CHAPTER1
Fundamentals of Capillary Electrophoresis
Theory
Kevin D. Altria 1. Introduction This section will describe the fundamental theory, equations, and definitions necessary to comprehend the concepts involved in capillary electrophoresis (CE). This is not an exhaustive treatment, but is considered sufficient to comprehend and appreciate the principles of CE. More detailed theoretical background can be obtained from a number of reference books (I-6). Developments in the field of CE are reviewed in detail annually in the journal Analytical Chemistry. For example, the 80 1 papers published in 1992-l 993 were recently reviewed (7). CE can be broadly described as high-efficiency separations of sample ions in a narrow bore (25-100 pm) capillary tube that is filled with an electrolyte solution. A typical schematic of an instrument setup is shown in Fig. 1. The principal components are a high-voltage power supply, a capillary that passes through the optical center of a detection system connected to a data acquisition device, a sample introduction system, and an autosampler. Typically, the CE instrument is controlled by a personal computer. The capillary is first filled with the required buffer solution. Sample solution (typically l-20 nL) is then introduced at the end of the capillary away from the detector (usually the anode). The capillary ends are then dipped into reservoirs containing high-voltage electrodes and the required buffer solution. One electrode is connected to a cable leading to From
Methods m Molecular Bology, Vol 52 Capdary Electrophoresrs Ed&d by K Altrla CopyrIght Humana Press Inc , Totowa, NJ
3
Altria
4 n
High voltage supply
Empty vial
0
Fig. 1. Typical instrumental setup.
the high-voltage output, whereasthe other (situatedat the detector end of the capillary) is connectedto an earthing cable. Electrodesare composed of an inert material, such as platinum. Application of a voltage (for example, 10-30 kV) across the capillary causes electrophoretic and electroendosmoticmovements(discussedlater in this chapter) resulting in the ionic speciesin the samplemoving along the capillary and passing through the on-line detector. A plot of detector response(usually UV absorbance)with time is generated,which is termedan electropherogram. 1.1. Electrophoresis
This processis the movementof sampleions under the influence of an applied voltage. The ion will move toward the appropriateelectrodeand passthrough the detector. The migration rate, or mobility, of the solute ion is governed largely by its size and number of ionic charges. For instance, a smaller ion will move faster than a larger ion with the same number of charges.Similarly, an ion with two chargeswill move faster than an ion with only one charge and similar size. The ionic mobility (pE) is therefore related to the charge/massratio (Eq. [ 11). (1)
Fundamentals
of CE Theory
5
Detector response
Fig. 2. Theoretical separation of a range of catrons. where PE = electrophoretic mobility, CJ= number of charges, IJ = solution viscosity, and r = radius of the ion. Therefore, when we separate a hypothetical mixture of ions havmg different charges and sizes, the smaller, more highly charged ions will be detected first (Fig. 2). The actual electrophoretic velocity, or speedof the solute ions, is related to their mobilities and the magnitude of the applied voltage (Eq. [2]). v=pEE
(2)
where v = velocity of the ion and E = applied voltage (volts/cm). 1.2. Electra-Osmotic Flow (EOF) Application of voltage across a capillary filled with electrolyte causes a flow of solution along the capillary. This flow effectively pumps solute ions along the capillary toward the detector. This flow occurs because of ionization of the acidic silanol groups on the inside of the capillary when m contact with the buffer solution. At high pH, these groups are dissociated resulting in a negative charged surface. To maintain electroneutrality, cations build up near the surface. When a voltage is applied, these cations migrate to the cathode (Fig. 3). The water molecules solvating the cations also move, causing a net solution flow along the capillary (Fig. 3). This effect could be considered an “electric pump.” The extent of the flow is related (Eq. [3]) to the charge on the capillary, the buffer viscosity, and dielectric constant of the buffer: pEOF=(&&/q)
(3)
where pEOF = “EOF mobility,” IJ = viscosity, and 6 = Zeta potential (charge on capillary surface).
6
Altria
Ftg. 3. Schematicof electroendosmoticflow. The level of EOF is highly dependent on electrolyte pH, since the &, potential is largely governed by the ionization of the acidic silanols. Below pH 4, the ionization is small (8), and the EOF flow rate is therefore not significant. Above -pH 9, the silanols are fully ionized and EOF is strong. The pH dependence of EOF is shown in Fig. 4. The level of EOF decreases with increased electrolyte concentration as the 6 potential is reduced. The presence of EOF allows the separation and detection of both cations and anions within a single analysis, smce EOF is sufficiently strong at pH 7, and above, to sweep anions to the cathode regardless of their charge. Analysis of a mixture of cations, neutral compounds, and anions would result in the electropherogram shown in Fig. 5. The migration times correspond to the time the individual peaks pass through the detector. The smaller anions fight more strongly against the EOF and are therefore detected later than anions with a lower mobility. Multiply charged anions will migrate more strongly against the EOF and will be detected later. Therefore, pH is clearly identified as the major operating parameter affecting the separation of ionic species, smce it governs both the solute charged state and the level of EOF. The overall migration time of a solute is therefore related to both the mobility of the solute and EOF. The term apparent mobility @A) is measured from the migration time, and is a sum of both yE and pEOF: PA = pE + JJEOF= (ZL/ tV) (4) where I= length along the capillary (cm) to detector, V = Voltage, and L = total length (cm) of the capillary.
Fundamentals
of CE Theory
7
15
1
05
0 3
4
5
6
I
9
PH
Fig. 4. Varlatron of EOF with pH. Mobility values can be calculated from migration times when both ionic and neutral components are measured. For instance, in the separation of a five-component mixture shown in Fig. 5, the mobility values for the peaks are calculated and given in Table 1. Example peak 2 = IA = (1L/ Vt) = (50 x 57 / 30,000 x 500) = 1.9 x lOA vEOF (from peak 3) = (IL / Vt) = (50 x 57 / 30,000 x 600) = 1.58 x 10q j~E=pA-pEOF=0.32x IO4 The negative values of PE for peaks 4 and 5 indicate that they are anions. The separation of ions is the simplest form of CE and is often termed Free Solution Capillary Electrophoresrs (FSCE). The separations rely
Altria
8
Frg. 5 Theoretical
separation
Calculated Mobility
of a range of ionic and neutral solutes. Table 1 Values for the Peaks m Fig. 5
Peak no
Mlgratlon time, s
PA cm2/Vs
1 2 3 4 5
400 500 600 750 900
2.38 x lo” 190x1@ 1.58 x 1W’ 1.27 x lo” 1.06 x lti
PE 080x lOA 0 32 x lOA 0 -031 x lOA -052 x 10“
I = 50 cm, L = 57 cm, and V = 30,000 V
principally on the pH-controlled dissociation of acidic groups on the solute or the protonation of basic functions on the solute. In FSCE, all neutral compounds are swept, unresolved, through the detector together (Fig. 5). Separation of neutrals is generally achieved by incorporation of anionic surfactant, at sufficient concentration to form micelles. These anionic micelles migrate against the EOF and can chromatographically interact with neutral solutes. Solutes having a large interaction will migrate later than those having little or no interaction. Use of micellar solutions is known as micellar electrokinetic capillary chromatography (also called micellar electrokinetic chromatography) and is covered in depth in Chapter 12, which is coauthored by the originator of the technique. When dealing with large biomolecules, such as nucleic acids, their electrophoretic mobilities may be very similar, and FSCE is often insufficient for adequate resolution. In this case, separations are performed in
Fundamentals
of CE Theory
capillaries filled with gel solutions. In Capillary Gel Electrophoresis (CGE), a sieving effect occurs as solutes of various sizes migrate through the gel filled capillary toward the detector. Chapter 13 describes the exceptional, efficient separations that can be obtained in gel filled capillaries. The separation and quantitation of chiral samples are an important area in many industries. Highly efficient chiral CE separations (Chapter 14) can be obtained by the addition of chirally selective substances, such as cyclodextrins, into the electrolyte. Capillary electrochromatography (CEC), which is a hybrid between CE and HPLC, has been developed. In this technique, CE equipment is used to generate HPLC-type separations. Capillaries are filled with HPLC packing material, and the application of a voltage results in the EOF pumping the mobile phase through the capillary. The full details of this technology and some applications are given in Chapter 15, which is written by one of the initial developers of the technique. 1.3. Sample
Introduction
Sample can be introduced into the capillary by three techniques, all of which involve immersing the capillary end into the sample solution and exerting a force to inject sample into the capillary. The three mechanisms for introduction of sample solution into the capillary are hydrodynamic, gravity, and electrokinetic. All these methods are quantitative, and equations describing the volumes injected have been derived. Figure 6 shows the principles of operation for the three methods. 1.3.1.
Pressure Differential
In this method, the sampling end of the capillary is immersed in the sample solution and a pressure difference applied (positive pressure or vacuum). The volume of sample solution injected onto the column can be calculated: Volume = AP d411t / 128 q L (5) where AP = pressure difference (mbar), q = buffer viscosity, L = total capillary length, and d = capillary diameter (pm). Table 2 gives injection volumes (9) for l-s injections using 65-cm capillaries of varying bore, TJ= 1, and various AP values. These volumes generally correspond to sample plug lengths of
10
Al tria Hydrodynamic Pre
Siphoning
Electrokinetic
Fig. 6. Diagram
of the three sample mtroduction
methods.
Table 2 Inlectlon Volumes / s for Various Capillary Bores AP
50 pm
75 pm
100 pm
50 mbar 75 mbar 100 mbar
1 OnL 1 5 nL 2.0 nL
5 2 nL 7 8 nL 104nL
164nL 246nL 32.8 nL
cosity is inversely proportional to temperature. For example, raising the temperature from 20 to 25°C causesthe viscosity of water to change from 1.00 to 0.89 (IO). Therefore, it is essential to employ a constant temperature during routine operation to ensure reproducible injection volumes. The sampling pressure or vacuum setting is generally instrument-specific with the sampling variable being time. Lists are available from
Fundamentals
of CE Theory
instrument suppliers that equate a sampling time to the respective volume of sample (on the order of l-20 nL) introduced into the capillary. 1.3.2. Gravity Injection In this method (II), the capillary, while dipping into the sample vial, is mechanically raised above the height of the detector electrolyte vial. Typically, the sample vial may be raised 5 cm for 10 s. The volume injected (12) may be calculated: Volume = (pgAH d41YI t / 128 q 15) (6) where AH = height difference (cm), g = gravitational constant, and p = density of the liquid. An injection volume of 6.35 nL can be calculated for a 10-s injection at 5 cm using a capillary length of 67 cm and capillary diameter of 75 urn. The following values are employed in this calculation assuming water as the buffer at 20°C density is 0.99707g/mL, viscosity is 0.8904 x 1O-2 g/cm/s, g is 980 cm/s2. The sampling variables with this technique are time and the height the sample is raised. 1.3.3. Electrokinetic In this method, the sampling end of the capillary and the high-voltage electrode are inserted into the sample solution. A voltage is then applied, causing solute ions to enter the capillary by electrophoretic migration and EOF. A greater number of more mobile ions enter the capillary, which can lead to sample bias effects. This effect can be turned to advantage (1.3), especially when attempting to quantify trace levels of small ions. The amount introduced during electrokinetic sampling is related to a variety of factors (Eq. [6]). Q=[(~E+uEOF)VlY-Ir*Ct/L] (7) where Q = amount injected, C = concentration of sample, and r = capillary radius. The sampling variables are the level and polarity of voltage, and sampling time. Various modifications to these sample introductory procedures can be used to increase the volume of sample solution introduced onto the capillary dramatically. This has considerable benefit in terms of increased sensitivity. The various schemes reported are reviewed in Chapter 16 by the two most active developers of these techniques.
12
Altria
1.4. Peak EfCciency The capillary format employed in CE minimizes most sources of band broadening that occur in conventional electrophoresis or m HPLC. l
l
l
Joule heating-the heat generated within the capillary during the voltage appllcatlon IS effectively dlsslpated through the capillary walls, which reduces convection-related band broadening encountered m conventional electrophoresrs; On-column detection-a portion of the capillary is used asthe detector, which ehmmates postseparation broadening effects owing to connections, and EOF flow dynamics-the flow profile of EOF is plug-like m nature, which mmlmlzes sample dispersion during solute transport along the capillary, compared to the lammar flow encountered m pumped systems, such as HPLC.
The major dispersive effect remaimng in CE is that of molecular diffusion of the solute as it passesalong the capillary. This diffusion IS lowest for large molecules, such as proteins, which have small diffusion coefficients. Therefore, it is possible to obtain theoretical plate counts (N) of several million for biomolecules, such as nucleotldes and proteins (Chapters 17 and 19, respectively). The theoretical plate count can be calculated: N= pEV/2D
(8)
where D = diffusion coefficient. References 1 Ll, S F. Y , ed (1992) Capillary Electrophoresls, Prmclples, Practice and Applecatzons Elsevler. 2 Kuhn, R and Hoffstetter-Kuhn, eds. (1993) Capillary Electrophoresls Prznclples and Practzce. Springer-Verlag, Berlin. 3. Wernberger, R , ed ( 1993) Practical Capdlary Electrophorew. Acadermc, London 4. Grossman, P. D. and Colburn, J C., eds. (1992) Capillary Electrophoreszs’ Theory and Practice Academic, London 5 Vmdevogel, J. and Sandra, P., eds, (1992) Introduction to Mlcellar Electrobnetrc Chromatography Huthlg, Heidelberg 6. Camllleri, P., ed. (1993) m Capillary Electrophoresls Theory and Practrce CRC, Boca Raton, FL 7. Monmg, C A. and Kennedy, R T. (1994) Capillary electrophoresls Anal Chem 66,28OR-3 14R 8 Altna, K. D. and Simpson,C. F (1987) High voltage capillary zoneelectrophoresis: operating parameter effects upon electroendosmotlc flows and electrophoretic mobllities. Chromatographla, 24, 527-530
Fundamentals
of CE Theory
9 Heiger, D N. (1994) m High Performance
10 11 I2 13
13
Capdaly Electrophoresu, Hewlett Packard, Waldbronn Rush, R S and Karger, B. L. (1990) Beckman Techmcal Bulletm TIBC 104 Rose, D. J and Jorgenson, J. W. (1988) Characterisation and automation of sample mtroduction methods for capillary electrophoresis. Anal Chem 60,642-648 Olechno, J D , Tso, J M Y , Thayer, J , and Wamwright, A (199 1) Znt Lab May, 42-48. Jackson, P E and Haddad, P R. (1993) Optimisatron of inJection technique m capillary electrophoresis for the determination of trace levels of anions m environmental samples J Chromatogr 640,48 l-487
CHAPTER2
Standard
Commercial Description Kevin
Instrument
D. Altria
1. Introduction Commercial instruments have been available since 1988, and systems are currently available from over 20 suppliers. Recent survey papers have considered the relative performance and features of these commercial systems (1,2). Systems are available in both single bench-top units or in modular form. Figure 1 shows a photograph of a commercial unit. Currently the CE equipment market is dominated by the major HPLC instrument supply companies, such as Beckman (Fullerton, CA), Hewlett Packard (Waldbronn, Germany), Perkin Elmer (Foster City, CA), and Waters (Milford, MA). 2. High-Voltage Supply Separations are normally performed employing voltages in the region of 5-30 kV. Electrolyte ionic strengths are generally selected during method development, such that application of these voltages generates currents of 10-100 p-IA. Operations with currents above this level may lead to unstable, irreproducible operating conditions. On many instruments, it is possible to operate by applying constant voltage (most common), constant current, or constant power across the capillary. However, constant voltage is the most commonly employed operational mode. 3. Capillaries Generally, these are composed of fused silica with typical dimensions 25-100 pm wrde and 25-100 cm long. The exact length and bore of the From
Methods III Molecular B/ology, Vol 52 CapNary Electrophoresls Edtted by K Altna Copyright Humana Press Inc , Totowa, NJ
15
16
Altria
Fig. 1. Photographof commercial instrument.
capillary are optimized during method development.The capillaries are covered with a protective layer of polyimide, which is strongly UV absorbent.A sectionof this polyimide coatingis removedto allow on-capillary detection. Details regarding capillary treatment are given later in Section 7.4. of Chapter3. Capillaries can also have an internal coating to alter selectivity, and thesemay have specific handling requirements. Capillary volumes are on the order of a few microliters. For example, the approximate volume of a 50-pm wide, 50-cm long capillary is 1 pL. The volume can be calculated: Volume = (7cL S / 4)
(1)
4. Capillary Cartridge Many commercial instrumentshave a cartridge-like device in which the capillary is retained. This arrangementgives improved mechanical
Standard
Commercial
Instrument
17
support and guarantees consistent alignment of the capillary m the optical center of the detector. The ends of the capillary protrude from the cartridge and dip into the autosampler vials. The cartridge is manually inserted into the instrument and is then clamped into place to ensure optical alignment. Cartridges are used in conjunction with cooling systems. The cartridge may be either filled with a coolant fluid, or cooled air may be blown through the cartridge. In instruments with liquid coolant, the cartridge is filled with a fluorocarbon liquid that is maintained at the required temperature. Longer lengths of capillary are coiled on a spool within the cartridge. Capillaries are located and held by screws in noncartridge-based instruments. Particular attention should be paid to the alignment of the capillary in the optical center of the detector (see Section 7.4. of Chapter 3). If the capillary becomes blocked or broken, the cartridge should be fitted with a new capillary. Detailed specific instructions for this procedure are given in the instrument manuals. 5. Air Supply The majority of instruments employ a regulated air supply that is used to perform rinses and hydrodynamic sample injection. Individual instruments have specific requirements (detailed in the operating manuals) m terms of the type of gas required and the pressure settings. 6. Temperature Control Both the sample volume injected and solute migration times are strongly dependent on temperature. Therefore, it is important that a consistent temperature be maintained throughout an analytical sequence to ensure good migration time and peak area precision. Approaches to this are instrument-specific and involve the use of Peltier cooling devices, forced air cooling, and heating ovens. Temperature ranges typically employed are 20-50°C. Several instruments also offer the option of cooled autosamplers, which may be of benefit when analyzing temperature-sensitive samples. 7. Commercial Detection Systems All commercial instruments have W absorbancedetection as standard, and several now have diode array detection. The majority of CE methods employ UV absorbance detection. A few selected instruments also offer the possibility of fluorescence or laser-induced fluorescence detection.
18
Altria
The UV absorbance detectors are srmilar in format to those routmely encountered in HPLC. Operating wavelengths range between 190 and 760 nm, depending on the mstrument. The UV lamps employed are similar to those employed in HPLC. Some instruments are fitted with filters, which limits the choice of wavelengths. The typical portion of capillary exposed for detection purposes is 200 x 50 pm. These tmy dimensions require excellent focusing of light from the UV lamp through the capillary. This focusing is achieved through appropriate use of high-quality lens and/or fiber optics. 8. Safety Given the possible hazards associated with voltages of this level, particular attention is paid to ensuring adequate safety precautions by electrical interlocks. Inadvertent attempts to access the autosampler during operation will cause shutdown of the instrument. 9. Data Acquisition Device The output from a standard CE instrument is similar to that of an HPLC detector, i.e., plot of UV absorbance with time. Peak integration is the preferred method of quantitatton. 10. Personal Computer Controller The majority of CE systems are controlled by an external PC that externally controls all functions of the instrument. The operating parameters for each injection are preprogrammed into the PC controller. For routine analysis, an identical set of parameters can be applied to a number of samples. The flexibility of a PC controller is of great benefit during method development, where the instrument can be preprogrammed to evaluate several operating parameters m an unattended sequence. Less elaborate systems are available that are manually programmed from a front panel on the instrument. 11. Autosampler Typical autosampler capacity ranges from 20-50 samples. Sample volume requirements range from as little as 10 pL to 5 mL. Numerous injections can be made from a single vial, since the injection volumes are on the order of a few nanoliters. Autosamplers can also be preprogrammed to perform mrcropreparative fraction collection (see Chapter 9).
Standard
Commercial
Instrument
19
12. Consumables Autosampler vials and caps are instrument-specific and may only be available from the equipment suppliers. However, suitable alternatives may be available from other sources. Several reagent suppliers, such as Fluka, and CE equipment suppliers are now developing reagent ranges specifically purified and certified for use as CE reagents. Reagents available include prepared buffers, SDS solutions, and electrolyte systems for performing inorganic anion and cation determinations. References 1 Oda, R P., Spelsberg, T C , and Landers, J P (1994) Commercial capillary electrophoresis mstrumentatlon. LC-GC 12, 50,5 1 2 Watzlg, H. and Dette, C (1993) Precise quantitatwe capillary electrophoresls. Methodological and instrumental aspects. J Chromatogr 636,3 l-38
CI-I.~PTER
Typical
Operating Kevin
3
Procedures
D. Altria
1. Introduction The CE instrument is controlled by a PC. The instrument settings are defined on the PC in method files. Settings, such as temperature, autosampler vial positions, injection parameters, and rinse cycles, are defined in the steps within the method. A typical method is given below. The exact settings for the parameters are determined during method development. A method can be run for a single sample. To analyze several samples, a sequence is created in which the method, number of samples, and injection repetitions are defined. A typical method listing is given below. Additional information concerning each process is then described. Set temperature:Select appropriatetemperature Rinse 1: 1-min 0. 1MNaOH (or equivalent to regenerate capillary surface) Rinse 2: 2-min operating electrolyte Set detector: autoselection of selected h and AUFS range Injection: selected time and mode of injection
Separate:selectedtime and applied voltage level, detectorautozero, and integrator autostart 2. Set Temperature It is important to set an operating temperature, since both injection volume and migration time are temperature-related. Typical operating ranges are 20-50°C. From
Methods m Molecular Brology, Vol 52 Capillary Electrophoresls Ed&d by K Altrla CopyrIght Humana Press Inc , Totowa, NJ
21
Al tria
22 3. Rinse
1 (Regenerating
Rinse)
In this step, the capillary is flushed with a solution to clean the capillary surface. This may be performed to remove unquantified components of the previous injection, adsorbed materials, or to rehydrate surface silanols on the capillary. The settings will be defined durmg the method development. Typically, a 2-min rmse with O.lMNaOH may be performed with the drops of liquid displaced from the capillary collected into an empty vial. The rinse time and autosampler positions for the empty vial and regenerating solution vial are entered into the method. 4. Rinse
2 (Electrolyte
Rinse)
The capillary is rinsed and filled with the run electrolyte. This allows the capillary to adjust to the pH of the electrolyte. A typical rinse time may be 3 min with the displaced drops from the capillary again being collected in an empty vial. The rinse time and autosampler posittons for the empty vial and electrolyte vial are entered. 5. Set Detector
The wavelength, absorbance range, detector rise time, and autozero details are set during this step. If indirect detection is employed, then this option is selected at this stage. 6. Injection
The time and mode of injectron are designated in this step along with the autosampler position of the sample vial. See Chapter 1 for a description of the various sample injection techniques available. Hydrodynamic injection is routinely employed, although electrokinetic injection may be selected for sample matrix or sensitivity reasons. Since different instruments have varied hydrodynamic sampling settings, it may be appropriate to define an mjection volume, in nanoliters, in addition to sampling time. 7. Separate
Details regarding analysis time, autosampler positions of electrolyte vials, and the operating voltage are specified in thts step. It is normal to operate with a constant voltage. However, it is possible to perform analysis employing constant power or current.
23
Typical Operating Procedures 8. Preanalysis
Procedures
1. Install capillary and turn lamp on to warm up for 20 min. 2. Filter the electrolyte, If appropriate, using a 0.45~pm filter or eqmvalent. 3. Fill autosampler vials with electrolyte, regeneration solution, and sample as required. Ensure vials are adequately filled and that air bubbles are avoided. 4. Place the vials m the positions designated in the method and/or sequence. 5. Assign integrator to collect detector signal. 6. Perform a test separation to ensure appropriate performance. This may involve reference to systemsuitability criteria detailed in a specific method.
9. Postanalysis
Procedures
1. Rinse the capillary, typically using a 2-min rinse from a vial contammg distilled water. 2. Flush the capillary with air, typically employmg a 2-min rinse from an empty vial. 3 Remove the capillary and store. If the instrument 1sto be used again with the same capillary, it 1sbest to leave the Instrument setup, since this avoids the possibility of damage to the capillary ends. 4. Turn off the lamp and instrument. 5. Remove vials and dispose of accordingly.
10. Good Working
Practices
The following points are noted, which, if adopted, will maximize the quality of the data. 1. Reagent quality-Use only AnalaR-grade reagents or better. Specific reagents are available in specially purified form, specifically for electrophoresis. Water should be doubly distilled or HPLC grade to minimize background UV absorbance and ionic strength. Prepared electrolytes are now available from suppliers (see Chapter 2, Section 12.). 2. Electrolyte preparation and storage--Electrolyte should be prepared and stored in plastic bottles. If pH adjustment of the electrolyte is reqmred, the pH should be checked prior to use. Alternatively, the pH-unadjusted electrolyte should be stored and pH adjusted immediately prior to use. Glass storage bottles should be avoided since material (principally metal ions) leaches from the glass with time, causing an increase m pH. A suitable shelf life may be assigned as 1 mo, unless specifically determined. 3. Capillary storag+Following use,the capillary should be rinsed with dlstllled water and flushed with air. To flush with air, an empty autosampler vial is placed on the autosampler and a rinse step performed from the empty vial.
24
Altria
4. Conditionmg analyses-It is useful to conduct analyses of blanks or equtvalent at the start of a sequence to allow the system to settle. 5. Electrolyte filtration-Electrolyte should be filtered through a 0.45~pm filter or simrlar to remove particulates. Particulates would appear as apparent noise spikes on the detector baseline. Samples should also be filtered if particulate matter is presented. If sample filtration is performed, adequate recovery should be demonstrated during method validation. 6 Lamp warm-up-It is advisable to allow 15 mm for the lamp to achieve a steady baseline. If a filter mstrument 1s employed, the appropriate filter should be selected to allow the titter to reach the operating temperature. 7. Sample warm-up-The temperature within the autosampler tray can be significantly higher than room temperature. Therefore, it 1s advisable to place the samples on the instrument prior to the mltiation of the sequence This allows samples to reach a constant temperature, which is needed for consistent inlection volumes. This requirement can be avoided if conditioning analyses are conducted (see step 4). 8. Maintenance-To reduce the likelihood of instrument failure owing to voltage leakage, the instrument should be regularly cleaned with cotton wool or tissue dampened with water to remove traces of electrolyte. 9. Blanks-Blanks of dissolvmg solvent and/or matrix should be analyzed at the start of each sequence to allow the system to equilibrate. 10. Contammation-The cleanlmess of glassware and purity of dissolvmg solvent or water should be checked, especially if attempting to quantify trace levels of ions. 11. Buffer depletion-Electrolysts ofthe electrolyte during electrophorests can significantly alter the concentratton and nature of the electrolyte, This effect, termed “buffer depletton,” will be exhibited by shifts m migration time and peak area. Buffer replenishment, or the use of different buffer vials during a sequence, will mimmtze the effect of buffer depletion. Buffers should be used within their buffering pH range. 12. Capillary dedication-Separate capillaries should be reserved for each method to avoid cross contamination effects.
11. Capillary
Preparation
The following points should be considered during the preparation of a capillary prior to commencing use. These are essential to ensure that a repeatable separation is possible when switching between capillaries. 11 .l. Capillary Dimensions The internal and external diameter and the length (L) of the capillary should be specified in the method (see Fig. 1). A length of capillary
25
Typical Operating Procedures I L where I = length to detector L = total length
Fig. 1. Capillary lengths. should be measured and cut to a length slightly over that required. This allows some excess for trimrning to the exact length required. The length to the detector window (1) should also be marked prior to polyimide removal. Capillaries coated with a UV transparent material are available from Supelco (Bellefonte, PA) and do not require polyimide removal. However, these are unsuitable for use with liquid coolant-based cartridges. 11.2. Capillary
Coating
Removal
The fused silica capillaries generally used in CE are coated with polyimide to provide mechanical strength, because the exposed capillary is fragile. This coating strongly absorbs UV light and it is therefore necessary to remove the coating in the area of capillary that is to be used as the window for on-column detection. The exact distance along the capillary to the window (I) is instrument-specific and will be given in the instrument manual. The coating can be removed by several methods. The simplest method is to place the window area of the capillary briefly (ml s) in a low-heat flame from a burner or match. Avoid unnecessary time in the flame because this may cause the capillary to bend. The coating will char and the burned material can be gently removed employing a tissue dampened with methanol. If all the coating is not removed, further exposure to the flame is required. Alternatively, the coating can be removed by gently scraping the window area with a scalpel blade. An eyepiece or lens should be used to ensure that all the coating has been removed from the window, Commercial capillary window burning units have recently become available that employ a wire resistor. Application of a current across the resistor causes it to heat, which is sufficient to
26
Altria
Fig. 2. Capillary cartridge assembly.
burn off the coating material. Use of these, or similar units, is recommended because these produce a reproducible capillary window with minimal effort. It is recommended that you carefully wipe the window area with a methanol-dampened tissue to remove any remaining burned polyimide. The capillary window is fragile following removal of the coating and should be handled as little as,possible to avoid breakage. In addition, the window area should not be touched, because grease will get onto the capillary window thereby reducing light throughput.
11.3. Capillary
Installation
The capillary housing is different for individual equipment manufacturers. In several instruments, the capillaries are held in cartridges whereas in others the capillary is held by retaining devices. The cartridges often incorporate a spool around which the capillary is spooled (Fig. 2). In all cases it is essential to align the window area of the capillary in the optical centre of the capillary housing device. Exact details of this aligning process will be given in the instrument manual.
27
Typical Operating Procedures
The alignment should be checked wherever possible using an eyepiece or lens. The ends of the capillary should be trimmed to the required length using a cutting stone to ensure a neat edge. The cutting stone should be used to lightly score the capillary and cut through the polyamide protection. Use of excessive force will result in the capillary end being crushed. If the capillary end is not cut correctly, peak tailing and poor precision may be obtained because sample solution can become irreproducibly trapped in the jagged end of the capillary. Depending on the instrument, this trimming may be needed prior to window alignment. Careful measurements of the capillary length are required to ensure correct alignment of the capillary during installation. Use of too long a length can result m the capillary becoming broken during operation. If the capillary is too short, it will not dip into the electrolyte reservoirs or sample solution. 11.4. Capillary
Surface
Rehydration
The majority of capillaries employed in CE are composed of uncoated fused silica. Fused silica is treated in a high-temperature oven and the surface therefore has no residual silanols. To rehydrate this surface, a rinse with a high pH solution is required (I). Following this rinse the surface silanols are restored and a consistent electroendosomotic flow (EOF) is possible. It is recommended that a rinse with O.lMNaOH for at least 20 min should be used to regenerate the capillary prior to its initial use. The capillary should then be filled with the method-specific electrolyte. A voltage should be applied across the capillary for 20-30 min prior to injection of the first sample. This procedure is to allow time for the capillary to stabilize and to achieve a consistent EOF level. This stabilization period could alternatively be accomplished by incorporation of appropriate blank injections at the beginning of a sequence. Use of rehydration steps for capillary coatmgs should be avoided unless prescribed in the instructions supplied with the capillary. Once the capillary is regenerated, it is not necessary to rehydrate it with an extended NaOH rinse before each subsequent use, providing that it is stored correctly. The capillary should be labeled and reserved for use solely for a selected electrolyte. Failure to do this may result in severe problems with irreproducible separations.
