NEW INSIGHTS INTO MEMBRANE SCIENCE AND TECHNOLOGY: POLYMERIC AND BIOFUNCTIONAL MEMBRANES
Membrane Science and Technology Series Volume 1 : Pervaporation Membrane Separation Processes Edited by R.Y.M. Huang (1991) Volume 2 : Membrane Separations Technology, Principles and Applications Edited by R.D. Noble and S.A. Stern (1995) Volume 3 : Inorganc Membranes for Separation and Reaction By H.R Hsieh (1996) Volume 4 : Fundamentals of Inorganic Membrane Science and Technology Edited by A.J. Burggraaf and L. Cot (1996) Volume 5 : Membrane Biophysics Edited by H. Ti Tien and A. Ottova-Leitmannova (2000) Volume 6 : Recent Advances in Gas Separation by Microporous Ceramic Membranes Edited by N.K. Kanellopoulos (2000) Volume 7 : Planar Lipid Bilayers (BLMs) and their Applications Edited by H.T. Tien and A. Ottova-Leitmannova (2003) Volume 8 : New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes Edited by D. Bhattacharyya and D.A. Butterfield
Membrane Science and Technology Series, 8
NEW INSIGHTS INTO MEMBRANE SCIENCE AND TECHNOLOGY: POLYMERIC AND BIOFUNCTIONAL M E M B I ~ N E S Edited
by
Dibakar Bhattacharyya Department of Chemical and Materials Engineering and Center of Membrane Sciences University of Kentucky, Lexington, KY 40506-0046, USA and
D. Allan Butterfield Department of Chemistry and Center of Membrane Sciences University of Kentucky, Lexington, KY 40506-0059, USA
2003 ELSEVIER Amsterdam
- B o s t o n - L o n d o n - N e w Y o r k - O x f o r d - Paris
San Diego - San Francisco.
Singapore - Sydney - Tokyo
ELSEVIER SCIENCE B.V. Sara Burgerhartstraat 25 P.O. Box 211,1000 AE Amsterdam, The Netherlands © 2003 Elsevier Science B.V. All rights reserved.
This work is protected under copyright by Elsevier Science, and the following terms and conditions apply to its use: Photocopying Single photocopies of single chapters may be made for personal use as allowed by national copyright laws. Permission of the Publisher and payment of a fee is required for all other photocopying, including multiple or systematic copying, copying for advertising or promotional purposes, resale, and all forms of document delivery. Special rates are available for educational institutions that wish to make photocopies for non-profit educational classroom use. Permissions may be sought directly from Elsevier's Science & Technology Rights Department in Oxford, UK: phone: (+44) 1865 843830, fax: (+44) 1865 853333, e-mail:
[email protected]. You may also complete your request on-line via the Elsevier Science homepage (http://www.elsevier.com), by selecting 'Customer Support' and then 'Obtaining Permissions'. In the USA, users may clear permissions and make payments through the Copyright Clearance Center, Inc., 222 Rosewood Drive, Danvers, MA 01923, USA; phone: (+1) (978) 7508400, fax: (+1) (978) 7504744, and in the UK through the Copyright Licensing Agency Rapid Clearance Service (CLARCS), 90 Tottenham Court Road, London W1P 0LP, UK; phone: (+44) 207 631 5555; fax: (+44) 207 631 5500. Other countries may have a local reprographic rights agency for payments. Derivative Works Tables of contents may be reproduced for internal circulation, but permission of Elsevier Science is required for external resale or distribution of such material. Permission of the Publisher is required for all other derivative works, including compilations and translations. Electronic Storage or Usage Permission of the Publisher is required to store or use electronically any material contained in this work, including any chapter or part of a chapter. Except as outlined above, no part of this work may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise, without prior written permission of the Publisher. Address permissions requests to: Elsevier's Science & Technology Rights Department, at the phone, fax and e-mail addresses noted above. Notice No responsibility is assumed by the Publisher for any injury and/or damage to persons or property as a matter of products liability, negligence or otherwise, or from any use or operation of any methods, products, instructions or ideas contained in the material herein. Because of rapid advances in the medical sciences, in particular, independent verification of diagnoses and drug dosages should be made.
First edition 2003 Library of Congress Cataloging in Publication Data A catalog record from the Library of Congress has been applied for. British Library Cataloguing in Publication Data A catalogue record from the British Library has been applied for.
ISBN: 0-444-51175-X ISSN: 0927-5193 ~ T h e paper used in this publication meets the requirements of ANSI/NISO Z39.48-1992 (Permanence of Paper). Printed in Hungary.
TO:
Bhattacharyya: Gloria and Anita, my wife and daughter, for their support and understanding," my graduate students, for making teaching and research life to be very exciting. Butterfield: Marcia and Nyasha, my wife and daughter, who have been so inspirational, encouraging, and loving.
This page intentionally left blank
Preface Membrane techniques provide a broad science and technology base with applications ranging from water purification to chemical/biomolecule synthesis, material recovery, medical devices, to nano-domain interaction- based sensors and highly selective separations. Although there are several books in the traditional membrane field, there is a great need for a highly comprehensive book containing advances in both synthetic and biofunctional/bimimetic membranes including non-invasive characterization to biomedical devices. Over the years the membrane field has advanced from the development of thinfilm composite membranes for desalination to recognition-based separation and reactions by taking advantage of biomolecular interactions in the nano-domain. This refereed book covers materials from highly respected researchers with topics ranging from membrane transport models to non-invasive characterization, functionalized material, biomedical devices, to sensors and environmental applications. Nineteen chapters in this special book are invited, refereed papers, and mostly based on the 2001 North American Membrane Society Annual Meeting held in Lexington, KY, organized and hosted by the University of Kentucky Center of Membrane Sciences. The book is divided into two sections. Section I, is subdivided into three areas, Advances in Membrane Transport~ouling, Imaging Techniques, and Contacting Devices. Section II deals with Functional Membranes and Materials for Biocatalysis, Separation, and Analysis and is further subdivided into three areas ranging from biofunctional membranes to sensors. Authors and co-authors are from various fields including chemistry, chemical engineering, mechanical engineering, biomedical engineering, biotechnology, and environmental engineering. This book is highly multidisciplinary in nature and should be highly valuable to scientists and engineers involved in activities ranging from separations/reactions, to advanced biofunctional materials, to contactor designs, to the general field of membrane science and technology. Students and faculty members around the world should find this to be an excellent reference book for courses ranging from traditional separation, to bioscience/engineering, to formal membrane courses. Each chapter of this book was peer-reviewed, and we would like to give special thanks to the reviewers for this book, including various authors and co-authors of the chapters, and Drs. Kloos (Osmonics Corporation) and Hestekin (Argonne National Lab). Thanks also go to the NAMS Board of Directors, and in particular to NAMS presidents Drs. Pushpinder Puri and Glenn Lipscomb for their encouragement of the book. Mollie Fraim of the University of Kentucky °. Vll
Center of Membrane Sciences did an extraordinary amount of work with us to bring this book to fruition, and we would like to express our sincere appreciation and thanks to Ms. Fraim. The editors would also like to thank Mike Lundin (chemical engineering student) for his hard work in putting the book in right format.
Dibakar Bhattacharyya, Ph.D. Department of Chemical and Materials Engineering and Center of Membrane Sciences University of Kentucky Lexington, KY D. Allan Butteriield, Ph.D. Department of Chemistry and Center of Membrane Sciences University of Kentucky Lexington, KY
viii
About the Editors Dibakar Bhattacharyya (DB) is the University of Kentucky Alumni Professor of Chemical Engineering and a Fellow of the American Institute of Chemical Engineers. He obtained his B.S. (Jadavpur University) and M.S. ( Northwestern University, Evanston, IL) in Chemical Engineering, and his Ph.D. in Environmental Engineering from the Illinois Institute of Technology. He is the Co-Founder of the Center for Membrane Sciences at the University of Kentucky. He and his students have advanced passive membrane applications to functionalized poly-ligand membranes for high capacity metal capture to site-specific biocatalysis for organic degradation to membrane-based nanoparticle synthesis for dechlorination reactions. He has published over 140 refereed journal articles and book chapters, and has recently received three U.S. Patents (two on Functionalized Membranes for Toxic Metal Capture, and one on hazardous waste destruction technology). Dr. Bhattacharyya has mentored many graduate and undergraduate students in the area of water and wastewater related research and membrane separation. For his research, Dr. Bhattacharyya has received funding from U.S. EPA, DoD, NSF, NIEHS-SBRP, Dow Chemical, Glaxo SmithKline, Eastman Chemical Co., Daramic, Inc., etc. He has received a number of awards for his research and educational accomplishments, including the Larry K. Cecil AIChE Environmental Division Award, the Kentucky Academy of Sciences Distinguished Scientist Award, Henry M. Lutes Award for Outstanding Undergraduate Engineering Educator, AIChE Outstanding Student Chapter Counselor Awards, and the University of Kentucky Great Teacher (1984 and 1996) Awards. D. Allan Butterfield received his B.A. in Chemistry from the University of Maine in 1968. Following three years of teaching Africans chemistry and mathematics in Zimbabwe, he entered Duke University, receiving the Ph.D. in Physical Chemistry in 1974. A NIH Postdoctoral Fellowship at the Duke University Neuroscience Program followed. Dr. Butterfield joined the Chemistry Department at the University of Kentucky soon after and was quickly promoted to Full Professor in 1983. In 1986, he and several others, but principally Professor Dibakar Bhattacharyya, formed the University of Kentucky Center of Membrane Sciences, and Professor Butterfield has been Director from its inception. Continuous federal funding from NIH, NSF, and DoD has supported his research on membrane structure and function in neurodegenerative disorders and on applications of biofunctional membranes to important societal problems. He and his students have published over 260 refereed papers. Professor Butterfield received the Southern Chemist Award from the American Chemical Society. He has directed the graduate careers of over 50 Ph.D. and M.S. students, and in 1998 Professor Butterfield received the Presidential Award for Excellence for Science, Mathematics, and Engineering Mentoring from President Clinton in the White House. This award was his efforts to increase the number of female and Appalachian Ph.D. students in Chemistry, both groups being highly under represented in this discipline. In 2002, the University of Kentucky Board of Trustees appointed Dr. Butterfield the Alumni Professor of Chemistry.
ix
This page intentionally left blank
Contents Preface ................................................................................................
vii
About the Editors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ix
List of Contributors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xv
A d v a n c e s in M e m b r a n e T r a n s p o r t / F o u l i n g , I m a g i n g T e c h n i q u e s , a n d C o n t a c t i n g D e v i c e s .............................................................................. 1
Membrane Transport Models, Fouling, and Formation ..................... 3
Mass Transfer in Axial Flows through Randomly Packed Fiber Bundles ......................................................................... 5 L. Bao, G.G. Lipscomb Fouling Phenomena during Microfiltration: Effects of Pore Blockage, Cake Filtration, and Membrane Morphology ....... 27 A.L. Zydney, C. Ho, W. Yuan Differential Scanning Calorimetry and rheological experiments to study membrane formation via thermallyinduced phase-separation .................................................................. 45 P.C. van der Heijden, M.H.V. Mulder, M. Wessling Non-lnvasive Characterizations o f Membrane Fluid Transport and Fouling ........................................................................................ 63
Study of Membrane Fouling and Cleaning in Spiral Wound Modules Using Ultrasonic TimeDomain Reflectometry ...................................................................... 65 Zh.-X. Zhang, A.R. Greenberg, W.B. Krantz, G.-Y. Chai Nonintrusive Characterization of Fluid Transport Phenomena in Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach ......................... 89 C.K. Poh, P.A. Hardy, W.R. Clark, D. Gao Membrane Contactors and Environmental Applications ................ 123
Industrial applications and opportunities for membrane contactors ....................................................................................... 125 R. Klaassen, P.H.M. Feron, R. van der Vaart, A.E. Jansen xi
Membrane Contactors: Recent Developments ................................ 147 A.S. Kovvali, K.K. Sirkar
Membrane Aromatic Recovery System (MARS)A New Process for Recovering Phenols and Aromatic Amines from Aqueous Streams ....................................... 165 F.C. Ferreira, A. Livingston, S. Han, A. Boam, S. Zhang Functional Membranes and Materials for Biocatalysis, Separation, and Analysis ............................................................................................ 183
Biofunctional Membranes and Biomedical Devices ........................ 185 Membrane Bioreactors for Biotechnology and Medical Applications ................................................................... 187 L. Giomo, L. De Bartolo, E. Drioli Structural and Performance Characteristics of Hemodialysis Membranes ............................................................ 219 D. Gao, W.R. Clark Biofunctional Membranes: Site-Specifically Immobilized Enzyme Arrays ........................................................ 233 D.A. Buttertield, D. Bhattacharyya Biocatalytic Membrane Reactor with Continuous Removal of Organic Acids by Electrodialysis .............................. 241 H.C. Ferraz, T.L.M. Alves, C.P. Borges Use of Micro-Porous Affinity Membranes for Protein Purification: A Case Study ............................................... 263 F. Cattoli, G.C. Sarti Economic Production of Biopharmaceuticals by High-Speed Membrane Adsorbers ............................................... 283 W. Demmer, S. Fischer-Frtihholz, D. NuBbaumer, D. Melzner Functionalized Membranes f o r Separations and Reactions ............ 297 Polymer Grafted Membranes ....................................................... 299 S.M.C. Ritchie
xii
Functionalized Membranes for Tunable Separations and Toxic Metal Capture ............................................................... 329 A.M. Hollman, D. Bhattacharyya The Design of High Performance, Gel-Filled Nanofiltration Membranes ........................................................... 353 R.F. Childs, A.M. Mika Sensors ............................................................................................. 377 Membranes for the Development of Biosensors ........................... 379 V.G. Gavalas, J. Wang, L.G. Bachas
Ion-Partitioning Membranes as Electroactive Elements for the Development of a Novel Cation-Selective CHEMFET Sensor System ................................ 393 E.A. Moschou, N.A. Chaniotakis
Index ........................................................................................................ 415
xiii
This page intentionally left blank
List of Contributors
T.L.M. Alves Chemical Engineering Program- COPPE Federal University of Rio de Janeiro, P.O. Box: 68502 21945-970, Rio de Janeiro R J - Brazil Leonidas G. Bachas Department of Chemistry and Center of Membrane Sciences University of Kentucky Lexington, Kentucky 40506-0055 USA E-mail: bachas@pop, uky. edu Lihong Bao Chemical & Environmental Engineering Department Mail Stop 305 The University of Toledo Toledo, Ohio 43606-3390 USA Dibakar Bhattacharyya Department of Chemical and Materials Engineering and Center of Membrane Sciences University of Kentucky Lexington, KY 40506-0046 USA E-mail: db@engr, uky. edu Andrew Boam Membrane Extraction Technology Ltd Imperial College London SW7 2BY UK C.P. Borges Chemical Engineering Program - COPPE Federal University of Rio de Janeiro, P.O.Box: 68502 21945-970, Rio de Janeiro R J - Brazil E-mail: cristiano@peq,coppe, ufrj.br
XV
D. Allan Butterfield Department of Chemistry and Center of Membrane Sciences University of Kentucky Lexington, KY 40506-0059 USA E-mail: dabcns@uky, edu Francesca Cattoli DICMA - Dipartimento di Ingegneria Chimica, Mineraria e delle Tecnologie Ambientali Alma Mater Studiorum - Universit~t di Bologna, viale Risorgimento 2 40136 Bologna Italy G.-Y. Chai Department of Chemical Engineering University of Colorado Boulder, Boulder CO 80309-0424 USA Nikolas A. Chaniotakis Laboratory of Analytical Chemistry Department of Chemistry University of Crete, 71409 Iraklion, Crete Greece Ronald F. Childs Department of Chemistry McMaster University Hamilton, ON, L8S 4M1 Canada E-mail: rchilds@mcmaster, ca William R. Clark, M.D. Renal Division Baxter Healthcare Corporation McGaw Park, IL Nephrology Division Indiana University School of Medicine Indianapolis, IN USA E-mail: bill_clar~axter, com
xvi
L. De Bartolo Research Institute on Membranes and Modelling of Chemical Reactors, IRMERC-CNR c/o University of Calabria via P. Bucci cubo 17/C, 87030 Rende (CS) Italy W. Demmer Sartorius AG G/Sttingen Germany E. Drioli Research Institute on Membranes and Modelling of Chemical Reactors IRMERC-CNR c/o University of Calabria via P. Bucci cubo 17/C, 87030 Rende (CS) Italy E-mail: e. drioli@irmerc, cs. cnr. it
P.H.M. Feron TNO Institute of Environmental Sciences Energy Research and Process Innovation Department Chemical Engineering PO Box 342 7300 AH Apeldoorn The Netherlands H.C. Ferraz Chemical Engineering Program - COPPE Federal University of Rio de Janeiro, P.O.Box: 68502 21945-970, Rio de Janeiro R J - Brazil S. Fischer-Frtihholz Sartorius AG Gfttingen Germany Frederico Castelo Ferreira Department of Chemical Engineering and Chemical Technology Imperial College of Science Technology and Medicine Prince Consort Road London SW7 2BY UK
xvii
Dayong Gao Department of Mechanical Engineering Center for Biomedical Engineering University of Kentucky Lexington, KY USA E-mail: dgao@engr, uky. edu Vasilis G. Gavalas Department of Chemistry University of Kentucky Lexington, Kentucky 40506-0055 USA L. Giomo Research Institute on Membranes and Modelling of Chemical Reactors, IRMERC-CNR c/o University of Calabria via P. Bucci cubo 17/C, 87030 Rende (CS) Italy A.R. Greenberg Department of Mechanical Engineering University of Colorado Boulder, Boulder CO 80309-0427 USA E-mail: greenbea@,spot, colorado, edu Shejiao Han Department of Chemical Engineering and Chemical Technology Imperial College of Science Technology and Medicine Prince Consort Road London SW7 2BY UK Peter A. Hardy Center for Biomedical Engineering University of Kentucky Lexington, KY USA Chia-Chi Ho Department of Chemical Engineering University of Cincinnati Cincinnati, OH 45221 USA
xviii
Aaron M. Hollman Department of Chemical & Materials Engineering University of Kentucky Lexington, Kentucky 40506-0046 USA A.E. Jansen TNO Institute of Environmental Sciences Energy Research and Process Innovation Department Chemical Engineering PO Box 342 7300 AH Apeldoom The Netherlands R. Klaassen TNO Institute of Environmental Sciences Energy Research and Process Innovation Department Chemical Engineering PO Box 342 7300 AH Apeldoorn The Netherlands E-mail: r. klaassen@mep, tno.nl A. Sarma Kovvali Department of Chemical Engineering New Jersey Institute of Technology Newark, NJ 07102 USA W.B. Krantz Department of Chemical Engineering University of Cincinnati Cincinnati, Ohio 45221-0171 USA G. Glenn Lipscomb Chemical & Environmental Engineering Department Mail Stop 305 The University of Toledo Toledo, Ohio 43606-3390 E-mail:
[email protected]
xix
Andrew Livingston Membrane Extraction Technology Ltd. Imperial College London SW7 2BY UK and
Department of Chemical Engineering and Chemical Technology Imperial College of Science, Technology and Medicine Prince Consort Road London SW7 2BY UK E-mail: a. livingston@ic, ac. uk
D. Melzner Sartorius AG Gfttingen Germany E-mail: dieter.melzner@sartorius, com
Alicja M. Mika Department of Chemistry McMaster University Hamilton, ON, L8S 4M1 Canada Elizabeth A. Moschou Laboratory of Analytical Chemistry Department of Chemistry University of Crete, 71409 Iraklion Greece Current address: Department of Chemistry University of Kentucky Lexington KY 40506-0055 USA E-mail: l_moschou@yahoo, com
M.H.V. Mulder University of Twente PO Box 217 7500 AE Enschede The Netherlands D. NuBbaumer Sartorius AG Grttingen Germany
XX
Chum K. Poh Department of Mechanical Engineering University of Kentucky Lexington, KY USA S.M.C. Ritchie Department of Chemical Engineering University of Alabama Tuscaloosa, AL, 35487-0203 USA E-mail: sritchie@bama, ua. edu Giulio C. Sarti DICMA - Dipartimento di Ingegneria Chimica Mineraria e delle Tecnologie Ambientali Alma Mater Studiorum - Universith di Bologna viale Risorgimento 2 40136 Bologna Italy E-mail: giulio.sarti@,mail, ing. unibo, it Kamalesh K. Sirkar Department of Chemical Engineering New Jersey Institute of Technology Newark, NJ 07102 USA E-mail: sirkar@adm, nj it. edu P.C. van der Heijden University of Twente PO Box 217 7500 AE Enschede The Netherlands E-mail: pcvdheijden@hetnet, nl R. van der Vaart TNO Institute of Environmental Sciences Energy Research and Process Innovation Department Chemical Engineering PO Box 342 7300 AH Apeldoom The Netherlands Jianquan Wang Department of Chemistry University of Kentucky Lexington, Kentucky 40506-0055 USA xxi
M. Wessling University of Twente PO Box 217 7500 AE Enschede The Netherlands Wei Yuan Innovasep Technologies 420 Maple Street Marlborough, MA 01752 USA Shengfu Zhang Membrane Extraction Technology Ltd Imperial College London SW7 2BY UK Zh.-X. Zhang Department of Chemical Engineering University of Colorado Boulder, Boulder CO 80309-0424 USA Andrew L. Zydney Department of Chemical Engineering The Pennsylvania State University University Park, PA 16802 USA E-mail:
[email protected], edu
xxii
Advances in Membrane Transport/Fouling, Imaging Techniques, and Contacting Devices
This page intentionally left blank
Membrane Transport Models, Fouling, and Formation
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 1
Mass transfer in axial flows through randomly packed fiber bundles L. Bao and G, G. Lipscomb* Chemical & Environmental Engineering Department, Mail Stop 305, The University of Toledo, Toledo, Ohio 43606-3390 *Corresponding Author, e-mail:
[email protected], Phone: (419) 530-8088, Fax: (419) 530-8086 1.
ABSTRACT
The literature contains numerous correlations of mass transfer coefficients for axial flows through randomly packed fiber bundles. The predictions of these correlations can differ by an order of magnitude. This large variation severely limits the usefulness of the correlations for design purposes and confounds an understanding of the underlying physics. While the literature suggests randomness in fiber packing may contribute to the variation, a rigorous analysis of mass transfer in randomly packed bundles has not been reported. The results of such analyses are summarized here for uniform wall concentration and uniform wall mass flux boundary conditions. The results indicate that channeling through randomly packed bundles can dramatically reduce mass transfer coefficients relative to regularly packed bundles, especially m the welldeveloped limit. However, experimental mass transfer coefficients are significantly different from predicted values for regularly and randomly packed bundles; the literature contains correlations that predict much higher and much lower values. This suggests that other factors can control performance such as cross-flow regions that exist near shell ports or poor fluid distribution from the shell inlet port. 2.
INTRODUCTION
Hollow fiber membrane modules are used for a wide range of applications including dialysis [1], filtration [2], eontaetors [3], and gas separations [4]. Typically, the modules consist of a bundle of randomly packed hollow fibers. The ends of the fiber bundle are potted to form tubesheets - slicing the tubesheets provides access to the fiber interior or lumens. The bundle is inserted into a case and a seal formed between the tubesheet and ease. Ports on the
Mars TranafcrIn Axial Flm Tlmugb Rmdaoly Packed Fiber B d e d - Lipscomb
periphery of the case permit fluid introduction and removal from the shell space (the region outside the fibers) while headers on either end permit fluid introduction and removal from the lumen space (the region inside the fibers). The tubesheet prevents the direct mixing of the shell and lumen fluids. Figure 1 illustrates a typical module. In operation, mass transfer occurs between the fluids flowing through the lumen and shell and the module functions llke the mass exchange equivalent of a shell-and-tube heat exchanger. The material used to produce the hollow fiber membranes can facilitate the desired separation - ideally, the membrane would selectively permeate one or more components of the feed streams at high rates. Through proper choice of membrane material and operating conditions, one can separate the components of a feed stream into hgh-purity product streams. Unfortunately, the efficiency and throughput of these devices can be limited not by the membrane material but by concentration boundary layers that develop in the lumen and shell spaces. The literature [ 5 ] indlcates that the effect
Figure 1. Schematic of a typical hollow fiber membrane module and the case that holds it.
of lumen boundary layers is described well by the Graetz solution [6] for mass transfer coefficients in pipe flows. However, there is much greater uncertainty about the effect of shell boundary layers. Theoretical analyses and experimental measurements of shell-side mass transfer coefficients have been confounded by the complex geometry of the shell and the flows within it. This complexity arises primarily from four factors: 1) fibers are packed randomly in the module whch leads to irregular flow domains that can be described only statistically (i.e., one can speclfy a packing fraction but not the exact location of the fibers) in contrast to the single, well-defined domain that exists in regularly packed bundles; 2) the fibers may not lie parallel to each other which can lead to a transverse flow (cross-flow) component and a velocity field with three components that varies in all three coordinate duections in contrast to a uniaxial velocity field that varies in two coordinate directions in regularly packed bundles; 3) the need to introduce and remove fluid from the shell leads to cross-flow regions and axial-flow regions - concentration boundary layer growth depends strongly on the relative size of these regions but most analyses of heat and mass transfer assume only axial flow; and 4) fluid
6
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
distribution into the bundle may not be uniform if the entrance and exit ports are not well designed- locally flows may be higher or lower than expected based on the assumption of a uniform axial pressure gradient in the bundle (This has been seen experimentally in hemodialyzers [7] where flows appear higher in that part of the bundle closest to the inlet port.). The last two factors can affect performance in both randomly and regularly packed fiber bundles. All four are illustrated in Figure 2. a) cross-sections of bundles with random (left) and regular square (fight) fiber packings o• -N • • o•i O %
/_
. I o og
/oo
•
/000000% ~ooooooo% Ioooooooo Iooooooool
;o_O
i• • OoOp I gO • o o oa,/ gO | ooa= % o - g o • ,=3" \ o • o_O • • ~o % oe-o o //
!
ooooeo
\go•o••/ %~ooo/
b) transverse flows that arise from flow past non-parallel fibers (left) that do not exist when fibers are parallel (right) I
,
''
'
,
'
....
,'
,'
,
, , l
I'
""
, , ,
'
I
i
,
I
I
c) cross-flow regions near shell entrance and exit ports due to fluid distribution from the shell port across the fiber bundle
I
i
"
OOi' I
,
"' gOl
I
,
~:>
got
I
,
,
i
i
i
,,i,,!
, i
i i
ii
ii
ii
d) more fluid may flow through that portion of the bundle closest to the shell inlet port than in other regions .Shell
Ftber
• "
oundle
~
: " ~ = ~ '
inlet
port
hell fluid distribution collar
Figure 2. Four complexities associated with shell flow through hollow fiber membrane modules: a) random fiber packing, b) transverse flows arising from non-parallel fibers, c) cross-flow regions near shell inlet port, and d) poor distribution from shell inlet port.
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles -
Lipscomb
Despite these challenges, the literature offers a number of mass transfer coefficient correlations nominally for axial flows through randomly packed hollow fiber bundles [8-13]. Lipnizki and Field [14] provide an excellent summary and discussion of the reported expressions. The proposed correlations are summarized in Table 1 in terms of the mean Sherwood number: Shr,,=2Rkm/D where R is the fiber radius, km the mean mass transfer coefficient and D the diffusion coefficient. The Sherwood number is a function of the Reynolds number (Re=2RVb,O/lt), Schmidt number (Sc=ldp/D), fiber length (L), and fiber packing fraction (¢~= total fiber cross-sectional area / module cross-sectional area). Here, V5 is the bulk fluid velocity in the shell, ,o the fluid density, and/t the fluid viscosity. Table 1. Literature mass transfer coefficient correlations for axial flows through hollow fiber bundles and the reported range of conditions used to obtain the correlations. In all cases the Schmidt number is approximately 500-1000. Note [15] that the correlation for Prasad and Sirkar [9] has been modified to provide a better fit to their data. Additionally, (1-¢)/~ is the ratio of the hydraulic diameter to the fiber diameter. Correlation
[8]
/, ~'xo.s6/,,,,,~o.93 Sh,,, = 1.25~2-~ l ~ ~ Re°93Sc °'33
[9]
Shin = 11.5
ll-',/[
(1_ +t_._~. j
[10] Shr,, = (0.53-0.58~tL~) [11] Sh,,, = 8.71
ReO.6ScO.33
0001
0
0001
,('1- ~'~-°'2-°16~ Re°'S-°'16~Sc °'33
Sh,,, = (0.31¢2 -0.34~+0.10
Re°'9S¢°'33
0.003
0.004 30
Re°'74Sc °'33
2R/L 0.0020.006
0 00
Re°SaSc°'33
[12] Sh. : 0.090-¢~-7J [13]
100~ Re(1-~)/ # 3-26 0.5-500
0-0.2
0.002
35-80 0-0.5
0.0030.005
10-70 -~0-200
-~0.006
Figure 3 compares the predictions of the mean Sherwood number as a function of Graetz number (defined here as Gz = Re*Sc*(rcR/2L)*(1/~-1) to facilitate comparison with the theory described in the next section) for correlations valid over the range of conditions shown or that require limited extrapolation. The predictions for a regular square packing (constant wall mass flux boundary condition) are given as well; values for triangular packing are just
Lipscomb
Mass Transfer In Axial Flows Through Randomly Packcxl Fiber Bundles -
slightly higher. Clearly, there is significant variation between the correlations. Additionally, the correlations predict values that are b o t h much high and much lower than that for a regular packing. The literature attributes the large differences in mass transfer coefficients to the above four factors. However, a rigorous analysis of the effect of each is not available, and only the first (the effect of random fiber packing) has received much attention. Rogers and Long [16], Wu and Chen [13], and Lipnizki and Field [14] describe similar ad hoe approaches to predicting the effect of random fiber packing. Each set of authors introduces randomness by surrounding a representative set of fibers with an annular region of varying cross-sectional area; the first two pairs of authors use Voronoi tessellation to generate polygonal regions while the last assumes concentric circular regions, see Figure 4. The 10 2
101
Sh~ 10 0
1 0 -1 10
.
.
.
.
.
.
.
.
. 10 2
.
.
.
.
.
. 10 3
Gz
Figure 3. Comparison of literature correlations for mean shell-side mass transfer coefficients in randomly packed bundles with theoretical predictions for regular square packing for []=0.3, Sc=1000, and 2R/L=0.003. The sources ofthe correlations are: o Yang and Cussler [8], + Prasad and Sirkar [9], * Costello et al. [10], x Viegas et al. [11], 0 Gawronski and Wrzesinska [12], and small dot Wu and Chen [13]. The [] represent theoretical predictions for a square packing with uniform wall mass flux from Miyatake and Iwashita [23]. flow through each region is calculated from an appropriate friction factorReynolds number relationship using the region's hydraulic diameter and assuming the pressure drop across each region is the same. Each pair of authors predict a reduction in mass transfer coefficient due to randomness, but the effect of their ad hoe approach on the results is unclear. Results from rigorous analyses of the effect of random packing on mass transfer coefficients for axial flows through fiber bundles are summarized here. As found previously, randomness reduces mass transfer coefficients due to
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lip~omb
channeling in the fiber bundle. The results provide a baseline for evaluation of experimental results and suggest most experimental correlations do not reflect mass transfer in axial flows - either the shell-side flow possessed a significant cross-flow component or flow distribution from the shell ports into the bundle was poor. b) Surrounding each fiber with concentric circular regions
a) Using Voronoi tessellation to surround individual fibers with polygons
Figure 4. Approximate approaches for evaluating the effect of random packing on module performance - mass transfer rates are calculated for each fiber by using the hydraulic diameter of the surrounding shell region to calculate a flow rate and mass transfer coefficient for that region.
3. THEORY Commercial modules contain from 10,000 to over 1 million fibers. Unfortunately, finite computational resources limit the ability to calculate velocity and concentration fields within such large bundles. To reduce computational complexity, we use a unit cell containing a finite number of fibers to represent an infinite, spatially periodic bundle. This unit cell is translated in each co-ordinate direction to generate the bundle; the fibers lie parallel to the zaxis and the unit cell is translated along the x- and y-axes as illustrated in Figure 5. The value of a field variable at a point (x, y) in the unit cell is the same as the value at a point (x + ilx, y + fly) in the infinite bundle where i a n d j are integers and lx and ly represent the unit vectors in the x and y co-ordinate directions, respectively. Fiber packing fraction and type of packing (i.e., a regular or random packing) determine unit cell dimensions. Fiber centers are placed on a regular grid to produce regular packings while fibers centers are placed in a random, sequential fashion to produce random packings. The size and properties of all fibers are assumed to be identical. The random media produced by this method are representative of infinite random media only if the unit cell contains a sufficient number of fibers. 10
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
Calculations with increasing fiber numbers are used to verify that the calculated results do not depend on fiber number- up to 100 fibers were used in the work
lb-
Figure 5. Use of a unit cell to represent an infinite, randomly packed fiber bundle. The unit cell is indicated by the bold square and contains nine solid fibers. The solid lines emanating from the unit cell indicate the required translation to produce an infinite medium. The numbers 1 and 2 correspond to a pair of periodic boundaries: a denotes the top and bottom while b denotes the left and right.
reported here. However, given sufficient computational resources, the algorithms could be used for arbitrarily large fiber numbers. Additionally, one may use this approach to generate bundles containing non-uniform fibers if the unit cell contains a sufficiently large number of fibers to accurately represent the statistical distribution. This would further increase the required computational resources. The governing conservation of momentum and mass equations are solved given the following assumptions: 1. Fluid flow through the fiber bundle is axial, laminar, isothermal and welldeveloped; velocity field entrance lengths are more than an order of magnitude smaller than concentration field entrance lengths [14]. 2. Mass transfer rates are sufficiently low that shell flow rates are constant along the fiber bundle. 3. The wall concentration or wall mass flux along the surface of each fiber in the mass transfer region is constant. 4. Mass transfer by axial diffusion is negligible relative to convective mass transfer. 5. Diffusion is Fickian. 44
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles -
Lipscomb
6. Physical properties of the fluid are constant. These assumptions are equivalent to those used to obtain the Graetz solution for mass transfer in pipes and allow one to write the conservation of momentum and mass equations as"
02V 02V
--+
ax 7
=1
(1)
0¢ 02¢ 02¢
v-- = --+--
Oz
Ox2
(2)
03,2
where v is the dimensionless axial velocity, c the dimensionless concentration, and x, y, and z the dimensionless coordinates. The dimensionless variables are defined as
V V= ~R2 , 0p
X
Y
x=-~-,
y=-~,
/t0Z
ZD Z=R40p
(3 /
p0Z
where P is the pressure. For a constant wall concentration boundary condition, c is given b y C -C w Co - C w
(4a)
where Co is the inlet fluid concentration and Cw the specified wall concentration. For a constant wall flux boundary condition, c is given by: C -C O c=~
(4b)
J~R/D
where Jw is the specified wall mass flux. The boundary conditions for Eq. (1) and (2) are: c-1 v = 0,
or
c=0
c = 0,
(5)
z=0, f o r a l l x , y or
ac/an = 1 along fiber surface
19
(6)
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
=q=, cl, =cl=
(7a) (7b)
1
2
1
2
where n is the unit outward normal direction along the boundary of the solution domain. The first of the two values for the concentration boundary condition given in (5) and (6) is for the constant wall concentration case while the second is for the constant wall mass flux case. The periodicity of the unit cell is captured by Eq. (7). The velocity and concentration are equal along a pair of periodic boundaries (the subscripts 1 and 2 refer to a pair of periodic boundaries as indicated in Figure 5) while the gradients of velocity and concentration are equal in magnitude but opposite in sign because of the difference in unit outward normal directions for the two boundaries. The axial velocity field, v, is independent of axial distance, z, because of the well-developed velocity field and low mass transfer rate assumptions. Consequently, one can solve for v, substitute it into Eq. (2), and avoid solving the two equations simultaneously. Since an analytical solution cannot be found, a numerical approximation to the solution for v is obtained using the Finite Element method as described in the next section. If shell flow rate changes significantly due to flow across the fiber wall, one may still use the theoretical approach described here but must solve the conservation of momentum equation simultaneously with the conservation of mass equation - this greatly increases the computational complexity. Solving the conservation of mass equation, Eq. (2), for the concentration field allows one to calculate mass transfer coefficients. The effective, overall mass transfer coefficient, k, is given by: N
Z I (Cb -cw),Rao k(2n'RN ) i=1 2xRN
N
= ~, I2x D ( O C / O r ) ~ , i R d O i=1
(8)
where N is the number of fibers in the unit cell, Cb the mixing cup average bulk concentration given by: CVdA Cb = A
(9)
IvdA
A
and A the area available for flow. Eq. (8) can be written in terms of a local Sherwood number as: 13
Mass TransferIn Axial Flows ThroughRandomlyPacked Fiber
Bundles - L i p s c o m b
N
2Rk
Sh = ~ =
~,f2o~(a¢/a¢)w,iao
2 i=l
D
iv y, f2~r(c b _ c w l i d O
(10)
i=1
where ~ is the dimensionless radial coordinate (r/R). For a constant wall concentration boundary condition, Eq. (10) reduces to:
Sh e = 2 * 0¢/0.....~ Co
(11)
while for a constant wall mass flux boundary condition it reduces to: 2 Shf = (cb _cw )
(12)
where the overbar indicates the average value along the surface of each fiber in the unit cell. Values for the Sherwood number will be computed as a function of the Graetz number: (13)
Gz = Q/DL = Re*Sc*(xR/2L)*(1/~- 1)
where Q is the flow rate per fiber. In the large Graetz number limit (either high flow rates or short mass transfer regions), concentration boundary layers between adjacent fibers are much smaller than the fiber spacing. This limit is referred to as the entry mass transfer limit. In the small Graetz number limit (either low flow rates or long mass transfer regions), concentration boundary layers between adjacent fibers completely overlap. This limit is referred to as the well-developed mass transfer limit. 4.
NUMERICAL APPROXIMATIONS The Finite Element method [17] is used to obtain a numerical approximation to the governing conservation equations. To discretize the shell domain, V oronoi tessellation is used to surround each fiber in a unit cell with a polygon. The external boundary of these regions forms the boundary of the unit cell as illustrated in Figure 6. The region within each polygon is further discretized to form the finite element mesh. Note that the unit cell boundary consists of a series of line segments, instead of the four sides of the square unit cell in Figure 5, thus the periodic boundary consists of a sequence of line 14
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles -
Lipscomb
segment pairs (see Figure 6). The use of this boundary facilitates the calculations, because one does not have to identify where fibers cross the boundary. The velocity field is calculated first from the Finite Element discretization of Eq. (1) and substituted into Eq. (2). Although the velocity field is independent of axial position, z, the concentration field is not and the Finite Element discretization of Eq. (2) leads to the following set of order differential equations: M -dc - + Kc = F dz
(14)
where c is now a vector containing the concentration at each node in the Finite Element mesh, M and K are coefficient matrices and F is a coefficient vector. lb
1~ \''~ S.
Figure 6. Use of Voronoi tessellation to generate the boundary of the unit cell and discretize the shell domain. Two pairs of periodic boundaries are indicated: (la, 2a) and (lb, 2b). A small portion of the actual mesh is also shown. These equations are solved by approximating dc/dz with a second-order, backward Finite Difference and marching along the z-direction to calculate the concentration field from the start to the end of the mass transfer region. A variable axial step size (Az) is used to speed up the calculations - the smallest values are used at the start of the mass transfer region where large concentration gradients exist. The calculations are stopped when the difference between the last two mass transfer coefficients is less than 0.001 in magnitude (approximately 0.1% change) or the bulk concentration is less than 10-6. One can avoid this computationally intensive solution for the concentration field as a function of z in the entry mass transfer limit. For large Graetz numbers, the relationship between Sh and Gz is given by [18]:
15
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles -
Lipscomb
(15)
Sh = a G z 1/3
where the constant a depends on ~b and the wall boundary condition. As demonstrated previously [19], a can be calculated from the velocity field alone for both constant wall concentration and mass flux boundary conditions, so one does not have to obtain a numerical approximation to Eq. (2). In the welldeveloped mass transfer limit, the relationship between Sh and Gz is of the form
[2o]: (16)
Sh = fl
where the constant fl also depends on ¢~and the wall boundary condition. As demonstrated previously [21], flf (the subscript f indicates a constant wall mass flux boundary condition) can be calculated from that portion of the welldeveloped concentration field that is independent of z. One must obtain a numerical approximation to the well-developed portion of c, but marching along the z-axis is not required. Unfortunately, a similar solution does not exist for tic (the subscript c indicates a constant wall concentration boundary condition) and one must solve for the concentration field as a function of x, y, and z to evaluate Values for the Sherwood number calculated using each of these approaches are summarized in the next section. Note that one may use the marching solution procedure to obtain values for a and fl for both boundary conditions. The results obtained by marching are in excellent agreement with the results for ac, o?, and flfobtamed using the simplifications just described. Mesh refinement analyses are performed to evaluate discretization error. The number of nodes is increased by 50% until the calculated mass transfer coefficients change by less than 5% over the range of Graetz numbers considered. Additionally, mass transfer coefficients are calculated for regular arrays to validate the computational approach. 5.
RESULTS AND DISCUSSION Figure 7 illustrates the calculated variation of the local Sherwood number with Graetz number for triangular packings and a constant wall concentration boundary condition using the marching solution approach. The results are compared to those reported by Miyatake and Iwashita [22] for the equivalent heat transfer problem. The agreement between the two is very good and is comparable to that observed for square packings. We believe the minor differences that exist are due primarily to differences between Miyatake and Iwashita's actual results and the correlation they propose to represent the results; the correlation was used to generate Figure 7. 46
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles Lipscomb -
Table 2 contains the values of a and fl calculated for triangular and square packings as a function of packing fraction and wall boundary condition. The computational approach used to evaluate each is indicated. Miyatake and Iwashita's results [22, 23] are also given; note that these values correspond to specific results they report and are not generated from the correlation they propose. The agreement between the two is excellent which helps validate the work summarized here. .
.
.
.
.
.
.
.
i
.
.
.
.
.
.
.
.
i
.
.
.
.
.
.
.
.
,
.
.
.
.
.
.
.
.
,
.
.
.
.
.
.
.
.
0.75 102
O.e3 0.40 0.23
Sh 10
.......
o
101
10 2
10 3
10 4
10 s
Gz
Figure 7. Variation of local Sherwood number with Graetz number for triangular fiber packings and constant wall concentration boundary condition. The solid lines are the results obtained using the marching algorithm described here and the dashed lines are the correlation proposed by Miyatake and Iwashita [22]. The values of ~ are indicated on the curves.
In-between the entry and well-developed mass transfer limits, the following expression provides a simple correlation of the relationship between Sh and Gz [24]:
Sh = (fl3 + ot3Gz~/3
(17)
Figure 8 compares the Sherwood numbers generated by this correlation and the actual calculated values for regular triangular arrays and a constant wall concentration boundary condition. Differences between the two are small except for the highest fiber packing, ~b= 0.75, which is outside the range of commercial interest; note that fiber packing ranges from 0.4 to 0.6 in most commercial modules. Consequently, mass transfer coefficients calculated for random packings will be reported in terms of a and fl and Eq. (17) will be used to calculate Sh. Figure 9 illustrates the differences between velocity fields calculated for a square packing and typical random packing with 0~=0.50. Regions of high flow where fiber packing is lowest are dearly visible in the randomly packed bundle. 17
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
This flow channeling is undesirable and reduces mass transfer coefficients. Where fibers are dose, only a few contours are visible indicating low velocities and flows. Such regions are nearly stagnant and also undesirable. Table 2. Values of a and fl calculated for triangular and square packings. Values for a~ and af were taken from [ 19]. Values for/~ were taken from [21]. Values for tic were obtained using the marching algorithm described in the text. Triangular army - Constant wall concentration boundary condition ot~ This Work Miyatake-Iwashita [22] This Work 0.823 3.15 3.15 9.49 0.750 2.61 2.56 9.94 0.630 1.93 1.88 9.89 0.403 1.12 1.11 6.92 0.227 0.70 0.71 4.21 0.057 0.31 0.32 1.98
Miyatake-Iwashita [22] 9.52 9.97 9.89 6.86 4.17 1.96
Triangular army - Constant wall mass flux boundary condition a? This Work Miyatake-Iwashita [23] This Work 0.823 3.74 3.64 4.89 0.750 3.14 3.02 8.78 0.630 2.33 2.28 11.8 0.403 1.35 1.39 7.59 0.227 0.84 0.87 4.48 0.057 0.37 0.37 2.05
Miyatake-Iwashita [23] 4.92 8.75 11.6 7.49 4.41 2.04
Square array - Constant wall concentration boundary condition o~ This Work Miyatake-Iwashita [22] This Work 0.712 1.80 1.81 4.05 0.649 1.67 1.68 4.13 0.545 1.41 1.41 4.27 0.349 0.95 0.94 4.19 0.196 0.62 0.62 3.34 0.049 0.28 0.28 1.81
Miyatake-Iwashita [22] 4.07 4.17 4.33 4.21 3.29 1.82
Square array - Constant wall mass flux boundary condition o? ¢~ This Work Miyatake-Iwashita [23] This Work 0.712 2.07 2.06 2.29 0.649 1.97 1.92 3.12 0.545 1.70 1.63 4.42 0.349 1.14 1.14 4.98 0.196 0.75 0.77 3.68 0.049 0.34 0.35 1.89
Miyatake-Iwashita [23] 2.31 3.11 4.32 4.74 3.45 1.67
18
Mass Transfer In Axial Flows
Through
Randomly
Packed
Fiber Bundles
Lipscomb
-
Figure 10 illustrates the concentration field in the well-developed limit for the same randomly packed unit cell as shown in Figure 9. The concentration is nearly zero everywhere except in the lowest fiber packing region which corresponds to the highest flow region in Figure 9. Fluid residence time is lower in the low packing regions than in the high packing regions resulting in smaller concentration changes - one may envision a low packing region as a channel that allows fluid to "by-pass" the mass transfer region. The low packing regions dominate overall mass transfer performance because flow rates are much higher through them than high packing regions. ........
,
........
|
........
|
........
|
........ 0.75
102 0.63 0.40
0.23
Sh 101
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
........
i
104
.......
105
Gz
Figure 8. Comparison of the proposed local Sherwood number correlation (dashed lines) with calculated results (solid lines) for triangular packings and constant wall concentration boundary condition. The values of # are indicated on the curves.
F4 (a) (b) Figure 9. Contours of constant velocity in (a) regular square and (b) random fiber packings.
19
Mass Transfer In Axial Hows Through Randomly Packed Fiber Bundles - Lipscomb
Figure 10: Contours of constant concentration in a randomly packed bundle in the welldeveloped mass transfer limit for a constant wall concentration boundary condition. Table 3. Values of a and fl calculated for random packings. Random array - Constant wall concentration boundary condition a~ [19] Fiber Min. Max. Ave. Fiber Number Number 0.30 9 0.580 0.709 0.643 32 0.40 9 0.799 0.889 0.860 32 0.50 9 0.971 1.19 1.08 9 0.50 16 1.01 1.21 1.08 16 0.50 32 32 0.50 64 64 0.50 100 1.12 1.15 1.14 100 0.55 9 1.09 1.33 1.22 32 0.575 9 1.17 1.46 1.32 32 Random array - Constant wall mass flux af [19] Fiber Min. Max. Number 0.30 9 0.604 0.808 0.40 9 0.50 9 1.08 1.43 0.50 16 1.09 1.31 0.50 32 0.50 64 0.50 100 1.23 1.28 0.55 9 1.28 1.40 0.575 9 1.11 1.53
Min.
Max.
Ave.
0.484 0.480 1.11 0.914 0.694 0.758 0.676 0.713 0.615
0.840
0.623 0.918 1.71
1.23 2.83 2.08 1.27
1.27
1.19
0.968 0.987 0.800 0.928 0.904
]~ [21] Max. Min.
Ave.
0.185 0.329 0.225 0.153 0.165 0.142 0.175 0.154 0.098
0.213 0.355 0.460 0.342 0.331 0.247 0.252 0.197 0.144
1.20 0.908
1.20
boundary condition Ave.
Fiber Number 64 64 9 16 32 64 100 64 64
0.742 1.25 1.20 1.27 1.34 1.35
20
0.275 0.387 0.657 0.504 0.452 0.373 0.415 0.256 0.199
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
Values calculated for a and fl for randomly packed bundles are summarized in Table 3. The maximum, minimum, and average values for at least five different unit cells are given. Table 3 also indicates the effect of unit cell fiber number. For 50% fiber packing, a and fl were calculated for unit cells containing 9, 16, 32, 64, or 100 fibers. An analysis of the variance (ANOVA [25]) indicates there is no significant difference between the results at the 95% confidence level if a sufficiently large number of fibers is used: at least 9 for c~c and o~, 32 for tic, and 64 for fl~ The results of this analysis were used to specify the fiber number for calculations with other ~ values. A comparison of Tables 2 and 3 indicates the detrimental effect of random fiber packing on mass transfer performance. Relative to a square packing, the random packing values for ac and af are-~20% lower. The decrease in the welldeveloped limit is much more dramatic: tic is 80% lower while flf is 90% lower. This detrimental effect is graphically illustrated in Figure 11 for ~=0.40 where the rapid decline in performance with decreasing Gz is readily apparent. 102
....................................
• .......
101
J J
Sh
J J J 10 0 ___.
1 0110_ 1
J
__-.-
10 °
101
10 2
10 3
04
GZ
Figure 11. Variation of local Sherwood number with Graetz number for regular square (solid line) and random (dashed line) fiber packings with ~=0.40 and constant wall concentration boundary condition. The following correlations accurately capture the dependence of a and fl on fiber packing fraction for randomly packed bundles - for a constant wall concentration boundary condition:
(18a) ( 8b)
ac = 2.44~- 0.097 tic = - 1 0 . 2 g + 9.92~- 1.43 and for a constant wall mass flux bounda~ condition:
21
Mass Transfer In Axial
Flows T h r o u g h
R a n d o m l y Packed Fiber Bundles -
Lipscomb
(19a)
off= 2.30~ + 0.061 f l f = - 8 . 1 5 g + 6 . 8 2 ~ - 1.09
(19b)
Substituting Eq. (18) and (19) into Eq. (17) gives a correlation that can be used to predict local shell-side mass transfer coefficients for randomly packed bundles as a function of G z and ~k (0.30< ~<0.575). The m e a n mass transfer coefficient, or equivalently Sherwood number is given by [26]:
Sh,,, = Lo
(20)
az
where the subscript m indicates mean value. Evaluating the m e a n Sherwood number in the entry and well-developed mass transfer limits leads to the following modification of Eq. (17) as a correlation for Shr,,:
+ 0 s ) oz] where a and fl are given by Eq. (18) and (19), respectively. The predictions of Eq. (21) for a constant wall concentration are compared to various literature correlations in Figure 12. Only those correlations 10 2
.
.
.
.
.
.
I
.
.
.
.
.
.
101
Sh.
10 0
1 0 "1
101
.
.
.
.
.
.
.
.
i
10 2
.
.
.
.
.
.
.
.
10 3
Gz
Figure 12. Comparison of calculated Sherwood numbers for randomly packed bundles with experimental correlations for 4~=0.50, Sc=1000, and 2R/L=0.003. The sources of the correlations are: * Costello et al. [10], 0 Gawronski and Wrzesinska [12], and small dot Wu and Chen [13]. The n represent theoretical predictions for a square packing and the line without symbols represents the random packing results both for constant wall concentration. valid for the chosen operating conditions are shown, and the predictions for a constant wall mass flux are just slightly higher than the constant wall
22
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
concentration values. Although the experimental boundary condition is neither constant wall concentration or mass flux, one expects the theoretical predictions to provide bounds for the experimental values as observed for the lumen mass transfer coefficient. Clearly, the available correlations predict mean Sherwood numbers that are both much higher and much lower than theoretical predictions for either regular or random fiber packings. Why is this the case? One may attribute the positive (higher values) deviations to the presence of significant cross-flow effects that arise from either transverse flows engendered by non-parallel fibers or fluid distribution from the shell port (see Figure 2). As in heat exchange, concentration boundary layers do not increase as rapidly in cross-flow as in axial flow. The reduction in concentration boundary layer thickness increases mass transfer coefficients. One may attribute the negative (lower values) deviations to a combination of the other complexities of shell-side flow illustrated in Figure 2. First, poor fluid distribution from the shell inlet port may lead to higher flows in some regions and lower flows in others than one would expect if the axial pressure gradient was uniform throughout the bundle. Radial variations in the axial pressure gradient would be expected if the shell port and associated fluid distribution collar are not designed well: one would expect slightly higher axial pressure gradients closer to the port due to pressure drop through the distribution collar around the bundle. Evidence for higher flows in that portion of the bundle closest to the inlet port (see Figure 2) is given in the literature [7]. Poor packing of the fibers near the case would exacerbate this effect, since gaps between the bundle and case serve as conduits for fluid bypassing or channeling. Finally, the presence of large cross-flow regions may reduce the effective mass transfer coefficient due to poor utilization of the chemical potential driving force for transport. The lower driving force would be partially countered by the reduction in concentration boundary layer thickness discussed previously. Thus, mass transfer coefficients might decrease or increase depending on the extent and location of cross-flow regions. 6.
