases however, when no form of CCC may be achieved. These include capsules that are heavily fibrosed and shrunken, as in certain congenital, secondary, and traumatic cataracts. In these patients, a continuous, curvilinear opening often still may be achieved by using a capsule scissors to cut through the fibrosed part of the anterior capsule. In the nonfibrosed area, the smooth edge border of the CCC is achieved by the usual methods. A Kraff-Utrata forceps can be used to perform CCC. The initial puncture in the anterior capsule is made with a bent needle, (or with the forceps itself, if the tips of the forceps are sharp enough), and the rest of the CCC is continued with the forceps. Using the forceps requires a larger opening into the anterior chamber as compared to a bent needle, and viscoelastic is usually necessary. Two-staged CCC In this procedure, the original capsulotomy is just large enough to admit the smaller endocapsular phaco probe and a second instrument for lens manipulation. After the
Figs 10.2A and B: Two-staged continuous curvilinear capsulorrhexis (CCC)
CAPSULORRHEXIS (CCC): A B EGINNER’ S GUIDE
97
lens material is removed a small initial opening is converted to a larger one, of the desired diameter, while still maintaining the continuous tear edge. The second capsulotomy is started with a tangential snip on one side of the opening with a Vannas scissors (Figs 10.2A and B). This requires a viscoelastic agent in the anterior chamber and lens capsule. It is important to prevent the side of the capsule opening from folding over or under at the scissors tip so as to prevent a V-cut. Also, the snip is not taken to the scissors point, because the point may create an irregularity in the line of the cut. Such a notched cut destroys the integrity of the continuous tear. Once the tangential cut is successfully achieved, the second continuous tear is then extended, using Utrata forceps to complete a larger circle, which is centered in the pupil, and is of the desired diameter. The forceps enlarges the original capsulotomy by removing a strip or ribbon of additional capsule. Two-staged CCC is Particularly Useful • In patients with small pupils, when an originally small CCC requires subsequent conversion to a larger CCC. • When the original CCC is made inadvertently small. • For corneal endothelial protection in intercapsular and endocapsular cataract extraction. A similar technique may be used in blunting or turning back, short inadvertent tears of the anterior capsular border. Posterior CCC (PCCC) Posterior CCC uses the principle of CCC and is used to advantage, when a small linear or triangular tear inadvertently occurs in the posterior capsule, in order to convert it into a smooth CCC that is resistant to radial extension. The PCCC is accomplished just as one does an anterior CCC. The size of the PCCC is kept as small as possible to preserve the maximum support by the posterior capsule. Equatorial capsular tears, posterior capsule tears near the equator or with extensions to the equator are not suitable for PCCC. PCCC may also be used, for making primary posterior capsulectomies, in removing posterior plaques, or to blunt, small extending tears with posterior continuous. Intumescent Cataracts and CCC Intumescent cataracts are a special problem for any form of capsulotomy, especially CCC. The internal lens pressure simply splits any rent towards the equator. A useful technique in such cases, is to decompress the lens by first aspirating some lens cortex. Another difficulty is the poor visualization of the tearing edge, in milky white cataracts, where there is no red reflex. In such instances, Lee has found it extremely useful to perform CCC under air. The distortion of the tearing capsule under an air-lens interface is apparently, exceptionally clear.
98
THE ART
OF
PHACOEMULSIFICATION
Advantages of CCC There are several advantages of CCC: • In-situ phacoemulsification is facilitated, and the ultrasonic turbulence is contained within the lens capsule. • IOL implantation and verification in the bag is greatly facilitated because of the smooth-edge visible rim. • IOL rotation with no chance of decentration caused by loops coming out of the bag is allowed. • No capsular tags or V-shaped tears are left that can extend into the posterior capsule, under even minimal, mechanical stress. • A diaphragm quality of the capsule for sulcus placed lenses is maintained or preserved in the event of a ruptured posterior capsule. • Chances of posterior synechiae are reduced. • In-the-bag IOL implantation in the very elastic capsule of children is facilitated. ECCE and CCC Most surgeons recommend one or two “relaxing incisions” in the CCC, prior to nuclear expression in extracapsular cataract extraction (ECCE). The relaxing incision helps prevent untoward complications as zonular tears, vitreous loss, unintended ICCE, or even prolapse of lens into the vitreous cavity. Hydrodissection, hydrodelineation, and hydroexpression techniques are useful adjuncts, when using CCC along with ECCE. Complications of CCC These include: • Shrinkage of anterior capsular opening • Capsular bag hyperdistention • Epithelial cell hyperproliferation on the posterior capsule. CAPSULAR CONTRACTION SYNDROME Capsular contraction syndrome is seen after ECCE using can-opener capsulotomy or CCC. While rarely with can-opener capsulotomies, with anterior radial capsular tears, it is relatively frequent with CCC. The syndrome consists of an exaggerated reduction in the anterior capsulotomy opening and equatorial capsular bag diameter. This may also
Fig. 10.3: (Davison) Capsular contraction syndrome with capsulorrhexis opening displaced inferiorly and open to only 2.0 mm several months after surger y in eye with pseudoexfoliation
CAPSULORRHEXIS (CCC): A B EGINNER’ S GUIDE
99
lead to a malposition of the IOL. These effects seem more exaggerated in small CCC openings and the older patient (Figs 10.3 to 10.6). It is due to capsular bag contraction from fibrous dysplasia of residual lens epithelial cells, countered by relatively unopposed weak zonular support. This is particularly seen, in pseudoexfoliation, advanced age, and in association with uveitis, pars planitis, and myotonic muscular dystrophy. Fig. 10.4: (Davison) Excessive capsular contraction has shrunk the capsulorrhexis opening to approximately 2.0 mm. The opening is decentered as is the silicone optic and the capsular bag; so the zonular fibers are visible below. Pseudoexfoliative material can be seen on the fibers
Role of IOL Material