Al tria
28 11.5. Capillary
Storage
Prior to storage, capillaries should be flushed with water and then air to remove electrolyte from the capillary. Storage filled with air prevents formation of a gel layer inside the capillary that can form if the capillary is stored filled with solution (2). If the capillaries are stored filled with electrolyte, blockages can form as the electrolyte crystallizes out. References 1 Coufal, P., Stulik, K., Claessens, H A , and Cramers, C. A. (1994) The magmtude and reproducibllrty of the electroosmotic flow in sihca capillary tubes JHRCC 17, 325-334. 2. Schwer, C and Kenndler, E. (1991) Electrophoresls m fused-silica capillaries: the
influence of orgamc solvents on the electroosmotlc velocity and the zeta potential Anal. Chem. 63,180 l-1807.
CT~APTER
Method
4
Development/Optimization Kevin D. AZtria
1. Introduction This chapter provides some possible starting points for method development and further optimization. A later subsection in this chapter presents some starting points and guidelines for method development of chiral separations. A more extensive treatment of chiral separations is given in Chapter 14. A detailed background to the parameters affecting MECC separations is given in Chapter 12. Before proceeding with any CE experimental work, the following physicochemical information on the sample should be obtained if available. 1. The aqueoussolubihty of compound over a pH range; 2. Suitable dissolving solvent, preferably water; dtlute acid or base (> 10 rr&I), or water with the mimmum level of organic solvent required to solubke the compound; 3. pK, data;
4. UV spectralinformation. Many of the variables used to adjust selectivity are similar to those employed in HPLC method development. However, additional variables are also exploited (Table 1). The majority of these parameters must be optimized during method development. Resolution and analysis time are dependent on several factors (I). Therefore, it is necessary to follow a logical method development path, such as that given below: 1. Selectelectrolyte; 2. Select capillary and dimensions; From
Methods m Molecular Bology, Vol 52 Capillary Electrophoresls Edlted by K Altrla Copyright Humana Press Inc , Totowa, NJ
29
30
Al trta Table 1 Variables AvaIlable to Develop/Optlmlze
Variable
Typical range
Voltage Current Capillary length Capillary bore
5-30 kV 5-250 PA 20-100 cm 25-100 pm
Capillary coating PH
Various 1 5-11.5
Surfactant m MECC
1O-200 n-&l
Organic solvent
l-30% v/v
Urea
l-7M
Ion-pair reagent
l-20 M
Amme modifiers Cyclodextrms
l-50 mM l-100 n04
Vlscoslty Electrolyte cone
Various 5-200 mM
Catiomc surfactant Injection time
I-20 mA4 I-20 s
3. 4. 5. 6. 7. 8. 9. 10.
Separation
Effect of increasmg vanable Reduced analysts time and some loss m resolution Reduced analysis time and some loss m resolution Increased analysis time and gam m resolution Increased current, better sensltlvlty, and possible efficiency reduction Change of EOF and selectlvlty Increased EOF, Increased lomzatlon of acids, and reduced lomzatlon of bases Increased retention, lower EOF, and selectivity changes Increased solublllzatlon, reduced EOF (except acetomtnle), and alteration m selectivity Increased solublllzatlon of hydrophobic solutes and Increased migration times Can reduce or increase solute charge, and can alter selectivity Reduced surface charge and reduced peak tailing Increased vlscoslty, reduced EOF, and increased solute migration if complexation occurs Reduce EOF and longer migration times Increased current, EOF and solute lomzatlon, and reduced peak tailing Reversal of EOF dfrectlon and MECC condltlons Improved signal, some loss of resolution, and loss of peak symmetry
Optimize temperature; Optimize wavelength selection; Optimize sample concentration and compositlon; Determine voltage/current; Determme rinse cycle selection, Optimize injection selectlon; Optlmlze precision (see Chapter 6); Optlmlze sensitivity (see Chapter 7).
2. Select Electrolyte
Choice
There are three main separation mode options available in CE, and selection largely depends on the charge of the test solute in solution (Fig. 1).
Development /Optimization
YES
YES
0
u
No
0 u Low pH Uncoated +ve voltage
NO
Yes 0 Low pH Coated +ve voltage
0
MECC
FREE SOLUTION CE
No
NO
0
High pH Uncoated +ve voltage
Yes
0
High pH MECC Uncoated +ve voltage
Yes
0 Low pH MECC Coated -ve voltage
High pH Coated -ve voltage
Fig. 1. Method development options flowchart. Option 1: Low-pH free solution CE (FSCE): Cationic solutes separateby virtue of their differing mobilities. Option 2: High-pH FSCE: Anionic speciesseparateby virtue of their differing mobilities. Option 3: Micellar electrokinetic capillary chromatography: Soluteschromatographically interact with a migrating micellar phase; this mechanism allows separation of both charged and neutral species.
Capillaries can be internally coated to alter selectivity or to reduce EOF (see Section 3.). These coatings can be either permanent (chemically bonded) or temporary (use of electrolyte additives). Permanent coatings are usually achieved by derivatization of the capillary wall silanols followed by a covalent binding or material, such as polyacrylamide. Electrolyte additives, suchaspolyvinyl alcohol or ethylene gylcol, are used to reduce or eliminate EOF. Surfactant additives, such as tetradecyltrimethylammoniumbromide, can be employed as additives to reversethe EOF direction (this is usedespecially in the analysis of small anions--see Section 1. of Chapter 20). If p& data are available on the test compound,an informed selection of electrolyte can be made.If the compoundis basic,then option 1 would be preferred. If the compound is acidic, option 2 would be most appro-
Al tria
32
priate. Option 3 would be applicable in all cases. However, if no p& data are available, it may be appropriate to perform preliminary method scouting experiments to cover all of the above options. The test solution should be prepared in water or in water with the minimal amount of organic solvent to solubilize the analyte (see Section 6.). Sample concentration should be adequate to produce a measurable response; typically an initial sample concentration of -0.1 mg/mL may be appropriate. The sample solution must be soluble in the electrolyte, or it will precipitate out in the column, resulting in no peaks being detected. The test solution could then be analyzed under the following three starting conditions: Electrolyte A: 25 mMNaH,PO, adjusted to pH 2.5 with cont. H3P04 Electrolyte B: 25 mM borax Electrolyte C 20 mM Na2HP04, 10 mM borax, with 50 mM SDS Typically, set a run-time of 30 mm, and apply +25 kV across a 57 cm x 50 pm capillary for each separation. Set the detector to 200 nm or an appropriate UV absorbance maxima (if known) for the selected compound. If no peak is obtained using electrolyte A, then the compound is uncharged or only marginally ionized at low pH, which effectively eliminates option 1. When using electrolyte B, the electroendosmotic flow (EOF) in the system will result in a negative peak at about 2-4 min. This position is similar to the dead volume marker in HPLC. If the sample comigrates with the EOF, then it is neutral under these conditions, which would effectively eliminate option 2. Using option 3, a peak will be obtained between 3 and 30 mm. It may then be necessary to alter conditions to achieve the required selectivity. For the method scouting experiments, use typical method setup as given below: Step I: Rinse cycle 1: O.lMNaOH (2 mm). Step II: Rinse cycle 2: electrolyte A, B, or C (2 mm). Step III: Set detector: select desired iL. Step IV: Sampling. 1 s hydrodynamic. Step V: Operating voltage: + 25 kV. Run time: 30 mm. Capillary dimensions: 50 urn x 57 cm (50 cm to detector).
Development / Optimizatton
33
Table 2 Commonly Used Electrolytes Electrolyte Phosphate Citrate Acetate MES PIPES ACES MOPS0 MOPS HEPES Trls Borate CHES CAPS
2 12 pK,l 7.21 pK,2 12 32 pK,3 3.06 pK,l 5 A0 pK,2 4 75 6 15 6.80 6 90 6 90 7 20 7 55 8 30 9 24 9 50 1040
2.1. Free Solution CE A list of common buffers and pH ranges used in FSCE is given in Table 2. The inorganic electrolytes, such as phosphate and borate, are commonly used at concentrations in the area of 10-50 mM. Higher concentrations are not generally possible because of internal heating problems within the capillary, but can improve peak shape. The blological or “Goods buffers,” such as CAPS, TRIS (Tris-[hydroxymethyl]aminomethane), and so on, can be used at higher concentrations because of their low conductivities, but problems can occur owing to their high UV absorbances compared to inorganic electrolytes, such as phosphate or borate. 2.2. Low-pH FSCE (Option 1) The principal variables that can be employed to alter separation selectivity at low pH for the separation of cationic solutes are listed below: 1. Electrolyte
pH: This IS the most useful parameter
to vary since it can be
usedto alter the chargednatureof the solute.To ensurefull protonation of the compound, a pH of 1 or more units below its p& should be selected.
34
Altria
2. Cyclodextrm concentratton: These carbohydrate addmves are available m a range of native and derivatized forms and may differentially complex wtth the migrating solute, thereby increasing its migration time and separation selectivity. Typically l-1 00 rnA4hydroxy- propyl+-cyclodextrm (2) or similar are utilized. 3. Organic solvent. Solute pK,lpKbs are altered by the addition of orgamc solvents. Typically acetonitnle, methanol, or isopropanol at levels of l-30% v/v are employed often to improve the solubility of the solute (3,4). 4. Ion-pair reagents: These additives are used to alter the net charge of the solute. Typically, l-40 mA4sodium heptanesulfomc acid (2,5,6) or similar ton-pair reagent are added to the electrolyte. 5. Electrolyte nature and concentration: Higher ionic strength buffers improve peak shape, but generally necessitate a reduced applied voltage level since the current increases with buffer concentration. It 1srecommended that the voltage be reduced sufficiently to maintain current levels below 100 PA. The use of Goods buffers (7) at high concentrations can be advantageous, since these generate relatively low currents compared to inorganic electrolytes, such as phosphate or borate. The choice of electrolyte at a particular pH also affects both the selectivity and EOF level. For instance, the level of EOF decreases acetate > phosphate > citrate > borate (8). Also, the level of current generated by a specific electrolyte can be reduced by switching to an electrolyte counterton having a smaller ionic radius, 1.e, using lithium acetate instead of sodium acetate (lithium dodecyl sulfate is commercially available, which can be used at higher concentrations m MECC than SDS). Higher ionic strengths generally lead to improved peak efficiencies (Fig. 2), because sample “stacking” is improved. Figure 2 shows the analysts of a peptide mixture using various electrolyte concentrattons. 6. Zwitterionic additives: Compounds, such as betame, gylcine, and taurine, can be added to the electrolyte to reduce tailing effects. Given their zwttteriomc nature they can be used at lOO+ rnJ4 concentrations with no significant impact on overall current (9). 7. Flow reversal by hydrodynamic coating of the capillary-the direction of EOF can be reversed by addition of catronic surfactants (Z&I I) or polybrene (12). These additives form a double layer at the capillary wall, resulting in a net positive charge. Application of a voltage therefore results in a reversal of the conventional EOF direction. Therefore, a negative applied potential is employed to cause a flow m the direction of the detector. 8. Amine modifier: Excessive peak tailmg for highly basic solutes, such as peptides, can occur when they interact with the acidic silanols on the cap-
35
Development/Optimization
O.lOO M
0.060 M
4
II
Time
10 ( Minutes
la
14
)
Fig. 2. Effect of electrolyte concentration on resolution. illary surface. Additives, such as diaminopropane (13), can be employed to reduce this interaction by removing the active sites. 2.3. High-pH
FSCE
(Option
2)
This separation option is useful for the separation of anionic solutes. The variables are similar to those considered for low-pH FSCE. To ensure full ionization of the compound a pH, 1 or more units above its pK,, should be selected. EOF is more important at higher pH values and is decreased by decreases in pH, increases in organic solvent content (except acetonitrile), and the addition of cyclodextrins. Since the solutes
36
Altria
are now anionic, the nature of the ion-pair reagent should be changed to typically l-10 mM tetraethylamtnonium bromide (TEAB).
2.4. MECC
(Option
3)
This option is useful for the separation of charged and neutral species (a more detailed background to MECC [also known as micellar electrokinetic chromatography] is given in Chapter 12). The options available in optimization of MECC separation have been summarized (14,1.5). The selectivity variables and effects are identical to high-pH FSCE. However, there are also several additional parameters that greatly affect both selectivity and migration times: 1. Surfactant type: Altering the nature of the surfactant greatly affects the chromatographic mteractlons with the mlcelle. The prmclpal alternative to SDS 1sto employ bile salt surfactants (16,17), such as sodium cholate. If the solute is anionic under the separation conditions, it may be appropriate to employ a catlomc surfactant, such as cetyltrimethylammontum bromide, and to apply a negative potential (18). Noniomc surfactants, such as Brig-35, can also be added to SDS-based MECC electrolytes to alter selectivity (f 9). 2. Surfactant concentration: Increased concentration results m a higher number of micelles, and therefore, the solute 1smore retained, resulting in a longer migration time. There is an optimal surfactant concentration for each separation, and a range should be exammed during method development. 3. Cyclodextrins: These are useful as selectivity manipulators m MECC, especially when separating hydrophobic compounds. 4. Urea: This additive can be employed to increase solubihzation of hydrophobic compounds if they are poorly retained in the mlcelle (20). If the separation 1sgood, but analysis times are too long, increase pH and/or decrease SDS concentration. If the migration times are long with poor resolution, use an organic modifier or additive. If the migration times are short with poor resolution, increase SDS concentration. If the migration times are short with moderate resolution, decrease pH, and increase SDS concentration. l
l
l
l
Given that optimization can be achieved through adjustment of an assortment of variables, the appropriate use of experimental design procedures can significantly reduce the number of experiments required. A detailed section on experimental design is given in Chapter 20. Overall,
Development/Optimization
37
the electrolyte should be chosen that gives low UV absorbance at the detection wavelength, good buffering capacity at the pH required, and sufficiently low conductance to give stable operating currents. 3. Select Capillary and Dimensions Capillaries are almost exclusively composed of fused silica material, which is relatively cheap and readily available for a number of capillary suppliers, such as Polymicro (Phoenix, AZ), SGE (Ringwood, Victoria, Australia), and Supelco (Bellefonte, PA). Generally, these capillaries are not internally coated and have bare internal walls. However, there are a number of examples where capillaries are internally coated to modify the level of EOF and to alter selectivity. This area has recently been reviewed (21). Coatings are often with polymers, such as cellulose or dextran (22), which effectively suppress EOF and can reduce sample adsorptton onto the capillary wall. Alternatively, the capillaries can be internally coated with such substances as polyethylenimine (23), which can induce an effective positive charge on the capillary wall, resulting in a reversal of electroendosmotic flow direction. Alternatively, the capillary walls may be coated with long-chain (C6-Cl8) hydrocarbons (24,25). This is achieved following reaction of the capillary wall with appropriate silanes. A detailed procedure for preparation of bonded phase capillaries has been published (26). Alternatively, capillaries internally coated with bonded polyacrylamide are also employed. Coated capillaries are available from both a number of CE instrument suppliers and capillary manufacturers. The coated capillaries are generally supplied with handling/rinsing instructions and often with recommended electrolytes for particular applications. However, it is stressedthat the majority of CE applications are performed in bare fused silica capillaries. The choice of capillary bore and length largely governs the speed and sensitivity of the method. 3.1. Capillary Length Use of a longer capillary increases migration times for two reasons. First, the length of capillary to the detector area is increased, and in addition, the voltage gradient (V/cm) is decreased moving to a longer capillary while maintaining the same applied voltage. Therefore, a doubling in capillary length dramatically increases migration time, but also gives improved resolution. Figure 3 shows use of the same separation performed on both a 27- (Fig. 3A) and a 57-cm capillary (Fig. 3B). Limita-
Altria
38
ma
am Mtgrabon
7m
coo
aoa
xl00
bme (minutes)
c
I
I iso
I t4a Mgrabon
bme
I lm
I 2m
(minutes)
Fig. 3. Effect of capillary length reduction on resolution. (A) 57 cm; (B) 27 cm.
tions to the minimum capillary length possible are specific to individual commercial instruments. If attempting to employ a relatively high electrolyte concentration, it may be advisable to apply a low voltage (i.e., l-5 kV) across a short capillary to avoid problems of excessive current generation.
Development /Optimization
39
3.2. Capillary Bore The choice of capillary internal diameter largely governs the sensittvity of the method. If sensitivity is not an issue, then it is advisable to employ a 50 pm capillary to avoid any problems with excessive current generation. For maximum sensitivity, the bore may be extended to 100 pm if necessary, but at the expense of a reduced voltage and/or electrolyte concentration since internal heating problems increase with capillary diameter. 4. Optimize Temperature Temperature plays an important part in many separations, becauseboth solute mobility and the level of EOF are temperature-related. The majority of commercial instruments have capillary thermostatting facilities, and typical operating ranges are on the order of 25-6O”C. Temperature can have a marked effect on selectivity in MECC, where increased partitioning occurs at higher temperatures. On-column chelation, for example, interactions between borate ions and sugars, may also be increased at higher temperatures (2 7). 5. Optimize Wavelength Selection The maJority of electrolyte systems employed in CE have only limited U activity, and it is therefore possible to operate in the 190-220 nm UV region, which is generally inappropriate in HPLC. Many compounds have significantly greater UV activity in this region, and these UV wavelengths are widely employed in CE. Alternatively, it may be appropriate to employ indirect detection. 6. Optimize Sample Concentration and Composition This parameter should also be optimized during method development. Excessive sample concentration can lead to severely distorted peaks. Generally, peak shapebecomes more triangular as sample concentration is increased excessively. When attempting to determine trace impurities, these high concentrations may not necessarily be avoided. However, if possible, use of lower concentrations will result in more symmetrical peaks, which will result in improved resolutions. In addition, excessive sample concentrations may lead to distorted peaks if the sample has only a limited solubility in the run buffer. Operation at high temperatures can reduce on-column solubility problems, but will alter selectivity.
Altria The efficiency and performance of a CE separation can also be greatly affected by the presence of undesired sample components, such as high levels of salt or organic solvent. The most appropriate sample solutions are water or a 1: 10 dilution of the run buffer. If the ionic strength of the sample solution is lower than the run buffer, a focusing of the sample solution, known as “stacking,” occurs during the initial seconds of the separation. Stacking improves peak efficiency and can greatly improve resolution. A full treatment of sampling stacking and its limitations is given in Chapter 16. Samples having a high ionic strength present the most difficulties m CE. In these circumstances, the stacking process works against the technique, and results m band broadening and loss of resolution. This can result in severe difficulties, since many biological samples analyzed may contain high levels of salt. To minimize this disruption, the use of short mjection times is recommended. If sensitivity permits, dilution of the sample will also reduce the effect. If these approaches are inappropriate, then a sample pretreatment, such as solid-phase extraction, may be required to remove ionic interferences. When attemptmg longer injection ttmes of high-ionic-strength samples, run failures may occur. These events are caused by boiling of the sample solution during the initial secondsof separation, because most of the heat initially generated would be in the sample zone area. If this occurs, it IS necessary to reduce the sampling time and reinject. High levels of organic solvent in the sample solution can also have an undesired impact on the quality of a separation (28). Again, the best strategy is to minimize the sample injection time. Poor water solubility may require high levels of organic solvents. Therefore, higher sample concentrations may be appropriate. Alternatively, it may be appropriate to dissolve the sample m dilute acid or base (29) if possible. Sample stability in the drssolving solvent will need to be evaluated. If excessive sampling times are attempted with samples containing high organic solvents, run failure can occur. This failure is owing to out-gassing of dissolved gases in the solvent. Ultimately, sample solvent composition is dependent on the solubility of the sample. When performing quantitative analysis, it is important to match the viscosities of all samples with each other and the standards, because the volume injected is related directly to the sample viscosity. If
Development /Optimization
41
4 6
10 a Voltage
JI
c!ONDn-XON.9
l+!l!.L
CnpUhr&Ocm(Ld)x63Bcm~t~ x '76um (id) I 376um bd) FS Temperature= 26’C BuITer 0 03M SBS and O.OBM N&rate PA - 6 82 in 66% Waler/16% Me011 PEAK IDENTIF-ICATION l- NIACINAMIDE 2 - CYANOCOBALAhUN (812) 3 - PYRIDOXINFl HCL (BE)
26 kV
016
)(I 7s (kv)
4-NJACM 6 - TiaAMINE
01
tiCL
(B1)
16 kV
4
-A.-
1
00%
1
2
3
(4 10 kV
CI
I.
i 3
6
13
TIME
18 ( Minutes
\ 23
28
9
)
Fig. 4. Influence of voltage on resolution (From ref. Z). it is impossible to match viscosities, use of an internal standard will compensate for this problem. 7. Determine Voltage/Current This factor (I) largely affects the speed and quality of a separation. Application of a high voltage reduces analysis time, but may lead to significant losses of resolution and peak efficiencies if excessive heating occurs within the capillary. The choice of operating voltage should be optimized in conJunction with the choice of electrolyte concentration, capillary dimensions, and temperature to produce an acceptable level of current. Figure 4 shows that resolution is improved at lower voltages, but at the expense of increased analysis time.
42
Altria
Many instruments may be operatedin constant current mode, which may be an advantageif temperaturefluctuateswithm a separationsequence.However, it is advisable to operate generally in constant voltage mode, since slight mterday variations in electrolyte composition will have an impact on the conducttvity, which may alter the run current significantly. 8. Determine
Rinse
Cycle
Selection
It is important to maintain a consistent EOF run-to-run since any variation results in poor migration time precision (30). Sample components, such as proteins, can become adsorbed onto the capillary surface and change the effective charge on the wall, resulting in a reduction m EOF. The adsorped material can also chromatographically interact with the solute and cause tailing. To prevent difficulties owing to adsorption and to ensure a consistent EOF, the capillary 1s flushed between injections with a dilute sodium hydroxide solution that effectively strips the top surface of the capillary wall. Typically, a 1-min rinse with 0. IM NaOH is sufficient. The capillary is then flushed with the buffer prior to injection. Other rinsing regimens may involve use of a dilute acid solution or a strong buffer solution. For example, if the run buffer is 20 mA4 phosphate, it may be appropriate to rinse with 100 mk! phosphate.It is not obvious whether a rinse stepshould be included in a particular method. Generally, it is better to include one to prevent potential problems. If the sample is an uncomplicated matrix, then it may be possible to avoid a rinse step. During method development, the rinse step should be optimized to give a good migration time precision. Having established the optimum buffer conditions, ten replicate analyses should be conducted of a typical sample solution using the selected rinse step(s) between injections. If the migration time precision is poor, it may be necessary to extend the rinse times or include additional steps. Alternatively, if good migration time precision is obtained, the possibility of reducing the time or number of rinse steps should be evaluated. During robustness testing, the time and concentration of rinse solutions should be varied to assesstheir impact on the performance of the method. Following this robustness, testing limits can be put on the rinse time(s) and rinse solution concentrations. It is also important to allow a capillary to become adjusted to new buffer conditions. Therefore, it may be appropriate to employ an extended rinse (i.e., 5 min) with buffer at the start of a sequence or change in buffers during method development.
Development /Optimization
43
9. Optimize Injection Selection The means of sample injection should be selected and optimized for the particular application. Generally, electrokinetic sampling should only be conducted when quantifying trace levels of very mobile ions or when employing gel-filled capillaries where it is not possible to perform hydrodynamic injection. Injection times should be optimized together with sample concentration to give an acceptable peak height and shape. Generally, injection times of l-10 s may be possible. More elaborate mjection schemes employing sample stacking techniques may be employed to improve sample loading--see Chapter 16 for more details. 10. Chiral Separation Methods Development This important application areaof CE is covered in greater depth in Chapter 14.However, the following section representssomeinitial starting condition suggestionsbasedon the electrophoretic characteristics of the solute in electrolyte options l-3 (Le., low-pH FSCE, high-pH FSCE, or MECC). 10.1. LowpH FSCE This is the most commonly employed condition for chiral basic compounds. A useful starting point would be a pH 2.5 phosphate buffer containing 15 mM hydroxypropyl+cyclodextrin. The method should be developed and optimized as shown in Fig. 5. 10.2. High-pH FSCE Under high-pH conditions, the EOF sweeps along the cyclodextrin, while the anions migrate after the EOF flow. Interactions with the cyclodextrin reduce migration time. An initial electrolyte may be 15 mM borax (natural pH -9.3) containing 15 mM hydroxypropyl+cyclodextrin. Figure 6 shows a schematic for the possible optimization of a method. 10.3. MECC Combinations of SDS micellar electrolyte containing cyclodextrins have been shown to achieve chiral selectivity for both neutral and charged compounds. Optimization (Fig. 7) largely involves selection of the appropriate cyclodextrin and its concentration. 10.4. MECC for Hydrophobic Chiral Compounds For large hydrophobic compounds, it may be advisable to employ a micellar electrolyte containing bile salts, such as sodium cholate, which
44
Altria
r-l
Fig. 5. Optimization of the chiral separationof a basic compound.
Fig. 6. Optimization of the chiral separationof an acidic compound. are naturally occurring chirally selective compounds. The addition of organic solvents, such as isopropanol (IPA), can have a beneficial effect on chiral resolutions (Fig. 8).
45
Development /Optimization
Fig. 7. Optimization of the chiral MECC separation.
~~
0 ~~I~/1
0 7
&, ‘
LIzzl
~,o,~,
Fig. 8. Optimization of the chiral MECC separationfor hydrophobic compounds.
11. Method
Protocol
Figure 9 gives a detailed method protocol that, when fiAly completed, would contain all the information needed to fully document the details of a particular method.
Method no. Method purpose Method condltlons Rmse 1 Rmse 2
mm with mm with
seconds employmg
InJection parameters
Detector settmgs
. .
(equivalent to nl) nm AUFS rise time
Separate
nunutes at . ,, constant
Instrument
.
supplier
Capillary
mode model no
pm,
. cm(
. . cm todetector)
(Regenerate fresh captlkuy using 0 5M NaOH for 20 mmutes prior to use) Electrolyte composition
* .
. . .
. ... Sample composltlon
. .
. sample into
Typxal weight . Sample treatment
mgJml m .. .
.
* .
.
sonicate/filter/centfuge . .
Method development details reference
page
employmg
of
Fig. 9. Typical method protocol. 46
. ..
47
Development/Optimization References
1. McLaughlm, G. M., Nolan, J. A, Lmdahl, J. L., Morrrson, J A , and Bronze& T. J. (1992) Practical drug separations by HPCE. practical considerations. J Lzquzd Chrumatogr 15,961-102 1 2 Yeo, S K., Ong, C. P., and Li, S. F. Y (1991) Optimisation of high-performance capillary electrophoresrs of plant growth regulators usmg the overlappmg resolution mapping scheme Anal Chem. 63,2222-2225. 3. Chadwick, R. K. and Hsteh, J. C. (1991) Separation of CIS and trans double bond isomers using capillary zone electrophoresrs. Anal Chem 63,2377-2380. 4. Wemmann, W., Mater, C., Baumeister, K , Przybylski, M , Parker, C E., and Tomer K. B. (1994) Isolatton of hydrophobic hpoprotems in organic solvents by pressure assisted capdlary electrophoresis for subsequent mass spectrometrtc charactertzatton. J. Chromatogr 664,271-275 5. Sctacchitano, C. J , Mopper, B., and Specchio, J. J (1994) Identificatton and separatton of five cephalosporms by mtcellar electrokmettc capillary chromatography J Chromatogr. 657,395-399. 6. Rush, R. S., Derby, P. L., Strickland, T. W., and Rhode, F. (1993) Peptide mapping and evaluation of glycopepttde mtcroheterogenelty derived from endoprotemase digestion of erythroporetin by affimty high-performance capillary electrophorests Anal. Chem 65,1834-1842. 7. Nesi, M , Chrart, M., Carrea, G , Ottolma, G., and Rrghettt, P G (1994) Caprllary electrophorests of mcotmamtde-adenine dmucleotide and nicotmamtde-adenme dinucleottde phosphate derivatrves in coated tubular columns. J Chromatogr. 670, 215-221 8. Atamna, I. Z , Metral, C J , Muschik, G. M , and Issaq, J (1990) Factors which influence the mobility, resolutron and selectrvtty in capillary zone electrophoresls III The role of the buffer cation J Lzquzd Chromatogr 13,320 l-32 10 9. Bushey, M. M and Jorgenson, J W. (1989) Capillary electrophorests of proteins in buffers containing high concentrations of zwttteriomc salts. J Chromatogr 480, 301-3 10. 10. Altrra, K. D., Goodall D. M., and Rogan, M. M. (1994) Quantitative determinatron of drug counter-ton stolchtometry by capillary electrophoresls. Chromatographza 38,637-642. 11. Lm, Y.-M. and Sheu, S.-J. (1994) Separatron of aromatic acids by reversed electroosmottc flow capillary electrophoresis. J. Chromatogr 663,239-243. 12. Honda, S., Taga, A , Kakeht, K., Koda, S., and Okamoto, Y. (1992) Determination of cefixtme and its metabohtes by high-performance caprllary electrophoresis. J Chromatogr. 590,364-368. 13. Bullock, J. A. and Yuan, L.-C. (1991) Free solution capillary elecrophoresis of basic protems m uncoated fused-silica captllary tubing. J Micro Sep 3,241-248 14. Foly, J. P. (1990) Opttmtsatton of micellar electrokmetrc chromatography Anal Chem. 62, 1302-1308. 15. Strasters, J. K. and Khaledr, M. G. (1991) Migration behaviour of cationic solutes in micellar electrokmetic capillary chromatography. Anal Chem 63, 2502-2508.
Altria
48
16 Cole, R. 0 , Sepamak, M J , Hmze, W L., Gorse, J., and Oldlges, K. (1991) Bile salt structures III micellar electrokmetlc capillary chromatography Apphcation to hydrophobic molecule separations J Chromatogr 557, 113-123. 17 Ingvardsen, L , Mlchaelsen, S., and Sorensen, H. (1994) Analysis of individual phosphohplds by high performance capillary electrophorests J Am Od Chem Sot 71, 183-188 18 Crosby, D and El Rassl, Z (1993) Mlcellar electrokmetlc capillary chromatography wtth catlomc surfactants. J Liquid Chromatogr 16,2 16 l-2 187 19 Goebel, L. K and McNalr, H M (1991) Optlmlsatlon of resolution m mlcellar electrokinetic chromatography JHRCC 14,25 20 Terabe, S., Ishlhama, Y., Nlshl, H , Fukuyama, T , and Otsuka, K (199 1) Effect of urea addltlon m mlcellar electrokinetic chromatography J Chromatogr 545,359368 2 1 Wehr, T (1993) Recent advances in capillary electrophoresls columns LC-CG 11,
14-20. 22 HJerten, S. (1993) Electrophoreszs
14, 390
23 Smith, J. T. and Rassi, El (1993) Electrophoresrs 14,396 24 Chen, M and Cassldy, R. M (1992) Bonded-phase capdlarles and the separation of morgamc ions by capillary zone electrophoresis. J Chromatogr 602,227-234 25 Dougherty, A M , Wooley, C L , WIlllams, D L , Swaile, D F , Cole, R 0 , and Sepamak, M J. (1991) Stable phases for capillary electrophoresls J Llquld Chromatogr 14,907-912 26 Cobb, K. A , Dolmk, V , and Novotony, M (1990) Electrophoresls of protems m capillaries with hydrolytically stable surface structures Anal Chem 62, 24782483. 27 Hoffstetter-Kuhn, S , Paulus, A , Gassmann, E., and Wldmer, H M (1991) Influ-
ence of borate complexation on the electrophoretlc behavlour of carbohydrates m capillary electrophoresis. Anal Chem. 63, 1541-1547 28 Ackermans, M. T , Everaerts, F M., and Beckers, J. L (1991) Determination of some drugs by mlcellar electrokmetlc capillary chromatography The pseudoeffective moblhty as parameter for screening. J Chromatogr. 585, 123-l 3 1, 29. Altria, K. D and Chanter, Y (1993) Validation of a capillary electrophoresls method for the determination of a qumolme antlblotlc and its related impurities J Chromatogr
652,459-463
30 Coufal, P , Stulik, K , Claessens, H. A., and Cramers, C. A (1994) The magnitude and reproduclblhty of the electroosmotic flow in silica capillary tubes. JHRCC 17, 325-334.
