Experimental Methods in the Physical Sciences VOLUME 40 CAVITY-ENHANCED SPECTROSCOPIES
EXPERIMENTAL METHODS IN THE PH...
25 downloads
1017 Views
16MB Size
Report
This content was uploaded by our users and we assume good faith they have the permission to share this book. If you own the copyright to this book and it is wrongfully on our website, we offer a simple DMCA procedure to remove your content from our site. Start by pressing the button below!
Report copyright / DMCA form
Experimental Methods in the Physical Sciences VOLUME 40 CAVITY-ENHANCED SPECTROSCOPIES
EXPERIMENTAL METHODS IN THE PHYSICAL SCIENCES Robert Celotta and Thomas Lucatorto, Editors in Chief FoundingEditors L. MARTON C. MARTON
Volume40
Cavity-Enhanced Spectroscopies Edited by Roger D. van Zee J. Patrick Looney
National Institute of Standards and Technology Gaithersburg, Maryland, USA
ACADEMIC PRESS An imprint of Elsevier Science Amsterdam Boston London New York Oxford San Francisco Singapore Sydney Tokyo
Paris
San Diego
This book is printed on acid-free paper. 9 2002 Elsevier Science (USA) All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, recording, or any information storage and retrieval system, without permission in writing from the Publisher. The appearance of the code at the bottom of the first page of a chapter in this book indicates the Publisher's consent that copies of the chapter may be made for personal or internal use of specific clients. This consent is given on the condition, however, that the copier pay the stated per copy fee through the Copyright Clearance Center, Inc. (222 Rosewood Drive, Danvers, Massachusetts 01923), for copying beyond that permitted by Sections 107 or 108 of the U.S. Copyright Law. This consent does not extend to other kinds of copying, such as copying for general distribution, for advertising or promotional purposes, for creating new collective works, or for resale. Copy fees for pre-2002 chapters are as shown on the title pages. If no fee code appears on the title page, the copy fee is the same as for current chapters. 1079-4042/01 $35.00 Explicit permission from Academic Press is not required to reproduce a maximum of two figures or tables from an Academic Press chapter in another scientific or research publication provided that the material has not been credited to another source and that full credit to the Academic Press chapter is given. Chapters 1 and 3 of this book were contributed by the United States government and are subject to Section 105 of the U.S. Copyright Law within the United States of America. Academic Press An Elsevier Science Imprint 525 B Street, Suite 1900, San Diego, California 92101-4495, USA http://www.academicpress.com International Standard Book Number: 0-12-475987-4 International Standard Serial Number: 1079-4042/01 Printed in Great Britain by MPG Books Ltd, Bodmin, Cornwall 02 03 04 05 06 07 MP 9 8 7 6 5 4 3 2
1
To the one-hundred-year-long tradition of accuracy and precision i n physical and chemical measurements at the National Bureau of Standards/ National Institute of Standards and Technology. " . . . when you can measure what you are speaking about, and express it in numbers, you know something about it; but when you cannot measure it, when you cannot express it in numbers, your knowledge is of a meagre and unsatisfactory kind," it may be the beginning of knowledge, but you have scarcely in your thoughts advanced to the state o f science . . . . " Lord Kelvin
This Page Intentionally Left Blank
CONTENTS CONTRIBUTORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xi
VOLUMES IN SERIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xiii
PREFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xvii
.
Stabilizing Diode Lasers to High-Finesse Cavities by RICHARD W. FOX, CHRIS W. OATES, and LEO W. HOLLBERG 1.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1
1.2. I n t r o d u c t i o n to Diode Laser Locking . . . . . . . . . . . . . . . . . . . . . . . .
2
1.3. G e n e r a t i n g the E r r o r Signal: The P o u n d - D r e v e r - H a l l Method .....................................................
5
1.4. The L o o p Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19
1.5. Locking the Laser and L o o p Optimization . . . . . . . . . . . . . . . . . .
29
1.6. Repetitive Locking for Cavity R i n g - D o w n Spectroscopy . . . .
37
1.7. Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
44
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
44
2. Spectroscopic Applications Using Ring-Down Cavities by GIEL BERDEN, GERARD MEIJER, and WIM UBACHS 2.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
47
2.2. 02: A B e n c h m a r k System for R i n g - D o w n Techniques . . . . . . .
47
2.3. Spectroscopic Applications of C R D
63
........................
2.4. C R D and Alternative Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . .
73
Acknowledgment ...............................................
74
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
74
3. Absorption Detection at the Quantum Limit: Probing High-Finesse Cavities with Modulation Techniques by JUN YE and JOHN L. HALL 3.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
83
3.2. Cavity E n h a n c e m e n t : A Simple Physics Picture . . . . . . . . . . . . .
86
vii
viii
CONTENTS 3.3. Signal E n c o d i n g a n d Extracting: Modulation-plus-Heterodyne Detection . . . . . . . . . . . . . . . . . . . . .
90
3.4. W e a k A b s o r p t i o n M e a s u r e d by Field D e c a y (Time-Domain) ............................................
95
3.5. W e a k A b s o r p t i o n M e a s u r e d by Field Phase (Frequency-Domain) ......................................
105
3.6. C o n c l u d i n g R e m a r k s on Signal E x t r a c t i o n a n d Cavity E n h a n c e m e n t . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
121
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
122
Cavity-Enhanced Absorption: Intracavity Laser Spectroscopy
,
by GEORGE H. ATKINSON 4.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
129
4.2. Principles of ILS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
132
4.3. D e v e l o p m e n t of ILS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
134
4.4. C o n c l u d i n g R e m a r k s . . . . . . . . . . . . . . . . . . . . . .
145
. ...............
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
146
The Continuous-Wave Hydrogen Raman Laser
.
by PETER A. ROOS, LEI S. MENG, a n d JOHN L. CARLSTEN
.
5.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
157
5.2. Semiclassical T h e o r y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
162
5.3. N d : Y A G - P u m p e d System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
166
5.4. D i o d e - P u m p e d Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
170
5.5. Practical C o n s i d e r a t i o n s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
177
5.6. S u m m a r y . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
181
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
181
2D Microcavities: T h e o r y a n d E x p e r i m e n t s by JENS U. NOCKEL a n d RICHARD K. CHANG 6.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
185
CONTENTS
.
ix
6.2. Dielectric Microcavities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
186
6.3. Whispering-Gallery M o d e s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
189
6.4. Scattering Resonances a n d Q u a s i b o n d States . . . . . . . . . . . . .
192
6.5. Cavity R i n g - D o w n and Light Emission . . . . . . . . . . . . . . . . . . .
195
6.6. Wigner Delay Time and the Density of States . . . . . . . . . . . .
197
6.7. Lifetime Versus Linewidth in Experiments . . . . . . . . . . . . . . . .
200
6.8. H o w M a n y M o d e s Does a Cavity Support? . . . . . . . . . . . . . .
201
6.9. Cavities W i t h o u t C h a o s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
204
6.10. Chaotic Cavities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
207
6.11. Phase Space R e p r e s e n t a t i o n with Poincar6 Sections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
208
6.12. U n c e r t a i n t y Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
210
6.13. Husimi Projection . . . . . . . . . . . . . . . . .
211
.......................
6.14. Constructive Interference with Choatic Rays . . . . . . . . . . . . . .
213
6.15. Chaotic Whispering-Gallery M o d e s . . . . . . . . . . . . . . . . . . . . . . .
216
6.16. D y n a m i c a l Eclipsing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
218
6.17. Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
219
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
220
Microparticle Photophysics: Fluorescence Microscopy and Spectroscopy of a Photonic Atom by STEPHEN ARNOLD and STEPHEN HOLLER 7.1. I n t r o d u c t i o n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
227
7.2. P h o t o n i c A t o m M o d e l . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
229
7.3. S p o n t a n e o u s Emission in a Microsphere: Cavity Q E D . . . . .
232
7.4. S p o n t a n e o u s Emission f r o m Molecules at a Microsphere Surface: A Semiclassical A p p r o a c h . . . . . . . . . . .
236
7.5. Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
244
Acknowledgment ..............................................
250
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
251
x
CONTENTS .
Single Atoms Moving in a High-Finesse Cavity
by P. W. H. PINKSE and G. REMPE
8.1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
255
8.2. Basic Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
259
8.3. Cavity-Mediated Light Forces . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
270
8.4. Experimenta! Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
282
8.5. O u t l o o k . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
288
Acknowledgments
289
.............................................
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
289
The Cavity-Enhanced Optical-Frequency Comb Generator and
its Applications
by MOTONOBU KOUROGI, BAMBANG WIDIYATMOKO, a n d MOTOICHI OHTSU 9. l. I n t r o d u c t i o n
..............................................
297
9.2. G e n e r a t i n g a n O p t i c a l - F r e q u e n c y C o m b . . . . . . . . . . . . . . . . . . .
298
9.3. U s i n g O p t i c a l - F r e q u e n c y C o m b s
305
9.4. S u m m a r y
..........................
.................................................
References ..................................................... INDEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
314 315 321
CONTRIBUTORS Numbers in parentheses indicate the pages on which authors' contributions begin.
STEPHEN ARNOLD (227), Polytechnic University, Brooklyn, New York, USA GEORGE H. ATKINSON (129), University of Arizona, Tucson, Arizona, USA GILL BERDEN (47), FOM Institute for Plasma Physics, Nieuwegein, The
Netherlands JOHN L. CARLSTEN (157), Montana State University, Bozeman, Montana,
USA RICHARD K. CHANG (185), Yale University, New Haven, Connecticut, USA RICHARD W. F o x (1), National Institute of Standards and Technology,
Boulder, Colorado, USA JOHN L. HALL (83), National Institute of Standards and Technology and
University of Colorado, Boulder, Colorado, USA LEO W. HOLLBERG (1), National Institute of Standards and Technology,
Boulder, Colorado, USA STEPHEN HOLLER (227), Los Gatos Research, Mountain View, California,
USA MOTONOBU KOUROGI (297), Tokyo Institute of Technology, Yokohama,
Japan GERARD MEIJER (47), FOM Institute for Plasma Physics, Nieuwegein, and
University of Nijmegen, Nijmegen, The Netherlands LEI S. MENG (157), Montana State University, Bozeman, Montana, USA JENS U. NOCKEL (185), University of Oregon, Eugene, Oregon, USA CHRIS W. OATES (1), National Institute of Standards and Technology,
Boulder, Colorado, USA MOTOICHI OHTSU (297), Tokyo Institute of Technology, Yokohama, Japan P. W. H. PINKSE (255), Max Planck Institute of Quantum Optics, Garching
bei Mfinchen, Germany G. REMPE (255), Max Planck Institute of Quantum Optics, Garching bei
Mfinchen, Germany PETER, A. ROOS (157), Montana State University, Bozeman, Montana, USA WIM UBACHS (47), Free University, Amsterdam, The Netherlands BAMBANG WIDIYATMOKO (297), Tokyo Institute of Technology, Yokohama,
Japan JUN YE (83), National Institute of Standards and Technology and University
of Colorado, Boulder, Colorado, USA
xi
This Page Intentionally Left Blank
VOLUMES IN SERIES EXPERIMENTAL METHODS IN THE PHYSICAL SCIENCES (formerly Methods of Experimental Physics)
Editors-in-Chief Robert Celotta and Thomas Lucatorto Volume 1. Classical Methods Editedby Immanuel Estermann Volume 2. Electronic Methods, Second Edition (in two parts) Editedby E. Bleuler and R. O. Haxby Volume 3. Molecular Physics, Second Edition (in two parts) Edited by Dudley Williams Volume 4. Atomic and Electron Physics~Part A: Atomic Sources and Detectors; Part B: Free Atoms Editedby Vernon W. Hughes and Howard L. Schultz Volume 5. Nuclear Physics (in two parts) Editedby Luke C. L. Yuan and Chien-Shiung Wu Volume 6. Solid State Physics-Part A: Preparation, Structure, Mechanical and Thermal Properties; Part B: Electrical, Magnetic and Optical Properties Editedby K. Lark-Horovitz and Vivian A. Johnson Volume 7. Atomic and Electron Physics--Atomic Interactions (in two parts) Editedby Benjamin Bederson and Wade L. Fite Volume 8. Problems and Solutions for Students Editedby L. Marton and W. F. Hornyak xiii
xiv
VOLUMES IN SERIES
Volume 9. Plasma Physics (in two parts) Edited by Hans R. Griem and Ralph H. Lovberg Volume 10. Physical Principles of Far-Infrared Radiation Edited by L. C. Robinson Volume 11. Solid State Physics Edited by R. V. Coleman Volume 12. Astrophysics~Part A: Optical and Infrared Astronomy Edited by N. Carleton Part B: Radio Telescopes; Part C: Radio Observations Edited by M. L. Meeks Volume 13. Spectroscopy (in two parts) Edited by Dudley Williams Volume 14. Vacuum Physics and Technology Editedby G. L. Weissler and R. W. Carlson Volume 15. Quantum Electronics (in two parts) Edited by C. L. Tang Volume 16. Polymers~Part A: Molecular Structure and Dynamics; Part B: Crystal Structure and Morphology; Part C: Physical Properties Edited by R. A. Fava Volume 17. Accelerators in Atomic Physics Edited by P. Richard Volume 18. Fluid Dynamics (in two parts) Edited by R. J. Emrich Volume 19. Ultrasonics Edited by Peter D. Edmonds Volume 20. Biophysics Editedby Gerald Ehrenstein and Harold Lecar
V O L U M E S IN SERIES
Volume 21. Solid State Physics: Nuclear Methods EditedbyJ. N. Mundy, S. J. Rothman, M. J. Fluss, and L. C. Smedskjaer Volume 22. Solid State Physics: Surfaces Editedby Robert L. Park and Max G. Lagally Volume 23. Neutron Scattering (in three parts) Editedby K. Skold and D. L. Price Volume 24. Geophysics--Part A: Laboratory Measurements; Part B: Field Measurements Editedby C. G. Sammis and T. L. Henyey Volume 25. Geometrical and Instrumental Optics Editedby Daniel Malacara Volume 26. Physical Optics and Light Measurements Editedby Daniel Malacara Volume 27. Scanning Tunneling Microscopy Editedby Joseph Stroscio and William Kaiser Volume 28. Statistical Methods for Physical Science Editedby John L. Stanford and Stephen B. Vardaman Volume 29. Atomic, Molecular, and Optical Physics--Part A: Charged Particles; Part B: Atoms and Molecules; Part C: Electromagnetic Radiation Editedby F. B. Dunning and Randall G. Hulet Volume 30. Laser Ablation and Desorption
Editedby John C. Miller and Richard F. Haglund, Jr. Volume 31. Vacuum Ultraviolet Spectroscopy I
Editedby J. A. R. Samson and D. L. Ederer
Volume 32. Vacuum Ultraviolet Spectroscopy II
EditedbyJ. A. R. Samson and D. L. Ederer
Volume 33. Cumulative Author Index and Tables of Contents, Volumes 1-32
XV
xvi
VOLUMES IN SERIES
Volume 34. Cumulative Subject Index Volume 35. Methods in the Physics of Porous Media Edited by Po-zen Wong Volume 36. Magnetic Imaging and its Applications to Materials Editedby Marc De Graef and Yimei Zhu Volume 37. Characterization of Amorphous and Crystalline Rough Surface: Principles and Applications Editedby Yi Ping Zhao, Gwo-Ching Wang, and Toh-Ming Lu Volume 38. Advances in Surface Science Edited by Hari Singh Nalwa Volume 39. Modern Acoustical Techniques for the Measurement of Mechanical Properties Editedby Moises Levy, Henry E. Bass, and Richard Stern Volume 40. Cavity-Enhanced Spectroscopies Edited by Roger D. van Zee and J. Patrick Looney
PREFACE
While the proposition that lasers have revolutionized spectroscopy is unlikely to generate controversy, the claim that cavity-enhanced methods have effected comparable change might provoke protest. But we believe that this assertion can be successfully argued too. Witness the ubiquity of build-up cavities in atomic spectroscopy, the incredible sensitivity of cavity-enhanced, frequency-modulation spectroscopy, or the explosive growth in the laboratory use of the ring-down techniques. This volume is aimed at practising spectroscopists who are interested in cavity-enhanced methods but may lack the photonics expertise of an electrical engineer. In this vein, and because this volume is dedicated to cavity-enhanced methods, the authors were encouraged to highlight the enabling role the resonator plays. This volume is not, however, a monograph on optoelectronic theory, and any such theoretical ruminations are counterbalanced with plenty of practical, experimental discussions. Thus, the introductory chapter is a practical review of locking lasers to optical resonators. This volume then presents an eclectic sampling of spectroscopic methods that use resonators for signal enhancement or otherwise exploit light/matter interactions within a resonator. The second and third chapters discuss the simplest cavity-enhanced method: the use of a passive resonator to do spectroscopy on atoms and molecules in a cavity. The next chapter deals with a similar concept, absorption measurement in a cavity, but an active one. The book closes with chapters that discuss microresonators and phenomena unique to or enhanced by the resonator environment, such as the hydrogen Raman laser, cavity QED, and optical-frequency comb generators. Other subjects could certainly have been included in this book, and the ones we selected are based partly on our fancy. Equally important in our selection, however, was the criterion that a method should have matured beyond conceptualization and proven its mettle, but not be in common use. In this regard, these chapters should prove a timely aid, for much of what is described in this volume is known mostly to expert practitioners. Hopefully, this volume will become a standard and authoritative reference and a useful guide to more widespread implementation of these important techniques. Finally, we would also like to take this opportunity to acknowledge the contributions of the many who helped us during the course of our work, in
xvii
xviii
PREFACE
particular Dr. Joseph T. Hodges, Dr. James R. Whetstone, and Dr. Hratch G. Semerjian.
National Institute of Standards and Technology Gaithersburg, Maryland, USA July 2002
xviii
R. D. van Zee J. P. Looney
1. STABILIZING DIODE LASERS TO HIGH-FINESSE CAVITIES Richard W. Fox, Chris W. Oates, Leo W. Hollberg National Institute of Standards and Technology, Boulder, Colorado, USA
1.1 Introduction This chapter is written for the chemist, physicist, or engineer who is interested in locking a diode laser to an optical cavity. There already exist many good references on the locking of lasers [1, 2], including some on diode lasers in particular [3, 4]. In this chapter we aim to provide a practical hands-on guide, with most or all of the theory left to cited references. The motivations for locking lasers to high-finesse optical cavities include applications as varied as laser cooling, length metrology, and analytical and precision spectroscopy. In some cases it may be the stabilization of the laser's frequency to a narrow cavity resonance that is of interest, while in others it may be the huge build up of intracavity power or the long effective path length between the mirrors that is important. The case of locking diode lasers is of particular interest because of their unique characteristics, relatively low cost and widespread use in many applications. Before proceeding further, we should offer a note of caution that the term "diode laser" now encompasses many devices whose electrical and optical characteristics differ greatly. Included in this category are the Fabry-P6rot laser, vertical-cavity surface-emitting lasers, and distributed feedback lasers. While the basic approach for locking these different types of lasers is the same, the performance of the lock and the details in the electronic feedback loops will be quite different. Rather than overwhelm the reader with details for all these different types, we instead focus on one type in particular, the widely used extended-cavity Fabry-P6rot laser. For those interested in locking other types of diodes to cavities, this chapter should still be a good place to start. This chapter consists of the following. We begin with a general discussion of the issues involved in diode laser locking and introduce the reader to some of the terminology. We then describe in detail the various steps needed to lock the laser to a cavity resonance: (1) derivation of the error (locking) signal, (2) design of the electronic feedback circuitry, (3) initial locking of the laser, (4) adjustment of the feedback design, and (5) evaluation of the lock performance. We illustrate this discussion by frequency locking an 1 EXPERIMENTAL METHODS IN THE PHYSICAL SCIENCES Vol. 40 ISBN 0-12-475987-4
Published by Elsevier Science (USA) All rights of reproduction in any form reserved. ISSN 1079-4042/01 $35.00
2
STABILIZING DIODE LASERS
extended-cavity diode laser, reducing the linewidth to a few hertz relative to the cavity. We conclude with an example in which we modified the locking apparatus for a cavity ring-down demonstration. Included are results showing the laser repetitively locking and unlocking to the cavity.
1.2 Introduction to Diode Laser Locking Regardless of the application, the basic goal of locking the frequency of a laser to a cavity is to reduce the frequency fluctuations between the laser and cavity. The noise spectrum of the laser's frequency fluctuations leads to an effective "linewidth" of the laser, which conceptually describes the broadening of the laser's spectrum around its central frequency. The concept of laser linewidth can be a little confusing but is valuable, so it is worth a brief discussion here. The basic confusion arises from the fact that the laser linewidth is dependent on the timescale over which it is evaluated, as a laser's noise spectrum typically contains very different fast and slow components. A useful intuitive picture is that the fast fluctuations give rise to what is called the "fast linewidth" (or sometimes just "linewidth"); by "fast" we usually mean in a time less than the effective spectroscopic interaction time (typically tens or hundreds of microseconds). With this definition, the linewidth defines the narrowest feature that the laser is capable of resolving. The low-frequency fluctuations (usually of larger magnitude) then cause jitter of this narrow spectral line in frequency space. Even slower changes often occur because of thermal effects, which can cause the laser's central frequency to drift. In order to reduce the laser linewidth, one needs a stable frequency reference suitable for measuring the laser's frequency fluctuations. One can then construct a feedback loop that attempts to compensate these fluctuations. A Fabry-P6rot cavity, which usually consists of two highreflectivity mirrors separated by a spacer, is a convenient choice for a frequency reference. With mirror reflectivities > 99% over a given range of wavelengths, a cavity has a series of evenly spaced, sharp resonances (typically with linewidths < 10 MHz). The resonance or "fringe" separation is c/2L, where c is the speed of light and L the length of the optical cavity (i.e. the distance between the mirrors). Radio-frequency modulation techniques can be used to derive an electronic error signal that represents the deviations of the laser frequency from a given cavity reference fringe. One then uses electronic feedback (primarily to the laser current in the case of diode lasers) to control the laser frequency and minimize its deviations relative to the cavity fringe.
I N T R O D U C T I O N TO DIODE LASER L O C K I N G
3
If one succeeds in making the electronic feedback bandwidth (i.e. range of frequencies over which corrections are effectively imposed) wider than the dominant noise spectrum of the laser, the laser's fluctuations can be controlled such that the laser appears tightly locked to the peak of the cavity mode. In this case, the linewidth relative to the cavity's center frequency can be a small fraction (~ 10-4) of the cavity's linewidth. (In this case, the actual laser linewidth will almost certainly be determined by the mechanical stability of the cavity itself.) Likewise, even though the power is not actively stabilized, the power transmitted from the cavity can be stable to better than 1%. On the other hand, if the electronic feedback stabilizes only the lowfrequency portion of the laser's fluctuations, the center frequency will be fixed, but no linewidth narrowing will occur and relatively little of the laser power will be coupled into the mode. Thus we see that there are two basic determining factors in the final performance of the lock: the initial noise spectrum of the laser and the electronic bandwidth achievable in the feedback or "servo" system. The short cavity lengths and low facet reflectivity found in most diode structures result in "low-Q" cavities, and generally the optical output will have a minimum spectral linewidth of 10 MHz, and often much larger. Thus, if we expect to narrow the linewidths of these lasers significantly, we will need feedback loops with sufficient gain up to 10 MHz and beyond. However in most cases (for most commercial diodes), a feedback bandwidth of 10 MHz is quite difficult to achieve because of the characteristics of the diode laser itself. At low modulation frequencies, the frequency shift with injection current is primarily caused by thermal effects. That is, because the injection current causes heating and the index of refraction is temperature dependent, we find that the subsequent frequency modulation (df/d/) is negative. This effect diminishes with modulation frequency, and at times on the order of 1 gs, another smaller effect, caused by the electronic charge carriers, starts to dominate. For this case we have the opposite dependence, i.e., df/dI is positive. The resulting net phase response of the diode is troublesome for the servo designer, making electronic feedback bandwidths beyond a few megahertz increasingly difficult to achieve. To achieve tight locking, it is therefore often necessary to reduce the frequency fluctuations of the laser by other means before locking it to the cavity. A common technique is to increase the optical cavity's Q by adding optics external to the laser chip in order to form a longer cavity. A laser cavity built in this manner is often called an ECDL, "extended-cavity diode laser," or "external-cavity diode laser." The use of frequency-dependence reflection for the external cavity (e.g. with a diffraction grating) can offer the additional benefits of frequency pulling and reduced mode competition. There are many extended-cavity lasers available commercially, and several
4
STABILIZING DIODE LASERS
AR Coating
l
--
PZT
Mirror
Output Beam
Fio. 1.1. A diode laser arranged in a Littman-Metcalf configuration. The beam is incident at grazing incidence on the diffraction grating, and the first order is reflected back by a mirror. The zero order serves as the laser output. Coarse wavelength tuning can be done by adjusting the mirror angle, and fine-tuning by using a PZT to adjust the cavity length. However, true single-mode scanning over a large range requires that the laser cavity length be changed synchronously with the mirror angle.
good references for their design and construction [5, 6]. The laser used as an example here is an off-the-shelf commercial diode that was antireflection coated and built into a Littman-type extended cavity about 8cm long (see Fig. i.1). The extended cavity is formed with a diffraction grating (for wavelength-selective optical feedback) and a high-reflectivity mirror mounted on a piezo-electric transducer (PZT). The laser operates around 830nm and yields an output power of 6 m W at an injection current of 70 mA. We can tune the laser frequency quickly over small excursions with the laser current, or slowly over larger excursions with the voltage on the PZT. The resulting fast linewidth for this laser is roughly 50 kHz, while there are low-frequency vibrations of the mechanical structure (caused by room acoustics), which cause the laser output to jitter in frequency on the order of a megahertz. We note that in many cases, one must use an off-the-shelf commercial diode (without an additional antireflection coating), which usually requires that a shorter extended cavity be used to reduce laser mode competition. Unfortunately, shorter extended cavities yield less frequency noise reduction, leaving more work to be done by the electronic feedback system. Nonetheless, the techniques described in this chapter should still be directly applicable to this case as well and provide good locking performance [7]. We should also add that while the use of electronic feedback is convenient, it is not the only option for stabilizing the diode laser's frequency to a cavity. The use of resonant optical feedback directly from a cavity is a wellestablished technique that can be used to stabilize even a simple Fabry-P6rot diode to certain cavities [8-10]. A requirement is that the
G E N E R A T I N G THE ERROR SIGNAL
5
(a)
(b)
FIG. 1.2. Two configurations to optically lock a diode laser to an optical cavity. The cavity must return light to the laser when on resonance, while reflecting away the off-resonance beam. Diagram (a) is a three-mirror V-shaped cavity, while diagram (b) is a confocal optical cavity aligned in a V-configuration. For stable operation, the path length from the laser to the cavity must be servo controlled. Not shown are mirrors and a coupling lens to align the laser to the cavity.
optical alignment be such that light is reflected or returned to the laser diode only when the cavity is in resonance. Two possible configurations include a three-mirror V-shaped cavity, and a confocal cavity aligned in the c/4L configuration (see Fig. 1.2). Additionally, it is necessary to electronically control the phase of the light returning to the diode laser, which can be accomplished with a mirror mounted on a PZT [11].
1.3 Generating the Error Signal:
The Pound-Drever-Hall Method
The first step in the locking procedure is to generate an electronic error signal that can be further processed for locking the laser to the cavity. Perhaps the two most convenient approaches for generating this signal are side-locking and the Pound-Drever-Hall (PDH) method [1]. With the simpler technique, side-locking, the laser is stabilized to the side of the cavity fringe without the use of modulation techniques. While this technique is straightforward to implement and has some advantages (e.g. simplicity, modulation-free), it suffers from several serious disadvantages. First, because it is modulation-free, one necessarily detects the error signal
6
STABILIZING D I O D E LASERS
at DC, where there can be significant amplitude noise. Second, side-locking has a much smaller acquisition range, which means the laser and cavity frequencies need to be nearly coincident before the system will lock. Third, the lock is less robust because perturbations (e.g. vibrations) that drive the laser to the opposite side of the resonance will cause the feedback loop to push the laser further from resonance. Fourth, because one locks to the side of the resonance rather than the top, there is reduced buildup of optical power in the cavity, and there may be increased noise on the transmitted intensity. The PDH technique circumvents most of these drawbacks by modulating the frequency of the light, enabling detection of the error signal at a high frequency where the technical noise is near the shot-noise limit. The resulting demodulated error signal has a high signal-to-noise ratio and a large acquisition range, which can produce robust locks. Furthermore, this error signal has odd symmetry about the line center that enables locking to the top of a cavity fringe. For these reasons, we prefer the PDH method and focus on its implementation throughout the remainder of this chapter. The PDH technique is described in detail in many references [1-4], but we will review the general approach here. Consider a purely frequency modulated (FM) laser beam impinging on the input mirror of an optical cavity and reflecting back to a detector. For low modulation index (as we typically use), one can view the frequency spectrum of the modulated light as consisting of a carrier with two sidebands: one at higher frequency with a phase relative to the carrier that is in phase with the modulation, and one at lower frequency that is out of phase by 180 ~. As long as there is no absorption or phase shift of the laser carrier or modulation sidebands with respect to one another, the detector photocurrent will not have a signal at the modulation frequency. (Only the optical phase is modulated, not the optical power.) A simple view of this fact is that the beating between the carrier and the upper frequency sidebands creates a photocurrent modulation that is exactly canceled by the out-of-phase modulation from the lower frequency side. If a sideband is attenuated or phase shifted, or the carrier's phase is shifted, the photocurrents will not cancel and RF power at the modulation frequency will appear on the detector signal. Near a cavity resonance, the resultant optical reflection of the carrier from the cavity is phase shifted with respect to the sideband components that are further away from the cavity resonance. Consequently, the detector photocurrent will show power at the modulation frequency. The laser's frequency noise will then appear as noise sidebands centered around the modulation frequency. When this signal is mixed to base-band (using phasesensitive detection with the appropriately chosen phase), the result is
GENERATING THE ERROR SIGNAL
7
Fro. 1.3. A schematic representation of the Pound-Drever-Hall error signal as if the laser was scanned over a cavity resonance. The cavity reflection is detected and demodulated by a mixer using a phase-delayed local oscillator. Proper adjustment of the phase delay results in a sharp central frequency discriminator, with secondary peaks spaced by the modulation frequency.
a frequency discriminator with odd symmetry that may be used to correct the frequency of the laser as shown schematically in Figure 1.3. Here we note that the light seen by the detector actually consists of two components: the fraction of the input beam that is reflected, plus the fraction of the internal cavity wave that is transmitted back out of the input coupler. The detected photocurrent represents the interference of these two components. For Fourier components of the laser's frequency noise below the cavity linewidth, this system acts like a frequency discriminator as described above. At frequencies above the cavity linewidth, the input field is essentially heterodyned with the cavity wave. Thus, for these Fourier components, the system acts as a phase discriminator, which causes the system response to faster frequency fluctuations to decrease as 1If. In electronic terms, the P D H technique gives us a frequency error signal with a sensitivity that can be measured in volts per hertz of optical frequency. At the Fourier frequency corresponding to the cavity linewidth, the sensitivity starts to decrease, and continues to decrease as 1/f. At some higher frequency, the error signal will cease to be useful as the magnitude decreases to the level of the background noise, although this limit is usually well above the attainable servo bandwidth. For those who are perhaps puzzled by some of terminology above, recall that phase-sensitive detection is the process that a simple lock-in amplifier uses, although most commercial lock-in devices do not operate at the
8
STABILIZING DIODE LASERS
frequencies required here (10-50 MHz). In this chapter we show how to build the necessary circuitry using RF amplifiers and filters, a simple mixer, and subsequent operational amplifier (op-amp) filtering and amplification. Our experimental layout is shown in Figure 1.4. The light goes from our ECDL through an optical isolator and electro-optic modulator to the optics used for coupling to the optical cavity. While this setup would be suitable for most types of lasers, let us emphasize a couple of aspects particular to diode lasers. Because diode lasers are extremely sensitive to optical feedback, good optical isolation (> 50dB) is necessary to attenuate the light reflected from the cavity. Additionally, the laser's spatial mode is not round, but is instead oval because the laser's output aperture is asymmetric. Although a spherical collimating lens is often used at the diode output, correcting the beam's aspect ratio with anamorphic prisms or cylindrical lenses would result in better spatial matching to the cavity mode. This would in turn increase the power coupled to the cavity. In the examples here, a single lens and an adjustable aperture are used for spatial mode-matching to the cavity, and two alignment mirrors are provided between the lens and the cavity. The cavity reflection is returned to the photodetector using a quarter-wave plate along with a polarizing beam splitter. The detector signal is amplified and filtered to pass the modulation frequency, and phasesensitively demodulated using a balanced mixer. The resulting signal is then
Polarizing Beamsplitter
X/4
i LaserI "'--" I Phase-Modulator
IP'FSynthezerI I
Cavity
I Delay I Mixer
~
r Band-pass Detection Amplification
LoopFilter [~ Error signal
Fla. 1.4. A basic Pound-Drever-Hall locking scheme. A modulation frequency in the range 10-40 MHz is typically used. The phase of the RF local oscillator input to the mixer is adjusted to give a symmetrical frequency discriminator at the mixer output. In the absence of residual amplitude modulation or RF pickup, the center of the cavity resonance will correspond to zero at the output. The error signal must be properly filtered and amplified to control the laser.