Hayashi et al showed the reduction in the area of anterior capsule opening at various postoperative intervals after continuous capsulorrhexis and compared any differences in the area reduction between polymethylmethacrylate (PMMA), silicone, and soft acrylic IOLs. Figure 10.7 shows retroillumination photographs, which illustrate the typical postoperative changes in the anterior lens capsule of the three optic materials. Figure 10.7 top left, shows an eye after undergoing PMMA IOL implantation. A slight degree of Fig. 10.5: (Hansen) Red reflex of an eye with dense contraction and fibrosis of the anterior capsular fibrosis and well-centered, but anterior capsule opening occurred extensively constricted capsular opening (1.3 mm). Note the radial stress lines and the broad fibrous rim of over the PMMA optic. Figure 10.7, top the circular capsulotomy. The three-piece posterior right, shows an eye after undergoing chamber IOL is completely within the capsular sac and silicone IOL implantation. shows a mild inferotemporal decentration The contraction of the anterior capsule progressed markedly up to 3 months after surgery. The degree of fibrosis in the anterior capsule was also extensive. Figure 10.7, bottom shows an eye after soft acrylic IOL implantation. The anterior capsule contraction was very slight. Surprisingly, no fibrosis in the anterior capsule was evident in some cases in the soft acrylic IOL group. The study clarified that the area of the anterior capsule opening gradually decreased for up to 3 months after surgery. However, after 3 months, the area reduction in the anterior capsule opening showed no further progression. The percentage of the area reduction with the silicone IOL
100
THE ART
OF
PHACOEMULSIFICATION
Fig. 10.6: (Hansen) Photograph of a case immediately after Nd: YAG laser anterior capsulotomy in a cross pattern, cutting through the fibrous rim of the constricted opening
was greater than that with the polymethylmethacrylate and soft acrylic intraocular lenses. No significant difference was observed between the PMMA and soft acrylic intraocular lenses. Furthermore, although it could not be quantitated in this study, the degree of fibrosis in the anterior capsule was most extensive in the silicone IOL, followed by the PMMA IOL. On the other hand, the anterior capsule fibrosis over the soft acrylic optic was extremely slight.
Prevention Capsular fibrosis is caused by metaplastic lens epithelium. The more epithelium that is left, the greater the potential for capsule contraction. Since twice as much epithelium is removed with a 5.5-mm capsulectomy as with a 4.00-mm capsulectomy,
Fig. 10.7: Retroillumination photographs showing postoperative changes in the anterior lens capsule. (Top left) after a polymethylmethacrylate (PMMA) IOL implantation. A slight degree of contraction and fibrosis in the anterior capsule was observed over the PMMA optic. (Top right) An eye after a silicone IOL implantation. A marked contraction of the anterior capsule opening occurred for up to 3 months. The degree of fibrosis in the anterior capsule was also prominent, especially along the capsulorrhexis edge. (Bottom) An eye after soft acrylic (AcrySof) IOL implantation. The anterior capsule contraction over the soft acrylic optic was very slight. It was surprising that no fibrosis in the anterior capsule was evident
CAPSULORRHEXIS (CCC): A B EGINNER ’ S GUIDE
101
one should make the larger anterior capsulectomy in vulnerable eyes. The sphincter effect of an intact capsulorrhexis is important in creating significant capsule shrinkage. A can-opener capsulotomy with deliberately created anterior radial capsular defects may be appropriate in lose eyes with zonular weakness. Another influence in maintaining capsular bag size and shape and good functional IOL position is the use of a one-piece, all-PMMA IOL with a relatively firm broad haptic structure. Nishi suggested vacuuming the undersurface of the anterior capsule to significantly reduce the amount and effect of residual lens epithelial cells.
Fig. 10.8: Six anterior YAG capsulotomies are almost not visible, three weeks after creation in a patient with pseudoexfoliation
Fig. 10.9: Same eye as in Figure 10.8 six anterior YAG capsulotomies have been redefined with more YAG separation
Role of Nd:YAG Laser Early anterior radial YAG laser, relaxing capsulotomy helps resolve the ultimate contraction of the anterior capsulectomy opening (Figs 10.8 and 9). These anterior capsulotomies may also reduce the incidence of more rare complications of excessive zonular traction and its sequelae, IOL dislocation and retinal detachment. If capsule contraction is noted, consider YAG laser relaxing anterior capsulotomies at 2 to 3 weeks postoperatively. Active capsular fibrosis and attendant contracture of the anterior capsule and indeed the entire capsular diameter can be influenced with early YAG laser intervention, whereas later intervention may not help really as much. CAPSULAR
BAG
HYPERDISTENTION
“Capsular block” implies fluid hyperdistention of the capsular bag from occlusion of the circular anterior capsule opening by the IOL optic; the resulting anteriorplacement of the optic induces an artificial myopia; the source of the fluid is unclear, although some have suggested that it is retained viscoelastic (Fig. 10.10). Other possibilities include transudation through the lens capsule or exudation from the lens epithelial cells. Capsular block is generally self-limiting and is only associated with capsulorrhexis and IOLs flexible or mobile enough to move forward against the capsule opening; it has not been observed with the traditional can-opener
102
THE ART
OF
PHACOEMULSIFICATION
Fig. 10.10: (Masket) Schematic representation of the capsular block phenomenon. The highly flexible one-piece lens depicted in this model is bowed forward, the optic portion blocks the capsulorrhexis, and the capsular bag is distended
Fig. 10.11: (Masket) String of pearls observed as the hyperproliferation of lens epithelial cells (Elschnig pearls) surrounding a previously formed Nd: YAG laser posterior capsulotomy
capsulotomy or with one-piece PMMA IOLs. A small Nd:YAG laser opening in the anterior or posterior capsule permanently relieves capsular block. HYPERPROLIFERATION
OF
LENS
EPITHELIAL
CELLS
Hyperproliferation of lens epithelial cells in the form of Elschnig pearls on the central posterior capsule, surrounding a previously made laser capsulotomy, has been reported (Fig. 10.11). The “string of pearls” that forms around the capsulotomy reduces the opening, making the capsulotomy potentially too small to be optically adequate. The pearls appear to form on the anterior hyaloid and capsulotomy edge, using either as scaffolding, and then progress centrally. The epithelial cells surrounding the capsulotomy are dense, but became more sparse peripherally. Perhaps a stimulus to epithelial cell growth is present in the anterior vitreous, since the string of pearls does not form until after the posterior capsule has been opened. The “string of pearls” has only been observed in cases in which the IOL diameter was larger than the anterior CCC. In this situation, the anterior capsule overlaps the IOL edge, and the anterior and posterior capsules cannot fuse in a true Soemmering’s ring because the optic is sandwiched between the capsule leaflets. The reduced effective size of the posterior capsulotomy associated with the hyperproliferation of lens epithelial cells, has required Nd: YAG laser “retreatment” in many patients.