CONCLUSIONS Calculations of mass transfer coefficients for axial flows through randomly packed fiber bundles are presented. The results indicate randomness in fiber packing reduces mass transfer coefficients relative to structured, regular fiber packings. The reduction is most pronounced in the well-developed mass transfer limit, Gz
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
regions due to shell port design or non-parallel fibers may lead to higher than expected mass transfer coefficients. Poor fluid distribution from the shell inlet port into the bundle or excessive cross-flow may lead to lower than expected values. Additional work is required, though, to quantitatively evaluate these effects. 7.
ACKNOWLEDGMENT The authors would like to acknowledge partial support of this work by the National Science Foundation through Grant CTS-9408414.
So A=
NOTATION 2 flow area, m 6,= dimensionless concentration C = concentration, mol/m 3 D = diffusion coefficient, m2/s vector from source term, generated by Finite Element discretization F= Gz = Graetz number (Q/DL) i= number of unit cell translations j = number of unit cell translations Jw = mass flux, mole/m:/s k= mass transfer coefficient, m/s K = coefficient matrix 1= unit vector L = module length, m M = coefficient matrix n= unit outward normal direction N = number of fibers in the unit cell P = pressure, Pa Q = flow rate, m3/s r= radial coordinate, m R = fiber radius, m Re = Reynolds number (2RVbp/,u) Sc = Schmidt number (/1/pD) Sh = Sherwood number (2Rk/D) dimensionless axial velocity V-axial velocity, m/s V = dimensionless coordinate x:" coordinate value, m X= y= dimensionless coordinate y= coordinate value, m z'-dimensionless axial coordinate Z = coordinate value, m 24
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles - Lipscomb
At = p= 0=
8°1.
0= b= ¢=
f= m= w = X =
y=
coefficient in mass transfer coefficient correlation coefficient m mass transfer coefficient correlation fiber packing fraction fluid viscosity, Pa- s density, kg/m 3 angular coordinate dimensionless radial coordinate
Subscripts and superscripts initial value bulk value evaluated for a constant wall concentration evaluate for a constant wall mass flux mean value wall value x-coordinate value y-coordinate value
REFERENCES [1]
[21 [31 [4]
[5] [6]
[7]
[8] [91 [10]
[111
J.T. Daugirdas, T.S. Ing, P.G. Blake, Handbook of Dialysis, 3rd Ed., Little Brown & Co., Boston, 2000. M. Cheryan, Ultmfiltration and Microfiltration Handbook, Technomic Publishing Co. Inc., Lancaster-Basel, 1998. A. Gabelman, S.T. Hwang, Hollow fiber membrane contactors, J. Membr. Sci. 159 (1999) 61. J. Henis, Commercial and practical aspects of gas separation membranes, pp. 441-512, Polymeric Gas Separation Membranes, D.R. Paul and Y.P. Yampol'skii, Eds., CRC Press, Boca Raton, 1994. L. Dahuron, E.L. Cussler, Protein extraction with hollow fibers, AIChE J. 34 (1988) 130. E.L. Cussler, Diffusion Mass Transfer in Fluid Systems, Cambridge University Press, Cambridge, 1997. A. Frank, G.G. Lipscomb, M Dennis, Visualization of concentration fields in hemodialyzers by computed tomography, J. Membr. Sci. 175 (2000) 239. M.C. Yang, E.L. Cussler, Designing hollow fiber contactors, AIChE J. 32 (1986) 1910. R. Prasad, K.K. Sirkar, Dispersion-free solvent extraction with microporous hollowfiber modules, AIChE J. 34 (1988) 177. M.J. Costello, A.G. Fane, P.A. Hogan, R.W. Schofield, The effect of shell side hydrodynamics on the performance of axial flow hollow fiber modules, J. Membr. Sci. 80 (1993) 1. R.M.C. Viegas, M. Rodriguez, S. Luque, J.R. Alvarez, I.M. Coelhoso, J.P.S.G. Crespo, Mass transfer analysis of Wilson-plot methodology, J. Membr. Sci. 145 (1998) 129.
25
Mass Transfer In Axial Flows Through Randomly Packed Fiber Bundles -
[12] [13] [14]
[15] [16]
[17] [18] [19] [20] [21]
[22]
[:3] [24]
[25] [26]
Lipscomb
R. Gawronski, B. Wrzesinska, Kinetics of solvent extraction in hollow-fiber contactors, J. Membr. Sci. 168 (2000) 213. J. Wu, V. Chen, Shell-side mass transfer performance of randomly packed hollow fiber modules, J. Membr. Sci. 172 (2000) 59. F. Lipnizki, R.W. Field, Mass transfer performance for hollow fibre modules with shell-side feed flow: using an engineering approach to develop a framework, J. Membr. Sci. 193 (2001) 195. A.S. Sangani, Private communication. J.D. Rogers, R.L. Long, Modeling hollow fiber membrane contactors using film theory, Voronoi tessellations, and facilitation factors for systems with interface reactions, J. Membr. Sci. 134 (1997) 1. J.N. Reddy, D.K. Gartling, The Finite Element Method in Heat Transfer and Fluid Dynamics, CRC Press, Boca Raton, FL, 1994. L.G. Leal, Laminar Flow and Convective Transport Processes, ButterworthHeinemann, Boston, 1992. L. Bao, B. Liu, G.G. Lipscomb, Entry mass transfer in axial flows through randomly packed fiber bundles, AIChE J. 45 (1999) 2346. W.M. Deen, Analysis of Transport Phenomena, Oxford University Press, New York, 1998. L. Bao, G.G. Lipscomb, Well-developed Mass Transfer in Axial Flows Through Randomly Packed Fiber Bundles with Constant Wall Flux, Chem. Eng. Sci. 57 (2002) 125. O. Miyatake, H. Iwashita, Laminar flow heat transfer to a fluid flowing axially between cylinders with a uniform surface temperature, Int. J. Heat Mass Transfer 33 (1990)417. O. Miyatake, H. Iwashita, Laminar flow heat transfer to a fluid flowing axially between cylinders with a uniform wall heat flux, Int. J. Heat Mass Transfer 34 (1991) 322. H. Kreulen, C.A. Smolders, G.F. Versteeg, W.P.M. van Swaaij, Microporous hollow fiber membrane modules as gas-liquid contactors. Part 1. Physical mass transfer processes, J. Membr. Sci. 78 (1993) 197. E.A. McBean, F.A. Rovers, Statistical Procedures for Analysis of Environmental Monitoring Data & Risk Assessment, Prentice Hall, Upper Saddle River, NJ, 1998. W.M. Kays, M.E. Crawford, Convective Heat and Mass Transfer, McGraw-Hill, New York, 1980.
26
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 2
Fouling phenomena during microfiltration: effects of pore blockage, cake filtration, and membrane morphology Andrew L. Zydney *+, Chia-Chi Ho s, Wei Yuan # +Department of Chemical Engineering, The Pennsylvania State University, University Park, PA 16802 SDepartment of Chemical Engineering, University of Cincinnati, Cincinnati, OH 45221 #Innovasep Technologies, 420 Maple Street, Marlborough, MA 01752 * Corresponding Author, Phone: 814-863-7113, Fax: 814-865-7846, e-mail:
[email protected] 1.
INTRODUCTION
Fouling remains a major problem in many membrane filtration processes. Particulate matter can deposit on or within the membrane pore structure, significantly increasing the overall hydraulic resistance to flow. This can lead to dramatic reductions in the filtrate flux, ultimately requiring mechanical/chemical cleaning or complete replacement of the filter media. The reduction in filtration flux (or velocity) during constant pressure filtration is typically described using one of the classical fouling models. These blocking laws were originally developed by [1] by assuming that the filtering medium was composed of a parallel array of capillary cylindrical pores of constant and equal diameter. The complete blocking law assumes that particulate matter deposits on the membrane surface and completely blocks (or seals) the capillaries. The standard blocking law assumes that the particles deposit throughout the pore volume, causing a uniform constriction m the pore radius. It is also possible for the particles to form a growing deposit or cake layer on the external surface of the filter. The cake filtration model is developed by assuming that the rate of cake growth is uniform across the membrane surface and is directly proportional to the filtration velocity. Although these fouling models have been used quite extensively to analyze filtration data, the assumptions that underlie these models are almost never satisfied in actual practice. Most filtration media have a highly irregular pore structure formed by the void space between the polymer fibers or globules that make up the membrane or filter. The effective size of these pores can vary dramatically both laterally and with depth through the membrane filter. In
27
Fouling Phenomena During Microfiltrafion: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
addition, particle deposition on the upper surface of such a membrane will be unable to completely block or seal the pores since the fluid can flow under and around any surface blockage due to the highly interconnected nature of the pore structure. A cake layer may form on the external surface of these membranes, but the cake growth rate is likely to be non-uniform, with those regions of the filter that were fouled first having the thickest cake layer. Finally, the actual fouling in any given system is likely to be due to a combination of mechanisms, with pore blockage and cake growth occurring simultaneously during the filtration process. Researchers [1-3] showed that the cake filtration and blocking laws could all be written in a common mathematical form as: d:t- k(dt n dV2 dV )
(1)
where t is the filtration time and V is the total filtered volume. The exponent n characterizes the filtration model, with n = 0 for cake filtration, n = 3/2 for standard blocking, and n = 2 for complete blocking. Hermans and Bred6e [1] also coined the term "intermediate blocking" to describe the filtration behavior when n = 1, although no physical interpretation of this situation was provided. Hermia [4] subsequently showed that this intermediate blocking law could be derived from the complete blocking model by allowing for the superposition of deposited particles on the filter surface. More recently, derived the governing filtration equations for 11 different values of n ranging between 0 and 2, including n = 1/4, 1/3, 1/2, 2/3, 3/4, 5/4, and 4/3 were derived [5], although no physical interpretation was provided for these model equations. Equation (1) has often been used to analyze experimental results during membrane filtration; however, much of the flux decline data obtained in these studies is difficult if not impossible to explain using this simple theoretical framework. For example, analyzed filtrate flux data during microfiltration of bovine serum albumin (BSA) through track-etch polycarbonate membranes with straight-through cylindrical pores were analyzed [6], and these researchers found a distinct maximum in the plot of dZt/dV2 versus dt/dV. The slope on the log-log plot thus becomes negative at large dt/dV, corresponding to a negative value of n in Eq. (1). Similar behavior was seen for BSA filtration through nitrocellulose microfiltration membranes having a more tortuous and irregular pore structure [7]. In this case, the value of the slope on the log-log plot decreased throughout the filtration, going from a value of more than 6 at the start of the experiment to a value less t h a n - 2 at long filtration times. Such extreme values of n have never been explained, nor has there been any quantitative explanation for the large variation in n observed over the course of the filtration. This type of unusual filtration behavior is in no way limited to solutions of proteins like bovine serum albumin. For example, studied the flux decline during filtration of 28
Fouling Phenomena During Microfiltrafion: Effects Of Pore Blockage, Cake Fillrafion, And Membrane Morphology - Zydney
a sodium acetate suspension in methanol through a kaolin membrane preformed on a stainless steel support was reported [8]. Data were analyzed using Eq. (1), with the calculated values of the slope being consistently less than zero. Values as small as -15 were reported for experiments performed at high temperature and pressure. This type of filtration behavior was denoted as "decelerating resistance" [8], although no clear physical explanation was provided for this phenomenon. We have recently developed a new approach for describing the flux decline during membrane microfiltration that addresses many of the shortcomings in the classical fouling models and is able to explain much of the unusual flux decline phenomena reported in the literature. The following section of this Chapter describes the basic principles for this combined pore blockage - cake filtration model, including the effects of the underlying pore morphology on the flux decline. Section 3 examines the behavior of this theoretical model, using experimental data from the literature to help illustrate many of the key phenomena. The final section summarizes the key results and discusses the implications of this new understanding of membrane fouling for the design and operation of improved filtration processes. 2.
PORE B L O C K A G E - CAKE FILTRATION MODEL
Although fouling can occur by specific chemical interactions between feed components and the membrane material, our focus is on the fouling associated with the deposition of particulate matter on the upper surface of the membrane filter. This physical situation will certainly be appropriate for the filtration of suspensions in which the particles are larger than the membrane pore size. In addition, several studies have demonstrated that fouling during protein microfiltration occurs primarily by the physical deposition of large protein aggregates on the external membrane surface [9]. This type of aggregate deposition is also the primary cause of humic acid fouling during microfiltration [10]. The rate of surface coverage (pore blockage) is assumed to be proportional to the mass flow rate of large particles/aggregates to the open (unfouled) surface of the membrane: dA
open = _ aQopenCb dt
(2)
where Qopenis the volumetric flow rate through the open pores and Cb is the bulk concentration of large aggregates, o~ is a pore blockage parameter and is equal to the membrane area blocked per unit mass of aggregates convected to the membrane. Cb depends upon the properties of the protein and buffer
29
Fouling Phenomena During Microfillration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
solution. Unlike most prior work, we allow for the possibility that the blocked pores may be partially permeable to filtration, e.g., the fluid may be able to flow through the interstitial space between individual macromolecules in the aggregate. As the membrane surface becomes more heavily fouled, the large particles or aggregates will also begin to deposit directly on the fouling layer, increasing its hydraulic resistance to flow. This is exactly what occurs in the classical cake filtration model, although in this case the cake growth is assumed to occur simultaneously with the coverage of the remaining open area of the membrane. The rate of deposit growth over each blocked region of the membrane is assumed to be proportional to the particle mass flux to that particular region: dm dt
(3)
P = Obloeked - J b ~ k ) f ' b
where mp is the protein mass deposited per unit membrane area and Jb,ck accounts for any back particle flux associated with shear induced diffusion, inertial particle lift, surface migration, or long-range repulsive interactions between the particles and cake layer [ 11 ]. C'b is the concentration of particulate matter that adds to the growing deposit. C'b can be greater than Cb if the deposit is able to retain a greater fraction of the feed stream during the filtration pr'ocess. In order to integrate Eqs. (2) and (3), we first need to develop expressions for the volumetric flow rate or filtrate flux (J = Q/A) through the open and blocked pores. These flow rates will not only be a function of the extent of membrane fouling, they will also depend upon the underlying pore structure and morphology of the membrane. In the following sub-sections we consider three distinct membrane morphologies: an isotropic membrane with noninterconnected (straight-through) pores, a composite membrane in which the upper ("skin") layer has non-interconnected pores and the lower layer (substructure) has very highly interconnected pores, and a homogeneous membrane with an interconnected pore structure throughout the membrane. 2.1
Non-Interconnected Pore Structure For a membrane with non-interconnected pores, the volumetric filtrate flow rates through the open and blocked pores are given as:
AP Qown - t.tRm Aopen
(4)
30
FoulingPhenomenaDuringMierofiltration:EffectsOfPoreBlockage,CakeFiltration,And Membrane Morphology - Zydney
A
Qblo~k~d
=
blocked
J'O ' ( R m + R p o + R ' m p
)dAbloeked
(5)
where ~t is the solution viscosity, Rm is the membrane resistance, Rpo is the initial resistance of the first particle/aggregate, and R' is the specific resistance of the particle deposit or cake layer on the membrane surface. Eq. (5) is expressed as an integral over the blocked area to account for the spatial inhomogeneity in the thickness (or mass) of the protein deposit arising from the time-dependent blockage of the membrane surface. Thus, the protein deposit on a given region of the membrane only grows over the time interval t-tp where tp is the time at which that particular region was first covered or blocked by a protein aggregate [12,13]. Those regions of the membrane surface that have been blocked most recently will have the smallest values of mp and thus the greatest local filtrate flux. Equations (2) - (5) can be integrated numerically over time using the approach presented by Ho and Zydney [12,13]. It is also possible to integrate these equations analytically by assuming that the mass of the protein deposit is approximately uniform over the membrane surface at its maximum value [12-14]. The net result is: Q - ex Qo
-
+ ~Rm
1 - ex Rm + Rpo + R' mp
-
t ~R m
(6)
with:
J back
~baek
/~
-- ~ b a e k
(Rm + Rp-~)
]
(7)
The first term in Eq. (6) is equivalent to the classical pore blockage model and gives a simple exponential decay in the volumetric flow rate. At long times, the volumetric flow rate is dominated by the second term and is thus proportional to the ratio of the membrane resistance to the total resistance. The mass of the cake layer is a complex fimction of time, although it is relatively easy to show that Eq. (7) reduces to the form given by the classical cake filtration model in the limit of Jback= 0 [ 14]. 2.2
Composite Membrane Structure The situation is more complex for a composite membrane due to the flow through the interconnected pores in the membrane substructure (Figure 1). The flux through the open and blocked pores in the upper layer of the membrane are expressed in terms of the pressure drop across the upper layer of the composite 31
Fouling Phenomena During MicrofUtralion: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
Figure 1" Schematic drawing of flow through a composite membrane structure. structure (AP1) using Eqs. (4) and (5) but with AP replaced by AP1. The total pressure drop across the composite membrane is expressed in terms of the total filtrate flow rate as: AP = AP1 +
gQRsub
(8)
A
where Rsub is the resistance of the membrane substructure. The area A in the denominator of Eq. (8) is the total membrane area since the effects of any pore blockage on the surface of the upper layer are "lost" in the bottom layer due the lateral fluid flow. This is discussed in more detail elsewhere [ 15]. Equations (4), (5), and (8) can be combined to evaluate the filtrate flux through the open and blocked pores as:
J open _ R m(Rm + R~ub) + (Rw + R' mp )(R m + R~ub) J o - Rm (Rm + R~ub) + (Rpo + R' m p)(R m + XR~ub) J blocked
Rm(Rm+ R~ub)
J0
Rm (Rm + Rsub)-b (Rpo + R' mpXRm -b XRsub )
(9)
(10)
where X = Aop~/A is the fraction of open pores at any time t. Note that Joppa actually becomes greater than J0 as the pores become blocked, i.e. as X decreases. This occurs because the fluid flow through the composite membrane is shunted away from the blocked pores and through the open pores. This effect is not seen in a homogeneous (single layer) membrane (i.e. when R~ub- 0), since there is no opporttmity for the fluid to flow laterally as it moves through such a pore structure. Equations (2) and (3) can be integrated numerically over time, with the filtrate flux through the open and blocked pores given by Eqs. (9) and (10) [15]. It is also possible to develop an analytical solution to these equations [15]. In this case the open pore area is evaluated iteratively as an implicit function of
32
Fouling Phenomena During Microfiltration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
time, with the mass of the deposit expressed analytically in terms of the fractional open pore area.
2.3
lsotropic Membrane Structure Particle deposition on the upper surface of an isotropic membrane will only disturb the fluid flow over a small penetration depth into the pore structure since the fluid can flow laterally through the interconnected pores as it percolates through the membrane (Figure 2). The size of the surface blockage and the relative ease of lateral flow within the pore structure will determine the actual depth of the penetration.
....
Figure 2: Leftpanel: schematicof flow percolation through isotropic membrane Right panel: schematicof upper surface of partially fouled membrane In order to evaluate the filtrate flux through such a membrane structure, we assume that particle/aggregate deposition occurs uniformly and randomly over the upper surface membrane, giving the situation shown schematically in the fight hand panel of Figure 2. At low surface coverage the flow through this membrane can be described by a Krogh-cylinder model, with the surface blockage located at the center of a larger cylindrical region that sub-divides the membrane. The radius of this outer cylinder is directly related to the extent of surface blockage:
1--
Aopen A
m
/ 12 rbl 3eka~e
(11)
\ r~,,lindo~
As the membrane becomes more heavily fouled, the fluid will flow primarily through the gaps (or holes) between the fouled regions on the membrane surface. In this case, the membrane can be described using a similar model with the blockage occupying an outer annular region around a central void [ 16,17]. The velocity profiles within the porous membrane are described by Darcy's law:
33
FoulingPhenomenaDuringMicrofiltration: Effects OfPoreBlockage,CakeFiltration,AndMembraneMorphology- Zydney
c~ Vr = - K r
m
6].
c~ Vz = - K z - -&
(12)
where Kr and 1(7. are the Darcy permeabilities in the radial and transverse directions, respectively. A membrane with non-interconnected (straightthrough) pores would have Kr = 0 while a truly isotropic membrane would have Kr = Kz. The velocities Vr and Vz must satisfy the continuity equation within the membrane, leading to the following second-order partial differential equation for the pressure:
~P
+
~2
7
-0
(13)
where the ratio of the permeabilities (Kd~Kz) provides a quantitative measure of the ease of lateral flow. A novel experimental approach to evaluate this permeability ratio, which is really just a measure of the extent of pore connectivity, based on the relative fluid flow rate through a membrane that is partially covered with an impermeable tape was developed [ 12,13]. Equation (12) can be solved numerically for the pressure profiles within the homogeneous membrane. Symmetry conditions are applied in the radial direction at the center and outer edge of the Krogh cylinder. The boundary condition at the downstream surface of the membrane (z = ~Sm)is simply P = Pfiltr.te. A split boundary condition is applied on the upper surface to account for the hydraulic resistance of the deposit covering the blocked region of the membrane:
z-O" z=0:
Pf~ed -- P~=0
Vz--Kz"& =g(Rpo +R'mp) forO___r_
for rbloekage < r ___l'cylinder
(15)
The filtrate flux through the open and blocked regions are then calculated directly from the pressure profiles by numerical integration of the Darcy expression for Vz over the membrane area:
reylinder
_ J open
-1
.
/~/
[.2~rKz
'¢g~cylinder ~ 2 " - ~blockage)rblockage'I
dr z=O
34
(16)
FoulingPhenomenaDuringMicrofiltration:Effects Of Pore Blockage,CakeFiltration,And MembraneMorphology- Zydney rbloekage
Jblockea --
dr
2 ]/rblockage 0
z=0
(17)
The volumetric flow rate is then evaluated as a function of time by numerical integration of Eqs. (2) and (3) with the flux through the open and blocked pores given by Eqs. (16) and (17). 3.
F O U L I N G D A T A A N D ANALYSIS
The effects of the membrane morphology on the flux decline during the constant pressure filtration (AP = 14 kPa = 2 psi ) of 2 g ~ solutions of the protein bovine serum albumin (BSA) are shown in Figure 3. These BSA solutions contained 0.03 % protein aggregates, thus Cb and C'b in Eqs. (2) and (3) were both 0.6 mg/L. The data were obtained with three different membrane morphologies. The polycarbonate (PCTE) membrane is made by bombardment and etching of a polycarbonate film, resulting in a membrane with straightthrough (non-interconnected) nearly cylindrical pores. The polyvinylidene fluoride (PVDF) membrane is made by immersion casting and has a highly irregular and interconnected pore structure defined by the void space between polymer fibers. The PCTE/PVDF is a composite membrane formed by placing a PCTE membrane directly on top of an isotropic PVDF membrane. The data are plotted as the normalized flux (J/J0) as a function of the cumulative filtrate volume, where J0 is the initial flux through the membrane (values summarized in Table 1). The PVDF membrane has a smaller initial flux than the PCTE membrane due to its much greater thickness. The composite membrane has the smallest initial flux due to the additive nature of the resistances of the two membranes in series. Jo (m/sX 10 4)
Thickness (~Lm)
¢z (me k g -1 x 10 4)
(m -1 x 10 -11)
R' (m k g -1 x 10 -is)
PCTE
4.0
10
1.4 4-0.1
4.0 4- 0.2
8.0 4- 0.7
PVDF
2.2
120
0.7 4-0.1
4.0 4- 0.2
8.0 4- 0.7
Membrane
Rpo
PCTE/PVDF 1.7 130 1.1 4- 0.1 4.0 4-0.2 8.7 4-0.7 Table 1: PhysicalProperties and Fouling Parameters for Different Membranes The rate of flux decline for the PCTE membrane is very rapid, with the flux declining to less than 10% of its initial value within the first 60 mL of filtration. In contrast, the flux through the isotropic PVDF membrane remains greater than 80% of its initial value over the first 100 mL of filtration. The net result is a significantly greater capacity for the PVDF membrane, even though
3S
Fouling Phenomena During Microfiltration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydaey
that membrane has a much lower initial flux. For example, the cumulative filtrate volume obtained before the flux drops below J = 4 x 10-5 m/s is about 60 mL for the PCTE membrane compared to more than 180 mL for the PVDF. The addition of the PVDF membrane beneath the PCTE membrane also reduces the extent of flux decline compared to that for the PCTE membrane alone, although the fouling remains much more dramatic than that for the PVDF membrane. Note that these differences in flux decline were not due to differences in the total pore volume as SEM micrographs demonstrated that all of the fouling in these experiments occurred on the upper surface of the membranes. The much smaller flux decline seen with the isotropic PVDF membrane is a direct result of the highly interconnected pore structure of this membrane. Aggregate deposition on the surface of the PCTE membrane leads to complete blockage of the pores, although the blocked pores do allow some fluid leakage through the interstices within the protein aggregate. In contrast, aggregate deposition on the surface of the PVDF membrane only disturbs the fluid flow over a small penetration depth into the membrane, with the fluid simply flowing under and around the surface blockage (Figure 2). The flow in the composite membrane is somewhat more complex, with the surface blockage effectively sealing the pores in the upper (PCTE) layer while the entire (PVDF) substructure remains available for flow. In this case, the fouling reduces the pressure drop across the lower layer of the membrane, leading to an increase m AP1 and a corresponding increase in the flux through the open pores [15]. This increase in Jo~ partially offsets the reduction in the area of the open pores, thus reducing the extent of flux decline compared to that for the PCTE membrane alone. The solid curves m Figure 3 are model calculations developed by numerical solution of Eqs. (2) and (3). The filtrate fluxes through the open and blocked pores were given by the appropriate equations for a membrane with straight-through non-interconnected pores (Eqs. 4 and 5), for a composite membrane structure (Eqs. 9 and 10), and for an isotropic membrane with highly interconnected pores (Eqs. 16 and 17). In each case the best fit values of the model parameters were determined by minimizing the sum of the squared residuals between the filtrate flux data and the model calculations using the method of steepest descent, with the results summarized in Table 1. Jbackwas set equal to zero for these calculations, although a slightly better overall fit to the data could be achieved using a small positive value for this parameter. Additional details on the model calculations are provided elsewhere [12,13,15]. The fairly sharp change in slope for the PVDF membrane around V = 180 mL occurs when the membrane surface is nearly completely covered by protein aggregates and reflects the transition from a pore blockage to a cake filtration mechanism. The model calculations are in very good agreement with the experimental data for all 3 membranes, demonstrating the ability of this simple
36
Fouling Phenomena During Microfillration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydncy
O PCTE
~
0.8
PVDF PCTE/PVDF
0.6
0.4
r-! I-I
0.2
0
50
100
150
290
250
Filtrate Volume. V (mL) Figure 3: Effects of membrane morphology on the flux decline during the constant pressure filtration of 2 g/L solutions of BSA. Data are shown for the polycarbonate (PCTE), polyvinylidene fluoride (PVDF), and composite PCTE/PVDF membranes.
theoretical framework to describe the flux decline behavior of porous membranes with very different underlying pore morphologies. The best fit values of Rpo and R' for the PCTE, PVDF, and PCTE/PVDF membranes were essentially identical, with the actual values in good agreement with independent estimates of these fouling parameters obtained from measurements of the hydraulic resistance and mass of the protein deposits [16,17]. The values of the pore blockage parameter, (z, show some differences between the three membranes which is likely due to small batch-to-batch variations in the properties of the BSA solutions. The dashed curve in Figure 3 shows the model calculations for the PCTE membrane evaluated using the simple analytical solution (Eqs. 6 and 7) using the same parameter values given in Table 1. This approximate solution is in excellent agreement with both the experimental data and the full numerical results, even though the approximate solution was developed by assuming that the mass of the protein deposit is uniform across the membrane surface at its maximum value. This effect is shown more explicitly in Figure 4, in which the
37
Fouling Phenomena During Microfillralion: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydncy
calculated values of the deposit mass per unit membrane area are plotted as a function of the fractional surface coverage of the membrane at four different filtration times. 30 t = 32 min
25 20
15
t = 16 min
10 t = 8 min t=4
0
0.2
min
0.4
0.6
0.8
1.0
Fractional Blocked Area, Ab~oCkedA Figure 4: Model calculations for the deposit mass over the fouled region of the membrane. Solid curves are full numerical solution. Dashed curves are approximate analytical solution.
The dashed curves show the calculations using the approximate analytical solution, in which case the deposit mass is uniform across the fouled region of the membrane at its maximum value. The solid curves are the calculations using the full numerical solution and show the variation in deposit mass over the fouled region of the membrane. Even at long filtration times, the deposit mass remains relatively uniform over the membrane surface due to the inherent "selfleveling" character of the fouling process. Those regions of the membrane that are fouled first have the smallest filtrate flux, leading to the slowest rate of cake growth. Those regions that were fouled most recently have the thinnest cake but the greatest rate of cake growth. The net result is that the average protein layer resistance over the fouled region of the membrane at t = 32 mm is only 6% smaller than the maximum resistance, with this difference decreasing to less than 3% after 100 mm of filtration. This self-leveling phenomenon is the reason why the simple approximate solution provides such accurate predictions for the filtrate flux as seen in Figure 3.
38
Fouling Phenomena During Microfiltration: Effccta Of Pore Blockag©, Cake Filtration, And M©mbran¢ Morphology - Zydacy
Additional insights into the fouling behavior can be obtained by analyzing the flux decline data using the functional form suggested by Eq. (1). The required derivatives can be calculated directly from the filtrate flux data versus cumulative filtrate volume data in Figure 1 as: dt
m
dV dZt dV 2 -
1
(18)
JA 1 (dJ) j2 A ~,~'~)
(19)
where dJ/dV was evaluated numerically [16,17]. The results are shown in Figure 5, with the solid curves representing the model calculations developed using the equations in Section 2 using the same values of the model parameters shown in Table 1. The model calculations are again in very good agreement with the data for all three membranes. The small discrepancy at large dt/dV for the PVDF membrane is due to the uncertainties associated with the transitions from pore blockage to cake filtration and from flow around a centrally located blockage to flow through a cylindrical region between aggregates. At low dt/dV (corresponding to high filtrate flux), the data for the PCTE membrane yield a linear relationship with slope approximately equal to 2.0, which is consistent with the pore blockage mechanism. In contrast, the data for the PVDF and the composite PCTE/PVDF membranes show much steeper slopes at the start of the filtration. All three membranes show a distract maximum in d2t/dV 2 at an intermediate dt/dV. Others [6,7] also have seen this maximum in d2t/dV 2, a result that has never been explained previously and is completely inconsistent with the behavior predicted by the classical fouling models. This behavior is very accurately described by the pore blockage - cake filtration model, with the transition between these two fouling mechanisms causing a sharp reduction in the rate of flux decline leading to the observed maximum in d2t/dV 1. The transition between the pore blockage and cake filtration fouling mechanisms can be seen more dearly by numerically evaluating the slope of the d2t/dV 2 versus dt/dV data in Figure 5. The results are shown in Figure 6 as a function of the filtration volume. The large scatter in some of the data is due to the errors involved in the numerical differentiation of the raw data. The solid curves again represent the model calculations using the parameter values given in Table 1. The slope for the PCTE membrane begins around 2, and remains nearly constant at that value over the first 60 mL of filtration. This behavior arises from dominance of the
39
Fouling Phenomena During Microfiltration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology
- Zydncy
1013
l"1
1012 E
N~
1011 0 PCTE r"l PVDF A PQ,TE/PVDF 10lo
. . . .
I
108
10 7
109
dt/dV (s m "3) Figure 5: Flux decline analysis for BSA filtration through the PCTE, PVDF, and composite PCTE/PVDF membranes. Experimental data are from Figure 3. Solid curves are model calculations.
pore blockage mechanism; the slightly smaller value of the slope compared to that for the classical pore blockage model is due to the small leakage flow through the interstitial space within the protein aggregates. The transition from pore blockage to cake filtration occurs when the membrane surface is nearly completed covered with protein aggregates, and it is marked by a rapid reduction m slope to a value less than -6. The slope then increases and attains a value of zero for V > 85 mL, consistent with the dominance of the cake filtration mechanism at long times. The slope for the composite membrane beings around 6 and decreases monotonically during the start of the filtration process. This reduction in slope reflects the shunting of the fluid flow towards the open regions of the membrane. There is again a sharp decrease m slope around V = 100 mL due to the transition between a pore blockage dominated fouling to a cake filtration mechanism. The initial slope for the isotropie PVDF membrane is around 9 and decreases in value throughout the filtration run. Similar extreme values of the slope have also been reported previously [7,8], but they never have been explained fundamentally. Our new theoretical framework provides a good qualitative prediction of the observed behavior,
40
Fouling Phenomena Dunng ~.~crofiitration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology -Zydney
10
,~~,
8 c:
6
a-
4
.=-
2
O
0
..O
.,--.,.
u,.,
DF
O
O PCTE r l Pvog ~ PCTE/PVDF
Z~ A
%~'~A E]
-2
PCTE/PVDF
,
0
O
E!
-4.
-8
~,n ,-,
PCTEt~:~ #~)O O
II--
O
,.~ '
I
50
,
,
100
I
150
,
I
200
n ,
250
Cumulative FiltrateVolume, V (mL) Figure 6: Slope determined from the plot of d2t/dV 2 versus dt/dV for the PCTE, PVDF, and composite PCTE/PVDF membranes. Experimental data are from Figure 3. Solid curves are model calculations. with the quantitative discrepancies seen in Figure 6 arising from the multiple derivatives that have to be taken to evaluate the slope. Thus, there is no need to invoke any new or complex fouling phenomena to explain these experimental observations. The key is the proper inclusion of both the pore blockage and cake filtration behavior and the proper analysis of the flow properties of membranes with different underlying pore morphologies. Although the experimental data presented in Figures 3 - 6 were all obtained with bovine serum albumin, the pore blockage- cake filtration model can also explain the flux decline behavior seen in a wide variety of filtration systems. For example, Figure 7 shows results for the constant pressure filtration of BSA, lysozyme (a 14 kD molecular weight protein), and a soil-based humic acid through the 0.2 gm pore size polycarbonate track-etched membranes Additional details on the protein and humic acid experiments are provided elsewhere [14,18], respectively. In each case, the derivatives were evaluated numerically directly from data for the filtrate flux as a function of the cumulative filtrate volume. The solid curves are calculations using the simple approximate solution (Eq. 6) with the best fit values of the parameters determined by comparison of the model and data. Jbackwas set equal to zero for both BSA and lysozyme, while the best fit value of Jbackwas 5.3 x 104 m s1 for
41
Fouling Phenomena During Mierofiltration: Effects Of'Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
lo 3
A ,,f,
.
.
.
.
.
.
.
.
i
.
.
.
.
.
.
.
.
i
.
.
.
.
.
.
.
.
,~
1012
E m
"o ol
"o 1011
LJ 10lo 10 6
113_ t.yso me
~
..~.. Hum.ic~.Ac.id
10 7
10 8
10 9
dt/dV (s m "3) Figure 7: Flux decline analysis for the constant pressure filtration of solutions of BSA (2 g/L), lysozyme (2 g/L), and a soil-based humic acid (2 ppm) through polycarbonate track-etch membranes. Solid curves are model calculations.
the 2 ppm humic acid. The initial slope for all three systems was approximately equal to two, which is again consistent with the dominance of the pore blockage phenomenon at the start of the filtration. The flux decline data for the humic acid solution show a distinct maximum in d2t/dV2 reflecting the transition between the pore blockage and cake filtration mechanisms. At long filtration times, dZt/dV2 for the humic acid solution approaches zero as the flux approaches its steady-state value J = Jb,ck. This sharp decline in d2t/dV2 provides a convenient diagnostic for determining the presence of a steady-state flux in any given filtration process. The lysozyme data show a much smoother transition from the pore blockage to cake filtration mechanisms, with no distinct maximum observed in the dZt/dV2 data. This occurs because of the greater rate of cake growth relative to pore blockage for the lysozyme [ 18]. 4.
SUMMARY
Membrane fouling remains a major problem in applications of microfiltration for particle removal, sterile filtration, and clarification. The theoretical framework presented in this chapter provides a completely new approach for the analysis and interpretation of flux decline data obtained in these membrane systems. The model explicitly accounts for fouling due to surface (pore) blockage and cake filtration, with the cake forming over those
42
Fouling Phenomena During Microfiltration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
regions of the membrane that have first been covered by large aggregates. Just as importantly, this new theoretical framework provides the first quantitative analysis of the effects of the pore morphology on the rate of flux decline. Membranes with straight-through pores show the greatest rate of flux decline since the particles completely cover (block) the non-interconnected pores in these membranes. In contrast, particle deposition on the surface of a homogeneous membrane with highly interconnected pores will disturb the filtrate flow only over a relatively small penetration distance into the membrane pore structure. This implies that very thick membranes should exhibit a slower rate of flux decline due to the smaller relative disturbance in the flow, an effect that has been confirmed experimentally [ 16,17]. The flux decline for composite membrane structures shows a more complex behavior. The surface fouling completely blocks the pores in the upper skin layer, but the fluid rapidly redistributes itself through the very highly interconnected pores within the membrane substructure. This causes a shunting of the fluid flow towards the remaining open pores, leading to a reduction in the rate of flux decline compared to that for a single membrane layer. This theoretical analysis also provides a framework that can be used for the design and development of new membrane structures having reduced rates of fouling. This would include the proper choice of pore connectivity, membrane thickness, and the specific properties of the individual layers in more complex composite structures. These phenomena have often been neglected in membrane design/development, although they can clearly have a dramatic effect on the flow distribution, fouling, and overall selectivity for the particular separation. REFERENCES [1]
P. H. Hermans and H. L. Bred6e, "Principles of the Mathematical Treatment of Constant-Pressure Filtration," J. Soc. Chem. Intl. 55T (1936) 1.
[21
V. E. Gonsalves, "A Critical Investigation on the Viscose Filtration Process," Ree. Trav. Chim. Des Pays-Bas, 69 (1950) 873.
[3]
H. P. Grace, "Structure and Peformance of Filter Media," AIChE J., 2 (1956) 307.
[4]
J. Hermia, "Constant Pressure Blocking Filtration Laws - Application to Power Law Non-Newtonian Fluids," Trans. Inst. Chem. Eng., 60 (1982) 183.
[5]
K. Luckert, "Model Selection Based on Analysis of Residue Dispersion Using SolidLiquid Filtration as an Example," Int. Chem. Eng., 34 (1994) 213.
[6]
W. R. Bowen, J. I. Calvo, and A. Hemandez, "Steps of Membrane Blocking in Flux
Decline During Microfiltration," d. Membrane Sei., 101 (1995) 153. [7]
E. Iritani, Y. Mukai, Y. Tanaka, and T. Murase, "Flux Decline Behavior in Dead-End Microfiltration of Protein Solutions," J. Membrane Sci., 103 (1995) 181.
43
Fouling Phenomena During Microfiltration: Effects Of Pore Blockage, Cake Filtration, And Membrane Morphology - Zydney
[8]
J-Y. Wang, K-S. Chou, and C-J. Lee, "Dead-End Filtration of Solid Suspension in Polymer Fluid through an Active Kaolin Dynamic Membrane," Sep. Sei. Tech., 33 (1998) 2513.
[9]
S. T. Kelly, W. S. Opong, and A. L. Zydney, "The Influence of Protein Aggregates on the Fouling of Microfiltration Membranes During Stirred Cell Filtration," J. Membrane Sei., 80 (1993) 175.
[10]
W. Yuan and A. L. Zydney, "Humic Acid Fouling During Microfiltration," J. Membrane Sci., 157 (1999) 1.
[11]
G. Belfort, R. H. Davis, and A. L. Zydney, "The Behavior of Suspensions and Macromolecular Solutions in Crossflow Microfiltration," J. Membrane Sei., 96 (1994) 1.
[12]
C. C. Ho and A. L. Zydney, "A Combined Pore Blockage and Cake Filtration Model for Protein Fouling during Microfiltration." J. Coll. Interf. Sei. 232 (2000) 389.
[13]
C. C. Ho and A. L. Zydney, "Measurement of Membrane Pore Interconnectivity," J. Membrane Sci., 170 (2000) 101.
[141
W. Yuan, A. Kocic, and A. L. Zydney, "Analysis of Humic Acid Fouling using a Combined Pore Blockage - Cake Filtration Model," J. Membrane Sei. (in press 2002).
[15]
C. C. Ho and A. L. Zydney, "Protein Fouling of Asymmetric and Composite Microfiltration Membranes," Ind. Eng. Chem. Res., 40 (2001) 1412.
[16]
C. C. Ho and A. L. Zydney, "Effect of Membrane Morphology on Protein Fouling During Microfiltration," d. Membrane Sei., 155 (1999) 261.
[17]
C. C. Ho and A. L. Zydney, "Theoretical Analysis of the Effect of Membrane Morphology on Fouling during Microfiltration." Sep. gel. Teeh. 34 (1999) 2461.
[18]
L. Palacio, C. C. Ho, and A. L. Zydney, "Application of a Pore Blockage - Cake Filtration Model to Protein Fouling during Microfiltration," Bioteeh. Bioen~ (submitted 2002).
44
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 3
Differential scanning calorimetry and rheological experiments to study membrane formation via thermallyinduced phase-separation P.C. van der Heijden, M.H.V. Mulder, M. Wessling University of Twente, PO Box 217, 7500 AE Enschede, The Netherlands, Fax: xx31-53-4894611 1.
INTRODUCTION
1.1
Thermally-Induced Phase-Separation Porous polymer membranes can be prepared by different techniques: track-etching, stretching and phase-separation [ 1]. The phase-separation method is very common to prepare polymeric membranes and much attention has been paid to prepare membranes from various polymers and diluents [1,2]. The preparation of a porous structure consists of two steps. First, the homogeneous polymer solution has to undergo liquid-liquid demixing to obtain a polymer-rich continuous matrix and a dispersed polymer-lean phase of almost pure solvent. The second step is a fixation step to give the structure mechanical stability. Liquid-liquid demixing takes place when the solvent power is not sufficient anymore to keep the polymer dissolved. Two major methods can be distinguished to decrease the solvent power, either by adding a non-solvent which is called diffusion-induced phase-separation (DIPS) or by changing the temperature, defined as thermally-induced phase-separation (TIPS). The latter method is studied in this paper. Fig. 1 shows a schematic representation of the formation process of a porous structure with the TIPS method. The dashed arrow represents a typical cooling trajectory. When passing the binodal, the homogenous polymer solution liquid-liquid demixes in a polymer-rich phase with a polymer concentration given by the binodal and a polymer-lean phase of almost pure diluent. After liquid-liquid demixing of the binary polymer-diluent system, the structure has to be fixed. This can be achieved by crystallization, vitrification (glass transition), or gelation of the polymer-rich phase. After the structure fixation, it is possible to remove the polymer-lean phase by evaporation or extraction. An advantage of the TIPS method over the DIPS method is the ability to form porous structures with polymers which do not form a solution at room temperature, for example polyethylene and polypropylene.
45
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
,soo
c
H I binodal
. _ _
•
400 1
E(L) 300 -F-
"~pi~n°dalT
200
' 0.0
I 0.2
S
'
I 0.4
i
I 0.6
i
I 0.8
i 1.0
Volume fraction polymer (-) Figure 1. Example of a phase diagram of a binary polymer- diluent solution. The black arrow represents a cooling trajectory to obtain a porous structure with the TIPS method. It goes from the homogeneous solution (I-I) via the liquid-liquid demixing gap (L-L) into the structure fixation area (S). ~c is the critical concentration of the solution. The porous structures obtained with TIPS can be used for more applications than membranes, for example: separators m electrochemical cells, synthetic leather, 'breathable' rainwear, diapers, surgical dressings and bandages [3], inertial confmed fusion targets [4,5], biodegradable implants for cell transplantation [6], and scaffolds for tissue engineering [7]. Typical cell sizes obtained with this method are in the order of micrometers.
1.2
Historical Background
The formation of porous structures via TIPS was firstly reported by Castro [8]. In this patent, he investigated hundreds of polymer-diluent systems, both amorphous and crystalline polymers, on their ability to form a porous structure. Later on, Shipman [9] described a procedure to prepare porous structures with a certain shape, and Josefiak et al. [10] introduced a second liquid prior to the cooling step to adjust the morphology of the porous structure. The first papers on the TIPS method were published in 1984 [ 11 ]. Since then, the interest in the formation of porous structures of various polymerdiluent systems with the TIPS method has increased rapidly. Table 1 gives a
46
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
summary of crystalline polymer-diluent systems used in the TIPS method and Table 2 gives an overview of the formation of amorphous porous structures via TIPS. Besides the studies on new polymer-diluent systems, studies were performed on variations of the TIPS method by introducing, for example, an evaporation step to obtain an asymmetric structure [ 12-16].
Table 1. Crystalline polymer-diluent systems which have been used to form a porous structure with TIPS. Polymer Diluent Isotactic polypropylene (iPP) Diphenylether [ 17-19] n,n-Bis(2-hydroxylethyl)tallowamine [11,20-231 Polyvinylidene fluoride Cyclohexanone [ 11 ] (PVDF) Butrolactone Propyl enecarbonate Carbitolacetate Isotactic polystyrene (iPS) Nitrobenzene [24] Cyclohexane [25] Dioxane / isopropanol Poly(tetrafluorethylene-co-per- Chlorotrifluorethylene [26] fluoro-(propyl vinylether)) (Teflon®PFA, NeoflonT~FA) Poly(2,6-dimethyl- 1,4Cyclohexanol [ 12] phenylene ether) (PPE) Nylon 12 Poly(ethylene~lycol) [271 Dioxane / water [28] Poly(7-benzyl-l-glutamate) Benzene (PBIG) Dichloroethane Polyethylene (HDPE) Diphenylether [29] 1,2-Ditridecylphtalate / hexadecane[30] Polysilastyrene (PSS) Cyclohexane [31 ] Benzene 4-Methyl- 1-pentene (PMP) Diisopropylbenzene [5] Polylactide 1,4-Dioxane [6] As mentioned before, the formation of a porous structure with the TIPS method consists of two steps: liquid-liquid demixing to obtain a porous structure and the fixation step to obtain mechanical stability. Both steps have been studied separately for quite a time. The first papers on liquid-liquid demixing of polymer solutions were already published in 1950's (for references see [43,44]).
47
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
The existence of crystalline polymers has been known from the mid 1940's (a historic overview can be found m [45]) and the vitrification temperature depression (or glass transition temperature depression), for example, was already quantified in 1961 [46]. Table 2. Amorphous polymer-diluent systems which have been used to form a porous structure with TIPS. Polymer Diluent Atactic poly(methylmethacrylate) Cyclohexanol [32,33] (aPMMA) Tetramethylene sulfone (sulfolane) [34,35] 1-Butanol [32] Tert-butylalcohol [36] Atactic polystyrene (aPS) Cyclohexane [37] Diethylmalonate Cyclohexanol [38-41] Trans-decahy dronaphtalene (decalin) [42]
Liquid-liquid demixing of polymer solutions is extensively studied since the last half of the twentieth century (for references see [43,44]). Phase diagrams of many polymer-diluent systems have been determined visually or with other optical techniques such as optical microscopy and light scattering [21,23,27,28,35,38,43]. These techniques are also used to follow the time dependency of the demixing process to study the kinetics of demixing [17,27,33,47-50]. Also other experimental techniques have been occasionally used to compose phase diagrams or to follow the demixing process like viscometry [51 ], NMR [34], X-ray scattering [52] and DSC [32,42,53,54]. 1.3
Aim of this Paper In this paper, DSC is used to study liquid-liquid demixing and the structure fixation step of an amorphous polymer solution. In addition, rheological experiments are carried out to study liquid-liquid demixing as well. Using DSC and theology, new insights may be obtained in the membrane formation process. However, practical reasons can be mentioned to use these techniques instead of the frequently used light based techniques. For example, with light based techniques the refractive index between both the diluent and polymer has to differ significantly, and transparency of the experimental setup is required to observe a signal. This is not a problem with DSC and rheological experiments. Furthermore, with DSC and rheology it is possible to detect both the liquid-liquid demixing step and the structure fixation step in one experiment. 48
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
With optical microscopy for example, liquid-liquid demixing can be observed but it is impossible to observe the glass transition. Finally, high temperatures are involved in the TIPS process and to exclude evaporation of diluent, closed sample holders are necessary to use, and this is very easy to achieve with DSC. Q
DIFFERENTIAL SCANNING CALORIMETRY [55]
2.1
Introduction With DSC, the heat flow is measured as a function of time or temperature to observe physical transitions. After the introduction of the first differential scanning calorimetry (DSC) apparatus polymers have been studied extensively. See Refs. [56,57] for a historical overview about the development of thermal analysis devices in general and for polymers in particular. DSC is a well-known technique to study solid-liquid demixing (crystallization) [20,23,24,29] and vitrification [32,38,54] of liquid-liquid demixed polymer solutions. Berghmans and co-workers [32,42,53,54] used DSC as well to determine the liquid-liquid demixing temperature. They carried out DSC experiments for the systems atactic polystyrene / decalin and atactic polymethylmethacrylate / 1-butanol and cyclohexanol respectively. Upon cooling (cooling rate 5 K.min-1), an exothermic heat flow shift was observed and the onset was taken as the liquid-liquid demixing temperature. This signal agreed very well with optical observations. With one DSC run they could determine both the liquid-liquid demixing temperature and the glass transition temperature of the polymer-rich phase. However, DSC is hardly used for the determination of liquid-liquid demixing because the heat effect is very small and disappears easily in the base line drift. Recently, a rather new technique has been developed: temperature modulated differential scanning calorimetry (TMDSC) [58,59]. In spite of some discussion about the interpretation of the measured signals [60-67], TMDSC is a very useful device to measure small heat effects and to separate overlapping thermal events. In the next sections, TMDSC experiments are used to examine the cooling trajectory of binary polymer solutions in the concentrated region (at the right side of the critical point in Fig.l). The experimental work described in this work is carried out with the polymer-diluent system atactic polystyrene in 1dodecanol. A number of reasons can be mentioned to choose this system. This system forms a porous polymer structure with the TIPS method at room temperature [8], 1-dodecanol is a non-toxic solvent, polystyrene has solvents at room temperature and therefore easy to use, and many physical data are available for both components.
49
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
2.2
Temperature Modulated Differential Scanning Calorimetry
An extensive description of this technique can be found in Refs. [58,62,68]. Here, only a short description is given. With TMDSC, a second function (for example a sine wave) is superimposed onto the conventional linear or isothermal temperature ramps (see Fig. 2).
200
-
0 o v
"~ 1 9 5 O.
E I--
"
190 I
0
'
I
1
'
I
'
2
I
3
'
I
4
'
5
Time (min)
Figure 2: Linear cooling trajectory (thick black line) with superimposed temperature modulations.
The temperature ramp can then be described as: (1)
T = To + bt + A sin cot
where To is the initial temperature, b the underlying scanning rate, A the temperature amplitude, co the angular frequency, and t the time. The TAinstruments user manual [68] suggests a value of the underlying scanning rate (b) below 5 K.min "1 and typical values for A and co are 1 K and 60 s. The resulting heat flow consists of two contributions. The first part is caused by rapid processes and is proportional to the scanning rate. The second part is caused by kinetically hindered or irreversible processes and hence independent of the scanning rate: dQ dt
= Cp
dT -~
(2)
+ f (t,T) "
50
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
dQ/dt is the heat flow, Cp the heat capacity, and fit, T) a contribution to kinetic effects. The resulting heat flow is a wave function as well. With the help of discrete Fourier transformation software which resides in the DSC module, the heat flow signal is deconvoluted in a cyclic signal and an underlying signal (which is equivalent to the conventional DSC). In this work the modulus of the complex heat capacity is used to present the TMDSC results. This quantity is calculated from only the amplitude of both the temperature and the heat flow modulation. It is thus completely derived from the cyclic signal. Physical processes, which take place at time scales smaller or comparable to the modulation period, contribute to this complex heat capacity. Examples of such physical processes are vibrations and rotations of atoms. Slow physical processes, like the mobility of vitrified polymers with time scales much larger than a modulation period does not contribute to the complex heat capacity.
(Icp*l)
(IcjI)
2.3
Experimental
2.3.1 Materials Atactic polystyrene (aPS) (Styron* 686E) was kindly supplied by Dow Benelux NV (Mw and Mw/M,: 2.3.105 g.mol 1 and 2.1 respectively, determined with GPC). The diluent, 1-dodecanol (purity > 98%, Merck-Schuchardt), was used without further purification.
2.3.2 Sample preparation A homogeneous solution of aPS and 1-dodecanol was prepared in a threeneck bottle under nitrogen at 200°C. 1-Dodecanol vapor was allowed to evaporate during stirring with a mechanical stirrer. Small amounts of various polymer concentrations were poured in Petri-dishes and cooled in air. The compositions of the samples were determined by thermogravimetric analysis. About 20 mg of the sample was inserted in a platinum sample pan of a TGA 2950 Thermogravic Analyzer of TA Instruments and heated up to 200°C with a heating rate of 10 K-min-1. Afterwards the temperature was kept constant at 200°C for maximum 2 h to evaporate all the 1-dodecanol. From the ultimate weight loss the polymer concentration was determined.
