ME T H O D S
IN
MO L E C U L A R BI O L O G Y
Series Editor John M. Walker School of Life Sciences University of Hertfordshire Hatfield, Hertfordshire, AL10 9AB, UK
For other titles published in this series, go to www.springer.com/series/7651
TM
Light Microscopy Methods and Protocols
Edited by
Hélio Chiarini-Garcia Laboratory of Structural Biology and Reproduction, Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil
Rossana C.N. Melo Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil
Editors Hélio Chiarini-Garcia Department of Morphology Federal University of Minas Gerais Belo Horizonte, MG 31270-901, Brazil
[email protected]
Rossana C.N. Melo Department of Biology Federal University of Juiz de Fora Juiz de Fora, MG 36036-900, Brazil
[email protected]
ISSN 1064-3745 e-ISSN 1940-6029 ISBN 978-1-60761-949-9 e-ISBN 978-1-60761-950-5 DOI 10.1007/978-1-60761-950-5 Springer New York Dordrecht Heidelberg London Library of Congress Control Number: 2010936902 © Springer Science+Business Media, LLC 2011 All rights reserved. This work may not be translated or copied in whole or in part without the written permission of the publisher (Humana Press, c/o Springer Science+Business Media, LLC, 233 Spring Street, New York, NY 10013, USA), except for brief excerpts in connection with reviews or scholarly analysis. Use in connection with any form of information storage and retrieval, electronic adaptation, computer software, or by similar or dissimilar methodology now known or hereafter developed is forbidden. The use in this publication of trade names, trademarks, service marks, and similar terms, even if they are not identified as such, is not to be taken as an expression of opinion as to whether or not they are subject to proprietary rights. Printed on acid-free paper Humana Press is part of Springer Science+Business Media (www.springer.com)
Preface Of all scientific instruments, probably none has had more applications in the life sciences than the light microscope. Advances in microscope instrumentation, sample preparation and imaging techniques have been producing fundamental insights into the functions of cells and tissues. The protocols in Light Microscopy: Methods and Protocols cover a variety of brightfield and fluorescence microscopy-based approaches central to the study of a range of biological questions. The book provides information on how to prepare cells and tissues for microscopic investigations, including detailed staining procedures and how to analyze images and interpret results accurately. Techniques are presented in a friendly, step-by-step fashion with helpful information and useful tips. Section I covers selected applications of bright-field microscopy to the study of animal and plant biology. Section II covers the fundamental principles of fluorescence microscopy as well as its applications to multiple fields including immunology, ecology, cancer biology and cell signaling. Light Microscopy: Methods and Protocols addresses different needs of researchers, who are exploring the microscopic and intriguing world of the cell. We thank Prof. John M. Walker and the staff at Humana Press for their invitation, editorial guidance, and assistance throughout the preparation of this book for publication. We also would like to express our sincere appreciation and gratitude to the contributors for sharing their precious laboratory expertise with the microscopy research community. Hélio Chiarini-Garcia Rossana C.N. Melo
v
Contents Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
v
Contributors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ix
SECTION I 1.
2.
BRIGHT-FIELD MICROSCOPY APPLICATIONS
Glycol Methacrylate Embedding for Improved Morphological, Morphometrical, and Immunohistochemical Investigations Under Light Microscopy: Testes as a Model . . . . . . . . . . . . . . . . . . . . . . . . . . . Hélio Chiarini-Garcia, Gleydes Gambogi Parreira, and Fernanda R.C.L. Almeida
3
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . Gerluza Aparecida Borges Silva, Adriana Moreira, and José Bento Alves
19
3.
Large Plant Samples: How to Process for GMA Embedding? . . . . . . . . . . . Élder Antônio Sousa Paiva, Sheila Zambello de Pinho, and Denise Maria Trombert Oliveira
37
4.
Image Cytometry: Nuclear and Chromosomal DNA Quantification Carlos Roberto Carvalho, Wellington Ronildo Clarindo, and Isabella Santiago Abreu
. . . . . . .
51
5.
Histological Approaches to Study Tissue Parasitism During the Experimental Trypanosoma cruzi Infection . . . . . . . . . . . . . . . . . . . . Daniela L. Fabrino, Grazielle A. Ribeiro, Lívia Teixeira, and Rossana C.N. Melo
6.
Intravital Microscopy to Study Leukocyte Recruitment In Vivo . . . . . . . . . . Vanessa Pinho, Fernanda Matos Coelho, Gustavo Batista Menezes, and Denise Carmona Cara
SECTION II
69
81
FLUORESCENCE MICROSCOPY APPLICATIONS
7.
Introduction to Fluorescence Microscopy . . . . . . . . . . . . . . . . . . . . . Ionita C. Ghiran
8.
Using the Fluorescent Styryl Dye FM1-43 to Visualize Synaptic Vesicles Exocytosis and Endocytosis in Motor Nerve Terminals . . . . . . . . . . . . . . 137 Ernani Amaral, Silvia Guatimosim, and Cristina Guatimosim
9.
Imaging Lipid Bodies Within Leukocytes with Different Light Microscopy Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149 Rossana C.N. Melo, Heloisa D’Ávila, Patricia T. Bozza, and Peter F. Weller
vii
93
viii
Contents
10.
EicosaCell – An Immunofluorescent-Based Assay to Localize Newly Synthesized Eicosanoid Lipid Mediators at Intracellular Sites . . . . . . . . . . . 163 Christianne Bandeira-Melo, Peter F. Weller, and Patricia T. Bozza
11.
Nestin-Driven Green Fluorescent Protein as an Imaging Marker for Nascent Blood Vessels in Mouse Models of Cancer . . . . . . . . . . . . . . . . 183 Robert M. Hoffman
12.
Imaging Calcium Sparks in Cardiac Myocytes . . . . . . . . . . . . . . . . . . . 205 Silvia Guatimosim, Cristina Guatimosim, and Long-Sheng Song
13.
Light Microscopy in Aquatic Ecology: Methods for Plankton Communities Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215 Maria Carolina S. Soares, Lúcia M. Lobão, Luciana O. Vidal, Natália P. Noyma, Nathan O. Barros, Simone J. Cardoso, and Fábio Roland
14.
Fluorescence Immunohistochemistry in Combination with Differential Interference Contrast Microscopy for Studies of Semi-ultrathin Specimens of Epoxy Resin-Embedded Samples . . . . . . . . . . . . . . . . . . 229 Shin-ichi Iwasaki and Hidekazu Aoyagi
Subject Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241
Contributors ISABELLA SANTIAGO ABREU • Laboratory of Cytogenetics and Cytometry, Department of General Biology, Federal University of Viçosa, Viçosa, MG, Brazil FERNANDA R.C.L. ALMEIDA • Laboratory of Structural Biology and Reproduction, Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil JOSÉ BENTO ALVES • University of Uberaba, Uberaba, MG, Brazil ERNANI AMARAL • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil HIDEKAZU AOYAGI • Advanced Research Center, School of Life Dentistry at Niigata, The Nippon Dental University, Niigata, Japan CHRISTIANNE BANDEIRA-MELO • Laboratory of Inflammation, Carlos Chagas Filho Institute of Biophysics, Federal University of Rio de Janeiro, Rio de Janeiro, RJ, Brazil NATHAN O. BARROS • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil PATRICIA T. BOZZA • Laboratory of Immunopharmacology, IOC, Oswaldo Cruz Foundation, Rio de Janeiro, RJ, Brazil SIMONE J. CARDOSO • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil CARLOS ROBERTO CARVALHO • Laboratory of Cytogenetics and Cytometry, Department of General Biology, Federal University of Viçosa, Viçosa, MG, Brazil HÉLIO CHIARINI-GARCIA • Laboratory of Structural Biology and Reproduction, Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil WELLINGTON RONILDO CLARINDO • Laboratory of Cytogenetics and Cytometry, Department of General Biology, Federal University of Viçosa, Viçosa, MG, Brazil FERNANDA MATOS COELHO • Laboratory of Immunopharmacology, Department of Biochemistry and Immunology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil DENISE CARMONA CARA • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil HELOÍSA D’ÁVILA • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil DANIELA L. FABRINO • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil IONITA C. GHIRAN • Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA CRISTINA GUATIMOSIM • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil
ix
x
Contributors
SILVIA GUATIMOSIM • Department of Physiology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil ROBERT M. HOFFMAN • AntiCancer Inc and Department of Surgery, University of California, San Diego, CA, USA SHIN-ICHI IWASAKI • Advanced Research Center, School of Life Dentistry at Niigata, The Nippon Dental University, Niigata, Japan LÚCIA M. LOBÃO • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil ROSSANA C.N. MELO • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil GUSTAVO BATISTA MENEZES • Laboratory of Immunopharmacology, Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil ADRIANA MOREIRA • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil NATÁLIA P. NOYMA • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil DENISE M.T. OLIVEIRA • Department of Botanic, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil ÉLDER ANTÔNIO SOUSA PAIVA • Department of Botanic, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil GLEYDES GAMBOGI PARREIRA • Laboratory of Structural Biology and Reproduction, Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil SHEILA ZAMBELLO DEPINHO • Department of Biostatistics, Institute of Biosciences, UNESP – Universidade Estadual Paulista, Botucatu, SP, Brazil VANESSA PINHO • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil GRAZIELLE A. RIBEIRO • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil FÁBIO ROLAND • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil GERLUZA APARECIDA BORGES SILVA • Department of Morphology, ICB, Federal University of Minas Gerais, Belo Horizonte, MG, Brazil MARIA CAROLINA S. SOARES • Laboratory of Aquatic Ecology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil LONG-SHENG SONG • Division of Cardiovascular Medicine, Department of Internal Medicine, University of Iowa Carver College of Medicine, Iowa City, IA, USA LÍVIA TEIXEIRA • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil LUCIANA O. VIDAL • Laboratory of Cellular Biology, Department of Biology, ICB, Federal University of Juiz de Fora, Juiz de Fora, MG, Brazil PETER F. WELLER • Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA
Section I Bright-Field Microscopy Applications
Chapter 1 Glycol Methacrylate Embedding for Improved Morphological, Morphometrical, and Immunohistochemical Investigations Under Light Microscopy: Testes as a Model Hélio Chiarini-Garcia, Gleydes Gambogi Parreira, and Fernanda R.C.L. Almeida Abstract Glycol methacrylate (GMA), a water and ethanol miscible plastic resin, is a medium handy to use for light microscopy embedding that has a number of advantages than paraffin embedding. The GMA improves the histological, morphometrical, and immunohistochemical evaluations, mainly due to the accurate assessment of cytological details. This chapter focuses on our experience in the GMA processing and describes in detail the fixation, embedding, and staining methods that we have been using for testes evaluations. Key words: Glycol methacrylate, fixation, embedding, light microscopy, testes.
1. Introduction The first studies that described the seminiferous epithelium structure as we know today emerged at the end of 1950 and starting of 1960 (1–3). These studies demonstrated that germ cells in different steps of the spermatogenic process – spermatogonial, spermatocytary, and spermiogenic – are distributed in a wellorganized way along the seminiferous tubules. These initial studies were developed mainly in testes fixed in Zenker or Bouin solutions and embedded in paraffin. Although these histological processing were adequate to describe the development of the acrossomal system (stained in purple by periodic acid-Schiff), to H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_1, © Springer Science+Business Media, LLC 2011
3
4
Chiarini-Garcia, Parreira, and Almeida
differentiate steps of the spermatocyte and able to distinguish some spermatogonia, accurate morphological details of all process, mainly those related to spermatogonial subtypes, were not clearly demonstrated. At the same time, another morphological method was also standardized; this was very important for the spermatogonial biology studies. It was a whole mount method that analyzed the testis in toto and classified the spermatogonial subtypes by their grouping, that is, if they were single, in pair or aligned in 4 up to 32 spermatogonia together in the same clone (4). However, this method was not accurate to distinguish the spermatogonial subtypes by their morphology. More recently, it was demonstrated that a morphological method used in the past for important male reproduction researches (5), applying glutaraldehyde fixation, araldite embedding, and semi-thin sections, could be used for high-resolution light microscopic evaluations of the spermatogonial cell in different morphological and morphometrical approaches (6–13). However, as this method is a pre-preparation for transmission electron microscopy process, fragments have to be very small (2 mm2 ), allowing adequate penetration of fixatives and resins into tissues. A method that combines different advantages of the method exposed above and allows morphological, morphometrical, and immunohistochemical studies in semi-thin or thick sections, in large fragments and with satisfactory morphology, is the one that uses fixation with paraformaldehyde and/or glutaraldehyde and embedding in plastic resin based in glycol methacrylate (GMA). The GMA embedding has been used to present some advantages over the usual methods (14–16), namely (a) fast processing, (b) hydrosoluble, (c) easy handling, (d) infiltration and polymerization at room temperature, (e) possible to obtain semi-thin section (0.5 μm), (f) less distortion and artifacts, and (g) better resolution over light microscopy. We have been using GMA embedding since 1990 for different studies, such as mast cells (17–22), male reproduction (23, 24), equine endometrium (25), and aquatic organisms (26). Here, we present, in detail, our experience in GMA processing, and its tricks, focusing on testis preparation for its high performance studies.
2. Materials 1. Heparin (Liquemine, Roche). 2. Sodium thiopental (Thiopentax, Cristalia) intravenous bottle. 3. Three-way stopcock.
GMA for Improved Investigations Under Light Microscopy
5
4. Catheter (Angiocath, BD). 5. Saline (sodium chloride at 0.9%). 6. Phosphate buffered solution of 0.1 M and pH 7.4: Dissolve 1.38 g NaH2 HPO4 .H2 O (0.1 M) in 100 mL distilled water (solution A) and 1.42 g Na2 H.HPO4 (0.1 M) in 100 mL distilled water (solution B). To prepare the buffer at pH 7.4, mix 19 mL of solution A in 81 mL of solution B. Adjust pH with the same solutions. 7. 8% Paraformaldehyde solution: Heat 70 mL distilled water at 60–70◦ C and add 8 g of paraformaldehyde. Mix well and add drops of NaOH (0.1 M) until a clean solution is obtained. Wait to get to room temperature before using. 8. 4% paraformaldehyde in phosphate buffer 0.05 M pH 7.4: Prepare 100 mL solution by mixing 50 mL of 8% paraformaldehyde, freshly prepared in 50 mL of phosphate buffer at 0.1 M and pH 7.4. 9. 5% glutaraldehyde in phosphate buffer 0.05 M pH 7.4: Mix 10 mL of glutaraldehyde (biological grade at 50%) in 50 mL of 0.1 M phosphate buffer at pH 7.4 and complete the volume to 100 mL with distilled water. 10. Karnovsky’s fixative − 2% paraformaldehyde, 2.5% glutaraldehyde in phosphate buffer 0.05 M pH 7.4: We have used the original formula diluted in a half as follows: mix 50 mL phosphate buffer (0.1 M), 5 mL glutaraldehyde (50%), 20 mL paraformaldehyde (10%) and complete the volume to 100 mL with distilled water. 11. Alcohol (from 70 to 100% in distilled water). 12. GMA kit (Historesin, Leica). 13. Plastic mold for embedding. 14. Wooden pin holder. 15. Dentist acrylic resin kit: Mix the powder and the liquid to prepare a viscous medium. Just after this, pour the mixture into the mold holes and immediately put a wooden pin. The resin takes only a couple of minutes to polymerize. 16. Razor blade. 17. Glass knife: prepared with glass strips (400 × 25 × 6.4 mm) from Leica in an LKB Knifemaker, model 7880B. 18. Toluidine blue-borate: 1 g toluidine blue O (Allied Chemical) and 1 g of sodium borate (Na2 B4 O7 anidrous) dissolved in 100 mL distilled water. 19. Erythrosine-Orange-Toluidine: solution A (0.2 g erythrosine, 1 g orange G in 100 mL distilled water); solution B (toluidine blue-borate, detailed above at item 18).
6
Chiarini-Garcia, Parreira, and Almeida
20. Harris’ hematoxylin: mix 1 g hematoxylin, previously dissolved in 10 mL ethanol, with 20 g aluminum potassium sulfate (AlKO8 S2 12H2 O) previously dissolved in 200 mL of heated distilled water. Immediately, add 0.5 g mercury oxide (HgO). Take out from the heater and cool the solution by immersing it cold water. To increase the nuclear contrast, 4% acetic acid can be added to the solution. 21. Mordent solution: add 2% solution of ammonium iron sulfate (FeH8 N2 O8 S2 .6H2 O) in distilled water. 22. Eosin solution: mix 1 g yellow eosin dissolved in 10 mL absolute ethanol in 0.5 potassium dichromate (K2 Cr2 O7 ) dissolved in 80 mL distilled water, followed by 10 mL of saturated solution of picric acid (for saturation, add 1.4 g picric acid in 100 mL distilled water). 23. Period acid: periodic acid at 0.5% in distilled water. 24. Differentiator solution: mix 6 mL of 10% sodium metabisulfite (Na2 O5 S2 ) and 5 mL of chloride acid 1 N (8.35 mL HCl up to 100 mL distilled water) in distilled water and complete the volume up to 100 mL. 25. Schiff reactive: dissolve 1 g basic fuchsin in hot water but without boiling it. Wait to cool down to 50◦ C and then add 10 mL of chloride acid (1 N). Wait to cool down to 25◦ C and add 1 g of sodium metabisulfite. Mix for 1 h and keep it in a dark place for 24 h at room temperature. Keep the final solution at 4◦ C. 26. 5-Bromo-2-deoxyuridine (Sigma) – diluted 6 mg/mL in phophate buffer solution PBS. 27. Colorfrost Plus Microscope Slides (Fisher-Scientific). 28. 0.6% Hydrogen Peroxide (H2 O2 ): Dilute 1 mL of 30% H2 O2 (Sigma-Aldrich) in 49 mL of distilled water. 29. 0.1% Protease (Sigma-Aldrich) diluted in PBS. 30. 10× phosphate-buffered saline: Dissolve 76.0 g NaCl, 3.6 g NaH2 PO4 and 9.94 g Na2 HPO4 in 1000 mL of distilled water. To make 1× PBS, dilute 10× PBS at a 1:9 ratio in distilled water. Adjust to pH 7.4 with HCl. Store in glass bottles at room temperature. 31. 2 N hydrogen chloride (HCl): Dilute 73 mL of HCl to 1 L of solution in distilled water. Store the solution in a glass bottle at room temperature. 32. 0.1 M sodium borate (Na2 B4 O7 ): Dissolve 38.14 g of Na2 B4 O7. 10H2 O (decahydrate) in distilled water, up to 1 L. Mix it in a hot plate, until the salt is completely dissolved. Store the solution in a glass bottle at room temperature.
GMA for Improved Investigations Under Light Microscopy
7
33. Triton X-100 (Sigma-Aldrich). 34. PBST solution: 0.2% Triton X-100 in 1× PBS. Add 200 μL of Triton X-100 to 1 L of PBS. Store the solution in a glass bottle at room temperature. 35. Normal horse serum (Sigma-Aldrich) diluted in PBST . 36. Primary antibody anti-BrdU B44 (BD Biosciences). 37. ABC kit, Mouse IgG (Vector Labs): This kit contains secondary biotinylated antibody and reagents A and B. 38. Normal goat serum (Sigma-Aldrich), diluted in PBST. 39. DAB (3,3 -diaminobenzidine) kit (Vector Labs). 40. Xylene. 41. Entelan medium (Merck). 42. Coverslip.
3. Methods To obtain a good tissue preparation, all the processing steps have to be carefully developed. The sum of minimal defect in some steps can impair the final results. Thus, care has to be taken regarding fixation, embedding, sectioning, and staining. Below, details about each of these steps will be described considering different experimental possibilities. 3.1. Fixation and Storage
Glutaraldehyde is a non-coagulant fixative known to cross-link protein, preserving very well cellular structures. As protein is a universal component of cells, found in membranes and in the cytosol, the glutaraldehyde can fix it as a whole, making the cell as a single interlocking structure (27). Besides, this aldehyde reaction is not limited to protein. It can also react in a lower degree with lipids, carbohydrates, and nucleic acids. Formaldehyde is also another aldehyde used for cell fixation. However, it makes less cross-linking, reducing the tissue meshwork. Considering that the rate of penetration of glutaraldehyde into tissues is very low and the formaldehyde penetration is about five times faster than glutaraldehyde, a combination of both is frequently used, mainly in tissues of difficult penetration and/or in fixation by immersion. This method was described by Karnovsky (28). Fixatives using picric acid and alcohol are coagulant. They can denaturate proteins, permanently modify their structure and affect the tissue resolution. Thus, for morphological evaluation of testes under light microscopy, aldehydes are normally chosen. To avoid lower pH reduction during the fixation procedure and the introduction
8
Chiarini-Garcia, Parreira, and Almeida
of artifacts, a buffering system should be used with the fixative. The main buffers used are phosphate and cacodylate buffer at pH 7.2–7.4. 3.1.1. Testes Fixation by Perfusion of Whole Body
When small animals are used, like rodents, marmosets, opossums or cats, the best method to preserve whole testes is fixing them by the injection of the fixative into the whole body through the circulatory system (11). This method of fixation by perfusion is reached by introducing a catheter into the left ventricle, in the direction of the aorta (Note 1). This needle is also connected to two vials, one with saline and the other with fixative, through a triway plastic pipe. These vials are positioned ∼1.2 m above the heart, reaching a liquid pressure of ∼80 mmHg inside the vascular system. Just before the beginning of the perfusion process, the right atrium is cut for draining the blood and solutions (saline and fixatives) that will be injected. First, the circulatory system is perfused with saline for 5–10 min, cleaning of blood cells (Note 2). Immediately after, it initiates the perfusion with the fixative for about 25–30 min (Note 3). Fifteen minutes before perfusion, heparin is injected intraperitonealy in the proportion of 125 IU/kg of body weight (Note 4). After perfusion, testes should be cut in thin slabs and fixed by immersion for 12–24 h at 4◦ C.
3.1.2. Fixation by Perfusion of Isolated Testis
When testes of large animals are used, like bull, boar, and ram, the whole body perfusion method becomes very expensive. In this way, the orchiectomy is made and only the testes are perfused. The perfusion is made by introducing the catheter into the testicular artery, once this artery is easily identified (Note 5). To avoid blood clumps during the perfusion process, heparin should be added to the saline solution in a proportion of 10 IU/L. The time of saline perfusion should be enough to clean the testis from blood cells (∼5 min) and can be verified visually by the clear liquid running from the testicular vein. The fixation time should be between 20 and 30 min (Note 3). Both, saline and fixative, should be perfused at a pressure of ∼80 mmHg. After perfusion, testes should be cut in thin slabs and fixed by immersion for 12–24 h at 4◦ C.
3.1.3. Testis Fixation by Immersion
If for some reason it is not possible to fix the testis by perfusion, they can be fixed by immersion. However, for good morphological preservation, some cares have to be taken. The use of a fixative solution with combined aldehyde (glutaraldehyde and formaldehyde) can be an alternative solution, once formaldehyde penetrates faster and temporarily stabilizes cellular structures. Glutaraldehyde, which penetrates slowly, arrives later and permanently stabilizes cellular components. In case of the use of glutaraldehyde only as a fixative, only thin slabs (∼1 mm thickness)
GMA for Improved Investigations Under Light Microscopy
9
should be cut from the fragment surface (Note 6). The fixative time by immersion should be for 12–24 h at 4◦ C. 3.1.4. Comments
Regardless of the method used, testes fragments can be kept in buffer after fixation at 4◦ C for a long time (Note 7). Testes fixed by perfusion are used to keep the seminiferous tubules together, preserving the interstitial tissues and possibly making more accurate morphometrical evaluation of testes compounds. Otherwise, testes fixed by immersion normally show seminiferous tubules dispersed. The artifactual spaces among them are provoked by the pressure in a soft tissue during the cut processing in small slabs. For immunohistochemistry evaluations, fixatives with glutaraldehyde are avoided once the resulted cross-link into tissues could block the antibody receptors. The most common fixatives used for immunohistochemical is the formaldehyde, which in spite of not preserving very well morphological details, keeps the tissue receptors exposed for antibodies.
3.2. Embedding
Testes fragments should be progressively dehydrated with alcohol at crescent concentrations (see Note 8), infiltrated and embedded in GMA, as described below: Ethanol 70% Ethanol 85% Ethanol 95% Absolute ethanol Infiltration resin Pure resin Embedding (see Note 9, Fig. 1.1)
3.3. Sectioning
30 min 30 min 30 min 30 min (2×) overnight overnight
The tissue embedded in GMA blocks can be cut from 0.25 μm up to 15 μm thickness (see Note 10). For the best sectioning, excessive resin should be trimmed with a razor blade, keeping a border of approximately 1–2 mm around all tissue. To obtain a nice section, glass knife can be used in a microtome. For distention, sections are floated in distilled water just after the collection at room temperature (see Note 11). The section should be picked up with a slide and transferred to a hot plate (70–80◦ C) until the water droplet over the section evaporates (Fig. 1.1).
3.4. Staining 3.4.1. For Morphological and Morphometrical Studies
For different histological evaluations, testes embedded in GMA can be stained by different methods, namely, toluidine blueborate, erythrosine-orange-toluidine, hematoxylin-eosin, and PAS-hematoxylin. As tissues embedding in GMA present weak staining, when compared with those embedded in paraffin
10
Chiarini-Garcia, Parreira, and Almeida
Mi
HP WB A
Re BL
T
GK
B
Mo
7
C
4
1
WP BL
D
Fig. 1.1. In a, microtome (Mi, Reichert Jung, model 1140/Autocut) setup with GMA block (BL) and glass knife (GK) for sectioning, evident in the circle (detail in B). Moreover, a room temperature water bath (WB) for section distension and a hot plate (HP) for drying and section adherence on the slide. In b, detail of the circle in a, showing also the tissue (T) facing the glass knife. In c, a slide with histological sections of 7, 4, and 1 μm thickness, showing the different staining intensity. d shows components used for GMA embedding: plastic mold (Mo), wooden pin (WP), dentist resin (Re), and tissue into the molds under polymerization (black arrows) and block with tissue (BL) after polymerization and with wooden pin adherence.
(13–15), the methods described below were modified from those originally used to stain testes embedded in paraffin, with the purpose of obtaining good staining and contrast and, consequently,
GMA for Improved Investigations Under Light Microscopy 7µm
4µm
1µm
s
AT
s
s
A
11
C
B
s
HE s
s
D
E
s
s
s
EOGAT
F
H
G
I s
s s
PAS
J
K
L
Fig. 1.2. Photomicrographies of the human seminiferous epithelium stained with toluidine blue-borate (a–c), hematoxilin-eosin (d–f), erythrosine-orange G-toluidine (g–i), and PAS-hematoxylin (j–l). These pictures show the relationship between resolution and section thickness. In the first column (a, d, g, j), sections were obtained at 7 μm and the spermatocyte (S) heterochromatin were not easily visible. In the second column (b, e, h, k), sections at 4 μm thickness allowed the observation of details from the spermatocyte (S) heterochromatin. One micrometer is the thickness of sections in the third column (c, f, i, l) and more details can be observed in the spermatocyte (S) nuclei. The PAS stained in purple grains in the cytoplasm of the germ cells, which were differently observed depending on the section thickness. While in J they were compactly observed (arrows), in k some grains can be seen and in L the PAS grains were individually observed. Bar: 10 μm.
best resolution under light microscopy (Fig. 1.2). All the staining methods presented below can be changed depending on the tissue, fixative applied, and the section thickness. Tests should be made to standardize them for each specific experimental approach. 3.4.1.1. Toluidine Blue-Borate
Although it presents just one color, exceptions for metachromatic cells/tissues (mast cells, goblet cells, mucus), it is elected as one of the best staining for tissues embedded in GMA for good contrast and resolution, mainly in black & white pictures (see Note 12, Fig. 1.2). 1. Put several drops of toluidine blue-borate over the sections on a slide for 1 min.
12
Chiarini-Garcia, Parreira, and Almeida
2. Rinse the slide with running water to clean the excessive staining (see Note 13). 3. Remove excess water by gently pressing the slide, with the tissue facing down, over a piece of filter paper. 4. Allow the slide to dry completely at room temperature and mount with a coverslip using any conventional medium. 3.4.1.2. Hematoxylin-Eosin
Even after some methodological modification, the H&E staining does not present very nice contrast in GMA-embedded sections. The contrast can be increased on thicker sections; however, the resolution can be impaired. In order to intensify the H&E staining, the use of a mordant solution (see Note 14) and the increase of the staining time (Fig. 1.2) are recommended. 1. Put several drops of mordent solution over sections on a slide for 10 min. 2. Rinse the slide with running water for 5 min. 3. Put several drops of hematoxilin solution over sections for 15 min. 4. Rinse the slide with running water for 5 min. 5. Put several drops of eosin solution over sections for 30 s. 6. Rinse the slide with running water to clean the excessive staining (see Note 13). 7. Remove excess water by gently pressing the slide, with the tissue facing down, over a piece of filter paper. 8. Allow the slide to dry completely at room temperature and mount with a coverslip using any conventional medium.
3.4.1.3. ErythrosineOrange-Toluidine
In paraffin, the trichrome stains are used for cytoplasmic stains combined with nuclear stains. However, these methods do not work in GMA. We have applied an alternative method which has been used with relative success for tissues embedded in GMA (Fig. 1.2). 1. Put several drops of solution A (erythrosine-orange) over sections on a slide for 10 min. 2. Rinse the slide with running water to clean the excessive staining. 3. Put several drops of solution B (toluidine blue-borate) over sections on a slide for 1 min. 4. Rinse the slide with running water to clean the excessive staining (see Note 13). 5. Remove excess water by gently pressing the slide, with the tissue facing down, over a piece of filter paper. 6. Allow the slide to dry completely at room temperature and mount with a coverslip using any conventional medium.
GMA for Improved Investigations Under Light Microscopy
3.4.1.4. PAS-Hematoxylin
13
Used for neutral amino-glycol localization in tissues. Although this method has been commonly used for acrosomal identification in the testis, here it was used to show PAS-positive granules into the cytoplasm of germ cells (Fig. 1.2). 1. Put several drops of periodic acid solution over sections for 20 min. 2. Rinse the slide with distilled water for 5 min. 3. Put several drops of Schiff reactive over sections for 60 min. 4. Rinse in three baths of differentiator solution in a total of 3 min. 5. Rinse the slide with running water for 30 min. 6. Put several drops of hematoxylin for 10 min. 7. Rinse the slide with running water for 30 min. 8. Remove excess water by gently pressing the slide, with the tissue facing down, over a piece of filter paper. 9. Allow the slide to dry completely at room temperature and mount with a coverslip using any conventional medium.