GENERATING THE ERROR SIGNAL
9
amplified and filtered as desired for the feedback loop, and finally sent to the diode laser frequency control elements. Let us now see how to design and construct these various pieces, starting first with the optics before moving on to the electronics.
1.3.1 Coupling Light into the Cavity In this section, we describe how to choose and align the optics used for coupling the light from the diode laser into the optical cavity. The optimization of this procedure requires consideration of two distinct and important concepts, mode-matching and optical impedance-matching. Modematching refers to adjustment of the input beam's size, shape, and wavefront curvature to match the cavity mode. Impedance-matching refers to the adjustment of the cavity parameters (e.g. mirror reflectivities) to maximize coupling into the cavity. This occurs when the transmission through the input mirror equals all the other losses of the resonator. For instance, a two-mirror cavity would be impedance-matched if the loss from the input mirror equaled the sum of the second mirror's loss plus the loss resulting from scatter and absorption on the mirrors and anything else in the cavity. If this ideal condition is met, the reflection of a perfectly spatially matched input beam will destructively interfere completely with the cavity wave transmitted back through the input coupler. Thus, no net power will be reflected off the input mirror on resonance in steady state. Usually this condition is not met, and even a perfectly spatially matched beam will not couple all the power into the cavity. However, coupling efficiencies of 80% are not uncommon. Instead, the reflected signal from the cavity on resonance will show a dip to a minimum, but not a dip to zero. In some applications, however, such as ring-down spectroscopy, impedancematching may be less important than some other consideration such as maximizing the ring-down time. In such a case, one might minimize the loss at each mirror, for example, even if it means that less power is coupled. Matching the spatial mode of a diode laser's beam to that of a spherical mirror cavity is a challenge because the laser's output beam is neither round nor "diffraction limited." Conceptually, think of a beam emanating from a waist in the cavity towards the laser, and let us call this the "cavity wave." The beam waist and radius of curvature of the cavity wave may be calculated from the mirror geometry [12]. The mode-matching task is to shape and focus the diode beam such that it approaches the conjugate of this cavity wave, i.e., a wave that is exactly like it except moving towards the cavity. Typically we start with a beam that is collimated with an additional lens, if necessary, after it emerges from the optical isolator or electro-optic
10
STABILIZING DIODE LASERS
modulator. A first step is often then to make the collimated diode beam's cross section somewhat round by using a pair of anamorphic prisms. Next, at the fictitious point where the diode beam and the cavity wave are the same size, a piano-convex lens of the proper power will focus the diode beam to a waist at the same position as the cavity's waist. If the collimated diode beam is larger than the cavity wave, use either a collimating lens of shorter focal length on the diode, or a multiple lens solution to reduce the beam's cross section. For instance, a positive lens can be used to create a converging diode beam that at some point will have the same cross-sectional size as the cavity wave. A negative lens of the proper focal length at this point will reduce the wavefront curvature, and cause the diode beam to reach a waist at the same position as the cavity waist. See Reference [12] for more details on mode-matching. A brief description on exactly how to align the diode beam to a highfinesse, two-mirror Fabry-P6rot cavity is warranted here. Much of this discussion applies to ring cavities as well. An adjustable diaphragm prior to the lens is useful, both for alignment and for reduction of the power coupled to higher order modes. It is easiest to align the diode beam to the cavity if there are two adjustable turning mirrors of good quality between the cavity and the focusing optics. Adjust the two mirrors such that the input beam is incident at the center of the cavity mirror, and also such that the cavity reflection hits the turning mirrors at the same points as the input diode beam. Observe the cavity reflection on the diaphragm; if the beam size is approximately the same size as the input beam at the diaphragm, then input wavefront curvature is well chosen. If not, either the input diode beam's radius of curvature is incorrect, or the position of the waist is not near the cavity wave's waist. Careful calculation or some experimentation should enable one to get close. Now, with a detector placed just after the cavity, one can monitor the cavity transmission on an oscilloscope, while sweeping the laser frequency slowly (~ 30Hz) with a triangle drive signal fed to the PZT-controlled mirror. Alternatively, the cavity length can be swept if one of the cavity mirrors is mounted on a PZT. Once modes are observed, the laser's wavelength sweep should be adjusted so that about two cavity flee-spectralranges are covered. Fine adjustment of the alignment (and mode-matching) should then lead to good coupling to the fundamental TEM00 mode. One common problem that arises is to figure out which peak actually corresponds to the TEM00 mode. To this end we offer several suggestions. Closing the adjustable aperture to a small size will attenuate the coupling to other modes more than the TEM00 mode, because it is spatially symmetric and the smallest mode. Also, all the mode peaks except for the TEM00 mode are actually two or more modes with approximately the same resonance
G E N E R A T I N G THE E R R O R SIGNAL
11
frequencies. This degeneracy is broken in high-finesse cavities by any departure of the mirror surfaces from a spherical shape [13]. This leads to a fine splitting of all the higher order transverse modes that is easily seen by zooming in on an unknown mode by reducing the laser sweep and increasing the oscilloscope gain. The fundamental mode does not have this degeneracy and will appear as a single peak. Observation of the transmitted spatial mode profile by eye or with a video camera after the cavity can help to optimize alignment into the TEM00 mode as well. With some patience, one can usually get to the point where the TEM00 peak is a factor of five or ten times larger than any of the competing modes. When working with very high-finesse cavities, one may notice that the peak heights vary drastically from sweep to sweep. Sweeping more slowly may increase the heights of the peaks, but may or may not reduce the variation. These effects result from the laser frequency not staying in resonance long enough (i.e. the cavity decay time) to reach equilibrium (recall that the cavity buildup time is long for high-finesse cavities). Alignment under these conditions requires patience, although we will see later on that when the servo system is operating, one can easily optimize the alignment by working with the laser locked to the cavity fringe. We also note that mirror contamination can affect the loss of the fundamental mode relative to higher order modes, especially the first odd transverse mode. It is well known that the mirrors of high-power dye laser systems will become contaminated by material deposited on the mirror surface just where the beam power is maximum. With good low-loss mirrors, the ring-down cavity power can be 10 W continuous-wave, even for milliwatt input powers, so one can expect this problem occasionally if the cavity is operated with high power. A characteristic sign of this occurring is a change in the spatial coupling such that the TEM00 mode power is degraded in relation to the transverse modes.
1.3.2 Modulating the Laser Frequency There are several competing factors to consider when choosing the modulation frequency for the PDH lock. The frequency needs to be sufficiently high so that the process of filtering and demodulating to base-band does not yield a significant phase shift within the desired feedback bandwidth. We find that choosing a modulation frequency an order of magnitude higher than the required bandwidth is more than sufficient. Although modulating at even higher frequencies (hundreds of megahertz) is possible, this requires more skill to avoid distorting the DC baseline because of RF pickup. It also makes the local-oscillator phase more
12
STABILIZING DIODE LASERS
susceptible to variations through changes in temperature of the cables, and more susceptible to residual amplitude modulation (AM) offset errors associated with variations in the optical path length. In practice, we find that modulation frequencies ranging from 15 MHz to 40 MHz work well. To modulate the laser frequency, we either modulate the laser's injection current directly or use an electro-optic modulator (EOM) placed between the laser and the cavity. Of these two schemes, using an EOM is preferable because it yields much less residual AM. Residual AM can be quite troublesome for the locking process, as it leads to DC offsets after demodulation, which in turn shift the lock point away from the peak of the cavity transmission. In principle such offsets can be compensated, but often the factors responsible for the residual AM are unstable, and lead to a drifting offset that is more difficult to null. Thus, we use an EOM when frequency drift relative to the cavity is important, but employ direct-current modulation for less critical locks. Commercial EOM units are generally configured either as resonant circuits or as broadband modulators. In order to generate sidebands of sufficient amplitude (5-10% of the carrier peak intensity), a resonant configuration makes sense because it reduces the required RF power by about a factor of ten. The addition of a low loss inductor and capacitor external to a broadband EOM (as shown in Fig. 1.5) works well and Q's > 10 are easily achieved. In order to know what fraction of the light is coupled into the sidebands by the modulator, one can couple light into the cavity and monitor the transmission spectrum. The sidebands can then be observed by sweeping the laser's frequency over a cavity resonance, or alternatively by sweeping the cavity if its length is adjustable with a PZT. This signal can then be used to adjust the RF power sent to the EOM in order to generate sidebands of the desired amplitude. Of course, an EOM adds considerable cost to the system, and it is often possible to lock by modulating the laser directly. The success of this approach depends on the magnitude of the amplitude modulation that accompanies the desired frequency modulation. This in turn depends on the laser design, specifically on parameters such as the length of the extended cavity and the output coupling of the laser. Unfortunately the same process that serves to reduce the laser's frequency noise and emission linewidth , relative to the bare diode laser also reduces the laser's frequency deviation for a given modulation current. Consequently, modulating the injection current of a laser with a long extended cavity and a high reflectivity grating will produce large AM sidebands, with very little associated frequency modulation. One can check the modulation characteristics by aligning the laser to an optical cavity and observing the amplitudes of the induced sidebands. If direct modulation produces obviously asymmetric sidebands,
GENERATINGTHE ERRORSIGNAL
13
(b)
(a)
lOpF T
?
1 uH
-li
C (crystal)@
Fla. 1.5. A broadband nonresonant electro-optic modulator can produce sidebands adequate for locking if it is driven resonantly. With an inductor mounted externally but close to the modulator, the tank circuit shown will resonate at 2rcf~ (LC) -1/2. In configuration (a), an additional external capacitor is necessary for coupling (an airspaced adjustable capacitor works well). Tuning the inductor will affect primarily the resonance frequency, while tuning the coupling capacitor will change the strength of coupling and hence the input impedance. Stronger sidebands will be obtained with configuration (b), however, this requires that neither side of the modulator crystal be grounded. The highest Q will be obtained by using nonadjustable air-core inductors and tuning the modulation frequency into resonance. this is an indication that the relative amounts of amplitude and frequency modulation are comparable, and that the degree of frequency modulation is sufficient to create an error signal for locking. The asymmetry of the sidebands is caused by the AM and F M components being in-phase on one side of the carrier and out-of-phase on the other side. For an ECDL, if the direct modulation sidebands are symmetrical, it is possible that they are primarily AM sidebands, and the mixer output will have a DC component that is large with respect to the error signal. The amount of AM is also easily detected by a high-speed photodiode. Direct modulation can be accomplished by coupling an R F signal to the injection current with a small capacitor (_< 50pF), as shown in Figure 1.6. The standard electrostatic precautions and warnings apply to this current input terminal. For instance, one needs to ensure that no current pulse is delivered to the laser when connecting a cable or turning on the RF function generator. When in doubt, the safe practice would be to connect the function generator while it is on, but with the output at zero. Regardless of the modulation technique, the amplitude of the sidebands imposed on the laser is not critical, and it is possible to trade some modulation depth for gain elsewhere in the loop. A good starting point is to set the sideband amplitude at approximately 5-10% of the carrier. If the modulation sidebands are too weak, then the noise floor of the resulting system will be higher than necessary, and the resulting laser linewidth relative to the cavity will be higher than necessary. Also, the acquisition range will be poor, as the wings of the P D H error signal will be degraded to the noise level. If the modulation sidebands are too large, the power in the carrier and the subsequent cavity power buildup and transmittance will suffer.
14
S T A B I L I Z I N G D I O D E LASERS 50 pF 1[
r",'-'c--ra__ Current Source Laser Diode
Fro. 1.6. A simple circuit to apply RF modulation to the laser through a coupling capacitor.
1.3.3 Detecting the Reflected Light Light reflected off the input mirror of the high-finesse cavity can be picked off with a simple beam splitter, or a polarizing beam splitter and either a quarter-wave plate or a Faraday rotator. The latter choices use the light more efficiently, while the former is cheaper and reduces the number of optics in the beam path (which can cause residual AM). Attention must be paid to multiple reflections in the path between the cavity and the detector, because they can lead to ~talon effects that are all too familiar to spectroscopists, l~talon effects usually manifest themselves as ripples in the mixer output as the laser is swept in wavelength. These baseline ripples can be quite large and seriously compromise the locking performance. Thus, it is important to take some care with the optics to avoid normal reflections that can lead to unintentional 6talons. Slight tilting of the detector will avoid sending the reflection from the detector surface back towards the cavity. Likewise, defocusing or using a longer focal length lens to capture the light on the detector will result in less scattered light being reflected back towards the cavity. The detector monitoring the cavity reflection should be a lownoise detector with a reasonable bandwidth and also have a low-frequency DC output. Although a simple photodiode and a load resistor may be used to capture the reflected signal, the signal size will be limited because the load resistor needs to be relatively small in order for the detector to have sufficient bandwidth. The signal then must be routed through an RF amplifier to reach levels optimum for the mixer input (see Fig. 1.4). Serious consideration should be given to using a transimpedance amplifier to convert the photocurrent output to a voltage. There are wide-band transimpedance amplifiers available with built-in feedback resistors and impressively low input noise. Although not necessary for locking purposes, the DC output is useful for monitoring the laser power and to diagnose laser-cavity coupling problems. The low-noise aspect of the detector is important if the reflected light power from the cavity is expected to be low. To have optimum signal-tonoise for the error signal, the photocurrent fluctuations must be larger than the detector noise. The amount of light power needed for this to be the case
GENERATING THE ERROR SIGNAL
15
can be calculated assuming the photocurrent fluctuations are attributable to shot noise. Of course if the photocurrent noise is actually higher than the shot level, the detector noise level will be exceeded with even less light than this calculation indicates. With no light incident, the detector-amplifier combination will exhibit some output noise density at the modulation frequency, measured in V/~/Hz which can easily be measured by an RF spectrum analyzer. One can calculate the power P0 that would be necessary for the photocurrent noise to exceed the detector noise by setting this background noise level equal to the shot noise,
Gv/2qPoR In this expression G is the transimpedance gain in V/A, q is the electronic charge, 1.6x10 -19 C, and R is the detector's responsivity in A/W. We note that there are commercial photoreceivers sold as "low noise" that will require more than 1 mW of power on the detector before the photocurrent shot level is larger than the detector noise. However, there are also transimpedance amplifiers with more than 100 MHz of bandwidth that require less than 50 gW of power before the light noise is larger than the detector's noise.
1.3.4 Demodulation Once a satisfactory signal is obtained from the detector, it is time to demodulate the error signal so that it will be ready to be used in the servo system. The effect of frequency modulation was to encode the laser frequency fluctuations around the modulation frequency (rather than DC); now it is time to "decode" this error information by mixing it down to DC. This demodulation process is easily accomplished by combining our error signal from the detector with a phase-shifted version of the local-oscillator signal used to generate the modulation in a doubly balanced mixer. We note that it can be beneficial to add an RF band-pass filter between the detector and the mixer, as signals outside of the anticipated servo bandwidth only add excess noise at the mixer input. However, care needs to be taken that the band-pass filter is not too narrow, as this will lead to phase shifts that will limit the servo bandwidth. Unless there is significant noise outside the servo bandwidth, a good starting point is to use no filter prior to the mixer. For good mixer performance, the local-oscillator (LO) signal power should be enough to fully turn on the mixer diodes. (While mixers are available for a number of different power levels, we find that those requiring an LO power of + 7 d B m to be a convenient choice.) The size of the resonance signal from the detector into the mixer's R F port should be within
16
STABILIZING DIODE LASERS
a few decibels of the maximum given in the manufacturer's specifications (typically the RF port level will be a maximum of +1 dBm for a +7 dBm LO). If the output of the detector is much less than this, a low-noise RF amplifier can be inserted before the mixer's RF port. The RF signal, which is centered around the modulation frequency, is down-converted by the mixer to base-band (i.e. DC), with the resultant output accessible at the mixer IF port. This is done in a coherent manner, and with some conversion efficiency, typically about - 6 dB. The IF port output also contains the same information at twice the modulation frequency, as a doubly balanced mixer converts the input signal at fRF to fLO--fRV and fLO+fRF. In addition, if the mixer IF port output is observed with a spectrum analyzer, a smaller but often significant narrowband component at lEo will be found. These high-frequency outputs of the mixer can be sufficiently attenuated by the loop filter, which as we shall see is actually a low-pass filter. However, it is possible that these residual RF signals can generate a small DC signal through rectification by any nonlinearity in the circuit. This can wreak havoc with the frequency stability of a system that was designed to closely track the laser to the center of the cavity. It thus makes sense to attenuate the strong high-frequency components with resonant LC notch filters at the mixer output. The resonances should be of sufficient Q to avoid unnecessary phase shifts within the desired servo bandwidth [14]. In Figure 1.7, we show circuit components to reduce f and 2f (30 MHz and 60 MHz) while introducing only I~ of phase shift in the region from 1MHz to 10 MHz. The lineshapes one detects at the output of the mixer IF port depend on the relative phase between the RF signal and the LO. In order to generate a frequency discriminator with odd symmetry and maximum amplitude, it is essential to set this relative phase to be dispersion-sensitive (i.e. 90 ~ outof-phase). The phase of the LO fed to the mixer is easily adjusted with phase-shifting IC, a delay box, or simple lengths of cable. One could instead adjust the phase of the RF signal, but this may lead to undesired delay o~ phase shifts in the servo loop. To set the LO phase to the ideal value, it is easiest to start by using an oscilloscope to monitor the signal from the IF port while sweeping the laser or cavity over a cavity resonance. A low-pas~, filter (e.g. with a corner at ~ 100 kHz) will attenuate the 2fsignal and help tc see the features more clearly. One then adjusts the LO phase and looks foJ the appropriate discriminator line shape (see Fig. 1.3). There will in fact b( two such "ideal" settings; these yield mirror images of the desirec discriminator signal (switching between these two settings gives an eas3 way to change the overall sign of the feedback signal). If the laser frequenc) jumps around too much to find the phase setting that optimizes th( discriminator amplitude, an alternative approach is to first adjust the phas~ to minimize the central feature. Then shift the phase by 90 ~ by adding o:
17
G E N E R A T I N G THE E R R O R SIGNAL m
"0
10
09
0
09
-10
n,' '-
-20
C 0 O.
LL
J
ooi i o
-30
120
E3
v
09 r 0 El. 09
n~
9 09 r
r
t
,
,
T
i
106
2
3
4
5
T
~
i
,
I
678,tn7 iu
,
,
,
F
2
3
4
5
T
I
,
,
I
6784n8 IlL/
-120
n
Frequency(Hz) Fla. 1.7. Series LC traps at the mixer IF port can reduce the high-frequency components while adding very little phase shift within the servo bandwidth. The curves show measured data for a two-notch filter that adds only I~ of phase shift in the region from 1-10 MHz. Once the filter is installed in the circuit, tunable elements are used to set the frequency.
subtracting the appropriate time delay, which equals one-quarter of the modulation period. The signal will appear larger as the laser is swept past the cavity resonance more slowly, because there is more time for the light to build up in the cavity mode. The error signals for a system using EOM modulation and one using injection-current modulation are shown in Figure 1.8. Note that the error signal observed when current modulation is employed contains a large DC component. This is attributable to the strong intensity modulation (residual AM) that must be compensated at the integrator input. Subsequent changes of the optical system (alignment for instance) will change this DC level, and require some adjustment of the offset compensation. The optimum locking will be obtained when the highest possible signalto-noise (S/N) exists on the error signal. The error signal's discriminator will have a linear slope measured in units of volts per laser frequency, for instance V/kHz. The factors that determine this slope include the cavity's linewidth, the sideband amplitude, the power on the detector, detector gain (V/A), postdetector amplification, and the mixing process. The noise on this discriminator should be attributable to the laser frequency noise and ultimately at a much lower level by shot noise and detector noise. The best performance comes when the system is shot-noise-limited, i.e., the S/N is
18
STABILIZING
DIODE
LASERS
0.100 30 MHz
Sweep turn around
modulation
L -0.000
L
.......
r-
.
.
.
.
.
"--'...........-"~I. . . . . . . . .
EO Modulation
~..... -~ ' i - - ....... -~:i--: . . . . . . . . . . . . . -"
> C
-O.lOO o uJ
-0.200
-0.300
............
I
............
I ........ r
i
....
[ . . . . . . . . . [ ................................. [ . . . . . . . . l
I
. . . . . . . [ ................................ l . . . . . .
~
0.005
__
[
r ...... 1 .....
Current Modulation -.....
I
I
0.015
0.025
Time (ms) (From trigger point) FIG. 1.8. The low-pass filtered output from the mixer is the error signal. The local oscillator phase is set to maximize the amplitude, slope, and symmetry of the central frequency discriminator. Intensity modulation in the lower example results in a large DC offset, which must be compensated by adding a DC current to the integrator input.
limited by the intrinsic noise on the photocurrent. However this depends on whether the laser intensity fluctuations at the modulation frequency are at the shot-noise limit. If the laser exhibits technical noise in excess of the shot-noise level, noise reduction is possible by using a subtracting detector configuration. In this case, a fraction of the laser beam is sampled by a photodiode prior to the cavity, and the photocurrent is subtracted from the reflection signal. With proper attention to the phase of the two signals, much of the excess common-mode noise can be canceled. We will not dwell further on this approach, because ECDLs are usually within 10dB of the shot level at practical modulation frequencies. Furthermore, even though a little tighter frequency control is to be gained, there is less light available for the cavity ring-down signal. A good approach is to ensure that the noise on the signal reaching the mixer is indeed caused by the light. This is most easily checked by sending the RF input to a spectrum analyzer and monitoring the baseline noise levels
THE LOOP FILTER
19
near the modulation frequency. With the laser off-resonance (and the light level set to the approximate value that exists when the system is locked), one simply compares the noise levels with the light blocked and unblocked. If the noise does not increase when the light is unblocked, then the system is likely to be limited by the detector-amplifier noise and either the load resistance or the light level should be increased.
1.4 The Loop Filter The loop filter connects the error signal to the laser, thus completing the feedback loop. The goal of the feedback electronics is to supply enough gain to drive the laser's frequency fluctuations to the noise floor over as much of the servo bandwidth as possible. With proper design, the electronics should neither limit the residual frequency noise level, when the laser is locked, nor limit the maximum servo bandwidth. The residual frequency noise should be limited by the light's amplitude fluctuations (technical or shot noise), and the correction speed of the loop should be limited primarily by the laser chip itself. In this section we describe the construction of a loop filter that attempts to achieve these objectives. We start with a very simple introduction to servo system theory and then proceed to a concrete example, namely, a loop-filter design suitable for locking the frequency of a diode laser to a high-finesse cavity.
1.4.1 Introduction to Feedback Systems A short introduction to the stability of negative feedback systems may be useful a t this point. Our objective here is not to treat general control systems [15], but simply to provide some basic understanding and a few rules-of-thumb to aid the construction of a feedback loop to lock a diode laser. With that in mind, we note that a feedback loop supplies corrections to some device (e.g. the laser) within some electronic bandwidth. By "loop" we refer to a signal path that goes from the laser to the cavity, to a detector, to an electronics gain stage, and back to the laser. At low frequencies, the loop will have very high gain, so that small deviations of the error signal are greatly amplified and the device is forced to move in the proper,direction, effectively suppressing the deviation. In general, the perturbations will be reduced by the factor of [l+A(f)] -1, where A(f) refers to the system gain at the Fourier frequency f. However, consider a small sine-wave perturbation of the laser frequency occurring at a somewhat higher rate, say 100 kHz. As this signal traverses the loop,
20
STABILIZING DIODE LASERS
various components will shift its phase. For instance, a simple low-pass filter made from a resistor and capacitor will begin to introduce a noticeable phase shift at one-tenth of the filter - 3 d B frequency, (2rcRC) -1. As the signal frequency increases, the filter starts attenuating the signal, and the resultant phase shift increases. This phase shift stops accumulating at a frequency about ten times (2TcRC)-1, with the filter output lagging the input by 90 ~ Primarily as a consequence of the various loop components that act as low-pass filters, the correction signal will be phase shifted from the original perturbation. (Any additional time delay, such as lengths of cable, will further increase the phase lag.) We can safely predict that at a high enough frequency, the correction signal will in fact be 180 ~ away from the negative feedback; in other words, positive feedback! In this situation, if the correction signal causes the laser to move more than the original perturbation, an oscillation will of course quickly develop. To avoid this scenario, one must ensure that the loop gain is less than 0 dB, or unity gain, in the frequency range where the phase shift of the loop is approaching 180 ~. The phase margin of the loop refers to exactly how far (in degrees) from 180 ~ the phase shift is when the loop gain falls below unity. If the phase margin is less than about 45 ~, the system will tend to oscillate. In the time domain, insufficient phase margin is associated with an under-damped system, and a ringing response to transients will be observed. The task of the servo design is to ensure that there is sufficient gain at low Fourier frequencies while keeping the unity-gain frequency low enough such that the loop is stable. This naturally leads to a gain versus frequency dependence with negative slope. In fact, a plot of this dependence is a standard part of the servo designer's toolbox. These plots, called Bode plots, are drawn with log-log axes to simplify the interpretation of the various shapes. An example of such a plot is shown in Figure 1.9. A simple picture of the open loop transfer function is the gain as a function of frequency that a signal would experience if it were injected into the diode laser and detected after it travels through the whole system. The bandwidth of the loop refers to the frequency at which the gain falls to 1 (or 0 dB). This is commonly called the unity-gain point, and is often in the range of several megahertz for a servo controlling a diode laser via the injection current. A rule-of-thumb is that at the unity-gain point, the slope of the loop transfer function may not be decreasing by much more than 20dB per decade of frequency, which corresponds to a factor of 10 per decade or an f - 1 slope on the log-log plot. At lower frequencies the slope may be steeper, as shown in Figure 1.9, but the transition from a steeper slope to the f - 1 slope may not occur near the unity-gain frequency. This model
21
THE LOOP FILTER 120 110
Cavity Av
100 S
90 rn
80 Integrator switch closed
"o
60
Integrator RC roll-off
70
40 dB/decade slope l.-- 40
20 10
Unity gain -10 102
r
r
rsle+]
2 3 4 567 103
t
q
T
,
~ , ,,
2 3 4 567 104
................ 2 3 4 567 105
2 3 4 567 106
2 3 4 567 107
Frequency (Hz) FIG. 1.9. The loop gain transfer function, or Bode Plot, shows loop gain as a function of frequency. Transitions to a steeper slope are caused by RC roll-off elements and the optical cavity. The transition from -40dB/decade to -20dB/ decade near 500 kHz is caused by a differentiating stage. Note that the loop crosses the 0 dB axis with a slope of 20 dB/decade.
transfer function fulfills the requirements discussed earlier, namely, large gain at low frequencies while maintaining sufficient phase margin at unity gain to protect against loop oscillation.
1.4.2 Characterizing the Loop Transfer Function We attempt to achieve this stable transfer function by compensating the response inherent in the actual components such as the laser, cavity, and detector. First, the transfer function of the elements in the loop (except for the loop filter itself) needs to be estimated and/or measured. Then a loop filter is designed using op-amps and filters to provide gain and adjust the shape of the actual loop to approach the desired form. In this section we will investigate the loop transfer function before the addition of the loop filter, while the next section focuses on the loop filter itself.