Shrikant Kelkar
Capsulorrhexis: Principles and Advanced Techniques
ANTERIOR
11
CAPSULORRHEXIS
History The problems related to the sulcus implantation of posterior chamber intraocular lenses (PC IOLs) during the period of about 1975 to early 1980s lead to the idea in favor of placing the intraocular lens implant into the capsular bag which lead to a better centration of the implant. The new design of Simcoe loops (modified C loop) improved better centration. However decentration was not uncommon. The analysis of this decentration showed that in spite of the fact that the lenses were in correct endocapsular situation tears of the anterior capsule originated, luxating one loop into the sulcus which was considered a precondition for subsequent decentration. Realizing the above difficulty in the centration of lens the idea of continuous curvilinear capsulorrhexis (CCC) was born. The credit goes to Tobias H Neuhann and his brother Thomas in developing the technique of capsulorrhexis. At the same time totally independent development by Howard Gimbel who was working on the same idea called this technique as “continuous tear capsulotomy”. Finally, Neuhann and Gimbel called their development continuous curvilinear capsulorrhexis which soon gained popularity all over the world. Physics of Capsulorrhexis If we consider a strip of paper to be torn into two pieces it can be done in two ways: one is as shown in Fig. 11.1A—what can be considered as shearing technique or Fig. 11.1B—ripping.
104
THE ART
OF
PHACOEMULSIFICATION
Figs 11.1A and B: (A) Shows the shearing technique as compared to the ripping technique in (B). Note shearing is a much more controlled technique
Capsulorrhexis with Shearing Capsulorrhexis starts when the cystitome enters point a and proceeds to point b in a radial manner (Fig. 11.1C). At point b the cystotome is pulled in the direction of the arrow to reach point c to create a capsular crack which gets folded over to lie on the top of intact anterior capsule. At the position d the flap is engaged with cystotome or forceps and is pulled in the direction of curved arrow. Figure 11.1D shows the progress of capsulotomy. This flap is a mirror
Figs 11.1C to E: Rhexis by shearing. Note the position of the dot indicates successive placement site for needle rhexis
CAPSULORRHEXIS: PRINCIPLES
AND
ADVANCED TECHNIQUES
105
Figs 11.1F and G: Capsulorrhexis by ripping technique: Dot indicates point where the tear is held with the forceps
image of the area of capsulorrhexis performed so far and its edges will indicate in which direction the capsulorrhexis will proceed. Figure 11.1E shows one-third of the rhexis completed. Take a note of the point at which the instrument engaged the capsule which has remained 2 to 3 O’clock hours away from the point of (n) shearing. Instead if the instrument was placed to say at point m, an artificial stress line would be created that would compromise the predictability of the direction of shear. In shearing technique the flap must be spread out flat otherwise it would convert the shear to modified rip. Capsulorrhexis with Ripping Figures 11.1F and 11.1G demonstrate the ripping technique. The surgeon must note that the direction of pulling is much more towards the center of the capsule and the flap is engaged by pulling the instrument at a point which is much closer to the tear. The ripping technique has the tendency to extend peripherally. It has been found that ripping techniques are more difficult to control and are more likely to extend peripherally compared to shearing technique. Yet one must know its ability in changing the direction of the tear. Initiation of Capsulorrhexis The capsule can be incised at point a, proceeding to point b. The triangular flap thus created is grasped at point c and drawn in the direction of the arrow in a curvilinear fashion. Shearing forces will be equally acting at point b and d. In Figures 11.1H and I the pulling motion has changed to a curvilinear direction shown by the arrow. Figure 11.1J shows its further progress.
106
THE ART
OF
PHACOEMULSIFICATION
Figs 11.1H to J: Triangle tear technique of capsulorrhexis
Methods • • • •
Needle technique Forceps technique Capsulostripsis Diathermy capsulotomy
Capsulorrhexis can be performed by number of methods. All these methods give practically the same results. It involves a continuous symmetrical linear opening of the anterior capsule. In order to get good postoperative results capsulorrhexis must satisfy: Site For optical and functional results the capsulotomy must be centered on the pupil which reduces the possibility of irido-capsular-synechial formation. Suitability (for IOL implantation) Since there is a huge range of IOLs available. The surgeon must first choose the lens and perform the capsular opening accordingly.