3.4.2. For Immunohistochemical Studies
Some immunohistochemical studies can be performed using GMA. As an example, 5-bromo-2-deoxyuridine (BrdU) method has been used to study cellular cycle in the testis (29). During cellular division, BrdU incorporates into the DNA chain and can be followed during the cellular cycle using antibody against BrdU. BrdU was injected intraperitonealy in a dose of 60 mg/kg of body weight one hour before killing the mice. The spermatogonia that divided during this time incorporated the BrdU in their nuclei. After one hour, the mice were fixed by perfusion and the testes embedded in GMA as described above. Testes sections of 5 μm thickness were used for the present evaluation. Examples of immunohistological staining of BrdU in spermatogonia are presented in Fig. 1.3.
A
B
Fig. 1.3. a and b: Sections of the seminiferous epithelium of a mouse showing immunohistochemical staining for BrdU in spermatogonia (arrows). In b, note the staining concentrated in the inner border of the envelope nuclear. Bar: 10 μm.
14
Chiarini-Garcia, Parreira, and Almeida
To perform the BrdU immunostaining in GMA, the following steps must be taken: 1. Put slides in distillated water for 1 min. 2. Prepare 0.6% H2 O2 in distilled water and immerse slides in a 50 mL Coplin jar, at room temperature, for 5 min. 3. Wash slides in distilled water for 1 min. 4. In a moisture chamber, incubate sections with 0.1% protease in 1× PBS, for 60 min, at room temperature. 5. Rinse slides two times with 1× PBS for 5 min each time. 6. Denature sections with 2 N HCl for 50 min, in a Coplin jar, at room temperature. 7. Neutralize the acid with 0.1 M Na2 B4 O7 for 2 min. 8. Wash slides two times in PBST , for 5 min each. Prepare the blocking solution while washing slides. 10% NHS (normal horse serum) in PBST is used as blocking solution, before the primary antibody incubation step. 9. Incubate slides with the blocking solution, in a moisture chamber, during 30 min at 37◦ C. While incubating with the NHS, prepare the 1◦ antibody (B44) in 10%NHS in PBST , following 1:200 dilution. 10. Pipet off the blocking solution from sections to be tested, keeping one section as a negative control. Add 100 μL of the antibody prepared to each section on the slide. Incubate slides with the 1◦ antibody for 60 min, at 37◦ C, in a moisture chamber. 11. After incubation with the B44, wash slides twice with PBST , for 5 min each. While washing slides, prepare the secondary biotinylated antibody in 10% NGS (normal goat serum) with PBST , following 1:1000 dilution. 12. Add 100 μL of secondary antibody to each section and keep it at 37◦ C, for 30 min, in a moisture chamber. Prepare the ABC reagent while incubating with the 2◦ antibody. Add 20 μL of Reagent A plus 20 μL of Reagent B in 800 μL 1× PBS. Keep the same proportion for all reagents when preparing more than 1000 μL. 13. Wash slides twice with PBST , for 5 min each. 14. Add 100 μL of the prepared Reagent ABC to sections and incubate at 37◦ C, for 30 min, in a moisture chamber. 15. Wash slides twice with 1× PBS, for 5 min each. Prepare the DAB while washing slides for the last time in 1× PBS. The DAB should be prepared following the manufacter’s instructions. When using Vector Labs DAB kit, add two drops of buffer, plus four drops of DAB, plus two drops of H2 O2 to 5 mL distilled water.
GMA for Improved Investigations Under Light Microscopy
15
16. Add one drop of DAB to each section, for 30 s, then wash in distilled water. Keep slides in distilled water while staining the other slides with DAB. 17. Stain the section with hematoxylin, diluted 1:1 in distilled water, for 1 min. Wash with tap water for 5 min. 18. Dehydrate slides using 95% ethanol and 100% ethanol for 3 min each and emerge slides in xylene for 5 min. 19. Mount slides with conventional mounting media and coverslip. 3.5. Photomicrography
Photomicrographies were obtained using a microscopy BX-51 in which a Q-Color 3 digital camera from Olympus was connected. The obtained images were transferred to a computer through the Image-Pro Express (Media Cybernetics) software and adjusted for resolution (1000 dpi), sharpness, contrast, brightness, and gray levels using Photoshop (Adobe System, Inc., Mountain View, CA). Plates were organized and characters added using the Adobe Illustrator software.
4. Notes 1. The circulatory system should be closed and under pressure for adequate entrance of the fixative in the aorta, reaching testes afterwards and exiting by the right atrium. The most common mistake during the perfusion is the needle placing. If a hole is made in the interventricular septum – a thin wall that divides the two ventricular chambers – the pressure comes down and the fixative will enter the right ventricle reaching the pulmonary circulatory system. This mistake will decrease the pressure inside the general circulatory system and impair testes fixation. 2. The effective time for saline perfusion is the one necessary for cleaning the blood vessels. When the saline that comes out of the right atrium is clean, it is time to stop the saline and start the fixative perfusion. 3. As the fixative penetrates the testis by the blood vessels, it reaches all testicular compartments and cells by diffusion. These processes take a long time and require a fixation time of at least 25 min, even if the animal body is apparently well perfused. 4. The use of heparin is very important for a successful testes perfusion success (30), as it avoids blood clumps into blood vessels during the fixative perfusion.
16
Chiarini-Garcia, Parreira, and Almeida
5. For testes fixation directly through the testicular artery, to avoid reflux of solutions the testicular artery has to be tied with a line, but not too strong to avoid cutting the vessel wall. 6. For testes fixation by immersion, the albuginea capsule has to be taken out once it is a dense connective tissue that avoids fixative penetration by diffusion. If it is not possible to take the albuginea completely out, small holes or cuts should be done on it, around the entire testis. Another alternative is to cut the testis in large slabs and put them in the fixative by immersion. After 12–24 h, only thin slabs from the surface (1–2 mm thickness) of the big slabs should be taken. The rest has to be discharged. Fixation by immersion requires a fixative volume of at least 30 times greater than the tissue volume, for adequate fixation. 7. If it is necessary to keep tissues in phosphate buffer for a long time, even under 4◦ C, add one drop of glutaraldehyde in the flask to avoid fungi growth. 8. If for any reason it is not possible to dehydrate with alcohol, the water can be eliminated by crescent concentration of resin in water, such as 50, 70, 80, 90% and finally pure resin. 9. Testes fragments should be smaller than the cutting surface of the block and centrally positioned (Fig. 1.1). During the cutting procedure, the border of blocks could be damaged, impairing the histological analyses of the whole tissues. After resin polymerization, a support should be attached to each resin block to firmly attach them in the microtome for sectioning. Originally aluminum support has been used. However, in our laboratory we have used wooden pins as support, which are attached to the resin block using dentist resin (Fig. 1.1). 10. The best section thickness depends on the type of the tissue and the researcher’s interest. When the research requires low magnification under microscopy (×2 to ×10 objectives), the tissue should be sectioned at a range thickness of 4 to 8 μm, once the stain intensity of the biological tissues is proportional to the section thickness. Otherwise, sections of 0.5–3 μm are adequate for high magnifications (×40 to ×100 objectives), once there is less over superposition of the cellular structures. As the correct thickness depends on the tissue, if it has more cells or connective tissues, we previously define the correct thickness by cutting sections of the same block from 1 up to 6 μm and collect them in a same slide. Afterwards, two thicknesses are chosen from them to develop the project, one thicker and the other thinner. We have frequently used slides with sections of 2 and 4 or 3 and 5 μm.
GMA for Improved Investigations Under Light Microscopy
17
11. During the sectioning procedure, the section should be taken individually with a forceps and laid down in a water with distilled water at room temperature. We should wait a couple of minutes for the section distension and collect it over a clean slide. 12. The green filter has been used in microscopy to increase the contrast of black and white micrographies. 13. When the staining is excessive or if it is necessary to take the stain out of the tissue, slides can be immersed several times and quickly in acid–water to clean tissues. Acid–water solutions are made with chloride or acetic acid at the proportion of 0.5–2%. Acid–alcohol solution should not be used once the alcohol wrinkles the GMA. 14. The mordant acts by increasing the electrostatic forces of tissue macromolecules, intensifying stains attachment.
Acknowledgments These methods were standardized during the development of different projects that were partially supported by Brazilian financial foundations (CAPES, CNPq, FAPEMIG, PRPq-UFMG). We thank Ana Luiza Drumond for helping in the immunohistochemistry processing. References 1. Clermont, Y., Perey, B. (1957) Quantitative study of the cell population of the seminiferous tubules of immature rats. Am J Anat 100, 241–268. 2. Clermont, Y., Perey, B. (1957) The stages of the cycle of the seminiferous epithelium of the rat: practical definitions in PA-Schiffhematoxylin stained sections. Rev Can Biol 16, 451–462. 3. Clermont, Y. (1962) Quantitative analysis of spermatogenesis of the rat: a revised model for the renewal of spermatogonia. Am J Anat 111, 111–129. 4. Clermont, Y., Bustos-Obregon, E. (1968) Re-examinations spermatogonial renewal in the rat by means of seminiferous tubules mounted ‘in toto’. Am J Anat 122, 237–247. 5. Russell, L. D., Clermont, Y. (1977) Degeneration of germ cells in normal, hypophysec-
6.
7.
8.
9.
tomized and hormone treated hypophysectomized rats. Anat Rec 187, 347–366. Chiarini-Garcia, H., Russell, L. D. (2001) High-resolution light microscopic characterization of mouse spermatogonia. Biol Reprod 65, 1170–1178. Chiarini-Garcia, H., Hornick, J. R., Griswold, M. D., Russell, L. D. (2001) Distribution of type-A spermatogonia in the mouse is not random. Biol Reprod 65, 1179–1185. Russell, L. D., Chiarini-Garcia, H., Korsmeyer, S. J., Knudson, C. M. (2002) Bax-dependent spermatogonia apoptosis is required for testicular development and spermatogenesis. Biol Reprod 66, 950–958. Chiarini-Garcia, H., Raymer, A. M., Russell, L. D. (2003) Non-random distribution of spermatogonia in rats: evidence of niches in
18
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
Chiarini-Garcia, Parreira, and Almeida the seminiferous tubules. Reproduction 126, 669–680. Bolden-Tiller, O. U., Chiarini-Garcia, H., Poirier, C., Alves-Freitas, D., Weng, C. C., Shetty, G., Meistrich, M. L. (2007) Genetic factors contributing to defective spermatogonial differentiation in juvenile spermatogonial depletion (Utp14b jsd ) mice. Biol Reprod 77, 237–246. Nascimento, H. F., Drumond, A. L., França, L. R., Chiarini-Garcia, H. (2008) Spermatogonial morphology, kinetics and niches in hamsters exposed to short- and longphotoperiod. Int J Androl, 32, 486–497. Doi:10.1111/j.1365–2605.2008.00884.x. Chiarini-Garcia, H., Meistrich, M. L. (2008) High-resolution light microscopic characterization of spermatogonia, in (Hou, S. X., Singh, S. R., eds.), Germline Stem Cells, vol 450. Humana Press, Totowa, NJ, Methods in molecular biology, pp. 95–107. Chiarini-Garcia, H., Alves-Freitas, D., Barbosa, I. S., Almeida, F. R. L. C. (2009) Evaluation of the seminiferous epithelial cycle, spermatogonial kinetics and niche in donkeys (Equus asinus). Anim Reprod Sci, 116, 139–154. Doi:10.1016/j.anireprosci.2008.12.019. Bennett, H. S., Wyrick, A. D., Lee, S. W., McNeil, J. H. (1976) Science and art in preparing tissues embedded in plastic for light microscopy, with special reference to glycol methacrylate, glass knives and simple stains. Stain Technol 51, 71–97. Cole, M. B., Jr., Sykes, S. M. (1974) Glycol methacrylate in microscopy: a routine method for embedding and sectioning animal tissues. Stain Technol 49, 387–400. Woodruff, J. M., Greenfield, S. A. (1979) Advantages of glycol methacrylate embedding systems for light microscopy. J Histotechnol 2, 164–167. Chiarini-Garcia, H., Machado, C. R. S. (1992) Mast cell types in the lymph nodes of the opossum Didelphis albiventris (Marsupialia, Didelphidae). Cell Tissue Res 268, 571–574. Chiarini-Garcia, H., Ferreira, R. M. A. (1992) Histochemical evidence of heparin in granular cells of Hoplias malabaricus Bloch. J Fish Biol 41, 155–157. Chiarini-Garcia, H., Pereira, F. M. A. (1999) Comparative studies of lymph nodes mast cell populations form five different marsupials species. Tissue Cell 31, 318–326.
20. Chiarini-Garcia, H., Santos, A. A. D., Machado, C. R. S. (2000) Mast cell types and cell-to-cell interactions in lymph nodes of the opossum Didelphis albiventris. Anat Embryol 201, 197–206. 21. Santos, A. A. D., Chiarini-Garcia, H., Oliveira, K. R., Machado, C. R. S. (2003) Development of different mast cell types in the opossum Didelphis albiventris. Anat Embryol 206, 239–245. 22. Rocha, J. S., Chiarini-Garcia, H. (2007) Mast cell heterogeneity between two different species of Hoplias sp (Characiformes: Erythrinidae): response to fixatives, anatomical distribution, histochemical contents and ultrastructural features. Fish Shellfish Immun 22, 218–229. 23. Paula, T. A. R., Chiarini-Garcia, H., França, L. R. (1999) Seminiferous epithelium cycle and its duration in capybaras (Hydrochoerus hydrochaeris). Tissue Cell 31, 327–334. 24. Neves, E. S., Chiarini-Garcia, H., França, L. R. (2002) Comparative testis morphometry and seminiferous epithelium cycle length in donkey and mules. Biol Reprod 67, 247–255. 25. Amaral, D., Chiarini-Garcia, H., Vale Filho, V. R., Allen, W. R. (2004) Effects of formalin and bouin fixation upon the mare’s endometrial biopsies embedded in plastic resin. Braz J Vet Anim Sci 56, 7–12. 26. Melo, R. C. N., Rosa, P. G., Noyma, N. P., Pereira, W. F., Tavares, L. E. R., Parreira, G. G., Chiarini-Garcia, H., Roland, F. (2007) Histological approaches for highquality imaging of zooplanktonic organisms. Micron (Oxford) 38, 714–721. 27. Bozzola, J. J., Russell, L. D. (1999) Electron Microscopy: Principles and Techniques for Biologists, 2nd edn. Jones and Bartlett, Sudbury, MA. 28. Karnovsky, M. J. A. (1965) A formaldehydeglutaraldehyde fixative of high osmolarity for use in electron microscopy. J Cell Biol 27, 137A–138A. 29. Shuttlesworth, G. A., de Rooij, D.G., Huhtaniemi, I., Reissmann, T., Russell, L.D., Shetty, G., Wilson, G., Meistrich, M.L.. (2000) Enhancement of a spermatogonial proliferation and differentiation in irradiated rats by gonadotropin-releasing hormone antagonist administration. Endocrinol 141, 37–49. 30. Russell, L. D., Ettlin, R.A., Sinha Hikim, A. P., Clegg, E. D. (eds.) (1990) Histological and Histopathological Evaluation of the Testis. Cache River Press, Clearwater, IL.
Chapter 2 Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis Gerluza Aparecida Borges Silva, Adriana Moreira, and José Bento Alves Abstract It is possible to obtain histological preparation of teeth and periodontium with satisfactory levels of quality by means of routine histological techniques, since specific cares are implemented during the sample processing. The formation of access ducts for the quick penetration of the fixative solution, the complete removal of the demineralizing agent and the increase of the time of dehydration, clearing, and paraffin embedding are some of these cares. A variety of fixing and demineralizing solutions have been proposed in the literature for teeth and periodontium processing. The author’s’ experience along the years demonstrated the possibility of satisfactory results with 10% buffered neutral formalin as fixative solution and 10% pH 7.3 EDTA as demineralizing solution. Sections of 6 μm in thickness obtained from paraffin-embedded samples, stained with hematoxylin and eosin, comply with the most morphological and morphometric evaluations. Besides, this routine protocol allows the use of serial sectioning for more specific techniques such as histochemical and immunohistochemical analyses, which are suitable for cellular constituent and extracellular matrix evaluation of teeth and periodontium. For the study of mineralized phases of isolated human teeth, ground sections can be obtained by the cutting–grinding technique. Though it is a recognized method of study, there are some technical difficulties involved, which are little exploited in the literature. This chapter presents a detailed cutting–grinding protocol for the histological evaluation of undecalcified isolated teeth and routine histology, which can be easily reproduced in any research or teaching support laboratory. Key words: Teeth, periodontium, histological processing, buffered neutral formalin, EDTA, cutting–grinding technique.
1. Introduction The light microscopy evaluation of teeth and periodontium is often performed by means of two basic histological techniques: 1. Cutting–grinding technique: a method for the study H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_2, © Springer Science+Business Media, LLC 2011
19
20
Silva, Moreira, and Alves
of undemineralized samples and 2. Routine histological technique for the study of demineralized teeth and periodontium. The cutting–grinding technique is an appropriate method in the evaluation of undecalcified isolated teeth, bone, and other hard tissues through macroscopic and light microscopic investigations. The fact that teeth and periodontium have both tissues of very different consistencies, constituted by mineralized connective tissues juxtaposed to the loose connective tissues, makes more complex the histological processing of these samples. Artifacts, as displacements, can be easily produced and may interfere in samples’ evaluation. Some researchers have been using the cutting–grinding technique for simultaneous evaluation of soft and hard tissues of teeth, periodontium, bone, and bone-anchored implants with good histological results (1–5). However, when the study refers only to evaluation of mineralized portion of isolated teeth, the cutting–grinding technique is a simple method that allows results with excellent quality for histological and morphometrical analysis. Though the cutting–grinding technique is a recognized method of study, there are some technical difficulties involved, which are little exploited in the literature. This chapter presents a detailed cutting–grinding protocol for isolated human teeth that can be easily reproduced in any research or teaching support laboratory. The soft tissues and organic matrix present in the teeth and periodontal structures can be evaluated from demineralized sections, prepared according to routine histological technique. A large variety of histological methods for demineralized samples have been suggested in the literature, with use of different associations between fixing and demineralizing solutions (6–14). However, their protocols are poorly described, making its repeatability as well as their comparative analysis difficult. The quality of histological sections is conditioned to the technique selected and the incorporation of some specific cares during histological processing of samples, aiming at the simultaneous preservation of tissues that constitute both the teeth and periodontium. Observations related to several cellular events, such as presence of inflammatory infiltrate and resorption processes, require material of excellent histological quality obtained by standardized and reproducible technique. At least four factors must be considered for the selection of the histological method: (a) the case urgency, (b) the tissue mineralization stage, (c) the purpose of the research, and (d) the staining technique that will be used (14). For example, the more rapid the decalcifier, the more injurious are its effects on subsequent staining likely to be. The effect is most noticeable in nucleic acids and manifests itself chiefly in the failure of nuclear chromatin to take up hematoxylin and basic dyes as readily as undecalcified soft tissues (15).
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
21
Our laboratory experience along the years demonstrated that histological technique considered as routine (use of 10% buffered neutral formalin as fixative solution, 10% EDTA as demineralizing solution, paraffin embedding and hematoxylin–eosin staining) allow to obtain satisfactory histological sections to the most qualitative and morphometric evaluations of teeth and periodontium. The staining with hematoxylin and eosin is the most commonly used for general purpose in the histological laboratory. Hematoxylin can be thought of being a basic dye when combined to aluminum, iron, copper, and tungsten salts, having an affinity to nucleic acids of the cell nucleus (16). It binds to acidic structures, structures yielding a blue-purple color. As such, the nucleus stains blue. Eosin is an acidic dye, which stains basic structures resulting from electrostatic combinations with tissues. The cytoplasm, collagen, and muscle are usually stained in red or pinkish red by eosin (17, 18). The staining with Gomori’s trichrome can be an alternative for the analysis of teeth and periodontium processes of repair, since it shows up the type I collagen – the main component of the organic matrix of these tissues. The expected results are collagen in blue and nuclei in blue to black (19, 20). Besides, the Gomori’s trichrome can be also indicated for the visualization of vascular alterations, such as hemorrhagic areas, hyperemia, and processes of vascular neoformations, since it enhances vessels and erythrocytes (cytoplasm in red), with a higher contrast than that of hematoxylin and eosin staining. This chapter presents, in details, protocols of cutting– grinding technique for undemineralized human teeth and routine histology for the analysis of demineralized samples, both standardized in our laboratory, with notes and orientations, which make possible its reproduction with predictable results.
2. Materials 2.1. Cutting–Grinding Technique: Mineralized Samples
1. Wax blocks for dental carving measuring 1.5 cm × 4 cm × 1.5 cm. 2. Carton cuts, scotch tape and petroleum jelly. 3. Synthetic rubber of white silicone 8001; room temperature vulcanizing (base and HS II catalyser) – Prepare a mixture with 3% of the catalyser. In the pattern, e.g., we used 100 g of silicone and 3 g of catalyzer, mixed with a wooden spatula to form a homogeneous mixture. Silicone 8001 has a mold durability of 2 years.
22
Silva, Moreira, and Alves
4. Crystal polyester resin (3061) prepared with 10% of styrene and one drop of hardener liquid (MEKP) for each 5 mL of solution (see Note 1). As a pattern, for three blocks of resin, we use 27 mL of resin + 3 mL of styrene + 6 drops of hardener liquid. 5. Pins for disposition of teeth during resin embedding. These pins can be made either of wood, plastic or acrylic. Examples: toothpicks cuts or cottontail rods. R 6. Rapid glue – Super Bonder .
7. Apparatus: (a) IsoMet 1000 – Precision saw with diamond wafering blade series 15 LC diamond (6 diameter × 0.220 – 152 mm × 0.5 mm)/Buehler-USA. (b) Motored polisher sanding machine. 8. Wet sand papers (600, 1000, and 1200 granulation/mm2 ). 9. Wood/aluminum blocks for supporting of dental sections during the final grinding. 10. Xylene. Inflammable: exhibits neurological effects. Can also cause irritation of the skin, eyes, nose, and throat. Requires fume hood for safe usage. 11. Glass slides and histological cover glass (24 × 50 mm). R 12. Mounting medium: Entellan .
2.2. Histological Processing for Demineralized Samples
1. Double-face diamond disk adapted in a low rotation equipment (microrectifier with flexible axle). 2. Discardable steel blades or surgical knifes. 3. 10% Neutral buffered formalin solution: (a) 36% Formaldehyde solution; Carcinogenic: toxic by inhalation, in contact with skin and if swallowed. Cause burns. May cause sensitization by skin contact. Use gloves to avoid bare hand contact. May cause heritable genetic damage. Use only in well ventilated areas. (b) Disodium hydrogen orthophosphate anhydrous – Na2 HPO4 – 0.65% w/v; (c) Sodium dihydrogen orthophosphate monohydrate – NaH2 PO4 .H2 O – 0.4% w/v. Prepare solution 0.65% w/v disodium hydrogen orthophosphate anhydrous and 0.4% w/v sodium dihydrogen orthophosphate monohydrate in distilled water. Stir up until obtaining a homogeneous solution. Add 36% formaldehyde solution – 10% v/v. Store it in a dark vial (see Note 2).
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
23
4. Ethylenediaminetetraacetic acid disodium salt 2-hydrate (EDTA). Prepare 10% w/v EDTA aqueous solution pH 7.2–7.4. Dissolve EDTA under stir in distilled water at 59◦ C. Wait until the solution reaches room temperature and adjust pH using sodium hydroxide pastilles. Complete with distilled water to final volume of solution. 5. Absolute ethanol – The preparation of alcohol solutions 70, 80, 90, and 95% by dilution in distilled water is performed by using a Gay-Lussac alcoholmeter. 6. Xylene. 7. Embedding agent for histology – pastilles-solidification point 56–58◦ C. 8. Microtome blades PTFE coated – Low profile blades. 9. Glass slides and cover glasses (24 × 50 mm). 10. Mayer’s albumin: Two white eggs, glycerin 87% and thymol. Beat the white eggs to resemble firm snow, then reserve for a 24 h period at 4◦ C, filter through filter paper and add glycerin 1:1 v/v. Add thymol crystals 0.1% w/v (see Note 3). 11. Staining methods: (a) Routine staining with hematoxylin and eosin: i. Harris hematoxylin solution – ready for use. ii. Putt’s eosin: eosin yellowish p.a.; potassium dichromate p.a.; ethanol p.a. and saturated picric acid solution (1.2%). Dissolve 1 g of eosin yellowish in 10 mL of ethanol. Dissolve 0.5 g of potassium dichromate in 80 mL of distilled water and add the eosin solution. Add 10 mL of saturated picric acid solution (see Note 4). (b) Staining for collagen fibers: Gomori’s trichrome i. Harris hematoxylin solution – ready for use. ii. Gomori’s trichrome solution: chromotrope 2R, fast green, phosphotungstic acid p.a., acetic acid p.a. Prepare the solution by dissolving 1.8 g of chromotrope 2R + 0.9 g of fast green + 1.8 g of phosphotungstic acid in 300 mL of distilled water. Heat up to dissolve. Wait until the solution reaches room temperature and add 3 mL of acetic acid. 12. Quantitative filter paper – rapid filtration/white band (15 cm diameter) R 13. Mounting medium: Entellan .
24
Silva, Moreira, and Alves
3. Methods 3.1. Cutting–Grinding Technique: Mineralized Samples
1. Preparation of the mold for silicone (see Note 5): Draw a mold in a carton sheet, in a proportional size to the number of resin blocks (see Fig. 2.1a–c). After drawing the box, the carton is cut with a scissor and walls fitted with adhesive tape (see Fig. 2.1d). 2. Glue one (or more) block of wax for dental carving inside of the carton box (see Note 6). Fix a little support (about 1.0 cm) on one of the surfaces of the block. This fixation is mechanically performed, only with a small pressure of the
A
B
C
D
E
F
G
H
I
J
K
L
Fig. 2.1. Sequence of stages for cutting–grinding technique: a histological method for mineralized dental samples. a–h Preparation of silicone mold for attainment of blocks in resin. i–k Preparation of teeth for embedding. l Dental sample embedded in crystal polyester resin.
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
25
pin on the wax surface. Such block(s) will act as mold for the cavity to be filled with resin in other step. The coupled pin generates an orifice that will work as a guide for the placement of teeth (see Fig. 2.1 h and j). Lubricate the bottom of the carton box with petroleum jelly before applying the silicone. 3. Wearing gloves, handle the silicone and throw it inside of the fitted box (see Fig. 2.1e), covering about 1 cm above the block of wax. Wait 24 h for rubber vulcanization (hardening). Dismount the box of paper (see Fig. 2.1f). Remove blocks of wax. 4. Preparation of teeth for embedding: (a) Glue a support with a measure such like that used in blocks of wax, on the palatine/lingual surface of teeth (see Fig. 2.1i). Such procedure is performed with rapid R glue (Super Bonder ) and aims at standardizing the embedding position (see Fig. 2.1 k). (b) Using a paint brush, overlay the tooth surface with the resin mixture before the embedding so as to reduce the surface tension. 5. Locate the tooth (teeth) on the center of cavity(ies) for embedding (see Fig. 2.1j, k). Prepare the solution of resin– styrene (see Note 1) and pour it into the silicone mold, in cavities molded by blocks of wax. 6. After a 24 h period of resin cure, withdraw the set resin + tooth from the silicone mold (see Fig. 2.1 l). 7. Make a guide for the block cutting, by drawing a dotted line on the resin surface passing along the axis of the embedded tooth (detail in Fig. 2.2c). With the aid of the cutting set IsoMet (slicing machine fitted with a diamondimpregnated cutting disc of 0.5 mm thick (see Fig. 2.2a, c)), regulated to the speed of 300 rpm with a loading of 100 g (see Note 7), one can section the block, under refrigeration (see Note 8). 8. Using a motorized polisher sanding machine (see Fig. 2.2b) – polish the halves attained, with the surface of teeth turned to the wet sand paper sheet (see Fig. 2.2d), using all the sequence of granulation (600, 1000, and 1200 granulation/mm2 ) (see Note 9). 9. Return with the block to IsoMet cutting set for a new cutting, thus obtaining a slice of about 1.0 mm thick (see Fig. 2.2e). 10. Glue the tooth slice onto a piece of aluminum or plane wood (support for gripping), with a mounting medium R Entellan (see Fig. 2.2f). The polished side, already treated, must be turned to the support, leaving free the
26
Silva, Moreira, and Alves
A
B
C
D
E
H
F
G
I
Fig. 2.2. a IsoMet 1000. b APL-04 motorized sand polisher machine. c First section of resin block and detaching of the tooth in two halves. d First finish stage of dental face in refrigerated wet sand paper sheets. e Second section of resin block for obtaining slices of 1 mm thick. f Gluing of slice onto an aluminum support. g Trimming of the slice and finishing in sand paper sheets 600, 1000, 1200. h Image of slice of the tooth imbedded in resin, after trimming in the polisher. i Ground section prepared from half of the tooth, using a cutting–grinding system. Section ready for mounting and analysis at light microscope.
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
27
side not yet submitted to the polishing. Keep inside of a drying kiln at 35–37◦ C during a 24-h period for resin polymerization. 11. Perform the same treatment described in item 8 (see Fig. 2.2 g), until obtaining a smooth surface and a thickness of 30 μm (see Fig. 2.2 h) that allows the microscope light to pass through the specimen (see Note 10). 12. Set free the cut obtained, by immersing in xylene the surface glued onto the wood. 13. Wash dental sections in two baths of absolute ethanol, under stir or ultrasonic cleaning to remove debris. Immediately after, immerse in two baths of xylene (5 min each). 14. Mount the dental cuts on glass slides, with mountR ing medium Entellan , overlapping a cover glass (see Fig. 2.2i). Transfer slides to a drying kiln (35–37◦ C) during a 24–48 h period. 15. Evaluate at light microscope with partial closing of condenser diaphragm. Results are shown in Fig. 2.3. 3.2. Routine Histological Processing of Demineralized Samples 3.2.1. Samples Collection
1. Rats maxillae fragments: After decapitation of animals by guillotine, heads must be rapidly dissected for skin removal, separation of maxillae and brain removal in order to facilitate the access of the fixative solution to the remaining tissues. 2. Isolated human teeth: Immediately after exodontics, the radicular apex (apices) must be removed aiming at enlarging the apical orifice to facilitate the access of the fixative solution to pulpal tissue. Apicectomy must be performed under refrigeration with new double-face diamond disk.