CHAPTER5 Quantitation
Procedures
Kevin D. Altria 1. Introduction The options for conducting quantitative analysis are similar to those adopted in HPLC (I). The output format is similar to an HPLC chromatogram, i.e., a plot of UV absorbance vs time. Therefore, HPLC data handling and peak integration packages are generally applicable to CE. Main component assay may be performed using external and internal standardization or by standard addition. Impurity data can be calculated and reported as either % w/w or % area/area. Although all aspects of data handling are essentially the same as tn HPLC, it is important to normalize peak areas when calculating impurities (or enantiomers) as % area/area (2). This necessity arises smce all peaks do not pass through the detector at the same velocity. Therefore, the slower-moving peaks spend longer m the detector, giving rise to longer response times and larger peak areas.This is unlike HPLC, where all solutes are pumped through the detector at a constant flow rate. To obtain a time-independent peak area, the peak area of each peak (2,3) is divided by its corresponding migration time. The calculated areas are termed “normalized” areas. This simple manipulation is conducted by most of the commercial GE software packages on instrumentation. Calculation of results without using normalization can result in calculation of incorrect results (2). For example, a synthetic precursor to the antiulcer drug ranitidine was spiked into a solution of ranitldine at the 9.3% w/w level. The two compounds have identical UV response at the detection wavelength used for the CE separation shown in Fig. 1. From
Methods in Molecular hology, Vol 52 Cap/&y Nectrophoress Edtted by K Altna Copyright Humana Press Inc , Totowa, NJ
49
Altria
50
LlB
ClAcl AJ
40
1 l.40
Mtgratlon time (minutes)
I i6rJ
I 180
Fig. 1 Separation of a ramtldme sample spiked with a known amount of a synthetic impurity (from ref. 2)
However, the calculated % area/area indicated only 6.3% area/area impurity content. Use of normalized areas gave a result of 9.1% area/
area, confirming the splkmg level. In this example, the impurity 1sunderestimated. In the case of an impurity migrating after the main peak, it would be overestimated, 2. Main Peak Assay 2.1. External Standardization Calibration solutions of weighed amounts of the standard material are prepared and analyzed. The areas of the integrated peaks are used to calculate the response factor. Sample solutions are then analyzed, and the
recorded peak areas are multiplied by the response factor to calculate the concentration in the sample solution. Using a well-controlled validated method, it is reasonable to expect RSD values of better than l-2% for peak area (4). Typical steps are: 1. Accurately weigh and transfer a portion of the reference material and samplesto appropriatevolumetric flasks
Quantitation
Procedures
51
2. Dilute to volume with the dissolving solvent. 3 Shake or somcate as required to ensure dissolution of the maternal. 4. Falter or centrifuge the sample tf requu-ed, and transfer the sample solution to autosampler vials 5. Place the vials contammg the electrolyte, rinse solution(s), calrbrations, blank, and samples mto the approprtate positions on the autosampler. 6 Perform a test injection of a test mixture to confirm the system 1soperating correctly, 7. Analyze the samples under the conditions specified m the method. A typical sequence will contain analysis of a solution of the dissolving solvent, and a number of injections of the calibration solutions and sample solutions. An appropriate sequence for four samples may be: 1. 2 3. 4. 5 6 7. 8. 9. 10. Il.
Blank. Calibration Calibration Sample 1 Sample 2 Calibration Calibration Sample 3. Sample 4. Calibratton Calibratron
1. 2. 1. 2 1. 2.
Each solution should be injected in duplicate resulting in 22 injections for this example. The integrated peak areas for the calibrations are then used to calculate the response as given below: Response factor = [weight calibration (mg) x % purity / (1) peak area x dilution volume (mL) x 1001 The response factors for all the calibration solutions are calculated and averaged to give a Mean Response Factor (MRF). The concentration of analyte in the sample, if supplied as a solution, is calculated by: mg/mL = peak area x MRF
(2)
If a solid sample was sampled, then the % w/w purity is calculated as: % w/w = [MRF x peak area x dilution volume (mL) / wt (mg)] (3) An example of the use of an internal standard is shown in Fig. 2 (5) m the determination of the drug sumatriptan in injection solutions. The use of
Altria
52
3.00
Retention
5.00
7.00
8.00
time h mhutee
Fig. 2. Separation of sumatriptan employmg an internal standard (from ref. 5). an internal standard allowed an average RSD value of < 1.O%for response
factors to be obtained. 2.2. Internal
Standardization
Steps l-7 (see Section 1.1.) for external standardization are followed, replacing the dissolving solvent by a solution of an appropriate internal standard. The concentration of the internal standard in the final sample solution should approximately match that of the sample. Calculations are modified to incorporate the internal standard. Peak area ratio (PAR) = (peak area sample / peak area mtemal standard)
(4)
Response factor = [wt (mg) x % purity / PAR x dilution volume (mL) x 1001
(5)
Solution concentration = PAR x MRF % w/w = [PAR x MRF x dilution volume (mL) / wt (mg)] 2.3. Standard Addition
(6)
This calibration procedure involves addition of known amounts of standard to the sample solutions (6). Thesespiked samplesare then analyzed and
Quantitation
53
Procedures
the peak areas are plotted against spiking level. The sample content is then calculated from the intercept and slope of the line. The sample spiking can be performed manually or can be preprogrammed into the separation method. 2.4. Manual
Standard
Addition
This procedure (6) is of particular benefit when dealing with samples of varying viscosity, such as biofluids. Since injection volumes are related to sample viscosity, it is essential to match closely the viscosity of the samples and calibration solutions. 1. Takealiquotsof the samplesolutions,typically four, into appropriate vessels. 2. Prepare four calibratron solutions over the required calibration range. 3. Add a consistent volume of each of the standard solutions to the sample solution (for example, add 1.O mL of standard solution to 10.0 mL of sample ahquot). 4. Analyze, at least in duplicate, each of the spiked samples and an unspiked sample. 5. Plot the peak areas agamst spiking level (typlcally recorded m mg/mL or mg addition). 6. Using the slope and intercept values, calculate the sample concentration. A typical calculation is shown below where spiking level IS recorded as mg: mg/mL = [intercept x sample volume (mL) / slope] (7) This approach is demonstrated in Fig. 3 (6), where standard addition is
used to determine trace levels of fluoride. 2.5. Automated
Standard
Addition
In this approach (7), two injection steps are incorporated into the separation method as shown below: Rinse 1: Xmin Rinse 2: Y min Set detector: desired 1 InjectIon 1: x s from standard Injection 2: x s from sample
Separate:requiredmin at selectedvoltage To cover a calibration range, sampling times should be varied appropriately. For example, the time for injection 2 could be set at 5 s, and sampling times for injection 1 set at 0, 1,3, and 5 s. This approach is of use when quantifying a component in a complex mixture, as component identification is simultaneously confirmed. Fig. 4A shows separation of
Altria
54 1600. 1400
-
1200
-
lcaO800
-
600400
-
200
0,. 0.0
, 0 2
I‘
0.4
) 0.6
.
I 0.8
'
I 1.0
-
J 1.2
Fluoride Concentration Ftg. 3. Standard addition graph for fluoride deterrnmatton (from ref. 6). a 5-s injection of a solution of a pharmaceutical. Fig. 4B shows a 5-s injection of a solution of a pharmaceutical followed immediately by a 5-s mjection of a solution of a known impurity. It can be clearly seen which peak in Ftg. 4A is attributable to the specified impurity.
3. Impurity
Contents
Impurity levels may be quoted as % w/w (8) or % area/area (9,10) depending on the application. The impurity level may be calculated as % w/w if an appropriate standard of the impurity is available (8). More commonly, impurity results are calculated as % total area, since standards for all impurities may not be available.
3.1. Percent
(wlw) Impurity
Determination
Procedure
1 Prepare impurity caltbratton soluttons of approprtate concentratton to match likely levels present in samples. 2. Prepare sample soluttons rn the same dissolving solvent as the standards. 3. Analyze both samples and standards in an injection sequence. 4. Calculate impurity levels using response factors from calculatton soluttons. RF = [wt calibration (mg) x % purity / dilutton volume (mL) x 1001 (8)
Quantitation
3 c i5 m 88
Procedures
55
B.OO-
B.OoII 7.00-
pp
6 lO.OO-
B.OD-
I
8.at3-
6.0
Ill
II
1 8.00
7.00 Migrallon
time
8.00
(mmutes)
Fig. 4. Separation of salbutamol tmpurities (from ref. 7). (A) 5-s injection of salbutamol solution; (B) 5-s injection of salbutamol solution + 5-s injection of salbutamol tmpurlty solution.
Altria % w/w = [MRF x peak area x dllutlon volume x 100 / wt sample (mg)] (9)
This approach has been employed for the determination of the dimeric impurities of salbutamol(8). Fig. 4A shows the separation. The calculated results in % w/w were then directly compared to HPLC and TLC results. 3.2. Percent
Area/Area
Impurity
Procedure
This procedure 1smost commonly employed for ease of operation and simplicity (9). In addition, isolated standards of all impurities may not be available. This procedure infers that all impurities have similar response factors at the detection wavelength to the main component. If the impurity response factors differ significantly, appropriate adjustment of data prior to reporting is necessary 1. Analyze sample solutions and a blank at least m duphcate. 2. Integrate all sample related peaks, ignoring all peaks attributed to the blank. 3. Normalize each peak area to Its correspondmg mlgratlon time: Normalized peak area = (peak area / migration time) (10) 4. Add all the normalized peak areas to give a total normalized area for the electropherogram. 5. Calculate specific impurity levels as: % area = (normalized peak area x 100 / total normahzed areas) (11) 6. If the identity of a peak 1sunknown, it 1soften useful to calculate the relative migration time (RMT) as an identifier. RMT 1scalculated by* RMT = (migration time Impurity / migration time of main component) (12)
Impurity profiles for the duplicate injections of each sample solution should be compared. If the profiles do not differ significantly, impurity results for either injection or an average can be quoted. Typical reporting format may include: No. of lmpurltles Total % area lmpurltles Greatest impurity (% area) Identity of greatest impurity Second greatest impurity (% area)
Identity of secondgreatestimpurity Figure 5 shows the separation of related impurities in salicylamide by MECC (9), which were calculated as % area/areawith a detection limit of
Quantitation
Procedures
57
4 h
Ftg. 5. MECC separation of sahcylamide-related impurities (from ref. 9).
If the response factor for selected impurities differs significantly from the mean component (i.e., typically below 80% or above 120% of the main component response), adjustments should be performed. These corrected data give more representative impurity data. For example, If an impurity gives only a 40% response relative to the main component, it will be considerably underestimated. Conversely, an impurity with a
higher UV response will appear to be overestimated. Corrected % area = (% area x 100 / % relative response factor)
(13)
Data manipulation to generate % area data is usually performed employing validated software programs. 4. Enantiomeric Purity Determination When quantifying enantiomers (1 I), it is common to quote % enantiomerit purity. 1. 2. 3. 4.
Analyze the sample employing the chirally selective CE method. Integrate the peaks attributed to each enantiomer, and ignore all other peaks. Normalize the peak areas to their mtgration times. Calculate % enantiomeric purity by: % enantiomeric purity = (total normalized area- normalized areaen1 / (14) total normalized area) / 100
58
I 00
7 60
8 20
a a0
9.10
10.00
RETENTION
Fig. 6. Determmatton (from ref. 12).
TIME
10.60
11.10
11.80
11 to
4 11 00
(MINUTES)
of 0.1% of the undesrred enantromer
m a choral drug
Table 1 Effect of Peak Area Normalizatton on Peak Area Data for an Enanttomertc Separatrona Enanttomer 1 Peak area, observed Peak area, normahzed Mrgratton time, mm
1,199,781
103,827 11.55
Enantromer 2
Peak area ratio
I ,246,293 103,841 12 00
1.04 1 00
n Normallzatton of the dataconfirmsthe correctarearatlo for a racemlccompound where total area = sum of two enantiomers peak areas and area en1 = normalized area of undestred enantromer (or second migrating enanttomer if racemic).
If the compound is racemic, an enantiomeric purity of 50% would be expected. If 1% of the undesired enantiomer is present, then a percent enantiomeric purity of 99% would be obtained. Figure 6 shows the determination of a chiral drug containing 0.1% of the undesired enantiomer (12). Irrespective of which method is used, it is essential to use normalized peak areas (2). This is exemplified by the data obtained for the repeated analysis of a racemic drug (Table 1). A peak area ratio of 1.OOshould be
Quantitation
Procedures
59
obtained for a racemrc compound. The unnormalized areas indicate a ratio of 1.04, which suggests that more of the second detected enantiomer is present. However, the expected ratio of 1.OOis obtained when calculating using normalized areas (2). The racemic drug, clenbuterol, was chirally resolved by seven mdependent pharmaceutical companies in an crossvalidation exercise (13). The peak area ratio, when employing normalized peak areas, gave a value of 49.8:50.2 with an RSD of 0.6% on the results from the seven companies. 5. Determination of Small Ions by Indirect Detection This detection principle is widely employed in the determination of small ionic species, such as inorganic anions and metal ions (see Chapter 20 for further details). For example (‘I#), metal ions are analyzed using a low-pH electrolyte containing imidazole at 5 rnM. When monitoring at 214 nm, this provides a high UV background signal. The passage of the metal ions through the detector displaces the lmidazole molecules, resulting in a reduction of the background signal giving negatrve peaks. The output signal of the detector is reversed to give apparently positive peaks to simplify integration. Chapter 7 contains a section on the princrples of indirect UV detection. Quantitative analysrs using these methods (IS) generally employs external standards of Analar-grade inorganic materials, such as NaCI, to obtain response factors for the individual ions of interest. Use of appropriate internal standards has been shown to improve peak area precision (I 5). References 1. Altrta, K D (1993) Quantnative apphcattons of capillary electrophorests to the determination of pharmaceuticals and drug related Impunties. J Chromatogr 646, 245-257 2. Altrta, K. D. (1993) Essential peak area normalisation m captllary electrophorests Chromatographra 35, 177-182. 3. Ewmg, A G., Wallingford, R. A., and Olefirowlcz, J. (1989) Capillary electrophoresis. Anal Chem 61,292A-303A 4. Watztg, H and Dette, C (1993) Precise quantitative capillary electrophorests. Methodologtcal and mstrumental aspects. J. Chromatogr 636, 3 l-38 5. Altria, K. D. and Ftlbey, S. D. (1993) Quantrtattve analysts of sumatrrptan by capillary electrophoresis J, Lrquld Chromatogr. 16,228 l-2292. 6. Jackson, P E. and Haddad, P (1993) Optlmtsatton of inJectton techmque m captllary electrophorests for the determmation of trace levels of anions in envnonmental samples J Chromatogr 640,48 l-487
60
Al tria
7. Altria, K. D and Luscombe, D C. M (1993) Apphcatron of captllary electrophoresrs as a quantrtattve identity test for pharmaceuticals employing on-column standard additton. J Pharm Bzomed Analysis 11,415-420 8. Altrta, K D (1993) Quantttattve analysts of salbutamol related impurities by capillary electrophorests. J Chromatogr 634,323-328 9 Swartz, M E. (1991) Method development and selecttvny control for small molecule pharmaceutrcal separations by capillary electrophoresls J Lzquzd Chromatogr 14,923-938. 10. Korman, M., Vindevogel, J., and Sandra, P. (1993) Separation of codeme and its by-products by capillary zone electrophoresls as quahty control tool m the pharmaceuttcal mdustry J Chromatogr 645,3&S-370 11, Altrta, K D., Goodall, D. M , and Rogan, M M (1994) Quantttatrve appltcattons and validation of the resolutton of enantiomers by capillary electrophoresrs Electrophoreszs 15,824-827. 12 Werner, A., Nassauer, T., Krechle, P., and Erm, F (1994) Choral separation by capillary zone electrophorests of an optmally-acttve drug and ammo acids by hostguest complexatton with cyclodextrms J Chromatogr 666,37!&379. 13. Altrta, K D., Harden, R. C , Hart, M , Hevtzi, J , Halley, P A , Makwana, J V , and Portsmouth, M J (1993) An inter-company cross-validation exerctse on capillary electrophorests 1. Chrral analysis of clenbuterol J Chromatogr 641, 147-153. 14. Beck, W and Engelhardt, H. (1992) Capillary electrophoresls of orgamc and inorganic cations with mdu-ect UV detection Chromatographza 33,3 13-3 16 15 Altria, K D., Rogan, M. M , and Goodall, D M. (1994) Quantrtatrve determmatton of drug counter-ton stotchiometry by caprllary electrophoresis Chromatographza 38,637-642.
CHAPTER6
Optimization of Precision in Quantitative Analysis Kevin
D. Altria
1. Introduction Early reports of quantitative analysis by CE on homemade systems indicated that peak area precisions of -5% relative standard deviations (RSD) could be expected. The advent of reliable commercial mstruments and improved methodology can now enable RSD figures of l-2% or better to be routinely obtained (1). The following chapter provides some guidelines toward optimizing precision. The major source of imprecision remaining when using commercial instrumentation is injection volume variability. Reproducible nanoliter sample volumes present a formidable engmeering challenge. Use of an internal standard will improve precision (2). The method should be well controlled in terms of consistent migration times (3), since peak area is related to both sample concentration and migration time. If there is a slight drift in migration times, peak area normalization (4) can be conducted to improve peak area precision (5). Injection precision (1) is also highly related to sample concentration, being worse at low sample concentrations (Fig. 1). This is because of increased variance contributions from such factors as integration errors and solute adsorption onto the capillary surface. It is therefore suggested that whenever possible, a high sample concentration and injection volume should be employed. This has the effect of reducing separation efficiencies, but improving precision. This use of high concentrations also allows the simultaneous determination of related impurities. From
Methods Edlted by
m Molecular Bology, Vol 52 CapHary Electrophoresrs K Altrla Copyrtght Humana Press Inc , Totowa, NJ
61
AZtria
62
* *
* * I
500
”
‘1
b
1000
”
”
mg/l
!
+
1500
Frg. 1. Varlatron of precrsronwith sampleconcentratron(from ref Z). 2. High Sample Concentration This is the most important consideration, since several studies (I, 67) have shown a correlation of improved peak area precision with increased sample concentration. This is attributed to a reduction in integrator errors and minimizes the effects of any solute adsorption that may occur onto the capillary wall. Sample solubility, peak shape, and resolution requirements must all be considered when optimizing the sample concentration. 3. Temperature Control Temperature affects both viscosity and electrophoretic mobility (8). Therefore, it is important to maintain a constant temperature (9) throughout a seriesof injections. This need for temperature control applies to the capillary, sample vials, and electrolyte reservoirs. Viscosity decreases with increased temperature. Therefore, when temperature increases, more sample solution enters the capillary for the same injection time. Electrophoretic mobility increases with temperature, which reduces migration times. It is important to allow the sample solutions to attain a constant temperature on the CE autosampler, since the volume injected is related to the sample solution viscosity (IO). Since in CE the area of peak is related to both sample concentration and migration time, temperature differences
Precision in Quantitative
63
Analysis
during an analytical sequence will cause shifts in migration times and apparent peak areas, leading to poorer precision. 4. Injection
Times
Using hydrodynamic injection, a pressure difference is generated across the capillary for a set period of time (l-20 s). The pressure dlfference is instrument-specific, and the exact pressure level is monitored durmg injection. Most commercial instruments are self-regulating in that, if the pressure difference 1s below the set level, the injectlon time is extended to compensate. Conversely, if the pressure difference 1sabove the set limit, the injection time is reduced accordingly. In all instances, if the monitored pressure difference is too far from the prescribed level, no injection would be made. If short injection times, such as 1 or 2 s, are employed, the ability to self-regulate injection is reduced. Therefore, it is advisable to employ longer injection times, such as 5-10 s. To achieve this, it may be necessary to decrease sample concentration. When using a positive pressure injection system, liquid is displaced from the capillary during injection. This liquid appears as drops at the detector end of the capillary. If the drop falls durmg injection, the pressure difference across the capillary is altered, causing imprecision. To eliminate this possibility, it is advisable to perform the injection step with the detector end of the capillary immersed in a vial containing separation electrolyte. 5. Capillary
Rinsing
It has been shown (I, II) that rinsing the capillary between injections can improve (12) migration time precision to give migration time precisions of
64
Al tria Table 1 Preclslon Data Obtamed for Different InJectIon Times (from ref 13)
Sample K+Na K+Na K+Na
Injection time, s
No of mjectlons
RSD %, K
RSD %, PAR
1 2 2
9 9 45
36 11 2.8
11 03 02
6. Internal Standard Much of the variance in precision is attributable to injection volume fluctuations. Therefore, incorporation of an appropriate internal standard(2) will mnumize this source of error. It is important to match appropriately sample and internal standard concentrations.It is necessaryto ensure that the internal standard does not corn&ate with any sample-related compounds. The purity and stability of the internal standard should be selected, such that no internal standard-related impurities interfere with the sample component peaks. Use of an internal standard also partially compensates for variations in area owing to interrun migration time variation. The advantages of employing an internal standard are highlighted by data from the validation (13) of an assay method for potassium using sodium as an internal standard (Table 1). The precision of injection for the potassium peak alone ranged from l . l-3.6% RSD. However, the precision for peak arearatios (PAR) was improved and ranged from 0.2-l. 1% RSD. 7. Electrolyte Considerations Peak tailing can lead to poor precision because of variable integration. Peak tailing can be reduced by increasing the electrolyte concentration. Excessive electrolyte concentration can, however, lead to a generation of unacceptably high levels of current, which lead to poor precision of migration time. Therefore, rt is necessary to optimize electrolyte in terms of composition and concentration. When performing multiple analysis, electrolysis of the electrolyte in the buffer reservoirs can occur (14,15), resulting in a pH gradient being formed across the capillary (this effect has been termed buffer depletion). These changes may affect both migration times and selectivity. Figure 2 shows the first and fourth injection (14) using a low-capacity buffer. It should be noted that the migration time and selectivity vary, A stable method was obtained by selection of a more appropriate buffer.
Precision in Quantitative
Analysis
65
B I V
0
I
I
5
10
j. 1s 0
I I
I
5
10
15
ThWmin
Fig. 2. First and fourth nqectlons of an atenolol test mixture usmg a lowcapacity buffer.
To minimize buffer depletion, large volumes of electrolyte (5 mL or greater) should be employed. Higher ionic strength reduces the significance of this depletion. It may be appropriate (‘II) to specify use of an electrolyte vial for only a selected number of analyses. After this number of injections, the analysis is performed employmg vials containing fresh electrolyte. It may be necessary, for an extremely long sequence, to utilize several changes of electrolyte. Switches of buffer vials can be preprogrammed into the PC controlling the CE instrument. 8. Preconditioning Preconditioning of the system in certain methods (I, I I, I@, particularly high-pH free solution capillary electrophoresis (FSCE) and micellar electrokinetic capillary chromatography (MECC), can lead to improved precision. It is our experience that the first injection in a sequence can produce significantly different peak area and selectivity to subsequent analyses. This may be the result of an equilibration of the capillary surface and a stabilization of electrolyte composition under electrophoretic conditions. It may therefore be appropriate to include a precondition analysis in the sequence. This precondition may simply
Altria comprise analysis of sample dissolving solvent. Alternatively, the capillary could be filled with the electrolyte, and the appropriate voltage applied for 20-30 min. Incorporation of these preconditioning steps also allows the solutions on the autosampler to reach a constant temperature. Use of temperaturecontrollable autosampler tray may assist in this aspect. 9. Sample Consideration The following points should be considered when performing quantltative analysis: 1. Match lomc strength and vlscoslty of samples and standards.Viscosity influences the volume Injected mto the system (8) Therefore, It 1sfimdamental to match the vlscosltles of both standards and samples Samples, such as bioflmds, will have very dtfferent vlscositles compared to standard solutions. If this 1s lmposslble, standard addition calrbratlon procedures could be performed. 2. Evaporation losses should be mmlmlzed by employmg sealed vials and aqueous-based sample solvents with the mmlmal level of volatile organic solvents possible. Often, small sample volumes of Cl mL are employed using mlcrovial sample contamers. Evaporation losses from these mlcrovials can be very significant (Z 7) over the course of a few hours. The losses may be reduced by using a cooled autosampler tray and ensurmg that the mlcrovlals are stored m humid condltlons (this can be achieved by inserting the microvials into a large vial containing a few drops of water m the bottom). Use of an internal standard alleviates any potential difficulties
owing to evaporation. 10. Capillary Washing Poor precision may result (Z6) if sample solution is transferred onto the outside of the capillary during the injection process. When the separation voltage is applied, some of this transferred material may migrate into the capillary, giving a nonreproducible higher sample loading. To prevent this effect from occurring, an additional step in the method may be incorporated: Step 1: Rinse. Step 2: Rinse electrolyte. Step 3. Inject sample. Step 4: Inject electrolyte or water (1 s). Step 5: Separate.
Precision in Quantitative
Analysis
67
The purpose of step 4 is to wash the capillary exterior to avoid any crosstransfer of sample solution. 11. Capillary Cutting If the capillary end is jagged, sample solution may be irreprodwcibly transferred from the sample solution. It is therefore important to break the capillary using a cutting stone or a diamond-tipped cutter. If this source of imprecision 1s suspected, examine the capillary tip using an appropriate eyepiece. Replace the capillary if necessary. 12. Solution Height Considerations Dipping the capillary into the solution (18,19) causes a small amount of sample solution to enter by capillary action. The amount of this “inadvertent injection” is related to the internal diameter of the capillary and the height of the sample solution in the vial. Siphoning of sample solution into the capillary can occur if the height of the sample solution vial is higher than the level in the detector end electrolyte vial. This would give a higher sample loading than if the levels were identical. Conversely, a lower sample solution level would result in a siphoning flow in the opposite direction, giving a lower on-column sample loading. Unless controlled, this can be a further source of imprecision. This problem will be increased moving to wider-bore capillaries. To reduce this effect, it may be necessary (II) to pipet a set volume of sample solution into each sample vial to ensure a constant solution level. A siphoning flow can occur if a height exists between the solution levels in the high-voltage and low-voltage electrolyte vials. For example, the siphoning flow will be toward the detector if the level is higher in the high-voltage electrolyte vial. In this instance, the migration time will be reduced as the siphoning flow will be superimposed on the mobilities and EOF governing the separation. The magnitude of this flow is such that a 2-mm height difference may have a 2% change in migration time for a 50-pm capillary, whereas a 10% difference may occur with 100-pm capillaries. Repeated rinses from an electrolyte vial will cause a gradual decreasein solution height level, resulting in an increased siphoning flow against the direction of migration. This will be observed as a drift in migration times and corresponding peak areas.This effect can be avoided by performing rinses from a separate vial from those used for separation purposes.
Al tria
68 Table 2 Method Settmgs for Optlmal Preclslona Cycle Set temperature Rinse Injection Injection Separate
Settmg Constant temperature Typically 2 mm Typically 5 s IS
As appropriate
Vlal number, at samphng end
Vlal number, at detector end
1 3 1 2
5 4 4 4
“See Rg 3 for vial numbers From ref 21
13. Voltage
Ramp When the separation voltage is initially applied across the capillary containing the sample solution, much of the load is applied across the sample zone, which usually has a higher electrical resistance. This situation leads to accelerated heating of the sample zone and thermal expansion. This expansion can in turn lead to a small, nonreproducible amount of the sample to be forced out of the capillary at the onset of the separation (20), To minimize this effect, it is recommended to ramp up the voltage gradually to the final level over the course of 20-30 s. This facility is available with many commercial systems. In addition, this effect may be minimized (16) by incorporating a water or electrolyte slug after the sample zone. Therefore, if some material is lost, it will be unimportant, since it will only be water or buffer. 14. Speed of Sampling It has been shown (19) that delays between sample introduction and reinsertion of the capillary tip into the electrolyte vial can cause a loss of precision. Therefore, the timing of this procedure should be well controlled. This feature is built into fully automated CE systems. Use of an internal standard will eliminate this problem. 15. Method Settings for Optimal Precision on Automated Equipment The best settings for a CE method to obtain maximum precision (21) is given in Table 2. The vial numbers refer to those shown in Fig. 3.
Precision in Quantitative
69
Analysis High voltage supply
Buffer vlal
Buffer vial
3 @2 as Sample 0
vial
Fig. 3. Optlmlzed
Empty vial
4
*
0
5
system setup (from ref. 21).