22
STABILIZING DIODE LASERS
The two elements that should have the greatest effect upon the shape of the transfer function are the cavity and the diode laser. Other elements such as the detector have been designed to have minimal effect on the loop transfer function. (We also assume that the cables in the loop path are as short as possible to minimize loop time delay.) As mentioned previously, the reference cavity acts as a frequency discriminator at frequencies below the cavity linewidth and as an integrator ( f - 1 slope) for frequencies above the cavity linewidth. The transfer function of the diode laser itself is considerably more complicated and varies significantly among devices. However, most common low-cost diode lasers are similar in that the phase response rolls off sharply near 1 MHz. As we will see, to compensate this phase roll-off we use phase-lead circuits. While we often determine the component values for the compensation of the laser empirically, it is certainly possible (and probably preferable when time and equipment are available) to measure the complete phase and amplitude transfer function of the laser's frequency modulation response. This can be accomplished with a frequency discriminator such as that provided by the side of an optical cavity resonance. A plot of the laser's transfer function can be obtained by modulating the current of the laser and recording the amplitude and relative phase of the reflected or transmitted intensity modulation. A number of issues must be addressed in order to make sure the measured phase shift actually represents the laser's phase response. The detector's response must be known, along with that of the frequency discriminator [16]. The laser frequency must loosely be controlled such that it remains on a linear portion of the fringe during the measurement. Also, the laser's amplitude modulation must be removed from the measurement, which can be done with a subtracting-detector arrangement. Careful adjustment and matching of the DC intensity of the beams can be avoided by using negative feedback to match the photocurrents at low frequencies [17]. Although we have mentioned the shape of the servo loop's response curve, another quantity that we need to determine is the amount of gain required from the loop filter. Insufficient gain could easily bring the 0 dB point (the unity-gain point) up to the steep portion of the gain curve (i.e., the f - 2 portion of the curve), rendering the system unstable. Conversely, too much gain could push the unity-gain point up to a frequency with insufficient phase margin, again rendering the system unstable. The gain of the loop depends on the slope of the discriminator and on the modulation characteristics of the laser Rself. For Fourier frequencies less than the cavity linewidth, the discriminator slope is independent of frequency, and can be calculated by slowly sweeping over the cavity line while observing the error signal. If you know the cavity linewidth, simply
THE LOOP FILTER
23
divide the peak-to-peak error-signal magnitude by the cavity linewidth to obtain an estimate of the slope in volts per hertz. 1 The modulation function of the laser can be determined at a single frequency (we suggest 10 kHz) by driving the laser current with a sine-wave function generator applied through a coupling resistor. (Our circuit, shown later, is coupled to the laser via a 2 kS2 resistor.) A measurement of the deviation of cavity resonance will then result in a hertz per volt measurement of the laser modulation characteristic at the 10 kHz rate. 2 When multiplied together, these two measurements yield a dimensionless number, call it A', that indicates the loop gain at 10 kHz without the loop filter. Expressed in decibels, this would be 20 log[A]. One can then use this to scale the overall gain needed to achieve the desired transfer function shown in Figure 1.9. As we will see in the next section, there are many ways to adjust the overall gain in the circuit.
1.4.3 Loop-Filter Electronics The loop transfer function described in the previous section is realized by building the proper electronic loop-filter circuit. Fortunately, a fairly simple circuit based on op-amp electronics is sufficient to provide the gain and shaping required for locking a diode laser. The circuit diagram for the loop filter we have implemented for our test laser is shown schematically in Figure 1.10. The basic feedback circuit consists of a first stage op-amp that filters and amplifies the error signal coming from the mixer. The output from the first stage is split into two channels, a fast channel that goes to the laser's injection current, and a slower channel that goes to the PZT controlling the length of the extended cavity. The circuit also contains resistors and capacitors used to shape the transfer function and control the overall gain, as well as six switches (these can be controlled manually or by TTL signals) that are used to enhance the performance of the servo system.
lIf the approximate cavity linewidth is unknown, an oscilloscope with cursors can be used to measure the time between peaks of the central discriminator. Then the measurement can be calibrated using the known frequency offset of the sidebands as a frequency scale. 2Care is necessary to avoid electrically damaging the laser. It is safe to connect the function generator while the power is on, but the output should be adjusted to its lowest setting. With the laser frequency sweeping slowly so that the cavity resonance is visible, one can turn up the drive voltage until a measurement of the deviation can be made with calibrated oscilloscope cursors.
Laser Current Source
39 p~
82 Q
Laser
To PZT Amplifier
FIG.1.10. The loop-filter circuit. The switches across the first amplifier and the PZT amplifier are used for increasing the DC gain, and the input and output switches can be used to rapidly turn the servo off and on. The offset on the second-stage amplifier is used to bias the output such that the unlocked laser is not perturbed by the circuit. See text for circuit details, most of the switches shown are optional.
THE LOOP FILTER
25
Let us consider this circuit in more detail, with the initial assumption that all the switches are closed. The first stage consists of an op-amp with parallel R C feedback channels. At DC, the effective op-amp gain is 220K/1K or 46.8 dB, but at the frequency (2TcRC)-1 or 8 kHz, the gain starts decreasing and falls as f -a. This stage is responsible for one half of the - 4 0 dB/decade slope shown in our model transfer function in Figure 1.9. The remaining half is attributable to the cavity. (As mentioned earlier, the cavity roll-off that occurs beyond the [5 kHz] cavity linewidth contributes a f -1 or - 2 0 dB/ decade response.) Additional shaping of the transfer function with the first stage is achieved by including a capacitor in parallel with the input resistor. The result is a decreasing input impedance starting at the frequency (2~RC) -1 or 480 kHz for the values shown. The purpose is to bring the system gain slope from a f - 2 or - 4 0 dB/decade slope to a f - 1 or - 2 0 d B / decade slope, prior to the unity-gain frequency. Note this inflection point is shown in Figure 1.9. A much smaller resistor is included in series with the input capacitor in order to limit the very high frequency gain. Depending on the op-amp used, this resistor may not be necessary, and in any event should be as small as possible. An extra boost in low-frequency gain (below Fourier frequencies of 8 kHz) can be achieved by opening switch $3, because then the op-amp functions as an integrator. In fact, the gain at lower frequencies increases as f - a until the amplifier "open-loop" gain is achieved. This level will depend on the particular op-amp and supply voltage, but could easily be 100 dB. Thus, along with speed and noise performance, the open-loop gain should be considered when selecting an op-amp. Because of the increased gain at low frequencies, opening the integrator switch $3 forces the error signal ever closer to zero. However, with $3 open, if the laser is not locked to a cavity mode then amplifier "a" will integrate to a voltage supply rail and relocking will not necessarily occur. With $3 closed, the gain is moderate and the system can respond if the laser frequency is tuned near a cavity mode. Therefore, a good strategy is to acquire the locked condition with $3 closed; one can then open it after the lock is established. In Figure 1.9 we show (with a dashed line), the system gain through the injection current channel with $3 closed. We emphasize however, that this transfer function does not include the low-frequency gain supplied by the PZT channel. As we will describe, for Fourier frequencies under 1 kHz the PZT channel gain becomes significant and is in fact dominant near DC. Another feature of the first stage is the external offset current that is fed in via switch $2. The op-amp itself should be balanced using the standard offset compensation circuit that can be found in the amplifier datasheet. We have chosen to include an additional offset current channel because the circuit may be used with injection current modulation instead of an
26
STABILIZING DIODE LASERS
electro-optic crystal. In this case, substantial current offsets are necessary to compensate for the laser's amplitude modulation. We stress once again that current modulation is not recommended if drift of the locked laser relative to the exact center of the cavity is critical, because the amount of AM from the laser typically changes with time. The voltages that the offset current is derived from (shown in Fig. 1.10 as 4-Vo) should be well filtered and no larger than necessary to compensate for the mixer offset. The output from the first-stage amplifier is then sent through separate paths to the two correction elements. First let us consider the fast path, which passes through a variable attenuator and goes to op-amp "b" that feeds current to the laser diode. If gain beyond that provided by the first stage is needed, one can appropriately set the feedback resistance on op-amp "b." Otherwise, one can simply omit op-amp "b," which will enable a slight increase in the attainable servo bandwidth. Regardless, the resulting "fast" correction signal is sent to the diode laser through switch $6 and a 2 kS2 resistor. This resistance together with any capacitance inherent in the laser will form a low-pass filter, possibly contributing phase lag that would affect the system bandwidth. Another possible coupling configuration would be to use a smaller coupling resistor and limit the voltage of the amplifier output, but this is not recommended as it could expose the laser to dangerously large current spikes. We emphasize here that caution is required to avoid damaging the laser, because the feedback is coupled directly through the resistor to the laser anode or cathode (depending on the diode's polarity). Note that, although Figure 1.10 is drawn with a coupling to the anode of a positively biased laser, no changes are necessary to couple to the cathode of a negatively biased laser. At the end of this section, we describe how to use $6 to protect the laser against potentially fatal current spikes from the loop-filter electronics. In order to provide phase-lead to compensate the diode laser, we have added parallel capacitance on the second stage input resistance and on the coupling resistor to the laser (see Fig. 1.10). While the optimum values for these components will depend on the particular laser, the values shown are probably good starting points. In addition, in both of these places and on the first-stage input there is also a second resistor in series with the capacitor. These resistors will limit the circuit's highfrequency gain in the region above 30 MHz. The purpose is to prevent parasitic loop oscillations that can develop at very high frequencies. Again the values shown are probably good starting points, although decreasing the resistances may allow a wider bandwidth as long as the loop remains stable. Also included on the input to op-amp "b" is an offset control. Its purpose is to bias the second amplifier's output such that under steady-state
T H E LOOP F I L T E R
27
conditions (or for the unlocked state) there is very little current through the coupling resistor. In other words, we adjust the pot so that the amplifier's output is close to the laser's junction voltage. This gives the circuit the advantage that it can be connected to the laser by closing switch $6 without appreciably perturbing the laser frequency. The second path, or "slow" feedback, goes to op-amp "c" that controls the voltage sent to the PZT driver for the laser cavity end mirror (see Figure 1.1). The advantage of this configuration is that the low-frequency corrections are dominated by the large gain in the PZT channel so that the laser current DC level remains constant (large changes in this current can lead to laser mode hops). In fact, under the locked condition the DC level after the first stage should be equal to zero when this stage is configured as an integrator. This highlights one of the aspects of using integrators in these feedback loops--a nonzero correction can be applied (after amplifier "c" in our circuit) while a zero-error signal can be maintained. We also benefit from the increased low-frequency gain in the overall loop transfer function. For added flexibility in this stage we have included a switch, $5, across the feedback capacitor, so we can switch between flat gain and a full integrator (as we did with $3 on the first stage). We have assumed that the PZT driver has a polarity switch, so one can find the right sign for the feedback. If the PZT driver does not have a polarity switch, a simple inverting amplifier following the integrator may be required. We note also that for more accurate locking (a smaller offset from the center of the cavity resonance), amplifier "c" should have its inherent offset adjusted prior to any input. This can be done with a trim pot wired as shown in the amplifier datasheet. This adjustment is not important for the second-stage amplifier "b," because the gain of this stage is small relative to that of the first stage. We control the overall loop gain with a potentiometer that varies the attenuation between the first and second amplifier stages. A one-turn pot is used to minimize stray capacitance that might limit the feedback bandwidth. Care should be taken to minimize the length of the signal path by positioning the pot close to the amplifiers, preferably by mounting it on the board. The choice of amplifiers for the various stages is based on bandwidth, noise density, and offset performance. For the two amplifiers in the current correction path ("a" and "b"), we want to have amplifiers with little time delay (recall that time delay leads to unwanted phase shifts) and good noise performance. Thus we usually work with op-amps with gainbandwidth products > 30 MHz and a noise level substantially less than that of the input signal. Of course, the level of the input noise most likely depends upon how much light is incident on the detector (recall the
28
STABILIZING DIODE LASERS
discussion in Section 1.3.3). For light power above 100 gW, the noise requirement on the op-amp is not severe and should not present a problem. The op-amp in the PZT channel ("c") does not need to be fast, so a standard field-effect transistor op-amp with good DC characteristics should suffice. As we have already seen, there are several analog switches in the circuit, which we use to increase the low-frequency gain, and to allow quick, remote unlocking and relocking via TTL signals. Most of these switches are optional, with the exceptions being the PZT integrator switch $5 and the output protection switch $6. However, these may be simple manual toggle switches. The TTL-controlled switches are available from many manufacturers with many different features. Because speed is relatively important, we implement switches that can respond in about lOOns. Low-leakage current is also important in the integrator configurations; the switches we chose are specified at 100pA. Minimizing switching transients is another consideration, with the pertinent specification showing up on many datasheets as "charge injection." In order to maximize performance and flexibility, it is important to consider the layout of the electronics. The electronics board should be located close to the laser to avoid unnecessary time delays and cable capacitance. To achieve higher loop bandwidths we typically place the loop filter in a small box adjacent to the laser on the optical table. We have also chosen to place the coupling resistor and capacitor to the laser on the electronics board instead of physically placing them next to the laser. This allows an additional TTL-controlled switch to be placed on the laser-side of the coupling resistor, for the purpose of rapidly shunting a few milliampere of current away from the laser to quickly shift the laser frequency. Such switching may be useful in cavity ring-down spectroscopy, and is discussed in the following sections. We conclude this section with some words of caution. The output switch $6 is necessary in the present implementation in order to protect the laser against current spikes from the loop-filter electronics during cable connect/disconnect episodes or power up/down. When connecting or disconnecting the cable between the circuit and the laser, one first opens $6. Similarly, before turning the circuit power off, it is wise first to open this switch and disconnect the laser. While there are certainly more elegant solutions that might be included to protect the laser without contributing unwanted phase shift in the signal channel, we have found that this approach works fine in the laboratory setting. A more robust solution may be necessary, however, if the circuit is to be handled by less-experienced users.
LOCKING THE LASER AND LOOP OPTIMIZATION
29
1.5 Locking the Laser and Loop Optimization 1.5.1 Initial Locking With the loop filter designed, all the pieces are in place to lock the laser to a cavity fringe. It is easiest to start by sending a triangle wave to the laser's or cavity's PZT driver to sweep slowly over a cavity free-spectral range and view the transmitted signal on an oscilloscope. A sweep rate of ~ 3 0 H z is a good compromise between sweeping too quickly which reduces the power buildup in each mode, and too slowly which makes alignment difficult. First we need to compensate any DC offsets in the signal coming from the mixer. With the laser nominally aligned as described in Section 1.3 and the output switch $6 open, the signal at the output of op-amp "a" should yield a low-pass filtered version of the error signal, centered at zero volts DC. Any offset from zero can be adjusted with the offset current through switch $2. We usually check for locking action by closing switch $6 and turning up the gain knob (with the PZT feedback gain turned off for the moment). An increase in transmission, widening of the resonance, or even a flat section on top of the fringe are all indications that the polarity of the system is correct and the feedback loop is attempting to keep the laser locked to the fringe (see Fig. 1.11). If the overall sign of the servo is wrong, the fringe will be suppressed and the system will try to lock to the sidebands instead. This situation can be corrected by choosing the opposite slope for the discriminator (e.g. by changing the demodulation phase by 180~ Adjusting the overall gain by turning the potentiometer will give some clues as to whether more gain is required or not. If the servo seems to be trying to lock, one can then reduce the sweep to zero and the laser should stay in lock for increasing fractions of the sweep. If this is indeed the case, now is a good time to optimize the alignment to the optical cavity, because the sweep-to-sweep amplitude variations will be greatly reduced. If the servo clearly perturbs the laser but no locking action is observed, even on the sidebands, it is possible that the gain is too low and the servo is unstable. Remember that a portion of the servo loop's gain curve is decreasing as f - 2 , or - 4 0 dB per decade. If the overall loop gain is such that the servo's unity-gain point occurs on this steep slope then no locking will occur. Determining the system gain as detailed in Section 1.4.2 will offer a clue as to whether this is the problem. When the laser is locked to the cavity, increasing the gain past the optimum point will cause a servo oscillation that will be evident on the error signal. Another indication that this is occurring is a reduction of
30
STABILIZING DIODE LASERS 0.5
--
Sweep Up
-------------
Sweep Down
-
1
O --h
"~
0.4
O
V
E
......
~
Reflection
.5 O
v
t_
0.3
Q_ "o 0.2
E c
0
f
O
0.1
t_
!-
0.0 I
2
~
I
4
,
A I
6
O O
3 v
Transmission I
I
8 Time (ms)
I
I
10
i
I
12
14
Fla. 1.11. Traces of the laser power transmitted through, and reflected from a highfinesse cavity as the laser frequency is swept up and down. The injection-current portion of the electronic servo is on, and briefly locks the laser to the TEM00 mode as the laser passes the resonance from each direction. Here the asymmetry is attributable to the offset adjustment of the high-gain stage in the feedback loop, but similar behavior is possible from other causes (see text). the power transmitted through the cavity as power is taken out of the carrier and put into oscillation sidebands. Reducing the gain slightly (just below oscillation) provides a good initial setting. In a similar way, the PZT gain can be increased until instability ensues and then reduced slightly. If turning on the PZT gain drives the laser out of lock, the sign of the PZT feedback may need to be reversed. With the PZT gain turned up, the signal observed after the first amplifier stage ("a" in Fig. 1.10) should be driven to zero. As an aside, we note that with high-finesse cavities there is often an asymmetry in the locking performance in regard to the direction of the frequency sweep (see Fig. 1.11). This usually is the result of a nonzero offset in the locking electronics. Such asymmetries can also be caused by localized heating of the high-finesse cavity mirrors, caused by absorbed light as the mode power builds up. In this case, the thin films will expand and the cavity resonance will shift slightly toward higher frequencies. This effect is similar to the nonlinearity observed with gas-filled cavities [18]. Regardless of the cause, the net result is that the effective sweep rate of the laser relative to the cavity is effectively higher in one direction than the other and/or the locking range is reduced. Consequently, the laser spends less time locked to the cavity, possibly to the extent that maximum transmission is never achieved.
LOCKING THE LASER AND LOOP OPTIMIZATION
31
1.5.2 Adjusting the Loop Parameters In the previous section we identified the lock condition simply by observing the light level transmitted through the cavity. In order to optimize the feedback system, it is necessary to have a better diagnostic. Fortunately, much of this information can be obtained from the same error signal one uses for the locking. A directional coupler placed just before the mixer can be used to extract a small fraction of the error signal without significantly changing the overall loop gain. One can then use an RF spectrum analyzer to view the noise sidebands, which are centered at the modulation frequency. The spectrum analyzer display should resemble the trace shown in Figure 1.12, which consists of a center peak at the modulation frequency with noise sidebands on either side. If the gain is increased, the sidebands will increase in amplitude as the servo starts to oscillate; if the gain is then reduced slightly so that the system is well behaved, the width of the noise spectrum is an indication of the servo bandwidth. One can now attempt to increase the bandwidth and reduce the level of the low-frequency noise by changing component values in the loop-filter circuit. We recall that there are three resistors in the circuit of Figure 1.10 that serve as convenient points to place a parallel capacitor, thus creating a high-pass element with its corresponding phase advance. These resistors include the first- and second-stage amplifiers' input resistors, and the coupling resistor to the laser diode. One can use the spectrum analyzer display to monitor the error signal's noise spectrum and adjust these component values in an iterative fashion to obtain the best performance. It is worth spending some time on the process by changing components and values to arrive empirically at a wider system bandwidth. A simple technique that often helps is to place a finger on one of the phase-lead capacitors, thereby increasing the actual capacitance slightly. This may offer a clue as to whether more or less capacitance is desired. Not only can the capacitor values be adjusted, but different op-amps can also be inserted to determine whether amplifier speed is playing a role in the system bandwidth. Similarly, the length of the cable between the detector and mixer can be adjusted to determine whether time delay is an issue. A spectrum analyzer monitoring the error signal can be useful in improving the performance of the servo loop as illustrated using our demonstration system. Figures 1.12a and 1.12b show a noise spectrum for the locked ECDL using EOM modulation. These data are referenced to the mixer input, and are measured by using a - 1 0 dB coupler prior to the mixer and subsequently correcting the data plots by 10dB. The lowest level depicted in Figure 1.12a is the laser's amplitude noise level. This represents the lowest level to which we can reduce the frequency noise--further gain
32
STABILIZING DIODE LASERS
only writes amplitude noise onto the frequency of the laser. The AM noise was measured by unlocking the system and attenuating the beam reflected from the cavity such that the power incident on the detector was the same as when the system was locked. The signal when no light was incident on the detector is not shown, but is approximately 10 dB below the light level. A number of technical noise spikes are evident in the laser's AM spectrum, along with a peak at the modulation frequency, 29 MHz. The top trace shows the error signal, which contains several important features. One can see at 4-4 MHz from the center, large "servo bumps," which are characteristic of the servo system. Here the phase for the correction signal is approaching 180 ~ (positive feedback), so the noise is increased rather than suppressed. If one turns up the loop gain, the system will oscillate (displaying tall, sharp peaks) in this region. The oscillation frequency indicates roughly the bandwidth of the servo--if the bandwidth is much less than anticipated, then further evaluation of the loop filter and laser/detection system is needed. Not surprisingly, we have observed that the bandwidth can change with the laser's operating parameters. This is to be expected, because we know that the frequency noise spectrum of the laser -50 ,~,
-60
-r" 1::
-70
N
rn
-80 ::3 Q. t.m
-90
x -100 E -,--' -110
0
n
i,_
~6 oo
System Locked
-120
-130 Laser A M
-14o -15o -160
,
I
-10
,
I
-8
~
I
-6
~
I
-4
]
I
-2
T
I
-0
~
I
2
~
I
~
4
I
6
t
I
8
r
I
10
Frequency (MHz) (relative to 29 MHz)
Fla. 1.12a. The noise spectrum at the mixer input, measured with a RF spectrum analyzer using a 100 kHz resolution bandwidth. The noise sidebands are peaking at 4 MHz, indicating the approximate servo bandwidth nearly this wide. The noise with the laser unlocked and shifted away from the resonance is shown in the lower trace. An expanded scale is shown in Figure 12b.
33
L O C K I N G THE LASER AND LOOP OPTIMIZATION -50 .-. N ZE "~ rn "o -~, 23 r~ c"
-60
! !
-70 -80 -90
.--
Q) X -100
E
N
-110
I..
0 a. --
o El.
O0
System Locked
-120 -130
Flashlight
-140 -150 -160 -250
t -150
i -50 Frequency (kHz)
t 50
i 150
J 250
(relative to 29 MHz)
FIG. 1.12b. The noise spectrum at the mixer input, under the same conditions as Figure 1.12a, but measured with a 3 kHz resolution bandwidth. The lower trace, generated by a flashlight, shows the noise fluctuations of a shot-noise-limited photocurrent of the same DC level as when the system is locked. The laser's amplitude noise shown in the preceding figure is essentially at the same level. Note that the error signal approaches the intensity noise at low frequencies, indicating that the system is suppressing frequency fluctuations to the shot-noise limit in the kilohertz range. changes as the laser sees optical feedback, or approaches a mode hop, for instance. An increase in the magnitude of the F M noise spectrum would be accompanied by an increase of the laser's F M modulation coefficient, which in turn will affect the servo loop's gain. Similarly, optical feedback characteristics (e.g. alignment of the feedback mirror) of the laser cavity will change the operating parameters, so tweaking up the alignment can sometimes restore the locking performance. The error-signal noise at low frequencies appears to be well above the amplitude noise in Figure 1.12a, however this is a measurement artifact caused by the wide spectrum analyzer bandwidth. Taking a closer look, as in Figure 1.12b, the error-signal noise when locked is approaching the amplitude noise at approximately 20 kHz from the carrier. At even lower frequencies where the servo has higher gain, the error-signal noise will appear to be less than the amplitude noise level, however this is simply an indication that the laser is being frequency modulated by the amplitude noise. Note that we have calibrated our noise measurements to the shot-noise-limit with the aid of a flashlight to supply a shot-noise-limited photocurrent. As in the case
34
STABILIZING DIODE LASERS
of the laser amplitude noise measurement in Figure 1.12a, the shot-noise measurement was accomplished by adjusting the DC photocurrent to the same level as exists when the system is locked. The data traces show that the laser's AM noise is essentially at the shot-noise level in the region near 30 MHz. We noted the power peak at the modulation frequency in Figure 1.12a. There is almost always some residual power detected at the modulation frequency because of RF pickup by the laser cables or housing. Another possible source of AM power at the modulation frequency that often occurs is attributable to improper alignment of the modulator crystal with respect to the laser's polarization axis. However, the residual level attributable to these sources is indicated by the lower trace. The coherent spike increases by 20 dB when locked, and is most likely caused by one or more transverse cavity modes near the modulation frequency. The affected sideband would be slightly phase shifted if any power was coupled into a transverse mode. Consequently, the RF modulation detected on the cavity reflection would not cancel at the exact center of the cavity. For applications in which the small offset from the center of the cavity resonance is important, attention to the proper choice of modulation frequency, cavity geometry, and spatial coupling is required. Note that the coherent peak is still a very small fraction of the power at the mixer input when the laser is unlocked but near the cavity mode. The power when the system is unlocked may be measured with an oscilloscope as the laser is swept over a cavity fringe, and for this system it is on the order of 0dBm, or - 5 0 dBm/Hz when referenced to the 100 kHz spectrum analyzer bandwidth of Figure 1.12a. At this point the performance of the laser lock should be adequate for most applications that require narrow linewidths and high spectral resolution. It is now appropriate to expand the capabilities of the lock and start to take advantage of the switches to make the lock more robust and more flexible.
1.5.3 Evaluating the Locked System With the servo working reasonably well, one might be curious as to the linewidth of the locked laser and how to measure it. This problem is actually trickier than one might guess. The best way is to beat it against a more stable laser on a photodiode and evaluate the resulting fluctuations. The strength of this technique is that it reveals both of the major contributors to the laser's frequency-noise spectrum: the residual noise of the laser lock, and the fluctuations of the cavity length. However, this requires access to a more stable laser of the same color. One can of course build a replica of the first locked laser system and make a comparison. An independent Fabry-P6rot
LOCKING THE LASER AND LOOP OPTIMIZATION ,
35
cavity can also be used as a frequency discriminator. However, both of these approaches will suppress any noise that is common-mode to the two systems under comparison, so the resulting beat-note noise spectrum may be artificially clean. This is a serious problem because a frequency fluctuation of kilohertz magnitude corresponds to a vibration of the cavity length below the picometer level (25cm cavity length, 800nm wavelength). The selfheterodyne techniques involving long optical fibers are impractical, given the long coherence length of a laser locked to a high-finesse cavity, and the additional phase noise caused by the fiber. For many applications including evaluation of the lock itself, it is useful to measure the noise within the servo loop relative to the cavity. It is very important to emphasize that measurements of the noise within servo loops are not good indications of the actual laser linewidth or lineshape. One reason is that the servo gain acts to minimize the noise at the integrator input, even if the noise is not caused by frequency fluctuations. For instance if some of the signal fluctuations at the integrator input are actually amplitude fluctuations, the servo will suppress this signal by increasing the frequency fluctuations. Nonetheless, the noise measurements are useful in determining the performance of the electronics, transducers, and servo components. In the absence of excess noise of the electronics or cavity (usually not the case), one can make an estimate of the performance of the servo system. In particular, consider the case in which the noise spectrum is flat out to some bandwidth, and whose noise level is small compared to that bandwidth (typically the case for a well-stabilized laser). Then the contribution of this noise to the laser line shape is a Lorentzian with a linewidth equal to A v = rc(Av~ms)2/B, where A vrms is the root-mean-squared fluctuation derived from the spectrum analyzer noise level divided by the discriminator slope, and B isthe bandwidth for the spectrum analyzer measurement. Many spectrum analyzers give the noise level as spectral power, in dBm/Hz; one needs only to.convert dBm to volts 2 and then divide by the square of the discriminator slope to convert the measurement into hertz. While our noise spectrum is not usually completely flat, one can nonetheless use this formula to make estimates that are reasonable enough, especially in light of the fact that this approach neglects fluctuations of the cavity itself, which for most good locks, probably are the dominant noise source. There are also mathematical techniques for mapping spectra that are more complicated into laser line shapes [19, 20].
1.5.4. Making the Lock More Robust While the method described thus far should produce a good laser lock, there are several minor modifications we can add to make the locking
36
STABILIZING DIODE LASERS
system more robust and reliable. In particular, the system will be easier to work with if it can keep the laser locked in the face of moderate disturbances and quickly recover from more severe events. Increased servo gain is the answer to acoustic perturbations, as the loop will reduce the error-signal fluctuations by nearly the inverse of the gain. For examPle , at acoustic frequencies the system gain may be 100 dB, or 105. The frequency noise of the laser caused by voices will be reduced to 10.5 of what it would have been in the unlocked case. More challenging problems come with sharp raps on the mechanical system caused, for instance by dropping a tool on the optical table, as the system has much less gain at a few hundred kilohertz. As a first step toward reducing the sensitivity to such effects, it is advisable to use solid mechanical structures for the optical components in the system. Good vibration isolation of the reference cavity and the laser box will reduce the instability of the system and hence the servo requirements. Other effective passive approaches include the use of robust mounts (i.e. no laboratory posts) for the mirrors and beam splitter. Even with good mechanical mounting, the system will still be susceptible to out of lock events caused by sharp mechanical perturbations. Here the inclusion of two electronic switches in the control circuit, one manual, and one electronic, serves to improve the lock and make it more robust. The first of these, $5 in the circuit diagram, simply turns the op-amp "c" from a proportional gain stage to a pure integrator. This increases the lowfrequency gain and drives the signal after the first stage to zero as desired. During acquisition it is preferable to have this switch closed (i.e. integrator off); otherwise, this op-amp will sit on the rail (because of even tiny offsets, an integrator will drift to rail quickly when the laser is out of lock) and make acquisition unidirectional at best. When lock is achieved, one can simply flip the switch to increase the low-frequency gain. A second switch, $4 in the circuit diagram, can be inserted to make the lock considerably more robust. This switch may be configured so that it simply turns off the input to the PZT integrator when the laser momentarily goes out of lock for any reason. The TTL signal that controls the switch is derived from a comparator that monitors whether the transmission through the cavity (i.e. the voltage from a detector placed after the cavity) is greater than a reference voltage (set to be roughly half of the detector voltage when the laser is locked to the cavity). When the laser is pulled into resonance by the current, the switch is closed and the PZT channel is turned on. If a perturbation momentarily knocks the laser out of lock, the switch automatically opens and the value on the capacitor is held, so that when the perturbation is over the laser frequency should be close to the previous value and relocking to the cavity can occur easily. We have found that the addition of this switch will keep the laser locked in the presence of fairly
REPETITIVE LOCKING FOR CAVITY RING-DOWN SPECTROSCOPY
37
significant noise. This makes it possible to work on the table without knocking the laser out of lock. Note that if the laser stays out of lock for more than a short time, the integrator will start to drift away from the "lock" value toward a rail and relocking is less likely. To reacquire the lock, usually one needs to close $5 to turn off the integrator. In principle, this can pull the laser frequency quite far from that of the desired cavity mode, especially if the previous "lock" value was large. We have found that with occasional monitoring, this correction voltage from the amplifier can be kept near zero simply by tuning the DC knob on the PZT driver. Then relocking to the same cavity mode is usually straightforward. During momentary unlock events the input switches (S1 and $2) to op-amp "a" could also be opened in a manner analogous to using $4 as discussed above. This would allow the amplifier feedback switch $3 to remain open, in integrator mode. However, the RC time constant of this amplifier is much shorter than amplifier "c," and our experience has been that relocking is much less likely if switch $3 remains open. There are other approaches to building "smart" servos that automatically relock the laser. For a processor-controlled system, it is possible to control the locking electronics and the laser in such a manner as to mimic manually relocking of the laser. Alternatively, this can be achieved using analog electronics. For instance, one approach upon an accidental unlock is to sweep the laser back and forth to once again "find" the cavity mode. Our implementation uses a single-pole, double-throw switch on the input to the first op-amp, in place of S1. The normally closed switch setting would connect the mixer output to the op-amp input. The other switch position connects the op-amp to a small oscillating current, generated by a low-pass filtered 555-timer circuit. When the cavity transmission falls below threshold, S1 is set to sweep the laser. Once the laser relocks, S1 is switched and the sweep is no longer sent to the circuit. This works quite well, as pounding on the optical table will cause the laser to unlock, reacquire, and lock within a fraction of a second. The circuit is shown in Figure 1.13.