3.2.2. Sample Fixation and Demineralization
1. Keep specimens in 10% buffered neutral formalin, into dark vials, of large mouth, at 4◦ C, for 24 h (see Note 11). After this period of time, replace the fixative solution and keep samples at room temperature for 24 h (see Note 12). The volume of the fixative solution must be at least 20 times larger than the tissue to be fixed (21, 22). 2. Wash in running water (3 × 10 min). 3. Demineralization with 10% EDTA aqueous solution (see Note 13) in a shaker at room temperature. The decalcifying solution is changed every two days. Demineralization is controlled by means of superficial cuttings performed with a
28
Silva, Moreira, and Alves
Fig. 2.3. Light photomicrographs of human permanent teeth sections obtained by the cutting–grinding technique. a Ground section of a canine. b Image of the tooth crown showing the Enamel (E); Mantle Dentin (MD); Dentinal tubules (DT), and Interglobular dentin (arrows). c Dentin (D); Incremental lines (arrows). d Dentin (D); Enamel (E); Dentin-enamel Junction (DEJ). e Pulp chamber (PC); Tertiary dentin (arrow). f Secundary dentin (arrow). g Dentinal tubules (thin arrows); Enamel lamella (thick arrows); h Dentinal canalicules (tubules) in the radicular wall (RC); Sharpey’s fiber (arrows); Acellular cementum (∗ ). i Root dentin (RD); Cellular cementum (∗ ). j Granular layer of Tomes (arrows); Cellular cementum (∗ ). Bars = 400 μm.
steel blade in fragments edges. The attainment of cuts of firm texture, but without resistance to the steel blade indicates the ideal point of decalcification. After 15 days of demineralization, it is possible to perform the trimming of samples – reduction of fragments locating the area of interest, trimming fragments so as to standardize the embedding position
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
29
Fig. 2.4. Schematic drawing showing the preparation of maxillae of rats for histological evaluation of demineralized sections of molars and periodontium in longitudinal plane. a After initial demineralization stage in EDTA, maxillae are reduced by means of two cuts: Section 1. Crosscut with discard of pre-maxilla. Section 2. Reduction of the palatal faces in a plane parallel to the imaginary line traced over molars occlusal surfaces. b The inclusion of fragments, with palatal surface turned to the plane of microtomy (arrows), allows the simultaneous visualization of teeth (crown and roots) and periodontium. Due to tipping of first molar mesial root of rodents, it is recommended to maintain part of the diastema region (∗ ), so as to make possible the visualization of this root and associated periodontal structures.
(Fig. 2.4). Samples of human teeth also can be reduced, cut or divided after 30 or 45 days of demineralization, depending on dental type. After this preparation, it is advisable to keep fragments in the demineralizing solution for a week more, in case of maxillae, and for two more weeks in case of a human teeth, in order to assure the complete demineralization of samples. 3.2.3. Tissue Processing for Paraffin Embedding and Microtomy
1. Removal of EDTA: wash in running tap water (3 × 30 min + 1× 14–18 h). 2. Dehydration. 3. Clearing. 4. Paraffin infiltration: see Table 2.1. 5. Embedding in paraffin. 6. Microtomy: Sections of 6 μm collected in glass slides covered by Mayer’s albumin. Allow it to dry inside a kiln (35–37◦ C) for 24 h before staining.
3.2.4. Staining
It is possible to perform staining in serial sections with hematoxylin and eosin (H&E) and Gomori’s Trichrome, according to protocols described in Table 2.2. Figure 2.5 shows results obtained with histological routine staining technique (H&E) and Gomori’s trichrome staining.
30
Silva, Moreira, and Alves
Table 2.1 Protocol for dehydration, clearing and paraffin infiltration of maxillae of rat and isolated human teeth Maxillae of rat
Isolated human teeth
–
1 × 30 min
70%; 80%; 90% ethanol
2 × 30 min
2 × 30 min
Absolute ethanol
3 × 30 min
3 × 60 min
1 × 20 min
3 × 20 min
2 × 10 min
(see Note 14)
Xylene: paraffin (1:1)
–
1 × 30 min
Paraffin
3 × 30 min
3 × 60 min
Step 2. Dehydration – series of alcohol solutions 50% ethanol
3. Clearing Xylene (In exhaustion hood, with gloves) 4. Infiltration in paraffin (inside kiln 58◦ C)
For evaluation of processing quality, some parameters can be taken as reference: (a) preservation of odontoblasts layers without displacement of predentin (Fig. 2.5a–c); (b) preservation of endothelial wall of blood vessels (Fig. 2.5a); (c) preservation of collagen fibers and extracellular matrix (Fig. 2.5a–c, h); and (d) bone trabeculae with osteocytes and endosteum layer well preserved (Fig. 2.5i).
4. Notes 1. Formation of small blisters in the preparation of resin is common due to its viscosity. The incorporation of styrene to the resin formula induces to lower viscosity and, as a result, smaller quantity of blisters in the mixture. Styrene prolongs the resin cure stage a little, allowing blisters to flow together and reach the surface so as to be eliminated before resin hardening. 2. The formaldehyde decomposes itself easily in formic acid by light action and may negatively interfere in the staining of sections. The use of buffers neutralizes the action of these acids (23). For this reason, 10% neutral buffered formalin solution is preferably used and must be stocked into amber vials, for protection against luminosity that can favor acid formic formation.
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
31
Table 2.2 Protocols for routine staining (hematoxylin and eosin) and for type I collagen fibers (Gomori’s trichrome) Step
H&E
Gomori’s trichrome
1. Deparaffinization Xylene Xylene
1 × 30 min 2 × 15 min
1 × 30 min 2 × 15 min
2. Hydration Absolute ethanol 90% ethanol 80% ethanol 70% ethanol Running tap water
3 × 2 min 1 × 2 min 1 × 2 min 1 × 2 min 1 × 20 min
3 × 2 min 1 × 2 min 1 × 2 min 1 × 2 min 1 × 20 min
Filtered Harris hematoxylin solution 1 × 1 min
Filtered Harris hematoxylin solution 1 × 1 min
4. Running tap water
1 × 20 min
1 × 20 min
5. 2ª staining solution
Putt’s eosin solution 40 s to 1 min (see Note 15)
Gomori’s Trichrome solution 1 × 15 min (see Note 16)
6. Stain washing Running tap water
(10–20 s)
(10–20 s)
7. Dehydration 70% ethanol 80% ethanol 95% ethanol Absolute ethanol
1 × 10 s 1 × 10 s 1 × 30 s 3 × 2 min
– – 1 × 30 s 3 × 2 min
8. Clearing Xylene Xylene
2 × 2 min 1 × 10 min
2 × 2 min 1 × 10 min
R Cover slip using Entellan mounting medium
R Cover slip using Entellan mounting medium
3. 1ª staining solution
9. Mounting (see Note 17)
3. The glycerin prevents the slide from completely drying while the egg albumin, a protein, is denatured when immersed in 70% ethanol and fixes sections onto the glass slide. The addition of thymol crystals allows the solution preservation, without contamination by fungi, at 4◦ C even for 12 months. 4. Putt’s eosin solution intensely stains the acidophilic structures when recently prepared. Dilution 1:1 in distilled water is recommended to reduce the staining intensity. 5. The silicone form can be prepared either for one detached block or can be amplified with several cavities for simultaneous embedding of several teeth. Preparation of the resin
32
Silva, Moreira, and Alves
Fig. 2.5. a–e Dentin-pulp complex of human tooth. a Panoramic view. HE. Bar = 140 μm. b Layers of pulp. HE. Bar = 50 μm. c Dentin-pulp complex stained with Gomori’s trichrome. Bar = 160 μm. d Collagen of dentin bridge (arrowheads) induced by direct pulp capping with calcium hydroxide. Bar = 140 μm. e Presence of extravasated erythrocytes (arrow) in inflammated dental pulp. Gomori’s trichrome. Bar = 160 μm. f Components of periodontium stained with Gomori’s trichrome. Bar = 350 μm. g Panoramic view of the first and second molars region in a longitudinal section of rat maxilla. HE. Bar = 600 μm. h Higher magnification of rat periodontium. HE. Bar = 80 μm. i Resorption area of alveolar bone in higher magnification. Osteoclasts (arrows). HE. Bar = 80 μm. Dentin (D); Predentin (Pd.); Vessel (V); Odontoblast layer (Od); Cell-free zone (CFZ); Cell-rich zone (CRZ); Central layer (CL); Pulp (P); Root dentin (RD); Alveolar bone (AB); Periodontal Membrane (PM) or (∗ ); Interradicular septum (IS).
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
33
for the embedding of several teeth is desirable, since the proportion of the resin components for a larger volume is more reliable. 6. The wax blocks can be replaced by wood blocks prepared in the same dimensions or larger than the wax blocks. In this case, they must be lubricated with petroleum jelly before applying the silicone. 7. Faster speed as well as heavier weight tends to damage the specimen surface. With such careful management (200 or 300 rpm and 100 g), the detailed structure of specimens will be preserved from damage. 8. The cut of resin blocks can be handmade by using a doubleface diamond disc, in a mandrel, adapted either to an electric micromotor or to a microrectifier with flexible axle. 9. Wear and finishing process of dental sections embedded in resin can be either manually performed or by means of other types of electric polisher sand machines (orbital), since it allows the replacement of sand paper sheets. In the case of manual processing, it is important to fix the wet sand paper sheets onto a plane surface and moisten them constantly. 10. The thickness of sections is variable. It is important to obtain a homogeneous, semi-transparent, whitish texture. It has been reported that 100 μm thick sections are satisfactory for light microscopy (24). 11. The buffered neutral formalin is considered a universal fixative, with good results for most tissues (25). Formalin is not significantly harmful to any tissue types and for this reason it allows viewers to observe the tissue in its almost natural state. 12. Samples can be stocked in buffered neutral formalin for several days without significant tissue alterations for the routine histological analyses. However, for both histochemical and immunohistochemical techniques, the fixation of specimens for at the most 72 h is recommended. We have observed that a 48 h period is sufficient for a good fixation and this information is in agreement with the literature (14). 13. A variety of methods has been recommended for decalcifying the hard tissues. Decalcification can be performed through immersion into acids (weak or strong solutions) or into chelant components. All acids, in some way, interfere with the tissue stability, in reliance on the acidity of solution and the time of sample demineralization. In whole tooth specimens, the distortion of the pulp tissue can be easily demonstrated if a stronger decalcifying agent is used.
34
Silva, Moreira, and Alves
The removal of calcium by the acid produces carbon dioxide. The pressure of this gas and its movement through the tissue can be one factor causing the separation of connective tissues observed after decalcification (26). The EDTA is a chelant substance that captures metallic ions, among which is calcium, removing them slowly with excellent preservation of histological details (16). Hard tissues, particularly dental tissues, decalcified by chelators (e.g. 10% EDTA neutralized with sodium hydroxide, pH 7.2), show a minimum of artifact. Staining is of an acceptable standard after using this chelating agent and good results can be obtained (17). Although these chemical agents are the slowest, it preserves cells and matrix constituents, particularly mucopolysaccharides (27), and gives the best antigenic/enzymatic preservation. In case of urgent dental and periodontal analysis, we suggest the utilization of: (a) Planck Rychol’s solution (mixture composed by 7% aluminum chloride (w/v); 8.5% hydrochloric acid (v/v); and 5% formic acid (v/v) or (b) Anna Morse’s solution (mixture composed by 50% formic acid solution (v/v) and 20% sodium citrate (w/v) aqueous solution, used 1:1 (14, 28). Good results can be reached, but samples must be monitored every day, because these acid solutions demineralize in a short time, depending on the size of samples. There are risks of irreparable degradation of the organic matrix, in case the acid solution is kept in a higher time than that required for demineralization of hard tissues of tooth and periodontium. 14. The third bath of xylene, particularly for dental samples, can extend for even 2 h, for total clarifying of samples. Dental tissues clarify very well thus coming to the diaphanous aspect; therefore, any whitish spot inside the sample is a sign of the requirement of extending the clarification period. Nevertheless, the monitoring of the third bath is extremely significant, because the excessive time in xylene can harden the piece and make the microtomy stage difficult. 15. After the eosin staining and rapid washing in running tap water, one can proceed the regressive staining (differentiation), when the tissue presents itself extremely stained. Regressive staining consists in dipping slices rapidly (for some seconds) in 1% chloridric acid solution in 70% ethanol and washing in running tap water for 2–3 min. 16. Formalin-fixed tissues may benefit from secondary fixation of sections in Bouin’s fluid, which enhances the red color of muscle fibers and epithelial cells in staining with trichromics (29). The Bouin’s fluid considerably enhances trichrome
Histological Processing of Teeth and Periodontal Tissues for Light Microscopy Analysis
35
intensity and brilliance due the action of picric acid in the tissue. After hydration step, plunge sections into Bouin’ solution heated at 56◦ C (saturated picric acid 750 mL; formaldehyde 250 mL; glacial acetic acid 50 mL) for 1 h. Wash slides in running tap water for 10 min before application of the advocated stains for Gomori’s technique (20). R 17. The Entellan is a rapid cure resin and polymerizes in a maximum of 24 h. Alternatively, one can opt for synthetic Canada balsam due to the significantly lower cost when R compared to Entellan ; however, this medium requires about 4–5 days for polymerization.
References 1. Donath, K., Breuner, G. (1982) A method for the study of undecalcified bones and teeth with attached soft tissues. The SägeSchliff (sawing and grinding) technique. J Oral Pathol 11(4), 318–326. 2. Lan, W. H., Kwan, H. W., Sunada, I. (1986) Slicing technique for tooth specimens in histological preparation. Bull Tokyo Med Dent Univ 33(4), 129–136. 3. Rohrer, M. D., Schubert, C. C. (1992) The cutting-grinding technique for histologic preparation of undecalcified bone and bone-anchored implants. Improvements in instrumentation and procedures. Oral Surg Oral Med Oral Pathol 74(1), 73–78. 4. Günhan, M., Günhan, O., Celasun, B., Safali, M. (1996) Examination of periodontal tissues by a cutting-grinding technique. Aust Dent J 41(3), 173–175. 5. Cano-Sánchez, J., Campo-Trapero, J., Gonzalo-Lafuente, J. C., Moreno-Lopes, L. A., Bascones-Martínez, A. (2005) Undecalcified bone samples: a description of the technique and its utility based on the literature. Med Oral Pathol Oral Cir Bucal 10 Suppl 1, E74–E87. 6. Abreu, E. M. (1990) Processo de reparação de feridas de extração dentária, após implante de colágeno microcristalino: estudo histológico em ratos. Rev Port de Est e Cir Maxilofac 31(2), 95–102. 7. D’Souza, N. R., Bachman, T., Baumgardner, K. R., Butler, L. M. (1995) Characterization of cellular responses involved in reparative dentinogenesis in rat molars. J Dent Res 74(2), 702–709. 8. Lekic, P., Sodek, J., Mcculloch, C. A. (1996) Osteopontin and bone sialoprotein expression in regenerating rat periodontal ligament and alveolar bone. Anat Rec 244(1), 50–58.
9. Lamano Carvalho, T. L., Bombonato, K. F., Brentegani, L. G. (1997) Histometric analysis of rat alveolar wound healing. Braz Dent J 8(1), 9–12. 10. Terai, K., Takano-Yamamoto, T., Ohba, Y., Hiura, K., Sugimoto, M., Sato, M., Kawahata, H., Inaguma, N., Kitamura, Y., Nomura, S. (1999) Role of osteopontin in bone remodeling caused by mechanical stress. J Bone Miner Res 14(6), 839–849. 11. Gonçalves, E. L., Pavan, A. J., Tavano, O., Guimarães, S. A. C. (2002) Morphogenetic activity of demineralized dentin matrix: a study in dogs. Rev Fac Odontol Bauru 10(1), 51–56. 12. Silva, G. A. B., Lanza, L. D., Lopes-Júnior, N., Moreira, A., Alves, J. B. (2006) Direct pulp capping with a dentin bonding system in human teeth: a clinical and histological evaluation. Oper Dent 31(3), 297–307. 13. Sawada, T., Sugawara, Y., Asai, T., Aida, N., Yanagisawa, T., Ohta, K., Inoue, S. (2006) Immunohistochemical characterization of elastic system fibers in rat molar periodontal ligament. J Histochem Cytochem 54(10), 1095–1103. 14. Fernandes, M. I., Gaio, E. J., Rosing, C. K., Oppermann, R. V., Rado, P. V. (2007) Microscopic qualitative evaluation of fixation time and decalcification media in rat maxillary periodontium. Braz Oral Res 21(2), 134–139. 15. Bancroft, J. D., Stevens, A., Turner, D. R. (1990) Bone, in Theory and Practice of Histological Techniques, 3rd edn. Ed. Churchill Livingstone, New York, NY, pp. 309–342. 16. Tolosa, E. M. C., Rodrigues, C. J., Behmer, O. A., Freitas-Neto, A. G. (2003) Técnica histológica, in Manual de Técnicas para
36
17.
18.
19. 20.
21.
22.
Silva, Moreira, and Alves Histologia Normal e Patológica, 2nd edn. Ed. Manole, Barueri, SP, pp. 19–86. Bancroft, J. D., Cook, H. C. (1994) Principles of tissue demonstration and routine morphological staining, in Manual of Histological Techniques and Their Diagnostic Application. Ed. Churchill Livingstone, New York, NY, pp. 425–428. Burkit, H. G., Stevens, A., Lowe, J. S., Young, B. (1996) Notes on staining techniques, in Wheater’s Basic Histopathology: A Colour Atlas and Text. Ed. Churchill Livingstone, New York, NY, p. 277. Gomori, G. (1950) A rapid one-step trichrome stain. Am J Clin Pathol 20(7), 661–664. Behmer, O. A., Tolosa, E. M. C., FreitasNeto, A. G. (1976) Coloração do tecido conjuntivo, in Manual de T´ecnicas para Histologia Normal e Patol´ogica. Ed. EDART – USP, São Paulo, pp. 109–131. Michalany, J. (1980) Fixação, in T´ecnica Histol´ogica em Anatomia Patol´ogica: com Instruc¸˜oes para o Cirurgi˜ao, Enfermeira e Citot´ecnico. Ed. EPUP, São Paulo, pp. 40– 51. Junqueira, L. C. U., Junqueira, L. M. M. S. (1983) Fixação e descalcificação, in T´ecnicas ´ basicas de Citologia e Histologia, Instituto de
23.
24.
25. 26. 27. 28.
29.
Ciˆencias Biom´edicas e Faculdades de Medicina da USP, Ed. Santos, São Paulo, pp. 13–20. Burck, H. C., Carilla, P. C. (Transl.) (1969) Técnica Histológica: Manual para realizar preparaciones microscópicas en el laboratorio, Ed. Paz Montalvo, Madrid. Souza, E. M. D., Stott, G. G., Alves, J. B. (1999) Determination of age from cemental incremental lines for forensic dentistry. Biotech Histochem 74(4), 185–193. Grimaldi-Filho, G. (1981) Manual de Técnica Histológica. Fiocruz–Centro de Microscopia Eletrônica, Rio de Janeiro, 70p. Brain, B. E. (1966) The Preparation of Decalcified Sections. Charles C Thomas, Springfield, IL, pp. 69–135. Bélanger, L. F., Copp, D. H., Morton, M. A. (1965) Demineralization with EDTA by constant replacement. Anat Rec 153(1), 41–47. Morse, A. (1945) Formic acid-sodium citrate decalcification and butyl alcohol dehydration of teeth and bones for sectioning in paraffin. J Dent Res 24, 143–153. Tolosa, E. M. C., Rodrigues, C. J., Behmer, O. A., Freitas-Neto, A. G. (2003) Coloração para tecido conjuntivo, in Manual de T´ecnicas para Histologia Normal e Patol´ogica, 2nd edn. Ed. Manole, Barueri, SP, pp. 111–142.
Chapter 3 Large Plant Samples: How to Process for GMA Embedding? Élder Antônio Sousa Paiva, Sheila Zambello de Pinho, and Denise Maria Trombert Oliveira Abstract It is often necessary to process large plant samples for light microscopy studies, but due to structural characteristics of plant tissues, especially intercellular spaces, large vacuoles, and phenolic substances, results are often unsatisfactory. When large samples are embedded in glycol methacrylate (GMA), their core may not polymerize, remaining soft and moist and making it difficult to cut microtome sections. This situation has been erroneously interpreted as the result of poor infiltration, when the soft core of these samples is actually the result of incomplete polymerization. While GMA is in fact present inside samples, unsatisfactory polymerization results from rapid external polymerization that does not allow sufficient hardener to reach the sample core, while the relatively large volume of GMA inside the tissue block also dilutes the hardener. In this chapter we propose a new method for processing large plant specimens that avoids these problems by: (1) slowing the polymerization process through cooling in order to permit the penetration of hardener into the sample core and (2) increasing the hardener:GMA ratio to aid polymerization of the sample core. Key words: (2-Hydroxyethyl)-methacrylate, historesin, embedding, GMA, HEMA, plant anatomy, plant tissue, resin polymerization.
1. Introduction Paraffin has traditionally been used in routine histological work with animal and plant tissue samples, but resin embedding media commonly used in electron microscopy has been increasingly employed in light microscopy in the last few decades because it preserves morphological features far better than paraffin mixtures (1). This resin embedding media was developed to resist damage H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_3, © Springer Science+Business Media, LLC 2011
37
38
Paiva, Pinho, and Oliveira
by the electron beam and high vacuum, and to preserve cellular and subcellular details (1). The glycol methacrylate (GMA, synonymous with (2-hydroxyethyl)-methacrylate [HEMA]) (1) is currently employed for light microscopy purposes throughout the world. Among the advantages of the use of GMA as an embedding medium is the ease of preparing heterogeneous tissue sections (1). Glycol methacrylate embedding also has disadvantages, including difficulties encountered in embedding large samples. Instructions provided with the historesin kits are sufficient for standard samples, usually animal tissue that are relatively homogeneous, and when plant tissue samples with similar characteristics are used embedding is largely successful. Plant tissues generally have greater structural diversity; however, with unique physical and chemical properties they demand specific procedures and methodological adaptations to assure good results. Several tissues are found in plant organs, making them a mosaic of cells with different physical–chemical properties. Plant samples used for anatomical studies vary widely in their proportions of intercellular spaces, dimensions, the chemical contents of their cell vacuoles, and the mechanical properties of their cell walls. These characteristics can negatively affect procedures of dehydration, infiltration, and embedding. The most common problems encountered in processing plant samples are principally related to difficulties with GMA infiltration, although these can usually be minimized (or solved) by simply reducing the sample size. Sample size reduction, however, can make the posterior analysis of the material more difficult, and must often be avoided. In the specific case of plant tissues, the use of a vacuum in at least one of the processing steps will contribute to minimizing infiltration problems; the vacuum should preferably be applied during fixing as this procedure removes air from the intercellular spaces and quickly replaces it with fixative. When studies demand whole-organ interpretation (e.g. studies of floral vasculature, fruit and seed structural organization, and nodal anatomy), large samples are essential. These large samples require more time at each dehydration step and according to Feder and O’Brien (2), the specific dehydration time intervals will depend on the size and permeability of the specimen. In the case of infiltration and polymerization of the GMA, however, simple time adjustments alone may not be sufficient to solve the problem. This chapter presents solutions to some of the problems encountered in processing large plant samples that can significantly increase GMA polymerization efficacy.
Large Plant Samples: How to Process for GMA Embedding?
39
2. Materials 1. Ginger rhizomes (Zingiber officinale Roscoe): cubes (0.8 cm on each side) without periderm and cortical portions were prepared using a razor blade. 2. Formalin-aceto-alcohol (FAA): 50 mL of 37% paraformaldehyde, 50 mL acetic acid, and 900 mL of 50% aqueous ethanol. 3. Ethanol. 4. Histomolds (1 cm3 cells). 5. HEMA (LeicaTM Historesin kit). 6. Wood stubs and adhesive (we used epoxy-based glue – AralditeTM ). 7. Steel knife: Stainless Steel C-Profile Knife. 8. 0.1 M pH 6.8 Phosphate buffer (1.38 g NaH2 HPO4 .H2 O [0.1 M] in 100 mL of distilled water [solution A] and 1.42 g Na2 H.HPO4 [0.1 M] in 100 mL of distilled water [solution B]). To prepare the pH 6.8 buffer, 51 mL of solution A was mixed with 49 mL of solution B and the pH was adjusted with the same solutions. 9. Toluidine Blue O: (to prepare a 100 mL solution, add 0.05 g of Toluidine Blue O [Allied Chemical] to 0.1 M pH 6.8 100 mL phosphate buffer). 10. Glass slides. 11. Coverslips.
3. Methods 3.1. Fixation
Samples were obtained from mature ginger rhizomes. The periderm and the cortical portion were removed with a razor blade in order to prepare tissue cubes (0.8 × 0.8 × 0.8 cm) of the central cylinder region. This region presents fairly homogeneous structural characteristics, so sample size is a major obstacle to good embedding. Fifteen samples were employed in this experimental assay and these were fixed by immersion in FAA 50 (3) and submitted to moderate vacuum (180 mmHg) using a vacuum pump for 10 min in order to remove all air from the intercellular spaces and vessel lumens (see Note 1). Samples remained in the fixative for 48 h (see Note 2).
40
Paiva, Pinho, and Oliveira
3.2. Dehydration
3.3. Pre-infiltration, Infiltration, and Embedding Treatments
Samples were dehydrated in a 50–95% graded ethanol series. The dehydration stages were conducted at room temperature, as detailed below. It is important to emphasize that the times required for dehydration vary according to samples’ sizes and their densities. Due to difficulties involved in evaluating parameters related to absorption rates, we recommend at least 2 h for each dehydration step for large samples (∼1 cm3 ). All alcohol changes must be performed with care to avoid exposure to the open air and minimize air penetration into samples. Summary of the dehydration steps: Step
Duration
Conditions
50% ethanol
2h
Room temperature
70% ethanol (see Note 2)
2h
Room temperature
85% ethanol
2h
Room temperature
95% ethanol (see Note 3)
2h
Room temperature
After dehydration, samples were submitted to infiltration with GMA historesin; we only recommend the use of GMA activated by dibenzoylperoxide, as indicated in the historesin embedding kit. The alcohol is first substituted by pre-infiltration of a mixture of 95% ethanol: GMA (1:1) for several hours. This use of the alcohol: GMA solution as an intermediate step with plant samples facilitates infiltration and reduces problems related to resin viscosity (1). During the embedding stage when the hardener (dimethyl sulfoxide) is added to the GMA, resin polymerization time is short and positioning samples into histomolds must be done efficiently and quickly. When several samples are to be embedded, we suggest putting the histomold in an ice bath, which will delay polymerization and provide enough time to adjust positions of samples (see Note 4). Steps referred to here are crucial to processing large samples that frequently show embedding problems due to incomplete resin polymerization. We experimented with different preinfiltration/infiltration/embedding repertoires, altering times, temperatures, and hardener concentrations, with five repetitions (five samples) each: 1. Treatment 1 (T1) – The manufacturer’s instructions that accompany the resin kit (Leica Historesin Embedding Kit) suggested longer infiltration times for plant specimens, but details of each step are not specified. According to the manufacturer, an intermediate infiltration solution (95% ethanol: GMA, 1:1) is recommended to ensure even sample penetra-
Large Plant Samples: How to Process for GMA Embedding?
41
tion. Two hours of intermediate infiltration solution (preinfiltration) were employed here and samples were then transferred to a full-strength infiltration solution (GMA) for 24 h. The embedding solution used the GMA: hardener proportion suggested by the manufacturer (15 mL:1 mL) and it was placed into histomolds together with the tissue samples. All steps were conduced at room temperature. This treatment sought to establish the effectiveness of the standard protocol with large specimens. T1 Protocol:
Step
Duration
Conditions
Pre-infiltration (ethanol: GMA, 1:1)
2h
Room temperature
Infiltration (GMA)
24 h
Room temperature
Embedding (GMAa : hardener, 15:1)
∼30 min
Room temperature
a The resin volume must be sufficient to assure complete sample submersion in the histomold cell.
2. Treatment 2 (T2) – Increasing the pre-infiltration exposure time to 48 h, followed by extended cold infiltration in GMA (24 h in a refrigerator and 48 h in a freezer). Samples were kept for 48 h in a freezer to increase infiltration time under low temperatures. After this time period, embedding was proceeded using the GMA: hardener proportions suggested by the manufacturer (15 mL:1 mL). Samples were placed into histomolds and transferred to room temperature for polymerization. This treatment sought to determine effects of increasing infiltration time under cold conditions. T2 Protocol:
Step
Duration
Conditions
Pre-infiltration (ethanol: GMA, 1:1)
48 h
Room temperature
Infiltration (GMA)
24 h
Refrigerator (±5◦ C)
Infiltration (same solution as in previous step)
48 h
Freezer (±18◦ C)
Embedding (GMAa : hardener, 15:1)
∼30 min
Room temperature
a The resin volume must be sufficient to assure complete sample submersion in the histomold cell.
42
Paiva, Pinho, and Oliveira
3. Treatment 3 (T3) – Increasing the pre-infiltration exposure time to 48 h followed by cold infiltration in GMA (24 h under refrigeration) as in T2, plus cold infiltration of the embedding solution (48 h). The T3 protocol examined effects of prolonged cold exposure (more than 48 h) to the embedding solution (GMA + hardener). After this cold infiltration period, histomolds were transferred to room temperature for polymerization. It is important to carefully arrange samples in histomolds, as polymerization can be started during this stage since the hardener has already been added. This treatment sought to determine effects of correcting hardener volume by considering the amount of resin present in the tissue samples and extending GMA + hardener penetration time. T3 Protocol:
Step
Duration
Conditions
Pre-infiltration (ethanol: GMA, 1:1)
48 h
Room temperature
Infiltration (GMA)
24 h
Refrigerator (±5◦ C)
Infiltration: Embedding (GMAa : hardenerb )
48 h
Freezer (±18◦ C)
Embedding (same solution from previous step)
∼ 30 min
Room temperature
a The resin volume must be sufficient to assure complete sample submersion in the histomold cell. b The proportion of hardener must be calculated considering the resin volume outside the tissue sample plus 80% of the tissue sample volume (assuming that the sample is fully infiltrated with resin). In the present treatment, the sample has 0.512 cm3 (0.8 × 0.8 × 0.8 cm), which corresponds to a volume of 0.410 mL (80% of 0.512 mL) of resin inside each block (i.e. 2.05 mL of resin in the five blocks). This value must be considered in calculating the hardener volume.