References 1 Watzlg, H and Dette, C (1993) Precise quantltatlve capillary electrophoresis Methodological and instrumental aspects J Chromatogr 636,3 l-38 2. Dose, E. V and Gulochon, G A (1991) Internal standardlsatlon technique for capillary zone electrophoresls. Anal Chem 63, 1154-l 158. 3. Ewing, A G , Walhngford, R. A, and Olefirowicz, T. M (1989) Capillary electrophoresis. Anal Chem 61,292A-303A. 4. Altria, K D. (1993) Essential peak area normalisation in capillary electrophoresrs Chromatographla 35, 177-l 82 5 Altria, K. D., Harden, R. C., Hart, M., Hevizl, J., Hailey, P A., Makwana, J. V., and Portsmouth, M. J. (1993) An inter-company cross-validation exercise on capillary electrophoresls 1 Choral analysis of clenbuterol J Chromatogr 641, 147-153 6. Ryder, S. (1992) Determination of sodium vmyl sulphonate in water-soluble polymers usmg capillary zone electrophoresls. J Chromatogr. 605, 143-147. 7. Altria, K. D , Rogan, M. M., and Goodall, D M. (1994) Quantitative determmatlon of drug counter-ion stoichlometty by capillary electrophoresls Chromatographza 38,637-x542. 8. Rose, D. J and Jorgenson, J. W. (1988) Characterlsation and automation of sample
introduction methods for capillary electrophoresls Anal Chem 60,642--648 9. Goodall, D. M , Wllhams, S. J., and Lloyd, D. K. (1991) Quantttattve aspects of capillary electrophoresis TrAC 103, 272-279. 10. Rush, R. S. and Karger, B. L. (1990) Sample inJection with P/ACE system 2000. Importance of temperature control with respect to quantltation. Beckman Techmcal Bulletzn TIBC 104, I,2
Al tria 11 Thomas, B R., Fang, X G , Chen, X , Tyrell, R J , and Ghodbane, S (1994) Vahdated mlcellar electrokmetrc caprllary chromatography method for the quality control of the drug substances hydrochlorothtazide and chlorothrazrde. J Chromatogr 657,383-394. 12 Coufal, P , Stuhk, K , Claessens, H A., and Cramers, C A (1994) The magnitude and reproductbihty of the electroosmottc flow m sthca captllary tubes. JHRCC 17, 325-334 13 Altrta, K. D , Wood, T , Kttscha, R , and Roberts-McIntosh, A (1995) Valtdatlon of a capillary electrophorests method for the determmatton of potassmm counterton levels in an acrdrc drug salt. J Pharm Bzomed. Analyszs 13, 33-38 14 Shafaatr, A and Clark, B J (1993) Development of a caprllary zone electrophoresis method for atenolol and its related tmpurtttes m a tablet preparatton Anal Proc 30,48 1483. 15 Sun, Y -L , Zhang, C -X , Ling, D -K , and Sun, Z -P (1994) Vartatton of pH of the background electrolyte as a result of electrolysts in CE JHRCC 17,563,564. 16 Altria, K D , Clayton, N G , Harden, R C , Hart, M , Hevtzt, J , Makwana, J V., and Portsmouth, M J (1994) An inter-company cross-vahdatton exercise on captllary electrophorests testing of dose uniformity of paracetamol content m formulations Chromatographza 39, 180-l 84 17 Mormg, S. E., Colbum, J. C , Grossman, P. D , and Lauer, H H (1990) Analyttcal aspects of an automated capillary electrophorests system. LC-GC 8,34-40 18 Dose, E V. and Gurochon, G A (1992) Problems of quantitative inJection in capillary zone electrophoresis Anal Chem 63, 123-128 19. Frshman, H. A., Amudi, N M., Lee, T. T , Scheller, R H , and Zare, R N. (1994) Spontaneous mJectton m microcolumn separattons Anal Chem 66,23 18-2329 20. Knox, J. H and McCormack, K A (1994) Volume expansion and loss of sample due to mittal self-heating in capillary electrophoresis (CES) systems
Chromatogruphza 38,279-282 2 1. Altria, K D. and Fabre, H (1995) Approaches to optimrsation of precision m capillary electrophorests. Chromatographza 40, 3 13-320
CHAPTER7 Optimization Kevin
of Sensitivity D. Altria
1, Introduction The injection volumes involved in CE are very small (in the order of l-10 nL). In addition, the area of capillary employed for on-column detection may be only 50 x 200 pm. Both these factors influence the detector sensitivity to a large extent. CE is less sensitrve when directly compared to HPLC with typical injections of 10-50 pL and l-cm detector cells. The difference may be up to an order of magnitude (2) when comparing at the same UV wavelength. Several strategies may be employed (Table 1) to maximize the sensitivity in CE. These include use of low UV wavelengths, increased capillary bore, and optimized sampling procedure. 2. Detection Wavelength Many solutes have considerably enhanced UV activity at low UV wavelengths (i.e., 190-220 nm). Excessive background UV absorbance from the organic solvents employed in HPLC generally prevents operation at these wavelengths. However, since the electrolytes employed in CE are predominantly aqueous-based,the background absorbance is not excessive and methods can be routinely operated at low wavelengths. Adopting this approach can allow direct detection of solutes that may have required derivatization prior to HPLC. Examples include the direct detection of sugars (2) and amino acids employing detection at 195-200 nm. Equivalent concentration sensitivity for salbutamol and related impurities was obtained (3) for CE and HPLC employing detection wavelengths of 200 and 276 nm, respectively. From
Methods m Molecular Ebology, Vo/ 52 Capillary Electrophoresrs Edtted
by K Altna
Copynght
Humana
71
Press
Inc , Totowa,
NJ
Al tria
72 Table 1 Summary of Approaches Available for Increasmg Sensitivity Action to Improved sensmvtty Employ low-UV
wavelength
Increase capillary bore Increased mlection time Approprtate use of electrokmettc mlection Increased electrolyte strength Optimize electrolyte composmon Decrease operating voltage Decrease temperature Captllary modrficattons Sample derrvattzation Indirect detection Wide-bore capillaries
Increased detector slit wtdth
Drawback(s)
to consider
Increased background notse-detennme wavelength to gave optimum stgnal-to-noise ratio Increased current, reduced EOF givmg possible alteration m selecttvtty Reductton m separatton effictency; excesstve ttme will result m run failure Samphng bias for more mobtle tons, sample matrix effects on qectton amount Increased current and associated notse Effects on selectivity and current Increase m analysis trme Increase m analysis time Reduction m separation efficiency and resolution, Increased cost Addmonal samplmg handhng Extra method development considerations EOF profile disturbed, adjustments to rmse and mjection times, stphonmg effects more pronounced Reduction m separatton effictency and resolution
3. Capillary Bore Increasing the capillary bore improves detector sensitivity considerably. This gain is caused primarily by an increase in the volume of caplllary employed as the on-column detection area and a similar increase in the volume of sample solution introduced into the capillary. Typically, 50- and 75-pm capillaries are routinely employed, since problems of reduced separation efficiency and increasing current occur moving to wider-bore capillaries. Band broadening resulting from nonideal EOF increases with capillary bore. Levels of EOF also decrease significantly with increased capillary bore, which will influence separation selectivity. The ability to dissipate heat is also reduced with increased capillary bore as the effective surface area is reduced. This increased heat can lead to reduced separation efficiency, and if too excessive, can cause outgassing or boiling of the electrolyte. The sample volume introduced hydrodynamically into the capillary is related to the pressure difference across the capillary. Wider-bore capillaries have considerably
73
Sensitivity
Fig. I. Separationof drug-relatedimpurities using a 180~pmcapillary (from ref. 4). reduced back pressures, and rinse regimes require careful adjustment (see Section 12). Capillaries with bores as wide as 180 pm (4) have been successfully used to enhance sensitivity greatly. Figure 1 shows the detection of impurities in a drug substance using a 180~pm capillary. The impurities at the 0.1% level are almost off the full-scale deflection of detector. However, attention should be given to optimizing injection time in conjunction with capillary bore (see Section 12). 4. Injection
Time
The injection time employed in hydrodynamic sampling determines the volume of sample introduced into the capillary. The increase in peak area is linear with injection time. Peak height increases are not necessarily in as linear a fashion. Separation efficiency is reduced with increased injection time, which will have implications on resolution achieved. Excessive sampling times can lead to a breakdown in the conductance along the capillary, causing a run failure. Therefore, injection times can only be increased to a limit that would be determined in method optimization. This breakdown in conductance is most pronounced when
74
Al tria
samples are dissolved m organic solvents, since these are nonconductive. Samples prepared in water, although giving good separation efficiencies for sample components, have low conductivity. Therefore, to extend sampling times, it may be appropriate, for example, to dilute the sample with a low-tome-strength solution (5). Typically, samples may be dissolved in a solution of electrolyte employed for separation, diluted 1 to 10 with distilled water. The sample-dissolving solvent should be optimtzed during method development. 5. Injection Procedure Several means of optimizing sampling to increase sensitivity have been reported. These mclude the use of electrokinetic mjection and multiple electrolyte systems. Chapter 16 discusses some advanced means of Increasing sensitivity. In electrokinetic injection, sample ions enter the capillary by virtue of their mobility and any EOF present. A bias exists with this injection procedure in that a greater number of more mobile ions enter than less mobile ones. When attempting to maximize sensitivity, this effect can be advantageous. For example, to determine trace levels of small inorganic cations, application of a voltage of +5 kV for 10 s will result in a sensitivity improvement of 2-3 orders of magnitude (6) compared to pressure injection. Using this approach, it is possible to determine low-ppb levels of metal cations with indirect detection. The limit of detection for these cations using hydrodynamic injection with similar detection is -1 ppm. To optimize injection of anions, it is necessary to suppress EOF or to alter the direction of EOF. Anions are electrokmetically injected employing a negative voltage, and similar gains in sensitivrty to that described for cations can be achieved. Electrokinetic injection does not give significant sensitivity increases for less mobile tons and is less effective for neutral species, since these enter the capillary only by virtue of the generation of any EOF. 6. Electrolyte Composition The ionic strength and pH of the electrolyte should be optimized to reduce peak tailing. Higher ionic strengths reduce tailing and give improved separation efficiencies. If possible, the srgnal-to-noise ratio should be obtained for electrolytes under consideration, since the background absorbance will vary between electrolytes. For example, when
Sensitivity operating at low UV detection wavelengths, such as 200 nm, borate and phosphate buffers are especially useful, because their background UV absorbances are low compared to electrolytes, such as citrate or acetate. The major source of zone dispersion is molecular diffusion. Diffusion is related to sample molecular weight, electrolyte viscosity, and time. The electrolyte viscostty can be increased to maintain peak sharpnessby addition of agents, such as cellulose, PVA, or glycerol. These additions do, however, suppress EOF, which is also viscosity-dependent 7. Operating Voltage Molecular diffusion is related to analysis time. Therefore, it is useful to operate at higher applied voltages. This can be achieved by either mcreasing the voltage or reducmg the capillary length. Limitations concerning current and electrolyte ionic strength will need to be considered. 8. Temperature Thermal vibration of the capillary can occur at increased temperatures. This vibration will be shown by an increase in detector noise. Operating at high current levels can induce a similar effect owing to internal heatmg within the capillary and associated refractive index changes 9. Capillary
Modifications Several approaches have been described for extending the effective pathlength of the area of capillary employed for detection. These have included use of a Z-shaped portion of capillary (7) and creation of a bubble within the capillary at the detection window. Bubble cells are commercially available from Hewlett Packard (Palo Alto, CA), whereas Z-cells are available from Perkin Elmer (ABI, Foster City, CA) or LC Packings (Zurich, Switzerland). Figure 2 shows the possible sensitivity gain when employing a Z-cell capillary. Rectangular capillaries have also been utilized (8). However, these are not yet commercially available. Capillaries are also available with a small portion of reversed-phase packing material inserted in the injection end of the capillary (9). Extended electrokinetic injection of the sample causes sample ions to migrate into the capillary and become adsorbed onto the packing. After this injection a short, l-s injection of solvent, such as methanol, removes the material from the packing. Using this approach, electrokinetic injection times of several minutes can be employed, and good
76
Al tria
I
2
HS OPTICAL CELL
3 3 f-,
0
;-~~
2
4
4
6 TIME
0.18%
8
10
12
(min)
Fig. 2. Sensitlwty gains possible using Z-cell capillary (from ref. 22).
peak shape and efficiency were still maintained. Detection limits of up to a IOOO-fold over conventional hydrodynamic injection were shown. 10. Derivatization Sensitivity can also be improved by appropriate derivatization of the sample. The derivative may have a considerably higher UV absorbance or may be selected to allow fluorescence detection. The majority of reports (IO) have centered on preseparation denvatlzation. This obviously involves additional sample handling steps. Appropriate derivatization can enable use of laser-induced fluorescence monitoring, which may be up to 1000 times more sensitive than UV absorbance. Some recent reports have indicated use of on-column derivatization (II, 12) in which the sample and derivatizatlon chemicals are injected
Sensitivity
77
FS
0 1 1. Empty capillary
FS A
FS A
0
0 2. Filled caplllary
3. Separation
Fig. 3. Prmclples of mdlrect detection.
onto the capillary and allowed to react prior to separation. This approach is particularly attractive if the derivative has poor stability. On-column chelation is another useful option to improve detection. For example, UV-transparent metal ions are complexed on-column with EDTA to produce strongly UV absorbance anionic chelates, which are well resolved (13) with good detection limits. 11. Indirect Detection This method of detection is widely employed in CE, since simple inorganic or small organic ions are widely analyzed. These species have only limited chromophores and cannot be readily detected by conventional UV absorbance detection. Indirect detection involves the addition of strongly UV active species into the electrolyte to give a constant high UV signal in the detector. This detector signal decreases when solutes have a limited chromophore pass through the signal, resulting in the generation of negative peaks (Fig. 3). Commercial instruments have the ability to reverse detector output such that peaks are easily integrated. The choice and concentration of the UV-absorbing species have a large impact on the performance and sensitivity of the method (14-18). The concentration of the background absorber primarily governs the sensitivity obtainable and the linear dynamic range possible. If the concentration is too high, the UV responsechange for a lower concentration sample will be insufficient to be detectable. Conversely, if the concentration is too low, the linearity will be limited to an unacceptable degree. Therefore, compromise concentrations of l-10 mA4 are typical. The mobility of the absorber in the run buffer should approximately match that of the sample constituents. If the mobility of the solute exceeds that of the UV-
78
Altria 2 69-l
1 2 00
1 3 00
I 2 50
/ 3 50
mlnutas
Fig. 4. Separationof a rangeof metal cations usmg indirect UV detectron (from ref. 18). absorbing ion, then a fronting peak is obtained (K+, Fig. 4). If the mobilities of the solute ion and UV-absorbing ion are exactly matched, then a sharp, Gaussian peak will be obtained @a+, Fig. 4). If the sample is slower, then an apparently tailing peak will be obtained (Li+, Fig. 4). In this separation, the mobility of the background absorber imidazole 1s0.44 cm2/kV/s, which exactly matches the mobility of Ba+. If an exact mobility match IS not immediately available (19), it may be appropriate to alter the mobility of the UV absorber to match the mobility of the sample more closely. For example, P-cyclodextrin was added to match the mobility of 2,6 napthelenedicarboxylate to agree more closely to that of selected aliphatic acids. Indirect detection is extensively employed in the detection of small inorganic and organic ions. Reference to these sections in Chapter 20 will cover the range of possible UV absorbers and applications more comprehensively. Generally, the buffer system should be optimized to give a maxrmum background absorbance of 0.1 aufs (17). The electrolyte composition should be considered to produce a minimal amount of internal heating, smce this 1sexhibited as increased baseline noise owing to thermal vibra-
Sensitivity tion of the capillary. Therefore, methods are typically operated at conditions that produce 5-10 PA. Optimal sensitivity m indirect detection is also obtained at a wavelength where there is a maximum difference between the absorbance of the solute and the UV absorber. Therefore, detection wavelength should be optimized to give the greatest signal-to-noise ratio.
12. Use of Wide-Bore
Capillaries
This is an effective means of increasing sensitivity, smce the detection pathlength is the size of the capillary diameter. Although attractive (Section 3.), several complications must be considered when increasing capillary bore. 1. EOF IS strongly related to capillary bore. A change m capillary bore may, therefore, have sigmficant impact on separation selectlvlty, especially at high pH. 2 A doubling of capillary bore increases the current level obtamed for a set applied voltage by a factor of four. This may result m excessive current or may require operation at a reduced voltage level. 3. Flow rate through the caprllary 1srelated to the fourth power of the dlameter, because the back pressure along the capillary 1s greatly reduced by mcreasing capillary bore. Therefore, sampling and mJectlon times must be adjusted accordingly when capillary bore 1sincreased 4. Moving to capillary bores of 100 pm or greater can cause serious distortions of the separation achieved usmg conventional bore capillaries. The EOF profile becomes more laminar, which causesband broadening. Height differences between electrolyte levels in the buffer vials can result m a siphoning flow. Siphoning flow has a lammar profile, but 1sminimal when employmg conventional bore capillaries. Siphoning can become a slgmficant factor usmg wide-bore capillaries, and careful attention should be paid to electrolyte levels in the vial. 5. When employmg wide-bore caplllanes, it may be beneficial to eliminate EOF by using coated capillaries or by the addition of polymeric additives to the electrolyte. Polymeric additives, such as hydropropyl methyl cellulose or polyvinyl alcohol, are water-soluble, and small additions (< 1% w/w) significantly increase electrolyte vlscoslty (20). Increased viscosity 1sbeneficial for several reasons; EOF flow decreases with viscosity, siphoning flow rate IS reduced with viscosity, and rinse and injection times may be increased as the increased electrolyte viscosity increases the back pressure across the capillary. The disadvantage of these approaches is that only anions or cations may be quantified within a single analysis and that MECC cannot be performed.
80
Altria
Despite these apparent drawbacks, tt is posstble to operate routmely with IOO-pm capillaries (21). 13. Detector
Slit Width
Viewing a longer portton of the capillary will also lead to gains in sensitivity. For example, changing from a 200~pm slit width to 800~pm will generally double the signal-to-noise for a given peak. The expected fourfold increase is not achieved, since the noise is also increased to a lesser extent. Since each peak IS monitored by the detector for a longer time, observed separation efficrenctes will decrease and resolutton of closely migrating peaks will be reduced. References 1 Altrra, K D (1993) Sensmvrty optrmlsation for use of caprllary electrophoresrs m pharmaceutical analysrs LC-GC Znt 11,438-442 2 Hoffstetter-Kuhn, S , Paulus, A , Gassman, E , and Wrdmer, H M (1993) Influence of borate complexatron on the electrophoretrc behavlour of carbohydrates m caprllary electrophoresls Anal Chem 63, 154 l-l 547 3. Altrra, K D. (1993) Quantrtatlve analysis of salbutamol related rmpurrties by capillary electrophoresls J Chromatogr 634, 323-328 4 Altrra, K. D (1993) Capillary electrophoresrs for pharmaceutrcal research and development LC-GC Int 6,6 16-620 5. Chlen, R. L and Burghl, D S (1992) On-column sample concentration using field amplificatton m CZE. Anal Chem 64,489A-496A 6 Jackson, P E and Haddad, P. (1993) Optrmrsatron of mJectlon technique m capillary electrophoresrs for the determmatron of trace levels of anions m envnonmental samples J Chromatogr 640,48 l-487 7. Chervet, J P., Van Soest, R. E J., and Ursem, M (1991) Z-shaped flow cell for UV detectton m caprllary electrophorests. J Chromatogr 543,43!+-449 8 Tsuda, T., Sweedler, J. V , and Zare, R. N (1990) Rectangular caplllarles for capillary zone electrophorests. Anal Chem 62,2 149-2 152 9 Swartz, M. E. and Merron, M (1993) On-lme sample preconcentration on a packedinlet caprllary for lmprovmg the sensrttvrty of caprllary electrophoretlc analysrs of pharmaceutmals J Chromatogr 632,209-2 13 10. Amankwa, L. N , Albmk, M , and Kuhr, W. G (1992) Fluorescence detectron m caprllary electrophoresls TRAC 11, 114-120. 11 Ruyers, H. and Van der Wal, S J. (1994) Fully automated analysts of ammo acid enantlomers by derlvatisatron and chiral separatron on a capillary electrophoresrs instrument J Lzquzd Chromatogr 17, 1883-l 897. 12 Reinhold, N J , TJaden, U. R , and Vand der Greef, J (1994) Automated on-caprllary isotachophoretrc reaction cell for fluorescence derrvatrsatron of small sample volumes at low concentratrons followed by capillary zone electrophoresrs J Chromatogr 673,255-266
Sensitivity
81
13 Motomrzu, S , Oshrma, M., Matsuda, S., Obata, Y , and Tanaka, H (1992) Separatton and determmatron of alkalme-earth metal Ions as UV absorbing chelates wrth EDTA by capillary electrophoresrs Determination of calcmm and magnesium in water and serum samples. Anal Scz. 8,6 19-624. 14 Wang, T. and Hartwtck, R. A (1992) Norse and detectron hmrts of indirect absorption detectton m capillary zone electrophoresrs J Chromatogr 607, 119-125. 15 Foret, F , Fanali, S , Ossrcmr, L , and Bocek, P (1989) Hindu-ectphotometrrc detection m capillary zone electrophorests. J Chromatogr 470,299-308. 16. Vorndan, A G., Oefner, P J , Scherz, H., and Bonn, K (1992) Indirect UV detectton of carbohydrates in caprllary zone electrophoresis. Chromatographla 33, 163-168. 17 Nrelen, M W. F. (1991) Quantttative aspects of mdtrect UV detection m caprllary zone electrophoresrs. J Chromatogr 588, 321-326 18 Beck, W. and Engelhardt, H (1992) Capillary electrophorests of orgamc and morgame cations wrth indirect UV detection Chromatographza 33, 3 13-3 16. 19 Tindall, G. W , Wilder, D R., and Perry, R L (1993) Optimrsmg dynamrc range for the analysis of small ions by captllary zone electrophoresrs J Chromatogr 641,163-167
20 Belder, D and Schomburg, G (1992) Enantromer separatron of tocannde analogues by cyclodextrm modified electrokmettc chromatography JHRCC 15,686-693 2 1 Thomas, B R , Fang, X. G., Chen, X , Tyrell, R. J , and Ghodbane, S. (1994) Vahdated mtcellar electrokmetrc capillary chromatography method for the quahty control of the drug substances hydrochlorothraztde and chlorothrazrde. J Chromatogr 657,383-394. 22 Mormg, S. E , Parraud, C , Albm, M , Locke, S , Thrbault, P , and Tmdall,
G W (1993) Enhancement of UV detectron sensitrvrty for captllary electrophoresis Am Lab July, 22
CHAPTER8
Method Kevin
Validation D. Altria
1. Introduction The criteria for validation applied to CE methods (1-3) are similar to those employed for other quantitative separative techniques, such as HPLC (4). The extent of tests applied will vary according to the nature of the application and the extent of intended usage. Generally, such aspects as accuracy, precision, reproducibility, linearity, robustness, and method of transfer are evaluated during validation. Suggestedmethod validation approacheswill be given for the three key application areas of purity testing, main peak assay, and chiral analysis. All tests do not necessarily need to be performed for each method, and those most appropriate should be conducted and recorded by the analyst. Table 1 gives a suggested range of testing. 2. Selectivity 2.1. Purity Determination Appropriate selectivity of the method should be demonstrated for all known likely synthetic or degradative impurities (1,&d). This would be performed by analyzing test solutions spiked with all available impurities at expected levels. If appropriate, the method may be further challenged by analyzing stressed samples. The stressed samples may be generated by exposure to high temperatures, pH extremes, and both natural and UV light. The repeatability of selectivity should be assessed by repeated injections of a suitable sample or test mixture (.5), since electrolysis (depletion) of the buffer can occur leading to changes in selectivity. Selectivity should be demonstrated at the levFrom
Methods m Mokcular Biology, Vol. 52 Caprllary Electrophoresa Edited by K Altna Copynght Humana Press Inc , Totowa, NJ
83
84
Altria Table
1
Vahdatlon Requwements Test
Identity confirmation
Choral and achlral purity
Main component assay
J J J
J J J J J J J J J
J J J J J J J J J
Accuracy Precwon Linearity Sensitwlty Selectwty LOD
LOQ
Solution stablhty Robustness
J J
domperidoneR33812
I
I
a
0
Fig. 1. Separation
I
I
u
I
I
I
II
IS
4
5
I
I
I
I
17
Time (min)
of all major impurities
I
of dompertdone
in a batch (A) and
a sample solution spiked with 0.1 (B) and 1.O%(C) of each lmpurtty (reproducedfrom ref. 20). els required for quantitation. Figure 1 shows separation of all major impurities of domperidone in a batch and a sample solution spiked with 0.1 and 1.O% of each impurity.
Method Validation
10 00
85
12 00
14 a0
16 00
Mlgratlon
8.00
time (mm )
2000
2200
2400
Fig. 2. Separation of a pure (+) enantiomer of fluparoxan spiked with I % of the (-) enantlomer to confirm migration order (reproduced from ref lo).
2.2. Main Component Assay Selectivity for likely interferents should be demonstrated in addition to the likely sample-related impurities. For example, when determining drug content in a formulation, samples of placebo formulattons should be prepared and analyzed to confirm no interfering peaks. 2.3. Chiral Purity Determination During validation of a chiral determmation method, tt 1snecessary to demonstrate that no synthetic or degradatrve tmpurrties would interfere with the determination of either enantiomer at the required levels. The migration order of the enantiomers should be confirmed by spiking experiments with the pure enantiomer and the racemate if available. Figure 2 shows separation of a pure (+) enantiomer of fluparoxan spiked with 1% of the opposite (-) enantiomer to confirm migration order. 3. Detector Linearity It is necessary (3) to demonstrate detector lineartty over the intended operating range for the method. For example, if the intention of a specific method is to quote impurity levels at the 0.1% level, then tt may be appropriate to prepare standards at 0.1, 1, 10, 50, 100, and 150% of the
Al tria
86
target concentratton. These standards should be injected in duplicate, and the peak area obtatned plotted against percent of nominal concentration. Acceptable correlation coefficients (0.99 or greater) and an intercept close to the origin should be achieved. Many reports (7-9) have shown that this performance is possible with standard CE mstruments. 4. Rectilinearity
When determining impurity levels or trace enantiomer content, tt is also necessary to demonstrate detector rectilinearity. This involves maintaining the main component at a constant concentration and varying the content of the minor component to mimic real samples; where samples containing variable levels of impurities will be analyzed. Acceptable rectilmearities have been reported for lmpurrtres of an quinolone anttbiotic (1). Correlation coefficients of >0.99 have been reported (IO) for fluparoxan enantiomer levels of l-10% of the desired enantiomer content. A further example of this validation aspect has been provided (II) in the determination of levels of the undesired enantiomer of a drug in which the main component was held constant and the trace enantiomer spiked over the range 0. l-l 5% (correlatton coefficient 0.9995). Detector rectilinearity should be demonstrated by analyzing samples spiked with known levels of the appropriate impurities of interest. The main component concentration should be held constant, whereas the spiked impurity levels should be varied over the required range. For example, if the expected content of individual impurities is likely to range from O.l-l.O%, then a sample at the concentration specified in the method should be prepared and aliquots of this sample solution should be spiked with levels of the impurity of interest. For this example, it may be appropriate to spike with 0.1,0.25,0.5,0.75, and 1.O%. An unspiked sample should also be analyzed to quantify the residual level of the selected impurity present in the sample. The peak area of the impurity should be plotted against % w/w spiking. The correlation coefficient of the line should be in excess of 0.9, and the intercept value should be similar to the peak area obtained for the unspiked sample. 5. Response
Factors
The detection wavelength is typically a maximum for the main component. This may not necessarily be the case for the impurities whose chemical structures and UV spectra may differ greatly from the parent
Method Validation
__--
--.____--Impurity
I-
t x a
,
, z E
87
-c
--------.
F
‘8
--..
I I I I ’
, PD w
,
!
Impurity D ‘...--~
-._... .---..._ ---.______----
--..
Impurity I,,
, s 0,
I,
I s z
I
I
(I, s z
--.==.
L
G 1
( z z
Wavelength
Fig. 3. Spectra of ramtldine and three of its major impurities (reproduced from ref. 12).
molecule. If the UV absorbances of both the main component and impurity are different, then the quoted % area/area may not be a true indication of the level of impurity present. If the impurity has a significantly lower UV activity at the detection wavelength, then it will be underestimated. Conversely, if the impurity has a significantly higher UV activity, it will be overestimated. If the impurities have similar UV spectra, then the response factors are similar. Figure 3 shows the spectra of ranitidine and three of its major impurities (12) obtained from a diode array CE detector. To measure response factors, standards of the available impurities and the main component are prepared in duplicate, and analyzed in duplicate, employing the method conditions. The normalized areas are used to calculate response factors (RQ Rf = [normalized peak area / P x wt (mg)]
(1)
Altria where P = purrty of material, which accounts for all impurities, including moisture, solvents, inorganic residue, counterion content (if prepared as a salt), and related Impurities, i.e., 106total impurity content. Typically, rf the Rf of an impurity is 80-120% of that of the main component, it may be considered to be equivalent to the main component and no correction would be necessary. However, rf the response factor difference exceeds these hmits, rt may be necessaryto adjust the reported impurity content, i.e., if the impurity has an Rf of only 40% of the rmpurity, the reported value would be: (normahzed peak area lmpurlty / total normalized area) x (1 / 0.40) x 100 (2)
6. Limit of Detection CLOD) Thus value denotes the mnumum detectable level of impurities. The LOD is often defined (2) as the sample concentration that produces a peak with a height three times the level of the baseline noise. The LOD may alternatively be expressed as the smallest peak that can be detected as a % area/area of the electropherogram when determining purity, i.e., the LOD of 0.1% of one enantiomer in the presence of the main component may be possible. These LOD values will be dependent on the UV activity of the solute, the capillary bore, and the sample loading. Frgure 4 shows the LOD for a quinolone antibiotic analyzed using a low-pH CE method. 7. Limit of Quantitation (LOQ) The LOQ value refers to the lowest level of impurity that can be precisely and accurately measured. The LOQ may be calculated (IO) as ten times the signal-to-noise. Typically, replicate analysis of samples at the LOQ should give an RSD value of 10% or better. 8. Precision Repeated analysis (typically 10 sequential injections) of both a sample and a calibration solution should be conducted. Acceptable precision in terms of migration times and peak areas should be demonstrated to a predetermined level. Precision in CE is typtcally on the order of 0.5-2% RSD for main peak assay (7,&I.?) (with or without internal standard). Repeatable preparation of samples and standards should also be demonstrated. For example, 10 individual standards should be prepared and analyzed in duplicate. The precrsion for the response factors obtained
89
Method Validation
1
I
1
14.20
I
14.40 Retentlon
Fig. 4. LOD for a qumolone (reproduced from ref I).