CAPSULORRHEXIS: PRINCIPLES
AND
ADVANCED TECHNIQUES
107
Integrity of the zonules In order to avoid damage to zonular fibers the capsulotomy should be performed carefully in the pre-equatorial zones. Aperture The aperture must be wide enough to allow emulsification of the nucleus easily. A large capsular aperture facilitates easy and quick aspiration of the cortical material. At the same time it should not be so wide as to allow the nucleus to prolapse into the anterior chamber during hydrodissection procedure which usually happens in grade I and grade II cataracts and hence between 5 mm and 6 mm is the ideal size of capsulotomy. Shape If we consider the shape of the eyeball as a round clock, the movement while doing the capsulotomy should be like a radiating arm moving in a circular fashion. Similarly the opening should be round. Last but very important aspect is to have continuous margin. The advantages of the continuous margins are: • Correct centration of the lens in the bag. • The edges of the incision are strong and elastic enough and reduce the risk of capsuler rupture extending to the periphery which can happen in a canopener technique. Principles To achieve the above results the surgeon has to keep in mind the following principles: One must keep in mind before proceeding to surgery that any mistake made during this procedure may face subsequent problems. • It is desirable to use high magnification so that surgeon can control every step of the procedure, adjust the sight and width of the incision and avoid traction process as and when possible. • The light beam must be angled to provide good red reflex of the fundus. • The intensity of the microscope light must be sufficient to facilitate clear view of the capsule and the red reflex. • The pupil must be widely dilated which makes it easier for the surgeon to perform capsulotomy; small pupil can bring contact between cystitome and iris. Dilated pupil also facilitates good red reflex. • A deep chamber reduces the risk of endothelial and iris damage hence the depth of the chamber must be maintained during the entire capsulotomy procedure. The high molecular weight viscoelastic substances are the best tools for this. • One should keep in mind the attachment of the zonuler fibers. On the safer side the capsular aperture should not extend more than 3 mm peripherally from the center of the anterior pole of the lens. • The shape and size of the capsulotomy must be planned by the surgeon prior to beginning the capsular aperture. • The entrance of the instrument into the anterior chamber must be well planned. One must keep in mind the anatomical variations of the individual eye like myopia, hypermetropia, etc. Thus it is desirable to have an optimal working
108
THE ART
OF
PHACOEMULSIFICATION
position with respect to the planes of the anterior capsule and the iris. Correct use of the instrument and the released hands of the surgeon makes the job easier. The forceps or the cystitome must be handled gently and carefully. They must be used on the surface of the capsule with practically no pressure applied on the cataractous lens. The surgeon must avoid touching the iris because every such touch can reduce the size of the pupil. The surgeon must keep in mind that in younger patients the capsule is thin and elastic and the capsulotomy tends to extend easily towards the equator. Any time surgeon anticipates the difficulty he should stop what is being done and inject the viscoelastic substance to deepen the chamber which allows him a greater margin for maneuvers and errors. Advantages of CC Capsulorrhexis Intraoperative Advantages • CCC limits the risk of tears extending to the periphery and to the posterior capsule during surgery especially in young patients. • Hydrodissection becomes safe and easy. • It restricts the intraocular turbulence inside the capsule. • Reduces the stress on the zonules during surgery. • Most important is that it allows easy aspiration of the cortex and does not let any anterior capsular tags get caught into the aspiration port. • It permits the correct positioning of the IOL in the bag because the surgeon has excellent view of capsular rim. Last but the most important is that since the rim is can be stretched even a slightly large lens can also be manipulated with a comparatively small capsular opening. Postoperative Advantages • Due to extensive contact area between the loop of the IOL the anterior capsule the possibility of decentering is reduced. • Not being in contact with the ciliary body pigment dispersion, hyphema and inflammation are reduced. • In the presence of ruptured posterior capsule after doing anterior vitrectomy a large optic diameter lens can be inserted over the anterior capsule. • Since the lens is put in the bag there is no need to use miotics or the need for iridectomy. It also avoids pupillary capture. Difficulties during Rhexis Rhexis Escape In some cases while the procedure of capsulorrhexis is proceeding on the aperture goes on increasing towards the periphery. The moment this happens the surgeon must ensure that the chamber is adequately deep by injecting more viscoelastic
CAPSULORRHEXIS: PRINCIPLES
AND
ADVANCED TECHNIQUES
109
substance. It is advisable to use a cystotome attached to a viscoelastic syringe so that in case the chamber becomes shallow more viscoelastic can be injected into the chamber immediately. The chamber will deepen and the tendency for the rhexis to extend will be minimized. At this moment the flap should be caught close to the point in the capsule, the surgeon should then extend controlled pressure directed towards the center of the pupil. Alternative to this technique, instead of forceps or needle one can take deeply curved scissors, deepen the anterior chamber with the viscoelastic and cut the capsule with scissors at the escape point and redirect the opening back to the initial route. One can also choose a new point to start rrhexis in another position operating in a counterclockwise direction and try to join to the first rrhexis escape point. In spite of these maneuvers if the rrhexis extends, abandon the idea of continuous capsulorrhexis and continue with can-opener capsulotomy. Capsulorrhexis in Special Cases Pseudoexfoliation The surgeon must be aware that in this condition the capsule is fragile so the opening must be small so that it does not reach the zonules or create traction forces. Such a small opening can subsequently be extended after the insertion of IOL or subsequently by YAG laser. Young Patients In very young patients proper pupillary dilatation is a big problem. In addition reduced scleral rigidity increases the tendency to positive vitreous pressure. This leads to loss of control in the direction of capsulotomy with an increased risk of peripheral escape. In children therefore the initial incision must be small which can be better controlled by forceps than needle. Viscoelastic substances must be kept in hand. Hypermature or Intumescent cataract Capsulotomy can be done easily when the red reflex is seen. In such situations like hypermature cataract there is no red reflex so capsulotomy becomes extremely difficult as there is no perception of the capsular flap. In addition when the cataract is swollen there are more chances of peripheral escape because of the greater tension on the anterior capsule. When the cataract is intumescent, the contents are usually very liquid like milk, it is advisable to let the semiliquid cortex escape through a central incision, this should be removed with I/A and chamber should be filled with viscoelastic substance, the flap must be taken under a high magnification with very high illumination. This is a good indication to use forceps instead of needle. Instead of red reflex illumination it is advisable to use high intensity oblique illumination. Some surgeons prefer air in the anterior chamber instead of viscoelastic substance. It is advisable to use cystitome instead of forceps when the air is injected into the anterior chamber. Black cataracts also pose the same problem. In small pupils it is advisable to use viscoelastic to deepen the anterior chamber. Use of iris hooks or bimanual stretching is also recommended.