3.4. Microtomy and Staining
The following methodology was employed: 1. After 24 h of polymerization in histomolds, the resin blocks were glued to wooden stubs (Fig. 3.1) using commercial epoxy glue (AralditeTM ). These blocks must be kept in a low-humidity environment; we recommend storage under silica gel. Using this procedure, samples can be maintained indefinitely before sectioning (see Note 5). 2. Transverse sections (10 μm thick) were obtained using a LeicaTM rotary microtome equipped with a steel knife (Cprofile). This microtome has a section counter and we discarded the first 400 sections, collecting four sections from the median region of each block. Twenty sections were therefore obtained from each treatment (five blocks, with four sections from each block).
Large Plant Samples: How to Process for GMA Embedding?
43
Fig. 3.1. Blocks obtained from the three treatments; from front to back, T1–T3, each row shows the five repetitions per treatment.
3. Each section was carefully transferred to a droplet of distilled water on a glass slide. 4. Slides were dried on a hot plate (50◦ C–60◦ C) until the water droplet completely evaporated (see Note 6). 5. Sections were stained with 0.05% Toluidine Blue, pH 4.7 (4) and mounted in distilled water under a clean coverslip. 3.5. Analysis
The 60 sections obtained (20 per treatment) were examined with a light microscope coupled to a camera lucida; the external outlines of sections were drawn, as were their internal outlines, when non-polymerized areas were present. Illustrations of the polymerized areas were analyzed using the Planimetry System software package (SPLAN) developed by CINAG, São Paulo State University, Botucatu, São Paulo, Brazil, to quantify the polymerized areas. For statistical analysis, we considered the average polymerized area per treatment. Averages were compared using the Tukey test (p < 0.05).
3.6. Main Results and Discussion
The tissue samples processed according to the manufacturer’s recommendations (T1) demonstrated non-polymerized areas in the core portion of blocks corresponding to 49.36% of their crosssection (Fig. 3.2a). During microtomy, these incompletely polymerized samples tended to fragment, making it very difficult to obtain whole sections. Additionally, these sections did not spread well on the glass slides and sections were only rarely suitable for microscopic analysis. This situation is often erroneously interpreted as being due to poor infiltration. Infiltration proved to be satisfactory and resin could be seen inside samples when using the standard manufacturer’s techniques. On the other hand, although resin penetrated
44
Paiva, Pinho, and Oliveira
Fig. 3.2. Blocks and sections obtained from the three treatments. a–c Treatment 1. a General aspect of the block (arrow indicates the non-polymerized area); this is the first repetition of T1, as indicated on the side of the wood stub (1/1). b General view of the section; note limits of hardened resin around the section (arrowhead). (c) Detail of the section, showing the hollow core (asterisk). d–f Treatment 2. d General aspect of the block (arrow indicates the non-polymerized area); this is the third repetition of T2, as indicated on the side of the wood stub (2/3). e General view of the section; note limits of hardened resin around the section (arrowhead). f Detail of the section, showing the hollow core (asterisk). g–i Treatment 3. g General aspect of the block without any non-polymerized areas; this is the second repetition of T3, as indicated on the side of the wood stub (3/2). h General view of the section; note limits of hardened resin around the section (arrowhead). i Detail of the perfectly extended section. Dotted lines in a, d, and g indicate the sample border.
to the sample core, polymerization was not observed in this region. If infiltration was therefore adequate, why did polymerization not occur in the sample core? Apparently, although the sample was well infiltrated, there was not enough hardener present for polymerization to occur rapidly and homogeneously. This is a crucial point, for once hardener is added to the activated resin and poured into histomolds, the polymerization occurs quite rapidly, completing in about 30 min at room temperature. This period apparently is not sufficient to permit hardener penetration into the tissue sample and polymerization only occurs in the external resin pool and on a small superficial portion of the sample, generating a decreasing gradient of polymerization towards the sample core (Fig. 3.2b, c).
Large Plant Samples: How to Process for GMA Embedding?
45
As explained below, we do not recommend using the manufacturer’s protocols for processing large samples. In T2, the pre-infiltration stage was extended from 2 to 48 h and the infiltration stage itself was conducted at low temperatures and divided into two stages: the first under refrigeration (5◦ C) for 24 h and the second in a freezer (−18◦ C) for 48 h. After infiltration, the embedding and hardening media was prepared following the manufacturer’s recommendations (15 mL GMA:1 mL hardener) and samples were placed into histomolds at room temperature for polymerization. Results of this procedure were similar to those of T1, with polymerization being restricted to the superficial portion of the tissue blocks; the non-polymerized area occupied 44.61% of the sample core (Fig. 3.2d). In addition to poor polymerization, sections demonstrated problems during mounting and staining, resulting in low quality material for microscopic analysis (Fig. 3.2e, f). We therefore conclude that increasing the infiltration stage under cooling improves infiltration but does not facilitate resin polymerization in the sample core upon addition of the hardener. In T3, the times of pre-infiltration and infiltration were identical to T2, but the hardener was added in the beginning of the second infiltration step when samples were kept in freezer, slowing the polymerization reaction. The hardener volume was also modified to compensate for the fact that large samples contain significant resin volumes (∼80% for the ginger samples in the present work). Thus the calculation of the amount of hardener to be added must consider the volume of resin used to compose the block around the sample plus 80% of the volume of the sample. For details of the hardener calculation, see Note 7. The T3 procedures yielded 100% polymerization of the five samples (Fig. 3.2g), resulting in sections with excellent quality for histological analysis (Fig. 3.2h, i). What caused these large differences in polymerization with this treatment? Two factors were critical: the addition of extra hardener in proportions appropriated for resin polymerization inside the sample and the extension of the infiltration/embedding times by maintaining histomolds with complete embedding media in a freezer. The addition of extra hardener would normally produce even faster polymerization than the conventional approach if maintained at room temperature, but maintaining a very low temperature allowed the GMA: hardener solution to penetrate into the tissue material without polymerizing. For polymerization, see Note 8. Statistical analyses presented in the following table confirm observations discussed here, showing the average polymerized area in each treatment, compared by Tukey test (averages followed by same letter do not differ statistically at a 5% level of probability):
46
Paiva, Pinho, and Oliveira
Treatment
Polymerized area (%)
T1
50.64b
T2
55.39b
T3
100.00a
a,b The average values followed by different letters, in the same column,
differ at 5% level of probability in the Tukey test.
It is important to note that this T3 procedure is only necessary when bulky or heavy materials are used, for the GMA: hardener solution penetrates very rapidly into small and porous materials, ensuring good results with this material while using standard procedures.
4. Notes 1. Several authors have noted fixation problems and the occurrence of artifacts due to the air present in the intercellular spaces in plant tissues. Unfortunately, air removal is not a common practice in plant anatomy laboratories, due in large part to the lack of vacuum pumps in plant anatomy laboratories. We were able to develop an alternative, very inexpensive and portable method that can substitute the use of a vacuum pump – a simple disposable syringe. The tissue samples were immersed in fixative solution inside a syringe (we found the 20 mL variety, without a needle, to be most useful) and the plunger was then pushed in to remove any air remaining in the syringe itself. When a small amount of fixative can be seen emerging from the syringe opening, seal that aperture with your forefinger (do not forget to wear gloves!) and then pull the plunger back until the inner syringe volume doubles; maintain the plunger in this position and observe the air bubbles exiting the sample. Repeat this process twice to ensure complete air removal. Samples are now air-free and the fixative is in contact with the innermost cells. Additionally, air removal facilitates resin infiltration. In addition to being extremely economical, air removal using syringes (our portable ‘vacuum pump’) allows adequate fixation of plant samples in the field, immediately after their collection. 2. According to Johansen (3), the minimum fixation time for FAA is 18 h and the material can then be left in this solution almost indefinitely without appreciable damage. This author also remarked that ‘this property of nearly perfect
Large Plant Samples: How to Process for GMA Embedding?
47
preservation makes FAA the ideal fluid to take on collecting trips’. Jensen (5) noted that tissues fixed in FAA for a minimum of 4 h can be stored indefinitely, but considered this solution to be only a moderately good fixative for most plant tissues: ‘In most tissues, it causes considerable shrinkage, hardens cells and occasionally makes sectioning difficult’. Formalin-aceto-alcohol is the most popular fixative in plant anatomy because it readily penetrates most tissues and is very cheap and easy to use in the field. To avoid problems detected by Jensen, samples can be conserved in 70% ethanol after FAA fixation. Experience has shown that fixation times should not be longer than 48 h and that samples should then be washed with 50% ethanol (which has the same concentration as the fixative and helps remove its residues) before storing in 70% ethanol. If samples will be processed quickly, they can be left overnight in 60 or 70% ethanol. 3. As GMA is water soluble, complete sample dehydration is not necessary (as stated in the Leica Historesin Embedding Kit) and the ethanol series was run only to commercial ethanol levels (92–98%). Do not forget to immediately wash (with ethanol) all the glasswares that come into contact with GMA + hardener. 4. Important: all materials employed during embedding (including hardener, histomolds and GMA, and most glassware) must be kept at low temperatures (∼5◦ C). Pipettes must be used at room temperature to avoid imprecision. If the glassware or histomold are warm, polymerization occurs more rapidly, making it difficult for the hardener to penetrate deep inside samples, which will result in unsatisfactory polymerization in cores of large tissue samples. Additionally, rapid polymerization makes it difficult to correctly position samples inside the histomold cells. Maintaining histomolds over ice will guarantee enough time to adjust the tissue sample positions. 5. The sample blocks embedded in GMA must be kept under very low-humidity conditions in a sealed recipient with silica gel crystals because GMA easily hydrates, becoming soft and very difficult to section. 6. We recommend temperatures of between 50 and 60◦ C for drying the glass slides on hot plates. At higher temperature (>60◦ C), air present in the water can form bubbles between the slide and the thin-section, in detriment to staining and microscopic analysis. 7. To calculate the amount of additional hardener needed for blocks with large specimens, assume that about 80% of the sample volume is resin. The sample volume must therefore
48
Paiva, Pinho, and Oliveira
be multiplied by 0.8 to calculate the amount of resin it contains (volume A). Add to volume A the amount of resin required to enclose the tissue sample in the histomold (volume B) and then calculate the quantity of hardener needed at the rate of 15 mL of GMA to 1 mL of hardener (volume C). Quickly mix the volume of resin for the block (volume B) with the hardener (volume C) in a beaker and quickly transfer it to the histomold with the sample; then move the mold immediately to the freezer. This operation must be performed quickly to avoid the beginning of polymerization. Cover histomolds with plastic film in order to protect the freezer from the volatile historesin, especially after adding the hardener. A sample calculation: a tissue sample that is 0.8 × 0.8 × 0.8 cm (i.e. 0.512 cm3 ) that is properly infiltrated with resin will have ∼80% of its volume filled by GMA. Volume A is therefore 0.512 (sample size) times 0.8 (to reach 80%), which equals 0.4096 mL. If histomolds hold 1.0 cm3 , then 0.488 mL of resin (volume of histomold minus the volume of the sample, i.e. 1.0–0.512) will be needed to top-off each cell (this is the volume B). Adding volume A (0.4096) to volume B (0.488) yields the total volume of resin (0.8976 mL), to be used to calculate the hardener volume in a proportion of 15:1. The volume of hardener (volume C) is therefore 0.05984 mL (the total volume of resin [0.8976 mL] divided by 15). Remember that you will use only 0.488 mL of new resin in each histomold because the rest is already inside the tissue sample. Thus for the embedding media you will mix 0.488 mL of GMA to 0.05984 mL of hardener per histomold cell. Remember, also, that these values are for each cell, and must be multiplied by the number of blocks to be made. If it is difficult to estimate the tissue sample volume because it has a very irregular contour, increase the volume of hardener by 40–80% in tests to determine the most efficient proportions. 8. When the embedding mixture becomes viscous at room temperature (25◦ C), the polymerization process should be finished in an oven at 40◦ C until final hardening (∼30 min) as suggested by Igersheim and Cichocki (6).
Acknowledgments Denise M.T. Oliveira and Elder A.S. Paiva thank CNPq for their research grants. This method was standardized during the development of projects that were partially supported by Brazilian foundations (CNPq, CAPES, FAPEMIG, and FAPESP).
Large Plant Samples: How to Process for GMA Embedding?
49
References 1. Gerrits, P. O. (1991) The Application of Glycol Methacrylate in Histotechnology: Some Fundamental Principles. Heidelberg, German: Leica Gmgh, 80p. 2. Feder, N., O’Brien, T. P. (1968) Plant microtechnique: some principles and new methods. Am J Bot 55, 123–142. 3. Johansen, D. A. (1940) Plant Microtechnique. McGraw-Hill, New York, NY, 523p. 4. O’Brien, T. P., Feder, N., McCully, M. E. (1964) Polychromatic staining of plant cell
walls by Toluidine Blue O. Protoplasma 59, 368–373. 5. Jensen, W. (1962) Botanical Histochemistry: Principles and Practice. W.H. Freeman, San Francisco, CA, 408p. 6. Igersheim, A., Cichocki, O. (1996) A simple method for microtome sectioning of prehistoric charcoal specimens, embedded in 2-hydroxyethyl methacrylate (HEMA). Rev Palaeobot Palynol 92, 389–393.
Chapter 4 Image Cytometry: Nuclear and Chromosomal DNA Quantification Carlos Roberto Carvalho, Wellington Ronildo Clarindo, and Isabella Santiago Abreu Abstract Image cytometry (ICM) associates microscopy, digital image and software technologies, and has been particularly useful in spatial and densitometric cytological analyses, such as DNA ploidy and DNA content measurements. Basically, ICM integrates methodologies of optical microscopy calibration, standard density filters, digital CCD camera, and image analysis softwares for quantitative applications. Apart from all system calibration and setup, cytological protocols must provide good slide preparations for efficient and reliable ICM analysis. In this chapter, procedures for ICM applications employed in our laboratory are described. Protocols shown here for human DNA ploidy determination and quantification of nuclear and chromosomal DNA content in plants could be used as described, or adapted for other studies. Key words: Image cytometry, C value, nuclear DNA content, chromosomal DNA content, picograms, quantitative microscopy, integrated optical density, Feulgen reaction, genome size.
1. Introduction Image cytometry (ICM) is an association of microscopy and digital-image-based techniques (1), particularly useful for spatial and densitometric measurements in the cytological area (2, 3). This tool has been widely used in quantitative applications for DNA ploidy studies in cancer pathology (4–6) and in absolute C value determination in plant (1, 7–9) and animal species (10, 11). Considering these approaches, the microscope, camera, and software industries have developed new integrated technologies in order to simplify the ICM laboratorial routine. With H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_4, © Springer Science+Business Media, LLC 2011
51
52
Carvalho, Clarindo, and Abreu
the evolution in digital cameras and system stations for highend laser analysis, ICM can be performed in almost all cytology laboratories using a good digital camera coupled to a light microscope (12). As main advantages in comparison with others methodologies, like flow cytometry (FCM), ICM carries out analysis in samples with relatively small number of nuclei (13), allowing visual recognition, in a semi-interactive way, in order to monitor structures of interest (14) and is cost-accessible for most specialized laboratories. Various examples of ICM studies in qualitative and quantitative DNA analysis have been reported in the following areas: (i) human pathology, mainly in tumorigenesis screening based on aneuploidy and genome size instability, such as diagnosis and monitoring of malignant melanomas and squamous cell carcinoma (5, 15), neoplasia (15, 16), cervical intraepithelial lesions and invasive carcinoma (5), and primary achalasia (17); (ii) animal genome size quantification, especially applied in taxonomic and evolution studies (10, 11); and (iii) plant science, for determination of absolute nuclear DNA values and ploidy level in agronomical (1, 7–9) and ecological species (18). Besides nuclear genome size, chromosomal DNA content has also been quantified by ICM, such as in Zea mays (19) and Capsicum annuum (20), resolving the genome size at the chromosomal level. The nuclear and chromosomal DNA contents, reported in picograms (pg) and/or base pairs (bp) have become useful in genome sequencing projects and genetic mapping (21). For nuclear and chromosomal DNA densitometric measurements by ICM, the system components basically include a computer with image analysis software, linked to a digital video camera coupled to a microscope (2, 14, 22, 23). Digital images are captured from slides subjected to Feulgen reaction (24) and then stored on hard disk. As images are recorded in pixel matrix, the system needs to be calibrated so as to grab images in known spatial and densitometric parameters (2). Since pixels do not have an intrinsic value, which depends on the system resolution, spatial and optical density (OD) values from micrometric and gray reference scales are attributed to them, respectively. To find the measurement of interest, the nucleus or chromosome area (μm2 ) is multiplied by their average OD, resulting in the integrated optical density (IOD). Generally, IOD values are automatically estimated by software algorithms (2, 23). Seeing that the IOD of the sample is equivalent to the DNA content, this value can be converted to pg or bp (1). As this methodology combines the high technology of optical, electronic, and digital instruments, additional technical details may be found in a wide variety of specialized references. Good theoretical references on digital cam-
Image Cytometry: Nuclear and Chromosomal DNA Quantification
53
era technology and optical microscopy have been overviewed (2, 12, 25). In this chapter, we describe our ICM protocols for system calibration and setup and for DNA content quantification in the following materials: (a) human and plant nuclei and (b) plant chromosomes. Protocols presented here are reproducible in our routine and were transcribed to contribute as a useful guideline for cytogenetic and cytometric laboratories. Given the need to assure accurate measurements, mainly in clinical practice, the application of ICM methodology requires important quality control of instrumentation and procedures (26–32).
2. Materials The equipments, reagents, labware trademarks and models cited in protocols correspond to those used in our laboratory routine, but similar ones can also be employed. We strongly recommend observing the hazard classification on the label and the safety data sheet of reagents used, according to the manufacturer’s warnings. All laboratorial precautions should be taken to avoid accidents. 2.1. Calibration and System Setup 2.1.1. Slide Preparation
R 1. A 5 mL tube with sodium heparin (Vacutainer ) for taking chicken red blood cells (CRBCs).
2. CRBCs are obtained from healthy male and female individuals (Gallus domesticus). 3. Micropipette and tips. 4. Glass microscope slides (26 × 76 × 1 mm), pre-cleaned with ethanol to remove dirt and oil. 5. Hot plate (surface temperature of 50◦ C). 6. Coplin jar with screw cap. 7. Freezer (−20◦ C). R ). 8. Fixative solution 1: 70% ethanol (Merck
9. Fixative solution 2: 4% phosphate-buffered formaldehyde, freshly prepared from 37% formaldehyde and stored at −20◦ C. 2.1.2. Feulgen Solution
1. Distilled water (dH2 O). 2. 5 M HCl: 42.5% (v/v) 37% HCl. Caution in the preparation: first add the dH2 O, then the acid. Store at room temperature.
54
Carvalho, Clarindo, and Abreu
3. Schiff’s reagent (see Section 2.2). 4. 0.5% SO2 water: 5% (v/v) 1 N HCl, 5% (v/v) 10% metabisulphite. Store at room temperature. 5. Immersion oils. Use a refractive index oil (nD = 1.525 − 1.540) between the slide and the coverslip, and a refractive index oil (nD = 1.515) between the coverslip and the microscope objective lens. 6. Glass coverslips (24 × 50 mm). 7. Nail polish. 2.1.3. System Setup
1. Trinocular Photomicroscope (OlympusTM BX-60) with: (a) stabilized light, (b) UPlanFI objective magnification ×40 with 0.75 numeric aperture, (c) PlanApo ×60 oil immersion objective with 1.40 numeric aperture, (d) PlanApo ×100 oil immersion objective with 1.40 numeric aperture, (e) Aplanat Achromat condenser with aperture 1.4, and (f) neutral density filter (ND6). 2. Monochromatic charge-coupled device (CCD) digital video camera of 12 bits gray and frame grabber card (Photometrics R CoolSNAP Pro – Roper Scientific, Tucson, AZ, USA) (see Note 1). 3. Computer Pentium 4 HT, CPU 3.2 GHz, 1 GB RAM, R 80 GB ROM, Microsoft Windows XP Professional V 2002 Operational System. R R 4. Image Pro – Plus 6.1 software (Media Cybernetics ).
5. Slide Micrometer (1000 μm, OlympusTM ) 6. Interference filter (green color, 550–570 nm). 7. Neutral density filters (ND6): 0.15, 0.30, 0.40, 0.60, 0.90, R and 2.50 (Edmund Industrial Optics , Barrington, NJ, USA). 8. Linear 11 stepped density filter (Edmund Industrial R , Barrington, NJ, USA). Optics 2.2. Preparation of Schiff’s Reagent
To prepare 100 mL: 1. Dissolve the reagent: in a glass bottle, dissolve 0.5 g basic fuchsin in 85 mL boiling dH2 O, with the help of a magnetic shaker (see Note 2). 2. Decolorize with bisulfite: when the solution temperature reaches 50◦ C, add 15 mL 1 N HCl and 2.23 g K2 S2 O5 and shake until dissolving. 3. Allow fuchsin to decolorize: after it cools down, store at 4◦ C in a dark glass bottle for 24 h, shaking occasionally to dissolve any pink precipitate. 4. Remove organic impurities: add 0.703 g (up to 1.0 g/100 mL is acceptable if demanded) activated charcoal and shake.
Image Cytometry: Nuclear and Chromosomal DNA Quantification
55
Then, pour through filter paper using a Büchner funnel as long as needed. Use only enough charcoal to decolorize the stain solution, which should be colorless or champagne color (23, 33). 5. Storage of Schiff’s reagent: store in a tightly capped dark bottle in a refrigerator (4◦ C), and use exposing as little as possible to light (see Note 3). 2.3. Slide Preparation for Ploidy Determination of Human Nuclear DNA by ICM
Items 3–9 under Section 2.1.1, items 1–7 under Section 2.1.2 and the following additional materials are needed: 1. Endocervical swab and tube kit for collection of cellular material. 2. Cervical cellular material (sample), collected by an authorized professional. 3. Cellular material of female buccal mucosa (standard). 4. Refrigerator (4◦ C). 5. Vortex. 6. Microcentrifuge tubes of 2 mL. 7. Microcentrifuge. 8. Diamond pencil. 9. Cytocentrifuge. 10. Filter paper.
2.4. Slide Preparation for Plant Nuclear DNA Quantification by ICM
Items 4–7, 9 under Section 2.1.1, 1–7 under Section 2.1.2, 6, 8 under Section 2.3, and the following additional materials are needed: 1. Plant nuclei material: root tips from germinated seeds of the standard (see Note 4) and sample species. 2. Petri dishes. 3. B.O.D incubator, with temperature ranging from 28 to 30◦ C. 4. Enzymatic solution (see Note 5). 5. Water bath 34◦ C. 6. Stereoscopic microscope. 7. Scalpel. 8. Fixative solution 3: methanol and glacial acetic acid in proportion of 3:1 (v/v), freshly prepared and stored at −20◦ C. 9. Fixative solution 4: 95% ethanol.
2.5. Slide Preparation for Plant Chromosomal DNA Quantification by ICM
Items 4–7 under Section 2.1.1, 1–4 under Section 2.1.2, 6 under Section 2.3, 2–9 under Section 2.4, and the following additional materials are needed: 1. Plant chromosomal material: root tips from germinated seeds.
56
Carvalho, Clarindo, and Abreu
2. Metaphase blocking solution: a chemical inhibitor for metaphase arrest, such as colchicine, oryzalin, trifluralin, 8-hydroxyquinoline and amiprophos-methyl. 3. Fixative solution 5: methanol, 37% formaldehyde and glacial acetic acid in proportion of 17:5:1 (v/v/v), freshly prepared.
3. Methods 3.1. Calibration and Setup of System 3.1.1. Slide Preparation
1. Prepare the slide using the smear technique: drop 5 μL male or female CRBC near one end of a slide and smear using another slide inclined at an angle of approximately 45◦ (see Note 6). 2. Air-dry the slide through fast arm waving and place it on a hot plate. 3. Store the slide in the fixative solution 1 in a Coplin jar at −20◦ C for at least 12 h, then fix it in solution 2 for 1 h at 25◦ C.
3.1.2. Feulgen Reaction (see Note 7)
1. After fixation, wash the slide in running water for 10 min and air-dry. 2. Hydrolyze in 5 M HCl at 25◦ C. The time should be adjusted from 10 to 60 min, according to cell type. Wash the slide three times (for 3 min each time) in dH2 O. 3. Stain the slide with Schiff’s reagent for 12 h at 4◦ C, in the dark. Subsequently, wash it three times (for 3 min each time) in 0.5% SO2 water and three times (for 1 min each time) in dH2 O and air-dry. 4. Drop 10 μL of immersion oil on the slide, add a coverslip, seal with nail polish and store it in the dark until image analysis, but no longer than two days.
3.1.3. Microscopy and Digital System Setup
1. All optical parts (filters, objectives, condenser and eyepieces) should be extremely clean. 2. Turn on the camera, computer and microscope, and wait for 15 min (see item 3 under Section 3.1.6) for light stabilization. 3. Adjust the microscope prior to each slide capture session by applying the Köhler method to obtain an optimal light path and, consequently, to reduce stray light. Attention! Do the setup for each objective magnification.
Image Cytometry: Nuclear and Chromosomal DNA Quantification
3.1.4. Spatial Calibration – From Pixel to Micrometer
57
1. Check your camera and software performance according to instruction manuals and make sure that particular functions are correct, especially the black and incident (white) level settings. 2. Use the slide micrometer scale and the software spatial calibration tools to establish the unit of measurement of the image from pixel to micrometer and record it in the preferences.
3.1.5. Density Calibration – From OD to IOD
1. Couple interference filter to field diaphragm. 2. Mount a “blank” slide, without cell material, add oil and lay a coverslip. Place it on the stage and focus at slide level. 3. Close field of the iris diaphragm to a slightly larger size than the image size. 4. Determine OD range (minimum–maximum) by opening the histogram live-window of the software and adjusting the microscope light intensity knob, so that the highest gray level (peak moves to the right) is slightly lower than the maximum value on the gray scale (Fig. 4.1). 5. Set 1/125 (8 ms) of exposure time for capture, or determine another value (i.e. 5–30 ms). 6. Use eight accumulated frames and divide by 8, in the accumulate settings command.
Fig. 4.1. Histogram showing the 12 bits gray range (0 to 4095 – coordinate x) and the light intensity peak at the position slightly lower than maximum saturation value on the gray scale.
58
Carvalho, Clarindo, and Abreu
7. Mount a step tablet with the standard neutral density filters set to calibrate the OD scale. For that, capture a series of empty images, interposing each one of these filters (or in combination to generate other OD values) over the interference filter in the light path (Fig. 4.2). 8. Use these step tablets and the density calibration tool of the software to fit the curve of intensity to OD and save it in preferences. This calibration process should be carried out for each objective (Fig. 4.2). 9. Remake the illumination settings before the capture session of every new slide.
Fig. 4.2. Plot (left) showing a linear density before calibration. Plot (right) of the calibrated intensity curve obtained from stepped tablet values mounted with standard neutral density filters (ND6 – 0.15, 0.30, 0.40, 0.60, 0.90, and 2.50, R , Barrington, NJ, USA). Edmund Industrial Optics
3.1.6. Calibration Tests
Calibration and evaluation of the image analysis system is performed based on three tests: stability (34), linearity (2, 23, 34), and uniformity (22). 1. Stability test: use the frame coordinates x, y on the center of the field to measure gray level of one pixel. Capture images every 3 min during 1 h. Plot the data on a spreadsheet to evaluate light source variations. Attention! In capture routine, use the analysis system only after its stabilization time (Fig. 4.3). 2. Linearity test: place the linear 11 stepped density filter on the microscope stage and capture each step. Use the density command to convert the captured image into a calibrated one and save. Apply the appropriate tool to segment each
Image Cytometry: Nuclear and Chromosomal DNA Quantification
59
Fig. 4.3. Stability test obtained from OD mean values (coordinate y) of one pixel (e.g. frame position: x = 150, y = 150) monitored every 3 min during 1 h (coordinate x). Note that this analysis system stabilized after 12 min.
step region and calculate the OD. Use the spreadsheet to compare, by linear regression, the obtained measurements with the known OD certified for each step (see Note 8 and Fig. 4.4). 3. Uniformity test: grab images, using the ×40 objective (×60 or ×100) of one single CRBC nucleus at distinct and well-distributed visual field positions (e.g. 36). Attention! Be careful to maintain the moved nucleus at the same focus. Calculate its OD values and coefficient of variation (CV) (see Note 9). 3.2. Slide Preparation for Ploidy Determination of Human Nuclear DNA by ICM 3.2.1. Protocol Employing Air-Drying Technique
1. Transfer 1 mL of cellular material (sample or standard) in a microcentrifuge tube and fill up to 2 mL with fixative solution 1. 2. Centrifuge the material at 100g for 5 min. Discard the supernatant and add 2 mL of fixative solution 1. Repeat this step three times, with intervals of 10 min and store at −20◦ C for at least 24 h.
60
Carvalho, Clarindo, and Abreu
Fig. 4.4. Linear 11 stepped density filter (above), showing 11 discrete density steps R , Barrington, NJ, USA), used in the with increments of 0.1 (Edmund Industrial Optics linearity test. Stepped tablet (below) exhibiting the 11 square images captured from each step of the filter and showing the 11 OD calibrated. The OD calculated (0.04–0.95) was compared with the certified for each step (0.04–1.00) by linear regression, using the spreadsheet. The R2 and SE obtained were 0.999 and 1.14, respectively.
3. Centrifuge the cellular suspension at 100g for 5 min. Discard the supernatant, add 500 μL of fixative solution 1 and store at 4◦ C. 4. Vortex at medium speed for 40–60 s. 5. Carefully drop 50 μL of sample cellular suspension near the left slide end; air-dry it through fast arm waving and place it on a hot plate. Repeat this step for the standard material near the right slide end (see Note 10). 6. Place the slide in fixative solution 2, in a Coplin jar, at 25◦ C for 1 h. 7. Repeat steps 1–4 of Section 3.1.2. 3.2.2. Protocol Employing Cytocentrifuge Technique
1. Repeat Steps 1–4 of Section 3.2.1. 2. Cytocentrifuge 200 μL of sample cellular suspension, so that the centrifuged material is imprinted centrally on the slide (on the right half), at 1800 rpm for 5 min. 3. Air-dry the slide by fast arm waving and place it on a hot plate. 4. Take the same slide and mount the centrifuge apparatus for standard cellular suspension, in a way that the centrifuged material is imprinted centrally on the slide, on the opposite end (i.e. on the left half). Cytocentrifuge at 1800 rpm for 5 min.