I
14.60 14.80 tlme In minutes
antlblotx
I
15 00
I
15.20
analyzed using a low-pH CE method
Table 2 Reporting Format for Related Impurity Content Injection 1
Injection 2
Average
Total no. of lmpuritles Total % impurities % Principal impurity RMT of principal impurity % Second greatest impurity RMT of second greatest impurity
should be within acceptable levels (typically l-2% RSD). Similarly, 10 individual samples should be prepared and analyzed in duplicate, and the precision for the pooled assay values should be acceptable. Repeated analysis of the same sample solution should give consistent results (Table 2 shows the reporting format) in terms of number of impurities, levels and relative migration time for each impurity, and total level of impurities. For trace impurities, the precision would be expected to be ~10% RSD (I, 14). Migration time and relative migration times should be about 1% RSD and
90
Altria
method for determining the enantiomeric ratio of clenbuterol (23), peak area precision was l-2% RSD, peak area ratio precision was
Method Validation
91
lished on a case-by-case basis decided by the companies and laboratories concerned. Generally, this activtty will be concerned with the simultaneous testing of identical samples. Comparisons of results generated by both laboratories should not differ stgmficantly. This aspect has been shown in CE by a series of mtercompany crossvalidatton exercises (13,17), in which methods have been repeated by seven independent pharmaceutical companies. 10. Freedom from Interference Samples of the dissolvmg solvent and matrix-related components should be analyzed to confirm absence of interference. For example, if analyzing drug and drug-related impurities in a tablet formulation, a solutron of a placebo formulatron should be prepared and analyzed. 11. Recovery It is necessary to demonstrate full extraction of components from the sample matrix. This is achieved by spiking drug and related rmpurttres into the matrix of concern (18). For example, m the case of a pharmaceutical formulation, it would be appropriate to add known levels of drug and related impurities to a placebo mixture (2,19). Analysis of the spiked sample should confirm acceptable extraction. Demonstration of acceptable recovery is especially important in bioanalysis (20), where sample pretreatments can be extensive. 12. Crossvalidation It is useful to demonstrate the accuracy of the method by performing the same determination using an alternative method of analysis. Typical combinations may be CE and HPLC or CE and TLC. If similar results are obtained by two enttrely different separation techniques, then greater confidence can be placed m the validity of the results achieved by the method under examination. 12.1. Related Impurities CE and HPLC, when used together (21,22), are very complementary techniques for related impurity determinations, since common wavelengths can be selected and results quoted as % area/area.Comparisons with other separative techniques, such as TLC, become more complicated, since response factors need to be calculated and applied for each impu-
92
Altria
rity. Results obtained by CE and HPLC for salbutamol impurities were statistically compared (22) and shown not to be statistrcally different. 12.2. Main Peak Assay Combinations of CE with a variety of analytical techniques have been reported for main component analysis. This crosscorrelation has been predominantly shown using CE and HPLC (7-9). Correlation studies between HPLC and CE results from the testing of a range of drug formulations produced correlatron coefficients of >0.999 between the data sets. 12.3. Chiral Analysis Currently, this testing is predominantly performed by HPLC. Therefore, it is most appropriate to crossvalidate wrth extstmg HPLC methods if possible. Several reports (23) have shown good agreement between the two techniques. 13. Robustness (Ruggedness) Robustness relates to the sensitivity of the method to small deliberate deviations from the method. For instance, if the method states an operating temperature of 3O”C, will acceptable performances be maintained at either 25 or 35”C? Two approaches to method robustness are possible: univariate or multivariate. The univariate approach involves systematically varying each parameter sequentially. Thus “one-by-one” approach has been performed for the determination of enalapril content in tablets by CE (24). The multivariate assessment involves simultaneous evaluation of several parameters using a predefined matrix. Typical experimental designs that may be employed include Placket-Burman (2.5)and Central Composites. See Section 2., Chapter 20 for further details of the applications of experimental designs in CE. CE has been employed to determine levels of drug-related impurities (26). Both fractional factorial and central composite designs were employed in the robustness testing of this method. Typically, each parameter may be varied by 5-10% above and below the value set in the method. Parameters examined may include temperature, pH, electrolyte concentration, rinse times, additive concentration, detector wavelength, and sample loading. The responses measured may include resolutions, peak efficiencies, relative and absolute migration times, peak areas, and migration order. For example, in the determination of impurities by an MECC method employing 20 Wborax, pH 8.5,
Method Validation containing 30 mA4 SDS at 25”C, preseparation rinse with electrolyte for 2 min, and detection at 240 run with a sample concentration of 0.5 mg/mL, it may be appropriate to explore the effects of the following ranges: Rinse time: 1.5,2.0,2.5 min Borax concentration: l&20, 22 mA4 pH: 8.3,8.5, 8.7 SDS concentration: 28,30, 32 mA4 Temperature: 23,25, 27OC Wavelength: 235,240,245 Sample concentration: 0.45,0.5,0.55 mg/mL
These values approximate to slight variations to the method. If one of these extremes is highlighted as having a significant effect on selectivity or on quantitative results, then tight limits for that parameter would be specified in the method. In order to minimize the number of analyses, experimental designs, such as fractional factorials and central composites, may be employed (26,27). The results from robustness testing should
be statistically examined to identify the tolerance limits that should be prescribed on each parameter in the method. 14. Peak Homogeneity Comigration of peaks is possible in CE as in any other separative technique. Therefore, it is useful to investigate the purity of separatedpeaks. Approaches to peak purity determination in CE are similar to those employed in HPLC. The principal methods are fraction collection (28) and spectral characterization. Common approaches would be used m related impurity determinations, main component assay, and enantiomerit separations. 14.1. Fraction Collection Micropreparative fraction collection is possible in CE (see Chapter 9). Fractions can be taken from the CE separation and analyzed using an alternative separative technique, such as HPLC (28,29). If a single peak is indicated by the secondary test method, this gives further confirmation of peak purity. 14.2. Spectral Analysis Many commercial CE instruments are now equipped with diode array facilities, which can be used in assessing peak purity. Standards of the main component and impurity can be analyzed, and their UV spectra
94
Altria
recorded and compared (30). If good correlation between the spectra is not achieved, this can indicate comigration of peaks. This approach is llmlted in that closely related impurities will have similar UV spectra to each other and the main component. 15. System Suitability Testing The performance of the instrument 1s normally assessed prior to initiation of an analytical test sequence. As in HPLC (3), this testing commonly involves an assessment of parameters, such as selectivity, precision, and separation performance. For example, a test mixture contaming components that represent the critical resolutions may be repeatedly analyzed, for example, five times. In this instance, the system suitability may be specified as: 1. Resolution of C2.0 for all components 2. Precwon of inJectIon for mam component of <2% 3. Peak effhency of MOO0 for mam component. Other system suitability parameters less commonly employed may include linearity and LOD. Linearity may be assessed by employing three calibration solutions, for example, covering 7%125% of the target sample concentration. A measurement of LOD may be of significance when performing trace-level quantitatlon. A maximum permissible peak tailing factor may also be specified in the method. Provldmg that the system meets these mimmum requirements, the analysis may proceed. In some instances, test injections are also positioned at the end and/or throughout the sequence to demonstrate the continued suitability of the instrument and reagents through the analytical sequence. 16. Solution Stability The shelf-life of all reagent, calibration, and sample solutions should be determined. Until this is performed, all solutions should be made up fresh daily. Typically, the stability is established by performing fresh solutions to establish the purity of the sample and calibrations and to generate assay results for the samples. This testing represents d 1 of the stability determmatlon. The solutions should be stored under recorded conditions (i.e., protected from light at room temperature or as appropriate). The analysis is repeated after, for example, 3 d, using freshly prepared sample, calibration, and reagent solutions. Two analytical sequences
Method Validation
95 Table 3 Method Vahdatlon Checkhst
Test
Result
Selectivity Linearity (range -to - nominal cone ) Rectllmeanty (range - to - nommal) Response factors Limit of detection (- mg/L, % nominal cont.) Limit of quantitatlon (- mg/L, - % nominal cone ) Preclslon of migration time (- RSD, n = -) Peak area (- RDS, n = -) Reproduclblhty between analysts between capillaries between mstruments Freedom from Interference Recovery Crossvalidatlon Ruggedness/robustness Peak homogeneity
would be performed in which both stored and fresh calibrations and sample solutions are initially analyzed with fresh reagents and then reanalyzed in a second sequence using the stored reagents. The impurity content for stored and fresh samples should be determined, and the assay results should be calculated using response factors from both the stored and fresh calibrations.
All solutions would be deemed fresh if: 1. Assay results calculated usmg response factors from fresh and stored cahbrations were shown to be identical.
2. Assay results for stored and fresh samples were identical. 3. Selectlvlty should be identical for stored and fresh reagents. 4. Impurity levels m stored sample and cahbratlon solutions should not be significantly increased compared to fresh solutions.
If d 3 results are acceptable, then further testing at 7- and 14-d, and l- and 3-mo time-points may be appropriate. The shelf-life for calibrations/samples would generally be shorter than for reagents. The shelflife and storage conditions for all solutions should be documented in the method. Table 3 gives a checklist for the tests performed in validation protocols.
96
Altria References
1 Altrra, K. D and Chanter, Y L (1993) Vahdatton of a caprllary electrophorests method for the determmatton of a qumoline anttbtottc and Its related tmpurmes J Chromatogr 652,459-463. 2. Thomas, B. R,, Fang, X G., Chen, X , Tyrell, R. J., and Ghodbane, S (1994) Vahdated mtcellar electrokmettc captllary chromatography method for the qualny control of the drug substances hydrochlorothiazide and chlorothrazide J Chromatogr 657,383-394. 3 Clarke, G S (1994) The validatton of analyttcal methods for drug substances and drug products m UK pharmaceuttcal laboratortes J Pharm Boomed Anal 12, 643-652 4 Edwardson, P. A D., Bhaskar, G , and Fanbrother, J E (1990) Method vahdatton m pharmaceuttcal analysrs J Pharm Boomed Anal 8,929-933 5 Shafaatt, A and Clark, B J (I 993) Development of a caprllary zone electrophoreSISmethod for atenolol and Its related impurines m a tablet preparatton Anal Proc 30,48 I-483. 6 Korman, M , Vindevogel, J., and Sandra, P (1993) Separation of codeme and Its by-products by capdlary zone electrophorests as quahty control tool m the pharmaceuttcal mdustry J Chromatogr 645,3&k370 7 Altrta, K. D and Ftlbey, S. D. (1993) Quantttattve pharmaceutrcal analysrs by captllary electrophoresls J Llquld Chromatogr 16, 228 1-2292 8 Ackermans, M. T., Beckers, J. L , Everaerts, F M., and Seelen, I G J A (1992) Compartson of tsotachophorests, captllary zone electrophorests and htgh-performance lrquld chromatography for the determmatton of salbutamol, terbutalme sulphate and fenoterol hydrobromtde m pharmaceuttcal dosage forms. J Chromatogr 590,341-353 9 Tsat, E W , Smgh, M. M , Lu, H H , Ip, D. P., and Brooks, M A (1992) Apphcanon of captllary electrophorests to pharmaceuttcal analysts. Determinatton of alendronate m dosage forms J Chromatogr 626,245-250. 10 Altrra, K. D , Walsh, A R., and Smtth, N. W. (1993) Vahdation of a captllary electrophoresrs method for the enanttomerrc purity testmg of fluparoxan. J Chromatogr 645, 193-l 96. 11 Werner, A , Nassauer, T., Krechle, P., and Erni, F. (1994) Choral separation by capillary zone electrophorests of an optically-acttve drug and ammo acids by host-guest complexatton w&h cyclodextrms J Chromatogr 666,375-379 12. Altria, K. D. and Grace, G (1994) m preparation 13 Altrta, K D , Harden, R C., Hart, M., Hevtzt, J , Harley, P A., Makwana, J V , and Portsmouth, M J (1993) An Inter-company cross-vahdatron exercise on caprllary electrophoresis 1. Choral analysis of clenbuterol. J Chromatogr 641, 147-l 53 14 Swartz, M E. (1991) Method development and selectivrty control for small molecule pharmaceutical separattons by captllary electrophorests J LIqutd Chromatogr 14,923-938. 15 Flurer, C. L and Wolmk, K. A (1993) Quantttatton of gentamrcm sulfate m mJectable soluttons by capillary electrophorests. J Chromatogr. 663, 259-263
Method Validation 16 Madrup, G. (1992) Rugged method for the determination of deaminatron products in msulm solutions by free zone captllary electrophorests usmg an untreated fusedsilica caprllary. J Chromatogr 604,267-28 1. 17. Altrra, K. D., Clayton, N. G., Harden, R. C., Hart, M., Hevtzi, J., Makwana, J. V., and Portsmouth, M. J (1994) An inter-company cross-validation exercise on capillary electrophorests testing of dose uniformity ofparacetamol content m formulations. Chromatographiu 39, 180-l 84. 18. Sun, P., Mariano, G J., Barker, G., and Hartwtck, R A (1994) Compartson of micellar electrokinettc capillary chromatography and high-performance liquid chromatography on the separation and determination of caffeine and tts analogues in pharmaceutical tablets Anal Lett 27,927-937 19. Thomas, B. R. and Ghodbane, S. (1993) Evaluation of a mixed mtcellar electroklnetic capillary electrophorests method for validated pharmaceutical quality control. J Liquid Chromatogr 16, 1983-2006 20. Prunonosa, J., Obach, R , Dtez-Cascon, A., and Gouesclou, L (1992) Compartson of high-performance hqutd chromatography and high-performance capillary electrophoresis for the determmatton of cicletanme in plasma J Chromatogr 581, 219-226 21 Pluym, A., Van Ael, W , and De Smet, M (1992) Capillary electrophoresrs m chemtcal/pharmaceutrcal quality control TRAC 11,27-32 22. Altrta, K. D. (1993) Quantitative analysts of salbutamol related lmpurtttes by capillary electrophorests. J Chromatogr 634, 323-328 23. Rogan, M. M., Altrta, K D , and Goodall, D. M (1994) Enanttoselectrve separations using capillary electrophoresis. Chwality 6,2-O. 24. Thomas, B. R. and Ghodbane, S (1993) Evaluation of a mixed mtcellar electrokrnetrc captllary electrophorests method for validated pharmaceuttcal quality control. J. Lzqurd Chromatogr 16, 1983-2006 25. Vindevogel, J. and Sandra, P (1991) Resolutton optlmtsatton m mlcellar electrokinetic chromatography. use of Plackett-Burman statistical design for the analysis of testerone esters Anal Chem 63, 1.530-l 536. 26 Altria, K D and Filbey, S. D. (1994) The apphcatton of experimental design to the robustness testing of a method for the deterrnmatton of drug related impurities by capillary electrophoresis. Chromatogruphaa 39, 306-3 10 27. Filbey, S. D and Altria, K D (1994) Robustness testing of a capillary electrophorests method for the determmation of potassium content m the potassium salt of an acidic drug. J Cap Electrophoresis 1, 190-195. 28. Camillerr, P., Okafo, G. N , Southan, C., and Brown, R (1991) Analytical and mrcropreparattve capillary electrophoresis of the peptides from calcitomn Anal Bzochem. 198,36-42. 29. Altria, K D. and Dave, K (1993) Peak homogeneity determination and mtcropreparative fraction collection by capillary electrophoresis m pharmaceuttcal analysis. J Chromatogr 633,22 l-225. 30. Beck, W., Van Hoeck, R., and Engelhardt, H. (1993) Application of a diode-array detector m capillary electrophorests Electrophoresls 14, 540-546
CI-IAPTER9
Fraction Kevin
Collection D. Altria
1. Introduction Fraction collection by CE is a potentially important area in which there are several approaches necessary (I-8) to yield sufficient material following fraction collection. Even using optimized conditions, the amounts collected employing preparative CE are extremely small and are on the order of nanogram quantities. However, these levels can be useful to perform operations such as: (1) peak purity verification by reinjection on CE or onto another analytical method, such as HPLC, and (2) sample identity confirmation by characterization of the material by another technique, such as amino acid sequencing or off-line mass spectrometry. The key advantage to preparative CE compared to, for example, preparative HPLC, is that the fraction collection can be readily performed from tiny sample volumes, since typical CE injection volumes are only l-50 nL/injection. The technicalities of performing preparative CE are demanding and require use of a programmable autosampler. In principle, however, the operation of preparative CE is identical to normal CE, and standard equipment is employed. Sample solutions are injected, separated, and detected in a conventional fashion. The separated components are then collected as they emerge from the detector end of the capillary. To allow this collection, it is necessary to stop the separation voltage and position the autosampler tray such that the capillary then dips into a collection vial, containing a few microliters of water or dilute buffer. The voltage is then restarted, and the required sample components migrate out of the capillary into the collection vial. When all the sample zone has emerged from the capillary, the voltage is stopped. From
Methods m Molecular Biology, Vol 52 Caprllary Necfrophoresrs Edited by K Altna CopyrIght Humana Press Inc , Totowa, NJ
99
Al tria
100
The times required for calculating the exact time for the swttching between vials can be calculated from the total capillary length, the distance to the detector, and the migration time of the peak of interest (I). Therefore, the switching time to collect a peak detected at 7.8 min on a 57-cm long capillary (50 cm to the detector) is: x 7.8 = 8.9 mm
(1) The exact time of the switch should, however, be programmed to be at least 0.1 min before the expected time to ensure the entire peak is collected. The length of peak collection should also be suffictent to cover collection of the required peak only. Typically, if the required peak is well resolved from interfering peaks, then the collection step may be for 0.5 min. Using this example of a peak that was detected at 7.8 min, the following method may be appropriate: 57150
Step 1: Rinse with generatlonsolution (0.IM NaOH or slmllar). Step 2: Rmsewith electrolyte solution. Step 3: Set temperature. Step 4: Inject sample. Step 5: Apply voltage for 8.8 mm (capillary dipping into electrolyte vial at detectorend). Step 6. Apply voltage for 0.5 mm (capillary dtpping into fraction collection vral at detectorend). There are three main approaches to optimizing the amount collected by preparative CE. These are to use concentrated sample solutions, perform several sequential runs, or employ relatively large-bore capillaries. 2. Use of Concentrated Sample Solutions The limitation 1s that resolution may be lost with increased sample loading, For example, 25 mg/mL fluparoxan solutions (I) and 5 mg/mL calcitonin solutions (2) have been used in micropreparative work. 3. Use of Wide-Bore Capillaries Increasing the bore dramatically increases the volume of sample injected into the capillary. However, the required resolution may be lost at these higher injection volumes, and considerations must be given to joule heating effects. A 100 1-1x 57 cm polyacrylamide-coated capillary has been used (3) to collect peptlde components from a four-component
Fraction Collection
Fig. 1.The preparativeseparationof samplecontaining four peptldes(reproducedfrom ref. 3). text mix of peptides. The fractions were collected into a IO-PL microvial of separation buffer. The separation was operated under stacking conditions (see Chapter 16), which allowed a 20-s injection on this relatively large-bore capillary, which is equivalent to a 500-nL injection volume. Figure 1 shows the preparative separation of the sample. The collected fractions were reexamined by CE to confirm purity levels. Fifty nanograms of the individual peptides were collected (3). This was sufficient material to conduct tirther analyses,such as amino acid sequencing. 4. Repeated Injections Repeated injections in an automated sequence would also increase the total amount collected. This facility IS possible with well-controlled methods where very consistent migration times are obtained. Therefore, particular attention should be paid to the use of appropriate rinse cycles and to operation using a constant temperature. It may be advisable to perform two test Injections prior to starting a preparative sequence to
10.2
Altria
allow the system to equilibrate and to reach a constant temperature (see Chapter 3, Section 10. on good working practices). For example, a fraction sample of a fluparoxan impurity was obtained following 27 injections of a drug substance solution. Injection sequences of up to 30 injections were shown (4) to give collectron yields of 70-100% for injected amounts of various peptides and proteins. The expected amount collected can be calculated for a given number of injections using a particular set of conditions (I). For example, 27 replicate injections of a 25 mg/mL solution containing a 2.5% impurity, using injection settings that give a 11.8-nL injection volume with collection of the fraction mto 20 l..tLof water, yielded a lo-ppm (mg/L) solutton. Each injection is equivalent to a loading of 295 ng. However, the impurity is only present at 2.5%. Therefore, each injection gives a loading of 7.4 ng of the impurity. Twenty-seven replicate injections give a total amount of impurity collected as 198 ng into 20 ltL of water. This gives a final solution concentration of 10 pg/mL (ppm), which is suitable for further mvestigations. 5. Notes 1. The configuration on the ABI mstrument requnes that the detector end reservoir is fixed and cannot be used for fraction collection purposes. Therefore, a scheme has been devised (4) m which sample solution IS vacuum-injected and a rinse cycle is used to suck the sample slug to the detector end of the capillary. A negative voltage is then apphed, causmg the sample components to mtgrate back along the capillary, pass through the detector, and emerge from the capillary at the sample mtroduction end of the capillary. Again the voltage IS stopped immediately prior to
emergence of the sample zone from the capillary, and the capillary is dipped into a collection vial and the voltage restarted. A user bulletin (user bulletm number 4) is available from ABI that gives precise details of this operation. 2. Bundles of capillaries m an array format (5) have been shown to be of use m particular applications. Beckman Instruments are expected soon to launch a clmical CE mstrument (Tradename, Paragon) in which several capillaries are simultaneously employed, This instrument would have obvious application to preparative operations.
References 1 Altrra, IL D. and Dave, Y. K. (1993) Peakhomogeneitydetermination and mrcropreparativefraction collection by capillary electrophoresisin pharmaceuticalanalySW J. Chromatogr
633,221-225.
Fraction
Collection
103
2. Camilleri, P , Okafo, G. N , Southan, C , and Brown, R (1991) Analytical and micropreparative capillary electrophoresrs of the peptides from calcitomn Anal Blochem 198,36-42 3. Schwer, C and Lottspeich, F. (1992) Analytical and mrcropreparatrve separatron of peptrdes by capillary zone electrophoresis using drscontinuous buffer systems J Chromatogr. 623,345-355. 4. Albin, M., Chen, S -M., Loure, A, Pairaud, C., Colburn, J., and Wrktorowrcz, J. (1992) The use of caprllary electrophoresis m a micro-preparattve mode methods and applicatrons. Anal Blochem 206,382-388. 5. Huang, X. C , Quesada, M. A , and Mathres, R A (1992) DNA sequencing usmg capillary array electrophoresis. Anal Chem 64,2 149-2 154 6. Herold, M. and Wu, S -L (1994) Automated peptide fraction collection m CE, LCGC Int 7, 554-558. 7. Lee, H. G. and Desiderro, D. M. (1994) Preparatrve caprllary zone electrophorests of synthetic peptides. Conversion of an autosampler mto a fractron collector J Chromatogr 686,309-3 17. 8 Lee, H. Ci. and Desideno, D. M (1994) Optrmrsation of the loading hmrt for capillary zone electrophoresrs of synthetic oprod and tachykinm peptrdes. a study of the interactions among the amount of peptide, resolutron, saturation, mJectton volume and the capillary diameter. J Chromatogr 662,35-45.
CHAPTER 10
Troubleshooting Kevin
D. Altria
1. Introduction This chapter covers the possible solutions to problems that may be encountered in routine operation. Further details may be available in specific instrument manuals or by contacting appropriate service specialists. The most likely problem areas are: 1. Voltage/currentproblems-Chart 1 2. No peaksobtained-Chart 2. 3 Poor peak shape/height-Chart 3. 4. Poor peak time precision-Chart 4. 5 Poor peakareaprecision--Chart 5. 6. Poor sensitivity--Chart 6. 2. Chart l-Voltage/Current Problem 2.1. Blocked
or Broken
Capillary
Capillaries can block if stored filled with electrolyte, which will crystallize out with time. To test if a blockage has occurred, an empty vial should be placed on the autosampler. A rinse cycle is then performed from this empty vial, which should result in a stream of air bubbles in the “detector end” of the electrolyte reservoir. If no bubbles appear after 1-2 min, then the capillary is blocked. A series of rinse steps may clear the blockage. If this is unsuccessful, then a new capillary should be installed. If the capillary is broken, no current will be generated when a voltage is applied across the capillary when filled with an electrolyte. Visual inspection of the capillary will confirm this problem. A new capillary should be installed. From
Methods I# Molecular Bology, Vol 52 Capllary Electrophoresti Edlted by K Altrta CopyrIght Humana Press Inc , Totowa, NJ
105
Al tria
106
Chart 1. Voltage/current problem.
2.2. Correct VoltagelCurrent The level and polarity of the voltage specified in the method should be checked.Application of too high a voltage will causea rapid increasein the temperaturewithin the capillary, possibly boiling the electrolyte or causing out-gassing of air bubbles. Either of thesetwo occurrenceswill result in breakdown of the electrical circuit. Application of the incorrect voltage polarity will result in sampleions migrating in the oppositedirection along the capillary to that required. 2.3. SamplelElectrolyte Vial Levels If the vials are insufficiently filled, a rinse or injection cycle will cause air to enter the capillary, resulting in an electrical breakdown. This problem should be consideredif a long injection sequenceis performed with rinsing from a single electrolyte reservoir.
Troubleshooting
107
pil +,
m sample?
~~1
Chart 1. continued.
2.4. Voltage Leak If all areasaround the high-voltage connectionsare not kept clean and free from electrolyte, the voltage may escapeto earth rather than pass acrossthe capillary. Principal problem areasinclude electrolyte drops on buffer vial caps.Theseshould be checkedprior to an analysis. Operation at subambienttemperaturecan lead to condensationproblems. Urea is occasionally used as an electrolyte additive, which is known to “creep” acrosssurfaces,leading to voltage leakages.Regular careful cleaning of electrode areas with cotton wool dampenedwith water will minimize voltage leakageproblems. Specific maintenanceadvice will be included in instrument manuals. 2.5. Injection Time Typically, samplesare preparedin water or may contain someorganic solvent. Therefore, the sample zone is relatively nonconductive and, if
108
Al tria
the zone length is too long, will causea breakdown tn the conductivity along the capillary This effect is reduced by shortening the sample injection time. 2.6. Buffer If the buffer is too concentrated, then excess current will be drawn, which will result m breakdown in the circuit owing to electrolyte boiling or out-gassing. The electrolyte should be diluted accordingly or a lower voltage applied. 2.7. Correct Vials The positioning and contents of the vials on the autosampler should be verified and replaced if incorrect. 2.8. SampZe Solvent If the sample 1s dissolved in an inappropriate solvent (such as pure organic solvent or of high ionic strength), this may cause a breakdown owing to localized heating in the sample region. This effect can be reduced by reducing the sampling time or by dilution of the sample with water. 2.3. Air Bubble in Sample If only small sample volumes are available, microvials are used to hold the samples. When using these microvials, it is possible to trap air in the bottom of the vial. This will result in injection of air mto the capillary, resulting in electrical breakdown. The sample microvials should be refilled with sample solution to remove the trapped air. 3. Chart ~-NO Peaks Obtained Blocked or broken capillary--see Chart 1. Correct voltage/current set--see Chart 1. Sample vial level--see Chart 1. Voltage leak--see Chart 1. 3.1. Analysis Time Too Short The analysis time should be extended (or voltage increased) and the sample reanalyzed. 3.2. Sample SoZu bili ty If the sample is insoluble in the electrolyte, it may drop out of solution within the capillary. If this is suspected, an alternative electrolyte should be selected in which the sample is sufficiently soluble.
Troubleshooting
109
Chart 2. No peaks obtained.
110
Altria
1311 Chart 2. continued. 3.3. Correct
Buffer
and Vials
The positioning and contentsof the vials on the autosamplershould be verified and replaced if incorrect. 3.4, Air Bubble
in Sample-See
3.5. Capillary
Chart
1.
Not AZigned
If the capillary detection window area is not correctly aligned within the detector, no peaks will be observed, since the UV light will be adsorbedby the polyimide coating on the capillary. This can be checked following guidance from specific instrument manuals and the capillary realigned if necessary. 3.6. Lamp
Burned
Out
This can be checked following guidance from specific instrument manualsand the lamp replaced if necessary.
111
Troubleshooting
0
Change and reanalyze
Chart 3 Poor peak shape/height. 3.7. Data
Collection
The connections to the data system should be checked. The correct operation of the data system should be checked, and the system replaced if appropriate. 4. Chart 3-Poor Peak Shape/Height 4.1. Correct
EZectroZyte
Used
If the electrolyte concentration is too high, excessive current will be drawn, possibly leading to temperature-related peak distortion. Conversely, if the electrolyte concentration is too low, the conductivity of the sample will lead to peak distortion or tailing. The electrolyte concentration should be optimized during method development. 4.2. Correct
VoltagelCurrent
Excessive voltages will result m temperature-related peak distortion. However, too low a voltage will result in extended migration times and diffusion-related peak broadening.
Altria
Chart3. continued. 4.3. Injection
Time
Excessivesamplevolumes will causemixing of the samplewithin the capillary, resulting in loss of resolution and separation efficiency. The injection time should be optimized during method development. 4.4. Sample
Matrix
Extreme peak distortion can occur if the sample is present in a highionic-strength solution or high organic solvent levels (I). Reduction in sampleinjection time or dilution of the samplewill reducepeak distortion. 4.5. Temperature
Too high a temperaturecan causepeak splitting and distortion, and lower temperatures increase analysis time, therefore the temperature should be optimized during method development.
Troubleshooting
113
Chart 4. Poor peak time precision. 4,6. Sample
Adsorption
Peak tailing owing to sample adsorption may be reduced by several approaches,such as use of pH extremes,high-ionic-strength buffers, or coatedcapillaries.Appropriateuseof rinsecyclesbetweeninjectionscanbe usedto removeunquantifiedmaterial from the capillary, which may cause adsorptionproblems.An increasein electrolytestrengthor decreasein pH can reducesample/surfaceinteractions,which may be causingpeak tailing. 4.7. Sample
Concentration
Sampleoverloading will dramatically reducepeak efficiency and resolution. The sampleshould be diluted or a shorter injection time selected. 5. Chart 4-Poor Peak Time Precision Chapter 6 discussesfactors affecting precision in greater detail.
114
Altria
5.1. Temperature Electrophoretlc mobihty 1sdirectly related to temperature, and therefore, fluctuations in temperature during a sequence will cause variations in mlgration times. The operating temperature should be specified in the method. 5.2. Capillary Surface Variations m the capillary surface will lead to changes m the EOF, resultmg in mlgratlon time vanablhty. Appropriate capillary conditionmg and use of rinse cycles of base, acid, water, or electrolyte should ensure a consistent EOF during a sequence. These cycles should be establashed durmg method development. 5.3. Electrolyte Depletion Extensive apphcatlon of a voltage across an electrolyte solution can cause electrolysis of the buffer to occur (2,.?). This may result in drifts m mlgratlon time and/or changes in selectivity. Therefore, it may be approprlate to program use of fresh vials of electrolyte during an extended series of injectlons (4) 5.4. Standard/Sample Composition The ionic strength and organic solvent content of the sample may alter EOF, leading to changes in migration time. Therefore, it is necessary to match the composition of samples and standards.
6. Chart
B-Poor
Peak Area Precision
Chapter 6 discusses factors affecting precision in greater detail. 6.1. Temperature Sample viscosity decreases with increased temperature. The sample volume introduced into the capillary is also related to viscosity. Therefore, an increase m temperature will result in a higher sample loading onto the capillary, giving a higher peak area. A constant temperature should be specified in the method. 6.2. Migration Time In CE, the area of a peak is proportional to both the solute concentration and its migration time (5) Therefore, variations in migration time will produce apparent variations m peak area. The factors controlling migration time precision are covered in Chart 4.
115
Troubleshooting
Chart 5. Poor peak area precision.
6.3. Peak Size Peak areaprecision is improved with peak size. Use of higher sample concentrations or longer injection times produces larger peaks. Errors relating to peak integration and solute adsorption are reduced with increasedsample loading. 6.4. Evaporation Evaporation during a sequencewill lead to a gradual increasein areas during a sequence.Samplevials should be sealedif possible or an internal standardemployed. 7. Chart 6-Poor Sensitivity Chapter 7 discussesfactors affecting sensitivity in greater detail.
116
Al tria
Chart 6. Poor sensitivity. 7.1. Sample Load Sample injection time should be increased as much as possible while still retaining required resolutions. 7.2. Capillary Alignment If the detection window area of the capillary is not correctly aligned, a loss in sensitivity will result. The alignment should be checked according to the instrument manual. 7.3. Capillary Bore An increase in capillary bore will improve sensitivity, but may have implications relating to resolution and selectivity (see Section 2., Chapter 7). 7.4. Noise Spikes Particulates in the sample or electrolyte may migrate along the capillary and pass through the detector. Theseparticulates will appear as needle-sharp
Troubleshooting
117
Chart 6. continued peaks. The electrolyte should be filtered through an appropriate filter. A sample filtration step would require further additional validation. 7.5. Lamp Poor lamp intensity or intermittent lamp output will be observed as extreme baseline noise. Instrument-specific lamp performance diagnostic tests should be performed and the lamp replaced if necessary. 7.6. Voltage Increased voltage causes higher capillary vibration, and both shortand long-term baseline noise. A reduction in voltage, although extending analysis time, will reduce this source of noise. 7.7. Temperature Capillary vibration noise is also related to temperature, being more pronounced at higher temperatures. A reduction in temperature will reduce this effect, but may unduly alter selectivity and resolution and extend analysis time.