2.3.3 Temperature (TMDSC)
Modulated Differential
Scanning Calorimetry
The TMDSC used is a DSC 2920 of TA Instruments. Calibration with indium and high density polyethylene (HDPE) (for calibration of the heat capacity) was carried out. About 5 mg of the sample was put in the aluminum closed sample pan. The TMDSC was heated to 200°C and kept isothermally for
51
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
30 min to ensure homogeneity. The cooling rate was set to 2 K-min 1 to 0°C and after an isothermal step of 5 min the sample was heated again with 2 K.min -1. The amplitude of the superimposed sine wave was 1 K with a period of 60 s [68]. The glass transition temperature Tg and the liquid-liquid demixing temperature TL_Las well as the heat capacity shift at TL_Lwas determined with the TA Universal Analysis software.
2.3.4 Optical Microscopy (OM) To compare the TMDSC results with a well-known technique for studying liquid-liquid demixing, optical microscopy experiments are carried out. The polymer sample was placed on an object glass within an aluminum ring (thickness 0.1 mm, inner diameter 5 mm) and covered by a second glass. To prevent diluent loss caused by evaporation, laboratory grease was used to stick the aluminum spacer to the object glasses [23]. The sample was placed in a hot stage (Linkam THMS 600) which was controlled by the Linkam TMS92 hot stage controller. The sample was heated and cooled with a rate of 2 K.min -1 and demixing was observed visually with an Olympus BH2 microscope (magnification 200x).
2.4
Results
Cooling and subsequent heating curves of aPS - 1-dodecanol are plotted in Fig. 3 for two polymer concentrations (weight fractions of 0.38 and 0.69). The modulus of the complex heat capacity (Icp'l) is plotted at the y-axis. Two transitions can be observed: the glass transition and a small baseline shift at higher temperatures, which is assumed to be the liquid-liquid demixing temperature. In the following, the onset of this signal upon cooling is defined as the liquid-liquid demixing temperature (TL-L) comparable with the observations of Amauts et al. and Vandeweerdt et al. with the conventional DSC [32,42,53,54]. The glass transition temperature (Tg) is chosen as the onset upon cooling because below this temperature influences can be expected of vitrification on the liquid-liquid demixing behavior. The difference in heat capacities ([Cp*l) between the results for the two polymer concentrations can be explained by the heat capacity difference between both single components. Literature values of the heat capacity for pure polystyrene and 1-dodecanol at T = 180°C are 2.1 J.gl.K-1 [69] and 3.1 J.gl.K-1 [70], respectively. Therefore, a polymer solution that contains the higher content of polystyrene should have a smaller heat capacity than a diluted polymer solution. At least, when it is assumed that the heat capacity of a solution of aPS in 1-dodecanol is between the values of both pure components.
52
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
3.0
'7, '7
--
2.5-ETJ
-..3
"
0
0.
2.0
--
..,.=,_.-
69 wt-%
1.5
;
I
;
40
I
80
:T,, ;
I'
120
'
I
160
'
200
Temperature (°C) Figure 3. Cooling and subsequent heating curves (weight fractions of polymer: 0.38 and 0.69). Gray lines" heating curves, black lines" cooling curves.
With conventional DSC experiments, an optimum has to be found between a high scanning rate which results in a high intensity and a low scanning rate which results in a high resolution of a thermal signal. Consequently, a small thermal transition such as liquid-liquid demixing is very difficult to observe with conventional DSC. With TMDSC, both a low scanning rate (the underlying scanning rate) and higher scanning rates (the temperature modulations) are present within one experiment and therefore it is very suitable to detect liquid-liquid demixing. In the cooling curves the L-L phase transitions at TL_Lare represented by a steep heat capacity shift. The heating curves have the same slopes as the cooling curves only at the liquid-liquid demixing temperatures the transition is not as distinct. This difference is discussed later in this section. In the following, the details of the cooling curves are further used and discussed only. Performing such TMDSC cooling experiments over a large concentration range allows the construction of the phase diagram of the polymer-diluent system. To support the assumption of the base line shift to stem from the L-L demixing, the TMDSC results are compared with optical microscopy (OM) data indicating visually the phenomenon of L-L demixing. In Fig. 4, the TL-Land Tg determined with TMDSC and OM are plotted. The open circles represent the TMDSC liquid-liquid demixing data whereas the filled black squares are the OM data. The closed circles are the glass-transition temperature data points. The liquid-liquid demixing data obtained from TMDSC
53
Differeraial Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
and OM agree well and therefore it can be concluded that the observed TMDSC signals are indeed caused by liquid-liquid demixing. 200
~,
160
o o v
::}
120
(1.) (:}.
E
80
!--
40
0.0
0.2 Weight
0.4 fraction
0.6
0.8
polymer
1.0 (-)
Figure 4. Phase diagram a P S - 1-dodecanol. Open circles: TMDSC data L-L demixing. Closed squares: OM data L-L demixing (cloud points). Closed circles: TMDSC data glass transition. Lines are drawn to guide the eye.
The shift of the TMDSC curve at the liquid-liquid temperature is assumed to be completely caused by the enthalpy of demixing. The difference in values in the heat capacity shift (Acp* at Tt.t, defined in Fig. 3) between the different concentrations is caused by the interaction between polymer and diluent. This can be quantified by calculating the enthalpy of mixing with the help of the Flory-Huggins theory [54]. To minimize the error in the calculation of the modulus of the complex heat capacity with the TMDSC software, it is recommended that at least 4 complete superimposed cycles fit within a phase transition [68]. This requirement is satisfied for the glass transition because this transition covers a temperature range of at least 10 K. However, in case of liquid-liquid demixing the heat capacity shift only covers a temperature interval of 2 K, so only one modulated cycle fits within this transition. By lowering the underlying cooling rate, the number of cycles within the transition can be increased; the resulting TMDSC curves are shown in Fig. 5. From Fig. 5 it can be concluded that cooling rates of 2 K.min -1 and lower have no significant influence on the measured modulus of the complex heat capacity. As already mentioned the measured complex heat capacity is only measured from the amplitudes of the modulated heat flow and modulated temperature, hence it is not influenced by the cooling rate. However, the shape of the curve can be influenced by time-dependent effects like the growth of the
54
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
demixing domains and the increasing viscosity of the polymer solution (see section 3) upon cooling. The growth of the demixed domains can be the reason for the difference between the cooling and heating curve (see Fig. 3 for a 38 wt.% polymer solution). In the demixed polymer solution regions of almost pure diluent grow when the temperature of the solution is in between the TL-L and the Tg. The TMDSC heating curve was measured after the cooling trajectory, so, the demixed domains already had a long time to grow. Therefore, it is more difficult for the polymer solution to follow the temperature modulation because of larger distances the diluent molecules must diffuse. ,,
,
,,
0.2 K.min~ 2 . 8
,,
,
,
2 K-min"~
-"
"7 "7
2.4 - -
v
-1
•
..
(3 J
2.0--
~/
5 K.min
1.6 J
50
!
I 100
'
I
'
150
I 200
Temperature (°C) Figure 5. Influence of cooling rate on modulus of the complex heat capacity. Weight fraction of polymer is 0.48.
The shape of the curves at low cooling rates (between 0.2 K.min 1 and 2 K.min1) are comparable, this means that influence of time depending effects play a negligible role in the cooling curves at these cooling rates. 2.5
Conclusions and Outlook With Temperature Modulated DSC, liquid-liquid demixing of polymerdiluent systems can be determined as well as the glass transition of the polymerrich phase in one run at a relatively low scanning rate. Liquid-liquid demixing observed with TMDSC agrees well with visually observed cloud points.
55
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
Upon quantifying the heat capacity shift at the liquid-liquid demixing temperature additional information with respect to the thermodynamics of the polymer solutions can be obtained. Furthermore, it offers the possibility to follow the heat capacity shift as a function of temperature both theoretically and experimentally to study the kinetics of liquid-liquid demixing. Especially this is interesting in the region of the glass transition temperature in which the structure fixation step can be studied. 3.
VISCOMETRY AND RHEOLOGY
3.1
Introduction
During the formation of porous structures with the TIPS process, the homogenous solution is demixed in two liquid phases and afterwards the polymer-rich phase is vitrified. It is expected that these phase transitions can be observed with rheological experiments. The possibility of studying liquid-liquid demixing with viscometry and rheology experiments is briefly discussed in this section. The experiments are carried out for the polymer-diluent system aPS diisodecylphthalate (DIDP). The reason to study this system is the following. The high vapor pressure of many common diluents in relation to the liquidliquid demixing temperature is often a problem studying the thermally-induced phase-separation method experimentally. Special care has to be taken to prevent evaporation of the diluent during an experiment. The diluent DIDP shows a cloud point with the polymer of about 50°C and has a very low vapor pressure. (For comparison, the vapor pressure of water at T = 20°C is 2.10 3 Pa and the vapor pressure of DIDP at T = 100°C is 0.1 Pa [70].) Therefore, for this system no special care has to be taken to avoid evaporation of diluent.
3.2
Experimental Homogeneous solutions of aPS in DIDP (Merck-Schuchardt, purity>99%) were prepared in the same way as described in section 2.3.2. The viscosity of the polymer diluent system aPS-DIDP was measured with a Brabender Viscotron, Mod. Nr 8024. Starting at high temperatures (in the homogeneous solution), the viscosity was measured for different shear rates at a fixed temperature. Subsequently the temperature was lowered and the complete procedure was repeated. Furthermore, polymer solutions of aPS in DIDP were studied in a Bohlin Rheometer CS50. Upon cooling with a cooling rate of 2 K.min -1, the complex viscosity was measured for different strains at a constant frequency of 0.1 Hz.
56
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
3.3
Results 200
150 --t~ ¢~
13_ >"100 0 ¢o
>
50-
~'~1 I 20
i
40
I 60
s~ = 80
T e m p e r a t u r e (°C) Figure 6. Viscosity of a 30 wt.% solution of aPS in DIDP as a function of the temperature upon cooling for different shear rates (indicated in the figure). Results of the viscosity measurements are visualized in Fig. 6 for a polymer concentration of 30 wt.%. Experiments in the concentration range of polymer between 20 and 40 wt.% show a comparable trend'as plotted in Fig. 6. The observation of an increasing viscosity upon cooling as the temperature approaches the liquid-liquid demixing temperature is in agreement with the work of Wolf et al. [51]. In fact, the location of the sharp decrease in viscosity (between T = 45 and 50°C) is in good agreement with cloud point experiments of this polymer-diluent system (T = 49°C). The observation of the sudden decrease at lower temperatures inside the liquid-liquid demixing gap is not in agreement with the expectation. Upon cooling a polymer solution, the viscosity increases. At the liquid-liquid demixing temperature, a polymer-rich matrix is formed enclosing the polymerlean phase. The viscosity is expected to be dominated by the viscosity of the continuous, polymer-rich, phase of the system. Therefore, it should be expected that upon liquid-liquid demixing the viscosity increases further. An explanation of the sharp decrease in viscosity could be the loss of contact between the polymer rich matrix and the rotating cone of the viscometer. Upon storage below the cloud point temperature the demixed polymer solution displays synereses; DIDP is expelled out of the continuous phase. This process is probably enhanced by the applied shear rate resulting into the formation of a slip
57
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van I)er Heijden
layer of DIDP between the polymer rich matrix and the rotating cone. To apply less mechanical force on the polymer solution, an oscillating rheometer was used instead of the rotating viscometer. 2 D1
?
t
Q... >,
.~_ f/)
o o
U)
•~
0.5
•
x
1
Q.,
E o
20
30
40
50
60
70
Temperature (*C) Figure 7. Complex viscosity for a 30 wt.% solution of aPS in DIDP. The frequency is 0.1 Hz. The strain varies from 0.1 to 0.001. The cooling rate is 2 K-min-1. In Fig. 7, a small plateau can be observed upon cooling at the liquid-liquid demixing temperature in the complex viscosity at a low strain (0.001). However, as expected, the complex viscosity increases upon further cooling. With increasing strains the plateau becomes longer and the complex viscosity even shows a depression at a strain of 0.1. The reason of this depression is probably the same as in the viscosity experiments. Diluent is pushed out of the demixed solution and forms a slip layer. When using rheology or viscometry experiments to determine the liquidliquid demixing temperature, sufficient mechanical force should be applied to achieve immediate synereses of the demixed solution. The synereses results into an apparent viscosity drop that could be used as a marker for the onset of demixing. However, for studying the mechanical properties of the demixed solution, very small strains are recommended in order to avoid shear enhanced synereses. Otherwise, the measurement of the mechanical properties of the demixed solution is unreliable.
58
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
3.4
Conclusions and Outlook Liquid-liquid demixing temperatures can be observed with rheology and viscometry by using high shear rates. To study the mechanical behavior of demixing polymer solutions, care has to be taken to avoid altering of the liquid structure due to shear forces. It should be of much interest when rheological experiments can be carried out with a polymer solution at different frequencies. These results can be compared with results of emulsions to obtain information about the structure belonging to the demixed solution. Furthermore, rheological experiments can be very suitable to study the mechanical behavior of the polymer-diluent system at a temperature in the region of the structure fixation step. For example, the system aPS - DIDP forms a gel at room temperature because of the high viscosity of the polymer solution. After the gelation of the solution, no growth of the demixed domains is observed. Hence, gelation is sufficient to stop the liquid-liquid demixing process. The polymer-diluent system aPS - 1-dodecanol, which was studied in section 2, shows an onset of the glass transition temperature at about 65°C. The glass transition temperature is known to be sufficient to stop the liquid-liquid demixing process. However, it is very interesting to study whether at higher temperatures than the glass transition temperature, the liquid-liquid demixing process is influenced by or even stopped by gelation. Rheological experiments have the potential to clarify the structure fixation step of a polymer solution.
4.
ACKNOWLEDGEMENTS
E. Schomaker, J. Bos en R. Lammers of Akzo Nobel in Amhem are acknowledged for discussions with respect to the TMDSC work, and for giving us the possibility to carry out the TMDSC experiments. B. Reuvers (Akzo Nobel in Arnhem) and M. van Egmond (University of Twente) are acknowledged for carrying out the rheological experiments and for the discussions. REFERENCES
[1] [2] [3] [4] [5] [6]
M. Mulder, Basic principles of membrane technology, Kluwer Acedemic Publishers, Dordrecht, (1996). L.J. Zeman and A.L. Zydney, Microfiltration and Ultrafiltration, Marcel Dekker, Inc., New York, (1996). D.R. Lloyd, K.E. Kinzer and H.S. Tseng, J. Membrane Sci., 52 (1990) 239. A.T. Young, J. Vac. Sci. Techn., A 4 (1985) 1128. J.M. Williams and J.E. Moore, Polymer, 28 (1987) 1950. C. Schugens, V. Maquet, C. Grandfils, R. Jerome and P. Teyssie, Polymer, 37 (1996) 1027.
59
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
[7] [8] [9] [10] [11] [12] [13] [14] [15] [16] [17]
[18] [19] [201 [21] [22] [23]
[24] [25] [26] [27]
[28] [29] [30] [31] [32] [33] [34]
[35] [36] [37]
[38] [39]
[40] [41]
[42] [43] [44] [45]
Y.S. Nam and T.G. Park, J. Biomed. Mat. Res., 47 (1999) 8. A.J. Castro, US-patent 4247498 (1981). G.H. Shipman, US-patent 4539256 (1985). C. Josefiak and F. Wechs, GB-patent 2115425B (1985). W.C. Hiatt, G.H. Vitzthum, W.K.B., K. Gerlach and C. Josefiak, in Material science of synthetic membranes, Lloyd, D.R.(Ed.) American Chemical Society, Missouri (1984) 229. S. Berghmans, H. Berghmans and H.E.H. Meijer, J. Membrane Sci., 116 (1996) 171. H. Matsuyama, S. Berghmans and D.R. Lloyd, J. Membrane Sci., 142 (1998) 213. H. Matsuyama, S. Berghmans and D.R. Lloyd, Polymer, 40 (1999) 2289. P.M. Atkinson and D.R. Lloyd, J. Membrane Sci., 175 (2000) 225. P.M. Atkinson and D.R. Lloyd, J. Membrane Sci., 171 (2000) 1. A. Laxminarayan, K.S. McGuire, S.S. Kim and D.R. Lloyd, Polymer, 35 (1994) 3060. K.S. McGuire, A. Laxminarayan and D.R. Lloyd, Polymer, 35 (1994) 4404. K.S. McGuire, K.W. Lawson and D.R. Lloyd, J. Membrane Sci., 99 (1995) 127. K.E. Kinzer and D.R. Lloyd, Polym. Mater. Sci. Eng., 61 (1989) 794. D.R. Lloyd, S.S. Kim and K.E. Kinzer, J. Membrane Sci., 64 (1991) 1. S.S. Kim and D.R. Lloyd, J. Membrane Sci., 64 (1991) 13. S.S. Kim and D.R. Lloyd, Polymer, 33 (1992) 1047. J.H. Aubert, Macromolecules, 21 (1988) 3468. J.H. Aubert and C.R.L., Polymer, 26 (1985) 2047. M.R. Caplan, C.-Y. Chiang, D.R. Lloyd and L.Y. Yen, J. Membrane Sci., 130 (1997) 219. B.J. Cha, K. Char, J.-J. Kim, S.S. Kim and C.K. Kim, J. Membrane Sci., 108 (1995) 219. C.L. Jackson and M.T. Shaw, Polymer, 31 (1990) 1070. L. Aerts, M. Kunz, H. Berghmans and R. Koningsveld, Makromolek. Chem., 194 (1993) 2697. H.C. Vadalia, H.K. Lee, A.S. Myerson and K. Levon, J. Membrane Sci., 89 (1994) 37. L.L. Whinnery, W.R. Even, J.V. Beach and D.A. Loy, J. Polym. Sci. Polym. Chem., 34 (1996) 1623. P. Vandeweerdt, H. Berghmans and Y. Tervoort, Macromolecules, 24 (1991) 3547. P.D. Graham, A.J. Pervan and A.J. McHugh, Macromolecules, 30 (1997) 1651. G.T. Caneba and D.S. Soong, Macromolecules, 18 (1985) 2538. F.J. Tsai and J.M. Torkelson, Macromolecules, 23 (1989) 775. F.J. Tsai and J.M. Torkelson, Macromolecules, 23 (1990) 4983. S.-W. Song and J.M. Torkelson, Macromolecules, 27 (1994) 6389. R.M. Hikmet, S. Callister and A. Keller, Polymer, 29 (1988) 1378. J.H. Aubert, Macromolecules, 23 (1990) 1446. S.-W. Song and J.M. Torkelson, Polym. preprints, 34 (1993) 496. S.-W. Song and J.M. Torkelson, J. Membrane Sci., 98 (1995) 209. J. Amauts, H. Berghmans and R. Koningsveld, Makromolek. Chem., 194 (1993) 77. R. Koningsveld and A.J. Staverman, J. Polym. Sci., Part A-2, 6 (1968) 349. R. Koningsveld, W.H. Stockmayer and E. Nies, Polymer phase diagrams, Oxford University Press, Oxford, (2001). F. Khoury and E. Passaglia, in Treatise on solid state chemistry, Plenum Press, Hannay, N.B. (Ed.), New York-London, (1976) 335.
60
Differential Scanning Calorimetry And Rheological Experiments To Study Membrane Formation Via Thermally-Induced Phase-Separation Van Der Heijden
[46] [47]
[48] [49]
[50] [51] [52]
[53] [54] [55] [56] [57]
[58] [59]
[60] [61] [62] [63] [64] [65] [66] [67]
[68] [69] [70]
F.N. Kelley and F. Bueche, J. Polym. Sci., 50 (1961) 549. S. Nojima, K. Shiroshita and T. Nose, Polym. J., 14 (1982) 289. J. Lal and R. Bansil, Macromolecules, 24 (1991) 290. K.S. McGuire, A. Laxminarayan and D.R. Lloyd, Polymer, 36 (1995) 4951. J. Szydlowski and W.A. Van Hook, Macromolecules, 31 (1998) 3255. B.A. Wolf and M.C. Sezen, Macromolecules, 10 (1977) 1010. Y. Xie, K.F.J. Ludwig, R. Basnil, P.D. Gallagher, C. Kon~ik and G. Morales, Macromolecules, 29 (1996) 6150. J. Arnauts and H. Berghmans, Polym. Comm., 28 (1987) 66. J. Arnauts, R. De Cooman, P. Vandeweerdt, R. Koningsveld and H. Berghmans, Thermochim. Acta, 238 (1994) 1. P.C. van der Heijden, M.H.V. Mulder and W. M., Thermochim. Acta, 378 (2001) 27. B. Wunderlich, Thermal Analysis, Academix Press, Inc., London, (1990). V.B.F. Mathot, Calorimetry and thermal analysis of polymers, Hanser Publishers, Munich, (1994). M. Reading, Trends Polym. Sci., 1 (1993) 248. M. Reading, B.K. Hahn and B.S. Crowe, US-patent 5224775 (1993). J. Schawe, Thermochim. Acta, 260 (1995) 1. T. Ozawa and K. Kanari, Thermochim. Acta, 253 (1995) 183. K.J. Jones, I. Kinshott, M. Reading, A.A. Lacey, C. Nikolopoulos and H.M. Pollock, Thermochim. Acta, 304/305 (1997) 187. G.W.H. H/Shne, Thermochim. Acta, 304/305 (1997) 121. J.E.K. Schawe, Thermochim. Acta, 304/305 (1997. B. Wunderlich, A. Boiler, I. Okazaki and K. Ishikiriyama, Thermochim. Acta, 304/305 (1997) 125. R. Scherrenberg, V. Mathot and A. Van Hemelrijck, Thermochim. Acta, 330 (1999) 3. R. Scherrenberg, V. Mathot and P. Steeman, J. Therm. Anal., 54 (1998) 477. TA Instruments, Modulated DSC (MDSC) How does it work ? (1998). B. Wunderlich, ATHAS Databank, http://web.utk.edu/~athas/databank/intro.html (2000). T.E. Daubert, R.P. Danner, H.M. Sibul and C.C. Stebbins, Physical and thermodynamic properties of pure chemicals. Data compilation, Taylor&Francis, Pennsylvania, (1989).
61
This page intentionally left blank
Non-Invasive Characterizations of Membrane Fluid Transport and Fouling
63
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes
D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 4
Study of membrane fouling and cleaning in spiral wound modules using ultrasonic time-domain retlectometry Zh.-X. Zhang a, A.R. Greenberg b, W.B. Krantz ¢, and G.-Y. Chai" aDepartment of Chemical Engineering, University of Colorado at Boulder, Boulder CO 80309-0424, USA bDepartment of Mechanical Engineering, University of Colorado at Boulder, Boulder CO 80309-0427, USA CDepartment of Chemical Engineering, University of Cincinnati, Cincinnati, Ohio 45221-0171, USA NSF Membrane Applied Science and Technology Center, University of Colorado, Boulder, CO 80309-0424, USA ABSTRACT Better detection and control of fouling in liquid-separation processes is essential if membranes are to find increased use in a variety of industrial applications. Ultrasonic time-domain reflectometry (UTDR) is an in situ, noninvasive real-time technique that has been successfully utilized to quantify fouling and cleaning of fiat sheet membranes. This study describes the extension of UTDR for the measurement of fouling and cleaning in commercial membrane modules employing spiral-wound elements. Experiments were conducted using a high-pressure separation system incorporating a commercial spiral-wound module. Calcium sulfate was utilized as the primary scalant in concentrations ranging from 1.2-1.8 g/L; ferric hydroxide colloidal fouling also occurred during the tests. A multi-cycle protocol was adopted whereby each cycle consisted of three stages: pure water equilibration, fouling and cleaning. Permeation rate as well as acoustic amplitude and arrival time measurements were made at regular intervals during three complete cycles on two modules. A new signal analysis protocol was developed such that systematic changes in the entire acoustic spectrum as a function of module operation time could be represented in terms of shift factors. In addition, gravimetric, microscopic and spectroscopic measurements were made at the end of the experiments so that the extent of the membrane fouling and cleaning could independently assessed. Results indicate that the overall decline in permeation rate is reasonably correlated with changes
65
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
in both the amplitude and arrival time shift factors. Overall, this work confirms the effectiveness of the UTDR technique and provides a sound basis for continued development with an ultimate goal of providing on-line sensors for the timely detection of fouling and cleaning of commercial liquid-separation membrane systems. 1.
INTRODUCTION
As generally def'med, membrane fouling involves the deposition of retained particulates, colloids, macromolecules, crystallization of dissolved inorganic salts on or within the membrane. Fouling is the most critical problem limiting further growth and wider application of membrane-based liquidseparation processes [ 1]. Although the occurrence of fouling can be traditionally inferred from a marked decline in permeate flux or product quality, such an inference is unreliable since other phenomena such as concentration polarization, membrane compaction and degradation may also lead to similar changes in permeate flux or quality [2]. Moreover, using currently available techniques, it is not possible to quantify when a commercial membrane module has been adequately cleaned via chemical or mechanical de-fouling. Consequently, there is a compelling rationale for the development of an in situ, noninvasive technique that can monitor the growth and removal of a membrane-fouling layer in real time under realistic operating conditions. Such a technique would not only provide for identification of the specific phenomena underlying the flux decline, but also enable the improvement of fouling remediation strategies. Ultrasonic time-domain reflectometry (UTDR) has been used in a number of applications that involve the measurement of the location and topographical characteristics of a moving or stationary interface [3]. When the operating requirements of commercial high-pressure separation modules are considered, UTDR appears to be more suitable for real-time monitoring of membrane fouling than other noninvasive techniques, such as nuclear magnetic resonance microimaging [3]. Indeed, previous studies have demonstrated that UTDR can be successfully utilized to quantify fouling in a flat-sheet membrane module [2-4]. Recently, the UTDR technique was extended to the much more complex situation that occurs with the multiple layers that comprise spiral-wound reverse-osmosis (RO) modules. As in the flat-sheet studies, this work utilized calcium sulfate as the primary scalant due to its importance as a membrane-fouling agent [5-8]. Whereas the initial efforts to interpret the complex acoustic spectra utilized only a few of the many signal peaks [9,10], the present study employed a signal
66
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Refleetometry - Greenberg
analysis protocol that accounts for systematic shifts in the entire acoustic spectrum as a function of module operating time. This paper also reports the first results obtained for UTDR detection of deposits formed by multiple foulants. 0
2.1
ULTRASONIC TIME-DOMAIN REFLECTOMETRY (UTDR)
Basic Concepts
The basic principles of UTDR and their application to the measurement of fouling on a fiat-sheet membrane have previously been described [2,3]. As shown in Figure 1, the ultrasonic wave pulses are transmitted by the transducer and propagate through the module. When the pulses encounter an interface, such as the one formed by the feed solution and top surface of the membrane, they will be reflected back to the transducer. The amplitude of the reflected waves depends on the acoustic impedance difference between the media on either side of the interface and the topography of the interface. The acoustic impedance is a function of the physical characteristics of the medium, and is given by the product of the density and the acoustic velocity within the medium. In addition, the time required for the transmission and reflection of a wave from the interface (i.e. arrival time) is a function of the path length and wave velocity: 2As AT = ~
(1)
where AT is the arrival time, As is the distance between the transducer and the interface and c is the acoustic velocity in the medium. Since the impedance, interface properties and path length may change with the growth of a fouling layer on the membrane surface, the change in the amplitude (AV) and the shift in the arrival time (AT) of the interface echoes can be analyzed and used to monitor fouling quantitatively in real-time.
2.2
Adaptation for Spiral-Wound Membrane Modules When the UTDR technique is applied to a spiral-wound RO membrane module contained in a cylindrical fiber-reinforced polymer housing, the signal analysis is complicated by two factors. First, a much more complex signal pattern will be obtained due to multiple reflections from the outer and inner surfaces of each layer of the multiple windings in the module assembly (Figure 2) [ 10]. Second, a significant loss of acoustic information will be caused by signal attenuation through these layers as well as through the module shell. The various reflections from these interfaces correspond to the many peaks in the waveform
67
Figure 1. Schematics that represent the principles of UTDR measurement: (Left) cross-sectional view of the separation cell, the externally mounted ultrasonic transducer, and the primary reflections, 1, 2, and 3 from the cell plate/fluid, fluidmembrane, and fluidfouling layer interfaces, respectively; (Right) the general ultrasonic response in the time-domain for each of these reflections.
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
obtained by plotting the signal amplitude as a function of the arrival time (Figure 3). Here the peaks in the waveform represent very complex superposed reflections from the many close-packed interfaces within the spiral-wound module instead of the simpler patterns from the largely separated interfaces that are characteristic of the fiat-sheet geometry. Within this complex spectrum, certain peaks are more responsive to the growth and subsequent removal of fouling layers than others. In addition, the shape of the waveforms depends on the position and mounting mode of the transducer on the external surface of the module shell. In the previous studies [10], considerable effort was required to select and analyze those peaks in the amplitude versus arrival time spectra that appeared most sensitive to fouling. The technique employed was time-intensive and made real time analysis difficult. In addition, all of the acoustic information obtained could not be fully used. In order to overcome these limitations, a more sophisticated signal acquisition and analysis protocol was developed. This procedure accounts for systematic shifts in the entire acoustic spectrum as a function of module operating time and enables information about the extent of fouling to be obtained in real time. The specific steps of the signal analysis protocol are summarized below: • All acoustic spectra are digitized. • A reference spectrum is obtained for a particular set of operating conditions with the spiral-wound module of interest. The reference spectrum corresponds to a "zero" operating time at which fouling has not occurred. Changes in the acoustic spectra as a function of operating time are determined based upon a systematic departure or shift from the reference spectrum. Such changes are shown in Figure 3 for the RO module employed in this study. Here, two waveforms corresponding to the reference spectrum and an updated spectrum, i.e. a spectrum obtained at a later time, are superposed to demonstrate the changes in the signal amplitude and arrival time. • Two shift factors are computed using a root-mean-square calculation such that values are always positive. One shift factor, QA, is based upon changes in the signal amplitude and the other factor, QT, is based upon changes in the signal arrival time. The two relationships are: N
"QT = '
~(Tiu_Tir) 2 1/2 N
(2)
Here, Viu and Vir are the amplitude of peak i in the updated spectrum and the reference spectrum, respectively; Tiu and Tir are the arrival time of peak i in the updated spectrum and the reference spectrum respectively; and N is the total number of peaks in the spectrum under consideration.
69
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
Transducer
Shell Feed Channel Spacer Membrane Permeate Collection Film
II II II [Layer1
~l ~]
Layern
-~ [
[
Figure 2. Schematic showing a cross-sectional view of a spiral-wound RO module assembly, including the fiberglass shell, the position of an externally mounted ultrasonic transducer, and reflections from the multiple interfaces. The width and length of the arrows are proportional to the reflected amplitude and arrival time, respectively.
2.50E-02
;~ 0.00E+00 3.7:E-0
J
E-05
t
AT (s)
°~,,i
-2.50E-02
Figure 3. Representative results showing a portion of the acoustic response in terms of signal amplitude versus arrival time (AT) from a commercial spiral-wound RO module. Changes in the UTDR signal peaks that occur as function of operating time are observed as a shift from the reference spectrum (heavy line) to the updated spectrum (light line).
The effectiveness of the new methodology was assessed by quantifying the signal fluctuations obtained during 30-hr runs using pure-water, and then representing
70
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
the effects of module fouling in terms of the shift factors based upon anticipated changes in the signal amplitude and arrival time. An "ideal" response is shown in Figure 4 and indicates the expected relationship between the permeation rate and the amplitude shift factors. Here, the shift factors and the permeation rate are plotted as a function of operating time for a pure water equilibration step, a period of CaSO4 fouling, and a pure-water cleaning step. The signal analysis protocol is designed to distinguish the relatively small-scale fluctuations due to normal operating condition variations from the large-scale changes that are due to fouling. The dashed lines represent a "small-scale fluctuation" confidence interval for each parameter. From a practical point of view, a successful technique should indicate when the signal has moved out of the normal operating window, i.e. the confidence interval, as a result of module fouling. The new methodology also allowed instantaneous calculation of the shift factors so that the occurrence and removal of fouling in the RO spiral-wound module could be detected and monitored quantitatively in real-time. Moreover, the portion of the spectrum that provides the basis for the shift factor calculations can be easily altered. For example, Figure 5 shows a representative waveform in which the spectrum is divided into three regions, each designated by a different rectangle. The waveform within each rectangle contains information from different layers within the module. Since fouling does not occur to the same extent in all regions of the module at the same time, particular portions of the spectrum will be most responsive to its occurrence and thus provide the most sensitive detection of the growth and removal of fouling layers. 3. EXPERIMENTAL 3.1. RO System and UTDR Instrumentation Based on previous work [9,10], the experiments were conducted on a bench-scale RO flow system that allowed for the control and measurement of important system operating parameters including temperature, pressure, flow rate and calcium-sulfate feed concentration. The major components of the RO system consisted of a Koch 2521 spiral-wound RO membrane module, a feed tank with temperature control, a low-pressure pump used to transport the feed through a 2micron prefilter, and a high-pressure booster pump to drive the RO membrane separation process (Figure 6). The system was designed to enable both the retentate and permeate from the membrane module to be returned to the feed tank to enable stable operation and conservation of chemicals. The acoustic hardware included two specially designed high-frequency (3.5 MHz) transducers (Research Institute of Acoustics, Chinese Academy of Science), a pulsar-receiver (Panametrics 5052 PRX) and a digital oscilloscope (Nicolet Pro 50) that enabled a 20-ns sampling rate. A custom mounting assembly was utilized to allow the transducers to be securely attached at various
71
Operating Time (hr) Figure 4. Schematic showing the anticipated change in the UTDR shift factors (QA and QT)and the permeation rate using the new signal analysis protocol during the water equilibration (WE), fouling (F) and cleaning (C) phases of module operation. The solid lines represent the overall shift factor and permeation rate behavior as determined from values obtained at discrete time intervals (solid circles). The dashed lines indicate a confidence interval for normal operation. Fluctuations within the confidence interval are acceptable but departures outside the dotted lines are taken to indicate module fouling.
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
positions on the external surface of the module. Secure attachment is essential to maximize changes in the acoustic waveforms that correspond to membrane fouling and cleaning. In the current experiments the transducers were mounted on the midstream and downstream portion of the module. Such multiple-point sampling can provide important information about the relative uniformity of the fouling and cleaning processes as a function of position. The UTDR signals were continuously monitored and recorded (waveform records containing 1000 data points) using a data-acquisition system that consisted of a personal computer (Pentium II 300 MHz) with a GPIB board (CyberResearch). Custom dataacquisition software enabled interfacing with the measurement instrumentation, real-time analysis of the UTDR signals, calculation of the two shift factors, and graphical display of both the shifts in the acoustic spectrum and changes in the values of the two shift factors. The acoustic hardware included two specially designed high-frequency (3.5 MHz) transducers (Research Institute of Acoustics, Chinese Academy of Science), a pulsar-receiver (Panametrics 5052 PRX) and a digital oscilloscope (Nicolet Pro 50) that enabled a 20-ns sampling rate. A custom mounting assembly was utilized to allow the transducers to be securely attached at various positions on the external surface of the module. Secure attachment is essential to maximize changes in the acoustic waveforms that correspond to membrane fouling and cleaning. In the current experiments the transducers were mounted on the midstream and downstream portion of the module. Such multiple-point sampling can provide important information about the relative uniformity of the fouling and cleaning processes as a function of position. The UTDR signals were continuously monitored and recorded (waveform records containing 1000 data points) using a data-acquisition system that consisted of a personal computer (Pentium II 300 MHz) with a GPIB board (CyberResearch). Custom dataacquisition software enabled interfacing with the measurement instrumentation, real-time analysis of the UTDR signals, calculation of the two shift factors, and graphical display of both the shifts in the acoustic spectrum and changes in the values of the two shift factors.
3.2. Experimental Procedure As indicated in Table 1, the testing protocol utilized three cycles, each of which generally contained water equilibration, fouling and cleaning phases. During the water-equilibration phase (phase A), the membrane module was operated using pure deionized water until the membrane module performance (permeation rate) and the UTDR signals (amplitude and arrival time) reached steady-state values. Then the feed was switched to an aqueous calcium sulfate solution that was prepared by adding anhydrous calcium sulfate powder (Aldrich, 325 mesh) to deionized water (equilibrium calcium sulfate solubility: 2 g/L).
73
0.30
0.20
0.10
0.00
-
0 0
0.10
c :
a
-0.20
0
~ o . Y o o o o . . . o o . o o o o o o o o o o o o o o o o o o o o o o o ~ o o o . o o o ~
-0.30
~
1
-0.40
I
Arrival Time (ps) Figure 5. The regions (A, B, and C) of the representative waveform each correspond to reflections from different locations within the spiral-wound module: The portion of the signal indicated by A (within square) is contained within the larger region given by B (within rectangle); both A and B correspond to portions of C. The methodology employed in this study enables selection of the particular region of the spectrum from which the shift factors are calculated. This provides the means to enhance the sensitivity for fouling detection.
v1
Control Valve
L
Y
s
fP
I
d I FeedTank
I
Filter
F
Flow Feed Pump 2
E
Transducers
5’
00
L
‘
Feed Pump 1
1
:
System Electronics
0
c
.
:
5’
m
F
t-
Transducer
s5’
.............................................................................................
110
f
R’
4
I’
Pulser Receiver Digital Oscilloscope Computer ............................................................................................
B
I
Figure 6. Schematic of the high-pressure RO separation system that shows the major components including the integrated capabilities for UTDR measurement. The arrows indicate the directions for the recycled retentate and permeate flows. The system electronics for acoustic measurements is detailed in the enlarged view at the bottom.
fff.
I B P
d
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Refleetometry - Greenberg
This initiated phase B in which concentration polarization and fouling occurred. This phase was generally allowed to continue until the ultrasonic responsed permeation rate approached steady-state behavior. In the subsequent cleaning phase (phase C), the feed solution was switched back to pure deionized water in order to remove the fouling layer from the membrane surfaces. Overall, long-term experiments were conducted on two modules in order to accommodate the multiple fouling and cleaning cycles; the modules were run for a total of 654 and 1108 hours, respectively. In each experiment the fouling cycles generally utilized systematically higher concentrations of calcium sulfate. Table 1. Details of the Fouling and Cleaning Protocol
Module Cycle/Phase I Cycle 1:CaSO4 (1.2 g/L) Phase A: Water Equilibration Phase B: Fouling Phase C: Cleaning Cycle 2:CaSO4 (1.6 g/L) Phase A: Water Equilibration Phase B: Fouling Phase C: Cleaning Cycle 3:CaSO4 (1.6 g/L) Phase A: Water Equilibration Phase B: Fouling Total Operating Time (hr) II Cycle 1:CaSO4 (1.2 g~) Phase A: Water Equilibration Phase B: Fouling Phase C: Cleaning Cycle 2:CaSO4 (1.6 g/L) Phase A: Water Equilibration Phase B: Fouling Phase C: Cleaning Cycle 3:CaSO4 (1.8 g/L) Phase A: Water Equilibration Phase B: Fouling Phase C: Cleaning Total Operating Time (hr)
Duration (hr) 37 164 82 30 100 71 70 100 654 72 102 150 75 126 110 33 296 144 1108
All experiments were conducted under the same operating conditions. The upstream pressure of the module was maintained as 0.7MPa, and the feed
76
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
solution flow rate was maintained as 0.9 L/min. The temperature of the feedsolution in the tank was controlled at 20°C and the inlet temperature of the module was maintained at 22°C. The CaSO4 concentration of the feed, permeate and retentate were measured at regular intervals using a conductivity meter (Yellow Spring Instruments), and the membrane rejection was calculated via standard methods. In addition, the permeate and retentate flow rates were determined via a digital balance and timer. In order to confirm independently the extent of membrane fouling indicated by the values of the acoustic shift, each of the two RO modules was opened after the appropriate phase in the operating protocol for visual inspection, morphological studies, and gravimetric and elemental analysis. Module I was opened at the end of the Cycle 3 fouling phase to observe the extent of coverage as well as the distribution of the fouling layer on the membrane surfaces, while Module II was opened at the end of Cycle 3 cleaning phase to determine the degree of fouling deposit removal via pure water cleaning (Table 1). The morphological features of the fouled and water-cleaned membrane samples were characterized using light microscopy (Nikon Labophot) and scanning electron microscopy (Aspex Instruments). Gravimetric analysis was performed o n sections of the unrolled membrane using a microbalance, and the elemental composition and distribution of the deposits were determined via energy dispersive spectroscopy (EDS). 4. RESULTS AND DISCUSSION The important aspects of the UTDR response are demonstrated in the data obtained for module I after the completion of each of the three cycles. Results for the first cycle are shown in Figure 7 for the two shift factors, QA and QT, from the downstream transducer as well as the permeation rate as a function of module operating time. In order to compare directly the sensitivity and effectiveness of the UTDR technique with the new signal analysis protocol and the traditional fouling detection mode, changes in both the shift factors and the permeation rate are represented as percentage changes from their respective reference values, i.e. the values at the start of phase A. During the water-equilibration phase, the permeation rate is relatively constant, whereas there is a gradual but distinct increase in QA. During the fouling phase, the permeation rate declines by approximately 12% and QA increases by over 1500%. By completion of the cleaning phase, the permeation rate has returned to its reference value; however, QA remains almost unchanged. In addition, no apparent change in QT is observed over the course of cycle 1. Note that the spikes in QT at the end of the fouling run and the beginning of the cleaning run are believed to be due to a problem with the power supply. Similar results for the shift factors were obtained from the midstream transducer. Overall, the observed behavior in QA and QT was not as expected. Nevertheless, we continued with the protocol indicated in Table 1.
77
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
Results for the second cycle are shown in Figure 8. Here each of the shift factors is represented by three different curves, each of which corresponds to the different regions of the signal spectrum shown in Figure 5. These data demonstrate that the shift factors from each region of the spectrum lead to a similar response although the differences suggest that particular regions of the spectrum may indeed be more sensitive to the presence and removal of a fouling layer. After establishing steady-state behavior during the equilibration phase, there is a marked decline in the permeation rate (~20%) and an increase in QA and QT during the fouling phase. During the subsequent cleaning phase, the permeation rate and QT return to their reference values. Although QA clearly trends in the same direction, this factor has not reached the reference value at the completion of the cleaning cycle. The results for cycle 2 obtained from the midstream transducer, however, were considerably different than those from the downstream transducer. Indeed, the signals from midstream transducer provided shift factor responses that were rather similar to those observed during cycle 1 (Figure 7). This emphasizes a fundamental difference between the permeation and ultrasonic measurements whereby the former represents an average response for the entire module while the shift factors can reflect the condition of the module on a local basis. A third operating cycle was then conducted under the same conditions as for cycle 2. However, in this case the protocol was modified so that instead of a cleaning phase, the membrane module would be opened at the end of the fouling phase so that microscopic, spectroscopic and gravimetric analyses could be performed. Results for the permeation rate and shift factor responses during cycle 3 are shown in Figure 9. As expected, the three response parameters are reasonably constant during the water equilibration phase. With the initiation of the fouling phase, there is a relatively rapid decrease in the permeation rate (~20%) over the first 10 hours to an approximately steady-state value that is maintained for the next 90 hours until the end of the run. In contrast, both shift factors increase throughout the duration of the fouling phase. Whereas QA increases at a relatively constant rate, the initial rate of increase of QT becomes more modest after the first 30 hours of fouling. Overall, these results confirm the trends observed during cycle 2 (Figure 8). At the end of the fouling phase in cycle 3, the module was opened and unwound. Visual examination indicated the presence of a reddish-brown deposit on the membrane surface, with the color turning progressively lighter (whiter) from the entrance to the exit end of the module. Microscopic examination confirmed the coexistence of the reddish-brown and white deposits where the latter evidenced the same characteristics previously reported for crystalline CaSO4 (Figure 10) [2,3]. The CaSO4 crystals were mainly deposited along the regularly distributed convex lines formed on the membrane sheet during
78
Study O f Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
800 2500 600
z
o~'1500
< O "O 500 N t~ E O z -500
3 ~°
,~
i
0
.
.
.
.
.
.
.
.
.
.
.
. vl.........................................................t
..................
,I . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ............................................................................................... 0
•
i
..................
1..............................................................................................................
1 400N -I
4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
i ~.
,,
200~
~
I
-1500
o "ql-WE ~'q
F
" "
C
i1,'- -'ql
"
120 v
"~ 110 n,"
P,
&l& •
100
E • 90
••
• ••~4t.-AA• •
•••~••
• N
~
0 I I I
8o
I
E L_
0
z
I I
70
|
|
|
|
|.
50
|
| .
|
|
.
|
100
i
. i
i
i
i
I
I
I
i
Ii
150 200 Operating Time (hr)
i
i
i
i
~
250
.
.
.
.
300
Figure 7. Results obtained for cycle 1, module I (Table 1) including the water equilibration (WE), fouling (F) and cleaning (C) phases and shown as percent changes from the respective reference values. Top: Amplitude and arrival-time shift factors (downstream transducer); bottom: permeation rate. Permeation rate values in F and C phases based upon reference value that incorporated the membrane compaction that occurred during the WE phase.
79
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
800 I
1500
I I I I
!
"
500
.,L__
~
...........................................................................................
<
600
''r~'~':S~~::':~:~ ...........---.,,~,...................................... I~,,,:( ~
. . . . . . . ..V-.=c:_:.:;.
z o 3
i ............................................................
(J
-
-o
-500
N
400~"
tl)
I I
D
.-t
E t,_
- 200~
o
Z -1500
-2500
',
I
I
!
!
1
o~' 120 ll)
co 110 n,' tO
co 100 E 1,-(I) Q.
90-
"o tl) ..-N
E I,_
,
80 I
,,
O
z
,!
!
70
'
0
'
J
! '
I
50
'
'
'
'
I
'
'
"
100
'
I
150
~
'
'
'
I
200
Operating Time (hr) Figure 8. Results obtained for cycle 2, module I (Table 1) including the water equilibration (WE), fouling (F) and cleaning (C) phases, and shown as percent changes from the respective reference values. Top: Amplitude and arrival-time shift factors (downstream transducer); the three lines for each shift factor correspond to the different regions of the spectrum (Figure 5). Bottom: permeation rate.
manufacturing process while the reddish-brown deposits were observed on the entire membrane surface. EDS analysis indicated that the surface deposits primarily consisted of an iron-containing portion (colloidal) and CaSO4 (Figure 11). We later determined that this iron deposit was caused inadvertently by the installation of a new section of pipe during the modification of the flow system
80
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
that originally utilized plastic or stainless steel components. We believe that the introduction of air into this aqueous, open loop system caused oxidation of iron ions (Fe+2 ---> Fe ÷3) and the subsequent formation of ferric hydroxide. Previous work has shown that insoluble ferric hydroxides (colloidal) can produce fouling of RO systems [12]. The presence of this mixed foulant provided an unanticipated degree of complexity to the experiments and also accounts for the unexpected shift factor response previously described. Formation and deposition of the iron-containing foulant on the membrane surface undoubtedly occurred during the initial exposure of the new section of pipe to water (cycle 1). Fouling due to the deposition of iron-containing layers most likely became more extensive during phase B because the introduction of CaSO4 would facilitate electrolytic corrosion of the iron pipe [ 13]. Correspondingly, QA evidenced only a slight increase during the cycle 1 water-equilibration phase but a much more pronounced increase during the subsequent fouling phase (Figure 7). In contrast, QT was relatively unchanged throughout cycle 1. This behavior would be expected from the deposition of a very thin layer with a relatively loose structure since QA is sensitive to the topographical characteristics whereas QT is most responsive to the thickness of a sufficiently thick deposit. In the present case this suggests that QA and QT respond preferentially to the deposition of the ironcontaining and CaSO4 layers, respectively. If correct, this indicates that the deposition of the iron-containing colloidal material accounted for a significant portion of the overall fouling that occurred during phase B of cycle. This hypothesis is supported by the results of the retentate concentration measurements that indicated values of only 1.6 g/L at the end of phase B. As previously described, the permeation rate and QA responses during the cleaning phase of cycle 1 were puzzling since the former returned to its reference value but the latter decreased only slightly. However, such behavior would be consistent with the fact that while the pure water quickly cleaned the CaSO4, the cleaning protocol was not effective in removing the iron-containing layer. These results also suggest that the iron-containing layer had only a relatively small effect on the permeate flux due to its non-dense colloidal structure. During cycle 2, QA remained relatively stable during phase A in contrast to the increase observed during cycle 1 (Figure 7). The behavior of QA in conjunction with the relatively constant values of QT during the waterequilibration phase of both cycles suggests that any additional iron-based foulinglayer deposition during phase A was not extensive. Results obtained during the cycle 2 fouling phase, when a higher concentration of CaSO4 (1.6 g/l) was used in the feed solution, are consistent with a fouling layer that consisted of more significant amounts crystalline CaSO4 as well as the iron-containing colloidal material. In this case the permeation rate decreased and both shift factors increased. During the subsequent pure-water cleaning phase, the permeation rate and QT return to their respective reference values. 81
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
2500
800
1500
600 Z O
'-'1
<
O
:=1
¢D
500
n.
"o
N
400
N t'tl
O
E -500 t_
--I
O
z
200
-1500
-2500 -,,
WE'
F
120 v
*-' m
110
n,
¢0 :~ t'tl
100
E t,_ ft. "o N "--
I
90
11
t
80
E t,,,. O Z
70
0
20
40
60
80
100
120
140
160
180
Operating T i m e (hr)
Figure 9. Results obtained for cycle 3, module I (Table 1) including the water equilibration (WE) and fouling (F) phases, and shown as percent changes from the respective reference values. Top: Amplitude and arrival-time shitt factors (downstream transducer); bottom: permeation rate. The module was opened aider the fouling phase was completed.
82
w3 0
Figure 10. Representative micrographs of the fouling deposits on the membrane surface of module I. Left: Low magnification view showing the calcium-sulfate regions (two dark gray ovals along the picture diagonal) and the iron-containing regions (surrounding widely distributed light areas); Right: Higher magnification view of the needle-like structures characteristic of crystalline calcium sulfate.
D
5.
C ,'
p 3
Figure 11 . Representative results obtained from energy-dispersive spectroscopy of the fouling deposits from module I. The backscattered image indicates the specific area from which the iron, calcium and sulfur scans were obtained. The relative brightness of the image indicates the concentration of the respective element. The images indicate that the composition of the oval areas is consistent with calcium sulfate whereas the deposits in the surrounding regions contain significant iron.
