FLOW MEASUREMENT HANDBOOK Flow Measurement Handbook is an information-packed reference for engineers on flow-measuring techniques and instruments. Striking a balance between laboratory ideal and the realities of field experience, it provides a wealth of practical advice on the design, operation, and performance of a broad range of flowmeters. The book begins with a brief review of essentials of accuracy and flow, how to select a flowmeter, and various calibration methods. Following this, each chapter is devoted to a class of flowmeter and includes detailed information on design, application, installation, calibration, operation, advantages, and disadvantages. Among the flowmeters discussed are orifice plates, venturi meters, standard nozzles, critical flow venturi nozzles, variable area and other devices depending on momentum of the flow, volumetric flowmeters such as positive displacement, turbine, vortex shedding, swirl, fluidic, electromagnetic and ultrasonic meters, and mass flowmeters including thermal and Coriolis. More than 80 different types and 250 applications are listed in the index. There are also chapters covering probes, a brief introduction to modern control, and manufacturing implications. For those readers who want more background information, many chapters conclude with an appendix on the mathematical theory behind the techniques discussed. The final chapter takes a look at directions in which the technology is likely to go in the future. Engineers will use this practical handbook to solve problems in flowmeter design and application and to improve performance. Roger C. Baker is a Visiting Industrial Fellow in the Manufacturing and Management Division of the Department of Engineering, University of Cambridge; Visiting Professor, Cranfield University; and Director of Technical Programmes, the Gatsby Charitable Foundation.
Flow Measurement Handbook INDUSTRIAL DESIGNS, OPERATING PRINCIPLES, PERFORMANCE, AND APPLICATIONS
ROGER C. BAKER
CAMBRIDGE
UNIVERSITY PRESS
CAMBRIDGE UNIVERSITY PRESS Cambridge, New York, Melbourne, Madrid, Cape Town, Singapore, Sao Paulo Cambridge University Press The Edinburgh Building, Cambridge CB2 2RU, UK Published in the United States of America by Cambridge University Press, New York www. Cambridge. org Information on this title: www.cambridge.org/9780521480109 © Cambridge University Press 2000 This book is in copyright. Subject to statutory exception and to the provisions of relevant collective licensing agreements, no reproduction of any part may take place without the written permission of Cambridge University Press. First published 2000 This digitally printed first paperback version 2005 A catalogue recordfor this publication is available from the British Library Library of Congress Cataloguing in Publication data Baker, R. C. Flow measurement handbook : industrial designs, operating principles, performance, and applications / Roger C. Baker. p. cm. Includes bibliographical references. ISBN 0-521-48010-8 1. Flow meters - Handbooks, manuals, etc. I. Title. TA357.5.M43B35 2000 681'.28-dc21 99-14190 CIP ISBN-13 978-0-521-48010-9 hardback ISBN-10 0-521-48010-8 hardback ISBN-13 978-0-521-01765-7 paperback ISBN-10 0-521-01765-3 paperback DISCLAIMER Every effort has been made in preparing this book to provide accurate and up-to-date data and information that is in accord with accepted standards and practice at the time of publication and has been included in good faith. Nevertheless, the author, editors, and publisher can make no warranties that the data and information contained herein is totally free from error, not least because industrial design and performance is constantly changing through research, development, and regulation. Data, discussion, and conclusions developed by the author are for information only and are not intended for use without independent substantiating investigation on the part of the potential users. The author, editors, and publisher therefore disclaim all liability or responsibility for direct or consequential damages resulting from the use of data, designs, or constructions based on any of the information supplied or materials described in this book. Readers are strongly advised to pay careful attention to information provided by the manufacturer of any equipment that they plan to use and should refer to the most recent standards documents relating to their application. The author, editors, and publisher wish to point out that the inclusion or omission of a particular device, design, application, or other material in no way implies anything about its performance with respect to other devices, etc.
To Liz, Sarah and Paul, Mark, John and Rachel
Contents
Preface
page xix
Acknowledgments Nomenclature CHAPTER 1
1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8
Introduction
Initial Considerations Do We Need a Flowmeter? How Accurate? A Brief Review of the Evaluation of Standard Uncertainty Sensitivity Coefficients What Is a Flowmeter? Chapter Conclusions (for those who plan to skip the mathematics!) Mathematical Postscript i.A Statistics of Flow Measurement Introduction The Normal Distribution The Student t Distribution Practical Application of Confidence Level Types of Error Combination of Uncertainties Uncertainty Range Bars, Transfer Standards, and Youden Analysis
APPENDIX
l.A.l 1.A.2 1.A.3 1.A.4 I.A.5 1.A.6 I.A.7
xxi xxiii l
1 2 4 7 9 9 13 15 15 15 16 17 19 20 21 21
CHAPTER 2 Fluid Mechanics Essentials
24
2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10
24 24 24 27 30 32 34 36 38 39
Introduction Essential Property Values Flow in a Circular Cross-Section Pipe Flow Straighteners and Conditioners Essential Equations Unsteady Flow and Pulsation Compressible Flow Multiphase Flow Cavitation, Humidity, Droplets, and Particles Gas Entrapment
CONTENTS
2.11 2.12
Steam Chapter Conclusions
CHAPTER 3 Specification, Selection, and Audit 3.1 Introduction 3.2 Specifying the Application 3.3 Notes on the Specification Form 3.4 Flowmeter Selection Summary Tables 3.5 Other Guides to Selection and Specific Applications 3.6 Draft Questionnaire for Flowmeter Audit 3.7 Final Comments APPENDIX 3.A Specification and Audit Questionnaires 3.A.1 Specification Questionnaire 3.A.2 Supplementary Audit Questionnaire CHAPTER 4
4.1
4.2 4.3
4.4
4.5 4.6 4.7 4.8 4.9
Introduction 4.1.1 Calibration Considerations 4.1.2 Typical Calibration Laboratory Facilities 4.1.3 Calibration from the Manufacturer's Viewpoint Approaches to Calibration Liquid Calibration Facilities 4.3.1 Flying Start and Stop 4.3.2 Standing Start and Stop 4.3.3 Large Pipe Provers 4.3.4 Compact Provers Gas Calibration Facilities 4.4.1 Volumetric Measurement 4.4.2 Mass Measurement 4.4.3 Gas/Liquid Displacement 4.4.4 pvT Method 4.4.5 Critical Nozzles 4.4.6 Soap Film Burette Method Transfer Standards and Master Meters In Situ Calibration Calibration Uncertainty Traceability and Accuracy of Calibration Facilities Chapter Conclusions
CHAPTER
5.1 5.2 5.3 5.4 5.5
Calibration
5 Orifice Plate Meters
Introduction Essential Background Equations Design Details Installation Constraints Other Orifice Plates
39 41 42 42 42 43 46 53 55 55 56 56 58 61
61 61 64 65 66 69 69 72 74 74 77 77 79 80 80 81 81 82 84 91 92 93 95 95 97 100 102 106
CONTENTS
5.6 5.7 5.8 5.9 5.10 5.11 5.12 5.13 5.14 5.15 5.16 5.17
Deflection of Orifice Plate at High Pressure Effect of Pulsation Effects of More Than One Flow Component Accuracy Under Normal Operation Industrially Constructed Designs Pressure Connections Pressure Measurement Temperature and Density Measurement Flow Computers Detailed Studies of Flow Through the Orifice Plate, Both Experimental and Computational Application, Advantages, and Disadvantages Chapter Conclusions
106 109 113 117 118 119 122 124 124
APPENDIX 5.A Orifice Discharge Coefficient
128
CHAPTER
6.1 6.2 6.3 6.4 6.5 6.6 6.7
Introduction Essential Background Equations Design Details Commercially Available Devices Installation Effects Applications, Advantages, and Disadvantages Chapter Conclusions
CHAPTER
7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 7.10 7.11
6 Venturi Meter and Standard Nozzles
7 Critical Flow Venturi Nozzle
Introduction Design Details of a Practical Flowmeter Installation Practical Equations Discharge Coefficient C Critical Flow Function C* Design Considerations Measurement Uncertainty Example Industrial and Other Experience Advantages, Disadvantages, and Applications Chapter Conclusions
124 127 127
130 130 131 134 135 135 137 138 140 140 141 143 145 146 147 148 149 151 152 152
CHAPTER 8 Other Momentum-Sensing Meters
153
8.1 8.2
153 153 154 154 155 156 157
Introduction Variable Area Meter 8.2.1 Operating Principle and Background 8.2.2 Design Variations 8.2.3 Remote Readout Methods 8.2.4 Design Features 8.2.5 Calibration and Sources of Error
CONTENTS
8.3 8.4 8.5 8.6 8.7 8.8 8.9 8.10 8.11 8.12 8.13 8.14
8.2.6 Installation 8.2.7 Unsteady and Pulsating Flows 8.2.8 Industrial Types, Ranges, and Performance 8.2.9 Computational Analysis of the Variable Area Flowmeter 8.2.10 Applications Spring-Loaded Diaphragm (Variable Area) Meters Target (Drag Plate) Meter Integral Orifice Meters Dall Tubes and Devices that Approximate to Venturis and Nozzles Wedge and V-Cone Designs Differential Devices with a Flow Measurement Mechanism in the Bypass Slotted Orifice Plate Pipework Features - Inlets Pipework Features - Bend or Elbow Used as a Meter Averaging Pitot Laminar or Viscous Flowmeters Chapter Conclusions APPENDIX 8.A History, Equations, and Accuracy Classes for the VA Meter 8.A.1 Some History 8.A.2 Equations 8.A.3 Accuracy Classes
157 158 158 159 159 159 162 163 163 165 167 168 168 169 170 173 176 177 177 178 180
CHAPTER 9 Positive Displacement Flowmeters
182
9.1
182 182 183 184 184 184 185 185 187 189 190 190
9.2
9.3
Introduction 9.1.1 Background 9.1.2 Qualitative Description of Operation Principal Designs of Liquid Meters 9.2.1 Nutating Disk Meter 9.2.2 Oscillating Circular Piston Meter 9.2.3 Multirotor Meters 9.2.4 Oval Gear Meter 9.2.5 Sliding Vane Meters 9.2.6 Helical Rotor Meter 9.2.7 Reciprocating Piston Meters 9.2.8 Precision Gear Flowmeters Calibration, Environmental Compensation, and Other Factors Relating to the Accuracy of Liquid Flowmeters 9.3.1 Calibration Systems 9.3.2 Clearances 9.3.3 Leakage Through the Clearance Gap Between Vane and Wall 9.3.4 Slippage Tests 9.3.5 The Effects of Temperature and Pressure Changes 9.3.6 The Effects of Gas in Solution
191 192 194 194 196 197 197
CONTENTS
9.4 9.5
9.6 9.7 9.8 9.9
Accuracy and Calibration Principal Designs of Gas Meters 9.5.1 Wet Gas Meter 9.5.2 Diaphragm Meter 9.5.3 Rotary Positive Displacement Gas Meter Positive Displacement Meters for Multiphase Flows Meter Using Liquid Plugs to Measure Low Flows Applications, Advantages, and Disadvantages Chapter Conclusions
198 199 199 200 202 203 205 205 206
APPENDIX 9.A Theory for a Sliding Vane Meter 9.A.I Flowmeter Equation 9.A.2 Expansion of the Flowmeter Due to Temperature 9.A.3 Pressure Effects 9.A.4 Meter Orientation 9.A.5 Analysis of Calibrators 9.A.6 Application of Equations to a Typical Meter
207 207 209 210 210 211 213
10 Turbine and Related Flowmeters
21^
Introduction 10.1.1 Background 10.1.2 Qualitative Description of Operation 10.1.3 Basic Theory Precision Liquid Meters 10.2.1 Principal Design Components 10.2.2 Bearing Design Materials 10.2.3 Strainers 10.2.4 Materials 10.2.5 Size Ranges 10.2.6 Other Mechanical Design Features 10.2.7 Cavitation 10.2.8 Sensor Design and Performance 10.2.9 Characteristics 10.2.10 Accuracy 10.2.11 Installation 10.2.12 Maintenance 10.2.13 Viscosity, Temperature, and Pressure 10.2.14 Unsteady Flow 10.2.15 Multiphase Flow 10.2.16 Signal Processing 10.2.17 Applications 10.2.18 Advantages and Disadvantages Precision Gas Meters 10.3.1 Principal Design Components 10.3.2 Bearing Design 10.3.3 Materials 10.3.4 Size Range 10.3.5 Accuracy
215 215 215 216 221 221 223 224 224 225 225 226 227 228 228 229 231 232 232 232 233 233 234 234 234 235 236 236 236
CHAPTER
10.1
10.2
10.3
CONTENTS
10.4
10.5
10.6
10.3.6 Installation 10.3.7 Sensing 10.3.8 Unsteady Flow 10.3.9 Applications 10.3.10 Advantages and Disadvantages Water Meters 10.4.1 Principal Design Components 10.4.2 Bearing Design 10.4.3 Materials 10.4.4 Size Range 10.4.5 Sensing 10.4.6 Characteristics and Accuracy 10.4.7 Installation 10.4.8 Special Designs Other Propeller and Turbine Meters 10.5.1 Quantum Dynamics Flowmeter 10.5.2 Pelton Wheel Flowmeters 10.5.3 Bearingless Flowmeter 10.5.4 Vane-Type Flowmeters Chapter Conclusions 10.A Turbine Flowmeter Theory Derivation of Turbine Flowmeter Torque Equations Transient Analysis of Gas Turbine Flowmeter
APPENDIX
10.A.1 10.A.2
237 238 238 240 241 241 241 242 243 243 243 243 244 244 244 244 244 245
245 245 246 246 251
CHAPTER 11 1Cortex-Shedding, Swirl, and Fluidic Flowmeters
253
11.1 11.2 11.3
253 253 254 255 257 259 260 263 264 264 264 267 267
11.4
Introduction Vortex Shedding Industrial Developments of Vortex-Shedding Flowmeters 11.3.1 Experimental Evidence of Performance 11.3.2 Bluff Body Shape 11.3.3 Standardization of Bluff Body Shape 11.3.4 Sensing Options 11.3.5 Cross Correlation and Signal Interrogation Methods 11.3.6 Other Aspects Relating to Design and Manufacture 11.3.7 Accuracy 11.3.8 Installation Effects 11.3.9 Effect of Pulsation and Pipeline Vibration 11.3.10 Two-Phase Flows 11.3.11 Size and Performance Ranges and Materials in Industrial Designs 11.3.12 Computation of Flow Around Bluff Bodies 11.3.13 Applications, Advantages, and Disadvantages 11.3.14 Future Developments Swirl Meter - Industrial Design 11.4.1 Design and Operation 11.4.2 Accuracv and Ranges
268 269 270
271 272 272 273
CONTENTS
11.5
11.6 11.7
11.4.3 Materials 11.4.4 Installation Effects 11.4.5 Applications, Advantages, and Disadvantages Fluidic Flowmeter 11.5.1 Design 11.5.2 Accuracy 11.5.3 Installation Effects 11.5.4 Applications, Advantages, and Disadvantages Other Proposed Designs Chapter Conclusions li.A Vortex-Shedding Frequency Vortex Shedding from Cylinders Order of Magnitude Calculation of Shedding Frequency
APPENDIX
11.A.I 11.A.2
273 273 273 274 274 275 276 276 276 276 278 278 279
CHAPTER 12 Electromagnetic Flowmeters
282
12.1 12.2 12.3 12.4
Introduction Operating Principle Limitations of the Theory Design Details 12.4.1 Sensor or Primary Element 12.4.2 Transmitter or Secondary Element Calibration and Operation Industrial and Other Designs Installation Constraints - Environmental 12.7.1 Surrounding Pipe 12.7.2 Temperature and Pressure Installation Constraints - Flow Profile Caused by Upstream Pipework 12.8.1 Introduction 12.8.2 Theoretical Comparison of Meter Performance Due to Upstream Flow Distortion 12.8.3 Experimental Comparison of Meter Performance Due to Upstream Flow Distortion 12.8.4 Conclusions on Installation Requirements Installation Constraints - Fluid Effects 12.9.1 Slurries 12.9.2 Change of Fluid 12.9.3 Nonuniform Conductivity Multiphase Flow Accuracy Under Normal Operation Applications, Advantages, and Disadvantages 12.12.1 Applications 12.12.2 Advantages 12.12.3 Disadvantages Chapter Conclusions
282 282 284 286 286 289 292 293 295 296 296 297 297
APPENDIX 12.A Brief Review of Theory 12. A.I Introduction
305 305
12.5 12.6 12.7
12.8
12.9
12.10 12.11 12.12
12.13
297 298 299 300 300 300 300 301 301 302 302 303 303 304
CONTENTS
12.A.2 12.A.3 12.A.4 12.A.5 12.A.6 12.A.7
Electric Potential Theory Development of the Weight Vector Theory Rectilinear Weight Function Axisymmetric Weight Function Performance Prediction Further Extensions to the Theory
307 307 308 310 310 311
CHAPTER 13 Ultrasonic Flowmeters
312
13.1 13.2
312 315 315
13.3 13.4 13.5 13.6
13.7
13.8
13.9 13.10
13.11
13.12 13.13