1.6 Repetitive Locking for Cavity Ring-Down Spectroscopy We conclude this chapter with some modifications to the locking circuits that make it well suited for cavity ring-down spectroscopy (CRDS) experiments. Many CRDS measurements are realized with pulsed lasers
Error signal from mixer IF port To amplifier "a", Figure 2.10
* ,qK TRIG
DIS
CVolt
- THR
\4
I$+zQ L
Cavity Transmission
d
1M
U
3 Way Switch
FIG. 1.13. An analog solution to automatically relocking the laser after it becomes unlocked because of some perturbation. The cavity transmission is monitored with a threshold circuit, and when the laser is off-resonance this circuit injects a small, slowly sweeping current generated by a 555 timer into the feedback electronics. The three-way switch can be set such that the input stage always receives the error signal, or always receives the sweep current, or receives the error signal only when some light is transmitted by the cavity.
REPETITIVE L O C K I N G FOR CAVITY R I N G - D O W N SPECTROSCOPY
39
or continuous-wave (CW) lasers that are swept through the cavity fringe. CW locking of the laser to the cavity has advantages that have been previously recognized and discussed in the literature [21-24]. These advantages come from the ability to excite a single-cavity mode, and with more power than is possible by simply sweeping past the mode. Such experiments use repeated locking and intentional unlocking to generate ringdown decays from the cavity, improving the signal-to-noise through averaging. While we briefly discussed relocking issues in the previous section, the process of rapid unlocking and relocking of a laser to a cavity requires a more detailed understanding of the reacquisition process. In this final section, we will discuss repetitive locking and show how the addition of a few more switches to the loop filter can accommodate its requirements. A limitation during reacquisition of the lock is the cavity ring-up time, which of course is the same as the ring-down time. In other words, one expects that the system cannot lock instantly, and that the time period required for the cavity transmission to reach a stable intensity will be no shorter than the cavity-decay time. For a cavity with a high Q, we expect that the contribution of the servo to the time required to reach an equilibrium state is relatively minor, because the servo bandwidth is approximately 1 MHz, while the timescale to equilibrium for a typical highQ cavity for ring-down spectroscopy is on the order of hundreds of microseconds. As expected, we find that during the relocking process the error signal returns to zero well before the cavity transmission stabilizes. There are two caveats to this; the first being that switching the servo "on" does not mean that the laser will immediately be driven with high gain to the resonance center. In fact if the laser frequency is off resonance by more than a few cavity linewidths (but less than the modulation frequency), the laser could still be driven in the proper direction, but relatively slowly, as the error signal is near zero. Thus, one expects that acquisition will take longer than the microsecond timescale that corresponds to the servo bandwidth. The second caveat is that for a system with a frequency control "knob" other than the laser injection current (such as the piezoelectric length of an extended cavity), the laser frequency and the transmitted power through the cavity may not be perfectly correlated. For instance, in response to a transient that shifts the cavity resonance towards the blue, the servo may quickly return the laser frequency to the center of the cavity resonance by decreasing the injection current and hence also the laser power. With the laser frequency fixed once again on the cavity resonance, the circuit in Figure 1.10 will slowly bring the injection current back up to the original value by acting on the PZT, decreasing the laser cavity's mechanical length. Consequently, we expect the "ring-up" of power in the cavity to the previous stable value may take slightly longer than the ring-down, because
40
STABILIZING DIODE LASERS
the laser power was lowered to relock the system. This process occurs in the reverse sense also, causing a slightly faster return to equilibrium if the laser power is increased to relock. In order to switch the laser in and out of lock repetitively, we need some fast way to turn the servo off and back on. Simply switching the error signal off (with S 1) at the first amplifier's input serves to unlock the laser, because the servo loses access to the error (i.e. correction) signal. Switching the output switch $6 has the same effect. We keep the PZT integrator switch ($5 in Fig. 1.10) open during the unlocking and locking cycle so that the laser's PZT drive voltage remains appreciably unchanged. Furthermore, for the same reason the integrator input switch $4 remains open during "unlock" and closed during "lock." If the input switch $4 is not utilized, the output of the integrator may be dependent on the duty-cycle of the unlocking and locking. This is a function of the current through the integrator's input resistor during the "unlock" portion of the cycle. If it is nonzero, the integrator output will attempt to steer the laser away from the resonance, which will result in the system breaking lock after only a few cycles of the relocking sequence. To allow the cavity ring-down time constant to be measured, CRDS experiments have most often used acousto-optic modulators to switch the input beam away from the cavity mirror. Switching the laser to a subthreshold value has also been employed [25]. Rapid frequency shifting of the laser to a stable off-resonance frequency has been used to allow heterodyne ring-down measurements in reflection [26]. Observing the ring-down decay in transmission by frequency shifting is also possible for some cavity configurations. The shift must be accomplished in a manner such that the cavity is a high-attenuation filter. This requires attention to the cavity's transverse-mode structure in the vicinity of the fundamental mode, and the cavity's vibrational stability. Here we have used a half-symmetric resonator constructed from a plane mirror and a 30cm radius mirror, separated by about 25 cm. There is a 15 MHz wide-frequency region to the blue side of each fundamental mode of this cavity that appears free from transverse modes. A modal analysis indicates that in this region there are no modes with combined transverse-mode orders (n+m) of less than 100. Thus by shifting the laser frequency so that the laser sits in this "dark" frequency region during cavity ring-down, we should have minimal contamination of the ring-down signal attributable to leakage of the carrier into the cavity. This approach may require neglecting some data points at the beginning of the ring-down decay if the laser does not shift frequency quickly enough. To ensure the locking sidebands at 4-29 MHz do not leak into the cavity, they are turned off with another CMOS analog switch at the start of each ringdown event. The switch was arranged in the center of a "pi"-shaped 50
REPETITIVE LOCKING FOR CAVITY RING-DOWN SPECTROSCOPY
41
configuration, which provided 30 dB of "off" attenuation and about 1 dB of "on" attenuation [14]. We note that more attenuation can be accomplished by using three switches, one in series with each resistor of the "pi" attenuator. The laser-frequency shift is accomplished with another analog switch, which is not shown in Figure 1.10. The switch is positioned on the diode laser side of the 2 kS2 coupling resistor, and, when closed, connects the diode anode to ground through a variable resistor. The switch is closed by the same TTL signal that unlocks the laser. Severallhundred microseconds later, after the ring-down measurement is completed, the switch is opened and the laser frequency returns to the vicinity of the cavity fringe (i.e. within the P D H capture range). A single cycle of this unlock, relock sequence is illustrated by Figure 1.14. During this cycle the laser frequency is monitored by using an auxiliary reference cavity, positioning the resonance such that
6.0
-5
5.0 eO 0o
4
4.0 3
E
E r
i
e--
i,_ 3.0 I->., > 2.0 0 s._ (/')
r O
.m
~
i 2 9
t q e
O
#
or
1.0
Reference cavity line width -8.7 MHz, FWHM 0
0.0
I
-100
0
~
I
100
~
I
200
,
I
~
300
I
400
~
I
500
~
I
600
T
1
700
Time (~s)
FIO. 1.14. Data from a single ring-down and relocking cycle. The exponential decay from the high-finesse measurement cavity corresponds to the left scale, while the laser's optical frequency is indicated by using the transmission from an auxiliary reference cavity. Prior to the ring-down event (triggered at t = 0), the laser is locked to the high-finesse cavity and the reference cavity transmission is approximately 75%. The laser is shifted to the blue approximately 3-4MHz by shunting 0.3 mA of current away from the diode when it is unlocked (see text). The relock signal happens at 355 ps, and reacquisition and locking occurs in the next 100 gs.
42
STABILIZING DIODE LASERS
the laser frequency is on the side of a fringe. The sample rates of both the ring-down data and the reference cavity data are 0.4 gs per point. Within three data points (~1.2 gs), the laser frequency has shifted about 3-4 MHz away from the high-finesse resonance. We find that the repetitive locking works better with only one full integrator in the system, so we keep switch $3 closed for both locking and unlocking. A system undergoing this repetitive locking is shown in Figure 1.15. Although only six cycles are shown, the system is capable of indefinitely unlocking and relocking. Using software to set the TTL "unlock" and "lock" delays, we have cycled through 106 ring-down sequences. However, some caution must be exercised before adding a large number of decays together to measure the exponential decay time. The limitation on the measurement time and subsequent number of averages will be given by any nonrandom process that may change the cavity loss, for instance a subtle change in gas pressure or temperature. In such a situation, although the random noise will appear to decrease with averaging, the
300
c
250
r
r
c
,
r
200
,
2 its sample interval
cycle: 215 las unlock, 1000 las lock, 6 repetitions |
-
15o
9
!
9
i
i
9
~
0
..
100
|
!
i
9
i
I
0
:
:
i
,
i
,
I
1
2
3
i
i ~i
9
i
9
t
.=
ii
U jU U
i
50
9
!
I
4
,
I
I
I
I
5
6
7
8
~
I
9
,
I
10
I
I
11
,
I
12
Time (ms) Fla. 1.15. Repetitive locking of the laser is possible by opening and closing the input switches or the output switch. At t = 0, the first ring-down cycle is triggered, and the laser is simultaneously unlocked and frequency shifted to the blue by decreasing the current to the diode. The laser is shifted back near the cavity resonance and the input switch closed 215 gs later. The transmitted power through the cavity returns to an equilibrium value, and the process is triggered again at a 820 Hz repetition rate.
REPETITIVE LOCKING FOR CAVITY RING-DOWN SPECTROSCOPY
43
measurement repeatability will not improve. Shown in Figure 1.15 is an arbitrary 1000 gs delay between the end of each "unlock" period and the start of the next "unlock" period. Higher throughputs (faster averaging times) are possible by reducing this dead time. As may be observed in Figure 1.14, this system returns to the steady-state transmission in about 350 gs from the TTL "lock" signal. Even though the system will relock to the cavity center in a shorter amount of time, this does not necessarily mean that the cavity transmittance has stabilized at its previous value. A set of twenty-five sequential decays is averaged together and is shown in Figure 1.16. The exponential decay appears linear on a log scale, until the trace reaches the noise floor set by the detector noise level. Fitting an exponential curve to this data results in a 1/e time constant of 24.2 gs. The exponential fit was weighted for shot noise (square root of the signal level), and includes a constant term for the offsets that account for the analog-todigital converter and detector noise. 10.0
cO
1.0
E t'-
t~
o
0. 1
0.01 T
i
0
T
i
20
,
I
I
40
60
i
I
80
i
Time (gs)
I
100
i
f
120
t
I
140
i
I
160
f
I
t80
I
I
200
Fro. 1.16. The average of twenty-five sequential cavity ring-downs, shown on a log scale. The cavity was in a closed chamber, with the loss caused by a coincidence between the cavity mode and an absorption line in the water band near 830 nm. The data was taken with a 16 bit analog-to-digital converter with a sample period of 0.4gs. The trigger point is at t =0, and the subsequent exponential decay time constant is 24.2 gs. When an exponential fit is performed on each of the ring-down decays individually, resulting in twenty-five time constants, the standard deviation is 0.01 gs.
44
STABILIZING DIODE LASERS
1.7 Conclusions In this chapter we have described in some detail a fairly simple approach for locking an ECDL to a high-finesse Fabry-P6rot cavity. In our lab, we have used this approach to lock diode lasers for many different applications including precision metrology of atom-trapping experiments, high-resolution atomic spectroscopy, cavity ring-down spectroscopy, and length metrology. While the different lasers required different values for the electronic components, we have found the overall approach to be quite flexible and yield good results on a fairly short time scale. Our experience has shown that laser linewidths of well under 1 kHz can be readily achieved with this technique. We note that many other laboratories have used a similar approach to achieve excellent results as well. We have attempted to make this chapter "user-friendly" in the sense that it includes details not normally included in the standard research papers. We even added descriptions of a few "bells and whistles" in the form of switches that might make the locking system more reliable and easy to work with. Of particular interest and importance in the present context is the demonstration of rapid locking and unlocking of an extended-cavity diode laser to a high-finesse optical cavity. This technology can greatly enhance the signal-to-noise ratio and data rate in cavity ring-down spectroscopy.
References I. R. W. P. Drever, J. L. Hall, F. V. Kowalski, J. Hough, G. M. Ford, A. J. Munley, and H. Ward, Laser phase and frequency stabilization using an optical resonator, Appl. Phys. B 31, 97-105 (1983). 2. M. Zhu and J. L. Hall, Stabilization of optical-phase frequency of a laser system--application to a commercial dye-laser with an external stabilizer, J. Opt. Soc. Am. B 10, 802-816 (1993). 3. H. R. Telle, Narrow linewidth laser diodes with broad, continuous tuning range, Appl. Phys. B 49, 217-226 (1989). 4. H. R. Telle, Stabilization and modulation schemes of laser diodes for applied spectroscopy, Spectrochim. Acta Rev. 15, 301-327 (1993). 5. A. S. Arnold, J. S. Wilson, and M. G. Boshier, A simple extended-cavity diode laser, Rev. Sci. Instrum. 69, 1236-1239 (1998). 6. P. Zorabedian and W. R. Trutna, Jr., Alignment-stabilized gratingtuned external-cavity semiconductor laser, Opt. Lett. 15, 483-485 (1990).
REFERENCES
o
o
10.
11.
12. 13. 14. 15. 16.
17. 18.
19. 20. 21. 22.
45
A. Schoof, J. Grunert, S. Ritter, and A. Hemmerich, Reducing the linewidth of a diode laser below 30Hz by stabilization to a reference cavity with a finesse above l0 s, Opt. Lett. 26, 1562-1564 (2001). B. Dahmani, L. Hollberg, and R. Drullinger, Frequency stabilization of semiconductor lasers by resonant optical feedback, Opt. Lett. 12, S76-S78 (1987). H. Li and H. R. Telle, Efficient frequency noise reduction of GaA1As semiconductor lasers by optical feedback from an external high-finesse resonator, IEEE J. Quantum. Electron. 25, 257-263 (1989). H. Patrick and C. E. Wieman, Frequency stabilization of a diode laser using simultaneous optical feedback from a diffraction grating and a narrow-band Fabry-P6rot cavity, Rev. Sci. Instr. 62, 2593-2595 (1991). S. Ohshima and H. Schnatz, Optimization of injection current and feedback phase of an optically self-locked laser diode, J. Appl. Phys. 71, 3114-3117 (1992). A. E. Siegman, "Lasers." University Science Books, Mill Valley, California, 1986. C. Fabre, R. G. Devoe, and Brewer, R. G. Ultrahigh-finesse optical cavities, Opt. Lett. 11, 365-367 (1986). C. Hutchinson, Ed., "The ARRL Handbook for Radio Amateurs." The American Radio Relay League, Newington, Connecticut, 2002. R. C. Dorf, "Modern Control Systems." Addison-Wesley Publishing, Reading, Massachusetts, 1989. P. Tremblay and R. Ouellet, Frequency response of a Fabry-P~rot interferometer used as a frequency discriminator, IEEE Trans. Instrum. Meas. 40, 204-207 (1991). P. C. D. Hobbs, Ultrasensitive laser measurements without tears, Appl. Opt. 36, 903-920 (1997). P. Dub6, L.-S. Ma, J. Ye, P. Jungner, and J. L. Hall, Thermally induced self-locking of an optical cavity by overtone absorption in acetylene gas, J. Opt. Soc. Am. B 13, 2041-2053 (1996). D. S. Elliott, R. Roy, and S. J. Smith, Extracavity laser band-shape and bandwidth modification, Phys. Rev. A 26, 12 (1982). R. W. Fox, Trace Detection with Diode Lasers. Ph.D. Thesis, University of Colorado, Boulder, 1995. D. Romanini, A. A. Kachanov, N. Sadeghi, and F. Stoeckel, CW cavity ring-down spectroscopy, Chem. Phys. Lett. 264, 316-322 (1997). B. A. Paldus, C. C. Harb, T. G. Spence, B. Wilke, J. Xie, J. S. Harris, and R. N. Zare, Cavity-locked ring-down spectroscopy, J. Appl. Phys. 83, 3991-3997 (1998).
46
STABILIZING DIODE LASERS
23. C. R. Bucher, K. K. Lehmann, D. F. Plusquellic, and G. T. Fraser, Doppler-free nonlinear absorption in ethylene by use of continuouswave cavity ring-down spectroscopy, Appl. Opt. 39, 3154-3164 (2000). 24. T. G. Spence, C. C. Harb, B. A. Paldus, R. N. Zare, B. Willke, and R. L. Byer, A laser-locked cavity ring-down spectrometer employing an analog detection scheme, Rev. Sci. Instrum. 71,347-353 (2000). 25. A. A. Kosterev, A. L. Malinovsky, F. K. Tittel, C. Gmachl, F. Capasso, D. L. Sivco, J. N. Baillargeon, A. L. Hutchinson, and A. Y. Cho, Cavity ring-down spectroscopic detection of nitric oxide with a continuouswave quantum-cascade laser, Appl. Opt. 40, 5522-5529 (2001). 26. M. D. Levenson, B. A. Paldus, T. G. Spence, C. C. Harb, R. N. Zare, M. J. Lawrence, and R. L. Byer, Frequency-switched heterodyne cavity ring-down spectroscopy, Opt. Lett. 25, 920-922 (2000).
2. SPECTROSCOPIC APPLICATIONS USING RING-DOWN CAVITIES Giel Berden
FOM Institute for Plasma Physics, Nieuwegein, The Netherlands
Gerard Meijer
FOM Institute for Plasma Physics, Nieuwegein, and University of Nijmegen, Nijmegen, The Netherlands
Wim Ubachs
Free University, Amsterdam, The Netherlands
2.1 Introduction The present chapter deals with the applications of cavity ring-down spectroscopy and its variants. In the first section, a rather large number of studies involving the weak transitions in the oxygen molecule, including the effects of magnetic fields and collisions, are described. This molecule has served as a benchmark system to demonstrate the opportunities of ringdown techniques. In a subsequent section, we give an overview of the studies in which ring-down techniques have been applied to produce new spectroscopic data on molecules. In the last section, we have listed some pros and cons of ring-down techniques and the alternative techniques commonly used in molecular spectroscopy: laser-induced fluorescence, resonantly enhanced multi-photon ionization, coherent anti-Stokes Raman spectroscopy, degenerate four-wave-mixing and the non-laser Fourier transform spectroscopic technique.
2.2 02: A Benchmark System for Ring-down Techniques In many studies performed to demonstrate aspects of ring-down techniques the oxygen molecule is chosen as the target; this choice is connected to the properties of a somewhat peculiar set of low-lying electronic states in this molecule. Its lowest molecular orbital configuration ,2 ,2 ,2 (lSO'g)2 (ls%) (2SO'g)2 (2Set.) (2pCrg)2 (2pZru)4 (2pZrg) gives rise to the three wellknown lowest electronic states of the molecule: the electronic ground state X 3Eg and the excited states a 1Ag and b 1Eg.+ Transitions between the ground 47 EXPERIMENTAL METHODS IN THE PHYSICAL SCIENCES Vol. 40 ISBN 0-12-475987-4
Copyright 9 2002 by Academic Press All rights of reproduction in any form reserved. ISSN 1079-4042/01 $35.00
48
APPLICATIONS USING RING-DOWN
SPECTROSCOPY
state and these excited states are well studied: the b 1Ng-X + 3~2 transition, in the visible range is called the Atmospheric system, while the a 1Ag-X3Ng transition is called the Infrared atmospheric system. Also known is the Noxon (b 1Ng-a + 1Ag) system connecting the two electronically excited states. As shown in Figure 2.1 besides these states there are more bound states correlating with two ground state oxygen atoms O(3p)4-O(3p): the c 1~C~-, A t3Au, and A3~C+ state. The three systems connecting the electronic ground state to these states, observed in the ultraviolet, A 3I ]+, - X 3~g, At3 A,--X 3leg, and c 1]C~--X3~C2, are referred to as the Herzberg I, II, and III systems. The lowest excited state, a lag, is connected to these three excited states as well, but only the Chamberlain bands (A'3A,-a 1Ag) and the Richard-Johnson system (cl~C-2-alAg) were experimentally observed. As follows from an ab initio study of the low-lying states of 02, apart from the six observed
".l
'
4O
'
.
'
'
"
'
--
'
i,:i::,:_ 9 9 ..... ... :- .....
,.,:
,. ....
,-;, .
I
'
.,.
~:
.
'
I
,,~.,
A:~. ~ u
q
-"
~,
...........................................................
30
'
J
:. :\'-",i-::..:,.::. 9 . ,(, ...........
. . . . . . .
'
.............=9":"
t .
t
"7E
I
i.:,~ - .........
I.
35
~.','
1/
.................................................. ' ~
% ,,r.-,.
--.... >-
25
(.9 rr"
uJ Z IJJ
2o
J
-- 15 hz LIJ pO
g
10
a~A g X32; g-
0.10
0.14
INTERNUCLEAR
0.18
0.22
DISTANCE
/ nm
Fro. 2.1. Potential energy curves and vibrational energies of the important low-lying electronic states correlating with O(3p)+O(3p) dissociation limits. Figure kindly provided by Dr. R. Copeland of SRI International, Menlo Park, USA.
02: A BENCHMARK SYSTEM FOR RING-DOWN TECHNIQUES
49
bound states there are a number of additional states, 3FI,, l I-[g, 51-[u, 5~u, and 5rig all correlating with O(3p)+o(3p); these states are basically repulsive at the first few Angstroms of internuclear separation and are all very weakly bound at large internuclear separation [1]. The 5rig state, with the largest well-depth of the quintet states (0.16 eV), is included in Figure 2.1. Between the lowest six of these states fifteen radiative transitions are conceivable, but the peculiarity lies in the fact that they are all forbidden in the electric dipole approximation. Hence the electronic transitions are very weak and only eight of them have been observed. The excited state lifetime of the b 112g+ state is about 11 s; the transition, connecting to the ground state, is threefold forbidden in view of I~+-Y:-, triplet-singlet, and g-g transitions. The b-X system, allowed for magnetic dipole transitions, is a factor of 109 weaker than typical electric dipole allowed transitions. The higher lying v > 0 vibrational levels in the b 1YZg + state suffer from much lower Franck-Condon overlap with the xayZg, v - 0 ground state level, resulting in oscillator strengths offl0 - 1.6 x 10-11 and f 2 0 - 6.3 x 10-13. Nevertheless it was these two weak transitions, the Atmospheric B and y-bands, that was recorded in the first experiment demonstrating cavityring-down (CRD) spectroscopy [2]. Particularly because of the weakness of the bands, and in view of the easy availability of oxygen gas, and the atmospheric transitions being in the range of commercially available tunable lasers, these transitions have served as the prototypical test systems in many subsequent studies, demonstrating the power and sensitivity of ring-down techniques. In fact before the invention of the powerful CRD method, some physically related techniques using the enhancement properties of cavities, also focused on the forbidden optical transitions of the oxygen molecule to demonstrate their power and feasibility [3, 4].
2.2.1 The bl ~]g-FX3~g Atmospheric System Naus and Ubachs and co-workers have measured the CRD absorption spectra of the b 112g+-X312g (v, 0) bands (v = 0-3) of 1602, 160180, 160170, 1802, 170180, and 1702 isotopomers of oxygen [5-8]. In these works the focus was on spectroscopy and the generic pulsed CRD technique, with a fitting of the decay transients. As a result new or improved molecular constants for the b 1Y;+, v states were determined for all isotopomers, while for 170 2 the constants in the X ly;+, v - 0 ground state were determined. Note that the (3, 0) band near 590 nm, also referred to as the 3-band is 104 times weaker than the (0, 0) band; part of a recorded spectrum for the ~-band of 1602 at a pressure of 183 torr of natural oxygen is shown in
50
APPLICATIONS USING RING-DOWN SPECTROSCOPY I
I
19
20.0
I
17
I
15
I
I
17
I
13
I
15
!
11
I
13
I
1
I
9 9
7
I
7
....... I
I
5
I
5
PQ--
3
I
3
I
1
Pp
1602
E
% o
10.0 o E O x_ O
0.0 _ I
17165.0
l
~
,
i
I
17190.0
f
,
,
i
I
17215.0
17240.0
i
i
17265.0 cm q
FIG. 2.2. CRD spectral recording of the b-X(3, 0) band of 1602 at a pressure of 183 torr. The baseline was corrected for the loss rate related to the mirror reflectivity. For the P-branches an assignment is given. Reproduced with permission from Reference [8].
Figure 2.2. The same 3-band was also investigated by means of intracavity laser absorption spectroscopy (ICLAS) [9] and this allows for a comparison of this method with the pulsed laser CRD technique; the ICLAS spectrum yielded a similar signal-to-noise ratio (SNR), but was recorded at 3.5 times higher pressure, resulting also in somewhat collisionally broadened lines. The SNR in the generic CRD experiment depends of course strongly on the reflectivity of the mirrors used, and this was R > 99.99% in Reference [8]. The CRD studies of References [5-8] focused on an improvement of the spectroscopy of the b - X system of 02, in particular for the less abundant isotopomers. In view of the inherent Doppler broadening of the CRD technique, the accuracy in the old work of Babcock and Herzberg [10] could not be significantly improved. In fact also the sensitivity, previously obtained from long atmospheric path lengths, is comparable to that of the recent CRD studies; some lines of the weakest 3-band were already recorded back in 1948. In Reference [5], using pulsed CRD near 760 nm with mirrors of R ~ 99.998%, a quadrupole transition in the b--X(0, 0) band, with an intensity of 2 x 10 .9 cm -1 could be observed, demonstrating the high sensitivity. It is noted, however, that such transitions were observed, even with higher SNR, from atmospheric absorption measurements over 43 atm km absorption path lengths [11].
02:A BENCHMARKSYSTEMFORRING-DOWNTECHNIQUES
51
In the pulsed C R D experiments some of the isotopic species were observed in natural abundance, as previously in atmospheric absorption, but also isotopically enriched samples could be used. This demonstrates a notable advantage of the ring-down setup: long effective path lengths, and therewith high spectroscopic sensitivity can be obtained from small cavities for which only minute gas samples are required. Here the C R D technique is favored over the alternative application of FT spectrometers attached to White-multipass-cells, where much higher gas inputs are required. Particularly when applied to expensive isotopically enriched gas samples this advantage is of importance. With the use of a cell of 40 cm length, a diameter of 1 cm and a pressure of 180 torr, a spectrum of the very weak g-band of 1802 could be recorded [8] from a sample of ~10 atm cm 3 or 4 x 10 -4 M. Similarly, in an application of phase-shift C R D the v-band of 1802 was measured using a minute sample cell [12]. Also the use of the cavity-enhanced-absorption (CEA) variant of ringdown techniques (discussed below) in application to an absorption measurement of oxygen in a slit jet demonstrates a specific advantage over FT-techniques. The ~P](1) line of the b 1Eg-X+3Eg(0, 0) band of 1602 could be measured with a cavity of only 10 cm length [13]. The slit-jet configuration resulted in a linewidth of less than 0.01 cm -1 on this resonance shown in Figure 2.3, which is due to residual Doppler-broadening
Oxygen Argon
0.08
1 DiodeLaser]I
SlitN~ 1
0.06
.~ ~' 0.04
E" . .
. .
. .
. .
. .
. .
. .
I[~L' ~-~J~A4 b.~_~I
ToPumps[
~
Oscilloscope Sweep Generator
i
I
0.02 0.00 ,
I'
f
I
f
'o.; o]o o.s .o Frequency [GHz]
Fro. 2.3. Left: Scheme of the experimental setup [13]; Right: CEA spectrum of the ePl(1) transition of the A band of 1602 measured in a 10 cm long optical cavity positioned around a slit-nozzle expansion. The observed linewidth of 270 MHz is due to residual Doppler broadening.
52
APPLICATIONS USING RING-DOWN SPECTROSCOPY
in the planar jet-geometry. This example constitutes the first measurement of an optical transition in 02, which is not limited by the full Dopplerbroadening encountered when probing static gas samples. In this example only the lowest rotational energy level could be probed, due to the jet conditions, where a mixture with strongly cooling argon gas was used; this technique might be used in future to improve upon the electronic spectroscopy of the electronic transitions in 02. Xu et al. [14] performed quantitative CRD studies on three atmospheric bands b-X(v', 0) for v ' = 0-2, with a laser of 0.09 cm -1 bandwidth, just somewhat broader than the molecular absorption feature. In the analysis of the decay transients the signal was assumed to originate from two contributions, giving rise to a biexponential transient: one part was assumed to be exactly on resonance and absorbing proportional to the strength of the spectral line, while the remaining part was assumed to be off-resonant. Within this model-assumption, pressure dependent rotational line strengths were analyzed as a function of pressure for a number of lines in the three bands. Resulting values for the line strengths confirm the quoted values in the HITRAN database. Pressure broadening, involving self-broadening effects as well as collisions with N2, was investigated by CRD on the b--X(1,0) band at 688 nm [15]; there is a vast literature on this subject coming from FT studies.