P
B
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Refleetometry - Gteenberg
However, although QA clearly decreases, the reference value is not attained by the end of the cleaning phase. These responses are consistent with the removal of the CaSO4 scalant but the continued presence of the ironcontaining layer. An interesting aspect of the phase C shift factor responses is that both QA and QT evidence a rather sharp decrease after approximately 20 hr of pure water cleaning. This behavior is in contrast to that of the permeation rate, which evidences a more gradual increase over this same time period. These characteristics underscore a fundamental difference between the two types of measurements: whereas the permeation rate represents the average condition of the whole module, the acoustic responses are site specific, i.e., they indicate the condition of a relatively small portion of the module. For example, the focused transducers used in this study have a "spot" size of only a few millimeters in the focal plane. Similarly, the failure of an acoustic parameter to return to the appropriate reference value is consistent with the continued presence of a fouling layer at that specific location even though the permeation rate indicates that the reference condition has been achieved. Moreover, differences in the shift factor responses obtained from transducers attached at different locations on the module can indicate that conditions at these locations are not the same. In the present study, gravimetric measurements indicated that the iron-containing deposit was concentrated on the upstream end of the module, whereas the crystallized CaSO4 scalant was concentrated on the downstream end. This situation accounts for the differences in the responses from the midstream and downstream transducers whereby the former reflects the greater influence of the iron-containing deposit, while the latter is more response to the CaSO4 scalant. The ability to obtain information from multiple locations represents an important advantage of the acoustic measurements since they can provide additional insight regarding module performance. In order to evaluate the reproducibility of the results, a second module was tested using the protocol described in Table 1. Although similar results were obtained for the first cycle in which the lowest concentration of CaSO4 (1.2 g/L) was used, the acoustic signals obtained during the second and third cycles were much noisier than those from module I. We believe that this problem was due to inconsistencies in remounting the transducers that in tuna produced excessive vibrations during module operation. Despite these difficulties, the basic trends described for module I were still evident for module II. As indicated in Table 1, the experimental protocol for module II was modified somewhat in that the experiment was ended and the module was opened after the completion of the cycle 3 cleaning phase. As shown in Figure 12, microscopic examination confirmed the presence of a reddish-brown
85
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
deposit on the membrane surface; however, here there was no evidence of the white crystalline deposit usually indicative of CaSO4 fouling. The presence of this iron-containing deposit on the surface after cleaning correlates well with the behavior of the shift factors during this cycle of module II operation whereby the factors decreased from their "fouled" values but did not reach their "clean" (reference) values. In contrast, the permeation rate did return to its reference value after the cleaning cycle. Previous studies have shown that at a pH value close to the isoelectric point of ferric hydroxide, fouling occurred via the deposition of loosely packed large aggregates that did not result in a significant decrease in membrane flux [11]. The micrographs shown in Figure 12 indicate the ferric deposits are loosely packed on the membrane surface. Consequently, the permeation rate completely recovers, whereas as the shift factors do not. In this situation the ultrasonic technique appears more sensitive to fouling layer formation and removal than the fluxdecline behavior. Based upon the previous successful UTDR studies of flat sheet membranes, the selection of calcium sulfate as a model fouling system for a first UTDR investigation of fouling and cleaning in spiral-wound membrane modules is a reasonable choice. However, calcium sulfate scaling is generally not considered to be the most common problem in commercial separations. Indeed, fouling due to the deposition of a colloidal layer is a much more common phenomenon, and can cause flux decline in almost all membranebased liquid separations [2]. The rate at which such flux decline occurs will depend upon the specific system and operating conditions and the particular characteristics of the resulting fouling layer. Within this framework, the capability of the UTDR technique to detect mixed-mode fouling comprised of colloidal as well as inorganic deposits is of considerable significance with respect to the application of this technique in many commercial situations in which mixed fouling occurs. 5. CONCLUSION This research continues the efforts to extend the UTDR technique to the measurement of fouling and cleaning in commercial membrane modules. The results of this study indicate that the new signal-analysis methodology along with corresponding software for real-time monitoring is an effective approach for quantitative detection of fouling and cleaning in spiral-wound modules. This research provides the first evidence that the UTDR technique is capable of detecting mixed-mode fouling comprising colloidal as well as inorganic deposits. Overall, the insights obtained from this study provide a strong basis for the continued development of the UTDR technique with an ultimate objective of successfully monitoring membrane fouling and cleaning under the 86
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
!bt
(dt
tot
Figure 12. Representative micrographs of the membrane surface after fouling and water cleaning: (a) section from module I indicates the presence of both calcium-sulfate crystals and grainy reddish-brown iron deposits after the cycle 3 fouling phase; (b) section from module II shows that only iron deposits remain after cycle 3 cleaning phase; (c) another section from module I shows the presence of both deposits after the cycle 3 fouling phase; and (d) same section as (c) but after overnight immersion in w a t e r - the calcium-sulfate deposits are removed but the iron deposits remain on the membrane surface.
rigors of commercial operating conditions. The research emphasizes the complexity of the fouling process and the necessity of a robust experimental design. In retrospect this study would have benefited from a larger scale effort involving the simultaneous operation of multiple modules, each outfitted with appropriate acoustic hardware. Although difficult to accomplish within cost and space limitations, such an approach represents a more efficient way to successfully evaluate all of the factors that influence the ultrasonic measurements during fouling and cleaning. Despite these limitations, the results clearly indicate the potential advantages of UTDR measurement including enhanced sensitivity and
87
Study Of Membrane Fouling And Cleaning In Spiral Wound Modules Using Ultrasonic Time-Domain Reflectometry - Greenberg
additional insight regarding module conditions as compared with information obtained from permeate measurements alone. ACKNOWLEDGEMENT
The authors gratefully acknowledge the support of this research by the NSF Membrane Applied Science and Technology (MAST) Center at the University of Colorado. REFERENCES [1] [21
[3] [4] [5] [6] [7] [8] [9] [10] [11] [12] [13]
G. Belfort, in: G. Belfort (Ed.), Synthetic Membrane Processes, Chapter 1, Academic Press, Orlando, 1984. A. P. Mairal, A.R. Greenberg, W.B. Krantz, L.J. Bond' J. Membr. Sci. 159 (1999) 185-196. A. P. Mairal, A.R. Greenberg, W.B. Krantz, Desalination 130 (2000) 45-60. L.J. Bond, G.Y. Chai, A.R. Greenberg, W.B. Krantz, Method and Apparatus for Determining the State of Fouling/Cleaning of Membrane Modules, U.S. Patent 6,161,435, December 19, 2000. E. Fountonkidis, Z.B. Maroulis, D. Marinos-Kouris, Desalination 79 (1990) 47-63. M. Brusilovky, J. Borden, D. Hasson, Desalination 86 (1992) 187-222. A. Malik, F.A. Aleem, Desalination 96 (1994) 409-419. F.H. Butt, F. Rahman, U Baduruthanal, Desalination 109 (1997) 323-336. G. Y. Chai, A.R. Greenberg, W.B. Krantz, Euromembrane 99 2 (1999) 357-359. G. Y. Chai, W.B. Krantz, A.R. Greenberg, "Measurement of Membrane Fouling in Spiral-Wound Modules," 6th World Congress of Chemical Engineering, Melbourne, Australia, September 2001. A. P. Mairal, Ph.D. dissertation, University of Colorado at Boulder, 1998. K. Scott, Handbook of Industrial Membranes, 1st ed., Elsevier Advanced Technology, Oxford, UK, 1995. P.L. Mangonon, The Principles of Materials Selection for Engineering Design, Prentice Hall, Upper Saddle River, NJ, 1999.
88
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 5
Nonintrusive characterization of fluid transport phenomena in hollow-fiber membrane modules using MRI: An innovative experimental approach C. K. Poh 1, P. A. Hardy 2, Z. Liao 1, W. R. Clark 3, and D. G a o 1'2'4. 1Department of Mechanical Engineering, University of Kentucky, Lexington, KY, USA. 2Center for Biomedical Engineering, University of Kentucky, Lexington, KY, USA. 3Renal Division, Baxter Healthcare Corporation, McGaw Park, IL, USA. 4Department of Thermal Science and Energy Engineering, University of Science and Technology of China, Hefei, Anhui, China *Corresponding author, Phone: (859) 257-6336 ext. 80666, Fax: (859) 2573304, e-mail:
[email protected] ABSTRACT A nonintrusive experimental tool is useful for a better physical understanding of fluid transport phenomena in hollow-fiber membrane modules and as test cases for validating, improving, or developing numerical models. In this chapter, we introduced two innovative, nonintrusive flow-imaging techniques using magnetic resonance imaging (MRI) for characterizing fluid transport phenomena in hollow-fiber membrane modules. These flow-imaging techniques are called the 2-D Phase-Contrast (2DPC) and 2-D FourierTransform (2DFT) techniques. The principles, validation, advantages, limitations, and some examples of experimental results are presented. We used the 2DPC technique to study the spatial flow distribution and the 2DFT technique to characterize the flow profile and quantify the local ultrafiltration rates in hollow-fiber artificial kidneys (also known as hemodialyzers). These flow-imaging techniques are equally applicable to other hollow-fiber membrane modules. 1.
INTRODUCTION
Knowledge of experimental flow distribution in hollow-fiber membrane modules provides a valuable insight into the real fluid transport phenomena in these modules, which cannot be predicted by numerical models, and it may lead
89
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
to an improved engineering design of these modules. Any flow maldistribution such as shunt or occluded flow will reduce the separation performance of these modules. Ideally, a uniform flow distribution is preferred in both the shell and lumen sides of these modules for optimal separation performance. Nonintrusive flow visualization in hollow-fiber membrane modules is impossible by the use of conventional flow visualizing tools. Conventional flow visualizing tools are intrusive, where they require seeding particles or probes to be inserted into the flow, and have limited spatial resolution. The small inner diameter of the hollow fibers (200 to 500 ~tm in hemodialyzers, 100 to 400 ~tm in permeators, and 200 to 400 ~tm in contactors) and the presence of large number of hollow fibers (10,000 to 12,000 in hemodialyzers, 300 to 3,500 in permeators, and 900 to 12,500 in contactors) in a given module complicates the flow distribution and poses a great challenge in visualizing the flow in these modules. In most cases, small-scale countercurrent flow is realized in these modules due to the flow inside the hollow fibers (lumen-side flow) moves in the opposite direction to that outside the hollow fibers (shell-side flow). Conventional flow visualizing tools, with limited spatial resolution, are incapable of capturing this small-scale countercurrent flow. Recent advancement in MRI-based flow imaging technique has made it an attractive, nonintrusive tool for studying the fluid transport phenomena in hollow-fiber membrane modules. However, the current application of MRI in the study of fluid transport phenomena in hollow-fiber membrane modules remains minimal. Hammer et al. [1], Heath et al. [2], and Donoghue et al. [3] used MRI to investigate flow distribution in hollow-fiber bioreactors. Pangrle et al. [4], Zhang et al. [5], Laukemper-Ostendorf et al. [5], Osuga et al. [7], and Poh [8] used MRI to study flow distribution in hollow-fiber hemodialyzers. In this chapter, we introduced two innovative, nonintrusive flow-imaging techniques called the 2-D Phase-Contrast (2DPC) and 2-D Fourier-Transform (2DFT) techniques. We used the 2DPC technique to investigate the spatial flow distribution and the 2DFT technique to characterize the flow profile and quantify the local ultrafiltration rates in hollow-fiber hemodialyzers [8-14]. Some examples of these experimental results are presented in this chapter to illustrate the capabilities of these flow-imaging techniques. An extensive discussion on the principles of MRI is available in a number of books [ 15-18]; therefore, it will not be presented in this chapter. We will only discuss the commonly used pulse sequences in MRI. In addition, we will discuss the principles, advantages, limitations, validation, and applications of the 2DPC and 2DFT techniques in hollow-fiber membrane modules. Other MRI-based flow imaging techniques are also available, such as the Delays Alternating with Nutations for Tailored Excitation (DANTE) [19], Multiple Overlapping ThinSlab Acquisition (MOTSA) [20], Real-Time Acquisition and Velocity Evaluation (RACE) [21], and n Echo-Planar Imaging (PEPI) [22]. The 90
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
discussion of these flow-imaging techniques is beyond the scope of this chapter. In general, different pulse sequences of MRI can be designed or modified for different flow-imaging techniques. 2.
ADVANTAGES OF THE 2DPC AND 2DFT TECHNIQUES
The 2DPC and 2DFT techniques introduced in this chapter are nonintrusive or nondestructive, where they do not require any seeding particles or insertion of probes. Therefore, these flow-imaging techniques do not disturb the natural occurrence of the flow in hollow-fiber membrane modules. The nonintrusive nature of these flow-imaging techniques allow us to visualize and quantify the natural spatial and temporal behavior of the flow in hollow-fiber membrane modules that is otherwise impossible by means of conventional experimental tools such as hot-wire sensors. These flow-imaging techniques also do not require the object of interest to be transparent; hence, they are particularly advantageous for investigating fluid transport phenomena in hollowfiber membrane modules because these modules are usually opaque, where optical fluid visualizing tools cannot be used. Experimental data acquired from these flow-imaging techniques can be used as test cases to develop, validate, or improve realistic numerical models for use in predicting the separation performance of hollow-fiber membrane modules under different engineering designs and operating conditions. Countercurrent flow is commonly used in hollow-fiber membrane modules to optimize their separation performance. The 2DFT technique is capable of measuring simultaneously countercurrent flow in these modules that is otherwise impossible by means of other nonintrusive fluid-visualizing tools such as the Laser Doppler Velocimetry (LDV) and Particle Imaging Velocimetry (PIV). The LDV and PIV are incapable of measuring countercurrent flow in hollow-fiber membrane modules due to the presence of several thousands of hollow fibers with a relatively small inner diameter of about 100 to 500 ~tm. These hollow fibers allow countercurrent flow to occur simultaneously in a small region of about 250 by 250 btm2, where one fluid flows inside the lumen and the other fluid flows countercurrently outside the lumen of these hollow fibers. The LDV and PIV are incapable of even measuring unidirectional flow in such a small flow region or small-scale flow field [23]; therefore, they are incapable of measuring countercurrent flow in such a small-scale flow field. As in all areas of fluid mechanics, the nonintrusive 2DFT technique, capable of measuring countercurrent flow in such a small-scale flow field, is of great value for studying the flow distribution in hollow-fiber membrane modules.
91
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
3.
OVERVIEW OF STANDARD IMAGING PULSE SEQUENCES
The magnetic resonance signal emitted from a fluid of interest is a function of several imaging parameters such as the proton density, T1- and T2relaxation times, chemical shift, and fluid motion. The Tl-relaxation time is the time for the longitudinal magnetization to grow from zero to 63 % of its initial value, and the T2-relaxation time is the time for the transverse magnetization to decay to 37 % of its initial value. The relative contributions of each imaging parameter determine the final quality of the reconstructed MR images, and they are controlled by the radio frequency pulses, magnetic-field gradient pulses, and data acquisition timing. A pulse sequence is a set of radio frequency and magnetic-field gradient pulses used in conjunction with the data acquisition timing to produce the desired MR images. The two most commonly used imaging pulse sequences in the clinical MRI are the spin-echo and gradient-echo pulse sequences. The time for a complete cycle of a pulse sequence is called the repetition time, TR, and the time between the first maximum echo signal emitted is called the echo time, TE.
3.1.
Spin-Echo Pulse Sequence The standard spin-echo pulse is shown in Figure 1. The radio frequency pulses are denoted as RF, the slice-selective gradient pulses are denoted as Gstice,the phase-encoding gradient pulses are denoted as Gehase,and the frequency-encoding gradient pulses are denoted a s GReadin Figure 1. The magnetic resonance signal, an echo signal that emitted from the fluid of interest after the application of the 90-degree and 180-degree radio frequency pulses is shown on the last line in Figure 1. The frequency-encoding gradient is denoted a s aReadbecause it is frequently referred to as the read-out gradient. The spatial information in the MR images comes from the frequencyencoding gradient and phase-encoding gradient. By convention, the main static magnetic field in a MR imager is always aligned with z-axis. The frequencyencoding gradient is applied along the x-axis and is commonly denoted as Gx instead of GReta. The phase-encoding frequency is applied along the y-axis and is commonly denoted as Gy instead of Gpha~e. The frequency-encoding gradient creates a position-specific frequency distribution along the x-axis, and the phase-encoding gradient creates a positionspecific phase distribution along the y-axis. To create a MR image with Nx by Ny pixels, we need to sample Nx data points from the frequency-encoding gradient and to change the phase-encoding gradient Ny times with different gradient values. A MR image is then reconstructed from the recorded raw data, containing both position-specific frequency and position-specific phase distribution, through an inverse two-dimensional Fourier transformation where 92
Nonintrusive Characterization O f Fluid Transport P h e n o m e n a In Hollow-Fiber M e m b r a n e Modules Using MRI: A n Innovative Experimental A p p r o a c h - Poh
90"
180"
90' .................
I
GSlice I
I
GPhase
GRead 0 I
Signal
.................
½TE v
,_.. ~-~
I
'
TR
I
.--II
v I
I
DataSampling Figure 1. The standard spin-echo pulse sequence. Reprinted with permission from Poh [8]. the acquired signal is assigned to the appropriate pixel in the MR image. The acquired raw data set is commonly called the k-space. In a standard spin-echo pulse sequence as shown in Figure 1, the sliceselective gradient, together with the 90-degree radio frequency pulse, is used to select an image slice. The image slice thickness is controlled by the magnitude of the slice-selective gradient, and the bandwidth and shape of the radio frequency pulse. The steeper the slice-selective gradient or the smaller the bandwidth of the radio frequency pulse, the smaller the image slice thickness. The 90-degree radio frequency pulse shown in Figure 1 is also used to flip the net magnetization of nuclear spins into the transverse plane. Prior to the 180degree radio frequency pulse, a stepped phase-encoding gradient is applied to encode the spatial information along the y-axis. After the 90-degree radio frequency pulse, the transverse magnetization of nuclear spins started to decay
93
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
due to the observed transverse (T2*)relaxation. T2* relaxation is the dephasing of nuclear spins caused by the spin-spin interactions and magnetic field inhomogeneity. A 180-degree radio frequency pulse is then applied at exactly one-half TE. The 180-degree radio frequency pulse reverses the dephasing effect of nuclear spins, caused by the magnetic field inhomogeneity, by flipping them in a mirror-like fashion. The 90-degree and 180-degree radio frequency pulses are sometimes called the excitation and refocusing pulses, respectively. After the 180-degree radio frequency pulse, the rephased nuclear spins cause the transverse magnetization to grow to a maximum value, and then it decays again due to the T2 relaxation instead of T2* relaxation. The dephasing effect of nuclear spins due to magnetic field inhomogeneity is time independent, and that due to spin-spin interactions is time dependent. The 180-degree radio frequency pulse corrects the dephasing effect due only to magnetic field inhomogeneity, but not that due to spin-spin interaction. Therefore, after the 180-degree radio frequency pulse, the recovered transverse magnetization experiences T2 relaxation instead of T2* relaxation because the dephasing effect due to magnetic field inhomogeneity has been corrected by the 180-degree radio frequency pulse. The recovery of the transverse magnetization by a 180-degree radio frequency pulse is called a spin echo, and it generates an echo signal as shown in Figure 1. This echo signal is acquired in the presence of the frequencyencoding gradient. The frequency-encoding gradient provides spatial information along the x-axis. A similar frequency-encoding gradient is applied before the 180-degree radio frequency pulse to ensure that any stationary nuclear spins present are rephased during the spin echo. This process is repeated Ny times using different values of the phase-encoding gradient to acquire Ny data points along the y-axis. The frequency-encoding gradient lobe applied after the 180-degree radio frequency pulse has twice the area of the frequency-encoding gradient lobe applied before the 180-degree radio frequency pulse, allowing the full echo to be refocused. The advantage of the spin-echo pulse sequence is that it is independent of magnetic field inhomogeneity due to the use of a 180-degree refocusing radio frequency pulse. On the down side, however, the time TR in this pulse sequence is relatively long, typically 200 to 2000 ms. Therefore, the spin-echo pulse sequence generally requires a long image-acquisition time. Its TR is relatively long due to the use of a 180-degree radio frequency pulse and to the time required for the longitudinal magnetization to recover before the application of another 90-degree radio frequency pulse. The reason for waiting for the longitudinal magnetization to recover before the application of another 90degree radio frequency pulse is to make sure that sufficient amount of signal is emitted when subsequent 90-degree radio frequency pulses are applied.
94
Nonintrusive Characterization Of Fluid TransportPhenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
Alternative to the spin-echo pulse sequence is the gradient-echo pulse sequence that uses a smaller flip angle (< 90°), and this pulse sequence does not require a 180-degree refocusing radio-frequency pulse. 3.1.2. Gradient-Echo Pulse Sequence The standard gradient-echo pulse sequence is shown in Figure 2. In a gradient-echo pulse sequence, a radio frequency pulse with a smaller flip angle (< 90 °) is applied to flip the net magnetization of nuclear spins compared to that applied in the spin-echo pulse sequence. In addition, a bipolar frequencyencoding gradient pulse instead of a 180-degree radio frequency pulse is used to rephase or refocus the dephasing nuclear spins after the first radio-frequency pulse compared to that in the spin-echo pulse sequence. The gradient-echo pulse sequence is also known as the gradient-recalled pulse sequence. Initially, a radio frequency pulse is applied to flip the net magnetization of nuclear spins by an angle, a with respect to the axis of the applied static magnetic field, where a is usually between 0 and 90 °. This radio frequency is commonly denoted as an a pulse. Similar to the spin-echo pulse sequence, the image slice is selected using the slice-selective gradient together with the a pulse as shown in Figure 2. After the application of an a pulse, a stepped phase-encoding gradient is applied. The stepped gradient modulates the phase of the signal along the y-axis, and hence provides the spatial information along the y-axis. At the same time, a bipolar frequency-encoding gradient, as shown in Figure 2, is applied. The first negative lobe of the frequency-encoding gradient dephases the transverse magnetization of the nuclear spins caused by the a pulse. The second positive lobe of the frequency-encoding gradient then rephases the transverse magnetization and hence increases its magnitude. Subsequently, a gradient echo is produced at the time TE. The echo signal is acquired during the rephasing of the transverse magnetization. The frequency-encoding gradient also provides the spatial information along the x-axis. This process is repeated Ny times using different values of the phase-encoding gradient to acquired Ny data points along the y-axis. One of the advantages of the gradient-echo pulse sequence compared to the spin-echo sequence is its short image-acquisition time. With a smaller flip angle, longitudinal magnetization takes shorter time to recover compared to a 90 ° flip angle, and the pulse sequence requires a shorter TE and TR times. Therefore, the gradient-echo pulse sequence has a shorter total imageacquisition time compared to the spin-echo pulse sequence. On the other hand, the gradient-echo pulse sequence is sensitive to magnetic field inhomogeneity. The bipolar frequency-encoding gradient used in the gradient-echo pulse sequence does not reverses the dephasing effect of the magnetic field inhomogeneity as does the 180-degree refocusing radio95
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
G
RF-Gs,= I GPhase
GR~d
Signal TE TR I
Data Sampling
Figure 2. The standard gradient-echo pulse sequence. Reprinted with permission from Poh [8].
frequency pulse used in the spin-echo pulse sequence. Therefore, the signal generated from the gradient echo in the gradient-echo pulse sequence is weaker than that generated from the spin echo in the spin-echo pulse sequence. 4.
OVERVIEW OF FLOW IMAGING TECHNIQUES
Flow phenomena in MRI can be categorized into the phase shift phenomenon and time-of-flight (TOF) phenomenon [24]. The phase shift phenomenon is used in the 2-D Phase-Contrast (2DPC) technique; a combination of phase shift phenomenon and TOF phenomenon is used in the 2D Fourier-Transform (2DFT) technique. A full discussion on these flow phenomena and imaging techniques can be found in the references cited throughout the rest of this chapter.
96
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
4.1.
2-D Phase-Contrast Technique The 2-D Phase-Contrast (2DPC) technique incorporates flow-encoding gradients into a standard gradient-echo imaging sequence and makes use of flow-induced phase shift to encode velocity information of a constant flow. The precessional frequency, co of a nuclear spin in a static magnetic field is given by the Larmor equation:
(1)
co = yB o
where ~, is the gyromagnetic ratio of hydrogen atoms (42.57 MHz/Tesla), and Bo is the strength of the applied static magnetic field.
Suppose that the nuclear spin is moving in a spatially homogeneous magnetic field, Bo, and when a magnetic field gradient, G is introduced, the precessional frequency of the moving nuclear spin becomes positional dependent, and it is given by: co(,')= coo + 7G.r
(2)
where COo is the precessional frequency of the nuclear spin in Bo with the absence of the magnetic field gradient, G, ?, is the gyromagnetic ratio, and r is the positional vector of the nuclear spin. The change of phase angle or phase shift, ~ with respect to the time of the moving nuclear spin relative to its initial position at r = 0 can be defined by rearranging Eq. (2): d¢
dt
= co(*')-coo =7G'r
(3)
Integrating Eq. (3) with respect to time, t, we have: ¢(t)= Y ~G(t).r(t)dt
(4)
where ~ (t) is the phase shift of the nuclear spin as a function of time relative to its initial position at r = 0. In addition, the positional vector of the nuclear spin, r can be expressed in terms of time derivatives of its value at t = 0 using the Taylor series expansion:
97
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
"(') = "o +
,=o t + - =L J,=o
= ro + Vot + laot2
+"" + "!l or" Jt
=o
(5)
+ ...
where to, Vo, and ao are the position, velocity, and acceleration vectors of the nuclear spin, respectively, at an initial time, t = 0. Substituting Eq. (5) into Eq. (4), we have: 1
m • j'c(t)a, +,,o. j'c(t),a, + r,o.
+...
(6)
Alternatively, Eq. (6) can be written in terms of moments of the magnetic field gradient as:
~b(t)= y M
1 Z
o . r o + 7"M 1 . v o + -~ y M 2 . a o +
o oil
(7) and
M i - ~G(t)tidt
where M o , M 1 , and M 2 are the zeroth, first, and second moments of the magnetic field gradient, respectively. The higher order terms (+...) in Eq. (7) are usually very small and are negligible. The magnetic-field gradient moments in Eq. (7) can be manipulated accordingly during the experiment by choosing the proper shape of the magnetic-field gradient waveform. For instance, the zeroth moment of the magnetic field gradient can be set to zero by choosing a magnetic field gradient that has two opposite gradient lobes with equal areas as a function of time, called the bipolar gradient [30]. In theory, any moments can be set to zero, and we can generate either velocity images by setting the zeroth moment of the magnetic field gradient to zero or acceleration images by setting the zeroth and first moments of the magnetic field gradient to zero. For a uniform flow i.e. without acceleration, Eq. (8) can be rewritten as: (9)
~(t)-- ~ M 0 "1"o + ~M1 "Vo
911
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
My G(t)
g Spin
I'
~
Phase
1st GradientPulse
1
Stationary'7 Mx
2nd GradientPulse
¢~sl
9
Position
¢~s2!
StationarySpin
MovingSpin
Figure 3. The effect of a bipolar gradient pulse on the phase shifts of stationary and moving spins. Reprinted with permission from Poh [8].
The first term on the fight-hand side of Eq. (9) is proportional to the position and area under the gradient waveform and is called the positional encoding term. The second term on the right-hand side of Eq. (9) is proportional to the velocity and is called the velocity encoding term. In the 2DPC technique, to measure the axial velocity of the flow, in this case in the z-direction, we choose a bipolar magnetic-field gradient pulse, or sometimes called a bipolar gradient pulse, Gz(t) that has two opposite gradient lobes with equal areas as shown in Figure 3 such that its zeroth moment is zero (positional independent), i.e. the net phase shift is independent of any stationary nuclear spins. As shown in Figure 3, the bipolar gradient pulse has no effect on the phase shifts induced by the stationary nuclear spin because the these phase shifts at the first and second gradient lobes of the bipolar gradient pulse cancel out each other (¢s2-¢s~ =0). Nevertheless, the phase shifts induced by the moving nuclear spin at the first and second gradient lobes of the bipolar gradient pulse do not cancel out each other (¢M2-¢M~ ~ 0), and the net phase shift, ev, is 99
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Experimental Approach - Poh
a
RF
Using
MRI:
An
Innovative
a
i
~q~
~~(?:~.....
]") FlowCompensated
i
. . . . ~1 ............
,,
GS,~,tF~
. . . . . . . . . . . . . .
RF
FlowEncoded __1 ............
.............
i i i
rE i
Figure 4. The 2-D Phase-Contrast (2DPC) pulse sequence. The radio frequency pulses are denoted as RF, and the flow-encoding gradient pulses are denoted as Gsl~ce/Flow. The phase-encoding gradient pulses Geh~se and the frequency-encoding gradient pulses Ggead are not shown in this figure. An echo signal (not shown) is generated at the time TE and a is usually between 0 and 90 °. Reprinted with permission from Poh [8].
proportional to the velocity of the moving nuclear spin, where #v =¢M2--~MI" Therefore, by using a bipolar gradient pulse that has two opposite gradient lobes with equal areas, the moving nuclear spin has a net phase shift proportional to its velocity, and this net phase shift is independent of any stationary spins that may have been present in the system. The net phase shift due to the moving nuclear spin in the z-direction, ~,.=(t) is: ~v,z (t) -- y M 1v z + (some error source terms)
100
(10)
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
The phase shift, #v.z(t) is now velocity dependent and positional independent. This phase shift is sometimes called the flow-induced phase shift. However, this flow-induced phase shift is associated with some error sources or background phase variations such as magnetic field inhomogeneity, imperfections in pulse sequence tuning, and eddy-current effect. The eddy-current effect will be discussed in the latter part of this chapter. Two data sets with different flow sensitivities in an interleaved fashion are acquired during a single experiment to eliminate the background phase variations. The imaging pulse sequence for the 2DPC technique is shown in Figure 4. The first data set acquired is flow compensated [25-28], i.e. flow desensitized, where the first moment of the magnetic-field gradient pulse is set to zero; and the second data set is flow encoded, i.e. flow sensitized, to record the axial velocity (z-direction) of the flow. Flow compensation in the first data set is also known moment nulling. The flow sensitivity in the second data set is controlled by the strength or amplitude of the magnetic-field gradient pulse. Since both data sets have an equal background phase variations for a given experiment, the first data set is used to correct for the background phase variations in the second data set by subtracting the first data set from the second data set. Therefore, the net phase shift, in radian, is given by:
(11)
A ~v,z -- y'AI~l Vz
where the net phase shift, aq~v.,is proportional to the velocity in the z-direction and is independent of background phase variations. AM1 is the difference in the first moment of the z-direction gradient pulses. An important parameter in the 2DPC technique is the encoding velocity, Ve,c. The encoding velocity defines the dynamic range of velocities measurable by the imaging technique. It is controlled by the difference in the first moment of the gradient pulses, zJM1 and is defined as:
Venc
7,~1
(12)
By combining Eq. (11) and Eq. (12), and rearranging; we have the measurable velocity, Vz given by :
(13)
vz=A~v,z venc
101
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
The net phase shift, a#v,z can only take values between -zr to +zr for countercurrent flows, and as a result, the measurable velocity, Vz is between - Venc to + Ve~c. The encoding velocity, Ve#c has to be determined a priori, so that the measurable velocity, Vz is within the velocity range of the flow of interest. Phase shift aliasing will occur when the measured net phase shift exceeds the limits of -n: to +zr (i.e. the measured velocity exceeds the range of the encoding velocity) resulting in a potential ambiguity in determining the flow velocity. The measure net phase shift is then converted into axial velocity using Eq. (13) with a predetermined encoding velocity. The precision of the velocity determined from the measured net phase shift depends on the noise in the phase shift measurement. The standard deviation of the velocity determined from the measured net phase shift, trv is given by:
Ve c
(14)
SNR
where S N R is the signal-to-noise ratio in the acquired image. Eq. (14) shows that the noise or standard deviation of the velocity can be reduced by increasing the S N R or by decreasing the Venc. 4.1.2. 2-D Fourier-Transform Technique The 2-D Fourier-Transform (2DFT) technique, on the other hand, incorporates flow-encoding gradients into a standard spin-echo imaging sequence. This imaging technique makes use of the combined effects of time-offlight (TOF) and phase-shift imaging techniques, proposed by Moran [29], called the flow-velocity zeugmatographic imaging. In addition to the incorporation of flow-velocity zeugmatographic imaging technique into a standard spin-echo pulse sequence as introduced in this chapter, the flowvelocity zeugmatographic imaging technique can be incorporated into any MR imaging pulse sequences without interfering with the spatial encoding and reconstruction algorithm of the imaging technique [29]. Basic understanding of the TOF phenomenon is required for the 2DFT technique. Reviews on TOF phenomenon can be found in [24,30-34]. In a standard spin-echo pulse sequence, TOF phenomenon demonstrates two flow effects: the inflow and outflow effects. In this 2DFT technique, we make use of the inflow effect from the TOF phenomenon. Therefore, only the inflow effect will be discussed in this chapter. Inflow effect refers to a moving of fresh or unexcited nuclear spins through an imaging slice perpendicular to the flow direction. Since these fresh 102
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
nuclear spins have not experienced any radio frequency pulses before, they emit a stronger signal when they are first excited by a radio frequency pulse in the imaging slice than any surrounding saturated stationary nuclear spins. The stationary nuclear spins within the imaging slice have experienced several excitation pulses, and the TR time is too short for their longitudinal magnetization to recover. Therefore, these stationary nuclear spins are either partially or fully saturated, and they emit either very little or no signal at all during the appearance of the spin echo. The inflow effect is dependent on the timing of the radio frequency pulse (TR and TE) and the velocity of the flow. In a given imaging slice, for moving nuclear spins to emit a maximum signal during the spin echo or data acquisition, these moving nuclear spins must experience both the 90-degree and 180-degree pulses in a standard spin-echo pulse sequence and move perpendicular to the imaging slice. Any nuclear spins that move through the imaging slice in the time less ½ TE will not be excited by the 180-degree pulse and hence will not emit any signals. The maximum velocity allowable, Vmaxfor a moving nuclear spin to experience both the 90-degree and 180-degree pulses for a standard spin-echo pulse sequence is given by: V~a~ = Z~Slice TR
(15)
where Zst~ceis the thickness of the image slice, and TR is the repetition time. In general, the signal emitted from the moving nuclear spins increases linearly with increasing velocity [24], provided the velocity is less than Vmax.For nuclear spins that move faster than Vm~x,a significant fraction of the moving nuclear spins may have experienced the 180-degree pulse before the 90-degree pulse. Since these fast moving nuclear spins have no transverse magnetization, they do not contribute to the spin echo, and the signal will eventually decreases as velocity increases beyond Vmax. A detailed analysis by Wehrli [24] shows that four different populations of moving nuclear spins can coexist in an imaging slice of a standard spin-echo pulse sequence. The moving nuclear-spin populations are fresh nuclear spins, fully relaxed nuclear spins, inverted nuclear spins, and saturated nuclear spins as shown in Figure 5. Assuming a plug flow with velocity, v in a circular pipe and taking an imaging slice with slice thickness, Zst~ce as shown in Figure 5. A 90degree pulse is applied at the time, t = 0, and a bolus of fluid is tracked with increase in time. At the time t = ½ TE, the 180-degree pulse is applied, followed by another 90-degree pulse at the time, t = TR. Finally, at the t = TR + ½ TE, that is at the time the 180-degree pulse is applied after the application of the second 90-degree pulse, the bolus of fluid has traveled a total distance of v(TR + ½ TE), 103
Nonintrusiv¢ Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using M R I : Experimental Approach - Poh
II t t !
I I I
!
t
An Innovative
t
!i ................... , v(rm) t
i
t
v(m)
i
v(TR * 'A TE}
Figure 5. Four different populations of nuclear spins in an imaging slice of a plug flow after a 90°-180°-90°-180 * spin-echo pulse sequence cycle. Region 1 represents fresh nuclear spins, region 2 represents fully relaxed nuclear spins, region 3 represents inverted nuclear spins, and region 4 represents saturated nuclear spins. Zslice is the imaging slice thickness, TR is the repetition time, TE is the echo time, and v is the flow velocity. Reprinted with permission from Poh [8].
and fresh nuclear spins have entered into the imaging slice. In addition, the signal intensities from the moving nuclear-spin populations that emit signal during the spin echo can be determined from the Bloch equations [24]: S Relaxed = Mt=oe (-TE//T2)
Slnverted = Mt= 0 1 - 2e k
T1
.I e(-TE//T2)
(16)
Ssaturate d = M t = 0 1 -
2e L
--
+e(_rR//n) e(-rE//~2)
104
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
where SRelaxed, Slnverted, and Ssaturated a r e the signal intensities for the fully relaxed, inverted, and saturated nuclear spins respectively, M ~ is the magnetization of nuclear spins at time, t = 0, TE is the echo time, and TR is the repetition time. In the 2DFT technique, the velocity information is encoded by a stepped flow-encoding gradient as shown in Figure 6. A 90-degree radio frequency pulse is initially applied followed by a 180-degree radio frequency pulse at ½ TE. An echo signal is generated after the 180-degree radio frequency pulse. TE is the time between the first 90-degree radio frequency pulse and the maximum echo signal. TR is the time between successive 90-degree radio frequency pulses. ~ is the period of the stepped flow-encoding gradient pulses. During data acquisition, unlike in the 2DPC technique where the flowencoding gradient pulse is constant, the flow-encoding gradient in the 2DFT technique is stepped N times, and the resulting velocity data is Fourier transformed into N velocity images, each with a specific velocity bin size, Av. As the name of the imaging technique implied, these N velocity images are the Fourier-velocity images of the flow, and each Fourier-velocity image covers a specific velocity range or bin size. The velocity bin size, Av of the Fouriervelocity images reconstructed from the 2DFT technique is controlled by the thickness of the image slice, number of reconstructed Fourier-velocity images, and period of the stepped flow-encoding gradient pulses as shown by the following equation:
A v = Z Slice
(17)
Nr
where Zst~ is the thickness of the image slice, N is the number of reconstructed Fourier-velocity images, and r is the period of the stepped flow-encoding gradient pulses. In contrast to the 2DPC technique, which encodes velocity in the phase images, the 2DFT technique encodes velocity in the reconstructed Fourier-velocity images as magnitude images. The 2DFT technique is less sensitive to phase dispersion within a single pixel compared to the 2DPC technique and hence is more accurate in measuring less steady flow. Phase dispersion within a single pixel is due to the presence of a velocity distribution in a single pixel. The 2DPC technique, on the other hand, is extremely sensitive to phase dispersion within a single pixel because this flow imaging technique makes use of flow-induced phase shift to encode velocity information. On the contrary, the 2DFT technique requires a longer total imageacquisition time compared to the 2DPC technique due to the requirement of multiple flow-encoding steps in the 2DFT technique to reconstruct multiple Fourier-velocity images. 105
Nonintrusive Characterization Of Fluid Transport Phenomena E x p e r i m e n t a l A p p r o a c h - Poh
90" |
In Hollow-Fiber Membrane
M o d u l e s Using
180"
MRI:
An
Innovative
90" !
RF |
N
! ! ! e
Gs~lo~
i ! i
t
' i ! ! !
Gp,~
i I
G~
I
:
i
I | ! i ! i
Signal '
'ATE
:i
!
'
| i
rE !
i
i !
' t
,,el
.................
i i
TR
I
I
Data Sampling F i g u r e 6. T h e 2 - D Fourier-Transform (2DFT) pulse sequence. The radio frequency pulses are shown as RF, the flow-encoding gradient pulses are shown as Gsti,/~'lo,,,,, the phase-encoding gradient pulses are shown as Gehase, and the frequency-encoding gradient pulses are shown as GR~aa.Reprinted with permission from Poh [8].
e
VALIDATION OF THE 2DPC AND 2DFT TECHNIQUES
We validated the accuracy of the 2DPC and 2DFT techniques using a flow phantom or test flow in a Siemens 1.5T MAGNETOM Vision whole-body MR imager (Siemens AG, Erlangen, Germany). The MR imager is identical to that used for human imaging. It has a maximum gradient strength of 25 mT/m and a rise time of 600 ~ts. All image processing was performed using userwritten MATLAB programs together with the MATLAB Image Processing Toolbox 2.2 software (The MathWorks Inc., Natick, MA, USA) rtmning on a Linux workstation. 106
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
ElevatedReservoir ~ ) ~ " ~ Pump ,I i i
I
Rotameter
' ImagingPlane
PlexiglasTubes
Figure 7. The experimental setup of a flow phantom. Reprinted with permission from Poh [8].
5.1.
Experimental Methods The flow phantom was made of countercurrent laminar flows in two identical Plexiglas tubes at a steady flow rate as shown in Figure 7. The Plexiglas tubes were 1.8 m in length with inner and outer diameters of 1.6 and 1.9 cm, respectively. These Plexiglas tubes were positioned parallel to each other and placed along the axis of the main magnetic field inside the bore of the MR imager. A solution made of distilled water doped with cupric sulfate pentahydrate, CuSO4"5H20 (ACS Reagent, Sigma Chemical Co., St. Louis, MO, USA), with a final concentration of three millimolar (mM), was gravimetrically fed to one of the tubes from an elevated constant-head reservoir. A calibrated rotameter (Cole-Parmer Instrument Co., Vernon Hills, IL, USA) was used to control the desired flow rate. The flow out of the tube was channeled back to the other tube to create a countercurrent flow. The purpose of adding cupric sulfate pentahydrate to the distilled water was to increase the signal-to-noise ratio and to reduce the total image-acquisition time of the experiment. A peristaltic pump (Cardiovascular Equipment, Sams Inc., Ann Arbor, Michigan, USA) was then used to recycle the flow out of the flow phantom into the elevated constant-head reservoir. The flow rate used in validating the accuracy of the flow-imaging techniques was 100 mL/min as measured from the calibrated rotameter. This flow rate corresponds to a Reynolds number of 135 and a mean velocity of 0.9 cm/s, assuming a laminar pipe flow, in each Plexiglas tube.
107
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
All imaging was performed, at the mid-length cross sections, perpendicular to the flow direction, of the Plexiglas tubes (Figure 7). Long Plexiglas tubes (1.8 m) were used to ensure that fully developed laminar, countercurrent flows occur at the mid-length of these tubes. In addition, two test tubes filled with the similar prepared solution were placed beside the Plexiglas tubes at the imaging plane. These test tubes served as a reference for stationary flow. All velocities were measured in the axial direction; the direction along the length of the Plexiglas tubes. The flow phantom was first imaged using the 2DPC technique with the maximum velocity encoded at 10 cm/s followed by the 2DFT technique. Other imaging parameters for the 2DPC technique were TR = 45 ms, TE = 12 ms, a = 15 °, and slice thickness = 4 mm. TR is the repetition time, TE is the echo time, and a is the flip angle. The total image-acquisition time was approximately one minute. All phase images acquired had a matrix dimension of 512 x 512 with a field-of-view (FOV) of 300 mm. For the 2DFT technique, twenty-four Fouriervelocity images with velocities ranging from -2.5 to +2.3 cm/s, each with a velocity bin size of 0.2 cm/s, were acquired. Other imaging parameters for the 2DFT technique were TR = 200 ms, TE = 111 ms, r = 100 ms, and slice thickness = 4.8 mm. The total image-acquisition time was approximately ten minutes. All Fourier-velocity images acquired had a matrix dimension of 64 x 256 with a FOV of 50 x 100 mm.
5.1.2. Experimental Results We defined the flow moving into the MR imager as the positive flow and that moving out of the MR images as the negative flow.
Figure 8. A raw phase image of the flow phantom. Reprinted with permission from Poh [8]. 108
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using Experimental Approach - Poh
Flow Rate = 100 mL/min
MRI:
An
Innovative
Re = 135
1.5 ........... ;.................. ...................................
i
i ..... : .....
.... "
1 ...........
U
"6 O >
0.5
i
i
i
..,-;.: .......
....... '".............
""
o.s "~
-. i. . ....... . ......
. J
~
............... i
0
.... .,. ........... !
0
4." ........
._~ o0
>
-0.5
-0.5 ~
-1
!;~iiilziiii:i~i~ili ~ii~,iiiii;~ii
~ii~iiii!iii!!~;i
i 150
~
0
0
......:-~'--'--" SO
Y-Position In Space
150 100
II
-1.5
X-Position In Space
Figure 9. 3-D representation of velocity profile acquired from the mid-length of the flow phantom. Reprinted with permission from Poh [8].
5.2.2. 2DPC Technique Figure 8 shows the raw phase image acquired by the 2DPC technique from the flow phantom. The image artifact was removed during image processing, and the image processing methodology has been discussed in Poh [8]. The processed phase image was then converted into a velocity image using Eq. (13). A 3-D representation of the velocity profile (Figure 9) acquired from the mid-length of the flow phantom shows that the countercurrent flows had the characteristics of a fully developed laminar pipe flow, and the maximum positive and negative velocities were approximately equal in magnitude. These countercurrent flows were fully developed laminar flows because at these cross sections of the tubes, the ratio of the tubes' length to their inner diameters (L/d) was 56. Generally, a laminar pipe flow is fully developed when the L/d ratio [35] is greater than 0.06Re, Re being the Reynolds number determined from the inner diameter of the pipe. In our flow phantom, its 0.06Re had a value of eight,
109
NonintrusiveCharacterizationOf FluidTransportPhenomenaIn Hollow-FiberMembraneModulesUsingMRI:An Innovative ExperimentalApproach- Poh T a b l e 1. Comparison of experimental data measured from the 2DPC technique and rotameter. Reprinted with permission from Poh [8] Positive-Flow Tube in a Flow Phantom
Flow Rate (mL/min) Maximum Velocity a (cm/s)
2DPC
Rotameter
% Relative Error b
97.27
100.2
-2.9
1.65
1.8
-8.3
Negative-Flow Tube in a Row Phantom
Flow Rate (mLJmin) Maximum Velocity a (cm/s)
2DPC
Rotameter
% Relative Error b
94.52
100.2
-5.7
1.68
1.8
-6.7
a Calculated from the Poiseuille's equation based on the flow rate measured by the calibrated rotameter. b Determined from the following equation" % Relative Error =
2DPC Data - Rotameter Data
x 100 %
Rotameter Data
which was much less than its L/d ratio; therefore, the countercurrent flows at the mid-length of the flow phantom were fully developed laminar flows. In addition to the parabolic velocity profile of the flow phantom, the velocities measured from the 2DPC technique were in good agreement with the analytical velocities determined from the Poiseuille's equation based on the flow rate measured from the calibrated rotameter (Table 1). To further validate the accuracy of the 2DPC technique, we quantified the flow rates and determined the flow areas in each tube from the velocity image. The flow rates in each tube quantified from the 2DPC technique were calculated by multiplying the area of a single pixel with the sum of all velocities in each tube, and they were in good agreement with that measured from the calibrated rotameter (Table 1). The flow areas in each tube, on the other hand, were determined from the velocity image by multiplying the area of a single pixel with the sum of all pixels in each tube; they were 2.04 and 1.96 cm 2 in the tubes with positive and negative flows, respectively, compared to the actual flow area of each tube, 1.96 c m 2. Therefore, the accuracy of the 2DPC technique, with the imaging parameters used in this study, has a six to eight percent relative error in the measured velocity, a three to six percent relative error in the quantified flow rate, and a four percent relative area in the measured flow area of the flow phantom. The accuracy of these measurements may be further improved by increasing the spatial resolution of the velocity image using a smaller pixel size
110
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
F l o w R a t e = 100 m L / m i n
R e = 135
O ..........
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
,.,., .
.
.
.
.
i .io............, .....ii ,
Oi
iiil,i~i4i i,!~i~i~!:ii.i!~i~ !i~:
O
e ~:~,,
12 mm/s
= + 14.:~
Figure 10. Sixteen of the twenty-four raw Fourier-velocity images of countercurrent flows acquired from the mid-length of the flow phantom. Reprinted with permission from Poh [8]. (the pixel size used in this study was 0.34 mm2). The readers should take note, however, this accuracy does not necessary represent the accuracy in measuring flow in hollow-fiber membrane modules, which depends upon the inner diameter of the hollow-fiber membrane and the spatial resolution of the MR images. 5.2.1.2DFT Technique We acquired twenty-four Fourier-velocity images, each with a velocity bin size of two mm/s, across the mid-length of the flow phantom. Figure 10 shows sixteen of the twenty-four raw Fourier-velocity images acquired. The image artifact was caused by the Gibbs ringing [36-38] and noise [39,40] generated from the receiver coil during image acquisition and was removed during image processing [8]. The percent flow rates in each velocity bin relative to the total flow rate in each tube determined from these Fourier-velocity images were in good agreement with the analytical values derived from the Poiseuille's equation for a fully developed laminar pipe flow for mean velocities ranging from 2 to 14 mm/s (Figure 11). For a fully developed laminar pipe flow, this percent flow rate relative to the total flow rate is predicted from the Poiseuille's 111
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
FI0w Rate = 100 mLJmin
Re = 135
25 2DFT Data (Positive Flow) 2DFT Data (Negative Data) Analytical Data
20-
i]
;)i i
(1)
rr"
15
\ \ i:~:i \
o LL ¢..
\ i:i:i
a~ 10 o !,_
fl_
\
0
!
!
i
!
!
2
4
6
8
10
12
!
i
!
14
16
18
20
Mean Velocity (mm/s)
Figure 11. Comparison of the percent flow rate as a function of mean velocity at the mid-length of the flow phantom. Reprinted with permission from Poh [8].
equation to increase linearly with increasing mean velocity (Appendix). The discrepancies between the experimental and analytical values at the mean velocities of 16 and 18 mm/s may have been caused by experimental errors such as inhomogeneity in sensitivity profile of the radio frequency coil, finite velocity resolution in the Fourier-velocity images (each image covers a specific velocity bin size, and the velocity resolution decreases with increasing velocity bin size), Gibbs artifact [36-38], noise [39,40] generated from the receiver coil, and variations in flow velocity during the image acquisition period. @
EXAMPLES OF EXPERIMENTAL RESULTS FROM HOLLOWFIBER HEMODIALYZERS
All experiments were performed on a Siemens 1.5T MAGNETOM Vision whole-body MR imager (Siemens AG, Erlangen, Germany) with a maximum gradient strength of 25 mT/m and a rise time of 600 ItS. All image processing was performed using user-written MATLAB programs together with the MATLAB Image Processing Toolbox 2.2 software (The MathWorks Inc., Natick, MA, USA) running on a Linux workstation.
112
Nonintrusive Characterization Of Fluid Transport PhenomenaIn Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
POLYAMIDE HEMODIALYZER Flow Rate: 400 mL/min
Blood Compartment
..
o ?,-
.
150 0 Y-Po~
in Space
X - P o ~ n in Space
Figure 12. Blood-side velocity distribution at the middle cross section of a polyamide hemodialyzer. Reprinted with permission from Poh [8].
We characterized the flow distribution in the blood compartment of hemodialyzers by introducing a three-millimolar cupric sulfate solution, prepared from distilled water, through the blood compartment using a nonpulsatile pump, while the dialysate compartment was filled with the same solution and sealed. Similarly, we characterized the flow distribution in the dialysate compartment by introducing the solution through the dialysate compartment, while the blood compartment was filled with the solution and sealed. In addition, we characterized the flow profile in each compartment and quantified the local ultrafiltration rates by introducing the solution through the blood compartment; we recycled the outflow from the blood compartment to the dialysate compartment countercurrently. The flow rates were between 200 to 1000 mL/min, and five imaging planes were acquired from different transverse cross sections of each hemodialyzer. Detailed experimental methods and image processing methodology have been reported in Poh [8]. Figures 12 and 13 show the velocity distributions in the blood and dialysate compartments of a polyamide hemodialyzer. These velocity images were acquired from the middle cross section (mid-length) of the hemodialyzer using the 2DPC technique with nominal blood and dialysate flow rates of 400 and 600 mL/min, respectively. The polyamide hemodialyzer contains 10,000 hollow fibers, each with an inner diameter of 215 ~m and a wall thickness of 50 113
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
POLYAMIDE
HEMODIALYZER
Dialysate Compartment
Flow Rate: 600 mL/min
....... •.
...; ..-:"i;~i!~.' ~'
i!, '
"
'
"
:
.
.
.
4
"}i. .
.
2 o
._~
1
0
150 100
50
> '~
150 0 Y-Position in Space
X-Positionin Space
t/" Figure 13. Dialysate-side velocity distribution at the middle cross section of a polyamide hemodialyzer. Reprinted with permission from Poh [8].
~tm. It has an ultrafiltration coefficient of 71 mL/Hr/mmHg and a total membrane surface area of 1.7 m 2. Figures 12 and 13 show that the flow distribution in the blood compartment of the polyamide hemodialyzer was uniform and that in the dialysate compartment was nonuniform. Figure 14 shows the dialysate-side axial velocity distributions acquired from the imaging planes right across the dialysate ports of the polyamide hemodialyzer at nominal flow rates between 200 and 1000 mL/min. Figure 14 shows that the axial dialysate-side flow near to the dialysate ports moved preferentially at the top region of the cross section of the hemodialyzer. Figure 15 shows that the percent flow rate increased linearly with increasing mean velocity at the mid-length of a cellulose triacetate hemodialyzer, indicating that the flow in the blood compartment of the hemodialyzer had a fully developed laminar flow profile (Appendix). On the other hand, the flow in the dialysate compartment of the hemodialyzer at the same imaging plane was slightly skewed towards higher mean velocities (positive skewness in mean velocities), indicating that slightly more flow moved at lower mean velocities than at higher mean velocities. The data in Figure 15 were obtained from the Fourier-velocity images reconstructed by the 2DFT technique at a nominal flow rate of 280 mL/min. The cellulose triacetate hemodialyzer contains 12,000 hollow fibers, each with an inner diameter of 200
114
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
CELLULOSE TRIACETATE HEMODIALYZER Dialysate C o m p a r t m e n t
Dialysate In
Dialysate Out
200 mLtrniN
400 mLtmir~
600 mLlmir;
800 mL!min
t000 mLimin
~i~#iii~i!iiii~i~i~i~i~ii~i~i~i~i~i~i~!i~!iii!!!i !i!~i!ii!?__ 0
5
10 Axial Velocity in cml$
15
20
Figure 14. Dialysate-side axial velocity distributions acquired from the imaging planes
right across the dialysate ports of a polyamide hemodialyzer at nominal flow rates between 200 and 1000 mL/min. Reprinted with permission from Poh [8]. lam and a wall thickness of 15 pm. It has an ultrafiltration coefficient of 36 mL/hr/mmHg and a total membrane surface area of 1.9 m 2. Figure 16 shows the local maximum velocities and ultrafiltration rates quantified from a polyethersulfone hemodialyzer using the 2DFT technique at nominal flow rates of 300 and 600 mL/min, respectively. The local maximum velocities and ultrafiltration rates decreased monotonically along the length of the hemodialyzer, indicating that the transmembrane pressure decreased along the length of the hemodialyzer. The polyethersulfone hemodialyzer contains 11,200 hollow fibers, each with an inner diameter of 200 ~tm and a wall thickness of 30 ~tm. It has an ultrafiltration coefficient of 84 mL/hr/mmHg and a total membrane surface area of 1.9 m 2. These experimental results served as an illustration of the capabilities of the 2DPC and 2DFT techniques. Detailed experimental results and discussions have been published elsewhere [8-14].
115
N o n i n t r u s i v e Characterization O f F l u i d Transport P h e n o m e n a In H o l l o w - F i b e r Membrane Modules U s i n g E x p e r i m e n t a l A p p r o a c h - Poh
Flow Rate = 280 mL/min 20
MRI: An Innovative
Re(single hollow fiber) = 2
,
,
,
,
,
Hemodialyze
Cellulose Triacetate
0
0 '
4
8
12
16 20 Mean Velocity (minis)
24
'
20
'
:
:
28
'
.
32
'
;
I"
I Cellulose Triacetate Hemodialyzer N 15
I1
................... ~..................... i ..................... !..................... :.................... D i a l y s a t e C o m p a r t m e n t
°
•
!
rrlO
.