Introduction Transit-Time Flowmeters 13.2.1 Simple Explanation 13.2.2 Flowmeter Equation and the Measurement of Sound Speed 13.2.3 Effect of Flow Profile and Use of Multiple Paths Transducers Size Ranges and Limitations Signal Processing and Transmission Accuracy 13.6.1 Reported Accuracy - Liquids 13.6.2 Reported Accuracy - Gases 13.6.3 Manufacturers' Accuracy Claims 13.6.4 Special Considerations for Clamp-On Transducers Installation Effects 13.7.1 Effects of Distorted Profile by Upstream Fittings 13.7.2 Unsteady and Pulsating Flows 13.7.3 Multiphase Flows General Published Experience in Transit-Time Meters 13.8.1 Experience with Liquid Meters 13.8.2 Gas Meter Developments Applications, Advantages, and Disadvantages Doppler Flowmeter 13.10.1 Simple Explanation of Operation 13.10.2 Operational Information 13.10.3 Applications, Advantages, and Disadvantages Correlation Flowmeter 13.11.1 Operation of the Correlation Flowmeter 13.11.2 Installation Effects 13.11.3 Other Published Work 13.11.4 Applications, Advantages, and Disadvantages Other Ultrasonic Applications Chapter Conclusions APPENDIX 13.A Simple Mathematical Methods and Weight Function Analysis Applied to Ultrasonic Flowmeters 13.A.1 Simple Path Theory 13.A.2 Use of Multiple Paths to Integrate Flow Profile 13.A.3 Weight Vector Analysis 13.A.4 Doppler Theory
316 319 322 325 325 327 327 327 328 328 330 330 334 335 335 335 338 344 345 345 346 346 346 346 347 348 349 349 350 351 351 353 355 355
CONTENTS CHAPTER 14 Mass Flow Measurement Using Multiple Sensors for Single- and Multiphase Flows
14.1 14.2
14.3 14.4
14.5
Introduction Multiple Differential Pressure Meters 14.2.1 Hydraulic Wheatstone Bridge Method 14.2.2 Theory of Operation 14.2.3 Industrial Experience 14.2.4 Applications Multiple Sensor Methods Multiple Sensor Meters for Multiphase Flows 14.4.1 Background 14.4.2 Categorization of Multiphase Flowmeters 14.4.3 Multiphase Metering for Oil Production Chapter Conclusions 14.5.1 What to Measure If the Flow Is Mixed 14.5.2 Usable Physical Effects for Density Measurement 14.5.3 Separation or Multicomponent Metering 14.5.4 Calibration 14.5.5 Accuracy
CHAPTER
15.1 15.2
15.3 15.4
15.5
15.6
15 Thermal Flowmeters
357 357 357 359 359 360 361 361 362 362 363 365 367 367 368 369 369 370 371
Introduction Capillary Thermal Mass Flowmeter - Gases 15.2.1 Description of Operation 15.2.2 Operating Ranges and Materials for Industrial Designs 15.2.3 Accuracy 15.2.4 Response Time 15.2.5 Installation 15.2.6 Applications Calibration of Very Low Flow Rates Thermal Mass Flowmeter - Liquids 15.4.1 Operation 15.4.2 Typical Operating Ranges and Materials for Industrial Designs 15.4.3 Installation 15.4.4 Applications Insertion and In-Line Thermal Mass Flowmeters 15.5.1 Insertion Thermal Mass Flowmeter 15.5.2 In-Line Thermal Mass Flowmeter 15.5.3 Range and Accuracy 15.5.4 Materials 15.5.5 Installation 15.5.6 Applications Chapter Conclusions
371 371 371 374 374 374 375 376 376 376 376 377 378 378 378 379 381 381 381 381 382 383
15.A Mathematical Background to the Thermal Mass Flowmeters 15.A.I Dimensional Analysis Applied to Heat Transfer 15.A.2 Basic Theory of ITMFs
384 384 385
APPENDIX
CONTENTS
15.A.3 General Vector Equation 15.A.4 Hastings Flowmeter Theory 15.A.5 Weight Vector Theory for Thermal Flowmeters 16 Angular Momentum Devices
391.
Introduction The Fuel Flow Transmitter 16.2.1 Qualitative Description of Operation 16.2.2 Simple Theory 16.2.3 Calibration Adjustment 16.2.4 Meter Performance and Range 16.2.5 Application Chapter Conclusions
391 392 394 394 395 396 396 397
CHAPTER
16.1 16.2
16.3
386 388 389
CHAPTER 17 Coriolis Flowmeters
398
17.1
Introduction 17.1.1 Background 17.1.2 Qualitative Description of Operation 17.1.3 Experimental Investigations Industrial Designs 17.2.1 Principal Design Components 17.2.2 Materials 17.2.3 Installation Constraints 17.2.4 Vibration Sensitivity 17.2.5 Size and Flow Ranges 17.2.6 Density Range and Accuracy 17.2.7 Pressure Loss 17.2.8 Response Time 17.2.9 Zero Drift Accuracy Under Normal Operation Performance in Two-Component Flows 17.4.1 Air-Liquid 17.4.2 Sand in Water 17.4.3 Pulverized Coal in Nitrogen 17.4.4 Water-in-Oil Measurement Industrial Experience Calibration Applications, Advantages, Disadvantages, and Cost Considerations 17.7.1 Applications 17.7.2 Advantages 17.7.3 Disadvantages 17.7.4 Cost Considerations Chapter Conclusions
398 398 400 402 402 404 407 407 408 408 409 410 410 410 412 413 414 414 414 414 415 416 416 416 418 419 419 420
17.A A Brief Note on the Theory of Coriolis Meters 17.A.I Simple Theory 17.A.2 Note on Hemp;s Weight Vector Theory 17.A.3 Theoretical Developments
421 421 423 424
17.2
17.3 17.4
17.5 17.6 17.7
17.8
APPENDIX
CONTENTS
18 Probes for Local Velocity Measurement in Liquids and Gases
CHAPTER
18.1 18.2 18.3 18.4 18.5
Introduction Differential Pressure Probes - Pitot Probes Differential Pressure Probes - Pitot-Venturi Probes Insertion Target Meter Insertion Turbine Meter 18.5.1 General Description of Industrial Design 18.5.2 Flow-Induced Oscillation and Pulsating Flow 18.5.3 Applications 18.6 Insertion Vortex Probes 18.7 Insertion Electromagnetic Probes 18.8 Insertion Ultrasonic Probes 18.9 Thermal Probes 18.10 Chapter Conclusions
CHAPTER
19 Modern Control Systems
19.1
427 427 428 430 431 431 431 433 434 435 435 436 437 437 438
Introduction 19.1.1 Analogue Versus Digital 19.1.2 Present and Future Innovations 19.1.3 Industrial Implications 19.1.4 Chapter Outline 19.2 Instrument 19.2.1 Types of Signal 19.2.2 Signal Content 19.3 Interface Box Between the Instrument and the System 19.4 Communication Protocol 19.4.1 Bus Configuration 19.4.2 Bus Protocols 19.5 Communication Medium 19.5.1 Existing Methods of Transmission 19.5.2 Present and Future Trends 19.5.3 Options 19.6 Interface Between Communication Medium and the Computer 19.7 The Computer 19.8 Control Room and Work Station 19.9 Hand-Held Interrogation Device 19.10 An Industrial Application 19.11 Future Implications of Information Technology
438 439 439 440 440 441 441 442 443 444 444 445 446 446 446 447 448 448 448 449 449 449
CHAPTER 20 Some Reflections on Flowmeter Manufacture, Production, and M a r k e t s
451.
20.1 20.2 20.3 20.4 20.5
Introduction Instrumentation Markets Making Use of the Science Base Implications for Instrument Manufacture The Special Features of the Instrumentation Industry
451 451 453 454 454
CONTENTS
20.6
20.7
20.8 20.9
Manufacturing Considerations 20.6.1 Production Line or Cell? 20.6.2 Measures of Production The Effect of Instrument Accuracy on Production Process 20.7.1 General Examples of the Effect of Precision of Construction on Instrument Quality 20.7.2 Theoretical Relationship Between Uncertainty in Manufacture and Instrument Signal Quality 20.7.3 Examples of Uncertainty in Manufacture Leading to Instrument Signal Randomness Calibration of the Finished Flowmeters Actions for a Typical Flowmeter Company
455 455 456 456 457 457 459 461 461
CHAPTER 21 Future Developments
463
21.1 21.2 21.3
463 463 465 465 467 467 469 470 470 470 470 471 471 471 471 471 471 472 472 472
21.4 21.5
21.6 21.7
21.8
Market Developments Existing and New Flow Measurement Challenges New Devices and Methods 21.3.1 Devices Proposed but Not Exploited 21.3.2 New Applications for Existing Devices 21.3.3 Microengineering Devices New Generation of Existing Devices Implications of Information Technology 21.5.1 Signal Analysis 21.5.2 Redesign Assuming Microprocessor Technology 21.5.3 Control 21.5.4 Records, Maintenance, and Calibration Changing Approaches to Manufacturing and Production The Way Ahead 21.7.1 For the User 21.7.2 For the Manufacturer 21.7.3 For the Incubator Company 21.7.4 For the R&D Department 21.7.5 For the Inventor/Researcher Closing Remarks
Bibliography A Selection of International Standards
473 475
Conferences
479
References
483
Index
515
Main Index Flowmeter Index Flowmeter Application Index
515 518 521
Preface
This is a book about flow measurement and flowmeters written for all in the industry who specify and apply, design and manufacture, research and develop, maintain and calibrate flowmeters. It provides a source of information on the published research, design, and performance of flowmeters as well as on the claims of flowmeter manufacturers. It will be of use to engineers, particularly mechanical and process engineers, and also to instrument companies' marketing, manufacturing, and management personnel as they seek to identify future products. I have concentrated on the process, mechanical, and fluid engineering aspects and have given only as much of the electrical engineering details as are necessary for a proper understanding of how and why the meters work. I am not an electrical engineer and so have not attempted detailed explanations of modern electrical signal processing. I am also aware of the speed with which developments in signal processing would render any descriptions which I might give out of date. In the bibliography, other books dealing with flow measurement are listed, and my intention is not to retread ground covered by them, more than is necessary and unavoidable, but to bring together complementary information. I also make the assumption that the flowmeter engineer will automatically turn to the appropriate standard; therefore, I have tried to avoid reproducing information that should be obtained from those excellent documents. I include a brief list that categorizes a few of the standards according to meter or application. I also recommend that those involved in new developments keep a watchful eye on the regular conferences, which carry much of the latest developments in the business. I hope, therefore, that this book will provide a signpost to the essential information required by all involved in the development and use of flowmeters, from the field engineer to the chief executive of the entrepreneurial company that is developing its product range in this technology. In this book, following introductory chapters on accuracy, flow, selection, and calibration, I have attempted to provide a clear explanation of each type of flowmeter so that the reader can easily understand the workings of the various meters. I have then attempted to bring together a significant amount of the published information that explains the performance and applications of flowmeters. The two sources for this are the open literature and the manufacturers' brochures. I have also introduced, to a varying extent, the mathematics behind the meter operations, but to avoid disrupting the text, I have consigned this, in most cases, to the appendices at the end of many chapters. This follows the approach that I have used for technical review papers on turbine meters, Coriolis meters, and to a lesser extent earlier papers on
PREFACE
electromagnetic flowmeters, positive displacement flowmeters, and flowmeters in multiphase flows. However, when searching the appropriate databases for flowmeter papers, I quickly realized that including references to all published material was unrealistic. I have attempted to select those references that appeared to be most relevant and available to the typical reader of this book. However, the reader is referred to the list of journals and conferences that were especially valuable in writing this book. In particular, the Journal of Flow Measurement and Instrumentation has filled a gap in the market, judging by the large number and high quality of the papers published by the journal. It is likely that, owing to the problems of obtaining papers, I have omitted some that should have been included. Topics that I do not consider to be within the subject of bulk flow measurement of liquids and gases, and that are not covered in this book, are metering pumps, flow switches, flow controllers, flow measurement of solids and granular materials, open channel flow measurement, hot-wire local velocity probes or laser doppler anemometers, and subsidiary instrumentation. In two areas where I know that I am lacking in first-hand knowledge - modern control methods and manufacturing - 1 have included a brief review, which should not be taken as expert information. However, I want to provide a source of information for existing and prospective executives in instrumentation companies who might need to identify the type of products for their companies' future developments. This requires a knowledge of the market for each type of flowmeter and also an understanding of who is making each type of instrument. It requires some thought regarding the necessities of manufacturing and production and the implications for this in any particular design. I have briefly referred to future directions for development in each chapter where appropriate, and in the final chapter I have drawn these ideas together to provide a forward look at flow metering in general. The techniques for precise measurement of flow are increasingly important today when the fluids being measured, and the energy involved in their movement, may be very expensive. If we are to avoid being prodigal in the use of our natural resources, then the fluids among them should be carefully monitored. Flow measurement contributes to that monitoring and, therefore, demands high standards of precision and integrity.
Acknowledgments
My knowledge of this subject has benefited from many others with whom I have worked and talked over the years. These include colleagues from industry and academia, and students, whether in short courses or longer-term degree courses and research. I hope that the book does justice to all that they have taught me. In writing this book, I have drawn on the information from many manufacturers, and some have been particularly helpful in agreeing to the use of information and diagrams. I have acknowledged these companies in the captions to the figures. Some went out of their way to provide artwork, and I am particularly grateful to them. Unfortunately, space, in the end, prevented me from using many of the excellent diagrams and photographs with which I was provided. In the middle of already busy lives, the following people kindly read through sections of the book, of various lengths, and commented on them: Heinz Bernard (Krohne Ltd.), Reg Cooper (Bailey-Fischer & Porter), Terry Cousins (T&SK Flow Consultants), Chris Gimson (Endress & Hauser), Charles Griffiths [Flow Automation (UK) Ltd.], John Hemp (Cranfield University), Yousif Hussain (Krohne Ltd.), Peter lies-Smith (Yokagawa United Kingdom Limited), Alan Johnson (Fisher-Rosemount), David Lomas (ABB Kent-Taylor), Graham Mason (GEC-Marconi Avionics), John Napper (formerly with FMA Ltd.), Kyung-Am Park (Korea Research Institute of Standards and Science), Bob Peters (Daniel Europe Ltd.), Roger Porkess (University of Plymouth), Phil Prestbury (Fisher-Rosemount), David Probert (Cambridge University), Karl-Heinz Rackebrandt (Bailey-Fischer & Porter), Bill Pursley (NEL), Jane Sattary (NEL), Colin Scott (Krohne Ltd.), John Salusbury (Endress & Hauser), Dave Smith (NEL), Ian Sorbie (Meggitt Controls), Eddie Spearman (Daniel Europe Ltd.), J. D. Summers-Smith (formerly with I.C.I), and Ben Weager (Danfoss Flowmetering Ltd.). I am extremely grateful to them for taking time to do this and for the constructive comments they gave. Of course, I bear full responsibility for the final script, although their help and encouragement was greatly valued. I am also grateful to Dr. Michael Reader-Harris for his advice on the orifice plate discharge coefficient equation, and to Prof. Stan Hutton for his help and encourgement when Chapter 10 was essentially in the form of technical papers. I acknowledge with thanks the following organizations that have given permission to use their material: ASME for agreeing to the reproduction of Figures 5.10(a), 10.11, 10.16, 17.1, 18.2, and 18.4.