Altria
118 References
1 Ackermans, M T , Everaerts, F M , and Beckers, J L. (1991) Determmatton of some drugs by mrcellar electrokmetrc captllary chromatography The pseudoeffecttve mob&y as parameter for screening. .I Chromatogr. 585, 123-13 1 2 Shafaatr, A and Clark, B J (1993) Development of a caprllary zone electrophoreSISmethod for atenolol and rts related rmpurmes m a tablet preparatron. Anal Proc 30,48 1-483 3 Zhu, T , Sun, Y -L., Zhang, C -X , Lmg, D -K , and Sun, Z -P (1994) Varratron of pH of the background electrolyte as a result of electrolysis in CE JHRCC 17,563,564. 4 Thomas, B R., Fang, X G , Chen, X, Tyrell, R J , and Ghodbane, S (1994) Valrdated micellar electrokmetrc captllary chromatography method for the quahty control of the drug substances hydrochlorothtaztde and chlorothrazrde. J Chromatogr 657,383-394 5 Altrta, K D. (1993) Essentral peak area normahsatron m caprllary electrophorests Chromatographta 35, 177-182
CHAPTER11
to Running
Quick Guide a Successful Separation Kevin
D. Altria
1. Introduction This chapter is intended to act as a brief guide to the operations required to perform a series of CE analyses and subsequent care of the capillary. 2. Preanalysis Procedures 1. Specify all parameters to Include sample concentration and dissolving solvent, electrolyte type and concentration, capillary length and bore, injection volume, rinse settings, wavelength, applied voltage, and analysis time. All these wrll have been optimized during method development (Chapter 4). 2. Check all connections and settings. The majority of mstruments perform rinses and sample injection by means of an external gas supply. In addition, most instruments involve use of a cartridge-type device to house the capillary. The appropriate capillary, or cartridge, designated for use for the specific separation should be inserted mto the Instrument. 3. Self-diagnostic tests are performed when the power IS turned on. Accuracy of wavelength, gas pressure, temperature, and voltage are generally tested. The lamp should be independently switched on if necessary. It is recommended to allow 15 mm for the mstrument and lamp to warm up. If the instrument employs a filter-based detector, the appropriate filter should be selected to allow it to warm up. Failure to preselect the appropriate filter wtll result in a driftmg baseline for the first 15 mm of operation 4. All of the operatmg parameters (as listed m step 1) should be programmed into a method file on the controlling personal computer. From
Methods m Molecular Biology, Vol 52 Caplllary Electrophoresds Edited
by K Altrra
CopyrIght
Humana
119
Press Inc , Totowa,
NJ
120
Altria
5. The electrolyte and/or regeneration solution should be filtered through a 0.45+m filter to remove any parttculates Failure to do this can result in notse spikes as the parttculates migrate through the detector. 6 Fill a sufficient number of autosampler vials with regeneration solutton and electrolyte. Fill the vials equally to the prescribed level m the specific mstrument manual. If the vials are unevenly filled, a siphoning flow can occur, which causesmtgration ttme variability If a large number of analysesare performed, levels m the vials being rinsed from would be constderably lowered. Therefore, if a long sequence 1s to be performed, rmses should be conducted from different vials to those used during separation. If the vials are not filled enough, the electrode and capillary will not reach the solutton, resulting m a run failure. Ensure that the caps of the vials are clean and free from electrolyte. Contamination can cause electrical arcing and would result m a run failure. Repeated analyses from a single electrolyte reservoir can alter the nature of the electrolyte owing to electrolysts. This occurrence ts termed “buffer depletion,” and may cause changes m electrolyte concentratton and pH Buffer depletion may result m changes in selectivity and drifting mtgration times. To reduce or prevent this depletion, frequent changes m electrolyte vials should be programmed mto the sequence. Place a sufficient number of vials onto the autosampler m the posittons designated m the method and/or sequence. Ensure sufficient levels of solution are in rinse vials, since repeated rmses lower the level and eventually the level may be reduced such that the capillary cannot reach the liquid. A rinse would then fill the capillary with air and would result in a run failure. 7. Fill the sample vials with sufficient sample solution to allow the captllaty to reach the ltqutd. Filter the sample solutton as necessary (this would require addttional valtdatton). If sample volumes are limrted, plastic mtcrovials may be employed with mnumum sample volumes of -10 pL. If these mtcrovtals, or similar, are employed, ensure that no au bubbles are present. These bubbles are often caused by trapped an at the bottom of the vial. If au bubbles are present, the vials should be refilled or the vial should be gently tapped until the au-bubble 1sshifted. Place the sample vials m the autosampler positions designated m the method/sequence. 8. Condttton the capillary as necessary, The first ever use of a capillary wtll involve a 20-mm rinse with 0.M NaOH to regenerate surface stlanols Coated captllaries may have specific regenerating requirements and reference should be made to the mstructions supplied with the capillary Followmg the conditioning, rinse and fill the capillary wtth the required electrolyte.
Successful Separation
121
9. Close the autosampler door/cover to activate the safety interlocks The mterlocks ensure that it is not possible to accessthe capillary/electrodes while the voltage is being applied. 10. An appropriate test mixture should be analyzed to ensure that the system is suitable for operation. This injection also allows the system to settle. 11. The integrator should be assigned with the correct analysis time and number of data files corresponding to the number of injections m the sequence.The detector output should be reversed if indirect detection is to be employed.
3. Postanalysis
Procedures
1. The capillary should be rinsed for 2-5 min with water to remove electrolyte. A vial of water should be placed on the instrument to allow this operation. 2. The capillary should be flushed with an prior to storage. This is to prevent any residual electrolyte crystallizing out Inside the capillary to form a blockage. An empty capped vial is placed on the autosampler and a 2-5 min rinse performed from the empty vial. 3. Carefully remove the capillary for storage. The capillary, or cartridge, should be labeled with an identifier indicating the history and dimensions of the capillary. If the capillary 1shoused in a cartridge, it should be stored in the box provided. Loose capillary can be coiled and stored m plastic boxes available from the instrument suppliers. The fragile capillary wmdow section may be protected using a plastic tubing sheath. A 2-in. section of narrow-bore plastic tubing, as used in HPLC, could be cut and the capillary threaded through until the capillary window section is covered. 4. If the instrument is to remain unused for an extended period, it may be appropriate to turn the mstrument off. However, if further imminent use is intended, the lamp alone may be switched off to preserve lamp life. 3.1. Capillary It is recommended that:
Care
The capillary be rinsed for 20 min with 0. IM NaOH on its initial use. The capillary be flushed with 0.M NaOH (or acid) and equilibrated with the running buffer prior to each injection as appropriate. The capillary be rinsed with water and air blown through it at the end of each day. This prevents buffer precipitation and subsequent capillary blockage. Individual capillaries be dedicated to specific methods to avoid “memory effects” that may lead to nonreproducible separations. An initial injection of a blank, or sample solutton, be performed prior to commencing an analytical sequence to allow the capillary wall to equilibrate to the run buffer.
122
Altria Table 1 Preanalvsls ChecklIst Actlon
Comphes (J)
Date Analyst System Correct capillary size/length Correct analysis time on CE instrument Correct analysis time on Integrator Correct voltage/current selected Correct mjection parameters Buffer and rinse vials sufficiently full Correct number of sample vials All sample vials full with no air bubbles Vials correctly posltloned on autosampler Safety door closed Blank solution included in sequence Correct wavelength selected Correct temperature specified Integrator parameters correct Test mjectlon satisfactory
Table 1 provides a preanalysis checklist that could be completed in order to ensure that all the correct procedures have been performed prior to starting an injection or series of injections.
CHAPTER
Micellar
Electrokinetic Koji
Otsuka
and
12 Chromatography
Shigeru
Terabe
1. Introduction Recently, capillary electrophoresis (CE) or high-performance capillary electrophoresis (HPCE) has become popular as a high-resolution separation method. The technique was first Introduced by Mikkers et al. (I), Jorgenson and Lukacs (2), and Hjerten (3), as an instrumental version of electrophoresis. Although CE can generally give a higher resolution within a shorter time compared with conventional high-performance liquid chromatography (HPLC), only ionic or charged solutes can be separated by CE, in principle. This was a serious limitation of CE, but the development of electrokinetic chromatography (EKC) (4) has been able to solve such problems. EKC is based on chromatographic principles using homogeneous solutions containing an ionic “carrier” and the same apparatus as CE. The unique characteristic of EKC is that both neutral and charged analytes can be separated electrophoretically. Among various modes of EKC, micellar EKC (MEKC) (5-7), which uses micellar solutions of ionic surfactants, has become the most popular technique for the separation of small neutral molecules. Many papers on fundamental characteristics and applications of MEKC have been published (8), and some reviews of MEKC are available (9-12). In this chapter, the separation principle and chromatographic consideration of MEKC will be described tirst as basic properties of MEKC. Then, strategies for selectivity manipulation will be discussed briefly, followed by the description of some applications. From
Methods m Molecular Biology, Vol 52 Cap/llary Electrophoresm Edlted by K Altria CopyrIght Humana Press Inc , Totowa, NJ
125
Otsuka and Terabe
126 --_-_------___--___---
-m----------m-
O-J = Surfactant (negatrve charge) I= I = Solute
w
Electroosmotlc = Electrophoresls
Flow
Fig. 1 Schematic lllustratlon of the separation prmclple of MEKC (ZI) Reprinted with permIssIon from Terabe (1992), Mxellar Electrokmetlc Chromatography; Beckman, Cahfomla.
2. Principle
of Separation
The separation principle of MEKC is schematically shown in Fig. 1 (I 1). A capillary or fused silica 1sfilled with an ionic surfactant solution, in which the concentration of the surfactant 1s higher than its crltical mlcelle concentration (CMC), so that mlcelles are formed. When an anionic surfactant, such as sodium dodecyl sulfate (SDS), is employed, the micelle is forced toward the positive electrode by electrophoresls. The electroosmotic flow migrates toward the negative electrode owing to the negative charge of the capillary surface. The electro-osmotic flow 1s larger than the electrophoretic migration of the micelle under neutral or basic conditions, and therefore, the anionic SDS micelle also migrates toward the negative electrode at a retarded velocity. When a neutral analyte 1s injected into the micellar solution, it will be distributed between the micelle and bulk solution. The analyte will migrate at the same velocity of the micelle when it is incorporated mto the micelle, and at the electro-osmotic velocity, since it is free from the micelle and exists m the bulk solution. Thus, the migration velocity of the analyte depends on the distribution coefficient of the mlcellar solubllizatlon. As long as the analyte is neutral, It must migrate at a velocity between the two extremes, i.e., the electro-osmotic velocity, v,,, and the velocity of the micelle, v,,, as shown m Fig. 2A (6). In other words, the migration time of the analyte, tR, 1slimited between the migration time of the bulk solution, to, and of the micelle, tmc(Fig. 2B) (6).
MEKC
127
A
Micelle E;I
L inj.
Water
Solute lpil column
Water w det.
Solute
Micelle
I
1 tlllC
Bi I to
I 0
f Fl
N Time
Fig. 2. Schematicof the zone separationin MEKC (A) and chromatogram(B) (6). Reprintedwith permissionfrom Terabeet al. (1985)Anal. Chem.57,834. When an acidic solution or pH below 5.0 is employed, the electroosmotic flow becomes smaller than the electrophoretic velocity of the SDS micelle, and then the micelle migrates toward the positive electrode (13). When a catiohic surfactant, e.g., dodecyltrimethylammonium bromide, is employed instead of SDS, the direction of the electro-osmotic flow will be reversed or toward the positive electrode through the adsorption of the surfactant molecule to the inside wall of the capillary (14). 3. Chromatographic
3.1. Chromatographic
Properties
Parameters
3.1.1. Capacity Factor Capacity factor, K, can be defined in the case of conventional chromatography (6): k’ = (%lc / %J (1) where y1,, and naq are the amount of the analyte incorporated into the micelle and in the aqueous solution, respectively. Then we can obtain the relationship between the capacity factor and the migration time as: k’ = (tR- to) / [to (1 - ttJtmc)] It can be rewritten as:
tR= [(l + k’) / 1 + (t,,/t,,)k’] to
(2)
128
Otsuka and Terabe 2
0004AU
6 AL I 0
I
i. 1
12 Capacity
I
I
/‘I
5 10 2050~ Factor
1'0 Time (mln)
Fig. 3. An example of MEKC separation of the test solutes (6) (1) methanol, (2) resorcmol, (3) phenol, (4) p-mtroamlme, (5) nltrobenzene, (6) toluene, (7) 2-naphthol, (8) Sudan III. Mlcellar solution, 50 mM SDS m 100 mM borate-50 mM phosphate buffer, pH 7 0, capillary, 50 pm id x 650 mm (effective length, 500 mm); applied voltage, 15 kV; current, 33 PA; detectlon wavelength, 2 10 nm; temperature, 35°C. Reprinted with permission from Terabe et al (1985) Anal Chem 57,834. Here, the reciprocal of to/t,, or t,,ltO is a parameter representing the migration time window. If the analyte is not incorporated into the micelle or does not interact with the micelle at all, the migration time of such a solute 1s equal to t0 and hence k’ = 0. On the other hand, when the analyte 1s totally incorporated into the micelle, the migration time becomes tR, and k’ becomes infinity. Thus, the migration time window is limited between to and t,,. A typical example of MEKC separation of neutral compounds 1s shown in Fig. 3 (6). In this figure, the scale of the capacity factor is inserted to show the relationship between the migration time and capacity factor.
MEKC
When tmcis infinity or the micelle never comes out from the capillary, that condition is attained only when the absolute value of the electroosmotic velocity is identical to that of the electrophoretic velocity of the micelle in opposite directions. Equation (3) becomes: (4) This is the same situation as in conventional chromatography, that is, the parameter of the migration time window is equal to infinity. When to is infinity or the electro-osmotic flow is completely suppressed, Eq. (3) becomes: tR = (1 + l/k’)t,, (5) tR=(l
+/cyto
In this case, the aqueous phase never comes out from the capillary, and only the micelle migrates toward the positive electrode through the aqueous phase. Thus, electro-osmotic flow is not essential m MEKC. According to Eq. (2), to, tR, and t,, are required to obtain the capacity factor. As a marker of the electro-osmotic flow or to, methanol is usually used becausethe distribution coefficient of methanol between the micelle and aqueous phase is negligibly small. Although methanol is transparent to UV, it can be detected as a baseline deflection with a UV detector owing to a refractive mdex change. Sudan III or IV is often employed as a tracer of the SDS micelle (6), which is completely incorporated into the micelle. Timepidium bromide is also useful as a tracer for an anionic micelle (15). It should be noted, however, finding good tracers for to and tmc applicable to every condition is difficult. 3.1.2. Resolution
Resolution, R,, in MEKC is given as: R, = (N’“/4)
[(a-
1)/a] [Ii241 +I?;)]
{[(l -to)/&,,,]
/ [l + (to/t&‘,]}
(6)
Here, N is the theoretical plate number, a the separation factor equal to k;lk’,, and k’i and k; are the capacity factors of analytes 1 and 2, respectively. Effects of these parameters on resolution are briefly discussed below. 3.1.2.1. PLATE NUMBER Resolution increases proportionally with an increase in square root of the plate number. Usually average plate numbers for most analytes are lOO,OOO-200,000.Normally, the higher the voltage applied, the higher
Otsuka and Terabe
130
the plate number that can be attained, unless excessive Joule heating 1s generated at the higher apphed voltage. Since the diffusion coefficrent of the micelle 1s small, solutes having larger capacity factors can yield higher plate numbers. 3.1.2.2. SEPARATIONFACTOR The separation factor reflects the relative drfference of the distrrbutron coefficrents between two analytes and is a unique variable to a given separation conditron. Thus, we can manipulate the value by changing the type of micelles or the bulk solution using modifiers. 3.1.2.3. CAPACITYFACTOR The optimum value of the capacity factor IS represented as (t,,,lt0)1’2. Under neutral conditions, the optimum value is close to 2 for most long alkyl chain surfactants. The capacity factor can be related to the distribution coefficient, K, between the micelle and aqueous phase as: (7) Here, V,,,, and Vaq are the volume of the micelle and aqueous phase, respectively. The phase ratio, V,,,,/V,,, can be wrrtten by using the concentration of the surfactant, C,, and specific volume of the mtcelle, V as: (V,, / V,,) = [ v (C,,- CMC) / 1 -V (CT,,- CMC)] (f-9 Then, at low mrcellar concentratrons, Eq. (7) can be rewritten as: k’ = Kt; (C,,- CMC)
(91 This reveals that the capacity factor increases linearly wrth an increase in the surfactant concentratton. The capacity factor can be easily adjusted by manipulating the surfactant concentratron through Eq. (9) If the CMC is known. 3.1.2.4. ELECTROOSMOTICFLOW The effect of the electroosmotic flow can be discussed m terms of the migration time ratio, to/&,,., or the migration time window, t,,ltO. The velocity of the mrcelle is given as: V
mc= [ k. + cL&Nl E
(10)
where E is the electrical field strength. Then:
t0~4tlc= [ 1+ rep
/ pealE
(11)
MEKC
131
16
Fig. 4. Dependenceof the capacityfactor (k’) on the SDS concentration(csus) (6). Reprintedwith permission from Terabeet al. (1985)Anal. Chem. 57, 834. The mobilities peaand I.t&rnc) usually have different signs, and the ratio ,uep(mc)/~eois between 0 and -1. Therefore, to/t,, is smaller than 1. The smaller the value of toltmc becomes, the larger the resolution will be. If a negative sign for to/t,, is assumed when peP(mc)/ peais smaller than -1, an extremely high resolution is expected for an analyte having the capacity factor close to -(t,,ltJ according to Eq. (6), although a quite long migration time is required (13). 3.2. Thermodynamic Parameters As mentioned above, the capacity factor increases linearly with an increase in the concentration of the surfactant. The dependence of the capacity factor on the SDS concentration is shown in Fig. 4 (6). This reveals that the distribution coefficient is almost constant regardless of the SDS concentration. The distribution coefficient measured at different temperatures should follow the van’t Hoff equation: InK = - (AHOIRT) + (ASOIR) (12)
132
Otsuka and Terabe Table 1 CMC of SDS and the Partial Specific Volume (V) of the SDS Mlcelle Buffela CMCImA4
i/mLlg
TempYC
B-P
PIPES
BES
Urea
B-P
20 22 25 30 35 40 45 50
28 29 25 26 3.0
38 42 43 42 44 48
44 45 53 5.9 59 64
0 8562
31 33 3.3 35 36 38
-
0 0 0 0
8610 8686 8710 8758
Urea 0 8126 0 8160 0 8242 0.8248 0 8290
-
OB-P, 100 mMborat+50 mMphosphate buffer, pH 7 0, PIPES, 20 mMPIPES-20 mMNaOH, pH 7 0, BES, 100 mA4 BES-100 mM NaOH, pH 7 0, urea, 5M urea m 100 mA4 boratc50 mA4 phosphate buffer, pH 7 0
where AHo is the enthalpy change associated with micellar solubilization or the transfer of the solute from the aqueous phase to the micelle, Aso the corresponding entropy change, R the gas constant, and T the absolute temperature. Thus, Eq. (12) allows the calculation of the enthalpy and entropy changes in micellar solubilization from the temperature dependence of K. In various buffer systems, such as borate-phosphate (B-P), piperazine-N,N’-bis(2-ethanesulfonic acid) mono sodium salt (PIPES)sodium hydroxide, N,N’-bis(2-hydroxyethyl)-2-aminoethanesulfonic acid (BESksodium hydroxide, and B-P with urea, CMC and the partial specrtic volume were measured for the SDS mrcelle (16) as shown in Table 1. Table 2 lists distribution coefficients of test solutes at different temperatures between the SDS micelle and a B-P buffer. Corresponding enthalpy, entropy, and Gibbs free energy changes were calculated as shown m Table 3. Obviously, the effect of the buffer solution is not significant on these thermodynamic quantities. The entropy changes for resorcinol, phenol, p-nitroanilme, and 2-naphthol were negative, and the contribution of A.S” to AGO was significant except for phenol. These solutes favored the aqueous phase from the viewpoint of entropy.
MEKC
133 Table 2 Dlstrlbutlon Coefficients of Alkylphenols Between the SDS Mlcelle and a 100 mM Borat+SO mM Phosphate Buffer, pH 7.0” Temp I “C
Solute o-Cresol m-Cresol p-Cresol 2,6-Xylenol 2,3-Xylenol 3,4-Xylenol 2,4-Xylenol p-Propylphenol p-Butylphenol p-Amylphenol
30
35
40
45
100 104 112 203 233 250 269 788 2320 7120
93.0 96.5 104 187 214 230 248 729 2140 6660
86.4 89.9 96 8 173 197 212 229 668 1940 5870
81.1 84 5 90.7 161 183 197 213 622 1800 5580
‘Separation solution, 50 mM SDS m 100 mM boratc50 tiphosphate
50 76.3 79 5 85 5 150 170 183 198 571 1620 4900
buffer, pH 7 0
Table 3 Enthalpy, Entropy, and Gibbs Free Energy Changes m Mlcellar Solublhzatlon of Alkylphenols by the SDS Mlcelle” Solute o-Cresol m-Cresol p-Cresol 2,6-Xylenol 2,3-Xylenol 3,4-Xylenol 2,4-Xylenol p-Propylphenol p-Butylphenol p-Amylphenol
AH”lkJlmol -11.1 -10 8 -109 -12 4 -12.7 -12 7 -12 4 -13 1 -14.4 -150
AS”/J/mollK 16 29 33 3.2 3.3 39 55 12 3 16 9 24.3
AGO/kJ/mol, 35’C -11 6 -117 -119 -13.4 -13.7 -13 9 -14.1 -169 -19 6 -22 5
OTheseparation solution was the same as m Table 2.
4. Selectivity Manipulation Selectivity in chromatography can be discussed by using the separation factor, a. The separation factor can be manipulated with chemical considerations. The micelle in MEKC corresponds to the stationary phase in reversed-phasehigh-performance liquid chromatography (RP-HPLC),
Otsuka and Terabe whereas the bulk solution or the surrounding aqueousphase correspondsto the mobile phase, from the viewpoint of selectivity manipulation. Mainly, the following four factors can be controlled to manipulate selectivity: 1 2. 3. 4.
The mlcellar structure; Temperature, pH, and Additives to the aqueousphase. 4.1. Effect of the Micellar Structure 4.1.1. Surfactants A surfactant molecule has a hydrophobic and hydrophilic group, and both groups affect selectivity in MEKC. Since most analytes interact with the micelle on its surface, the hydrophilic or ionic group is generally more important than the hydrophobic one in determining selectivity. For example, SDS and cetyltrlmethylammomum bromide (CTAB) show considerably different selectivity as shown m Fig. 5 (17). The structural difference between SDS and CTAB is mainly ionic groups: sulfate in SDS and quaternary ammonium in CTAB. The hydrophobic groups are slmilar, only different in the length of alkyl chains. Similar results have also been published (18). In Table 4, differences in distribution coefficients of the test solutes among three surfactants, such as SDS, sodium tetradecyl sulfate (STS), and sodium dodecanesulfonate (SDDS), are shown. The distribution coefficients for resorcinol, phenol, p-nitroaniline, and nitrobenzene are virtually identical between SDS and STS, which have the identical ionic groups, but different alkyl chain lengths. On the other hand, significantly different distribution coefficients are observed between SDS and SDDS, which have the ldentlcal alkyl chains, but different ionic groups. These results suggest that many compounds are adsorbed on or at least strongly interact with the surface of the micelle. Bile salts, such as sodium cholate, sodium deoxycholate, and sodmm taurodeoxycholate, which form helical mlcelles (19), can have slgnificantly different selectivity compared with the long alkyl chain surfactants (20). Bile salts are naturally occurring chiral surfactants, and hence, they can also be used for enantiomeric separation (19,21-23). 4.1.2. Mixed Micelles In MEKC, ionic micelles are usually used, and they can be easily modlfied by adding ionic or nonionic surfactants to form mixed micelles. The
MEKC
135 A
I
1
OOOZAU
J 0
I
6
B
I
I
4
I, 0
0OWaAU
I,,
12
,
,
20
16
I 0
I
I 4
8
Time I mln
I I I 0 12 Time / min
I
I
16
Fig. 5. Effect of the surfactant structures on selectivity (17). (1) water, (2) aniline, (3) nitrobenzene, (4) m-nitroaniline, (5) p-nltroamlme, (6) o-nitroamline. Micellar solution, (A) 100 rmI4 SDS in 100 mM borate50 mM phosphate buffer, pH 7.0, (B) 50 mM CTAB in 100 mM Tns-HCl, pH 7 0; capillary, 50 pm id x 650 mm (effective length, 500 mm); applied voltage, 15 kV; detection wavelength, 230 run; temperature, ambient. Table 4 Coeffictents at 3Y’C
Distribution
Dlstrlbution Solute Resorcinol Phenol p-Nitroamline Nitrobenzene Toluene 2-Naphthol
coefficient
SDS0
STSb
SDDSC
21.6
20.8 52.3
27.7 56.1 84.3
52.1 103
135 318 656
%odlum dodecyl sulfate. “Sodium tetradecyl sulfate cSodium dodecanesulfonate
100 138
111
345 789
288 698
136
Otsuka and Terabe
mixed micelle consisting of ionic and noniomc surfactants 1s usually larger than the original iomc micelle, and has a lower charge density, and hence, a lower electrophoretic mobility. Consequently, a narrower migration time window is obtained (24), and also different selectivity is expected (24) since the surface of the mixed micelle is different from that of the original one.
4.2. Effect
of Temperature
The distribution coefficient is dependent on temperature: It decreases with an increase in temperature, which causes a reduced migration time. The increase in temperature also causes increases m v,, and vep(mc) to the same extent because of reduced viscostty. Dependencies of the distribution coefficients on temperature are different among solutes. Therefore, temperature affects selectivity as shown in Fig. 6 (26). It should be noted that temperature seriously affects the migration time, although its effect on selectivity is not remarkable, and hence, it is important to maintain temperature precisely to obtain reproducible results.
4.3. Effect 0fpH Effect of the constituents of the buffer is not significant, whereas the pH is a critical factor for ionizable analytes. If the ionized form of the solute has the same charge as the micelle, it will be incorporated into the micelle less than its neutral form. Figure 7 demonstrates the dependence of the apparent capacity factor on the buffer pH for some chlorinated phenols (2.5). Here, the apparent capacity factor was calculated by Eq. (2) regardless of whether the solutes were ionized or not. For acids, the increase in pH will promote ionization. Then the dtstribution coefficient to the anionic micelle or SDS will be reduced. It should be noted that the change of the buffer pH, especially in the lower pH region, causes a significant change in the electro-osmotic velocity as mentioned in the previous section (13).
4.4. Effect of Additives
to the Aqueous
Phase
The most versatile and effective methods to manipulate selectivity in MEKC are the use of additives to the aqueous phase as well as the choice of surfactants. In conventional HPLC, modifications with additives to the mobile phase is well established, and the knowledge can be applied in a similar manner to MEKC. Of course, we must understand the difference between the micellar phase in MEKC and the stationary phase in
MEKC
137
6
4i /-/-y-q7 l-4 31
33
T-1 , ,03-L
Fig. 6. Van’t Hoff plots of alkylphenols (16). (I) o-cresol, (2) m-cresol, (3)p-cresol, (4) 2,6-xylenol, (5) 2,3-xylenol, (6) 3,4-xylenol, (7) 2,4-xylenol, (8) p-propylphenol, (9) p-butylphenol, (10) p-amylphenol. Micellar solution, 50 mMSDS in 100 mMborate50 Wphosphate buffer, pH 7.0; apphed voltage, 10kV. Reprinted with permission from Terabe et al. (1993) J. Microcol. Sep. 5,23.
HPLC to use such additives. There are four main categories of additives that are useful in MEKC: 1. Cyclodextrins (CDs); Ion-pair reagents;
2. 3. 4.