110 POSTERIOR
THE ART
OF
PHACOEMULSIFICATION
CAPSULORRHEXIS
Posterior capsulorrhexis is a linear continuous circular capsulotomy with all the advantages of elasticity, stability and long-term resistance. This procedure is done in children which avoids second surgery or YAG laser. The surgeon proceeds in the same way as in the anterior CCC. The phacoemulsification is carried out, the anterior segment is filled with viscoelastic substance, the bag is also filled with viscoelastic substance. The first step is to perforate the posterior capsule. Viscoelastic is again introduced behind the posterior capsule which can prevent vitreous prolapse. The second step is to carry out posterior CCC with needle or forceps. In some cases when the vitreous prolapses anteriorly, vitrectomy is done. The remaining tyre like capsular residue provides the stable and secure site for IOL fixation. This posterior capsulorrhexis done in pediatric cataract surgery can avoid secondary membrane formation. Disadvantages of CCC The introduction of capsulorrhexis has led to a new problem called capsular shrinkage syndrome or capsular phimosis. This problem is observed more frequently in patients suffering from pseudoexfoliation syndrome, uveitis, retinitis pigmentosa, in combination with polymethylmethacrylate (PMMA) or silicone IOL implantation. All the above mentioned diseases have reduced number of zonular fibers as common observation. Special Techniques • • • •
Fluorescein blue light assisted capsulorrhexis for mature cataract. Staining the capsule with indocyanine green for white cataracts. Trypan blue capsule staining for better visualization in capsulorrhexis. Diathermy capsulotomy.
Trypan Blue Staining When retroillumination is absent, e.g. dense cataract, it is difficult to discriminate the anterior capsule from the underlying lens tissue and capsulorrhexis carries a high risk of radial capsule tears. To overcome this difficulty and better visualization of capsulorrhexis during surgery, one can utilize: (i) use of side illumination, (ii) hemocoloration of the capsule with autologous blood, and (iii) staining the capsule with gentian violet 0.1 percent or methylene blue 1 percent. All these procedures are relatively time consuming and require an adjustment of the surgical technique and may have endothelial toxicity. On the other hand trypan blue stain enables the surgeon to visualize the capsulorrhexis in the absence of red reflex and does not affect the endothelium. Before injecting the dye anterior chamber is injected with air, removing the aqueous through the cannula. Trypan blue, 0.1 ml in a conccentration of 0.1 percent in
CAPSULORRHEXIS: PRINCIPLES
AND
ADVANCED TECHNIQUES
111
phosphate buffered sodium chloride is applied to the anterior capsule. After few seconds anterior chamber (AC) is thoroughly irrigated, viscoelastic substance is injected into the AC (Fig. 11.1K). Because of the blue stain the outline of the capsulorrhexis is easily visible. There is an additional benefit that the peripheral anterior capsule ring remains stained and clearly visible during phacoemulsification. Diathermy Capsulotomy In order to overcome the difficulty in doing capsulorrhexis under circumstances such as mature 11.1K: The edges of the rhexis cataract and absence of red reflex, many surgeons Fig. can be stained with trypan blue for better use diathermy capsulotomy. One must keep in visibility mind that some fundamental research from Denmark showed that the extensibility of the capsule edge after diathermy capsulotomy was reduced to half than that of the CCC technique as done by needle or forceps. It was also noticed that the breaking force required to break the diathermy capsulotomy was one-fifth of that required to break the CCC edge. So, the surgeon must keep in mind the risk of inadvertent capsule opening subsequently by tearing during phacoemulsification or on implantation of IOL. CONCLUSION Every surgeon dreams to provide greatest degree of stability of vision to his or her patients. In other words, the surgeon tries to restore the patient’s vision very close to the normal healthy eye without glasses. The new technique of CCC has introduced a biggest breakthrough in performing small incision or phaco surgery. The most important advantage of continuous capsulorrhexis is that it holds the nucleus down in the bag during phacoemulsification surgery. It makes the surgeon keep the instrument tip in the bag and work on the nucleus. The second advantage is, it helps to maintain intact capsular bag. The circular opening into the anterior capsule opens the window over the nucleus. The structural rigidity and integrity of the capsular bag are almost identical to a completely intact bag. In the can-opener technique, the purpose was to permit the nucleus to get out of the bag while CCC holds the nucleus inside. The principle of capsulorrhexis is well established now and its technical performance is being refined and advanced everyday. Capsulorrhexis is a fundamental surgical principle. These days the ophthalmic instrument manufacturing companies are bringing out many tools for the maximum comfort of the individual surgeon. Stability of the capsular bag and centration of IOL by this technique make emmetropia possible.