ACKNOWLEDGMENTS
Elsevier Science Ltd. for permission to use Figures 4.18, 5.5, 5.10(b), 5.12, 8.6,11.7, 11.11,11.13,11.14,11.17,13.9, 21.1, 21.2 and for agreement to honor my right to use material from my own papers for Chapters 10 and 17. National Engineering Laboratory (NEL) for permission to reproduce Figures 4.9, 4.14-4.16, 4.19, 5.11, 11.4-11.6, and 14.5. Professional Engineering Publishing for permission to draw on material from the Introductory Guide Series of which I am Editor, and to the Council of the Institution of Mechanical Engineers for permission to reproduce material identified in the text as being from Proceedings Part C, Journal of Mechanical Engineering Science, Vol. 205, pp. 217-229, 1991. Extracts from BS EN ISO 5167-1:1997 are reproduced with the permission of BSI under licence no. PD\ 19980886. Complete editions of the standards can be obtained by mail from BSI Customer Services, 389 Chiswick High Road, London W4 4AL, United Kingdom. I am also grateful for the help and encouragement given to me by many in the preparation of this book. It would be difficult to name them, but I am grateful for each contribution. The support of my family must be mentioned. In various ways they all contributed - by offering encouragement, by undertaking some literature searches, by doing some early typing work, and by helping with some of the diagrams. I particularly thank my wife whose encouragement and help at every stage, not to mention putting up with a husband glued to the word processor through days, evenings, holidays, etc., ensured that the book was completed. My editor has been patient, first as I overran the agreed delivery date and then as I overran the agreed length. I am grateful to Florence Padgett for being willing to overlook this lack of precision!
Nomenclature
d Diameter of tube bundle straightener tubes Q Sensitivity coefficient g Acceleration due to gravity f{x) Function for Normal distribution Hodgson's number K K factor in pulses per unit flow quantity H K Pressure loss coefficient k Coverage factor M Mach number M Mean of a sample of n readings n Index as in Equation (2.4) m Index p Pressure N(/JL, a2) Normal curve po Stagnation pressure n Number of measurements, Exponent APioss Pressure loss across a pipe fitting p Probability, Index qv Volumetric flow rate q Mean of n measurements qj, Exponent qm Mass flow rate qj Test measurement R Radius of pipe qv Volumetric flow rate Reynolds n u m b e r qvo Volumetric flow rate at calibration point Re r Radial coordinate (distance from pipe r Exponent axis) s Exponent T Temperature s(q) Experimental standard deviation of To Stagnation temperature mean of group qj V Velocity in pipe, Volume of pipework s(qj) Experimental standard deviation of qj and other vessels between t h e source t Student's t of t h e pulsation a n d t h e flowmeter U Expanded uncertainty position u(xi) Standard uncertainty for the zth quantity Vb Velocity on pipe axis uc(y) Combined standard uncertainty Kms Fluctuating component of velocity x Coordinate V Mean velocity in pipe Xi Result of a meter measurement, Input z Elevation above datum quantities y Ratio of specific heats x Mean of n meter measurement li Dynamic viscosity y Output quantity y Kinematic viscosity z Normalized coordinate (x — /x)/a p Density l± Mean value of data for normal curve SUBSCRIPTS v Degrees of freedom 1,2 Pipe sections o Standard deviation (a2 variance) 4>(z) Area under Normal curve [e.g., 4>(0.5) is the area from z = -oo to z = 0.5] CHAPTER 3 >(x) Function for normalized Normal Cj Sensitivity coefficient for the zth distribution quantity famin Lubrication film thickness CHAPTER 2 n Bearing rotational speed p Bearing load A Cross-section of pipe ui Standard uncertainty for the / th quantity Local speed of sound c Specific heat at constant pressure r) Friction coefficient Specific heat at constant volume k Specific film thickness Diameter of pipe ix Fluid viscosity D
CHAPTER 1
NOMENCLATURE
a
Combined roughness of the two contacting surfaces of the bearing
CHAPTER 4 Q Concentration of tracer in the main stream at the downstream sampling point Cdmean Mean concentration of tracer measured downstream during time t Q Concentration of tracer in the injected stream Cu Concentration of tracer in the main stream upstream of injection point (if the tracer material happens to be present)
cx Mn p qv qVi R T t V v p
Sensitivity coefficient Net mass of liquid collected in calibration Pressure Volumetric flow rate in the line Volumetric flow rate of injected tracer Gas constant for a particular gas Temperature Collection time during calibration, Integration period for tracer measurement Amount injected in the sudden injection (integration) method Specific volume Liquid density
f
H h K Li
h '2
M'2 n Pd
Pu Ap qm qv
Re
r t V
CHAPTER 5
A
a\ a€ b b€ C CRe Claps
c C\ C*
D D1 d E ET E* e F
Function of p and Re Expression in orifice plate bending formula Constant Constant Constant Discharge coefficient Part of discharge coefficient affected by Re Part of discharge coefficient which allows for position of taps Discharge coefficient for infinite Reynolds number Expression in orifice plate bending formula Constant Pipe diameter (ID) Orifice plate support diameter Orifice diameter Velocity of approach factor (1 - £4)~~1/2:, Thickness of the orifice plate Total error in the indicated flow rate of a flowmeter in pulsating flow Elastic modulus of plate material Thickness of the orifice Correction factors used to obtain the mass flow of a (nearly) dry steam flow
V
vWs X
a
P y
Hm €1 K
Pi
°y 4>
Frequency of the pulsation Hodgson number Thickness of orifice plate Loss coefficient, Related to the criterion for Hodgson's number = h/D = l'2/D (The prime signifies that the measurement is from the downstream face of the plate) Distance of the upstream tapping from the upstream face of the plate Distance of the downstream tapping from the downstream face of the plate (The prime signifies that the measurement is from the downstream face of the plate) = 2L'2/(1 - p) Index Downstream pressure Upstream pressure Differential pressure, pressure drop between pulsation source and meter Mass flow rate Volumetric flow rate Reynolds number usually based on the pipe ID Radius of upstream edge of orifice plate Time Volume of pipework and other vessels between the source of the pulsation and the flowmeter position Mean velocity in pipe with pulsating flow Root-mean-square value of unsteady velocity fluctuation in pipe with pulsating flow Dryness fraction Flow coefficient, CE Diameter ratio, d/D Ratio of specific heats Small changes or errors in qm, etc. Expansibility (or expansion) factor Expansibility (or expansion) factor for orifice
Isentropic exponent Density at the upstream pressure tapping cross-section Yield stress for plate material Ratio of two-phase pressure drop to liquid flow pressure drop Maximum allowable percentage error in pulsating flow
CHAPTER 6
C
cRe
Coefficient of discharge Part of coefficient of discharge affected by Reynolds number
NOMENCLATURE
Ctp Coo D d E
Coefficient for wet gas flow equation Discharge coefficient for infinite Reynolds number Pipe ID Throat diameter Velocity of approach factor (l-/*4)-l/2
Frg g k n Ap qg qi qm qtp qv Re Red Vsg X £ e p pg p\
Superficial gas Froude number Gravitational acceleration Roughness Index Differential pressure Gas volumetric flow rate Liquid volumetric flow rate Mass flow rate Apparent volumetric flow rate when liquid is present in the gas stream Volume flow rate Reynolds number based on D Reynolds number based on d Superficial gas velocity Lockhart-Martinelli parameter diameter ratio d/D Expansibility (or expansion) factor Density Gas density Liquid density
CHAPTER 7
A2 A* a ac az b bc bz C CR C* c cp cv d dt d2 f M Mi M n po pi p2i p2max p*
Outlet cross-sectional area Throat cross-sectional area Constant Constant Constant Constant Constant Constant Discharge coefficient = C*^Z Critical flow function Sound speed Specific heat at constant pressure Specific heat at constant volume Throat diameter Diameter of tapping Outlet diameter Obtained from Equation (7.17) Mach number Mach number at inlet when stagnation conditions cannot be assumed Molecular weight Exponent in Equation (7.12) Stagnation pressure Pressure at inlet when stagnation conditions cannot be assumed Ideal outlet pressure Actual maximum outlet pressure Throat pressure in choked conditions
R Red To % X Z ZQ
p y K V
PO
Mass flow Universal gas constant Reynolds number based on the throat diameter Stagnation temperature Throat temperature in choked conditions Mole fraction of each component of a gas mixture Compressibility factor Compressibility factor at stagnation conditions d/D Ratio of specific heats Error Isentropic exponent Kinematic viscosity Density at stagnation conditions
CHAPTER 8
A A' Af Ax A2 a B C Cc D d E F g K L M p qv R Re V V v x /x fig /xstd p Pi
Cross-sectional area of the pipe, Constant Constant Cross-sectional area of float Cross-sectional area of tapering tube at height x Annular area around float, Annular area around target Area of target Constant Coefficient Contraction coefficient, Constants in curvefitfor target meter discharge coefficient Pipe diameter Throat diameter for pipe inlet Full-scale or upper range value of flow rate used in precision calculation Summation error in flow rate Gravity Loss coefficient, Precision class, Bend or elbow meter coefficient Length of laminar flow tube Actual flow rate used in precision calculation Pressure Volumetric flow rate Radius of bend or elbow Reynolds number Velocity, Volume of float Mean velocity in tube Specific volume of gas Height of float in tube Viscosity Viscosity of calibration gas at flowing conditions Viscosity of reference gas at standard conditions Density Density of float material
NOMENCLATURE
Viscosity Angular position of arm of calibrator
CHAPTER 9
E F g
Young's modulus of elasticity Friction force Acceleration due to gravity
L
Length of clearance gap in direction of flow Axial length of clearance gap Axial length of measuring chamber Mass of rotor Rotational speed of rotor Rotational speed of cam disk of calibrator (clutch system) Rotational speed of shaft into calibrator (epicyclic system) M a x i m u m rotational speed of rotor Rotational speed of outer ring of calibrator (clutch system) Rotational speed of shaft out of calibrator (epicyclic system) Slippage in NIN N u m b e r of teeth o n gear Downstream pressure Upstream pressure As u s e d in Equation (9.A.6) Leakage flow rate Bulk volumetric flow rate
I £ax M N N[ NIN Nmax No NOUT Ns n pa pu
<7siip qv R r rx ro rs T TIN TOUT 7b T\ t t0 u y am «i A
8
Volumetric flow rate through meter at no rotation Volumetric flow rate Radius of friction wheel Radius of point of friction wheel contact Inner radius of measuring chamber Outer radius of measuring chamber Shaft radius Temperature, Torque Torque on input shaft to calibrator Torque out from calibrator Constant drag torque Speed-dependent drag torque Clearance between stationary and moving members Thickness of outer casing Fluid velocity Position coordinate across clearance gap Coefficient of linear expansion of metal Coefficient of volume expansion of liquid Change in quantity, Distance between center of rotation of arm and center of discs in calibrator Reduction in area of the measuring chamber due to blade sections
9Q
Angular position of center disk of calibrator
0[ 0o
Angular position of cam disk of calibrator Angular position of outer ring of calibrator
OTHER SUBSCRIPTS / 1
m 1,2 1-8
Dummy suffix for summation Liquid Metal Different materials Epicyclic gears
CHAPTER 1 0 A Cross-sectional area of the effective
a do, U\, #2/ #3 B b
C
cD
cD ct ch CL c D D\ D2 £>3
d\, d2 e F
f
I i
i(r) K
K(r) Ki,K2 Kh L N n
P poi P02
annular flow passage at the rotor blades Turbine meter aerodynamic torque coefficient Constant coefficients Axial length of rotor Turbine meter aerodynamic torque coefficient, Bearing length, Frequency response coefficient unless used as an exponent Proportional to 1/K Drag coefficient Drag coefficient adjusted to allow forC h Fluid drag coefficient Constant for a particular design (Behaves like an adjustment to the main aerodynamic drag term coefficient) Lift coefficient Chord Drag force, Pipe diameter, Impeller diameter Rotor response parameter Fluid drag parameter Nonfluid drag parameter Additional geometric variables Meter error Nonfluid forces Frequency of blade passing Mass moment of inertia of rotor system about rotor axis Incidence angle Incidence at radius r Lattice effect coefficient, K factor (pulses/unit volume) Lattice coefficient at radius r Turbine meter resisting torque coefficients Constant used in equation for helical blade angle Lift force, Helical pitch of blades Number of blades Rotational speed (frequency) Index in error equation Stagnation pressure at inlet Stagnation pressure at outlet
NOMENCLATURE pi p2 q
Static pressure at inlet Static pressure at outlet Actual average flow rate, Index in error equation Time average flow rate over period T Initial flow rate Final flow rate Base flow rate
a
q\
Indicated average flow rate Maximum flow rate for which the meter is designed
r
<7min
Minimum flow rate for which the meter is designed Normal flow rate Flow rate at time t Flow rate at change of precision Volumetric flow rate Pipe bend radius Reynolds number Radial position, Index in error equation Journal bearing radius Hub radius Meter bore radius Tip radius Slope of no-flow decay curve at standstill Blade spacing Temperature, Fundamental period of pulsating flow, Torque
qo q\ qi qb
qn qt ^trans qv R Re r rj rh ro rt S s T
TB
Bearing drag torque
Tc Td Tpo Th Tn 7^> Tr Tt Tw t
Temperature at calibration Driving t o r q u e Mechanical friction torque o n rotor at zero speed H u b fluid drag torque Nonfluid drag torque Temperature at operation Retarding torque Blade tip clearance drag torque H u b disk friction drag torque Time
£B
Blade thickness
£R V Vo V\ V2 ^rnax ^rnin
Relaxation t i m e N o n d i m e n s i o n a l fluid velocity Time average value of Vz over period T Inlet relative flow velocity Outlet relative flow velocity M a x i m u m value of Vz M i n i m u m value of Vz
Vz W Wh Y Z
Axial velocity (instantaneous inlet fluid velocity of pulsating flow) Rotor blade velocity, Nondimensional instantaneous rotor velocity under pulsating flow Relative velocity at the hub Tangential force Axial force
Pi
p2
Pm 8
n V
p X
Wave shape coefficient, Thermal coefficient of expansion, Angle between inlet flow direction and far field flow Blade angle at radius r measured from axial direction of meter Relative inlet angle of flow Relative outlet angle of flow Mean of the inlet and outlet flow field Full-flow amplitude relative to average flow Deflection of flow at blade outlet from blade angle Flow deviation factor Dynamic viscosity Kinematic viscosity Fluid density Nondimensional time (t/T), Period of modulation Rotor coast time to standstill Instantaneous rotor angular velocity
CHAPTER 11
A
a B
D
f
H h K L A/7 Patmos Pgmin
Pv qv
Re S s V Vmax W
8 Q CO
Pipe full flow area, Constant of value about 3.0 used in equation for avoiding cavitation in flows past vortex meters Flow area past bluff body Area when integrating vorticity Constant of value about 1.3 used in equation for avoiding cavitation in flows past vortex meters Pipe ID Shedding frequency Streamwise length of bluff body Parallel flats on sides of bluff body at leading edge Calculation factor for shedding frequency (Zanker and Cousins 1975), K factor = pulses/unit volume Length of bluff body across pipe between end fittings Pressure drop across vortex meter (about 1 bar at 10 m/s) Atmospheric pressure Minimum back pressure 5D downstream of vortex meter Saturated liquid vapor pressure Volumetric flow rate Reynolds number Strouhal number Length along curve when integrating velocity around a vortex Velocity of flow Velocity of flow past bluff body Diameter or width of bluff body Shear layer thickness Injected swirling flow rate/total flow rate Vorticity
NOMENCLATURE CHAPTER 12 A Used for area of electrode leads forming a loop causing quadrature signals in the electromagnetic flowmeter a Pipe radius B Magnetic flux density in tesla B Magnetic flux density vector Bo Maximum value of magnetic flux density Be 6 component of magnetic flux density b Inner radius of an annulus of conducting fluid in two phase annular flow D Diameter of pipe E Electric field vector f Frequency of magnetic field excitation j Current density vector / Length of wire traversing magnetic field Ap Pressure of the pipeline fluid above atmospheric Volumetric flow rate qv r Radial coordinate 5 Electromagnetic flowmeter sensitivity t Pipe wall thickness U Electric potential AI/EE Voltage between electrodes AC/p Voltage across a wire P moving through a magnetic field (similarly for Q and R) V Velocity
V
Velocity vector
Vm W W W W" Wz a € 6 \x a
Mean velocity in the pipe in meters per second Weighting function Weight function vector Rectilinear flow weight function Axisymmetric weight function Axial component of W Void fraction Strain Cylindrical coordinate Electric permeability Conductivity
Scalar potential
CHAPTER 1 3 Weighting factors for Gaussian A
a c D
f /d
ft fr /u
quadrature Radius of pipe Velocity of ultrasound in the medium, Coefficients of a polynomial Pipe diameter Frequency of stable quartz oscillator for ultrasonic measurement system Frequency for the downstream sing-around pulse train Transmitted doppler frequency Reflected doppler frequency Frequency for the upstream sing-around pulse train
fit) h I L
PR PI
Re Ryxir) T t tm At
V Vm V(X) V5 Vo V
W X
y
z
z
(3 7
e
X p Pm T
xm
Difference between the sing-around frequencies, Frequency shift in doppler flowmeter Function of t Displacement of ultrasonic beam from axis of pipe Driving current in ultrasonic flowmeter Adiabatic compressibility Distance along path in transit-time flowmeter, Distance between transducers in ultrasonic correlation flowmeter Counters for ultrasonic measuring system \/n is the index for turbulent profile curve fit, Index Ultrasound power reflected Ultrasound power transmitted Mass flow rate Reynolds number Correlation coefficient Correlation time for integration Pipe wall thickness Downstream wave transit time Mean wave transit time Upstream wave transit time Difference between these two wave transit times Beam positions for ultrasonic flowmeter using Gaussian quadrature Received voltages for ultrasonic flowmeter Flow velocity in pipe Mean velocity in pipe Flow velocity profile along pipe chord Ultrasonic flowmeter created flow Acoustic field for cases (1) and (2) Relative flow velocity in the direction of the acoustic beam Vector weight function Axial length between transducers in ultrasonic flowmeter Length of chord in the plane of the ultrasonic path Distance along chord used in ultrasonic flowmeter Acoustic impedance Ultrasonic beam deflection distance on opposite wall Ultrasonic beam deflection angle Ratio of specific heats for a particular gas Ultrasound beam angle Ultrasound wave length Density of material Mean density in ultrasonic meter Measurement period for ultrasonic system, Time period between ultrasonic wave peaks Mean ultrasonic correlation time
NOMENCLATURE CHAPTER 14
A c K ks p A/?A A/?B ApAB etc.