2.2.2 The a 1Ag--X3~g Infrared Atmospheric System The a 1Ag-X3]Eg (0, 0) band at 1.27 gm was investigated independently by two groups using the generic pulsed CRD method [16, 17]. This nearinfrared atmospheric system is, with an upper state lifetime exceeding 1 h, even weaker than the b--X system. In the study of Newman et al. [16] CRD was compared with long path Fourier-transform spectroscopic investigations involving 60 atmm effective path lengths. As for the sensitivity, the CRD and FT methods are found to be equally suited as spectroscopic tools. In this region of the near-infrared, where the combined Doppler and collision broadened width (for atmospheric pressure) on oxygen lines is typically 0.1 cm -1, most commercially available pulsed laser systems have bandwidths beyond this value. As in the case of Newman et al., the bandwidth of 0.25 cm -1 gives rise to problems in deducing quantitative information on the line intensities. This bandwidth problem has been recognized and treated formally by various groups [18-20] and is in a sense similar to the slit-width problem in classical spectroscopic investigations. In CRD problems arise because the measured transients are no longer monoexponential if the bandwidth of the laser exceeds the widths of the
02: A B E N C H M A R K SYSTEM FOR R I N G - D O W N TECHNIQUES
53
molecular resonance. Note that Xu et al. [14] assessed this problem slightly differently with the assumption of a two-component decay, as discussed above. Newman et al. followed the estimates and the analysis of Zalicki and Zare [19], who showed that the deviation is reasonably small and quantifiable if the absorbance per cavity round-trip remains small. In such case the integrated intensity, i.e. across a spectral line, deviates only by an amount of less than 10%. However, to ensure the smallness of the single round-trip absorption only a 0.5 gs portion from the decay transient could be used. We note that in the experiment, the cavity decay time was 7 gs (mirror reflectivity R = 9 9 . 9 3 % and 1.5 m cavity length) and hence the majority of data points, usually giving rise to improved signal-to-noise, were omitted. As a final result the total integrated band intensity on the a 1Ag-X3Ng (0, 0) band could be brought into agreement with the Fouriertransform spectroscopic measurements to within 2%, yielding a value for the Einstein coefficient of A = 2.19(7) x 10 .4 s -1 [16]. In the study of Miller et al. [17] a more state-of-the-art laser system was used with a nearly Fourier-transform limited bandwidth of 0.017 cm -1 at 1.27 gm, therewith avoiding the aforementioned bandwidth problems in the quantitative analysis of the CRD-transients. However, as was demonstrated previously by Martin et al. [21], when the coherence time exceeds the roundtrip time in the cavity, disturbing oscillations on the ring-down transients will occur, that prohibit a straightforward interpretation of the decay transient in terms of an absorbance. Miller et al. overcame this problem by lowering the detector response to 600 ns, therewith washing out the oscillations. It is noted that such a procedure will decrease the obtainable sensitivity in the experiment, but here the focuswas on a determination of the band integrated intensity and the Einstein A-coefficient. Measurements were performed over a range of pressures, varying from 15 to 557 torr, resulting in a value of A = 2.3(3) x 10 -4 s -1, which is in good agreement with the average from a number of classical experiments, and confirming the study of Newman et al. [16].
2.2.3 The Herzberg Bands in the Ultraviolet The extension of ring-down techniques to the ultraviolet wavelength domain is relatively straightforward, and again the oxygen molecule provides a benchmark test system for C R D studies. Frequency doubling techniques, converting the output of commercially available tunable lasers into the ultraviolet (UV), usually reduce the laser output by an order of magnitude, but since only millijoules of laser intensity are required in
54
APPLICATIONS USING R I N G - D O W N SPECTROSCOPY
a generic C R D experiment, this has no adverse effect. A major point of concern is the reduced reflectivity of avaliable mirrors to a maximum of typically 99.8%; this severely limits the sensitivity of C R D in the UV. Huestis et al. [22] reinvestigated the spectra of the Herzberg bands close to the dissociation threshold by C R D and by an alternative technique of probing the oxygen atoms, produced by collisional dissociation involving an excited oxygen molecule. In an initial study, a laser with a bandwidth of 0.5 cm -1 was applied, and it was demonstrated that C R D had a superior sensitivity and new features were observed. Also the first direct information on oscillator strengths of the Herzberg II and III systems was generated in these experiments. In a subsequent study [23] focusing on the 242-244 nm region, where new features had been identified, a laser of improved bandwidth (0.2 cm -1) was used for the purpose of better resolving the features and for addressing the bandwidth issue in quantitative CRD. It should be noted that this narrower bandwidth is still a factor of three above t h e Doppler width on the molecular oxygen resonances. In this region of the ultraviolet again a comparison can be made between the C R D technique and the concomitant developments in Fourier transform spectroscopy. A UV Fourier transform spectrometer was developed [24, 25] that could be set to an effective resolution of 0.06 cm -1 yielding an accuracy of 0.005 cm -1 in the line positions. In terms of resolution, this F T setup is indeed superior to the C R D setup used in References [22, 23] although it should be noted that there do exist UV lasers that would match this resolution. In terms of sensitivity the C R D experiment, here limited b y the mirror reflectivity of 99.6%, made possible the observation of five new excited vibrational levels A3~ +, v - 12, A'3A,, v - 12, 13 and clI~2, v 17, 18; moreover of the A - X ( l l , 0) band 12 branches were observed via CRD, and only 8 with FT spectroscopy. Apart from the newly identified, features, the C R D experiment yielded also some 100 additional--yet unidentified--spectral lines just below the dissociation threshold in O2. As for the oscillator strengths, an attempt was made in the C R D studies to correct for the effect of the laser bandwidth, but a comparison with t h e higher resolution FT study [25] indicates that there is still an underestimate of 30% on the line strengths. The developments in C R D and F T techniques have progressed simultaneously. After the work of Slanger et al. [23], Jenouvrier et al. [26] performed FT studies in the UV, again focusing on the energy region below the first dissociation threshold in O2. With an improved sensitivity some lines reported in the C R D experiments were not observed in F T and vice versa. Nevertheless, based on the resolution of 0.12 cm -1 many hitherto unidentified lines could be given an assignment. F r o m a~ deperturbation analysis, some transitions could be assigned to a weakly
02: A BENCHMARK SYSTEM FOR RING-DOWN TECHNIQUES
55
bound state of 3rI, symmetry of which vibrational levels v = 0 and v = 1 were identified and for which a dissociation energy of De -- 140 cm -1 was determined.
2.2.4 The (02)2 Complex Since the end of the 19th century it was known that high-density oxygen gas features some additional absorption phenomena. Broad resonances, investigated in the condensed phase and in the high-pressure regime, are understood as caused by 0 2 - 0 2 collisional complexes. During a collision an electronic transition is induced in the (02)2 system. This is an effect of symmetry breaking by the collision partner, giving rise to a change of the selection rules. As a result both 02 partners may leave in an electronically excited state after absorption of a single photon. At the 630 nm resonance both molecules are excited to the a lag, v = 0 state. Naus and Ubachs [27] employed the sensitivity of pulsed CRD to investigate the (02)2 features at 630 and 580 nm in the pressure regime below 1 atm. Density-dependent cross sections were determined at discrete wavelength settings from an analysis of measured C R D extinction curves as a function of pressure as shown in Figure 2.4. The derived cavity-decay times have a quadratic dependence on gas pressure because of the collisional phenomenon involving two molecules; however the extinction is so weak that a linear contribution, related to Rayleigh scattering had to be included in the analysis; also a constant term related to the mirror reflectivity was included in the fits to the data. The C R D method appears to be ideally suited for such measurements: (i) for the broad resonances there is no bandwidth problem and the extinction can be straightforwardly interpreted in terms of cross sections; (ii) the contributions of Rayleigh scattering and collision-induced absorption can be unraveled; (iii) by choosing discrete frequency positions of the laser the features of the (O2)2-complex at around 630 nm can be disentangled from the linearly absorbing ?,-band in the O2-monomer, also at 630 nm. The latter entails a marked advantage over the FT technique, which was also applied to investigate (02)2 [28]. While the (02)2 collisional complex is associated with the repulsive part of the O2-O2 intermolecular potential, van der Waals bound states do exist below the dissociation threshold within this potential. Sharp and wellresolved absorption spectra of the O2-dimer have been recorded in a supersonic slit-jet expansion using the continuous wave (CW)-CRD technique [29]. Here the slit-jet configuration provides sub-Doppler resolution on the resonances of below 0.01 cm -1 as shown in Figure 2.5.
56
APPLICATIONS USING RING-DOWN SPECTROSCOPY (a) '
'
"
'
' 600
(b) 5{ X' ',it '.,~.t~,,2~2,11 5~-R '.a,.l,,_3~,I;,a,i~t-~ II
~
51-L.,~.',.A~ ' ' . . ~
7.5 ~,E~ 400
.
,' . , 205s 400
, % . ' i g
600
J
/1' ,,
II~ f
J
b
g
200
2.5
~ 0.
<
0
Ob
g 5.0
b
- 51-,
/ -tIII/
l/I /
.o 15000
I
,
I
,
I
16000 17000 18000 Frequency (cm 1)
0 0
200
400 600 Pressure (Torr)
FIG. 2.4. Observation of two prominent (02)2 collisional induced features with CRD spectroscopy; (a) density-dependent absorption cross sections as a function of wavelength (dots)and a fit to skewed Voigt profiles (full line); (b) Extinction measured as a function of pressure at three frequency positions (I, II and III) also shown in spectrum (a); while at position III extinction is nearly linear (only Rayleigh scattering) at position I the quadratic behavior is evident; in the inset the residuals from a least-squares fit. Reproduced with permission from Reference [27]. From a rotational analysis of these spectra, the geometrical structure could be determined in the ground and excited states. For the [X3~:g, v - 012 ground state configuration a geometry in the form of an H, with an intermonomer separation of 6.04-6.14 a0 was determined as the most stable one. In this study [29] also a comparison was made between the intracavity laser absorption technique and CW-CRDS. Although sensitivities were similar the grating spectrograph in ICLAS limits the resolution, rather than the laser as in CW-CRDS; hence the resolution in CW-CRDS was better by more than a factor of two.
2.2.5
02
in Magnetic Fields
Even though the light in most of the C R D experiments is polarized, there are not many spectroscopic studies which make use of this property. Berden et al. [30] have recorded the rotationally resolved C R D spectra of the A band of molecular oxygen in magnetic fields up to 20 T. Spectra were measured with linearly and circularly polarized light, leading to different
57
02: A B E N C H M A R K SYSTEM F O R R I N G - D O W N T E C H N I Q U E S
10-
RQo !j~)
14 5 - 1 ....
16 !
i12 I
10 [
14 I
8 I
6 I
4 2 I I
12 j R
!0 J~
8 !
d: Ji
J
4 2 I
i
Qo(J)*
O v---t
O 9
|
15804
|
15806
15808
|
15810
'
i
i
15812
Wavenumber (cm -1 ) Fla. 2.5. Part of the CW-CRDS spectrum of the (02)2 van der Waals complex. The assigned lines pertain to two different sub-bands of the [alAg, v--012 +-[X3l~g, v =012 transition. In the lower panel a calculated/fitted spectrum is shown. Reproduced with permission from Reference [29].
A M selection rules, thereby aiding in the assignment of the spectra. The Bitter-type magnet has an 80 cm long bore with a diameter of 3.1 cm. Because the magnetic field was homogeneous only over a few centimeters at the center of the bore, a very short cavity has been used in order to prevent the spectra from becoming too complex. Therefore, the mirrors were mounted on a 3 cm long tube, in a fixed position relative to each other. This short cavity was placed at the center of the magnet. The axis of the cavity was parallel to the magnetic field lines, i.e. the so-called F a r a d a y configuration. Figure 2.6 shows the spectra recorded with right-hand circularly polarized light (a+) as a function of the magnetic field strength. Only the central part of the A band is displayed. Due to the specific polarization of the light, only transitions with A M = + 1 are allowed. Although not shown here, Berden et al. [30] also measured the C R D spectra using left-hand polarized light and linearly polarized light, containing transitions with A M = - 1 and A M = + 1, respectively.
58
APPLICATIONS USING RING-DOWN SPECTROSCOPY i i
i
20
f
i
18
i
16 i
14 Itl
n
3
12 :~
o
,.,.
o 10~
> o
o.. i
83
i
t t t - - -
!
~-
. . . . . . . . . . . . . .
t
131 O0
13120
13140
frequency (cm 1 )
Fla. 2.6. Right-hand circularly polarized light has been used to measure the CRD spectrum o f the c e n t r a l part o f the b 1]Eg+(v -- 0) +-- j(3 ]E~-(v -- 0) t r a n s i t i o n o f molecular oxygen as function of the magnetic field. Two transitions with the same upper state are indicated by the dashed lines. Figure reproduced from Reference [30] with permission.
These spectra have been used to validate a theoretical model that is used to describe the interaction between the oxygen molecule and the magnetic field, by comparing the calculated and measured line positions and intensities. This model has further been used to calculate the spin contribution to the magnetic susceptibility of molecular oxygen. It also shows that the oxygen molecules are aligned in high magnetic fields, and that this alignment is caused by the spin-spin interaction via a coupling of the spin angular momentum to the magnetic field direction [30]. By placing a polarization selective optical element in front of the detector (e.g. a polarizer), it is possible to measure the rotation of the plane of polarization of the incoming linearly polarized beam upon passage through the ring-down cavity. If (part of) the cavity is placed inside a magnetic field, the plane of polarization can rotate owing to dispersion (magnetic birefringence) or polarization dependent absorption (magnetic dichroism). This has been exploited by Engeln et al. [31], who have placed the cavity
02: A BENCHMARK SYSTEM FOR RING-DOWN TECHNIQUES 4.0
M=+I
N=I, J=l
M=I
59
N=I, J=2 M=+I M=I
2.0 Q
"~ c" p,
0.0 +
2.0
4.0
i
13117.0
,
i
13119.0
,
!
13121.0
frequency ( c m l )
FIG. 2.7. PD-CRD dispersion spectrum in the region of the PP(1, 1) transition in the A band of molecular oxygen at 200 mbar. Measurements are performed in a 3 cm cavity which is placed in a magnetic field of 1.5 T. The magnetic field, which is parallel to the cavity axis, splits the transition into two components. The two small dispersion curves belong to the PO(1,2) transition, which becomes allowed through magnetic field induced mixing with the ~P(1, 1) transition. The dispersion curves are labeled with the quantum numbers of the X3E 2 state. Figure reproduced from Reference [30] with permission.
between two polarizers. The light exiting the cavity was split into two mutually orthogonal polarized components, and the time dependence of these components was measured simultaneously. In this so-called polarization dependent cavity ring down (PD-CRD) scheme, the information is obtained from the difference between the cavity losses in the two directions. It has been theoretically shown and experimentally demonstrated [31] that this difference is proportional to the magnetic dichroism in the case that the magnetic field lines are perpendicular to the cavity axis (Voigt configuration), and it is proportional to the magnetic birefringence in the case that the magnetic field lines are parallel to the cavity axis (Faraday configuration). An example is shown in Figure 2.7, where part of the P D - C R D dispersion spectrum (magnetic birefringence) of molecular oxygen is shown [30]. This spectrum is recorded at 1.5 T by using the aforementioned 3 cm long cavity in a Bitter magnet. It is evident that P D - C R D is only suitable to record spectra of paramagnetic molecules. Absorptions of other species are detected in both polarization directions and cancel out in the difference spectrum. In principle, P D - C R D spectroscopy is the C R D variant of magnetic rotation spectroscopy (MRS). Just like MRS, P D - C R D can be used to simplify spectra because both techniques are more sensitive for transitions with low rotational quantum
60
APPLICATIONS USING RING-DOWN SPECTROSCOPY
numbers, and to distinguish spectral features of paramagnetic species from those of nonparamagnetic species.
2.2.6 Spectroscopy of Oxygen with CW Lasers In CW-CRD spectroscopy, the cavity is excited with a CW laser. Just as in pulsed CRD spectroscopy, the ring-down time, and thus the absorption coefficient of the molecules present in the cavity, is determined by measuring the decay of the light exiting the cavity (see e.g. Ref. [32]). However, CRD spectroscopy with CW lasers can also be performed without temporal analysis of the ring-down transient. In phase shift cavity ring down spectroscopy (PS-CRD), the absorption coefficient is extracted from a measurement of the wavelength dependent phase shift that an intensity modulated CW light beam experiences upon passing through a high-finesse optical cavity. The phase shift method has been used to record the absorption spectra of the g-band of 1802 [12]. In CW-CRD spectroscopy an optical switch is required to prevent the in-coupling of light into the cavity while the decay transient is recorded. In PS-CRD spectroscopy an intensity modulator is required. In CEA spectroscopy no such device is needed. In the latter spectroscopic method the absorption coefficient is determined from a measurement of the timeintegrated intensity of the light leaking out of the cavity, while light is coupled into the cavity via accidental coincidences of the laser frequency with the frequency of one of the multitude of modes of the cavity [13]. It has been shown that this time-integrated intensity is proportional to the cavity ring-down time r and thus inversely proportional to the total cavity losses [13, 33]. Note that the name CEA spectroscopy is also frequently used to label all techniques using a cavity to enhance the absorption (like the title of this book), while in this chapter CEA is only used to describe the aforementioned technique (see also Reference [34]). Figure 2.8 shows the bandheads of the RR and RQ branches of the v-band of 1602 recorded with the CEA technique. The spectrum consists of three partly overlapping measurements, each covering about 1.5 cm -1, and have been recorded with a CW single mode ring dye laser in about 1 min. The absorption coefficient is expressed in units of ( 1 - R)/d, with R the mirror reflectivity and d the distance between the mirrors. Therefore CEA spectroscopy is not a self-calibrating technique in contrast to the generic form of CRD spectroscopy; the mirror reflectivity has to be known in order to make the absorption axis absolute (the mirror reflectivity can be determined in a CRD experiment, or in a CEA experiment with a known molecular absorption; see Reference [13]),
0 2"
A BENCHMARK SYSTEM FOR RING-DOWN TECHNIQUES
f
19
13
1715
RR(N) I 19
,0
61
13
II 17 15 RQ(N)
!
30
20
10
0
t I
15925.0
~
I
15926.0
i
I
15927.0
15928.0
frequency (cml)
Fro. 2.8. CEA spectrum of the oxygen ?,-band recorded in a cell of 12 cm length with 200 mbar pure oxygen, displaying the bandheads in the RR and RQ branches near 628 nm. Figure reproduced from Reference [13] with permission.
The CEA method can also be combined with MRS which results in cavity enhanced magnetic rotation (CEMR) spectroscopy [13]. As an example, we will discuss a C E M R spectrum of the PPI(1) transition of the v-band of 1602. The cavity is placed inside a homogeneous magnetic field of 0.88 T in such a way that the magnetic field lines are perpendicular to the cavity axis (Voigt configuration). The polarization of the CW light that enters the cavity makes an angle of 45 ~ with respect to the magnetic field. Therefore, transitions with A M = 0, + 1 are allowed. A polarizer between the exit of the cavity and the detector transmits the light with a polarization direction of ~bD relative to the polarization direction of the incoming light.
62
APPLICATIONS USING RING-DOWN SPECTROSCOPY
0.1 95 ~
0.0 -0.1
92 ~
"o 91 ~
rr (tl
o .m
==
O
88 ~
I
,
15899.0
I
15899.5
,
I
,,
15900.0
frequency (cm 1 )
FIG. 2.9. CEMR spectra of the PPI(1) transition of the y-band of 02 recorded in a magnetic field of 0.88 T (Voigt configuration), as function of the angle ~bD (with crossed polarizers 4~D=90~ Figure reproduced from Reference [13] with permission. In Figure 2.9 the inverse of the time-integrated light intensity passing through the analyzer is shown as function of frequency in the spectral region of the ePl(1) transition for seven different values of the angle q~I~[13]. For a value of qSD sufficiently far away from 90 ~ the inverse of the time-integrated signal is proportional to the polarization dependent absorption (magnetic dichroism). The spectrum measured with crossed-polarizer geometry, exclusively shows the effect of optical rotation due to dispersion [13]. Note that the spectrum with qSD= 90 ~ measured with the Voigt configuration can directly be compared with the dispersion spectrum shown in Figure 2.7 which has been measured for the same transition (but then for the A band) with the Faraday configuration.
SPECTROSCOPIC APPLICATIONS OF CRD
63
2.2.7 Toward Ultimate Sensitivity and Accuracy In the above section the application of the generic pulsed CRD technique and some variants using CW-laser sources were described in the measurement of weak absorption in 02. Along with detection of the multiply forbidden quadrupole bands at 760 nm in a pulsed CRD experiment, with mirrors of R = 99.998%, a test of the detection limit on the even rotational lines in 02 was performed [5]. These lines are strictly forbidden by the symmetrization postulate of quantum mechanics and they serve as a test for this fundamental principle. In the experiment of Reference [5] a sensitivity of 1.5 x 10.6 was obtained. It was already mentioned that in the pre-CRD era some important developments in intracavity dye-laser spectroscopy were pursued in which nearly comparable absorption sensitivity was achieved on the forbidden 02 bands in the visible [3, 4]. Following up on these CRD-precursor techniques sophisticated cavity-enhanced techniques have been developed, using locking of the cavity to a CW laser, phase modulation and heterodyne signal detection, in which the absorption sensitivity was increased to 7 x 10-11 cm -1 Hz-1/2; using this setup a test on the even lines in 02 was performed at the 5 x 10.8 level [35]. Further progress in this area of highly sensitive cavity-enhanced techniques is discussed elsewhere in this book. Van Zee et al. [36] have assessed the issue of quantitative precision in a CRD experiment and have argued that a single-mode approach yields the highest precision. For most spectroscopic purposes, where the focus is on easy tunability of the spectrometer, the CRD cavity is set such that it can be excited by a large number of modes giving the cavity a white spectrum; the whiteness of a high-finesse cavity, set far from confocal, was experimentally demonstrated [37]. The more cumbersome single-mode approach, where the cavity length has to be kept at the same mode of the resonator and in lock with the laser mode, is however shown to produce a quantitative reproducibility with a 3 x 10-4 standard deviation in the ring-down time in a measurement of the P Q(9) line in the oxygen A band. The authors have discussed on the possibility of using such a CRD setup as a pressure standard.
2.3 Spectroscopic Applications of CRD The CRD technique is especially powerful in gas phase spectroscopy for measurements of either strong absorptions of species present in trace amounts or weak absorptions of abundant species. For spectroscopic applications of the CRD technique this roughly means that species with relatively weak transitions are studied in static gas cells (for example
64
APPLICATIONS USING RING-DOWN SPECTROSCOPY
molecular oxygen), because one can simply increase the number density of the species. On the other hand, species with strong transitions are studied in supersonic jets, discharges, or flames, since the number densities in these environments are usually relatively low. In the remaining part of this chapter, an overview will be given of pure spectroscopic applications of CRD spectroscopy in static gas cells, flames and discharges, and supersonic jet expansions.
2.3.1 In Static Gas Cells
In 1988 O'Keefe and Deacon [2] reported the first CRD spectra. They measured the spectra of the B and y-band of molecular oxygen in a static gas cell. As exemplified in the previous section, molecular oxygen has always been one of the favorite molecules for CRD spectroscopy, but many other molecules have been investigated as well. Besides the determination of spectroscopic constants from the frequency position of the observed transitions, an important feature of static cell experiments is the determination of absolute line and band intensities (which is made possible due to the accurately known number density). In a series of publications, Romanini and Lehmann reported the absorption spectra of HCN and its isotopomers in the 435-571 nm spectral region [38-41]. This spectral region contains many very weak overtone bands. Figure 2.10 shows the CRD spectrum of the 106+-000 transition of HCN (1 quantum in the CN stretch, 0 in the bending mode, and 6 quanta in the CH stretch). By recording this spectrum at different pressures, it was shown that there is a large collisional line-mixing effect in the proximity of the R branch heads. Campargue et al. [42] recorded the 5vl band of propyne by using the pulsed CRD technique. The same method was used by Kleine et al. [43] to measure the absolute intensity of the fifth CH stretching overtone of benzene and by Naus et al. [44], who reinvestigated the weak overtone spectrum of water vapor in the wavelength range 555-604 nm. Brown et al. [45] determined the absorption cross sections for the third and fourth O - H overtone transitions on nitric acid and hydrogen peroxide, while the O-H stretching vibrational overtones of 2-butanol were studied by Xu et al. [46]. With a tunable CW diode laser He et al. [47] recorded the spectra of the Vl-+-3v3 band of N20 and CHC13 around 1.3 gm.
2.3.2 In Hostile Environments
Triggered by the success of the detection of radicals in laser ablation sources [48], laser photolysis reactors [49] and flames [37], CRD
SPECTROSCOPIC APPLICATIONS OF CRD
65
2.5-
2.0-
r" ('0 r f... o 03 c'~
t.5-
:t.0-
0.5-
0.0-
2o~5o
2o~7o
2o~9o
2o~io
2o~ao
2o~o
2o~;,o
Frequency i n cm-~FIG. 2.10. The CRD absorption spectrum of the 106+--000 transition of HCN. The sample pressure was 100 torr. Figure reproduced from Romanini and Lehmann [38] with permission.
spectroscopy has become a powerful tool to perform spectroscopic measurements in hostile environments. Zalicki et al. [50] first investigated CRD absorption of methyl radicals (CH3) in a flow reactor, where CH4/H2 mixtures flow over a hot tungsten filament at 2300 K. The broad absorption feature at 216 nm was used for a quantitative analysis of CH3 densities. Similarly the silyl radical (Sill3) was probed, via its broad resonance at 215 nm, in a remote Ar-H2-SiH4 plasma during hydrogenated amorphous silicon thin film growth [51]. Booth et al. [52] probed various fluorcontaining radicals (CF, CF2,A1F and SiF2) in radio-frequency plasmas in fluoro-carbon gases; these molecules have strong absorption features in the 215-230 nm region as well. The application of ring-down techniques for the spectroscopy of species in harsh environments requires some special precautions to protect the highly reflecting mirrors. Deposition of contaminants, that cause a reduction of their reflectivity, can be circumvented by the use of diaphragms inside the vacuum chamber and helium curtains near the mirrors [45, 53, 54]. The study of predissociation in excited states of molecules using CRD has become a topic in itself. Spaanjaars et al. [55] have determined the predissociation rates of OH (A2Y;+, v ' = 3) for N' up to 17 by simultaneously
66
APPLICATIONS USING R I N G - D O W N SPECTROSCOPY
measuring laser-induced fluorescence (LIF) and CRD absorption in the same laminar CH4/air flame. The relative intensities in the LIF and CRD spectra are different, which is due to the rotational dependence on the predissociation rate; with varying rotational quantum number, the competition between predissociative and radiative decays has a strong effect on the intensity of the fluorescence signal. CRD measures direct absorption and the signal intensity is independent on the relaxation of the excited state. Hence the combination of the two techniques, CRD and LIF, provides a sensitive tool for predissociation phenomena. Extensive predissociation studies of a large number of molecules have been reported by Orr-Ewing and co-workers (reviewed by Wheeler et al. [56]). A nice example is the molecule HNO. In Figure 2.11, a LIF spectrum and a CRD spectrum measured in the same spectral region are shown. Much of the structure present in the CRD spectrum is absent in the LIF spectrum because of predissociation of the A' state [57]. Besides missing lines in the LIF spectrum, predissociation can also be observed via an increased linewidth of the rotational transitions (because of the short lifetime of the predissociating upper state). A detailed study of predissociation offers a valuable approach to explore state-selected unimolecular chemical reaction dynamics on coupled potential energy surfaces [56]. The group of Orr-Ewing
I
I 7
I
15
I
1
16140
I
5
I
I
i
15
I
.
I
I
I
I
P,(J'~
10
N fl n n I Ii1/f1111
i
i
i
I
16150
w
I
I
10
I
i
5
I I I III
5
.
I
I
16160
.
.
5
Q.(J") R.(J']I
I
I
I
"R(J") , ~ 5I I I
Q(J") .
I
16170
II loI'~ 0204 000
}
10
100-000
I I NllllInNNll
I
I
I
I
I
I
I
I
16180
I
I
I
l
16190
W a v e n u m b e r / cm"
FIG. 2.11. LIF (top)~ and CRD (bottom) spectra of the HNO molecule with transitions form the X' ~4' (000) ground state to vibrational levels (100) and (020) of the A ~4'~ excited state. Because of predissociation part of the structure is absent in the LIF spectrum. Reproduced with permission from Reference [57].
SPECTROSCOPIC APPLICATIONS OF CRD
67
has studied many other molecules, which were generated in a flow reactor, like SH and SO [58, 59], $2 [60], BrO [61], IO [62], C10 [63], HC1CO [64], and FCO [65]. Kotterer and Maier [54] have detected the 2FI-X2rI electronic transition of the carbon chain C6H by pulsed CRD spectroscopy at 526 nm. This radical, which is of astrophysical interest, was generated in a hollow cathode discharge of acetylene in helium. They used the same experimental setup to study weak vibrational bands of the electronic transition of the molecular ion N~- [66]. Additional bands were measured in a similar setup by Aldener et al. [67]. CRD spectroscopy has also been used to determine the dipole transition moment of photochemically produced sodium hydride [68]. In this experiment, Nail was formed by laser excitation of sodium, generated in a heat-pipe oven, in a hydrogen atmosphere. Strong saturation effects (power broadening and dips at the center frequency) were observed [69]. Similar saturation effects have been observed for Li2 which was also produced in a heat-pipe oven [70].
A2FI,+-X212+g
2.3.3 In Jet Expansions Supersonic molecular beams and jets are very important in molecular spectroscopy. Expanding molecules seeded in a carrier gas (mostly a noble gas) into vacuum produces a cooling of the vibrational and rotational degrees of freedom. The advantage for spectroscopy is two-fold. On the one hand, only the lowest rotational and vibrational levels in the electronic ground state are populated, leading to less congested excitation spectra. On the other hand, the low internal temperatures permit the stabilization of structural variants (tautomers or conformers) and the stabilization of molecular clusters (van der Waals and hydrogen bonded clusters). Already in 1990, Saykally and co-workers reported their first results on the electronic spectroscopy of jet-cooled metal clusters which were produced by laser vaporization of a metal rod [48]. Since then, many metal clusters were investigated such as PtSi [71], AuSi [72], AgSi [73], CuSi [74], A12 [75], WO [76], and Cr2 [77]. This metal cluster research has been reviewed by Scherer et al. [78, 79]. While this work was performed in the visible region of the spectrum, their more recent work is focused on the infrared region, where vibrational transitions can be probed. Pulsed tunable infrared light (2-7 lam) is efficiently generated by Raman shifting of the pulsed dye laser output. The extension of the CRD technique towards the near and mid-infrared has been used to measure the O-H stretching vibrations in clusters of water [80]
68
APPLICATIONS USING R I N G - D O W N SPECTROSCOPY
(H20)n n=
>66
5 4
3
2
I
Free O-H bands
400-
E
{3. (/)
300-
if) 13. I-
(1) c 0
"= o_
200-
t
0 o9 .o
~J
-100-
0-
I
3000
I
3200
i
I
3400
3600
frequency
(crY)
FIG. 2.12. CRD spectra of the O-H stretch in water clusters obtained for various source backing pressures; n indicates the number of H20 molecules in a cluster. Figure reproduced from Reference [80] with permission.