112
THE ART
OF
PHACOEMULSIFICATION
Keiki R Mehta Cyres K Mehta
Hydrodissection and Hydrodelineation
12
HYDRODISSECTION Hydrodissection is the creation of a cleavage plane between the nucleus and the cortex. It can also be defined as the separation, by a fluid dissecting wave, of the nucleus from the external cortex adhering to the capsule. It is important to appreciate that the cleavage plane is not between the capsule and the cortex. If that were so, then there would never be any need to do cortical aspiration. Hydrodissection is a very important step for endocapsular phacoemulsification. Its biggest advantage is that it permits free maneuvers on the nucleus in the bag without, in any way transgressing on the safety of the capsule. It must be clearly noted that effective and safe hydrodissection can only be done after a good rhexis. It is unsafe to do hydrodissection in a can-opener capsulotomy, or if the rhexis has, inadvertently, run away into the periphery. Injecting fluid at this time will cause the tear to spread backwards. Hydrodissection Technique The technique involves injecting a small amount of fluid (Ringer lactate or BSS) under the anterior capsule with a fine blunt cannula connected to a 3.00 ml syringe. Because of the fluid pressure and the dissecting ability of the fluid to take the path of least resistance, the fluid separates the cortex and the epinucleus and only partly between the capsule and the cortex. During the hydrodissection, the fluid wave can be seen clearly (unless it is a very hard cataract or an opaque one) to separate the cortex from the nucleus and is indicative of a successful hydrodissection. More hydrodissection is usually carried out in three sites commencing with the 4.00 O’clock position (Fig. 12.1) followed by the 2.00 O’clock position (Fig.12.2) and finally followed
HYDRODISSECTION
Fig. 12.1: Hydrodissection at 8 O’clock position
AND
HYDRODELINEATION
113
Fig. 12.2: Hydrodissection at 2 O’ clock position
by hydrodissection at 8.00 O’clock position. It is important that small aliquots of fluid be utilized, as excess fluid especially in a hard brown cataract is liable to balloon the capsule posterior, rather than spreading as a wave, and may, if more fluidic pressure is applied, rupture the capsule. It is important to visualize while injecting the fluid diffusion wave. The ideal syringe is a 3.00 ml Luer-Lok, plastic, disposable, Teflon-coated or siliconized. This type of a syringe permits a better control, prevents too much pressure from being applied, the plunger moves very smoothly, and does not stick. Too thin a cannula, (ideal is 24-26 G, flat cannula), even if it is blunt is liable to puncture the capsule if accidentally inserted too far into the periphery. Also a thin cannula permits the fluid to emerge in a sharp jet, at high velocity, which is not required. The one way to be sure the hydrodissection is complete is to check whether the nucleus rotates freely in the bag. The ideal technique of cannula placement for effective hydrodissection is to place the cannula just within the capsulorrhexis edge, slightly tenting it or lifting it upwards. This technique termed as cortical cleaning hydrodissection was originally conceived by Dr Howard Fine. Injecting the fluid along the rhexis edge permits the fluid wave, literally to shear close to the capsule thus, significantly diminishing the quantum of cortical remnants which will need to be aspirated after the primary nucleus is removed by phacoemulsification. In all cases hydrodissection should be followed by mechanical rotation (Fig. 12.3) of the nucleus to be sure that the nucleus rotates freely. Rotation confirms that all the adhesions between the epinucleus and the cortex have been broken. It is important to appreciate that if the lens does not rotate freely one must do hydrodissection again, till smooth rotation is achieved. It is important that after every injection of fluid the lens should be gently pressed backwards. This technique is termed as compression hydrodissection, and works by causing the fluid to disperse and spread out as a flat lamellar zone at the back of the nucleus and thereby enhance the hydrodissection. This technique should be conducted gently following each injection of fluid under the capsular flap. Compression hydrodissection thus, decompresses a filled capsular bag, and at the same time hydrodissects or shears off any adhesions.