qm qvp V 7 f>
Area of duct Speed of sound Constant Adiabatic compressibility Pressure Pressure drop to throat of venturi A Pressure drop to throat of venturi B Differential pressure between venturi throats, across limbs of the hydraulic Wheatstone bridge, or across diagonals of t h e hydraulic Wheatstone bridge Total mass flow Metering p u m p volumetric transfer flow Velocity in t h e meter Ratio of specific heats Density of the fluid
CHAPTER 15
A B C,D cp cv D d g / K k kf I Nu n Pe Pr p Qh <j a q\y qh qm qv R Re 5 So St T Ti,Tz,Tc AT t V V
Area of the duct, Constant equal to 1/n Constant equal to (2/TT) 0 - 5
Functions of temperature Specific heat at constant pressure Specific heat at constant volume Pipe diameter Heating element diameter Gravitational constant Current through resistance R Constants Thermal conductivity of the fluid Constant allowing for heat transfer and temperature difference at zero flow Finite difference dimension Nusselt number Index Peclet number Prandtl number Pressure Heat transfer Rate of heat addition per unit volume Heat flux Heat flux vector Mass flow rate Volumetric flow rate Resistance of heating element Reynolds number Flow signal Flow signal at start Stanton number Absolute temperature of the fluid Temperatures used in finite difference approximation Measured temperature difference in Kelvins Time from start of flow change Fluid velocity Vector velocity
W p [i v T
Weight vector Fluid density Dynamic viscosity of the fluid Kinematic viscosity Time constant of flowmeter
CHAPTER 16
qm R s X 0 T co
Mass flow Radius of the annulus of angular momentum meter Spring constant Angular momentum Angular deflection Time difference between markers on the angular momentum meter rotating assembly Angular velocity of the rotor
CHAPTER 17 A Cross-sectional area of the pipe d Width of U-tube
F h
/u
K
Ks Ku I 8m qm
r 8r 8r' T t V V v (!) v (2)
w 0 p X
Q
Force due to Coriolis acceleration Inertia in plane of twisting oscillation Inertia in plane of normal oscillation Constant, Allows for the fact that the twist of the tube will not form a straight integration Spring constant of the U-tube in twisting oscillation Spring constant of the U-tube in normal oscillation Length of the U-tube Element of mass equal to pA8r' Mass flow Radius Elementary length of tube Elementary length of fluid Torque Time Flow velocity Vector velocity Oscillatory velocity field set up in the stationary fluid by the driving oscillator Oscillatory velocity field set up in the stationary fluid by the Coriolis forces Vector weight function Twist angle of U-tube Amplitude of twist angle of U-tube when in sinusoidal motion Density Difference in transit time of two halves of twisted U-tube Phase difference between the total velocities at the two sensing points Angular velocity of the pipe caused by the vibration
NOMENCLATURE
CO
cos COu
Amplitude of the angular velocity of the pipe caused by the vibration Driving frequency Natural frequency of U-tube in twisting oscillation Natural frequency of U-tube in normal oscillation
CHAPTER
18
k M p p0 Ap V y p
Coefficient Mach number Pressure Stagnation pressure Dynamic pressure Velocity Ratio of specific heats Density
A At Anin
Ax A2 A* a
0, 01/
ac
CHAPTER 20
C cp D d eq K Ks ma me Pi, pi, p3, p4 qm qw 5 At tm 0 r co cos
Cost Component precision Pipe diameter Width of Coriolis flowmeter U-tube Electronics quality Constant Spring constant of the U-tube in twisting oscillation Manufacturing accuracy Material consistency Dimensional and other factors Mass flow rate Volumetric flow rate Flow signal Time difference between upstream and downstream waves or pulses Mean time of transit used to obtain sound speed Ultrasound beam angle Difference in transit time of two halves of twisted U-tube Driving frequency Natural frequency of U-tube in twisting oscillation
B Bo
bz
h C
CD
COMBINED ALL CHAPTERS
A
Cross-section of pipe, Function of p and Re, Cross-sectional area of the pipe, Constant, Cross-sectional area of the effective annular flow passage at the rotor blades, Constant of value about 3.0 used in equation for avoiding cavitation in flows past vortex meters, Used for area of electrode leads forming a loop causing quadrature signals in the electromagnetic flowmeter, Weighting factors for Gaussian quadrature, Area of the duct, Constant equal to 1 In
Constant Cross-sectional area of float Flow area past bluff body Cross-sectional area of tapering tube at height x Outlet cross-sectional area, Annular area around float, Annular area around target Throat cross-sectional area Constant, Area of target, Turbine meter aerodynamic torque coefficient, Area when integrating vorticity, Pipe radius
Qmean Q
Q CL CR
Constant coefficients Expression in orifice plate bending formula Constant Constant Constant Constant, Axial length of rotor, Constant of value about 1.3 used in equation for avoiding cavitation in flows past vortex meters, Magnetic flux density in tesla, Constant equal to (2/TT) 0 - 5 Magnetic flux density vector Maximum value of magnetic flux density 0 component of magnetic flux density Inner radius of an annulus of conducting fluid in two-phase annular flow Constant, Turbine meter aerodynamic torque coefficient, Bearing length, Frequency response coefficient unless used as an exponent Constant Constant Constant Discharge coefficient, Coefficient, Proportional to 1/K, Function of temperature, Cost Contraction coefficient, Constants in curve fit for target meter discharge coefficient Drag coefficient Drag coefficient adjusted to allow for C h Concentration of tracer in the main stream at the downstream sampling point Mean concentration of tracer measured downstream during time t Fluid drag coefficient Constant for a particular design (Behaves like an adjustment to the main aerodynamic drag term coefficient) Concentration of tracer in the injected stream Lift coefficient
NOMENCLATURE
CRe Craps Qp Cu Coo C* c c\ Q cp cp cv c€ D D' D\ D2 D3 d
dt d\ dz E
E Ej E* e eq F
Frg f
fy
Part of discharge coefficient affected by Re °f discharge coefficient that allows for position of taps Coefficient for wet gas flow equation Concentration of tracer in t h e main stream upstream of injection point (if the tracer material happens to be present) Discharge coefficient for infinite Reynolds n u m b e r Critical flow function Sound/ultrasound speed, Chord, Coefficients of a polynomial Expression in orifice plate bending formula Sensitivity coefficient Specific heat at constant pressure C o m p o n e n t precision Specific heat at constant volume Constant Pipe diameter (ID), Drag force, Impeller diameter, Function of temperature Orifice plate support diameter Rotor response parameter Fluid drag parameter Nonfluid drag parameter Diameter of tube bundle straightener tubes, Orifice diameter, Throat diameter, Heating element diameter, Width of Coriolis flowmeter U-tube Diameter of tapping Additional geometric variable Outlet diameter, Additional geometric variable velocity of approach factor (1 - £ 4 )~ 1 / 2 , Thickness of the orifice plate, Full-scale or upper range value of flow rate used in precision calculation, Young's modulus of elasticity, Fluid compressibility Electric field vector Total error in the indicated flow rate of a flowmeter in pulsating flow Elastic modulus of plate material Thickness of the orifice, Meter error Electronics quality Correction factors used to obtain the mass flow of a (nearly) dry steam flow, Summation error in flow rate, Friction force, Nonfluid forces, Force due to Coriolis acceleration Superficial gas Froude number Frequency of the pulsation, Obtained from Equation (7.17), Frequency of blade passing, Shedding frequency, Frequency of magnetic field excitation, Frequency of stable quartz oscillator for ultrasonic measurement system Frequency for the downstream sing-around pulse train
Part
ft ft fu Af f{t) fix) g H h
frmin /
Is Ju i /(r) j K
Kir) K\,Ki Kh Ks Ku k k! ks L
Transmitted doppler frequency Reflected doppler frequency Frequency for the upstream sing-around pulse train Difference between the sing-around frequencies, Frequency shift in doppler flowmeter Function of t Function for Normal distribution Acceleration due to gravity Hodgson number, Streamwise length of bluff body Thickness of orifice plate, Parallel flats on sides of bluff body at leading edge, Displacement of ultrasonic beam from axis of pipe Lubrication film thickness Mass moment of inertia of rotor system about rotor axis, Driving current in ultrasonic flowmeter, Current through resistance R Inertia in plane of twisting oscillation Inertia in plane of normal oscillation Incidence angle Incidence at radius r Current density vector Pressure loss coefficient, Related to the criterion for Hodgson's number, Precision class, Bend or elbow meter coefficient, Lattice effect coefficient, K factor (pulses/unit volume), Calculation factor for shedding frequency (Zanker and Cousins 1975), Constant, Allows for the fact that the twist of the tube will not form a straight integration Lattice coefficient at radius r Turbine meter resisting torque coefficients Constant used in equation for helical blade angle Spring constant of the U-tube in twisting oscillation Spring constant of the U-tube in normal oscillation Coverage factor, Roughness, Thermal conductivity of the fluid, Coefficient Constant allowing for heat transfer and temperature difference at zero flow Adiabatic compressibility Length of laminar flow tube, Length of clearance gap in direction of flow, Lift force, Helical pitch of blades, Length of bluff body across pipe between end fittings, Distance along path in transit-time flowmeter, Distance between transducers in ultrasonic correlation flowmeter, Finite difference dimension
NOMENCLATURE
L'2
V2
I M
M Mi M2 Mn m 8m ma me N
Ns Nd,N n Nt,Nu N(/JL,
Nu
PR
Pr Pe Pr Patmos
a2)
h/D V2/D (The prime signifies that the measurement is from the downstream face of the plate) Length of wire traversing magnetic field, Length of the U-tube Distance of the upstream tapping from the upstream face of the plate Distance of the downstream tapping from the downstream face of the plate (The prime signifies that the measurement is from the downstream face of the plate) Axial length of clearance gap Axial length of measuring chamber Mean of a sample of n readings, Mach number, Actual flow rate used in precision calculation, Mass of rotor Molecular weight Mach number at inlet when stagnation conditions cannot be assumed = 2L'2/(1 - P) Net mass of liquid collected in calibration Index Element of mass equal to pASr' Manufacturing accuracy Material consistency Rotational speed of rotor, Number of blades Rotational speed of cam disk of calibrator (clutch system) Rotational speed of shaft into calibrator (epicyclic system) Maximum rotational speed of rotor Rotational speed of outer ring of calibrator (clutch system) Rotational speed of shaft out of calibrator (epicyclic system) Slippage in N|N Counters for ultrasonic measuring system Normal curve Nusselt number Number of measurements, Exponent in equations, Bearing rotational speed, Number of teeth on gear, Rotational speed (frequency) Probability, Pressure, Bearing load, Exponent in error equation Ultrasound power reflected Ultrasound power transmitted Peclet number Prandtl number Atmospheric pressure Downstream pressure Minimum back pressure 5D downstream of vortex meter
Po
Stagnation pressure Upstream pressure Saturated liquid vapor pressure Stagnation pressure at inlet P02 Stagnation pressure at outlet Pressure at inlet when stagnation Pi conditions cannot be assumed, Static pressure at inlet, p\pi dimensional and other factors Static pressure at outlet P2 Ideal outlet pressure P2i Actual maximum outlet pressure P2max Throat pressure in choked conditions P* Pressure drop to throat of venturi A Pressure drop to throat of venturi B ApB ApAB etc . Differential pressure between venturi throats, across limbs of the hydraulic Wheatstone bridge, or across diagonals of the hydraulic Wheatstone bridge Pressure loss across a pipe fitting Apioss Differential pressure, Pressure drop Ap between pulsation source and meter, Pressure drop across vortex meter (about 1 bar at 10 m/s), Pressure of the pipeline fluid above atmospheric, Dynamic pressure Heat transfer Qh Actual average flow rate; Exponent in q error equation Mean of n measurements q\ q Time average flow rate over period T qo Initial flow rate q\ Final flow rate qi Rate of heat addition per unit volume q* Base flow rate Bulk volumetric flow rate
NOMENCLATURE
qv\ R
R Re Red Ryx(r)
r
n r ) ro rs
8r 8r' S
So St 5
s(q) s(q\) T 7B
Tc Td TFO
Th TIN
Tn TOUT
To Tr Tt TW 7i
Volumetric flow rate at calibration point Metering pump volumetric transfer flow Volumetric flow rate of injected tracer Radius of pipe, Gas constant for a particular gas, Radius of bend or elbow, Radius of friction wheel, Resistance of heating element, Radius of the annulus of angular momentum meter Universal gas constant Reynolds number usually based on the pipe ID Reynolds number based on the throat diameter Correlation coefficient Radial coordinate (distance from pipe axis), Radius of upstream edge of orifice plate, Radius of point of friction wheel contact, Exponent, Radius Hub radius Inner radius of measuring chamber Journal bearing radius Tip radius Outer radius of measuring chamber, Meter bore radius Shaft radius Elementary length of tube Elementary length of fluid Slope of no-flow decay curve at standstill, Strouhal number, Electromagnetic flowmeter sensitivity, Flow signal Flow signal at start Stanton number Exponent, Experimental standard deviation, Blade spacing, Length along curve when integrating velocity around a vortex, Spring constant Experimental standard deviation of mean of group q^ Experimental standard deviation of q^ Temperature, Torque, Fundamental period of pulsating flow Bearing drag torque Temperature at calibration Driving torque Mechanical friction torque on rotor at zero speed Hub fluid drag torque Torque on input shaft to calibrator Nonfluid drag torque Torque out from calibrator Stagnation temperature, Constant drag torque, Temperature at operation Retarding torque Blade tip clearance drag torque Hub disk friction drag torque Speed-dependent drag torque
Ti,T2,Tc T* AT
t
£B
td tm to £R
tu At
t\,t2 U AC/EE
AC/p
My) V
V *4ms
V v (l)
v(2)
vm nnax ^min
vs Vo
vz