Image Cytometry: Nuclear and Chromosomal DNA Quantification
61
5. Repeat Step 3. 6. Repeat Steps 6 and 7 of Section 3.2.1. 3.3. Slide Preparation for Plant Nuclear DNA Content Quantification by ICM
1. Germinate seeds (sample and standard) in Petri dishes with dH2 O in a B.O.D., at 28–30◦ C. 2. Place the root tips in fixative solution 3, making three changes with 10 min intervals. Store at −20◦ C for 24 h. 3. Replace this solution with fixative solution 4, change three times and store at −20◦ C for 24 h. 4. Put the root tips in a microcentrifuge tubes and macerate them in an enzymatic solution (see Note 5) at 34◦ C in a water bath. 5. Wash the root tips for 10 min in dH2 O, place them in the fixative solution 4, changing it three times and store at −20◦ C for at least 24 h. 6. Using the diamond pencil, trace a soft transversal line across the middle of the microscope slide and mark “a” for sample and “b” for standard material, on the end corners. 7. Place the slide inclined at an angle of 25–30◦ under a stereoscopic microscope, keeping the slide center in focus. 8. Place the root tip (e.g. sample) right below the traced line. 9. Apply the cellular dissociation and air-drying techniques (35): quickly dissociate the meristem with a scalpel, while dripping one to three drops of fixative solution 4. Air-dry the slide through fast arm waving and place on a hot plate to completely dry up (see Note 11 and Fig. 4.5). 10. Incline the same slide at 25–30◦ under the stereoscopic microscope, this time with the other (empty) slide half in focus. 11. Place another root tip (e.g. standard) left below the traced line (see Note 10). 12. Repeat Step 9. 13. Place the slide immediately in fixative solution 2, in a Coplin jar, for 12 h at 25◦ C. 14. Repeat Steps 1–4 of Section 3.1.2.
3.4. Slide Preparation for Plant Chromosomal DNA Quantification by ICM (see Note 12)
1. Germinate the sample seeds in Petri dishes with dH2 O in a B.O.D, at 28–30◦ C. 2. Treat the root tip meristems with a blocking solution, with treatment time and concentration being established for each species. 3. Wash roots for 20 min in dH2 O and place them in fixative solution 3, changing it three times and store at −20◦ C for 24 h.
62
Carvalho, Clarindo, and Abreu
Fig. 4.5. Start position for cellular dissociation procedure of the plant nuclei and chromosomes slide preparation. Note that the root tip (white draw) is placed on the slide inclined at 25–30◦ .
4. Replace it with fixative solution 4, changing it three times and store at −20◦ C for at least 24 h (20). 5. Wash the root tips for 10 min in dH2 O, transfer to microcentrifuge tubes containing enzymatic solution (see Note 5) and incubate at 34◦ C in water bath. 6. Wash roots for 10 min in dH2 O, place them in fixative solution 4 and store at −20◦ C. 7. Prepare slides using the cellular dissociation and air-drying techniques (35): place each root tip on a slide inclined at 25–30◦ under a stereoscopic microscope. Quickly dissociate it with a scalpel, while dripping one to three drops of fixative solution 4. Air-dry the slide by fast arm waving movements and place it on a hot plate (see Note 11 and Fig. 4.5). 8. Place the slide in a fixative solution 5, in a Coplin jar, at 25◦ C for 12 h. 9. Repeat Steps 1–4 of Section 3.1.2. 3.5. Image Cytometry 3.5.1. Human Nuclear ICM
1. Place the slide on a microscope, recheck the microscope setup and recall the spatial and OD calibration settings. 2. Capture (×40 or ×60 objective) 20–40 isolated and wellpreserved nuclei of the standard.
Image Cytometry: Nuclear and Chromosomal DNA Quantification
63
3. Move to the other slide half (without changing the microscope setup) and capture over 200 isolated and wellpreserved target nuclei of the sample. 4. Make sure that the captured images are saved as calibrated ones. When not, open the customized spatial and density commands to convert the captured images to calibrated ones and save them. 5. Mount a frame with a collection of the standard nuclei as a “family portrait”, using copy and paste, macro-customized functions or programming sources. Repeat the procedure for the sample nuclei, placing them below the standard collection. 6. Select the IOD in the measurement parameters list. If the software does not have this parameter, alternatively select AREA and OD and multiply these values to obtain the IOD. 7. Apply the selection tool around the standard nuclei collection, adjust the appropriate segmentation level and apply the count/size tool to measure the IOD. Repeat this step for the sample nuclei collection. 8. Export the data to a spreadsheet for statistical analyses. 9. Find the IOD values MODE for the standard nuclei, divide each one of these values by the MODE value and multiply by two, in order to convert them into C value (36). 10. Open a histogram window and customize its X coordinate for 0, 2, 4, 6, 8 units and so on (2 corresponds to G1 diploid nuclei). 11. Plot these C values on the X coordinate of the histogram and use as reference. 12. Repeat Step 9 for sample nuclei, plot these C values on the X coordinate of the histogram and compare with the reference data (see Note 13 and Fig. 4.6). 3.5.2. Plant Nuclear ICM
1. Place the slide under the microscope, recheck the microscope setup and recall the customized spatial and OD calibration settings. 2. Capture (×40 or ×60 objective) ten isolated and wellpreserved early prophase nuclei of the standard and repeat the session for late telophase. 3. Move to the other slide end (without changing the microscope setup) and repeat the capture sessions for the sample. 4. Repeat Steps 4–8 of Section 3.5.1 (Fig. 4.7). 5. Measure the nuclear DNA content, using the formula:
64
Carvalho, Clarindo, and Abreu
Fig. 4.6. Histogram obtained from DNA ploidy values measurement of cervical nuclei. Note that images of nucleus (above) were selected to show some DNA ploidy level (C = 1.56–7.91).
Fig. 4.7 Collection of nuclei at the early prophase (P) and late telophase (T) captured with ×40 objective. a Nuclei of Pisum sativum L. “Ctirad” used as internal standard (2C = 9.09 pg). b Nuclei of Capsicum annuum L. “Fortuna Super” (2C = 8.10 pg). The CV and 4C/2C ratio obtained were 3.45 and 2.0, respectively. Bar 5 μm.
(I) 1 Cs = (IODs × ICp )/IODp where 1Cs = 1C nuclear DNA content of the sample; 1Cp = 1C nuclear DNA content of the standard; IODs = nuclear IOD value of the sample; IODp = nuclear IOD value of the standard; 3.5.3. Plant Chromosomal ICM
1. Place the slide under the microscope, recheck the microscope setup and recall the customized spatial and OD calibration settings.
Image Cytometry: Nuclear and Chromosomal DNA Quantification
65
2. Capture (×100 objective) the ten metaphases with all non-overlapping and well-preserved chromosomes. 3. Repeat Step 4 of Section 3.5.1. 4. Open a metaphase image file and use the software tools to assemble the karyogram, in the same frame, according to cytogenetic rules. 5. Repeat Step 6 of Section 3.5.1. 6. Select chromosomes, adjust the appropriate segmentation level and apply the count/size tool to measure the IOD of each of them. 7. Export the data to a spreadsheet and do the statistical analyses. 8. Measure the mean DNA content of each chromosome, using formulas: (I) 1Cn = (2Cn /r)/2 (II) IODc = (IODpc × n)/4 (III) IODt = IODc (IV) 1Cc = (1Cn × IODc )/IODt where 1Cn = mean 1C nuclear DNA content; 2Cn = 2C nuclear DNA content; r = number of FCM replications; IODc = mean IOD value of the chromosome with one chromatid (1C); IODpc = IOD of the chromosome with two chromatids (2C); n = number of metaphases; IODt = IOD of all chromosomes; 1Cc = mean 1C chromosomal DNA content.
4. Notes 1. Besides the PhotometricTM camera, other high-performance CCD digital cameras can be used (i.e. SonyTM , OlympusTM , AndorTM , and HamamatsuTM ). A good technical reference for a choice of the most appropriate digital camera for a particular interest can be found at the Scientific Digital Camera solutions page (AndorTM technology catalog – www.andor.com). 2. Alternatively, another reagent can be used: pararosaniline R (prepare the solution as recommended by the Merck microscopy support).
66
Carvalho, Clarindo, and Abreu
3. When a white precipitate begins to form, the stain solution should be discarded. Schiff’s reagent is commercially available, but sometimes it is not as reliable as freshly prepared solutions and, therefore, it is not recommended (23, 33). 4. An ideal DNA reference standard should be known in absolute value, genetically stable, easy to manipulate, available in sufficient quantities for analyses, and should have a genome size close to that of the target species. Different plant standards have been used, including Arabidopsis thaliana, Raphanus sativus, Pisum sativum (37, 38). 5. Carry out several test sessions to find the optimal time and concentration of the enzymatic treatment, according to cell type and species. Test the enzymatic solution prepared with pectinase, hemicellulase, cellulase, driselase, separately or combined. 6. This method is recommended for providing undamaged monolayers of blood cells and greatly preferred over “pushing” the blood, which can damage cells (23). 7. The Feulgen reaction principle basically includes: one hydrolysis step, using strong acid to generate free aldehyde groups in the DNA molecule, specifically by splitting up the purine bases A and G, thus producing “apurinic acid”, and another step, in which the fuchsin molecule decolorized with SO2 can bind and recover its pink color (24, 33). 8. In our laboratory, the software of the image analysis system automatically calculated R2 = 0.999 in the linearity test. The R2 and standard deviation considered adequate in the medical area are above 0.99 and below 5%, respectively (34). 9. Coefficient of variation below 3% is recommended for ICM analysis in cancer diagnosis (23, 34). Our CV value is lower than others of similar tests described in the literature. 10. In the internal standardization procedure, sample and standard nuclei are strictly processed on the same slide (7, 39, 40), in order to avoid random instrument drift and variation in the preparation and staining. 11. Alternatively, slides can be prepared using the squashing technique: meristems are squashed onto glass slides, coverslips removed over a cold plate, and slides air-dried (1, 7–9, 40). However, the cellular dissociation technique provides target material preparations flattened on the slide, well spread and showing little cytoplasmic background, overlaps or structural deformations of the chromatin (20, 35). 12. In this case, there is no direct reference standard. The nuclear DNA content of the same target species is used,
Image Cytometry: Nuclear and Chromosomal DNA Quantification
67
whose value can be established through nuclear ICM or by FCM, which also uses an internal standard (19, 20). 13. Basic performance standards for sampling, measurement, and scaling in ICM analyses must follow guidelines, consensuses, and reports of the European Society for Analytical Cellular Pathology – ESACP (27–32) and algorithms described by (26).
Acknowledgments We thank CNPq – Conselho Nacional de Desenvolvimento Científico e Tecnológico and CBP&D/Café – Consórcio Brasileiro de Pesquisa e Desenvolvimento do Café, Brazil, for their financial support. References 1. Vilhar, B., Greilhuber, J., Koce, J. D., Temsch, E. M., Dermastia, M. (2001) Plant genome size measurement with DNA image cytometry. Ann Bot Fenn 87, 719–728. 2. Chieco, P., Jonker, A., Van Noorden, C. J. F. (2001) Image Cytometry. Microscopy Handbooks 46. Springer, New York, NY, p. 116. 3. Pektas, Z. O., Keskin, A., Ömer, G., Karslio˘glu, Y. (2006) Evaluation of nuclear morphometry and DNA ploidy status for detection of malignant and premalignant oral lesions: quantitative cytologic assessment and review of methods for cytomorphometric measurements. J Oral Maxillofac Surg 64, 628–635. 4. Biesterfeld, S., Reus, K., Bayer-Pietsch, E., Mihalcea, A. M., Böcking, A. (2001) DNA image cytometry in the differential diagnosis of endocervical adenocarcinoma. Cancer Cytopathol 93, 160–164. 5. Böcking, A., Nguyen, V. Q. H. (2003) Diagnostic and prognostic use of DNA image cytometry in cervical squamous intraepithelial lesions and invasive carcinoma. Cancer Cytopathol 102, 41–54. 6. Grote, H. J., Nguyen, H. V. Q., Leick, A. G., Böcking, A. (2004) Identification of progressive cervical epithelial cell abnormalities using DNA image cytometry. Cancer Cytopathol 102, 373–379. 7. Greilhuber, J., Ebert, I. (1994) Genome size variation in Pisum sativum. Genome 37, 646–655.
8. Greilhuber, J., Obermayer, R. (1997) Genome size and maturity group in Glycine max (soybean). Heredity 78, 547–551. 9. Baranyi, M., Greilhuber, J. (1999) Genome size in Allium: in quest of reproducible data. Ann Bot Fenn 83, 687–695. 10. Gregory, T. R. (2001) The bigger the C-value, the larger the cell: genome size and red blood cell size in vertebrates. Blood Cell Mol Dis 27, 830–843. 11. Gregory, T. R. (2003) Genome size estimates for two important freshwater molluscs, the zebra mussel (Dreissena polymorpha) and the schistosomiasis vector snail (Biomphalaria glabrata). Genome 46, 841–844. 12. Russ, J. C., Russ, J. C. (ed.) (2008) Introduction to Image Processing and Analysis. CRC Press – Taylor & Francis Group, Upper Saddle River, NJ, p. 355. 13. Greilhuber, J. (2008) Cytochemistry and C-values: the less-well-known world of nuclear DNA amounts. Ann Bot Fenn 101, 791–804. 14. Rodenacker, K., Bengtsson, E. (2003) A feature set for cytometry on digitized microscopic images. Anal Cell Pathol 25, 1–36. 15. Bollmann, R., Méhes, G., Torka, R., Speich, N., Schmitt, C., Bollmann, M. (2003) Human papillomavirus typing and DNA ploidy determination of squamous intraepithelial lesions in liquid-based cytologic samples. Cancer Cytopathol 99, 57–62.
68
Carvalho, Clarindo, and Abreu
16. Motherby, H., Pomjanski, N., Kube, M., Boros, A., Heiden, T., Tribukait, B., Böcking, A. (2002) Diagnostic DNA-flow vs. image-cytometry in effusion cytology. Anal Cell Pathol 24, 5–15. 17. Gockel, I., Kämmerer, P., Brieger, J., Heinrich, U. R., Mann, W. J., Bittinger, F., Eckardt, V. F., Junginger, T. (2006) Image cytometric DNA analysis of mucosal biopsies in patients with primary achalasia. World J Gastroenterol 12, 3020–3025. 18. Temsch, E. M., Greilhuber, J., Krisai, R. (1998) Genome size in Sphagnum (Peat Moss). Bot Acta 111, 325–330. 19. Rosado, T. B., Carvalho, C. R, Saraiva, L. S. (2005) DNA content of maize metaphasic A and B chromosomes determined by image cytometry. Maize Genet Cooperation Newsl 79, 48–49. 20. Abreu, I. S., Carvalho, C. R., Clarindo, W. R. (2008) Chromosomal DNA content of sweet pepper determined by association of cytogenetic and cytometric tools. Plant Cell Rep 27, 1227–1233. 21. Doležel, J., Bartoš, J. (2005) Plant DNA flow cytometry and estimation of nuclear genome size. Ann Bot Fenn 95, 99–100. 22. Puech, M., Giroud, F. (1999) Standardization of DNA quantitation by image analysis: quality control of instrumentation. Cytometry 36, 11–17. 23. Hardie, D. C., Gregory, T. R., Hebert, P. D. N. (2002) From pixels to picograms: a beginners’ guide to genome quantification by Feulgen image analysis densitometry. J Histochem Cytochem 50, 735–749. 24. Feulgen, R., Rossenbeck, H. (1924) Mikroskopisch-chemischer Nachweis einer Nukleinasäure von Typus der Thymonukleinsäure und die darauf beruhende selective Färbung von Zellkernen in mikroskopischen Präparaten. Hoppe Seylers Z Physiol Chem 135, 203–248. 25. Andor TM Technology (2006) Scientific Digital Camera Solutions Catalog, p. 316. 26. Böcking, A., Adler, C. P., Common, H. H., Hilgarth, M., Granzen, B., Auffermann, W. (1984) Algorithm for a DNAcytophotometric diagnosis and grading of malignancy. Anal Quant Cytol 6, 1–8. 27. Chieco, P., Jonker, A., Melchiorri, C., Vanni, G., Van Noorden, C. J. F. (1994) An user’s guide for avoiding errors in absorbance image cytometry: a review with original experimental observations. Histochem J 26, 1–19. 28. Böcking, A., Giroud, F., Reith, A. (1995) Consensus report of the ESACP task force
29.
30.
31.
32.
33. 34. 35.
36.
37.
38.
39.
40.
on standardization of diagnostic DNA image cytometry. Anal Cell Pathol 8, 67–74. Haroske, G., Dimmer, V., Meyer, W., Kunze, K. D. (1997) DNA histogram interpretation based on statistical approaches. Anal Cell Pathol 15, 157–173. Giroud, F., Haroske, G., Reith, A., Böcking, A. (1998) ESACP consensus report on diagnostic DNA image cytometry. Part II: specific recommendations for quality assurance. Anal Cell Pathol 17, 201–208. Haroske, G., Giroud, F., Reith, A., Böcking, A. (1998) ESACP consensus report on diagnostic DNA image cytometry. Part I: basic considerations and recommendations for preparation, measurements and interpretation. Anal Cell Pathol 17, 189–200. Haroske, G., Baak, J. P. A., Danielsen, H. et al, Giroud, F., Gschwendtner, A., Oberholzer, M., Reith, A., Spieler, P., and Böcking, A. (2001) Fourth updated ESACP consensus report on diagnostic DNA image cytometry. Anal Cell Pathol 23, 89–95. Chieco, P., Derenzini, M. (1999) The Feulgen reaction 75 years on. Hystochem Cell Biol 111, 345–358. Vilhar, B., Dermastia, M. (2002) Standardization of instrumentation in plant DNA image cytometry. Acta Bot Croat 61, 11–26. Carvalho, C. R., Clarindo, W. R., Almeida, P. M. (2007) Plant cytogenetics: still looking for the perfect mitotic chromosomes. Nucleus 50, 453–463. Gonçalves, S., Haas, P., Spada, C., Fontana, C. S., Rangel, L., Mendonça, M. A. C., Carvalho, C. R. (2007) Citometria de imagem do conteúdo de DNA nuclear de células ´ epiteliais do colo uterino. Rev Bras de Anal Cl´ın 39, 71–78. Doležel, J., Sgorbati, S., Lucretti, S. (1992) Comparison of three DNA fluorochromes for flow cytometric estimation of nuclear DNA content in plants. Plant Physiol 85, 625–631. Johnston, J. S., Bennett, M. D., Rayburn, A. L., Galbraith, D. W., Price, H. J. (1999) Reference standards for determination of DNA content of plant nuclei. Am J Bot 86, 609–613. Thunnissen, F. B. J. M., Perdaen, H., Forrest, J. (1992) Influence of different cell extraction methods on cytometric features. Cytometry 13, 485–489. Greilhuber, J., Temsch, E. M. (2001) Feulgen densitometry: some observations relevant to best practice in quantitative nuclear DNA content determination. Acta Bot Croat 60, 285–298.
Chapter 5 Histological Approaches to Study Tissue Parasitism During the Experimental Trypanosoma cruzi Infection Daniela L. Fabrino, Grazielle A. Ribeiro, Lívia Teixeira, and Rossana C.N. Melo Abstract During acute infection with the parasite Trypanosoma cruzi, the causal agent of Chagas’ disease, tissue damage is related to intense tissue parasitism. Here we discuss histological approaches for an optimal visualization and quantification of T. cruzi nests in the heart, the main target organ of the parasite. These analyses are important to evaluate the course of the infection in different experimental models and also can be used to investigate parasite colonization and inflammatory processes in other infected tissues and biopsies. Key words: Chagas’ disease, Trypanosoma cruzi, histology, heart, inflammation, glycol methacrylate, animal models.
1. Introduction The flagellated protozoa T. cruzi is the causal agent of Chagas’ disease (also known as American trypanosomiasis), discovered around a century ago by the Brazilian physician Carlos Chagas (1). This disease remains a major problem with a great impact on public health in Latin America. Acute infections are usually asymptomatic, but the resulting chronic T. cruzi infections lead to high ratios of mortality, associated mainly with heart lesions (2, 3). This parasite has an obligate intracellular, proliferative, nonflagellate form, termed amastigote. Once the infection has established in vertebrate hosts, intracellular parasite replication occurs H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_5, © Springer Science+Business Media, LLC 2011
69
70
Fabrino et al.
as amastigotes, followed by the release of flagellate forms (termed trypomastigotes) that can be carried by the bloodstream to infect different organs, especially the heart (reviewed in (2, 4, 5)). Because of the intense parasitism in the myocardium induced by the acute Chagas’ disease, histological analyses of the heart are frequently performed during experimental studies to evaluate the course of the infection. However, the value of the histological techniques in enabling high-quality imaging of parasite nests can be limited by inadequate fixation, handling and/or processing/analysis of the tissue of interest. Here we report on the application of appropriate histological approaches for improved tissue morphology and optimal visualization and quantification of parasite nests in the heart. These approaches include the right choice and preparation of the fixative, embedding with a plastic resin for improved tissue resolution and adequate analysis of the semi-serial sections. These basic methodologies can also be used to investigate parasitism in other infected organs as well as to evaluate the presence of inflammatory infiltrates, which generally occurs in parallel to the tissue colonization by the parasite.
2. Materials 2.1. For Fixation
1. Phosphate buffer, 0.2 M, pH 7.3 (stock solution). For buffer preparation, dissolve 2.75 g of monobasic sodium phosphate (Na H2 PO4 .H2 O, MW = 137.99) into 50mL of distilled water and complete the volume up to 100 mL (final volume). Prepare another solution with dibasic sodium phosphate (Na2 HPO4 .7H2 O, MW = 268.14): dissolve 5.36 g of this salt into 50 mL of distilled water and complete the volume to 100 mL (final volume). To obtain the buffer at the pH 7.3 mix 20 mL of phosphate monobasic solution with 77 mL of phosphate dibasic solution. Adjust to pH 7.3 (see Note 1). 2. Paraformaldehyde (4 g) is dissolved in 40 mL of distilled water. Dilutions should be made in fume hood (see Note 2) under heating at a hot plate. Add 3–5 drops of 0.1 N sodium hydroxide and heat the solution until 60◦ C (do not allow going over this temperature). When the paraformaldehyde solution is completely mixed, cool to room temperature and complete the volume to 50 mL (final volume). Mix 50 mL of the paraformaldehyde solution with 50 mL of phosphate buffer 0.2 M. Adjust to pH 7.3. The final concentration of the fixative will be 0.1 M. Fresh solutions of paraformaldehyde should be used in each experiment for optimal fixation.
Histological Approaches to Study Tissue Parasitism
2.2. For Processing and Embedding
71
1. Glycol methacrylate (GMA) embedding kit prepared according to the manufacturer’s instructions. The GMA kit is composed of basic resin (solution of GMA monomer: 2-hydroxyethyl methacrylate) and other solutions such as polyethylene glycol and benzoyl peroxide. 2. Ethanol (from 70 to 100% in distilled water). 3. Plastic molds for embedding.
2.3. For Sectioning
1. Microtome. 2. Knife maker. 3. Glass knives − 400 × 25 × 6.4 mm. 4. Histological bath or large beaker. 5. Slides and coverslips.
2.4. For Staining and Mounting
1. Hematoxylin solution (Harris’ hematoxylin) can be purchased as a ready solution or alternatively be prepared. For preparation, dissolve 1 g of hematoxylin powder in 10 mL of 95% ethanol (solution I). Prepare the solution II: in a flask, to 200 mL of distilled water, add 20 g of potassium alum. Place the flask on a heater/stirrer, turn on the heater and allow mixing until the alum dissolves – this takes about 15 min. Mix solutions I and II quickly and allow boiling for 1 min. Add 0.5 g of red mercury oxide. The solution becomes dark purple. Take out the flask from the heater and allow cooling by immersing the flask in a container with cold water. After that, the solution must be stored for at least 48 h before use. The indicative sign that the solution is ready to use is the uprising of a metallic layer on the surface. Hematoxylin solution lasts for years, but eventually can deteriorate. The hematoxylin solution can be used as concentrated or diluted 1% in distilled water. To increase the nuclear contrast, 4% acetic acid can be added to the solution. 2. Two percent ammonium iron sulfate (also known as ferric ammonium sulfate or iron alum) dissolved in distilled water. 3. Yellow eosin solution is prepared by dissolving 1 g yellow eosin in 10 mL absolute ethanol and mixing this solution in 0.5 potassium dichromate dissolved in 80 mL distilled water. Just after, add 10 mL of saturated solution of picric acid (for saturation, add 1.4 g of picric acid in 100 mL distilled water). Eosin preparation should be made in fume hood. Alternatively, eosin solution can be purchased as a ready solution. 4. Acid–alcohol solution is prepared by dissolving 1 mL hydrochloric acid (HCl) (0.5–2%) plus 99 mL of 70%
72
Fabrino et al.
alcohol. Alternatively, an acid–water solution prepared with HCl or acetic acid (0.5–2%) can be used. 5. Toluidine blue-borate solution is prepared by dissolving 1 g sodium tetraborate (Na2 B4 O7 ) and 1 g toluidine blue O in 100 mL of distilled water. 6. Basic fuchsin solution (stock solution) prepared by dissolving 1 g basic fuchsin powder in 100 mL of 50% ethanol. The working solution is prepared by adding 3 mL of this solution to 60 mL of distilled water. 7. Mounting medium for histology.
3. Methods 3.1. Fixation, Processing, and Embedding
Fixation is a crucial step during sample preparation. This process is intended to preserve cell structure by avoiding tissue autolysis. In addition, fixation inhibits bacterial and fungal growth, and makes the tissue resistant to damage during the subsequent processing. Aldehyde solutions fix tissues by introducing cross-links between different tissue components (proteins, nucleic acids, and lipids). Generally, tissue fragments are fixed at least for 12 h. However, it is not recommended to fix more than 24 h. Sample collection for fixation requires a special care. Tissues must be collected as fast as possible, cut into small pieces (to enable optimal fixation) and promptly transferred to the fixative. If necessary, samples may be quickly washed in saline before cutting to clean excessive blood. It is very important to handle samples very gently to prevent mechanic damage that can lead to morphological alterations. For histological studies of the heart, it is important to define the anatomical region of interest (if atrium or ventricle, for example) before organ fragmentation. After this procedure, organ pieces are then carefully transferred to previously and adequately labeled vials containing the fixative. Fixation is followed by sample dehydration, a mandatory step before embedding. Embedding is performed by using a resin (glycol methacrylate) instead of paraffin (6). Glycol methacrylate is a water and ethanol-miscible plastic resin. One advantage of this resin is to avoid tissue damage induced by heating required for the embedding step with paraffin. Another critical advantage of plastic resin is that it permits increased tissue resolution, which is crucial to visualize fine cellular morphological details. Better resolution is due to plastic polymerization that causes less shrinkage and retraction compared to conventional techniques. Embedding with GMA has other important advantages when compared with
Histological Approaches to Study Tissue Parasitism
73
usual methods. This plastic resin is hydro-soluble, easily handled, and enables faster processing compared to paraffin embedding. Yet, it allows microtomy with section thickness varying from 0.5 to 12 μm (7, 8). The use of GMA embedding for improved histological, morphometrical, and immunohistochemical evaluations is also discussed in Chapter 1. Fix samples overnight at 4◦ C (see Note 3). Refer to Section 2.1 for fixative preparation. 1. Transfer samples to vials with cold phosphate buffer 0.1 M and, if necessary, keep at 4◦ C. Refer to Note 4 for sample storage. 2. Dehydrate samples. Dehydration step consists of successive baths in graded ethanol (70, 80, 90% – 20 min in each solution – followed by 100% ethanol (two changes of 20 min). Samples are kept in the same vial and the alcohol solution quickly replaced. Keep vials closed during the dehydration time. 3. Infiltrate samples in a first bath of resin (infiltrating resin) at room temperature, overnight. 4. Discard the resin (since it will contain alcohol) as the manufacturer’s instructions. 5. Infiltrate samples in a second bath of resin (pure resin), at room temperature, overnight. Use an amount of resin just to cover the sample. Refer to Note 5 for resin re-use. 6. After labeling molds with an identification code of your material, carefully transfer samples to within them and fill out with the embedding resin. 7. Keep molds inside a hood on a steady and plane surface for 24 h at room temperature. 8. Unblock and store blocks in a dry place. 3.2. Sectioning
For parasitism analyses, it is strongly recommended preparing semi-serial sections from the organ of interest to enable a reliable evaluation of the parasite nests within the organ. The indicated section thickness is usually 5 μm and it is important to keep a 70 μm-interval between sections to avoid parasite recounting (9, 10). 1. Set the microtome with the appropriate glass knife and knife holder for plastic resin. 2. Carefully cut the sample block. Sections come out as single sections. If you are performing semi-serial sections, collect just sections between intervals of interest. 3. Carefully transfer the selected sections to a recipient with tap water (histological bath or large becker) at room temperature. Use preferentially an antistatic tweezer to transfer
74
Fabrino et al.
sections to the water surface. Sections will stretch on the water and stay floating. 4. Collect sections with a clean slide and transfer it to a hot plate (70–80◦ C) until the water droplet evaporates (approximately 2 min). 3.3. Staining and Mounting
3.3.1. Hematoxylin–Eosin
Tissues embedded in GMA usually stain weakly with conventional methods when compared to paraffin-embedded tissues, and for this reason, some minor staining changes are required (11). Here we describe two useful and basic staining methods for tissue parasitism evaluations: hematoxylin/eosin and toluidine blue/basic fuchsin (12). Staining patterns obtained with these techniques are similar (Fig. 5.1). Both methods stain cell nuclei in dark-blue and cytoplasm and connective tissue fibers in various shades and intensities of pink (Fig. 5.1). However, different degrees of contrast can be obtained when toluidine blue/basic fuchsin is used (13). 1. Wash slides with sections in tap water for 5 min. 2. Immerse slides in the solution of iron–ammonium sulfate for 10 min. This solution acts as a mordant to increase the intensity of the hematoxylin staining. 3. Wash in tap water for 5 min. 4. Stain with Hematoxylin for 15–25 min (Note 6). 5. Rinse slides for 10 min in running tap water. 6. Immerse slides several times and quickly in an acid–alcohol or acid–water solution to enable nuclear differentiation (30 s to 1 min). Refer to Section 2.4, Section 4, for preparation of this solution (Note 7). 7. Rinse in tap water for 5 min. 8. Stain with eosin for 1–2 min. 9. Rinse in tap water for 2–5 min (Note 8). 10. Blot excess water from the slide by gently pressing the slide with a filter paper. Be careful not to touch sections. 11. Dry slides at room temperature (Note 9). 12. Mount using a conventional mounting medium for histology and let dry at room temperature or 37◦ C. (see Fig. 5.2 for the slide-mounting technique). Refer to Note 10 for removing excess mounting medium.