"E
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
!
::
:
4
8
12
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
5
o
0
16 20 Mean Velocity (ram/s)
24
28
32
Figure 15. Percent flow as a function of mean velocities at the mid-lengthof a cellulose triacetate hemodialyzer. Reprinted with permission from Poh [8]. 7.
LIMITATIONS OF THE 2DPC AND 2DFT TECHNIQUES
Like all other experimental tools, these flow-imaging techniques have their own limitations. Both flow-imaging techniques are only sensitive to flow moving in the flow-encoding direction. In the examples of experimental results presented earlier, the flow-encoding direction was in the axial direction of the hemodialyzer. In reality, the flow in the dialysate compartment of a hemodialyzer moves in all other directions in addition to the axial direction due to the complex flow pathway created by the presence of several thousands of hollow fibers in the hemodialyzer. Moreover, these flow-imaging techniques are only applicable to flow with steady velocity due to their long acquisition times (approximately one minute in the 2DPC and ten minutes in the 2DFT). Any changes in flow velocity during the acquisition period can lead to artifacts in the acquired MR images and subsequently can lead to inaccuracy in the recorded velocities. However, these flow-imaging techniques can be modified by using rapid imaging techniques to account for pulsatile flow seen in all hemodialysis therapies.
116
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
120
100
-r-
UltrafUtration Rate I - - - 3 Ultrafiltration Rate ~ Maximum Velocity -0Maximum Velocity
-r-
|
t°~
E E
80
r'F
60
I I !'
at Q=300 at Q=600 at e=300 at Q=600
mL/min mL/min mL/min mL/min
(~
5 ~" E 4 o,~
v
I
"g ~_.
40
!
20 -
~i~;
0
.
. 0.00 (Blood Inlet)
o
l
3 2
.~
1
. 0.25
.
. 0.50
0.75
0 1.00 (Blood Outlet)
Normalized Distance Along the Length of Hemodialyzer
Figure 16.
Local blood-side maximum velocities and ultrafiltration rates along the length of a polyethersulfone hemodialyzer at nominal flow rates of 300 and 600 mL/min. Reprinted with modification from Hardy et al. [14].
It is also worthwhile pointing out that the 2DPC technique is only capable of measuring unidirectional flow in a hollow-fiber membrane module. Any countercurrent velocities contained in a single pixel are averaged over the whole pixel. Moreover, if the flow is oblique instead of perpendicular to the imaging plane, the velocity can be underestimated [34] by a factor of cos(B), where fl is the oblique angle between the flow direction and the imaging plane. For a small r, the error caused by the obliqueness of flow is small since cos(r) is approximately one for a small ft. If fl is known and is smaller than 45 °, the underestimated velocity can be corrected by dividing it by cos(r). In addition to the obliqueness of flow, eddy-current [41 ] induced by the changes in gradient waveforms during data acquisition from the 2DPC technique causes some bias in the velocity determined from the measured phase shift. The eddy current causes a small variation in the phase shift and hence a small variation in the velocity. The eddy-current effect is often seen as a variation in phase shift in the phase images with stationary flow. The eddycurrent effect can be corrected after image acquisition by examining the variation in phase shift of a stationary flow, and it has been corrected in the
117
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
experimental results shown in this chapter. The correction for the eddy-current effect has been discussed in Walker et al. [41] and Poh [8]. In the 2DPC technique, the measured velocities are also distorted by the partial volume effect [42-45]. The measured velocity in each pixel is determined by the phase of the resultant magnetization, which is affected by all MR-visible protons within the voxel (volume element). In the case of the hemodialyzers, each voxel contains a mixture of protons in three environments: inside the hollow fibers, within the wall of the hollow fibers, and outside the hollow fibers. Different voxels will have different percentages of protons in each environment because of the random distribution of the hollow fibers relative to the imaging grid. In the characterization of the blood-side flow distribution, the fluid inside the hollow fibers was moving, while that outside the hollow fibers was stationary; the opposite occurred in the characterization of the dialysate-side flow distribution. Thus, in each voxel, the signal arises from both stationary and moving protons; the measured velocity in each pixel is therefore distorted and is generally lower than the true velocity. Nevertheless, the region-to-region variation in velocity can be estimated by measuring the mean velocities in different region-of-interest (ROI) large enough to include sufficient number of pixels to average out the variation arising from differences in the number of hollow fibers in each voxel. That is, by determining the mean velocity in one region, we will have derived a quantity that is proportional to the total flow rate of that region. This means velocity can then be used to compare the differences in flow in different ROI of the hemodialyzers. On the other hand, in the 2DFT technique, the measurable velocity resolution is limited by the velocity bin size selected for each Fourier-velocity image. Theoretically, we can improve the measurable velocity resolution by decreasing the velocity bin size. However, there is a tradeoff between velocity resolution, acquisition time, and signal-to-noise ratio for a given range of velocities. For a given range of velocities, a higher velocity resolution causes a lower signal-to-noise ratio and requires a longer acquisition time due to the increase in the number of Fourier-velocity images required for a smaller velocity bin size. 8.
CONCLUSIONS
We introduced two innovative, nonintrusive flow imaging techniques using MRI called the 2DPC and 2DFT techniques with the hope that these flowimaging techniques will provide a greater insight into the fluid transport phenomena in hollow-fiber membrane modules. These flow-imaging techniques provide a valuable nonintrusive experimental tool for studying flow distribution in hollow-fiber membrane modules. The experimental results obtained from these flow-imaging techniques give the actual fluid transport behavior in the 118
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
hollow-fiber membrane modules that cannot be predicted from numerical models and provide us a better physical understanding of the fluid transport phenomena in these modules. Moreover, an improved understanding of fluid transport phenomena in these modules may lead to the development of more efficient hollow-fiber membrane modules. These flow-imaging techniques can also be used to validate, improve, or develop numerical models capable of predicting separation performance of hollow-fiber membrane modules under different engineering designs and operating conditions. Some examples of experimental results obtained from hollow-fiber hemodialyzers are presented to illustrate the capabilities of these flow-imaging techniques. The readers are encouraged to refer to the references cited throughout this chapter for a full appreciation of these and other MRIbased flow-imaging techniques.
REFERENCES [ 1] [2] [3] [4]
[5] [6] [7] [8] [9] [10] [11] [12] [13]
B.E. Hammer, C.A. Heath, S.C. Mirer, G. Belfort, Quantitative flow measurements in bioreactors by nuclear magnetic resonance imaging, Biotechnology, 8 (1990) 327-30. C.A. Heath, G. Belfort, B.E. Hammer, S.D. Mirer, J.M. Pimbley, MRI and modeling of flow in hollow-fibre bioreactors, AIChE J, 36 (1990) 547-58. C. Donoghue, M. Brideau, P. Newcomer, B. Pangrle, D. BiBiasio, E. Walsh, Use of magnetic resonance imaging to analyze the performance of hollow-fiber bioreactors, Ann N Y Acad Sci, 665 (1992) 285-300. B.J. Pangrle, E.G. Walsh, S. Moore, D. DiBiasio, Investigation of fluid flow patterns in a hollow fiber module using magnetic resonance imaging, Biotech Tech, 3 (1989) 67-72. J. Zhang, D.L. Parker, J.K. Leypoldt, Flow distributions in hollow fiber hemodialyzers using magnetic resonance fourier velocity imaging, ASAIO J, 41 (1995) M678-82. S. Laukemper-Ostendorf, H.D. Lenke, P. Blumler, B. Blumich, NMR imaging of flow in hollow fiber hemodialyzers, J Membr Sci, 138 (1998) 287-95. T. Osuga, T. Obata, H. Ikehira, S. Tanada, Y. Sasaki, H. Naito, Dialysate pressure isobars in a hollow-fiber dialyzer determined from magnetic resonance imaging and numerical simulation of dialysate flow. Artif Organs, 22 (1998) 907-9. C.K. Poh, Characterization of fluid transport in artificial kidneys - A MR/approach, MS Thesis, University of Kentucky, Lexington, KY, USA, 2001. C.K. Poh, P.A. Hardy, W.R. Clark, D. Gao, Characterization of flow in hemodialyzers using MRI, Proc Intl Soc Mag Reson Med, 9 (2001) 1984. C.K. Poh, P.A. Hardy, D. Gao, W.R. Clark, Measurement of local ultrafiltration rate in a hemodialyzer: a novel MRI approach (abstract), ASAIO J, 47 (2001) 160. C.K. Poh, P.A. Hardy, Z. Liao, Z. Huang, D. Gao, W.R. Clark, Characterization of flow distribution in hemodialyzers using MR/(abstract), ASAIO J, 48 (2002) 179. C.K. Poh, P.A. Hardy, Z. Liao, Z. Huang, W.R. Clark, D. Gao, Effect of spacer yams on the dialysate flow distribution of hemodialyzers: A MRI study, ASAIO J, in press. C.K. Poh, P.A. Hardy, Z. Liao, Z. Huang, W.R. Clark, D. Gao, Characterization of flow distribution in hemodialyzers - A MR/approach, in preparation.
119
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
[14]
[151 [16] [17] [18] [19] [20] [21] [22] [23]
[24] [25] [26] [27] [28] [29]
[30] [31]
[32] [331
P.A. Hardy, C.K. Poh, Z. Liao, D. Gao, W.R. Clark, The use of magnetic resonance imaging to measure local ultrafiltration rate in hemodialyzers, J Membr Sci, 204 (2002) 195-205. Abragam, The Principle of Nuclear Magnetism, Oxford University Press Inc., New York, USA, 1983. B.R. Friedman, A.P. Salmon, G. Chaves-Munoz, C.R. Merritt, J.P. Jones, Principles of MR/, McGraw-Hill Co., USA, 1989. P.T. Callaghan, Principles of Nuclear Magnetic Resonance Microscopy, Oxford University Press Inc., New York, USA, 1995. M.A. Brown, R.C. Semelka, MR/: Basic Principles and Applications, 2nd ed., Wiley, John & Sons Inc., USA, 1999. T.J. Mosher, M.B. Smith, A DANTE tagging sequence for the evaluation of translational sample motion, Magn Reson Med, 15 (1990) 334-9. D.D. Blatter, D.L. Parke, R.O. Robinson, Cerebral MR angiography with multiple overlapping thin slab acquisition. Part I. Quantitative analysis of vessel visibility, Radiology, 179 (1991) 805-11. E.M. Haacke, A.S. Smith, W. Lin, J.S. Lewin, D.A. Finelli, J.L. Duerk, Velocity quantification in magnetic resonance imaging, Top Magn Reson Imaging, 3 (1991) 34-49. P. Mansfield, M. Bencsik, Fluid flow in porous systems, Magn Reson Imag, 16 (1998) 451-4. R.B. Miles, W.R. Lempert, Quantitative flow visualization in unseeded flows, Annu Rev Fluid Mech, 29 (1997) 285-326. F.W. Wehrli, Time-of-flight effects in MR imaging of flow, Magn Reson Med, 14 (1990) 187-93. A. Constantinesco, J.J. Mallet, A. Bonmartin, C. Lallot, A. Briguet, Spatial or flow velocity phase encoding gradients in NMR imaging, Magn Reson Imag, 2 (1984) 335340. P.M. Pattany, J.J. Phillips, L.C. Chiu, J.D. Lipcamon, J.L. Duerk, J.M. McNally, S.N. Mohapatra, Motion artefact suppression technique (MAST) for MR imaging, J Comput Assist Tomogr, 11 (1987) 369-377. P.J. Keller, F.W. Wehrli, Gradient moment nulling through the Nth moment. Application of binomial expansion coefficients to gradient amplitudes, J Magn Reson, 78 (1988) 145-149. J.G. Pipe, T.L. Chenevert, A progressive gradient moment nulling design technique, Magn Reson Med, 19 (1991) 175-179. P.R. Moran, A flow velocity zeugmatographic interlace for NMR imaging in human, Magn Reson Imag, 1 (1982) 197-203. F. St~hlberg, B. Nordell, A. Ericsson, T. Greitz, B. Persson, G. Sperber, Quantitative study of flow dependence in NMR images at low flow velocities, J Comput Assist Tomogr, 10 (1986) 1006-1015. G.T. Gullberg, S. Margaret, and F.W. Wehrli, A mathematical model for signal from spins flow during the application of spin echo pulse sequences, Magn Reson Imag, 6 (1988) 437-461. A. Caprihan, and E. Fukushima, Flow measurements by NMR, Physics Reports (Review Section of Physics Letters), 198 (1990) 195-235. J.M. Pope, S. Yao, Quantitative NMR imaging of flow, Concepts in Magnetic Resonance, 5 (1993) 281-302.
120
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
[34]
[351 [36] [37] [38] [39] [40]
[41] [42] [431 [44] [45]
N.J. Pelc, G. Sommer, K.C.P. Li, T.J. Brosnan, R.J. Herfkens, D.R. Enzmann, Quantitative magnetic resonance flow imaging, Magnetic Resonance Quarterly 10 (1994) 125-147. D.C. Wilcox, Basic Fluid Mechanics. DCW Industries Inc., California, USA, 1997, pp. 214. R.M. Henkelman, and M.J. Bronskill, Artifacts in magnetic resonance imaging, Rev Magn Reson Med, 2 (1987) 1-126. D. L. Parker, G.T. Gullberg, P.R. Frederick, Gibbs artifact removal in magnetic resonance imaging, Med Phys, 14 (1987) 640-5. S. Amartur, E.M. Haacke, Modified iterative model based on data extrapolation method to reduce Gibbs ringing, J Magn Reson Imaging, 1 (1991) 307-17. C.N. Chen, V.J. Sank, S.M. Cohen, D.I. Hoult, The field dependence of NMR imaging I. Laboratory assessment of signal-to-noise ratio and power deposition, Magn Reson Med, 3 (1996) 722-9. A. Macovski, Noise in MRI, Magn Reson Med, 36 (1996) 494-7. P.G. Walker, G.B. Cranney, M.B. Scheidegger, G. Waseleski, G.M. Pohost, A.P. Yoganathan, Semiautomated method for noise reduction and background phase error correction in MR phase velocity data, J Magn Reson Imaging, 3 (1993) 521-30. R.L. Wolf, R.L. Ehman, S.J. Riederer, P.J. Rossman, Analysis of systematic and random error in MR volumetric flow measurements, Magn Reson Med, 30 (1993) 8291. C. Tang, D.D. Blatter, D.L. Parker, Accuracy of phase-contrast flow measurements in the presence of partial-volume effects, J Magn Reson Imaging, 3 (1993) 377-85. C. Tang, D.D. Blatter, D.L. Parker, Correction of partial-volume effects in phasecontrast flow measurements, J Magn Reson Imaging, 5 (1995) 175-80. R.M. Hoogeveen, C.J. Bakker, M.A. Viergever, MR phase-contrast flow measurement with limited spatial resolution in small vessels: value of model-based image analysis, Magn Reson Med, 41 (1999) 520-8.
APPENDIX
This appendix shows the derivation of the linear relationship between the percent flow rate in each velocity bin and the mean velocity for a fully developed laminar, circular tube flow from the Poiseuille's equation. From the Poiseuille's equation, we have:
vi
2Q
=~(R n.R4
2 -r/2)
(A.1)
where v~ is the axial velocity at radius, r~ of the tube, Q is the total flow rate in the tube, and R is the inner radius of the tube. Rearranging Eq. (A. 1), we have:
121
Nonintrusive Characterization Of Fluid Transport Phenomena In Hollow-Fiber Membrane Modules Using MRI: An Innovative Experimental Approach - Poh
/
V i ;r~ 4
t~ = aiR 2 V
(A.2)
2Q
The concentric flow area, Ai of a laminar tube flow at radii, r~ and Fi+1 is:
Ai --;rg'(ri21--ri 2 )
(h.3)
Substituting Eq. (A.2) into Eq. (A.3) and simplifying, we have:
;rt.2R4 A i "- ~ ( v
20
i
-Vi+l)
(m.4)
From our 2DFT technique, the velocity bin size, ( v i - Vi+l) is a constant. Therefore, A; is independent of (vg- V;+l) and is a constant for a given R and Q from Eq. (A.4). The flow rate, Q; at a given concentric flow area, A; is given by:
Qi =(v)iAi
(A.5)
where
; is the mean velocity corresponding to A; and is given by: 1
(v)i = --~(vi + vi+, )
(A.6)
Eq. (A.5) shows that Qi is linearly proportional to ;. The percent flow rate, %Q; in each velocity bin, (v;- V~+l) is determined by dividing Eq. (A.5) with Q and then multiplying by 100 percent:
=~
Q
x 100%
(A.7)
Therefore, from Eq. (A.7), the percent flow rate in each velocity bin, %Q is linearly proportional to the mean velocity, ; where A~ is a constant for our 2DFT technique, and Q is a constant for a fully developed laminar, circular tube flow.
122
Membrane Contactors and Environmental Applications
123
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 6
Industrial applications and opportunities for membrane contactors R. Klaassen, P.H.M. Feron, R. van der Vaart, A.E. Jansen
TNO Institute of Environmental Sciences, Energy Research and Process Innovation, Department Chemical Engineering PO Box 342, 7300 AH Apeldoorn, The Netherlands phone: +31 55 493196 fax: +31 55 5493410 email: [email protected] SUMMARY
In a membrane contactor membrane separation is fully integrated with another separation technology, like extraction or absorption, in order to exploit the benefits of both technologies fully. Membrane contactor applications can be found both in water and gas treatment. Several recently developed applications that have been introduced in industry - aromatics recovery from a process water, selective removal of heavy metals from a galvanic process bath and ammonia recovery from an off gas s t r e a m - and new applications under developmentCO2 membrane gas absorption and membrane stripping- will be discussed technically and economically to show the wide scope of the technology. 1.
INTODUCTION
Stringent demands are put onto new separation technologies due to stricter product quality requirements, environmental legislation, and energy efficiency demands and needs for cost reduction. In order to meet these ever increasing needs there is a tendency to combine processes to a hybrid process. In a membrane contactor a membrane separation is not only combined with an extraction or absorption process but both processes are fully integrated and incorporated into one piece of equipment. In this way advantages of both processes can be fully exploited.The membrane offers a flexible modular energy efficient device with a high specific surface area. The extraction or absorption process can offer a very high selectivity and a high driving force for transport even at very low concentrations. Applications of membrane contactors can be found in gas and water
125
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
treatment. In the membrane gas absorption process the gas stream to be treated is brought very efficiently into contact with an absorption liquid in the membrane contactor. Typical examples are CO2 removal from flue gas or indoor air, flue gas desulphurisation, indoor air conditioning, ammonia recovery from off gasses or mercury removal from natural gas. In the pertraction process water is treated in a membrane contactor with an extraction liquid. Typical applications are hydrocarbons (e.g. aromatics, phenol, haloginated hydrocarbons) removal and recovery from waste and process water and selective recovery of heavy metals from e.g. galvanic bath liquids or waste water. In a membrane contactor also ammonia can be removed and recovered from waste water in a combined stripping and absorption process. Many different developments in the field of membrane contactor technology are going on worldwide. Development of a specific application normally starts with a proof of principle followed by a feasibility study, further development, pilot plant tests and finally full-scale demonstration on site. After a general introduction on membrane contactors some examples of successful developed full-scale industrial membrane contactor installations will be discussed in this paper. These industrial reference installations help to gain confidence of end users in the membrane contactor technology. Thus further market introduction will be facilitated as well as development of new applications for membrane contactors. 2.
MEMBRANE GAS ABSORPTION
2.1
Principle
Membrane gas absorption is a gas-liquid contacting operation [ 1,2]. The key element in the process is a microporous hydrophobic hollow fibre membrane. The process is illustrated in figure 1 for removal of component X from a gas stream. The gas stream is fed along one side of the membrane where an absorption liquid is flowing at the other side of the membrane. The hydrophobic membrane wall keeps gas phase and absorption liquid separated from each other. The absorption liquid is chosen in such a way that it has a high affinity for component X. Component X will now diffuse through the gas filled pores of the membrane to the other side of the membrane where it is absorbed in a liquid phase. Absorption in the liquid phase takes place either by physical absorption or by a chemical reaction. This determines the selectivity of the process. The membrane used gives no contribution to the selectivity: the membrane's role is to keep the two phases separated and to provide a large contacting surface area.
126
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
Hollow fiber
gas
=,.,ou,v,u,, liquid ~
gas
Membrane
Figure 1: Membrane gas absorption For operation of membrane gas absorbers it is essential that liquid phase and gaseous phase do not mix. This means that the absorption liquid is not allowed to enter the pores. This is influenced by pore size, pressure difference across the membrane and the interaction of the absorption liquid with the membrane material. This is described by the Laplace equation: d P = - 2 (y/r) . c o s (~o)
( 1)
where: d P - pressure difference between liquid and gas [Pa] 7 - surface tension of the absorption liquid [N/m] r - pore radius [m] (p - contact angle for liquid and membrane material The membrane pores will not be wetted if the contact angle is greater than 90 ° and the pressure difference is limited for a given pore size. Suitable membrane materials for aqueous absorption liquids are non-polar polymers such as polypropylene, polyethylene and Teflon. 2.2
Membrane module The design of a membrane absorber cannot be based on conventional
127
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
available hollow fibre membrane modules. These module designs have historically been based on the use of filtration duties and consist of a bundle of hollow fibres in a tubular housing. This gives ill-defined flow conditions on the outside of the fibre bundle. When membranes are used for contacting duties it is necessary that flow conditions are well def'med on both sides of the membrane to achieve good mass transfer. Furthermore it is essential that, given the often large volume flows of flue or off gas streams, the module design gives low pressure drops and is easy to scale up. TNO has patented a new membrane module design to overcome the problems with conventional membrane modules. Figure 2 shows this new membrane module design [3] for contacting duties. The module consists of a rectangular housing with fibres at well-def'lned positions. The gas flow is perpendicular to the fibres. The absorption liquid is fed through the fibres. Elements of the module can be coupled parallel and in series. In this way the module can be easily adjusted to the flow rate of the gas stream to be treated and the desired removal efficiency.
inlet
outlet
inlet
Figure 2: New membrane module
Important features of the module are: well defined flow conditions on both sides of the membrane, easy scale up, high mass transfer, low pressure drop and high specific surface area. Specific surface area's of well over 1000 rn2/m3 can be reached. 2.3
Membrane gas absorption versus conventional contactors
Conventional G-L contactors, such as packed towers, preferably operate
128
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
with relatively large gas flows under steady-state conditions. Membrane gas absorption can be an eminently suitable alternative for 'smaller' gas flows and operation at variable loads. Especially by using hollow fibre membranes it is possible to develop a very compact gas-liquid contactor. Hollow fibre membrane modules combine a high specific surface area, a well-def'med gas flow regime and short diffusion ways in a modular piece of equipment. The membrane gas absorption technology is able to treat gas streams with extreme variations in gas flow and or concentrations of the components to be removed. This is possible at low costs in very compact equipment. The reduction in volume and weight compared to a conventional absorber can be more than a factor 10. Through the use of hollow fibre membranes the membrane gas absorption process can offer operational and economical advantages over conventional spray towers or packed columns. Operational advantages include independent Gas/Liquid control, flexible operation, optimal load of the absorption liquid, no entrainment, flooding or foaming and modular low weight and very compact equipment. Economical advantages include low investment costs, low pumping power for the absorption liquid and the fact that no expensive civil engineering work is required. That gas absorption membranes can result in very compact equipment can be illustrated by the following example: For a coal fired power plant of 645 MWe a design has been made for a conventional absorber (spray tower) and a membrane gas absorption unit. A flue gas flow of 2"106 nm3/hr has to be treated in this case. The calculated overall volumes of the absorbers are given below. Conventional SO2 absorber Calculated membrane gas absorption
2.4
- volume 9000 m 3 - volume 250 m 3
Applications
Membrane gas absorption can be applied for removal of those components from a gas stream where a suitable absorption liquid is available. A suitable absorption liquid has a high affinity for the components to be removed and does not wet the membrane. The majority of absorbents used in conventional gas absorption processes can also be used for membrane gas absorption. Applications [4,5] for the use of membrane gas absorption are listed below. Flue-gas and off-gas containing e.g. SO2, HC1, NH3 or HES can be treated to meet emission standards in many industrial situations. Recovery and re-use of CO2 from flue gas, biogas and off-gases in horticultural industry, beverage production and other industrial
129
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
applications. Upgrading and desulphurisation of biogas from anaerobic digesters and landfills. Acid gas removal and dehydration of natural gas. Acid gas removal from fuel gas mixtures. Mercury removal from natural gas, flue gas or glycol overheads. Alkene-alkane separation in petrochemical industry. Indoor air treatment for e.g. tobacco smoke components. Due to the modular equipment used in membrane processes application of gas absorption membranes is also possible on a small scale. This is particularly valuable for applications in the field of environmental technology, as it enables an efficient and economical treatment of small-scale emission sources. If regeneration of the absorbent cannot be done one site, central regeneration is another possibility. In paragraph 2.5 the recovery of ammonia with membrane gas absorption is discussed. In paragraph 2.6 developments with respect to CO2 absorption are shown. Another industrial application of membrane gas absorption is the transfer of unstable C102 to potable water [27]. Also the recovery of VOC's from a paint booth exhaust is recently shown under minipilotplant conditions [28].
2.5.
Membrane gas absorption installation for ammonia recovery An industrial membrane gas absorption unit for ammonia recovery has been installed at a dyes intermediates production plant of Aliachem in Pardubice the Czech Republic (Figure 3). In this plant dyes intermediates are produced batch wise in pressurised reactors. Ammonia is used as a reactant in the reactor. Depressurisation of the reactor results in an ammonia containing off gas stream, which gives a significant loss of ammonia. In the membrane gas absorption unit the total off gas stream is treated with water as absorption liquid. The ammonia is recovered as an aqueous ammonia solution of > 20 wt% which can be re-used in the dyes intermediates production process. The membrane gas absorption unit is successfully in operation since June 1999; aqueous ammonia solutions of 27 wt% ammonia have been produced and the emission of ammonia to the environment is reduced with 99.9%. The installation has a capacity to absorb 50 kg/hr ammonia and has proved to be very easy in operation. Variation in gas flow rate or ammonia concentration can be handled without any problems.
130
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
Figure 3: MGA unit on site in Pardubice, Czech Republic 2.6
C02 removal
2.6.1 Introduction CO2 membrane gas absorption has been under development at TNO for a number of years. The potential of this technology has been identified in a number of assessment studies [6,7]. The ensuing customer-driven development has been focusing on the production of carbon dioxide from flue gas with the aim of supplying high purity carbon dioxide to the horticultural industry to increase crop yield [8,9]. Other commercial applications are in the field of life support (spacecrafts, submarines, operating theatres, consumer products, biogas treatment). In recent years there has also been an increased interest in CO2-removal and subsequent storage in view of the enhanced greenhouse effect. Figure 4 shows an overview of the atmospheric pressure applications currently being investigated. The process is based on novel membrane absorption method patented by TNO [10], which employs dedicated aqueous absorption liquids, called CORAL, based on amino-acids and alkaline salts. Benefits of these liquids over conventional amines are: - Stable operation with cheapest membranes available viz. polyolefin membranes. - Oxygen stability. - Improved corrosion resistance. - No vapour pressure, hence no loss of absorption liquids through evaporation.
131
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
CO~-MGA application range 100 --0-
x "
~
• O~ (J
~
--
Greenhouse Blogas Submarine Space Medical Consumer
0.1 --o--
0.01 0.001
1
1000
Capacity [kg C021h ]
Figure 4:CO2 membrane gas absorption applications (atmospheric pressure) currently under development.
2.6.2 C02-production for greenhouses In the Netherlands carbon dioxide is used to promote plant growth in greenhouses. Depending on the crop, the production can be increased by 25% by increasing the CO2-concentration to 500 ppm. In temperate climates it is necessary to heat these greenhouses, which makes the horticultural industry in the Netherlands a large consumer of natural gas. Future heat supply systems will be largely based on cogeneration plants (gas engines and gas turbines) as significant energy savings can be achieved. Energy savings can be increased if the heat supply is coupled to carbon dioxide supply. This will incite growers to buy heat from the cogeneration plant rather than to use their own boilers. These boilers are needed to cover peak demand and emergencies. Also the amount of available carbon dioxide per unit heat will be larger in case of combined production of heat and electricity, thus allowing for an improved production. The heat demand and CO2 demand are generally anti-cyclic. During periods of high heat demand (during winter and at night) little or no CO2 is needed to maintain the desired CO2 levels in the greenhouse. During periods of high CO2 demand (during summer and during the day) little or no heat is needed. The thermal energy of the cogeneration plant can either be used for greenhouse heating or for the production of carbon dioxide. This will also extend the operating time of the cogeneration plant to daytime hours, which is an additional benefit in terms of capacity credits. A schematic drawing of a cogeneration plant supplying heat and carbon dioxide to greenhouses is shown in Figure 5.
132
IndustrialApplicationsAndOpportunitiesForMembraneContactors- Klaassen
Greenhouses Flue gas
I
ico2 (loo°/o) ,
l
H.t
N,ura,
Electricity 1
T Natural gas
Grid
Figure 5: Cogeneration plant supplying heat and carbon dioxide to greenhouses
2.6.3 Pilot plant Pilot plant experiments were carried out to achieve the following objectives: - Assessment of mass transfer under a variety of conditions with several liquids, - Assessment of heat consumption, - Assessment of CO2-quality. During the feasibility study a pilot plant has been designed and built for testing in the lab as well as for testing on location with a real flue gas. The basic specifications of the pilot plant are shown in table 1. Table 1: Basic specifications of COz-removal pilot plant CO2-concentration 3.5 % Gas flow rate 25 m3/hr CO2-removal 80% Liquid flow rate 20-1001/hr The pilot plant flow sheet is shown in Figure 6. In case of the lab experiments CO2 and air are mixed via mass flow controllers and fed to a prototype cross-flow membrane absorber (DAM-module). In this absorber is CO2 is transferred from the gas stream to a liquid stream in a counter current fashion. On location a small stream has been taken from the chimney, cooled down and pumped directly to the membrane absorber. The rich liquid is fed to the top of a stripper via a plate heat exchanger. The stripper temperature is maintained at a temperature of around 120°C by an electric heating element. As a result of the temperature increase of the absorption liquid CO2 is liberated from the solution. Water is recovered in the condensor and trickles back into the stripper. The lean liquid exits the stripper at the bottom and is fed back to the membrane absorber via the heat exchanger. Temperatures, pressures, dew
133
IndustrialApplicationsAndOpportunitiesForMembraneContactors- Klaassen
points, flow rates and heat consumption have been measured and are used for the performance assessment. C02
Steam Gas in
Air r v
~Mi i
mo Ju
Rich I exchanger I =' I Heat =I
I "~
! Stripper
Lean
Cooler
~
Heatero
Gas out
.= Gas analyser
Figure 6: Pilot plant flow sheet
The specifications of the prototype cross-flow membrane module designed according to [3] are shown in Table 2. Table 2
Specifications prototype cross-flow membrane module Membrane area 5.0 m 2 Fibre material Polypropylene (Accurel® Q3/2) Nominal pore size 0.3 i.fin Fibre diameter (inside/outside) 0.6/1.0 mm
The pilot plant as operated on location is shown in Figure 7.
Figure 7: Pilot plant on location.
134
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
2.6.4
Mass transfer performance
An important performance parameter is the mass transfer in the membrane gas absorption process, as this will determine the membrane area requirement. Membrane area is expected to represent the largest investment component. The mass transfer, incorporating a chemical reaction, is influenced by: - The type of absorption liquid. The reaction kinetics and reaction path are dependent on the nature of the active components and their concentration levels. - The absorption temperature. Higher temperature leads to faster reaction kinetics but the thermodynamic equilibrium will be less favourable. A trade-offhas to be made here. - Fluid dynamics. The fluid dynamic conditions (gas and liquid) might influence the mass transfer to some extent. The effect is however small. - Membrane type. The membrane type should be sufficiently porous not to impose any additional mass transfer barriers. - CO2-10ading in liquid and CO2-concentration levels in gas. If the liquid already has a considerable CO2-content this will limit the mass transfer. A proper regeneration process can achieve this, but a trade-off with the energy consumption has to be made. Also at high CO2-concentration levels mass transfer will be limited because local saturation of the active components might OCCUr.
The mass transfer performance, expressed as a carbon dioxide flux through the membrane, is shown in figure 8 for the CORAL20 liquid at room temperature as a function of the average carbon dioxide partial pressure. The data have been gathered from a variety of experiments carry out with a small pilot plant including regeneration of the absorption liquid. It clearly shows that over a wide range of the carbon dioxide partial pressure the membrane fluxes will increase with increasing partial pressure. A feasibility study has demonstrated that the CO2 can be produced economically from flue gas on a large scale [11]. The economical analysis in this study indicates that at a production capacity of 10 tonnes CO2 per hour the productions costs will be around 50 US$ per tonne CO2. At this cost level the process is cheaper than e.g. CO2 delivered by truck or current CO2-production processes. A development project is currently underway in which a pilot plant producing around 150 kg CO2 per hour will be built. Successful completion of the development will then pave the way towards a demonstration on a large scale (10 tonnes CO2 per hour) for a greenhouse.
135
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
800
700
600
500
.~ 400 300
200
100
0
v
,
'
re [kPal
Figure 8: Measured CO2 membrane mass flux for CORAL20 at room temperature as a function of the logarithmic mean CO2 partial pressure. 3.
PERTRACTION
3.1
Principle
Pertraction is a non-dispersive membrane based liquid-liquid extraction process. Process or wastewater flows contaminated by organic substances such as aromatics or chlorinated hydrocarbons can be cleaned by pertraction. The pollutants are removed from the water by extraction into an organic extractant, which is immiscible with water. The membrane itself has no selectivity. The process selectivity must come from the extractant. The interface between wastewater and extractant is immobilized using a hydrophobic microporous hollow fiber membrane by means of a small transmembrane pressure gradient (-0.1 bar)[ 12,13], see figure 9. Contrary to conventional extraction processes there is no direct mixing of extractant in the wastewater stream. This offers important advantages over conventional extraction: The often difficult and time-consuming separation between water phase and extractant is not necessary. Flows of waste water and extractant are flexible, and can be adjusted
136
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
independently of another, making process optimisation simple and allowing a highly efficient contact between a large volume of (waste) water and a very small quantity of extractant. Pertraction installation can be of very compact construction, thanks to the high specific surface area and good mass transfer of the pertraction membrane modules. No differences in density between the water phase and the organic extractant are required. extractant hollow fibre /^\
J '!
waste water
/
i
'
i
:
!
' i
i
\
/
\
/
\
! s s
o,
i! i! i
~
waste water
t
, l i
!
i
i
'
t
~r
T
extractant
extractant
Figure 9: Principle of pertraction
An advantage of pertraction over stripping and carbon adsorption is that both volatile and non-volatile components can be extracted in one step. In the pertraction process pollutants are concentrated in an extractant (Figure 10). The latter can be either directly re-used or regenerated in e.g. a vacuum film evaporator, and can thus continue to circulate in the system, while the pollutants are released in pure form. pertraction module P(P(P~,.-r., cleaned
" "
contaminated water
... _ ~
~
"
water
~<
loaded extractant
film evaporator regenerated extractant
Figure 10: Pertractie-regeneration process, with pollutants released in pure form.
137
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
The choice of the proper extractant is an essential step in development of new pertraction applications. In -
the selection procedure requirements for the extractant are: A high affinity for the components to be removed from the waste water. Very low solubility in water. Non-toxic. Low viscosity. Possibilities for re-use or regeneration.
Pertraction is especially suitable for removal of hydrophobic organic compounds like halogenated, aromatic and aliphatic organics from aqueous streams. Water streams containing these hydrocarbons can be found at many places: Effluents/process streams from pharmaceutical-, off shore-, automotive-, paint-, petrochemical- and chemical-plants. Water originating from tank cleaning and tank storage firms. Effluents from garages dry cleaning and wood preservation firms. Contaminated ground water. Landfill leachate from sites containing chemical wastes. 3.2
Economics
of Pertraction
A comparative cost model [ 14] has been developed for pertraction and the conventional technologies air stripping and activated carbon treatment. The cost models for all the technologies are based on the same assumptions to obtain comparable results. The results of the cost models are total treatment costs per cubic meter water treated. These costs include depreciation of the installation, energy costs, maintenance and operation. In table 3 results of the cost calculations are shown for trichloroethene based on a waterflow of 10 m3/hr with an inlet concentration of 10.000 ~tg/1. The outlet concentration of trichloroethene is a variable and lies in the range between 1000 ~tg/1 and 1 ~tg/1. The results show that in all the cases calculated pertraction is cheaper than the alternatives (air stripping or activated carbon filtration). Other substances, which can be economically treated by pertraction, include chlorinated solvents (e.g. carbon tetrachloride, chloroform, tetrachloroethene, trichloroethene), PCB's, di- and tri-chlorobenzene, pesticides and higher polycyclic hydrocarbons.
138
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
Table 3: Cost evaluation: trichloroethene. flow rate: 10 m3/hr trichloroethene inlet concentration 10.000 ,g/1 u'-' Outlet concentration
Total costs in USD per m3 treated water for the different selected processes Air stripping; with air treatment
Activated carbon filtration
Pertraction
1000 ~tg/1
USD 0,53
USD 0,83
USD 0,28
100 ~xg/1
USD 0,53
USD 0,89
USD 0,36
10 ~tg/1
USD 0,60
USD 0,92
USD 0,44
l~g~
USD 0,60
USD 0,94
USD 0,51
3.3 Pertraction installation for recovery of aromatic compounds from waste water
A full-scale pertraction installation has been successfully put into operation in 1998 at the chemical firm KoSa Netherlands BV in Vlissingen the Netherlands. The pertraction installation is equipped with Liqui-Cel ® Membrane Contactors [29]. It treats 15 m3/hr waste water originating from a chemical reactor (Figure 1 I). The water is heavily polluted with the product formed in the reactor. This product is an aromatic compounds. In the pertraction installation the organic compound is extracted from the waste water with a feedstock for the reactor as extractant. In this way the water is not only cleaned but also lost product is recycled back to the reactor. Next to this the existing wastewater incineration installation could be put out of operation thanks to the pertraction installation. In this way KoSa saves 5-million m 3 natural gas annually. The unit is more then three years in operation now and is a critical element in the production process.
Figure 11" Industrial pertraction installation.
139
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
4.
EMULSION PERTRACTION
4.1
Principle
Emulsion pertraction [15] is a process evolved from a combination of permeation through membranes and solvent extraction. In the emulsion pertraction process the water phase is kept apart from the emulsion phase by a hydrophobic microporous membrane. The emulsion phase consists of an organic solvent with a dissolved extractant as continuous phase with aqueous droplets of strip liquid dispersed in it. The contact surface between water phase and emulsion phase lies in the pores of the membrane. The metal to be removed from the water phase is bound by the extractant present in the pores of the membrane. At the other side of the membrane the strip liquid continuously regenerates the extractant. The strip liquid is dispersed in the organic phase. There is no need to make the dispersion very stable because the hydrophobic nature of the membrane keeps the waste water and aqueous strip liquid always separated. Therefore no aqueous phase can pass the membrane. As shown in figure 12, the metal indicated as Me is bound to extractant L to form a complex Me-L which is regenerated at the interface of the droplets strip liquid to form for example a concentrated metal sulphate solution. There is no selectivity needed of the membrane itself. The selectivity of the process is provided by the extractant. Hollow fiber
"~
i ........... '-' ...... 'i
i
i
i
~,~j =
/
!
,,_,_27,,__,,_.,_,:
..,.,,-.,.
I I I
Membrane
>
t t '
I
!
-
~
t
t
I
I
~
water
~
emulsion
Me
Metal
L
Extractant
I
water
Figure 12: Principle emulsionpertraction The emulsion used in emulsion pertraction is analogue to the emulsion used in the Liquid Membrane Process (LMP) developed by the university of Graz [16]. The main advantage of pertraction over the LMP process is that the emulsion may be unstable while LMP requires a stabilised emulsion that is difficult to break. In the Stabilized Liquid Membrane (SLM) process only the pores of the 140
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
membrane are filled with the organic extractant. This process suffers from stability problems due to depletion of the extractant phase, which only has the very small volume of the membrane pores. These problems are over come with the emulsion pertraction process where the extractant that is flowing through the hollow fiber membranes can easily make up for any small losses that might occur from the membrane pores. These losses can be fatal to the SLM process. Advantages of emulsion pertraction over normal extraction are: - Extraction and stripping can be carried out in one operation. - No possibility of emulsion formation in the water phase. - Volume of extractant is relatively small. - Process parameters are very flexible. - Phase separation is not necessary. - Large specific surface area because of hollow fibre membranes. - Short diffusion ways. - Compact modular equipment. - Low energy consumption. Emulsion pertraction is suitable for selective removal of heavy metals like copper, zinc, iron, chromium, cadmium or lead from aqueous solutions. The heavy metals can be recovered in a concentrated solution. Lifetime extension of passivating bath liquids is an industrial application for emulsion pertraction. This application is discussed in paragraph 4.2. The principle of emulsion pertraction is also used by Commodore Separation Technologies for the recovery of Cr(VI) from waste water [30].
4.2 Emulsion pertraction installation for life time extension of passivating bath liquids An emulsion pertraction installation (Figure 13) has been tested on site at Rogal in Enschede the Netherlands to extend the lifetime of their passivating bath. Passivating is an operation frequently used in galvanic industry to improve corrosion resistance of galvanized objects. The lifetime of the passivating bath liquid is limited due to accumulation of zinc and iron. After the zinc and iron concentration reach a certain limit the colour and corrosion resistance of the passivated objects falls no longer within the specs and the bath liquid has to be replaced. Selective removal of iron and zinc from the passivating bath with an emulsion pertraction installation can extend the lifetime of the bath. Care must be taken that no of the active elements chromium or cobalt are being removed. Tests on site have shown that the emulsion pertraction installation is easy
141
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
to operate. Operation takes no more then 20 minutes a week. Zinc and iron(III) are removed from the passivating bath with a high selectivity; there is a
Figure 13: Emulsion pertraction installation at site of Rogal
negligible loss of chromium and cobalt. The extractant used is continuously regenerated with sulphuric acid. The membranes are resistant to the on-site conditions and transport rates are sufficiently high that a compact installation can be built. Due to a reduction in the use of raw materials (flesh bath liquid) and less metal sludge to be disposed of the installation gives annual benefits of USD 10,000 whereas the total annual operational costs are only USD 6,000. 5.
M E M B R A N E STRIPPING
5.1
Introduction
The removal of gases, especially oxygen, from water is important in several industrial sectors, e.g., in the microelectronics industry (rinse water) [17], in the electricity production sector (boiler feed water) [18, 19], in the brewing of beer (heavy brewing) [20, 21] and in the oil & gas industry (injection water) [22]. In the above examples, the oxygen needs to be removed to prevent oxidation of substrates, equipment or dissolved components or to prevent biological growth. Several methods are commercially available for oxygen removal [23]. Classical methods involve the (thermal) vacuum stripping of water in a voluminous packed bed, often combined with the application of a stripping gas. Also, the addition of chemicals like hydrazine or sulphite to the water system is in use [24]. Drawbacks of these methods are high energy consumption, the harsh chemical nature of hydrazine, the increase of chemicals concentration (conductivity) in the product-water in the case of sulphite addition, and the 142
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
consumption of chemicals. More sophisticated methods are based on reduction of oxygen with hydrogen over a Palladium catalyst or stripping via a membrane contactor. These methods are intrinsically clean. A drawback of the hydrogen method is the difficult mixing/dissolving of hydrogen in water, requiring voluminous equipment [ 19]. 5.2
Catalytic membrane contactor The use of membrane contactors to improve the dissolution of hydrogen has been described. The catalytic phase may be present on ion exchange beads in between hollow membrane fibres [25] or in a separate subsequent contactor. Drawbacks of these methods are the use of expensive Pd-loaded ion exchangers and the fact that the product water contains excess hydrogen gas. Recently, the proof-of-principle of a fully integrated catalytic membrane contactor (CMC) has been reported [26]. The CMC system is based on the integration of the catalyst phase with a hydrophobic microporous membrane that keeps the hydrogen and water separated. The catalyst is present at the interface where the reaction takes place on the membrane external surface (Figure 14). This system minimises mass transfer resistances, resulting in optimal performance and minimal required membrane surface area, amount of catalyst and amount of hydrogen lost to the water. No pre-mixing of hydrogen with the water stream is required, which results in a much more compact installation. Moreover, this type of membrane is based on commercially available and relatively low-cost hollow fibre membranes. The CMC was evaluated for use in the production of ultrapure water for the semiconductor industry. It is expected that the CMC should remove oxygen at 23-34% lower investment costs, 20-38% lower fixed costs and 35-75% lower operating costs.
Figure 14: Back-scatterelectron microscope image of a Palladium-coatedporous hydrophobic membrane fibre. The highlighted area indicates the Palladium coating. 6. CONCLUSIONS 143
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
Membrane contactors combine properties like modular concept with high specific surface area from membrane technology with the selectivity from conventional separation processes like extraction or absorption. This creates a new generation of highly efficient, compact and flexible separation technologies. Applications for membrane contactors can be found in gas-liquid and liquidliquid contacting operations. Examples are removal of specific compounds from gas streams (e.g. NH3, CO2, SO2, H2S, H20) and selective recovery o f hydrocarbons and heavy metals from aqueous streams. The first full scale industrial membrane contactor installations are in operation in industry now. The success o f these applications depends on the unique advantages o f membrane contactors (flexible capacity, no phase separation, low volume, low weight, extreme phase ratios etc.). On the other hand it is also very important to demonstrate the reliability on the long run under industrial conditions. This will give more confidence in the m e m b r a n e contactor technology for other end users and can lead to a whole new world o f membrane contactor applications.
REFERENCES [1]
[2] [3] [4]
[5] [6]
[7] [8] [9] [10] [11]
Zhang, Q., Cussler, E.L. - Microporous hollow fibres for gas absorption I. Mass transfer in the liquid, J. Membr. Sci., 23(1985) 321-332. Klaassen, R. et. al, - SO2 removal from flue gas with gas absorption membranes, Proceedings Large Chemical plants, p 75-77, October 1992 Antwerpen Belgium. Meulen, B. Ph. ter - Transfer device for the transfer of matter and/or heat from one medium flow to another medium flow., Eur patent 0509031, US patent 5230796. Klaassen, R., Feron, P.H.M., Jansen, A.E. - Membrane gas absorption environmental applications, Proceedings ICOM 96 p 916-917 Feron, P.H.M., JansenA.E., Klaassen, R. - Membrane gas absorption: a new membrane technology for gas separation and purification 6th Aachener Membran Kolloquium 3-5 March 1997 Feron, P.H.M., JansenA.E., Klaassen, R. - Membrane Technology in Carbon Dioxide Removal, Energy Convers. Mgmt., Vol. 33, No. 5-8, pp. 421-428, 1992 Feron, P.H.M - Membranes for carbon dioxide recovery from power plants, "Carbon Dioxide Chemistry: Environmental Issues",, Ch. 24, pp. 236-249 eds. J. Paul, C.M. Pradier, The Royal Society of Chemistry, Cambridge, 1994 Feron, P.H.M., JansenA.E. - Capture of carbon dioxide using membrane gas absorption and reuse in the horticultural industry, Energy Convers. Mgmt. 36(411-414) 1995 Feron, P.H.M., JansenA.E . - The production of carbon dioxide from flue gas by membrane gas absorption, Energy Convers. Mgmt. 38($93-$98)1997 Feron, P.H.M., et al. - Method for gas absorption across a membrane, USP 5749941, EP 0751815 Feron, P.H.M., JansenA.E. - Techno-economic assessment of membrane gas absorption
144
Industrial Applications And Opportunities For Membrane Contactors - Klaassen
[12] [13]
[14] [15]
[16] [17]
[18] [19] [20] [21] [22] [23] [24]
[25] [26]
[27]
[28]
[29] [30]
for the production of carbon dioxide from flue gas, Greenhouse Gas Control Technologies, pp. 53-58 Pergamon, 1999 Kiani, A., Prasad, R., Bhave, R.R., Sirkar, K.K. - Solvent extraction with immobilized interfaces in a microporous hydrophobic membrane, J. Membr. Sci., 20(1984) p125-145 Klaassen, R., Jansen, A.E., Bult, B.A., Oesterholt, F.I.H.M., Schneider, J., - Removal of hydrocarbons from waste water by pertraction - Proceedings 7th International Symposium on Synthetic Membranes in Science and Industry. Ttibingen Germany August 29- September 1, 1994 p 316 - 319 Klaassen, R., Jansen, A.E., Bult, B.A., Oesterholt, F.I.H.M., - Pertraction of hydrocarbons from waste water streams; Final report, ref. nr. 94-039, March 1994. Klaassen, R., Jansen, A.E., - Selective removal of heavy metals with emulsion pertraction, Minerals, Metals and the Environment II p207-216 The Institution of Mining and Metallurgy London 1996 Draxler, J., Marr, R. - Emulsion liquid membranes, Chem.Eng.Process., 20 (1986) 319329 Yagi, Y., Y. Kamasa, Ohmi, T. - Advanced ultrapure water systems with low dissolved oxygen for native oxide-free wafer processing, IEEE Transactions on semiconductor manufacturing, 1992.5: p. 121-127. d'Angelo, P., - Oxygen removal: theory and potential use of deoxygenation membranes in the utility industry, Ultrapurewater, 1995. July/Aug: p. 60-63. Matt, K., - Preventing corrosion by catalytic reduction of oxygen in boiler and recirculation water, Reports Sci. Technol., 1997.59: p. 52-57. Spielberger, F., - Neue Wege der Sauerstoff entfemung im brauwasser. Brauwelt, 1991.23: p. 975-979. Kohnke, H.-V., - Wasserentgassung und Einsatz entgasten Processwassers. Brauwelt, 1991: p. 1357-1359. Henriksen, N., - Deaeration of water, 1985. Butler, I.B., M.A.A. Schoonen, Rickard, D.T., - Removal of dissolved water from water:a comparison of four common techniques, Talanta, 1994.41: p. 211-215. Chabria, M.C., Sharma, M.M., - Kinetics of absorption of oxygen in aquous solutions: (1) acidic chromous chloride, (2) acidic titanous chloride and (3) ammoniacal cuprous chloride. - Chem. Eng. Sci., 1974.29: p. 993-1002. Li, K., I. Chua, Teo, W.K., - Removal of dissolved oxygen in ultrapure water production using a membrane reactor, Chem. Eng. Sci., 1995.50: p. 3547-2556. Vaart, R. van der, Hafkamp, B., Koele, P.J., Querreveld, M., Jansen, A.E., Volkov, V.V., Lebedeva, V.I., Gryaznov, V.M., - Oxygen removal from water by two innovative membrane techniques, Conference on Ultra pure water Asia 2000. Mutto, T., Dean, D . - Development of Chlorine dioxide gas transport contactor (ERCO R101 Generator) for use in drinking water disinfection. Proceedings 12th Annual NAMS Meeting Lexington, KY, May 15-20 2001, p53. Majumdar, S., Bhaumik, D., Sirkar, K.K., Simes, G. - A pilot-scale demonstration of a membrane based absorption stripping process for removal and recovery of volatile organic compounds. Environ.Prog.,20(1) April 2001, p. 27-35. Celgard website: www. Celgard.com. Ho, W.S.W. and Poddar, T.K., - New membrane technology for removal and recovery of chromium from waste waters, Environ. Prog., 20(1) April 2001, p. 44-52.
145
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 7
Membrane contactors: recent developments A. S. Kovvali + and K. K. Sirkar*
Department of Chemical Engineering New Jersey Institute of Technology, Newark, NJ 07102 +Present Address: Baker Petrolite, Sugar Land, TX, 77478; (Tel.) 281-276-5740; (Fax) 281-276-5492; email: [email protected] *Corresponding author: (Tel.) 973-596-8447; (Fax) 973-642-4854; email: [email protected] 1.