Temperatures used in finite difference approximation Throat temperature in choked conditions Measured temperature difference in Kelvins Student's t, Collection time during calibration, Integration period for tracer measurement, Time, Clearance between stationary and moving members, Pipe wall thickness Blade thickness Downstream wave time Mean time of transit used to obtain sound speed Thickness of outer casing Relaxation time Upstream wave time Time difference between upstream and downstream waves or pulses Beam positions for ultrasonic flowmeter using Gaussian quadrature Expanded uncertainty, Electric potential Voltage between electrodes Voltage across a wire P moving through a magnetic field (similarly for Q and R) Received voltages for ultrasonic flowmeter Standard uncertainty for the /th quantity Combined standard uncertainty Fluid velocity, Amount injected in the sudden injection (integration) method, Volume of pipework and other vessels between the source of the pulsation and the flowmeter position, Volume of float, Nondimensional fluid velocity Mean velocity in pipe with pulsating flow Root-mean-square value of unsteady velocity fluctuation in pipe with pulsating flow Velocity vector Oscillatory velocity field set up in the stationary fluid by the driving oscillator Oscillatory velocity field set up in the stationary fluid by the Coriolis forces Mean velocity in the pipe in meters per second Maximum value of Vz, Velocity of flow past bluff body Minimum value of Vz Velocity on pipe axis Ultrasonic undisturbed flow Acoustic field for cases (1) and (2) Superficial gas velocity Axial velocity (instantaneous inlet fluid velocity of pulsating flow)
NOMENCLATURE Vo V\
T i m e average v a l u e of Vz over p e r i o d T I n l e t r e l a t i v e flow v e l o c i t y
Vz V v
Outlet relative flow velocity Mean velocity in pipe Specific volume of gas, Relative flow velocity in the direction of the acoustic beam Rotor blade velocity, Nondimensional instantaneous rotor velocity under pulsating flow, Weighting function Weight function vector Rectilinear flow weight function Axisymmetric weight function Relative velocity at the h u b Axial component of W Diameter or width of bluff body Axial length between transducers in ultrasonic flowmeter, Mole fraction of each component of a gas mixture, Angular m o m e n t u m , LockhartMartinelli parameter Coordinate, Dryness fraction, Height of float in tube Result of a meter measurement Mean of n meter measurements Length of chord in the plane of the ultrasonic path, Tangential force Output quantity, Wetness fraction, Position coordinate across clearance gap, Distance along chord used in ultrasonic flowmeter Compressibility factor, Axial force, Acoustic impedance Compressibility factor at stagnation conditions Normalized coordinate (x — ii)/o, Elevation above datum, Ultrasonic beam deflection distance on opposite wall Flow coefficient CE, Wave shape coefficient, Thermal coefficient of expansion, Angle between inlet flow direction and far field flow, Void fraction Coefficient of linear expansion of metal Coefficient of volume expansion of liquid Diameter ratio d/D, Blade angle at radius r measured from axial direction of meter, Ultrasonic beam deflection angle Relative inlet angle of flow Relative outlet angle of flow Mean of the inlet and outlet flow field Ratio of specific heats Full-flow amplitude relative to average flow Change in quantity, Distance between center of rotation of arm and center of discs in calibrator
W
W W W" W\i Wz w X
x Xi x Y y
Z Zo z
a
am a\ 0
01 02 0m y r A
8qm etc.
fig
p Pi pm po o
r
rm
Reduction in area of the measuring chamber due to blade sections, Deflection of flow at blade outlet from blade angle, Shear layer thickness Small changes or errors in qm, etc. Expansibility (or expansion) factor, Error, Strain Expansibility (or expansion) factor for orifice Friction coefficient, Flow deviation factor Cylindrical coordinate, Ultrasound beam angle, Angular deflection, Twist angle of U-tube Angular position of center disk of calibrator Angular position of cam disk of calibrator Angular position of outer ring of calibrator, Amplitude of twist angle of U-tube when in sinusoidal motion Isentropic exponent Specific film thickness, Ultrasound wave length Mean value of data for normal curve, Dynamic viscosity, Fluid viscosity, Electric permeability Viscosity of calibration gas at flowing conditions Viscosity of reference gas at standard conditions Degrees of freedom, Kinematic viscosity Density, Fluid density Density of float material Mean density in ultrasonic meter Density at stagnation conditions Standard deviation (a2 variance), Combined roughness of the two contacting surfaces of the bearing, Conductivity Yield stress for plate material Nondimensional time (t/T), Period of modulation, Measurement period for ultrasonic system, Time period between ultrasonic wave peaks, Time constant of flowmeter, Time difference between markers on the angular momentum meter rotating assembly, Difference in transit time of two halves of twisted U-tube Mean ultrasonic correlation time Rotor coast time to standstill Angular position of arm of calibrator, Scalar magnetic potential Area under Normal curve [e.g., 0(0.5) is the area from z= - oo to z = 0.5]
NOMENCLATURE
Q
Ratio of two-phase pressure drop to liquid flow pressure drop Maximum allowable percentage error in pulsating flow Function for normalized Normal distribution Phase difference between the total velocities at the two sensing points Injected swirling flow rate/total flow rate, Angular velocity of the pipe caused by the vibration Amplitude of the angular velocity of the pipe caused by t h e vibration in sinusoidal motion
CO
cos cou
Instantaneous rotor angular velocity, Vorticity, Angular velocity of the rotor, Driving frequency Natural frequency of U-tube in twisting oscillation Natural frequency of U-tube in normal oscillation
OTHER SUBSCRIPTS Dummy suffix for summation / / Liquid m Metal 1,2 Pipe sections, Different materials 1-8 Epicyclic gears
CHAPTER
Introduction
1.1
INITIAL CONSIDERATIONS
Some years ago at Cranfield, where we had set up a flow rig for testing the effect of upstream pipe fittings on certain flowmeters, a group of senior Frenchmen were being shown around and visited this rig. The leader of the French party recalled a similar occasion in France when visiting such a rig. The story goes something like this. A bucket at the end of a pipe seemed particularly out of keeping with the remaining high tech rig. When someone questioned the bucket's function, it was explained that the bucket was used to measure the flow rate. Not to give the wrong impression in the future, the bucket was exchanged for a shiny new high tech flowmeter. In due course, another party visited the rig and observed the flowmeter with approval. "And how do you calibrate the flowmeter?" one visitor asked. The engineer responsible for the rig then produced the old bucket! This book sets out to guide those who need to make decisions about whether to use a shiny flowmeter, an old bucket, nothing at all, or a combination of these! It also provides information for those whose business is the design, manufacture, or marketing of flowmeters. I hope it will, therefore, be of value to a wide variety of people, both in industry and in the science base, who range across the whole spectrum from research and development through manufacturing and marketing. In my earlier book on flow measurement (Baker 1988/9), I provided a brief statement on each flowmeter to help the uninitiated. This book attempts to give a much more thorough review of published literature and industrial practice. This first chapter covers various general points that do not fit comfortably elsewhere. In particular, it reviews recent guidance on the accuracy of flowmeters (or calibration facilities). The second chapter reviews briefly some essentials of fluid mechanics necessary for reading this book. The reader will find a fuller treatment in Baker (1996), which also has a list of books for further reading. A discussion of how to select a flowmeter is attempted in Chapter 3, and some indication of the variety of calibration methods is given in Chapter 4, before going in detail in Chapters 5-17 into the various high (and low) tech meters available. Chapter 18 deals with probes, Chapter 19 gives a brief note on modern control systems, and Chapter 20 provides some reflections on manufacturing and markets. Finally, Chapter 21 raises some of the interesting directions in which the technology is likely to go in the future.
INTRODUCTION
In this book, I have tried to give a balance between the laboratory ideal, the manufacturer's claims, the realities of field experience, and the theory behind the practice. I am very conscious that the development and calibration laboratories are sometimes misleading places, which omit the problems encountered in the field (Stobie 1993), and particularly so when that field happens to be the North Sea. In the same North Sea Flow Measurement Workshop, there was an example of the unexpected problems encountered in precise flow measurement (Kleppe and Danielsen 1993), resulting, in this case, from a new well being brought into operation. It had significant amounts of barium and strontium ions, which reacted with sulfate ions from injection water and caused a deposit of sulfates from the barium sulfate and strontium sulfate that were formed. With that salutary reminder of the real world, we ask an important - and perhaps unexpected - question.
1.2
DO WE NEED A FLOWMETER?
Starting with this question is useful. It may seem obvious that anyone who looks to this book for advice on selection is in need of a flowmeter, but for the process engineer it is an essential question to ask. Many flowmeters and other instruments have been installed without careful consideration being given to this question and without the necessary actions to ensure proper documentation, maintenance, and calibration scheduling being taken. They are now useless to the plant operator and may even be dangerous components in the plant. Thus before a flowmeter is installed, it is important to ask whether the meter is needed, whether there are proper maintenance schedules in place, whether the flowmeter will be regularly calibrated, and whether the company has allocated to such an installation the funds needed to achieve this ongoing care. Such care will need proper documentation. The water industry in the United Kingdom has provided examples of the problems associated with unmaintained instruments. Most of us who are involved in the metering business will have sad stories of the incorrect installation or misuse of meters. Reliability-centered maintenance recognizes that the inherent reliability depends on the design and manufacture of an item, and if necessary this will need improving (Dixey 1993). It also recognizes that reliability is preferable in critical situations to extremely sophisticated designs, and it uses failure patterns to select preventive maintenance. In some research into water consumption and loss in urban areas, Hopkins et al. (1995) found that obstacles to accurate measurements were • • • • •
buried control valves, malfunctioning valves, valve gland leakage, hidden meters that could not be read, and locked premises denying access to meters.
They commented that "water supply systems are dynamic functions having to be constantly expanded or amended. Consequently continuous monitoring, revisions and amendments of networks records is imperative. Furthermore, a proper
1.2 DO WE NEED A FLOWMETER?
programme of inspection, maintenance and subsequent recording must be operative in respect of inter alia: • • • • • •
networks, meters, control valves, air valves, pressure reducing valves, non-return valves."
They also commented on the poor upstream pipework at the installation of many domestic meters. So I make no apology for emphasizing the need to assess whether a flowmeter is actually needed in any specific application. If the answer is yes, then there is a need to consider the type of flowmeter and whether the meter should be measuring volume or mass. In most cases, the most logical measure is mass. However, by tradition and industrial usage, there are places where volume measurement may be the norm, and as a result, the regulations have been written for volume measurement. This results in a Catch-22 situation. The industry and the regulations may, reasonably, resist change to mass flow measurement until there is sufficient industrial experience, but industrial experience is not possible until the industry and the regulations allow. The way forward is for one or more forward-looking companies to try out the new technology and obtain field experience, confidence in the technology, and approval. In this book, I have made no attempt to alert the reader to the industry-specific regulations and legal requirements, although some are mentioned. Some regulations are touched on by the various authors, and Miller (1996) is a source of information on many documents. The main objective of the Organisation International de Metrologie Legale (OIML) is to prevent any technical barriers to international trade resulting from conflicting regulations for measuring instruments. With regard to flow measurement, it is particularly concerned with the measurement of domestic supplies and industrial supplies of water and gas (Athane 1994). This is because there are two parties involved, the supplier and the consumer, and the consumer is unlikely to be able to ascertain the correct operation of the meter. In addition these measurements are not monitored continually by the supplier, the meters may fail without anyone knowing, the usage is irregular and widely varying in rate, the measurements are not repeatable, and the commodities have increased in value considerably in recent years. In order to reduce discussions and interpretation problems between manufacturers and authorized certifying institutes, the European Commission is mandating the European standardization body (CEN/CENELEC) to develop harmonized standards that will give the technical details and implementation of the requirements based on OIML recommendations. These are such that a measuring instrument complies with essential requirements, assuming that the manufacturer has complied with them (Nederlof 1994). The manufacturer will also be fully aware of the electromagnetic compatibility (EMC), which relates to electromagnetic interference. In particular, the EMC
INTRODUCTION
characteristics of a product are that • the level of electromagnetic disturbance generated by the instrument will not interfere with other apparatus, and • the operation of the instrument will not be adversely affected by electromagnetic interference from its environment. In order to facilitate free movement within the European area the CE mark identifies products that conform to the European essential requirements, and all products must be so marked within the European Economic Area (DTI 1993, Chambers 1994). First, we consider the knotty problem of how accurate the meter should be.
1.3
HOW ACCURATE?