3.3.2. Toluidine Blue-Basic Fuchsin
1. Cover sections with a few drops of toluidine blue solution for 3–5 min on a hot plate at 55–60◦ C (see Note 11). Refer to Section 2.4, Section 5, for toluidine blue preparation. Wash in running tap water to remove excess stain. 2. Quickly dry on a hot plate (see Note 11).
Histological Approaches to Study Tissue Parasitism
75
Fig. 5.1. Light micrographs of the heart stained with hematoxylin-eosin (a, c) and basic fuchsin-toluidine blue (b, d) from uninfected (a, b) and infected (c, d) rats at day 12 of the acute infection with Trypanosoma cruzi. Histopathological analyses show parasite nests (arrows) and inflammatory processes (encircled in d). The boxed area in c shows a nest with amastigote forms of the parasite in higher magnification. Note in d a clear visualization of the cross striations in the cytoplasm of cardiomyocytes. Scale bar, 15 μm (a–c), 10 μm (d).
76
Fabrino et al.
Fig. 5.2. Procedure for slide-mounting. (1) Place the slide on a plain surface and apply a drop of a mounting medium in the center of the slide using an eyedropper. (2) Hold the cover slip at an angle of 45◦ so that it rests on the slide and touches one side of the medium drop. (3) Once the medium spreads along the coverslip edge, carefully lower the cover slip over the section and medium. Make sure there are no bubbles under the cover slip carefully. (4) Wait drying. Full drying at room temperature may require up to 48 h; however, drying can be accelerated at 37◦ C.
3. Cover sections with few drops of basic fuchsin solution (working solution) for 1 min at room temperature. Refer to Section 2.4, Section 6, for basic fuchsin solution preparation. 4. Wash well in running tap water. 5. Blot excess water from the slide by gently pressing the slide with a filter paper. Be careful not to touch sections. 6. Dry slides completely at room temperature. 7. Mount using a conventional mounting medium for histology and let dry at room temperature (see Fig. 5.2 for the slide-mounting technique). Refer to Note 10 for removing excess mounting medium. 3.4. Parasite Analysis and Quantification
Studies of tissue parasitism are dependent on histopathological analyses of a considerable number of sections of the organ of interest so as to have a reliable picture of the infection. In experimental research, it is also crucial to evaluate material obtained from at least three animals per group for optimal sampling. Parasite quantification can be performed by different methods. Our group has been using two methodologies for this purpose: enumeration of parasite nests and/or quantification of tissue areas occupied by the parasite. For both methods, we use 5 μmthickness sections. Moreover, considering that a single T. cruzi nest can reach up to 75 μm depth (9), enumeration of parasite nests must be performed on semi-serial sections with an interval
Histological Approaches to Study Tissue Parasitism
77
of 70 μm between them to avoid parasite recounting. By keeping this interval and using rats as experimental models, we found that 20–40 sections of the heart per animal (around 10–15 obtained from atrium and 15–20 obtained from ventricle) provide a consistent analysis of the organ. Trypanosoma cruzi nests can greatly vary in size. Enumeration of these nests is performed by analyzing successive fields (see Note 12) and counting the numbers of nests on a light microscope at magnification of 400× (Fig. 5.3). To compare different experimental groups, the same number of fields must be analyzed in each group. For example, in our experimental studies in rats, we generally evaluate a total of 1600 fields (800 from atria and 800 from ventricles) for each group of animals (3–4 animals). The tissue area occupied by parasites is obtained by using a microscope equipped with an integrating eyepiece usually with 100 squares at a magnification of 400× (14). The numbers of squares partially or totally occupied by the parasite are then enumerated and the proportion is calculated based on the total number of counted squares (Fig. 5.4). For example, by analyzing 145 heart sections from infected rats we found 1098 squares occupied by the parasite after counting a total of 43,000 squares. This represents an area of 2.20% occupied by the parasite. Using this methodology, it is also possible to determine the area occupied by cardiomyocytes and inflammatory processes. After enumeration, statistical analysis is usually done with a non-parametric test such as the Mann-Whitney Test (also known as U test or Wilcoxon rank-sum test) that is used to compare two independent groups of sampled data (15). Inflammation analyses can also be performed using the same sections prepared for parasitism studies. However, the quantification of inflammatory infiltrates is usually done at the magnification of 200×, which enables the observation of a higher area of tissue and a better identification of the intensity of the infiltrate and if it is focal or diffuse. On the other hand, cell features of the inflammatory and other cells are observed at 400× (Fig. 5.1).
4. Notes 1. To raise pH, add more phosphate dibasic solution. To low down pH, add more phosphate monobasic solution, always under magnetic stirring. 2. Paraformaldehyde is volatile and its fumes are very toxic (causes severe irritation to eyes, skin, and respiratory tract).
78
Fabrino et al.
Fig. 5.3. Schematic representation of the parasitism analysis on a light microscope. Correct enumeration (a) is achieved by counting parasite nests on successive fields at 400×. An incorrect analysis of fields (b) may provide a deficient picture of the infection.
Thus, any manipulation involving this chemical must be performed in a fume hood and wearing gloves. 3. To fix appropriately, aldehyde-based fixatives must be freshly prepared to avoid formic acid formation.
Histological Approaches to Study Tissue Parasitism
79
Fig. 5.4. Schematic representation of the parasitism evaluation using an integrating eyepiece on a light microscope. The numbers of squares partially or totally occupied by the parasite (marked with an “x” in the figure) are enumerated at 400× and the proportion is calculated based on the total number of counted squares.
Alternatively, these fixatives may be stored for a couple of months under minus 20◦ C. 4. After fixation, the tissue samples are transferred to vials containing phosphate buffer 0.1 M and can be stored in this solution at 4◦ C until processing. It is not recommended storing for more than 2 months. 5. The second resin can be re-used. Store at 4◦ C and re-use for the first bath of resin in another embedding procedure. 6. Filter the staining solutions immediately before their use. This procedure will avoid precipitation of sediments on sections. 7. Differentiation with acid solutions requires some practical experience to ascertain the correct end-point, since these solutions are used to bleach or decolor samples and the color of the tissue can be modified to red. The correct endpoint is when, after bluing up, the background is almost colorless. 8. Eosin is highly soluble in water. Over-staining is removed by washing in running water. 9. After staining, dehydration of sections is done just by letting sections dry at room temperature. Alcohol dehy-
80
Fabrino et al.
dration must be avoided since this solvent can damage sections. 10. After drying, any excess medium can be removed with a cotton swab moistened with a bit of xylene. 11. Slide drying on a hot plate can sometimes induce a light blue background on the resin. In this case, drying can be done at room temperature. 12. To avoid an incorrect enumeration of parasite nests, be careful not to overlap fields during the histopathological analysis on the light microscope (Fig. 5.3). References 1. Chagas, C. (1909) New human trypanosomiasis. Morphology and life cycle of Schysotrypanum cruzi, the course of a new human disease. Mem I Oswaldo Cruz 1, 159–218. 2. Melo, R. C. N. (2009) Acute heart inflammation: ultrastructural and functional aspects of macrophages elicited by Trypanosoma cruzi infection. J Cell Mol Med 13, 279–294. 3. Teixeira, A. R., Nitz, N., Guimaro, M. C., Gomes, C., Santos-Buch, C. A. (2006) Chagas’ disease. Postgrad Med J 82, 788–798. 4. Ropert, C., Ferreira, L. R., Campos, M. A., Procopio, D. O., Travassos, L. R., Ferguson, M. A., Reis, L. F., Teixeira, M. M., Almeida, I. C., Gazzinelli, R. T. (2002) Macrophage signaling by glycosylphosphatidylinositolanchored mucin-like glycoproteins derived from Trypanosoma cruzi trypomastigotes. Microbes Infect 4, 1015–1025. 5. Teixeira, A. R., Nascimento, R. J, Sturm, N. R. (2006) Evolution and pathology in Chagas’ disease – a review. Mem I Oswaldo Cruz 101, 463–491. 6. Melo, R. C. N., Rosa, P. G., Noyma, N. P., Pereira, W. F., Tavares, L. E., Parreira, G. G., Chiarini-Garcia, H., Roland, F. (2007) Histological approaches for high-quality imaging of zooplanktonic organisms. Micron 38, 714–721. 7. Bennett, H. S., Wyrick, A. D., Lee, S. W., McNeil, J. H. (1976) Science and art in preparing tissues embedded in plastic for light microscopy, with special reference to glycol methacrylate, glass knives and simple stains. Stain Technol 51, 71–97. 8. Cole, M. B., Jr., Sykes, S. M. (1974) Glycol methacrylate in light microscopy: a routine
9.
10.
11.
12.
13.
14.
15.
method for embedding and sectioning animal tissues. Stain Technol 49, 387–400. Hanson, W. L., Roberson, E. L. (1974) Density of parasites in various organs and the relation to numbers of trypomastigotes in the blood during acute infections of Trypanosoma cruzi in mice. J Protozool 21, 512–517. Melo, R. C. N., Machado, C. R. S. (2001) Trypanosoma cruzi: peripheral blood monocytes and heart macrophages in the resistance to acute experimental infection in rats. Exp Parasitol 97, 15–23. Cerri, P. S., Sasso-Cerri, E. (2003) Staining methods applied to glycol methacrylate embedded tissue sections. Micron 34, 365–372. Fabrino, D. L., Leon, L. L., Parreira, G. G., Genestra, M., Almeida, P. E., Melo, R. C. N. (2004) Peripheral blood monocytes show morphological pattern of activation and decreased nitric oxide production during acute Chagas’ disease in rats. Nitric Oxide 11, 166–174. Abreu, M., Baroza, L., Rossi, M. (1993) Toluidine blue-basic fuchsin stain for glycolmethacrylate embedded tissue. J Histotechnol 16, 139–140. Melo, R. C. N., Machado, C. R. S. (1998) Depletion of radiosensitive leukocytes exacerbates the heart sympathetic denervation and parasitism in experimental Chagas’ disease in rats. J Neuroimmunol 84, 151–157. Mann, H. B., Whitney, D. R. (1947) On a test of whether one of two random variables is stochastically larger than the other. Ann Math Statistics 18, 50–60.
Chapter 6 Intravital Microscopy to Study Leukocyte Recruitment In Vivo Vanessa Pinho, Fernanda Matos Coelho, Gustavo Batista Menezes, and Denise Carmona Cara Abstract The intravital microscopy is a valuable tool to capture images of cells in living organisms and to make studies of molecular determinants of leukocyte trafficking easier. Using this technique, we can directly visualize and measure each step of the leukocyte recruitment paradigm, including leukocyte rolling flux, rolling velocity, adhesion, and emigration. Thus, it is possible to understand the process involved in leukocyte homing as well as the cell recruitment to inflammatory tissues. Nowadays, two types of intravital microscopy are used routinely. The light microscopy is used to assess migration of intravascular cells in thin, tissues which must be sufficiently translucent. Epifluorescence microscopy allows the visualization of the microcirculation while permitting the distinction of leukocyte subpopulations in solid organs. Key words: Intravital microscopy, leukocyte, recruitment, inflammation, light microscopy, epifluorescence microscopy.
1. Introduction Intravital microscopy is an extremely useful tool used as a qualitative and quantitative method to observe leukocyte–endothelial cell interactions in vivo. For many years, the involvement of microcirculation in inflammation has been a fascinating and widely studied subject. The major point of concern at the outset was the now known involvement of the microcirculation in the inflammatory insult, but as time and techniques have progressed, the interest has moved onto leukocyte–endothelial interactions. One of the best techniques that have enabled scientists to understand these processes is intravital microscopy. The study H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_6, © Springer Science+Business Media, LLC 2011
81
82
Pinho et al.
of the microcirculation using intravital microscopy represents a sophisticated research tool to analyze complex biological interactions and disease mechanisms. This technique is also important in studies that aim to develop novel prophylactic and therapeutic approaches intending to modify microvascular disorders and cellular dysfunction associated with inflammatory diseases. Leukocyte recruitment is a multistep recruitment process that is optimally visualized and studied in vivo using intravital microscopy. Making use of this approach, we directly visualize and measure each of the steps of leukocyte recruitment paradigm, including leukocyte rolling flux, rolling velocity, adhesion, and emigration (1). The initial use of intravital microscopy date back century by Dutrochet in 1824 (reviewed in (1)). This researcher observed that leukocytes emigrated across small blood vessel walls. This finding that leukocyte emigration occurred due to tissue injury was finally reported in 1843 by Addison (reviewed in (1)). Several animals can be used for theses studies; however, rodents, especially mice due to genetic manipulations, have been more used. Using light microscopy, it is possible to assess the vasculature and leukocytes within some thin and transparent tissues without contrast-enhancing methods, such as hamster cheek pouch (2), mesentery (3), and cremaster muscle (4). In these preparations we can see leukocyte rolling, adhesion, emigration, and measure blood flux. Also, microcirculation of nontransparent organs (called generically as solid) and tissues can be studied using epifluorescence microscopy, including liver (5), spleen (6), brain (7), lymph node (8), intestinal wall (9), knee joint (10), and others. Epifluorescence microscopy can be useful for studies of leukocyte rolling and adhesion and vascular permeability (11). The intravenous injection of FITC-labeled BSA (fluorescein isothiocyanate-labeled bovine serum albumin) allows the evaluation of vascular albumin leakage from the microcirculation to the extravascular space as a parameter of vascular permeability in several tissues (12, 13). To study leukocyte recruitment, leukocytes can be labeled with dyes such as rhodamine (14) or with fluorochromes-conjugated antibodies (15). These strategies make possible the visualization of microcirculation while permitting the distinction of leukocyte subpopulations. Also, it is possible to visualize transfected cells that express the green fluorescent protein derived from transgenic mice (16). Concerning the intravital studies developed in leukocyte recruitment field, several cell structures and movements were clarified. A recent illustration of this situation is a new step on the leukocyte recruitment cascade, when cells “crawl” into endothelial layer, looking for a more adequate way to emigrate to extravascular spaces. Strikingly, the inhibition of certain subtypes of adhesion molecules can modify the way used by leukocytes to emigrate, especially, increasing the rate of transcellular
Intravital Microscopy to Study Leukocyte Recruitment In Vivo
83
emigration (through the endothelial cell) instead of the conventional intercellular extravasation (between the endothelial cells), forming indeed “dome-like” structures expressed by the endothelial cell that never had been visualized before (17, 18). The most recent advanced observation of microcirculation includes the spinning disk and the multi-photon confocal microscopy. The confocal laser scanning microscopy was an option to get rid of the out-of-focus haze of objects that can emit fluorescence. In order to improve the scanning velocity of image acquisition, a new different type of scanning method, called initially as Nipkow disk was developed and it is now currently known as spinning disk confocal microscopy. As a consequence of this improvement, a fast phenomenon would be more clearly observed as leukocyte movements under a blood flow. Indeed, lasers can have different wavelength, culminating in different fluorescence production by the stained cell or molecule, allowing different colors and co-localization of structures in a merged picture. The optical microscopy is still the only way to examine the four dimensions of a live phenomenon (the three conventional dimensions − x, y, z – allied with time course evaluation), which is very close to the realistic conditions found under physiological states. In addition to the spinning disk confocal technology, it is widely used nowadays in the acquisition of images using more than one laser wavelength simultaneously. Nowadays a twophoton microscopy has been used to study several physiological phenomena, such as for imaging electrical activity in deep tissues, for assessing the rate of blood flow, and for tracking immune-cell motility and morphology (19). The outlook of confocal intravital microscopy is very promising for physiological studies and many important advances are currently underway, like portable twophoton micro-endoscope for tridimensional imaging of mouse brains and lungs. Additional developments have been achieved on fluorescent probes and new fluorophores designed to provide a more intense bright using a lower time and a lower power of laser incidence, and improving the protection of the organ or tissue. In this chapter, we outlined techniques using a cremaster muscle preparation, as an example of light intravital microscopy, and a joint knee preparation, as an example of epifluorescence intravital microscopy.
2. Materials 2.1. Materials for Both Videomicroscopy
1. Ketamine (200 mg/kg). 2. Xylazine hydrochloride (10 mg/kg). 3. Bicarbonate-buffered saline (131.9 mM NaCl; 4.7 mM KCl; 1.2 mM MgSO4 ; 20 mM NaHCO3 ; pH 7.4).
84
Pinho et al.
4. P10 canule (Polyethylene tube). 5. Board with an optically clear viewing pedestal for cremaster (20) or knee joint preparation (21). 6. Surgery tools. 7. Cautery. 8. 4-0 suture thread. 9. Rhodamine 6G chloride and/or coupled antibodies. 10. LPS (lipopolysaccharide; sigma; B4:0111). 2.2. Light Microscopy Set
1. Light microscope with ×25 objective lens and ×10 eyepiece. 2. Video camera. 3. TV monitor. 4. DVD. 5. Optical Doppler velocimeter (Microcirculation Research Institute, Texas A&M University, College Station, TX, USA).
2.3. Epifluorescence Microscopy Set
1. Epifluorescence microscope. 2. Video camera. 3. Filter 45 MM NCP11. 4. Computer 2 DUO 8 GB. 5. Software NIS ELEMENT.
3. Methods Methods described below outline the preparation of an experimental model (mouse) for intravital microscopy (1), the surgery for cremaster preparation (2), and preparation of knee joint (3). The basic set up of intravital equipment for light intravital microscopy (see Fig. 6.1a) and epifluorescence intravital microscopy (see Fig. 6.1c) is described. 3.1. Mouse Preparation for Intravital Microscopy
Male mice are used for intravital observations. The process of leukocyte recruitment in cremaster muscle and knee joint (and also in other tissues) may be induced by administration of different inflammatory agents. Here we use LPS or specific antigens (21). 1. Anesthetize the animal with an intraperitoneal injection of a mixture of xylazine and ketamine hydrochloride. Although other anesthetics may be used, this approach generates very steady hemodynamic parameters, which is absolutely necessary to avoid blood flow effects on leukocyte recruitment.
Intravital Microscopy to Study Leukocyte Recruitment In Vivo
A
C
B
D
85
Fig. 6.1. Schematic representation of an intravital setup. a and c show the setup for light and epifluorescence intravital microscopy, respectively. In b and d is observed a cremaster muscle preparation (b) and the intra-articular synovial tissue (d) of the knee joint. Arrow indicates the synovial microcirculation. Scale bar = 0.5 cm.
2. Cannulate the left jugular vein to administer additional anesthetic or drug if necessary. 3. Administrate the inflammatory agent. For intravital microscopy in cremaster muscle, LPS is administered by subcutaneous injection beneath the left scrotal skin. For intravital microscopy in knee joint, an intra-articular injection with antigen (model of antigen-induced arthritis) is administered (21). 3.2. Intravital Microscopy of Cremaster Muscle
1. Cut the scrotal skin to expose the left cremaster muscle. 2. Carefully dissect this muscle free of the associated fascia. 3. Cut the cremaster muscle longitudinally with a cautery. 4. Separate the right testicle and the epididymis from the underlying muscle. 5. Move them back into the abdominal cavity. The muscle is held flat on an optically clear viewing pedestal and is secured along edges with 4-0 suture (Fig. 6.1b). 6. Superfuse the exposed tissue with 37◦ C warmedbicarbonate-buffered saline (0.15 M, pH 7.4) (see Note 1). 7. Examine the cremasteric microcirculation by using a light microscope equipped with ×10 or ×20 objective and a ×10 eyepiece. A video camera is used to project images onto a
86
Pinho et al.
monitor, and images are recorded for playback analysis using a DVD recorder. Single unbranched cremasteric venules (25–40 μm in diameter) are preferred and to minimize variability, the same section of cremasteric venule should be observed throughout the experiment (Fig. 6.2a).
AA
B
C
D
E
F
Fig. 6.2. Evaluation of the interaction between leukocytes and endothelial cells in the cremaster muscle microvasculation. a, b Images captured from the cremaster muscle microcirculation of mice, 4 h after intrascrotal injection with saline (a) or E. coli lipopolysaccharide (LPS, 0.05 μg/kg) (b). LPS induces a decrease of leukocyte velocity (e) and an increase of the number of rolling cells (c), adherent cells (d), and emigrated cells in the extravascular space (f). ∗ indicates a statistical difference in relation to saline group. p < 0.05, Student T test, n = 5/group. Scale bar = 50 μm.
3.3. Intravital Microscopy of Knee Joint
Unlike the cremaster muscle, the intravital microscopy is performed in the synovial microcirculation of the mouse knee. 1. Anesthetize the mouse and place the hind limb on a stage, with the knee slightly flexed and the patellar tendon mobilized and partly resected (see Fig. 6.1d). The intra-articular synovial tissue of the knee joint is then visualized for the determination of leukocyte rolling and adhesion (see Fig. 6.1d).
Intravital Microscopy to Study Leukocyte Recruitment In Vivo
87
2. Select 2–4 regions of interest (unbranched venules with 25–40 μm in diameter) in each mouse by using an ×20 objective and a ×10 eyepiece. 3. To measure the leukocyte–endothelial cell interactions, inject the fluorescent marker rhodamine 6G intravenously throughout caudal vein as a single bolus of 0.15 mg/kg immediately before measurements. Rhodamine epilumination is achieved with a 103 W/2 variable HBO mercury lamp in conjunction with a filter set (see Note 2). 4. Capture the microscopic images with a video camera and record them on a computer. Data analysis is performed offline using the imaging software NIS ELEMENT. The number of rolling, adherent, and emigrated leukocytes is determined offline as cited above. Rolling leukocytes is defined as those cells moving at a velocity less than that of erythrocytes within a given vessel. The flux of rolling cells is measured as the number of rolling leukocytes passing a given point in the venule per minute. Leukocyte rolling velocity is measured for the first 20 leukocytes entering the field of view at the time of recording and calculated from the time required for a leukocyte to roll along a
3.4. Evaluation of Leukocyte Recruitment
B
A
C
D
Rolling
Adhesion
10.0
72
*
cells/min
36
*
18
cells/100µm
* 54
7.5 5.0 2.5 0.0
0
saline
antigen
saline
antigen
Fig. 6.3. Evaluation of the interaction between leukocytes and endothelial cells in the synovial microvasculature. a, b Images captured after intra-articular administration of saline (a) and antigen (b) in mice. Rolling cells (c) and adherent leukocytes (d) to the synovial endothelium were assessed after injection of antigen or sterile saline (control) into the knee joint of mice. ∗ indicates a statistical difference in relation to saline group. p < 0.05, Student T test, n = 5/group. Scale bar =50 mm.
88
Pinho et al.
100 μm length of a venule. A leukocyte is considered to be adherent if it remained stationary for at least 30 s, and total leukocyte adhesion is quantified as the number of adherent cells within a 100 μm length of venule in 5 min (see Figs. 6.2 and 6.3). For light intravital microscopy, it is possible also to count leukocyte emigration (see Note 3 and Fig. 6.2). Leukocyte emigration is quantified as the number of cells in the extravascular space on the visible field adjacent to the observed venule. Only cells adjacent to and clearly outside the vessel under study should be counted as an emigrated leucocyte. All these parameters can be altered, for example, 4 h after the intrascrotal injection of LPS, as shown in Fig. 6.2. Mean red blood cell velocity can be measured using an optical Doppler velocimeter, which uses a pair of photodiodes to generate a voltage from an image of moving red cells that is a linear representation of red cell velocity. Wall shear rate is calculated based on the Newtonian definition as (mean red blood cell velocity/diameter) × 8 (s−1 ) (22).
4. Notes 1. The prevention of bleeding is crucial. Excessive bleeding will affect mainly the number of circulating leukocytes, and so it is crucial to assess the total leukocyte count after every experiment. During the whole experimental protocol, the animal should be kept under physiological conditions. So, dehydration, volume lost, hypothermia should always be avoided, especially on those preparations where the peritoneum cavity is exposed (liver, intestine, mesentery, etc.) and an important modification on animal parameters is induced. The cover of exposed organs that will not be observed with a soaked gaze, napkin of a PVC-wrap will minimize the water and temperature loss, and even reduce the movement of organs. Importantly, frequent and strong movement caused by animal breathing or instability of the microscope components will make the correct analyses of the data difficult. A warmer lamp (infrared) is strongly recommended in rooms where the temperature is lower than the mice (about 25–30◦ C) and in peritoneum-opened procedures. 2. Especially related to fluorescence microscopy, the continuous exposure of the same field for many minutes can cause “bleaching” of the fluorescence and incorrect analyses. The stability of the fluorophore should be evaluated
Intravital Microscopy to Study Leukocyte Recruitment In Vivo
89
in accordance with exigencies of the experimental protocol. Finally, as the leukocyte recruitment will begin since the first cells are injured, the improvement of the operatory skills will avoid unexpected leukocyte recruitment and will afford mainly the reproducibly of the data. An exceedingly traumatic operatory approach and even an incorrect administration of the pro-inflammatory mediators (dose and vial) will interfere in the local blood flow, which is a sine qua noncondition to leukocytes to get vessels on the field of view. 3. Light microscopy illuminates a translucent tissue and this facilitates the emigration enumeration. However, in the epifluorescence evaluation of solid organs, it is hard to clearly see cells outside microcirculation. Additionally, the leakage of rhodamine from the microcirculation to the extravascular space makes it difficult to track cells present in extravascular space. References 1. Springer, T. A. (1994) Traffic signals of lymphocyte recirculation and leukocyte emigration: the multistep paradigm. Cell 76(2), 301–314. 2. Duling, B. R. (1973) The preparation and use of the hamster cheek pouch for studies of the microcirculation. Microvasc Res 5(3), 423–429. 3. Kvietys, P. R., Perry, M. A., Gaginella, T. S., Granger, D. N. (1990) Ethanol enhances leukocyte–endothelial cell interactions in mesenteric venules. Am J Physiol 259, 578–583. 4. Goldberg, M., Serafin, D., Klitzman, B. (1990) Quantification of neutrophil adhesion to skeletal muscle venules following ischemiareperfusion. J Reconstr Microsurg 6(3), 267–270. 5. Vollmar, B., Glasz, J., Menger, M. D., Messmer, K. (1995) Leukocytes contribute to hepatic ischemia/reperfusion injury via intercellular adhesion molecule-1mediated venular adherence. Surgery 117(2), 195–200. 6. Schmidt, E. E., MacDonald, I. C., Groom, A. C. (1990) Interactions of leukocytes with vessel walls and with other blood cells, studied by high-resolution intravital videomicroscopy of spleen. Microvasc Res 40(1), 99–117. 7. Mooradian, A. D., McCuskey, R. S. (1992) In vivo microscopic studies of age-related changes in the structure and the reactivity of cerebral microvessels. Mech Ageing Dev 64(3), 247–254.
8. Von Andrian, U. H. (1996) Intravital microscopy of the peripheral lymph node microcirculation in mice. Microcirculation 3(3), 287–300. 9. Sekizuka, E., Benoit, J. N., Grisham, M. B., Granger, D. N. (1989) Dimethylsulfoxide prevents chemoattractant-induced leukocyte adherence. Am J Physiol 256, 594–597. 10. Veihelmann, A., Szczesny, G., Nolte, D., Krombach, F., Refior, H. J., Messmer, K. (1998) A novel model for the study of synovial microcirculation in the mouse knee joint in vivo. Res Exp Med 198(1), 43–54. 11. Hulström, D., Svensjö, E. (1979) Intravital and electron microscopic study of bradykinin-induced vascular permeability changes using FITC-dextran as a tracer. J Pathol 129(3), 125–133. 12. Kubes, P., Gaboury, J. P. (1996) Rapid mast cell activation causes leukocyte dependent and -independent permeability alterations. Am J Physiol 271, 2438–2446. 13. Cara, D. C., Ebbert, K. V., McCafferty, D. M. (2004) Mast cell-independent mechanisms of immediate hypersensitivity: a role for platelets. J Immunol 15, 4964–4971. 14. Baatz, H., Steinbauer, M., Harris, A. G., Krombach, F. (1995) Kinetics of white blood cell staining by intravascular administration of rhodamine 6G. Int J Microcirc Clin Exp 15(2), 85–91. 15. McDonald, B., McAvoy, E. F., Lam, F., Gill, V., de la Motte, C., Savani, R. C., Kubes, P. (2008) Interaction of CD44 and hyaluronan is the dominant mechanism for neutrophil
90
Pinho et al.
sequestration in inflamed liver sinusoids. J Exp Med 205(4), 915–927. 16. Stein, J. V., Rot, A., Luo, Y., Narasimhaswamy, M., Nakano, H., Gunn, M. D., Matsuzawa, A., Quackenbush, E. J., Dorf, M. E., von Andrian, U. H. (2000) The CC chemokine thymus-derived chemotactic agent 4 (TCA-4, secondary lymphoid tissue chemokine, 6Ckine, exodus-2) triggers lymphocyte function-associated antigen 1-mediated arrest of rolling T lymphocytes in peripheral lymph node high endothelial venules. J Exp Med 191(1), 61–76. 17. Phillipson, M., Kaur, J., Colarusso, P., Ballantyne, C. M., Kubes, P. (2008) Endothelial domes encapsulate adherent neutrophils and minimize increases in vascular permeability in paracellular and transcellular emigration. PLoS One 3(2), 1649. 18. Phillipson, M., Heit, B., Colarusso, P., Liu, L., Ballantyne, C. M., Kubes, P. (2006) Intraluminal crawling of neutrophils to emigration sites: a molecularly distinct process
19.
20.
21.