INTRODUCTION
Contacting two phases for efficient transfer of one or more solute species between the two phases is a primary objective of most conventional mass transfer operations. The contacting phase pairs can be gas-liquid, liquid-liquid, solid-liquid or solid-gas. The contacting area provided between the phases and the mass transfer coefficients of the solutes in each phase influence directly the transfer rates of solute species. Conventional phase contactors develop the contact area by forming of micro-phase domains (bubbles for gases, droplets for liquids and particles for solids) of one phase dispersed in the other phase. The interphase mass transfer rates are influenced strongly by the size and population of droplets, bubbles, particles etc. For superior rates of mass transfer to occur, the size of the droplets, bubbles etc. should be small while their number should be large. Another class of conventional contactors operates by forming thin films of one phase over which the other phase passes, accomplishing the species transfer. However, in each of the above mentioned cases, the dispersive contacting of phases necessitates the separation of phases after contacting. Further, scaling of such devices is problematic. Given such a context, membranes have found a niche role in phase contacting operations by providing the platform for nondispersive phase contacting. Simultaneously, they provide much larger value of volumetric mass transfer coefficient (kla, kga or ksa) over a much broader range of phase flow ratios without flooding, loading, channeling, weeping etc. Often such devices are called by a generic designation, membrane contactors; specific designations include membrane absorbers or membrane strippers or membrane extractors. Membrane contactors are gaining industrial recognition and commercial importance because of their unique features. Introduction to membrane 147
Membrane Contactors: Recent Developments - Sirkar
contactors and many of the unique advantages of membrane contactors have been identified in a number of book chapters and reviews [ 1-4]. New applications of membranes as phase contactors are being rapidly developed for a variety of applications. Membranes are being used or studied as contactors for: - physical transfer of one or more components from one phase to another phase in a nondispersive manner; - transfer of one or more components from one phase to the other phase where the transfer is enhanced by chemical reactions; - providing semipermeable barriers for one or more species while allowing contact between two phases for transfer of other species; and - providing complexing/reactive sites on a membrane surface for the transfer of one or more species from one phase, to be followed by elution/stripping. New and improved materials are being used in membrane contactors. As the range of applications expands, the demands on the membrane materials to withstand the conditions of the applications (temperature, pressure, chemical resistance) are different. New polymeric membranes are being used for membrane contactors. In some applications, the membrane contactor surfaces are modified to suit the particular hydrodynamic and chemical conditions imposed. Surface modification by forming a layer of another polymer is being attempted to provide stability to the phase interface and/or to provide chemical and fouling resistance rather than to provide any separation ability. Similarly, the ability of appropriate polymer chains grafted on membrane pore surfaces to provide large phase contact area for transfer of a species is also being explored [5]. As the demands on the membrane contactors grow, there is a need to optimize the contactor design to maximize its mass transfer capabilities. In this regard, new designs are being pursued, while some of the successful designs in other conventional areas are being adapted for new applications. This article attempts to review some of the key developments in membrane contactors in recent years; the emphasis will be on the new directions of membrane-based contacting technologies. The mechanistic aspects of the interface immobilization of two phases are discussed for their relevance to stable operation of membrane contactors. Recent advances in membrane materials and module configurations are discussed for their enhanced features. The applications of membranes as contactors can be broadly categorized as:
1) Liquid-gas contactors 2) Liquid-liquid contactors 3) Supercritical fluid-based contactors 4) Adsorption-based contactors 148
Membrane Contactors: Recent Developments- Sirkar
5) Phase contactors as reactors. Even though immobilized (supported) liquid membranes provide immobilized phase interfaces on each side of the porous/mesoporous/microporous membrane and may qualify as membrane contactors, the issues and the problems associated with their stability and performance are quite different from those relevant for membrane contactors. For this reason, they are not considered in this article. 2. MECHANISTIC ASPECTS OF THE INTERFACE IMMOBILIZATION IN MEMBRANE CONTACTORS Figure 1 presents a schematic representation of fluid-fluid interface in a membrane contactor. Unlike interfaces in conventional contacting equipment, the fluid-fluid interface and its location in membrane contactors is clearly defined. Fluid-Fluid Inerface ..q
Fluid 2
Fluid 1
I,
Polymer Support
Figure 1. Fluid-Fluid Interface in Membrane Contactors In most practical applications, the interface between two phases in a membrane contactor lies on one side of the membrane surface. It also implies that one phase usually occupies the porous region of the membrane. (However, this situation is quite different from "gas membranes" wherein the gas phase fills the pores of the membrane while the desorbing and absorbing solutions are in contact with the two surfaces of the membrane). Under these circumstances, the pore fluid (fluid 1) can not be displaced with the outside fluid (fluid 2) until the pressure difference between the external phase (fluid 2) and the phase in the pores (fluid 1) is greater than the breakthrough pressure (APb). The relationship is governed by Young-Laplace equation: ae~
=
2 y Cos O
(1)
r
where 7 is the interfacial tension between the two fluids, 0 is the contact angle measured from the pore wall to the tangent of the fluid-fluid interface and r is 149
Membrane Contactors: Recent Developments - Sirkar
the pore radius [ 1]. When 0 is zero, the substrate is completely wetted by the liquid. If a finite contact angle is formed (30 ° <0 < 89°), the substrate would be "partially wetting" and for 90 ° and above, the substrate is nonwetting. Moreover, the roughness of the surface plays a significant role in altering the contact angle and hence the breakthrough pressure [6]. A most important aspect in a successful operation of a membrane contactor is the ability of the phase interface to withstand as high a phase pressure difference between the two phases (AP) as possible without displacing the pore fluid. This is achieved mainly by a control of pore size and pore size distribution. As can be seen from equation 1, small and uniform pores in a membrane contactor ensure large and defined breakthrough pressures. The smaller the pore radius, the higher the breakthrough pressure. A narrow pore size distribution is helpful for phase interface immobilization in a given membrane contactor. For this reason, it appears that most membranes used in contactor applications are produced by techniques other than phase inversion and often by controlled stretching of melt extruded sheets/fibers or by thermally induced phase separation (TIPS) process. One aspect of the pore geometry to keep in mind is that the pore geometries in most membranes used for membrane contactors are considered as cylindrical. However, the actual pore geometry usually is non-cylindrical with slit like opening etc. Another important consideration in achieving a stable interface is the interfacial tension between the two fluids in contact. In liquid-liquid phase contacting, some applications involve very low interfacial tensions [7]. This has to be taken into account. For applications involving degassing of liquids, it is advantageous if the liquid does not wet the pores. In that scenario, ~ is equivalent to the surface tension of the liquid. The liquid will not enter the pores as long as the transmembrane pressure is less than the critical entry pressure. Young-Laplace equation govems this behavior as well [8]. An additional polymeric (usually dense and thin) layer can be deposited on the membrane surface; this layer will act as a physical barrier for the liquid to enter the pores while permeable species like volatile organic compounds (VOCs) permeate rapidly through such a coating [9]. Recently surface modification of the membrane has been attempted to address the wet-out problem. A specific example is a hydrogel layer to prevent wetting out of the membrane by oily components in fruit juice concentration by osmotic distillation [10,11]. As the thickness of this additional layer is usually small, the reduction in solute fluxes is expected to be minimum.
150
Membrane Contactors:
Recent Developments -
Sirkar
3. DEVELOPMENTS IN MATERIALS AND MODULE CONFIGURATIONS FOR MEMBRANE CONTACTORS The need for improved chemical resistance, appropriate wetting/nonwetting properties, good flow distribution in the contactor module, higher species transfer coefficients and reduced cost have led to a number of developments in substrate materials, membranes as well as module configurations. The wetting of a porous hydrophobic polymeric material by a liquid is determined by the difference in the surface tension of the liquid (~,) and the critical surface tension 0'c) of the polymer [ 12]. When ~/-~/ctends to zero, the membrane tends to get wetted. It is known that organic solutes in water reduce its surface tension. Further, organic solutes (volatile or nonvolatile) containing both polar and nonpolar functional groups may be adsorbed on the hydrophobic pore surfaces through the nonpolar functional groups and increase the value of ~,~ of the membrane material. Therefore a recent development has adopted the Goretex membranes of porous PTFE for gas absorption with aqueous solution of amines e.g. MEA, which are volatile since the value of ~ for polytetrafluoroethylene (PTFE) is quite low around 18 dynes/cm. Such a membrane does not get wetted with aqueous MEA solutions used to absorb CO2. Modules of such membranes are beginning to be used in large scale [13]. A second development utilizes conventional porous hydrophobic substrates (e.g. polypropylene, polysulfone etc.) but provides a thin coating of a high free volume fluoropolymer on one of its surfaces where the gas-liquid interface is located [14,15]. The nonporous coating on the membrane has sufficiently large free volume dimensions which provide very little resistance to species like H2S, CO2, N2, 02, CH4, H20, etc.; yet it essentially eliminates wetting of the porous substrate. The thin coating/skin of this new composite membrane provides an entirely new functional platform compared to the skin/coating of an asymmetric/composite membrane used in conventional membrane separation processes for separation of gas mixtures and liquid solutions. Since the notion of ozonation of aqueous streams was introduced by us using nonporous and microporous membranes with or without an intervening fluorocarbon phase [16-20], membrane-based ozonation is being practiced commercially. Contacting the highly reactive ozone with water in the presence of pollutants (e.g. semiconductor processing) creates a highly reactive environment. One needs chemical resistance, non-wetting membrane, higher species transfer coefficients amongst others. The new fluorinated polymer-based ozonator (INFUZOR) has all such features [21 ].
151
Membrane Contactors: Recent Developments - Sirkar
The choice of the polymer chosen as the membrane material, either for the substrate and/or for any coating, ultimately determines the long term stability of the membrane contactor for a given application. For most applications where the process streams are chemically compatible with polypropylene (PP) and where the liquid surface tensions are not too low, microporous PP hollow fibers appear to be the membrane of choice. Particular advantages with microporous polypropylene membrane contactors are their compactness (high surface area per unit volume) and their cost. In addition, new membrane materials are being developed to handle low surface tension liquids [22], harsh chemicals such as 03, NH3, HC1 [21] or to prevent wetting [23]. The applications for membrane contactors in medical applications essentially center around introduction of oxygen to blood and removal of CO2. The main additional feature required for these applications is the compatibility of the membrane material with the components of blood. Again, microporous PP is the membrane of choice [24]. Recent developments also focus on silicone-coated microporous fibers for preventing plasma leakage in long term use [25,26] The design of the membrane contactor also plays a significant role in the overall rates of species transfer and the membrane contactor size. For non-blood applications, LiquiCel Extraflow ® modules [22] appear to have become the standard configuration for membrane contactors. Currently transverse flow modules are available mainly with baffles. This configuration increases the mass transfer on the shell side significantly over earlier shell and tube membrane contactors [1]. Some of the new developments in membrane contactors using new membrane configurations [21,23] employ proprietary geometries to maximize the liquid side mass transfer coefficients. For medical applications, the primary issues in module design are low priming volume, low shear rates and low pressure drops [24]. For most applications involving membrane contactors, there are two steps involved: absorption of a solute from one phase into a second phase and desorption into a third phase to concentrate the solute. Examples of such applications are SO2 removal from gas streams [27], heavy metal removal from waste waters [28], isomer separation [29] using liquid membranes. Practically all these applications are pursued using two membrane contactors, one for absorption of the species and another for desorption (concentration). However, the two contactors can be integrated into one device with two sets of fibers. Such a configuration is called hollow fiber contained liquid membrane (HFCLM). An extensive review of this concept and applications was conducted by Sirkar [30]. One limitation of the concept is potential for maldistribution of the shell-side fluid, influencing the mass transfer rates of solute. In spite of the obvious advantages of compactness and economics, commercialization of this concept has not yet taken place. The reasons could be the difficulty in making a
152
Membrane Contactors: Recent Developments - Sirkar
device with two sets of fibers separated at the ports but interspersed intimately and uniformly in the device. 4.
CATEGORIES OF MEMBRANE CONTACTORS
Here we will identify different classes of contacting operations and briefly highlight important applications/developments. The different categories are listed in Table 1. Table 1. Categories of Membrane Contactors Category of Contactor Gas-Liquid Liquid-Liquid Supercritical Fluid-Liquid Liquid-Solid Contactors as Reactors All of the above categories of contacting operations except contactors as reactors are directed toward transferring of species from one phase to the other phase for the purpose of separation/purification whereas the last one utilizes such transport solely for the purpose of carrying out one or more chemical reactions. Generally both phases are mobile in the contactor except in the case of liquid-solid systems.
4.1
Liquid-Gas Contactors Gas-liquid contacting operations and devices are used to accomplish 1) degassing of liquids, 2) gassing of liquids, 3) absorption of gas/vapor, 4) gas/vapor stripping from absorbents, 5) membrane distillation, 6) supported gas membrane-based processes and 7) osmotic distillation. Such operations invariably use a porous hydrophobic membrane, most often in the hollow fiber form for large scale applications; the pores are filled with gas/vapor. In processes 1,2,3, and 4, the gas stream flows on one side of the membrane. The other side has a non-wetting liquid stream. In processes 5,6 and 7, both sides generally have a non-wetting liquid stream except in the processes of vacuum membrane distillation/air-gap membrane distillation. Tranverse-flow membrane contactors built out of hydrophobic microporous PP hollow fibers (Celgard ®) and possessing as much as 130 m E membrane surface area (based on 224640 fibers of O.D. 300 ~tm and I.D. 240 ~tm) in a module of dimensions 10" diameter and 28" long are routinely used to degas ultrapure water streams of dissolved oxygen; the liquid is in transverse flow on the shell side and the fiber bores are subjected to a combination of high vacuum and inert N2 purge [31]. The O2 level in the degassed water may be 153
Membrane Contactors: Recent Developments - Sirkar
reduced to less than 1 ppb. Another application of such devices/processes is to strip the volatile organic compounds (VOCs) from wastewater using air as the sweeping gas [32]. Although degassing of ultrapure water is one of the largest industrial applications of membrane contactors, the biomedical application of these types of hollow fiber membranes in blood oxygenators / artificial lungs of suitable design for removing CO2 from blood continues to be the largest degassing application. Concurrently, supplying 02 to blood via such membranes from a gas stream is the other major goal in the blood oxygenators. (A recent application of this technique is to use an intravenous membrane oxygenator [33] as an intracorporeal artificial lung as opposed to conventional extracorporeal devices). Gassing of streams in the same mode is a major thrust area of industrial membrane contactor applications. Current applications include carbonation of beverages [34], adding hydrogen to wastewater prior to its entering a membrane reactor [35] for removal of nitrates from drinking water via heterogeneous catalysis with catalysts embedded in a microporous polyetherimide membrane. Absorption of gases/vapors into an absorbent liquid in a membrane contactor for industrial gas separation/purification was one of the earliest areas of investigation [36]; the species to be absorbed were primarily CO2, SO2, NH3 amongst others. Absorption of CO2 in aqueous amines solutions containing monoethanolamine (MEA), diethanolamine (DEA) etc. in membrane contactors containing microporous hydrophobic PP fibers with/without a thin silicone coating encountered problems due to the wetting of the pores by the liquid within 7-15 days. Kvaerner in collaboration with W.L.Gore & Co have developed membrane contactors of porous polytetrafluoroethylene (PTFE) such that the membrane pores do not get wetted [ 13]. A large-scale CO2 absorption unit has been put up. These are particularly useful for CO2 removal in offshore applications due to drastic weight and height reductions. Klaassen and Jansen [37] have reported the installation of an ammonia absorption unit at a dyes intermediates production plant of Aliachem in Pardubice (Czech Republic). The ammonia-containing off gas stream is contacted with an aqueous solution which recovers NH3 from the gas stream as an aqueous ammoniacal solution (>20 wt%) to be reused in the dyes intermediate production process. The unit in operation since June 1999 has reduced NH3 emission to the environment by 99% and currently has a capacity to absorb 50 kg/hr. The volume and weight of the membrane absorbers were more than 10 times smaller than conventional absorbers. Additional applications being exploited include SO2 absorption [38]. The above applications involve liquids which are primarily aqueous and under ordinary circumstances many not wet the hydrophobic microporous membrane. Organic liquids (e.g. silicone oil) which are substantially nonvolatile and spontaneously wet the pores of such fibers have been
154
Membrane Contactors:
Recent Developments -
Sirkar
successfully employed as an absorbent for the removal of VOCs from air streams in LiquiCel Extraflow ® membrane contactors under minipilot plant conditions using an actual paint booth exhaust stream [39]. This effort followed the laboratory scale studies of Poddar et al. [40] where they had used both porous membrane as well as a porous membrane with a nonporous silicone coating for absorption of VOCs in an oil. Stripping of absorbed gases from the liquid absorbent at a higher temperature with or without vacuum or an inert gas (e.g. N2) / vapor (steam) can also be implemented using membrane contactors. If a higher temperature is involved, special attention has to be paid to the materials of the membrane and the device. Efforts are ongoing with the microporous Goretex PTFE microporous membranes to strip the CO2 from the amine-absorption solution at a higher temperature [41]. Stripping of absorbed VOCs from the silicone oil absorbent was successfully implemented in a minipilot plant scale using vacuum and hydrophobic microporous polypropylene membranes having an ultrathin nonporous plasmapolymerized silicone membrane coating facing the silicone oil [39]. The coating was highly permeable to the absorbed VOCs but was impermeable to the silicone oil absorbent. Laboratory studies [9] carried out similar stripping successfully at a much higher temperature of 50-70 °C. The first successful laboratory studies of membrane distillation were by Gore [42]. Schneider et al. [43] studied the process of direct contact membrane distillation (DCMD) using larger units made from hydrophobic PP hollow fibers. Lawson and Lloyd [44] had demonstrated using very small fiat membranes that very high water vapor flux is possible in desalination of water by DCMD as much as 2-3 times that in reverse osmosis. However, this DCMD process is plagued by water vapor flux reduction and distillate contamination by brine with time due to wetting of the pores of the hydrophobic porous PP membranes normally employed. Further, high water vapor fluxes have to be achieved in a compact but high surface area contactor device. Vacuum membrane distillation process is not favored as such due to the need for a separate condenser. In DCMD, a hot saline solution produces water vapor which diffuses through the gas filled pores of a porous hydrophobic membrane and then condenses in the cold distillate water stream on the other side of the membrane. The water vapor pressure difference on the two sides of the membrane is the driving force through the gas in the pores, the so called "gas membrane" or the "supported gas membrane" (SGM). When there is a volatile species in an aqueous solution and an aqueous solution without the volatile species on the other side, then a porous hydrophobic membrane may be used to desorb the volatile species from the feed solution and get it absorbed in the receiving solution (for earlier references, especially the studies by Edward Cussler, see [2]). Chlorine dioxide (C102) has been stripped from an aqueous electrolytic 155
Membrane Contactors: Recent Developments - Sirkar
solution of an electrochemical reactor through a highly chemically resistant porous hydrophobic membrane stack and absorbed in the potable water to be disinfected and treated; large-scale plant operation has successfully taken place [45]. Osmotic Distillation is said to occur when the receiving solution on one side of the gas membrane is a highly concentrated saline solution having a vapor pressure of water which is much less than that of an aqueous feed solution on the other side of the membrane. Both solutions are at the same temperature just as in the C102-separation example above and unlike that in DCMD. Consequently, there is a partial pressure-based driving force for water transport from the feed to the receiving solution. Large-scale tests using two 19.2 m 2 LiquiCel ® modules and continuous production have been carried out for fruit juice concentration using such a technique [46]. Wetting of the pores due to hydrophobic components in the feed solution can be prevented by having a hydrogel layer on the surface of the hydrophobic membrane facing the feed fruit juice solution [ 11 ].
4.2
Liquid-Liquid Contactors Membrane-based liquid contactors are primarily employed to extract a nonionic or ionic solute selectively from one liquid phase into another immiscible liquid phase. The phases often encountered in an immiscible twophase system are aqueous and organic. (References to immiscible phase pairs which are organic-organic or biphasic aqueous are available in [1]). A large number of studies have been conducted exploring membrane solvent extraction of metals, organic acids, organic pollutants, antibiotics etc. from an aqueous solution into an organic extractant. Back extraction into an aqueous phase has also been conducted especially for metals, organic acids, antibiotics etc. A number of pilot plant studies were implemented. Almost all such studies employ Celgard hydrophobic microporous PP hollow fibers/Liquicel ® modules with the organic solvent phase in the pores and the aqueous phase outside at a higher pressure [47]. The Liquicel ® modules have the tube sheet made out of a lower Tg olefinic resin and have considerable solvent resistance. However, a number of solvents including chlorinated solvents, dimethyl formamide (DMF), dimethylsulfoxide (DMSO) etc. are not to be used. Back extraction studies of the solute into an aqueous phase also employ the same type of hydrophobic membranes/modules even though hydrophilic fibers with the aqueous phase in the pores is much more efficient from a mass transfer perspective [ 1]. A full-scale membrane-based solvent extraction plant has been operating since 1998 at Kosa, Vlissingen, Netherlands [37]. It employs Liquicel ® modules and an organic feedstock to extract an aromatic compound formed in a reactor
156
Membrane Contactors: Recent Developments - Sirkar
and being discharged in an aqueous stream. The organic feed stock recycles the product back into the reactor. The waste water volume treated is 15 m3/hr. The waste water incineration facility was eliminated resulting in an annual saving of 5 million m 3 of natural gas. An additional example of large-scale hydrophobic microporous membrane-based solvent-extraction involves extraction of Cr 6+ from an aqueous solution into an organic extractant containing a secondary amine present in the pores of a Celgard fiber. The aqueous phase is maintained at a higher pressure than the organic extractant phase to immobilize the aqueous-organic interface in this plant operated by Commodore Separation Technologies, Kennesaw, GA [48,49]. However, there is a novelty in this process wherein the organic extractant phase has dispersed in it drops of the aqueous back extraction liquid which strips the Cr6÷ from the complex into a highly basic stripping solution. A similar technique is being used in large scale for removing zinc and iron from a passivating bath at Rogal, Enschede, The Netherlands [37]. A large enantiomer resolution plant at Tanabe Seiyuku, Osaka, Japan employs lipase enzyme-based resolution of the enantiomers present in an organic stream; the enzymes are reversibly immobilized in the pores of substrate of a hydrophobic asymmetric polyacrylonitrile (PAN)-based hollow fiber ultrafiltration membrane. The skin of this fiber is in the I.D.; an aqueous solution flows through the fiber bore. The aqueous-organic interface is immobilized on the fiber O.D. with the shell-side organic phase outside the pores maintained at a pressure higher than that of the aqueous phase. Total membrane area employed is 1440 m2. This technology is from Sepracor, Marlborough, MA. [50]; however the microporous membrane-based solvent extraction aspect of it was licensed from Hoechst Celanese Inc. (Current successor is Celgard Inc., Charlotte, NC). Additional techniques are being developed wherein liquid-liquid extraction takes place at one side of a porous membrane whereas a sweep gas/vacuum is maintained on the other side. It is called supported liquid membrane pervaporation (SLMPV) [51]. Highly stable technologies are currently being developed in our laboratory since conventional SLMs are generally considered unstable.
4.3
Supereritieal Fluid-based Contactors PoroCrit LLC, Berkeley, CA has started using Celgard fiber-based Liquicel ® contactors and other porous PP fiber-based devices to achieve species transfer from a liquid phase into supercritical CO2 or vice versa. They claim that "the process is conducted with the pressure on both sides of the membrane in the module being essentially the same" [52]. As the two phases flow in a countercurrent fashion in the PP hollow fiber module, they will undergo considerable pressure drops. So there will be considerable pressure difference 157
M e m b r a n e Contactors: Recent D e v e l o p m e n t s - Sirkar
between the two phases along the module length. Such an operation has been successfully used to extract flavors from wines and juices, recover aromas from vegetable and marine oils and nuts and fermentation broths and for industrial organic solvent recovery and recycle [53]. 4.4
Adsorption-based Contactors
When a gas or liquid phase containing a solute is contacted with a packed bed of adsorbents/ion exchange resin beads, adsorption or ion exchange-based transfer of solutes occur to the bed particles. There is significant~igh pressure drop in flow through packed beds, especially of fine particles. Further there is considerable resistance to diffusion of solute molecules (especially larger ones) through the porous resin beads; as a result, the utilization of the ligands etc. in the porous resin beads is low. Porous membranes used in the microfiltration separations and having pores in the range of 0.02-10 gm have been used as a substrate: ligands/biofunctional groups of appropriate types have been grafted onto the surfaces of pores in microfiltration membranes. The liquid phase containing solutes has been passed through the pores in the microfiltration mode of operation. The liquid convection through the membrane pores conveniently brings the solutes directly to the ligands on the pore surfaces eliminating the diffusional resistance encountered in the conventional resin beads. The process of adsorption/complexation takes place throughout the pore volume. Fractional ligand utilization is very high. The solute uptake process takes place rapidly. After some time, the feed flow is stopped and the ligands on the pore surfaces are regenerated by passing an eluent solution since they get saturated. Such devices are called membrane adsorbers but are in effect, liquidsolid contactors. Their application to protein separation is commercially practiced, albeit in a smaller scale. An earlier reference to the literature is available in Thommes and Kula [5]. Such a technique has now been adopted to heavy metal adsorption from waste waters with or without disposable membrane materials [54]. Disposable and cheap membrane materials (e.g. of cellulose) are useful especially if regeneration efficiency for the heavy metals is poor. 4.5
Contactors as Reactors
Whenever a reactor employs two immiscible phases which have to be contacted intimately, a membrane contactor may be employed. The two immiscible phases will flow/be on the two sides of the membrane and appropriate phase pressure differences have to be maintained for nondispersive operation. The reaction may take place in the bulk of one or both phases, at the interfaces on the pore mouths and/or in the fluid layers immediately adjacent to the phase interface [20]. Membrane-based ozonators developed by W.L.Gore and Pall Inc. are commercialized examples of such reactors: ozone-containing oxygen/air flows
158
Membrane Contactors: RecentDevelopments- Sirkar
on one side of the membrane and the water to be treated flows on the other side. The materials of the device are chemically resistant perfluoropolymers. The membranes provide very little resistance to gas transport but do not usually allow liquid to come to the other side of the membrane. This area has an enormous potential for growth in the near future. The limitations are: selection of an appropriate example amenable to reaction in a membrane contactor; chemical resistance of the membrane material; controlling the thermal effects; flow maldistribution etc. Additional examples of membrane contactors employed as reactors are identified below: - Destruction of VOCs in a gas phase by their absorption in a fluorocarbon phase which is supplied with ozone from a gas phase via a separate membrane [55]. - Elimination of mercury from a gas phase by oxidative membrane absorption in an oxidizing aqueous liquid solution of H202/H2SO4 or K2Cr207 o r K 2 S 2 0 8 , K M n O 4 etc. [56]. The review paper by Sirkar et al. [57] provides additional examples especially in the production of chemicals/biochemicals. Amongst these examples are cases of hydrogenation of organic liquids via a porous ceramic membrane reactor which is essentially a membrane contactor. Porous ceramic membranes hold special promise especially due to their excellent thermal resistance. Membrane contactors have also been employed by Porocrit Co. with supercritical CO2 to sterilize juices and liquids by destroying microorganisms and inactivating enzymes [52]. 5.
MASS TRANSFER CORRELATIONS AND CONSIDERATIONS
The resistance to mass transfer of a species in a hollow fiber-based membrane contactor is the sum of three resistances: lumen side, pore and shell side. This can be represented as 1
=
1
1
1
~ + ~ + ~ (2) gtotat kl .... kpore kshett where Ktotal is the overall mass transfer coefficient, klumen, kpore, kshell are the local mass transfer coefficients on the lumen side, membrane pore, and shell side respectively. Depending on the application, one or more of these resistances may become negligible compared to the others. For example, for gas-liquid systems, gas phase resistance is usually negligible compared to liquid phase resistance. So, the side where the gas flows will not offer much resistance to mass transfer and almost all of the resistance is offered by the liquid side. For membranes wetted by the liquid phase, resistance offered by the liquid in the pore is constant and does not change with the hydrodynamics.
159
Membrane Contactors: Recent Developments - Sirkar
In hollow fiber membrane contactors, lumen side mass transfer is usually well defined whereas the shell side mass transfer coefficient is a strong function of module geometry, fiber distribution and the resulting hydrodynamics. Recent efforts to understand the effect of shell side flows on the membrane contactor performance [31,58-60] highlight the complex nature of shell side hydrodynamics. In addition, these efforts may lead to a scaleup parameter for designing membrane contactors for different conditions. While membrane contactors are commercially available with different membrane areas [22] for handling a range of flow volumes, the blood oxygenation devices come in a limited number of sizes. Each device is presumably optimized for the required shear rates, pressure drops etc. 6.
FUTURE DIRECTIONS
Membrane contactors are finding increasing commercial acceptance in recent years. This trend is expected to continue with new applications attempted using membrane contactors. The successful use of membrane contactors will depend on one or more of its unique advantages (small footprint, flexible capacity, broad range of phase flow rates/ratios without associated problems, chemical resistance of materials etc.) which are not offered by conventional contacting equipment. The main class of applications where membrane contactors are used still appears to be for effective gas-liquid contacting in a nondispersive manner using aqueous streams. New applications in gas-liquid contacting and other areas are gaining ground. New materials and geometries are being used for making microporous substrates and membranes. The new materials are pursued mainly for their non-wetting properties and chemical resistance. Characterization of membrane contactors for their mass transfer is still empirical. A better understanding of the hydrodynamics (particularly shell side) and ways to quantify it in terms of mass transfer correlations is needed for effective scaleup of these contactors. Similarly, designs of membrane contactors for conventional applications are limited in number. As the range of applications grows, more designs are to be expected to cater to various aspects of contacting applications. 7.
ACKNOWLEDGEMENT
One of the authors (ASK) wishes to acknowledge many discussions with Dr. P.V. Shanbhag during the preparation of this manuscript.
160
Membrane Contactors: Recent Developments - Sirkar
REFERENCES [1]
Prasad, R. and K. K. Sirkar, Membrane-based Solvent Extraction, in Membrane Handbook, W.S.W. Ho and K.K. Sirkar (Eds.), Chapman and Hall, New York (1992) Chap 41.
[2]
Sirkar, K.K., Other New Membrane Processes, in Membrane Handbook, W.S.W. Ho and K.K. Sirkar (Eds.), Chapman and Hall, New York (1992) Chap 46.
[3]
Reed, B.W.; Semmens, M.J.; and Cussler, E.L. Membrane Contactors in Membrane Separations Technology, Principles and Applications, R.D. Noble and S.A. Stem (Eds.), Elsevier Sci., (1995) 468-498.
[4]
Gableman, A.; Hwang, S.-T. Hollow Fiber membrane Contactors, J. Membr. Sci., 159 (1999) 61-106.
[5]
Thommes, J. and M.R. Kula, Membrane Chromatography- An Integrative Concept in the Downstream Processing of Proteins, Biotechnol. Prog. 11 (1995) 357.
[6]
Myers, D., Surfaces, Interfaces and Colloids. Principles and Applications, VCH Publishers, New York, (1991) 349-356.
[7]
Prasad, R. and K.K. Sirkar, Hollow Fiber Solvent Extraction of Pharmaceutical Products: A Case Study, J. Membr. Sci., 47 (1989) 235-259.
[8]
Kim, B.-S. and P. Harriott, Critical Entry Pressure for Liquids in Hydrophobic Membranes, J. Colloid and Interface. Sci. 115 (1987) 1-8.
[9]
Xia, B., S. Majumdar and K.K. Sirkar, Regenerative Oil Scrubbing of Volatile Organic Compounds from a Gas Stream in Hollow Fiber Membrane Devices, Ind. Eng. Chem. Res. 38 (1999) 3462-3472.
[10]
Mansouri, J. and A.G. Fane, Membrane Development for Processing of Oily Feeds in IMD, Presented at the Workshop on Membrane Distillation, Osmotic Distillation and Membrane Contactors, Cetraro, Italy, July 2-4, 1998.
[11]
Mansouri, J. and A.G. Fane, Osmotic Distillation of Oily Feeds, J. Membr. Sci., 153 (1999) 103-120.
[12]
Zisman, W.A., "Relation of the Equilibrium Contact Angle to Liquid and Solid Constitution", in "Contact Angle, Wettability and Adhesion", Advances in Chemistry Series 43, ACS, Washington, D. C. (1964) 1-51. p 32.
[13]
Falk-Pedersen, O. and H. Dannstrom, Method for Removing Carbon Dioxide from Gases, US Patent 6,228,145, May 8, 2001.
[]4]
Shanbhag, P.V., J.Cissel, M. Gramerb and S. M. Nemser, Bioreactor Oxygenation Applications with CMS Membrane Contactors, paper presented at American Institute of Chemical Engineers, 2000 Spring National Meeting, Atlanta, GA, March 5-9, 2000.
[15]
Stutman, M.B., P.V. Shanbhag, J.J. Bowser and S.M. Nemser, Membrane Contactors Featuring Non-wetting Perfluorinated Membranes for Efficient Gas/Degasification of Process Fluids, paper Presentated at the AIChE Annual Meeting, Reno, NV, November 8 (2001).
[16]
Guha, A.K. and K.K. Sirkar, A Novel Separation-Reaction-Separation System, Paper Presented at the AIChE Annual Meeting, Chicago, IL, Nov. 13 (1990) paper 58g.
161
M e m b r a n e Contactors: Recent D e v e l o p m e n t s - Sirkar
[17]
Guha, A.K., P.V. Shanbhag K.K. Sirkar, C.H. Yun, D.H. Trivedi and D.A. Vaccari, Novel Membrane-Based Separation and Oxidation Technologies, Waste Management 13 (1995) 395.
[18]
Shanbhag, P.V., A. K. Guha and K. K. Sirkar, Single-phase Membrane Ozonation of Hazardous Organic Compounds in Aqueous Streams, J. Hazard. Mater. 41 (1995) 95104.
[19]
Shanbhag, P.V. and K.K. Sirkar, Ozone and Oxygen Permeation Behavior of Silicone Capillary Membranes Employed in Membrane Ozonators, J. Appl. Polymer. Sci. 69 (1998) 1263-1273.
[201
Shanbhag, P.V., A.K. Guha and K.K. Sirkar, Membrane-Based Ozonation of Organic Compounds, Ind. Eng. Chem. Res. 37 (1998) 4388-4398.
[21] [22] [23] [241
Pall Corporation website, www.pall.com (2001). Celgard website, www.celgard.com (2001). Killer, A. and C. Bier, Tubing Unit, WO Patent 01/70375 Voorhees, M.E. and B.F. Brian, Blood-gas Exchange Devices, International Anesthesiology Clinics, 34(2) Spring 1996, 29-45.
[25]
Mueller, X.M., B. Marty, H.T. Teveaerai, P. Tozzi, D. Jigger and L.K. von Segesser, A Siliconized Hollow Fiber Membrane Oxygenator, ASAIO J., (2000) 38-41.
[26]
Kawahito, S., T. Maeda, T. Motomura, T. Takano, K. Nonaka, J. Linneweber, M. Mikami, S. Ichikawa, M. Kawamura, J. Glueck, K. Sato and Y. Nose, Development of a New Hollow Fiber Silicone Membrane Oxygenator: In Vitro Study, Artif. Organs, 25 (2001), 494-502.
[27]
Majumdar, S., A. Sengupta, J.S. Cha and K.K. Sirkar, Simultaneous SO2/NO Separation from Flue Gas in a Contained Liquid Membrane Permeator, Ind. Eng. Chem. Res., 33 (1994) 667-675.
[28]
Guha, A.K., C.H. Yun, R. Basu and K.K. Sirkar, Heavy Metal Removal and Recovery by Contained Liquid Membrane Permeator, AIChE J. , 40 (1994) 12231237.
[29]
Mandal, D., A.K. Guha and K.K. Sirkar, Isomer Separation by a Hollow Fiber Contained Liquid Membrane Permeator, J. Membr. Sci. 144 (1998) 13-24.
[30]
Sirkar, K.K., Hollow Fiber-Contained Liquid Membranes for Separations: An Overview, in Chemical Separations with Liquid Membranes, R.A. Bartsch and J.D. Way (Eds.), ACS Symp. Ser. 642, American Chemical Society, Washington, D.C., Chapter 16.
[311
Sengupta, A., P.A.Peterson, B.D.Miller, J.Schneider, C.W. Fulk Jr., Large-scale Application of Membrane Contactors for Gas Transfer from or to Ultrapure Water, Separation and Purification Technology, 14 (1998) 189-200.
[32]
Mahmud, H., A. Kumar, R.M. Narbaitz and T.Matsuura, A Study of Mass Transfer in the Membrane Air-Stripping Process using Microporous Polypropylene Hollow Fibers, J. Membr. Sci., 179 (2000) 29-41.
[33]
Hattler, B.G., G.D. Reeder, P.J. Sawzik, L.W. Lund, F.R. Waiters, A. Shah, J. Rawleigh, J.S. Goode, M. Klain and H. Borovetz, Development of an Intraveneous Membrane Oxygenator: Enhanced Intraveneous Gas Exchange through Convective
162
Membrane Contactors: Recent Developments - Sirkar
Mixing of Blood Around Hollow Fiber Membranes, Artif. Organs, 18(11) (1994) 806812. [34]
Sirkar, K.K. Membrane Separations: Newer Concepts and Applications for the Food Industry, in Bioseparation Processes in Foods, R.K.Singh and S.S.H. Rizvi (Eds.), Marcel Dekker, New York (1995) pp 351-374.
[35]
Peinemann, K.V., K. Ludtke, D. Fritsch and K. Kleine, Water Purification with Catalytically Active Membranes, paper I1-04-IN presented at the International Conference on Materials for Advanced Technologies, Singapore, July 1-6, 2001.
[36]
Qi, Z., and E.L. Cussler, Microporous Hollow Fibers for Gas Absorption. I Mass Transfer in the Liquid II. Mass Transfer Across the Membrane, J. Membr.Sci. 23 (1985) 321-340.
[37]
Klaassen, R. Jansen, A.E. The Membrane Contactor: Environmental Applications and Possibilities. Environ. Prog. 20(1) April (2001) 37-43.
[38]
Klaassen, R. , P.H.M. Feron and A.E. Jansen, Membrane Gas Absorption: Environmental Applications, Proceedings ICOM'96, pp 916, 1996.
[39]
Majumdar, S., D. Bhaumik, K.K. Sirkar and G. Simes, A Pilot-scale Demonstration of a Membrane-based Absorption-Stripping Process for Removal and Recovery of Volatile Organic Compounds, Environ. Prog., 20 (1) April (2001) 27-35.
[40]
Poddar, T.K., S. Majumdar and K.K. Sirkar, Membrane-based Absorption of VOCs from a Gas Stream, AIChE J., 42 (1996) 3267-3282.
[41]
Herzog, H. and O. Falk-Pedersen, The Kvaerner Membrane Contactor: Lessons from a case study in how to reduce captm'e costs, paper presented in 5th International Conference on Greenhouse Gas Control Technologies, Australia, 13-16 August, 2000.
[42]
Gore, D.W., Gore-Tex Membrane Distillation, in Proc. 10th Annual Convention of Water Supply Improvement Association, Honolulu, Hawaii, July 25-29 (1982).
[43]
Schneider, K., W. Holtz, R. Wollbeck and S. Ripperger, Membranes and Modules for Transmembrane Distillation, J. Membr. Sci. 39 (1998) 25.
[44]
Lawson, K. and D.R. Lloyd, Membrane Distillation. II Direct Contact MD, J. Membr. Sci. 120 (1996) 123.
[45]
Muttoo, T. and D. Dean, Development of Chlorine Dioxide Gas Transport Contactor (ERCO R 101 Generator) for Use in Drinking Water Disinfection, Paper 1-2 presented at the 12th Annual NAMS Meeting, Lexington, KY, May 16, 2001.
[46]
Hogan, P.A., R.P. Canning, P.A. Peterson, R.A. Johnson and A.S. Michaels, A New Option: Osmotic Distillation, Chem. Eng. Prog., July (1998) 49-61.
[47]
Kiani, A.K., R.R. Bhave and K.K. Sirkar, Solvent Extraction with Immobilized Interfaces in a Microporous Hydrophobic Membrane, J. Membr. Sci. 20 (1984) 125.
[48]
Ho, W.S.W. and T.K. Poddar, New Membrane Technology for Removal and Recovery of Chromium from Waste Waters, Environ. Prog., 20(1) April (2001) 44-52.
[49]
Ho., W.S.W. and T.K. Poddar, New Membrane Technology for Removal and Recovery of Metals from Waste Waters and Process Streams, Paper Presented at the Topical Conference on Membrane and Extraction Science and Technologies for Environemental Applications, AIChE 2000 Spring National Meeting, Atlanta, GA, March 5-9 (2000).
163
Membrane Contactors: Recent Developments - Sirkar
[50]
Lopez, J.L. and S.L. Matson, A Multi-phase Extractive Enzyme Membrane Reactor for Production of Diltiazem Chiral Intermediate, J.Membr. Sci., 125 (1997) 189.
[51]
Qin, Y., J. P. Sheth and K. K. Sirkar, Supported Liquid Membrane-Based Pervaporation for VOC Removal from Water, Ind. Eng. Chem. Res. 41 (2002) 34133428.
[52]
Robinson, J.R. and M.J. Sims, Method and System for Extracting a Solute from a Fluid Using Dense Gas and a Porous Membrane, U.S. Patent, 5,490,884, February 13 (1996).
[53]
Towsley, R.W., J.A.Turpin, M. J. Sims and J. R. Robinson, "Porocritical Fluid Extraction using COE for Industrial Organic Solvent Recovery and Recycle", Paper Presented at CISF-99, Fifth Conference on Supercritical Fluids and their Applications, June 13-19, Garda (Verona), Italy (1999).
[54]
Hestekin, J.A., L.G. Bachas and D. Bhattacharyya, Poly(amino acid) Functionalized Cellulosic Membranes: Metal Sorption Mechanisms and Results, Ind. Eng. Chem. Res. 40 (2001) 2668-2678.
[55]
Shanbhag, P.V., A.K.Guha and K.K. Sirkar, Membrane-based Integrated AbsorptionOxidation Reactor for Destroying VOCs in Air, Env. Sci. Tech. 30 (1996) 3435.
[56]
van der Vaart, R., J. Akkerhuis, P. Feron and B. Jansen, Removal of Mercury from Gas Streams by Oxidative Membrane Gas Absorption, J. Membr. Sci., 187 (2001) 151-157.
[57]
Sirkar, K.K., P.V. Shanbhag and A.S. Kovvali, Membrane in a Reactor: A Functional Perspective, Ind. Eng. Chem. Res., 38(10) (1999) 3715-3737.
[58]
Lemanski, J. and G.G. Lipscomb, Effect of Shell-side Flows on Hollow-Fiber Membrane Device Performance, AIChE J. 41 (1995) 2322-2326.
[59]
Lipnizki, F. and R.W. Field, Mass Transfer Performance for Hollow Fiber Modules with Shell-side Axial Feed Flow: Using an Engineering Approach to Develop a Framework, J. Membr. Sci. 193 (2001) 195-208.
[601
Schoner, P., P. Plucinski, W. Nitsch and U. Daiminger, Mass Transfer in the Shell Side of Cross Flow Hollow Fiber Modules, Chem. Eng. Sci., 53 (1998) 2319-2326.
164
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 8
Membrane Aromatic Recovery System (MARS) - A new process for recovering phenols and aromatic amines from aqueous streams .
•
F. C. Ferreira*, A. Llvlnt[ston
+q¢
, S. Han*, A. Boam +, S. Zhang
+
*Department of Chemical Engineering and Chemical Technology, Imperial College of Science, Technology and Medicine, Prince Consort Road, London SW7 2BY, Unity Kingdom; +Membrane Extraction Technology Ltd, Imperial College, London SW7 2BY, UK Corresponding author: Tel +44-20-75945582; fax +44-20-75945629, E-mail address: [email protected] 1.
INTRODUCTION
Phenolic compounds (chemicals such as phenol and its derivatives) are used in phenolic resins, polycarbonates, biocides and agrochemicals. Aromatic amines (chemicals such as aniline and its derivatives) are used in a wide range of consumer products, including polyurethane foam, dyes, rubber chemicals and pharmaceuticals. The factories that manufacture and/or use these types of chemicals often create aqueous waste streams containing significant concentrations (0.1-10wt%) of these chemicals. Both aromatic amines and phenolic compounds are toxic and many of them are also carcinogenic. Tightening legislation in many countries calls for dramatic reductions in emissions of these species. A variety of processes have been proposed for treatment of these aromatic amine or phenolic compound containing wastewaters. Off-site disposal (landfill, deepwell injection) and biodegradation result in the compounds, which have typical values in the range US$0.75-US$20 per kg, being lost. These compounds have high boiling points and low vapour pressures. Hence, processes that rely on liquid-gas phase transition, such as distillation and pervaporation, have high-energy requirements. The use of adsorbents, such as activated carbon [1-4] or resins [58], is usually expensive due to difficulties and complexity in the regeneration stage. Problems associated with the use of solvent extraction [9-13] arise with phase separation [14] and contamination of the wastewater with solvent [15] due to the intermediate polarity of the compounds, which require moderately water soluble solvents. 165
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
Emulsion liquid membranes [16-19] and surfactant liquid emulsion membranes [20-22]systems were used to extract aromatic acids and bases from wastewaters. In particular Li and co-workers [23] used an emulsion liquid membrane containing caustic as reactive agent to increase phenol mass transfer. The main drawback of the use of liquid membranes is its inherently instability due to the leakage of the internal phase to the wastewater and the easy swelling of the liquid membrane by the wastewater phase. To overcome this problem, porous membranes have been used to support liquid membranes [24]. To avoid breakthrough of the immobilized phase in the pores to the aqueous phase, Ho and et al. [25-29] report development of a supported polymeric liquid membrane, in which the pores of a microfiltration or ultrafiltration membrane are filled with polyamphiphilic polymeric (oligomeric) liquids. These membranes were successful used, for phenolic compounds extraction from wastewater to an alkaline stripping solution at lab scale. However the use of liquid membranes, including supported polymeric liquid membrane, always requires an additional step for membrane phase regeneration, which often requires difficult demulsifications and/or the use of additional chemical agents [16]. The use of a non porous solid dense membrane avoid this step, hence aromatic acids and bases extraction from wastewaters have been tried in systems such as pervaporation [30,31 ], and aqueous/membrane/organic perstraetion [32]. The use of an acid-base reaction to maintain the driving force for phenol [33] or aniline [34] dialysis through several membranes was demonstrated using, sodium hydroxide or sulphurie acid respectively. In these studies, polymethylsilane-polyearbonate eopolymer appears to be the more promising membrane, although the values for permeability appear lower than those for reported for PDMS (phenol 4.5x10 -11 mE.s-1 [35] o r 2 . 5 x 1 0 -11 mE.s-l[36] against lxl0 "11 mE.s-1 [33] and aniline 20.9x10 -11 mE.s-1 [37] against 2.5x10 "11 mE.s 1134]). In the work reported by Klein et al. [33,34], the molar concentration of phenolate or anilinium is kept much lower than the concentration of base or acid used as the stripping solution in order to ensure a high driving force, and the stated intention is to dispose of the extracted aromatics. In this present report, working under conditions where the molar concentration of the phenolate or anilinium is carefully maintained near to the concentration of acid or base, we are able to show that the recovery of relatively pure phases of phenol and aniline is possible. This chapter describes the Membrane Aromatic Recovery Systems (MARS), a stable membrane process able to recover valuable aromatic amines [37] and phenolic compounds [38] from aqueous streams and its successful application at pilot scale for aniline recovery. The aromatic compound is extracted from the k,wastewater to a stripping solution, through a nonporous membrane selectively permeable to aromatic in its non ionic form and
166
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
impermeable to charged species. The driving force is maintained using an acidbase reaction in the stripping solution, where the aromatic is accumulated in its ionic form until a high enough concentration to allowed aromatic recovery by neutralization and consequentially phase separation. The MARS process is shown schematically in Figure 1.
1.1
MARS Operating Principle Wastewater out stripped of dissolved Amines (phenols)
HCI (NaOH)
NaOH (HCl)
R-NH4 +
(R-O-)
,~!~i ~:~:, ,,i~,
R-NH4+ (R-O-)
Nonporous Membrane
~: ~
R-NH 3 recovered amines (ROH recovered phenols)
Saline aqueous phase Wastewater containing dissolved amines (phenols) Figure 1" Schematic diagram of MARS process showing operating principles. Diagram show configuration for aniline recovery- figures in brackets show phenol recovery configuration.
1.1.1 Extraction Stage The aromatic compound is continuously extracted from the wastewater through a nonporous membrane separating layer into a stripping solution, where pH is controlled. The stripping solution is acidic for aromatic amines, and alkaline for phenolic compounds. In the stripping solution, acid-base reaction takes place and the aromatic molecule is converted into an ionized form (anilinium chloride or sodium phenolate). The membrane separating layer is hydrophobic; therefore charged molecules (such as ionized aromatic rings) cannot cross back through the membrane into the wastewater. The driving force for mass transfer of the aromatic acid or base through the membrane from the wastewater is maintained because the aromatic reacts to form its ionised form, a chemically distinct species. Note that the pH control operates such that each mole of aniline or phenol diffusing across the membrane requires one mole of either acid or base for neutralisation in the stripping solution. 167
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
1.1.2 Recovery Stage The stripping solution is periodically collected and pH is adjusted in order to recover the nonionic form of the aromatic molecule by acid-base reaction (i.e. alkaline conditions for amines and acid conditions for phenolics). Anilinium and phenate (the ionized forms of the aromatic) have virtually infinite solubility in water. However the solubility of nonionic forms of aromatic amines and phenols are usually less than 5 wt%. Hence, when the aromatic ion (highly concentrated in the stripping solution) is neutralized to the nonionic form, the resulting concentration greatly exceeds the aqueous solubility limit, and the solution separates into two phases: an aromatic-rich phase and a saline aqueous underlayer. NaC1 is a by-product in both cases, but this salty underlayer can be simply recycled to the wastewater feed as shown in Figure 1. In many of the industrial wastewater samples studied by Membrane Extraction Technology Ltd, we have found that the salt created by the acid-base recovery mechanism is negligible compared to the concentration of salt already in the waste stream. For example, the MARS process typically produces around 0.5 g of salt per g of organic recovered. In a wastewater containing 10 g L -1 of aromatic, the process will remove the aromatic and add around 5 g L -~ of salt. Many chemical industry wastes contain in excess of 100 g L 1 inorganic salts which arise in reaction processes, so the additional salt is not usually an environmental issue. 2.
MATERIALS AND METHODS
Analytical techniques employed for concentration determinations are described elsewhere[37,38]. The membrane used was silicone rubber tube (70 controlle Pressure gauge(~ I
pH / probe
Acid or caustic for pH control stewater out ~
Wastewater container
Stripping Membrane tank with solution stirring and heating overflow
Membrane coil
Figure 2: Laboratory and pilot plant continuous extraction configuration, with wastewater inside membrane tubes.
168
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
wt% poly(dimethylsiloxane) or PDMS and 30 wt % fumed silica), supplied by Silex Ltd. The membrane tubes used in all the laboratory experiments and pilot trials were 3 mm internal diameter and 0.5 mm wall thickness, with the exception of results shown in Figure 4, where the membrane tube had 3mm internal diameter but 0.35 mm wall thickness. The continuous flow configuration used for extraction stage laboratory experiments and continuous extraction trials at Solutia, UK is shown in Figure 2. The batch extraction configuration used for trials at Solutia UK is shown in Figure 3. The membrane tank solution temperature was kept at 50°C in all laboratory experiments and pilot trials. pH controller pH probe
Stripping solution recirculation
Wastewater Membrane coil
Acid for pH control
Stripping solution Membrane tank with vessel, level rises stirring and heating as aniline transfers Wastewater outside and acid added membrane Figure 3" Pilot plant batch extraction configuration, with wastewater outside membrane tubes. 3.
LABORATORY RESULTS
3.1
Phenol and Aniline Extraction
In phenol experiments a 28 m length membrane tube was used, a synthetic wastewater containing 10g.L -1 phenol was fed to the process, and a 12.5wt% NaOH solution was used to control the pH of a stripping solution between pH 11-13. Figure 4 shows phenol inlet and outlet concentrations from the membrane tube and total phenol (phenol plus phenate) concentration in the stripping solution over time. It can be seen that total phenol concentration in the stripping solution increases from 0 g.L -1 to 200 g.L ~. Operation at steady state is shown in Figure 5, in which the total phenol concentration (phenol and phenate) in the stripping solution was 242 g.L-1, three times higher than the phenol solubility in water and twenty five times higher than phenol concentration in the wastewater. The measured phenol concentration corresponds well to the value
169
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
oo tO
~
30
i
25 =
i
8,--. =o
-
160~ Q-
i
O
-,~ "~15 o
200
~10
~
,,
;H-111iI RH- !
80 ~ .-
8e- ~-~40
5
E
~ o 120 8 ~
8 ~
u
(1)
e-
-
0
10
0
I
-o- inlet ~
20 30 Time (day)
40
outlet -,,- stripping solution I
Figure 4: Evolution of phenol concentrations in the inlet, outlet and stripping solution over time in the MARS process operated continuously at laboratory scale: non steady state (reprinted with permission from J.Membr.Sci[38], page 224).
calculated for steady state assuming 12.5 wt% NaOH is neutralised by phenol to pH 13 of 227 g L 1. Figure 4 shows outlet concentration increasing with increasing phenol concentration in the stripping solution. From calculations based on the pKa of phenol, at pH 11, 9% of the phenol in the stripping solution is present as nonionic phenol and 91% as phenate ion. Therefore, when the concentration in the stripping solution increases the nonionic phenol u} to
300
15-
t-
_J
Q.
t--
250 ==
=o310-
200
"5 .o_
8 ~
-150 pH=13
~
_
5-
-
_.¢ _c
100
50 0
|
0 o inlet
5
i
|
10
~
i
,
==
G) O
Q. 0.
~ o
"r.
~
""
I--
c
|
15 20 Time (day)
r~ outlet
~E
25
× stripping solution
I
I
Figure 5: Evolution of phenol concentrations in the inlet, outlet and stripping solution over time in the MARS process operated continuously at laboratory scale: steady state (reprinted with permission from J.Membr.Sci [38] page 228).