There continues to be inconsistency about the use of terms that relate to accuracy and precision. This stems from a slight mismatch between the commonly used terms and those that the purists and the standards use. Thus we commonly refer to an accurate measurement, when strictly we should refer to one with a small value of uncertainty. We should reserve the use of the word accurate to refer to the instrument. A high quality flowmeter, carefully produced with a design and construction to tight tolerances and with high quality materials as well as low wear and fatigue characteristics, is a precise meter with a quantifiable value of repeatability. Also, it will, with calibration on an accredited facility, be an accurate meter with a small and quantifiable value of measurement uncertainty. In the context of flowmeters, the word repeatability is preferred to reproducibility. The meanings are elaborated on later, and I regret the limited meaning now given to precision, which I have used more generally in the past and shall slip back into in this book from time to time! In the following chapters, I have attempted to be consistent in the use of these words. However, many claims for accuracy may not have been backed by an accredited facility, but I have tended to use the phrase "measurement uncertainty" for the claims made. Hayward (1977) used the story of William Tell to illustrate precision. William Tell had to use his cross-bow to fire an arrow into an apple on his little son's head. This was a punishment for failing to pay symbolic homage to an oppressive Austrian ruler. Tell succeeded because he was an archer of great skill and high accuracy. An archer's ability to shoot arrows into a target provides a useful illustration of some of the words related to precision. So Figure 1.1 (a) shows a target with all the shots in the bull's-eye. Let us take the bull's-eye to represent ±1%, within the first ring ±3%, and within the second ring ±5%. Ten shots out of ten are on target, but how many will the archer fire before one goes outside the bull's-eye? If the archer, on average, achieves 19 out of 20 shots within the bull's-eye [Figure l.l(b)], we say that the archer has an uncertainty of ± 1 % (the bull's-eye) with a 95% confidence level (19 out of 20 on the bull's-eye: 19 + 20 = 0.95 = 95 -=-100 = 95%). Suppose that another archer clusters all the arrows, but not in the bull's-eye, Figure l.l(c). This second archer is very consistent (all the shots are within the same size circle as the bull's-eye), but this archer needs to adjust his aim to correct the
1.3
HOW ACCURATE?
offset. We could say that the second archer has achieved high repeatability of ±1%, but with a bias of 4%. We might even find that 19 out of 20 shots fell within the top left circle so that we could say that this archer achieved a repeatability within that circle (b) of ± 1 % with a 95% confidence. Suppose this archer had fired one shot a day, and they had all fallen onto a small area [Figure l.l(c)], despite slight changes in wind, sunshine, and archer's mood, then we term this good day-to-day repeatability. But how well can we depend on the archer's bias? Is there an uncertainty related to it? Finally, a third archer shoots 20 shots and achieves the distribution in Figure l.l(d). One has missed entirely, but 19 out of 20 have hit the target somewhere. The archer has poor accuracy, and the uncertainty in this archer's shots is about five 1 times greater than for the first, even though t -2 -1 CENTRE the confidence level at which this archer OF TARGET performs is still about 95%. (e) (f) If the third archer has some skill, then Figure 1.1. Precision related to the case of an archery the bunching of the arrows will be greater target, (a) Good shooting — 10 out of 10 arrows in the bull's-eye than in the next circle have hit the bull's-eye. An accurate archer? (b) Good out, and the distribution by ring will be as shooting? —19 out of 20 arrows have hit the bull's-eye. An accurate archer and a low value of uncertainty shown in Figure l.l(e). (±1%) with a 95% confidence level, (c) Shots all fall We shall find that the distribution of in a small region but not the bull's-eye. Good repeatareadings of a flowmeter results in a curve bility (±1%) but a persistent bias of 4%. (d) Shots, approximating a Normal distribution with all but one, fall on the target — 19 out of 20 have hit the target. A ±5% uncertainty with 95% confia shape similar to that for the shots. Figure dence level, (e) Distribution of shots in (d) on a linear 1.1 (f) shows such a distribution where 95% plot, assuming that we can collapse the shots in a ring of the results lie within the shaded area and semicircle onto the axis, (f) The Normal distribution, the width of that area can be calculated to which is a good approximation for the distribution of give the uncertainty, ± 1 % say, of the read- flowmeter readings. ings with a 95% confidence level. In other words, 19 of every 20 readings fall within the shaded area. With this simplistic explanation, we turn to the words that relate to precision. Accuracy
It is generally accepted that accuracy refers to the truthfulness of the instrument. An instrument of high accuracy more nearly gives a true reading than an instrument of low accuracy. Accuracy, then, is the quality of the instrument. It is common to refer to a measurement as accurate or not, and we understand what is meant. However, the current position is that accuracy should be used as a qualitative term and that no numerical value should be attached to it. It is, therefore, incorrect to refer to
INTRODUCTION
a measurement's accuracy of, say, 1%, when, presumably, this is the instrument's measurement uncertainty, as is explained later. Repeatability
In a process plant, or other control loop, we may not need to know the accuracy of a flowmeter as we would if we were buying and selling liquid or gas, but we may require repeatability within bounds defined by the process. Repeatability is the value below which the difference between any two test results, taken under constant conditions with the same observer and with a short elapsed time, are expected to lie with 95% confidence. Precision
Precision is the qualitative expression for repeatability. It should not take a value and should not be used as a synonym for accuracy. Uncertainty
Properly used, uncertainty refers to the quality of the measurement, and we can correctly refer to an instrument reading having an uncertainty of ±1%. By this we mean that the readings will lie within an envelope ± 1 % of the true value. Each reading will, of course, have an individual error that we cannot know in practice, but we are interested in the relationship of the readings to the true value. Because uncertainty is referred to the true value, by implication it must be obtained using a national standard document or facility. However, because it is a statistical quantity, we need also to define how frequently the reading does, in fact, lie within the envelope; hence the confidence level. Confidence level
The confidence level, which is a statement of probability, gives this frequency, and it is not satisfactory to state an uncertainty without it. Usually, for flow measurement, this is 95%. We shall assume this level in this book. A confidence level of 95% means that we should expect on average that 19 times out of 20 (19/20 = 95/100 = 95%) the reading of the meter will fall within the bracket specified (e.g., ± 1 % of actual calibrated value). Linearity
Linearity may be used for instruments that give a reading approximately proportional to the true flow rate over their specified range. It is a special case of conformity to a curve. Note that both terms really imply the opposite. Linearity refers to the closeness within which the meter achieves a truly linear or proportional response. It is usually defined by stating the maximum deviation (or nonconformity, e.g., ± 1 % of flow rate) within which the response lies over a stated range. With modern signal processing, linearity is probably less important than conformity to a general curve. Linearity is most commonly used with such meters as the turbine meter.
1.4
A BRIEF REVIEW OF THE EVALUATION OF STANDARD UNCERTAINTY
Range and Rangeability
An instrument should have a specified range over which its performance can be trusted. Therefore, there will be upper and lower range values. This reflects the fact that probably no instrument can be used to measure a variable when there are no limitations on the variable. Without such a statement, the values for uncertainty, linearity, etc., are inadequate. The ratio of upper range value and lower range value may be called the rangeability, but it has also been known as the turndown ratio. The difference between upper and lower or negative range values is known as span. It is important to note whether the values of uncertainty, linearity, etc., are a percentage of the actual flow rate or of the full-scale flow [sometimes referred to as full-scale deflection (FSD), full-scale reading (FSR), maximum scale value, or upper range value (URV)].
1.4 A BRIEF REVIEW OF THE EVALUATION OF STANDARD UNCERTAINTY Kinghorn (1982) points out the problem with terminology in matters concerning statistics and flow measurement. To the engineer and the statistician, words such as error and tolerance may have different meanings. The word tolerance was used for what is now known as uncertainty. In providing an introduction to the terminology of uncertainty in measurement, I shall aim to follow the guidance in BIPM et al. (1993), which is usually known as the Guide, and also in a document consistent with the Guide, which provides the basis for uncertainty estimates in laboratories accredited in the United Kingdom (NAMAS 1997). The reader is strongly advised to consult this document, which is full of clear explanations and useful examples. Those wishing to pursue background arguments are referred to van der Grinten;s (1994, 1997) papers and other papers in the proceedings of the 1997 North Sea Flow Measurement Workshop. Random error, the random part of the experimental error, causes scatter, as the name suggests, and reflects the quality of the instrument design and construction. It is the part that cannot be calibrated out, and the smaller it is, the more precise the instrument is. It may be calculated by taking a series of repeat readings resulting in the value of the standard deviation of a limited sample n, and sometimes called the experimental standard deviation:
J
/2
where q is the mean of n measurements q}. The experimental standard deviation of the mean of this group of readings is given by ^
/n
(1.2)
Where too few readings have been taken to obtain a reliable value of s(g;), an earlier calculation of s{q{) from previous data may be substituted in Equation (1.2). In obtaining the overall uncertainty of a flowmeter or a calibration facility, there will be values of group mean experimental standard deviation for various quantities, and
INTRODUCTION
so NAMAS (1997) defines a standard uncertainty for the /th quantity as (1.3) where xt is one of the input quantities. For those with access to NAMAS (1997), this is, essentially, dealt with there as a Type A evaluation of standard uncertainty. Systematic error, according to flowmeter usage, is that which is unchanging within the period of a short test with constant conditions. This is, essentially, dealt with in NAMAS (1997) under the heading Type B evaluation of standard uncertainty. It is also called bias. However, in modern flowmeters and in calibration facilities, it is likely that this bias or systematic error will result in a meter adjustment, or a rig correction. The resulting uncertainty in that adjustment or correction will contribute to the overall uncertainty. The systematic uncertainty, therefore, may derive from various factors such as a. b. c. d. e.
uncertainty in the reference and any drift, the equipment used to measure or calibrate, the equipment being calibrated in terms of resolution and stability, the operational procedure, and environmental factors.
From these we deduce further values of u(xt). There has been debate about the correct way to combine the random and systematic uncertainties. We can combine random and systematic uncertainties conservatively by arithmetic addition. This results in a conservative estimate. NAMAS (1997) has followed the Guide in taking the square root of the sum of the squares of the standard uncertainties in consistent units. Thus the combined standard uncertainty is My) = y/VlcMxi)]*
(1.4)
where y is the output quantity. To ensure consistent units, a sensitivity coefficient Ci will be required for each input xif although in practice this may be unity in most cases (as in Figure 4.3). The final step (and we have glossed over many important details in NAMAS 1997) is to deduce from uc the bracket within which the reading of, say, the meter lies. In the past, bearing in mind that uc or its components have been derived from standard deviations, we have used Student's t value which for a number of readings n is given by n
t
10
2 .26
20
2 .09
>30
2 .0
for a 95% confidence level. The Guide replaces this, in general, with a coverage factor k, to obtain the expanded uncertainty U = kuc(y)
(1.5)
1.6 WHAT IS A FLOWMETER?
The recommended value is k = 2, which gives a confidence level of 95.45% taken as 95%, assuming a Normal distribution. If this assumption is not adequate, then we need to revert to Student's t. The net result is that the assumption of a factor of 2 has now been given a systematic basis. The reader who is interested in more details about the basis of Normal and t distributions is referred to Appendix l.A.
1.5
SENSITIVITY COEFFICIENTS
Suppose that output quantity, a flow rate, has the relationship y = xplxq2xr3xs4
(1.6)
then if x2, x3f and *4 are held constant, we can differentiate y with respect to X\ and obtain the partial derivative. This is the slope of the curve of y against X\ when the other variables are kept constant. It also allows us to find the effect of a small change in Xi on y. This slope (or partial derivative) is the sensitivity coefficient cx for xx and may be found by calculation. It will have the value cx — px[p~r]x\x\x\t where the values of xx, Xz, X3, and X4 will be at the calibration point and may be dimensional. In some cases, it may be a known coefficient (e.g., a temperature coefficient of expansion). For cases where it is difficult to calculate, it may be possible to find the coefficient by changing xx by a small amount and observing the change in y. In some cases, the sensitivity coefficient may provide a conversion between different sets of units (e.g., where output quantity or velocity may be obtained from a dimension, a pressure, a movement, or a voltage).
1.6
WHAT IS A FLOWMETER?
We take as a working definition of an ideal flowmeter: A group of linked components that will deliver a signal uniquely related to the flow rate or quantity of fluid flowing in a conduit, despite the influence of installation and operating environment. The object of installing a flowmeter is to obtain a measure of the flow rate, usually in the form of an electrical signal, which is unambiguous and with a specified expanded uncertainty. This signal should be negligibly affected by the inlet and outlet pipework and the operating environment. Thus the uncertainty of measurement of a flowmeter should be reported as y ± U, where U, the uncertainty band, might have a value of, say, 0.5%, and it should be made clear whether this is related to rate, full-scale deflection (FSD) or other value that might be a combination of these [e.g., in the form ±a (rate) ±b (FSD)]. The range should be given (e.g., Im 3 /h to 20m 3 /h). The statement of performance should include the coverage factor k = 2 and the level of confidence of approximately 95%, and, if appropriate, the authority that accredited the calibration facility (national or international).
10
INTRODUCTION
In addition, the ranges of properties for which it can be used should be specified, such as fluid, flow range (beyond calibration), maximum working pressure, temperature range of fluid, and ambient temperature range. It is useful to introduce two factors that define the response of flowmeters, although they are most commonly used for linear flowmeters with pulse output. The K factor is the number of pulses per unit quantity. In this book, we shall take it as number of pulses per unit volume when dealing with turbine and vortex meters: Pulses True volume whereas the meter factor is usually defined as Meter factor =
True volume Indicated volume
The reader should keep a wary eye for other definitions of meter factor such as the reciprocal of the K factor. Let us take a specific example of a fictitious, but reasonably realistic flowmeter. In Figure 1.2(a), a typical flowmeter envelope is shown. It defines an approximately linear flowmeter with a 10:1 turndown and an uncertainty of ± 1 % of rate with a confidence level of 95% against a traceable standard calibration. This is a reasonable performance for a flowmeter and probably satisfies most requirements in industry. This, let us assume, is the performance specification the manufacturer carries in its sales literature. Actually the characteristic of the flowmeter may be the curve shown in Figure 1.2(a). If the company works to a high standard of manufacture, then the company may know that this characteristic lies within close tolerances in all cases. It may, therefore, only be necessary for the manufacturer to calibrate each flowmeter at, say, 90% of FSD, or 50% and 90% of FSD, in order to make the claim that the characteristic falls within the envelope specified in the sales literature. If the meters are actually of this standard, it may well be feasible to calibrate them in much greater detail so that a 5-, 10-, or even 20-point calibration may provide a characteristic that ensures that the reading is known to, say, ±0.2%. The values obtained from the calibration will then be programmed into a flow computer, which will interpret each reading of the flowmeter against this look-up table. Since we are comparing the flowmeter's signal to a linear one, if it were without error it would also be linear. Consequently, companies sometimes record linearity within their literature. In this case, it would also be ± 1 % with a 10:1 turndown. The envelope that has been discussed gives the uncertainty at each flow rate in terms of the actual flow rate. Because of the physical basis of some flowmeters, this method is not appropriate, and the uncertainty may then be given in terms of the full scale. Figure 1.2(b) shows such an envelope where the performance of the flowmeter would be defined as ± 1 % FSD. It is apparent that the uncertainty in the flowmeter's reading at, say, half scale is ±2% of rate and at 20% of reading will be as much as ±5% of rate. The problem often arises wherein the user has a particular flow range that does not match that of the actual instrument. The user's full flow may only be at 60% of the instrument's range, and so for the user the instrument has an uncertainty, at best, of 1.7% of rate.
1.6 WHAT IS A FLOWMETER? Uncertainty in reading
FLOWMETER PRECISION ENVELOPE
-
(a)
50%
_
iqo% FULL SCALE
FLOWMETER CHARACTERISTIC RELATIVE TO IDEAL
-2%
Uncertainty in reading
, FLOWMETER PRECISION ENVELOPE
-5% -
Uncertainty in reading FLOWMETER PRECISION ENVELOPE ^MINIMUM
(c)
2% -2%
^TRANSITION
20%
^NORMAL
50%
100% FULL SCALE
Figure 1.2. Required envelope for a flowmeter. (a) Envelope as a percentage of rate, (b) Envelope as a percentage of FSD. (c) Stepped envelope with increased uncertainty at low range values.
A third type of envelope is shown in Figure 1.2(c). This is particularly common in the specifications for water and gas meters. In the example shown, the meter has an uncertainty of ±2% of rate from full flow down to 20%. Below this value of flow rate, the uncertainty is ±5% of flow rate down to 2% of range. In practice, a meter might have more steps in its envelope. In many cases, the specification of uncertainty by the manufacturer may be a combination of these. As indicated earlier, it is common to have an uncertainty that combines a value based on rate and another on full-scale deflection. In addition, there may be allowances to be added for zero drift, temperature change, and, possibly, even pressure change. In some flowmeters, viscosity is important but is probably accommodated by charts showing the variation in performance with viscosity. One note of caution! Clever electronics can take any signal, however nonlinear, and straighten the characteristic before the signal is output. Suppose such a procedure were used for the characteristic in Figure 1.2(a), below 10% of range.