Urea; and Organic modtfiers. 4.4.1. Cyclodextrins
Recently, CD has become a popular additive or stationary phase in chromatography. In most cases, CD’s capability of recognizing specific molecules that fit its hydrophobic cavity is used for chromatographic separations. The use of CDs is especially effective for the separation of
138
Otsuka and Terabe
1-
l
l I 6
I 6
I 7
I 9
PH
Fig. 7. Dependence of apparent capacity factors (Pap& of chlorinated phenols on pH (25). (1) phenol, (2) 2-chlorophenol, (7) 2,5dtchlorophenol, (14) 2,4,Wrtchlorophenol, (17) 2,3,4,5-tetrachlorophenol, (20) pentachlorophenol. Micellar solutton, 100 mM SDS in 50 mA4 phosphate-borate buffer; capillary, 50 ym id x 650 mm (effective length, 500 mm); applied voltage, 15 kV; detection wavelength, 220 nm; temperature, 35OC Reprinted with permission from Otsuka et al. (1985) J. Cizromatogr. 348, 39. aromatic isomers and aromatic enantiomers that have a chiral center close
to the aromatic ring. Originally, CD is electrtcally neutral and not affected by the electrophorests. This means that CD itself cannot be used as a carrier m EKC, unless an iomc group IS introduced into CD. The surface of CD is hydrophilic, and hence, we can assume that CD is not mcorporated into the mtcelle. A surfactant molecule, however, may be included into the CD cavity. The separation principle of CD-modified MEKC (CD-MEKC) is schematically shown in Fig. 8 (26). In this system, CD migrates at the same velocity as the electro-osmotic flow. The analyte molecule, which is assumed to be neutral, both included by CD and in the aqueous phase,
139
MEKC -0------_--
------------
G--
Anionic Surfactant
(
Electroosmotrc
-
Analyte
,-
Electrophoretvz Mrgratron
Flow
Cyclociextrrn
Fig. 8. Schematicof the separationprinciple of CD-MEKC (26) Reprinted wrth permission from Terabeet al (1990)J Chromatogr 516,23 migrates at the same velocity as the electro-osmotic flow. On the other hand, the analyte migrates at a different velocity from the electro-osmotrc flow when it is incorporated into the ionic micelle. In the case of highly hydrophobic analytes, they seem to be totally incorporated into the micelle in the absenceof CD. This means that the addition of CD reduces the apparent distribution coefficient of the analytes between the bulk phase and the micelle and makes rt possible to separatesuch solutes. The higher the concentration of CD becomes, the smaller the distribution coefficient will be observed. In CD-MEKC, therefore, the capacity factor can be manipulated by varying both the concentrations of CD and the micelle. An example of CD-MEKC separation of hydrophobic compounds is shown m Fig. 9 (26). CD-MEKC is also effective for enantiomerit separation, which will be discussed later. It should be noted that CD-MEKC is a different technique from CDEKC, although these two terms are somewhat confusing. In CD-MEKC, CD is added to micellar solutrons, and while in CD-EKC, an ionic CD derivative is used as a carrier of EKC in a solution without micelles. 4.4.2. Urea Urea is usually used to increase the solubility of hydrophobic compounds in water. In MEKC, a successful separation of highly lipophilic
Otsuka and Terabe
140
;o Time (mln)
Fig. 9. Separation of 11 trichlorobiphenyl isomers by CD-MEKC (26): BIPH = blphenyl. Separation solution, 60 rmI4 y-CD, 100 mM SDS, and 2M urea m 100 &borate50 &phosphate buffer, pH 8.0; capillary, 50 pm id x 650 mm (effecttve length, 500 mm); applied voltage, 15 4 kV; current 50 PA. Reprinted with permission from Terabe et al. (1990) J. Chromatogr. 516,23.
compounds was achieved with an SDS solution containing high concentration of urea, as shown m Fig. 10 (27). By adding urea to the micellar solution, the electro-osmotic velocity is slightly reduced, whereas the migration velocity of the micelle is considerably reduced, which causes the reduced capacity factors. Urea is also effective to improve peak shapes, especially m the separation of ammo acids (28). Although a remarkable change in the selectivity is not attained by the urea addition, a slight change in the selectivity can be recognized, especially for the separation of closely related compounds. 4.4.3. Organic Modifiers Similar to the case of HPLC, an organic solvent miscible with water can be used as an additive to the micellar solution to manipulate the capacity factors or selectivity. In HPLC, highly hydrophobic compounds can be analyzed by using a high concentration of the organic solvent, whereas in MEKC, a high concentration of the organic solvent cannot be employed because of the breakdown of the micellar structure. In general, the usable maximum content of the organic solvent is approx 20%. The use of the organic solvent usually provides an improved resolution and/or a change in the selectivity. In MEKC, methanol (2932),2-propanol (33), and acetonitrile (30) are used as the organic modifiers, and they contribute to reduce the electro-osmotic velocity and
MEKC
141
A
6
Time (mln)
Fig 10. The effect of urea addition to the SDS solution (27) (1) hydrocortlsone, (2) hydrocortisone acetate, (3) betamethasone, (4) cortisone acetate, (5) triamcmolone acetonlde, (6) fluocmolone, (7) dexamethasone acetate, (8) fluocinonide. Separation solution, 50 mA4 SDS In 20 mA4 borate-phosphate buffer, pH 9.0. (A) Without urea and (B) with 6M urea; capillary, 50 pm Id x 650 mm (effective length, 500 mm); applied voltage, 20 kV, detection wavelength, 210 nm. Reprmted with permission from Terabe et al. (1991) J Chromatogr 545,359. expand the migration time wmdow. An example of the use of methanol for the separation of aromatic sulfides is shown in Fig. 11 (29). Recently, Tanaka (34) has reported that the use of a fairly high concentration of methanol, e.g., 80 to almost 100% (v/v), in an SDS solution is effective for the MEKC separation of some hydrophobic compounds. In such circumstances, it is unclear whether the SDS micelle still exists in the solution, but some interactions between the solutes and the SDS molecule or micelle might occur. Imasaka and coworkers (3.5) have reported that the addition of N,Ndimethylformamide (DMF) to a bile salt micellar solution is effective for
142
Otsuka and Terabe l-
1
6
L 0 c I
I 10
1
I
30
50
Fig. 11 Separation of 11 aromatic sulfides (29) (1) benzyl methyl sulfide, (2) benzyl ethyl sulfide, (3) benzyl propyl sulfide, (4) benzyl tsopropyl sulfide, (5) methyl phenyl sulfide, (6) ethyl phenyl sulfide, (7) phenyl propyl sulfide, (8) isopropyl phenyl sulfide, (9) butyl phenyl sulfide, (10) tsobutyl phenyl sulfide, (11) s-butyl phenyl sulfide, (12) Sudan III. Separation solutton, 30 mM SDS, pH 7.0, contammg 20% (v/v) methanol, captllary, 50 urn td x 900 mm (effecttve length, 750 mm); applied voltage, 22 kV, current, 20 PA; detection wavelength, 210 nm, temperature, ambient Reprinted with permtsston from Otsuka et al. (1986) Ncppon Kagaku Kauh, 950 the separation of polyaromatrc hydrocarbons (PAHs) as shown in Fig. 12. Similarly, the use of a high concentratton of dimethyl sulfoxide (DMSO)
~IIthe SDS-MEKC systemISfound to be useful for the analysis of PAHs (36). 4.4.4.
Ion-Pau-
Reagents
In MEKC, the use of an ion-pair reagent causes a remarkable change m separation characteristics, which is mainly because of the charge of the micelle. When a tetra-alkylammonium salt is added to the SDS micellar solution, anionic analytes form paired ions with the ammonium
ion, and hence, the electrostatic repulsion between the anionic SDS micelle and the anionic analyte is reduced. That formation of the paired ion IS promoted with an increase of the concentration of the ammonium salt, that is, the higher the concentration of the ammonium salt, the larger
143
MEKC I2
I
I
I
0
4
8
Time
I
12
16
/ min
Fig. 12. Separation of PAHs by MEKC with DMF (3.5) (1) fluoranthene, (2) pyrene, (3) perylene, (4) benzo[a]pyrene, (5) 2,3-benztriphenylene, (6) dlbenz[a,h]anthracene. Separation solution, 70 mM sodium deoxycholate in Trls buffer containing 20% DMF; capillary, 50 pm rd x 600 mm (effective length, 500 mm); detection, He-Cd laser-Induced fluorescence. Reprinted wrth pernnssion from Kaneta et al.
the capacity factor of the anionic analyte. On the other hand, a cationic analyte competes with the ammonium ion in pairmg to the anionic micelle, so the migration time of the cation decreases with an increase in the concentration of the ammonium salt. The effect of the addition of tetra-alkylammonium salts to SDS micellar solutions on the selectivity is shown m Fig. 13 (1.5). The normal CZE separation, without SDS, using the buffer containing the salt is also shown. The effect strongly depends on the structure of the ion-paring reagent, e.g., the length of the alkyl chain. 4.4.5. Metal Salts Cohen et al. (37) reported the effect of the addition of metal salts to the SDS micellar solution. By adding magnesium, zinc, or copper(I1) to the
144
A
3
8 B
:I 4
5
I
Otsuka and Terabe
a
7 +
Time (mln)
,
I
r
I
0
5
10
15
Time (mm)
Time (mm)
Fig. 13. Separation of cephalosporin anttbtotlcs by (A) CZE, (B) MEKC with SDS, and (C) MEKC with SDS and tetramethylammomum salt (15). (1) C-TA, (2) ceftaztdtme, (3) cefotaxime, (4) cefmenoxrme, (5) cefoperazone, (6) cefpiramide, (7) cefptmlzole, (8) cefmmox, (9) ceftrtaxone. Separation solution, (A) 20 mM borate-phosphate buffer, pH 9.0, (B) 50 mM SDS added to (A), (C) 40 mM tetramethylammomum bromide added to (B); caprllary, 50 pm rd x 650 mm (effective length, 500 mm); applied voltage, 20 kV; detection wavelength, 2 10 nm Reprinted with permtssion from Ntshr et al. (1989) Anal. Chem 61,2434.
SDS micellar solution, the separation of oligonucleotides
was successfully achieved, and good selectivity could be obtained, as shown in Fig. 14.
5. Applications 5.1. General
Scope
Almost 10 years have passed since the first paper on MEKC (5) was published, and a number of applications of MEKC have appeared. Although separation characteristics of MEKC are similar to those of RPHPLC, the range of analytes applied to MEKC is limited compared to
MEKC
145
Fig. 14. Separation of 18 oligonucleotides, each with 18 bases, by MEKC with a metal ion (37). Separation solution, 50 mA4 SDS, 3 mM Zn(II), and 7M urea m 20 mM Tris-5 mM sodium phosphate buffer; capillary, 50 urn id x 850 mm; applied voltage, 22 kV; current, 10 PA; detection wavelength, 260 m-n;temperature, 25’C. Reprmted with permission from Cohen et al. (1987) Anal. Chem 59, 1021.
RP-HPLC: MEKC is mainly employed for the separation of small molecules, because the size of the micelle is relatively small and not able to incorporate big molecules, such as proteins, whereas RP-HPLC can treat such molecules as analytes. Regardless of such a limitation, MEKC has been recognized as a useful and powerful technique in various analytical fields, because of its advantages over RP-HPLC. The main advantage of MEKC is the higher separation efficiency. Nishi and Terabe (38) have shown some such examples, especially in the pharmaceutical analyses. Other advantages of MEKC over RP-HPLC are as follows: 1. MEKC analysis can be carried out with smaller amounts of sample and separation solutions. 2. Separation usually can be completed within a shorter time.
146
! 0
Otsuka and Terabe
/ 10
20
Time
/
mln
10
20 Time
30 / mln
Fig. 15. Separation of PTH-ammo acids by SDS-MEKC (A) wtthout urea (14) and (B) wtth urea (27) The peaks are labeled with one-letter abbrevtattons for amino acid. Separation solutton, (A) 50 mM SDS, pH 7.0, (B) 100 mM SDS contammg 4 3M urea, capillary, (A) 50 pm td x 650 mm (effecttve length, 500 mm), (B) 52 urn td x 500 mm (effecttve length, 300 mm); applred voltage, (A) 10 kV, (B) 10.5 kV; detectton wavelength, (A) 260 nm, (B) 220 nm, temperature, (A) 35’C, (B) ambient Reprinted with permtssions from (A) Otsuka et al. (1985) J Chromatogr 312,219 and (B) Terabe et al (1991) J Chromatogr
545,359
3. Maintenance of the separation capillary, e.g., cleaning or replacmg, can be easily operated. A typical application of MEKC is the separation of closely related compounds. A mixture of phenylthrohydantoin amino acids (PTH-AAs) was successfully separated by using an SDS, as shown m Fig. 15A, and a dodecyltrimethylamtnonium bromide (DTAB) solution (24). By adding urea to the SDS mrcellar solution, better resolution and selectivity could be obtained, as shown in Ftg. 15B (27). Separation of all isomers of chlorinated phenols, including phenol, could also be achieved with an SDS solution as shown in Fig. 16 (25). These separations cannot be carried
out by a simple Isocratic HPLC, i.e., a gradient method is required. The overall discussions on the MEKC applications are available in the review by Janmi and Issaq (9).
147
MEKC
14 11 7 I
h.li I
5
10 Ttme
4
I
15 lmm
Fig. 16. Separation of chlormated phenols by MEKC (25) Phenols, (1) phenol, (2) 2-chloro, (3) 3-chloro-, (4) 4-chloro-, (5) 2,3-dtchloro-, (6) 2,4-dtchloro-, (7) 2,5-drchloro-, (8) 2,6-drchloro- (9) 3,4-dichloro-, (10) 3,5-dtchloro-, (11) 2,3,4-trichloro-, (12) 2,3,5-trichloro-, (13) 2,3,6-trtchloro-, (14) 2,4,5-trrchloro-, (15) 2,4,6-trichloro-, (16) 3,4,5-trrchloro-, (17) 2,3,4,5-tetrachloro-, (18) 2,3,4,6-tetrachloro-, (19) 2,3,5,6-tetrachloro-, (20) pentachloro-, separation solutton, 70 mJ4 SDS, pH 7.0; capillary, 50 pm td x 650 mm (effecttve length, 500 mm); applied voltage, 10 kV, current, 17 PA; detection wavelength, 220 nm; temperature, 35°C. Reprinted with permlssron from Otsuka et al. (1985) J Chromatogr. 348,39.
5.2. Optical Resolution Recently, optical resolution has become one of the major objectives m chromatographic separation, and many papers on optical resolution by HPLC have appeared. A number of reports on chiral separations by MEKC have also been published at the present stage. In MEKC, the following two methods are usually employed to achieve optical resolution: (1) MEKC with chiral surfactants and (2) cyclodextrin-modified MEKC (CD-MEKC). Brief reviews on chiral separations by MEKC and also by CE have been pubhshed previously (39-41). A more extensive treatment is given in Chapter 14. 52.1. MEKC with Chiral
Surfactants
Chiral surfactants mainly used in MEKC for optical resolution are as follows: Amino acid derivatives, such as sodium Wdodecanoyl-L-
148
Otsuka and Terabe
I
,
0
20 Time
I 40 / mln
Frg 17. Chual separatton of six PTH-m-ammo actds by MEKC wrth SDVal (45): Corresponding AAs: (1) Ser, (2) Aba, (3) Nva, (4) Val, (5) Trp, (6) Nle. (0) Acetomtrile. Micellar solution, 50 nuI4 SDVal-30 mM SD%-O.SM urea, pH 9.0, containing 10% (v/v) methanol; capillary, 50 ym td x 650 mm (effective length, 500 mm); applied voltage, 20 kV; current, 17 PA; detection wavelength, 260 nm, temperature, ambient. Reprmted with perrmsston from Otsuka et al. (1991) J Chromatogr. 559,209.
(Z&42-45) and sodium N-dodecanoyl-L-glutamate (SDGlu) (46), which were effective in resolution of PTH-DL-AAS, as shown in Fig. 17 (45). Digitonin, which is a glycoside of digitogenin, could achieve the optical resolution of some dansylated m-amino acids (Dns-m-AAs), used as the mixed micelle with SDS or bile salts (4446). Bile salts are useful to chnal separations, as mentioned previously. By using sodium taurocholate (STC) and sodium taurodeoxycholate (STDC), some Dns-DL-AAs were optically resolved (21). Some chiral drugs, e.g., diltiazem hydrochloride and trimetoquinol hydrochloride, have also been valinate
(SDVal)
149
MEKC
resolved (22,23,4 7). Enantiomeric separation of binaphthyl analogs by MEKC with bile salts was reported by Cole et al. (48). As other chiral surfactants, saponins, such as glycyrrhizic acid and p-escin, have been used for optical resolution of Dns- or PTHDL-AAs (49). 5.2.2. Cyclodextrin-Modified
MEKC (CD-MEKC)
As mentioned previously, CD-MEKC is capable of optical resolution, especially of aromatic and related enantiomers. Some Dns-oL-AAs were optically resolved by CD-MEKC using SDS solutions containing p- or ?/-CD (50). Not only the underivatized CDs but also some CD derivatives, e.g., 2,6-di-O-methyl-P-CD, in SDS solutions can be used for the resolution of the optical isomers (51). Recently, the CD-MEKC system has become one of the most popular techniques for chiral separations in HPCE: Optical resolution of some labeled amino acid enantiomers (52) and RS-chlorpheniramine (53) has been reported. It should be noted that the CD-added CE system without micelles is usually more effective than CD-MEKC for chiral separation of ionic compounds, especially for the analyte having a high electrophoretic mobility, and CD-MEKC and CD-added CE are complementary techniques to each other. 6. Conclusion As mentioned previously, many papers on MEKC, which include fundamental characteristics and applications, are available at this present stage. Since only brief discussion on some aspects of MEKC are described in this chapter, it is necessary to refer to some of that literature where more detailed information is available. Especially for optimization of MEKC, which was not discussed in this chapter, theoretical discussions by Foley (54), Vindevogel and Sandra (55), and Khaledi and coworkers (56,57) should be cited, along with the review article (II). There are some EKC techniques other than MEKC: Cyclodextrin EKC (CDEKC) (58), ion-exchange EKC (IXEKC) (59), and microemulsion EKC (MEEKC) (60,61). In CDEKC, a cyclodextrin derivative having an ionic function is used instead of the micelles in MEKC. Similarly, polymer ions and microemulsions are used in IXEKC and MEEKC, respectively.
150
Otsuka and Terabe
Electrokinetic chromatography, which is indeed a branch of HPCE, has become the normal technique for high-resolution separation of neutral species by electrophoresis, and will be used in wider fields in the future. Notes Added in Proof Recently, mass spectrometry (MS) has become one of the powerful detection schemes in CE. The development of the MEKC-MS system, however, did not successfully progress, since most surfactants normally used in MEKC often deteriorate ionization efficiency and cause high background signals in an electrospray ionization-MS (ESI-MS). One solution to these limitations is to use high-mol-wt surfactants, instead of a conventional one, such as SDS, as pseudostationary phases in MEKC Terabe and coworkers (62,63) used butyl acrylate-butyl methacrylatemethacryhc acid copolymer sodium salts (BBMA), and investigated their use in MEKC-MS systems. In an ESI-MS system, BBMA was successfully used for the separation and detection of some quaternary ammomum salts, alkaloids, and sulfamids (64). In an atmospheric pressure chemical ionization-MS (APCI-MS) system, some xanthine derrvatives were able to be separated and detected (65). The use of high-mol-wt surfactants in MEKC is also useful for the alteration of the selectivity and theoretical treatments on separation characteristics becausethe high-molwt surfactant forms the molecular micelle or the micelle with a single molecule. As mentioned m Section 4.4.3., separation of hydrophobic compounds, such as polycyclic aromatic hydrocarbons (PAHs), by MEKC is still in progress. Some PAHs were successfully separated with a CD-MEKC mode, i.e., using and y-CD-SDS solution (66), and also by MEKC with an SDS-acetone solution (67). As new pseudostationary phases in EKC other than micelles, starburst dendrimers were introduced by Tanaka et al. (34,68,69). They showed remarkably different selectivity from that m the SDS-MEKC system and could be used for the separation of various PAHs. Other techniques for the separation of hydrophobic compounds, nonaqueous CE (70), and hydrophobic interaction electrokinetic chromatography (HI-EKC) (71) were demonstrated. References 1. Mikkers, F E. P , Everaerts, F M , and Verheggen, Th. P. E. M. (1979) Highperformance zone electrophorem J Chromatogr 169, 1 l-20.
MEKC
151
2 Jorgenson, J W and Lukacs, K D. (198 1) Zone electrophoresis in open-tubular glass capillaries. Anal. Chem 53, 1298-l 302 3 HJerten, S. (1983) High-performance electrophoresis: the electrophorettc counterpart of high-performance liquid chromatography J Chromazogr 270, l-6. 4 Terabe, S (1989) Electrokmetic chromatography: an interface between electrophoresis and chromatography. Trends Anal. Chem. 8, 129-134 5. Terabe, S , Otsuka, K , Ichikawa, K., Tsuchiya, A., and Ando, T. (1984) Electrokinetic separations with mtcellar soluttons and open-tubular capillaries. Anal. Chem 56, 111-113 6 Terabe, S , Otsuka, K., and Ando, T. (1985) Electrokmetic chromatography with mxellar solution and open-tubular capillary. Anal Chem 57,834841 7 Terabe, S , Otsuka, K., and Ando, T. (1989) Band broadenmg in electrokmetic chromatography wtth micellar solutions and open-tubular captllaries. Anal Chem 61,25 l-260 8 Kuhr, W G. and Monnig, C A. (1992) Capillary electrophoresis Anal Chem. 64, 389R-407R 9. Janini, G. M. and Issaq, H J (1992) Micellar electrokinettc capillary chromatography: basic considerations and current trends. J Lzquzd Chromatogr 15,927-960 10. Vindevogel, J and Sandra, P (1992) Introduction to Mzcellar Electrobnetcc Chromatography Huthrg, Heidelberg 11 Terabe, S (1992) Mzceilar Electrokinetzc Chromatography Beckman, Cahforma 12 Terabe, S (1993) Micellar electrokmetic chromatography, m Capillary Eiectrophoreszs. Theory, Methodology, and Appbcatlons (Guzman, N A , ed ), Marcel Dekker, New York, pp. 65-87 13 Otsuka, K and Terabe, S. (1989) Effects of pH on electrokinetic velocities in micellar electrokinettc chromatography J Mzcrocol Sep 1, 150-154 14. Otsuka, K , Terabe, S , and Ando, T. (1985) Electrokmettc chromatography with micellar solutions* separation of phenylthiohydantoin-ammo acids J Chromatogr 312,219-226 15 Nlshi, H., Tsumagari, N , and Terabe, S (1989) Effect of tetraalkylammonmm salts on mtcellar electrokinettc chromatography of ionic substances Anal Chem 61,2434-2439. 16 Terabe, S., Katsura, T., Okada, Y., Ishihama, Y., and Otsuka, K. (1993) Measurement of thermodynamic quantities of micellar solubihzation by micellar electrokinetic chromatography wnh sodium dodecyl sulfate. J Mcrocol Sep 5,23-33. 17 Ishthama, Y. and Terabe, S , unpublished data 18. Nrshi, H., Fukuyama, T., Matsuo, M., and Terabe, S. (1990) Effect of surfactant structures on the separatton of cold medtcme Ingredients by micellar electrokinettc chromatography J Pharmaceut Sci 79,519523 19. Cole, R 0 , Sepamak, M J., Hinze, W. C., Gorse, J., and Oldiges, K. (1991) Bile salt surfactants m micellar electrokmetx capillary chromatography. apphcation to hydrophobic molecule separatrons. J. Chromatogr 557, 113-I 23. 20. Ntshr, H., Fukuyama, T., Matsuo, M., and Terabe, S. (1990) Separation and determination of lipophilic cortxosteroids and benzothtazepm analogues by mxellar electrokinetic chromatography using bile salts. J. Chromatogr. 513,279295.
152
Otsuka and Terabe
21 Terabe, S., Shlbata, M , and Mlyashlta, Y. (1989) Chiral separation by electroklnetlc chromatography with bile salt mlcelles J Chromatogr 480,403-4 11 22 Nlshl, H , Fukuyama, T., Matsuo, M., and Terabe, S. (1990) Chrral separation of dlltlazem, trlmetoqumol and related compounds by mlcellar electrokmetlc chromatography J Chromatogr 515,233-243 23 Nlshi, H , Fukuyama, T , Matsuo, M., and Terabe, S (1990) Chiral separation of trlmetoqumol hydrochloride and related compounds by mlcellar electrokmetlc chromatography using sodium taurodeoxycholate solutions and application to optical purity determination Anal Chum Acta 236, 281-286. 24 Rasmussen, H. T., Goebel, L K., and McNalr, H. M (1991) Optimization of resolution m mlcellar electrokmetlc chromatography J Hzgh Resolut Chromatogr 14,25-28
25. Otsuka, K , Terabe, S., and Ando, T (1985) Electrokmetlc chromatography with mlcellar solutions retention behavlour and separation of chlorinated phenols J Chromatogr 348,39--47. 26 Terabe, S , Miyashita, Y , Shlbata, 0 , Barnhart, E R., Alexander, L. R , Patterson, D G , Karger, B. L , Hosoya, K., and Tanaka, N (1990) Separation of highly hydrophobic compounds by cyclodextrm-modified mlcellar electrokinetlc chromatography. J. Chromatogr 516,23-3 1 27 Terabe, S , Ishlhama, Y., Nlshi, H , Fukuyama, T , and Otsuka, K (1991) Effect of urea addition m mlcellar electrokmetlc chromatography. J Chromatogr 545,359-368 28 Otsuka, K and Terabe, S. (1990) Effects of methanol and urea on optical resolution of phenylthiohydantom-m-ammo acids by micellar electrokinetlc chromatography with sodium N-dodecanoyl+valinate Electrophoresis 11,982-984 29 Otsuka, K., Terabe, S., and Ando, T. (1986) Separation of aromatic sulfides by electrokinetic chromatography with micellar solution (m Japanese). Nippon Kagaku Kalshl, 950-955 30 Gorse, J., Balchunas, A. T., Swaile, D F , and Sepamak, M J (1988) Effects of organic mobile phase modifiers in mlcellar electrokmetlc capillary chromatography. J Hugh Resolut Chromatogr Chromatogr. Commun. 11,554-559. 3 1 Bushey, M M and Jorgenson, J. W. (1989) Separation of dansylated methylamme and dansylated methyl-d3-amme by mlcellar electrokmetlc capillary chromatography with methanol-modified phase. Anal Chem 61,491-493 32 Bushey, M. M. and Jorgenson, J W (1989) Effects of methanol-modified mobile phase on the separation of isotoplcally substituted compounds by micellar electrokinetic capillary chromatography. J. Mlcrocol. Sep 1, 125-130. 33. Balchunas, A. T and Sepamak, M J. (1988) Gradient elutlon for mlcellar electrokinetic capillary chromatography. Anal. Chem 60,6 17-62 1. 34. Tanaka, N., Fukutome, T., Tanigawa, T., Hosoya, K., Klmata, K., Araki, T , and Unger, K. K. (1995) Structural selectivity provided by starburst dendrlmers as pseudostationary phase in EKC J Chromatogr A. 699,33 l-341 35 Kaneta, T., Yamashita, T., and Imasaka, T. (1992) Separation of polyaromatlc hydrocarbons by laser-induced fluorescence detection--mlcellar electrokmetlc chromatography, Abstracts ofPapers, 12th Symposmm on Capillary Electrophoresis, Himeji, Hyogo, Japan, Abstract 8.
MEKC
153
36 Otsuka, K , Koike, R., and Terabe, S. (1993) Effect of addmon of organic modlfiers in mrcellar electrokinetic chromatography, Abstracts of Papers, 65th Meetmg of the Chemrcal Socrety of Japan, Tokyo, Japan; Abstract 3A252. 37. Cohen, A. S , Terabe, S., Smith, J. A., and Karger, B. L. (1987) High-performance capillary electrophoretrc separatron of bases, nucleosides, and olrgonucleotrdes retentron manipulation vta mtcellar solutions and metal additives. Anal Chem 59, 1021-1027. 38 Nishi, H and Terabe, S. (1990) Applications of micellar electrokmetic chromatography to pharmaceutrcal analysis Electrophoreszs 11, 69 l-70 1 39 Otsuka, K. and Terabe, S (1993) Enantromerrc separation by mrcellar electrokrnetrc chromatography. Trends Anal, Chem. 12, 125-l 30 40 Otsuka, K. and Terabe, S (1993) Choral separation by capillary electrophoresis and electrokmetic chromatography, in Capillary Electrophoresis Theory, Methodology, and Appllcatlons (Guzman, N. A., ed.), Marcel Dekker, New York, pp 6 17-629 41. Terabe, S , Otsuka, K., and Ntshi, H. (1994) Separation of enanttomers by caprllary electrophoretrc techniques J Chromatogr A 666,295-3 19 42. Dobashi, A , Ono, T , Hara, S , and Yamaguchr, J. (1989) Optical resolution of enantiomers with chiral mixed micelles by electrokmetic chromatography Anal Chem 61,1984-1986. 43. Dobashi, A., Ono, T., Ham, S., and Yamaguchr, J (1989) Enantroselectrve hydrophobic entanglement of enantiomertc solutes wtth chnal functionahzed micelles by electrokinetic chromatography J Chromatogr. 480,413420 44. Otsuka, K and Terabe, S. (1990) Enantromeric resolution by mrcellar electroklnetrc chromatography with chual surfactants J Chromatogr 515,221-226. 45 Otsuka, K., Kawahara, J , Tatekawa, K , and Terabe, S. (1991) Chual separatrons by mrcellar electrokmetrc chromatography with sodium N-dodecanoyl+-vahnate. J. Chromatogr 559,209-214 46. Otsuka, K., Kashihara, M., Kawaguchr, Y., Koike, R., Huamttsu, T., and Terabe, S. (1993) Optical resolution by high performance capillary electrophoresrs. mrcellar electrokinetic chromatography with sodium N-dodecanoyl+-glutamate and drgitonin. J. Chromatogr A 652,253-257. 47. Nishr, H., Fukuyama, T , Matsuo, M., and Terabe, S. (1989) Chnal separatron of optical isomeric drugs usmg mrcellar electrokmetic chromatography and bile salts. J Microcol Sep 1,234-241 48. Cole, R. O., Sepamak, M. J , and Hinze, W. L. (1990) Optimizatron of bmaphtyl enantiomer separations by CZE using mobile phases contammg bile salts and organic solvent. J. Hugh Resolut. Chromatogr. 13,579-582 49. Ishihama, Y. and Terabe, S. (1993) Enantiomerrc separation by micellar electrokrnetrc chromatography using saponins J. Llquld Chromatogr. 16,933-944. 50. Mryashita, Y. and Terabe, S. (1990) Separation of dansyl ot.-amino acids by micellar electrokmetrc caprllary chromatography wrth and without cyclodextrms using P/ACE system 2000. Appkcatlon Data, Hugh Performance Capillary ElectrophoreSIS, Beckman, DS-767 5 1. Nrshr, H., Fukuyama, T., and Terabe, S (1991) Chiral separation by cyclodextrmmodified micellar electrokmetrc chromatography. J Chromatogr 553, 503-5 16
154
Otsuka and Terabe
52 Ueda, T , Krtamura, F , Mttchell, R , Metcalf, T., Kuwana, T , and Nakamoto, A (1991) Choral separation of naphthalene-2,3-dtcarboxaldehyde-labeled ammo acid enanttomers by cyclodextrm-modified mtcellar electrokmettc chromatography with laser-Induced fluorescence detection. Anal Chem 63,2979-298 1 53 Otsuka, K and Terabe, S (1993) Optical resolution of chlorphemramme by cyclodextrm added captllary zone electrophorests and cyclodextrm modttied mrcellar electrokmettc chromatography. J Lzquzd Chromatogr 16, 945-953 54 Foley, J P (1990) Optimization of mtcellar electrokmetic chromatography Anal Chem 62, 1302-1308 55 Vmdevogel, J. and Sandra, P (199 1) Resolutton opttmrzatton m mtcellar electrokinetic chromatography use of Plackett-Burman stattsttcal design for the analysis of testosterone ester Anal Chem 63, 1530-1536 56 Strasters, J K and Khaledi, M G (1991) Mtgration behavior of catiomc solutes in micellar electroktnetic caprllary chromatography Anal Chem 63, 2503-2508 57 Khaledt, M. G , Smith, S C., and Strasters, J. K (1991) Micellar electrokmetic capillary chromatography of acidic solutes: migration behavior and opttmtzation strategies Anal Chem 63, 1820-1830 58 Terabe, S , Ozakt, H , Otsuka, K , and Ando, T (1985) Electrokmetic chromatography with 2-O-carboxymethyl-P-cyclodextrm as a movmg “statronary” phase J Chromatogr 332,2 11-217 59 Terabe, S. and Isemura, T. (1990) Ion-exchange electrokmettc chromatography with polymer tons for the separation of tsomerrc ions having identical electrophoretic mobthttes Anal Chem 62,65&652 60 Watarar, H. (199 1) Mtcroemulston capillary electrophorests Chem Lett 39 1-394. 61. Terabe, S , Matsubara, N., Ishthama, Y , and Okada, Y (1992) Mtcroemulston electrokmettc chromatography* comparison with micellar electrokmettc chromatography J Chromatogr 608,23-29 62. Terabe, S., Ozakt, H., and Tanaka, Y. (1994) New pseudo-stationary phases for electrokmehc chromatography* a htgh-molecular surfactant and proteins J Chw Chem Sot 41,251-257. 63 Ozaki, H , Ichthara, A , and Terabe, S (1994) Mrcellar electrokinetic chromatography using high-molecular surfactants. use of butyl acrylate/butyl methacrylate/ methacrylic acid copolymers sodium salts as pseudo-stationary phases J Chromatogr A 680, 117-123. 64 Ozaki, H , Ito, N , Terabe, S , Takada, Y., Sakatrt, M , and Koizuml, H (1995) Micellar electrokmetic chromatography-mass spectrometry (MEKC-MS) using a high-molecular surfactant. on-line couplmg with an electrospray tomzatton mterface J Chromatogr A , in press 65. Ozaki, H., Ito, N., Terabe, S., Takada, Y ., Sakatrt, M., and Koizuml, H. (1995) MEKC-MS using high-molecular surfactants. Abstracts ofPapers, 56th Forum of Analytical Chemtstry, Osaka, Japan, Abstract 1DO 1. 66. Terabe, S., Mtyashita, Y., Ishthama, Y., and Shtbata, 0 (1993) Cyclodextrin-moditied mtcellar electrokmetlc chromatography. separation of hydrophobic and enantiomertc compounds J Chromatogr 636,47-55.