114
THE ART
OF
PHACOEMULSIFICATION
Fig. 12.3: Rotation following hydrodissection
Fig. 12.4: Viscodissection to permit the lens to rotate vertically prior phacoemulsificaiton
Viscohydrodissection The parameters change radically when a viscous material (Healon, Provisc, Viscoat or hydroxypropylmethylcellulose–HPMC) is used. Since it takes much more force to inject, one has to be totally sure of the quantity injected; otherwise rupture of the posterior capsule becomes inevitable. In addition, the viscous fluid will force the iris-lens diaphragm forwards, shallowing the anterior chamber, to almost a mere chink. The fluid stays back as it is too viscous to escape from the sides of the 22 G opening normally used for the hydrodissection cannula (Fig. 12.4). Though it is easier to commence, fluid wave is rarely seen unless it is a very immature cataract. In a hydrodissected nucleus with an adequate sized capsulorrhexis, it tends to push the nucleus forwards and prolapse it out of the rhexis opening. The authors utilize viscodissection only after hydrodissection is complete to rotate the edge of the lens forwards in their technique of vertical phacoemulsification. For safety purposes, viscodissection should only be done after hydrodissection with BSS is complete and it is certain that the lens is freely mobile. Caution would dictate that viscodissection, unless used for a specific purpose, may best be left in abeyance. HYDRODELINEATION OR HYDRODELAMINATION Hydrodelineation is the term coined by Anis Aziz to describe the cleavage of lens structures through the injection of fluid. It is also termed hydrodelamination or hydrodemarcation. The technique is fairly simple. A small bore cannula (26–28 G), blunt-tipped, attached to a Luer-lok 1.00 ml plastic syringe, filled with BSS or Ringer lactate, is placed in the middle of the nucleus, and pushed forward into the nucleus till it reaches the middle of the nucleus (in soft cataracts), or meets resistance (medium to hard cataracts). The point of resistance is where the soft outer nucleus meets the harder central nucleus. At the point of resistance, the cannula is pulled back a fraction of a millimeter, and the fluid is injected. The fluid passes into the body of the cataract, and the dissected plane is usually identified by the appearance of a golden ring around the nucleus. This golden ring may not always be visible nor always clear depending on the density of the cataract. Sometimes only a dark separation plane may be noticed. As with hydrodissection, hydrodelamination must produce a cleavage and
HYDRODISSECTION
AND
HYDRODELINEATION
115
Figs 12.5 and 6: Hydrodelineation showing golden rings
a good separation of the nucleus from the epinuclear zone. If only a portion of the ring appears (Fig. 12.5 and 12.6), it may be necessary to reintroduce the cannula in a different place and try to inject the fluid again. It is important to appreciate the differences between hydrodissection and hydrodelamination. Hydrodissection is done to permit phacoemulsification in the bag. If the nucleus did not rotate, it would not be possible to chop a lens, rotate the nucleus for further chopping, and it would also not be possible to allow each sequential piece of the nucleus to be rotated into its best place for removal. None of these techniques could be possible if the nucleus remained adherent to the capsule. On the other hand, hydrodelamination is performed for safety. Hydrodelamination was a necessary preliminary step in the phacoemulsification technique of four-quarter grooving. It tells the surgeon how far he could groove with ultrasound into the periphery without taking any risks. In addition, hydrodelamination involves separation of the peripheral softer epinucleus from the deeper harder nucleus. Thus, in essence, the harder nucleus sits on a softer epinucleus bed. One can phaco the harder part with impunity knowing that the peripheral softer nucleus acts as a buffer safe zone. To recapitulate, the firm nucleus can be worked on within the softer nucleus of almost rubber-like consistency. This particular technique is especially useful with endocapsular phacoemulsification with nuclear cleavage and with the chip and flip technique. Decompression of the Capsule Bag Decompression of the capsule bag means reducing the pressure by allowing the excess fluid to leak out of the sides of the capsular bag and out of the anterior chamber. If the nucleus is hard, hydrodelamination becomes an impossibility, and there even hydrodissection becomes technically difficult. In a dense nucleus the fluid wave is no longer visible and the fluid injection seems to have no effect. The surgeon is tempted to inject more and more at a higher pressure hoping to get a separation, but what does happen however is that the posterior capsule tents backward and as the pressures increases, the thin capsule gives way, rupturing. It is thus a disaster waiting to happen.
116
THE ART
OF
PHACOEMULSIFICATION
How does one recognize that this complication is impending? In a soft cataract even a single injection at a single site under the capsule while doing hydrodissection is adequate. However in the case of a hard or a suprahard cataract, it is important to inject at multiple points. This disperses the fluid and gives more sites for the fluid under pressure, to escape. The second important step is to compress on the nucleus, with the heel of the cannula after every injection, which disperses the fluid. In case the fluid does not come out, two important signs have to be recognized: the chamber shallows very much, and the eye pressure rises sharply. The tendency of the surgeon is to reform the chamber by either injecting a viscoelastic, or still worse, by pressing on the nucleus with a repositor or cannula. Both techniques will lead to a posterior rupture. The correct way of handling this situation is to insert a thin blade iris repository under the edge of the capsule, at 5 O’clock and simply sweep it in both directions. Almost immediately the surgeon will be rewarded with a gush of fluid (which had been entrapped) and the eye immediately softens. Hydrofracture It is a technique, which involves possible separation of the lamella of the internal nucleus by a combination of ultrasonic needle penetration and a pressure injection of BSS. This method is known as a hydrosonic technique and was commenced by Dr Anis of USA. It has the advantage that it fragments the nucleus into little fine bits permitting easier phacoemulsification. However, with the advent of chopping technique it is now rarely utilized and is purely of academic interest. CONCLUSION Both the techniques of hydrodissection and hydrodelamination have the advantage that they also help the surgeon assess the degree of hardness of the lens and therefore, indirectly assess the quantum of ultrasound time, which would be required. Both techniques are essential and though in standard chopping methods hydrodelamination is not utilized, it still is a useful method especially if one expects difficulties to occur during the surgery. It also has the advantage that if a rhexis has been done a little too small, hydrodemarcation reduces the nucleus into its component parts diminishing the size thus, permitting the hard nucleus to be chopped and removed within the soft epinucleus buffer zone. FURTHER READING 1. Fine IH: Cortical cleaving hydrodissection. J Cataract Refract Surg 18(5:) 508-12, 1992. 2. Mehta KR, SM Sathe, SD Karyekar: Computer Terminal Usage and Eye Fatigue, Xth Congress APAO. Soc Proc 2:946-48,1985. 3. Mehta KR: Pitfalls encountered in 1500 consecutive posterior chamber implant. All India Ophthl Soc Proc 165-6,1986. 4. Mehta KR: Phacoemulsification cataract extraction with foldable IOLS—first 50 cases. All India Ophthl Soc Proc 56-60,1989. 5. Mehta KR: Clear corneal phaco with injectable silicone IOL proc. All India Ophthl Soc Proc (Mumbai) 1995.