(see Fig. 2.12), methanol [81], ethanol, and butanol [82]. Other studies are, for example, the monomer bending vibration in water clusters [83], and the carbonyl stretch in arginine [84]. Aromatic molecules have been studied with CRD spectroscopy as well. Although a R E M P I scheme was much more sensitive to record the vibrationally resolved spectrum of the $1 +--So of diphenylamine (DPA), Boogaarts and Meijer additionally used C R D spectroscopy to determine the absolute number density of DPA molecules in their laser desorption jetcooling molecular-beam spectrometer [85]. Ruth et al. [86] recorded C R D
SPECTROSCOPIC
APPLICATIONS
69
OF CRD
spectrum of the TI+--So transition of jet-cooled 4-H-l-benzopyrane-4thione, which was compared with the R E M P I spectrum and the phosphorescence excitation spectrum. They also used the CRD technique to study the internal conversion rate as a function of excess energy for the azulene molecule [87]. In the 545-680 nm spectral region, Romanini et al. [88] recorded the absorption spectrum of naphthalene cations, which were produced in a slit jet coupled to an electronic discharge. The spectroscopy of azulene molecules and naphthalene cations is also interesting from an astrophysical point of view; the absorption spectra can directly be compared with the diffuse interstellar bands (DIB), the ubiquitous absorption features in the spectra of many stars, whose origin is still a mystery. A couple of hundred DIBs have been observed, primarily in the 400-900 nm region. Besides aromatic molecules, carbon chains are promising candidates as well. Linnartz and Maier and co-workers have investigated species which were generated by a discharge in a supersonic expansion of gas mixtures through a slit nozzle. With the CRD technique
i
I
i
i
i
I
i
i
t
i
i
I
i
i
i
i
IIIIIIIIIIIIIIIII 3/2
f2 = 3/2 71/2
5/2
Ilia,,,,I,,,,Jiilill,Jlliil,Jilirp ~ = 1/2
II
1/2
,,,,2
18980
_
_
18982
;
!
18984
;
!
,
!
;
!
18986 18988 18990 Frequency (cm 1 )
illllillllllllllll
e
_
18992
e
18994
l
18996
FIG. 2.13. The two spin-orbit components in the 2I-[+-X2I-[ transition of C6H measured with CRD spectroscopy in a supersonic slit-jet plasma. Reproduced with permission from Reference [89].
70
APPLICATIONS USING RING-DOWN SPECTROSCOPY
they measured the (rotationally resolved) electronic spectra of C6H and C6D [89] (see also Fig. 2.13), CsH and C10H [90], C6H~- and C 2 [53], CsH + [91], C4 [92], C5 [93], HC6N and HC7H [94] and various carbon containing chain molecules [95]. A similar setup has been used by Ball et al. [96, 97] to record electronic spectra of HCnH (n = 7, 9, 11, 13). Other CRD spectroscopic studies report on the T1 state of thiophosgene (ClzCS) [98], the S1 state of acrolein (CHzCHCHO) [99], the C IA' state of $20 [100], the first overtone of the free hydrogen stretch of several (HC1)2 isotopomers [101], and the infrared spectrum of the formic-acid dimer [102]. By using CW-CRD spectroscopy, Romanini et al. [103] investigated saturation effects in jet cooled NO2, and Hippler and Quack [104] studied the v2 + 2v3 combination band of methane around 7510 cm -1.
2.3.4 Quantitative Study of Rayleigh Scattering Rayleigh scattering is usually considered a nuisance in CRD experiments. It gives rise to a nonresonant background and is in a sense similar to the contribution of mirror reflectivity to the decay transient. Ultimately it is a factor limiting the sensitivity of CRD experiments. The Rayleigh-scattering cross section is closely connected to the index of refraction, scales with v4 (v being the frequency) and at the few percent level a correction factor is invoked for Rayleigh scattering off nonspherical particles. While the value of the Rayleigh cross section can be derived from these parameters, well known for most molecules, a direct measurement of the Rayleigh extinction was not performed until the sensitive technique of CRD was applied [105]. For the gases Ar, N2, and SF6, that undergo no absorption in the visible wavelength range, the extinction was measured in a CRD-cavity as a function of pressure; results are shown in Figure 2.14. The slope of the cavity-decay rate is proportional to the product caiN, where c is the speed of light, N the number density and a~ the frequencydependent Rayleigh cross section; the intercept from a fit to the decay rate, shown in Figure 2.14(a), again represents the mirror reflectivity, which can be subtracted in the analysis. For various frequencies the pressure dependent extinction yields values for a~, which can subsequently be invoked in a general expression for the frequency-dependent Rayleigh cross section, scaling as ~v4+~. Here e represents the frequency dependence of the index of refraction, giving rise to a deviation from a proper v4-dependence. In Figure 2.14(b) the resulting values for &, obtained from a fit to the derived values for a~ are compared with calculations from known indices of refraction. This CRD approach to Rayleigh scattering confirms these
71
SPECTROSCOPIC APPLICATIONS OF CRD 650/
"-"
'll
625
i
'
i
'
J
4.5
SF6 Ar
"'E
575
t:~ 3
(~
rr
,o I-10
,
0
o
'
o
I
200
~
,
~
9Nitrogen. Argon. Fit to data. Calculations.
%) 600 o
' I /
o
'
o
i
oo
'
400 Pressure (Torr)
~, o
600
*~ -1 o
"
J
2.5
, /
/
/ " 9 1 4 9 /~
f
:
.
./:/ 15000
,
I
16000
,
I
17000
,
I
18000
Frequency (cm "1)
FIG'. 2.14. Measurement of the Rayleigh scattering cross section; (a) CRD-loss rate
as a function of pressure for SF6, N2 and Ar at )~ = 579.83 nm; in the bottom part are shown the residuals of an unweighted linear regression for the Ar data with a resulting rms error of 3 x 10.9 cm-]; (b) Resulting cross sections as determined from CRD measurements at various wavelengths, and a comparison with cross sections calculated from known refractive indices. Reproduced with permission from Reference [105].
estimates, including depolarization correction factors. Moreover the experiment produces a similar level of accuracy (1%) as the calculated Rayleigh cross sections for the two gases N2 and Ar. For SF6 no accurate data on the refractive indices are available in this wavelength region; here such a value can be determined from CRD.
2.3.5 Spectroscopic Applications of Other Cavity Enhanced Methods In the previous sections an overview of applications of pulsed and CWC R D spectroscopy has been given. Also the other techniques which were illustrated before by an application involving oxygen, have been used in spectroscopic studies. For example, the PS-CRD technique has been used to record the spectra of the A v = 6 C - H stretching overtones of C2H4, C2H6, C3H8, n-C4H]0, and n-CsH12 [106]. The CEA technique has been used by Berden et al. [107] to measure the absorption spectra of rotationally cold ]4NH3 in the 1.5 gm spectral region. In the slit-nozzle expansion, the
72
APPLICATIONS USING R I N G - D O W N SPECTROSCOPY
linewidth of the recorded rotational lines is determined by the residual Doppler-broadening and amounts to 0.006 cm -1. As a result of this narrow linewidth and due to the low rotational temperature, the recorded transitions could easily be assigned. In this way, they were able to characterize the vl+v3 band at 6609 cm -1 and the Ill = 2 component of the Vl + 2v4 band at 6557 cm -1. In the same spectral region, Peeters et al. [108] have recorded the CEA spectra of water at room temperature and at 1100 K. The spectra are compared to simulations using data from the H I T R A N and HITEMP databases, showing several discrepancies. This is, however, not too surprising since the high temperature data is primarily based on calculations. These results show that CEA spectroscopy, and also CRD spectroscopy, is a promising technique to obtain experimental data at high temperatures. As a result of the fundamental differences between magnetically induced polarization effects and field free polarization effects, the PD-CRD approach cannot be applied to the study of optical activity induced by chiral compounds. Without a magnetic field, the overall polarization rotation will cancel out during each round-trip through the cavity, thereby precluding the study of field free optical activity. In order to measure the field free polarization rotation, the direction of the optical rotation must be reversed upon each reflection at the cavity mirror. This can be accomplished by placing a quarter-wave plate in front of each cavity mirror. The path length in which the optical rotation takes place is determined by the separation between the two intracavity quarter-wave plates. It is evident that 9these intracavity components must have a highly efficient antireflection coating. Vacarro and co-workers [109] used this approach to measure the optical rotation and the differential absorption induced by chiral compounds in the gas phase. A calcite prism and a multiple-order quarter-wave plate were used to produce laser light with a well-defined circular polarization. A quarter-wave plate and a prism placed behind the exit mirror of the cavity projected the light exiting the cavity into two orthogonal components, which were detected by two photomultiplier tubes. Both the detected signals displayed rapid oscillations superimposed on the decay curves. These oscillations result from the change in polarization upon each round-trip in the cavity, and their period is a measure for the amount of optical rotation. With this cavity ring-down polarimetry (CRDP) technique Vacarro and co-workers demonstrated a sensitivity of 4 x 10.8 rad cm -1 . They determined the specific rotation at 355 nm for oe-pinene, fl-pinene, cis-pinane, limonene, fenchone, and propylene oxide. From a comparison of their results with the results of solution-phase measurements, they concluded that solvent effects can be significant and nonintuitive.
CRD AND ALTERNATIVE TECHNIQUES
73
2.4 CRD and Alternative Techniques Cavity ring-down laser spectroscopy was invented at a time when laserinduced fluorescence (LIF), resonantly enhanced multiphoton ionization (REMPI), degenerate four-wave mixing (DFWM) and coherent anti-Stokes Raman spectroscopy (CARS) were fully established laser techniques for the detection of molecular species in the gas phase. An important aspect of CRD is that it measures the absorption cross section directly and is neither dependent on any excited state relaxation process (as in LIF and REMPI) nor on the influence of other electronic states in the molecule (CARS and REMPI); also the signal is proportional to the molecular density (as opposed to CARS and DFWM). CRDS works equally well in the infrared (as opposed to REMPI) as long as lasers, mirrors, and detectors are available. As shown in this chapter CRD is sensitive and versatile (all molecules have absorption bands and the long list of examples speaks for itself), but there are certain drawbacks in some experimental conditions. One shortcoming might be that the signal is integrated along the path length in the cavity and that no localized region of space can be probed, as in the focusing techniques (REMPI, DFWM, CARS). The absorption spectrum of a mixture of gases cannot be decomposed according to species such as is the case in REMPI, where mass selection can be applied. The sensitivity is related to the length of the ring-down transients, which in many cases extends over 100 gs in experiments, where short-lived species are to be detected, this absorbing period might be too long. Even for the very weak transitions, the CRD technique is yet powerful, but does not automatically outperform the Fourier-transform spectroscopic techniques in terms of sensitivity or accuracy, particularly if long absorption path lengths and large samples can be applied. In many cases both methods are equally powerful and the data from both methods can be used for validation of results; this issue was addressed in the section on 02 spectroscopy. Under some special conditions, where the experiment is limited for some reason, CRD might bring a strong advantage. Measurements on small samples, e.g. containing expensive isotopes, were mentioned. The availability of a strong homogeneous magnetic field over a limited length of 3 cm was favorably applied using cavity enhancing absorption techniques for the study of the Zeeman effect in 02 in fields up to 20 T [30]; here FT-spectroscopy would have failed. As for resolution, FT easily reaches the Doppler-limit in any wavelength region. The resolution in CRD studies depends on the bandwidth of the laser applied. For the generic pulsed CRD studies usually the easily available tunable pulsed lasers are
74
APPLICATIONS USING RING-DOWN SPECTROSCOPY
used with a typical bandwidth of 0.1 cm -1 in the visible range and larger in the UV. The development of CW-CRD techniques gives easy access to more narrowband lasers such as diode lasers. As for sensitivity, FT techniques depend on the possibility to obtain long absorption path lengths and here the availability of highly reflecting mirrors, required for multipass setups, is important. In CRD the sensitivity is directly linked to the mirror reflectivity; hence both techniques rely on the advances in coating technology.
Acknowledgment The Netherlands Foundation for Fundamental Research on Matter (FOM) and the Space Research Organization Netherlands (SRON) are gratefully acknowledged for financial support.
References 1. H. Partridge, C. W. Bauschlicher, and S. R. Langhoff, Theoretical study of the low-lying bound states of 02, J. Chem. Phys. 95, 8292-8300 (1991). 2. A. O'Keefe and D. A. G. Deacon, Cavity ring-down optical spectrometer for absorption measurements using pulsed laser sources, Rev. Sci. Instrum. 59, 2544-2551 (1988). 3. R. G. Bray, W. Henke, S. K. Liu, K. V. Reddy, and M. V. Berry, Measurement of highly forbidden optical transitions by intracavity CW dye laser spectroscopy, Chem. Phys. Lett. 47, 213-217 (1977). 4. M. Chenevier, M. A. Melieres, and F. Stoeckel, Intracavity absorption line shapes and quantitative measurements on 02, Opt. Commun. 45, 385-391 (1983). 5 9H. Naus, A. de Lange, and W. Ubachs, b 1E + g - X 3Eg (0, 0) band of oxygen isotopomers in relation to tests of the symmetrizationpostulate in 1602, Phys. Rev. A 56, 4755-4763 (1997). 6. H. Naus, S. J. van der Wiel, and W. Ubachs, Cavity-ring down spectroscopy on the b 1E + g - X 3Y;g (1, 0) band of oxygen isotopomers, J. Mol. Spectrosc. 192, 162-168 (1998). 7. H. Naus, K. Navaian, and W. Ubachs, The v-band of 1602,'160170 , 1702, and 1802, Spectrochim. Acta A 55, 1255-1262 (1999). 8. H. Naus and W. Ubachs, The b 1E+-X 3Y;g (3, 0) band of 1602 and 1802, J. Mol. Spectrosc. 193, 442-445 (1999). 9. L. Biennier and A. Campargue, High-resolution spectrum of the (3-0) band of the b 1E + g - X 3Y;g red atmospheric system of oxygen, J. Mol. Spectrosc. 188, 248-250 (1998).
REFERENCES
75
10. H. B. Babcock and L. Herzberg, Fine structure of the red system of atmospheric oxygen bands, Astrophys. J. 108, 167-190 (1948). 11. J. Brault, Detection of electric quadrupole transitions in the oxygen A band at 7600 A, J. Mol. Spectrosc. 80, 384-387 (1980). 12. R. Engeln, G. von Helden, G. Berden, and G. Meijer, Phase shift cavity ring down spectroscopy, Chem. Phys. Lett. 262, 105-109 (1996). 13. R. Engeln, G. Berden, R. Peeters, and G. Meijer, Cavity-enhanced absorption and cavity enhanced magnetic rotation spectroscopy, Rev. Sci. Instrum. 69, 3763-3769 (1998). 14. S. Xu, D. Dai, J. Xie, G. Sha, and C. Zhang, Quantitative measurements of 02 b +--X(2, 1 , 0 - 0) bands by using cavity ring-down spectroscopy, Chem. Phys. Lett. 303, 171-175 (1999). 15. N. Seiser and D. C. Robie, Pressure broadening in the oxygen b 1~;(v t - 1) +-- X3Eg(V " - 0) band by cavity ring-down spectroscopy, Chem. Phys. Lett. 282, 263-267 (1998). 16. S. M. Newman, I. C. Lane, A. J. Orr-Ewing, D. A. Newnham, and J. Ballard, Integrated absorption intensity and Einstein coefficients for the 02 a 1Ag-X"3~]g (0,0) transition: A comparison of cavity ring-down and high resolution Fourier transform spectroscopy with a long path absorption cell, J. Chem. Phys. 110, 10749-10757 (1999). 17. H. C. Miller, J. E. McCord, J. Choy, and G. D. Hager, Measurement of the radiative lifetime of 02 (a lag) using cavity ring-down spectroscopy, J. Quant. Spectrosc. Radiat. Transfer 69, 305-325 (2001). 18. R. T. Jongma, M. G. H. Boogaarts, I. Holleman, and G. Meijer, Tracegas detection with cavity ring-down spectroscopy, Rev. Sci. Instrum. 66, 2821-2828 (1995). 19. P. Zalicki and R. N. Zare, Cavity ring-down spectroscopy for quantitative absorption experiments, J. Chem. Phys. 102, 2708-2717 (1995). 20. J. T. Hodges, J. P. Looney, and R. D. van Zee, Laser bandwidth effects in quantitative cavity ring-down spectroscopy, Appl. Opt. 35, 4112-4116 (1996). 21. J. Martin, B. A. Paldus, P. Zalicki, E. H. Wahl, T. G. Owano, J. S. Harris, C. H. Kruger, and R. N. Zare, Cavity ring-down spectroscopy with Fourier-transform limited pulses, Chem. Phys. Lett. 258, 63-70 (1996). 22. D. L. Huestis, R. A. Copeland, K. Knutsen, T. G. Slanger, R. T. Jongma, M. G. H. Boogaarts, and G. Meijer, Branch oscillator strengths for the Herzberg absorption systems in oxygen, Can. J. Phys. 72, 1109-1121 (1994). 23. T. G. Slanger, D. L. Huestis, P. C. Cosby, H. Naus, and G. Meijer, 02 photoabsorption in the 40950-41300 cm -1 region: New Herzberg bands, new absorption lines, and improved spectroscopic data, J. Chem. Phys. 105, 9393-9402 (1996).
76
APPLICATIONS USING RING-DOWN SPECTROSCOPY
24. K. Yoshino, J. R. Esmond, J. E. Murray, W. H. Parkinson, A. P. Thorne, R. C. M. Learner, and G. Cox, Band oscillator strengths of the Herzberg I bands of 02, J. Chem. Phys. 103, 1243-1249 (1995). 25. K. Yoshino, J. R. Esmond, J. E. Murray, W. H. Parkinson, A. P. Thorne, R. C. M. Learner, G. Cox, and A. S.-C. Cheung, Fourier transform spectroscopy and cross section measurements of the Herzberg IlI bands of 02 at 295 K, J. Chem. Phys. 112, 9791-9801 (2000). 26. A. Jenouvrier, M.-F. M&ienne, B. Coquart, M. Carleer, S. Fally, A. C. Vandaele, C. Hermans, and R. Colin, Fourier transform spectroscopy of the 02 Herzberg bands, J. Mol. Spectrosc. 198, 136-162 (2000). 27. H. Naus and W. Ubachs, Visible absorption bands of the (O2)2-collision complex at pressures below 760 torr, Appl. Opt. 38, 3423-3428 (1999). 28. D. A. Newnham and J. Ballard, Visible absorption cross sections and integrated absorption intensities of molecular oxygen (O2 and 04), J. Geophys. Res. D 103, 28801-28815 (1998). 29. L. Biennier, D. Romanini, A. Kachanov, A. Campargue, B. BusseryHonvault, and R. Bacis, Structure and rovibrational analysis of the [O2(a 1Ag)v=0]2 <----"[02(X3~]g)v=012 transition of the 02 dimer, J. Chem. Phys. 112, 6309-6321 (2000). 30. G. Berden, R. Engeln, P. C. M. Christianen, J. C. Maan, and G. Meijer, Cavity ring down spectroscopy on the oxygen A band in magnetic fields up to 20 Tesla, Phys. Rev. A 58, 3114-3123 (1998). 31. R. Engeln, G. Berden, E. Van den Berg, and G. Meijer, Polarization dependent cavity ring down spectroscopy, J. Chem. Phys. 107, 4458-4467 (1997). 32. D. Romanini, A. A. Kachanov, N. Sadeghi, and F. Stoeckel, CW cavity ring down spectroscopy, Chem. Phys. Lett. 264, 316-322 (1997). 33. R. Peeters, G. Berden, A. Apituley, and G. Meijer, Open-path trace-gas detection of ammonia based on cavity-enhanced absorption spectroscopy, Appl. Phys. B 71, 231-236 (2000). 34. G. Berden, R. Peeters, and G. Meijer, Cavity ring-down spectroscopy: experimental schemes and applications, Int. Rev. Phys. Chem. 19, 565-607 (2000). 35. L. Gianfrani, R. W. Fox, and L. Hollberg, Cavity-enhanced absorption spectroscopy of molecular oxygen, J. Opt. Soc. Am. B 16, 2247-2254 (1999). 36. R. D. van Zee, J. T. Hodges, and J. P. Looney, Pulsed, single-mode cavity ringdown spectroscopy, Appl. Opt. 38, 3951-3960 (1999). 37. G. Meijer, M. G. H. Boogaarts, R. T. Jongma, D. H. Parker, and A. M. Wodtke, Coherent cavity ring down spectroscopy, Chem. Phys. Lett. 217, 112-116 (1994).
REFERENCES
77
38. D. Romanini and K. K. Lehmann, Ring-down cavity absorption spectroscopy of the very weak HCN overtone bands with six, seven, and eight stretching quanta, J. Chem. Phys. 99, 6287-6301 (1993). 39. D. Romanini and K. K. Lehmann, Cavity ring-down overtone spectroscopy of HCN, H13CN and HC15N, J. Chem. Phys. 102, 633-642 (1995). 40. D. Romanini and K. K. Lehmann, Calculation of the Herman-Wallis effect in the rI-Y; vibrational overtone transitions in a linear molecule: comparison with HCN experimental results, J. Chem. Phys. 105, 68-80 (1996). D. Romanini and K. K. Lehmann, Line-mixing in the 106+-000 41. overtone transition of HCN, J. Chem. Phys. 105, 81-88 (1996). 42. A. Campargue, L. Biennier, A. Garnache, A. Kachanov, D. Romanini, and M. Herman, High resolution absorption spectroscopy of the Vl = 2-6 acetylenic overtone bands of propyne: spectroscopy and dynamics, J. Chem. Phys. 111, 7888-7903 (1999). 43. D. Kleine, S. Stry, J. Lauterbach, K. Kleinermanns, and P. Hering, Measurement of the absolute intensity of the fifth CH stretching overtone of benzene using cavity ring-down spectroscopy, Chem. Phys. Lett. 312, 185-190 (1999). 44. H. Naus, W. Ubachs, P. F. Levelt, O. L. Polyanski, N. F. Zobov, and J. Tennyson, Cavity ring-down spectroscopy on water vapor in the range 555-604 nm, J. Mol. Spectrosc. 205, 117-121 (2001). 45. S. S. Brown, R. W. Wilson, and A. R. Ravishankara, Absolute intensities for the third and fourth overtone absorptions in HNO3 and H202 measured by cavity ring down spectroscopy, J. Phys. Chem. A 104, 4976-4983 (2000). 46. S. Xu, Y. Liu, G. Sha, C. Zhang, and J. Xie, The O-H stretching A v=3, 4, and 5 vibrational overtones and conformational study of 2-butanol, J. Phys. Chem. A 104, 8671-8676 (2000). 47. Y. He, M. Hippler, and M. Quack, High-resolution cavity ring-down absorption spectroscopy of nitrous oxide and chloroform using a nearinfrared cw diode laser, Chem. Phys. Lett. 289, 527-534 (1998). 48. A. O'Keefe, J. J. Scherer, A. L. Cooksy, R. Sheeks, J. Heath, and R. J. Saykally, Cavity ring down dye laser spectroscopy of jet-cooled metal clusters: Cu2 and Cu3, Chem. Phys. Lett. 172, 214-218 (1990). 49. T. Yu and M. C. Lin, Kinetics of phenyl radical reactions studied by the "cavity-ring-down" method, J. Am. Chem. Soc. 115, 4371-4372 (1993). 50. P. Zalicki, Y. Ma, R. N. Zare, E. H. Wahl, J. R. Dadamio, T. G. Owano, and C. H. Kruger, Methyl radical measurement by cavity ringdown spectroscopy, Chem. Phys. Lett. 234, 269-274 (1995). 51. M. G. H. Boogaarts, P. J. B6cker, W. M. M. Kessels, D. C. Schram, and M. C. M. van de Sanden, Cavity ringdown detection of Sill3 on the
78
APPLICATIONS USING RING-DOWN SPECTROSCOPY
A2A~ +-- )~2A1 transition in a remote Ar-H2-SiH4 plasma, Chem. Phys. Lett. 326, 400-406 (2000). J. P. Booth, G. Cunge, L. Biennier, D. Romanini, and A. Kachanov, Ultraviolet cavity ring-down spectroscopy of free radicals in etching plasmas, Chem. Phys. Lett. 317, 631-636 (2000). T. Motylewski and H. Linnartz, Cavity ring down spectroscopy on radicals in a supersonic slit nozzle discharge, Rev. Sci. Instrum. 70, 1305-1312 (1999). M. Kotterer and J. P. Maier, Electronic spectrum of C6H: 21--[--X21"I in the gas-phase detected by cavity ringdown, Chem. Phys. Lett. 266, 342-346 (1997). J. J. L. Spaanjaars, J. J. Ter Meulen, and G. Meijer, Relative predissociation rates of OH (A2E +, / = 3 ) from combined cavity ring down--laser induced fluorescence measurements, J. Chem. Phys. 107, 2242-2248 (1997). M. D. Wheeler, S. M. Newman, A. J. Orr-Ewing, and M. N. R. Ashfold, Cavity ring-down spectroscopy, J. Chem. Soc., Faraday Trans. 94, 337-351 (1998). J. Pearson, A. J. Orr-Ewing, M. N. R. Ashfold, and R. N. Dixon, Spectroscopy and predissociation dynamics of the A ~1" state of HNO, J. Chem. Phys. 106, 5850-5873 (1997). M. D. Wheeler, A. J. Orr-Ewing, M. N. R. Ashfold, and T. Ishiwata, Predissociation lifetimes of the (A2~;+, v = 1) state of the SH radical determined by cavity ring-down spectroscopy, Chem. Phys. Lett. 268, 421-428 (1997). M. D. Wheeler, A. J. Orr-Ewing, and M. N. R. Ashfold, Predissociation dynamics of the AzN + state of SH and SD, J. Chem. Phys. 107, 7591-7600 (1997). M. D. Wheeler, S. M. Newman, and A. J. Orr-Ewing, Predissociation of the B3E~- state of $2, J. Chem. Phys. 108, 6594-6605 (1998). M. D. Wheeler, S. M. Newman, T. Ishiwata, M. Kawasaki, and A. J. Orr-Ewing, Cavity ring-down spectroscopy of t h e A2r-[3/2-X21--[3/2 transition of BrO, Chem. Phys. Lett. 285, 346-351 (1998). S. M. Newman, W. H. Howie, I. C. Lane, M. R. Upson, and A. J. Orr-Ewing, Predissociation of t h e A21-[3/2 state of IO studied by cavity ring down spectroscopy, J. Chem. Soc., Faraday Trans. 94, 2681-2688 (1998). W. H. Howie, I. C. Lane, S. M. Newman, D. A. Johnson, and A. J. Orr-Ewing, The UV absorption of C 1 0 - P a r t 1. The AZr-[-X21-I spectrum at wavelengths from 285-320 nm studied by cavity ring down spectroscopy, Phys. Chem. Chem. Phys. 1, 3079-3085 (1999). broadband
52.
53.
54.
55.
56.
57.
58.
59.
60. 61.
62.
63.
REFERENCES
79
4o H. Ding, A. J. Orr-Ewing, and R. N. Dixon, Rotational structure in the 1A t t ~ I A l spectrum of formyl chloride, Phys. Chem. Chem. Phys. 1,
4181-4185 (1999). 65. W. H. Howie, I. C. Lane, and A. J. Orr-Ewing, The near ultraviolet spectrum of the FCO radical: re-assignment of transitions and predissociation of the electronically excited state, J. Chem. Phys. 113, 7237-7251 (2000). 66. M. Kotterer, J. Conceicao, and J. P. Maier, Cavity ringdown spectroscopy of molecular ions: A2I-[u+-X2~g§ (6-0) transition of N~-, Chem. Phys. Lett. 259, 233-236 (1996). 67. M. Aldener, B. Lindgren, A. Pettersson, and U. Sassenberg, Cavity ring-down laser absorption spectroscopy--nitrogen cation, Phys. Scr. 61, 62-65 (2000). 68. L. Lehr and P. Hering, Cavity ring-down spectroscopy of photochemically produced Nail for the determination of relative dipole transition moments, Appl. Phys. B 65, 595-600 (1997). 69. L. Lehr and P. Hering, Quantitative nonlinear spectroscopy: a direct comparison of degenerate four-wave mixing with cavity ring-down spectroscopy applied to Nail, IEEE J. Quantum Electron. 33, 1465-1473 (1997). 0o I. Labazan, S. Rudi6, and S. Milogevi6, Nonlinear effects in pulsed cavity ringdown spectroscopy of lithium vapor, Chem. Phys. Lett. 320, 613-622 (2000). 71. J. B. Paul, J. J. Scherer, C. P. Collier, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy and time-of-flight mass spectroscopy of jet cooled platinum silicides, J. Chem. Phys. 104, 2782-2788 (1996). 2, J. J. Scherer, J. B. Paul, C. P. Collier, A. O'Keefe, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy and time-of-flight mass spectroscopy of jet-cooled gold silicides, J. Chem. Phys. 103, 9187 (1995). 3. J. J. Scherer, J. B. Paul, C. P. Collier, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy and time-of-flight mass spectroscopy of jet-cooled silver silicides, J. Chem. Phys. 103, 113-120 (1995). 4~ J. J. Scherer, J. B. Paul, C. P. Collier, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy and time-of-flight mass spectroscopy of jet-cooled copper silicides, J. Chem. Phys. 102, 5190 (1995). 5, J. J. Scherer, J. B. Paul, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy of the jet-cooled aluminum dimer, Chem. Phys. Lett. 242, 395-400 (1995). 6~ D. Kraus, R. J. Saykally, and V. E. Bondybey, Cavity ringdown laser absorption spectra of tungsten oxide, Chem. Phys. Lett. 295, 285-288 (1998).