22.
from adhesion in the recruitment cascade. J Exp Med 203(12), 2569–2575. Benninger, R. K. P., Hao, M., Piston, D. W. (2008) Multi-photon excitation imaging of dynamic processes in living cells and tissues. Rev Physiol Biochem Pharmacol 160, 71–92. Cara, D. C., Kubes, P. (2004) Intravital microscopy as a tool for studying recruitment and chemotaxis. Methods Mol Biol 239, 123–132. Coelho, F. M., Pinh, V., Amaral, F. A., Sachs, D., Costa, V. V., Rodrigues, D. H., Vieira, A. T., Silva, T. A., Souza, D. G., Bertini, R., Teixeira, A. L., Teixeira, M. M. (2008) The chemokine receptors CXCR1/CXCR2 modulate antigen-induced arthritis by regulating adhesion of neutrophils to the synovial microvasculature. Arthritis Rheum 58(8), 2329–2337. Borders, J. L., Granger, H. J. (1984) An optical doppler intravital velocimeter. Microvasc Res 27(1), 117–127.
Section II Fluorescence Microscopy Applications
Chapter 7 Introduction to Fluorescence Microscopy Ionita C. Ghiran Abstract This chapter is an overview of basic principles of fluorescence microscopy, including a brief history on the invention of this type of microscopy. The chapter highlights important points related to properties of fluorochromes, resolution in fluorescence microscopy, phase contrast and fluorescence, fluorescence filters, construction of a fluorescence microscope, and tips on the correct use of this equipment. Key words: Fluorescence microscopy, fluorochromes, fluorescence filters.
1. A Brief History of Fluorescence Microscopy
The invention of fluorescence microscopy was the result of a series of fortunate but largely unrelated discoveries that culminated at the beginning of the 20th century. Long before, the curious properties of various minerals and especially plant extracts to either emit light when kept in the dark (phosphorescence) or reflect a Disclaimer: When preparing the manuscript I had to use examples that were pertinent to fluorescence microscopy and, therefore, to use actual fluorescence microscopes, objectives, cameras and image acquisition and analysis software. With one exception (the Nikon 63 × 1.49 objective, which was borrowed from Perkin-Elmer), all the other instruments and accessories were already present in the laboratory. The fact that I used those particular instruments to record images present in this chapter does not constitute a quality judgment, either positive or negative, about their optical and mechanical performances. The decision about purchasing a certain brand over another is a complex process that has to balance quality, price, previous experience, technical support and idiosyncratic personal preference.
H. Chiarini-Garcia, R.C.N. Melo (eds.), Light Microscopy, Methods in Molecular Biology 689, DOI 10.1007/978-1-60761-950-5_7, © Springer Science+Business Media, LLC 2011
93
94
Ghiran
certain color and transmit a different one when illuminated with sun light (fluorescence) were well known. A special place in the history of fluorescence is reserved for Nicolas Monardes, a 16th century Spanish botanist who wrote in “La historia medicinal de las cosas que se traen de nuestras Indias Occidentals” (“The medical history of the objects that are brought from our Occidental Indies”) about the peculiar properties of the yellow aqueous extract of the kidney wood (Lignum nephriticum) that had a distinct blue tint (fluorescence) in reflected light. However, the person most responsible for a scientific approach to the fluorescence phenomenon was George Stokes, an Irish physicist who noticed in 1852, while working at Cambridge University that the mineral fluorspar (or fluorite, CaF2 ) when illuminated with blue light, emitted red light. To describe this phenomenon he coined the term “fluorescence,” 1 based on the name of the mineral he used for his experiments. He was also the first to notice that fluorescent substances emitted light at a longer wavelength than the excitation light, describing for the first time the shift in wavelength that now bears his name. 1.1. Invention of Fluorescence Microscopy
At the same time physicists were trying to understand the phenomenon of fluorescence, a race for microscope objectives with improved resolution was underway in England and Germany. Before long (around 1886), the highest practical numerical aperture (NA) of objectives was reached based on Abbe’s (1872) published relationship between resolution and wavelength. It then became clear that one could produce higher resolution images with a given objective if light with a shorter wavelength (i.e. blue, violet and ultraviolet) was used to illuminate the object. Although invisible to the human eye, the existence of UV light had been known at that time for over 70 years due to research conducted in 1801 by the young German physicist, Johann Wilhelm Ritter (1776–1810). He noticed while studying the response of silver chloride (AgCl) to light that the intensity of the reaction increased as the interacting color was closer and even beyond the left end of the visible spectrum (blue–violet–ultraviolet) and decreased towards the right end (red and infra-red). Using blue and UV light appeared to be a good method to increase microscope resolution, but there were several significant drawbacks that made implementation difficult. First, ordinary glass is not very transparent beyond 380 nm; therefore, new quartz lenses had to be designed and employed for UV microscopy. Next, because the eye is not sensitive to UV light, images had to be focused using
1 The name fluorite describes the crystal’s low melting point, from the latin “fluo”, which means to flow that is also the origin of the word fluid.
Introduction to Fluorescence Microscopy
95
visible light. Due to uncorrected axial chromatic aberrations of objectives, out-of-focus images were recorded when structures of interest were photographed using UV light results. Finally, experimenting with UV imaging was potentially hazardous because all microscopes at that time used trans-illumination that exposed the observer directly to dangerous UV light. In 1904, while experimenting with blue and UV light, August Kohler (a German physicist who 10 years earlier, at only 27, described a new illumination method that became the gold standard for all microscopes produced since) noticed that certain tissues fluoresce when illuminated with UV light. He was the first to describe primary fluorescence (auto-fluorescence) of tissues, although he reported it initially as a nuisance rather than a discovery with possible practical applications. In 1913, a dedicated fluorescence microscope was built by Heinrich Lehmann and Stanislaus Josef Mathias von Prowazek that could observe and also measure fluorescence signals. However, the breakthrough that first demonstrated the exceptional potential of fluorescence microscopy in biology had to wait until 1941 when an American immunologist, Albert Hewett Coons, successfully used anthracene–isocyanate and later fluorescein to directly label pneumococcal anti-serum. Results obtained, although exceptional, were not well received by the scientific community due to low specificity (especially when using whole serum) and poor signal amplification when using monoclonal antibodies labeled with the rapidly fading fluorescein (1). A second significant breakthrough happened in 1954 when Thomas Weller (who received in the same year the Nobel prize in Medicine), working together with Albert Coons, developed and used succesfully an indirect method of staining, comprised of a non-labeled primary and a fluoresceinlabeled secondary antibody, to identify viral particles in culture (1). It is difficult to name another method that shaped cell biology, molecular biology and immunology more than the use of fluorescently labeled antibodies. Whether fluorescence microscopy, flow cytometry, real-time PCR or gene array, all these methods rely on the use of a fluorescently labeled probe (antibody, DNA, or RNA fragment) that specifically and quantitatively recognizes an epitope. Even the use of GFP-tagged proteins stems from the same basic idea.
2. What Is Fluorescence? Fluorescence is the property of atoms and molecules to emit light following excitation by an outside source of energy. A molecule capable of fluorescence is called a fluorochrome. Fluorescence
96
Ghiran
Fig. 7.1. Jablonski diagram of a theoretical fluorescent molecule showing most of the possible electronic transitions during excitation and emission (see text for details).
occurs when excess energy released by electrons returning to the ground state is released as photons (2). The illustration describing this process is known as the Jablonski diagram (Fig. 7.1), after the Polish physicist Aleksander Jablonski who used it first in 1935 to describe the transition of electrons during absorption and emission of light. Because fluorescence microscopy is based on fluorescence of organic molecules rather than atoms, we will use molecular fluorescence to discuss the physical basis of fluorescence. In steadystate conditions, electrons are found in the lowest energy level, S0 , or the ground state. When an organic substance absorbs photons with sufficient and appropriate energy, electrons from S0 transit to higher energy states called excited single states, S1 , S2 or Sn . The transition phase from S0 to S1 or S2 is very fast and takes about 10−15 s. The excited single states of organic molecules are not as discreet as those of atoms, so the energy levels become energy bands. The direct consequence of this is the very broad excitation and emission spectra of organic fluorochromes (Fig. 7.2) compared to the sharp peaks of the emission spectrum characteristic of individual atoms (see below). The ground or excited state is called single because all electrons on these bands are spin-paired. When the excited electron returns to the ground state it can take several paths depending on the molecular configuration of the molecule, and the presence of other species of molecules in the media, can result in 1. fluorescence 2. phosphorescence 3. radiation-less conversion (heat)
Introduction to Fluorescence Microscopy
97
Fig. 7.2. Excitation and emission spectra of a hypothetical fluorescent molecule illustrating the red shift (Stokes’s shift) of emission wavelength compared to excitation. The amount of the shift depends on the molecular composition of the molecule. Fluorochromes with larger shifts are desirable for fluorescence microscopy.
When returning from an excited state, electrons usually go from one energy level to the next energy level down, releasing excess energy as heat in a process known as internal conversion. Because the energy levels S2 and S1 are very close, the probability that an electron from S2 will bypass S1 and go directly to S0 is significantly lower than going from S2 to S1 . Therefore, most fluorescence happens when electrons return from S1 to S0 . The difference in energy between S1 and S0 is the energy of the emitted photon (Kasha’s rule). In other cases the energy lost by an electron going from S1 to S0 can be given up as heat. In this situation there will be a fluorescence-less process, where no photons are emitted by the organic molecule. The energy state of electrons depicted on the right side of the Jablonski diagram is labeled “triplet state,” denoting that one set of electron spins is unpaired. This happens when spin of an electron from one of the higher vibrational energy levels (S1 or S2 ) is reversed, switching the electron to the triplet state through a process called intersystem crossing. The triplet state is depicted below the corresponding singlet state with the same electronic configuration because it has a slightly lower energy level. Therefore, the lowest excited state of a molecule is often a triplet state. Although the return from this state to the ground state does not result in fluorescence but rather phosphorescence or heat, the triplet state is important because in this state electrons are quite reactive, leading to intermolecular reactions that cause a decrease in fluorescence efficiency (photobleaching) and to interactions with oxygen that produce cell-toxic, reactive intermediate species (3). 2.1. The Chemical Basis of Fluorescence
Certain organic molecules fluoresce whereas others do not. What makes molecules fluorescent is the presence of conjugated, alternate double bonds that allow electrons to be displaced through
98
Ghiran
the rigid and planar molecule. Therefore, the benzene ring (C6 H6 ) is a very common group among fluorescent molecules. Usually there is a correlation between the size of the molecule (the number of double bonds) and wavelengths of the excitation and emission states, specific for those particular molecules.
3. Properties of Fluorochromes 3.1. Fluorescence Quantum Yield
Fluorescence Quantum Yield (quantum efficiency, QE) is a measure of the efficiency of the fluorescence process. It is defined as the ratio between the number of emitted photons and the number of absorbed photons. This does not mean that in the case of fluorescein that has a quantum yield of 90%, the energy of the emitted photons is 90% of the energy of the absorbed photons. It means that number-wise, there is a 10 to 9 relationship between excitation and emission. The loss in energy is due to non-radiative processes (Stokes’s losses) that render the emitted photon to be red-shifted compared to the excitation photon. However, the QE of fluorescein varies greatly under different conditions, especially pH. At a pH of 8, fluorescein has the highest QE; whereas as the pH drops towards 6, the QE reaches 0.3. This is important because the same amount of fluorescein will fluoresce brightly outside the cell and be very dim once inside the cell in an acidic compartment (i.e. phagosome). Therefore, one should avoid fluorescein alone to track phagocytosed particles (or any biological processes where the pH varies) and choose a pH-insensitive dye. On the other hand, tagging a particle with both fluorescein and a pH-insensitive dye represents a very powerful method that allows one to track particles intracellularly and simultaneously monitor pH changes.
3.2. Quenching
Quenching is defined as a decrease in the QE due to interactions with molecular species nearby such as proteins or fluorochromes. The energy transfer is non-radiative and results in less efficient fluorescence. The emitted light of a quenched fluorochrome displays identical spectral characteristics to the non-quenched fluorochrome. Commonly, fluorochromes lose some of their QE following conjugation to proteins such as BSA (bovine serum albumin) or IgG (immune globulins) if the conjugation density of the fluorochrome is too high. Therefore, it is important to keep the molar ratio fluorochrome-to-protein low rather than high to prevent quenching due to fluorochrome–fluorochrome interactions. (Each company that sells fluorescence labeling kits has an optimal range depending on the size of the fluorochrome and the size of the protein.) Other molecules such as ANS
Introduction to Fluorescence Microscopy
99
(8-anilino-1-naphthalene sulfonate) behave differently, displaying low fluorescence in polar solvents and increased fluorescence upon interaction with various proteins. Quenching is not always detrimental; it may offer useful information about the nearby species as well as the distance between them, which is always at least an order of magnitude smaller than the resolution of a standard fluorescence method. 3.3. Photobleaching
Unlike quenching, which can be reversible, photobleaching is generally irreversible. This process is due to electrons transitioning from single to the triplet state; electrons will interact efficiently with other molecular species with triplet ground states such as the ubiquitously present oxygen. The net result is an irreversible change in the molecular structure of the fluorochrome, which results in a permanent loss of fluorescence. Another side effect of this interaction is the formation of reactive oxygen species in or around the fluorescently labeled cell that will chemically interact and damage cellular structures. In our experience, very low concentrations of ascorbic acid (nanomolar range) offer a cell friendly way to prevent photobleaching and oxygen-mediated cell damage.
3.4. Molar Extinction
Molar extinction (ε) (synonyms: molar absorption or molar absorptivity), as defined by the Beer-Lambert law, measures the efficiency of a fluorochrome (or any substance) to absorb light at a certain wavelength (λ). Importantly, the absorption depends on the molecular structure of the molecule, the wavelength of the absorbed light, pH value and temperature. Formula: A = εcl where A=absorbance of the homogenous, isotropic sample ε = molar extinction c = concentration of fluorescent dye (mol/L) l = thickness of the fluorescent solution, expressed in cm (although theoretically it should be expressed in meters)
3.5. Fluorescence Lifetime
Fluorescence lifetime represents the average time a molecule spends in the excited state before returning to the ground state. In Fig. 7.1, we saw that following excitation, molecules transit from the ground state to excited states (Sn ). The deactivation phase is a two-step process that includes a non-radiative (internal conversion) and a radiative (fluorescence) process that depopulate the excited phase. If there is no energy transfer to the environment (no acceptors), the fluorescence lifetime is constant for
100
Ghiran
a given molecule in a given solvent and is usually in the order of nanoseconds. However, when acceptors such as oxygen, calcium, magnesium or hydrogen are present, the lifetime of fluorochrome is significantly decreased. Measuring fluorescence lifetime (not general fluorescence) can offer precise information about the environment surrounding the fluorochrome that allow researchers to map cells with unsurpassed space and time resolution for parameters such as: calcium, magnesium and pH, as well as for the presence of certain quenching molecules. There is little doubt that fluorescence lifetime imaging will be a widespread fluorescence microscopy method in the years to come. 3.6. Stokes Shift
4. Resolution in Fluorescence Microscopy
Stokes shift represents the shifted emission of a fluorochrome compared to excitation (Fig. 7.2). When the excitation and emission spectra are displayed together the Stokes shift is represented by the distance between the two peaks. The curve on the left side of graph is called the excitation spectrum and in general (but not always) is identical with the absorption spectrum of the fluorochrome. When the spectrum of excitation is very different than that of emission, it means that non-fluorescent species, likely dimers, are present in the solution (4). A better representation of the excitation curve would be against the molar extinction coefficient, which is the wavelength-dependent absorptivity (see above) of that particular fluorochrome in the solvent used to generate the spectra. However, because excitation and emission spectra are generally displayed on a single graph, fluorescence excitation is used against various wavelengths. The emission spectrum is a plot of the fluorescence intensity versus various wavelengths. The shape of the emission plot does not depend on the excitation wavelength, but only on the molecular identity of the fluorochrome due to Kasha’s rule (see above).
Classically, resolution of an optical microscope is defined as the minimal distance (expressed in μm) at which two points in the sample are seen separately. Ernst Abbe showed in 1872 that the minimal distance between two points (d) that are imaged using transmitted light (a regular bright-field microscope) depends on the half-angle of the aperture of the objective (sinα), the wavelength (λ) of the light used to form the image and the refractive index (n) of the media between the frontal lens and the sample (or coverglass). While the aperture (opening) of an objective is measured in degrees, the NA is an adimensional value that depends on the aperture of the objective and the refractive
Introduction to Fluorescence Microscopy
101
index of the media between the sample and the frontal lens. The smallest detail resolved with an objective with a NA of 1.40 (the highest possible number until a few years ago without using special immersion oil and special cover glasses; the value now is 1.49, used with green light (λ = 550 nm) and with an immersion condenser (NA = 1.40), so called double immersion) on a perfectly aligned microscope would be about 220 nm. d=
λ λ = NAcond + NAobj 2NA
Anyone who has ever used a fluorescence microscope knows that when using fluorescence, it is possible to see light originating from structures that are about 10 times smaller than the theoretical resolution of the best immersion lens. Even with a dry objective (40×), it is possible to see quantum dots, which are fluorescent crystals the size of a ribosome (about 20 nm). So what is the explanation? In fluorescence microscopy, the resolution does not measure the size of the smallest self-luminous point it can detect but rather the radius of the image formed by the objective that represents that self-luminous point. If one images three particles having 20, 100 and 180 nm in diameter, the microscope will see them as having the same size. The intensity associated with particles may be different depending on their actual size but as far as the final size the microscope will show, all particles will be the same. In the same way, we can see on a dark night, a lit candle miles away or stars that are hundreds of light years away from us although our eyes do not have the angular resolution needed to resolve neither the flame nor the disk of the stars at that distance. What we see is the light coming from the flame or stars because of the dark background. Notably, the light seen with unaided eyes will contain no details regarding the shape or original size of the luminous points. Similarly, fluorescence microscopy allows us to detect the presence of fluorescent particles that are significantly below the resolution limit of the objective without any indication of structure, shape or actual size. The initial work regarding the resolution of optical instruments imaging self-luminous objects was done using astronomical instruments (stars are perfect selfluminous points). Therefore, not surprisingly, the theory explaining the resolution of fluorescent microscopy is based on the work of a British astronomer George Airy. He was the first to show that when in focus, the image of a star seen in a telescope is not a point but rather a small bright disk that is surrounded by larger and ever fainter bright disks alternating with dark disks. Based on Airy’s work, Lord John Rayleigh (who received the Nobel Prize in 1904) derived an empirical formula that shows that in fluorescence microscopy the smallest fluorescent structure will form a disk with a radius of
102
Ghiran
r=
0.61λ 1.22λ 0.61λ = = NAcond + NAobj NA n sin α
In this formula, λ is the wavelength of the fluorescence (emitted light), α is the half-angle of the objective aperture and n is the refractive index of the media between the frontal lens and the sample. In fluorescence microscopy, the objective is the condenser; therefore, the NA of the condenser will be equal to the NA of the objective. Rayleigh resolution criteria also states that in order for two self-luminous points to be resolved, the center of one object has to be on the first minimum (the first dark ring that surrounds the airy disk) of the adjacent point. Although this depends significantly on one’s ability to resolve different shades of gray, novel digital methods used to record and process microscopy images help standardize and even exceed this criterion (5).
5. The Construction of a Fluorescence Microscope
Current fluorescence microscopes are based on the design used by Johan Ploem that was developed while working together with Leitz in the early 1960s.2 Today, virtually all fluorescence microscopes are based on Ploem’s design, and, therefore, we will present it in the following pages. The simplest way to represent the basic fluorescence microscope design is shown in Fig. 7.3a. The light emitted by the fluorescence light source (excitation) bounces off of a dichromatic mirror, passes through the objective and excites fluorochromes in the sample. Due to the nature of the dichromatic mirrors that behave like true mirrors for shorter wavelengths and are transparent to longer wavelengths∗ , the fluorescent light (emission) originating in the sample, passes unreflected through the dichromatic mirror towards the detector, which can be the eye or an electronic device (i.e. CDD camera). Some spinning disk confocal microscopes have dichromatic mirrors that allow shorter wavelengths to pass through and reflect the fluorescent light to the detector. Figure 7.3b shows the fluorescence path of a commercial fluorescence microscope along with the bright-field in the lower part of the microscope. Although the
2
Due to poor circulation of scientific journals at that time and lack of translation of most scientific articles written in languages other than English, neither Ploem nor researchers from Leitz were aware that more than 5 years before, two Russian scientists, Brumberg and Krylova, did not only described the advantages of employing “dividing mirrors” in fluorescence microscopy but actually used them.
Introduction to Fluorescence Microscopy
103
Fig. 7.3. Diagram illustrating the basic principle of epi-fluorescence microscopy. The light source emits light that is reflected by the dichromatic mirror through the objective into the sample. The fluorochromes present in the sample following excitation will emit light of longer wavelength (fluorescence) that will go back through the objective and dichromatic mirror and form the final image on a detector that can be either the eye or a CCD camera.
presence of a series of lenses, diaphragms and filters in the fluorescence path make the design of the fluorescence microscope fluorescence path seem more complicated than the schematic, the basic principles apply. Therefore, to understand how a fluorescence microscope works, and more importantly be able to get the most out of it and troubleshoot potential problems, we shall start by describing components of a fluorescence microscope following the light path, as shown in Fig. 7.4a. Although it is not obvious, a fluorescence microscope is basically a folded bright-field microscope (see Fig. 7.4a), where the objective lens play the role of both the condenser and the objective. As a rule, in epi-fluorescence microscopy everything involved in image formation is located above the microscope slide. None of the controls that are located under the microscope stage, such as the ones that modify the condenser height, the size of the field and aperture diaphragms, the position of neutral density, or green interference filters affect the fluorescence image. In some particular instances, the height of the condenser may have a negative impact on the image (see below), but in general all of the adjustments for the fluorescence microscope are located on the fluorescence illuminator. On a bright-field microscope, the light that forms the image originates from the halogen light bulb located for upright microscopes in the base of the microscope. Less intuitive, in fluorescence microscopy the light that forms the final image is generated and emitted by fluorochromes present in the sample and not by the light source located at the back-end of the fluorescence illuminator. In a functional epi-fluorescent microscope,
104
Ghiran
Fig. 7.4. a Diagram of the light paths (epi-fluorescence and trans-illumination) of a modern Olympus microscope. All the optical and mechanical components involved in fluorescence image formation are located above the stage. The light source can be either attached directly to the illuminator or be connected to the microscope through a liquid light guide in which case it will be away from the microscope body (see text). b Comparison between the apertures of a 10× and 40× objective at the imaging distance. While a 10× objective has a significantly larger frontal lens than a 40× objective, it is also located further away from the sample, which translates into a smaller actual imaging aperture. An objective with smaller aperture will be less efficient in gathering photons and thus will render images with less resolution.
none of the light that originates from the light source will ever reach the detector. Because a fluorescent structure is formed by numerous self-luminous points, image formation in fluorescence microscopy have, in some respects, more in common with images formed by an astronomical instrument than with images formed by a bright-field microscope. Each fluorescence image is generated by a large number of individual self-luminous points of light or, more correctly, disks of light (see above) that scatter light in all directions around them. Only a fraction of this scattered light will be captured by the objective lens. The amount of light gathered by a lens is directly proportional to the opening (aperture) of the objective at the imaging distance of the lens. This is important because although a 10 × 0.25 objective has a significantly larger frontal lens than a 40 × 0.95 lens, the angle subtended by the lens at the working distance is only about 15◦ for the 10× lens and around 71◦ for the 40× lens (Fig. 7.4b). This is possible because the working distance of the 10 × 0.25 objective is about 6 mm, whereas for the 40 × 0.95 it is about 0.14 mm. In other words, the larger the angle of the objective
Introduction to Fluorescence Microscopy
105
frontal lens at the imaging distance, the more light will enter the objective and the more photons will contribute to the image (see Section 4). 5.1. Light Sources or Illuminators
A light source suitable for fluorescence imaging has to deliver a beam of light that is homogenous (uniform output), high power, constant power output in time (no flickering) and contains all wavelengths that are needed to excite fluorochromes present in a given specimen. Today, most fluorescence microscopes are equipped with one of the three main types of illuminators: mercury, xenon or metal halide lamp. Lamps are made out of fused silica quartz round or elliptical envelopes that are strong enough to sustain high temperature and extreme pressures. Mercury lamps and xenon lamps are very common with most microscopes 5–10 years and older, with their lamp housing physically attached to the microscope (see Fig. 7.3b). Newer microscopes are fitted with metal halide illuminators or LED light sources separated from the microscope stand using liquid light guides (3–5 ft long) to deliver the light to the fluorescence illuminator. Mercury and xenon lamps deliver a very bright and continuous spectrum from UV to infrared (Fig. 7.5a and b). The power of the lamps varies usually between 50 and 200 W for mercury lamps and 75 and 150 W for xenon lamps. A closer inspection of the mercury bulb output graph shows that at certain wavelengths the emission is significantly greater than others, such as 365, 405, 436, 546 and 579 nm. Most of the fluorochromes that were synthesized in the past and are still in use today were centered on these wavelengths to maximize the efficiency of the mercury light source. Additionally, most of the objectives used for wide-field fluorescence or confocal microscopy were also designed to offer maximum correction for these particular wavelengths. Interestingly, the most used fluorochrome, fluorescein, has its excitation maxima at 494 nm where the mercury lamp has no major emission lines. However, due to its high quantum efficiency and broad excitation spectra, fluorescein is bright even when excited with a mercury lamp. Fluorescein has several disadvantages, such as fast photobleaching and high pH sensitivity, which makes it less suitable for fluorescence microscopy but more appropriate for FACS (fluorescence-activated cell sorting), where the interaction of the laser beam with the fluorescein is brief enough such that bleaching is not an issue. Due to its characteristic larger distance between electrodes, mercury bulbs fill up the back focal plane of the objective significantly more compared to xenon sources, allowing a uniform illumination of the specimen even at small magnifications. On the other hand, mercury lamps have short life spans (about 200 h) and exhibit a steady decrease of power output due to carbon deposits that begin to appear after the first hours of use.
106
Ghiran
Fig. 7.5. Spectral characteristics of various light sources. a Mercury lamp, b xenon lamp and c metal halide.
Introduction to Fluorescence Microscopy
107
Xenon lamps have a more homogenous output throughout the spectrum without any obvious emission peaks (Fig. 7.5b). The emission in the visual range is greater than that of the mercury lamp, which is an advantage in certain fluorescence applications or when using particular fluorochromes. However, the output in the red and infra-red part of the spectrum is also considerable (almost half of the entire emission spectrum); therefore, the heat generated by these sources is also significant and creates problems during long time-lapse experiments. The heat produced by the light source (xenon or mercury) is transmitted by the microscope frame to the stage and is reflected as an “x−y” drift of the image. To minimize this problem, certain companies (i.e. Olympus) manufacture microscope stands out of a metal– ceramic alloy that has a minimal thermal expansion coefficient and allows for long time experiments with a minimal heat-dependent drift. Another efficient approach to correct for this problem is to use a liquid light guide that allows a complete separation between the heat source (halogen or mercury/xenon lamp) and the microscope body. A common problem for most fluorescence light sources is the variability in the power output (seen as flickering of the excitation and emission light) due to either variation in the output of the power source or lack of stable connections between electrodes when the lamp is on. While correcting for the latter is more difficult and requires the use of stabilized power sources (more expensive), the former is easier to prevent (never to correct) by simply allowing a new bulb to “burn in” when turned on for the very first time for at least one hour. By doing so, the discharge arc will etch stable contact points (small pits) on the electrodes. On these two points (one on each electrode), the resistance will be lower than on any other area of electrodes. Therefore, the next time the lamp is turned on, the discharge arc will form quickly and stably between the two points generating a steady output of light. Subsequently, if the lamp is turned off before “burn in,” the next time the lamp is turned on (and for the remaining of the 200 h) the arc will “look” for areas with lower resistance on the electrodes, producing quick drops in output as it jumps from one point to another. These drops in output are seen by the user as “flickering” of the fluorescence. The main disadvantage of flickering is the inability to generate any quantitative data using fluorescence microscopy, especially during time-lapse experiments, as flickering of the lamp is completely random. Metal halide lamps are becoming more and more popular because they have all of the advantages of mercury lamps and few, if any, of their disadvantages. The emission spectra is almost identical but appreciably brighter, lamps last significantly longer (2000 has opposed to 200 h) and are almost flicker-free due to
108
Ghiran
a stabilized power supply. Most lamps are sold with a liquid light guide that allows the power supply assembly containing the lamp to be stored safely away from the microscope, preventing any heat transfer to the stand or stage. Also, using a liquid light guide takes away the laborious task of centering the lamp after each replacement and provides a uniform light beam that will fill the back focal plane of each objective. Disadvantages of liquid light guides are the high susceptibility to mechanical strains that renders them inoperable seconds after stress (bending) and to the formation of air bubbles inside the guide after prolonged use that results in non-homogenous illumination. Taping the guide on the wall to prevent any accidental handling and making sure that the liquid guide does not form any sharp angles (especially behind the power source) will prolong its use. Every 2000 h or so (each time the lamp is replaced), unplugging the guide from the fluorescent illuminator and projecting it on the wall is a good practice to check the integrity of the light guide. If the resulting circle of light (very bright white) is not uniformly illuminated, it may be time to purchase a replacement. Meanwhile, performing the flatfiled correction again for each objective is desirable (see below). Never point the liquid light guide toward a person when the lamp is on and never point it toward reflective surfaces, as it is impossible to predict where the light will end up. Always prevent people from entering the room while checking the integrity of the liquid light guide by displaying a clear sign on the door.