HYDRODISSECTION
AND
HYDRODELINEATION
117
6. Mehta KR: Mehta tangential chop (MTC) technique for phacoemulsification. All India Ophthl Soc Proc (Chandigarh) 1996. 7. Mehta KR: Combined astigmatic annular keratotomy and phaco—a corneal topographic analytical technique. All India Ophthl Soc Proc (Chandigarh) 1996. 8. Mehta KR: Phaco-levitation—a peaceful way. All India Ophthl Soc Proc (Chandigarh) 1996. 9. Mehta KR: Lollipop phaco cleavage—a new technique for hard cataracts. All India Ophthl Soc Proc (Bangalore) 1991. 10. Mehta KR: SICS mon-phaco—hydroexpression with an irrigating vectis. Proc of SAARC Conference, Nepal, 1994. 11. Mehta KR: Management of subincisional cortex in small incision cataract surgery (SICS). Proc of SAARC Conference, Nepal, 1994. 12. Mehta KR: The new multiport phaco tip for safer, more effective phacoemulsification, with virtually zero capsular damage. Proc of SAARC Conference, Nepal, 1994.
118
THE ART
OF
PHACOEMULSIFICATION
Noshir M Shroff Ranjan Dutta Gurpreet Singh
Phacoemulsification: The Quadrantic Cracking, Chopping and Stuffing Technique
13
INTRODUCTION Phacoemulsification if it proceeds smoothly is an excellent procedure with early and extremely gratifying visual recovery. However, should a complication arise, the result can be disastrous with the patient’s sight under threat. A surgeon’s skills may range from excellent to average. An excellent surgeon will not have much difficulty in adapting to any new procedure including phacoemulsification. Unfortunately not many surgeons belong to this category, where there is little or no difficulty in adapting to the new technique of phacoemulsification. The vast majority of us belong to the second group of “average surgeons”. Not uncommonly an average surgeon begins phacoemulsification, has a few complications in the first few cases, loses confidence and gives up. Therefore, a nucleofractis technique which is useful for the vast majority of average surgeons, would be one which has a high order of safety with least chances of a posterior capsular rent and damage to the corneal endothelium, is easily reproducible and is easy to perform. Modern nucleofractis techniques can broadly be divided into two types: (i) four quadrant cracking, and (ii) stop and chop technique. Four-Quadrant Cracking (Shepherd’s Modification of Gimbel’s Divide and Conquer Technique) It is easier to handle four small quadrants rather than two large halves. Hence, this technique is easier for the beginner. Moreover, central sculpting debulks the hard central nuclear core. However, the disadvantage is that this technique uses more phaco power and phaco time.
QUADRANTIC CRACKING , CHOPPING
AND
STUFFING TECHNIQUE
119
Stop and Chop (Koch’s Modification of Nagahara’s Phaco Chop Technique) This technique has the advantage of using less phaco power and time as compared to quadrantic cracking. It is a very effective technique for hard nuclei. However it is an inherently difficult procedure for the beginner. A nuclear half is simply too big a piece to tackle and most complications of this technique arise due to this fact. The large nuclear cumbersome fragment, if not impaled exactly midway, tends to rock sideways on the phaco tip, particularly if the occlusion is not adequate. The lack of firm grip on the piece makes subsequent chopping frustratingly ineffective. Moreover, the absence of a second trench in the Stop and Chop technique makes it difficult to convert each “half” into “quarters”. As the chopper moves toward the center from the periphery, the final portion (i.e. the central bulk of the nucleus) is difficult to chop. Interconnecting fibrils may at times pose difficulty in separating the fragments, necessitating the use of excessive force with the chopper (Fig. 13.1). Significant lateral forces are thus required for separating the fragments. This puts stress on the capsular bag and makes it an unsafe procedure. In contrast to this, the pre-existent second trench perpendicular to the first, in the quadrantic-cracking technique, makes fragment separation possible with minimal lateral force.
Fig. 13.1: In the Stop and Chop technique (above), the final portion i.e. the central bulk of the nucleus is difficult to chop. Interconnecting fibrils make separation of the fragments difficult. Significant lateral forces are thus created which put stress on the capsular bag. On the other hand in the Quadrantic cracking technique (below), pregrooving leaves a thin posterior nuclear plate and minimum force is required to crack each nuclear half into quarters
120
THE ART
OF
PHACOEMULSIFICATION
The danger of damage to the lens capsule in the Stop and Chop technique is very real. A forceful phaco chop can easily cause a posterior capsule rent. A large fragment will require a longer centripetal chop, which may inadvertently tear the edge of the rhexis and extend it. One may avoid the rhexis edge with a shorter centripetal chop (i.e. by starting at the midperiphery rather than the equator), but this will eventually require wider lateral separation of the nuclear fragments and put stress on the capsular bag. Also, multiple phaco chops give rise to multiple fragments. These can act as splinters, and in conjunction with anterior chamber turbulence, cause endothelial damage. Keeping in mind that most of the above complications occur only due to the large nuclear “halves”, we have devised a technique that combines the best features of both techniques. QUADRANTIC CRACKING, CHOPPING AND STUFFING TECHNIQUE Our technique of quadrantic cracking, chopping and stuffing starts with four-quadrant fracture followed by tackling each quadrant using low power and high vacuum settings. Thus it combines the safety of quadrantic cracking with the efficacy of Stop and Chop and is suitable for most grades of nuclei. Preliminary Steps Following suitable ocular anesthesia, a small conjunctival flap is made and a bloodfree zone is created with a bipolar ca