80
APPLICATIONS USING RING-DOWN SPECTROSCOPY
77. D. Kraus, R. J. Saykally, and V. E. Bondybey, Cavity ringdown spectroscopy search for transition metal dimers, Chem. Phys. 247, 431-434 (1999). 8~ J. J. Scherer, J. B. Paul, A. O'Keefe, and R. J. Saykally, CRLAS: a new analytical technique for cluster science, Advances in Metal and Semiconductor Clusters 3, 149-180 (1995). 9o J. J. Scherer, J. B. Paul, A. O'Keefe, and R. J. Saykally, Cavity ringdown laser absorption spectroscopy: history, development, and application to pulsed molecular beams, Chem. Rev. 97, 25-51 (1997). 80. J. B. Paul, C. P. Collier, R. J. Saykally, J. J. Scherer, and A. O'Keefe, Direct measurement of water cluster concentrations by infrared cavity ringdown laser absorption spectroscopy, J. Phys. Chem. A 101, 5211-5214 (1997). 81. R. A. Provencal, J. B. Paul, K. Roth, C. Chapo, R. N. Caseas, R. J. Saykally, G. S. Tschumper, and H. F. Schaeffer III, Infrared cavity ring-down spectroscopy of methanol clusters: Single donor hydrogen bonding, J. Chem. Phys. 110, 4258-4267 (1999). 82. R. A. Provencal, R. N. Casaes, K. Roth, J. B. Paul, C. N. Chapo, R. J. Saykally, G. S. Tschumper, and H. F. Schaefer III, Hydrogen bonding in alcohol clusters: a comparative study by infrared cavity ringdown laser absorption spectroscopy, J. Phys. Chem. A 104, 1423-1429 (2000). 83. J. B. Paul, R. A. Provencal, C. Chapo, K. Roth, R. Casaes, and R. J. Saykally, Infrared cavity ringdown spectroscopy of water cluster bending vibrations, J. Phys. Chem. A 103, 2972-2974 (1999). 84. C. J. Chapo, J. B. Paul, R. A. Provencal, K. Roth, and R. J. Saykally, Is arginine zwitterionic or neutral in the gas phase? Results from IR cavity ring down spectroscopy, J. Am. Chem. Soc. 120, 12956-12957 (1998). 85. M. G. H. Boogaarts and G. Meijer, Measurement of the beam intensity in a laser desorption jet-cooling mass spectrometer, J. Chem. Phys. 103, 5269-5274 (1995). 86. A. A. Ruth, T. Fernholz, R. P. Brint, and M. W. D. Mansfield, The cavity ring-down absorption spectrum of the So~Ta and So--+SI transition of jet-cooled 4-H-l-benzopyrane-4-thione, Chem. Phys. Lett. 287, 403-411 (1998). 87. A. A. Ruth, E. K. Kim, and A. Hese, The So--+S1 cavity ring-down absorption spectrum of jet-cooled azulene: dependence of internal conversion on excess energy, Phys. Chem. Chem. Phys. 1, 5121-5128 (1999). 88. D. Romanini, L. Biennier, F. Salama, A. Kachanov, L.J. Allamandola, and F. Stoeckel, Jet-discharge cavity ring-down spectroscopy of ionized polycyclic aromatic hydrocarbons: progress in testing the PAH hypothesis for the diffuse interstellar band problem, Chem. Phys. Lett. 303, 165-170 (1999).
REFERENCES
81
89. H. Linnartz, T. Motylewski, O. Vaizert, J. P. Maier, A. J. Apponi, M. C. McCarthy, C. A. Gottlieb, and P. Thaddeus, Electronic ground and excited- state spectroscopy of C6H and C6D, J. Mol. Spectrosc. 197, 1-11 (1999). 90. H. Linnartz, T. Motylewski, and J. P. Maier, The 2FI +--X21-I electronic spectra of C8H and C10H in the gas phase, J. Chem. Phys. 109, 3819-3823 (1998). 91. D. Pfluger, T. Motylewski, H. Linnartz, W. E. Sinclair, and J. P. Maier, Rotationally resolved A21-[u+--X21Ig electronic spectrum of tetra-acetylene cation, Chem. Phys. Lett. 329, 29-35 (2000). 92. H. Linnartz, O. Vaizert, T. Motylewski, and J. P. Maier, The X3I]-2+--X3Ng electronic spectrum of linear C4 in the gas phase, J. Chem. Phys. 112, 9777-9779 (2000). 93. T. Motylewski, O. Vaizert, T. F. Giesen, H. Linnartz, and J. P. Maier, The 1FI~+-X1Eg+ electronic spectrum of C5 in the gas phase, J. Chem. Phys. 111, 6161-6163 (1999). 94. O. Vaizert, T. Motylewski, M. Wyss, E. Riaplov, H. Linnartz, and J. P. Maier, The A3N--X3N - electronic transition of HC6N, J. Chem. Phys. 114, 7918-7922 (2001). 95. Motylewski, H. T. Linnartz, O. Vaizert, J. P. Maier, G . A. Galazutdinov, F. A. Musaev, J. Krelowski, G. A. H. Walker, and D. A. Bohlender, Astrophys. J. 531,312-320 (2000). 96. C. D. Ball, M. C. McCarthy, and P. Thaddeus, Laser spectroscopy of the carbon chains HC7H and HC9H, Astrophys. J. 523, L89-L91 (1999). 97. C. D. Ball, M. C. McCarthy, and P. Thaddeus, Cavity ringdown spectroscopy of the linear carbon chains HC7H, HC9H, HCllH, and HC13H, J. Chem. Phys. 112, 10149-10155 (2000). 98. D. C. Moule, I. R. Burling, H. Liu, and E. C. Lim, The cavity ringdown spectrum of the visible electronic system of thiophosgene: an estimation of the lifetime of the TI(a3A2) triplet state, J. Chem. Phys. 111, 5027-5037 (1999). 99. K. W. Paulisse, T. O. Friday, M. L. Graske, and W. Polik, Vibronic spectroscopy and lifetime of $1 acrolein, J. Chem. Phys. 113, 184-191 (2000). 100. T. Mfiller, P. H. Vaccaro, F. P&ez-Bernal, and F. Iachello, Algebraic approach for the calculation of polyatomic Franck-Condon factors: application to the vibronically resolved absorption spectrum of disulfur monoxide ($20), Chem. Phys. Lett. 329, 271-282 (2000). 101. K. Liu, M. Dulligan, I. Bezel, A. Kolessov, and C. Wittig, Quenching of interconversion tunneling: the free HC1 stretch first overtone of (HC1)2, J. Chem. Phys. 108, 9614-9616 (1998).
82
APPLICATIONS USING R I N G - D O W N SPECTROSCOPY
102. F. Ito and T. Nakanaga, A jet-cooled infrared spectrum of the formic acid dimer by cavity ring-down spectroscopy, Chem. Phys. Lett. 318, 571-577 (2000). 103. D. Romanini, P. Dupr6, and R. Jost, Non-linear effects by continuous wave cavity ringdown spectroscopy in jet-cooled NO2, Vib. Spectrosc. 19, 93-106 (1999). 104. M. Hippler and M. Quack, CW cavity ring-down infrared absorption spectroscopy in pulsed supersonic jets: nitrous oxide and methane, Chem. Phys. Lett. 314, 273-281 (1999). 105. H. Naus and W. Ubachs, Experimental verification of Rayleighscattering cross sections, Opt. Lett. 25, 347-349 (2000). 106. E. K. Lewis, D. Reynolds, X. Li, De G. Villele, C. Leduc, D. L. Cedefio, and I. C. Manzanares, Phase shift cavity ring-down measurement of C-H (A v=6) vibrational overtone absorptions, Chem. Phys. Lett. 334, 357-364 (2001). 107. G. Berden, R. Peeters, and G. Meijer, Cavity-enhanced absorption spectroscopy of the 1.5 gm band system of jet-cooled ammonia, Chem. Phys. Lett. 307, 131-138 (1999). 108. R. Peeters, G. Berden, and G. Meijer, Near-infrared cavity enhanced absorption spectroscopy of hot water and OH in an oven and flames, Appl. Phys. B 73, 65-70 (2001). 109. T. Mfiller, K. B. Wiberg, and P. H. Vaccaro, Cavity ring-down polarimetry (CRDP): a new scheme for probing circular birefringence and circular dichroism in the gas phase, J. Phys. Chem. A 104, 5959-5968 (2000).
3. ABSORPTION DETECTION AT THE QUANTUM LIMIT: PROBING HIGH-FINESSE CAVITIES WITH MODULATION TECHNIQUES
Jun Ye, John L. Hall
JILA, National Institute of Standards and Technology and University of Colorado, Boulder, Colorado, USA
3.1 Introduction Two key ingredients are needed to achieve the highest sensitivity possible in the detection of atomic and molecular absorptions: enhancement of the absorption signal and elimination of technical noise. The absorption signal can be enhanced by using an optical cavity. However, care must be taken when measuring signals using an enhancement cavity: the sharp resonances of the cavity can introduce additional noise through frequency-to-amplitude noise conversion. Technical noise, caused by such things as cavity vibration and drifts, can also prevent one from reaching the fundamental, quantum-noise limit. In this chapter we discuss the application of various modulation techniques and appropriate experimental configurations that let one benefit from the signal enhancement aspect of a cavity while at the same time avoiding the introduction of additional noise in the detection process. As developers of sensitive absorption techniques, we would like to make these ideas and strategies available to the wider community, especially colleagues in other fields who have scientific problems that could be advanced if only a higher level of absorption sensitivity were available. We have a few such scientific problems and applications in mind ourselves, and most of this work has been motivated by two of these: an interest in the molecular vibration dynamics of weak transitions and the dream of a general method to produce an accurate and precise series of frequency references throughout the visible and nearinfrared wavelength regions. Molecular overtone transitions are suitable candidates for these studies because of the highly nonlinear characteristics of these vibrations and the generous spectral region over which they are found. With the development of the ultrasensitive absorption measurement techniques described in this chapter, such applications are headed towards fruition.
83 EXPERIMENTAL METHODS IN THE PHYSICAL SCIENCES Vol. 40 ISBN 0-12-475987-4
Published by Elsevier Science (USA) All rights of reproduction in any form reserved. ISSN 1079-4042/01 $35.00
84
ABSORPTION DETECTION TECHNIQUES
3.1.1 Signal Enhancement of Cavity Improving sensitivity by placing a sample inside an optical resonator is a well-known technique and is most commonly explained in terms of the multipass effect. In fact, it was realized in the early days of laser development that a laser cavity is useful for absorption enhancement [1], because of the multipass effect and the delicate balance between the laser gain and intracavity absorption [2, 3]. However, in current implementations, it is often preferred to separate the absorber from the laser, in order to extend the experimental flexibility and to better characterize working parameters. For example, enhancement of absorption sensitivity has been accomplished by using long, multipass absorption cells [4]. Kastler [5] suggested a Fabry-P6rot cavity could be used because its transmission is sensitive to small variations in absorption within the cavity. Cerez et al. [6] first applied this external cavity technique to record high-resolution, saturated-absorption spectra. Ma and Hall [7] were able to use an external resonator combined with the optical-heterodyne spectroscopy to achieve excellent signal-to-noise ratios. The use of an external cavity to detect molecular-overtone transitions has also been demonstrated [8]. In linearabsorption experiments, enhancement cavities have been used most extensively in the context of ring-down spectroscopy [9-11]. And finally, when the physical size of a cavity is brought down to a level such that the cavity-mode volume is near the atomic "radiative" volume, a whole new set of quantum dynamics associated with the full quantum susceptibility can be explored within the setting of cavity-quantum electrodynamics [12]. While we defer a detailed discussion of cavity-enhancement effects to Section 3.2, here we make a quick note of the advantages associated with an optical resonator. Of course, the well-known multipass effect leads to the enhancement of the effective absorption length by the factor of (2 x F/~), where F is the cavity finesse. Additionally, the cavity builds up the intracavity power, which allows for the study of nonlinear systems even with low-power laser sources. At the same time, however, the cavity reduces the output power level incident on the photodetector. Additionally, the geometrical self-cleaning and matching of the two counterpropagating waves inside the cavity are important both for eliminating pointing-direction related noise and for obtaining narrow, unshifted resonance lines [13]. Finally, a stable cavity can be used to stabilize a laser's frequency, by locking the laser to a cavity resonance, thereby reducing the detection noise.
INTRODUCTION
85
3.1.2 Issues Related to the Technical and Fundamental Noise
Direct absorption measurements often suffer from intensity noise on the laser. The phenomenon of increasing noise amplitudes towards the lowfrequency region of the intensity spectrum has motivated development of modulation schemes, such as frequency-modulation (FM) spectroscopy, to encode and recover signals in frequency intervals with minimal technical noise. The use of an enhancement cavity, while effective in enhancing a signal, can also introduce some extra technical noise. For example, when the relative frequency fluctuation between the laser and the cavity resonance is of the same order as the cavity linewidth, frequency noise will be converted to amplitude noise in the detected signal, if a naive, direct absorption approach is adopted. Because this noise conversion is inversely proportional to the cavity linewidth, the same enhancement factor that has improved the signal size can play an equal--but deleterious--role. Thus, while it is important to reduce noise amplitudes in the first place, it is usually equally imperative that some clever signal-recovery techniques are employed to minimize the influence of the noise to the detection result. For any absorption spectroscopy, the fundamental detection limit is reached when the minimum signal amplitude is fixed by the noise level associated with the discrete nature of the photon flux, which is a Poisson distribution for an ordinary laser output without technical noise. Such performance can be achieved when the technical noise is minimized by a differential measurement. For a true differential measurement to take place, quick comparisons of on-resonance and off-resonance information are required, that is, some sort of modulation technique should be employed. When combined with an effective modulation technique, the resonant cavity scheme approaches its potential: full signal enhancement without added noise. In the following sections, we will discuss two such experimental techniques.
3.1.3 Applications of Sensitive Detection Sensitive, cavity-enhanced detection schemes have found many different applications, including the characterization of dielectric stacks [14]; measurement of ultraslow reflector velocities [15]; atmospheric sensing [16, 17]; detection of trace gaseous species [18]; absolute determination of absorption band strengths and species concentration [19]; analysis of combustion and plasma dynamics [20, 21]; study of chemical kinetics and molecular dynamics [22, 23]; tests of fundamental physical postulates
86
ABSORPTION
DETECTION
TECHNIQUES
[2426]; improvements in laser stabilization and optical-frequency metrology [27-29]; and research into quantum dynamics and quantum information [30-32]. Extension of these methods to study surfaces and condensed matter [33], and potential applications as medical instruments only makes the field more exciting [34, 35].
3.2 Cavity Enhancement: A Simple Physics Picture In this section we will discuss the basic physics associated with the cavityenhancement effect. We will derive some useful equations from different approaches and show that, like many other subjects, the cavity-enhancement principle finds manifestation in different forms. It is useful to explore these different aspects as each highlights different practical consequences and can lead to effective experimental techniques.
3.2.1 Enhanced Radiation (Absorption) of Molecular Dipole Moments Inside a Cavity Perhaps cavity enhancement can be most simply understood from the cartoon picture shown in Figure 3.1(a). Suppose a single atom is present within the mode volume of a resonant light field of cross section a and length d. Denoting the atomic dipole moment/z and the transition wavelength 2 (angular frequency co = 2rcc/2, with c being the speed of light), we have the atomic decay rate y o ( / / 2 / ~ 3 and the atom-field coupling (Rabi) frequency g oc I~. v/co/(a 9d). In this last formula, we have substituted the electric field generated by a single (QED vacuum) photon. The absorption coefficient (o/) can be interpreted as arising from the atomic transition rate determined from the Fermi golden rule, g2/y, multiplied by the interaction time, d/c. Therefore O/o( (g2/y). d/c ~ 22/a ~ %/a, with % being the resonant absorption cross section of the atom. From this intuitive picture, we can immediately understand that if a cavity is placed around the atom with the intracavity photon lifetime resonantly increased to "Ccavity - - (2F/~). d/c, the absorption coefficient becomes O/enhanced - - ( g 2 / y ) 9"gcavity ~--- (2F/re)O/.
3.2.2 Derivation of the Length Enhancement Factor and Other Relevant Cavity Properties We can also start our discussion of cavity enhancement by looking into the cavity properties. The basic understanding is developed by placing the
CAVITY ENHANCEMENT: A SIMPLE PHYSICS PICTURE
87
Fla. 3.1. (a) An intuitive picture of atomic absorption: /x is the atomic dipole moment, 2 the transition wavelength, y the spontaneous decay rate, go the coherent coupling between the atomic dipole and the light field quantized within the volume of (a. d), and % represents the resonant atomic absorption cross-section. (b) Notation for the fields in and out of an optical cavity. The standing-wave field built-up inside the cavity is shown, along with the radiation field from the prepared atomic dipole.
absorbing atoms or molecules in an optical cavity and then comparing the absorption level to the intrinsic loss of the cavity. Thus low-loss mirrors (and consequently high-finesse cavities) help to enhance the signal contrast. Figure 3.1(b) shows a standing-wave optical resonator. The two mirrors are assumed to be identical, with power transmission of T and loss of A. The total empty cavity loss is Lcav--2(T + A). We denote the optical input power as Pin, the cavity-reflected power as Pr, and the cavity-transmitted power as Pt. The cavity finesse (F) is F .
27~
. . Lcav
7~
. T+ A
(3.1)
The resonant cavity-reflection efficiency, transmission efficiency, and intracavity buildup power, Pc, can be expressed, respectively, as
er _ ( A ) 2 Pin T + A '
et _ ( Z ) 2 Pin -
T+ A
Pc and
(l)2
Pi---~= T T + A
(3.2)
Consider a cavity of length d, filled with a weakly absorbing gas sample with an absorption coefficient of Olper unit length. (By weakly absorbing we mean the cavity-round-trip fractional power loss can be written as 1 - e x p ( - 2 o l d ) ~ 2old.) In a direct absorption measurement with a sample cell of length d, the output power is: Pout = Pin "e-~d ~ Pin (1 - old).
88
ABSORPTIONDETECTIONTECHNIQUES
Therefore, the absorption signal is (Pin.otd). When the sample is placed inside a cavity, the transmitted power is modified according to Eq. (3.2)
Pt ( Pin
r
)2 ( r)2(
T+/1+oed
~
T+A
old ) 1-2T+A
(3.3)
The detected signal contrast in the cavity transmission is therefore enhanced, ~Pt Pt
=
otd T +A
2 . . . . .
2F rc
(o~d)
(3.4)
In practice, the mirror parameters can be selected to maximize the resonant cavity transmission, by taking into consideration intracavity gas absorption. This can be especially advantageous if the anticipated sample absorption is of the same order or larger than the mirror transmission coefficients. It is also worth noting that, although the cavity finesse is determined by the sum of the mirror transmission T and loss A and neither quantity appears explicitly in Eq. (3.4), it is optimal for T to be dominant over A, because this increases the absolute size of the absorption signal (cf. Eq. (3.3)).
3.2.3 Constructive Interference of Atomic/Molecular Radiation
It is clear from Eq. (3.2) that the intracavity circulating power can be much larger than the input power. This power buildup is essential to have an appreciable level of saturation for very weak transitions, which enables resolution of sub-Doppler resonances. The strong light field drives phase coherently the atomic or molecular dipole moments. The radiation from these prepared dipole moments is the signal to be detected. However, the strong background of the unabsorbed incoming light sets the detection noise level, the shot-noise level in the ideal case (cf. Section 3.3). With the buildup cavity approach, this potentially large noise contribution from the intracavity field is reduced after the sample has been prepared and before the final detection. When the cavity is tuned onto a molecular line, the major part (determined by the cavity efficiency) of the molecular signal will leak out of the cavity to reach a detector, while a similar, or smaller, portion of the input power will be transmitted by the cavity and reach the same detector to set the shotnoise limit. The large intracavity buildup power, however, will remain trapped inside, after having prepared the phase-coherent molecular dipole moments of an enhanced magnitude. This result, although explained here
CAVITY ENHANCEMENT: A SIMPLE PHYSICS PICTURE
89
from yet another perspective, is the same cavity-enhancement effect discussed in the previous paragraphs. To mathematically clarify the above discussion, we note that the intracavity field Ec is related to the input field Ein by Ec = Ein'~/T/ (T + A) (cf. Fig. 3.1(b)). The radiation field of the prepared dipole moments can be written as Eatom = - E c " (oed/2), where the minus sign accounts for the negative interference between the sample radiation and the original field that leads to absorption. The integrated intracavity (single-pass) absorption signal is (c~d) and the factor of 1/2 reflects the fact that we are now dealing with the field amplitude. When the cavity is tuned to the atomic resonance, this atomic signal will be resonantly enhanced by constructive interference, by a factor of 2/(T + A). Here the factor of 2 accounts for the bidirectional effect in the cavity. Finally the field leaking out of the cavity is attenuated by ~/T before reaching the final photodetector. The final field can thus be expressed as
Eatom-final--~/-r'Ec(~d2) 9(T2_~_A) -
T+A
When this quantity is compared to the transmitted local-oscillator (LO) field, Ein" T/(T + A), the enhancement factor for absorption detection is again (2F/re). From a practical point-of-view, using a cavity is also attractive when the laser source has a relatively large amplitude noise. To achieve shot-noise limited detection in this case, the optimum intensity range tends to be pushed to a lower value. But when a buildup cavity is used, the detector does not have to receive the large intensity. Thus, one can take advantage of the large intracavity fields, while still having a transmitted power that allows one to operate in the shot-noise limited regime. This effect is similar to the result of polarization spectroscopy [36] or interference filtering.
3.2.4 Influence of the Cavity Physical Size Although our discussion has focused mainly on the semiclassical aspect of the light-matter interaction, it is nonetheless interesting to make a quick note on the connection between this classical picture of cavity-enhanced spectroscopy and optical cavity-based quantum electrodynamics (QED). When the cavity volume is chosen to be not far away from the critical radiation volume of the atom, namely Vcritical = 22 9c/?/, the atom-cavity interaction will have to be treated in a full quantum picture, considering that the interaction strength g can be larger than either the atomic-decay rate y or the cavitydecay rate ~c.In the optical bistability and cavity QED-literature [37, 38], there
90
ABSORPTION DETECTION TECHNIQUES
is an atomic co-operativity parameter defined as Ng2/Ky, where N is the number of atoms. This quantity signifies the level of modification of the empty cavity properties by the presence of atoms. Calculation reveals that the co-operativity parameter is nothing but the now familiar (2F/rc)oM.
3.3 Signal Encoding and Extracting: Modulation-Plus-Heterodyne Detection Parallel with signal enhancement, the issues of detection noise and how to reach the fundamental noise limit are equally critical to the implementation of sensitive spectroscopy. In this section, we will concentrate on the noise aspect of signal extraction. The ultimate detection sensitivity is achieved when we are able to observe each absorption event individually, and when the noise is limited by the uncertainty of an event occurrence. This is referred to as the shot-noise limit. From this we understand that the shot noise is associated with the discrete nature of the interaction between matter and the photon stream. The shot noise is fundamental in that it reflects the quantum nature of light from ordinary thermal sources and lasers far above threshold, sources that carry Poissonian statistics. Radiation fields can also be "squeezed" to have subPoissonian levels of fluctuation, providing an anomalously low-fluctuation level for the photocurrent (i.e. subshot noise). Here our task is just to be able to approach the standard, shot-noise level of absorption spectroscopy.
3.3.1 Fundamental Limits versus Technical Noise As noted in Section 3.2, the physical origin of the absorption process can be understood as the destructive interference between the incident radiation and the electric field generated by the coherently driven dipole moments of the sample. Therefore, direct absorption can be viewed as homodyne detection between the two interfering fields that have the same frequency. Maximum sensitivity occurs when the probe field has no amplitude noise beyond its intrinsic quantum fluctuations. Given a detection bandwidth of B and the average photocurrent idc, the associated shot-noise current is given by in = ~/2eidcB, where e is the electron charge. Taking the photodetector responsivity as rl (dimensions of electrical current per unit power incident on the detector) and P0 as the incident radiation power, then idc = ~P0 and the absorption signal isig = (oM)rIPo. The shot noise will set the fundamental sensitivity limit for straight absorption spectroscopy. When the signal-tonoise (S/N) ratio is one, i.e. when the molecular absorption is equivalent to
SIGNAL ENCODINGAND EXTRACTING
91
the shot noise in the measurement bandwidth, we obtain the minimum detectable absorption as 2_eB
(~
W~ 0
(3.6)
This direct absorption sensitivity can detect an integrated absorption of 2 x 10.8 for a 1 s averaging time for P0 = 1 mW and a reasonable rl(~ 0.8 A/W). Needless to say, this number seems optimistic when compared to experience. In practice, noise of various technical origins tends to become dominant in the low-frequency ranges, and actual laser systems display vastly more noise than the shot-noise limit. This extra technical noise may originate in inadequately smoothed power-supply potentials, laboratory vibrations that randomly dither the laser's alignment and mode structure, electromagnetic pickup from high-frequency or high-current devices, or unwanted optical feedbacks. A noisy background severely impacts the measurement of small signal changes, and typically the resultant SIN is several orders of magnitude worse than the shot-noise limit. In order to avoid excessive low-frequency noise of technical origins, one can use modulation techniques, either on the laser amplitude or frequency, to encode and then detect the absorption at a higher frequency and within a narrower bandwidth. Of course amplitude modulation of the laser will result in enhanced S/N only if the signal response is nonlinear. Otherwise one just has a high SIN way to see that the laser's power is unsteady. Reducedbackground detection techniques (such as polarization and interference spectroscopy) are also quite often used in avoiding excessive noises.
3.3.2 Introduction to OpticaI-Heterodyne Detection Before we get into the details of modulation techniques, it will be useful to become familiar with the approach of optical-heterodyne detection. The reason is that modulation detection can often be understood as a heterodyne interaction between one optical field (the carrier) and another (a modulation sideband). The optical power will generally have a time-dependent term if the applied laser field is the sum from two sources, i.e. [El(t)+ E2(t)] 2. For simplicity we have neglected important interference details by assuming the two contributing fields to be mutually mode-matched. With El(t) = E1 cos(colt) and similarly for E2 and co), when the difference of the two applied frequencies is within the detector's response bandwidth, we expect a detected photocurrent of the form i(t)--(i1-+- 2~/~i2 x sin(col -co2)t -+-/2). We refer to the cross term at the difference frequency as
92
ABSORPTION DETECTION TECHNIQUES
the heterodyne response. If we had chosen to think of the field E2 as somehow different, if for example it were a weak field being produced by a molecular sample, we can see one of the advantages of the heterodyne approach: the scale of the beat current can be increased by increasing the size of El, which is referred to as the LO field, following the convention of radio-frequency engineers. It is important that the S / N is not degraded by use of a larger LO power. Some technical remarks below return to this subject. Any real photodetector will have some output noise, even in total darkness. If we attempt direct detection of a weak signal, the incident power will need to be sufficient to overcome the detector's intrinsic noise. Consider the heterodyne case: now the signal-bearing light power is represented by the cross term between the LO field and the weak signal field, i.e. isig ~ E1E2. On the other hand, because E1 >>E2, the shot noise of the total photocurrent will be dominated by the LO power, i.e. in o( v/E~ -- El. It is only this noise term that needs to be adequately large to mask the detector noise. So by merely using a stronger LO field, we can overcome the appreciable noise produced by the amplifier circuit that converts the photocurrent into an output voltage. Ideally then, the S I N of the heterodyne detection depends only on the amplitude of the signal beam (E2) and the LO amplitude is canceled out in the final S I N ratio (see Fig. 3.2). Of course, this LO power can carry laser intensity noise to the detector as well. Any nonfundamental noise imposed on this intensity will be directly converted into unwelcome noise. But typically this noise, called "technical noise" to identify its origin, is concentrated at lower frequencies and arises from " l / f " noise processes. Technical noise seldom has a pure spectrum, and typically has spectral density that is concentrated at power-line harmonics and low frequencies associated with laboratory vibrations. Various noise processes in semiconductor devices are also restricted to audio
Det.
Atomic Dipole (g)
E
= - E. " a, afO~t~
2 + 2 E in "E atom
( E in + E atom ) 2 = E in S
2 E i n "(Ein " ~ ) oc
N
r
N E
#t
"C~
x] E 2I n
FIG. 3.2. Principle of optical (homodyne) heterodyne detection of an absorption signal.
SIGNAL ENCODING AND EXTRACTING
93
frequencies and below. So an important idea toward achieving low-noise performance is to place the information-carrying heterodyne signal at a frequency sufficiently high that a negligible level of technical noise is carried by the LO field. This is the usual motivation for using some form of modulation-based signal-recovery approach.
3.3.3 Motivation and Concept of Modulation From a time-domain point-of-view, signal detection schemes employing modulation (AC) methods permit comparison of two cases in quick succession. With suitable modulation schemes, these states can represent the on-resonant and off-resonant cases, being sampled in rapid succession. By simultaneously obtaining and subtracting these, one provides a signal channel with no output unless there is a resonance. The modulation approach therefore allows efficient extraction of weak signals from a noisy background. Lorentzian signal recovery with modulation methods has been well documented [39, 40]. The lock-in detection can provide the first-, second-, and third-derivative type of output signals. In this type of modulation spectroscopy, the modulation frequency is often chosen to be relatively low to avoid distortions of the spectral profile by the auxiliary resonances associated with modulation-induced spectral sidebands. However, this choice of low-frequency operation usually limits the achievable signal-to-noise ratio because of the excess noise of the laser source at low frequencies. To recover the optimum signal size, large modulation amplitudes (comparable to the absorption resonance width) can be employed. In this case, however, the intrinsic line shape is masked by this signal-acquisition process. Using high modulation frequencies, FM spectroscopy has become one of the most powerful techniques available for sensitive and high-speed detection of weak absorption signals [41-43]. In principle, FM spectroscopy using a high modulation frequency located in a spectral region where the amplitude noise of the laser source approaches the shot noise offers detection sensitivity close to that of Eq. (3.6). A reasonable modulation index is still needed to recover an adequate signal. However, the associated linewidth broadenings and line shape distortions associated with lowfrequency modulation processes are not present in high-frequency FM spectroscopy. Instead, one obtains added spectral features. Thus, the modulation frequency should be large compared to the width of the spectral feature under study. When scanning through an absorption resonance, each component of the FM spectrum, which consists of a central carrier and weaker sidebands, interacts with a spectral feature, thereby preserving the
94
ABSORPTIONDETECTIONTECHNIQUES
resolution of the laser. The high bandwidth, or equivalently speaking, a high Nyquist sampling frequency, associated with the high-frequency modulation enables rapid signal recovery. When received by a square-law photodiode, a pure frequency modulated spectrum will display no photocurrent at the modulation frequency. This is because the two beat signals between the carrier and either sideband are equal in magnitude but opposite in phase. Hence the net result is a perfect cancellation. However, the presence of an absorption feature is revealed as an attenuation and phase shift of one or several of the optical frequency FM components. The perfect FM balance is therefore upset, and the resultant optical signal has an amplitude-modulated component. The photocurrent signal can then be phase-sensitively detected using standard radio frequency techniques, yielding an absorption or dispersion line shape, depending upon the detection phase. The redistribution by deliberate FM modulation of some of the carrier power into FM sidebands causes only a slight penalty in the recovered signal size. When the modulation index/3 is on the order of unity or less, the FM spectrum can be approximated by a carrier [~ J0(/~)] and two first-order sidebands [~ J1 (fl)]. (Here Ji is the ith order Bessel function.) If only the carrier is tuned to interact with a narrow sub-Doppler resonance, then the detection process is intrinsically dispersion sensitive. Assuming the total probe power is still P0, the detected shot-noise current is given by v/2eB~Po(J2 + 2J 2) ~ ~/2eBrlPo. The signal current arising from the heterodyne beat between the carrier and the sidebands becomes ~/'-2rIPoJoJl" ~, with 4~ being the associated resonance phase shift. The magnitude relationship between dispersion and absorption (~b = An. cod/c and An = ~. 2/4rc, where An is the resonance-induced change of refractive index) sets the scale of the equivalent minimum absorption at the shot-noise limit:
(lYd)min-- ~
~
V -~0 Jo(fl)J1(fl)
(3.7)
The modulation-dependent function J0(/3) Jl(fl) has its maximum value of 0.34 at /3 ~ 1.1. Compared to Eq. (3.6) for the ideal case of homodyne detection, FM heterodyne detection suffers a factor of about four loss in sensitivity for fixed total optical power. This is a small price to pay for completely avoiding the laser's technical noise. The SIN loss arises in part because of the power reduction resulting from conversion of some of the main carrier to sidebands and in part from the down-conversion of shot noise from two additional spectral windows by the two sidebands. Carefully executed FM detection can often nearly reach the sensitivity limit set in Eq. (3.7).