11
12
INTRODUCTION
The characteristic is probably very sensitive to minor variations in this region, and any attempt to use the characteristic could lead to disguised, but serious, errors. Variation of temperature and pressure can have effects on the performance of a flowmeter, as can humidity, vibration, and other environmental parameters (Baker 1988/9). Often the units used in a manufacturer's catalog are not those that you have calculated. For this reason, conversion factors that provide conversions to four significant figures of flow rate, velocity, temperature, pressure, length, etc., have been included in Table 1.1. If not otherwise specified the International System of units (SI) based on meter, kilogram, second is assumed.
Table 1.1. Conversion for some essential measurements from Imperial, U.S., and other units to metric, to four significant figures 1 ina = 25.4 mm 1 ft = 0.3048 m Volume 1 ft3 = 0.0283 m3 1 ft3 = 28.321 (liter) lbbl (barrel) = 0.1590 m3 Mass l l b = 0.4536 kg 1 long ton (2,240 lb) = 1,016 kg 1 short ton (2,000 lb) = 907.2 kg 1 metric tonne (2,205 lb) = 1,000 kg Density 1 lb/ft3 = 16.02 kg/m3 Temperature (Temperature in°F — 32)/1.8 = Temperature in °C Pressure 1 psi = 6,895 N/m2 Viscosity Dynamic viscosity: SI (metric) unit is the Pascal second (Pas) to which the more common unit, the centipoise (cP), is related by 1 cP = 10~3 Pas. Kinematic viscosity: SI (metric) unit is m2/s to which the more common unit, the centistoke (cSt), is related by 1 cSt = 10~6 m2/s = 1 mm2/s. Velocity lft/s = 0.3048 m/s Volumetric flow rate 1 ft3/s (1 cusec) = 0.02832 m3/s 1 Imp gal/s = 0.004546 m3/s 1 Imp gal/s = 4.546 1/s 1 U.S. gal/s = 0.003785 m3/s 1 U.S. gal/s = 3.785 1/s 1 Imp gal/h = 0.004546 m 3 /h 1 Imp gal/h = 4.546 1/h 1 U.S. gal/h = 0.003785 m 3 /h 1 U.S. gal/h = 3.785 1/h Mass flow rate 1 lb/h = 0.4536 kg/h 1 lb/s = 0.4536 kg/s Length
a
An approximate conversion is 4 in. to 100 mm.
1.7 CHAPTER CONCLUSIONS
1.7 CHAPTER CONCLUSIONS (FOR THOSE WHO PLAN TO SKIP THE MATHEMATICS!) I have tried to bring together, within the scope of this book, essential information for all who may have dealings with flowmeters and flow measurement (although I have tried to avoid duplicating too much of the excellent and practical information given by Miller 1996). For this reason, the chapters not only address the technical aspects but also the selection, maintenance, calibration, and typical applications of the various meters. There is also a section on manufacture and production. I hope that the book will provoke the prospective entrepreneur, the small and medium-sized enterprises (SME), or the major instrument company to assess the market needs and the relevant development and production needs of their companies for new devices. One plea regarding manufacturers' brochures is in order. It would be helpful to have a standard format for the core information to which the manufacturer could add special features and presentation to reflect customer demands (e.g., from different countries and industries). I often find myself faced with two catalogs from the same manufacturer. I search them to find how they differ. They clearly are printed differently, and yet it is not clear whether they refer to different meters. Do just the catalogs differ, or is there a difference between the flowmeters described in each catalog? This may seem an unlikely problem, but the preparation of this book confirms that confusing catalogs are still with us. The corollary of the last plea (for a readable and complete brochure) is that the manufacturer needs to be told everything about the fluid and the installation in order to decide whether the meter can handle the requirements. We shall return to this point in much greater detail in Chapter 3. I have endeavored to reflect the requirements of standard documents, but the reader is encouraged to obtain relevant standards. A partial list is included together with a bibliography at the end of this book. Standards organizations and professional bodies have produced a range of essential documents that cover many aspects of flow metering. Upp (1993) provides a useful digest of some of the documents such as API's Manual of Petroleum Measurement Standards, AGA documents which for many flow measurement applications (e.g., Section 14.3) are the same as API, and ASME and ISA. Another example from the United Kingdom is IGE (1988), which recommends standards of good practice and provides an extremely thorough statement on gas meter installation, which, although not carrying the wider backing of a standard, is clearly very authoritative and should provide a companion volume. Such topics as commissioning and maintenance are covered, and the manual encourages procedures for routine maintenance and record keeping as do I (cf. Olsen 1974, Kulin 1975). As the reader moves into the following chapters, two sets of information may be useful. Table 1.1 lists the conversion factors for Imperial, U.S., and metric units, and Table 1.2 relates volumetric and mass flow rate to linear velocity in various sizes of tube (Baker 1988/9). It is common in flow measurement to require the velocity of flow, and Table 1.2 provides an order of magnitude. Finally, the whole matter of accuracy and the limits of accuracy, when related to all the parameters that influence a flowmeter's operation, remains an area with unanswered questions.
13
Table 1.2. Velocity in pipes for various flow rates to two significant figures 3
m /h° 1/min Very low
10~ lO-
3
2
0.1 1 10 100
1000 104 105
Very high
10 6
gal/min
0.017 0.17 1.7 17 170
1700 1.7 x 10 4 1.7 x 10 5 1.7 x 10 6 1.7 x 10
7
3.7 x lO" 3.7 x 10~ 0.37 3.7 37 370 3700 3.7 x 10 4 3.7 x 10 5
2
3.7 x 10
ft /min
gal/min 3
6
3
4.4 x 10~
3
4.4 x 10" 0.44 4.4 44 440 4400 4.4 x 10 4 4.4 x 10 5
2
4.4 x 10
6
10 mm
5.9 x lO-
5.9 59 590
25 mm
3.5 x 10"
3
2
3.5 x 10~ 0.35 3.5 35 350
5900 5.9 x 10 4
3500 3.5 x 10 4 3.5 x 10 5
5
6
5.9 x 10
3
4
5.9 x 10" 5.9 x 10" 2 0.59
3.5 x 10
Mean Velocity (m/s) in a Circular Pipe of Diameter 500 mm 50 mm 100 mm 200 mm 1000 mm
5.7 x lO-
4
3
1.4 x lO-
4
3.5 x 10
3
3.5 x 10 -4 3.5 x 10 - 3 3.5 x 10 -2 0.35
5.7 x lO" 5.7 x 10"2 0.57
1.4 x 10" 1.4 x 10" 2 0.14
5.7 57 570
1.4 14 140
3.5 35 350
2000 mm
-5
8.8 x lO- 5 8.8 x lO- 4 8.8 x l O - 3 8.8 x 10" 2 0.88
1.4 x lO- 5 1.4 x 10"4 1.4 x 10~ 3 1.4 x 10" 2 0.14
8.8 88 880
1.4 14 140
3.5 x lO- 5 3.5 x lO- 4 3.5 x l O - 3 3.5 x 10" 2 0.35 3.5 35 350
8.8 x 8.8 x 8.8 x 8.8 x 0.88
lO-5 lO- 4 lO-3 lO-2
8.8 88
Reproduced from Baker (1988/9) with permission of Professional Engineering Publishing. a Since water has a density of 1,000 kg/m3 (approximately), the mass flow rate in kilograms per hour of water may be obtained by multiplying this column
1.A.1 INTRODUCTION
1.8
MATHEMATICAL POSTSCRIPT
I have left this note to the end so that those who are not concerned with advanced mathematical concepts can ignore it. I have included essential mathematics in the main text of the book. In certain flowmeters, the mathematical theory is more complex (e.g., the turbine meter), and the theory has, accordingly, been consigned to an appendix after the relevant chapter. One important and interesting mathematical approach, which starts to develop a unified theory of flow measurement, was first suggested by Shercliff (1962) and significantly extended by Bevir (1970). Both applied it to electromagnetic flowmeters where it has been highly successful. Hemp (1975) has also applied this theory to electromagnetic flowmeters, but he has developed the theory for other types of flowmeter: ultrasonic (1982), thermal mass (1994a), and Coriolis (1994b, and Hemp and Hendry 1995). In Chapter 12, on electromagnetic flowmeters, an appendix describes the essential mathematics. The weight function developed in this theory provides a measure of the importance of flow in each part of the meter with respect to the overall meter signal. The flow at each point of a cross-section is weighted with this function. Ideally the weighting should result in a true summation of the flow in the meter to obtain a volume flow rate. It has been possible to approach this ideal for the electromagnetic flowmeter. For the other types of meters, the reader will be given only a brief explanation and will be referred to relevant papers. This is partly for want of space in this book and partly because the theory, although interesting, is still being developed and may not yet have reached a sufficiently significant stage for the designer. A second mathematical physics theory, first (to my knowledge) applied by Hemp (1988) to flow measurement, is reciprocity. This, essentially, states that if you apply a voltage to one end of an electrical network and measure the current at the other end, you find that by reversing the ends and hence the direction you obtain the same relationship. Hemp has proposed this as a means of eliminating some errors in flowmeters to which the theory is applicable.
APPENDIX
1.A
Statistics of Flow Measurement 1.A.1
INTRODUCTION
The main needs of the engineer are to • understand and be able to give a value to the uncertainty of a particular measurement; • know how to design a test to provide data of a known uncertainty; • be able to combine measurements, each with its own uncertainty, into an overall value; and
15
16
INTRODUCTION
• determine the absolute accuracy of an instrument at the end of a traceable ladder of measurement. The international and national documents set the recommended approach for flow metering. Most standard statistics books will provide the essentials (Rice 1988; cf. Campion et al. 1973, which is often quoted but may not be easy to obtain), but good school texts may be more accessible (Crawshaw and Chambers 1984, Eccles et al. 1993a, 1993b). Hayward (1977) is an extremely well written and elegant little book, which deserves to be updated and reprinted; Kinghorn (1982) provides a wellwritten and useful brief review of the main points; and Mattingly (1982) addresses some of the problems concerned with transfer standards.
1.A.2
THE NORMAL DISTRIBUTION
The Normal distribution, Figure l.A.l(a), is also known as the Gaussian distribution after Carl Friedrich Gauss, who proposed it as a model for measurement errors (Rice 1988). The notation used for the Normal curve is N(/x, a 2 ) which is the distribution under the curve (l.A.l)
f(*) =
where \x is the mean value of the data, and a 2 is the variance. Alternatively, a is the standard deviation for the whole population. We can simplify the curve (normalize it) by putting z = (x — fi)/a and obtaining [Figure l.A.l(b)] (0.5) = 0.6915
PROBABILITY OF READING LYING IN UNSHADED AREA = 0.3085
SHADED AREA - 0.950 ARE A =0.025
AREA - 0.025
Figure l.A.l. The Normal distribution.
0(z) =
1
(1.A.2)
With the form of Equation (1.A.2), the curve does not vary with the size of the parameters fi and o. What the curve tells us (in relation to instrument measurements) is that the statistical chance of an instrument reading giving a value near to the mean /x is high, but the farther away the reading is from the mean, the less the chance is of its occurring (indicated by the curve decreasing in height the further one moves from the mean), and as values of the reading get farther still from li, so the chance gets less and less. The area under the curve of Equation (1.A.2) [Figure l.A.l(b)], which reaches to infinity each way, is unity, and this is the
1 .A.3 THE STUDENT t DISTRIBUTION
probability of the reading lying within this curve (obviously). The area under the curve between z = —oo and some other value of z is given by
= —f
dt
(1.A.3)
y/2jT J-o
17
Table l.A.l. A selection of values from the normal distribution function (z) z
3Hz)
0 0.5
0.5000
Symmetrical Central Area Under Curve
0.6915 and is the probability that a reading will lie 0.8997 1.28 within that range and is obtained numeri1.282 0.9000 0.80 cally and given in Table l.A.l in normalized 1.29 0.9015 form. For instance, if z = 0.5, (z) = 0.6915, 1.64 0.9495 where (z) is the area under the curve from 1.645 0.9500 0.90 z = -oo to z = 0.5 in this case. So the 1.65 0.9505 chance of a reading lying beyond this point 1.96 0.9750 0.95 is 0.3085, or about 30%. 2.57 0.99492 We shall be interested in the chance that 2.576 0.99500 0.99 a flowmeter reading will fall between cer2.58 0.99506 tain limits each side of z = 0, the mean 3.29 0.99950 0.999 value. A chance of 95% is often used and is 0.99952 3.30 called a 95% confidence level. This means After D. V. Lindley and W. F. Scott, New Camthat 19 times out of 20 the reading will fall bridge Statistical Tables, 2nd ed., Cambridge: Cambridge Univ. Press. Table 4, pp. 34, 35. between the limits. This requires that the central area of the curve [Figure l.A.l(c)] has a value of 0.95, or 0.475 each side of the mean. To obtain z from this value, we need to add 0.475 + 0.5 = 0.975, and this gives a value (Table l.A.l) of z = 1.96. If we put this in terms of x, we obtain
x-ti
=
1.96a
(1.A.4)
or the band around the mean value of the reading within which 95% of the readings statistically should fall, is approximately ±2a, or two standard deviations from the mean. If we are interested, not in the spread of individual readings, but in the spread of the mean of small sets of readings, a statistical theorem called the Central Limit Theorem provides the answer. If a sample of n readings has a mean value of M, then the distribution of means like M is given by N(^, a2/n). This is intuitively reasonable because one would expect that the scatter of means of groups of n readings would have a smaller variance, o1In, than the readings themselves, as well as a smaller standard deviation, cr/y/n. In this discussion, we have skated over the need to know the value of the standard deviation of the whole normal population. If we do not know o, then we can approximate it with the value of the standard deviation s of the small set of n. So if n > 30, it is usually sufficiently precise to take a — s. If n < 30, the standard deviation should be taken as a = s^Jnj^Jn — 1.
1.A.3
THE STUDENT t DISTRIBUTION
We now need to look at one more subtlety of these estimates. The Normal distribution assumed that we had obtained many readings and could with confidence know
18
INTRODUCTION
that they formed a Normal distribution. We can agree that if the error is random, then it is a fair assumption that many readings would form a Normal distribution. However, often we have only a few readings, and these may not be uniformly distributed within the curve of Figure l.A.l. Too many may lie outside the 1.96 a limit. For this reason, we use the Student t distribution, -4 -3 -2 -1 which allows for small samples on the as1.96 2.23 sumption that the distribution, as a whole, Figure l.A.2. Student's t distribution curves com- is Normal. Figure l.A.2 shows the effect of pared with the Normal curve. Note p = 5% as related the small number of readings. Since, with to Table l.A.2 for both tails. a small number of readings, one has to be subtracted from all the others to obtain a mean, the number of independent values is one less than the number of readings, and so the statisticians say that there is one less degree of freedom than the number of readings. In Figure l.A.2, v is the symbol for the degree of freedom, and v = n - 1, where n is the number of readings. For v -* oo, the t distribution tends to a Normal curve with mean zero and variance unity. Figure l.A.2 shows clearly the larger area spreading beyond the Normal curve in which the readings may lie and the reason for the greater uncertainty. The curves are used in a similar way to the Normal curve, but, as an alternative, Table l.A.2 provides the information we need. If we have 10 readings, say, and so select the value of v = 9 for the degree of freedom, and if we wish to find the limits for a confidence level of 95%, we shall need to use the 5% column. We obtain t = 2.262, which we can apply to obtain the limits for a 95% confidence of ±2.262 a/y/n on the mean values of groups of readings, where n is the number of readings in the group. We should
Table l.A.2. A selection of values from the student t function p(%) (Total) 20
p/2 (%) n
V
10
5
1
5
10
2 3 5 10
1 2 4 9
3.078 1.886 1.533 1.383
6.314 2.920 2.132 1.833
20 30 61
19 29 60
121 oo
120 oo
1.328 1.311 1.296 1.289 1.282
0.1
(per tail)
2.5
0.5
0.05 636.6
12.71
63.66
4.303
9.925
31.60
2.776
4.604
8.610
2.262
3.250
4.781
1.729 1.699 1.671
2.093
2.861
3.883
2.045
2.756
3.659
2.000
2.660
3.460
1.658
1.980
2.617
3.373
1.645
1.960
2.576
3.291
After D. V. Lindley and W. F. Scott, New Cambridge Statistical Tables, 2nd ed., Cambridge: Cambridge Univ. Press, Table 10, p. 45.