MEKC
155
67 Otsuka, K , Hrgashimori, M , Korke, R., Karuhaka, K , Okada, Y , and Terabe, S. (1994) Separation of hpophthc compounds by mtcellar electrokmetrc chromatography wrth orgamc modttiers Electrophoreszs 15, 128CLl283 68. Tanaka, N , Tanigawa, T , Hosoya, K , Krmata, K , Arakr, T , and Terabe, S ( 1992) Starburst dendrimers as carriers in electrokinetrc chromatography Chem Lett 959-962 69 Tanaka, N., Fukutome, T , Hosoya, K., Ktmata, K., and Arakr, T (1995) Polymersupported pseudostatronary phase for EKC. EKC m a full range of methanol-water mixtures with alkylated starburst dendrimers. J Chromatogr , m press 70. Sahota, R. S and Khaledi, M G (1994) Nonaqueous capillary electrophorests AnaE Chem 66, 1141-l 146 7 1 Ahuja, E S and Foley, J P (1994) Separatron of very hydrophobrc compounds by hydrophobic mteractron electrokmetic chromatography J Chromatogr A 680, 73-83.
CHAPTER13 Capillary
Gel Electrophoresis And&s
Guttman
1. Introduction There is a great deal of interest in analytical biochemistry m the separation and identification of biologically important polymers, such as DNA protein and complex carbohydrate molecules (1,2). For relatively short single-stranded DNA units (i.e., oligonucleotides) and carbohydrate molecules, there is a need to separate by a single base difference (for DNA sequencing) (3) or even for identical chain length with a different sequence (identification of primers, probes, and antisense DNA molecules) (3,4). For the double-stranded DNA molecules, there is an interest to analyze and identify DNA molecules in the form of restriction fragments or PCR products. Using various types of sieving media allows us to do these kinds of separations. In capillary gel electrophoresis, crosslinked or noncrosslinked sieving matrices can be employed (5-7). The crosslinked gels, i.e., chemical gels, have a well-defined pore size. Noncrosslinked, or so-called physical gels, have a dynamic pore structure. This major difference provides the noncrosslinked linear polymer networks with much higher flexibility when compared to the crosslinked gels. One can operate at high temperatures (up to 5&7O”C) while applying extremely high field strengths (up to lo3 V/cm range) without any damage to the linear polymer network formulations (8’. It is important to note that the crosslinked gels are not usable under such extreme conditions (9). The other main advantage of the linear polymer network system is that it can be easily replaced in the capillary column by simply From
Methods Edrted by
m Molecular Biology, Vol 52 Caprllary Electrophoresrs K Altna Copynght Humana Press Inc , Totowa, NJ
157
Guttman
158
rinsing the gel matrix through the capillary by pressure or vacuum. Therefore, if the column becomes contaminated, the gel is easily replaced extending the lifetime of the system. Employing the replaceable concept, there is a possibility of the use of pressure injection compared to the crosslinked gels where electrokinetic injection mode is the only possibility (10). It is important to note that in addition to convenience, pressure injection permits quantitative analysis. 2. Materials 2.1. Apparatus
In all these studies, the power supply of the capillary electrophoresis apparatus (homemade or commercial) was used m reversed polarity mode, with the cathode on the injection side and the anode on the detection side. The separations were monitored on-column at 214 nm for the protein and carbohydrate, and at 254 nm for the DNA and the dansylated amino acid samples. The temperature of the gel-filled capillary columns was maintained in all experiments at 20°C * 0.5”C even at high field strengths by the Peltier device controlled cooling system (II). The electropherograms were acquired and stored on an Everex 386/33 computer using the System GoldTM software package (Beckman Instruments, Inc., Fullerton, CA). 2.2. Chemicals
The crude 70-mer and the slab-gel-purified 99-mer oligonucleotides were the gift of N. Bischoffer (Genentech, South San Francisco, CA). The human K-ras oncogenes (dGTTGGAGCT-C-GTGGCGTAG, dGTTGGAGCT-G-GTGGCGTAG, dGTTGGAGCT-T-GTGGCGTAG) were purchased from Pharmacia (Piscataway, NJ). The DNA restriction fragment mixture, @Xl 74 DNA-HaeIII digest, was purchased from New England Biolabs (Beverly, MA). The 102-mer was synthetized in-house. All the DNA samples were diluted to 50 pg/mL with water before injection. Ultrapure electrophoresis grade acrylamide, Tris, boric acid, EDTA, urea, ammonium persulfate, and TEMED were employed in the experiments (Schwartz/Mann Biotech, Cambridge, MA). Orange G (Sigma, St. Louis, MO) was used in the electrophoretic separations as an internal standard at 0.0 1% concentration. The dansylated n,L-amino acids (Dns-or,-AA) and the SDS protein test mixture (14,400-97,400 Dalton) were purchased from Sigma. ANTS-
Capillary
Gel Electrophoresis
159
labeled maltooligosaccharioles were from Glyko (Novato, CA). Before injection, the samples were diluted to 0.2-2 mg/mL with the eCAP SDS-200 sample buffer and were boiled m a water bath for 5 min after adding 2.5% P-mercaptoethanol as reducing agent and 0.0 1% Orange-G as the internal standard. All the DNA, amino acid, protein and carbohydrate samples were stored at -2OOC or freshly used. The buffer and gel solutions were filtered through a 1.2~pm pore size filter (Schleicher and Schuell, Keene, NH) and carefully vacuum degassed. 3. Methods 3.1. Denaturing Gels for Single-Stranded OZigonucleotide Separations and DNA Sequencing For the separation of single- and double-stranded DNA molecules, the polymerization of the different concentration (high- and low-viscosity) linear noncrosslinked polyacrylamides was accomplished in fused silica capillary tubing (Polymicro Technologies, Inc., Phoenix, AZ) m 100 mM Tris-boric acid, 2 m44 EDTA, pH 8.5, buffer (7). For stabilization, the high-viscosity linear polyacrylamide gel (6-l 2% acrylamide) was covalently bound to the wall of the column by means of a bifunctronal agent, (methacryloxypropyl)-trimethoxysilane (Petrarch Systems, Bristol, PA) (12). The polymerization was initiated by the amount of ammonium persulfate (2-4 pL, 10%) and catalyzed by tetramethylethylenediamine (TEMED) (2-4 FL), which caused full polymerization of the given percentage acrylamide solution within 20-40 min at room temperature. 3.2. Nondenaturing Gels for Single- and Double-Stranded DNA Separations In a second column type, the use of low-viscosity linear polyacrylamide (3-6% acrylamide) without binding to the capillary wall permits replacement of the gel-buffer system in the capillary column by means of the rinse operation mode of the P/ACE apparatus (i.e., replaceable gel). The samples were injected either electrokinetically (typically: 0.0 15-O.15 W) into the prepacked column or by pressure (typically: 5 s, 0.5 psi) into the replaceable polyacrylamide gel-filled capillary column. The denaturing polyacrylamide gel column contained 7M urea.
160
Guttman 3.3. ChiraZ
Gels for Capillary
EZectrophoresis
For the enantiomeric separation of Dansylated m-amino acids, the polyacrylamide gel contained specific chiral selectors, such as p-cyclodextrin. The 5% acrylamide, 0.17% N,l\r-methylene-bisacrylamide (BIS) gel was polymerized in the presence of the chiral agent. The buffer composition of that was: 0. IA4 Tris, 0.25M boric acid, pH 8.3, 7M urea, and 75 rnM P-cyclodextrin. 3.4. SDS-Protein
and Carbohydrate
Columns
In all the SDS-protein capillary electrophoresis experiments, the eCAPTM SDS-200 (Beckman Instruments, Inc., Fullerton, CA) capillary electrophoresrs size separation kit was used. For the carbohydrate separations, the eCAP Neutral coated capillary was used with 25 Macetate, pH 4.75, buffer containing appropriate amount of sieving polymer. In both instances, the 47-cm long (40 cm to the detector) and 0.1 -mm id coated capillary columns were washed with 1NHCl before each run. The sieving matrices in both cases were low-viscosity gel formulations that were not bound to the capillary wall. This permits replacement of the gel-buffer system in this coated capillary column by means of the pressure rinse operation mode of the P/ACE apparatus (i.e., replaceable gel). The samples were injected by pressure (typically 30-60 s, 0.5 psi) into the replaceable gel-filled capillary column. 3.5. Specific Applications 3.5.1. Denaturing Capillary Electrophoresis Separation of Single-Stranded Oligonucleotides
Denaturing polyacrylamide gel-filled capillary columns are utilized mainly in size separations of relatively short (up to several hundreds of bases) single-stranded DNA molecules. The most commonly used denaturing agents are urea and formamide. Figure 1 shows an example of using a denaturing polyacrylamide gel-filled capillary column containing 7M urea for the purity check of a synthesized single-stranded oligodeoxyribonucleotide (70- and 99-mer). The electropherogram shows full separation of all the failure sequences in addition to the peak of interest, the 70-mer, and the effectiveness of slab-gel purification of the 99-mer. This figure illustrates the very high resolving power of this technique, and the usefulness of the ability of fast separation and identification of single-stranded DNA molecules in molecular biology. These gels also
Capillary
161
Gel Electrophoresis 70-
B
mer
99-mer
crude
purlfled
Fig. 1. Denaturing capillary gel electrophoresis separation of synthettzed oligomers. (A) Crude 70-mer. (B) Slab-gel-purtfied 99-mer on a 5% acrylamide (crosslinked with 0.17% BIS) gel columns (see ref. 5). have micropreparative capability (13). It is possible to collect any of the peaks in Fig. 1A in sufficient concentration and purity to perform mlcrosequencing for identification of possible failures of the oligonucleotide synthesis. Several research groups are currently exploring the use of CE as an alternative to slab-gel electrophoresis for automated DNA sequence
162
Guttman
determination (14-Z 6). The large surface-area-to-volume ratio of the capillary permits higher electric fields than are used typically with slab gels (because of more efficient heat dissipation), resulting in very rapid and efficient separation of sequencing reaction products. The capillary format is readily adaptable to automated sample loading and on-line data collection. With CE, detection of separated DNA sequencing fragments is performed by laser-induced fluorescence (LIF) detection The sensitivity of the LIF method allows sequencing reactions to be performed on the same template and reagent scale as that of manual DNA sequencing with autoradiographic detection. The identity of the terminal base of each DNA sequencing fragment is encoded in the wavelength and/or the intensity of the fluorescent emission. 3.5.2. Nondenaturzng Capillary Gel Electrophoresis Separatbons of Single- and Double-Stranded DNA Molecules Nondenaturmg polyacrylamide gels are utilized when separation is based on the size and charge of the biopolymer. It is important to note that secondary structure differences can also be recogmzed utilizmg this kind of gel matrix. Figure 2 shows a nondenaturing polyacrylamide capillary gel electrophoretic separation of three 19-mers (human K-ras oncogene probes). Orange-G was used as internal standard in the separation in order to increase precision of migration time measurement (17). The three 19-mers in Fig. 2 differ only in a single change in sequence in the middle (position 10) of the molecules. This difference causes some secondary structure difference in the molecules. It is believed that these columns are able to separate these species based on that change. The same type of nondenaturing gels can be used in lower concentration for the separation of double-stranded DNA molecules, such as restriction fragments. Figure 3A shows a separation of a $X174 DNA HaeIII digest restriction fragment mixture by using replaceable polyacrylamide gel matrix. It is important to point out the baseline separation of the two closest size fragments, the 27 1 and 281 bp. This pair had proven difficult to separate by other types of sieving buffer systems, such as cellulose derivatives (IS). The separation of double-stranded DNA molecules can be improved by using special additives. The effect of an intercalator additrve, ethrdium bromide, on the separation of the previous test mixture can be seen in Fig. 3B. Ethidium bromide intercalates
Capillary
Gel Electrophoresis
163
)G
Fig. 2. Nondenaturmg capillary polyacrylamlde gel electrophoresls separation of a human K-ras oncogene probe mixture (19-mew). Peaks: 1 = dGTTGGAGCT-G-GTGGCGTAG, 2 = dGTTGGAGCTC-GTGGCGTAG, 3 = dGTTGGAGCT-T-GTGGCGTAG (see ref 14)
between the two strands of the DNA double helix. Since it is oppositely charged, it reduces the migration times of all the fragments. Because of this particular complexation phenomenon (one ethidium bromide molecule/5 bp) and the increasing rigidity of the complex, the larger DNA molecules move more slowly. Therefore, the separation time window opens up so the peak capacity is increased. Ethidium bromide is quite useful to manipulate the migration time and separation, making it possible to separate even the identical chain lengths of different sequences for double-stranded DNA molecules (IO). This enhanced separation is again based on the secondary structure differences caused by different primary sequences. Separation of DNA restriction fragments can be nnproved by using different temperatures during the electrophoresis (8), as well as using gradient methods, such asthe field strengthgradient separationtechnique (I 3). These
Guttman
164
g% CR
z 2
z 2
x Ei
=: ;:
x it
Fig. 3. Effect of the mtercalator additive, ethldmm bromide, on the separation of the $X 174 DNA-HaeIII digest restrlctton fragment mixture. Separatlon without (A) and with (B) 1 pg/mL ethldmm bromide m the gel-buffer system (see ref 10).
methods are useful in optimizatron of the time requirement of the separations, as well as the actual selectivity and resolution that can be achieved. 3.5.3. Chiral
Capillary
Gel Electrophoresis
Capillary polyacrylamlde gel columns can also be used with complexing agents to achieve special selectivittes. As m affinity electrophoresis (IO), the additive can either be covalently bound to the gel matrix or just incorporated into the polymeric fiber gel matrix. The latter method is always preferable, since it requires no special chemistry. Under an applied electric field, the complex will migrate according to its overall charge in the gel when the complexmg agent IS not bound to or entrapped in the matrix. As an example of selectivity manipulation, cyclodextrins (CD) can be used as complexing agents in polyacrylamide capillary gel electrophoresis (20). CDs are nonionic cyclic polysaccharides of glucose with the shape of a toroid or hollow truncated cone. The
Capillary
165
Gel Electrophoresis
‘: I
34
1
2
I 0
”
”
”
1”
! ” 10
”
”
’
”
I 20
”
“1
min
Fig. 4. High-effrclency capillary gel electrophoresls separation of Dns-m-amino acids usmg P-cyclodextrin m the gel as chnal selector. Peaks: 1 = Dns-L-Glu, 2 = Dns-n-Glu, 3 = Dns-L-Ser,4 = Dns-D-Ser,5 = Dns-L-Leu, 6 = Dns-D-Leu(seeref. 17). cavity IS relatively hydrophobic, whereas the external faces are hydrophilic. The torus of the larger circumference contains chiral secondary hydroxyl groups. By incorporation of j3-cyclodextrin into the gel matrix, the complexing agent is practically immobilized (since the CD has no charge) particularly when the pore size of the gel is smaller than the size of the P-CD (cl .4 nm). Cyclodextrins can be used for chiral separations, since chiral selectivity can arise from the entrance of the cavity with the chiral glucose moiety. Figure 4 shows a high-resolution chiral separation of a mixture of dansylated amino acids based on the chiral recognition of the P-cyclodextrin incorporated into the polyacrylamide gel matrix. 3.5.4. SDS-Protein
Capillary
Gel Electrophoresis
Figure 5 shows a separation of a protein test mixture in the mol-wt range of 14,00&98,000 Dalton in 16 min by means of a polymer network filled capillary column (eCAP SDS-200). An internal standard (Orange G) was used in order to increase migration time precision, and thereby improve the accuracy of the mol-wt determination of an unknown protein. The actual peak shape in the case of capillary gel SDS-protein elec-
Guttman
166
I
;
L
.ge cm
8 m
x N
:: m
x w
Fig. 5. Capillary SDS gel electrophoresls of SIX proteins on an eCAP SDS-200 capillary. Peaks: (1) lysozyme (14,400), (2) soybean trypsm mhlbltor (21,500); (3) carbomc anhydrase (mol wt 29,000); (4) ovalbumm (mol wt 45,000); (5) bovine serum albumin (mol wt 66,000); (6) phosphorylase B (97,400). A trackmg dye, Orange G (OG), was added to the sample. Condltlons: Injected amount, 100 ng protein. Detection 2 14 nm. Run temperature, 2OT; applied electric field, 300 V/cm; current 25-30 PA (courtesy of Beckman Instruments).
trophoresis IS a function of the complexation phenomena between the unfolded protein chain and the SDS micelles. Thus, depending on the actual protein molecule, peak efficiencies can be different. As an example m Fig. 5, the slower migrating phosphorylase B peak (#6) is sharper than the preceding ovalbumin (#4). The same column and sieving matrix can be used for even higher mol-wt proteins up to several hundred thousand Dalton (21). Figure 6 shows the mol-wt calibration curve of the standard protein mixture of Fig. 5. This particular gel can be used in fast screening methods for product purity checks, as well as mol-wt determination. The same replaceable concept applies here as previously described for the replaceable polyacrylamide gels. Depending on the purity of the sample, the polymer network can be replaced after each run
Capillary MOBILITY
167
Gel Electrophoresis (lOE-5)
1.1 i
I 100
07 10
30 LOG MW llOE3)
Fig. 6. Relationshrp between the logarithmic mol wt and the mobility. Data points from the electropherogram in Fig. 5. Correlatton coefficient: 0.995 (courtesy of Beckman Instruments) if necessary. This mmimtzes column contamination column lifetime.
3.5.5. Carbohydrate Separation by Capillary
while increasmg
Gel Electrophoresis
In recent years there has been considerable activity in the development of simple, rapid, and reliable separation methods for the analysis of complex carbohydrates released from a variety of glycoconjugates, including glycoproteins, glycolipids, proteoglycans, and glycosammoglycans (4). The high resolving power of CE for the separation of complex carbohydrates has been demonstrated earlier in capillary zone electrophoresis (22) and in capillary gel electrophoresis (23) modes. Since most carbohydrate molecules are not charged and have no significant absorbance of UV light at the common wavelength range of CE-detection systems (200600 nm), a derivatization procedure is required before analysis. This derivatization usually involves the stoichiometric labeling of the reducing end of the oligosaccharide with a labeling reagent (#,22). In most instances, the UVMuorescence dye disodium 8-amino- 1,3,6-naphtalene trisulfonate (ANTS) was employed as a labeling tag followed by reduction using NaCNBH,. The fluorescently labeled oligosaccharides are then separated and quantified by CE using UV (214 nm) or laser-induced fluorescent (LIF) detection (360 nm He-Cd laser).
Guttman
168 3
8 fii 2 P 2s LL 1 0 I
0
6
u I
I
I
I
5
IO
15
20
Migration
Time (min)
Fig. 7. Capillary electrophoretlc-LIF separation of the ANTS-labeled wheat starch digest Numbers correspond to the degree of polymenzatlon (see ref. 23). Figure 7 shows a fast (Cl 5 mm), high resolution separation of wheat starch digest at 20°C on a 40-cm long capillary column. The average plate count for these separations was higher than one million plate/meter, which corresponds to 3500 plate/m/s for the maltose (Fig. 7, peak/# 2). References 1 Landers, J. P (ed.) (1994) Handbook of Capzllary Electrophoreszs CRC, Boca Raton, FL 2 LI, S F. Y. (1993) Caprllary Electrophoresu. Elsevier, Amsterdam, The Netherlands. 3 Schwartz, H and Guttman, A (1995) Separation of DNA by Capillary Electrophoreszs Beckman Primer 607397, Fullerton, CA 4 El Rassi, Z (ed ) (1995) Carbohydrate Analyszs Elsevler, Amsterdam, The Netherlands 5 Cohen, A S , Najanan, D R , Paulus, A., Guttman, A , Smith, J. A., and Karger, B L. (1988) Rapld separation and purlficatlon of oligonucleotldes by high performance capillary electrophoresls Proc Nat1 Acad Ser. USA 85,9660-9663 6. Lux, J. A., Ym, H F , and Schomburg, G (1990) A simple method for the production of gel-filled caplllarles for capillary gel electrophoresis J Hzgh Resolut. Chromatogr
13,436,437
7 Helger, D. N , Cohen, A. S., and Karger, B. L. (1990) Separation of DNA restriction fragments by high performance capillary electrophoresls with low and zero
Capillary
Gel Electrophoresis
169
crosslmked polyacrylamide using continuous and pulsed electric fields J Chromatogr. 516,3348. 8. Guttman, A. and Cooke, N (1991) Effect of the temperature on the separation of DNA restriction fragments in capillary gel electrophorests. J Chromatogr 559, 285-294 9. Tanaka, T. (1981) Gels. Scl Am 244, 124-138 10 Guttman, A. and Cooke, N (1991) Capillary gel affinity electrophoresis of DNA fragments. Anal Chem 63,2038-2042. 11 Nelson, R. J., Paulus, A , Cohen, A S , Guttman, A., and Karger, B. L. (1989) Use of peltier thermoelectric device to control column temperature in high performance capillary electrophoresis. J Chromatogr 480, 111-127. 12 Cohen, A. S. and Karger, B L. (1987) High performance sodium dodecyl sulfate polyacrylamide gel capillary electrophoresis of peptides and proteins J Chromatogr 397,409-4 17 13. Guttman, A. and Mazsaroff, I (1992) Economical performance analysis m preparative capillary gel electrophoresis, m New Approaches In Lzquzd Chromatography, Intercongress, Budapest, Hungary, 63-75 14 Pentoney, S L., Konrad, K. D , and Kaye, W. (1992) A single-fluor approach to DNA sequence determmation using high performance capillary electrophoresis. Electrophoresrs 13,467-474. 15 Rutz-Martinez, M. C , Berka, J., Belenki, A., Foret, F , Miller, A W., and Karger, B L (1993) DNA sequencing by capillary electrophoresis with replaceable linear polyacrylamide and laser induced fluorescence detection. Anal Chem 65,285 l-2858 16. Dovichi, N. (1994) Handbook of Capzllary Electrophoresls (Landers, J P., ed ) CRC, Boca Raton, FL 17 Guttman, A , Nelson, R. J., and Cooke, N (1992) Prediction of migration behavior of ohgonucleottdes in capillary gel electrophorests J Chromatogr 593,297-303 18. Schwartz, H E , Ulfelder, K. J , Sunzeri, F J , Busch, F J., and Brownlee, R. G (199 1) Analysis of DNA restriction fragments and polymerase cham reaction products towards detection of the AIDS (HIV-l) virus m blood. J Chromatogr 559, 267-283 19. Guttman, A , Wanders, B., and Cooke, N (1992) Enhanced separation of DNA restriction fragments by capillary gel electrophoresis using field strength gradients. Anal Chem 64,2348-235 1. 20. Guttman, A , Paulus, A., Cohen, A. S., Grinberg, N., and Karger, B L. (1988) Use of complexing agents for selective separation m high performance capillary electrophoresis, chwal resolution via cyclodextrms incorporated with polyacrylamide gels. J Chromutogr 448,41-53. 21. Guttman, A , Shich, P , and Cooke, N. (1992) P/ACE SDS capillary gel electrophoresis of protcrns Beckman Technzcal Information Bulletrn, #DS-827. 22. Chiesa, C and Horvath, C. S. (1993) Capillary zone electrophoresis of maltoohgosaccharides derivatized with 8-ammonaphtalene-1,3,6-trisulfomc acid J, Chromutogr 645,337-352 23 Guttman, A , Cooke, N , and Starr, C. (1994) Capdlaty electrophoresis separation of ohgosacchandes I. effect of operational variables Electrophoresls 15, 15 18-l 522
CHAPTER 14
Chiral by Capillary Manus
Separations Electrophoresis
M. Rogan
and Kevin
D. Altria
1. Introduction Many papers have appeared concerning the application of capillary electrophoresis (CE) to the resolution of chiral compounds. Predominantly, separations have been achieved in Free Solution Capillary Electrophoresis (FSCE) employing cyclodextrins (CDs) as chiral selectors. This is because of their ready availability and broad applicability. Chiral resolutions have also been achieved m micellar electrokinetic chromatography (MEKC) through the use of chirally selective micelles. However, the variety of approaches has increased significantly. In this chapter, the use and applicability of these methods are considered. Chirally selective CE methods can have significant advantages over HPLC counterparts in terms of increased ruggedness and reductions in both method development time and cost. Acceptable performance levels are achievable, and validated methods are routinely used m many laboratories. In excess of 100 papers have appeared on this subject, and several reviews have been published (I-6). This chapter is divided into sections describing the various selectivity options, method development and optimization, and the application of chiral CE methods. 2. FSCE In chiral FSCE, optically active reagents are added to the electrolyte in order to achieve enantioselective interactions with the solute molecules. Choral recognition agents used in FSCE have included CDs, proteins, carbohydrates, and crown ethers. From
Methods
Edtted
m Molecular
by K Altna
Biology,
Copyright
Vol 52 Caprtlary E/8CtrOphOf8s/s Humana Press Inc. Totowa, NJ
171
172
Rogan and Altria
Fig. 1. Principles of chiral separation for basic compounds using CD additives. A,+ and Ad+ are “I” and “d” enantiomers,respectively. 2.1. CE Employing CDs The most widely reported chiral separation mode is operation at low pH utilizing CDs as chiral recognition agents for the resolution of basic compounds. There are currently wide ranges of both native and derivatized CDs commercially available. The native CDs, a, p, y, possess different numbers of glucose subunits, being six, seven, and eight, respectively. Enantioselective recognition is often explained by interactions between the CD (which possessesmany chiral centers) and the guest enantiomers. By derivatization of the surface hydroxyl groups on the cyclodextrin, the solubility and enantioselective properties can be significantly altered. Hydroxypropyl- and dimethyl-cyclodextrins are two such derivatives, which when compared to the native forms, differ significantly. In addition, CDs having acidic or basic ionizable groups are available for separation of uncharged solutes. These derivatived CDs are available from a number of chemical suppliers. The basis for the separation is the stereospecific interaction of the analyte with the CD cavity and surface (see Fig. 1). At low pH, EOF is minimal, and the uncharged CD does not migrate. Basic analytes are protonated at this pH and migrate toward the detector by virtue of their charge. When the solute interacts with the CD, its mobility is greatly reduced. If the two enantiomers have differing stability constants, then one enantiomer will migrate more slowly than the other, and chiral resolution will be achieved. Figure 1 shows the process diagrammatically. A number of groups have exploited CDs in the free solution mode. For example, Fanali (7,8) has reported the enantiomeric resolution of
Chiral Separations
6
173
.
i
i
mm
Fig. 2. Chiral resolutton of dtmethidine and a range of related compounds (reproduced with permission from ref. 9) Peak 1 = N-demethyl dimethldme, 2 = dlmethidme, 3 = N-demethyl-6-methoxy-dlmethtdme, 4 = 6-methoxy-dtmethyl-dlmethidme, 5 = dlmethldme-N-oxide, pH 3, electrolyte contammg hydroxypropyl+CD, ephedrine, norephedrine, epinephrine, norepinephrme, and tsoproterenol using dimethyl-P-CD. Dimethidine and a range of related compounds (9) have been simultaneously chirally resolved (Fig. 2). Free solution CE employmg CDs can also be performed at high pH
(10, II). Under these conditions, a different separation mechanism exists. The acidic solute is negatively charged and opposes the EOF. The strong EOF sweeps the neutral CD to the detector, and when the anionic solute interacts with the migrating neutral CD, its migration velocity is reduced. Therefore, the enantiomer having the most interactions migrates before the other stereorsomer, a result that is the reverse of that obtained at low PH (W. Alternatively, ionizable CDs can be employed for the separation of neutral solutes or for the improved separation of charged analytes (12). For example, sulfated CDs can be used to affect separation of neutral species at high-pH values. Carboxymethyl-P-cyclodextrin (C-Den) was used to resolve six enantiomeric dansylated amino acids (13).
174
Rogan and Altria
R
Fig 3. Complex of primary amine with crown etherwhere R ISa substituent group or groups
2.2. FSCE Employing Crown Ethers Crown ethers are macromolecules that can be employed to achieve chiral separation of primary amines. A typical structure of a crown ether/ primary amine complex is given in Fig. 3. The separation process is similar to CD-based FSCE in that the enantromers of prrmary ammes form complexes with the crown ether and their migration is retarded. The NH3+ group of the solute binds with the crown ether via hydrogen bonding, and enantrorecogmtion depends on the secondary interactions of the crown ether substituents with the functional groups on the analyte. Kuhn et al. (14) were the first to report the use of [18] crown-6-tetracarboxylic acid (18C6) as a selector in CE for separation of tryptophan and DOPA enanttomers. No buffer substituents were used in this work becausethe cations were found to interact compettttvely with the additive. The enantiomers of phenylalanine, phenylglycine methoxamme, and octopamine have been resolved using 18C6 (15). Crown ethers have also been employed for the resolution of a number of optrcally active amines of pharmaceutical interest (6,16, I 7). 2.3. FSCE Employing Enantioselective Metal Chelation In this separation mechanism, a transition metal/ammo acid (R or S) complex is added to the electrolyte. The coordination number of the complex is not fully satisfied, and therefore, solutes of an appropriate structure can form ternary diastereomeric complexes. If there are differences
Chiral Separations
175
in complex stability for the two stereoisomers, enantioselective recognition can be achieved. The more complexed enantiomers migrate faster than their lesser complexed counterparts. This approach has been used to resolve enantiomerically seven pairs of dansylated amino acids (18) using a Cu2+-aspartame complex as a background electrolyte additive. 2.4. FSCE Employing
Carbohydrates
It has been shown that additions of 2-10% of maltodextrins (mixtures of linear linked D-glucose polymers) can be used for chiral resolution of acidic drugs (‘19), such as nonsteroidal anti-inflammatories (NSAIDs), and anticoagulants, such as warfarin. In a second report, dextrins were employed for the chiral separation of a range of drug molecules and showed different enantioselectivities to CDs (‘20). 2.5. FSCE Employing
Proteins
Proteins are widely utilized in HPLC as chiral phases, since they possess multiple chiral centers. Similarly, proteins, such as bovine serum albumin (BSA) (21-23), and ovomucoid (OVM) (24) have been successfully employed in choral FSCE. Typical protem concentrations are on the order of a few milhgrams per milliliter, which makes them a realistic possibility from a cost perspective. The separations are typically performed at pH 7, where most proteins are negatively charged and migrate against the EOF. Favorable solute interactions with the proteins retard migration and lead to the possibility of chiral resolution. BSA has been successful in chiral separation of a range of drugs, including warfarin, benzoin and tryptophan (24), and leucovorin (21,22). Valtcheva et al. (25) successfully employed the protein cellobiohydrolase I (CBH I) as an additive in FSCE to resolve a series of P-blockers (propranolol, alprenolol, metoprolol, and pindolol). The addition of organic solvents, such as 2-propanol, was shown to be of importance for obtaining chiral separation. 2.6. FSCE Employing
GZycopeptides
In this separation m