WEAK ABSORPTION MEASURED BY FIELD DECAY (TIME-DOMAIN)
95
3.3.4 Combining Cavity Enhancement with Modulation: Introduction to the Techniques of AC Chop Ring-Down and Cavity-Enhanced FM Spectroscopy In Sections 3.4 and 3.5, we will describe two techniques that combine the cavity-enhancement approach with modulation techniques to reach the shot-noise limit. These techniques improve the detection sensitivity by a factor of (2F/re), without the added noise factors that can be associated with the use of an enhancement cavity. The two approaches are seemingly quite different, with the first one best understood from the time-domain picture in the context of cavity field ring-down, while the second one is more related to the FM spectroscopy discussed above. In the chopped ring-down spectroscopy [44], two slightly different ring-down time constants are compared, one associated with an empty cavity and the other with additional intracavity loss. Typically the comparison rate is in the audio-frequency range. The technique is suitable for linear absorption measurements. In the cavity-enhanced FM spectroscopy [45], the onresonance and off-resonance information is compared at a radio-frequency rate. This technique can be applied to both linear and nonlinear (subDoppler resolution) spectroscopy. The two techniques employ quite different experimental setups, but both achieve near shot-noise limited absorption sensitivity. As will be evident below, the two approaches are intimately related, through the common features of cavity enhancement and shot-noise limited signal recovery, and therefore the sensitivity can be calculated using (nearly) the same formula.
3.4 Weak Absorption Measured by Field Decay (Time-Domain) This section discusses a cavity-enhanced method for measuring very weak absorptions. In these measurements, two cavity modes, one probing the empty cavity and the other probing intracavity absorption, are present simultaneously but the intensities are temporally out of phase, with one mode decaying and the other rising, when viewed in cavity transmission. Heterodyne detection between the two modes reveals the dynamic time constants associated with the empty cavity and the intracavity gas absorption. Quick, differential measurement eliminates a great deal of technical noise. This is essentially an improved version of the cavity ringdown technique [10] and it yields a 1 x 10-l~ absorption sensitivity using microwatt-level laser powers.
96
ABSORPTION DETECTION TECHNIQUES
3.4.1 Brief Introduction to Cavity Ring-Down Spectroscopy Inside a high-finesse cavity, light completes many round-trips, and this effectively increases the path length by (2F/~). Cavity ring-down spectroscopy (CRDS) exploits this cavity-enhancement property by measuring the decay of the intracavity field during these many roundtrips. Because the decay rate increases when an absorber is inside the cavity, this method can be used as a spectroscopic tool. For example, an unknown absorption coefficient is determined by comparing the cavity losses when the absorber is in the cavity to the known mirror losses. Ideally, the mirror losses should be of the same order of magnitude of the absorption level to be measured, because this makes the "background" comparable to the "signal," and increases the measurement sensitivity. Because CRDS measures the decay dynamics after the field has built-up inside the cavity, it largely eliminates the technical noise on the radiation field. However, as commonly implemented, CRDS has two shortcomings. The first is that CRDS, in its ordinary form, is a long time-interval measurement. Two decay-time measurements are made, one on the empty cavity decay and the other on cavity-plus-sample. The difference between the two measured decay rates contains the desired information. However, the time between the two measurements allows slow drifts and other technical noise to contaminate the data. The second problem is that technical noise, rather than shot noise, dominates the tail-end of a ringdown signal. That is, technical noise is the dominant noise source in the signal beyond the point where the field's shot noise equals the instrument's technical noise. To overcome this limitation, a heterodyne technique was developed [11] that superimposes a large LO field onto the decay field so that the resultant beat signal can always be shot-noise limited. However some form of modulation strategy is still needed to achieve a rapid differential measurement. With a modulation technique to compare the signal against the background at short time intervals, one can accurately subtract noise in the background from the useful signal. In the context of CRDS, two decay constants are to be compared, one associated with the empty cavity loss (offresonance for the intracavity gas sample) and the other one associated with the total intracavity loss, in a quick succession. This concept can be realized if two different optical frequencies are fed into two cavity modes, one mode centered on the molecular transition and the other one far detuned from it. With the aid of cavity filtering and an intensity-stabilized laser, it is possible to reach the fundamental, quantum-noise limit to within a factor of four using this approach [44].
WEAK ABSORPTION MEASURED BY FIELD DECAY (TIME-DOMAIN)
97
3.4.2 Concept of AC Chop Ring-Down: Separating the Ring-Down Fields Associated with Cavity and Molecules Figure 3.3 illustrates the basic scheme of the AC chopped ring-down method. An intensity-stabilized, continuous-wave (cw) laser beam is split by two acousto-optic modulators (AOMs) into two beams with an impressed 1.3 GHz frequency offset, i.e. O1 - ( - f 2 2 ) = 1.3 GHz. This offset is chosen so that both beams can resonate with the cavity simultaneously (four mode orders apart, cavity free-spectral-range (FSR) splitting frequency 318.34 MHz, finesse F ~ 90,000). After being combined and mode-coupled into the cavity, the two beams give 3 laW in each cavity transmission. However, the intensities are chopped out-of-phase by the AOMs, such that only one beam is input at a time. Despite the switching, a detector viewing in cavity reflection is able to maintain the laser-cavity lock using the PoundDrever-Hall technique [46], with the radio-frequency (rf) sideband generated by the electro-optic modulator (EOM). Inside the cavity there is one decaying mode and one rising mode. That is, on the timescale of a ringdown time constant, the cavity mode that is being coupled in will rise exponentially while the other mode, with its input switched off, will decay exponentially. In transmission, the heterodyne beat waveform between the
7
Laser ]
[EOM]
Intensity .., ,.., ,._ L C o n ~
[AOMI~[AO
~
M 2] - ~ N ] K
BS
T
Laser-Cavity locking loop
F-"
[Demodulation]
[Signal]
Lens L/4
.4"
"
APD
I
PBS
C2H2
I High Finesse
PD 7 J
FIG. 3.3. Experimental setup for chopped heterodyne ring-down spectroscopy. The two AOMs provide the necessary frequency offset between the two modes and are chopped out-of-phase. The laser is locked to cavity. The heterodyne beat between the two modes at the cavity transmission is demodulated against the known carrier frequency to produce the decay signal. PD: photodiode; APD: avalanche photodiode; PBS: polarized beam splitter.
98
A BS O RP T I O N D E T E C T I O N T E C H N I Q U E S
two modes is detected. Demodulation against the known carrier frequency (1.3 GHz) then yields the heterodyne beat amplitude, which contains the information of dynamic variation of both modes. In the case of empty cavity, the beat amplitude waveform remains unchanged at neighboring chopping cycles. However, when a mode is tuned to a molecular resonance, the system exhibits two slightly different time constants. The beat waveform becomes asymmetric between the adjacent cycles, and the difference is related to the intracavity absorption. The period of chopping cycle can be chosen to roughly match the field decay time (1/e) of the empty cavity. This technique thus offers a quick comparison of on-resonance and off-resonance information and substantially suppresses technical noise. Because each ringdown waveform is measured during a chopping period that is on the order of the 1/e field decay time, shot noise dominates throughout the signal acquisition. To expand the foregoing discussion, consider the following theoretical model for this technique. Suppose the round-trip loss of the empty cavity is Lcav, the round-trip absorption of the intracavity medium is 2c~d, and the round-trip time of light within the cavity is tround-trip- For a cavity mode that is far detuned from the medium resonance, the characteristic time constant associated with the mode dynamics is given by "Ccav --
2tround-trip Lcav
(3.8)
We note that this is the 1/e decay time of the field, a relevant quantity because we deal with heterodyne detection between two fields. For the mode that is tuned to the absorption, the decay time constant becomes "Cabs --
2tround-trip Lcav -+- 2oed
(3.9)
Following the intensity-chopping scheme of Figure 3.3, let us assume that during the period of [0, At/2], mode 1 (El) of empty cavity is switched on, while mode 2 (E2), which sees the additional intracavity absorption, is switched off. After some cycles, the two field amplitudes evolve as,
[
( At)-exp(-r@vav)],E2--c2exp(--~abs)(3"10)
E1 -- Cl 1 + exp - 2rcav
Here cl and c2 are amplitude coefficients for E1 and E2, respectively. In the next half-cycle, [At/2, At], we reverse the two fields such that mode 1 is switched off and mode 2 on. The product of the two field amplitudes is what we detect in the demodulated signal of the cavity-transmitted heterodyne beat. If we compare the signal of the two neighboring half-cycles, we obtain
WEAK ABSORPTION MEASURED BY FIELD DECAY (TIME-DOMAIN)
99
the information of absorption in the following f o r m : (El E2)[o, At/2]--(E1E2)[At/2, At] = Cl c2[(1 -Jr-e(-At/2rcav))e-(t/rabs) --(1 4- e-(At/2rabs))e -(t/rcav)]
(3.11)
Figure 3.4 shows the results obtained from Eqs. (3.10) and (3.11). The chopping period, At, is chosen to be 4rcav, and the time axes have been normalized to rcav. In the case of an empty cavity, shown in the right column, the switched waveforms of modes 1 and 2 are totally symmetric, resulting in an equivalent signal of the beat amplitude during adjacent halfcycles and thus the difference gives a zero baseline. When modes 1 and 2 see different intracavity loss, as shown in the left column of the figure, there is a clear asymmetry in the heterodyne beat amplitudes between the adjacent half-cycles. The differences, as plotted in the bottom curve, show the level of With gas 1"03
switching I
~
NO gas
mode1
0.4 0.2
3.4 3.2 3.0 .
0.0
0.8 0.6 0.4 0.2 0.0 .
.
.
0.3 0.2 0.0
! i
1).8 3.6 3.4 (absorption) :).2 3.0 . 1).4
.
switching
1.0 \
model
.
.
.
.
.
i
/
(emptycavity)
.
m
3.2
.
.
.
3.1 Amplitude 3.0 ,
. !
~~
I
~'~ o x~'40~ -80 I 0
', ~
o.oo
No signal
signal I
1
I
2
I
3
I
4
I
0
I
1
2
T (normalized to cavity field (lie) decay time)
I
3
I
4
FIG. 3.4. Comparison of the demodulated ring-down curve between the empty cavity (right column) and the cavity with additional absorption (left column). The chopping period At = 4"Ccav, and the time axes are normalized to rcav. Shown from top to bottom are the switching dynamics of mode 1, mode 2, their product (beat), and the difference signal between the neighboring half-cycles. Reproduced with permission from Reference [44].
100
ABSORPTION DETECTION TECHNIQUES
additional absorption. In this case, the medium absorption is 10% of the empty cavity losses. The fundamental limit of the obtainable sensitivity of this method follows from Eqs. (3.10) and (3.11). Suppose the two modes have the same amplitude coefficients, Cl = c2 = ~/P0, in transmission, and the light is converted to a photocurrent, according to i = r/x P, where r/is as before the detector responsivity. The demodulated beat current is r/x 2E1E2/,,/-J.For simplicity of presentation, we use the small absorption limit, rcav ~ tabs, and assume At/z'cav _> 10. The difference signal of Eq. (3.11) becomes
2 P0 [e_(t/rabs)_ e-(t/rcav)] __ _rlx/~Poe-(t/rcav) /signal ~'~ q~ X [ 1 - exp(--t(r!bs
T!av) ) ] - -
rl~/-2P0 (T!bs
T!av) xt• (3.12)
Because the beat amplitude reaches the maximum when E] = E2, we have exp(-t/rcav) ~ 1/2, and t = "Ccav ln2. With the help of Eqs. (3.8) and (3.9), we obtain
/signal--r/~/2P0rcav ln2'--
2otd 2 2tround-trip
In 2 2ad =-r/P0-----
~ Lcav
The shot noise produced by the photocurrent iDC -- 7" 2 ( x / ~ / 2 ) 2 -is given as/noise = ~/2eBxrlPo/2.The resultant SIN is
/signal __ .~Po In 2 2old /noise shot noise- V ~ ~/2 Lcav
(3 13)
"
rlPo/2 (3.14)
In terms of the noise-equivalent sensitivity of single-pass integrated absorption, we set S/N =1, and 1 2~BLca v 2[feB~t (oed)min-- In 2 g ~ 0 2 = ln---2V oP0 2F
(3.15)
Compared to Eq. (3.6), we see that except for a numerical factor of about three, the shot-noise limited AC chop ring-down spectroscopy provides enhanced detection sensitivity by a factor of (2F/re) over the fundamental limit of direct absorption measurement.
3.4.3 Experimental Results of AC Chop Ring-Down Spectroscopy The concept discussed in the preceding section has been demonstrated in the measurement of rovibrational transitions of acetylene. The experiment
WEAK ABSORPTION MEASURED BY FIELD DECAY (TIME-DOMAIN)
101
used a Yb:YAG laser. The transition involved was the 3v3 R(29) overtone of C2H2, located at 1031.6528 nm, with an absorption coefficient of 4 x 10 .6 (torrcm) -1 In the 46.9 cm long cavity, a gas pressure of a few millitorr (1 t o r r = 133 Pa) was typically used, giving 1 x 10 .6 level of absorption. The beam chopping frequency was 1.4 kHz, corresponding to At = 714 Its. The cavity transmission was received by an avalanche photodiode (APD), and the beat signal was sent to an rf spectrum analyzer for demodulation. The frequency reference of the rf spectrum analyzer was tied to the rf signals used to drive the AOMs. In the linear amplitude display mode, the video output of the spectrum analyzer in zero-span-mode provides a phase-insensitive demodulation for the heterodyne beat. To measure the empty cavity finesse, both modes 1 and 2 were tuned out of the molecular resonance. Figure 3.5 shows a representative trace of the demodulated heterodyne beat ring-down waveform with an overlaid theoretical fit. As can be seen, the model presented above produces an excellent fit, giving the empty cavity field ring-down (1/e) time of 90 Its. This leads to a cavity linewidth (fullwidth at half-maximum) of 3.5 kHz and a finesse of 90,000. Within the detection bandwidth of 173 kHz, the recovered SIN was 150, about half the expected value. This was partly attributable to "ringing" of the spectrum analyzer's rf filter function, which is optimized for the frequency-domain analysis. It was not possible to use a smaller bandwidth on the instrument without distorting the signal.
c_..i c.-
300 kHz resolution BW 100 kHz video BW
0.4-
Fit cavity ringdown ( l i e field) 90 gs
v
o "o c-m
0.3-
data
0.2-
SIN =150 (calculated S/N=380)
theory fit
-o 0 i,._
t>.,
0.1-
>
..,.
O
fit residual 0.0i
i
I
i
I
I
I
i
0
50
100
150
200
250
300
350xl 0 .6
Time (s)
FIG. 3.5. A representative trace of the chopped ring-down curve of the empty cavity, obtained from the video output of the rf spectrum analyzer. A theoretical fit and the fit residual are also shown. Reproduced with permission from Reference [44].
102
ABSORPTION DETECTION TECHNIQUES
Mode 2 is then tuned to the center of the acetylene resonance, and the ring-down waveform becomes clearly asymmetric in the neighboring halfcycles. Figure 3.6 shows a set of experimental data where intracavity gas pressure was varied to generate four different intracavity absorption levels (expressed in terms of single-pass in the graph). The respective ring-down beat waveforms are shown in the left column of the figure. The absorption data (shown in the right column) were produced in the following way. First, a copy of the original data was shifted by a half-chopping cycle along the time axis. The differences between the original data and the shifted data gave the absorption signals. With a single-pass absorption of 1.7 x 10-6, the acquired SIN was 10 with a bandwidth of 173 kHz. The absorption sensitivity normalized to 1 s averaging time was then 1.6 x 10-1~ In steady state (no chopping), each mode has 3 gW (P0) in the cavity transmission. Given an r/for the APD of 0.3 A/W, the shot-noise limited sensitivity should have been ~ 1.2 x 10-11 at 1 s averaging. However, because the APD has an excess noise factor of about three, the expected minimum absorption sensitivity was ~ 4 x 10-11, which is within a factor of four of the experimental result. One can notice in Figure 3.6 that the recovered signal amplitude does not increase linearly with respect to the additional intracavity absorption. Of course one expects that if the gas absorption greatly exceeds the level of the empty cavity loss, signal saturation will occur. In Figure 3.7(a), the signal contrast is shown against the intracavity absorption normalized to the empty cavity loss. The dotted curve is calculated assuming that the coupling power to mode 2 (the absorbing mode) remains a constant. However, we know that with added loss inside, the power-coupling efficiency to the cavity changes, and the available power for mode 2 will decrease. Therefore for a fixed incident power, the signal saturation occurs sooner. This is the scenario plotted in the solid curve of Figure 3.7(a). Figure 3.7(b) illustrates saturation of the experimental data of Figure 3.6. Model from Figure 3.7(a) is used to fit the data. Remember the fraction round-trip losses of this empty cavity were 7 x 10-5. One solution to this problem is to increase the input power as the intracavity absorption level rises. A faster chopping cycle could also be used. But in any case this affects only the signal size scale; the recorded decay information is unaffected. Further improvement of the system includes the use of faster chopping cycles and the replacement of the APD with a sensitive PIN diode in a resonant matching circuit. At present the chopping frequency is limited by the locking loop between the laser and the cavity. After all, the sideband locking system also uses the heterodyne principle; in this case the beat is between the field of the direct reflection off the cavity-input
WEAK ABSORPTION MEASURED BY FIELD DECAY (TIME-DOMAIN) 103 Heterodyne ringdown waveform
Absorption signal
Empty cavity
80
20
60 40
0
20
-20 '
50.0
51.0
52.0
I
53.0 1.68 ppm
'
50.0
I
'
51.0
I
'
52.0
I
'
53.0
,o 40
20
30
0
20 9
"o9 -,-. ..13. I:::
10 0
'
50.0
51.0
52.0
I
'
53.0 4.5 ppm
"G9 >" o ,~.
c
50 40 30
9 .O
L_
20
c
c >~ -o
I0
s "~ "1"
0
9
.~
|
50.0
51.0
52.0
50
53.0 9
I
~ t.-
O 1D O~
._c
-20 I
'
51.0
I
52.0
'
I
'
53.0
20
N
o
"~
-20
9
o
c
'
50.0
I 50.0
~: i~
40
20
30
0
,
I 51.0
!
| l i 52.0 53.0 SIN = 33.3
|
20 10
-20
0 50.0
51.0
52.0
50.0
51.0
52.0
40 30
20
2O
0
10
-20
50.0
51.0
52.0
time (ms)
53.0
53.0 SIN = 37
53.0
time (ms)
FIG. 3.6. Demodulated heterodyne beat amplitudes between the two chopped cavity modes (left column) in the presence of intracavity absorption (labeled in the graph). Shown in the right column is the absorption signal obtained by differencing the beat amplitudes in the adjacent half-cycles. Reproduced with" permission from Reference [44].
104
ABSORPTION DETECTION TECHNIQUES .~_---- 0 . 2 5 c-
=
...........
(a)
......
~' 0 . 2 0 -
.--
.
,~ 0.15"~ (" o t,j
o.-~176
0.10 -
~9 0 . 0 5 r ._~ oo 0.00
,
,
.
.." ~
-"
..'"
Solid line: Cavity input coupling efficiency included
. ~
~ I 1
0
I 2
I 3
4
Intracavity gas absorption (normalized to empty cavity loss) 60x10 .3 50I--
Empty cavity loss (single pass) 35
xlO -6
40
O o
30-
c
20-
._~
(b)
f 10 - ~,
0-
--
I
10
t
20
9
signal contrast from data theoretical signal contrast I
30
Intracavity gas absorption (single pass)
I
40xl 06
FI~. 3.7. (a) Signal contrast vs. intracavity molecular absorption. The change
of cavity-input coupling efficiency is (not) taken into account in the solid (dotted) curve. (b) Experimental data from Figure 3.4, showing signal saturation when the molecular absorption approaches the empty cavity loss (35 x 10.6 single-pass). Reproduced with permission from Reference [44].
mirror and the field leaked out from the cavity storage. If the mode is being switched too fast, the cavity field does not have sufficient time to establish itself as the frequency/phase reference for the incident instantaneous laser field to compare against. The lock will then be based on equal contributions from two weaker cavity fields. Experimentally, there is a transient associated with the switching, and this leads to some problems with the lock's robustness. An alternative is to stabilize the laser on the cavity with a third mode, completely off from the molecular resonance and independent of the other two modes. The third mode can be left on all time to maintain lock while the switching can go on as before between the first two modes. The heterodyne detection rf system can conveniently filter out the contribution from the third mode. The only penalty in this arrangement is a somewhat increased level of shot noise, resulting from the added contribution of the third mode to the photocurrent. A hybrid of the on-resonance/off-resonance switch with transmission heterodyne detection against the third mode is another clear avenue for high-sensitivity detection.
WEAK ABSORPTION MEASURED BY FIELD PHASE
105
3.5 Weak Absorption Measured by Field Phase (Frequency Domain) In Section 3.4 the advantages of an enhancement cavity were discussed in a time-domain application. The signal exiting the cavity can also be measured using phase-sensitive detection methods, i.e. in frequency-domain applications. The phase of the light, along with the cavity-resonance structure, is perturbed by the molecular radiation, which leads to additional phase shifts. The objective of this section is to discuss phase-sensitive optical-heterodyne spectroscopy, using an enhancement cavity, as a tool for signal acquisition. The advantage of this approach lies in the characteristic property of FM spectroscopy: the simultaneous and continuous observation and subtraction of the signal and background optical phases. As one considers how to probe an external cavity signal with the FM technique, the first approach that comes to mind is to lock the laser frequency to a cavity resonance and then modulate that cavity mode around the desired molecular resonance while monitoring the cavity transmission. This approach is a simple lock-in derivative line shape recovery process. In order for this method to be successful, it is important to have a very tight frequency lock loop between the laser and the cavity because any laser frequency noise relative to the cavity will be converted to amplitude noise. To implement this scheme, a piezoelectric transducer (PZT) is mounted on one of the cavity mirrors. This assembly is used to modulate the cavity length, and the laser tracks the modulation. The modulation frequency is usually limited to the audio range on account of mechanical resonances and roll-off of the PZT and mirror assembly frequency response. Depending upon the laser's amplitude noise spectral distribution, the attainable modulation frequency may be too low to reach shot-noise limited detection. Moreover, the apparent line shape would also be broadened by this modulation process. However, promising results have been obtained. Using a solid-state Nd:YAG laser, locked to a high-finesse (100,000) cavity (corresponding to ~ 1 mHz relative linewidth), a cavity-dither (at 500 Hz), and lock-in detection, a detection sensitivity of 3 x 10-11(6.4 x 10-13/cm) has been measured at 1 s averaging [47].
3.5.1 Principle of NICE-OHMS To benefit from the full noise-reduction advantages of FM spectroscopy, one needs to increase the phase-modulation frequency of the probe field, usually to be much larger than the resonance linewidth under study. In addition to the laser-cavity locking issue, another obstacle remains; namely,
106
ABSORPTION DETECTION TECHNIQUES
the cavity-bandwidth limit. Specifically, F M sidebands at a high frequency are needed to eliminate low-amplitude noise, and at the same time the cavity must respond to the sidebands in exactly the same manner as it responds to the carrier. This will reduce the frequency-to-amplitude noise conversion process. This goal can be realized by frequency modulating the input laser beam at exactly the FSR of the cavity. We then detect and demodulate the cavity-transmitted light at the modulation frequency. Any small residualfrequency variations of the laser will still lead to some amplitude fluctuations and phase shifts of the transmitted carrier, but these will also lead to exactly the same amplitude fluctuations and phase shifts of the sidebands, which are transmitted on nearby cavity-axial orders. So the transmitted light still accurately represents an FM spectral triplet, with minimal AM conversion caused by the relative laser-cavity-frequency jitter. Thus the noise level can approach the intrinsic AM noise level of the laser at the FSR frequency. Figure 3.8 shows the case where the central component is used to detect the intracavity molecular resonance, illustrating how the saturated molecular dispersion causes an unbalance of the laser FM spectrum by a phase shift on the carrier component. Initially, all the FM components are lined up on their respective cavity modes. The central cavity mode will then be pulled because of the additional phase shift attributable to the molecular dispersion. After the phase-sensitive demodulation, the detector viewing the transmitted light will generate a dispersion signal in the rf beat. We refer
FIG. 3.8. The optical spectrum and the detection principle for NICE-OHMS: col denotes the laser frequency and A is the phase-modulation frequency that matches the cavity FSR. The original FM symmetry is upset when the molecular dispersion shifts a cavity resonance by COm.
WEAK ABSORPTION MEASURED BY FIELD PHASE
107
to this technique as (laser frequency-) noise-immune, cavity-enhanced, optical-heterodyne molecular spectroscopy (NICE-OHMS) [45, 29, 47, 48]. This modulation and detection scheme makes it possible to use a highfinesse cavity without introducing additional noise.
3.5.2 The Sensitivity of the NICE-OHMS Technique To estimate the sensitivity associated with NICE-OHMS, we notice that the cavity-enhancement effect applies only to the signal; no additional noise source has been introduced. Therefore the noise-equivalent absorption signal is that of Eq. (3.7) for ordinary FM spectroscopy, divided by the cavity-enhancement factor (2F/re). The power in the denominator should be that of the cavity-transmitted light, Pt. This argument can be supplemented with a more rigorous proof. Suppose the molecular dispersion changes the intracavity refractive index by An, with the shift of cavity resonance given by (An 9coo). Light going through the cavity will thus acquire an extra phase shift of r = arctan(An, coo/?'), where )i is the cavity HWHM (measured in radians/s). Following the previous treatment and under the assumption of a small 4~, we derive 2 coo cad ~ A n . coo/?' - ~-~a . . . . 7/ 2?' d
.
FSR (ad) . . . y
2F(~_d) rc
(3.16)
The signal amplitude is ~/r2rIPtJoJ 1 9r and the shot-noise limit is ~/2eBoPt. The minimum detectable absorption at S / N = 1 is thus,
(~
zc ~
~/2
2F V rlPtJo(fl)Jl(fl)
(317)
When/3 ~ 1, this sensitivity result is the same as that expressed in Eq. (3.15) for the AC chop ring-down method, except for a small numerical factor (~ 1.4), an expected result inasmuch as both methods are shot-noise limited. The difference arises from the fact that in the cavity-enhanced FM spectroscopy, some portion of the carrier power is converted to the sidebands, leading to a slight loss of sensitivity for a fixed total optical power. A numerical example helps demonstrate the potential of this technique. Suppose the modulation index,/3, is 0.5, and the photodiode responsivity, r/, is 0.85 A/W. Also, take the optical power, Pt, to be 5 mW and the detection bandwidth, B, to be 1/2n Hz, which corresponds to a 1 s time constant. Then, for a single-pass cell, the noise-equivalent integrated absorption,
108
ABSORPTION DETECTION TECHNIQUES
(o/d)min, is 2.2 x 10-8. Under the same conditions, a cavity with a finesse of 100,000 improves the sensitivity to 3.5 x 10-13. As discussed below, a noiseequivalent sensitivity of 5.2 x 10-13 of an integrated absorption at 1 s averaging has been achieved. This corresponds to an absorption of 1 x 10-14/cm for a cavity length of 50 cm [47].
3.5.3 Implementing a NICE-OHMS Experiment: Signal Line Shape, Size, and Sensitivity An experimental schematic is shown in Figure 3.9. Two electro-optic phase modulators are used to impose two sets of FM sidebands on the laser beam. Modulation at a low frequency, 3, is detected in the reflected cavity signal and is used to produce an error signal for locking the laser to the cavity. Sidebands at a higher modulation frequency, A, are set to the cavity's FSR frequency and are used to probe the intracavity molecular resonances. This signal is detected in transmission, with adequate optical isolation between the cavity and the photodiode. To study the resonance signal line shape and width, a precise scanning capability is important. A frequencyoffset locking loop is implemented to permit sweeps with rf resolution of the laser frequency relative to an independent stable reference. During the scan, the cavity FSR changes slightly. To maintain the noise-immune property, the sideband frequency must track this changing FSR.
FIG. 3.9. General schematic of the NICE-OHMS spectrometer, showing the major components of the laser-cavity locking, the transmitted sideband detection, and the precision tuning control. Reproduced with permission from Reference [47].
WEAK
ABSORPTION
MEASURED
BY FIELD
PHASE
109
As shown in Figure 3.8, the laser spectrum has three major components: the carrier at col and two phase-modulation sidebands at wi~ 4-A. Each of these three components has its own two sidebands located at 4-g away from itself, namely at col -t- g, col + A -t- ~ and col - A -t- ~ which are not shown in Figure 3.8. The three corresponding cavity-resonant peaks are denoted as coo and coo + F S R , with FSR = A. The beam reflected from the cavity carries all these frequency components and is detected by a photoreceiver. Demodulation at frequency ~ produces the servo error signal used to lock the laser frequency to the narrow linewidth cavity. Because of the additional modulation, this locking error signal now has three contributions. These are the two sideband resonances of strength J l ( f l ) 2 n e a r coo 4-A, and the carrier contribution J0(fl) 2 near coo. Together, these additively define the lock point. When none of the laser frequency components is affected by a resonance of the intracavity molecules, the servo error signal keeps the carrier, coL, and its two sidebands, coL 4-A, on the cavity resonances, coo and coo 4-A. Thus, the transmitted beam has the original, perfectly balanced FM spectra because the carrier (coL) and the sidebands (coL =k A) experience the same phase shifts and amplitude attenuations. However, when a molecular resonance affects any of these three components, the interaction converts part of the FM into AM, which is then detected by the photoreceiver viewing the transmitted light. For example, as shown in Figure 3.8, when a molecular resonance is n e a r the cavity resonance coo