LED (light emitting diode) technology was adopted by the microscopy industry only relatively recently, as in the past the low power output of these light sources made them best suited mostly for machine vision and the auto industry. Today, it is estimated that due to advances in LED technology the brightness of these devices doubles every couple of years, which makes LEDs the most promising illumination technology for fluorescence microscopy. The output wavelengths of LEDs available today cover most if not all of the fluorochromes that are currently in use. While the full width at half maximum (see above) is not as narrow as the laser light, ultimately this is an advantage because the use of certain fluorochromes is no longer prevented by the mismatch between the emission of the light source and the excitation of the fluorochrome. There are several advantages of the LED that make this technology superior to current illumination methods: LEDs are small, compact and have low power usage with little heat generated during “on” times. LEDs also require no warm-up time, last for 10,000 h or more with outstanding
Introduction to Fluorescence Microscopy
109
stability (experiencing less than 1% decrease in the power output over their lifetime). Another notable advantage over traditional fluorescence sources is their ability to be turned on and off within milliseconds, circumventing the need for mechanical shutters, a known source of vibration that becomes a serious imaging problem especially during fast time-lapse experiments. The downside of LED technology is the low power output that is equivalent (and that is true only for certain wavelengths) to a 75 W xenon lamp. However, it is just a matter of time before the LED will become bright enough to compete and likely replace most standard mercury and metal halide fluorescence sources. 5.1.1. Operating Rules of Fluorescence Source Power Supplies
When turning on a fluorescence light source, there is a short (less than a couple of seconds) discharge of about 50,000 V that will create a powerful magnetic and electric field that potentially could destroy electronic components found nearby (computer-related equipment). Even though all of the cables and the power sources are (theoretically) shielded, the possibility remains. Therefore, in a laboratory with multiple users, it is always useful to post a warning regarding the sequence in which components should be tuned on and tag each piece of equipment next to its power switch with a label that specify the order and the total number of power switches that have to be on for all the equipment to work. This is necessary because older pieces of equipment have the on/off switches hidden on the side or on the backside and are easy to miss. Therefore labels should read: 1/7, 2/7 etc. In any case, as a rule, the first component that has to be turned on and the last to be turned off is the fluorescence light source.
5.2. Fluorescence Illuminator
Fluorescence illuminator or epi-illuminator classically consists of the lamp housing located at the opposite part of the nosepiece, the collector lens assembly and the intermediate part that contains slots for balancer, neutral density filters, field diaphragm and the aperture diaphragm (see Fig. 7.6). The collector lens assembly is located between the light source and the balancer. The lens system of the assembly (which is located near the light source) can be adjusted i.e. focused, to collect (hence the name of the lens system) most of the light emitted by the source and send it to the filter cube turret. A second lens system (or several on newer microscopes) focuses the light originating from the collector lens into and fills the back focal plane of the condenser (represented by the objective in epi-fluorescence microscopy). High optical quality of the collector lens assembly is a must especially as new devices that provide confocal-like images by creating virtual pinholes such as DSU (disk scanning unit, Olympus) or Zeiss (Apotome) are used with increasing frequency. These devices create images that are formed by using slits that focus light with different wavelengths on the sample. Therefore,
110
Ghiran
Fig. 7.6. Picture of an epi-fluorescence illuminator with its main parts labeled: CL, collector lens assembly; ND, neutral density filters; AD, aperture diaphragm – or sometimes called aperture sto-p; FD, field diaphragm; FC, filter cubes. Before any imaging session, one has to check the status of all these accessories.
apochromatic corrected illuminators are essential. Each time the lamp is replaced it must be centered and adjusted to meet the criteria for Koehler illumination. Most new microscopes have in the lamp house either a concave or parabolic mirror. Therefore, the image of the lamp will be double when imaged without an objective in the optical path. It is important to know that the purpose of centering the lamp is not to overlap the two images of the lamp (one direct and the other one mirrored) but rather to align them beside each other. Once this is accomplished, the image must be slightly defocused by adjusting the collector lamp to even out the illumination of the field of view. Although this step may seem unimportant, it is crucial when using fluorescence images for quantitative purposes. When samples are imaged with low-to-intermediate magnification objectives (10× to 60×) and recorded using medium-to-large format CCD cameras (1.2– 3/4 ), an uneven illumination will cause unequal excitation of fluorochromes in the sample, rendering different readouts from virtually identical fluorescent structures. To make the situation worse, because of the nature of fluorescence imaging where most of the field of view is black, even significant variations in illumination are not readily identifiable. Therefore, flat-field correction must be performed before any quantitative analysis for each objective separately and, of course, after each bulb replacement. This operation will correct not only for uneven illumination due to the light source but also for dust particles located on optical surfaces between the objective and the CCD chip. Therefore, in theory,
Introduction to Fluorescence Microscopy
111
each filter cube should be flat-field corrected individually. Even on a homogenous illuminated field, the fluorescence originating from the sample will be absorbed by the dust particles located on the optical path, and the resulting fluorescence will be lower. If one quantifies changes in fluorescence of moving cells for a period of time, an artifactual, biologically irrelevant, increase or decrease in fluorescence could be recorded when cells start in or cross such areas. Only images that are corrected in such a way are suitable for any kind of image analysis when quantitative information is required. If parameters such as length, width, area etc., are of interest, flat-field correction may not be critical, but in any case one should always be aware of it. If flat-field correction cannot be done for various reasons and quantitative analysis must be performed, then comparing images that were acquired using the very same area of the field of view could bypass the need for flat-field correction. This can be done by using an eyepiece reticule and acquire images that were framed in the same quadrant of the reticule. Nonetheless, results generated by this method are an approximation at best and should not be used routinely. To perform flat-field correction one needs to acquire three images: (1) the image of interest, (2) a dark frame and (3) a flat-field image. The first image (I ) will contain structures of interest plus any artifacts induced by imperfect illumination, dust and electronic noise. Importantly, the exposure time should be recorded and used for the flat-field image and the dark frame. The second image, the dark frame (Df ), has to be acquired with no light reaching the detector (Fig. 7.7a) and is needed to subtract the contribution of the electronic noise from both the flatfield image and from the image of interest. The third image, the flat-field (Ff ) (Fig. 7.7b) can be the image of a concentrated solution of a fluorescent dye (fluorescein) that lacks any structures. It may help to slightly defocus the image to obtain a homogenous field although it would be ideal to acquire all images at
Fig. 7.7. Surface representation of the dark frame and uncorrected uniformly fluorescent surface. a The noise levels are less than 5% of the maximum value. b Surface representation of a homogenous fluorescein solution showing significant variation in illumination throughout the recorded field.
112
Ghiran
the same focus. It is also recommended that the exposure time of the flat-field image be long enough so that at least half of the maximum value of the pixel is reached (around 2400 for a 12-bit CCD). Because the exposure time is dictated by the image of interest, using neutral density filters and changing the concentration of the fluorescein may be needed. Also, acquiring several images (10–15) that will be averaged in the end are needed to limit the influence of noise in the flat-field image. Although the image will look rather uniform, any image analysis program will show significant differences in intensity throughout the image due to imperfect illumination and dust on the optical components. Pseudo-coloring the image or displaying it in 3D-mode based on the pixel value would make the unevenness of the field very apparent (Fig. 7.7b). The exposure time should match that of the image of interest as the electronic noise increases with exposure time. Electronic noise is present in every image and has more or less a constant value, which is about 5% (a pixel value of 180– 220) of the maximum pixel value (4095 for a 12-bit camera). Scientific-grade CCD chips found in virtually all cameras suitable for fluorescence microscopy will display, when imaging a perfectly uniform bright surface, a range of values that are within 5% of the mean pixel intensity. Consequently, when acquiring a background image (no illumination), the mean, lowest and highest pixel value, will be almost the same and around 5% of the maximum pixel value (see the “Y” axis in Fig. 7.7b).3 Therefore, when acquiring fluorescence images, one should aim for obtaining a pixel value of structures of interest between 3600 and 3900 to increase the signal-to-noise ratio. Saturated pixels, having a value of 4095, are useless when the image is needed for quantitative analysis because the actual value of the pixel, whether it is 4095 or any other value above that, would still be displayed as 4095, making further analysis inaccurate. Once all three images are available, most if not all of the image analysis programs have built-in capabilities that will use the following formula (or a variation of it) to generate the flat-field corrected image, Ic(x,y) : Ic(x,y) = M (I(x,y) − Df(x,y) )/Ff(x,y) − Df(x,y)
3 This is true for a certain exposure range that depends directly on the temperature of the chip. The lower the temperatures of the CCD chip, the lower the noise levels during longer exposures times. In general, cameras used for astronomical imaging require deep cooling because it is not uncommon to expose an object for hours at the time. Luckily, biological applications that require exposure time longer than several minutes are rather rare. CCD cameras that are cooled at −40 to −65◦ C are needed for applications where the fluorescence associated with biological structures or phenomena are very dim and short-lived. In these instances, it is crucial to lower the noise levels as much as possible so that faint fluorescence signal would become apparent.
Introduction to Fluorescence Microscopy
113
This operation will subtract the value of each pixel of the dark frame (Df(x,y) ) from the corresponding pixel in the actual image (I(x,y) ) and flat-field image (Ff(x,y) ) and divide the results. (M) represents the mean value of the corrected flat-field frame. Because the value of pixels from the original image will be changed irreversibly, this operation should be performed on a duplicate of the original image. The illuminator also houses the fluorescence shutter, as well as several important and often overlooked devices, such as the heat and neutral density filters, the excitation balancer, the aperture and the field diaphragms. Improper settings of any of these accessories could result in low-resolution, low-contrast images that are independent of the sample and the optical components used to generate images. The heat filter is located between the collector lens and the optical component of the epi-illuminator of all microscopes (Fig. 7.6) and is used to block wavelengths above 600 nm that could potentially overheat and shorten the lifespan of the coated optical elements. If near-IR or IR imaging is used, it may be useful to remove the filter to increase the excitation efficiency of IR fluorescent probes. Also, most cameras used for microscopy have a IR cut-off filter in front of the CCD/CMOS that could also decrease the efficiency of near-IR or IR imaging. One should always check with the microscope/camera manufacturer before attempting to remove any built-in optical or mechanical components from the microscope or camera head. It is important to remember that for 10–15 s or longer after the bright-field source (likely a 100 W halogen lamp) is turned off, the bulb will continue to transmit infrared radiation (heat) that could potentially affect the CCD and fog the resulting image. If a brightfield method is used to locate the cell or tissue of interest and the fluorescence imaging requires long exposure times, the IRinduced fogging could become a problem. There are several ways to avoid image fogging: waiting for 20–30 s once the halogen lamp is turned off to image in fluorescence mode, using a heat filter or using a mechanical shutter in the bright-field path. A shutter may be a better choice because all incandescent sources turn-on and die-off slowly. Without a shutter during time-lapse acquisition of alternating fluorescence and bright-field images, such as phase contrast or DIC (see below for analyzer-induced artifacts), there will be a bleed-through of the bright-field light into the next fluorescence image, especially during fast acquisition. Having a mechanical shutter will completely circumvent the problem, although it may transmit significant vibrations in the microscope stand that would blur the next fluorescence image. Most software dedicated to image acquisition have the option to delay the step that follows the closing or opening of a mechanical shutter (either on fluorescence or bright-field path) or change in X−Y or
114
Ghiran
Z-axis of the microscope stage, to prevent blurring of the image due to mechanical vibrations. This may not be a viable option if fast acquisition is required. Having liquid light guides for both light sources (bright-field and fluorescence) completely circumvents this problem, as the mechanical shutters are located away from the microscope stand. The excitation balancer (EB) slider (Fig. 7.6) is usually found in the fluorescence illuminator between the collector lens assembly and the neutral density filters and is useful only when used with a double or triple filter cube in the turret. It functions as a variable excitation filter, shifting the excitation wavelengths by either changing the angle of a dichromatic mirror from 0 to 45◦ (system employed by Olympus) or changing between a long and a short-pass excitation filter (Nikon). The balancer is very useful when an image with two or more fluorochromes is analyzed and certain features are stained very brightly with one fluorochrome whereas others stained with another fluorochrome are very dim. By gradually decreasing the excitation of the bright fluorochrome and increasing the brightness of the dim fluorochrome, the topological relationship between the two structures can easily be recognized without the need to repeat the experiment to correct for the unbalanced signal strength. Also, colocalization (if any) between two colors is easier to record because there is no “pixel shift” artifact, a common occurrence when using two or more single band cubes (see below). The disadvantage is that in order to record images rendered by a balancer, a color camera is needed and most of the scientific-grade cameras use monochrome CCD chips. Using a color filter in front of the camera can easily fix this problem. The downside of having the balancer installed on a microscope that is shared among several users is that more often than not, after they are done using it, most users leave the balancer engaged in the optical path. So when the next user images a sample using a single band filter cube, a significantly dimmer image than normal (or sometime no image at all) is obtained for reasons that have nothing to do with the sample itself. Again, a good practice is to always check the position of all accessories that could be in the optical path before any imaging session. Neutral density (ND) interference filters are used to decrease the intensity (amplitude) of the excitation light (Fig. 7.6) by reducing the intensity uniformly across the spectrum (from 400 to 700 nm). In fluorescence microscopy, filters are used mainly to protect the sample from excessive bleaching and, when imaging live samples, to diminish the generation of reactive oxygen species that depend directly on the amount of UV illumination. Most modern fluorescence microscopes have either two neutral filters in the optical path or a wheel with several ND filters of increasing absorbency from 20 to 90%. Depending on the mechanism employed by the filter to attenuate the incoming light, there are
Introduction to Fluorescence Microscopy
115
two types of ND filters: absorbing ND filters that have a smoked appearance or reflecting ND filters that look like a mirror on the reflecting side (which must face the light source). The aperture diaphragm (Fig. 7.6, AD) can usually be centered on the X−Y axis and is located in a conjugated plane with the rear aperture of the objective and, therefore, controls the intensity of excitation without affecting the size of the area that is illuminated. Closing the aperture diaphragm will reduce the glare of bright fluorescent images, although in our experience, it is less useful than the iris diaphragms that are fitted on some high NA, immersion objectives. A semi-closed or closed aperture diaphragm will render dim images, and, because it does not affect the size of the field of view but only the amount of fluorescence light that illuminates the sample, the problem is not obvious from the beginning. The status of this diaphragm should always be checked before any imaging session. The field diaphragm (Fig. 7.6, FD) is also centerable and is located between the aperture diaphragm and the filter cubes and can be either round or rectangular. It provides very useful means to control the size of the illuminated area to prevent unnecessary bleaching of the sample and to increase the contrast in samples with high background. For maximum contrast, the illuminated field should be as large as the structured imaged even if it means closing the diaphragm down all the way. However, this can also be a means to block-out the excitation of structures deemed “non-representative”; therefore, it should be used appropriately. Filter cubes (Fig. 7.6, FC) are located above the objective in either a manual or motorized turret (newer microscopes). Older microscopes used separate sliders for excitation and emission filters. Very few, if any, such microscopes are found today in research laboratories and, therefore, we will not discuss them any further. A fluorescence filter cube has three main components: an excitation filter, a dichromatic filter (dichroic mirror or beam splitter) and an emission filter (Fig. 7.8). Please note that arrows on the emission and excitation filters point in different directions depending on the manufacturer of filters. There are three main types of interference filters (Fig. 7.9) that are classified according to the range of wavelengths they allow to pass. The blocking range is usually not mentioned in the name of a filter. The three types are short pass (SP) filters that allow light with short wavelengths to pass while blocking everything else, long pass (LP) filters that allow light with long wavelengths to pass while blocking short wavelengths and band-pass (BP) filters. Band-pass filters allow only a range of wavelengths from the spectrum (a band or a window) for which the filter is perfectly transparent. Figure 7.9 shows the measured transmission profiles of the three main filters that are characterized by center wavelength (CWL, band filters), peak (P %T) and average
116
Ghiran
Fig. 7.8. The basic configuration of a fluorescent filter. Parts are shown before assembly for a better representation of the relationships between various components. Note the angled emission filter. The orientation of filters as shown by arrows on the emission and excitation filter are not universal. See text for details.
Fig. 7.9. Representation of the three simple filters used for fluorescence microscopy and their defining parameters. Long pass (LP) and short pass (SP) filters are defined by their cut-on and cut-off wavelengths. The bandpass filters (Bandpass) are defined by their peak transmission, center wavelengths (CWL) and full-width half maximum (FWHM) (Image courtesy of Chroma Inc.).
transmission (Avg. %T), or for bandpass filters, the full-width half maximum (FWHM), which is a measure of the filter slope (the steeper the better). Combinations of these three filter types are found in virtually all filter cubes. Figure 7.10 shows the diagrams of a high-quality simple-band DAPI filter. The novel methods employed to manufacture filters have reached almost the lowest limit for auto-fluorescence of the glass and the highest limit for
Introduction to Fluorescence Microscopy
117
Fig. 7.10. Diagram showing a high-quality single-band DAPI filter. Note the almost rectangular bandpass of the excitation and emission filters that have the 50% cut-on and cut-off values almost identical with the blocking range. The dichromatic mirror has an almost vertical slope and close to 0% transmittance for the excitation range and over 90% transmittance for the emission range (Diagram courtesy of Semrock, Inc.).
transmittance, barrier and slope of filters. Note the almost perfect wavelengths selection of the interference filters that ensures bright signals and black background. Excitation filters: In the past, selecting the desired wavelengths was realized by using colored solutions of specific dyes that were placed in glass containers in the light path. Colored gelatin sheets and colored glass filters later replaced containers. These approaches for selecting the excitation wavelengths worked well with trans-fluorescence methods where samples were illuminated through an immersion dark field condenser, which due to its specific design prevented most of the high-energy excitation light to enter the objective and allowed only the emitted fluorescence to form the image. While these filters were very affordable, they suffered from poor optical performance, low transmittance and mediocre wavelength discrimination. A significant problem was the high level of auto-fluorescence of dyes that were used in these filters, which considerably decreased the signal-to-noise ratio of most samples. In recent years, the majority of filters used in fluorescence microscopy are interference filters. The main technologies used today for constructing an interference filter use either soft or hard coating. Soft coating is based on deposition on several optically flat pieces of glass glued together by adhesive of multiple thin films of several different dielectric substances (metal
118
Ghiran
salts). The alternative, hard coating method, deposits dielectric layers on a glass disc using a special technology (ion-beam sputtering) that creates several layers on just one optically flat piece of glass avoiding the use of any adhesives, which are responsible for most of the auto-fluorescence of the soft-coated filters. Aside from being more resistant to light damage and moisture, the later filters are easier to manufacture with perfectly flat surfaces, a paramount characteristic when co-localization studies between two, three or more colors are performed. 5.2.1. Registration Artifacts When Using Two or More Filters
In Fig. 7.11a is shown the overlap image of a triple-labeled 10 μm-bead imaged by two microscopes, one using four standard DAPI, FITC, TRITC and triple-band filters and the other using filters of the same type but of “no image-shift” quality. In the first image, all images are shifted in respect with the optical axis creating the false impression of no or partial co-localization, although all images originate from the same structure. The “no image-shift” grade filters do not suffer from such an effect (Fig. 7.11c). To better represent the lack of co-localization, images in the left were processed using the Sorbell filter (ImageProPlus, Media Cybernetics) to show only the outline of the imaged bead (Fig. 7.11b). The origin of the problem can be the “wedge” shape of the emission filter but also the “wedge” shape of the dichromatic filter (dichroic mirror) that would create the same effect regardless of the quality of the emission filter. In some situations both filters suffer from the same defect. Therefore, it is not indicated to “mix and match” dichromatic and emission filters of different qualities if co-localization studies are needed. If standard filters are the only ones available, co-localization studies still can be accurately performed if the amount of shifting for two channels is calculated, assuming the third one to be perfectly aligned with the optical axis of the microscope and a back-shift
Fig. 7.11. Filter cube-induced artifact during co-localization. A triple-labeled 10 μm latex bead was imaged either through regular (a) or co-localization-quality (c) filter cubes. The white image of the bead (a, left) was recorded using a triple filter. The raw (a) and the processed image (b) show significant misalignment among filters, although the same structure contained all fluorochromes and there was no shift of the bead during acquisition. Co-localization-quality (c) filter cubes rendered a perfect overlapping image. The difference in bead size between the two images is due to the different pixel size of cameras used for the acquisition.
Introduction to Fluorescence Microscopy
119
correction is applied to the first two channels. All major image acquisition and processing pieces of software, as well as several freeware programs, have built-in modules for this option. Having triple-labeled beads on the slide along with cells of interest during acquisition can considerably simplify the task. The smaller the beads, the more precise the correction. If the amount of shifting between channels changes overtime, it could indicate a loose component (likely the dichroic mirror) in one of the filters. Types of interference filters: Depending on which range of wavelengths a filter allows to pass, there can be band-pass filters that allow only certain range of wavelengths to pass and can be either narrow or broad band-pass, short pass edge filters that allow only short wavelengths to pass while blocking long wavelengths or long pass edge filters that block short wavelengths and only allow long wavelengths to pass (see Fig. 7.8). Understanding the properties of a filter and the way it does or does not match the excitation and emission spectra of a certain fluorochrome is essential when buying a filter or, more importantly, when assembling a new filter cube from components available in the laboratory. Most filters must be mounted in a certain way and the direction is usually shown with an arrow that has to point to the dichromatic mirror. Also, it is important to know that most filters (so called soft-coated) are considered consumables with a limited lifespan that, depending on how much the microscope is used, have to be replaced every 1–2 years, as the transmittance will decrease below 20–30% (so called “burnout” filters). Figure 7.12a shows the picture of a soft-coated filter 9 months after installation followed by average to high usage. Figure 7.12b shows in detail the moisture damage centered on impurities found between the thin layers of coating. Using the microscope in areas with high humidity also favors the growth of fungus on all optical surfaces
Fig. 7.12. Damage of a soft-coated emission filter after less than a year of moderate to heavy use. a Image on the left shows the damage that started at the periphery (discolored areas) and then progressively moved toward the center. The first indication of the damage was the out-of-focus round shapes found mostly at the edges of the field of view noticed first during bright-field imaging. b Image on the right shows a low magnification detail of the affected area (see text for details).
120
Ghiran
(including filters) that drastically alters their optical properties. A visual inspection would show structures that look like hairs or vines branching from a center if there is fungal growth present or round darker areas if there is any delamination due to heat or impurities in the filter coat. All these problems could eventually lead to photo-darkening. Gradual decreases in light efficiency while imaging in fluorescence and the appearance of out-of-focus patterns when the filter is used in bright-field are signs that the filter has to be changed. It may be wise investing in hard-coated filters that, although are more expensive than the regular filters, do not “burn-out.” However, they are not impervious to moisturerelated damage (fungal growth). Most microscope manufactures use an antifungal treatment of all optical components (lenses and prisms) to slow down the growth of fungus that ultimately would etch the glass surface underneath rendering it useless. Dichromatic beam-splitters (dichroic mirrors) are special long or band-pass filters that are mounted at 45◦ degrees in the filter cube. Any variation of this angle will significantly change the intended optical properties of the filter. This property is used to change the range of wavelengths that will pass through the filter (see above, excitation balancers). Once the excitation filter has selected a certain range of wavelengths, the dichromatic filter performs like a mirror for the incoming shorter wavelengths, characterized by a percent reflection (close to 100% for certain wavelength range), reflecting the light through the objective into the sample. The fluorescence that originates from the sample has a longer wavelength than the excitation light (Stokes shift, see above) and, therefore, the dichromatic mirror allows the light to pass un-reflected (therefore characterized by percent transmission close to 100% for a certain wavelength range). Depending on the type of the filter cube that it is used, some dichromatic filters can be long pass filters (single excitation filter cubes) or band-pass filters (double and triple filter cubes). If the dichromatic mirror surfaces are angled (edge-shaped) instead of being parallel, the image formed by that particular filter cube will be shifted from the optical axis of the instrument. The same is true for the emission filters (see above). Figure 7.13 shows the diagram of a triple-band filters that allows simultaneous and specific excitation and emission of several (up to three) fluorochromes. Note that in a triple-band filter the dichromatic mirror and both the emission and excitation filters are band-pass type filters. Emission filters can be long pass (allows light with longer wavelength to pass), increasing the overall brightness of the signal but also increasing the background or band-pass type, in which case the brightness decreases slightly but the background is darker. They are easy to identify in a filter cube because they are mounted under a 5◦ angle to decrease the brightness of the background and prevent back-reflection.
Introduction to Fluorescence Microscopy
121
Fig. 7.13. Diagram of a high-quality triple-band-pass filter. The filter is designed to allow simultaneous imaging of up to three fluorochromes. Note that all the filters found in this cube are band-pass filters (Diagram courtesy of Semrock, Inc.).
5.2.2. Orientation of the Fluorescence Filters in a Filter Cube
The direction a filter has to be mounted in a filter cube is usually indicated with an arrow, but different companies use the arrow to point to different parts of the filter cube. In most cases, the arrow on the excitation filter points toward the dichromatic mirror (center of the cube), whereas the arrow on the emission filter points toward the dichromatic filter (Chroma) or towards eyepieces (Semrock). The situation is more complicated for the dichromatic filter, which has a coated surface that always has to face the light source for optimal performance. Because the dichromatic mirror is very thin, there are no arrows to show the proper orientation of the mirror. One can find the coated surface if one uses a light source (a desk lamp works well) to reflect it off from the dichromatic mirror; if the coated surface is pointing up, there will be only one image of the light source; if the coated surface points down, the image of the light source will be doubled and slightly shifted. If the performance of a filter set is less than expected, especially if the filter was custom built from preexisting parts in the laboratory, it is always a good idea to check the orientation of filters and consult the manufacturer.
5.2.3. Methods to Prevent Stray Light from Exiting the Filter Cube
Because there are no perfect filters, a certain percentage of the light that traverses through the excitation filter, instead of being redirected by the dichromatic mirror into the objective, passes through it and bounces off the back wall of the cube and then is reflected by the dichromatic mirror though the emission filter into the eyepiece or detector, increasing the brightness of the background and decreasing the dynamic range of the actual
122
Ghiran
image. Various methods have been devised over time to prevent this from happening, starting with coating the interior surface of the filter cube with a black, non-reflective paint (black baffled) or mounting the emission filter under an angle of 5◦ . In addition, increasing the efficiency of the filter also prevents reflection of the back-reflected light by the telan (tube) lens. Some microscope manufactures, in addition to an angled emission filter, use a slightly modified filter cube that contains either an absorbent neutral density filter on the opposite side of the excitation filter or an opening (Fig. 7.14) that allows stray photons to completely escape the filter cube (“Light-trap” first employed by Zeiss and Background terminator, employed by Nikon). The next component in the light path is the objective, which in all epi-illumination techniques plays the role of the condenser.
Fig. 7.14. Comparison between a standard fluorescence filter cube and light-trap (Zeiss). a In a standard fluorescence filter cube, the light that passes through the dichromatic filter (thin green arrows) is reflected between the back wall of the filter cube and the upper surface of the dichromatic filter, and crosses the emission filter increasing the background level of the final fluorescence image. b The light-trap filter cube (1) has the back wall removed so after passing through the dichromatic mirror (2), the stray light (3) is directed by the tapered wall of the fluorescence illuminator (4) outside the cube (5) thus improving significantly the signal-to-noise ratio (Diagram courtesy of Carl Zeiss MicroImaging, Inc.).
5.3. Objectives
Maybe the most important and specialized optical component of a microscope is the objective, which points to and forms the image of the examined object. In epi-fluorescence microscopy, the objective is also the condenser in the optical path, focusing the excitation light into the specimen. There is no other microscope component with a greater role in the image formation than the objective. In the diagram shown in Fig. 7.15, a cross-section through a microscope objective illustrates the mechanical and optical complexity of a fully corrected plan-apochromatic objective that can contain between 13 and 15 lenses. Therefore, the quality (or the lack thereof) of the objective will be crucial in determining the amount of information in the final image. That is not to say that a low-quality filter cube or an imaging device that is unsuitable for a certain application (due
Introduction to Fluorescence Microscopy
123
Fig. 7.15. Cross-section through a high-magnification planapochromatic objective. The image shows the mechanical and optical complexity of high-quality objectives that can contain up to 15 carefully polished and aligned lenses. Any mechanical stress (due to hitting or dropping of the objective) could potentially shift lenses from the optical axis of the objective significantly decreasing its optical qualities (Image courtesy of Leica Microsystems).
to speed, sensitivity, dynamic range or resolution) will not have a negative impact on the final image. However, irrespective of the quality of the other components involved in the image formation or acquisition, if the objective does not have the required NA or the appropriate chromatic and spherical aberration corrections for a particular application, no amount of correction or postacquisition processing will add the information that was missed by the objective. Therefore, for demanding specialized applications (calcium imaging etc.), building a new system starts with the objective followed by the camera, filters etc. If the resolution required for an application is known, for instance 0.25 μm, it is always a good idea to buy only that particular objective of the highest quality (such as PlanApo 60 × 1.40, 1.42 or 1.49) and the rest of the objectives (4×, 10×, 20× and maybe 100×) of standard quality (planacromats or planfluorite, see below). The premium one payments for fully corrected lenses are in the range of thousands of dollars and the money could be better used towards other components (camera, filters, image acquisition and
124
Ghiran
analysis software). If for most of the bright-field applications the type of objective (correction-wise) is not critical, for fluorescence microscopy, and especially for application such as co-localization of two or more colors in three dimensions (3D), the choice of objectives is paramount, due to the impact of the chromatic and spherical aberration on image formation (Fig. 7.16).
Fig. 7.16. Diagram showing the two main aberrations of microscope objectives. a Spherical aberrations are due to fact that light passing through the edge of a lens is refracted closer to the lens than the light going through the center. Therefore, a simple lens will not have a single focus but rather the focus will be present on a certain distance (area) along the optical axis of the lens between the two extremes formed by the periphery and central rays. The place along this area where focus is the “best” is called circle of least confusion and it was first described for astronomical refractors and photographic cameras. Its size depends on the focal distance of the lens and its diameter. b Chromatic aberrations are a direct consequence of the variability of the refractive index of a transparent material with the wavelength. Therefore, blue light will be refracted more (closer to the lens) and red light less (farther from the lens) inducing a halo around the imaged objects that changes color with focus.
There are several major optical aberrations that affect, in various degrees, all microscope objectives due to the geometry of lens and the wide variation of the refractive index of a particular media (glass or water) with wavelength and temperature: spherical aberration, chromatic aberration, astigmatism, field curvature and coma. We will address here only the first two as they are directly relevant for fluorescence imaging. Spherical aberrations are due to the fact that the thickness of a lens varies significantly between the center and the periphery. As a consequence, the light is focused in different points along the optical axis depending on the place where the incident light entered the lens. The light that is closer to the optical axis will pass through the lens with minimal refraction focusing farther from the lens, whereas the light entering the lens near the periphery will undergo a more significant change in a direction focusing closer to the lens (Fig. 7.16). The larger the frontal lens (aperture), the worst the aberrations. Images that result from an insufficiently spherically corrected objective are soft, lacking contrast and sharpness. It is important to remember that corrections that are built into any objective are calculated for specific conditions:
Introduction to Fluorescence Microscopy
125
certain refractive index and temperature of the media between the lens and the coverslip, specific distance between the top surface of the coverslip and the imaged objects, and the presence or absence of the coverslip. Any departures from such conditions would significantly degrade the final image. Efforts to decrease spherical aberration