170
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
concentration becomes high enough to have a significant negative effect on the driving force. At pH 13 only 0.1% of the phenol is in the nonionic form, the driving force is restored, and outlet concentrations drop to lower values, as shown in Figure 4. In aniline experiments, performed at laboratory scale using a 10.5 m membrane tube length, a synthetic wastewater containing 5 g.L-1 aniline was fed to the process, and 10.5 wt% HC1 solution was used to control pH in the stripping solution at pH 1. - 250
108
-
-
(D "7 200 ~._ ~.-J 0
L
15o
~6
loo 2-
...._~ ~._..; 50
'
"~. 0 ~ ' =Q..
~
2.4 2.0
0
0 0
2O
40
60
80
100
I Time (day) o inlet "-" outlet ~stripping solution - - --Predicted value at steady sate based on HCI concentration fed Figure 6: Evolution of aniline concentrations in the inlet, outlet and stripping solution over time in the MARS process operated continuously at laboratory scale.
Figure 6 shows the aniline inlet and outlet concentrations in the membrane tube, and total aniline (anilinium and aniline) concentration in the stripping solution, increasing over time. At steady state total aniline concentration in the stripping solution was 218.5 g.L-1, six times higher than the aniline solubility in water and forty four times higher than aniline concentration in the wastewater. These values illustrate the potential of MARS as a concentrating process. The steady state aniline concentration also compares well to the value obtained from stoichiometric neutralisation of 10.5 wt% HC1, 212 g L -1.
After total concentration in the stripping solution reached a steady state, the pH set point was varied in order to study the effect of pH on mass transfer flux (after day 85). The higher the pH in the stripping solution, the higher the percentage of nonionized aniline, and therefore the lower the driving force, resulting in lower mass transfer rates. It can be seen in Figure 6 that an increase in pH of the stripping solution clearly corresponds to an increase in outlet aniline concentration.
171
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
In the laboratory experiments a short membrane tube was used in order to quantify the outlet concentrations and estimate the overall mass transfer coefficient. Temperature was used to enhance mass transfer. The overall mass transfer coefficient obtained at 50°C across the 0.5 mm thickness membrane, was 1.4x10-7 m.s-1 for phenol and 4.8x10 -7 m.s-1 for aniline. Each mole of aromatic extracted from the wastewater requires one mole of acid or base to be fed to the stripping solution for neutralisation, ie each mole of aniline extracted requires one mole of HC1 and each mole of phenol extract requires one mole of NaOH. These ratios were tracked during the experiments and were usually close (within + 10%) to 1.0.
3.2 Recovery Stage Aromatic recovery was performed in batch. The aromatic compound accumulated in the stripping solution in ionized form. The overflow of the stripping solution, enriched with aromatic in the ionized form, was fed to the recovery stage and neutralized in order to obtain the neutral form of the aromatic. Neutralization was performed by adding 37% HC1 solution to the sodium phenate solution and 50% NaOH solution to the anilinium chloride solution. The aqueous solubility of the aromatic in the neutral form is low (8.2 wt% for phenol and 3.4wt% for aniline), hence when neutralization is completed the solution separates into two layers: a top phase that is the final product, and a underlayer aqueous phase that contains resultant NaC1 and traces of aromatic.
80
n Phenol in the aqueous saline phase • .,.Phenol,in the orqanic phase /otat pnenot Ted-to recovery stage
,° t
60
50 -~
40
a.
20
30 10 0
1
Overflow period (day)
Figure 7: Mass balances for phenol batch recoveries from laboratory data (reprinted with permission from J.Membr.Sci. [38], page 225).
172
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
Mass balances for recovery of phenol are shown in Figure 7. The organic phases were composed of aromatic compound and water. With phenol, the organic phase consists of 86.5% phenol and 13.5% water and, with aniline, 96.5% aniline and 3.5% water. No other organic was detected by GC analysis in the laboratory experiments. Recovery efficiency is defined as the ratio of mass of aromatic in the organic phase over the mass initially present in the stripping solution. Recovery efficiencies of over 92% were achieved for both systems. The balance of aniline or phenol is present dissolved in the aqueous layer resulting from the recovery process. The NaC1 in this saline aqueous underlayer is a by-product, which is responsible for a salting out effect that limits the solubility of the aniline or phenol in the aqueous layer. Hence, the higher the NaC1 concentration in the aqueous layer, the lower the aromatic concentration in the saline underlayer and the lower the water concentrations in the organic phase. Therefore the higher the NaC1 concentration, the higher the aromatic concentration in the final product and the higher the recovery efficiencies become. Figure 8 shows the salting out effect, with phenol aqueous solubility decreasing from 83 to 15.7 g.L1 as NaC1 concentration increases from 0 to 200 g.L 1 and phenol concentration in the organic layer increasing from 70.5 to 87.5 wt% for the same range of NaC1 concentration. Clearly there is a benefit to operating by adding as high a concentration of acid or base to the stripping solution as possible. However, this is balanced by the overall concentration of phenol or aniline rising and increasing the concentration of unionised aromatic in the stripping solution, leading to lower mass transfer fluxes. .~"~ 90 80 ; 70 c
k \
e Solubility o Phenol in organic _ phase
1 100 | | ~"
60
90
50
Q..
.~
~. 4o
"6
80 o
30
_c-
~= 20 10 0
0 ¢-
•c--
,
,
,
100
200
300
70 n 400
NaCI in water (g.L -~)
Figure 8: Phenol concentration in the aqueous phase, water and in the organic-rich phase as a function of NaC1 concentrations in the aqueous phase (reprinted with permission from J.Membr.Sci. [38], page 227). 173
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
3.3
Other Aromatics
In order to illustrate the applicability of MARS to recover compounds other than aniline and phenol, the overall mass transfer coefficients of a range of compounds were measured and results are shown in Table 1. Three independent non-coiled (randomly spaced) membrane tubes were submerged in a single stripping solution vessel, pH was kept below 1 using HCI solution for aromatic bases and over 14 using NaOH for aromatic acids. Under these conditions the unionised aromatic compound concentration in the stripping solution was assumed to be zero. In each run, three aromatic compounds were tested. The overall mass transfer coefficient for aniline and phenol in Table 1 are around twice (respectively 1.7 and 2.2 times higher than) the average value obtained in the laboratory MARS extraction. This is probably since in the mass transfer test the membrane was loosely arranged whereas the membrane in the MARS process was tightly rolled around a support, reducing the effective membrane-stripping solution interfacial area. Dimethylamine,4-nitrophenol,and 2,4,6 tris(dimethylaminomethyl)phenol exhibit very low overall mass transfer coefficients and their extraction from wastewaters will be difficult using silicone rubber tubes as membranes. Mass transfer rates for phenol, 4-nitroaniline, hydroquinone and pyridine have Table 1" Overall mass transfers for different compounds at 50°C across a 0.5 mm wall thickness silicone robber membrane Compounds
Kov x 107 (m.s-1)
Compounds
Kov x 107 (m.s"1)
Aniline 4-chloroaniline 2,4-chloroaniline 4-nitroaniline 4-fluoroaniline 2,4-fluoroaniline triethylamine dimethylamine benzyldimethylamin dicyclohexylamine
8.2 11.6 6.36 4.34 10.27 9.33 20 0.72 18 16.5
phenol 4-chlorophenol 2,4-dichlorophenol 4-nitrophenol 4-cresol hydroquinone Pyridine 2,4,6- tris (dimethy laminomethyl) phenol
3.1 9.3 14.7 0.48 7.04 2.37 4.4 0.64
intermediate values, and the extraction of these three compounds from wastewaters by MARS technology is possible using silicone rubber as membrane, but will use relatively large membrane areas. All other compounds tested have higher overall mass transfer coefficients, and so they can be
174
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
relatively easily removed from wastewaters by MARS technology using silicone rubber tubes as membranes. 4. PILOT PLANT RESULTS FOR ANILINE The MARS process has been scaled-up for recovery of aniline at Solutia, UK by Membrane Extraction Technology Ltd (UK). Further pilot trials are currently underway with phenol. Production of 4-nitrodiphenylamine (4NDPA) at Solutia UK results in a wastewater containing an aniline concentration that varies from 4.1 to 9.5 g.L -1, with an average value of 6 g.L 1. A pilot scale MARS unit was operated on site and recovered a total of 120 L of aniline from 29,500 L of wastewater over a period of some two months intermittent operation. 4.1
Continuous Extraction
Initially the pilot plant was configured for continuous operation, as for laboratory scale work (Figure 2), in which the wastewater flowed inside the membrane tube and aniline was accumulated out side of the tube, in an acidic stripping solution. At lab scale a membrane area of 0.099 m 2 (a single 10.5 m length tube) was used at a flow rate of 5.5 L.day -1, immersed in a membrane tank holding 1.5 L of stripping solution with pH controlled at 1. The pilot plant used a total area of 45 m" (48 tubes with 100m length), immersed in a 1000 L membrane tank and a pH in the stripping solution of 1.5. Flowrate was between 200-500 L.day 1. The temperature of the stripping solution was kept at 50°C and an 11% HC1 solution was fed to the stripping solution to control pH. Quantities of the stripping solution (80 L in the pilot plant against 0.25L in the lab scale) were periodically sent to batch recovery, which was achieved by addition of NaOH (33 wt% in the pilot plant, against 50 wt% in the lab scale). 12 11 10
60 - * - - i n l e t - - e - o u t l e t -e-stripping solution I
50
a) .-.. 9
=-~'
~
= o
6~
40 "~ ~, 88
8
"-"
7
o
15
30
4
20
2
10 ~-';
o ._ -~. m
~
1 0
.
.
.
.
,
5
.
.
.
.
.
.
10
.
.
.
.
.
15
.
.
.
.
.
.
.
.
,
20 25 Time (days)
.
.
.
.
.
.
30
.
.
.
.
.
.
35
.
.
i
0
40
Figure 9: Evolution of aniline concentrations in the inlet, outlet and stripping solution over time in the MARS process operated continuously at pilot plant scale.
175
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
Figure 9 shows aniline inlet and outlet concentrations, and total aniline concentration in the stripping solution, over time during continuous pilot plant operation. The average remo~cal achieved was 95% over the whole operating period, which corresponds to an overall mass transfer coefficient of 2.7 x 10-7 m.s 1, somewhat lower then the 4.8x10 -7 m.s -1value obtained at lab scale. 4.2
Interruption of Continuous Extraction The continuous operating period ended after day 36 when the backpressure of the wastewater being pumped down the membrane modules rose suddenly to over 0.5 bar. This caused a relief valve in the wastewater line to open and flow through the membranes ceased. Examination revealed that the membranes, and pipework downstream from the membranes, had blocked. The membrane tubes were removed from the membrane tank and some of them were opened up to reveal crystals of a black material, which were obviously blocking the flow. This black material dissolved readily in dichloromethane. When injected onto the GC, it was identified as 84% (by peak area) 4NDPA. The blockage corresponded to a dark-coloured organic phase being present in the wastewater due to insufficient upstream phase separation. This organic phase is probably organic product phase from the 4NDPA reaction. This mixture of aniline, 4NDPA and other compounds was initially liquid (aniline acts as a solvent for 4NDPA). In passing through the membranes, the aniline was stripped leaving the 4NDPA originally present in the liquid mixture as crystals in the membrane tubes. Filtration methods to avoid entry of this organic phase in the membrane tube were attempted, however frequent cleaning and filter changes are required and the filters were unable to prevent the passage of liquid solvents, and so were not sufficiently reliable or effective. 4.3
Batch Extraction To overcome these practical challenges, the pilot plant was reconfigured to extract aniline from wastewater batches, with the acidic stripping solution flowing inside the membrane tube and the wastewater in the membrane tank as shown in Figure 3. 770 L of wastewater was treated in each run. The same membrane area was used for batch and continuous extraction, and the membrane tank was kept at the same temperature (50°C). pH in the stripping solution was controlled at 1.5 using a 11% HC1 solution. A re-circulating flow rate of 80 L.h -1 was used in the stripping solution. In this configuration the wastewater, with the tar or organic phases when they are present, were pumped directly to the membrane tank without filtration. Membrane blockage was avoided since the solution inside membrane tube is the stripping solution in which no blockage material was present. Figure 10 shows aniline concentration initial and final
176
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
100
~'6o
- BO
°ii iI
r~
4"~
!
I
20
0
1 2 3
4
5 6
7 8
9 10 11 12 13 14 15 16 17 18 19 20 21
/ A n i l i n e inlet D Aniline outlet--*-Aniline removal
Batch Number
Figure 10" Batch operation pilot plant data: aniline initial and final concentration in each batch, aniline removal in each batch. concentration for each of the batch runs and the respective removal efficiency. Batch extraction was on average 22 hours in duration. An average of 90% removal over all the 21 batches was achieved. Aniline removal can easily be increased by increasing the membrane area/ membrane tank volume ratio (values of 100 m2m3 are easily obtainable). The stripping solution aniline concentration increases from 80g.L -1 to 160 g.L -1. The theoretical steady state of 240 g.L -1 which corresponds to 11% HC1 solution used was not reached in the time span of the trial. An average value for overall mass transfer coefficient of 4.2 xl0 -7 m.s 1 was calculated during the pilot plant batch extractions. This value is comparable to the average value obtained in laboratory scale continuous extraction (4.8x107 m.s-1), and higher than 2.7 x 10-7 m.s 1, the value obtained during continuous operation of the pilot plant. The mass transfer rate in the continuous pilot plant most probably decreased due to partial blockage of some tubes by tarry solids.
4.4
Recovery Stage
Aniline recovery from the acidic stripping solution enriched in anilinium chloride was achieved by neutralization using a 33% NaOH solution. A total of 128 L of aniline rich phase was recovered with a composition of 95 wt% aniline, 2.4 wt% Toluidine and 2.6 wt% water. The average aniline concentration in the salt under layer is 4 g.L -1. The purity of aniline obtained was high enough to allow it to be re-used in 4NDPA production.
177
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
5. PROCESS ECONOMICS Aniline price at time of writing is about $0.75 Kg -~, and hence it is not one of the more valuable aromatics in the market. Nevertheless due its Item
Comment
MARS Plant
Capital Charge Factor = 0.3 $45,000
Annual Cost (Benefit)
Membrane Replacement 2 year lifetime
$7,500
Acid (33% HC1)
17 ty-l@ $105 t-1
$1,785
Base (50 %)
12 t y -1 @ $210 t -1
$2,520
Power ( 3 kW)
8000 h y-~ @ $0.075
$1,800
Steam
82 t y-1 @ $15 t-1
$1,230
Labour
10% of one staff
$11,250
TOTAL
BENEFIT
($41,415)
Table 2: Case S t u d y - Process economics of MARS process using fluoroaniline as an example.
environmental impact, aniline removal is necessary and MARS provides a considerable economic advantage to alternative processes, although details cannot be provided here for commercial reasons. However, an example of MARS to recover a more valuable chemical is given in the Case Study contained in Table 2. The commercial value of fluoroaniline is ten times the aniline value (around $7.5 kg-1). In the example given the treatment of 10 m3 d-1 wastewater with a 5 g L 1 fluoroaniline concentration is considered. The MARS process in this case is able to deliver a net benefit through the value of the recovered fluoroaniline. 6. CONCLUSIONS MARS is a novel process coupling detoxification and recovery. It is capable of achieving high recovery efficiencies, and producing a relatively pure stream of recovered organics. At the laboratory scale, MARS has proven to be a successful process for removing and recovering aniline and phenol, with a recovered purities of 95% and 86.5% respectively, pH is an important parameter which controls the driving force across the membrane and consequently the removal efficiency of the aromatic from the wastewater. Anilinium and phenate salts were accumulated in the stripping solution to high concentrations, allowing further recovery by neutralization. The strip feed solution (HC1 or NaOH) concentration is a key parameter that controls the aromatic concentration at steady state in the stripping
178
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
solution. The salting out effect, resulting from high concentrations of NaC1 in the aqueous phase after neutralization, has a positive effect on the phase separation, decreasing the water concentration in the organic phase and the aromatic concentration in the aqueous phase. At pilot plant scale MARS was proven able to recover aniline from an industrial wastewater in a good purity. MARS was shown to be easily scaled up based on membrane area. Using different configurations, MARS was adapted to deal with a common problem in membrane technology, that is membrane blockage by tarry solids or organics precipitation. The MARS process can utilise very simple nonporous rubber tubes as membranes. Mass transfer is relatively low, but these membranes are relatively cheap. This is a key area where we expect to improve the process over the next months. MARS has low energy requirements because it exploits the acid-base functionality of aromatic acids and bases to produce a driving force based on the chemical energy contained in NaOH or HC1. The process can be carried out at conditions of pressure and temperature which are near ambient throughout all items of equipment. MARS does not rely on volatility of organics (or any phase transition), so can recover organics which membrane technologies such as pervaporation cannot reach. Finally, MARS has been shown to be promising for industrial application to recovery of a larger range of aromatics than phenol and aniline. Examples include amines, phenolics and pyridines. REFERENCES [ 1] [2] [3] [4] [5] [6] [7] [8] [9] [10]
ChiangP.C., Chang E.E., Wu J.S.: Comparison of chemical and thermal regeneration of aromatic compounds on exhausted activated carbon. Wat. Sci. Tech. 35-7 (1997) 279. OrshanskyF., Narkis N.: Characteristics of organic removal by PACT simultaneous adsorption and biodegradation. Wat. Res. 31-3 (1997) 391. Cha T., Glasgow: Sorptive reclamation of phenol from coal conversion wastewater. Ind. Eng. Chem. Res 22 (1983)198. Pahari P.K., Sharma M.M.: Recovery of heterocyclic amines from dilute aqueous waste streams. Ind. Eng. Chem. Res.30 (1991) 1980. Fox C.R. Plants uses prove phenol recovery with resins. Hydrocarbonprocessing 11 (1978) 269. Fox C.R. Remove and recover phenol. Hydrocarbon processing 7 (1975) 109. Fox C.R.: Removing toxic organics from waste water. Chem. Eng. Process. 8 (1979) 70. FergusonG.U.: Phenol recovery using polymeric adsorbent resins. Chem. Ind.8 (1982) 567. TeramotoM., Takihana H., Shibutani M.Yuasa T., et al: Extraction of amine by W/O/W emulsion system. J. Chem. Eng. Jap. 14-2 (1981) 122. Jagirdar G.C., Sharma M.M.: Separation of close boiling substituted Anilines: Gasliquid vs Conventional liquid-liquid dissociation extraction. J. Separ. Proc. Technol.24(1981)7.
179
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
[11]
[12] [13]
[14] [15]
[16] [17]
[18] [19]
[201 [21]
[22] [23]
[24] [25] [26] [27] [28] [29]
[301 [31]
Wadekar V.V., Sharma M.M.: Separation of close boiling organic acids/bases by dissociation extraction: Binary and Ternary Systems: Substituted anilines; Binary system with thermally regenerative extractant: Chlorophenols. J. Sep Proc. Tech.2-2 (1981)28. Medir M., Arriola A., Mackay D., et al: Phenol recovery from water effluents with mixed solvents. J. Chem. Eng. Data 30 (1985) 157. Jagirdar G.C., Sharma M.M.: Recovery and separation of mixtures of organic acids from dilute aqueous solutions. J. Separ. Proc. Technol. 1-2, (1980) 40. Netke S.A., Pangarkar V.G.: Extraction of naphthenic acid kerosene using porous and nonporous polymeric membranes. Sep. Sci. Technol. 31 (1996) 63. Krishnakumar V.K., Sharma M.M.: A novel method of recovering phenolic substances from aqueous alkaline waste streams. Ind. Eng. Chem. Process Des. Dev. 23(1984)410. Devulapalli R., Jones F: Separation of aniline from aqueous solutions using emulsion liquid membranes. J. Hazard. Mater. B.70 (1999) 157. Boyadzhiev L., Besenshek E., Lazarova Z.: Removal of phenol from waste water by double emulsion membranes and creeping film pertraction. J. Memb. Sci.21 (1984) 137. Thien M.P., Hatton T.A.: Liquid emulsion membranes and their applications in biochemical processing. (2001) 819. Gadekar P.T., Mukkolath A.V., Tiwari K.K.: Recovery of nitrophenols from aqueous solution by a liquid emulsion membrane system. Sep. Sci. Technol. 27 (1992) 427. Kataoka T., Nishiki T., Kimura S.: Phenol permeation through liquid surfactant membrane-Permeation model and effective diffusivity. J. Memb. Sci. 41 (1989) 197. Kakoi T., Goto M., Natsukawa S., et al: Recovery of phenols using liquid surfactant membranes prepared with newly synthesized surfactants. Sep. Sci. Technol.31-1 (1996) 107. Ulbrich M., Marr R., Draxler J.: Selective separation of organic solutes by aqueous liquid surfactant membranes. J. Memb. Sci.59 (1991) 189. Terry R.E., Li N.N., HO W.S.: Extraction of phenolic compounds and organic acids by liquid membranes. J. Memb. Sci. 10 (1982) 305. Kiani A., Bhave R.R, Sirkar K.K.: Solvent extraction with immobilized interfaces in a microporous hydrophobic membrane. J. Memb. Sci.;20 (1984) 125. Ho S.V.: A supported Polymeric Liquid Membrane Process for Removal of Carboxylic Acids from a Waste Stream. Environ. Prog. 18-4 (1999) 273. Ho S.V. Extracting organic compounds from aqueous solutions. (1996) US. Patent 5512180. Ho S.V. Solid poly-amphiphilic polymer having use in a separation process. (1996) US. Patent 5,552,053. Ho S.V. Supported liquid membrane and separation process employing some. (1996) US. Patent 5507949. Ho S.V., Sheridan P.W., Krupetsky E.: Supported polymeric liquid membranes for removing organics from aqueous solutions I. Transport characteeristics of polyglycol liquid membranes. J. Memb. Sci. 112 (1996) 13. Meckl K., Lichtenthaler R.N.: Hybrid process using pervaporation for the removal of organics from process and waste water. J. Memb. Sci. 113 (1996) 81. Lipnizki F., Hausmanns S., Ten P.K., et al: Organophilic pervaporation: prospects and performance. Chem. Eng. J. 73 (1999) 113.
180
Membrane Aromatic Recovery System (MARS) - A New Process For Recovering Phenols And Aromatic Amines From Aqueous Streams Livingston
[32] [33] [34]
[351 [36]
[37]
[38]
Ray S.K., Sawant S.B., Joshi J.B., et al: Perstraction pf phenolic compounds from aqueous solution using a nonporous membrane. Sep. Sci. Technol. 32-16 (1997) 2669. Klein E., Smith J.K., Weaver R.E.C., et al: Solute Separation from Water by dialysis II. The separation of Phenol by Downstream Conjugation. Separ. Sci. 8-5 (1973) 592 Klein E., Smith J.K., Wendt R.P., et al: Solute Separation from Water by dialysis I. The separation of Aniline. Separ. Sci. 7-3 (1972) 285. Brookes P.R, Livingston A.G.: Aqueous-aqueous extraction of organic pollutants through tubular silicone rubber membranes. J. Memb. Sci. 104 (1995) 119. Bennett M., Brisdon B.J., England R., et al: Performance of PDMS and organofunctionalised PDMS membranes for the pervaporative recovery of organics frm aqueous streams. J. Memb. Sci.137 (1997) 63. Castelo-Ferreira F., Han S., Livingston A.G.: Recovery of Aniline from Aqueous Solution using the Membrane Aromatic Recovery System (MARS). Ind. Eng. Chem. Res. (2002) In Press. Han S, Castelo-Ferreira F., Livingston A.G.: Membrane aromatic recovery system (MARS) -a new membrane process for the recovery of phenols from wastewaters. J. Memb. Sci. 188-2 (2001) 219.
181
This page intentionally left blank
Functional Membranes and Materials for Biocatalysis, Separation, and Analysis
183
This page intentionally left blank
Biofunctional Membranes and Biomedical Devices
185
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 9
Membrane bioreactors for biotechnology and medical applications L. Giorno, L. De Bartolo, E. Drioli* Research Institute on Membranes and Modelling of Chemical Reactors, IRMERC-CNR, c/o University of Calabria, via P. Bucci cubo 17/C, 87030 Rende (CS); *Corresponding author, Tel: +39 0984 492039; Fax. +39 0984 402103 e-mail: [email protected] ABSTRACT In this chapter we will discuss the properties and applications of bioartificial hybrid systems used in biotechnological and medical applications. In the first part we will focused on membrane bioreactors using biocatalysts such as enzymes, microrganisms and cells for the production of biological molecules or for separation of pharmaceutical products. In the second part we will discuss on membrane bioreactors using isolated mammalian cells (i.e., liver cells, pancreatic cells) as bioartificial organs in temporary or continuous substitution of injured organ. The properties of membranes to be used in these devices will be reported including mass transport, morphological and physico-chemical properties that influence their performance. The discussion of the biotechnological part will consider the effect of immobilization on functional stability and activity of biocatalysts, including membrane material and morphology, physico-chemical properties of reaction environment. The successful examples of bioreactors running at large scale, of which the authors are aware, will be also presented. In the medical application part, the development of membrane bioartificial organs, e.g. bioartificial pancreas and bioartificial liver, as well as the properties of membranes for these systems will be discussed. Particular attention will be given to the recent achievements in cytocompatibility and biocompatibility of membranes in bioartificial organs. 1.
INTRODUCTION
The combination of artificial membranes with macromolecules of biological origin allow to obtain hybrid systems capable to imitate what nature
187
Membrane
Bioreactors For Biotechnology And Medical Applications -
Drioli
has fine tune through aeons of evolution. In fact, these systems, based on their selective transport and catalytic properties, are able to carry out chemical reactions, selective separation, waste removal, etc. The use of membranes and biological tools for improving traditional production systems is a possible approach for maintaining a sustainable growth. Typical examples include novel pharmaceutical products with well-defined enantiomeric composition; new foods with improved nutritive properties; medical devices for the care of crucial diseases; and the treatment of wastes. The performance of the hybrid systems are governed by the properties of specific biomolecules and membranes and in particular by their reciprocal interactions [ 1]. When the biomolecule is an enzyme immobilized in a membrane, the overall performance of the biocatalytic membrane reactor is due to the preserved functional conformation of the enzyme as well as to the transport of reagent and product, through the membrane, to and from the enzyme, respectively. When mammalian cells are to be adsorbed on membrane to prepare artificial organs, the membrane properties are of crucial importance to keep them viable. The biomolecule can either be loaded to the membrane or just compartmentalized by the membrane. Reactors with segregated cells are suitable for therapeutic applications; for example, the bioartificial pancreas, the bioartificial liver, and the extracorporeal detoxification device [2]. 2.
M E M B R A N E BIOREACTORS IN B I O T E C H N O L O G Y
Biocatalytic membrane reactors can combine selective mass transport with chemical reactions and the selective removal of products from the reaction site increases the conversion of product-inhibited or thermodynamically unfavorable reactions. Many enzymes and micro-organisms have been used in membrane reactors to catalyze bioconversions for various application [4-24]. The possibility of integrating biotransformation into productive reaction cycles makes reliable the use of biocatalysts at large scale. The main parameter that influenced the development of biocatalytic processes at productive level is the stability of the biocatalyst. Immobilization has proved to increase stability of biocatalysts, although it may cause changes in the catalytic activity and enantioselectivity [25-27]. This is a common observation, nevertheless, it cannot be considered a general rule of the inverse relationship between stability and activity and enantioselectivity of immobilized enzymes. Most probably, these effects are related to the interactions between the chemical groups of enzyme and membrane: due to these interactions the molecule becomes more rigid, and therefore more stable, but the loss in flexibility damages the specificity and activity. 188
Membrane
Bioreactors For Biotechnology And Medical Applications - Drioli
Recently, studies on the structure of enzymes present in extremophiles (thermo-stable) and mesophiles (thermo-labile) micro-organisms have been reported [28-31 ]. Referring to the activity, it is observed that the enzyme that are resistant at high temperature are much less active (especially at low temperature). For immobilized enzyme, the challenge is to obtain the proper interactions to guarantee a balanced efficiency of the various properties (stability, enantioselectivity and activity). The efficiency of the overall immobilized system, depend on the biochemical (catalytic activity; half-life time; productivity; reaction kinetics; catalyst inhibition effects; concentration, viscosity, solubility and purity of substrate and product; sensitivity to pH and temperature; immobilization stability, etc.) mechanical (membrane configuration, structure and strength; size, morphology and pore size distribution; swelling and compaction effects; shear stress, etc.) and hydrodynamics parameters (flow regime; pressure drop; transmembrane pressure; flow velocity; residence time; easy to clean, etc). A comparative analysis of the performance of enzymes free and immobilized on membrane using various substrates is reported in Table 1. Hydrolases/esterases (such as lipases) are among the most studied biocatalysts [34-50]. Lipases (triacylglycerol-hydrolases, E.C. 3.1.1.3) act efficiently on water-insoluble substrates by binding to the water-organic interface. The natural organic interface consists of triglycerides. This binding places the lipase close to the substrate, and in particular increases the catalytic efficiency by interfacial activation. Due to their natural function at the oil-water interface, lipase exhibit high stability in contact with organic solvents, and beside triglycerides they are also active with a variety of non-natural substrates, showing generally a high stereo- and/or regioselectivity [41, 53]. Some author reported that the immobilized lipases used in organic solvent to catalyze esterification reactions, show higher activity and stability than lipase free in suspension [54, 55]. 2.1
Influence of operating conditions
2.1.1. Effect of immobilization
Lipases have on their surface hydrophobic and hydrophilic patches, therefore they can be immobilized by ionic interactions on ionic hydrophilic membrane support, or by van der Waals forces on hydrophobic membrane. Hydrophobic materials have shown to increase the performance of lipase [63, 64]; nevertheless, the support has to be chosen in agreement with the properties of the overall reaction system. For example, when using organic solvents, the resistance of the membrane to the solvent has to be taken into the account. In this case, hydrophilic membranes (such as polyamide,
189
MembraneBioreactorsFor BiotechnologyAnd MedicalApplications- Drioli
polyacrilonitrile) have shown to have higher stability than hydrophobie ones (e.g., polypropylene). Table 1" A comparative analysis of the performance of enzyme immobilized in various
membrane Enzyme (substrate)
Lipase (Olive oil)
Type of immobilization
Type of reactor
(Enzyme Free) a
STR E-EMR
Observed reaction rate (retool.1 "l h q)
Enzyme stability
ef.
-
6.95
Low
2
1.6
6.80
Medium
4.50
High
Axial Flow rate (ml-min "t)
3, 34
Entrapment by cross-flow filtration
TSP-EMR (organic) 1.6 (aqueous) 160 Cross-linkingby TSP-EMR (organic) 1.3 glutharaldehyde (aqueous) 167 Lipase (Cyano ester of (Enzyme free) b STR ibuprofen) Entrapment by TSP-EMR (organic) 160 cross-flow (aqueous) 400 filtration Lipase (Methyl ester (Enzyme free) STR of naproxen) Entrapment by TSP-EMR (organic) 300 cross-flow (aqueous) 300 filtration Fumarase STR (fumaric acid) (Enzyme free) ¢
4.50
1.12
Low
1.2x10 ~
High
2
Low 5.3x10 "3 8x10 "4
High
290
Low
Entrapment by UF-EMR 140 2375 High cross-flow filtration d aKm = 9 m m o l ' l "1, V m a x = 20 k + 2 - 6.86 s'l; aKm = 47 m m o l ' l l , V m a x = 4 mmol.1-1 h -1, k+2 = 1.37 s l ; aKin = 212 m M retool.l-l; V m a x = 9000 retool.1 -l h l ; k+2 = 5x1012 s "l, dk+2 = 8x1012 s -1
In addition to the membrane material, it is very important also the amount of enzyme immobilized [26, 55, 65, 66]. It is generally observed that the performance of immobilized enzyme as a function of density of enzyme on membrane (gprotein/Cm3) has an optimum. Increasing the gprotein/Cm3 the catalytic activity can be stable or increase until it reaches a maximum, after which, for further increase of protein loading no increase of activity is achieved. The initial increase of activity with the enzyme loading, indicates that the interaction with the support destabilizes the protein, and as the protein amount is increased it is more protected, and then more active. The constant activity is mainly due to mass transport limitations. It some cases, further increase of protein can even lead to decrease of activity. This happens when the protein-protein interactions
190
Membrane BioreactorsFor BiotechnologyAnd MedicalApplications- Drioli
at high concentration cause denaturation of protein conformation; either because of bound with the active site or because of proteolitic activity present in the crude enzyme preparation. The behavior of some enzyme as a function of loading amount is summarized in Figure 1, from which it can be seen that the same type of immobilized enzyme can show different behavior with different substrates. For
Activity
h.._
Enzyme loading
v
Figure 1. Qualitative behaviour of activity of immobilized enzymes as a function of enzyme loading: •
kipase C. rugosa in polyamide membrane (hydrolysis activity with olive oil) [26]
II
Lipase C. rugosa in polyamide membrane (hydrolysis activity with naproxen ester) [35] Lipase C. antarctica on Accurel0a)EP100 (Esterification activity) [55]
O
Lipase Humicola sp on Accurelc.)EP100 (Esterification activity) [55]
r]
Fumarase from porcine heart (hydrolysis activity)[ 19]
example, lipase from C. rugosa when used with triglycerides as substrate, shows constant behavior for initial increase of immobilized enzyme and after a decrease; while, when used with naproxen ester for initial increase of immobilized enzyme the activity increases and after, for further increase of enzyme, the activity decreases. This implies that the same enzyme conformation, achieved with the same type of immobilization, has different affinity for the different substrates. The use of proper conditions may allow to reduce the amount of enzyme with the result of reducing the mass transport limitations and cost.
2.1.2. The influence of hydrodynamics conditions In order to measure the kinetic properties of immobilized systems, it is necessary to work in conditions of reaction limited regime, where the mass transport rate is much higher compared to the reaction rate [67].
191
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
Among the various methods to control the mass transfer coefficient, the fluid dynamics conditions, such as axial velocity and transmembrane pressure, may help in achieving the proper mass transfer properties. In two-separate phase (or biphasic) membrane reactors (TSP-EMR), where the mass transport occurs mainly by diffusion, the effect of the flow rate of organic phase (which contains the substrate) can be very different from that of the aqueous phase (which extracts the product) on the reactor performance. An example of the mentioned behavior is reported in Figure 2 and refers to the lipase using olive oil and cyano ester of ibuprofen as substrate. For the studied values, the observed reaction rate decreases with increase of the organic flow rate, for both substrates. Vice versa, the observed reaction rate increases with increasing the aqueous flow rate. This behavior is due to the fact that the higher transport of substrate (achieved at higher axial velocity of the organic phase) a higher reaction rate is obtained. On the other hand, the fixed axial velocity of the aqueous phase was not able to promote the extraction of the reaction product, which therefore, inhibited the reactor performance. This show the importance of both transport of substrate and product through the enzyme-loaded membrane 0.14 0.12 .,,.... 0.1
.,=
0.08 E g 0.06
l e Olive oil &CNE
0.04 0.02 0 0
1O0
200
300
Qa (mmol/I h)
Figure 2. Effect of axial flow rate of organic and aqueous phases on the performance of two-separate phase enzyme membrane reactor
reactor. In ultrafiltration enzyme membrane reactors (UF-EMR), where the mass transport occurs by convection, the permeate velocity is the parameter that mostly affects the reactor performance, as it influences the residence time of reagents within the enzyme-loaded membrane. The residence time, T, is calculated as the ratio between the membrane thickness, L, and the permeate velocity, L/T. Figure 3 shows the influence of residence time on the conversion of fumaric acid into L-malic acid using fumarase immobilized in a ultrafiltration membrane reactor [19]. The effect of the amount of enzyme loading is also illustrated.
192
Membrane
Bioreactors For B i o t e c h n o l o g y
And
Medical
Applications
- Drioli
0.9 0.8 .-. 0.7
~
n-n
~o 0.6 •E
0.5 0.4 0.3 • 0.3 mg/cm 3
0.2 0.1
• 1.14 mg/cm3
0 0
i
i
i
~
i
100
200
300
400
500
600
Residence time (s) Figure 3. Influence of residence time on the efficiency of and ultrafiltration enzyme membrane reactors
2.1.3. The influence of reaction microenvironment on enzyme enantioselectivity Thanks to their properties, lipases have found applications in various fields of biotransformations. One of the most recent and attractive field is the synthesis of chiral compounds. The importance of using optical pure enantiomers became clear in the recent years thanks to the understanding of the different effects of enantiomers, where often one has the positive properties whilst the other can cause serious side effects [68]. The chiral market represents about the 25% of the drug market; on 1993 only the 3% of this was represented by pure enantiomers while the rest was sealed as racemic mixtures; on 1997 more than 75% of chiral drugs were marketed as pure enantiomers. This picture shows the rapid change towards the use of pure enantiomers in the pharmaceutical field, that had his first push in 1992 when FDA issued a policy statement. The same trend is expected to happen for food and agro-chemicals. On the basis of this context, many efforts are focused on the development of technologies to produce optically pure isomers. Although the most common method to produce enantiomers at industrial level is the chemical synthesis of a racemic mixtures followed by separation of the two enantiomers by diastereomeric crystallization, new resolution processes are being considered. Among others, membrane processes are today investigated for this purpose. On the basis of the current literature, there are two schemes on the use of membrane technology to produce enantiomers [69]. In one case, the membrane itself is intrinsically enantioselective and separates the wonted isomer on the basis of the pore spatial conformation. In the other case, a kinetic resolution catalyzed by an enantiospecific catalyst is combined with a membrane 193
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
separation process: the chiral system is represented by the catalyst, the membrane only separates the product from the substrate on the basis of their physical chemical properties. Studies on the use of lipase-catalyzed enantioselective biotransformations in multiphasic membrane reactors, for the production of pure (S)-naproxen have been carried out at our laboratory. Particular attention was focused on the parameters that influence the catalytic properties of the lipase immobilized on polymeric membrane. The performance of lipase immobilized with the same method on different type of membranes, which were then used in different reactor systems, has been investigated. The lipase was immobilized by cross-flow ultrafiltration, so that to obtain a stable irreversible lipase fouling within the sponge layer or on the lumen surface of asymmetric polymeric membranes. The lipase-loaded membranes were used to carry out the kinetic resolution of (R,S)-naproxen methyl ester to obtain the pure (S)-naproxen acid, in twoseparate phase enzyme membrane reactor (TSP-EMR) and in emulsion enzyme membrane reactor (E-EMR) systems. In the TSP-EMR the organic and aqueous phases are separated by the enzyme-loaded membrane and the mass transport through the membrane occurs by diffusion. In this system, it is difficult to obtain the interface at the level where the enzyme is immobilized. Therefore, the observed catalytic activity and enantioselectivity resulted lower compared to the free native lipase, because they were negatively affected by the fact that some enzyme worked in organic phase, some in aqueous phase and only part worked at the organic/water interface. Polyamide membranes allowed to obtain higher enantioselectivity compared to polysulphone membrane, and in general, for both types of membranes, the sponge layer resulted a better immobilization site, compared to the thin selective layer, due to its higher loading capacity. The catalytic stability of immobilized lipase was very good for both type of membranes [70]. In the E-EMR an organic in water (O/W) microemulsion, prepared by membrane technology, with a size distribution from 1 to 4 ~m, was fed to the lipase-loaded membrane from shell to lumen, so that pores into the sponge layer were filled with the microemulsion. The thin selective layer allowed only water to permeate through the membrane whilst the organic phase was retained. Therefore, while the substrate is kept in contact with the immobilized lipase, the product is removed in the permeated water. In this system mass transfer through the membrane occurs by convection, and between the two phases it occurs by diffusion. The great advantage of the E-EMR is that all the immobilized enzyme is able to work at the organic/water interface thanks to the presence of the microemulsion within the pores. This allows to achieve catalytic activity and enantioselectivity for the immobilized lipase comparable to the free native lipase, and much higher stability, with overcome of the product inhibition
194
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
effects. In this system, each finger of the sponge layer works as a continuous tank emulsion microreactor. The experiments with E-EMR demonstrated that the lower enantioselectivity of immobilized lipase observed in TSP-EMR is due to microenvironment reaction conditions rather than to deactivation of the enzyme due to immobilization. In fact, for comparable amount of enzyme loading, immobilized with the same procedure on the same type of membrane support, the E-EMR showed about 30% higher enantioexcess compared to the TSPEMR.
2.2
Applications Table 2 summarizes some of the applications of immobilized enzymes at large scale. Table 2. Examples of industrial application Purpose Ref. Biocatalyst Immobilization F,scherichia coli Entrapped in Production of L-aspartic acid 56 cells polyacrylamide Entrapped into the Hydrolysis of the milk and whey lactose 57 Lactase (13galactosidase) and fibres of cellulose acetate used for the Synthesis of the dipeptide AspartameTM 58 Termolysin Pseudomonas Immobilized with Production of L-alanine 59 dacunahe glutaraldehyde Glucose-isomerase Reticulate with Production of fructose concentrated 60 glutaraldehide syrups Amino acylase Immobilized on Production of L-amino acids from 61 DEAE-Sephadex racemic mixtures Brevibacterium Entrapped in Production of L-malic acid 62 ammoniagenes polyacrylamide Lipase Entrapped in hollow- Production of (2R,3S)-trans isomer of 49 fiber asymmetric methyl ester of 4-methoxyphenylglycidic membranes acid The major technological difficulties in using biological immobilized systems on an industrial level are related with the life-time of the enzyme; the availability of pure enzyme at an acceptable cost; the necessity for biocatalysts to operate at low substrate concentrations; microbial contamination; the lack of general assumption and the consequent need for individual development on a trial and error basis. Nevertheless, nowadays some new conditions (such as the higher attention of governments towards environment and health care; the need of innovative technologies for a sustainable growth, etc.) may force the industry to develop new, safe, clean, and low energy consumption technologies such as the membrane bioreactors. .....
195
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
Membrane technologies used in the production of optical pure isomers have been recently described [69, 71]. The major problems in the production of these compounds on a large scale concern: the requirements for expensive cofactors; the low water-solubility of the substrates; the separation and purification of the products from complex solutions. Studies carried out mainly on a laboratory scale indicate that in many cases the use of the appropriate type of membrane and membrane reactor design can overcome these difficulties. For example, UF-charged membrane reactors can be applied to retain cofactors [72]; two separate phase membrane reactors can be used for the bioconversion of low water-soluble substrates [49]. In addition to the use of enzymes in organicaqueous systems, their use in pure organic solvent is increasing. Enzymes in organic media are able to work in microenvironments that contain very little quantifies of water (usually less than the solubility limit). Immobilized enzymes operating in organic media show novel properties, such as enhanced stability and altered substrate specificity. An investigation on transesterification reaction in anhydrous tetrahydrofuran carried out with free and immobilized lipase demonstrated that the enzyme immobilized on zirconia membranes is more active compared with the enzyme in suspension and that the selectivity towards a reaction intermediate changed, resulting in the production of a non-naturally available flavanoid [54]. In many instances, when the product is obtained by fermentation, it is present as a component of a complex solution from which it needs to be separated and purified. In these cases, integrated membrane systems can be used for continuous production and downstream separation. For example, the production of L-lactic acid is obtained by continuous fermentation in a membrane fermentor. This consists of a traditional fermentor combined with an ultrafiltration unit. The solution recovered as permeate contains the product, Llactic acid, together with other small molecules that are not retained by the membrane, whilst the cells and macromolecules are recycled back to the bioreactor. The purification of the product can be achieved by membrane-based solvent extraction carried out through two membrane contactors. Electrodyalisis [73], nanofiltration and reverse osmosis [74, 75], membrane crystallization [76] are other separation processes that can be used in the downstream processing. In recent years, membrane technology has also attracted considerable attention for the treatment of waste-water [77,78]. The reason of this lies on the increase of potable water demand and on the increase of domestic and industrial wastewater discharges linked to the population growth. In addition, more restricted legislation towards environmental care, forces industries to minimize the input of energy and water. Two different membrane bioreactor configurations have been mainly investigated: MBRs with the membrane module outside the activated sludge tank, and MBRs with membrane submerged in the biological reactors. Due to
196
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
lower energy requirements, the submerged membrane bioreactors have found more success than the external ones. Membranes and membrane bioreactors have been studied for the removal of nitrogen from wastewater [79, 80]. Zoh et al. at University of California, combined a membrane bioreactor to a ceramic cross-flow ultrafiltration module for treating a synthetic wastewater containing hydrolysis byproducts of high explosive RDX (Hexahydro-l,3,5trinitro- 1,3,5-triazine) compound [81 ]. The state-of-the-art of submerged membranes has been recently depicted by R. Ben Aim who has discussed the potentialities (low energy input, long term operation without cleaning, less dependence on variation of rheology behavior with concentration) and limits (high membrane area due to low fluxes linked to the small transmembrane pressure) of the technology. Particular attention was devoted to the use of submerged membranes associated with aeration and on the effect of air bubble size on fouling [82]. Many studies in the field are oriented to the investigation of operating conditions and optimization, such as the effect of air sparging on cell viability; the nitrification rate and COD removal; the use of low cost membranes such as non-woven polypropylene to reduce capital costs; the conversion of organic wastes to energy and fertilizers. Others aim at showing the application at industrial level: Tazi-Pain reported the results of the first year operation of the BIOSEP (R) process, a full scale sequenced aeration MBR for municipal wastewater treatment. He claimed that this is the first plant in the world capable to achieve, in a single tank, the biodegradation of the organic matter, the nitrification-denitrification reactions and the physicochemical elimination of the phosphates [83]. The plant uses an immersed membrane system equipped with MF/UF hollow fibers supplied by Zenon Environmental Inc. Some examples of membrane bioreactors for waste water treatment at large scale are summarized in Table 3. 3.
MEMBRANE BIOREACTORS IN MEDICAL APPLICATIONS
Polymeric semipermeable membranes and membrane processes play a pivotal role in replacement therapy for acute and chronic organ failure and in the management of immunological disease. All clinical blood purification methods employ membrane devices [2].The next generation of artificial organs and tissue therapies is almost certain to be similarly grounded in membrane technology. Membranes of suitable molecular weight cut-off are used in bioartificial organs (e.g., pancreas, liver) using isolated cells, as selective barriers to prevent immune system components from getting into contact with the implant, while allowing nutrients and metabolites to permeate freely to and from cells. The use of membranes in bioartificial substitutes and tissue engineering date back to the 197
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
Table 3. Membrane bioreactors in water treatment at large scale application
Type of reactor
Type of wastewater
Application
BIOSEP (Submerged membrane bioreactor)
Municipal and industrial
Reduce BOD Removal of solids, nitrogen, phosphorous, etc.
Aerated MBR (Biofilms attached to membrane)
Industrial
COD and ammonia removal
Moving bed biofilm MBR (Submerged membrane)
Municipal
Separate biomass, colloids, soluble organic matter
Zeewed (Zenonsubmerged MBR)
Domestic and pharmaceutical effluents Urban
Reduce COD
Kubota membrane activated sludge
Reduce BOD
year 1933, when Vincenzo Bisceglie in Bail Italy encased mouse tumor cells in a nitro-cellulose membrane and inserted them into the abdominal cavity of guinea pig, to show that cells were not killed by an immune reaction in the pig [84]. Subsequently many researchers focused on the development of immunoprotective membranes to prolong the life of transplant. Currently two important areas of interest are the bioartificial pancreas for the treatment of insulin-dependent diabetes and the liver assist device for the temporary treatment of acute liver failure. Membrane capsules containing dopamine secreting cells also are being explored for treating of Parkinson's disease, a progressive brain disorder characterized by a deficiency of the neurotransmitter dopamine [85]. Immunoprotective membrane cell transplants are being investigated to treat other nervous system disorders. Polymer membranes also are being explored to block cell adhesion or scar tissue formation, for example after surgery, and thus improve wound healing. In addition, the membranes are being investigated for prevention of restenosis (coronary artery narrowing) after angioplasty [84]. In membrane bioartificial organs, cells are compartmentalized by means of semipermeable membranes that permit the transport of nutrients and metabolites to cells and the transport of catabolites and specific metabolic products to blood. The membrane must avoid the contact between xenogenic cells and patient's blood to prevent immunological response and rejection of xenograft. Membranes act as means for cell oxygenation and in the case of
198
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
anchorage-dependent cells as substrata for cell attachment and culture. As a result, the type of membrane to use in a bioartificial organ must be chosen on the basis of its permeability characteristics as well as on its physico-chemical properties related to the separation process. 3.1.
Membrane bioartificial pancreas
In patients with insulin-dependent diabetes, the cells that normally produce insulin don't function, so patients must receive insulin injections to regulate their blood glucose level. This approach in the long term result in the so-called "diabetic complications" such as kidney failure, neuropathy, which lead to the patient's severe incapacity or death. Transplantation of the whole pancreas, pancreatic tissue fragments or islets of Langerhans would assure the required glucose-related insulin secretion, but the rejection of implants and scarce availability of donor organs limits this therapeutic approach [86]. An alternative approach is the development of membrane bioartificial pancreas using isolated islets of Langerhans or single beta-cells, which are able of sensing the plasmatic glucose concentration and produce insulin amounts related to the actual glycemia, entrapped by means of membranes. In 1970 W.L. Chick and colleagues transplanted isolated islets protected by hollow-fibre ultrafiltration membrane (an acrylonitrile-vinyl chloride copolymer) into dogs made diabetic by surgically removing the pancreas. The device consists of a chamber through which passes a copolymer membrane connected to standard vascular grafts. Islets are placed inside the chamber, through ports in the housing into the cavity, but are outside of the blood stream. Nominal molecular porosity of 80,000 Dalton permits free diffusion of nutrients and insulin across the membrane but inhibits the entry of immunoglobulins and immunocytes from the blood stream into the chamber (Figure 4a) [87]. Since 1970 research efforts were devoted to the development of a hybrid bioartificial membrane pancreas [84-87]. Different bioartificial pancreas were designed in four physical types: hollow fibers, capsule, coatings and sheet. The many different types of prosthetic devices proposed so far can be grouped in three main categories" extravascular devices, intravascular devices and microencapsulated islets of Langerhans [89-98]. In the first case, the tissue is enclosed between membranes, if in a flat sheet configuration, or in the lumen of hollow fibre membranes, and then implanted in an extravascular site (see Figure 4). Other researchers [88] proposed hollow fiber device as extravascular bioartificial pancreas. This device consists of hollow fiber with 0.5 mm diameter, 20 mm length, containing 80,000 islets/ml (Figure 4b). The extravascular systems generally suffer from an intrinsically slow insulin response following changes of blood glucose concentration, limited by the purely diffusive mass transport and by the fibroblastic response of the host.
199
MembraneBioreactorsForBiotechnologyAndMedicalApplications- Drioli membrane i
m
membrane
~w
cavity
m
blood
housing ~x artificialvasculargraft
,ml "
m
m i
a)
m
~
i,m
mu
'~ ~
m l
I
i l
i i
i
mw i
-
pancreaticislets
am,
i
b)
membrane
~cis
u
Microencapsu pancreaticisle
c)
pancreatic islets
d)
~IB 0
~~ ....
blood
~ m
eassist *,,~ ,I~m r @
m e m b r a n e
qr Im ~
4' dmllllk~'$
:, I L Y ~ m Z ¢ . ~
~
17" abrane
insulin
~ ~'~1~,, "" • +
N~=,~
e)
200
J V
Pancreaticislets
M e m b r a n e B i o r e a c t o r s F o r B i o t e c h n o l o g y A n d M e d i c a l A p p l i c a t i o n s - Drioli
Corewithislets
.
,.
.
Meshforphysicalstr
Coating
. . . . . . . . . .