1.A.4
PRACTICAL APPLICATION OF CONFIDENCE LEVEL
note, however, that the 95% confidence level from Table 1.A.2 gives a t value that varies little from 2.0 if v > 20. The limits for 95% confidence will then be ±2.0 o/ Jn on the mean values of groups of readings. I have always been puzzled by the name Student, but Eccles et al. (1993b) explain that the originator of this technique was William S. Gosset, born 1876, who used the pseudonym Student. 1.A.4
PRACTICAL APPLICATION OF CONFIDENCE LEVEL
The method described in Section 1.4 leads to the following steps (cf. Hayward 1977): i. Write down systematic uncertainties and derive the standard uncertainty for each component, ii. Write down random uncertainties and derive the standard uncertainty for each component, iii. Calculate the combined standard uncertainty for uncorrelated input quantities (and refer to NAM AS 1997 if correlated), iv. Obtain the expanded uncertainty using k = 2 for 95% confidence. Taking a simple example, where we need to revert to t, suppose that we obtain a series of volumetric flow readings from a 50 mm ID flowmeter with the flow set at 10m 3 /h: 10.06, 10.01, 9.95, 9.99, 9.85, 10.02,10.03,10.12, 9.90, 9.98. The results are plotted in Figure l.A.3. The mean of these readings is 9.991, and the standard deviation is 0.07752. We could thus conclude that the true reading of this meter fell in the bracket 9.991 ± 2.262 x 0.07752/V9 = 9.991 ± 0.05845, or between 9.93 and 10.05 with a 95% confidence. This fairly brackets the value of 10 m 3 /h. The actual readings should have all fallen within 9.991 ± 0.07752, or 9.91 and 10.07. In fact three fell outside this bracket - rather higher than the 1 in 19 implied by the 95% confidence level. We might wish to look more closely at the procedure m3/h 10.15 - 10.10 - X
10.05 + 10.00 9.95 - 9.90 -
X
:
9.85 +
'
X
1
1
1
1
1
1
1
1
1
|—TIME
SUCCESSIVE CALIBRATION RUNS
Figure l.A.3. A set of flowmeter test readings (after Kinghorn 1982) for a fixed flow rate.
19
20
INTRODUCTION
for obtaining these results since this suggests a possible problem with the means for obtaining the data. 1.A.5
TYPES OF ERROR
There are essentially four types of error (Kinghorn 1982). • Spurious errors result from obvious failures, obvious in the sense that they can be identified and documented. Readings with these should be eliminated. • Random errors cause a variation in the output reading even when the input parameter has not changed. • Constant systematic error, which is also called bias, may vary over the range but is constant in time, and could, in principle, be corrected out of the reading. • Variable systematic error (bias) slowly varies with time, usually in a consistent direction, and may be caused by wear in bearings of a rotating meter, fatigue in components of a vibrating meter, erosion of geometry, etc. Figure 1.A.4 illustrates these errors. Clearly one of the readings is so far out that there must be some explanation other than randomness. It is comforting to know that some of the most eminent experimentalists of the past have had cause to discard readings in critical experiments! The scatter around the mean line will provide the basis of the calculation which we did in Section l.A.4. The constant systematic error (bias) can be seen and could be built into a flow computer. The change in the mean value with time shows the changing systematic error, which is, in part, the reason for regular recalibration of meters. The repeatability is related to the closeness of readings. If we expect a reading to lie within a band given by ±2s, the worst case difference between successive readings that fall within this band would be 45 [(2 x 2)5], but a less extreme working value is obtained from the root of the sum of the squares (or the quadrature; cf. Pythagoras and the length of the hypotenuse): J{2s)2 + (25)2
or
2V25 OUTLIER
RANDOM ERROR
T BIAS
4^
SPURIOUS ERROR TRUE VALUE
MEAN - MEASURED VALUE
X
NOTE DRIFT OF MEAN CAUSING A CHANGE IN BIAS
RANDOM UNCERTAINTY
TIME
Figure l.A.4. Diagram to show the various types of error (after Kinghorn 1982).
21
1.A.7 UNCERTAINTY RANGE BARS, TRANSFER STANDARDS, AND YOUDEN ANALYSIS
1.A.6
COMBINATION OF UNCERTAINTIES
If we combine uncertainties due to the nature of a flowmeter;s operating equation, then we take the following approach. Suppose that the flowmeter has the equation n v vnv Lit) — ^ \ ^0 ^3
/-i A r\ I 1«A« J I
vm
A
To obtain the uncertainty in qv, we need the partial derivative of qv with respect to X\, x2, etc. The required result can be achieved, either by differentiating the equation as it stands or by first taking logarithms of both sides. We shall skip this and go straight to the result: nu(x2)
uc(qv)
±
u(x3)
±
mu(x4)
(1.A.6)
The problem with this equation is that the arithmetic sum of the uncertainties is usually overpessimistic. It is, therefore, recommended that they be combined in quadrature, or by the root-sum-square (rss) method. This leads to the following equation:
(nu{x2)\
\
|
{ x2 )
\ x3
|V^V
(1.A.7)
FLOW RATE
There are complications beyond this equation. x2 may appear in the equation as Xs + *6- In this case, x$ + Xe will need to be dealt with first and will require careful consideration as to whether the actual errors in these quantities are combining, canceling, or random.
TRUE ' VALUE
(a)
1.A.7 UNCERTAINTY RANGE BARS, TRANSFER STANDARDS, AND YOUDEN ANALYSIS It is sometimes useful to indicate the range of uncertainty that has been estimated from the experimental method in each reading. This can be done by using bars that give uncertainty limits on each experimental point. This will then indicate, for a particular flow rate, the likely uncertainty in the reading. This is shown in Figure l.A.5(a). In some cases where flow rate varies, there may be an uncertainty in both primary flow rate measurement and reading of second meters. In this case, there will need to be uncertainty bars in both
TIME
SECONDARY METER READING
(b)
PRIMARY METER READING
Figure l.A.5. (a) Uncertainty limit bars on readings. (b) Uncertainty limit bars on both flow rate and readings.
22
INTRODUCTION 71.7r
71.6 = 71.5445 71.5
71.4
|
MAY MAY JUN SEP NOV DEC JAN FEB FEB 3 | 20 | 7 | 28 | 17 | 21 | 6 | 8 I 4 | DATE
(a) 1.005r
1.004
r
85 1.00372
T 2cr
1.003
1.002
MAY MAY JUN SEP NOV DEC JAN FEB FEB 1-3 1 20 1 7 1 28 | 17 I 21 | 6 I 2 I 4 I DATE
(b)
METER FACTOR
MTRN©1
(C)
Figure l.A.6. Turbine meter as a transfer standard (from Mattingly 1982; reproduced with the author's permission), (a) Typical turbine meter control chart for meter factor, (b) Typical turbine meter control chart for ratio, (c) Graphical representation of the Youden plot.
1.A.7 UNCERTAINTY RANGE BARS, TRANSFER STANDARDS, AND YOUDEN ANALYSIS
directions, and the rectangle [Figure l.A.5(b)] will define the limits of the possible uncertainty. In other words, the maximum uncertainty will have been obtained by quadrature. Mattingly (1982) describes the procedure for checking the validity of different flow measurement laboratories using a Measurement Assurance Program (MAP) [or Proficiency Testing as referred to by Mattingly in a recent draft report on the approach of the National Institute of Standards and Technology (NIST)] where a goodquality flowmeter acting as a transfer standard is exchanged between laboratories. Figure l.A.6(a) shows the results of such a cycle of checks, and the bars on the experimental points presumably indicate the uncertainty of the turbine meter at a particular laboratory. Various people have suggested the use of two flowmeters usually in series, but as a possible alternative in parallel, to enhance the accuracy of a transfer standard. In this case, the ratio of the signals from the two meters will give an indication as to whether there has been a shift in either, and, if there has not been, the reading of the meters will provide, with greater confidence, the accuracy of the facility. In Figure l.A.6(b), Mattingly (1982) gave a typical control chart for the ratio of a pair of National Bureau of Standards turbine meters. If the ratio falls outside an agreed tolerance band, the cause of the error will need to be found before confidence in the meter readings is restored. For laboratory comparison, the transfer package with two meters should be run at one or two agreed flow rates. The position of the two flowmeters may be interchanged to obtain a second set of data. Using the Youden analysis in Figure l.A.6(c), each laboratory is represented by a single point (1, 2, 3, etc.) resulting from plotting the meter factor of the two meters on the two axes. It is apparent that the position of the points relates to the type of error. Essentially, if the points lie in quadrants I or III, then the meters are reading the same, and the error can be attributed to the flow rig. If the readings are in quadrants II or IV, the flowmeters are not agreeing, and the error may be due to a malfunction in one of them. The reader is referred to Mattingly's (1982) article for further details of this procedure (cf. Youden 1959).
23
CHAPTER S
Fluid Mechanics Essentials
2.1
INTRODUCTION
In a recent book (Baker 1996), I provided an introduction to fluid mechanics and thermodynamics, particularly aimed at instrumentation. I do not, therefore, propose to repeat what is written there but rather to confine myself to essentials. In addition, Noltingk (1988) has provided a very valuable handbook on general instrumentation. In this book, I shall use the term fluid to mean liquid or gas and will refer to either liquid or gas only when the more general term does not apply. 2.2 ESSENTIAL PROPERTY VALUES Flowmeters generally operate in a range of fluid temperature from — 200°C (-330°F) to 500°C (930°F), with line pressures up to flange rating for certain designs. Typical values of density and viscosity are given in Table 2.1. It should also be noted that liquid viscosity decreases with temperature, whereas gas viscosity increases with temperature at moderate pressures. In common fluids, such as air and water, the value of viscosity is not dependent on the shear taking place in the flow. These fluids are referred to as Newtonian in their behavior as compared with others where the viscosity is a function of the shear taking place. The behavior of such fluids, known as non-Newtonian, is very different from normal fluids like water and air. Newtonian fluid behavior is a good representation for the behavior of the bulk of fluids. 2.3
FLOW IN A CIRCULAR CROSS-SECTION PIPE
An essential dimensionless parameter that defines the flow pattern at a particular value of the parameter is the Reynolds number
[I
where p is the density of the fluid, \± is the dynamic viscosity, and, when applied to flow measurement, V is the velocity in the pipe, and D is the pipe diameter. Typical values of Reynolds number are: for water with /x/p = 10~6 m 2 /s, V =1 m/s (3.3 ft/s), D = 0.1 m (4 in.) Re= 105; and for air at ambient conditions with /x/p = 1.43 x 10- 5 m 2 /s, V = 10 m/s (33 ft/s), D = 0.1 m (4 in.) Re = 0.7 x 105. 24
2.3 FLOW IN A CIRCULAR CROSS-SECTION PIPE
Table 2.1. Typical values of viscosity (approximate values at 1 bar) Viscosity
Water Benzene Oxygen Nitrogen
co2 Air
Temperature (°C)
Density, 9 (kg/m3)
Dynamic,
McP)
Kinematic, v (cSt)
20 20 0 0 0 0
1,000 700 1.43 1.25 1.98 1.29
1.002 0.647 0.019 0.017 0.014 0.017
1.002 0.92 13.3 13.6 7.1 13.2
Adapted from Kaye and Laby (1966).
If Re is less than about 2,000, and the fluid has had a sufficient length of pipe to reach a steady state, it all moves parallel to the axis of the pipe. At the pipe wall, the fluid "sticks" to the pipe in, what is known as, a nonslip condition. The velocity of the fluid increases, therefore, from zero at the pipe wall to a maximum at the center, and the shape of the profile is parabolic. In this case, the flow is called laminar. At about Re = 2,000, a major change takes place, and the smooth parallel nature of the flow gives way to eddies in theflow.These eddies mix the high velocity at the pipe axis with the lower velocity near the pipe wall. The resulting profile is flatter (although it still goes to zero at the pipe wall). When fully developed, this is known as a turbulent profile. It is of a well-defined shape and with a known range of eddy sizes. It may, therefore, be misleading to use the term turbulent profile for any other profiles created by upstream disturbances due to bends, etc. Between the turbulent flow and the laminar flow, there is a transition when the flow alternates randomly, in space and time, between laminar flow and turbulent flow. Any flowmeter application to non-Newtonian fluids will need advice from the flowmeter manufacturer. It is probable that clear bore flowmeters, if applicable, will be most suitable. A flowmeter with relatively low sensitivity to profile, such as electromagnetic or multibeam ultrasonic designs, will be worth consideration provided the fluid is conducting for the first type or transmits an adequate level of sound for the second. It is apparent from the calculations of Reynolds number for water and air that, in the majority of industrial applications, theflowswill be turbulent, and, therefore, the behavior of the turbulent profile, and of flowmeters subject to it, is of primary importance to us. For laminar flow, we can use the equation V=V0
1 - [L.
(2.2)
where Vo is the velocity on the axis of the pipe, r is the radial point at which we are measuring velocity, and R is the pipe radius and where V is zero at the wall of the pipe because of the nonslip condition for a fluid at a solid boundary.
25
26
FLUID MECHANICS ESSENTIALS
Table 2.2. Approximate turbulent velocity profiles from equation (2.4) 2.0 x 106 to 3.2 x 106
Re
4 x 103
2.3 x 104
1.1 x 105
1.1 x 106
n V/V[air = 0.75R) V/V[a.tr =0.758/?)
6.0
6.6
7.0
8.8
10
1.003 0.998
1.004 0.999
1.004 1.000
1.005 1.002
1.005 1.002
The mean velocity in the pipe is then given by (2.3)
" 2 For turbulent flow an approximate curve-fit is V=
V0(l-r/R)l/n
(2.4)
where Vo is, again, the center line velocity, and we can relate n to Re from experimental data (Table 2.2). Note that pipe roughness, if the pipe is not hydraulically smooth, will also affect the profile. This expression, therefore, provides not only a convenient, although approximate, representation of the boundary layer shape but also an index adjustment to allow for the change in Reynolds number. The resulting profile shapes for the laminar and the turbulent regimes, based on Equation (2.4), are shown in Figure 2.1. One interesting feature of these turbulent profiles is that the ratio of the velocity at about the 3/4 radius point to the mean velocity is approximately unity. The mean velocity is given by V
In2
(2.5)
Hence V/V =
2n2
-a-r/R)\/n
(2.6)
Table 2.2 provides the values of this ratio for various values of Re and two radial positions. Thus if a single measurement of velocity in a pipe is to be used to obtain the mean of the velocity in a turbulent flow, a point at 0.7S8R may be found best. This figure will clearly be of importance when we consider where to position the probes described in Chapter 18.
LAMINAR Re < 2000
TURBULENT
Re = 2.3x10 Re =1.1x10 Figure 2.1. Laminar and turbulent pipe profiles.