...........................................................................i............................................................................................................................................................................................................................. ................. Outer
layer
g)
h)
Figure 4 Schematicof bioartificial pancreas configurations: a) device developed by Chick et al., pancreatic cells are loaded outside the tubular membrane; b) pancreatic cells are loaded inside of hollow fiber in lumen compartment; c) microencapsulated cells are loaded in the lumen compartment of hollow fiber membrane; d) pancreatic cells loaded outside of hollow fibres in extracapillary compartment; e) cells loaded outside of tubular membrane; f) microencapsulate pancreatic cells; g) pancreatic cells cultured in islet sheets between fiat sheet membranes; h) pancreatic cells inside coating. Intravascular membrane devices are designed so that the membrane separate the graft directly from blood stream of the host. In the Figure 4d cells are cultured outside hollow fiber membranes arranged in a housing in a shelland-tube configuration [89]. These devices suffer from blood clotting at the interface between blood and the synthetic material of the membranes or the point of access, but they are extremely attractive in terms of flexibility of design and use. Additionally, the implant site can be chosen on the basis of reducing the response time of the prosthesis following an increase of blood glucose concentration. A device currently in preclinical development licensed from Circe Biomedical is the PancreAssist Bioartificial Pancreas System [90]. This device consists of a single tubular membrane surrounded by insulin-producing porcine islets, which are in turn, enclosed within a disk-shaped housing (Figure 4e). The porous tubular membrane permit the transport of nutrients and glucose to cells and the transport of insulin from cells to blood. Membrane prevents also the contact between immunological species present in the patient's blood and islets. This device should be implanted near the kidney and surgically connected directly to circulatory's system using vascular graft. In the case of microencapsulated islets, the membrane in the form of alginate gel is formed around the islets of Langerhans, obtaining thus microcapsules with diameters of 300-400 ~tm (Figure 4f) [91, 92]. Microcapsules may make better implants than hollow fibres because they offer better conditions for diffusion of nutrients to the insulin-producing cells and
201
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
waste products from them. Normoglycemia has been reported after intraperitoneal transplantation of alginate-polylysine microencapsulated allogenic and xenogenic islets in diabetic animal models and recently also in men. However, graft survival is always limited to several weeks. Graft failure is interpreted as non-specific immune response i.e., foreign body reaction against the microcapsules resulting in a progressive overgrowth of the capsule and subsequent necrosis of islets. Researchers focused study on highly purified alginate and other biochemicals more biocompatible [91] and on novel membranes able to prevent permeation of low molecular weight humoral molecules released by xenogenic islets. The Islet Sheet is a thin planar bioartificial pancreas, licensed from Islet Sheet Medical Company, contains live, functional islets in an artificial polymer matrix (Figure 4h) [93]. Each sheet is several cm in diameter and contains 2 to 3 million cells and contains islets microencapsulated within in a mesh to increase the physical strength between two layer of semipermeable alginate membrane. A semipermeable membrane such as 0.2 ~tm cellulose ester filter membrane is saturated with crosslinking solution. The cells 4 to 6 sheets contain enough islet tissue to cure diabetes in an adult. The sheet is so thin (the overall thickness is 250 ~tm) that diffusion alone allows sufficient nutrients to reach the center of the sheet. A coat on the exterior of the sheet prevents contact between the cells inside and immune effector cells of the host as well as inhibiting diffusion of antibody and complement. The alginate membranes show high permeability to different solutes and excluded immunocompetent species. No immune suppression drugs are needed. The sheet may be removed or replaced at any time. Experiments on islet sheet began at the University of Chicago in September 1998, and moved to the University of Cincinnati and the University of Alberta in 2000. Large animal studies given encouraging results. The implantation of islet sheet in omentum of pancreatectomised dog permitted to return blood sugar to normal and at 60 days the blood sugar was lower at every measurement time [94]. 3.1.2. Membrane bioartificial liver
Each year in the US, approximately 150,000 people are hospitalized with liver disease and over 43,000 people die from it [95]. Transplantation, the only effective means of treating liver failure, is not an option for many patients. Ironically, the liver is a highly regenerative organ [96]. Supportive therapies as dialysis, hemofiltration, hemoperfusion for patients with acute liver failure have been proven ineffective because physical methods are not sufficient for the management of severe biochemical disorders. The development of an extracorporeal liver assist device, using isolated liver cells, to which a patient would be temporarily connected until he/she recovered or received a liver transplant could be a promising approach for 202
Membrane
Bioreactors
For
Biotechnology
And
Medical
Applications
-
Drioli
patients with liver failure. Since fulminant liver failure is potentially reversible, the extracorporeal bridging of liver function would also be beneficial until the patient's own liver resumes functional activity.
hepatocytes
membrane
m
I i m
blood
m
-
J
mlaam mm ,m~
am*
*am *D .m am am
am
b)
a) microcamer-attached am
blood-
m
_ n
-.,m
I,m
. _ --
.m
m~
blo
he~atc~es
blood " - _ -
;
-.d)
c) _
med
_
~
_
_
membrane
blo
/
g:ltrapme
ocytes blood
_
-_---
m
-_-----_-
e)
203
_-
-
M e m b r a n e B i o r e a c t o r s F o r B i o t e c h n o l o g y A n d M e d i c a l Applications - Drioli
membrane for • oxy
non-woven polyester matrix
hepatocytes
0 !. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
"1
hepatocytes
g) Figure 5: Schematic of liver support device configurations: a) microencapsulated hepatocytes; b) hepatocytes loaded outside of hollow fibres in extracapillary compartment; c) hepatocytes loaded between flat-sheet membranes; d) microcarrier-attached hepatocytes in the extracapillary compartment; e) hepatocytes loaded in a spirally wound non-woven polyester matrix; f) hepatocytes entrapped in a three-dimensional contracted gel matrix inside of hollow fibres; g) hepatocytes loaded in a network formed by four separate membrane capillaries with different functions.
The first clinical report of bioartificial membrane liver was released in 1987 [97]. This device consisted of a hepatocyte suspension that was separated from the patient's blood by a cellulose acetate dialysis membrane. Since different devices have been proposed (Figure 5). These devices differ from one other for the type of material used to construct the bioreactor: membrane, glass or non-woven fabrics; for configuration: fiat or hollow fiber; for coating used: matrigel, collagen; for cell culture technique" microcarrier, spheroids, aggregates, etc, and for cell capacity. In devices using hollow-fibre membranes, isolated hepatocytes are usually loaded outside of the hollow fibres in the extracapillary compartment, while blood, plasma, or culture medium flow through the lumen of the hollow fibres (Figure 5b). In devices using fiat-sheet membranes, hepatocytes are loaded between fiat membranes in a sandwich way,
204
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
while blood or culture medium flow outside of the membranes as shown in Figure 5c. Cells may be free in suspension, attached to walls or attached to microcarrier (Figure 5d). A key issue concerning the development of bioartificial liver is the maintenance of long-term viability and functions of hepatocytes: oxygen transport resistance and catabolite accumulation may limit the hepatocyte viability and metabolism [98-99]. To overcome the problem of maintenance of viability and liver specific functions in vitro of liver cells, Sussmann [100] used immortalized liver cell line in the extracorporeal hollow fibre device as biological element. This device, developed as Hepatix/Baylor unit, consisted of 10,000 individual hollow fibres with cut-off of 70 kDa. This has been tested on 11 patients with fulminant hepatic failure: four patients were sustained until transplant and two survived without the need of transplant and five died for complications. A disadvantage of this unit is the potential of seeding tumor cells into patients. Thus, other researcher proposed new culture models to address the problem of long-term maintenance of liver specific functions of liver cells inside of the bioreactor. Flendrig et al. [101], developed a bioreactor based on a spirally wound non-woven polyester matrix, three-dimensional framework for hepatocyte immobilization and hydrophobic polypropylene membranes for decentralized oxygen supply and CO2 removal (Figure 5e). Medium or plasma is perfused through the extrafibre space and in direct hepatocyte contact. Generally many investigators inoculate cells in the extrafibre space. In Nyberg's design hepatocytes are cultured in the hollow fibre lumens [102]. In this device hepatocytes are entrapped in a three-dimensional gel matrix and the extrafibre space of the bioreactor is perfused for 24 h with recirculating medium, as is shown in Figure 5f. The blood is passing through the extrafibre compartment and the gel-entrapped cells are perfused by means of medium to improve the supply of oxygen, nutrients, and the removal of catabolites. This device is scaled-up by using hepatocyte spheroids inside polysulfone membranes at University of Minnesota and recently was licensed to Algenix Inc. and it was approved by FDA for Phase I human clinical trials. An interesting device was designed by Gerlach et al. [103] by mixing different membrane fibres in various directions, thus improving the oxygen and nutrients supply to hepatocytes located in between such a network. The Figure 5g shows the scheme of such device licensed to Hybrid Organ GmbH. Four separate capillary membranes with different functions are utilized: plasma inflow by polyamide membrane; plasma outflow by polysulfone membranes; decentralized oxygen supply and carbon dioxide removal with low gradients by polypropylene membranes; sinusoidal endothelial coculture by hydrophilic membrane. This device was used for preliminary clinical studies. Based on dialysis model, the liver support system developed by Demetriou et al. [ 104] of Cedar Sinai Medical Centre of Los Angeles, consists 205
Membrane
Bioreactors For Biotechnology And Medical A p p l i c a t i o n s - Drioli
of a bioartificial liver cartridge that contains billions of pig liver cells and a machine that controls the flow of blood through the cartridge (Figure 5d). The commercial name of the device is "HepatAssist System" of Circe Biomedical and includes cellulose nitrate/cellulose acetate hollow fibre bioreactor containing porcine hepatocytes attached to a collagen-coated microcarriers, two charcoal columns, a membrane oxygenator, and a pump. Plasma from a patient flows through the cartridge. Pig liver cells on the outer surface of the fibres are in contact with the plasma through large pores on the fibre surface. This device was used in USA for a treatment of 6-7 h of 25 patients with severe acute liver failure. These reports encourage the further studies to address the problems related to the development of a bioartificial liver. In fact, although several groups worldwide are creating systems there is not still an ideal device. This is not surprising, given the complexity of liver functions. Liver support system must supply various liver specific functions including synthetic and detoxification activities for a time sufficient to allow recovery of a patient or maintenance of patients until transplantation. Recently, we evaluated in vitro the performance of a full-scale fiat membrane bioreactor (FMB) developed by Bader et al., that permits the culture of liver cells under in vivo-like conditions and at high-density culture [105-109]. In such bioreactor porcine hepatocytes are cultured within extracellular matrix between oxygen permeable fiat-sheet membranes. Isolated liver cells are located at a distance of 10-20 ~tm of extracellular matrix. This bioreactor provides culture conditions that improve liver specific functions of liver cells in vitro. The FMB is able to provide an in vivo-like microenvironment for liver cells: hepatocytes are arranged as a plate in 3-D coculture with intermingled nonparenchymal cells. In contrast to other bioreactors the FMB is based on the organization of liver cells as a plate within extracellular matrix in which each individual hepatocyte has its own membrane support and thereby its own oxygen supply position. In vitro studies demonstrated that the performance of a scale-up FMB using porcine hepatocytes is stable over a period of about 3 weeks and compares well with that of other systems present in literature. Isolated hepatocytes cultured in the FMB reconstitute many of the features of the liver in vivo. The cell concentration inside of the FMB increased in the first days of culture and then remained constant until 18 days. Specific metabolic functions of hepatocytes in terms of albumin synthesis, ammonia elimination and urea synthesis are sustained for the investigated culture time demonstrating thus the long-term maintenance of functional integrity of hepatocytes cultured in the FMB.
206
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
3.1.3.~Cytocompatibility of membranes in bioartificial organs In membrane bioartificial organs using isolated cells as biological component, semipermeable membranes play more functions: they act as immunoselective barriers, as means for cell oxygenation and provide a large area for cell attachment. All these functions are important for the maintenance of cell viability and specific functions. In our experimental study we demonstrated that isolated rat liver cells cultured on oxygen-permeable membranes reconstitute many feature of the in vivo [98]. Cells cultured on polythetrafluoroethylene oxygen-permeable membranes maintained a morphological appearance of hepatocytes similar to their in vivo appearance and exhibited at high levels tissue specific functions in vitro in terms of albumin secretion, urea synthesis and drug biotransformation functions. This finding is of relevance for oxygen supply to cells in vitro both in batch and in large-scale bioreactor systems. In a membrane bioartificial organ, cells come into contact with the membrane surface. Therefore, the response of the biological components depends on surface properties of the used membrane. Physico-chemical properties including surface composition, surface charge, surface energy, and surface morphology, may affect cell adhesion and behavior. Recently, studies performed on isolated hepatocytes cultured on semipermeable polymeric membranes have indicated that wettable and rougher surfaces enhance adhesion and metabolism of isolated hepatocytes [ 110-112]. In particular, measurements of the wettability of membrane, expressed by the contact angle in the presence of different liquids permitted to evaluate and to compare surface free energy components of membranes with different physicochemical properties. Such measurements before and after modification of native membranes in culture medium resulted to be predictive index of their cytocompatibility and/or tissue biocompatibility. Therefore a material surface treatment might enable the adaptation of its surface free energy to biological requirements. The adhesion and activity of liver cells are affected by hydrophilic properties of membranes: surface free energy and its components. Isolated hepatocytes formed on membranes with high base parameters of surface free energy, such as cellulose acetate, polycarbonayte membrane, large cellular aggregates while on hydrophobic membranes such as polypropylene membranes cells spread to a large extent and the edge of cells were often indistinguishable in a network formed by fibrous proteins of extracellular matrix. For each membrane we observed that cell adhesion increased with increasing base parameter of membrane surface tension. In Figure 6 are reported the cell adhesion after 48 h of culture on membranes with different base parameter of surface tension. Liver cells adhere strongly on membranes with high value of base parameter. The absolute value of cell adhesion is higher in the presence of
207
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
serum proteins adsorbed on the membrane surface, which change the wettability by increasing base parameter of surface tension. Adsorption of proteins from surrounding liquids is one of the key steps for the cytocompatibility of the membrane. We observed that the extent of serum protein adsorption is relatively large on hydrophobic membranes and relatively small on more hydrophilic membranes [ 112].
e~
E 0
~
76-
(l) r,.)
,.n
5-
li
.". 4 o
•~
3-
.
2 I
0,5
I
2,9
I
7,4
12,4
30
y- [m Jim 2] Figure 6: Cell density after 48 h of culture on various membranes with different surface tension base parameter calculated according to Good van Oss's equation. The number of cells adhered on various membranes was estimated by counting at the light inverted microscopy after their removing from membrane surface. The cell adhesion values are the mean of eight experiments: + standard deviation.
The base parameter of surface free energy affected the metabolic functions of liver cells. We found a correlation between urea synthesis and base parameter of membrane surface tension. For each investigated membrane cells synthesise urea with a rate that increases with increasing of base parameter value of membrane surface tension (Figure 7). The metabolic activity is particularly expressed at high levels when cells are cultured on PC and CA membranes. Figure 7 shows that on CA and PC membranes the ability of liver cells to synthesize urea increases linearly with increasing the value of base parameter of surface tension and reached the value of about 120 mg/h in correspondence of the surfaces with respectively 46 mJ/m 2 for CA membrane and 54 mJ/m 2 for PC membrane. On such membrane surfaces cells exhibited rates of urea synthesis that significantly increase with time. These results suggest that there is a marked
208
MembraneBioreactorsForBiotechnologyAndMedicalApplications- Drioli
effect of the physico-chemical properties of membranes on cell adhesion and functions. Protein adsorption modified the surface tension components of the membrane and the consequent cell adhesion and functions. As a result, 100 r-""'l
""
90-
E
80-
.-~ 7 0 ¢~ 6 0 e-, t-
>, 5 0 t~ 4 0 -
= o
3020-
~
10-
(D
O-
r 0,5
r 2,9
T 7,4
12,4
-30
?-[m Jim 21 Figure 7: Urea synthesis rate of liver cells cultured on membranes with different base parameter of surface tension after 24h (full bar) and 48 h (shade bar) of culture. The metabolic rates were determined by culturing cells in the presence of 1 mM ammonia in the medium. The reported values are the mean of eight experiments' + standard deviation.
independently on the type of native polymeric membranes it is possible to improve cell adhesion and specific functions by changing their surface tension and their components. On the basis of our results, CA and PC membranes resulted to be good substrata for liver cell culture in vitro promoting cell adhesion and specific functions in term of urea synthesis. To this purpose, researches on interactions of cells with novel membranes made in our laboratory are running in order to develop more cytocompatible and biocompatible membranes.
3.1.4. Membrane biocompatibility When blood comes in contact with foreign substances by extracorporeal circulation, reaction to foreign body occurs in the blood. Thus, to evaluate the safety of the material is used a reaction to foreign body as an indicator of "biocompatibility". A biocompatible membrane is a membrane more affinous is to living body and that gives less reaction to foreign body.
209
Membrane Bioreactors For Biotechnology And Medical Applications -
Drioli
Materials in blood contact often initiate coagulation processes, affect platelet morphology and function and lead to an immunologic answer of the body. The first event of a complicated sequence that leads to thrombus formation and immunoreactions is the protein adsorption, which plays a mediating role in bioaccumulation, systemic foreign body reactions, tissue regeneration and is, therefore, crucial for biocompatibility and performance of medical devices. Thus, the development of biomedical polymers has to comprise the detailed investigation of protein adsorption. Several studies indicated that platelet activation increases in the positively charged membrane and decreases in the membrane having a microdomain structure in which hydrophilic groups and hydrophobic groups coexist randomly as molecules. Also the complement activation, which in the case of biomaterials proceeds via alternative pathway, is manly noted with cellulose membranes; free hydroxyl groups on the membrane surface is bonded with C3b and further with factor B which promotes the activation. Improvements of biocompatibility of cellulose membrane can be obtained by surface modification devoted to graft for example alkyl group to the hydroxyl group. Considering that functions of biointerfaces are mainly based on principles of molecular recognition, several powerful methods of surface functionalization including plasma treatments, grafting of reactive chains, coating of functional layers and others can be used to engineering news modified surfaces [113-114]. Furthermore, in addition to original cellulose membrane, a variety of synthetic polymers less potent in complement activation, can be used clinically. 30 ,.--, tN
E
25
t-
20
0 o) ::L ' - -
.O ~
2offl
15
c
10
0 c
"=
..1o ~ U_
5 0 <
tO
O9
U.I 13..
eo
~, Z
O
a.
O
•..._
~
a.
n
a.
U.I u.I
13_
Membrane Figure 8: Human plasma fibrinogen adsorption on commercial and novel membranes.
210
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
In this respect, study on biocompatibility of commercial and novel polymeric membranes made in our laboratory are running. Preliminary results show that human plasma fibrinogen and immunoglobulin G adsorb to a greatest extent on cellulose acetate membranes. Differently commercial membranes such as polypropylene and novel membranes such as modified polyurethane and polyetherketone membranes adsorb fibrinogen and immunoglobulin G to a leatest extent (Figure 8). On the basis of these preliminary results these membranes seem to have antithrombogenic properties. Further investigations are needed. 4.
CONCLUSIONS
The progresses made in the last years in membrane engineering and in the understanding of basic mechanisms for membrane transport phenomena, have a significant fall out not only in the industrial separation processes, in desalination, in gas treatments but also in biotechnology and in biomedical engineering. The potential advantages of membrane technology over more conventional approaches, include higher efficiency and reduced costs owing to the integration of bioconversion and product purification, thus reducing equipment costs and processing steps as today requested by a process intensification strategy. Enzymatic membranes is also contributing to the growth of new research area, such as nonaqueous enzymology, the use of antibodies as highly specificcatalysts, use of biomimetic catalysts (such as cyclodextrines) and the development of novel biosensor for diagnostic purposes. The results reached in the development of biocatalytic membrane reactors for the industrial applications and for the realization of artificial organs are important examples. Further progresses will be realized by the integration of these systems, by the study of new more selective and resistant organic and inorganic membranes, by their further miniaturization. REFERENCES [ 1]
[2]
[4]
Malmsten M. Protein adsorption at the solid-liquid interface, In Protein Architecture: interfacing, molecular assemblies and immobilization biotechnology Yuri Lvov and Helmuth Mohwald (eds.), pp. 1-24, Marcel Dekker Publisher, USA, 2000 De Bartolo L and Drioli E. Membranes in artificial organs. In Biomedical and Health Research vol. 16: New Biomedical Materials - Basic and Applied Studies, Haris PI and Chapman D, (eds), pp. 167-181, IOS Press: Amsterdam/Berlin/TokjoAVashington, 1998 Takamatsu S et al. Production of L-alanine from ammonium fumarate using immobilized microorganisms - elimination of side reaction Eur. J. Appl. Microbiol. Biotechnol. 1982; 15:147
211
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[5] [6]
[7]
[8] [9] [10]
[11]
[12]
[13] [14]
[15] [16]
[17]
[18] [19]
[201
[21]
[221 [23]
Boyaval P and Corre C. Continuous lactic acid fermentation with concentrated product recovery by ultrafiltration and electrodialysis Biotechnol. Lett. 1987; 9:207-212 Cantarelli C et al. Beverage stabilization through enzymatic removal of phenolics Food Biochem. 1989; 3:203-213 Illanes A et al. Immobilization of lactose for the continuous hydrolysis of whey permeate Bioprocess Eng. 1990; 5:257-262 Ikemi M et al. Sorbitol production in charged membrane bioreactor with coenzyme regeneration system:I. Selective retainment of NADP(H) in a countinuous reaction Biotechnol. Bioeng. 1990; 36:149-154 Battistel E et al. Enzymatic resolution of (S)-(+)-naproxen in a continuous reactor Biotechnol. Bioeng. 1991; 38:659-664 Livingston JA. A novel membrane bioreactor for detoxifying industrial wastewater: II. Biodegradation of 3-chloronitrobenzene in an industrially produced wastewater, Biotech. Bioeng. 1993; 41: 927-936 Timmer JMK and Krompkamp J. Efficiency of lactic acid production by Lactobacillus helveticus in a membrane cell recycle reactor, FEMS Microbiology Review 1994; 14: 29-38 Bouhallab S and Touze C. Continuous hydrolisys of caseino-macropeptide in membrane reactor: kinetic study and gram-scale production of antithrombotic peptides Le Lait 1995; 75:251-258 Moueddeb H et al. Membrane bioreactor for lactic acid production. J. Membr. Sci. 1996; 114:59-71 Alvarez V e t al. Microfiltration of apple juice using inorganic membranes: process optimization and juice stability Can. J. Chem. Eng. 1996; 74:156-159 Giorno L et al. Study of fouling phenomena in apple juice clarification by enzyme membrane reactor Sep. Sci. Technol. 1998; 33:739-756 Giorno L e t al. Study of enzyme membrane reactor for apple juice clarification Fruit Processing 1998; 8:239-241 Boyaval P, Bouhallab S, Lafforgue-Delorme, Goma G. Biorracteurs h membrane, In Les s~.parations par membrane dans les proc~.dds de L'industrie alimentaire, G. Daufin, F. Ren6 and P. Aimar (eds.), pp. 534-572, Lavoisier Tec & Doe, Publishers, London, New York, 1998 Burton SG, Boshoff A, Edwards W, Rose PD. Biotransformation of phenols using immobilized polyphenol oxidase, J. Molecular Catalysis B: Enzymatic 1998; 5: 411416 Giorno L, Drioli E, Carvoli G, Cassano A, Donato L. Study of an enzyme membrane reactor with immobilized fumarase for production of L-malic acid, Biotechnol. Bioeng. 2001; 72(1): 77-84 Liu W, Howell JA, Amot TC, Scott JA. A novel extractive membrane bioreactor for treating bio-refractory organic pollutants in the presence of high concentrations of inorganics. Application to a synthetic acidic effluent containing high concentrations of chlorophenol and salt, J. Membr. Sci 2001; 181 (1) 127-140 Giorno L, Chojnacka K, Donato L, Drioli E. Study of a cell-recycle membrane fermentor for the production of lactic acid by Lactobacillus bulgaricus, Ind. Eng. Chem. Res. 2002; 41:433-440 Matson SL, Quinn JA. Membrane reactors, In Membrane Handbook, WSW Ho and KK Sirkar (eds.), Chapman and Hall Publisher, New York, 1992 Drioli E, Giorno L. Biocatalytic Membrane Reactors: Applications in Biotechnology and the pharmaceutical industry, Taylor & Francis Ltd Pubis London, 1999
212
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[24] [25]
[26]
[27] [28]
[29]
[30] [31] [32]
[33]
[34]
[35] [36] [37]
[38] [39]
[40] [41] [42]
Sirkar KK, Shanbhag V, Kovvali S. Membrane in a reactor: a functional perspective, Ind. Eng. Chem. Res. 1999; 38:3715-3737 Chibata I, Tosa T, Sato T, Application of immobilized biocatalysts in pharmaceutical and chemical industries, in Biotechnology, H.-J.. Rehm and G. Reed, eds., VCHVerlags, Germany, Vol. 7a, Chapt. 12b, pp. 654-708, 1987 Giomo L, Molinari R, Drioli E, Bianchi D, Cesti P. Performance of a biphasic organic/aqueous hollow fiber reactor using immobilized lipase, J. Chem. Tech. Biotech. 1995; 64:345-352 Bower ST, Cuperus FP, Derksen JTP. The performance of enzyme-membrane reactors with immobilized lipase, Enzyme and Microbial Technology 1997; 21:291-296 Varley PG, Pain RH. Relation between stability, dynamics and enzyme activity in 3phosphoglycerate kinases from yeast and Thermus thermophilus, J. MoL Biol. 1991; 220:531-538 Zhang T et al. Enhancement of protein stability by the combination of point mutations in T4 lysozyme, Protein Eng. 1995; 8:1017-1022 Colombo G, Merz KM. Stability and activity of mesophilic subtilisin E and its thermophilic homolog: insights from molecular dynamics simulations, J. Am. Chem. Soc. 1999; 121: 6895-6903 Carrea G, Colombo G. Coupling high enzyme activity and stability: a challenging target, TIBTECH 2000; 18:401-402 Giorno L, Molinari R, Drioli E. Experimental studies on enzyme membrane reactors in oil treatment, In Advances in oils and fats, antioxidants, and oilseed by-products Vol II, S. S. Koseoglu, K. C. Rhee, R. F. Wilson (eds.), pp. 91-94, AOCS Press, Illinois, 1998 Molinari R, Santoro ME, Drioli E. Study and comparison of two enzyme membrane reactors for fatty acids and glycerol production, Ind. Eng. Chem. Res. 1994; 33: 25912599 Giomo L, Drioli E. Catalytic behaviour of lipase free and immobilized in biphasic membrane reactors with different low water-soluble substrates, J. Chem. Tech. Biotechnol. 1997; 64:345-352 Sakaki K, Giorno L, Drioli E. Lipase-catalyzed optical resolution of racemic naproxen in biphasic enzyme membrane reactors, J. Membr. Sci. 2001; 184:27-38 Hoq MM, Yamane T. Continuous hydrolysis of olive oil by lipase in microporous hydrophobic membrane bioreactor, J. Am. Oil Soc. 1985; 62(6): 1016-1021 Giorno L, Drioli E. Biocatalytic membrane reactors: applications and perpsectives, TIBTECH 2000; 18:339-348 Pronk W, Kerkhof PJAM, van Helden C, van't Riet K. The hydrolysis of triglycerides by immobilized lipase in a hydrophilic membrane reactor Biotechnol. Bioeng 1988; 32:512-518 Kang ST, Rhee JS. Characteristics of immobilized lipase-catalyzed hydrolysis of olive oil of high concentration in reverse phase systems, Biotech. Bioeng. 1989; 33: 14691476 Chen CS, Sih CJ. General aspects and optimization of enantioselective biocatalysis in organic solvents: the use oflipases, Angew. Chem. Int. Engl. 1989; 28:695 Klibanov AM. Asymmetric transformations catalyzed by enzymes in organic solvents Acc. Chem. Res. 1990; 23:114-120 Shaw JF, Chang RC, Wang FF, Wang YJ. Lipolitic activities of a lipase immobilized on six selected supporting materials Biotechnol. Bioeng. 1990; 35:132-137
213
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[43]
[44] [45] [46]
[471 [48]
[49]
[50]
[51]
[52] [53] [54]
[55] [56]
[57] [58] [59]
[60]
Yang D, Rhee JS. Continuous hydrolysis of olive oil by immobilized lipase in organic solvent, Biotech. Bioeng. 1992; 40, 748-752 Malcata FX, Garcia HS, Hill CG, Amudson CH. Hydrolysis of butteroil by immobilized liase using hollow-fiber reactor: PartI. Lipase adsorption studies Biotechnol. Bioeng. 1992; 39:647-657 Tanigaki H, Sakata M, Wada H. Hydrolysis of soybean oil by lipase with a bioreactor having two different membranes J. Ferment. Bioeng. 1993; 75:53-57 Nakajima M, Snape JB, Khare SK. Applications of enzymes and membrane technology in fat and oil processing, In Methods in non-Aqueous Enzymology, Munishwar Nath Gupta (ed.), pp. 70-90, Birkhauser Verlag Publisher, Basel-BostonBerlin, 2000 Warmuth W, Wenzig E, Mersmann A. Selection of a support for immobilization of a microbial lipase for the hydrolysis of triglycerides, Bioprocess Eng. 1995; 12:87-91 Lye GJ, Pavlou OP, Rosjidi M, Stuvkey DC. Immobilization of Candida cylindracea lipase on colloidal liquid aphrons (CLAs) and development of a continuous CLAmembrane reactor, Biotechnol. Bioeng. 1996; 51:69-73 Lopez JL and Matson SL. A multiphase/extractive enzyme membrane reactor for production of diltiazem chiral intermediate J. Membr. Sci. 1997; 125:189-211 Drioli E, Giorno L. Catalytic membrane reactors for bioconversion of low water solubility substrates, In Biocatalytic Membrane Reactors: application in the biotechnology and pharmaceutical industry, pp. 153-189, Taylor & Francis Publisher, London, 1999 Derewenda U, Brzozowski AM, Lawson DM, Derewenda ZS. Catalysis at the interface: the anatomy of a conformational change in a triglyceride lipase, Biochemistry 1992; 31:1532-1541 Derewenda U, Swenson L, Green R, Wei Y, Yamaguchi S, Joerger R, Haas MJ, Derewenda ZS. Current progress in crystallographic studies of new lipases from filamentous ftmgi, Protein Eng. 1994; 7: 551-557 Giorno L, Molinari R, Natoli M, Drioli E. Hydrolysis and regioselective transesterification catalyzed by immobilized lipases in membrane bioreactors, J. Membr. Sci. 1997; 125:177-187 Natoli M. Membrane ad alta specificit/l, Ph.D. Thesis, University of Calabria, Dept. Of Chemical Engineering, 1994 Bosley JA, Peilow AD. Immobilization of lipases for use in non-aqueous reaction systems, In Methods in Non-Aqueous Enzymology, Munishwar Nath Gupta (ed.), pp. 52-69 Birkhauser Verlag Publisher, Basel-Boston-Berlin, 2000 Chibata I et al. Immobilized aspartase-containing microbial cells Appl. Microbiol. 1974; 27:878-885 Pastore M and Morisi F. Lactose reduction of milk by fiber-entrapped fl-galactosidase Methods Enzymol. 1976; 44:822-830 Oyama K et al. On the mechanism of the action of thermolysis: kinetic study of the thermolysisn-catalised condensation reaction of N-benzyloxycarbonyl-L-aspartic acid with L-phenylalanine methyl ester J. Chem. Soc. 1981; 11: 356 Takamatsu S et al. Production of L-alanine from ammonium fumarate using immobilized microorganisms - elimination of side reaction Eur. J. AppL MicrobioL Biotechnol. 1982; 15:147 Carasik W and Carrol JO. Development of immobilized enzymes for the production of high-fructose corn syrup Food Technol. 1983; 37:85-91
214
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[61]
[62]
[63]
[64] [65]
[66]
[67] [68] [69]
[70] [71]
[72]
[73] [74] [75] [76] [77] [78] [79]
[80]
Sato T and Tosa T. Optical resolution of racemic amino acids by aminoacylase, In Industrial Application of Immobilized Biocatalysts (Tanaka A. et al. eds.) pp 3-14, Marcel Dekker 1993 Takata I and Tosa T. Production of L-malic acid, in Industrial Application of Immobilized Biocatalysts, A. Tanaka et al. (eds.) pp 53-55, Marcel Dekker Inc. Pubis 1993 Pencreatc'h G, Baratti JC. Activity of Pseudomonas cepacea lipase in organic media is greatly enhanced after immobilization on a polypropylene support, Appl. Microbiol. Biotechnol. 1997; 47:630-635 Bosley JA, Clayton JC. Blueprint for a lipase support: use of hydrophobic controlledpore glasses as model systems, Biotech. Bioeng. 1994; 43:934-938 Vishwanath S, Wang J, Bachas LG, Butterfield DA, Bhattacharyya D. Site-Directed and Random Immobilization of Subtilisin on Functionalized Membranes: Activity Determination in Aqueous and Organic Media. Biotech. & Bioeng. 1998; 60:608-616 Butterfield DA, Bhattacharyya D, Daunert S, Bachas L G. Catalytic Biofunctional membranes Containing Site-Specifically Immobilized Enzyme Arrays. J. Membrane Sci. 2001; 181:29-37 Bayley J. E., Ollis D. F., Biochemical Engineering Fundamentals, 2nd edn, pp. 202227, McGraw-Hill, New York, 1986 Sheldon RA, Chirality and biological activity, in Chirotechnology, Marcel Dekker, inc., pp. 39-72, New York, Basel, Hong Kong, 1993 Drioli E, Giorno L, Catalytic membrane reactors for production of pure optically active compounds, in Biocatalytic Membrane Reactors, E. Drioli and L. Giorno (eds), pp. 113-136, Taylor & Francis Ltd Pubis London, 1999. Li N, Giorno L, Drioli E. Effect of immobilization site and membrane materials on multiphasic enantiocatalytic enzyme membrane reactors, Submitted, 2002 Keurentjes JTF, Voermans FJM. Membrane separations in the production of optically pure compounds, In Chirality in industry II, AN Collins, GN Sheldrake, J Crosby, (eds.), pp. 157-180, John Wiley & Sons Ltd Publisher, 1997 Nidezky B, Schmidt K, Neuhauser W, Haltrich D, Kulbe K. D. Application of charged ultrafiltration membranes in continuous enzyme-catalyzed processes with coenzyme regeneration, In "Separation for Biotechnology 3", D. L. Pyle (ed.), pp. 351-357, Royal Society of Chemistry, London, 1994 Boyaval P, Corre C. Continuous lactic acid fermentation with concentrated product recovery by ultrafiltration and electrodialysis, Biotechnol. Lett. 1987; 9:207-212 Schlincher LR, Cheryan M. Reverse osmosis of lactic acid fermentation broths, J. Chem. Technol. Biotechnol. 1990; 49:129-136 Timmer J M K, van der Horst W C, Robbertsen T. Transport of lactic acid through reverse osmosis and nanofiltration membranes, J. Membr. Sci. 1993; 85:205-216 Curcio E, Criscuoil A, Drioli E. Membrane crystallizer, Ind. Eng. Chem. Res. 2001; 40:2679-2674 Rogella F. Cost reductions will make MBR realistic even for large plants, Desalination and Water Reuse 2002; 11(4), 9-11 van der Roest H. Dutch drive scale-up of MBR technology, Desalination and Water Reuse 2002; 11(4): 11-15 Chiemchaisri C and Yamamoto K. Biological nitrogen removal under low temperature in a membrane separation bioreactor Water Sci. Technol. 1993; 28:325-333 Magara Y e t al. Biological denitrification system with membrane separation for collective human excreta treatment plant Water Sci. Technol 1992; 25:241-251
215
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[81]
Zoh K-D et al. Treatment of hydrolysates of the high explosives hexahydro-l,3,5trinitro- 1,3,5-triazine and octahydro-1,3,5,7-tetrazocine using biological denitrification Water Environment Research 1999; 71 (2): 148-155 [821 Ben Aim R. Engineering of submerged membranes: State of the art and questions, In Proceedings of Engineering with membranes, S. Luque J. R. Alvarez (eds.), pp. 41-42, Granada, Spain, June 3-6, 2001 [83] Tazi-Pain A, Schrotter J-C, David N, Buisson H. First year operation of the first full scale sequenced aeration MBR for municipal wastewater treatment in Perthes-enGatinais, In Proceedings of Engineering with membranes, S. Luque J. R. Alvarez (eds.), pp. 43-48, Granada, Spain, June 3-6, 2001 [841 Hubbell JA and Langer R. Tissue engineering. C&EN, American Chemical Society 1995; March 13:42-53 [85] Aebischer P, Goddard M, Signore P, Timpson R. Functional recovery in hemiparkinsonian primates transplanted with polymer encapsulated PC12 cells. Exp Neurology 1994; 126:1 [86] Insulin-free world foundation www.insulin-free.org/ [87] Chick WL, Like AA, Lauris V. Beta-cell culture on synthetic capillaries :an artificial endocrine pancreas. Science 1975; 187:847-856 [88] Chang C-C. Analysis of transport phenomena and kinetics in an extravascular bioartificial pancreas. AICHE J 1996; Sep.: 2668-2682 [89] Catapano G, Iorio G, Drioli E, Lombardi CP, Crucitti F, Doglietto GB, Bellantone R. Theoretical and experimental analysis of a hybrid bioartificial membrane pancreas: a distributed parameter model taking into account starling fluxes. J. Mere. Sci 1990; 52: 351-378 [90] Circe Biomedical Technology. www.circe.bio.com/ [91] De Vos P, Hillebrands JL, De Haan BJ et al.Efficay of prevascularised expanded polytetrafluoroethylene solid support system as a transplantation site for pancreatic islets. Transplantation 1997; 63(6): 824-830 [92] Colton CK, Implantable biohybrid artificial organs. Cell Transpl 1995; 4(4): 415-436 [93] www.isletmedical.com/ [94] Storrs R, Dorian R, King SR, Lakey J, Rilo H. Preclinical developemnt of the islet sheet. Ann NYAcad Sci 2001; 944:252-266 [95] Liver foundation organisation, www.liverfoundation.org/ [96] Michapoulos GK. Liver regeneration: molecular mechanisms of growth control. FASEB J 1990; 4:176-187 [97] Matsumura KN, Robles Guevara G, Huston H, Hamilton WL, Rikimaru M, Yamasaki G, Matsumura MS. Hybrid bioartificial liver in hepatic failure: preliminary report. Surgery 1987; 101(1): 99-103 [98] Bader A, Fruhauf N, Tiedge M, Drinkgen M, De Bartolo L, Borlak JT, Steinhoff G, Haverich A.Enhanced oxygen delivery reverses anaerobic metabolic states in prolonged sandwich rat hepatocytes. Exp Cell Res, 1999, 246:221-232 [99] Catapano G, De Bartolo L, Lombardi CP, Drioli E. The effect of catabolite concentrations on the viability and functions of isolated rat hepatocytes. Int J Artif Organs 1996, 19 (4): 245-250 [100] Sussman NL, Chong MG, Kousayer T, HE D, Shang TA, Whisenand HH, Kelly JH. Reversal of fulminant hepatic failure using an extracorporeal liver assist device. Hepatology 1992; 6:60-65 [101 ] Flendrig LM, La Soe JW, Joerning GGA, Steenbeck A, Karlsen OT, Bov6e WMM, Ladiges NCJJ, te Velde AA, Chamuleau AFM. In vitro evaluation of a novel
216
Membrane Bioreactors For Biotechnology And Medical Applications - Drioli
[102]
[103] [104]
[105]
[106]
[107]
[108]
[ 109]
[ 110]
[111]
[112]
[ 113]
[114]
bioreactor based on an integral oxygenator and a spirally wound nonwoven polyestermatrix for hepatocyte culture as small aggregates. J Hepatology 1997; 26: 1379-1392 Nyberg SL, Remmel RP, Mann HJ, Peshwa MV, Hu W, Verra FB. Primary outperform hep G2 cells as source of biotransformation functions in a bioartificial liver. Ann Surg 1994; 220:59-67 Gerlach JV, Encke J, Hole O, Muller C, Ryan CJ, Neuhaus P. Bioreactor for larger scale hepatocyte in vitro perfusion. Transplantation 1994; 58:948-988 Demetriou AA, Whiting JF, Levenson AM, Chowdhury NR, Schechner R, Michalski S, Feldman D, Chowdhury JR. New method of hepatocyte transplantation and extracorporeal liver support. Ann Surg 1986; 204(3): 259-270 Bader A, Knop E, Boker K, Frfihauf N, Shuttler W, Oldhafer K, Burkhard R, Pychlmayr R, Sewing K-F. A novel bioreactor design for in vitro reconstruction of in vivo liver characteristics. Artif. Organs 1995; 19(4): 368-374 Bader A, FrOhauf N, Zech K, Haverich A, Borlak J. Development of a small scale bioreactor for drug metabolism studies maintaining hepatospecific functions. Xenobiotica 1998; 28:815-825 De Bartolo L, Jarosch-Von Schweder G, Haverich A, Bader A. A novel full-scale fiat membrane bioreactor utilizing porcine hepatocytes: cell viability and tissue specific functions. Biotechnology Progress 2000; 16( 1): 102-108 Bader A, De Bartolo L, Haverich A. High level benzodiazepine and ammonia clearance by fiat membrane bioreactors with porcine liver cells. J Biotech 2000; 81 (23):95-105 De Bartolo L and Bader A. Performance of fiat membrane bioreactor utilizing porcine hepatocytes cultured in an extracellular matrix. In: Berry MN, Edwards AM, eds. The Hepatocyte Review. Kluwer Academic Publishers, Dordrecht/Boston/London, 2000: 525-534 De Bartolo L, Catapano G, Della Volpe C, Drioli E. The effect of surface roughness of microporous membranes on the kinetics of oxygen consumption and ammonia elimination by adherent hepatocytes. J Biomaterials Sci-Pol Edn 1999; 10(6):641-655 De Bartolo L, Morelli S, Bader A, Drioli E. The influence of polymeric membrane surface free energy on cell metabolic functions. J Mater Sci: Materials in Medicine 2001; 12:959-963 De Bartolo L, Morelli S, Bader A, Drioli E. Evaluation of cell behaviour related to physico-chemical properties of polymeric membranes to be used in bioartificial organs. Biomaterials 2002; 23:2485-2497 Ansorge W, Spindler E, Vienken J, Baurrnester U. Membranes and polymer structures - biocompatibility aspects with respect to production limits. Transfus Sci 1993; 14: 199-209. Mandolfo S, Tetta C, David S, Gervasio R, Ognibene D, Wratten ML, Tessore E, Imbasciati E. In vitro and in vivo biocompatibility of substituted cellulose and synthetic membranes. Int J ArtifOrgans 1997; 20(11): 603-609
217
This page intentionally left blank
New Insights into Membrane Science and Technology: Polymeric and Biofunctional Membranes D. Bhattacharyya and D.A. Butterfield (Editors) © 2003 Elsevier Science B.V. All rights reserved.
Chapter 10
Structural and performance characteristics of hemodialysis membranes D. Gao I and W. R. Clark2, 3 .
1Department of Mechanical Engineering/Center for Biomedical Engineering, University of Kentucky, Lexington, KY, and Department of Thermal Science & Energy Engineering, University of Science and Technology of China, Hefei, Anhui, China 2
Renal Division, Baxter Healthcare Corp., McGaw Park, IL
3Nephrology Division, Indiana University School of Medicine, Indianapolis, IN *Corresponding author, Phone: (317) 613-2315; Ext. 327, Fax: (317) 613-2317 e-mail: [email protected] ABSTRACT Recent trends show a progressive increase in the use of modified cellulosic and synthetic dialyzers and a corresponding decrease in the utilization rate of unmodified cellulosic dialyzers. The purpose of this paper is to describe current membrane and dialyzer technology, with the focus almost entirely on modified cellulosic and synthetic membranes. A general overview of membrane-related determinants of dialyzer performance is first presented, followed by a discussion of specific characteristics of some of the more commonly used membranes and dialyzers. 1.
INTRODUCTION
End stage renal disease (ESRD) has been a major medical problem worldwide. According to the United States Renal Data System (USRDS), in the United States alone for year 2000, there were 372,407 people projected for the ESRD, and the life of 281,355 of these ESRD patients was sustained on hemodialysis using an artificial kidney (or so-called hemodialyzer). The total cost of Medicare for ESRD patients was projected to be $14.2 billion in year 2000 in USA.
219
Structural And Performance Characteristics Of H e m o d i a l y s i s M e m b r a n e s - Clark
Dialysate out
Dialysate in
t !
Blood out
Blood in
--? ...__~
I . . . . . . . . . . . .
iIiij11s~slsl
/ /
Header or Manifold 11
.-"
Cross-Sectional View -/ i i
I
I
I
I
I
'otting Region Hollow Fibers
Figure 1. A general schematic diagram of an artificial kidney. An artificial kidney is a countercurrent mass exchanger containing approximately 10,000 hollow fibers made up of porous membranes. This mass exchanger has two compartments: the blood compartment (total internal space of all hollow fibers) and the dialysate compartment (space among the hollow fibers). During the hemodialysis, blood is running through the blood compartment, while the dialysate is flowing counter-currently through the dialysate compartment. The uremic toxins are removed from the blood into the dialysate through both diffusion and convection across the hollow fiber membranes.
In engineering perspectives, an artificial kidney is actually a countercurrent mass exchanger containing approximately 10,000 hollow fibers made up of porous membranes. The schematic diagram of an artificial kidney/dialyzer is shown in Figure 1. This mass exchanger has two compartments: the blood compartment (internal space of the hollow fibers) and the dialysate compartment (external space of the hollow fibers). As their names imply, the blood is introduced through the blood compartment, while the dialysate (cleansing salt-balanced water solution) is flowing counter-currently through the dialysate compartment. The porous membranes (hollow fibers) act as an interface where solute mass transfer between the blood and the dialysate 220
Structural And Performance Characteristics Of Hemodialysis Membranes - Clark
takes place. The overall mass transfer area of an artificial kidney is maximized by the use of these hollow fibers. The countercurrent blood and dialysate flows are introduced to further optimize mass transfer efficiency of the artificial kidney to remove uremic toxins in the blood and excess plasma water from patients suffered from ESRD. As discussed recently (1), the first hollow fiber artificial kidney used clinically in the 1960's consisted of a 1.0 m 2 unmodified cellulosic membrane. Hollow fiber dialyzers with highly permeable polyacrylonitrile (2) and polysulfone membranes (3) were subsequently developed, as were dialyzers with modified cellulosic membranes. Over time there has been a progressive increase in the use of modified cellulosic and synthetic dialyzers, predominantly in the high-efficiency and high-flux segments, respectively, and a corresponding decrease in the utilization rate of unmodified cellulosic dialyzers (4). In fact, these latter dialyzers play a relatively limited role in the current global hemodialysis (HD) market, for which approximately 90 million units are now manufactured annually. The movement away from unmodified cellulosic dialyzers has been driven largely by the desire to extend the molecular weight spectrum of solute removal and to minimize complement activation. In addition, retrospective data suggest outcomes of patients treated with these types of dialyzers are inferior to those achieved with modified cellulosic and synthetic dialyzers (5, 6). The purpose of this paper is to describe current membrane and dialyzer technology. A general overview of membrane-related determinants of dialyzer performance is first presented, followed by a discussion of specific characteristics of some of the more commonly used dialyzers. 2.
H O L L O W FIBER MEMBRANE CHARACTERISTICS
In the assessment of the clinical effects of a particular dialyzer, the membrane justifiably receives the most scrutiny. In this section, several important characteristics determining membrane performance are discussed. These include hollow fiber dimensions, surface area, porosity, and water permeability. 2.1.
Hollow Fiber Dimensions An individual hollow fiber can be viewed as a solid cylinder in which a central region has been removed ("cored out") to form the blood compartment. The process of manufacturing ("spinning") a hollow fiber membrane is a complex one incorporating aspects of a broad array of scientific disciplines, including polymer chemistry, thermodynamics, reaction kinetics, and chemical engineering (7). Although clear differences exist among the various types of
221
Structural A n d Performance Characteristics O f Hemodialysis Membranes - Clark
membranes (see below), a number of common features are evident. From a structural perspective, most hollow fibers have a relatively standard inner (blood compartment) diameter (approximately 180-220 ~tm) and length (approximately 20-24 cm). These parameters are dictated essentially by the operating conditions used during HD and are the result of a compromise between opposing forces. On one hand, a relatively small hollow fiber inner diameter is desirable because it provides a short diffusive distance for solute mass transfer. At a given blood flow rate, a lower inner diameter also provides a higher shear rate, resulting in greater attenuation of blood-side boundary layer effects (8). However, a decrease in hollow fiber inner diameter also has undesirable effects. Fluid flow along the length of a cylinder (i.e., the axial flow) in many situations is governed by the Hagen-Poisseuile equation (9): QB = AP / (8~tL / r~r4)
(1)
In this equation, QB is blood flow rate, AP is axial pressure drop, ~t is blood viscosity, L is fiber length, and r is hollow fiber radius. A specific application of this equation is axial blood flow (i.e., from the arterial to the venous end) in a hollow fiber membrane during HD. A more general form of Eq. 1 is: QB = AP / R
(2)
From Equations 1 and 2, the resistance to blood flow (R) is: R = 8gL / nr 4
(3)
Due to the inverse relationship between R and r4, a small decrease in hollow fiber inner diameter induces a large increase in flow resistance. Eq. 3 also demonstrates that increases in fiber length and hematocrit (~t) are associated with an increase in flow resistance. In turn, as indicated by Eq. 2, an increase in flow resistance results in an increase in axial pressure drop at a constant blood flow rate. In contemporary HD practice, patients with progressively rising hematocrits are treated commonly with large surface area dialyzers of high water permeability. These dialyzers require relatively high blood flow rates (at least 350 mL/min) to derive maximum benefit from a solute removal perspective. The high flow resistance and associated large axial pressure drop in this specific scenario result in significant backfiltration of dialysate into the blood compartment under normal HD operating conditions (10). The combination of significant backfiltration and contaminated dialysate increases the likelihood of cytokine-inducing substance transfer, as has been discussed recently (11). 222
Structural And Performance CharacteristicsOf HemodialysisMembranes -Clark
Modifications in considerations.
hollow
fiber
dimensions
are
constrained
by
these
2.2.
Surface Area For a particular hollow fiber, the inner annular surface represents the nominal blood compartment surface area and is the theoretically maximal area available for blood contact. For the entire group of fibers comprising a dialyzer, total nominal surface area then depends on fiber length, inner diameter, and overall number, the latter of which varies generally from approximately 7,000 to 14,000. A frequently asked question relates to the manner in which membrane surface area is calculated. For the inner annular region of a single hollow fiber described above, the surface area (A) is given by the equation defining the surface area of a cylinder: Afiber
=
2nrL
(4)
Based on assumed values of r = 100 gm (10 -4 m) and L = 24 cm (0.24 m), the surface area of an individual hollow fiber can be calculated as: Afiber
(2n)(10 -4 m)(0.24 m) = 1.51x10 -4 m 2 "-
For the large surface area dialyzers routinely used now, the total number of fibers (N) typically used is approximately 12,000. Therefore:
(5)
Adialyzer-- ( A f i b e r ) ( N )
= (1.51x10 -4 m2)(12,000) = 1.81 m 2 2.3.
Membrane Pore-Related Characteristics To provide rough quantitative estimates of pore-related parameters for a hollow fiber membrane used in HD, the straight cylindrical pore model can be used (12). As shown in Figure 2, this model assumes a membrane's pores all have the same radius (rp), which is larger than the radius (rs) of the hypothetical solute shown. In addition, the directional orientation of the cylinders is assumed to be perpendicular to the flow of blood and dialysate. As noted above, the Hagen-Poisseuile equation governs fluid flow through cylinders in many situations, applying also to transmembrane ultrafiltrate flow through the pores in this membrane model. As such, the rate of ultrafiltrate flow is directly related to the fourth-power of the pore radius (i.e., 1"4) at a constant transmembrane
223
Structural And Performance Characteristics Of Hemodialysis Membranes -Clark
pressure. Thus, although the number of pores also influences water permeability (i.e. hydraulic conductivity), the membrane characteristic that most directly influences water permeability is mean pore size. On the other hand, the diffusive properties of a dialysis membrane are determined mainly by the the porosity (pore density) and, to some extent, the pore size (13). Based on the cylindrical pore model described above, membrane porosity is directly proportional to both the number of pores and the square of the pore radius (r2). Therefore, diffusive permeability is also strongly dependent on pore size but not as strongly as is water permeability. The major pore-related determinants of flux (r4) and diffusive permeability (number of pores, r 2) differ sufficiently that the two properties can be independent of one another for a
1 --
oo!101° .
10~ 10= Pore Radius Rp (A) 2Rs--~
O
O
lOP
]
2Rp
!
Figure 2: Pictorial representation of idealized, equal-sized pore membrane model (lower panel) and the associated relationship between pore size and solute permeability (upper panel). Reprinted with permission from (12) particular hemodialysis membrane. Specifically, cellulosic high-efficiency dialyzers typically have high diffusive permeability values for small solutes but low water permeability. On the other hand, there are examples of high-flux 224
Structural And Performance Characteristics O f Hemodialysis Membranes - Clark
dialyzers that have significantly lower small solute diffusive permeabilities than comparably sized dialyzers of much lower water permeability. It should be noted that HD membranes used in actual clinical practice demonstrate pores having a distribution of radii and tortuous (non-cylindrical) structures. As recently discussed (13), this pore size distribution and its influence on a membrane's sieving properties may differ significantly among the different dialysis membranes (Figure 3). The membrane represented by curve A has a large number of relatively small pores while the membrane represented by curve B has a large number of relatively large pores. Based on the relatively narrow pore size distributions, the solute sieving coefficient vs molecular weight profile for both membranes has the desirable sharp cut-off, similar to that of the native kidney. However, the molecular weight cut-off for membrane A (approximately 10 kDa) is
, .A
I0'
1.0
B
/", .~ a
!,
0.8
'
I
0
J: V \
0
0.,
@.
.o
....
0
lO
20
0,0
~
30
40
50
60
IO~
104
i0 ~
SotutcMolecularweight (Daltons)
Average pore Diameter (Angstroms)
Figure 3: Pore size distribution and solute sieving coefficient profiles for three hypothetical dialysis membranes. The left diagram shows the number of pores as a function of mean pore size while the fight diagram shows the relationship between solute sieving coefficient and molecular weight. Reprinted with permission from (13).
consistent with a high-efficiency membrane while that of membrane B (approximately 60 kDa) is consistent with a high-flux membrane. In addition, primarily due to the large number of pores, both membranes would be expected to demonstrate favorable diffusive transport properties. On the other hand, membrane C exhibits a pore size distribution that is unfavorable from both a
225
Structural And Performance Characteristics Of Hemodialysis Membranes - Clark
diffusive transport and sieving perspective.